HPLC Calibration Format
HPLC Calibration Format
S. Description of the
Acceptance Criteria Observation Remarks
No. calibration parameter
Calibration of flow rate
1.
Port A B C D ---
0.49 mL/min to 0.51
0.5 mL/ min
mL/min.
0.98 mL/min to 1.02
1.0 mL/min
mL/min.
1.96 mL/min to 2.04
2.0 mL/min
mL/min.
Calibration of GPV
2.
A/B/C A/B/D ---
The relative height of
19.0 to 21.0 %
peak
The measure value
Injection volume should be between ± 1
3.
accuracy µl of the injection
volume 50 µl.
4. Injector linearity
Injection volume (µL) 5 10 20 50 100 ---
%RSD of area NMT 1.0
%RSD of RT NMT 0.5
Correlation
coefficient NLT 0.999