ModWorX Software Manual
ModWorX Software Manual
WARNING This symbol identifies information about practices or circumstances that can
! lead to personal injury or death, property damage, or economic loss.
Caution Indicates actions or procedures which if not performed correctly may lead to personal
injury or incorrect function of the instrument or connected equipment.
IMPORTANT Indicates actions or procedures which may affect instrument operation or may lead to an
instrument response which is not planned.
Note Indicates additional information about specific conditions or circumstances that may affect instrument operation.
Sensia LLC
200 Westlake Park Blvd
Houston, TX 77079
+1-866-773-6742
sensiaglobal.com
*Mark of Sensia.
Other company, product, and service names are the properties of their respective owners.
Contents
Important Information............................................................................................................................................i
Symbols and Terms Used in this Manual.................................................................................................i
Technical Support Contact Information.....................................................................................................i
ii
ModWorX Pro Table of Contents
Archives............................................................................................................................................................. 38
Contract Hour.............................................................................................................................................. 39
Interval Period............................................................................................................................................. 40
Partial Records........................................................................................................................................... 40
Archive Parameter Setup.................................................................................................................................. 41
Alarms............................................................................................................................................................... 42
Setting Up Alarm Parameters............................................................................................................................ 43
Modifying Alarm Parameters....................................................................................................................... 44
User Modbus Registers..................................................................................................................................... 44
Section 9 - Flow Run Maintenance for Orifice Input (AGA-3 or ISO-5167 Orifice)...................119
Procedure: Change Plate Parameters...................................................................................................... 121
Procedure: Change Gas Composition...................................................................................................... 123
Procedure: Change Gross Method Parameters........................................................................................ 125
Procedure: Change Steam Properties (Steam Flow Runs Only).............................................................. 126
Procedure: Change MPMS Liquid Parameters......................................................................................... 128
Procedure: Change Generic Liquid Parameters....................................................................................... 131
iii
Table of Contents ModWorX Pro
Section 11 - Flow Run Maintenance for Averaging Pitot Tube (Annubar) Input.......................150
Procedure: Change Averaging Pitot Tube Parameters............................................................................. 152
Procedure: Change Gas Composition...................................................................................................... 154
Procedure: Change Gross Method Parameters........................................................................................ 156
Procedure: Change MPMS Liquid Parameters......................................................................................... 158
Procedure: Change Generic Liquid Parameters....................................................................................... 160
iv
ModWorX Pro Table of Contents
Appendix C - Configuring Liquid Parameters Using Scanner Firmware 4.10 and Earlier.......C-1
Flow Rate Calculation Methods........................................................................................................................C-1
Flow Run Maintenance.....................................................................................................................................C-2
Procedure: Change API Liquid Parameters...............................................................................................C-2
Procedure: Change Generic Liquid Parameters........................................................................................C-4
v
Table of Contents ModWorX Pro
vi
ModWorX Pro Section 1
Note To go immediately to the unzipped folder, check the box next to “Show extracted files when complete.”
Note Internet Explorer is required to install software from the Scanner Software Installations screen. To use a different
browser, start Windows Explorer, browse to the location where the setup files were unzipped (or the CD drive),
open the Install folder, and double-click on setup.exe.
IMPORTANT The USB adapter should not be connected to your computer during driver installation.
To access the driver installation file, click setup.htm from downloaded ModWorX Pro Software folder. Click Install
USB Driver to install the NuFlo USB Adapter driver.
Driver installation instructions are also included in the ModWorX Pro readme file. The readme file can be accessed from
the Start menu (Start>Programs>NuFlo>ModWorX Pro 4.3.0>Documents>ReadMe.htm).
1
Section 1 ModWorX Pro
Figure 1.1
2. Click Connect. The Connection Settings screen will appear (Figure 1.2).
Note If prompted to select a COM Port, see Changing Connection Settings, page 7 for more information.
Figure 1.2
2
ModWorX Pro Section 1
–– Express Connect – recommended for single instruments. The software automatically connects to an instrument
without a slave address or baud rate provided.
IMPORTANT If multiple devices are daisy-chained together in a network, do NOT select the “Express
Connect” connection method.
–– Discover Device – recommended for multiple-instrument networks when the slave address or baud rate is not
known. When the software attempts to connect to a Scanner, it will compile a list of instruments on the network,
and the user then chooses a device from the list (Figure 1.3, page 3). A Search Now button will appear, al-
lowing the user to perform a new search for additional connected devices, if desired. The devices are ordered by
slave address, with the lowest address at the top of the list. If two or more devices have the same slave address
and baud rate, the Change Device Parameters button allows a user to select a new slave address and/or baud
rate for any detected device without leaving the Discover Devices dialog (Figure 1.4, page 4).
–– Specify Parameters – recommended for multiple-instrument networks when the slave address and baud rate is
known. The user enters the instrument’s slave address and baud rate, and adjusts the time-out setting, if neces-
sary. (Time-out is the length of time the software will search for the instrument before giving up and generating
an error message.) The user may also enter the number of retries desired before a connection attempt is termi-
nated.
Figure 1.3
3
Section 1 ModWorX Pro
Figure 1.4
4. On initial login, the system will automatically connect to the Scanner using default security permissions (full ac-
cess). The “Use default permissions of connected device port” checkbox (shown in Figure 1.2, page 2) will be
checked. No user name or password is required for login.
If individual security controls are implemented, the default permissions setting must be changed to “no access,” and
users will instead login with a user name and password. If the “Use default permissions...” checkbox is checked,
deselect it to enter a user name and password (Figure 1.5).
5. If individual user security controls are implemented, the user name and password will be saved by default when a
user logs out of ModWorX Pro, eliminating the need to re-enter the information with each login (note the “Save
User Name and Password between Sessions” checkbox). If this password save feature is not desired, remove the
check from the checkbox.
Figure 1.5
6. If you will use the same connection method each time you connect to the device, select the “Use these settings to
automatically connect next time” checkbox at the bottom of the screen (Figure 1.6, page 5). The software will
attempt to connect to the device without prompting the user to select a connection method.
Note Connection settings can be changed on the Connection Settings screen shown above or in the Tools menu (Op-
tions submenu, Connection screen). See Appendix A - Tools Menu for details. For information about setting up
and changing security settings for individual users, see Software Interface Security, page 34.
4
ModWorX Pro Section 1
8. If the date and time setting in the Scanner differs from the date and time setting in your computer by more than
14,400 seconds (or other threshold specified by the user), the software will prompt you to synchronize the device
date and time with the computer date and time. Click Yes to synchronize the instrument date and time with your
computer (Figure 1.7, page 5). The default threshold is set at 14,400 seconds (4 hours) to prevent undesired
prompts when traveling between time zones.
See Appendix A - Tools Menu for detailed instructions on changing the setpoint for the time synchronization prompt.
Figure 1.6
Figure 1.7
Mechanical Checkpoints
• Are the connections between your computer and your RS-485 or NuFlo USB adapter secure?
• Are the connections from your adapter to the Scanner secure?
• Is the polarity of the wiring from the converter to the instrument correct? For wiring diagrams, see the Quick-Start
Guide or Hardware User Manual for your Scanner model.
• Is the Scanner receiving adequate power, and are the power connections secure?
5
Section 1 ModWorX Pro
Software Checkpoints
• Are the slave address and baud rate correct?
–– I f you are using a radio link, a virtual COM port, or an Ethernet serial converter, ensure that the baud rate set in
the device is supported by the intermediate link.
–– I f you have a long wiring network, have many devices on your network, or have line noise, try a slower baud
rate to achieve reliable communication to your devices.
Note If you change the communications parameters (slave address, baud rate) of the Scanner from the instrument
keypad, the ModWorX Pro software will not detect the change instantaneously. Return to the Welcome Screen
and reconnect to the device.
• Is your computer network speed too slow for the communication setting defaults? If the software repeatedly fails to
connect to the Scanner, consider the following adjustments. To locate these settings, click on Tools in the task bar at
the top of the screen, choose Options…, then choose Communications.
–– Change the number of retries allowed (‘3’ is the default). If the program does not receive a valid response from
the device within the specified time-out period, the program will automatically resend the message a number of
times before finally reporting a communications error. By default, the program will re-attempt each communica-
tion two times before it gives up. To increase the number of retries, increase the value in the Number of Retries
field in the Communications section of the Program Options dialog. Line noise can corrupt messages exchanged
between the program and the device, and increasing the number of retries can improve the chances of a success-
ful connection.
–– C
hange the time-out period (5000 msec is the default). Depending on your computer’s hardware and software
characteristics and the characteristics of your device network, the default time-out period may be too short. To
extend the time-out period, enter a higher value in the Time-out Period field in the Communications section of
the Program Options dialog. If the program does not receive a valid response from the device within the speci-
fied time-out period, the program will automatically resend the message a number of times before finally report-
ing a communications error.
–– T
o increase the time that the program waits between consecutive Modbus commands, increase the Presend
Delay value in the Communications section of the Program Options dialog.
–– Y
our RS-485 adapter may require that the RTS line of the COM port be toggled to control the direction of data
flow for transmit and receive. If so, ensure that the “Toggle RTS line” option is enabled in the Communications
section of the Program Options dialog. Alternatively, if you are connected to a modem device on your COM
port that requires the RTS line be used in the standard way for RS-232 flow control, the “Toggle RTS line” op-
tion should be disabled.
6
ModWorX Pro Section 1
Figure 1.8
A Select COM Port dialog will appear, with COM1 displayed as the default serial port (Figure 1.9). Select a new COM
port from the dropdown menu and click OK. If the NuFlo USB adapter is in use, the COM port will appear as “NuFlo
USB Adapter.”
Figure 1.9
7
Section 1 ModWorX Pro
To change the connection method, a user can select Return to Welcome Page from the Tasks menu (Figure 1.10) in the
task bar.
Figure 1.10
See Appendix A - Tools Menu for detailed instructions on changing the communications port, adjusting the time synchro-
nization prompt, changing connection methods when the auto-connect feature is enabled, and changing other settings.
8
ModWorX Pro Section 1
9
Section 2 ModWorX Pro
Main Display
The Scanner Main Display screen appears immediately upon connecting to a Scanner (Figure 2.1). This screen is the
central hub of the software interface - the point from which a user can view real-time data and access menus for all com-
monly performed tasks.
Note the navigating window centered at the top of the screen. This window, which changes to identify the screen in view
at any given time, is displayed on nearly all screens in the interface, and allows users to navigate in and out of menus
with ease. It also allows users to quickly verify the connected device by the user-assigned name, serial number, site, field,
and/or location.
Navigating window
Task-based menus
Current data
User documentation
10
ModWorX Pro Section 2
Expansion board
parameters
(circled)
Figure 2.2
Navigating window
identifies Scanner
model
Figure 2.3
11
Section 2 ModWorX Pro
–– Configuring system parameters, setting up a flow run or turbine run, and configuring inputs and outputs
• View current data
• Link to user documentation
Quick-start Guide (a basic reference to installation, wiring, calibration and configuration)
––
Log Downloads
The red Download button allows a user to download log data and device settings for viewing onscreen, printing, or ex-
porting. A Download Device Data screen will appear. From this screen, a user can select:
• The range of logs to download (new records only, or all records)
• The log types to include (daily, interval, and/or event)
• Additional data to include (configuration, calibration, holding registers)
See Section 14 - Downloading and Exporting Logs for detailed instructions.
Task-Based Menus
The four gray menu buttons provide direct access to the controls for calibrating and configuring the Scanner for specific
needs (Figure 2.4, page 13).
• Calibrate inputs—calibrate all input parameters from one screen
• Maintain flowrun—change plate or cone parameters, change gas composition or steam properties, change generic
liquid parameters or change MPMS liquid parameters, change a meter factor or flow coefficient (selections vary, de-
pending on flow run configuration)
• Maintain turbine—change a turbine flowmeter, change a K-factor
• Configure—configure basic system parameters (such as time/date and archive setup), as well as flow run, turbine,
and input parameters
12
ModWorX Pro Section 2
Figure 2.4
Units
The units of measurement displayed are read-only from the Main Display screen. Units for system parameters such as
heating value and device temperature, are displayed in US Standard units by default, but these settings can be changed to
metric units.
To change from US Standard to metric units, or vice versa, click on the Tools button in the task bar at the top of the
screen, and select Options from the dropdown menu. The units selection is on the General tab. The values will automati-
cally convert to the new unit selection.
Other units for measurement parameters such as volume, mass, energy, differential pressure, etc. are specified by the user
during the configuration process and can be changed in the Configure menu. These settings are discussed in Section 3 -
Configuring System Parameters.
Refresh Options
To manually refresh the values displayed on the Main Display screen, press the Refresh button near the top of the screen.
To enable the automatic refreshing of values, click the Auto-Refresh checkbox.
13
Section 2 ModWorX Pro
Diagnostic Indicators
Color-coded diagnostic indicators are provided for input parameters, as well as for configured alarms (Figure 2.5).
Figure 2.5
Input Data
The diagnostic indicator in the Input Data grid is activated (turns from green to red) when communication to an input
device is lost or the device reading goes outside the user-specified range. When an override value is set for an input pa-
rameter, a lock symbol appears in combination with the diagnostic indicator. The override setting can appear with either
a green or red indicator. Indicators may include the following symbols:
For a detailed view of input parameters and their diagnostic status, press the Details button in the Input Data grid. The
Input Data Detail screen will appear.
14
ModWorX Pro Section 2
Note If an optional expansion board is installed in a Scanner 2000 or Scanner 210x, the Input Data Detail screen will
display the operational status and live values for two analog inputs. If a Scanner 2200 is installed, this informa-
tion will be displayed without an additional expansion board.
Figure 2.6
Note To clear an override or make other input configuration changes, click on the Configure menu on the Main Dis-
play screen and choose the appropriate parameter from the Inputs section. See Section 6 - Configuring Inputs,
page 68 for more information.
System Data
The diagnostic indicator in the System Data grid is activated when the value of a parameter that has been linked to an
alarm exceeds a user-specified setpoint in the alarm configuration. If one or more configured alarms is active, a red
diagnostic indicator will appear in the System Data grid (Figure 2.7, page 16). If no alarms are active (or no alarms
are configured), a green diagnostic indicator will appear in the System Data grid. If the smart battery is installed, then
the system data screen will show the smart battery capacity (“Battery %”) and health (“Health %”) (in Figure 2.7, page
16).
Note Replace the smart battery if it’s battery health (“Health %”) drops to 80% or below.
15
Section 2 ModWorX Pro
Figure 2.7
For a complete list of configured alarms and the status of each, press the Alarms button in the System Data grid. The
Alarm Data Detail screen will appear. Alarms can also be downloaded for viewing in a report. See Section 14 - Down-
loading and Exporting Logs for more information.
The supply and battery voltage is also displayed in the System Data grid. The supply voltage is the voltage supplied to
the Scanner via the external power or analog output terminal blocks on the backplate board. The battery voltage is the
measured voltage supplied by the battery (either lithium or smart battery).
The values will vary, depending on whether the Scanner is powered by the battery option (lithium or smart battery) only,
by external power or loop power only (Scanner 2105 only), or by external power or loop power (Scanner 2105 only)
with a battery option as a backup power source.
Note External power (external supply or loop power) and Lithium battery voltage will be displayed in System Data.
When the smart battery is connected, System Data will show the battery percentage capacity and the battery
health instead of the smart battery voltage.
16
ModWorX Pro Section 2
Note If an optional expansion board is installed in a Scanner 2000 or Scanner 210x, the System Data screen will
include a live analog output value. If a Scanner 2200 is installed, this value will be displayed without an additional
expansion board.
Figure 2.8
Alarm diagnostic indicators may include the following symbols:
Alarms that are configured as “latched” alarms will remain active until the user clears them. To clear all active alarms,
eliminate the condition(s) causing the alarm(s), then press the Clear Alarms button.
Note To change an alarm setpoint or deadband or configure a new alarm parameter, click on the
Configure menu on the Main Display screen and choose Alarms from the System section. See
Section 3 - Configuring System Parameters for more information.
17
Section 2 ModWorX Pro
Details
A detailed view of current data is available by clicking on the Details button in the lower right corner of each grid. The
Details screen content varies, depending on the parameters monitored and the flow rate and calculation method selected.
Figure 2.9
18
ModWorX Pro Section 2
Figure 2.10
Turbine Detail
Like the Flow Run Detail screen, the Turbine Detail screen (Figure 2.11) includes current interval and previous interval
volumes, calculations used to determine flow rate, and parameters used in those calculations.
Figure 2.11
Note If an optional expansion board is installed in a Scanner 2000 or Scanner 210x, the Turbine Detail screen will
include data for a second turbine input (T2). If a Scanner 2200 is installed, this data will be displayed without an
additional expansion board.
19
Section 2 ModWorX Pro
Figure 2.12
System Detail
The System Detail screen (Figure 2.13) displays information for identifying a specific instrument, to include serial num-
bers, manufacture and sales dates, firmware version numbers, etc. The System Detail screen also indicates the status of a
device seal for custody transfer installations, and displays the power mode of the microprocessor. The Scanner optimizes
performance and battery life by changing from high to low power mode or vice versa, depending on configuration set-
tings and operational requirements. If the smart battery is supported on the device, then the installation status will be dis-
played. If the smart battery is installed, the System Detail screen will confirm if it is communicating with the ModWorX
Pro software. The smart battery capacity (“Battery %”) and health (“Health %”) will be shown in System Data (Figure
2.7, page 16).
Figure 2.13
Note If an optional expansion board is installed in a Scanner 2000 or Scanner 210x, the System Detail screen will in-
clude expansion board information and a status indication for all expansion board parameters. If a Scanner 2200
is installed, status information for the complete range of inputs/outputs will be displayed and no expansion board
information will be shown.
Note Replace the smart battery if it’s battery health (“Health %”) drops to 80% or below.
20
ModWorX Pro Section 2
Figure 2.14
Figure 2.15
21
Section 2 ModWorX Pro
When selecting either of the first two options (User Manuals tab or Help menu), the following screen will appear (Figure
2.16).
Figure 2.16
1. Select the Scanner product in use.
2. Then choose the desired document from the list of available documents for the selected product (Figure 2.17).
Figure 2.17
22
ModWorX Pro Section 2
To return to the Select Your Product screen to select a different document, click the Home icon.
To restore the view of ModWorX Pro configuration controls, click the Device Interface tab at the bottom of the screen.
The following user documents are available from within the ModWorX Pro program:
• a quick-start guide that covers the basics of installing, wiring, and configuring the Scanner
• a complete hardware user manual
• a complete software user manual
• NuFlo USB adapter hardware installation guide (Scanner 2000 or Scanner 210x users only)
• NuFlo USB adapter driver installation guide
• NuFlo USB adapter driver troubleshooting tips
• communications troubleshooting tips
Note The initial load of these documents can take several seconds. Please be patient the first time you access these
documents and allow the screens to load properly.
23
Section 2 ModWorX Pro
24
ModWorX Pro Section 3
Figure 3.1
2. Then, click on the appropriate parameter in the System section of the Configuration Menu screen (Figure 3.2, page
26).
25
Section 3 ModWorX Pro
Figure 3.2
Device Identification
To configure device identification parameters, click on the Configure button on the Main Display screen. Then, click on
the Device Identification button on the Configuration Menu screen (Figure 3.3).
Figure 3.3
26
ModWorX Pro Section 3
Figure 3.4
The data storage directories shown in Figure 3.4 are controlled via settings within the Options dialog. See Appendix A -
Tools Menu for more information.
The Configure Device Identification screen allows a user to enter information that distinguishes an instrument from other
networked instruments, including
• Device/meter name
–– by default, the device/meter name is included in the filename for each download report
• Company name
• Site/well name
–– by default, the site/well name is the name of the folder created on your hard drive for storing downloaded logs:
C:\NuFlo Log Data\ModWorX Pro\Field_Name\Site_Name
• Field/lease name
–– by default, the field/lease name is the name of the folder created on your hard drive for storing site name sub-
folders: C:\NuFlo Log Data\ModWorX Pro\Field_Name
• Location name
• Legal description
When the device name, site name, field name, and location name are entered, this information appears in the navigational
window at the top of each screen (Figure 3.5, page 28). The serial number of the Scanner is also displayed in this win-
dow.
27
Section 3 ModWorX Pro
Figure 3.5
Click Apply to save the new settings without leaving the Configure Device Identification screen. Or click OK to save the
new settings and return to the Configuration Menu screen.
Date/Time
To configure the instrument date and time, click on the Configure button on the Main Display screen. Then, click on the
Date/Time button on the Configuration Menu screen (Figure 3.6).
Figure 3.6
28
ModWorX Pro Section 3
The Configure Date and Time screen (Figure 3.7) will appear.
Figure 3.7
The Configure Date and Time screen allows a user to set the date and time two ways:
1. Click the Sync with Computer Time button to set the instrument date and time to the date and time displayed on
your computer.
2. Click the Edit buttons to change the date and time displayed using the calendars shown in Figure 3.8.
a. To change the date, click the Edit button for Device Date, and select the date from the calendar provided, or
click Today to select the current date. Click OK to save the setting and update the Device Date display.
b. To change the time, click the Edit button for Device Time, and enter the correct time using the -/+ buttons to
change the hour, minutes and seconds shown. Click the dropdown toggle button to select “AM” or “PM.” Click
OK to save the settings and update the Device Time display.
Figure 3.8
c. Click Apply Settings Now on the Configure Date and Time screen to write the values to memory.
d. Click Done to return to the Configuration Menu screen.
29
Section 3 ModWorX Pro
Communications
To configure the communications ports, click the Configure button on the Main Display screen. Then, click the Com-
munications button on the Configuration Menu screen (Figure 3.9). The Configure Communications screen (Figure 3.10)
will appear. If a Scanner 2200 is installed, this screen will also allow configuration of radio control settings. See Radio
Control, page 31, for details. If a Scanner 2100 wireless device is installed, you will be prompted to enter the Network
ID.
Figure 3.9
Figure 3.10
The Configure Communications screen allows users to change the settings that are required for Modbus communication.
The Scanner has two slave communications ports. The Scanner detects which of these ports is connected to the instru-
ment and indicates the connection status on this screen, as shown in Figure 3.10.
30
ModWorX Pro Section 3
Both COM ports are currently slave ports. To change the slave address, click in the Slave Address field and enter a new
number between 1 and 65535 (the default setting is 1; 252 through 255 and 64764 are reserved and should not be used).
The slave address allows the Scanner to communicate with other devices via Modbus. If a Modbus request message con-
tains the matching slave address, the device will respond to the request. In network arrangements, the slave device must
have a unique address.
Note For a Scanner 2105, Port 2 is reserved and not available for user configuration (Figure 3.11). Port 2 is exclusively
used for communication between the main circuit board and the backplate board.
Figure 3.11
1. To change the baud rate, select a number from the dropdown menu or click in the field and use the up and down
arrow keys on your computer keyboard to select the desired baud rate. The default setting is 9600. The baud rate is
the speed at which data is transmitted or received via the serial port. Baud rates supported by the Scanner range from
9600 to 38.4K. This setting must match the baud setting of the master device polling the Scanner serial port.
2. To change the bus delay, delete the existing value and enter a new value. Bus delay is the amount of time (in mil-
liseconds) that passes before the Scanner attempts to take control of the RS-485 bus and transmit a message back
to the requesting device. The Scanner responds very quickly to incoming Modbus requests—in some cases, too
quickly. A configurable delay allows the user to control this response time. The default setting of 10 msec is nor-
mally sufficient, but a longer delay may be necessary when the Scanner is communicating with a radio or other end
device that responds more slowly.
3. To change the bus time-out, delete the existing value and enter a new value. Bus time-out is the amount of time (in
milliseconds) that must pass to cause the Scanner to reset its internal Modbus message handler. The default value of
50 msec is normally sufficient.
4. Click Apply to save the new settings without leaving the Configure Communications screen. Or click OK to save the
new settings and return to the Configuration Menu screen.
31
Section 3 ModWorX Pro
Figure 3.12
2. In the Radio Settings section, configure the power mode as “always on” for continuous power, or “on/off times” to
select the time of day for turning the radio off and on.
–– If “always on” is used, no further configuration is needed. Click Apply to save the new settings without leaving
the Configure Communications screen. Or click OK to save the new settings and return to the Configuration
Menu screen.
32
ModWorX Pro Section 3
Figure 3.13
Security
ModWorX Pro features three types of security controls: (1) keypad security for preventing unauthorized personnel from
altering configuration settings with the keypad, (2) a custody transfer device seal which prevents all changes to configu-
ration settings when used with the appropriate circuit board jumper setting, and (3) password-protected access to the
software interface.
Keypad Security
The Scanner’s keypad security feature is disabled at the factory.
To enable keypad security for the device:
1. Click the Configure button on the Main Display screen. Then, click the Security button on the Configuration Menu
screen (Figure 3.14). The Configure Security screen (Figure 3.15, page 34) will appear.
Figure 3.14
33
Section 3 ModWorX Pro
Figure 3.15
2. Change the Keypad Lock Enable setting to Yes.
3. Enter a four-digit lock code in the field below the checkbox. (Do not use “0000.”)
4. To lock input values (static pressure, differential pressure, and process temperature) during a plate change so that
volume measurements are not lost, enable the Strict API Compliance setting. When this feature is enabled, all plate
changes must be made using ModWorX Pro. When this feature is disabled, plate changes can be made from the front
panel (however, the inputs will not be locked during the plate change, making the volume calculations inaccurate).
5. Click Apply to save the new settings without leaving the Configure Security screen. Or click OK to save the new
settings and return to the Configuration Menu screen. The next time a user attempts to access the keypad, he will be
prompted to enter the security lock code.
Custody Transfer
When a device seal is installed on a Scanner 2000 for compliance with Measurement Canada custody transfer regula-
tions, the Device Seal status on the Configure Security screen must be set to Enabled to prohibit configuration changes
that can affect measurements (Figure 3.15). The clearing of flow totals is also prohibited when the device seal is installed
and enabled. See the Scanner 2000 Hardware Manual, Part No. 9A-30165023, for instructions on installing the device
seal for compliance with Measurement Canada regulations.
The Scanner 210x and 2200 have not been evaluated for Measurement Canada approval.
34
ModWorX Pro Section 3
Archive Partials
Tools: Manage
Security Level
Configuration
Tools: Create
Tools: Clear
Tools: Clear
Tools: Copy
Description
Tools: View
Flow Totals
All Modbus
Registers
EEPROM
Memory
Access
Note Up to nine user accounts can be set up in the Security menu. When nine users have been added to the User
Login Accounts list, the Add User button will be grayed out. No more users can be added.
To change the default settings, click on the default entry (User 1 or User 2) on the Configure Security screen, then click
the Change button. A User Account dialog (Figure 3.16) will open, providing access to all user account information.
From this window, the default user name, password, and access level can be changed.
An optional Modbus user login code can also be set up within this box. The Modbus user login code is needed only if the
account holder will access ModWorX Pro remotely via a host network.
Figure 3.16
35
Section 3 ModWorX Pro
Display
To configure the parameters to be displayed on the LCD, click on the Configure button on the Main Display screen.
Then, click on the Display button on the Configuration Menu screen (Figure 3.17).
Figure 3.17
The Configure Display screen will appear (Figure 3.18).
Figure 3.18
Up to 12 parameters can be selected for display on the LCD. The display scrolls from one parameter to the next until all
parameters have been displayed and then repeats the sequence. From the Configure Display screen, a user can:
• change the period of time each parameter is displayed
• adjust the LCD contrast
• view parameters configured for display
• add or delete parameters from the display sequence
• enable or disable the scrolling of the display
• change the way each parameter is displayed (with or without description and/or units)
• change the order in which parameters are displayed
36
ModWorX Pro Section 3
Display Period
By default, each parameter is displayed for 5 seconds. To change the time period, click the -/+ buttons in the Display
Toggle Period field. Click Apply to implement the change.
LCD Contrast
LCD contrast is temperature-dependent and may require adjustment for optimum readout visibility. To change the con-
trast, adjust the slide bar provided. Click Apply to implement the changes.
–– If Text Message is selected, the user must enter the text to be displayed in the Description field. Up to 19 char-
acters (spaces included) can be entered.
–– If the description contains more than six characters, the display will automatically scroll. To disable the scroll-
ing feature, limit the description to six or fewer characters.
4. Edit the number of decimal places to be included in the display readout, if desired.
5. Edit the update period (seconds), if desired.
6. Select the way you want the parameter to be identified in the lower half of the LCD. The Example field in the fol-
lowing screen captures shows the various description and unit options available (Figure 3.19). The options are:
a. Description Only (left)
b. Units Only (center)
c. Description and Units (right)
Figure 3.19
37
Section 3 ModWorX Pro
To delete a parameter, click on a display item to select it, and click Delete.
To change the order in which the parameters appear in the LCD scroll, click on a display item and click Move Up or
Move Down.
Note When 12 parameters have been added to the Display Item list, the Add Item button will be grayed out. No more
parameters can be added without deleting or changing an existing parameter.
7. Click Apply to save the new settings without leaving the Configure Display screen. Or click OK to save the new set-
tings and return to the Configuration Menu screen.
Archives
Caution Before configuring archive parameters, download all existing log data. For detailed instructions on
downloading data, see Section 14 - Downloading and Exporting Logs.
To configure the archive settings, click on the Configure button on the Main Display screen. Then, click on the Archives
button on the Configuration Menu screen (Figure 3.20).
Figure 3.20
38
ModWorX Pro Section 3
Figure 3.21
Archive settings control which parameters are logged by the Scanner, and the frequency of those logs. The standard
Scanner 2000 or Scanner 210x will hold up to 768 daily logs, up to 2304 interval logs (configured in terms of seconds,
minutes, or hours) and up to 1152 event/alarm logs. When an expansion board is added to the Scanner 2000 or Scanner
210x or a Scanner 2200 is installed, the archive capacity for interval logs is increased to 6392 records. Once the log is
filled, oldest records will be over-written by new records.
Daily logs are recorded at the time of day designated by the user-specified contract hour.
From the Configure Archive screen, a user can:
• change the contract hour
• change an interval period
• enable a partial record to be stored when a calibration or configuration change is made
• change the parameters to be logged
Contract Hour
To change the contract hour, click on the Contract Hour dropdown list and select the desired hour.
39
Section 3 ModWorX Pro
Interval Period
The number in the Interval Period field determines how frequently the selected parameters are logged.
To change the interval period:
1. Click Edit. The Edit Synchronized Time Periods window will appear, and an arrow will point to the Interval Archive
setting (Figure 3.22).
Figure 3.22
2. Choose a time unit (seconds, minutes, or hours) for the archive logging period.
3. Click on the -/+ buttons to enter the number of seconds, minutes, or hours. (The minimum configurable archive
period is 5 seconds.) To disable the interval period, enter a zero in this field. The screen will automatically update to
show the number of interval periods that will be logged per day, based on the new settings.
4. Click OK to return to the Configure Archive screen.
Partial Records
As the name “partial” suggests, a partial record is a set of values from an incomplete daily or interval archive period. By
default, the Enable Partials setting is disabled (Figure 3.21, page 39). When the setting is disabled and a calibration
or configuration change is made that affects flow calculations, the daily log values recorded for that period will reflect a
combination of calculations performed with the previous settings and calculations performed with modified settings.
When the Enable Partials setting is enabled and a change is made to calibration or configuration settings, the current ar-
chive record is terminated and a new record is initiated, based on the new calibration or configuration settings saved. Any
flow recorded prior to the setting change will be displayed as Previous Interval and Previous Day totals, and any flow
recorded after the setting change will be added to the Current Interval and Current Day totals.
For example, assume the contract hour is set at 8 a.m., and Partials is enabled. A calibration or configuration change
made at 10 a.m. would cause a partial log of just 2 hours’ flow to be generated. This log will be displayed as the Previ-
ous Day total. The new log is also a partial, beginning at 10 a.m. and continuing until 8 a.m. the next day. This new log is
displayed as the Current Day total during this period from 10 a.m. to 8 a.m. the next day.
To enable the creation of partial records within the flow archive, change the Enable Partials setting on the Configure Ar-
chive screen to Yes. See also Create a Partial Record, page A-7.
Note If the Scanner is to be used with a third-party flow analysis system such as Flow-Cal or PGAS, confirm that the
system will support partial records before enabling the Partials setting. Some systems will not process partial
records correctly.
40
ModWorX Pro Section 3
Note The parameter assigned to Archive #3 is the parameter whose values are displayed in the top LCD readout on
the front panel of the Scanner 2000 or Scanner 210x when the user views the log data. By default, this param-
eter is Flow Run 1 Volume. The user can change the LCD readout parameter by assigning a different parameter
to Archive #3.
Figure 3.23
2. Select the parameter you want to display from the dropdown menu. The text in the box at the bottom of the screen
will be used to identify the parameter in download reports.
3. Click OK to return to the Configure Archive screen.
To delete a parameter, click on the parameter to select it and click Delete Parameter.
To change the order in which the archive parameters appear in log reports, click on an archive parameter and click
Move Up or Move Down.
Note When 16 parameters have been added to the Archive Parameter list, the Add button will be grayed out. No more
parameters can be added without deleting or changing an existing parameter.
4. Click Apply to save the new settings without leaving the Configure Archive screen. Or click OK to save the new
settings and return to the Configuration Menu screen. A warning message will appear, prompting the user to clear the
memory to remove old log records.
5. To clear the memory, click on Tools>Clear Memory.
a. Select which settings you wish to delete: device settings, archive settings, or both.
b. Click OK to delete the settings.
c. Click Done.
41
Section 3 ModWorX Pro
Alarms
To configure the alarms settings, click on the Configure button on the Main Display screen. Then, click on the Alarms
button on the Configuration Menu screen (Figure 3.24).
Figure 3.24
The Configure Alarms screen will appear (Figure 3.25).
Figure 3.25
42
ModWorX Pro Section 3
From the Configure Alarms screen, a user can select up to 16 parameters for triggering an alarm. Users can configure
an alarm to trigger based on a high setpoint, a low setpoint, or either a high or low setpoint. Alternatively, a user can
establish a two-level alarm by assigning a high/low alarm and a high-high/low-low alarm to the same parameter. Only
one alarm type can be assigned to a parameter at a time. Therefore, if a two-level alarm is desired, the parameter must be
entered twice—once to assign a high/low alarm, and once to assign a high-high/low-low alarm.
For example, to enter a two-level differential pressure alarm, the user would enter Differential Pressure as Alarm Item
1 and configure it as a high/low alarm. Then, he would enter Differential Pressure as Alarm Item 2 and configure it as a
high-high/low-low alarm.
Figure 3.26
2. Select an alarm type: high/low or high-high/low-low.
3. Enter high and/or low setpoints as desired and click the dropdown selection beside each setpoint to enable that
setpoint. If a single alarm setpoint is desired—a high setpoint, for example—the user should enable the high setpoint
and disable the low setpoint.
4. Enter a deadband range, if desired. A deadband is a value that determines the point at which an alarm will clear once
it is activated. For example, assume a differential pressure input alarm is assigned a low setpoint of 20 In.H2O, a
high setpoint of 180 In.H2O, and a deadband of 5 In.H2O. A low alarm will not clear until the DP increases to 25
In.H2O (20 plus a deadband of 5). A high alarm will not clear until the DP falls below 175 In.H2O (180 minus a
deadband of 5).
5. Enter the output type desired: latching or non-latching (A latching alarm will remain active even after the associated
parameter reading returns to the assigned range; only a user can clear it. A non-latching alarm will become inactive
after the associated parameter reading returns to the assigned range without any action by a user.)
6. Click OK to save the new settings and return to the Configure Alarms screen.
43
Section 3 ModWorX Pro
Figure 3.27
44
ModWorX Pro Section 3
The Configure User Modbus Registers screen will appear (Figure 3.28).
Figure 3.28
From this screen, a user can group up to 25 commonly polled Modbus holding registers into one block for maximizing
a host system’s polling efficiency. The user can also control the order in which the registers are read, so that the order
matches the host’s existing register structure.
1. To add a register, click on an unassigned register and click Change Register. A User Modbus Register dialog will
appear. Select the register from the dropdown menu provided.
2. To change the order in which the Modbus holding registers are read, click on a Modbus register and click Move Up
or Move Down.
3. Click Apply to save the new settings without leaving the Configure User Modbus Registers screen or click OK to
save the new settings and return to the Configuration Menu screen.
45
Section 3 ModWorX Pro
46
ModWorX Pro Section 4
To set up a flowrun:
1. Click on the Configure button in the task menu bar on the Main Display screen (Figure 4.1).
Figure 4.1
47
Section 4 ModWorX Pro
2. Then, click on Flow Run 1 in the Configuration Menu screen (Figure 4.2).
Figure 4.2
The Configure Flow Run screen will appear (Figure 4.3).
Figure 4.3
48
ModWorX Pro Section 4
Figure 4.4
b. Click on the Display Unit dropdown menus to select a display unit and/or volumetric rate unit. The conversion
factor used to convert the base unit to the display unit will automatically update. To enter a custom unit, select
Custom from the dropdown menu, and enter the desired conversion factor and offset.
c. Click OK to save your changes and return to the Configure Flow Run screen.
3. Under Mass Units, enter the unit in which you want the mass flow displayed. To change the unit, click Edit, change
the unit, and click OK to return to the Configure Flow Run screen.
4. Under Calculation Parameters, click Edit to change the frequency of flow rate calculations. The Edit Synchronized
Time Periods dialog will appear and an arrow will point to the flowrun calculation period setting (Figure 4.5, page
50). The flowrun calculation period is the time period from one flow rate and volume calculation to the next.
Increasing this period decreases the power consumption for the device.
Note The calculation period cannot be longer than the interval archive period.
49
Section 4 ModWorX Pro
Figure 4.5
5. To change the flowrun calculation period
a. Choose a time unit (seconds, minutes, or hours) for the flowrun calculation period.
b. Click on the “+” sign to select the number of units (minutes, seconds, etc.) from the dropdown menu. The
screen will automatically update to show the number of times the calculation will be performed in the interval
log period.
For example, if the interval archive is set to record data every hour, and the flowrun calculation period is once
per minute, the flow rate and volume will be calculated 60 times in a single interval log period (1 hour).
Caution Do not enter the time increment before entering the unit. Doing so will cause the time
increment to be changed when the unit of measure is entered.
c. Click OK to return to the Configure Flow Run screen.
6. If a rate damping factor is needed, click the “+” sign to set the appropriate value.
7. Under Energy Units, enter the unit in which you want the energy measurement displayed. To change the unit, click
Edit, change the unit, and click OK to return to the Configure Flow Run screen.
8. Click Apply to save your settings without leaving the Configure Flow Run screen.
50
ModWorX Pro Section 4
Figure 4.6
Natural Gas
If the flowrun is configured for gas measurement, the selections will appear as shown in Figure 4.7, page 52. Enter the
appropriate fluid properties calculation method and the flow rate calculation method, then click OK to return to the Con-
figure Flow Run screen.
Note When using the NuFlo Cone Meter, choose the Cone-Spool algorithm as the flow rate calculation method.
Steam
If the flowrun is configured for steam measurement, the selections will appear as shown in Figure 4.8, page 53. Enter
the appropriate fluid properties calculation method, flow rate calculation method, wet correction method, and flowrun ac-
cumulation method. Then, click OK to return to the Configure Flow Run screen. Note that the steam flowrun calculation
parameters are interdependent. For example, when a user selects a fluid properties calculation method (IF-97 or IF-97/
James correlation for steam), the available selections for flow rate calculation will change to reflect only those methods
that support the selected fluid property. Selections for wet correction and flowrun accumulation method will, in turn, be
determined by the fluid property and flow calculation method selections.
51
Section 4 ModWorX Pro
Liquid
If the flowrun is configured for liquid measurement, the selections will appear as shown in Figure 4.9, page 53. Enter
the appropriate fluid properties calculation method (generic or MPMS Chapter 11.1). If MPMS is selected, specify the
type of fluid from the dropdown list provided.
For temperature-pressure correction of hydrocarbon liquids having a density greater that 610 kilograms per cubic meter
(excludes LPGs and LNGs), select the MPMS 11.1 calculation.
• Refined Products include gasoline, diesel fuel, fuel oil, and jet fuel. The density at base conditions and the viscosity
at flowing conditions must be input.
• Special Products are other hydrocarbon liquids that don’t fall into the above classifications, e.g., Gasohol. For
these, the fluid’s expansion coefficient “alpha” is also required. Consult MPMS 11.1 for instructions on determining
alpha.
• Select Generic Liquid for measurement of liquids in which temperature-pressure correction is not required (e.g.,
disposal water). The fluid’s density and viscosity must be input.
Select the appropriate flow rate calculation method. Then, click OK to return to the Configure Flow Run screen.
Important If using firmware version 4.10 or earlier, the content of Change Flow Run Calculation Methods
screens will be different. These older firmware versions support API-2540 standards which have
since been replaced by API 11.1 standards. Please see Appendix C - Configuring Liquid Parameters
Using Scanner Firmware 4.10 and Earlier, page C-1 for details.
Fluid Property: Natural Gas (AGA-8 Detail) Fluid Property: Natural Gas (AGA-8 Gross)
Figure 4.7
52
ModWorX Pro Section 4
Fluid Property: Liquid (Generic) Fluid Property: Liquid (MPMS Ch. 11.1)
Figure 4.9
53
Section 4 ModWorX Pro
Flow Rate Calculation Parameters for an Orifice, Cone, or Pitot Meter Run
The flow rate calculation parameters on the Configure Flow Run screen (Figure 4.10) will vary, depending on the calcu-
lation method selected. Use the following steps to configure a Scanner for use with AGA-3 (1991 or 2012), ISO-5167
Orifice, Cone-Spool, Cone-Wafer, Small Bore Orifice and Averaging Pitot Tube (Annubar) calculation methods.
Note For gas measurement, ISO-5167 is applicable only to flow that remains subsonic throughout the measuring
section and where the fluid can be considered single-phase. It is not applicable to the measurement of pulsating
flow. It does not cover the use of orifice plates in pipe sizes less than 50 mm
(2 in.) or more than 1000 mm (39 in.), or with pipe Reynolds numbers below 5000.
Note Analog inputs can be selected as a pressure source or a temperature source only if they are installed and prop-
erly configured. Ensure that sampling is enabled for analog inputs used as pressure or temperature sources. See
Procedure: Configuring an Analog Input, page 81 for details.
–– Differential Pressure Source—If using the MVT, select Integrated Differential Pressure. If using a pressure
transducer via an analog input, select the analog input. The analog input must be configured before it will ap-
pear in the dropdown menu.
–– Static Pressure Source—If using the MVT, select Integrated Static Pressure. If using a pressure transducer
via an analog input, select the analog input. The analog input must be configured before it will appear in the
dropdown menu.
–– Temperature Source—If using the RTD on the Scanner circuit board, select Integrated RTD. If using a tem-
perature transducer via an analog input, select the analog input. The analog input must be configured before it
will appear in the dropdown menu.
Figure 4.10
To calibrate the sensor inputs, click on the blue Calibrate Inputs hyperlink and follow the screen prompts.
54
ModWorX Pro Section 4
Note The user must enter the Maintenance Mode to calibrate an input, and will be returned to the Configure Flow Run
screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs for complete instruc-
tions.
2. Verify the parameters for the differential pressure producer (orifice plate, cone meter). To change a parameter, click
on the blue hyperlink next to the parameter (example: Change Plate or Change Cone Meter) and follow the screen
prompts.
Note The user must enter the Maintenance Mode to change a parameter, and will be returned to the Configure Flow
Run screen upon exiting the Maintenance Mode.
–– Section 9 - Flow Run Maintenance for Orifice Input (AGA-3 or ISO-5167 Orifice)—change orifice plate param-
eters
–– Section 10 - Flow Run Maintenance for Cone Meter Input—change cone meter parameters
–– Section 11 - Flow Run Maintenance for Averaging Pitot Tube (Annubar) Input, page 150—change averaging
pitot tube parameters
3. Where applicable, update the pipe parameters. To change the pipe size, delete the current value and enter a new
value. Select the pipe material and tap type from the dropdown menus provided.
Note When the Averaging Pitot Tube flow rate calculation method is selected, all parameters (including pipe param-
eters) must be changed by clicking on the Change Meter link and navigating to the Maintain Flow Run screen.
See Section 11 - Flow Run Maintenance for Averaging Pitot Tube (Annubar) Input, page 150 for details.
4. Under Base Conditions, adjust the base temperature, base pressure, and atmospheric pressure by deleting existing
values and entering new values, as required.
5. Under Value Control, adjust the low-pressure cutoff, if desired.
6. Click Apply to save your settings.
7. If Cone - Spool or Cone - Wafer is selected as the Flow Rate Calculation Method, a Cone Flow Coefficients section
will be displayed. See Configuring Cone Flow Coefficients, page 60.
8. Proceed to the Fluid Properties section of the screen. See Configuring Fluid Properties, page 57.
55
Section 4 ModWorX Pro
Figure 4.11
2. Under Turbine 1 Information, verify the turbine meter model shown. To change the meter selection, click on the blue
Configure Turbine hyperlink and follow the screen prompts. See Section 13 - Turbine Maintenance for complete
instructions.
Note The user must enter the Maintenance Mode to change a meter selection, and will be returned to the Configure
Flow Run screen upon exiting the Maintenance Mode.
3. Under Input Assignments, verify the pressure source and/or temperature source to be used. If measuring liquids,
only a temperature source will be selectable (Figure 4.12, page 57). To calibrate the sensor inputs from this
screen, click on the blue Calibrate Inputs hyperlink and follow the screen prompts.
Note Analog input sensors are only available for selection as a pressure source or a temperature source if the input
is installed and has been properly configured as a pressure transducer or a temperature transducer. Ensure that
sampling is enabled for analog inputs used as sources. See Procedure: Configuring an Analog Input, page 81,
for details.
Note The user must enter the Maintenance Mode to calibrate an input, and will be returned to the Configure Flow Run
screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs for complete instruc-
tions.
56
ModWorX Pro Section 4
Figure 4.12
4. Under Base Conditions, adjust the base temperature, base pressure, and atmospheric pressure by deleting existing
values and entering new values, as required.
Note When the fluid properties calculation method is selected for an MPMS or generic liquid, the base temperature
and base pressure values are set by the configured base temperature selection in the Fluid Properties section of
the screen.
IMPORTANT The user can change the fluid property calculation method (or any other flow run calculation
method) by pressing the Change button. However, users should be aware that changing the
fluid properties calculation method may result in a change in the flow rate calculation method
selected. Always verify the flow rate calculation method before clicking OK.
57
Section 4 ModWorX Pro
Figure 4.13
2. The parameters displayed in the remainder of the Fluid Properties screen will vary, depending on the fluid property
calculation method and flow rate calculation method selected.
a. If the flow run is configured for gas measurement, the Fluid Property section of the Configure Flow Run
screen will display either Gas Composition information (AGA-8 Detail; see Figure 4.13), or Gross Method
Parameters (AGA-8 Gross; see Figure 4.14, page 59). To change a parameter, click on the blue Change Com-
position or Change Parameters hyperlink and follow the screen prompts.
b. If the flow run is configured for steam measurement, Steam Property settings will be displayed (Figure 4.15,
page 59). Edit the parameters as necessary. To change a parameter, click on the blue Change Properties
hyperlink and follow the screen prompts.
c. If the flow run is configured for liquid measurement, Liquid Parameter settings will be displayed (Figure 4.16,
page 60). Edit the parameters as necessary. To change a parameter, click on the blue Change Parameters
hyperlink and follow the screen prompts.
Note The user must enter the Maintenance Mode to change gas composition, gross method parameters, steam prop-
erties or liquid parameters, and will be returned to the Configure Flow Run screen upon exiting the Maintenance
Mode. See Sections 9, 10, 11, and 12 for complete instructions.
58
ModWorX Pro Section 4
Figure 4.14
Figure 4.15
59
Section 4 ModWorX Pro
Figure 4.16
Figure 4.17
To enter a linear or multi-point flow coefficient, click on the blue Change Flow Coefficients (Cd) hyperlink and follow
the screen prompts.
Note The user must enter the Maintenance Mode to change cone flow coefficients, and will be returned to the Con-
figure Flow Run screen upon exiting the Maintenance Mode. See Section 10 - Flow Run Maintenance for Cone
60
ModWorX Pro Section 4
61
Section 5 ModWorX Pro
Figure 5.1
62
ModWorX Pro Section 5
2. Then, click on Turbine Input 1 (or Turbine Input 2, if applicable) in the Configure screen (Figure 5.2). To configure
a pulse input for a Scanner 2000 or Scanner 210x, you must select Turbine Input 2. To configure a pulse input for a
Scanner 2200, either Turbine Input 1 or Turbine Input 2 may be used.
Figure 5.2
The Configure Turbine Input screen will appear (Figure 5.3).
Figure 5.3
63
Section 5 ModWorX Pro
Note This procedure may be used to set up a second turbine input or a pulse input if an optional expansion board is
installed in a Scanner 2000 or Scanner 210x, or if a Scanner 2200 is installed.
–– Enter a name for the turbine meter. This name will replace the default name “Turbine 1” or “Turbine 2” on
the Main Screen and all other screens involving the turbine input and in archive log reports.
–– Confirm the meter model. To select a different meter model, click on the blue Change Meter hyperlink and
follow the screen prompts.
Note The user must enter the Maintenance Mode to change the meter model, and will be returned to the Configure
Turbine Input screen upon exiting the Maintenance Mode. See Section 13 - Turbine Maintenance for complete
instructions.
–– To display flow volume or flow rate in a different unit, click Edit. The Edit Display Unit dialog will appear
(Figure 5.4). The base unit is the unit in which the Scanner measures flow from the meter. Click on the Dis-
play Unit dropdown menus to select a display unit and/or volumetric rate unit. The conversion factor used
to convert the base unit to the display unit will automatically update. To enter a custom unit, select Custom
from the dropdown menu, and enter the desired conversion factor and offset factor. Click OK to save your
changes and return to the Configure Turbine Input screen.
Figure 5.4
c. Adjust input settings, if desired.
–– To change the sampling period, click Edit. The Edit Synchronized Time Periods dialog will appear, and an
arrow will point to Turbine Input (Figure 5.5, page 65).
–– Select the time period unit desired (seconds, minutes, or hours) using the dropdown menu. Then use the
-/+ buttons to select the time period desired. The screen will automatically display the number of sampling
periods that will occur per flow run calculation period. Click OK to save the settings and return to the Con-
figure Turbine Input screen.
64
ModWorX Pro Section 5
Figure 5.5
–– To change the rate damping factor, click the -/+ buttons. The resulting “settling time” will always be a
multiple of the sampling period. For example, when the sampling period is 4 seconds, entering a damping
factor of 2 will result in a settling time of 8 seconds.
–– To change the input threshold setting, click on the Input Threshold dropdown menu and choose a sensitiv-
ity level for minimizing noise interference in the signal reception. There are four sensitivity levels to choose
from, ranging from 20 mV to 200 mV. A “low” threshold value will allow the Scanner to detect a signal of
approximately 20 mV peak-to-peak, while a “max” threshold value will allow the Scanner to detect only
signals that exceed 200 mV, peak-to-peak.
–– To change the low input frequency cutoff, delete the existing value and enter a new value. The frequency
cut-off is reserved for the turbine input only. When the low-frequency cut-off is configured, the Scanner
will ignore inputs that are less than the user-entered value.
–– To change the low flow rate cutoff, delete the existing value and enter a new value. The low flow rate cutoff
is a setpoint for the accumulation of flow time in the hourly and daily records, and can be applied to either a
turbine input or pulse input.
–– To override the live flow rate value, click the Enable Override field and change the setting to Yes. Then
type in the override value desired.
d. Click Apply to save the new settings.
2. Verify the meter K-factor displayed at the bottom of the Configure Turbine Input screen. To change the K-factor,
click on the blue Change K-Factors hyperlink and follow the screen prompts.
Note The user must enter the Maintenance Mode to change the K-factor, and will be returned to the Configure Turbine
Input screen upon exiting the Maintenance Mode. See Section 13 - Turbine Maintenance for complete instruc-
tions.
If using the turbine meter for a Flow Run with AGA-7, enter the K-factor supplied by the meter manufacturer in
terms of pulses/ACF. Do not correct the K-factor for average temperature and pressure.
65
Section 5 ModWorX Pro
Note SCANNER 2000 OR SCANNER 210X WITH EXPANSION BOARD. A single pulse input is available when an
expansion board is installed on a Scanner 2000 or Scanner 210x. The Turbine Input 2 configuration screen must
be used for configuring this input.
Note SCANNER 2200. Two pulse inputs are available on the Scanner 2200. They are configured using Turbine Input 1
and Turbine Input 2 configuration screens.
The following procedure is based on the configuration of a pulse input using Turbine Input 2. If a Scanner 2200 is in use,
Turbine Input 1 can also be used to configure a pulse input. The same procedure applies.
To set up a pulse input, perform the following steps:
1. Select Turbine Input 2 from the Configure screen. The Configure Turbine Input 2 screen will appear (Figure 5.6).
Figure 5.6
2. Follow the instructions described in Procedure: Setting Up a Turbine Input, steps 1a through 1c.
3. Click the Enable Pulse Input dropdown menu and change the setting to “yes.” The Input Threshold selection will
be disabled (grayed out).
4. Click Apply to save the new settings.
5. Verify the meter K-factor displayed at the bottom of the Configure Turbine Input screen. To change the K-factor,
click on the blue Change K-Factors hyperlink and follow the screen prompts.
Note The user must enter the Maintenance Mode to change the K-factor, and will be returned to the Configure Turbine
Input screen upon exiting the Maintenance Mode. See Section 13 - Turbine Maintenance for complete instruc-
tions.
If using the turbine meter for an AGA-7 flow run, enter the K-factor supplied by the meter manufacturer in terms
of pulses/ACF. Do not correct the K-factor for average temperature and pressure.
66
ModWorX Pro Section 5
67
Section 6 ModWorX Pro
Figure 6.1
68
ModWorX Pro Section 6
2. Then, click the appropriate input in the Inputs section of the Configuration Menu screen (Figure 6.2).
Figure 6.2
Differential Pressure
To configure a differential pressure input, click Differential Pressure on the Configuration Menu screen. The Configure
Differential Pressure screen will appear (Figure 6.3).
Figure 6.3
69
Section 6 ModWorX Pro
Figure 6.4
b. Click on the dropdown Display Unit menu and select a display unit. The conversion factor used to convert the
base unit to the display unit will automatically update.
c. To enter a custom unit, select Custom from the dropdown menu, and enter the desired conversion factor and
offset.
d. Click OK to save your changes and return to the Configure Differential Pressure screen.
Note When the display unit is changed, the units for other input settings (failure value, low pressure cutoff, and over-
ride) will update accordingly.
70
ModWorX Pro Section 6
Figure 6.5
ii. Select the time unit desired (seconds, minutes, etc.) using the dropdown menu.
iii. Use the -/+ buttons to select the number of seconds, minutes, etc. desired. The screen will automatically
divide the flow run calculation period by the input sampling period to determine the number of sampling
periods that will occur per flow run calculation period.
iv. Click OK to save the settings and return to the Configure Differential Pressure screen.
b. To change the damping factor, click the – or + button. The resulting “settling time” will always be a multiple of
the sampling period. For example, when the sampling period is 4 seconds, entering a damping factor of 2 will
result in a settling time of 8 seconds.
c. To change the failure value (the value that will be substituted for a live reading should the unit fail), select the
existing value and type in a new value. In the event of a sensor failure, this value will be substituted for a live
reading in flow calculations.
d. To change the low pressure cutoff, select the existing value and type in a new value. Any differential pressure
measurements below this value will be set to zero. This setting helps prevent inaccuracies in daily and interval
differential pressure records caused by periods of very low differential pressure.
e. To override the live input reading, click the Enable Override field and change the setting to Yes. Then type in
the override value desired.
f. Click Apply to save your settings.
71
Section 6 ModWorX Pro
Sensor Calibration
1. Observe the calibration settings at the bottom of the screen (Figure 6.6). To calibrate the sensor, click on the blue
Calibrate hyperlink in the Sensor Calibration section and follow the screen prompts.
Note The green band across the top of the screen indicates that you are in Maintenance Mode.
Figure 6.6
Note The user must enter the Maintenance Mode to calibrate differential pressure, and will be returned to the Config-
ure Differential Pressure screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying
Inputs for complete instructions.
2. Observe the verification settings at the bottom of the screen (Figure 6.6). To verify the sensor’s calibration, click on
the blue Verify hyperlink and follow the screen prompts.
Note The user must enter the Maintenance Mode to verify differential pressure, and will be returned to the Configure
Differential Pressure screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs
for complete instructions.
72
ModWorX Pro Section 6
Static Pressure
To configure a static pressure input, click on the Static Pressure button on the Configuration Menu screen. The Config-
ure Static Pressure screen will appear (Figure 6.7).
Figure 6.7
73
Section 6 ModWorX Pro
Figure 6.8
Note When the display unit is changed, the units for the other input settings (failure value, low pressure cutoff, and
override value) will update accordingly.
Figure 6.9
ii. Select the time period unit desired (seconds, minutes, etc.) using the dropdown menu.
iii. Use the -/+ buttons to select the number of seconds, minutes, etc. desired. The screen will automatically
divide the flow run calculation period by the input sampling period to determine the number of sampling
periods that will occur per flow run calculation period.
iv. Click OK to save the settings and return to the Configure Static Pressure screen.
74
ModWorX Pro Section 6
b. To change the damping factor, click the -/+ buttons. The resulting “settling time” will always be a multiple of
the sampling period. For example, when the sampling period is 4 seconds, entering a damping factor of 2 will
result in a settling time of 8 seconds.
c. To change the failure value, select the existing value and type in a new value. In the event of a sensor failure,
this value will be substituted for a live reading in flow calculations.
d. To change the low pressure cutoff, select the existing value and type in a new value. Any static pressure
measurements below this value will be set to zero. This setting helps prevent inaccuracies in daily and interval
differential pressure records caused by periods of very low differential pressure.
e. To override the live input reading, click the Enable Override field and change the setting to Yes. Then type in
the override value desired.
f. Click Apply to save the new settings.
Sensor Calibration
1. Observe the calibration settings at the bottom of the screen (Figure 6.10). To calibrate the sensor, click on the blue
Calibrate hyperlink in the Sensor Calibration section and follow the screen prompts.
Figure 6.10
Note The user must enter the Maintenance Mode to calibrate static pressure, and will be returned to the Configure
Static Pressure screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs for
complete instructions.
a. Observe the verification settings at the bottom of the screen (Figure 6.10). To verify the sensor’s calibration,
click on the blue Verify hyperlink and follow the screen prompts.
Note The user must enter the Maintenance Mode to verify static pressure, and will be returned to the Configure Static
Pressure screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs for com-
plete instructions.
75
Section 6 ModWorX Pro
Process Temperature
To configure a process temperature input, click on the Process Temperature button on the Configuration Menu screen.
The Configure Process Temperature screen will appear (Figure 6.11).
Figure 6.11
76
ModWorX Pro Section 6
d. Click OK to save the new settings and return to the Configure Process Temperature screen.
Figure 6.12
Note When the display unit is changed, the units for other input settings (failure, low temperature cutoff, and override)
will update accordingly.
Figure 6.13
ii. Select the time period unit desired (seconds, minutes, etc.) using the dropdown menu.
iii. Use the -/+ buttons to select the number of second, minutes, etc. desired. The screen will automatically divide
the flow run calculation period by the input sampling period to determine the number of sampling periods that
will occur per flow run calculation period.
iv. Click OK to save the new settings and return to the Configure Process Temperature screen.
77
Section 6 ModWorX Pro
b. To change the damping factor, click the -/+ buttons. The resulting “settling time” will always be a multiple of
the sampling period. For example, when the sampling period is 4 seconds, entering a damping factor of 2 will
result in a settling time of 8 seconds.
c. To change the failure value, delete the existing value in the Failure field and type in a new value. In the event
of a sensor failure, this value will be substituted for a live reading in flow calculations.
d. To change the low temperature cutoff, delete the existing value and type in a new value. This setting helps pre-
vent inaccuracies in daily and interval process temperature records caused by periods of very low temperature
by establishing the minimum temperature at which a signal will be recorded.
e. The input source allows the user to specify the temperature input logged. When the flow run is configured for
gas measurement, the setting defaults to RTD and is not configurable (Figure 6.11, page 76). When the flow
run is configured for steam measurement, the setting defaults to Flow
Run 1 Calculated Temperature and the override settings below the Input Source field are automatically disabled
(Figure 6.14). To log a different temperature input, change the Input Source to RTD.
Note Using the RTD setting with a steam flow run configuration DOES NOT CHANGE the temperature input used by
the steam measurement (a calculated input will still be used). It only changes the temperature that will be logged
by the Scanner.
f. To override the live input reading (when applicable), click the Enable Override field and change the setting to
Yes. Then type in the override value desired.
Note A change to the flow rate calculation method or fluid property calculation method will disable any override of
process temperature. If such a change is required, the process temperature override must be re-enabled.
Figure 6.14
Sensor Calibration
1. Observe the calibration settings at the bottom of the screen (Figure 6.15). To calibrate the sensor, click on the blue
Calibrate hyperlink in the Sensor Calibration section and follow the screen prompts.
78
ModWorX Pro Section 6
Figure 6.15
Note The user must enter Maintenance Mode to calibrate process temperature, and will be returned to the Configure
Process Temperature screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying
Inputs for complete instructions.
2. Observe the verification settings at the bottom of the screen (Figure 6.15). To verify the sensor’s calibration, click on
the blue Verify hyperlink and follow the screen prompts.
Note The user must enter Maintenance Mode to verify process temperature, and will be returned to the Configure Pro-
cess Temperature screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs
for complete instructions.
Analog Input
When the optional expansion board is installed on a Scanner 2000 or Scanner 210x or when a Scanner 2200 is installed,
up to two analog inputs can be used to provide pressure or temperature data for calculating flow in accordance with
AGA-7, or to log virtually any data read by an instrument with analog output capability. The transmitter providing the
analog output signal is powered by the Scanner only when the Scanner is externally powered. Wiring diagrams are pro-
vided in Scanner 2000 or Scanner 210x and Scanner 2200 hardware manuals and quick start guides.
The analog inputs can be configured for a 1-5 V, 0-5 V, or 4-20 mA signal. To configure an analog input, click on the
Analog Input 1 or Analog Input 2 button on the Configuration Menu screen (Figure 6.16, page 80). The Configure
Analog Input screen will appear (Figure 6.17, page 80).
Note Analog Input 1 and Analog Input 2 function identically and are configured using identical steps. As an example,
the following procedure is illustrated with screens showing configuration of Analog Input 1. The same steps
should be used to configure Analog Input 2.
79
Section 6 ModWorX Pro
Figure 6.16
Figure 6.17
80
ModWorX Pro Section 6
Note When the transducer type is changed, all calibration settings are deleted, since they will not apply to the newly
selected transducer. See Sensor Calibration, page 84 for further instruction on recalibrating.
Figure 6.18
2. Enter basic sensor information.
a. Enter a unique sensor name, if desired. This name will replace the default name “Analog Input 1” or “Analog
Input 2” on the Main Screen and all other screens involving the analog input and in archive log reports.
b. Enter the sensor model number, if desired.
c. Enter the serial number for the sensor, if desired.
3. Change the Display Unit setting, if desired. The default setting varies, depending on what type of transducer is being
used. Default display unit settings include psig for static pressure, inches H2O for differential pressure, degrees F for
temperature, inches for level, volts for custom voltage input, and milliamps for custom 4-20 mA input.
To change the unit setting for a pressure, temperature, or level sensor,
a. Click Edit. The Edit Display Unit dialog will appear (Figure 6.19).
Figure 6.19
The base unit is the unit in which the Scanner measures the signal output by the transducer or sensor.
b. Click on the dropdown Display Unit menu to select a display unit. The conversion factor used to convert the
base unit to the display unit will automatically update.
81
Section 6 ModWorX Pro
c. Click OK to save the new settings and return to the Configure Analog Input screen.
4. To change the unit setting for a Custom Sensor Type,
a. Click the Edit button next to Sensor Parameters (Figure 6.20).
Figure 6.20
b. The Edit Custom Transducer Range dialog will appear (Figure 6.21).
Figure 6.21
Note When the display unit is changed, the units for other input settings (failure, low reading cutoff, and override) will
update accordingly.
82
ModWorX Pro Section 6
b. The Transducer Range High setting is the output value that will represent the high end of the sensor’s range.
(For example, a high setting of 200 psig for the same 1-5V pressure transmitter would cause the Scanner to
record a 200 psig measurement when it receives a 5-volt input.
c. If a 4-20 mA transmitter is being used, a resistor is required in the field wiring of the analog input device. A
third data field—Shunt Resistance—will appear in the Sensor Parameters section of the screen. Enter the resis-
tance value (ohms) of the resistor installed in that field (Figure 6.22).
Figure 6.22
d. If a Custom Sensor Type is selected, click the Edit button next to Sensor Parameters to access the Transducer
Range Low and Transducer Range High settings (see the Edit Custom Transducer Range dialog in Figure 6.21,
page 82).
6. Adjust parameter values in the Input Settings section, if desired.
a. To change the sampling period, perform the following steps:
i. Click Edit. The Edit Synchronized Time Periods dialog will appear, and an arrow will point to Analog Input
1 or Analog Input 2 (Figure 6.23, page 84).
ii. Select the time period unit desired (seconds, minutes, etc.) using the dropdown menu.
iii. Use the -/+ buttons to select the number of seconds, minutes, etc. desired. The screen will automatically
divide the flow run calculation period by the input sampling period to determine the number of sampling
periods that will occur per flow run calculation period.
iv. Click OK to save the new settings and return to the Configure Analog Input screen.
83
Section 6 ModWorX Pro
Figure 6.23
b. To change the damping factor, click the -/+ buttons. The resulting “settling time” will always be a multiple of
the sampling period. For example, when the sampling period is 4 seconds, entering a damping factor of 2 will
result in a settling time of 8 seconds.
c. To change the failure value, delete the existing value in the Failure field and type in a new value. In the event
of a sensor failure, this value will be substituted for a live reading in flow calculations.
d. To change the low reading cutoff (which may be identified as low pressure cutoff, low temperature cutoff, etc.,
depending on the sensor type selected), delete the existing value and type in a new value. This setting helps
prevent inaccuracies in daily and interval analog input records caused by periods of very low activity by estab-
lishing the minimum output at which a signal will be recorded.
e. To override the live input reading (when applicable), click the Enable Override field and change the setting to
Yes. Then type in the override value desired.
f. Click Apply to save the new settings.
Sensor Calibration
IMPORTANT Analog Input 1 and Analog Input 2 are calibrated at the factory before each unit is shipped,
and typically, field calibration of an analog input is not required. However, if a periodic field
calibration is required or desired, be sure to enter the range BEFORE performing the calibration.
Adjusting the range after an analog input has been calibrated can cause erroneous
analog records.
1. Observe the calibration settings at the bottom of the screen (Figure 6.24, page 85). To calibrate the sensor, click
on the blue Calibrate hyperlink in the Sensor Calibration section and follow the screen prompts.
84
ModWorX Pro Section 6
Figure 6.24
Note The user must enter Maintenance Mode to calibrate an analog input, and will be returned to the Configure Analog
Input screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs for complete
instructions.
2. Observe the verification settings at the bottom of the screen (Figure 6.24). To verify the sensor’s calibration, click on
the blue Verify hyperlink and follow the screen prompts.
Note The user must enter Maintenance Mode to verify an analog input, and will be returned to the Configure Analog
Input screen upon exiting the Maintenance Mode. See Section 8 - Calibrating and Verifying Inputs for complete
instructions.
85
Section 6 ModWorX Pro
86
ModWorX Pro Section 7
Digital Output
To configure a digital output, click the Configure button in the task menu bar on the Main Display screen (Figure 7.1).
Figure 7.1
87
Section 7 ModWorX Pro
Then, click the Digital Output button in the Outputs section of the Configuration Menu screen (Figure 7.2).
Figure 7.2
The Configure Digital Outputs screen will appear (Figure 7.3). By default, the digital output is disabled.
Figure 7.3
To enable a digital output, click Change. The Change Digital Output Mode dialog (Figure 7.4, page 89) will appear,
displaying all of the possible configuration options.
88
ModWorX Pro Section 7
Figure 7.4
Figure 7.5
2. Click OK to save the setting and return to the Configure Digital Outputs screen (Figure 7.6, page 90).
89
Section 7 ModWorX Pro
Figure 7.6
3. In the Output Scaling field, enter the volume that each pulse will represent. The engineering unit that is selected for
displaying volume (as entered on the Configure Flow Run or Configure Turbine screen) will appear as the output
scaling unit on this screen.
4. Adjust the pulse duration (in milliseconds) using the -/+ buttons, if desired.
5. Select the appropriate output state (normally closed or normally open). The screen will automatically display the
minimum pulse period, maximum output frequency, and maximum uncached flow rate, based on the scaling, volume
units, and pulse duration selected.
Figure 7.7
2. Click OK to save the setting and return to the Configure Digital Outputs screen (Figure 7.8, page 91).
90
ModWorX Pro Section 7
Figure 7.8
3. Adjust the output period by clicking the Edit button (Figure 7.9). The output period can be entered in terms of sec-
onds, minutes or hours.
Figure 7.9
4. Click OK to save the setting and return to the Configure Digital Outputs screen.
5. Adjust the pulse duration using the -/+ buttons (default is 10 msec).
6. Change the output state, if necessary. Options are Normally Open and Normally Closed.
7. Click Apply to save the new settings.
91
Section 7 ModWorX Pro
Figure 7.10
2. Click OK to save the setting and return to the Configure Digital Outputs screen (Figure 7.11).
Figure 7.11
92
ModWorX Pro Section 7
3. Click the Select button beside the Parameter field and select a digital output parameter from the dropdown list on the
dialog screen (Figure 7.12).
–– Select “Either High or Low Alarm” to generate a pulse output whenever a configured high or low alarm is
activated. A list of configured alarms will appear in the Trigger Conditions field.
–– Select “Low Alarm Status” to generate a pulse output only when a configured low alarm is activated. A list of
configured alarms will appear in the Trigger Conditions field.
Figure 7.12
–– Select “High Alarm Status” to generate a pulse output only when a configured high alarm is activated. A list of
configured alarms will appear in the Trigger Conditions field.
–– Select “Diagnostic 1” to generate a pulse output when a user-selected failure or override condition occurs. A list
of 16 diagnostic bits will appear in the Trigger Conditions field. Up to four diagnostic parameters (“Diagnostic
1, 2, 3, or 4”) can be used to generate a digital output and up to 16 diagnostic bits can be assigned as triggers for
each diagnostic parameter.
4. To enable the trigger conditions that will generate a digital output, check the appropriate selections in the Trigger
Conditions field.
–– If one of the top three selections was chosen in step 3, only parameters that are configured as alarms will appear
in the Trigger Conditions field (Figure 7.11, page 92). To add a parameter, configure a new alarm in the Con-
figure Alarms menu. See Archive Parameter Setup, page 41, for more details.
–– If a diagnostic parameter is chosen in step 3, check the diagnostic bits that you want to generate a digital output
(Figure 7.13, page 94). Up to 16 diagnostic bits can be selected as triggers for each diagnostic parameter.
5. Click Apply to save the new settings.
93
Section 7 ModWorX Pro
Figure 7.13
•• value below setpoint– the alarm is activated when the assigned value is below the designated threshold (Figure
7.14)
•• value above setpoint – the alarm is activated when the assigned value is above the designated threshold
•• value out of setpoint range (above or below threshold) – the alarm is activated when the assigned value goes
above or below the designated threshold
Figure 7.14
94
ModWorX Pro Section 7
2. Click OK to save the setting and return to the Configure Digital Outputs screen (Figure 7.15). In this example, the
alarm output is configured to activate when the value is below the designated threshold.
Figure 7.15
3. Select the parameter to be assigned to the digital output. Click Select to view the options. A Select Digital Output
Parameter window will appear (Figure 7.16). Click on the dropdown arrow to view all available parameters. Click
on the parameter of your choice, and click OK to save the selection and return to the Configure Digital Outputs
screen.
Figure 7.16
4. Enter low and/or high setpoint values, as applicable. The unit displayed for the setpoint is the unit in which the
selected parameter is configured.
5. Adjust the Alarm Trigger Hold-off value, if necessary, by clicking the Edit button. An Edit Time Period window will
appear (Figure 7.17). The alarm trigger hold-off is the amount of time the Scanner will wait before activating the
alarm when the assigned parameter value exceeds the setpoint value. By default, the hold-off is set at zero. The time
period can be configured in terms of seconds, minutes, or hours.
6. Click OK to save the selection and return to the Configure Digital Outputs screen.
Figure 7.17
95
Section 7 ModWorX Pro
7. Change the output state, if necessary (Figure 7.15, page 95). Options are Normally Open and Normally Closed.
8. Change the Output Latching setting, if necessary (Figure 7.15, page 95). Options are Non-latching and Latching.
If the alarm is latched, it will remain active once it is activated, even if the assigned parameter value returns to a
normal range. The alarm must be manually cleared by the operator. If the alarm is non-latching, the alarm will clear
when the assigned parameter value returns to a normal range without intervention from the operator.
9. Click Apply to save the new settings.
Figure 7.18
2. Click OK to save the setting and return to the Configure Digital Outputs screen (Figure 7.19).
Figure 7.19
3. Note that the Pulse Out 1 Pulses holding register controls the enabling and disabling of this output. See the protocol
section of the appropriate Scanner Hardware User Manual for more information about register configuration.
96
ModWorX Pro Section 7
4. Change the output state, if necessary. Options are Normally Open or Normally Closed.
5. Click Apply to save the new settings.
Analog Output
An analog output is standard on all Scanner 2105 and 2200 devices. To use the analog output to transmit data to another
current-reading device, an analog readout device must be connected to the Scanner and the Scanner must be powered
either by an external power supply or loop power (Scanner 2105 only). Wiring diagrams are provided in the appropri-
ate Scanner Hardware User Manual, in the Scanner 2000 Expansion Board Quick Start Guide, and in the Scanner 2200
Quick Start Guide.
To configure an analog output, click the Configure button in the task menu bar on the Main Display screen (Figure 7.20).
Figure 7.20
Then, click the Analog Outputs button in the Outputs section of the Configuration Menu screen (Figure 7.21).
Figure 7.21
97
Section 7 ModWorX Pro
The Configure Analog Output screen will appear (Figure 7.22). By default, the analog output is disabled.
Figure 7.22
To enable an analog output, click Change. The Change Analog dialog will appear (Figure 7.23).
Figure 7.23
Select Track Measured/Calculated Parameter and click OK. The Configure Analog Output screen will automatically
update to display fields for entering configuration parameters.
98
ModWorX Pro Section 7
Figure 7.24
2. Click Select to choose a parameter from the Select Analog Output Parameter dialog (Figure 7.25).
Figure 7.25
3. Click OK to close the Select Analog Output Parameter dialog and return to the Configure Analog Output screen.
4. In the Low Setpoint (for 4 mA) field, enter the parameter value that the 4 mA output signal will represent. The unit
associated with this value will be the unit configured for the parameter selected in step 1.
5. In the High Setpoint (for 20 mA) field, enter the parameter value that the 20 mA output signal will represent. The
unit associated with this value will be the unit configured for the parameter selected in step 1.
6. Click OK to save the settings and exit the Configure Analog Outputs screen, or click Apply to save the settings and
proceed with a calibration of the analog output, as described below.
The output is proportional to the parameter value between the low and high setpoints.
99
Section 7 ModWorX Pro
Figure 7.26
4. Read the output value as read by the receiving device, and enter it in the screen (Figure 7.26, right). Click OK to
calculate the adjustment. The adjustment will appear in the box next to the 4 mA Setting button (Figure 7.28, page
101).
5. Click on the 20 mA Setting button to send a 20 mA signal from the Scanner to the receiving device. The Full Scale
(20 mA) Adjustment dialog will appear (Figure 7.27, left).
Figure 7.27
6. Read the output value as read by the receiving device, and enter it in the screen (Figure 7.27, right). Click OK to
calculate the adjustment. The adjustment will appear in the box next to the 20 mA Setting button (Figure 7.28, page
101).
100
ModWorX Pro Section 7
Figure 7.28
7. Click Apply to save the adjustments. The adjustments will appear under the Active Output Calibration section of the
screen. Click OK to return to the Configuration menu, or click Apply to save the new settings without leaving the
Configure Analog Outputs screen.
8. To clear adjustments and repeat the calibration process, click the Clear Adjustments button; then, repeat steps 3
through 6.
9. To test the analog output following adjustments, click the Test Output button at the bottom of the Configure Analog
Outputs screen. A Test Analog Output dialog will appear (Figure 7.29).
Figure 7.29
10. Enter the output value (in mA) you wish to apply and click Apply. The Scanner will send the user-specified signal
to the receiving device. If the trim was applied correctly, the readout of the receiving device should match the test
value entered in the Test Analog Output screen. If the values do not match, repeat steps 3 through 7. Reading an
input value from the receiving device incorrectly or transposing numbers typed into the screen may be the cause of
the discrepancy in values.
Note The Scanner will not allow adjustments of more than 500 uA. If the value read by the receiving device varies by
more than 500 uA from the value read by the Scanner, equipment failure should be investigated.
101
Section 7 ModWorX Pro
102
ModWorX Pro Section 8
Figure 8.1
The instrument will enter the Maintenance mode. Click Yes at the Enter Maintenance Mode dialog to proceed. (Mainte-
nance Mode is a requirement for all calibration, verification, and set zero offset tasks.)
The Change Calibration Task dialog will appear (Figure 8.2, page 104). On this screen, the user selects both the calibra-
tion task to be performed (calibrate, zero, or verify) and the input parameter to be calibrated (differential pressure, static
pressure, or process temperature). If the optional expansion board is installed in a Scanner 2000 or Scanner 210x, or if a
Scanner 2200 is installed, the listing of selectable input parameters will include Analog Input 1 and Analog Input 2.
Note If an input is disabled (sampling period is set to zero), the input cannot be calibrated, and the option will be
grayed out and marked as “disabled.”
103
Section 8 ModWorX Pro
Figure 8.2
Note For more information about the appearance of column headings on the Calibrate screen, see Change Format of
Newly-Saved Calibrations, page A-12.
104
ModWorX Pro Section 8
Figure 8.3
Note The current calibration data (in this example, factory calibration data) appears on the right half of the screen, and
the new calibration data will be entered on the left half of the screen.
4. In Figure 8.3, the “Calibration Working Units” located above the new calibration table shows the unit of measure to
which the differential pressure will be calibrated. To use a different unit, select the desired unit from the Calibration
Working Units dropdown menu.
The “working unit” setting is independent of the unit used to display differential pressure. The “working unit” ap-
plies only to the calibration process itself. Therefore, a user who configures his device to display differential pressure
as In H2O can calibrate the device in In H2O or in a different unit; in either case, differential pressure will continue
to be displayed on the LCD, on the Main Display screen of the software, and in calibration reports as In H2O. A user
can view current calibration values in any available unit simply by selecting a new “working unit” on the Calibrate
screen. The values will automatically convert to the unit selected.
5. Apply a known pressure to the calibration test device (for example, 25 In H2O) and click Acquire Point (Figure
8.3). An Enter Applied Value dialog will appear (Figure 8.4).
Figure 8.4
105
Section 8 ModWorX Pro
6. Enter the applied pressure value (for example, 25) in this field, and click OK. A screen will appear, showing the live
value, the measured value, and a stability bar showing the progression of the calibration (Figure 8.5). The live value
is the instantaneous reading of the pressure input, and the measured value is the average of all samples read.
Figure 8.5
7. To view the progression of the calibration process in a graph, click Show Trend (Figure 8.5). This view also shows
the spread of values and the number of sample readings taken (Figure 8.6).
Figure 8.6
106
ModWorX Pro Section 8
When the Scanner reads 10 consecutive readings (within the tolerance based on the range of the sensor), it will
“lock on” to that data set and display a “measured value” (Figure 8.7).
Figure 8.7
Note Should the Scanner fail to lock on to a calibration value within a reasonable time period, the user can terminate
the calibration process from the trend graph screen by clicking on the Force Lock On button in the lower left
corner of the screen. The user can then recalibrate the point by clicking the Recalibrate Point button on the
Configure Differential Pressure screen.
8. When a measured value is indicated, click OK to save the calibration point and return to the Calibrate Differen-
tial Pressure screen (Figure 8.8, page 108). The New Calibration table is updated with the applied pressure, the
measured value, and a percentage of change. The percentage of change is expressed as a percentage of full scale,
and is a measure of how the calculated value will change based on the new calibration with respect to the previous
calibration.
107
Section 8 ModWorX Pro
Figure 8.8
9. Repeat steps 5 through 8 to enter up to 12 differential points.
10. To store additional information about this calibration, enter a note in the Note field. The note will be saved with the
calibration data when the calibration data is written to memory.
11. Click Save Changes in the top right corner of the screen to save the differential pressure calibration. The new values
will not be applied until the user exits the Maintenance Mode.
12. To perform a different calibration task or to calibrate a different input parameter, click Change Task. The Change
Calibration Task dialog will reappear, allowing the user to make a new selection.
13. When all calibration tasks are completed and all changes saved, click Exit Maintenance Mode. Inputs will be re-
stored to live conditions and the Main Display screen will reappear.
Figure 8.9
3. Click OK to save the selection. The calibration points will be transferred to the New Calibration table on the Cali-
brate Differential Pressure screen.
To recalibrate a data point, click on the point, apply the designated amount of pressure to the Scanner using the calibra-
tion test device, and click Recalibrate Point. It is not necessary to re-enter applied values.
108
ModWorX Pro Section 8
Note Performing a multipoint calibration will clear the Set Zero Offset.
1. On the Change Calibration Task screen, click Set Zero Offset as the Calibration Task, and click the input parameter
of your choice. For this example, Static Pressure is the input parameter to be zeroed (Figure 8.10).
2. Click OK to save the settings.
Figure 8.10
3. The Set Zero Offset – Static Pressure screen will appear, and the input parameter and calibration task selections you
made will appear at the top of the screen (Figure 8.11, page 110).
Note that the current zero offset data appears on the right half of the screen, and the new zero offset data will be
entered on the left half of the screen.
109
Section 8 ModWorX Pro
Figure 8.11
4. Note the unit displayed as the Calibration Working Unit. If a different unit will be used to calibrate the zero offset,
select the desired unit from the Calibration Working Units dropdown menu.
The “working unit” setting is independent of the unit used to display static pressure. The “working unit” applies
only to the calibration process itself. Therefore, a user who configures his device to display static pressure as psi can
calibrate the zero offset in psi or in a different unit; in either case, static pressure will continue to be displayed on
the LCD, on the Main Display screen of the software, and in calibration reports as psi. A user can view current zero
offset values in any available unit simply by selecting a new “working unit” on the Calibrate screen. The values will
automatically convert to the unit selected.
5. Apply 0 psig pressure to the calibration test device and click Acquire Zero Offset. An Enter Applied Value dialog
will appear (Figure 8.12).
Figure 8.12
110
ModWorX Pro Section 8
6. Enter zero (0) in this field, and click OK. A screen will appear, showing the live value, the measured value, and a
stability bar showing the progression of the calibration (Figure 8.13). The live value is the instantaneous reading of
the pressure input, and the measured value is the average of all samples read.
Figure 8.13
7. To view the progression of the calibration process in a graph, click Show Trend (Figure 8.13). This graph view also
shows the spread of values and the number of sample readings taken (Figure 8.14).
Figure 8.14
111
Section 8 ModWorX Pro
When the Scanner reads 10 consecutive readings (within the tolerance based on the range of the sensor), it will
“lock on” to that data set and display a “measured value” (Figure 8.15).
Figure 8.15
8. When a measured value is indicated, click OK to save the calibrated zero offset value and return to the Set Zero
Offset – Static Pressure screen. The calibrated value will appear in the Zero Offset field (Figure 8.16).
Note The zero offset is the value that is added to the instrument’s reading to obtain the desired (applied) value. If the
sensor is reading high, the zero offset will be a negative value.
Figure 8.16
112
ModWorX Pro Section 8
9. To store additional information about this calibration, enter a note in the Note field. The note will be saved with the
calibration data when the calibration data is written to memory.
10. Click Save Changes to save the zero offset calibration. The new values will not be applied until the user exits the
Maintenance Mode.
11. To perform a different calibration task or to calibrate a different input parameter, click Change Task. The Change
Calibration Task dialog will reappear, allowing the user to make a new selection.
12. When all calibration tasks are completed and all changes are saved, click Exit Maintenance Mode. Inputs will be
restored to live conditions and the Main Display screen will reappear.
Figure 8.17
113
Section 8 ModWorX Pro
3. The Verify Process Temperature screen will appear, and the input parameter and calibration task selections you made
will appear at the top of the screen (Figure 8.18).
Figure 8.18
Note that the current verification data appears on the right half of the screen, and the new verification data will be
entered on the left half of the screen.
4. Note the unit displayed as the Verification Working Unit. If a different unit will be used to verify the device, select
the desired unit from the dropdown Verification Working Units menu.
The “working unit” setting is independent from the unit used to display process temperature. The “working unit”
applies only to the verification process itself. Therefore, a user who configures his device to display process tem-
perature as degrees F can calibrate the device in degrees F or in a different unit; in either case, process temperature
will continue to be displayed on the LCD, on the Main Display screen of the software, and in calibration reports as
degrees F. A user can view current verification values in any available unit simply by selecting a new “working unit”
on the Verify screen. The values will automatically convert to the unit selected.
5. Apply a known temperature to the calibration test device (for example, 50 degrees F) and click Acquire Point (Fig-
ure 8.18). An Enter Applied Value dialog will appear (Figure 8.19).
Figure 8.19
114
ModWorX Pro Section 8
6. Enter the applied temperature value (for example, 50) in this field, and click OK. A screen will appear, showing the
live value, the measured value, and a stability bar showing the progression of the verification (Figure 8.20). The live
value is the instantaneous reading of the temperature input, and the measured value is the average of all samples
read.
Figure 8.20
7. To view the progression of the verification process in a graph, click Show Trend (Figure 8.20). This graph view also
shows the spread of values and the number of sample readings taken (Figure 8.21).
Figure 8.21
8. When the Scanner reads 10 consecutive readings (within the tolerance based on the range of the sensor), it will “lock
on” to that data set and display a “measured value” (Figure 8.22, page 116).
115
Section 8 ModWorX Pro
Figure 8.22
Note Should the Scanner fail to lock on to a verification value within a reasonable time period, the user can terminate
the verification process from the trend graph screen by clicking on the Force Lock On button in the lower left
corner of the screen. The user can then re-verify the point by clicking the Re-verify Point button on the Verify
Process Temperature screen.
9. When a measured value is indicated, click OK to save the verification and return to the Verify Process Temperature
screen (Figure 8.23). The New Calibration table is updated with the applied temperature, the measured temperature,
and a percentage of error. The percentage of error is expressed as a percentage of full scale, and is calculated with
respect to the applied value.
Figure 8.23
10. Repeat steps 5 through 8 to enter up to 12 verification points.
116
ModWorX Pro Section 8
11. To store additional information about this verification, enter a note in the Note field. The note will be saved with the
verification data when the verification data is written to memory.
12. Click Save Changes to save the temperature verification. The new values will not be applied until the user exits the
Maintenance Mode.
13. To perform a different calibration or verification task, click Change Task. The Change Calibration Task dialog will
reappear, allowing the user to make a new selection.
14. When all calibration/verification tasks are completed and all changes are saved, click Exit Maintenance Mode.
Inputs will be restored to live conditions and the Main Display screen will reappear.
117
Section 8 ModWorX Pro
118
ModWorX Pro Section 9
Gas Measurement: AGA-3 (1992 or 2012), ISO-5167 Orifice, or Small Bore Orifice (based on ASME MFC-14M)
Liquid Measurement: Liquid Orifice AGA-3 (1992 or 2012), Liquid Orifice (based on ISO-5167), or Small Bore
Orifice (based on ASME MFC-14M).
See Section 4 - Setting Up a Flowrun for instructions on selecting the flow run calculation method.
For gas measurement, ISO-5167 is applicable only to flow that remains subsonic throughout the measuring
section and where the fluid can be considered single-phase. It is not applicable to the measurement of pulsating
flow. It does not cover the use of orifice plates in pipe sizes less than 50 mm or more than 1000 mm, or for pipe
Reynolds numbers below 5000.
For liquid measurement, the Small Bore Orifice flow rate calculation method is appropriate for 1/2-in. to 1-1/2-in.
meter sizes with a beta ratio between 0.1 and 0.75. The ASME small bore orifice meter can be used to measure
natural gas, steam, and liquids.
Figure 9.1
119
Section 9 ModWorX Pro
2. Click Yes at the Enter Maintenance Mode prompt. The Change Flow Run Maintenance Task screen will appear
(Figure 9.2).
Figure 9.2
3. Flow Run 1 is automatically selected.
4. Select the maintenance task to be performed. The maintenance tasks displayed will vary, depending on the type of
flow being measured—gas, steam or liquid. Examples are provided in the following procedures.
5. Click OK to save the selections. The Maintain Flow Run screen will appear (Figure 9.3).
Figure 9.3
By default, engineering units are provided in US standard format. If metric units are required, click on the Tools button
in the task bar at the top of the screen, and select Options from the dropdown menu. Click on the Units System drop-
down menu to change the units selection to Metric. See Figure A.17, page A-12.
120
ModWorX Pro Section 9
Natural Gas (AGA-8 Detail) Natural Gas (AGA-8 Gross) Steam (IF-97 or IF-97+James)
121
Section 9 ModWorX Pro
Figure 9.5
2. Note that current orifice plate parameters appear on the right half of the screen. If you are making a minor change to
current parameters, click Copy from Current. Current parameters will be pasted into the fields on the left. This will
help to minimize the number of fields that have to be changed.
3. Update the following plate parameters on the left half of the screen, as necessary:
a. plate size – click in the field and type in a value
b. reference temperature – click in the field and type in a value; this value is typically provided on a test report
issued by the laboratory that calibrated the orifice plate
c. plate material – choose a material from the dropdown list; use the scroll bar to view the entire list of materials
d. model number (optional)
e. serial number (optional)
4. Enter a note in the Note field, if desired. The note will be stored with the new plate parameter settings.
5. Click Save Changes in the upper right corner of the screen. The new settings will be saved and will be made effec-
tive when the user exits the Maintain Flow Run screen.
6. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
7. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
122
ModWorX Pro Section 9
Figure 9.6
a. Under the Maintenance Task heading, click Change Gas Composition.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 9.7, page 124).
2. Note that current gas composition appears on the right half of the screen and the new gas composition will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Gas Com-
position section. The calculation method is read-only. To change this method, see Configuring Fluid Properties, page
57.
4. In the Analysis Reference No. field, enter a reference number from the gas analysis report, if desired.
5. Update the Reference Conditions field, if necessary. This selection identifies the ideal mass heating value table for
use in calculating the heating value of the gas composition and sets the reference temperature and pressure used in
converting the user-supplied heating value into the heating value mass basis required for energy calculations.
6. In the Heating Value field, specify a calculated or manually entered value. If the value is entered manually, enter the
heating value in the field provided.
7. In the Specific Gravity field, specify a calculated or manually entered value. If the value is entered manually, enter
the specific gravity value in the field provided.
8. Enter the viscosity of the measured fluid in the field provided. If you are unsure of the value, enter 0.010268.
9. Enter the isentropic exponent of the measured fluid in the field provided. If you are unsure of the value, enter 1.3.
Note Disregard the Gas Fraction field. This setting is not currently used in flow calculations.
123
Section 9 ModWorX Pro
Figure 9.7
10. To change the mole percentages of individual gas constituents,
a. Locate the gas constituent to be changed.
b. Double-click the constituent or click on the Change Mole % button and enter the appropriate percentage in the
dialog provided.
c. Repeat steps 10a and 10b until all constituents are entered.
Note When the AGA-8 Detail method is used, all constituent percentages entered must total 100%.
11. To load the current gas composition or a previously saved composition, click Load Composition… and select the set
of gas composition values from the dropdown menu in the dialog provided. If only minor changes are required to an
existing composition, this can save time by minimizing the number of constituents that have to be changed.
12. Enter a note in the Note field, if desired (for example, to identify the gas composition for future reference). The note
will be stored with the gas composition settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gas Composi-
tion and will be put into effect when the user exits the Maintain Flow Run screen.
124
ModWorX Pro Section 9
14. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
15. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
Figure 9.8
a. Under the Maintenance Task heading, click Change Gross Method Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 9.9, page 126).
2. Note that current gross method parameters appear on the right half of the screen and the new gross method parame-
ters will be entered in the left half of the screen. If you are making a minor change to current parameters, click Copy
from Current. Current parameters will be pasted into the fields on the left. This will help to minimize the number of
fields that have to be changed.
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Gross
Method Parameters section. The calculation method is read-only. To change this method, see Configuring Fluid
Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the gas analysis report, if desired.
5. Update the Reference Conditions field, if necessary. This selection sets the reference temperature and pressure for
air density, which is used in converting the user-supplied heating value into the heating value mass basis required for
energy calculations.
6. Enter the heating value of the measured fluid in the field provided.
7. Enter the specific gravity of the measured fluid in the field provided.
8. Enter the viscosity of the measured fluid in the field provided. If you are unsure of the value, enter 0.010268.
9. Enter the isentropic exponent of the measured fluid in the field provided. If you are unsure of the value, enter 1.3.
10. Enter the carbon dioxide content (mole %) in the field provided.
125
Section 9 ModWorX Pro
Figure 9.9
11. Enter the nitrogen content (mole %) in the field provided.
12. Enter a note in the Note field, if desired (for example, to identify the gross method settings for future reference). The
note will be stored with the gross method settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gross Method
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
14. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
15. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
126
ModWorX Pro Section 9
Figure 9.10
a. Under the Maintenance Task heading, click Change Steam Properties.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 9.11).
Figure 9.11
2. Note that current steam properties appear on the right half of the screen and the new steam properties will be entered
in the left half of the screen. If you are making a minor change to current parameters, click Copy from Current. Cur-
rent parameters will be pasted into the fields on the left. This will help to minimize the number of fields that have to
be changed.
127
Section 9 ModWorX Pro
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Steam
Properties section. The calculation method is read-only. To change this method, see Configuring Fluid Properties,
page 57.
4. In the Analysis Reference No. field, enter a reference number from the steam analysis report, if desired.
Note When a steam measurement algorithm is selected, enthalpy, specific gravity, viscosity, and isentropic exponent
values are automatically calculated. Proceed to step 5.
Figure 9.12
a. Under the Maintenance Task heading, click Change MPMS Liquid Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 9.13, page 129).
128
ModWorX Pro Section 9
Figure 9.13
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: Special Liquids) appears near the top of
the New MPMS Liquid Properties section. The calculation method is read-only. To change this method, see Config-
uring Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters, or
with values supplied by the user. To use the set of default values, click Load Default Params near the bottom of
the screen.
5. In the Density Source field, select a source to be used for the density value: Absolute Density, Specific Gravity, or
API Gravity.
6. Enter the Base Density (Absolute, Specific Gravity or API Gravity) value of the liquid. Density is assumed to be at
base temperature.
7. If Special Products is selected as the fluid type for the MPMS fluid properties calculation method, enter the fluid’s
expansion coefficient “alpha” (Figure 9.14, page 130) in accordance with MPMS Chapter 11.1. You may enter a
numeric value, or click the Select Alpha button to select a predetermined alpha value for products such as ethanol
and gasohol.
129
Section 9 ModWorX Pro
Figure 9.14
8. For best accuracy, enter the viscosity value of the liquid at flowing temperature. If the viscosity is unknown, use the
recommended default value for the liquid type selected.
9. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
10. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current MPMS Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
11. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
12. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
130
ModWorX Pro Section 9
Figure 9.15
a. Under the Maintenance Task heading, click Change Generic Liquid Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 9.16, page 132).
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: Generic Liquid) appears near the top of
the New Generic Liquid Properties section. The calculation method is read-only. To change this method, see Config-
uring Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
131
Section 9 ModWorX Pro
Figure 9.16
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters
which are the values for water. While the default values are specific to water, they indicate the value range that
is applicable for these parameters. To load the default values, click Load Default Params near the bottom of the
screen.
5. In the Density Source field, select a source to be used for the density value: Absolute Density or Specific Gravity.
6. Enter the Base Density (Absolute or Specific Gravity) value of the liquid. If the base density of the liquid is un-
known, enter the flowing density into both the Base Density and Flowing Density fields.
7. Enter the Flowing Density (Absolute or Specific Gravity) value, if applicable.
8. Enter the viscosity of the liquid. Viscosity is assumed to be at flowing temperature. If the viscosity is unknown, a
value of 1.0 cP (viscosity of water) is recommended.
9. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
10. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Generic Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
11. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
12. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
132
ModWorX Pro Section 9
133
Section 10 ModWorX Pro
Figure 10.1
2. Click Yes at the Enter Maintenance Mode prompt. The Change Flow Run Maintenance Task screen will appear
(Figure 10.2, page 135).
134
ModWorX Pro Section 10
Figure 10.2
3. Select the maintenance task to be performed. The maintenance tasks displayed will vary, depending on the type of
flow being measured - gas, steam or liquid. Examples are provided in the following procedures.
4. Click OK to save the settings. The Maintain Flow Run screen will appear (Figure 10.3).
Figure 10.3
135
Section 10 ModWorX Pro
Natural Gas (AGA-8 Detail) Natural Gas (AGA-8 Gross) Steam (IF-97 or IF-97+James)
136
ModWorX Pro Section 10
Figure 10.5
2. Note that current cone parameters appear on the right half of the screen. To make a minor change to current pa-
rameters, click Copy from Current. Current parameters will be pasted into the fields on the left. This will help to
minimize the number of fields that have to be changed.
3. Enter new cone parameters as necessary:
a. model number (optional)
b. beta ratio – click in the field and type in a value (this value is typically recorded on the nameplate of the cone
meter and/or recorded in the calibration data supplied with the cone meter)
c. meter size (ID) – click in the field and type in the value – (this value is typically recorded on the nameplate of
the cone meter)
d. reference temperature – click in the field and type in a value; this value is typically provided on a test report
issued by the laboratory that calibrated the orifice plate; if the cone reference temperature is not known, enter
68ºF (US standard units) or 20ºC (metric units).
e. cone material – choose a material from the dropdown list; use the scroll bar to view the entire list
f. meter material – choose a material from the dropdown list; use the scroll bar to view the entire list
g. serial number (optional)
4. Enter a note in the Note field, if desired. The note will be stored with the new cone parameter settings.
5. Click Save Changes in the upper right corner of the screen. The new settings will be saved and will be made effec-
tive when the user exits the Maintain Flow Run screen.
6. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
7. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
Note In addition to entering cone meter parameters, it is important to verify that the flow coefficient(s) are appropri-
ate for the cone parameters you have entered. See Procedure: Change Flow Coefficients (Cd), page 138, for
details.
137
Section 10 ModWorX Pro
Natural Gas (AGA-8 Detail) Natural Gas (AGA-8 Gross) Steam (IF-97 or IF-97+James)
138
ModWorX Pro Section 10
Figure 10.7
2. Note that current flow coefficients appear on the right half of the screen. To make a minor change to an existing set
of flow coefficients, click Load Coefficients From… and select the set of coefficients from the dialog provided. The
coefficients selected will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. To enter a linear flow coefficient,
a. Click Linear Flow Coefficient in the Entry Mode box.
b. Click the Enter Linear Factor button.
c. Type in the new linear coefficient in the dialog provided.
d. Click OK to update the entry in the New Flow Coefficients section of the Maintain Flow Run screen.
4. To enter a multi-point meter factor,
a. Click Multi-point Flow Coefficient in the Entry Mode box.
b. Click Add Coefficient and enter the appropriate Reynolds number and flow coefficient values in the dialog
provided.
c. Click OK to update the entry in the New Flow Coefficients section of the Maintain Flow Run screen.
d. Repeat steps 4a through 4c as required to enter up to 12 calibration points.
5. Enter a note in the Note field if you want to capture additional information about the flow coefficient change. The
note will be stored with the new flow coefficients.
6. Click Save Changes in the upper right corner of the screen. The new settings will be saved and will be made effec-
tive when the user exits the Maintain Flow Run screen.
7. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
8. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
139
Section 10 ModWorX Pro
Figure 10.8
a. Under the Maintenance Task heading, click Change Gas Composition.
b. Click OK to save your selections and return to the Maintain Flow Run screen (Figure 10.9, page 141).
2. Note that current gas composition appears on the right half of the screen and the new gas composition will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Gas Com-
position section. The calculation method is read-only. To change, see Configuring Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the gas analysis report, if desired.
5. Update the Reference Conditions field, if necessary. This selection identifies the ideal mass heating value table for
use in calculating the heating value of the gas composition and sets the reference temperature and pressure used in
converting the user-supplied heating value into the heating value mass basis required for energy calculations.
6. In the Heating Value field, specify whether the heating value is to be calculated or entered manually. If the value is
entered manually, enter the heating value in the field provided.
7. In the Specific Gravity field, specify whether specific gravity is to be calculated or entered manually. If the value is
entered manually, enter the specific gravity in the field provided.
8. Enter the viscosity of the measured fluid in the field provided. If you are unsure of the value, enter 0.010268.
9. Enter the isentropic exponent of the measured fluid in the field provided. If you are unsure of the value, enter 1.3.
10. To change the mole percentages of individual gas constituents,
a. Locate the gas constituent to be changed.
b. Double-click the constituent or click on the Change Mole % button and enter the appropriate percentage in the
dialog provided.
c. Repeat steps 10a and 10b until all constituents are entered.
140
ModWorX Pro Section 10
Figure 10.9
Note When the AGA-8 Detail method is used, all constituent percentages entered must total 100%.
11. To load the current gas composition or a previously saved composition, click Load Composition… and select the
set of gas composition values from the dropdown menu in the dialog provided. If the desired composition is very
similar to an existing composition, loading a composition can reduce setup time by eliminating the need to enter
each constituent separately.
12. Enter a note in the Note field, if desired (for example, to identify the gas composition for future reference). The note
will be stored with the gas composition settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gas Composi-
tion and will be put into effect when the user exits the Maintain Flow Run screen.
14. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
15. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
141
Section 10 ModWorX Pro
Figure 10.10
a. Under the Maintenance Task heading, click Change Gross Method Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 10.11, page 143).
2. Note that current gross method parameters appear on the right half of the screen and the new gross method parame-
ters will be entered in the left half of the screen. If you are making a minor change to current parameters, click Copy
from Current. Current parameters will be pasted into the fields on the left. This will help to minimize the number of
fields that have to be changed.
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Gross
Method Parameters section. The calculation method is read-only. To change this method, see Configuring Fluid
Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the gas analysis report, if desired.
5. Update the Reference Conditions field, if necessary. This selection sets the reference temperature and pressure for
air density, which is used in converting the user-supplied heating value into the heating value mass basis required for
energy calculations.
6. Enter the heating value of the measured fluid in the field provided.
7. Enter the specific gravity of the measured fluid in the field provided.
8. Enter the viscosity of the measured fluid in the field provided. If you are unsure of the value, enter 0.010268.
9. Enter the isentropic exponent of the measured fluid in the field provided. If you are unsure of the value, enter 1.3.
10. Enter the carbon dioxide content in the field provided.
142
ModWorX Pro Section 10
Figure 10.11
11. Enter the nitrogen content in the field provided.
12. Enter a note in the Note field, if desired (for example, to identify the gross method settings for future reference). The
note will be stored with the gross method settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gross Method
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
14. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
15. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
143
Section 10 ModWorX Pro
Figure 10.12
a. Under the Maintenance Task heading, click Change Steam Properties.
b. Click OK to save your selections and return to the Maintain Flow Run screen (Figure 10.13).
Figure 10.13
2. Note that current steam properties appear on the right half of the screen and the new steam properties will be entered
in the left half of the screen.
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Steam
Properties section. The calculation method is read-only. To change this method, see Configuring Fluid Properties,
page 57.
144
ModWorX Pro Section 10
4. In the Analysis Reference No. field, enter a reference number from the steam analysis report, if desired.
5. If wet correction is enabled, enter the steam quality (percent) in the Steam Quality/Dryness field.
6. Enter a note in the Note field, if desired (for example, to identify the steam properties for future reference). The note
will be stored with the steam property settings.
7. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Steam Proper-
ties and will be put into effect when the user exits the Maintain Flow Run screen.
8. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
9. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
10. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
11. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
Figure 10.14
a. Under the Maintenance Task heading, click Change MPMS Liquid Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 10.15, page 146).
145
Section 10 ModWorX Pro
Figure 10.15
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: Crude Oil) appears near the top of the
New MPMS Liquid Properties section. The calculation method is read-only. To change this method, see Configuring
Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters, or
with values supplied by the user. To use the set of default values, click Load Default Params near the bottom of
the screen.
5. In the Density Source field, select a source to be used for the density value: Absolute Density, Specific Gravity, or
API Gravity.
6. Enter the Base Density (Absolute, Specific Gravity or API Gravity) value of the liquid. Density is assumed to be at
base temperature.
7. If Special Products is selected as the fluid type for the MPMS fluid properties calculation method, enter the fluid’s
expansion coefficient “alpha” (Figure 9.14, page 135) in accordance with MPMS Chapter 11.1. You may enter a
numeric value, or click the Select Alpha button to select a predetermined alpha value for products such as ethanol
and gasohol.
8. For best accuracy, enter the viscosity value of the liquid at flowing temperature. If the viscosity is unknown, use the
recommended default value for the liquid type selected. Enter a note in the Note field, if desired (for example, to
identify the liquid parameters for future reference). The note will be stored with the liquid parameter settings.
9. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current MPMS Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
10. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
11. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
146
ModWorX Pro Section 10
Figure 10.16
a. Under the Maintenance Task heading, click Change Generic Liquid Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 10.17, page 148).
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: Generic Liquid) appears near the top of
the New Generic Liquid Properties section. The calculation method is read-only. To change this method, see Config-
uring Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters, or
with values supplied by the user. To use the set of default values, click Load Default Params near the bottom of
the screen.
147
Section 10 ModWorX Pro
Figure 10.17
5. In the Density Source field, select a source to be used for the density value: Absolute Density or Specific Gravity.
6. Enter the Base Density (Absolute or Specific Gravity) value of the liquid. If the base density of the liquid is un-
known, enter the flowing density into both the Base Density and Flowing Density fields.
7. Enter the Flowing Density (Absolute or Specific Gravity) value, if applicable.
8. If Special Products is selected as the fluid type for the MPMS fluid properties calculation method, enter the fluid’s
expansion coefficient “alpha” (Figure 9.14, page 135) in accordance with MPMS Chapter 11.1. You may enter a
numeric value, or click the Select Alpha button to select a predetermined alpha value for products such as ethanol
and gasohol.
9. For best accuracy, enter the viscosity value of the liquid at flowing temperature. Enter the viscosity of the liquid. If
the viscosity is unknown, a value of 1.0 cP (viscosity of water) is recommended.
10. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
11. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Generic Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
12. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
13. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
148
ModWorX Pro Section 10
149
Section 11 ModWorX Pro
Figure 11.1
2. Click Yes at the Enter Maintenance Mode prompt. The Change Flow Run Maintenance Task screen will appear
(Figure 11.2, page 151).
150
ModWorX Pro Section 11
Figure 11.2
3. Flow Run 1 is automatically selected.
4. Select the maintenance task to be performed. The maintenance tasks displayed will vary, depending on the type of
flow being measured - gas, steam or liquid. Examples are provided in the following procedures.
5. Click OK to save the selections. The Maintain Flow Run screen will appear (Figure 11.3).
Figure 11.3
By default, engineering units are provided in US standard format. If metric units are required, click on the Tools button
in the task bar at the top of the screen, and select Options from the dropdown menu. Click on the Units System drop-
down menu to change the units selection to Metric. See Figure A.17, page A-12.
151
Section 11 ModWorX Pro
Natural Gas (AGA-8 Detail) Natural Gas (AGA-8 Gross) Steam (IF-97)
152
ModWorX Pro Section 11
Figure 11.5
2. Note that current meter parameters appear on the right half of the screen. If you are changing only pipe size, probe
material or pipe material, click Copy from Current. Current parameters will be pasted into the fields on the left. This
will help to minimize the number of fields that have to be changed.
3. Click Select Meter to display a list of meter selections. The Select Averaging Pitot Tube screen will appear (Figure
11.6).
Figure 11.6
4. The first selection in the list—Averaging Pitot Tube—is appropriate for any non-Annubar product. To select an
averaging pitot tube meter, choose this selection on the left side of the screen, and enter the appropriate probe width,
discharge coefficient (K), and reference temperature in the fields on the right side of the screen. Click OK to return
to the Maintain Flow Run screen.
153
Section 11 ModWorX Pro
5. The remaining selections are Annubar meter models, with pre configured probe width, discharge coefficients (C1
and C2) and reference temperature. When an Annubar model is selected, no other entries are required on this screen.
Click OK to return to the Maintain Flow Run screen.
6. Enter the following parameters as necessary:
a. pipe size – click in the field and type in a value.
b. probe material – choose a material from the dropdown list; use the scroll bar to view the entire list of materials
c. pipe material – choose a material from the dropdown list; use the scroll bar to view the entire list of materials
a. serial number (optional)
7. Enter a note in the Note field, if desired. The note will be stored with the new meter parameter settings.
8. Click Save Changes in the upper right corner of the screen. The new settings will be saved and will be made effec-
tive when the user exits the Maintain Flow Run screen.
9. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
10. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
Figure 11.7
a. Under the Maintenance Task heading, click Change Gas Composition.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 11.8, page 155).
2. Note that current gas composition appears on the right half of the screen and the new gas composition will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
154
ModWorX Pro Section 11
Figure 11.8
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Gas Com-
position section. The calculation method is read-only. To change this method, see Configuring Fluid Properties, page
57.
4. In the Analysis Ref. No. field, enter a reference number from the gas analysis report, if desired.
5. Update the Reference Conditions field, if necessary. This selection identifies the ideal mass heating value table for
use in calculating the heating value of the gas composition and sets the reference temperature and pressure used in
converting the user-supplied heating value into the heating value mass basis required for energy calculations.
6. In the Heating Value field, specify a calculated or manually entered value. If the value is entered manually, enter the
heating value in the field provided.
7. In the Specific Gravity field, specify a calculated or manually entered value. If the value is entered manually, enter
the specific gravity value in the field provided.
8. Enter the viscosity of the measured fluid in the field provided. If you are unsure of the value, enter 0.010268.
155
Section 11 ModWorX Pro
9. Enter the isentropic exponent of the measured fluid in the field provided. If you are unsure of the value, enter 1.3.
Note Disregard the Gas Fraction field. This setting is not currently used in flow calculations.
11. To load the current gas composition or a previously saved composition, click Load Composition… and select the set
of gas composition values from the dropdown menu in the dialog provided. If only minor changes are required to an
existing composition, this can save time by minimizing the number of constituents that have to be changed.
12. Enter a note in the Note field, if desired (for example, to identify the gas composition for future reference). The note
will be stored with the gas composition settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gas Composi-
tion and will be put into effect when the user exits the Maintain Flow Run screen.
14. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
15. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
Figure 11.9
a. Under the Maintenance Task heading, click Change Gross Method Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 11.10, page 157).
156
ModWorX Pro Section 11
2. Note that current gross method parameters appear on the right half of the screen and the new gross method parame-
ters will be entered in the left half of the screen. If you are making a minor change to current parameters, click Copy
from Current. Current parameters will be pasted into the fields on the left. This will help to minimize the number of
fields that have to be changed.
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Gross
Method Parameters section. The calculation method is read-only. To change this method, see Configuring Fluid
Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the gas analysis report, if desired.
5. Update the Reference Conditions field, if necessary. This selection sets the reference temperature and pressure for
air density, which is used in converting the user-supplied heating value into the heating value mass basis required for
energy calculations.
6. Enter the heating value of the measured fluid in the field provided.
7. Enter the specific gravity of the measured fluid in the field provided.
Figure 11.10
8. Enter the viscosity of the measured fluid in the field provided. If you are unsure of the value, enter 0.010268.
9. Enter the isentropic exponent of the measured fluid in the field provided. If you are unsure of the value, enter 1.3.
10. Enter the carbon dioxide content in the field provided.
11. Enter the nitrogen content in the field provided.
12. Enter a note in the Note field, if desired (for example, to identify the gross method settings for future reference). The
note will be stored with the gross method settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gross Method
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
14. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
15. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
157
Section 11 ModWorX Pro
If a flowrun is configured for measuring liquid, the liquid parameters can be configured from the Maintain Flow Run
screen.
1. If Change MPMS Liquid Parameters is not displayed in the Maintenance Task field at the top of the Maintain Flow
Run screen, click on Change Task. The Change Flow Run Maintenance Task dialog will appear (Figure 11.11).
Figure 11.11
a. Under the Maintenance Task heading, click Change MPMS Liquid Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 11.12, page 159).
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: Crude Oil) appears near the top of the
New MPMS Liquid Properties section. The calculation method is read-only. To change this method, see Configuring
Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters, or
with values supplied by the user. To use the set of default values, click Load Default Params near the bottom of
the screen.
158
ModWorX Pro Section 11
Figure 11.12
5. In the Density Source field, select a source to be used for the density value: Absolute Density, Specific Gravity, or
API Gravity.
6. Enter the Base Density (Absolute, Specific Gravity or API Gravity) value of the liquid. Density is assumed to be at
base temperature.
7. If Special Products is selected as the fluid type for the MPMS fluid properties calculation method, enter the fluid’s
expansion coefficient “alpha” (Figure 9.14, page 135) in accordance with MPMS Chapter 11.1. You may enter a
numeric value, or click the Select Alpha button to select a predetermined alpha value for products such as ethanol
and gasohol.
8. For best accuracy, enter the viscosity value of the liquid at flowing temperature. If the viscosity is unknown, use the
recommended default value for the liquid type selected. Enter a note in the Note field, if desired (for example, to
identify the liquid parameters for future reference). The note will be stored with the liquid parameter settings.
9. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
10. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current MPMS Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
11. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
12. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
159
Section 11 ModWorX Pro
Figure 11.13
a. Under the Maintenance Task heading, click Change Generic Liquid Parameters.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure 11.14).
Figure 11.14
160
ModWorX Pro Section 11
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: Generic Liquid) appears near the top of
the New Generic Liquid Properties section. The calculation method is read-only. To change this method, see Config-
uring Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters
which are the values for water. While the default values are specific to water, they can provide an indication of
the value range that is applicable for these parameters. To load the default values, click Load Default Params
near the bottom of the screen.
5. In the Density Source field, select a source to be used for the density value: Absolute Density or Specific Gravity.
6. Enter the Base Density (Absolute or Specific Gravity) value of the liquid. If the base density of the liquid is un-
known, enter the flowing density.
7. Enter the Flowing Density (Absolute or Specific Gravity) value, if applicable.
8. Enter the viscosity of the liquid. Viscosity is assumed to be at flowing temperature. If the viscosity is unknown, a
value of 1.0 cP (viscosity of water) is recommended.
9. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
10. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Generic Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
11. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
12. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
161
Section 11 ModWorX Pro
162
ModWorX Pro Section 12
Note Before making changes to the flow run configuration, make sure the flow run calculation method is configured as
AGA-7 or Liquid Turbine. See Section 4 - Setting Up a Flowrun for instructions on selecting the flow run calcula-
tion method.
Figure 12.1
2. Click Yes at the Enter Maintenance Mode prompt. The Change Flow Run Maintenance Task screen will appear (Fig-
ure 12.2, page 164). The maintenance task displayed will vary, depending on the fluid property calculation method
selected. For example, when calculating gas flow with the AGA-8 Detail method is used, the maintenance task will
be “Change Gas Composition.” When the AGA-8 Gross method is used, the maintenance task will be “Change
Gross Method Parameters.”
163
Section 12 ModWorX Pro
Figure 12.2
3. Click OK to proceed. The Maintain Flow Run screen will appear (Figure 12.3).
Figure 12.3
164
ModWorX Pro Section 12
Note When the AGA-8 Detail method is used, all constituent percentages entered must total 100%.
11. To load the current gas composition or a previously saved composition, click Load Composition… and select the set
of gas composition values from the dropdown menu in the dialog provided. If only minor changes are required to an
existing composition, this can save you time by minimizing the number of constituents that have to be changed.
12. Enter a note in the Note field, if desired (for example, to identify the gas composition for future reference). The note
will be stored with the gas composition settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gas Composi-
tion and will be put into effect when the user exits the Maintain Flow Run screen.
14. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
165
Section 12 ModWorX Pro
Figure 12.4
2. Note that current gross method parameters appear on the right half of the screen and the new gross method parame-
ters will be entered in the left half of the screen. If you are making a minor change to current parameters, click Copy
from Current. Current parameters will be pasted into the fields on the left. This will help to minimize the number of
fields that have to be changed.
3. Note that the fluid properties calculation method selected for the flow run appears near the top of the New Gross
Method Parameters section. The calculation method is read-only. To change this method, see Configuring Fluid
Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the gas analysis report, if desired.
5. Update the Reference Conditions field, if necessary. This selection sets the reference temperature and pressure for
air density, which is used in converting the user-supplied heating value into the heating value mass basis required for
energy calculations.
6. Enter the heating value of the measured fluid in the field provided.
7. Enter the specific gravity of the measured fluid in the field provided.
8. Enter the viscosity of the measured fluid in the field provided. If you are unsure of the value, enter 0.010268.
9. Enter the isentropic exponent of the measured fluid in the field provided. If you are unsure of the value, enter 1.3.
10. Enter the carbon dioxide content in the field provided.
11. Enter the nitrogen content in the field provided.
166
ModWorX Pro Section 12
12. Enter a note in the Note field, if desired (for example, to identify the gross method settings for future reference). The
note will be stored with the gross method settings.
13. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Gross Method
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
14. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
If a flowrun is configured for measuring MPMS liquid such as crude oil, gasoline, jet fuel, fuel oils, and lube oil, the liq-
uid parameters can be configured from the Maintain Flow Run screen (Figure 12.5).
Figure 12.5
1. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
2. The fluid properties calculation method selected for the flow run (Example: Crude Oil) appears near the top of the
New MPMS Liquid Properties section. The calculation method is read-only. To change this method, see Configuring
Fluid Properties, page 57.
3. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters, or
with values supplied by the user. To use the set of default values, click Load Default Params near the bottom of
the screen.
4. In the Density Source field, select a source to be used for the density value: Absolute Density, Specific Gravity, or
API Gravity.
167
Section 12 ModWorX Pro
5. Enter the Base Density (Absolute, Specific Gravity or API Gravity) value of the liquid. Density is assumed to be at
base temperature.
6. If Special Products is selected as the fluid type for the MPMS fluid properties calculation method, enter the fluid’s
expansion coefficient “alpha” (Figure 9.14, page 135) in accordance with MPMS Chapter 11.1. You may enter a
numeric value, or click the Select Alpha button to select a predetermined alpha value for products such as ethanol
and gasohol.
7. For best accuracy, enter the viscosity value of the liquid at flowing temperature. If the viscosity is unknown, use the
recommended default value for the liquid type selected.
8. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
9. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current MPMS Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
10. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
11. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
If a flowrun is configured for measuring generic liquid such as water or emulsions, the liquid parameters can be config-
ured from the Maintain Flow Run screen (Figure 12.6).
Figure 12.6
168
ModWorX Pro Section 12
1. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
2. The fluid properties calculation method selected for the flow run (Example: Generic Liquid) appears near the top of
the New Generic Liquid Properties section. The calculation method is read-only. To change this method, see Config-
uring Fluid Properties, page 57.
3. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters
which are the values for water. While the default values are specific to water, they can provide an indication of
the value range that is applicable for these parameters. To load the default values, click Load Default Params
near the bottom of the screen.
4. In the Density Source field, select a source to be used for the density value: Absolute Density or Specific Gravity.
5. Enter the Base Density (Absolute or Specific Gravity) value of the liquid. If the base density of the liquid is un-
known, enter the flowing density.
6. Enter the Flowing Density (Absolute or Specific Gravity) value, if applicable.
7. If Special Products is selected as the fluid type for the MPMS fluid properties calculation method, enter the fluid’s
expansion coefficient “alpha” (Figure 9.14, page 135) in accordance with MPMS Chapter 11.1. You may enter a
numeric value, or click the Select Alpha button to select a predetermined alpha value for products such as ethanol
and gasohol.
8. For best accuracy, enter the viscosity value of the liquid at flowing temperature. If the viscosity is unknown, a value
of 1.0 cP (viscosity of water) is recommended.
9. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
10. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current Generic Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
11. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
12. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
169
Section 12 ModWorX Pro
170
ModWorX Pro Section 13
A single turbine input is available with the standard Scanner 2000 or Scanner 210x. If an optional expansion board is in-
stalled in a Scanner 2000 or Scanner 210x or if a Scanner 2200 is installed, two turbine inputs are available.
To change the turbine flowmeter model or to change a turbine K-factor,
1. Click on the Maintain Turbine button in the task menu bar on the Main Display screen (Figure 13.1).
Figure 13.1
2. Click Yes at the Enter Maintenance Mode prompt to proceed. The Change Turbine Maintenance Task screen will
appear (Figure 13.2, page 172).
171
Section 13 ModWorX Pro
Figure 13.2
3. Select the appropriate flowmeter, if applicable. If an optional expansion board is installed in the Scanner 2000 or
Scanner 210x or if a Scanner 2200 is installed, select either Turbine 1 or Turbine 2.
4. Select the maintenance task you want to perform, and click OK to save your selection. The Maintain Turbine screen
will appear (Figure 13.3).
Figure 13.3
172
ModWorX Pro Section 13
Figure 13.4
2. If more than one turbine flowmeter is displayed, select the turbine flowmeter to be changed.
3. Click Change Turbine Flowmeter.
4. Click OK to save the settings and return to the Maintain Turbine screen (Figure 13.5).
Figure 13.5
173
Section 13 ModWorX Pro
5. Note that current meter information appears on the right half of the screen, and that new meter information is en-
tered in the left half of the screen. To make a minor change to current parameters, click Copy from Current. Current
parameters will be pasted into the fields on the left. This will help to minimize the number of fields that have to be
changed.
6. To select a new meter, click Select Meter. The Select Turbine Meter dialog will appear (Figure 13.6).
Figure 13.6
a. Select a meter model from the list provided. The corresponding model number, low flow rate, high flow rate,
and nominal K-factor for the selected meter will appear on the right side of the screen.
b. Enter the flowmeter serial number, if desired.
c. Click OK to save the new selections and return to the Maintain Turbine Flowmeter screen.
7. Enter a note in the Note field, if desired. The note will be stored with the new meter information.
8. Click Save Changes in the upper right corner of the screen. The new settings will be saved and will be made effec-
tive when the user exits the Maintain Turbine Flowmeter screen.
9. To change the K-factor or make changes to a different turbine input without leaving the screen, click on Change
Task and make a new selection.
10. To exit the Maintain Turbine Flowmeter screen and return to the Main Display screen, click Exit Maintenance
Mode in the upper right corner of the screen.
174
ModWorX Pro Section 13
Figure 13.7
2. If more than one turbine flowmeter is displayed, select the turbine flowmeter for which a new K-factor is desired
(Turbine 1 or Turbine 2).
3. Click Change K-Factors.
4. Click OK to save the new settings and return to the Maintain Turbine Flowmeter screen (Figure 13.8).
Figure 13.8
175
Section 13 ModWorX Pro
5. Note that current K-factors appear on the right half of the screen, and that new K-factors are entered in the left half
of the screen. To make a minor change to current parameters, click Load K-Factors From... and select the K-Factor
settings from the dialog provided. The selected K-factors will be pasted into the fields on the left. This will help to
minimize the number of fields that have to be changed.
6. To enter a linear K-factor,
a. Click Linear K-Factor in the Entry Mode box.
b. Click the Enter Linear K-Factor button.
c. Type in the new linear K-factor in the dialog provided.
d. Click OK to update the entry in the New K-Factors section of the Maintain Turbine Flowmeter screen.
7. To enter a multi-point K-factor:
a. Click Multi-point K-Factors in the Entry Mode box.
b. Click Add Factor and enter the appropriate frequency and factor values from the calibration data in the dialog
provided.
c. Click OK to update the entry in the New K-Factors section of the Maintain Turbine Flowmeter screen.
d. Repeat steps 7a through 7c to enter up to 12 calibration points.
8. Enter a note in the Note field, if desired. The note will be stored with the new K-factors.
9. Click Save Changes in the upper right corner of the screen. The new settings will be saved and will be made effec-
tive when the user exits the Maintain Turbine Flowmeter screen.
10. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
11. To exit the Maintain Turbine Flowmeter screen and return to the Main Display screen, click Exit Maintenance
Mode in the upper right corner of the screen.
176
ModWorX Pro Section 13
177
Section 14 ModWorX Pro
Downloading Data
A user can download flow and event logs and device settings using the red Download button in the upper left corner of
the Main Display screen or Configure screen.
Alternatively, a user can configure the software to automatically download the logs and/or settings each time the software
connects with a Scanner. See procedures for each method below.
Figure 14.1
178
ModWorX Pro Section 14
Figure 14.2
2. Choose a download range: “new records only” or “all records.”
3. Choose the type(s) of logs you want to download.
4. Choose any additional device settings to be included in the download (configuration settings, calibration settings, or
holding register data snapshot).
Note Basic configuration data will automatically be downloaded even when Configuration Settings is not selected for
download. However, a complete set of configuration settings will be downloaded only when the Configuration
Settings checkbox is selected.
Note If you plan to export the downloaded data to Flow-Cal or PGAS format, daily logs are not required. It is recom-
mended that all other data types (interval and event/alarm logs, configuration settings, calibration settings, and
holding registers data snapshot) be selected for download for use with these third-party software applications.
Note To change the parameters included in the flow logs, the order in which they appear, or the logging interval,
change the settings on the Configure Archive menu. See Archive Parameter Setup, page 41, for details.
The initial download will display all records captured by the device. On subsequent downloads, however, the user can
“mark” downloaded records as read, and easily distinguish between “old” and “new” records. See Marking Downloaded
Records, page 183.
179
Section 14 ModWorX Pro
Figure 14.3
2. When the Program Options dialog opens, click the Auto-Download option from the list at the left side of the screen
(Figure 14.4).
Figure 14.4
3. From the Auto-Download Options screen, check the checkbox at the top of the screen: Automatically begin down-
loading device data after connecting.
4. Check the range of log records, log types, and device settings you want included in each download.
5. Click OK to save the new settings.
6. The next time ModWorX Pro connects to the Scanner, the Download Device Data screen will automatically appear,
prompting the user to download all selected logs and/or device settings.
180
ModWorX Pro Section 14
Figure 14.5
Sort Order
By default, the logs will appear in the order in which they are downloaded, top to bottom. However, the user can change
the sort order by changing the selection in the dropdown Sort Order box on the Device Reports screen (Figure 14.6, page
182). Users can base sort order on the date/time stamp or record number, as shown.
Each sort order selection is specific to the log type view (tab) selected when the sort order is designated, and sort order
selections are retained between ModWorX Pro sessions.
181
Section 14 ModWorX Pro
Figure 14.6
Data Filtering
Log data is easily filtered using the Filter settings near the top of the screen. When the Filter checkbox is checked, data
filtering is enabled. Filtering options include filtering by month/day (Figure 14.7) or by a user-specified date range (Fig-
ure 14.8, page 183). When the Filter by Month/Day option is enabled, a dropdown menu will list all months and days
for which data is stored. The user can select multiple months and/or days by placing a checkbox next to each month/day
desired.
Figure 14.7
Note When data is filtered by month, the first day of the month following the “filter month” will be included in the filter
range because data stored on the last day of the month will be downloaded at the contract hour of the follow-
ing day. For example, when a user filters data from July, the download from August 1 is included. This download
includes data stored during the 24 hours between the July 31 contract hour and the August 1 contract hour.
182
ModWorX Pro Section 14
When the Filter by Date Range checkbox is checked, the Start Date and End Date fields to the right of the checkbox
will be enabled. To initiate a filter, click the dropdown arrow next to the Start Date/End Date field to display a calendar
dialog, select the appropriate start and end date/time, click OK to return to the Device Reports screen, and click Apply
(Figure 14.8).
Figure 14.8
Figure 14.9
183
Section 14 ModWorX Pro
Figure 14.10
Figure 14.11
184
ModWorX Pro Section 14
A dialog will appear, allowing the user to select the .sdf file to be opened from a specified directory and/or subdirectory
(Figure 14.12). Select the appropriate folder, then select the subdirectory and click on the filename to view the .sdf files.
Figure 14.12
File Properties
To view a description of a downloaded record before you open it, click on the .sdf file to select it. The type of log, date
and timestamp, date range for daily, interval and event logs, and firmware version is displayed in the window at the far
right of the screen (Figure 14.13).
Figure 14.13
Note Files generated with ModWorX Pro version 2.0.0 or later can be viewed with ModWorX Pro Version 4.2 software.
185
Section 14 ModWorX Pro
Figure 14.14
The trend view presents the downloaded logs in two graphs:
The top graph provides a high-level view of all data in a downloaded log. Click anywhere in the top graph to select the
group of records to be magnified in the large graph display at the bottom of the screen.
Note If the records in a download are too numerous to be displayed in the top graph, a set of gray bars will appear on
either side of the top graph. Click on either bar to move the graph and bring hidden logs into view.
The bottom graph allows a user to view records in detail. Screen controls allow a user to view multiple parameters on
one graph, or to isolate selected parameters for viewing.
186
ModWorX Pro Section 14
•• Click on a point in the top graph. The logs nearest that selection point will appear in the bottom graph. Adjust the
view, if necessary, by moving the scroll bar beneath the bottom graph.
•• Click on Set Time Axis Range and enter a Start date and time and an End date and time using the calendars pro-
vided. Figure 14.15, shows the calendar for selecting a range of interval logs. For daily logs, the calendar will
prompt the user to enter a Start and End date only (no times).
Figure 14.15
2. Select the parameter to be viewed in the graph by clicking on a parameter in the table at the left side of the screen
(Figure 14.16, page 188). The parameter is highlighted in blue, its title and configured unit is displayed at the top
of the graph, and the graph scale changes in accordance with the values of the parameter selected. By default, all
archive parameters are displayed in the table, and are displayed in the trend graph, but values are displayed only for
the trend line selected at any given time. The bright yellow light bulb icon indicates that the parameter trend line is
enabled for display (Figure 14.16, page 188).
•• To disable all parameter trend lines except the one that is actively selected (highlighted), click the Disable All
button.
•• To enable all parameter trend lines, click the Enable All button.
•• To hide a parameter trend line, click on the yellow light bulb icon next to the parameter to be hidden. The light
bulb icon will change from yellow to blue, and the trend line will disappear from the graph.
•• To display a hidden parameter trend line, click on the blue (disabled) light bulb icon. The icon will change to yel-
low, and the trend line will appear in the graph.
187
Section 14 ModWorX Pro
Figure 14.16
Note Turning off the light bulb of a parameter that is highlighted in the table will cause the trend line and the Y axis
to disappear, but the parameter name remains on screen. Click on a parameter in the table to restore the trend
view with the appropriate axis.
3. Navigate the trend graph to analyze parameter values. Note the dotted vertical white line in the bottom graph (Figure
14.16). This Record Index line remains in a fixed location, and marks the record being read. In Figure 14.16, the
Record Index line intersects the trend graph at log record number 1,996. The log number, timestamp, and individual
parameter values in that log are displayed in the table at the left of the screen. To view a different section of logs, use
the scroll bar beneath the graph to shift the graph to the left or right. To advance the graph one log at a time, click
the right and left arrow buttons on the computer keypad, or click on the right or left arrow in the scroll bar beneath
the graph.
188
ModWorX Pro Section 14
Event Logs
The Event Logs tab allows users to view user event logs in a table format (Figure 14.18). Event logs capture user chang-
es such as K-Factor, orifice meter factor, and cone meter factor changes, plate and cone size changes, gas composition
changes, input setting changes and resets.
Events are logged by date and time, and each record includes the Modbus register address and register name, the old
value of the parameter, and the new value.
Events and alarms are stored in one common log; up to 1,152 combined events and alarms can be stored and viewed.
Figure 14.18
189
Section 14 ModWorX Pro
Alarm Logs
The Alarm Logs tab allows users to view alarm logs in a table format. An alarm log is generated each time a selected
parameter goes outside its user-selected setpoint(s). Alarms are logged by date and time, and each record includes the
Modbus register address and register name, status or type of alarm activated, and the associated value and unit where
applicable (Figure 14.19, page 190).
Alarms and events are stored in one common log; up to 1,152 combined alarms and events can be stored and viewed.
Figure 14.19
Figure 14.20
190
ModWorX Pro Section 14
Figure 14.21
Note If the device firmware is version 4.18 or higher, the previous and current verification settings will be included in
the Calibration report.
Snapshot Report
The Holding Registers Data Snapshot report allows a user to capture live flow run and turbine input data, sensor input
readings, and system data (Figure 14.22). The report can be exported and e-mailed to a technician for diagnostic help.
Figure 14.22
191
Section 14 ModWorX Pro
Printing a Report
To print a hard copy of an on-screen report,
1. Click the tab of the Device Reports screen containing the logs or settings you want to print.
2. Click Print Table in the left column of the screen (Figure 14.23).
3. A Print Preview window will appear. If necessary, change the margins or other settings to fine-tune the presentation
of data.
4. Click File from the Print Preview screen, then click Print.
5. Enter your print settings in the print dialog and click Print.
Figure 14.23
192
ModWorX Pro Section 14
3. The Save SDF File dialog will appear (Figure 14.24). To change the file path and/or file name from the defaults
listed below, browse to the desired file path and/or type the desired entry in the field provided. See Log Directory
and Filenames, page A-16 for more information.
Figure 14.24
4. Click Save. The file will be saved as with an .SDF extension in the specified directory.
193
Section 14 ModWorX Pro
Figure 14.25
Figure 14.26
Emailing Data
Daily logs, interval logs, event logs alarm logs, configuration settings, calibration settings, and a snapshot report can be
transmitted via email directly from the ModWorX Pro interface using any one of four tools:
• E-mail Data allows a user to email logs of a specific file type (daily logs, for example).
• E-mail Selected allows a user to select multiple log types for emailing.
• Email All to Flow-Cal allows a user to email all data in a single Flow-Cal file.
• Email All to PGAS allows a user to email all data in PGAS files.
194
ModWorX Pro Section 14
Note If multiple log types are desired, select Email Selected. See Email Selected Data (Multiple Log Types), page
196, for further instruction.
Figure 14.27
3. An Export... window will appear (Figure 14.28). To change the file path and/or file name from the defaults listed
below, browse to the desired file path and/or type the desired entry in the field provided. See Log Directory and
Filenames, page A-16 for more information.
–– By default, the export file will be saved in the directory C:\NuFlo Log Data\ModWorX Pro\Field_Name\Site_
Name.
–– By default, the export filename prefix will be <Device_Name>_<date (YYYYMMDD)>_<time (hhmm)>, fol-
lowed by an initial to indicate the log type.
Figure 14.28
195
Section 14 ModWorX Pro
4. Click Save to export the data to the file. The file will automatically be attached to an email window
(Figure 14.29).
5. Address the email, add additional notes to the body if desired, and send.
Figure 14.29
Figure 14.30
196
ModWorX Pro Section 14
2. The Email Selected window will appear (Figure 14.31). Select the file format desired and select the log types/reports
desired using the checkboxes provided.
–– To change the export file, enter the export file path or click on the folder icon to browse to the export location.
See Log Directory and Filenames, page A-16 for more information. The default export file is C:\NuFlo Log
Data\ModWorX Pro\Field_Name\Site_Name.
–– To change the file name, enter the file name in the field provided. The default export file name will include
<Device_Name>_<date (YYYYMMDD)>_<time (hhmm)>. Each log type will be exported in a separate file,
and the filename will include an initial to identify the log type. For example, a daily log downloaded on March
20, 2008 at 3:35 p.m. would be named Device_Name_20080320_1535_D.xls.
Figure 14.31
3. Click OK to export the data to the file. The file will automatically be attached to an email window (Figure 14.32).
4. Address the email, add additional notes to the body, if desired, and send.
Figure 14.32
197
Section 14 ModWorX Pro
Figure 14.33
2. An Export All to Flow-Cal File or Export All to PGAS Files window will appear (Figure 14.34). To change the file
path and/or file name from the defaults listed below, browse to the desired file path and/or type the desired entry in
the field provided. See Log Directory and Filenames, page A-16 for more information.
–– By default, the export file will be saved in the directory C:\NuFlo Log Data\ModWorX Pro\Field_Name\Site_
Name.
Figure 14.34
3. Click Save to save the data to the file. The .CFX file will automatically be attached to an email window (Figure
14.35).
4. Address the email, add additional notes to the body, if desired, and send.
198
ModWorX Pro Section 14
Figure 14.35
Exporting Data
Daily logs, interval logs, event logs alarm logs, configuration settings, calibration settings, and a snapshot report can be
exported from ModWorX Pro to the following formats: .XLS, .CSV, .PDF, .RTF, and .HTML. Calibration settings can
be exported to any of the above formats except .CSV. Additionally, all archive data can be exported, collectively, to the
Flow-Cal (.CFX) format or to PGAS formats (.ANA, .VOL, and .EVT).
Data can be exported using any of four tools:
• Export Data allows a user to export logs of a specific file type (daily logs, for example).
• Export Selected allows a user to select multiple log types for exporting.
• Export All to Flow-Cal allows a user to export all data in a single Flow-Cal file.
• Export All to PGAS allows a user to export all data in PGAS files.
Export Data (Single Log Type)
To export logs of a specific type,
1. Click on the appropriate log type tab (Daily, Interval, Events, etc.) to display the reports desired (Figure 14.36, page
200).
2. Click the Export Data button at the left of the screen and select the export format desired (.XLS, .CSV, .PDF, .RTF,
or .HTML).
Note If multiple log types are desired, select Export Selected. See Export Selected Data (Multiple Log Types), page
201, for further instruction.
199
Section 14 ModWorX Pro
Figure 14.36
3. An Export... window will appear (Figure 14.37). To change the file path and/or file name from the defaults listed
below, browse to the desired file path and/or type the desired entry in the field provided. See Log Directory and
Filenames, page A-16 for more information.
–– By default, the export file will be saved in the directory C:\NuFlo Log Data\ModWorX Pro\Field_Name\Site_
Name.
–– By default, the export filename prefix will be <Device_Name>_<date (YYYYMMDD)>_<time (hhmm)>, fol-
lowed by an initial to indicate the log type.
4. Click Save to export the data to the file.
Figure 14.37
200
ModWorX Pro Section 14
Figure 14.38
2. An Export Selected window will appear (Figure 14.39). To change the file path and/or file name from the defaults
listed below, browse to the desired file path and/or type the desired entry in the field provided. See Log Directory
and Filenames, page A-16 for more information.
–– By default, the export file will be saved in the directory C:\NuFlo Log Data\ModWorX Pro\Field_Name\Site_
Name.
Figure 14.39
201
Section 14 ModWorX Pro
Figure 14.40
2. An Export All to Flow-Cal File or Export All to PGAS Files window will appear (Figure 14.41). To change the file
path and/or file name from the defaults listed below, browse to the desired file path and/or type the desired entry in
the field provided. See Log Directory and Filenames, page A-16 for more information.
–– By default, the export file will be saved in the directory C:\NuFlo Log Data\ModWorX Pro\Field_Name\Site_
Name.
–– If desired, change the filename of the export file. By default, the export filename for a Flow-Cal file will be
<Device_Name>_<date (YYYYMMDD)>_<time (hhmm)>.xxx (where xxx is .CFX for Flow-Cal or .ANA for
PGAS).
3. Click Save to export the data to the file(s).
Figure 14.41
202
ModWorX Pro Section 14
If exporting to PGAS format, three files will be generated (Figure 14.42). Configuration data is saved in a file with an
.ANA extension. Interval logs are saved in a file with a .VOL extension, and events are saved in a file with an .EVT ex-
tension.
Figure 14.42
203
Section 14 ModWorX Pro
204
ModWorX Pro Appendix A
Figure A.1
Functions supported by this menu include:
• save and load configuration settings
• clear flow totals
• clear the Scanner memory
• copy Scanner memory to a binary file
• manually create a partial record (recommended following commissioning of a device)
• change the power saving mode
• reset user security settings
• modify or disable the time synchronization prompt
• change the units system
• change communications options
• enable/disable auto connection
• enable/disable auto downloads upon connection
• change download options
• change directory for configuration files
• enable automatic logging of data polls
• view Modbus registers
A-1
Appendix A ModWorX Pro
IMPORTANT To successfully upload a configuration file to another Scanner, the file being uploaded must
be generated by a unit with the same firmware version or an older firmware version than the
firmware in the unit being configured. If the configuration file being uploaded was created with
newer firmware than that in the Scanner being configured, an error message will appear
when an upload is attempted.
Note If the use of a single directory is preferred for storing all configuration files, as opposed to the file structure shown
in step 2, see Log Directory and Filenames, page A-16.
3. Change the filename or file location, if desired, and click Save. A Communications status box will appear while the
configuration files are saved, followed by an Information box confirming that the files have been saved successfully.
4. Click OK to exit the Information box, then click Done to exit the Manage Configuration screen. The configuration
file will appear in the folder designated in step 2.
Figure A.2
A-2
ModWorX Pro Appendix A
Items previously selected for upload will automatically be selected. This is a timesaving feature for users who are
configuring multiple units with the same configuration settings, in that the settings need to be specified for the first
upload only.
Figure A.3
2. Select the type of settings you want to restore from the selections provided.
Note To save previous and current calibration data, select “All (Board Replacement).”
To save only current calibration data, select “Configuration and Current Calibration.”
To manually select the parameters to be uploaded from a checklist, select “Custom.”
3. Click Next.
4. Select the configuration file to be uploaded. Click on the folder icon in the Filename field to open the C:\NuFlo Log
Data\ModWorX Pro\ directory. Then navigate by field name and site name to locate the appropriate configuration
file for upload. Double-click on the file or click Open to select it. The selected file will be highlighted on the Load
Configuration screen.
5. Click Next. The Scanner will quickly read the firmware version of the configuration file selected and compare it to
the firmware version of the Scanner. If no compatibility problem is detected, a Device Settings screen will appear
(Figure A.4, page A-4). If configuration settings have been selected for upload, the Device/Meter Name and Site
Name settings will appear. Change these settings to represent the unit being configured.
Note By default, the device will be synchronized to the computer time following upload. To override this default, dese-
lect Synchronize Device Time at the bottom of the screen.
A-3
Appendix A ModWorX Pro
Figure A.4
6. Click Next. If Configuration settings were selected for upload, a Device Communications Settings screen will appear
(Figure A.5). Enter slave address and baud rate settings for Ports 1 and 2 of the device being configured. If only
calibration settings, gas composition settings, or custom settings are being uploaded, the Device Communications
Settings screen will not appear.
Figure A.5
7. Click Next. A summary screen will appear.
A-4
ModWorX Pro Appendix A
8. Click Next to initiate the upload . A progress screen will be displayed throughout the upload (Figure A.6), followed
by a confirmation screen when the upload has been successfully completed.
Figure A.6
A-5
Appendix A ModWorX Pro
Figure A.7
WARNING Users can clear either device settings or archive settings, or both, from the Scanner‘s memory.
! Clearing device settings will delete all calibration and configuration settings. Clearing the device archive
will delete all daily, interval, event, and alarm log records.
Figure A.8
A-6
ModWorX Pro Appendix A
Figure A.9
2. Click OK on the Copy EEPROM to File screen (Figure A.9). A Write EEPROM to File dialog will appear, prompt-
ing the user to verify or change the location where the file will be stored, and the name of the binary file. Un-
less changed on the Options>Directories screen, the file location is C: NuFlo Log Data\ModWorX Pro, and the
filename is EEPROM_YYYYMMDD_HHMMSS.bin where YYYYMMDD is the year, month and day that the
memory is saved, and HHMMSS is the hour, minute and second that the memory is saved. For example, EE-
PROM_20090118_230443.bin is the name of a file copied at 11:04:43 p.m. on January 18, 2009.
3. Click Save to initiate the copying of the memory. The Copy EEPROM to File dialog will reappear, and a progress
bar will monitor the percentage of completion as the memory is copied. A confirmation message will appear on the
screen when the binary file has been successfully saved.
A-7
Appendix A ModWorX Pro
Figure A.10
Figure A.11
A-8
ModWorX Pro Appendix A
Figure A.12
Caution Do not close the Reset User Security screen until the reset process is complete. If the screen is
closed and reopened after a key code is provided, a new key code will be generated and the reset
code provided by the technician will not reset the security controls.
3. Enter the code into the field provided, and click OK.
4. The software will automatically return to the Welcome screen. Login using the default permissions and reestablish
user accounts with security settings, if desired.
Figure A.13
A-9
Appendix A ModWorX Pro
Figure A.14
Figure A.15
A-10
ModWorX Pro Appendix A
Figure A.16
A-11
Appendix A ModWorX Pro
Figure A.17
• Original Format (Applied/Measured). This format includes the measurement value applied to the sensor
(“Applied”) and the resulting reading generated by the factory calibration (“Measured”). Displayed errors indicate
the difference between the new user calibration results and the original factory calibration.
• Extended Format (As Left/As Found). This format is the default (recommended) format. “Applied/As Left” is the
measurement value applied to the sensor. “As Found” is the result of the previous user calibration. Displayed errors
indicate the difference between the new user calibration results and the results of the previous user configuration.
IMPORTANT Calibrations will only be displayed in format in which they were saved. If the user chooses
“Original Format,” the calibration can only be viewed in the original format. However, if you
saved the calibrations in the original format but changed your settings to “Extended,” old
calibrations should still be displayed in the original format until they are saved again.
Figure A.18
A-12
ModWorX Pro Appendix A
Figure A.19
Presend Delay
The Communications Options screen allows a user to enter a presend delay period (milliseconds) to increase the chances
for a successful connection between the computer and the Scanner when the instrument is networked with a radio or
other low-speed device. The computer transmits a request to send, which powers the device on, and then waits the speci-
fied length of time before attempting to transmit data. By default, this parameter is set to 50 msec.
Number of Retries
By default, the computer will make two attempts to connect to the Scanner before providing an error message. In situa-
tions where time-outs are anticipated—for example, due to a low-speed computer—the user can increase the number of
communication attempts allowed for a connection by increasing the number of retries.
Time-out Period
The time-out period is the length of time allowed for a connection to a Scanner. By default, this setting is 5000 msec (5
seconds). When connecting to a slow computer, try increasing the time-out period to increase chances for a successful
connection.
A-13
Appendix A ModWorX Pro
Figure A.20
A-14
ModWorX Pro Appendix A
Figure A.21
Figure A.22
A-15
Appendix A ModWorX Pro
Figure A.23
Note If you do not select a custom filepath and filename, the files will be saved with the default storage path and file-
name (i.e. “C:\NuFlo log data\ModWorX Pro\Field_Name\Site_Name\Device_Name_Timestamp.sdf”).
By default, *.sdf files are saved to C:\NuFlo log data\ModWorX Pro\.To change the file location,
1. From the Main Display, click on the Tools button and choose Options from the dropdown menu (Figure A.1, page
A-1).
2. When the Program Options dialog opens, click the Directories option from the list at the left side of the screen (Fig-
ure A.24, page A-17).
A-16
ModWorX Pro Appendix A
Figure A.24
3. To change the directory to which non-configuration data is saved, click in the Data Directory portion of the dialog
(Figure A.25), type the desired filepath or click the Folder icon to browse to the desired directory.
Figure A.25
4. To change the directory to which configuration data is saved, click in the Configuration Save Directory portion of
the dialog (Figure A.26, page A-18) and type the desired filepath or click the Folder icon to browse to the desired
directory.
A-17
Appendix A ModWorX Pro
Figure A.26
Note To save configuration files to the root directory only, select “Save Configuration Files only in the root
directory - ignore subdirectory path options.”
Subdirectories
You can save data logs in up to three subdirectories, including field/lease name, site name, company name, or location
name, in whatever order you choose. Unless otherwise specified, the file will be saved by default to C:\NuFlo log data\
ModWorX Pro\Field_Name\Site Name.
However, you can choose the filepath to which the data logs are saved. For example, if you have multiple clients with
multiple sites, you may want to select “Company Name” as the first subdirectory, followed by “Site Name” as the second
directory. To select a custom subdirectory storage path,
1. From the Main Display, click on the Tools button and choose Options from the dropdown menu (Figure A.1, page
A-1).
2. When the Program Options dialog opens, click the Directories option from the list at the left side of the screen.
3. In the Directory Settings portion of the screen (Figure A.27, page A-19), select the subdirectory storage path(s) to
which you want the file saved. The file will be saved in the order the subdirectories are selected.
For example, if you want to store files by field name, site name, and then company name, you would select “In-
clude” and choose “Field/Lease Name” to set field name as the first subdirectory, then continue to the next field and
select “Site/Well Name” and so on.
A-18
ModWorX Pro Appendix A
Figure A.27
Filenames
In addition to saving files to user-specified subdirectories, you can also add up to three descriptive elements to the file-
name from the following options: device/meter name, field/lease name, site/well name, company name, location name,
or legal description. By default, the filename will include the device/meter name and the date/time stamp (i.e. “Device_
Name_Timestamp.sdf”), which is always the last part of the name unless you chose not to include the date/time stamp.
To remove the date/time stamp, uncheck the “Include” box next to the Time Stamp field.
However, you can customize the filename. For example, if you are using the same device name on multiple sites, you
may want to include the site/well name as part of the filename to identify the device and the location stored in the file.
The resulting filename would be “Device_Name_Site_Name_Timestamp.sdf.”
To specify the descriptors to be included in the filename, select “Include” under the File Name heading and select the
desired item to include in the filename. See Device Identification, page 26, for more information.
A-19
Appendix A ModWorX Pro
Figure A.28
A-20
ModWorX Pro Appendix A
Figure A.29
Note Access to the Modbus registers is not required for the routine operation of the Scanner. Novice users should not
access these screens unless instructed to do so by Cameron technical support.
A-21
Appendix A ModWorX Pro
A-22
ModWorX Pro Appendix B
IMPORTANT To act as a PID controller, the Scanner 2000 or Scanner 210x must be purchased with optional PID
firmware and expansion board.
The Scanner 2000 or Scanner 210x can be used to control parameters such as
The tuning values selected for the controller will determine if the process cotnrol is effective and stable.
All installations require tuning. Tuning by observing numeric values alone is a difficult task for even the most skilled op-
erator. The tuning tools within ModWorX Pro simplify the process. The software plots the process variable, setpoint, and
valve position in an easy-to-read, scalable graph. Output is updated in real time, so every tuning change is visible to the
operator, who can then determine when an optimum response is achieved (Figure B.1).
Figure B.1
B-1
Appendix B ModWorX Pro
Figure B.2
Controller Action
The controller action setting (direct or reverse) determines the type of change in the controller output (increase or de-
crease) that will be used to align the selected process variable with the control setpoint. If the wrong controller action is
selected, the controller’s response will be opposite the intended action (for example, increasing temperature rather than
decreasing it).
Direct action causes the output value to change in the same direction as the change in the process variable. For example,
an increase in process variable temperature will result in a corresponding increase in the controller output.
Reverse action causes the output value to change in the opposite direction as the change in the process variable. For ex-
ample, an increase in process variable temperature will result in a decrease in the controller output.
B-2
ModWorX Pro Appendix B
Controller Output
The type of controller output (increased or decreased) needed for a particular application depends on
• The valve configuration (air-operated valves are referred to as “fail open” or “fail close”)
• The way a change in valve position will affect the process variable (for example, opening a valve in certain installations
will increase the process value, whereas opening the valve in other installations will decrease the process variable
value.
The following examples demonstrate how a valve configuration can affect the action setting. In both examples, the pro-
cess is designed so that the process value will decrease when the valve is opened.
Direct Action
When the controller action is “direct,” the control valve will fail in the closed position. An increased controller output
opens the valve and decreases the process value. In Figure B.3, the error is initially positive (process value – setpoint >
0). The positive error is counteracted by an increased controller output. Therefore, the controller is direct-acting.
Figure B.3
Reverse Action
When the controller action is “reverse,” the control valve will fail in the open position. A decreased controller output
opens the valve and decreases the process value. In Figure B.4, the error is initially positive (process value – setpoint >
0). The positive error is counteracted by an decreased controller output. Therefore, the controller is reverse-acting.
Figure B.4
B-3
Appendix B ModWorX Pro
Note If the process variable to be controlled is based on an analog input, you must configure the input before configur-
ing the PID Controller. See Section 6 - Configuring Inputs, page 68 for information about configuring an analog
input.
Figure B.5
B-4
ModWorX Pro Appendix B
5. Enter the Range Low and Range High values for the process variable being controlled in the fields provided. In
Figure B.5, page B-4, the range is 0 to 200 Lit/hour.
6. Enter the desired setpoint in the field provided. Note that the setpoint must be within the established low and high
range.
7. In the Control Loop section of the screen, enter the known tuning values. If no values are known, continue to Step 8.
Note The Proportional Constant (Kp) is referred to as “Gain” in the ModWorX Pro software interface. Gain is funda-
mental to the PID controller, because it is used to determine the controller output value. For example, assume a
gain value of 1.0 (actual gain values will be much smaller due to normalization of the process variable operating
range), a setpoint of 50%, and a process variable reading of 60%. To determine the output value, the controller
multiples the error (the process variable reading minus the setpoint, so 10 in this case) by the gain factor (1), the
calculated output value would be 10 × 1, or 10 (percent of output change). The direction of the change (increase
or decrease) in the output is determined by the action set in Step 3.
8. Click Apply to save the settings. The 4 to 20 mA output will automatically be configured for PID control.
9. Tune the controller using the instructions in Tuning the Controller, page B-7.
Note If the process variable to be controlled is based on an analog input, you must configure the input before configur-
ing the PID Controller. See Section 5 - Setting up a Turbine Input or Pulse Input, page 62 for information about
configuring an analog input.
5. Enter the Range Low and Range High values (0 to 200 L/hr) for the process variable being controlled in the fields
provided.
6. Enter the desired setpoint in the field provided. Note that the setpoint must be within the established low and high
range.
7. In the Control Loop section of the screen, enter the known tuning values. If no values are known, continue to step 8.
Note The Proportional Constant (Kp) is referred to as “Gain” in the ModWorX Pro software interface. Gain is funda-
mental to the PID controller, because it is used to determine the controller output value. For example, assume a
gain value of 1.0 (actual gain values will be much smaller due to normalization of the process variable operating
range), a setpoint of 50%, and a process variable reading of 60%. To determine the output value, the controller
multiples the error (the process variable reading minus the setpoint, so 10 in this case) by the gain factor (1), the
calculated output value would be 10 × 1, or 10 (percent of output change). The direction of the change (increase
or decrease) in the output is determined by the action set in Step 3.
8. In the Pressure Variable Override section (see Figure B.6), click in the “Static Pressure Source” field and select the
static pressure source.
Note If the static pressure source is the integral MVT, the setting will read “Integrated Static Pressure.”
B-5
Appendix B ModWorX Pro
Figure B.6
9. To establish the pressure source’s range, enter the Range Low and Range High values in the fields provided in the
Pressure Variable Override section.
10. Enter the pressure override setpoint in the “Pressure Setpoint” field in the Pressure Variable Override section. Should
the pressure exceed the setpoint value, the controller will actuate the valve to reduce the pressure.
11. Enter the pressure threshold dead band in the field provided in the Pressure Variable Override section. The dead band
discourages valve position oscillation when values are within the entered dead band.
12. To control how quickly the valve responds to a pressure override, enter the pressure override for known Gain, Inte-
gral, and Derivative values in the Override Control Loop section.
13. To select the pressure override controller action desired (direct or reverse), select the action in the field provided in
the Override Control Loop section.
14. Click Apply to save the settings.
15. Tune the controller using the instructions in Tuning the Controller, page B-7.
B-6
ModWorX Pro Appendix B
Main Graph
To manually control the PID controller, adjust the PID Control Loop values (which can be optionally plotted on the
graph) against the setpoint and valve position, as shown in Figure B.7.
Figure B.7
Note If the firmware version is 4.34 or higher, the “Current Output Mode” value will display the live condition, rather
than the current configuration setting. Firmware versions 4.33 and lower will only display the current configura-
tion setting.
B-7
Appendix B ModWorX Pro
Figure B.8
Figure B.9
B-8
ModWorX Pro Appendix B
B-9
ModWorX Pro Appendix C
Figure C.1
C-1
Appendix C ModWorX Pro
If the flow run is configured for liquid measurement, the selections will appear as shown in Figure C.2. Enter the appro-
priate fluid properties calculation method (generic or API-2540). If API-2540 is selected, specify the type of fluid from
the dropdown list provided.
• The API-2540 flow rate calculation method allows the user to select from a list of commonly measured liquids. The
user supplies the appropriate base temperature, density and viscosity of the fluid used in his application.
• When the API-2540 fluids do not adequately represent the thermal expansion properties of an application’s liquid or
when precise thermal expansion coefficients are known, the Generic fluid property calculation method with the Indi-
vidual Applications fluid property calculation selection is recommended.
• When only the flowing density and viscosity of the fluid are known, the Generic fluid property calculation method
with the Manual Flowing Density Calculation selection is recommended. If a base liquid density is also known, the
Scanner will correct the liquid volumes from the static flowing conditions to the base conditions.
Select the appropriate flow rate calculation method. Then, click OK to return to the Configure Flow Run screen.
Figure C.2
a. Under the Maintenance Task heading, click Change API Liquid Parameters. The other selections will vary,
depending on the flow rate calculation method selected.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure C.3, page C-3).
C-2
ModWorX Pro Appendix C
Figure C.3
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: API-2540 Crude Oil) appears near the
top of the New API Liquid Properties section. The calculation method is read-only. To change this method, see Con-
figuring Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
5. In the Base Temperature Reference field, select either United States (default) or Canada and UK. Choose the region
that most closely matches the base condition requirements of the region where the Scanner will be used.
Note The density source, base density, and viscosity parameters can be populated with a set of default parameters, or
with values supplied by the user. To use the set of default values, click Load Default Params near the bottom of
the screen.
6. In the Density Source field, select a source to be used for the density value: Absolute Density, Specific Gravity, or
API Gravity.
7. Enter the Base Density (Absolute, Specific Gravity or API Gravity) value of the liquid. Density is assumed to be at
base temperature.
8. Enter the viscosity value of the liquid. Viscosity is assumed to be at flowing temperature. If the viscosity is un-
known, use the recommended default value for the liquid type selected.
9. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
10. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current API Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
11. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
12. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
C-3
Appendix C ModWorX Pro
Figure C.4
a. Under the Maintenance Task heading, click Change Generic Liquid Parameters. The other selections will
vary, depending on the flow rate calculation method selected.
b. Click OK to save the settings and return to the Maintain Flow Run screen (Figure C.5, page C-5).
2. Note that current liquid parameters appear on the right half of the screen and the new liquid parameters will be
entered in the left half of the screen. If you are making a minor change to current parameters, click Copy from Cur-
rent. Current parameters will be pasted into the fields on the left. This will help to minimize the number of fields that
have to be changed.
3. The fluid properties calculation method selected for the flow run (Example: Generic Liquid) appears near the top of
the New Generic Liquid Properties section. The calculation method is read-only. To change this method, see Config-
uring Fluid Properties, page 57.
4. In the Analysis Reference No. field, enter a reference number from the liquid analysis report, if desired.
5. In the Base Temperature Reference field, select either United States (default) or Canada and UK. Choose the region
that most closely matches the base condition requirements of the region where the Scanner will be used.
C-4
ModWorX Pro Appendix C
Figure C.5
6. Select the method for calculating flowing density in the Flowing Density Calculation field (Figure C.6, page C-6).
Select either Manual Entry of Flowing Density or Individual Applications (Table 6C).
–– Manual Entry of Flowing Density: Recommended for applications in which only the flowing density and viscos-
ity are known. If a base density is not known, resultant volumes will be at flowing temperature. If a base density
is entered, the Scanner will calculate a mass flow rate and a corrected liquid volume at the base temperature
condition.
–– Individual Applications (Table 6C): Recommended for applications in which API-2540 fluids do not adequately
represent the thermal expansion properties of the process liquid or when precise thermal expansion coefficients are
known. Utilizing Table 6C from API-2540, the Scanner will compute the mass flow rate and the liquid volume at
the base temperature condition. When this method is selected, the user is prompted to select a temperature coef-
ficient or “alpha” value rather than a flowing density value (see step 9).
C-5
Appendix C ModWorX Pro
Figure C.6
Note When the Flowing Density Calculation method is set to “Manual Entry,” the density source, base density, and
viscosity parameters can be populated with a set of default parameters which are the values for water. While the
default values are specific to water, they indicate the value range that is applicable for these parameters. To load
the default values, click Load Default Params near the bottom of the screen.
7. In the Density Source field, select a source to be used for the density value: Absolute Density or Specific Gravity.
8. Enter the Base Density (Absolute or Specific Gravity) value of the liquid. If the base density of the liquid is un-
known, enter the flowing density into both the Base Density and Flowing Density fields.
9. Enter the Flowing Density (Absolute or Specific Gravity) value, if applicable. If the Flowing Density Calculation
method is set to Individual Applications,this field will be displayed as “Alpha” and the user will enter the appropri-
ate thermal expansion coefficient that has been provided for his specific application, or experimentally determined to
be suitable (Figure C.7, page C-7). Thermal expansion coefficients must be derived in accordance with API-2540,
Section 11.1.6. Click the Select Alpha button to select a predetermined alpha value (Figure C.8, page C-7).
10. Enter the viscosity of the liquid. Viscosity is assumed to be at flowing temperature. If the viscosity is unknown, a
value of 1.0 cP (viscosity of water) is recommended.
11. Enter a note in the Note field, if desired (for example, to identify the liquid parameters for future reference). The
note will be stored with the liquid parameter settings.
12. Click Save Changes in the upper right corner of the screen. The changes will be saved as the Current API Liquid
Parameters and will be put into effect when the user exits the Maintain Flow Run screen.
13. To perform a new maintenance task without leaving the screen, click on Change Task and make a new selection.
14. To exit the Maintain Flow Run screen and return to the Main Display screen, click Exit Maintenance Mode in the
upper right corner of the screen.
C-6
ModWorX Pro Appendix C
Figure C.7
Figure C.8
C-7
Appendix C ModWorX Pro
C-8
WARRANTY - LIMITATION OF LIABILITY: Seller warrants only title to the products,
software, supplies and materials and that, except as to software, the same are free
from defects in workmanship and materials for a period of one (1) year from the date
of delivery. Seller does not warranty that software is free from error or that software
will run in an uninterrupted fashion. Seller provides all software “as is”. THERE ARE
NO WARRANTIES, EXPRESS OR IMPLIED, OF MERCHANTABILITY, FITNESS OR
OTHERWISE WHICH EXTEND BEYOND THOSE STATED IN THE IMMEDIATELY
PRECEDING SENTENCE. Seller’s liability and Buyer’s exclusive remedy in any case
of action (whether in contract, tort, breach of warranty or otherwise) arising out of the
sale or use of any products, software, supplies, or materials is expressly limited to
the replacement of such products, software, supplies, or materials on their return to
Seller or, at Seller’s option, to the allowance to the customer of credit for the cost of
such items. In no event shall Seller be liable for special, incidental, indirect, punitive
or consequential damages. Seller does not warrant in any way products, software,
supplies and materials not manufactured by Seller, and such will be sold only with the
warranties that are given by the manufacturer thereof. Seller will pass only through to
its purchaser of such items the warranty granted to it by the manufacturer.