Vacuum Tube Volt Meter VTVM How It Works How To Use It Rhys Samuel 1956 254p
Vacuum Tube Volt Meter VTVM How It Works How To Use It Rhys Samuel 1956 254p
V.T.V.M.
• how it works
• how to use it
by RHYS SAMUEL
Instrument characteristics
Low circuit loading in voltage measurements. Input capacitance and
a.c. resistance of the v.t.v.m. Wide-range frequency response on
a.c. measurements. High sensitivity. Burnout-resistant meter circuit.
Resistance readings over wide ranges. Accuracy.
Meter scales
Typical meter scales. Multiplication factor. Scale divisions. Scale
subdivisions. Scale units. How to read scales. Linear and nonlinear
scales. Scale compression. Zero-center scales. Types of zero-center
scales. Decibel scales.
37
TV troubleshooting
Analyzing TV receivers. Tube troubles. A.g.c. circuit troubles.
Troubleshooting components, tuners, picture and sound i.f. ampli
fiers. Intermittents. Plotting a bandpass curve. Trap alignment.
Troubleshooting in the video amplifier. Checking picture tubes, sync
and deflection circuits.
79
chapter page
adjustments.
Preliminary
test records.
Checking phonograph equalizers. Determining input and output
impedances.
123
fl Miscellaneous applications
Calibrating generator and oscillator output. Power-supply measure
ments. Checking battery voltages, bias cells, transformers, tubes,
potentiometers. Scope calibration. Null indications. Checking selen-
ium-rectifier circuits, TV antennas, transmission lines, reactance,
impedance. Q and inductance measurements. Direct db readings.
Checking r.f,-amplifier gain. V.t.v.m. preamplifier.
157
Index 219
introduction
Rhys Samuel
6
chaptei
instrument characteristics
Fig. 10I-a,-b. In the illustration at the left (a) the loading effect of the meter is
such that it reduces the voltage across the lower resistor from 100 to 80. The
equivalent circuit at the right (b) shows that the low impedance of the meter
upsets the voltage distribution in the network.
voltage which actually appears at the center point of the network
when the low-resistance voltmeter is connected. The voltage will
rise, however, as soon as the shunt resistance of the voltmeter is
removed. A high input resistance is desirable in a d.c. voltmeter,
therefore, to minimize loading effects and provide a more accu
rate indication of circuit operating voltages. The d.c. input re
sistances of service type vacuum-tube voltmeters range from ap
proximately 10 megohms to as high as 50 megohms.
When d.c. voltage measurements are made in circuits contain
ing a.c., such as in the plate circuit of an amplifier stage, capaci
tance loading may cause serious disruption of circuit operation
unless the input capacitance of the voltmeter is kept low. Re
duced a.c. loading in d.c. voltage measurements is achieved in
most vacuum-tube voltmeters by use of an isolating resistor in
the d.c. voltage probe. This resistor acts to reduce the effective
input capacitance of the instrument to a few micromicrofarads
(see Chapter 4). Low input capacitance and high resistance,
which increase the overall a.c. input impedance of the voltmeter,
are desirable to prevent excessive loading and detuning of high-
impedance circuits and to insure dependable a.c. voltage measure
ments. Commercial vacuum-tube voltmeters using a vacuum-tube
9
signal rectifier usually have an input capacitance of less than 50
p[if and an a.c. resistance of approximately 1 megohm or higher.
There are some vacuum-tube voltmeters, however, which em
ploy a copper-oxide or selenium type signal rectifier, have much
lower input impedance characteristics. Consequently, their a.c.
loading effect is much higher and their frequency response is usu
ally limited to power-line frequencies or a few kilocycles at best.
The majority of vacuum-tube voltmeters contain voltage di
vider networks and bridge circuits which make possible input
resistances of several megohms on d.c. voltage scales and generally
high input impedances on a.c. measurements. Because circuit
loading is greatly reduced, more accurate indications of actual
circuit operating voltages are possible.
High sensitivity
The v.t.v.m. is a high-sensitivity instrument primarily because
it employs vacuum tubes, which are highly sensitive amplifiers.
10
The signal voltage obtained from the circuit under test is applied,
in part, to one grid of a triode bridge circuit. Because of the
amplifying action in the vacuum tube, this relatively small voltage
can control a large amount of plate-current flow.
The sensitivity of a properly designed vacuum-tube bridge cir
cuit can be made so high that an indicating meter of relatively
low sensitivity may be used. This high sensitivity permits use of
a more economical and more rugged meter movement which is
better able to stand the rigors of daily service work. Meters hav
ing a basic sensitivity of 200 pa to 1 ma are ordinarily used in
v.t.v.m. circuits.
Dependability
The v.t.v.m. is a dependable instrument principally because it
is a simple instrument. Its principles of operation, circuit design
and construction are straightforward, requiring only that the user
be familiar with the circuitry he is testing and that he know a few
fundamentals important to electrical measurements. When prop
erly used, the v.t.v.m. can be expected to give long, trouble-free
operation. Design of these instruments is usually conservative.
Operating voltages on bridge tubes and power rectifiers are usually
well below maximum ratings, which greatly extends the life ex
pectancy of tubes, components and the instrument in general.
The v.t.v.m. also has some miscellaneous features which help
make it a good all-around instrument. These include the ability
to measure a.c. voltages in d.c. circuits and the ability to measure
the d.c. component of a.c. signals; a wide range of voltage meas
urements from approximately 0.1 to 1,000 or 1,500 volts both a.c.
and d.c.; resistance measurements from a few tenths of an ohm
to 1,000 megohms; good stability over a range of line-voltage con
ditions; light weight and portability and adaptability to numer
ous other measurements through the use of external accessories.
For example, high-voltage probes are available for extending the
d.c. voltage measurement ranges of the v.t.v.m. up to 30,000 volts
and higher. These relatively inexpensive probes contain a special
precision resistor which drops the high voltage to a value which
can be measured with the v.t.v.m. Well insulated, the probes can
be attached directly to the d.c. voltage input connector of the
v.t.v.m. No modification of the v.t.v.m. is required. High-fre
13
quency probes which either slip on or connect to the probes sup
plied with the v.t.v.m. (or which attach to the front-panel con
nector of the instrument) can be used to extend the high-frequency
measurement range of the v.t.v.m. well into the megacycle region.
These accessories and their principles of operation are discussed
in Chapter 4.
These, then, are the major features and advantages of the
v.t.v.m. Altogether, they make the v.t.v.m. one of the most ver
satile and reliable of all measuring instruments. When it is rea
lized that a single instrument incorporating all of these features
is within the price range of nearly everyone engaged in servicing,
experimenting or electronic construction, the v.t.v.m. has an even
greater attraction.
14
chapter
15
a vacuum-tube bridge circuit, a.c. signal rectifier, voltage divider
networks, power supply and, on some instruments, a compensat
ing section for contact potential in the a.c. signal detector, a spec
ial problem discussed further on in this chapter. A v.t.v.m. (the
RCA WV-77A Junior VoltOhmyst) may employ as few as two
tubes or as many as five (the Sylvania 221Z and 134A), including
the rectifier in the diode probe. The tubes are often dual-section
types, however, and may be used to perform more than one
function. In the case of the Junior VoltOhmyst, four-tube per
input divider networks for the three functions are designed so that
only a small part of the input voltage (1.5 to 3 volts maximum)
reaches the bridge circuit.
The power supply provides the heater and plate voltages for
the bridge tube and heater voltages for the a.c. signal rectifier
and contact-potential bucking tube, if one is used. Voltage for
ohms measurements is obtained from an internal low-voltage dry
cell. The compensating circuit for contact potential is used
when the v.t.v.m. is switched to a.c. measurements. An explana
tion of this circuit will be given later.
19
voltmeters is the familiar ohms control, located on the front
panel. This control usually requires adjustment whenever the
instrument is used. Other calibrating controls, however, are lo
cated inside the instrument case and are adjusted at the factory.
20
network, whether the instrument is to measure 1,500 or 1.5 volts.
On the WV-97A, approximately 1.35 volts is required at the grid
of the bridge input tube for full-scale meter deflection on all
ranges and functions. The d.c. voltage divider and other divider
networks are designed accordingly. It can be seen from Fig. 204
that as the voltage range increases, a smaller amount of the
applied voltage is tapped off for application to the bridge tube.
With 1,500 volts applied to the instrument, only about .09% or
Fig. 204. Voltage-divider network at the d.c. input to a v.t.v.m. The input
resistance, including the resistance of the external probe, is 11 megohms.
Some v.t.v.m.’s have an input rated at 25 megohms.
1.35 volts of the input voltage will be tapped off. The relation
ship of the individual resistances which make up the series net
work is such that when the maximum rated voltage for any par
ticular range is applied, the voltage developed between the cor
responding range tap and ground is always approximately 1.35.
Generally, commercial vacuum-tube voltmeters require between 1
and 3 volts at the grid of the input bridge tube for full-scale
deflection.
The input resistance of a v.t.v.m. on d.c.-voltage measurements
is calculated simply by adding the values of all the resistors in
series from the d.c. voltage input terminal to ground, plus the
value of the isolating resistor in the d.c. probe. The input resist
ance is constant for all d.c. voltage ranges. The purpose of the
isolating resistor (one megohm d.c. probe) is described in Chapter
4.
The values shown in Fig. 204 total 11 megohms (including 1
21
megohm in the probe) which is the input resistance and resist
ance-loading factor of the Senior VoltOhmyst. Current drain from
the circuit under test is about 100 pa on the 1,500-volt range and
only 0.14 pa on the 1.5-volt range! Circuit loading, therefore, is
not a serious problem.
The d.c. input resistance can be made higher, of course, by
employing a divider network made up of higher resistances. This
is not usually done, however, for two reasons. The first is that a
very high input resistance creates problems of stability. Leakage
resistance of switch wafers, wiring insulation and probes and
cables, all of which are affected by temperature, humidity and
dust, act to shunt the high-resistance divider network and change
its effective resistance. The second reason involves bridge circuit
design and vacuum-tube characteristics. It is highly desirable and
necessary to keep grid current as close to zero as possible. A direct
current as small as 1 pa flowing through a grid-to-ground resist
ance of 1 megohm can generate 1-volt d.c. Such a voltage is more
than enough to cause serious trouble in bridge-circuit action. If
a d.c. input resistance of 30 or 40 megohms is employed, extreme
care must be taken to restrict leakage resistance and to keep grid
current at a very low level. These measures require the most
careful circuit design and the use of especially high-quality com
ponents. However, these are expensive measures which experi
ence has shown are unnecessary in a service type v.t.v.m.
In Fig. 204, Rx and Cx comprise a simple filter network designed
to keep any high-frequency a.c. voltage picked up in the grid cir
cuit from being rectified by the input bridge tube and thereby
affecting meter indications. Rx also acts as a limiting resistor to
prevent overload voltages from driving the grid of the tube too
far in a positive direction. This filter network is part of the grid
circuit of the input bridge tube and is, therefore, in operation on
all functions. Similar networks are used on all commercial vac
uum-tube voltmeters.
The precision of the resistors used in the d.c. divider network
have a great influence on the accuracy of the voltage reading.
Wirewound or deposited carbon-film resistors having a tolerance
of ±1% or better are generally used since a change in value of
any resistor in the network can affect the accuracy of the voltage
reading on all ranges. Consequently, the resistance divider net
work is one of the first circuits to check when the v.t.v.m. starts to
give haywire d.c. voltage readings.
An important feature provided in most vacuum-tube voltmeters
22
is polarity switching on d.c. voltage measurements. The function
switch, which has positions marked -(-d.c. and -d.c., provides a
means of reversing connections of the meter terminals between
the two bridge tubes. By merely resetting the front-panel switch,
upscale deflection can be obtained on either positive or negative
d.c. voltage measurements. This convenient feature eliminates the
necessity of reversing test leads when changing the polarity of
measurement.
Resistance-measurement network
Two types of arrangements are used in the “ohms” measurement
section of the v.t.v.m. Both work on the same principle: the value
of the unknown resistance is compared with the value of a precis
ion standard resistor in the v.t.v.m. The answer is read directly in
~ JUV
o
—RLTER NETWORK
Fig. 205. Resistance measuring circuit of a v.t.v.m. Some v.t.v.m.fs use 1.5 volts
as the voltage source for resistance tests; others use 3 volts (see Fig. 206).
ohms from a v.t.v.m. scale specially calibrated in units of
resistance.
The comparison works on a principle in which the external
resistance is connected in series, by means of the ohms probe and
cable and the ground cable, with the standard resistor and the
internal battery. The most common arrangement is shown in
Fig. 205. The circuit is that used in the Eico model 214, which
has five resistance ranges for measuring values up to 1,000
megohms.
A 1.5-volt battery is used as the voltage source. When the range
switch is set to any of the five positions, one of the standard resist
ors is connected in series with the battery and the ohms probe. If
no external resistance is connected across the circuit, no direct-
current path will exist in the battery circuit, no current will be
drawn from the battery and a maximum amount of voltage will
be present at point A. This will cause the tube to conduct and,
in turn, cause the meter pointer to swing to the full-scale point.
The full-scale point, then, indicates an infinite amount of
resistance.
23
When an external resistance, shown as Rx, is connected across
the circuit, the d.c. path from the positive terminal of the bat
tery is completed through the standard resistor, Rx, the ohms
probe and cable and the ground cable. Current will flow, pro
ducing a voltage drop across both resistors. The ratio between
the values of the two resistances will determine the voltage avail
able at point A. This voltage will always be less than the full
battery voltage, causing a corresponding reduction in the voltage
reaching the grid and a smaller excursion of the meter pointer.
24
may be determined easily. (Typical ohms scales are shown in
Chapter 3.)
The circuit arrangement shown in Fig. 206 works in the same
manner as that shown in Fig. 205 except that, as the resistance
range is increased, the effective value of the standard resistor is
made up of two or more resistances connected in series. The dif
ferences in the merits of each arrangement are of no great impor
tance and both systems are used with good results in many vac
uum-tube voltmeters.
Most commercial units employ a single 1.5-volt flashlight cell as
the voltage source for ohms measurements. Some vacuum-tube
voltmeters, however, employ two batteries in series to give 3 volts.
25
Reliable a.c. voltage measurements can be assured if the tech-
nican is aware of these basic facts:
1. All vacuum-tube voltmeters are designed to provide a spe
cific type of a.c. voltage reading—either r.m.s., peak, or peak-to-
peak.
2. The shape and repetition rate of the waveform being meas
ured have a direct effect on the accuracy of the meter indication.
3. The v.t.v.m. is designed to measure a.c. signals over a fre
quency range defined by the manufacturer. Th« frequency re
sponse of the v.t.v.m. is not unlimited.
Service type vacuum-tube voltmeters are designed to indicate
directly eithers r.m.s. or peak-to-peak values of a.c. voltages. The
a.c. signal rectifier used in most r.m.s. indicating instruments is a
half-wave peak rectifier: the meter scales, however, are calibrated
TABLE 1
CONVERSION TABLE FOR SINE WAVES ONLY
MULTIPLY to get:
R.M.S. BY PEAK BY P-P BY
— 0.707 0.353 R.M.S.
1.414 — 0.5 PEAK
2.83 2.0 — PEAK-TO-PEAK
4Ht -Jdc
^symmetrical complex wave pulse
Fig. 208. Types of waveforms that you will encounter in radio and
TV servicing. Many v.t.v.m/s are designed for sine-wave measure
ments only.
sine waves, but the values of complex wave shapes as well. These
instruments employ special rectifier circuits however.
It is necessary, therefore, that the technician understand exactly
the types of waveforms his v.t.v.m. is designed to measure. This
information should be provided by the manufacturer in his pub
lished specifications for the instrument. In cases where no infor
mation is given regarding waveshape, it is safer to assume that on
a.c. the instrument is designed to measure sine waves only.
Frequency response
The frequency response, or the range of frequencies over which
the v.t.v.m. will provide accurate a.c. voltage readings, depends
upon the a.c. circuit and input characteristics of the instrument.
Mainly, the input resistance and capacitance are the governing
factors. These characteristics may be uniform throughout all the
voltage ranges of the instrument or they may vary from range to
range. All vacuum-tube voltmeters are limited to a certain range
of frequencies; the range may run from a few cycles per second
to several megacycles. Special high-frequency probes are often
available for extending the response into higher frequencies.
27
The input capacitance should be kept low and the resistance
made as high as possible to enable the v.t.v.m. to work at the
higher frequencies. The shunting capacitance may have a rela
tively high reactance at audio frequencies, but at radio frequen
cies the reactance may be low enough to attenuate the input sig
nal seriously. The meter reading, then, will be lower than the
normal value of the source voltage. The v.t.v.m. should be used
only for frequencies within its response range.
Fig. 209 shows a half-wave rectifier circuit which is similar in
Fig. 210. A.c. measuring circuit of a peak-to-peak type of v.t.v.m. This circuit
enables the instrument to measure the peak-to-peak values of complex waveforms.
beyond R4, R5. On the 300- and 1,200-volt ranges, however, at
tenuation occurs in the network made up of RI, R2 and R3 before
the voltage reaches the rectifier. This arrangement prevents the
inverse peak voltage rating of the tube from being exceeded.
Another advantage of this circuit arrangement is that a common
voltage divider network may be used on some a.c. and d.c. voltage
measurements.
A peak-to-peak rectifier circuit is shown in Fig. 210. This cir
cuit, which is used in the RCA WV-97A Senior VoltOhmyst, is
capable of rectifying sine waves and a variety of pulses and other
nonsymmetrical complex waves. The d.c. output voltage is pro
portional to the peak-to-peak value of the waveform.
Seven voltage ranges are provided. Measurement of peak-to-
peak values of waveforms other than sine waves is made possible
by the use of component values which give the rectifier circuit a
long time constant. When the applied a.c. signal swings negative,
C3 is charged through the bottom diode section to the negative
29
peak value of the applied voltage. As the voltage starts to swing
in a positive direction, this diode section becomes nonconducting
and the charge on C3 is prevented from discharging.
As the a.c. signal swings positive, the positive peak is added in
series with the charge on C3 and is applied to the plate of the
other diode section. C4 is then charged to a value equal to the
sum of the positive and negative peaks. Because of the relatively
long time constant of the circuit, the voltage across C4 will be
maintained at the peak-to-peak value of the applied a.c. voltage.
This charge is then applied as a d.c. voltage to the a.c. voltage
divider and the bridge tube.
RI, R2 and R3 are a voltage divider which limits the voltage
applied to the 6AL5 and prevents damage on the high-voltage
ranges. Cl is a blocking capacitor which prevents any d.c. volt
age in the test circuit from reaching the divider networks and
diodes. C2 is a frequency-compensating capacitor which is ad
justed at the factory to maintain a frequency response essentially
flat up to 3 me on all ranges up to and including the 500-voIt
range. Response may be extended to 250 me when an accessory
crystal rectifying probe is used. The crystal probe, however, does
not provide for peak-to-peak readings, and meter indications are
in r.m.s. values.
31
supply voltage divider. R is a variable resistor, usually a poten
tiometer, which is set correctly at the factory. In some instru
ments, the diode detector may be connected differently, causing a
negative voltage to be applied to the bridge. In these cases, a posi
tive bucking voltage is tapped from the power supply. The
method described in Fig. 212 is employed in the RCA WV-77A
Junior VoltOhmyst.
Another circuit arrangement is shown in Fig. 213. This a.c.
circuit is a full-wave peak-to-peak rectifier. The diode V2 is used
33
use, current flow through the meter is restricted to a maximum
of approximately 1 ma. The switch is connected in series with
the meter rather than in shunt so that the contact resistance will
not change the accuracy of the meter reading. The meter shunt
circuit is never broken so that no overload current surges will
occur when ranges are changed. The series resistance increases
on the higher current ranges and gives the meter some degree of
protection against current surges.
Instruments which use this type of circuit do not utilize the
bridge circuit and vacuum-tube functions of the instrument and
Power supply
Fig. 216. Power supply for the v.t.v.m. is small, uses selenium
rectifier and simple R-C filter. A vacuum-tube rectifier is
sometimes used in place of the selenium unit.
35
correcting circuit. Most power supplies are required to furnish
only the B+ and filament voltages. A resistance divider network
is used where it is necessary to furnish bias voltages or a contact
potential bucking voltage.
Commercial vacuum-tube voltmeters have simple power re
quirements; both the tube-heater and B-supply demands are
extremely small. Because the bridge circuit and a.c. signal recti
fier are required to furnish little power, B plus current drain is
usually limited to a few milliamperes. To minimize grid current
flow in the bridge tubes, a critical factor which was discussed on
page 22, some of the heaters are run at reduced voltages, often as
low as 5.8 for heaters which would normally be operated at 6.3
volts. Total power consumption of the v.t.v.m. is in the order of
5 to 10 watts.
36
chapter
meter scales
Fig. 301. In this illustration we have the dial scales of a modern v.t.v.m.
The use of a number of scales having various ranges gives the instrument
considerable versatility, extends its usefulness.
Scale divisions
Examples of the methods used to subdivide scales are shown in
Fig. 302. The simplified scale shown in Fig. 302-a has 10 major
divisions. Because no minor divisions are used, the operator must
mentally subdivide the scale. Fortunately, high precision is not
Fig. 302-a to e. Scales can be subdivided in a number of different ways. The technique
in reading a scale is the same as that used in reading a ruler.
required in most service measurements. The scale provides ade
quate spacing between divisions, however, and can be read without
difficulty to a half and a quarter of a division.
The division of meter scales may be compared to the methods
used to divide an ordinary ruler. As shown in Figs. 301 and 302,
scales are divided into major and minor divisions. The scales
of a v.t.v.m. are calibrated from left to right, with the full-
scale points on the right-hand end. On some units, however, the
“ohms” scale is reversed, with the zero or short-circuit point at the
right. This practice is conventional on many non-electronic
ohmmeters.
The scale shown in Fig. 302-b includes subdivision marks. Note
that these subdivision marks are shorter than the major division
39
marks and divide the major sections into five equal parts. Each
subdivision, therefore, represents one-fifth or 0.2 of a major
division. For example, the third subdivision mark to the right of
zero would represent 3 X 0.2, or 0.6. The second subdivision
mark to the right of 8 represents the setting for 8.4. Because it is
convenient to subdivide the scale mentally even further, it is pos
sible to read an indication between subdivisions. For example,
3.5 is located between the second and third subdivision marks to
the right of 3.
The scales on different makes of vacuum-tube voltmeters are not
always marked in the same units. Some scales have major divi
sions marked off in units of 10. Others may be calibrated with
decimal fractions or combinations of odd and even whole numbers
and decimal fractions. The subdivisions may be based on units of
one, tenths, quarters or other convenient fractional units. In all
cases, it is possible to determine easily the exact value of each
subdivision. This is done simply by dividing the quantity repre
sented by a major unit by the number of subdivisions between
major division points. For example, the scale shown in Fig. 302-c
has three types of division marks. Each major division mark repre
sents a step of 0.2. Because there are four subdivisions (formed by
three subdivision marks) between each numbered point, the sub
divisions represent steps of .05 (0.2 4- 4 = .05). The 1.7 point,
for example, is half-way distant between 1.6 and 1.8; the 0.65 point
is denoted by the first .05 mark to the right of 0.6.
Frequently, one scale base serves for two different types of scale
calibration. Examples of this type scale base are shown in Fig.
302-d and Fig. 302-e. In Fig. 302-d, one scale base serves for the
0 to 2.5 scale, with subdivision marks representing units of .05
(0.25 -r- 5 = .05). The other scale is calibrated from 0 to 50, with
minor division marks representing steps of 1. The scale in use is
determined by the setting of the range switch. When the 2.5 scale
is used, the third subdivision mark to the right of zero represents
3 X -05, or 0.15. This same point represents 3 on the 50 scale.
Points equally distant between subdivision marks would represent
steps of .025 on the 2.5 scale, and steps of 0.5 on the 50 scale.
Another method of sharing a scale base is shown in Fig. 302-e.
This type of scale is often found on units which read both the r.m.s.
and peak-to-peak value of a.c. waveforms. Because the peak-to-
peak value of a sine wave is always equal to 2.83 times the r.m.s.
value, it is possible to calibrate the same scale base in both r.m.s.
and peak-to-peak units so that both types of readings may be had
40
simultaneously. The scales are usually distinguished by different
colors.
Linear and nonlinear scales
All the scales shown in Fig. 302 have a common feature: the
calibration points are evenly spaced throughout the scale. These
are linear type scales. It is often necessary, however, to calibrate a
Fig. 303-a, -b. Typical resistance scales shown above are nonlinear. The right
hand end of such scales is so crowded that it is often difficult to read. It is
easier to make an error reading these than when reading voltage or current
scales.
scale in units which have nonuniform spacing between them.
These are called nonlinear scales. The ohms scale is always non
linear. Two representative ohms scales are shown in Fig. 303.
The calibration of the scale in Fig. 303-a, which is from the Syl
vania model 301 Polymeter, is representative of those used on the
majority of vacuum-tube voltmeters. The scale shown in Fig.
303-b is patterned after that used on the Hickok model 209A.
The most notable feature of these scales is the compression at
the high end. For example, in the scale in Fig. 303-a, the interval
between major division points becomes narrower as the scale
progresses. From 10 to 1,000 the scale intervals narrow down so
much that different calibration units must be used. From 0 to 10,
the major calibration unit is 1 with subdivisions at 0.5 intervals.
From 10 to 20, the major unit is 5, with subdivisions representing
steps of 2.5. As the scale continues toward the full-scale point, the
intervals become so narrow that all remaining major divisions are
41
marked off in progressively larger units of 50, 200, 500 and 1,000.
Subdivision marks in this upper region of the scale also represent
larger steps.
Although the scale shown in Fig. 303-b is calibrated differently
from that in Fig. 302-a, the nonlinear spacing between units can
be seen. On both these scales it is important to remember that
subdivision marks between major calibration points represent
different quantities at different points along the scale. Therefore,
it is important that each subdivision mark be understood if cor
rect readings are to be obtained.
The last mark on the scale does not have a number. This is the
Fig. >304. Some low a.c. voltage scales are also nonlinear, a situation caused
by the nonlinearity characteristic of the rectifying diode (used in the
v.t.v.m.) at low a.c. signal levels.
Zero-center scales
In many cases, the zero-adjust control of a v.t.v.m. has suffi
cient range to permit setting the meter pointer in the center of the
scale. When this is possible, the v.t.v.m. can be used as a null
indicator for making alignment checks on FM discriminators and
b C
Fig. 305-a, -b, -c. Typical zero-center scales used in some v.t.v.mfs. Some scales
are calibrated, others simply have a center marker and a plus-minus indica
tion. Calibrated scales are helpful in the alignment of circuits such as FM
detectors.
indicating circuit balance in bridge measurements. Many instru
ments have separate zero-center scales. These scales may be cali
brated directly in volts or be provided with reference markers to
indicate the degree of positive or negative unbalance in the cir
cuit. Applications which use the zero-center feature are discussed
in Chapter 7.
Some representative zero-center scales are shown in Fig. 305.
The scales shown in Fig. 305-a are used on the Jackson model 709
Tele-Volter. For the use of these scales, the v.t.v.m. must be set up
for +dc voltage measurements. The meter pointer is set to the
center 0 by means of the zero-adjust control. The d.c. voltage
range used should cover the overall voltage swing, from maximum
positive to minimum negative. For example, a circuit which can
be adjusted to provide a voltage ranging from -25 to +25 would
43
require that the 50-volt range be used. In most null-indicating
applications, however, voltages are low, often in the order of 2 or 3,
and a low-voltage d.c. range is employed. Two scales employ the
same base on this instrument.
An example of a zero-center scale which employs uncalibrated
reference marks is shown in Fig. 305-b. The scale shown in Fig.
305-c indicates only the polarity of the unbalance. In ordinary
applications, these types of scales are used where only a null indica
tion is required.
Decibel scales
A few commercial v.t.v.m.’s are equipped with special scales for
use in audio troubleshooting and measurement. These scales may
be calibrated in units of db, dbm, or VU, depending upon the
type of load with which the v.t.v.m. is used. In some v.tv.m.’s,
this load may be built into the instrument; in others it must be
set up externally. Because load characteristics are important in
44
chapter
CABLE
ADAPTER
PROBE BODY
Fig. 401-a, -b. Direct probe (a) shields the test lead almost up to the probe tip.
The lower illustration (b) shows slide-on type probes.
cause a minimum of disturbance to circuit operation. This re
quirement is met by means of different types of probes and cables,
which can be classified in this manner:
1. Direct probes and cables
2. Isolating probes
3. Rectifying probes
4. Multiplier (divider) probes
46
transfer electrical information from the test circuit directly to the
v.t.v.m. without attenuation or modification.
On most instruments, the direct probe and cable is used for
measurements of resistance, and low-frequency a.c. voltage. Some
vacuum-tube voltmeters also employ a direct probe and cable for
d.c. voltage measurements. On these instruments, the components
normally housed in the isolating probe or a.c. probe are placed
inside the instrument.
Fig. 401 shows two types of direct probes which are representa
tive of those used on all vacuum-tube voltmeters. That in Fig.
401-a is an insulated wire lead which has a needle point on the
probe end. The connector at the meter end of the cable may be a
phone tip, pin plug, spade lug or bare wire, depending on the
PROBE TIP METAL SHIELD
a.
Fig. 402-a, -b. Cutaway view of isolating probe (a) and its equivalent circuit (b).
v.t.v.m. The probe in Fig. 401-b is equipped with a length of
coaxial cable and a microphone type connector on the meter end.
The probe is designed to accommodate additional “slip-on” type
probes, such as a d.c. isolating probe, a switch probe, or a high-
frequency probe. The principal advantages of this type are the
shielding of the inner conductor from the probe tip to the con
nector and the fact that it is not necessary to change cables when
using different types of probes.
Isolating probes
Isolating probes are used for the measurement of d.c. voltages.
For example, d.c. voltage measurements from 0 to 1,000 are made
with a shielded cable terminated with an isolating resistor, as
shown in Fig. 402-a. The shielded cable eliminates the pickup of
any stray magnetic fields about the test bench, which might other
wise result in an inaccurate indication by the vacuum-tube volt
meter. The isolating probe resistance prevents the shielded cable
from acting as a capacitive shunt across tuned and high-imped
ance circuits, such as the local-oscillator in a receiver. If it were
not for the isolating effect of the resistance, connection of the
v.t.v.m. test leads might detune the test circuit so greatly that cir
cuit operation might cease altogether.
47
The equivalent circuit in Fig. 402-b shows that the cable capaci
tance C is isolated from the probe tip by the 1-megohm resistor,
which cuts down the effective value of C to approximately 1 or
2 p|if. This is a large reduction, as compared with the approxi
mately 75 pipf of cable capacitance. Hence, the d.c. probe can be
applied across a local-oscillator circuit to measure the self-bias on
the oscillator grid without disturbing circuit operation.
The equivalent circuit is also a low-pass filter. This is an essen
tial feature, since d.c. voltages at the grids and plates of tubes often
must be measured in the presence of large a.c. voltages. If the
a.c. signal is passed along to the bridge circuit, it might be rectified
and cause an erroneous d.c. reading. Most vXv.m.’s include an-
Fig. 403. Exploded view of a d.c. (or isolating) probe. The resistor, R, most often has
a value of one megohm. The probe, held together by a single screw and hex nut, can
easily be taken apart when repairs are needed.
other low-pass filter at the grid of the v.t.v.m. tube to make certain
that no a.c. voltage reaches the grid.
These probes consist of plastic housings containing the isolating
resistor in series with the probe tip and the output end of^the
probe. While the value of the resistor used depends upon the d.c.
input resistance of the v.t.v.m., a 1-megohm resistor having a 1^- or
1-watt rating is most generally employed. An exploded-view draw
ing of a d.c. probe is shown in Fig. 403.
The d.c. isolating probe should always be used for measure
ments of d.c. voltages because d.c. voltage scales are calibrated on
the basis of the d.c. input resistance. The isolating resistor in the
probe is in series with the voltage divider networks in the v.t.v.m.
and is part of the d.c. input resistance of the instrument. If the
48
isolating probe is not used, a considerable degree of error might
result.
The switch type probe, which combines the functions of direct
and isolating probes, is finding increased popularity among v.t.v.m.
manufacturers. An exploded-view drawing of one of these probes
is shown in Fig. 404. When set to the forward or direct position,
the sliding switch shorts the 1-megohm resistor so that the probe
can be used for ohms and a.c. measurements. When set to the rear
or d.c. position, the short is removed and the resistor is placed in
series with the test lead. This switch probe is similar in operation
Rectifying probes
Rectifying probes include probes which have several different
names, such as a.c., crystal-diode, demodulator, detector, high-fre
quency, r.f. and vacuum-tube probes. Because there is no pre
49
scribed standard of classification, probes which are almost iden
tical in electrical design and application may have different names.
All have a common purpose and that is to detect and rectify the
a.c. voltage at its source in the test circuit and to provide the
v.t.v.m. with a proportional d.c. voltage. All these probes contain
diode type rectifiers and filter circuits.
Although many rectifying probes employ similar designs, they
are not generally interchangeable. A rectifying probe is usually
designed for a specific v.t.v.m. by the instrument manufacturer.
Probe design depends upon a number of factors, including the
type of voltage reading required (peak, peak-to-peak or r.m.s.),
the frequency range over which measurements are to be made, the
Fig. 407. Crystal-diode probe. The ground clip should be fastened as close to the
test point as possible.
tremely small size (permitting compact probe design) and very
short lead lengths. Consequently, crystal diodes are especially well
suited to measurements above 50 me. Contact-potential difficulties
do not exist in these devices; probe cables and circuitry are sim
plified because no heater voltage is required. Crystal diodes have
very low input capacitances (which is desirable) but, conversely,
have a lower shunt resistance than vacuum-tube diodes, a factor
which lowers their overall input impedance.
Two differerit vacuum-tube rectifying probes are shown in Fig.
408. Note the metal enclosure which shields the tube and com
ponents. Because lead length is critical at high radio frequencies,
grounds must be made as near as possible to the test point. Ope
probe is equipped with an alligator type ground clip welded
directly to the metal probe housing. A short extension lead with
an attached alligator clip can be screwed onto the front end of the
probe for use in applications at lower frequencies.
R.f. voltages up to 300 can be measured in circuits containing
52
up to 500 volts d.c. with this probe. Because the output from this
probe is the rectified input signal, readings are indicated on the
d.c. voltage scales. For r.f. voltages from 0 to 3, however, a special
r.f. scale is provided because of the nonlinearity in tube perform-
Fig. 408-a, b. The vacuum-tube probe shown at the top uses a ground clip fastened
to the probe housing. The probe at the bottom uses a grounding pin.
ance at low signal voltages. The probe has an input resistance of
2.3 megohms and an input capacitance of only 3 upf. A special
subminiature tube, type 1247, is used as the rectifier.
Of particular interest is the vacuum-tube probe shown in Fig.
408-b. This is similar in application to that shown in Fig. 408-a
but has these important differences: The former is a peak-reading
probe which provides a d.c. output voltage substantially equal to
53
the positive peak value of the applied a.c. voltage. The polarity of
the output voltage from this probe, however, is negative. Because
the rectified peak voltage can be maintained at full value only
across an infinitely high d.c. input resistance, the output voltage
is slightly less than the peak value. The instruments for which this
probe is designed have a d.c. input resistance of 13.5 megohms on
all d.c. voltage ranges. While this input resistance is sufficiently
high to prevent excessive circuit loading, it presents a load to the
rectifying probe and causes a slight drop in the peak output volt
age. This voltage drop is small enough, however, to permit calibra
tion on the basis of the true peak voltage. For almost all conven
Fig. 409-a to d. Four different types of vacuum-tube probes. Those shown in a and
b are single diodes. The probe at the lower left (c) is a peak-to-peak (voltage
doubler) type. Although the probe shown at the bottom right (d) uses a triode,
it is actually worked as a diode with the plate grounded and the control grid
operating as the anode.
tional service applications, the readings are adequate. In the few
applications which may require a high degree of accuracy, a cor
rection chart is provided.
When the probe is connected to the v.t.v.m., contact potential
causes a slight deflection of the meter pointer. The contact poten
tial is counteracted on the low a.c. ranges by setting the zero
adjust control to zero the meter pointer. Little or no adjustment
is required to counteract the contact potential on ranges above
12 volts.
Schematic diagrams of four vacuum-tube probes are shown in
Fig. 409-a, -b, -c, -d.
Construction of a typical crystal-diode type probe, together with
the schematic diagram, is shown in Fig. 410. This probe can be
54
used for measurement of r.f. voltages from 50 kc to 250 me. The
probe requires no heater voltage, is compact and slips onto the
front end of the direct probe and cable shown in Fig. 401-b. Note
how the resistors are soldered in close to the terminals and how
the 1N34-A crystal is provided with a special clip type holder.
These features make for short lead lengths, especially important
above 30 me. A ground clip is attached directly to the probe to
provide a short r.f. return path to the test circuit.
The design and construction of other makes of crystal-diode
probes are similar. Most of these probes are equipped with a
Fig. 410. Internal construction of a crystal rectifier probe. The ground lead is an
important part of the crystal probe circuit.
special ground lead which should be used in conjunction with the
regular ground lead provided with the instrument.
Although probes of this type are peak rectifiers, design of the
R-C filter circuits and the v.t.v.m. input circuits result in a d.c.
output voltage equal to the r.m.s. value of a sine wave. The peak
value of the r.f. signal may be obtained by multiplying the meter
reading by 1.41.
Peak-to-peak probes
The peak-to-peak type of probe used to measure peak-to-peak
voltages of complex a.c. waveforms, is shown in Fig. 411. Tubes
and crystal diodes are used commercially, with the same general
considerations as have been noted for the half-wave type of probe.
The tube will withstand much higher voltages than the crystal
diode.
55
In some cases, the peak-to-peak indicating arrangement is built
into the v.t.v.m. This raises the input capacitance, but also pro
vides a much greater voltage-handling capability, since a compen
sated multiplier then precedes the rectifying tubes, and will usually
raise the top range of peak-to-peak voltage measurement to 1,000.
Such a range will accommodate the majority of complex a.c. wave
forms encountered in a TV chassis.
Multiplier probes
Multiplier probes are designed to extend the d.c. voltage ranges
of the v.t.v.m. to 30,000-50,000 volts. These probes are especial
ly useful in the measurement of high voltages in television re
ceivers, X-ray equipment, transmitters and similar electronic
equipment. The probes make possible such measurements with
complete safety for the operator.
Multiplier probes are also known as high-voltage probes and
divider probes. Although the terms “multiply” and “divide” have
opposite meanings, it is possible to justify the differences in
description if the functions of the probes are understood. Because
the majority of manufacturers prefer to call these accessories high-
voltage probes, this terminology will be used here. The ways in
which these probes “divide” and “multiply” are explained further
on.
Several high-voltage probes, designed for use with service type
vacuum-tube voltmeters are shown in Fig. 412- All these probes
56
are basically similar; they provide a well-insulated housing for a
multiplier resistor which serves as a series voltage divider.
One typical example of this type has a molded polystyrene body,
a tapered and rounded probe point to reduce corona and a series
of five anti-arc baffles which serve to increase the length of the
insulated path along the body of the probe. The heavy molded
flange on the front of the handle is equipped with a metal ground
Fig. 412. Two types of high-voltage probes. These are used to extend the d.c. range
of the v.t.v.m., should not be used for the measurement of a.c. voltages.
ring on the probe side to provide a protective barrier between the
point of voltage contact and the handle. The multiplier resistor
is completely enclosed and is suspended within the probe by its
57
ends only. Note that the probe is equipped with a separate ground
lead. The ground lead is connected to the outer shield of the high-
voltage cable and provides additional safety in event of a short
circuit in the cable.
The divider resistors used in high-voltage probes are of special
construction to reduce the possibility of voltage breakdown. They
are usually spiral-wound of metallic or carbon-film ribbon and
encased in a plastic having a high dielectric. Because these resistors
may have values of several hundred megohms, they must be care
fully constructed to resist the effects of temperature and humidity
changes. Resistors for typical high-voltage probes are shown in
Fig. 413.
The method by which the probe “multiplies” or “divides” can
be seen in Fig. 414. All measurements are made with the v.t.v.m.
set to one of the positive d.c. ranges. As shown on the simplified
58
voltage measurement, multiply the voltage reading on the v.t.v.m.
by the ratio, or multiplier factor, of the probe resistance. On
either the 300- or 1,200-voIt range, the meter will indicate 100
volts. Thus, 100 X 100 = 10,000 volts.
Multiplier probes may be used with almost any type of volt
meter to measure very high voltages. The value of the high-voltage
multiplier resistor required is computed as follows:
1000 MEG
Rhv
HIGH VOLTAGE
59
(multiplier M) must be computed on the basis of the input resis
tance of the meter and the resistance value of the multiplier
resistor. A new scale factor is computed as follows:
M = RiS Z (2)
“in
For example, assume that a v.t.v.m. has an input resistance of
16 megohms, including the 1-megohm d.c. isolating resistor, and
is to be used with a multiplier-probe resistor of 500 megohms.
The new scale factor, computed by substituting values in (2), is:
16.000.000 - 1.000.000 + 500.000.000
16,000,000
- . 32 (approx.)
16,000,000
When this combination of probe and v.t.v.m. is used, all voltage
readings taken with the high-voltage probe are multiplied by 32
to obtain the correct high-voltage reading. Although numbers
which are not multiples of 10 do not lend themselves to quick
multiplication, they can still be used to obtain accurate readings.
The measurement of high d.c. voltages, such as the second anode
voltage of a picture tube, with a v.t.v.m. and high-voltage d.c.
probe are rather well understood. Less understood is the measure
ment of relatively low d.c. voltages in the presence of high a.c.
pulse voltages, such as are found at the plate of the horizontal
output tube. Receiver manufacturers commonly insert a technical
note in their service data, warning against attempts to measure the
d.c. plate voltage of the horizontal output tube. This is because the
approximately 350 volts d.c. is accompanied by an approximately
6,000 volt kickback pulse. If this high-voltage a.c. pulse is applied
to the precision input circuit of the v.t.v.m., serious damage to
the v.t.v.m. can result.
Because the high-voltage d.c. probe provides a high degree of
lowpass filter action, it is possible to use the v.t.v.m. in conjunction
with the high-voltage probe for measurements in these circuits.
For example, the v.t.v.m. can be set to its 6-volt range, and when
used with a high-voltage d.c. probe having a 100-to-l attenuation
factor will indicate the 350 volts d.c. at approximately half-scale of
the v.t.v.m.; at the same time, the low-pass filter action of the high-
voltage d.c. probe effectively protects the v.t.v.m. against the
6,000 volt a.c. pulse.
Grounding the probe
Technicians sometimes overlook the fact that the ground lead
60
or ground connection on a probe is part of the measuring circuit,
and an important part at that. The significance of a ground con
nection depends upon just what it is you are measuring. When
measuring d.c. volts in a low-impedance circuit, the ground con
nection is not critical. When measuring voltage at a high-fre
quency point, not only does the ground connection become im
portant, but the length of the ground lead itself must be consid
ered. To keep the ground lead short, some probes have a ground
ing button or clip right near the tip end of the probe.
61
TABLE 2—V.T.V.M. APPLICATIONS AND LIMITATIONS
Feature Applications J-----------------------------
Misapplications
-------------------------------|
D.c. voltage measure Checking values of 5 Attempted measure- |
ment, with high input plate, screen, and grid i ment of plate voltage of j
impedance, even on the voltages; tracing prog- > horizontal-output tube; !
low-voltage ranges. ress of signal by meas- ' high-voltage a.c. pulse
(Measurements made urement of self-bias ; damages v.t.v.m. At-
• using shielded cable ter- (only in circuits using i. tempted measurement
i minated with 1-megohm overdrive grid bias); ;' of grid bias in vertical
1 isolating resistor.) checking local-oscillator i; blocking oscillator cir
operation by measure cuit when grid leak has
ment of self-bias; check very high resistance;
ing for amplifier distor v.t.v.m. will load circuit
tion in terms of shift seriously. Attempted
of operating point with measurement of pulsat
applied a.c. signal;mea ing d.c. voltage values;
surement of voltage in v.t.v.m. indicates only
high resistance circuits the average value.
such as a.g.c. systems,
etc.
62
Feature Applications Misapplications
—......... - . ...... p . .. ’----------———. II""—.......
|
( A.c. voltage measure- Cheeking peak voltages ' Attempted checking of
; ments, in peak values, of some oscillator cir complex waveform volt-
i using external diode cuits. Checking sine ! ages in sync and sweep
। probe. wave audio-frequency circuits with v.t.v.m.
voltages. Checking line which will measure sine
voltages, power-trans waveforms only. At-
former, and heater volt 1 tempted measurement
ages, being certain to | of voltage values ex
convert from peak to ceeding input rating of
r.m.s. values of sine probe. Attempted r.f.
waves. or i.f. signal tracing.
A.c. voltage measure Checking of peak-to- Attempted measure
ments, in peak-to-peak peak values of many of ment of a.c. voltage val
values, using external the complex a.c. wave ues exceeding input rat
peak-to-peak diode forms encountered in ing of probe. Attempted
probe. (Provides higher the sync and sweep sec measurements in high-
input impedance than tions of the TV chassis. impedance circuits, such
built-in peak-to-peak in Checking a.f. voltages. as ratio-detector cir
dicating circuit, but in Checking line voltages, cuits. Attempted signal
put voltage rating is power-transformer, and tracing of r.f. and i.f.
usually less.) heater voltages, being tuned circuits.
certain to convert from
peak-to-peak to r.m.s.
values.
Resistance measure Measurement of for Attempted measure
ments, when internal ward and back resis ment of high resistance
ohmmeter battery volt tance, and front-to-back values, such as the leak
age is other than 1.5. ratios of all crystal di age resistance of paper
(Some ohmmeters pro odes, including u.h.f. or mica capacitors (un
vide a low driving volt mixer crystals. Meas less high driving volt
age on the low-resist urement of most resis age is provided on the
ance ranges, and a high tance values encoun high-resistance ranges).
driving voltage on the tered in TV chassis. Attempted measure
high-resistance ranges.) ment of resistance
values in hot circuits,
or in circuits with capa
citors retaining residual
charge.
Resistance measure Measurement of for Attempted measure
ments, when internal ward and back resis ment of the filament re
ohmmeter battery volt tance of picture-detec sistance of small bat
age is higher than 1.5 tor crystals, but not of tery-operated tubes. At
(on low ranges). u.h.f. crystals (except tempted measurement
on higher ranges, to of the internal resist
limit terminal voltage ance of meter move
to 1.5). Measurement of ments. Attempted meas
all resistances found in urement of resistance
TV chassis, with excep values in hot circuits,
tion of leakage resist or in circuits with capa
ance of small capaci citors retaining residual
tors. charge.
63
v.t.v.m. was designed to measure. Unless manufacturer has speci
fied the type of waveshapes which can be measured with the
v.t.v.m., it should be assumed that the instrument can be used
only to measure sine waves.
4. On low a.c. measurements, questionable readings can be
obtained if the special “low a.c.” scale is not used.
5. A.c. voltage is too low to be read on v.t.v.m. This situation
is especially common in signal-tracing work in low-level r.f. and
i.f. amplifiers.
Resistance measurements
1. Measurement of same resistance on two different scales gives
two readings of considerable difference. Use resistance scale which
gives reading nearest the center-scale point.
2. Highly erratic action of meter pointer on resistance meas
urements. Be sure to remove power from circuit under test or to
discharge all capacitors before making ohms measurements.
3. Ohms measurement or components installed in circuit does
not give satisfactory indications. Shunting resistances of other
resistors and capacitors can be causing misleading reading. Dis
connect one or both ends of component from circuit.
64
thi'ihht
Front-panel controls
The function control is the front-panel switch used to select
the type of measurement to be made. Although this control is
labeled selector and circuit on some instruments, the purpose
is the same. The range control is the front-panel switch used to
select the voltage, current or resistance range on which the meas
urement is to be made. On some instruments a single switch which
has a large number of positions combines the duties of both the
function and range controls.
The zero-adjust control is used to set the meter pointer at the
left-hand zero mark on all functions after power is applied to the
instrument. The ohms-adjust control is used to set the meter
pointer to the infinity mark on the ohms scale after the selector
is set to the ohms position.
These four basic controls — range, selector, zero adj and
ohms adj — are provided on the RCA WV-87A Master Volt-
Ohmyst. Because operation of this v.t.v.m. is straightforward and
similar to other commercial units, it is an excellent example for
65
Fig. 501. The modern v.t.v.m. is one of the most useful and versatile test instruments in the service shop.
demonstration. The operating techniques and precautions for this
instrument apply, in general, to all vacuum-tube voltmeters.
The v.t.v.m. is equipped with a switch type probe and cable for
a.c. and d.c. voltage measurements, an ohms probe and cable, a
ground cable and two test leads for direct-current measurements.
A slip-on type switch probe is also provided for use with the direct
probe and cable to permit a.c. and d.c. voltage measurements with
a single probe and cable. The ground cable is used for all voltage
and resistance measurements.
The first two positions on the selector switch are for negative
and positive d.c. volts. When the switch is set to —volts, a nega
tive d.c. voltage applied to the d.c. probe tip will produce upscale
deflection of the meter pointer. When the switch is set to -(-volts,
the pointer will deflect upscale when a positive voltage is applied
to the d.c. probe. This polarity-reversal feature avoids the neces
sity for reversing the ground lead and d.c. probe when changing
the polarity of d.c. voltage measurements.
The v.t.v.m. will measure both negative and positive voltages
to 1,500, a.c. voltages to 1,500 r.m.s. or 4,200 peak-to-peak, resis
tance to 1,000 megohms and direct current to 15 amperes.
67
TABLE 3
5 Range- Meter Meter-
Function Quantity To be Switch Scale Scale
Measured Setting To Use Multiplier
D.C. VOLTS 0 - 1.5 v 1.5 V A 1
(use d.c. direct probe; 1.5 - 5 ” 5 ” C 1
set switch to D.C.) 5 - 15 ” 15 ” A 10
15 - 50 ” 50 ” C 10
50 - 150 ” 150 ” A 100
150 - 500 ” 500 ” C 100
500 - 1,500 ” 1,500 ” A 1,000
OHMS 0-30 Rx 1 R 1
(use ohms probe and 30 - 300 RX 10 R 10
cable) 300 - 3,000 RX 100 R 100
3,000 - 30,000 RX 1,000 R 1,000
30,000 - 300,000 RX 10,000 R 10,000
300,000 - 3 Meg RX 100,000 R 100,000
3 Meg - 1,000 Meg RX 1 Meg R 1,000,000
IQ 40
50 ”
++++
<
1 t-1 no
Illi
1
z
w o IQ
150 ”
z
05 t-
o Wo
77
500 ”
z
o
1,500 ”
68
scale point where accuracy is greatest. Referring to Table 3, we
see that meter scale A (Fig. 501) is the scale on which the reading
should be made. When the probe is connected to the cathode pin
of the amplifier tube, the meter pointer should swing up to the
1.4 mark on scale A if our voltage is correct.
Suppose we wTish next to check the operation of the horizontal
oscillator stage. A reliable indication of whether the stage is oscil
lating properly may be had by checking the developed grid bias
voltage. In our particular receiver, this voltage should be approxi
mately —9. The function switch, therefore, should be set to
—volts. Table 3 shows that scale A is used for this measurement
also, but the range switch should be set to the 15 v position to
provide for a reading of 9 volts. (Remember that the 1.5 mark
now corresponds to 15, and all readings on this scale must be mul
tiplied by 10 to obtain a correct reading.) This same procedure
applies to other scales and ranges. If the range switch is set to
150 v, for example, and 130 volts are applied to the test probes, the
meter pointer will rest at 1.3 on the 1.5 scale. The scale reading
should be multiplied by 100, therefore, to obtain the correct value.
The use of multipliers on the 5 v scale is the same. If, for ex
ample, the range switch is set to 500 v and a B-|- voltage of 240
is measured, the meter pointer will rest at the first division mark
to the left of 2.5 on scale C (Fig. 501). A multiplier factor of 100
must be used to obtain a correct reading.
The scales marked lo-a.c. (shown as scales E, F, G and H in Fig.
501) are used only for a.c. voltage measurements. Scale G is used
only for a.c. voltages (r.m.s.) from 0.1 to 1.5; scale E only for
a.c. voltages (r.m.s.) between 0.1 and 5. On this v.t.v.m., scales
B, D, F and H are printed in red and are used only for peak-to-
peak a.c. voltage readings. These scales are used simultaneously
with the black scales immediately above them. For example, an
r.m.s. voltage of 1.4 is read on the black scale G; the corresponding
peak-to-peak voltage of approximately 3.9 is indicated simultan
eously on scale H. An r.m.s. voltage of 14 is read on scale A; the
corresponding peak-to-peak voltage of approximately 39 is read
on scale B.
This same procedure applies to direct current and resistance
measurements. The use of multipliers is simple if it is remembered
the full-scale marking of the scale in use corresponds to the num
ber indicated on the range switch. A chart similar to that in
Table 3 can be made up easily for other types of v.t.v.m.’s and
may be extremely helpful in acquainting the technician with a
new instrument.
M
How to take measurements
70
Ohms
Here is the procedure for making ohms measurements:
1. Connect the ohms probe and cable to the ohms jack on the
front panel. The ground cable should be connected to the gnd
jack.
2. Estimate the value of the resistance to be measured and set
the range switch to the setting which includes this value. (See
Table 3 for the correct range setting.)
3. Turn the selector switch to ohms. The meter pointer should
deflect fully to the right. This is the normal setting for the pointer
when the v.t.v.m. is set up for ohms measurements and no load is
connected across the ohms probes. If the pointer does not rest
exactly at the full-scale infinity mark, set the ohms adj control for
the proper setting.
4. Short the ohms probe and the ground cable together. The
meter pointer should swing over to the left-hand zero point (zero
resistance). Adjust the zero adj control, if necessary, to bring the
pointer into exact alignment with the zero mark. Unshort the
probes and recheck the infinity setting of the pointer. Reset the
ohms adj control, if necessary.
5. Connect the ground clip to one side of the unknown resis
tance and the ohms probe to the other side. If the resistance is
connected into a circuit, it may be necessary to disconnect one end
from the circuit to prevent shunt resistances from affecting the
meter reading. CAUTION: Never attempt to make resistance
measurements in circuits which have power applied. Failure to
observe this precaution may result in burnout of the precision
resistors in the ohms-measurement network of the v.t.v.m.
6. Read the resistance from the top scale of the v.t.v.m.
A.c. voltage
1. With the power cord connected to an a.c. outlet, turn the
selector control to the A.c. volts position.
2. All a.c. voltage measurements are made with the ground
cable and the direct probe and cable. If the switch probe is used,
the sliding switch should be set to the direct position. Connect
these probes and cables, if necessary.
3. Estimate the value of the a.c. voltage to be measured and
set the range switch to the position which includes this value.
Table 3 lists the correct range settings. (If transformerless or
a.c.-d.c. equipment is involved, refer to the listing of special pre
cautions under “Measurement Tips.”) Make sure the frequency
71
and waveshape of the a.c. signal to be measured are within the
measurement capabilities of the v.tv.m.
4. Short the ground cable and direct probe together and zero
the meter pointer with the zero adj control, if necessary. The
probe switch should be set to direct.
5. Connect the clip of the ground cable to the chassis or
ground bus of the equipment under test. Connect the direct probe
tip to the test point. Do not touch the metal part of the tip with
the fingers.
6. Read the r.m.s. value of the a.c. voltage from the black scale
(indicated in Table 3). Read the peak-to-peak value from the
red scale.
Direct-current measurements
Current measurements* require several precautions not usually
necessary for other measurements. These precautions are recom
mended to protect the meter from burnout and to protect the
technician from electrical shock. They may be summarized briefly
as follows:
A. Take special care in setting up the instrument and the test
circuit because the meter is not protected against burnout
on current measurements.
B. Do not attempt to measure current in circuits containing
more than 1,500 volts d.c. Do not attempt to measure alter
nating current.
C. Always remove power from equipment under test before
connecting test leads. Discharge all capacitors in the test
circuit.
Current measurements are made as follows:
1. Turn the selector switch to the current position. (It is not
necessary that the instrument be connected to the power line for
direct current measurements.)
2. Connect the “negative” current cable (black) to the —cur
rent jack. Connect the “positive” current cable (red) to one of
the three -(-current jacks. The jack used depends upon the esti
mated value of the current to be measured. To protect the meter
against possible damage, the range selected for the initial measure
ment should be able to handle a current of at least twice the value
of the estimated current. The range switch is connected in the
♦Most v.t.v.m.’s measure resistance and voltage only, and do not have provisions
for measuring current.
72
circuit only when currents of 500 ma or less are measured. When
the +1.5 amp and +15 amp jacks are used, the setting of the range
switch may be ignored.
3. Remove power from the equipment under test. Short all
filter capacitors to prevent accidental shock.
4. Break the circuit wiring at the desired measurement point
and connect the two current cables in series with the circuit. The
black cable should be connected to the negative lead or terminal
of the opened circuit or to the point of lower potential, and the
red cable should be connected to the positive lead or terminal.
5. Apply poiver to the equipment and observe the meter read
ing. If necessary, readjust the range control or change connection
of the red cable at the v.t.v.m. to give a reading on a scale which is
nearest to full-scale deflection of the meter pointer (see Table 3).
6. Remove power from equipment and short the filter capaci
tors before disconnecting the cables.
Null indications:
The test procedure used in making null measurements varies
with the application. Null indications in bridge circuits, for exam
ple, may require that probes and cables be connected differently
than in other applications, such as FM discriminator tests. For
correct connection of probes and cables, the technician should
refer to the manufacturer's service notes or equipment instruc
tions. In general, the procedure is as follows:
1. Connect the power cord to an a.c. outlet and set the selector
control to + volts.
2. Connect the ground cable to the gnd jack and connect the
direct probe and cable, with the d.c. direct probe attached, to the
volts a.c.-d.c. connector. Set the sliding switch on the probe to
the d.c. position.
3. Rotate the zero adj control to position the meter pointer
exactly at center scale, as indicated by scale Z in Fig. 501.
4. Estimate the maximum voltage swing in the null-indicating
circuit. The maximum swTing is measured by adding the maximum
negative voltage to the maximum positive voltage. For the correct
range setting, see Table 3.
5. Remove power from the circuit under test. The ground
cable and d.c. probe should be connected as recommended by the
manufacturer of the equipment under test. Usually, the probe tip
is connected to a point which is positive with respect to the point
to which the ground cable is connected. If the meter pointer
73
swings off scale in either direction, switch the range control to a
higher range which gives on-scale deflection. Make circuit adjust
ments as required to center the pointer on the center-scale zero
mark.
6. Remove power from the test circuit and disconnect the test
leads.
Measurement tips
Here are some practical suggestions that will help you get the
most out of your v.t.v.m.
Ohms measurements
Because accuracy is the first consideration in any measurement,
keep the following suggestions in mind when resistance measure
ments are made.
As pointed out previously, nearly all ohmmeters are designed
to provide most accurate readings at the center-scale point. This
factor should be remembered when measurements are taken. Most
manufacturers, while specifying accuracy percentages for the a.c.
and d.c. voltage functions of their instruments, do not specify an
accuracy figure for resistance measurements. While this practice
may seem questionable, remember that resistance measurements
of high accuracy are not required in general service work. The
tolerance of most resistors used in conventional TV and radio
receivers is in the order of ±10 and ±20%. If the receiver can
tolerate deviations of these magnitudes, it is not necessary to read
the values to a close tolerance. Resistance measurements at very
tight tolerances should be made with calibrated resistance bridges
designed especially for such applications.
The value of the battery voltage used in the ohms section of the
v.tv.m. should be known and remembered by the technician.
Some sensitive devices, such as low-voltage low-wattage tube fila
ments, meter multipliers and thermocouples may be burned out
if too high an ohms test voltage is applied.
When resistance measurements are made in electronic circuits,
the shunt resistance may be very difficult to estimate. In these
cases, it is usually best to unsolder individual components or to
break up a complicated circuit by disconnecting major leads and
buses and checking resistances of parts and simple circuits indi
vidually. Some manufacturers provide resistance measurement
tables in addition to voltage tables in their service notes. Usually,
however, the technician has to work without this assistance.
74
In some circuits, especially those which contain large value capa
citors, several seconds may be required for the meter pointer to
reach its maximum position. This slow pointer swing is usually
most apparent on the high-resistance scales and is more pro
nounced in test circuits which contain both a large capacitance
and high resistance. This effect, caused by the charging of the
capacitor by the battery voltage in the v.t.v.m., is due to the fact
that the current must flow through the v.t.v.m. resistance in series
with the battery. This resistance is highest on the highest ohms
range and the charging time of the test-circuit capacitance, there
fore, is increased. Charging time may often be greatly reduced by
setting the range switch to the lowest ohms setting, where the
series resistance is least, and then switching up to the desired range.
V.t.v.m. ohms circuits are designed to provide extremely long
battery life. Aside from a minute leakage current, no battery cur
rent is drawn when the ohms probes or leads are disconnected.
If the v.t.v.m. is used over an extended operating period, the ohms
test leads should not be left connected in a shorted position.
A weak ohms battery is usually indicated when the meter
pointer will not deflect to the infinity mark on the ohms scale.
Batteries in this condition should never be left in the instrument.
The chemical action in an exhausted battery produces a highly
corrosive residue which quickly attacks such critical metal parts
as the switch contacts and solder joints. A badly corroded switch
section may require complete replacement of the switch and asso
ciated components. Units which require flashlight type batteries
should use steel-encased batteries to reduce effects of corrosion.
Voltage measurements:
A few tips on safety are of first importance in taking both a.c.
and d.c. voltage measurements. It is always best to work with one
hand in a pocket when a hot circuit is being worked on. The
ground cable of the v.t.v.m. should be connected first. If the probe
is connected first and the technician picks up the ground clip, he
is in series with the v.t.v.m., the voltage source and ground.
Special care should be taken when working with a.c.-d.c. or
transformerless equipment because the chassis, ground or B—
buses are usually tied to one side of the power line. If the ground
cable of the v.t.v.m. is connected to any of these points, the metal
case of the instrument will be at the same potential as the common
point in the test circuit. In some applications, the v.t.v.m. case
will be hot to the full value of the line voltage.
A situation of this type is shown in simplified form in Fig. 502.
75
When the line cords are connected to the outlets, one chassis will
be hot. Because the chassis are connected to different sides of the
line, a dead short across the line would result if the chassis should
touch or if they should be tied together through the ground cir
cuit of the v.t.v.m. Such a condition is, obviously, an extreme
shock hazard as well as a possible source of damage to expensive
test equipment. This common danger can be eliminated by use
of an isolation transformer which provides isolation from the
power line for a.c.-d.c. receivers.
When working with equipment carrying high B-j- voltages, a
shorting lead or screwdriver should be used to discharge capaci
tors after power has been removed. Large electrolytic capacitors
may store a charge for several minutes after power is shut off. Mica
capacitors, used in TV high-voltage supplies, may store a charge
for many hours.
One of the most common d.c. voltage measurement problems
arises in high-resistance high-impedance circuits which are sensi
tive to resistive and capacitive loading. The high input resistance
of the v.t.v.m. helps greately in overcoming the d.c. loading effects.
The isolating resistor in the d.c. probe aids in reducing the capaci
tive loading. This probe should always be used for d.c. voltage
measurements.
Greatest accuracy on both a.c. and d.c. voltage measurements is
usually obtained in the upper third of each scale. Overall v.t.v.m.
accuracy depends primarily upon two factors: (1) the accuracy
rating of the meter movement used and (2) the accuracy of the
resistors used in the voltage divider networks of the v.t.v.m. In the
case of the RCA Master VoltOhmyst, for example, the meter
movement is designed to provide a reading at the full-scale points
of within ±2% of the true voltage. The precision carbon-film
resistors used in the voltage divider networks have a tolerance of
±1%. The maximum combined reading error possible, therefore,
is ±3%. If a high-voltage multiplying probe is used, the possible
error is increased by the percentage tolerance of the high-voltage
multiplier resistor. Since the error is cumulative, with a high-volt
age probe having a resistance tolerance of ±2%, a maximum error
of ±5% is possible.
When measurements are made on scale points below full scale,
“tracking” error may be introduced. While a meter may be de
signed to have a high degree of accuracy at the full-scale point,
the amount of error may increase as readings are made elsewhere
on the scale. This effect is most noticeable in the lower quarter of
76
the scale. Vacuum-tube and meter movement nonlinearity and
meter-bearing friction often make readings in very low sections of
the scale untrustworthy. The degree of tracking error is seldom
specified by the v.t.v.m. manufacturer.
Probably the greatest source of error in a.c. signal measurements
is due to application of the v.t.v.m. The a.c. loading characteristics
of these instruments often are not as good as could be desired,
although they are vastly better than the a.c. loading characteristics
of nonelectronic meters. In TV signal circuits, for example, it is
often desirable to measure the a.c. signal drive supplied to the
grid circuit of a stage. Signal circuits are usually characterized by
a resistance of 1 megohm or better and usually have relatively low
shunt capacitance. Input characteristics of most vacuum-tube volt
meters are in the order of 0.75 to 1 megohm with from 50- to
100-quf capacitance. The effect of this added load upon the test
Fig. 502. If both a.c.-d.c. chassis arc plugged into the power line, as
shown, a connecting link between the two chassis (or if the two are tied
together through the v.t.v.m.) would result in a short circuit.
signal depends upon the impedance of the test circuit, the fre
quency and type of a.c. waveshape and the voltage value of the
signal. A good check of the loading effect of the v.t.v.m. may be
had by setting up a TV receiver in good operating condition and
tuning in a picture and accompanying sound signal. The a.c.
probe can be applied to different stages in the receiver and the
effect on picture and sound performance noted.
The discussion in Chapter 2 of the a.c. signal rectifiers used in
commercial vacuum-tube voltmeters demonstrates the importance
of frequency and waveshape upon the reading obtained with the
v.t.v.m. If measurements are made in TV circuits where special
types of waveforms are encountered, the meter should be able to
respond to the peak-to-peak value of the a.c. signal. Waveforms
which are narrow in shape and have a low repetition rate usually
cannot be measured accurately. Attempts to measure a.c. signals
77
which have higher frequencies than those the v.t.v.m. was designed
to handle may also lead to erroneous readings. The instrument
manufacturer usually states the frequency limits of his v.t.v.m.
and often specifies the type of waveform the instrument will meas
ure.
Poorly shielded instruments and cables may often lead to some
weird effects, especially when the instrument is used in a strong a.c.
field. Hum, noise and stray-field pickup may cause the meter
pointer to deflect regardless of how probes are connected. In high-
frequency circuits, especially where the v.t.v.m. is used on a low-
voltage range, good shielding and grounding are essential. When
r.f. circuits are involved, the ground cable should be connected as
closely as possible to the voltage test point. Probes and cables
should be shielded throughout to prevent the test leads from pick
ing up stray energy and feeding it to the meter circuit.
Most units which can be used for current measurement do not
require connection to the power line. The meter, together with its
associated shunting and dropping resistances, is inserted directly in
series with the test circuit. Three precautions are necessary in
these current-measurement applications. They are: (1) Use a
range setting considerably higher than the estimated value of the
current to be measured to avoid the possibility of meter burnout.
(2) Take extreme care if the meter case is connected to the minus
side of the current circuit because the case will then be at the
same voltage potential as the test circuit. (3) Be careful to connect
the current cables to the points of correct polarity in the circuit.
Because the meter must be connected in series with the test circuit,
all power should be removed and capacitors discharged before the
meter is connected.
The considerations discussed here apply in general to v.t.v.m
applications. The next few chapters, which deal with specific
applications, will point out some additional tips and considera
tions which can be used to advantage.
78
H
tv troubleshooting
Tube troubles
After the technician has checked operation and symptoms, he
should have a good indication of what section of the receiver is
defective. Tube failure, the most common difficulty, may be
checked easily by substituting new tubes in the appropriate sock
ets of the suspected section. Tubes should be substituted one at a
time. If it is necessary to remove several tubes of the same type,
they should be marked in some way so they can be replaced in
their original sockets. This is often important in the tuner section
where interchanging of the oscillator tube with the mixer tube
may require resetting of the oscillator channel adjustments.
Series-string receivers, which are rapidly gaining in popularity,
present a special problem because a single tube with an open
heater can remove power from all tubes in the string. Finding the
open tube can be time-consuming. The v.t.v.m. is an excellent
instrument for checking heater continuity. The “ohms” section
of the v.t.v.m. is used and the ohms probe and ground cable are
connected to the heater pins of individual tubes.
Often the technician will discover a set which has repeated
failure of one particular tube. While it may be a temptation
simply to install another tube, the careful technician will check
further for trouble. Persistent failures can usually be traced to a
defective component. For example, a common cause of repeated
failure of amplifier tubes may be traced to a leaky coupling capaci
tor, which causes the tube to draw an excessive amount of current.
The change in value of this capacitor may not be immediately
apparent in receiver performance but the tube may be consider
ably overloaded as a result. Similarly, changed-value resistors may
cause tube failure. Operation of the receiver may or may not be
80
seriously affected. Troubles of this type may be discovered quickly
by means of resistance measurements.
Surprisingly, repeated tube failures may sometimes be traced to
improper receiver design, although the technician should be ex
tremely careful in his diagnosis and troubleshooting before he
reaches this conclusion. Excessive line voltage, a condition which
can be checked quickly with the v.t.v.m., can cause circuit voltages
to exceed their proper values. Faulty oscillator operation and com
ponent breakdown may result from too low or too high a line
voltage. In such cases, the remedy may be an adjustable line-volt
age transformer which can step up or step down the line voltage
applied to the receiver. In a few cases, it is possible to substitute
tubes having higher ratings, although the technician should be
cautious in making any changes not specifically recommended by
the receiver manufacturer.
If tube substitution fails to correct the difficulty, the chassis
must be removed for troubleshooting. Always disconnect the a.c.
power cord and short the high-voltage output terminal to the
chassis before pulling the set. Modern receivers may store a high-
voltage charge for several days. While this charge may not be
lethal, a shock may result in dropping of the chassis or cuts and
bruises. A flexible wire, 18 inches long, equipped with alligator
clips on both ends is a good shorting device. Always clip the wire
to the chassis before clipping it to the high-voltage terminal.
If the suspected component cannot be located immediately, a
voltage check should be made. Whenever possible, the chassis
should be operated with the picture tube, yoke and focus coil con
nected. Receivers which have the tube and deflecting components
mounted on the chassis usually present no special problem because
the complete chassis can be removed and placed on its side on the
workbench. Other sets, however, have the picture tube and neces
sary components mounted in the cabinet separate from the chassis.
On many of these sets, the leads connecting to the chassis are
equipped with plugs which permit quick disconnection. Because
these leads are usually short, the technician requires a set of exten
sion leads or a “harness” to permit complete operation of the
receiver with the chassis removed. Pin connections for different
kinescopes are similar. Connection for the yoke, focus coil, high-
voltage lead and speaker vary with different manufacturers, and a
separate harness is required for each make. Ready-made harnesses
for Du Mont, Philco, RCA and others can be purchased from most
parts distributors or made up by the technician.
The correct circuit operating voltages are usually shown on the
81
manufacturer’s schematic design. In some service notes, a separate
voltage chart is provided which shows the correct d.c. voltage at
each tube pin in the set. These voltages are usually obtained at a
working line voltage of 117. A variation of ±20% from the manu
facturer’s voltage specifications is acceptable. A v.t.v.m. is almost
always used in making these tests because of its high input resis
tance. Before measurements are made, the antenna should be dis
connected and a shorting wire connected across the tuner input
Terminals to prevent any incoming signal from making voltage
changes in the receiver. On some receivers, the manufacturer will
specify that certain controls, such as brightness, contrast or a.g.c.,
be turned to a certain position.
Most d.c. voltages are measured with respect to the receiver
chassis, which serves as a common ground, and when this is the
case the ground cable of the v.t.v.m. should be clipped to the
chassis. Bias voltages and other negative d.c. voltages are also
ordinarily measured with respect to ground; it is necessary only
to set the function switch to —volts to switch polarity on the
v.t.v.m. probes. CAUTION: Remember that the chassis of trans
formerless or a.c.-d.c. type receivers may be hot. In some receivers,
the chassis is tied directly to one side of the a.c. line. In other sets,
the chassis may be connected to one side of the line through a
resistor. If the v.t.v.m. ground cable is clipped to the chassis, the
case of the v.t.v.m. will also be hot and there is danger of a severe
shock or short circuit. It is always best to use an isolation trans
former on receivers of this type.
With receiver cables properly connected and if the test precau
tions just described are complied with, the technician is ready to
take voltage measurements. The B plus voltage at the output from
the power supply should be checked first. This voltage will range
from about 120 to approximately 300, depending upon the type of
power supply. If B plus is considerably lower than it should be,
rectifiers and filter circuits should be checked for faulty operation.
If B plus voltage is excessively high, a bleeder resistor may be open
or a defective component elsewhere in the receiver may have
reduced the load on the power supply and caused a corresponding
increase in the supply voltage available.
Because of the complexity of TV receiver circuitry, several
B plus voltage buses may be used. The highest voltages are used
in the sweep and audio circuits; the lowest voltages in the tuner
and i.f. amplifier sections. D.c. operating voltages throughout the
receiver will generally vary with the value of the B plus voltage.
If a serious defect exists, it can cause radical changes in plate,
82
screen, grid and cathode voltages on most tubes in the receiver.
Because of this, voltage measurements throughout a defective re
ceiver may show that operating voltages are out of tolerance in
several stages which appear to have no direct relationship wTith
each other. A problem set of this type is best handled by breaking
up the B plus buses and checking them one at a time.
After the B plus supply voltage is checked, the operating volt
ages in the suspect section or stage of the receiver should be meas
ured. In most defective receivers, these measurements should bring
the technician very close to the source of trouble. Low voltage in a
section or stage may indicate a leaky or shorted capacitor. Too high
a voltage can point to an open resistor or defective tube. There
are many possible causes, of course, but these are the most com
mon.
83
connection of the capacitor from other parts of the circuit. A leaky
capacitor requires disconnection from the circuit before a positive
indication may be had. Disconnection is usually necessary because
other capacitors and resistors in the same circuit provide shunt
resistance around the capacitor being measured. A positive check
can be made by disconnecting one or both ends and measuring the
resistance across the capacitor.
The shorted capacitor can be detected by voltage measurements
in the associated section of the receiver. In any plate, screen,
cathode or grid-return circuit that uses a bypass capacitor, there is
normally some voltage across the capacitor. If there is not or if the
voltage is considerably lower than normal, the capacitor should be
disconnected from the circuit and its resistance checked. Also, if
the circuit voltage returns to normal after the capacitor has been
disconnected, a short in the component is indicated. Fig. 601 is
useful in demonstrating the effect of a faulty capacitor in a typical
amplifier stage. If any of the bypass capacitors are shorted, the
voltage at the circuit end of the capacitor will be zero. A leaky
capacitor will cause the circuit voltage to be lower than normal.
The letters A, B, C and D are typical test points.
When a plate or screen circuit capacitor becomes shorted, it
usually causes excessive current flow through any resistor con
nected in series between it and the voltage source. Consequently,
the resistor may change value considerably or become open. There
fore, whenever a shorted capacitor is discovered, it is always wise
to check out adjacent components with the v.t.v.m.
A shorted or leaky coupling capacitor can be detected by d.c.
voltage measurements. For example, a leaky coupling capacitor
will usually result in an abnormal voltage on the grid of the stage
being fed. The voltage may be higher or lower than normal. In
these cases, a new capacitor should be substituted in place of the
old one.
84
In some cases, a shorted or leaky capacitor may be extremely
difficult to detect because the short or leak may not show up until
after the full circuit voltage is applied across it. For example, a
plate or screen bypass capacitor may show infinite resistance when
checked with a v.t.v.m. which applies only 1.5 or 3 volts on the
ohms test. But, the capacitor may break down quickly when 200
or 300 volts are applied. Defects of this type are best discovered by
monitoring the d.c. voltage across the suspected capacitor with the
v.t.v.m. When the short or leak develops, the voltage read on the
v.t.v.m. should drop noticeably.
No prescribed resistance limits have been established by which
85
rent or contact-potential current from developing a voltage at the
measurement point.
Open capacitors are best detected by temporarily connecting a
good capacitor across each of the suspected capacitors while the
receiver is in operation and noting the effect on receiver operation
(Fig. 603). The substituted capacitor should have the same value
1--------- if—
COUPLING CAP
86
use to the technician in estimating the general condition of var
ious kinds of capacitors. Charging time depends upon the size of
the capacitor being tested and the input resistance of the v.t.v.m.
on the resistance range in use. The data given under the heading
of ‘"Charging Time” is based on the use of a v.t.v.m. having an
input resistance of approximately 10 megohms on the R X 1 meg
ohm resistance measurement range.
Resistors
In addition to faulty tubes and capacitors, defective resistors are
common causes of faulty receiver operation. Fixed composition
resistors frequently change value because of excessive current loads
or aging. These factors usually cause the resistor to increase in
TAB LE 4
RESIS
CAPACI TANCE*
TANCE INSULA VOLTAGE (Meg CHARGING
(/if) TION RATING ohms) SCALE TIME* SCALE
.001 paper 600 >1,000 R X 1 Meg. % Scale Flick R X 1 Meg.
99
.002 mica 600 >1,000 99
% ” ” 99
.006 paper 600 1,000 % ” ”
9» 99 99
.01 600 1,000 % ” ”
99 99
.04 600 >1,000 3 secs.
9V 99
.05 600 >1,000 4 ”
99 99
0.1 600 1.000 8 ”
99 99 99
0.25 400 200 20 ”
99 99
1.0 600 500 60 ”
99 99
1.0 600 500 6 ” R X 100K
99 99
4.0 electrolytic 450 20 25 ”
99
4.0 450 20 5 ” R X 10K
99 99 99
10 450 40 99
10 ** 99
»9
20 99
150 35 20 ”
»9
20 99
25 50 20 ” 99
25 450 10 20 ”
99 99 99
40 50 8 99
30 ”
99 9t
100 50 15 60 ”
99
100 12 15 10 ” R X IK
99
1000 — 3 R X 100K 15 ” R X 100
♦The values given for resistance and charging time are only typical values for the
guidance of the experimenter. They should not be considered as limits or quality end points.
The symbol > in the resistance column means “‘more than.”
Fig. 605. The v.t.v.m. and a d.c. probe can be used to test a
power supply. The d.c. voltage should increase as you move
the probe back toward the filament of the rectifier.
on the v.t.v.m. If R3 should open up, the voltages at points B, C
and D should increase noticeably. Any change which would cause
a marked difference in operating voltages is very likely to be
accompanied by a definite change in receiver sound or picture per
formance. When operation performance is affected, the technician
can check the reading on the v.t.v.m. to ascertain whether the
trouble lies in the power supply section or elsewhere in the re
ceiver.
The v.t.v.m. can, of course, be used to monitor d.c. supply and
signal voltages in all sections of the receiver. The monitor method
of checking intermittent resistors as well as capacitors has the
advantage of allowing the technician to work on another job while
waiting for trouble to show up.
Series voltage-dropping resistors which become defective should
not be replaced until the technician is satisfied that no other defec
tive components have caused the trouble. In Fig. 601, for example,
a shorted screen or plate bypass capacitor might cause an excessive
amount of current to flow through the screen and plate dropping
resistors. If one of these is faulty, the capacitors should be checked
for shorts or leakage.
Vertical and horizontal sweep and sync difficulties may often be
88
traced to a defective resistor in the oscillator stages. In multivibra
tor stages, for example, a changed value resistor may result in
inability to sync the picture.
&9
Shorts and leakage between windings and the core, shell or
ground (chassis) are best discovered by resistance measurements.
The v.t.v.m. is especially useful in these applications because of its
ability to read very high resistances. In many cases of leakage, the
resistance may be several hundred megohms or more. The sus
pected component should be disconnected from the circuit and
readings taken from each winding to the case, core, chassis and also
to other windings to detect leakage. In cases where suspected leak
age cannot be detected with the v.t.v.m., a higher test voltage may
be used, as shown in Fig. 607, to break down the leakage resistance.
Where there is reason to suspect leakage or breakdown from the
winding to the core, shell or chassis, a check can also be made by
temporarily insulating the unit from the chassis and checking
operation.
With the exception of leakage across a winding, the test setup
shown in Fig. 607 may be used to detect leakage in many cases.
This setup is especially useful in cases where leakage shows up
only when high voltages are present. These leakages can be greater
than 1,000 megohms, which is the highest resistance which can be
read on the “ohms” scales of the v.t.v.m. The external voltage
source can be any d.c. voltage from 50 to 500 or more. The supply
is connected to one end of the resistance and d.c. voltage readings
are taken with the v.t.v.m. at points A and B. Resistance is com
puted by Ohm’s law as follows:
R; (volts at A - volts at B)
ri xs . -■ .1 - ■ ............ _■-
x volts at B
where RJN is the input resistance of the v.t.v.m. on d.c. voltage
measurements.
When a power transformer operates hotter than usual, the
trouble may be caused by an overload in the receiver or power
supply circuits or by shorted turns in the transformer itself. Be
cause leaky or shorted filter capacitors are a frequent source of
trouble, the transformer can be checked by first disconnecting the
rectifier and filter circuits to see if an overload in the B plus cir
cuits is responsible. If the transformer continues to run in an over
heated condition, the transformer itself is probably defective. One
side of the filament circuit for all tubes except the power rectifier
is usually grounded to the chassis. Shorts in the filament circuit of
the receiver are best checked by unsoldering the grounded trans
former lead and taking resistance measurements to chassis.
90
Power supply filter chokes, vertical deflection output trans
formers and audio output transformers can be checked by measur
ing the d.c. resistance of the windings and employing some of the
same tests just mentioned.
Troubleshooting tuners
TV tuners are simple devices in principle but can be very com
plicated in construction. Wafer-switch types of tuners make
troubleshooting especially difficult, particularly if any of the small
components are suspected. Very often improper tuner operation
may result from incorrect supply and a.g.c. voltages being fed from
other parts of the receiver. There are, however, a number of ways
of checking voltages which do not require that the tuner be dis
mantled.
Most tuners are equipped with a readily accessible terminal
strip to which the filament voltage, a.g.c. voltage and one or more
92
B plus leads are connected. When tuner trouble is encountered,
voltage measurements should first be made at these accessible
terminals. If the voltages are abnormally high or low, these leads
should be disconnected and resistance measurements taken from
the tuner terminal points to ground. Useful voltage and resistance
checks may also be made by removing the tubes from the tuner
and measuring the pin voltages or resistances to ground from the
top of the socket. Remember that with the tube removed, B volt
ages may read somewhat high and proper allowance should be
TABLE 5
V.T.V.M.
Key Letter To Read Setting Probe Type
in Fig, 609 (Volts)
A A.g.c. bias voltage -D.C. D.c. isolating probe
B D.c. plate voltage 4-D.C. n
C D.c. screen voltage 4-D.C.
D R.f. plate-signal voltage ±0.0.* R.f. or high freq, probe
if
E ■ R.f. grid-signal voltage ±D.C.*
F ; Cathode-bias voltage 4-D.C. D.c. isolating probe
♦Polarity of v.t.v.m. setting depends on the polarity of the output voltage
from the high frequency probe.
Intermittent troubles
Intermittent troubles in the sound and picture i.f. amplifier sec
tions may often be detected by employing the v.t.v.m. as a voltage
monitor, as shown in Fig. 610. When intermittent picture trouble
is-encountered in a split-sound type receiver, the v.t.v.m. should
be set up to measure the d.c. voltage developed by the picture
94
signal across the load resistor of the second detector. If the inter
mittent is located in the video amplifier, the picture will be
affected, but the d.c.-voltage read on the v.t.v.m. should remain
essentially the same. If the intermittent trouble is located in the
picture i.f. section, both the picture and the reading on the v.t.v.m.
will be affected. This same technique can be applied to the sound
i.f. amplifier and audio sections in either a split-sound or inter-
carrier receiver. The v.t.v.m. should be connected across the dis
criminator or ratio detector load. If the intermittent affects the
sound from the loudspeaker and the voltage reading on the v.t.v.m.
Trap alignment
Alignment of the r.f. traps in the TV receiver is usually accom
plished before bandpass alignment is attempted. The purpose of a
trap is to act as a block to a signal of a specific frequency and keep
the signal out of a particular circuit. Most tuners utilize one or
more traps whose arrangement depends on the tuner design. Some
15
VOLTAGE OUTPUT
AT PIX DETECTOR
Fig. 612. Response curve of an i.f. amplifier using the technique shown in
Fig. 611.
of the most important traps are found in the i.f. and video sections
of the receiver. The older, split-sound type of receiver used sev
eral traps to attenuate the adjacent sound and picture carriers and
to block out accompanying sound from the picture i.f. channel.
Some split-sound sets also employ a 4.5-mc trap in the video section
to keep beat interference between accompanying sound and pic
ture carriers from affecting the reproduced picture.
The adjacent-channel sound and picture carrier traps and ac
companying sound traps are located in the picture i.f. amplifier
section. The frequencies to which these traps are tuned depends
upon the picture and sound intermediate frequencies used in the
receiver. These differ between different makes of sets.
The v.t.v.m. is the best instrument, when used in conjunction
with a TV signal generator, for indicating alignment. The actual
97
procedure is relatively simple and consists of feeding a signal of
the trap frequency into the tuner mixer stage and adjusting the
trap with an insulated screwdriver. The v.t.v.m. is set up to
measure the d.c. output voltage and is connected across the load
resistor of the second detector. When the resonant trap is tuned to
the exact frequency of the r.f. signal from the generator, the d.c.
output voltage will be minimum. An external bias voltage should
be used to override the normal a.g.c. voltage. Take care to prevent
too much signal from being injected into the mixer. The location
of the various traps in a typical split-sound receiver is shown in
Fig. 613.
TV receivers which use the intercarrier-sound system do not
employ the same traps as those used in the split-sound receiver.
Except for trap circuits in the tuner, the intercarrier set employs
only a 4.5-mc sound trap in the video amplifier. This trap must be
adjusted carefully, especially on sets having wide bandpass, to
reduce the possibility of 4.5-mc beat interference in the video
channel. On some intercarrier receivers, the sound i.f. takeoff point
is located ahead of the last picture i.f. amplifier. In these sets, the
last picture i.f. amplifier may contain a trap which is tuned to the
receiver sound intermediate frequency. The test signal for 4.5-mc
video traps is fed into the first video stage. The test signal for the
sound trap in the picture i.f. amplifier is fed into the grid circuit
of the mixer stage.
98
not be overlooked. They may be checked by voltage and resistance
measurements.
Fig. 614. Block diagram of the various sweep circuits used in a TV receiver.
These circuits can be checked through with a v.t.v.m. by making voltage
and resistance measurements. A.c. voltages are best measured with a peak-
to-peak reading v.t.v.m.
and horizontal sync circuits are among the most critical sections of
the receiver primarily because of their relative complexity and the
values and tolerances of the resistors and capacitors employed.
The integrating and differentiating networks, especially, can be a
perplexing source of trouble when components become defective.
Fortunately, trouble in sync stages is usually easy to diagnose
because of the effect of the sync action of the picture. Trouble in
any stage between the video amplifier and the second sync ampli
fier, if used, will probably result in loss of both vertical and hori
zontal sync. Trouble in the integrating network will affect vertical
sync; trouble in the differentiating network will have its greatest
effect on horizontal sync.
Troubles of this type are best attacked with a v.t.v.m. by use of
voltage measurements first and then resistance measurements.
Changes in resistance values or a leaky, changed-value or open
capacitor in any of the sync circuits can cause erratic sync action
or complete loss of sync. Because of the complexity of these cir
cuits, it is usually best to disconnect one end of the suspected part
100
from the circuit before taking resistance measurements. Capacitors
are best checked for opens by substituting a new capacitor.
Locating a defective capacitor in a sync circuit is usually a time
consuming job because of the large number of capacitors used.
Many experienced service technicians have found it more expedi
ent to clip out and discard several capacitors at once, replacing
.them with new units of the exact required values. The labor saved
is often worth several times the cost of the paper capacitors.
An oscillator stage may be checked for proper oscillation by
measuring the developed d.c. grid voltage. The voltage is negative
Fig. 615. Typical high-voltage circuit for a TV receiver. Various test points are
shown.
and may range from a few volts to 80 or 90. The voltage is usually
specified in the manufacturer’s sendee notes. If the developed volt
age is lower than normal, look for trouble in the oscillator stage.
Vertical deflection output stages, particularly those used in early
model receivers, may employ a “boosted” B plus voltage taken
from the horizontal output deflection stage. Any trouble in the
horizontal section which lowers the boosted B plus supplied to the
vertical output stage will probably affect the vertical deflection.
The horizontal deflection and high-voltage output stage requires
some special considerations when it comes to taking voltage meas
urements. A representative horizontal circuit is shown in Fig. 615.
Some of the common symptoms of trouble in this section are:
1. No raster or dim raster
2. Insufficient width
3. Insufficient height
4. Soft picture detail
The reader can see that these symptoms are related directly to
101
picture brightness (a high-voltage function) and to raster dimen
sions (scanning function). Failure of VI and V2 usually results
in no raster because no high voltage is available. When this
trouble is encountered, these tubes should be checked. Damper
tube failure (V3) may affect brightness and raster size. Sweep may
also be affected by VI and V2. Keystoning of the raster indicates
a defective yoke.
In cases where new tubes fail to clear up the difficulty, the B plus
voltage should be checked first to make sure that power supply
troubles are not responsible. D.c. voltage measurements can then
be made at points A, B, C, D, E, F and G. The voltage is usually
negative at the grid of the horizontal output tube, VI.
A.c. voltage measurements should not be attempted at points X,
Y and Z because of the large impedances or high a.c. or d.c. voltages
in these circuits. The high d.c. output voltage at point H can be
measured with the v.t.v.m. if a high-voltage probe is used.
Design of these circuits will vary with different receiver models
but the circuits are basically the same. Troubleshooting of FM
sound detectors and the audio amplifier sections is described in
Chapter 7.
102
am-fm alignment and repair
Am broadcast receivers
Radio receiver troubles are usually simple to diagnose and the
average repair can be made quickly. This is especially true with
the popular table-model a.c.-d.c. radio which uses readily available
tubes and standard components. Also, these receivers utilize well
established circuits which are simple in design. From a time
money standpoint alone, the radio repair can be as profitable as
the TV repair job.
Radio receiver troubles can be classified in this way:
1. Tube failures
2. Component failures
3. Intermittents
4. Misalignment
103
Tube failures
Tube failures, the most common source of trouble, can be dis
covered by different methods, depending upon the type of re
ceiver involved. A defective tube in the series-string type receiver
is usually found by tube substitution or with a tube tester. These
two methods are ordinarily employed when the chassis is left in
stalled in the cabinet. Either of these methods will often reveal a
tube which has an open heater or other defect. This approach has
the disadvantage, however, of requiring that each tube be removed
35W4 50C5 I2BA6 I2BE6 12AV6
«CT woo output ffAim cow oetavc audio amh.
Fig. 701. The v.t.v.m. can be used to check the heater wiring of filament
circuits.
and tested or else replaced by a good tube. These methods can be
time-consuming.
The v.t.v.m. provides a quicker means of finding an “open” tube
because the a.c.-d.c. receiver uses a series hookup for the tube
heaters. Failure of one tube heater will remove power from all
tubes in the string. The defective tube can be located quickly with
the v.t.v.m. if the chassis is removed from the cabinet so that volt
age measurements can be made. Fig. 701 shows the heater wiring
arrangement of a typical a.c.-d.c. radio receiver employing five
tubes. Suppose the heater in the fourth tube in the string, the
12BE6, should open. With the receiver chassis removed, the
v.t.v.m. can be used to locate the “open” tube quickly. Set up the
v.t.v.m. for a.c. voltage measurements. Connect the ground clip to
the minus B common bus wiring of the receiver. This can usually
be located at the power switch terminals. With the receiver power
switch turned on, voltage readings can be taken from points A
through E. Because of the exceedingly small current drawn by the
v.t.v.m„ the voltage drop through the tube heaters will be negli
gible and the v.t.v.m. should read the approximate value of die
104
line voltage at all points except E. At E, the reading will be zero
because of the open heater. If voltage is also read at E, it may be
concluded that the 12AV6 is defective.
Tube defects other than open heaters will usually cause a change
in the normal operating voltages at the tube pins. The d.c. bias
voltage developed at the cathode or grid pins is a good indication
of how the stage is functioning. Plate and screen voltages, which
are supplied by the B plus power supply section, will often1 in
crease or decrease considerably when circuit or tube troubles arise.
The v.t.v.m. may be used to check for a gassy tube which might
otherwise pass with a conventional tube tester. The presence of
Fig. 702. This test is effective for locating gassy tubes and
leaky coupling capacitors.
105
O
RFSIGGEH
Fig. 703. Simplified circuit arrangement of an AM broadcast receiver showing the various check points. The high-imped
ance feature of the v.t.v.m. permits tests to be made without disturbing circuit operation. Note that three different probes
are needed if a complete examination of the set is to be made. With this setup you can trace a signal completely through
from the front end to the speaker, quickly locate the point at which the signal disappears. At the same time this method
of testing helps you determine the effectiveness of operation of each particular stage.
a resistor of the proper value across the filament terminals of the
socket.
Shorts between tube elements can be detected by resistance
measurements with the v.t.v.m. Heater-to-cathode shorts and grid-
to-cathode shorts are the most common types of internal shorts
encountered. Remember, however, that these shorts may have a
very high resistance, often in the order of several hundred meg
ohms. Also, the shorts may not show up when the tube is cold. It
is sometimes helpful to connect the v.t.v.m. leads to the tube pins
and to tap the tube while watching the meter reading.
Component failures
Failure of capacitors, resistors and other circuit elements are
best discovered by first isolating the defective section or stage in
the receiver. When the trouble has thus been narrowed down, the
component tests discussed in Chapter 6 can be employed to ascer
tain the condition of individual components.
Because these component checks cannot be made until the faulty
section or stage is first located, it is important that the techniques
for isolating the defective stage be understood. There are several
ways of doing this, the correct way depending upon the type of
receiver trouble.
Signal tracing
Signal tracing with the v.t.v.m. provides a quick means of locat
ing a defective stage when the cause of trouble is not readily
apparent. An r.f. signal generator is required to feed a signal into
the antenna terminals of the receiver, as shown in Fig. 703. The
generator should be able to deliver approximately 0.1-volt output
and should have an adjustable output attenuator to permit adjust
ment of the output signal to the desired level. If the receiver uses
a loop antenna, it is usually sufficient to bring the “hot” clip of the
generator cable close to the loop, or to fasten it to the frame of the
loop.
This is the procedure for tracing the signal through the receiver:
1. Disable the a.v.c. circuit by disconnecting the a.v.c. bus at
the second detector stage or by applying an external bias. Since the
control grids of one or more tubes get their ground return through
the a.v.c. bus, disconnection of this line can leave such grids float
ing. If you do not use an external bias supply you should discon
nect the bus and ground it. Simply grounding the bus without
107
checking the a.v.c. circuit first will often result in shorting the
audio signal since in many circuits the volume control is made part
of the a.v.c. line. The word ground means either chassis or B
minus, depending upon the circuit arrangement of the receiver.
2. Connect the r.f. output cable from the generator to the
antenna terminal of the receiver. Connect a ground lead between
the receiver and the generator.
3. Turn the generator modulation switch on. Tune the gen
erator and the receiver to the same frequency in the broadcast
band. Set the receiver volume control for maximum volume.
4. Set up the v.t.v.m. for r.f. voltage measurements on its low
est range. In all measurements for the r.f. probe, always connect its
ground clip as near as possible to the test point. Connect the r.f.
probe to point 1 and adjust the receiver tuning for maximum
reading on the v.t.v.m. Use the high output from the generator
and adjust the generator output attenuator as required to obtain
a suitable meter deflection.
5. Move the r.f. probe to point 2. The signal voltage should be
greater at this point if the r.f. amplifier is functioning correctly.
6. Place the r.f. probe at point 3. Put the ground clip of the
probe to a ground point near the converter-tube socket. No signal
at this point indicates trouble in the transformer or in the con
verter input circuit.
7. Transfer the probe to point 4. Retune the receiver slightly
to obtain maximum reading on the meter. The ratio between the
voltages at this point and at point 3 is the approximate conversion
gain of the stage. The signal at point 4 will be the receiver inter
mediate frequency. No signal at this point indicates trouble in the
oscillator section of the converter.
8. Check the oscillator section by measuring the oscillator d.c.
bias voltage at point 5. Use a d.c. probe at this point.
9. Move the r.f. probe to point 6 and move the ground clip to
a new point near the socket of the first i.f, amplifier stage. If the
first i.f. transformer and the grid circuit of the stage are function
ing properly, a meter indication approximately equal to that ob
tained at point 4 will be read.
10. Transfer the r.f. probe to the plate of the i.f. amplifier
(point 7). The meter reading should increase, indicating ampli
fication in the stage.
II. Check performance at the grid (point 8) and plate (point
9) of the second i.f. amplifier. A considerable increase in gain
should be noted between these two points. Check r.f. signal at the
108
diode plate of the second detector to make sure the secondary of
the last i.f. transformer is not open or shorted.
12. Remove the r.f. probe and set up the v.t.v.m. for a.c. volt
age measurements. Connect the ground cable to the receiver
chassis. Connect the probe to the grid of the first audio amplifier
(point 11). Read the output directly from the a.c. voltage scales
on the v.t.v.m.
13. Check the audio signal at point 12 and on the output side
of the coupling capacitor (point 13). Move the probe to point 14.
A considerable increase in signal should be noted.
14. Measure the audio output signal from the receiver by con
necting the probe to the speaker voice-coil lead at point 15. Con
nection of the probe to the speaker voice coil will not give a volt
age reading unless one side of the voice coil is grounded. First,
run a continuity test between voice coil and chassis, using your
v.t.v.m. for this. If the voice coil is “floating,” that is, if one side is
not grounded, you can make a voice-coil voltage measurement in
either of two ways. You can put the probe on one of the voice-coil
terminals and the v.t.v.m. ground clip on the other, or else you
can ground one side of the voice coil (either side).
Signal tracing with the v.t.v.m. can be a valuable trouble shoot
ing method with proper experience. It has the advantage of pro
viding the technician with an overall picture of receiver perform
ance. Other methods of locating a defective stage, however, can be
used to find trouble quickly. After you are satisfied that tubes are
not responsible, start checking performance of the receiver from
the audio section back toward the r.f. stage. A quick check of the
audio section consists simply of touching the grid pin of the first
audio amplifier with a screwdriver. If the volume control is turned
up all the way and if the audio section is functioning properly, a
loud hum should be heard. It may be helpful to touch the metal
shaft of the screwdriver with a finger. The hum is introduced by
stray pickup into the grid circuit through the screwdriver. A rough
check of performance in the rest of the receiver may be had by
scraping the grid pins of the converter and r.f. stages with a screw
driver. In a properly operating receiver, this action should produce
noise from the loudspeaker.
Keep in mind that there is some hazard involved in checking a
direct line connected receiver without an isolation transformer.
In Fig. 703, for example, if the signal generator happened to be
grounded and the receiver power plug happened to be inserted so
that the high side of the line was connected to B minus, the test
leads might melt due to the direct short across the line that would
109
exist. A properly fused power line would prevent such an accident.
Although success of these rough checks depends somewhat on
Voltage measurements
Troubleshooting by measurement of circuit operating voltages
is probably the most used method of locating a defective stage or
section. The schematic diagram in Fig. 704 shows the signal-han
dling section of a six-tube receiver together with the test points at
which operating voltages can be measured. Table 6 describes the
TABLE 6
V.T.V.M.
Key Letter To Measure Setting Probe Type
in Fig. 704 (Volts)
A Cathode-bias voltage +D.C. D.c. isolating probe
B D.c. plate voltage +D.C.
C D.c. screen voltage +D.C. ft
Hum
Excessive a.c. hum is a common complaint in a.c.-d.c. receivers.
When hum is encountered, first determine whether it is coming
from the power supply or from the signal-handling section of the
receiver. If the hum originates in any section of the receiver ahead
of the volume control, it should change in level as the control is
112
rotated. If the hum is not tunable, use the v.t.v.m. to check the
ripple voltage at the output of the power supply. The instrument
should be set up for a.c. voltage measurements and the direct
probe connected to point A. As little as 2 or 3 volts of a.c. meas
ured here may cause considerable hum. If excessive ripple voltage
is found, the filter capacitors may be leaking or heater-to-cathode
leakage may exist in the rectifier tube. Capacitor leakage may be
detected by measuring the resistance of the capacitors when they
are disconnected from the circuit. Discharge the filter capacitor
before making resistance measurements. Heater-to-cathode leak
age in the rectifier tube as high as several hundred megohms can
cause considerable hum. If these checks do not indicate trouble in
Intermittents
Occasionally a tube heater will open up intermittently, usually
when the line voltage fluctuates. These intermittents may be very
difficult to detect by tube substitution. The v.t.v.m. can simplify
the repair, however, by using the instrument to monitor the volt
age drop across the heater of the suspected tube. For example, in
Fig. 701, if the I2BA6 is suspected, connect the v.t.v.m. directly
across the heater pins. Under normal operating conditions, ap
proximately 12.6 volts will be read. If the heater opens, however,
the v.t.v.m. will read almost the full value of the line voltage. If
the intermittent reduces the reading to zero, the v.t.v.m. should be
connected across other tubes until the faulty tube is discovered.
An adjustable-voltage transformer, which permits application of a
line voltage as high as 130, is often helpful in breaking down the
intermittent.
Other intermittent troubles may be localized by using the
v.t.v.m. to monitor signal and operating voltages in different sec
tions of the receiver. For example, if sound is intermittent, the
B plus voltage from the power supply can be monitored by using
the d.c. measurement feature of the v.t.v.m. If the signal stops and
the B plus changes, the trouble may be due to an open plate,
screen, or cathode circuit. The lower current drain will then cause
the voltage at point 4 to increase. The signal also may stop due to
a shorted screen or shorted cathode resistor. If B plus does not
change when the intermittent shows up, the fault is in the signal
handling section of the receiver. The v.t.v.m. should then be set
up for a.c. voltage measurements and connected to the grid pin
of the first audio amplifier stage (point 2). If the trouble is be
tween the r.f. stage and the first audio amplifier, the a.c. signal
114
voltage measured on the v.t.v.m. will change when the intermit-
tent trouble develops. If the reading is unchanged, look for
trouble between the v.t.v.m. test point and the loudspeaker. The
trouble source can be further narrowed down by moving the a.c.
probe to point 3, the grid of the audio output stage. See Fig. 709.
The r.f. probe can be used to monitor the signal at point 1 or
elsewhere in the r.f. section of the receiver. Because of the very low
signal level in this section of the receiver, however, it may be
necessary to employ an r.f. signal generator to obtain a signal of
suitable strength.
115
Fig. 710. Setup for the alignment of an AM broadcast receiver. The output signal of the r.f. generator is connected to the
input of the r.f. amplifier. If the receiver does not have such a stage, then the generator xwltage is fed into the signal grid
of the converter. The v.t.v.m. is set to read low a.c. volts and is connected across the voice coil, as shown. If one side of the
voice coil is not grounded, then it should be connected to the ground clip of the v.t.v.m. Note that we do not show the
use of an a.c. probe (rectifying probe). Whether such a probe is needed or not at this point depends entirely upon the
design of the v.t.v.m. Some v.t.v.m/s use rectifying probes for all a.c. measurements, while others use rectifying probes
simply to extend the frequency range of the instrument.
a few minutes by “touching up” the i.f. transformers and trimmers
in the oscillator and r.f. stages.
The alignment adjustments provided in a typical AM receiver
are shown in the simplified schematic of Fig. 710. This receiver
employs an r.f. amplifier, converter stage (combined oscillator and
mixer functions) and two i.f. amplifier stages. Complete r.f. and
i.f. alignment requires adjustment of trimmers in all four of these
stages. In addition, an antenna trimmer capacitor (Cl) is located
in the r.f. amplifier input section.
While receivers vary considerably in the design and the number
of alignment adjustments provided, the procedure and recommen
dations of the alignment technique which follows can be used
generally on any AM radio receiver. The v.t.v.m. and an r.f. signal
generator are the only test instruments required.
117
5. Set the generator dial to the intermediate frequency of the
receiver /'usually 455 kc). If the receiver is not completely mis
aligned, a small voltage will be read on the v.t.v.m.
6. Starting with the secondary of the last i.f. transformer and
using an insulated screwdriver or alignment tool, adjust the slug
in Lil (Fig. 710) for maximum voltage reading on the v.t.v.m.
If the i.f. amplifier stage is operating correctly, it should be pos
sible to tune the i.f. slug through a peak on the v.t.v.m. (The i.f.
alignment adjustments shown in Fig. 710 are adjustable slugs
which vary the inductance of the i.f. coils. Some receivers employ
small mica trimmer capacitors instead of slugs but the alignment
procedure is the same.)
7. Working toward the converter stage, similarly adjust the
slugs for LIO, L9, L8, L7 and L6 for maximum voltage output
across the voice coil. Keep the injected signal from the generator
ds low as possible while still obtaining a voltage reading on the
v.t.v.m. This precaution will prevent overloading of the i.f. ampli
fiers. The i.f. amplifiers are now aligned.
8. Remove the short -from the oscillator tuning capacitor (C6)
and move the r.f. output cable to the antenna terminal (point B).
(It is sometimes advisable to couple the signal loosely by laying
the cable near the antenna circuit.) On receivers employing built-
in antennas, the chassis, if possible, should be installed in the cab
inet before r.f. adjustments are made. Tune the signal generator
and the receiver to 1,400 kc.
9. Adjust oscillator trimmer C7 for maximum reading on the
v.t.v.m.
10. Retune the generator and the receiver to 1,400 kc and
adjust r.f. amplifier trimmer C3 and antenna trimmer Cl for maxi
mum voltage output.
11. Re tune the generator and the receiver to 600 kc. Adjust
the oscillator tuning slug in L5 for maximum output on the
v.t.v.m. Rock the tuning capacitor back and forth while making
this adjustment.
12. Repeat steps 8, 9, 10 and 11, if necessary, to get precise
alignment. If this procedure is followed, the receiver should track
well throughout the broadcast band.
Receiver manufacturers make different provisions for alignment
adjustments. Besides conventional trimmer capacitor and coil
slugs, “cut” and “slotted” rotor plates are used and the wire turns
of built-in antennas may be bent to accomplish the desired tuning
condition. The “cut” rotor, used in the oscillator section, consists
of plates having a smaller size than those used in the r.f. and an
118
tenna sections. The “slotted” rotor contains one or two plates with
slots cut into the metal to permit bending the plates to obtain the
desired tracking action at the low-frequency end of the dial.
FM receivers
Troubleshooting FM receivers with the v.t.v.m. utilizes the
techniques used in troubleshooting AM broadcast receivers. Ex
cept for the operating frequencies and the types of second detect
ors involved, the FM set is similar to the AM broadcast receiver.
Block diagrams of two superheterodyne FM and one AM receiver
are shown in Fig. 711. It is evident that these three superhets are
Fig. 712. Simplified circuit diagram of FM receiver using two limiter stages and discriminator-type detector. This arrangement is
gradually being replaced by FM sets using ratio detectors. Such receivers do not ordinarily have limiter circuits. Various test points
are shown in the FM circuit above. The audio stages have been eliminated since these are checked in the same manner for all
receivers.
Discriminator detectors
The test setup for alignment of FM receivers having a discrim
inator detector is shown in the simplified schematic in Fig. 712.
Many receivers will employ an r.f. amplifier stage ahead of the con
verter; others only a single limiter stage before the detector.
1. Tune the r.f. signal generator to the receiver intermediate
frequency (10.7 me) and connect the output cable to the shield of
the converter tube (point A). Lift the shield from the chassis to
prevent shorting the signal to ground. Set the generator output
control to a low level.
X
Fig. 713. Ratio-detector circuit.
121
to center the pointer. Connect the d.c. probe to point D in the
discriminator.
8. Carefully tune the secondary of the discriminator trans
former for an exact zero-center reading on the v.t.v.m. (The exact
setting of this trimmer may be critical. Turn the alignment tool
slowly.) Because the secondary winding is above ground or “float
ing,” the circuit may be very sensitive to hand capacitance. Use a
long-handled, nonmetallic alignment tool.
Ratio detectors
Receivers which employ ratio detectors do not usually employ a
limiter stage preceding the detector, although in some receivers
the last i.f. stage provides limiting to a small extent. The i.f. ampli
fier and ratio detector stages are aligned by observing voltage
changes only in the detector stage.
A balanced ratio detector circuit is shown in Fig. 713. Align
ment of this type receiver is as follows:
I. Feed the i.f. signal from the generator into the mixer (or
converter) as described in step 1 in the preceding section.
2. Set up the v.t.v.m. for negative d.c. voltage measurements
and connect the d.c. probe to point A.
3. Align the i.f. transformers as described earlier.
4. Tune the primary of the ratio detector transformer for
maximum voltage reading on the v.t.v.m. at point A. Disconnect
the d.c. probe.
5. Rotate the zero-adjust control on the v.t.v.m. to position the
meter pointer at the center-scale zero. (On some vacuum-tube volt
meters, it may be necessary to set the function switch to -J-volts
to center the pointer.) Connect the d.c. probe to point B.
6. Adjust the secondary of the ratio detector transformer for a
zero-center reading.
The unbalanced type of ratio detector is aligned with the same
procedure except for step 5. The v.t.v.m. is left at —volts and
the secondary of the discriminator transformer is adjusted for a
minimum d.c. voltage reading. This procedure is recommended
because unbalanced detectors usually cannot be adjusted for an
exact zero-center reading. For the unbalanced type, two resistors
of about 56,000 ohms, (use 1% resistors) are connected across the
single ratio detector load resistor in series, during adjustment. A
center tap is thus available for alignment.
122
chapter
125
If the v.t.v.m. readings at any point indicate that gain is reduced
or that the signal is lost, look for trouble between that point and
the last test point at which a signal was measured. D.c. operating
voltages should be checked first followed by resistance readings of
suspected components.
127
The v.t.v.m. leads should be connected so that the “negative”
current lead is in series with the ground side of the circuit. The
“positive” current lead connects to the section of the circuit near
est the cathode. Because this adjustment is for static tube current,
no input signal is required and it is usually advisable to short the
input terminals. Allow about ten minutes’ warmup time and then
adjust RI4 for approximately 125-ma current reading on the
v.t.v.m.
The individual tube currents must be set next by means of
potentiometer RI 3 to insure approximate d.c. balance between
the two tubes. Each tube should handle half the current load, or
62.5 ma. Readings in both cathode circuits should be obtained.
Do this by inserting the v.t.v.m. in series between the cathode of
V5 and the cathode resistor. Repeat this operation with V6.
With the tubes adjusted for equal currents, the rest of the ampli
fier can be checked for proper operation. The v.t.v.m. is used to
check individual stages for proper current drain, static operating
and dynamic (signal) voltages. In Fig. 802, d.c. voltages are indi
cated and the correct signal voltages obtained at the 15-watt power
level are underlined.
The current and d.c. voltage readings should be made with no
signal applied. For current readings, circuit leads will have to be
unsoldered to permit connection of the v.t.v.m. D.c. voltage read
ings are made with the d.c. probe.
The a.c. measurement section of the v.t.v.m. is used to check
signal voltages. A 400-c.p.s. signal from an audio oscillator should
be fed into the input stage and the direct probe used at the under
lined signal points. Perfect dynamic balance between the two
halves of push-pull stages is not usually obtainable with fixed com
ponents because of the slight differences between tubes and com
ponents. Consequently, the driving signals at the grids of the out
put tubes may be different. In the circuit of Fig. 802, good dy
namic balance may be obtained by adjustment of potentiometer
RI 2. The v.Lv.m. is set up for a.c. measurements on a low a.c.
range. The ground cable should be connected to the amplifier
chassis and the a.c. probe to the center tap of the output trans
former primary winding. Potentiometer RI2 should be adjusted
for minimum a.c. voltage reading on the v.t.v.m.
Eliminating hum
Hum is a “bug” which shows up in many amplifiers and fre
quently requires considerable experimenting to eliminate. It is
128
usually the result of insufficient or faulty power supply filtering or
is caused by faulty tubes or components or by improper lead dress
and parts placement. Power supply ripple (which can produce
hum in the output) can be measured directly with the v.t.v.m. If
excessive ripple is not present, the trouble usually lies in another
section of the amplifier.
When troubleshooting hum, set the v.t.v.m. for a.c. voltage
measurements on a low range and connect it across the speaker
voice coil. Operate the amplifier at maximum volume setting with
no signal applied. Turn the bass control to maximum and the
treble control to minimum. Because the v.t.v.m. is much more
sensitive to changes in hum level than is the human ear, the meter
reading should be noted whenever a corrective measure is made in
the amplifier.
Some typical causes of internal hum and their suggested reme
dies are listed next. Usually two or more factors are responsible.
Whenever a circuit change reduces the hum reading on the
v.t.v.m., the change should be incorporated in the amplifier and
further experimenting attempted until an acceptable hum level is
obtained. It is not always possible to eliminate completely a hum
voltage at the output and the experimenter will have to set his own
limits of acceptance. Here are some causes and remedies:
1. Heater-to-cathode leakage in power output stage which re
sults from the difference in heater-to-cathode voltage. This situa
tion may sometimes be corrected by applying a small voltage to the
tube heaters by means of a balancing potentiometer across the
heater winding of the power transformer. The center up or arm
of the potentiometer is connected to the cathodes of the output
tubes, as shown by the 500-ohm potentiometer in the power supply
of Fig. 801. When making this test, short the input jack and put
the v.t.v.m. across the speaker voice coil. Set the amplifier gain
control to maximum. Start with a high a.c. setting on the v.t.v.m.
and reduce the range setting as you adjust the potentiometer.
2. Excessive heater-to-cathode leakage in other tubes in the
amplifier can sometimes be reduced by changing tubes. In some
cases, where amplifier gain is high, it may be necessary to try more
than one new tube to obuin the desired reduction.
3. Inadequate power-supply filtering can cause hum. Addi
tional ripple filtering or a decoupling network may be required in
the B+ leads going to the preamplifier suge.
4. Poor lead dress can result in hum, especially in a low-level
high-gain sUge which has a sensitive lead placed in the power
129
field. Try moving grid and plate leads around while watching
the effect on hum level being measured by the v.t.v.m.
5. Inadequate shielding in high-gain stages. Make sure that
control-grid leads and switching circuits are shielded and ground
ed. Use grounded shields on high-gain amplifier tubes. Do not
assume that all metal tubes are properly grounded just because
they are metal. When in doubt, scrape the top surface of the tube
to expose the metal, and then make a connection between tube
and chassis. If hum goes down put a shield around the tube or
replace it. If the amplifier chassis has no bottom plate, try using
one.
6. Improper grounding. Try changing ground points of by
pass capacitors while watching the meter reading. If poor ground
ing is source of trouble, run a ground bus which is tied to chassis
only at input end of the amplifier.
7. Poor parts placement. Reorient power transformer, filter
chokes or output transformer to change direction of magnetic
field. This is not too easy to do, since chassis punchouts are usually
designed to hold transformers in one position. With these parts in
operation, turn them slowly and note effect on v.t.v.m. reading.
Frequency-response measurements
The frequency response of an amplifier, tone-control circuit or
equalizer is an important factor in evaluating any audio system.
Remember, however, that an audio system is only as good as its
poorest unit. Overall system fidelity depends not only on amplifier
characteristics but also on the quality of the sound source — phono
or tape pickup — and upon the speaker system. Because the sound
must pass through all of these units in the ordinary audio setup,
frequency response is limited by the unit which has the narrowest
response characteristics. The v.Lv.m. and an audio signal generator
can be used to check the frequency response of the audio system
from the output of the sound pickup, such as the phonograph
cartridge, to the input to the speaker system. Checks which can be
made with these two test instruments do not cover the response
characteristics of the pickup or the loudspeaker itself. Response
characteristics of phono pickups are best measured by use of a
special test record. The sound output of the loudspeaker system
can be checked only through use of a calibrated microphone-ampli
fier system. This latter test setup is usually costly and is conse
quently restricted to the laboratory or sound studio.
It is possible, however, for the audio experimenter to obtain a
130
complete overall check with the v.t.v.m., a test record and an audio
generator. If the response characteristics of the loudspeaker have
been accurately plotted by the manufacturer and if the speaker is
housed in the same type of enclosure used by the manufacturer
when he compiled the speaker-response data, the experimenter can
obtain an accurate evaluation of his entire audio system.
Before frequency-response tests are run on any equipment, check
the equipment to make sure that it is in proper operating condi
tion. Test amplifier response wherever possible with the loud
speaker mounted in its proper enclosure and connected to the
amplifier. This is desirable because the speaker impedance is
reflected back into the amplifier and has an effect on the amplifier
output signal. If the amplifier is tested at a high power level, the
reproduced sound may not be tolerable. In this case, a noninduc-
tive dummy load can be used. In any case, a high-power amplifier
should never be operated without a load connected io the output
transformer. Failure to observe this precaution may result in
burnout of the transformer. If a dummy load is used, it should
consist of a resistance equal to the voice coil impedance and should
have a wattage rating equal to approximately twice the amplifier
rated power output. Wirewound resistors are generally used be
cause of their high-wattage dissipation. The inductance of these
resistors is usually not high enough to cause undesirable reactance
at audio frequencies.
Preliminary considerations
The frequency response of amplifiers, tone-control circuits and
equalizers is measured by feeding a signal of fixed voltage level
from the audio oscillator into the input terminals of the unit
under test and measuring the voltage output with the v.t.v.m.
Readings are taken wdth the v.t.v.m. at different frequency inter
vals throughout the bandpass of the unit. It is necessary that the
voltage of the generator signal fed to the unit under test be kept
at the same level for all frequencies. This requires that the gen
erator have essentially flat output throughout its tuning range.
The v.t.v.m. can be used to check the output voltage, if desired,
whenever the frequency is changed. The output voltage control
on the generator can be used to set the voltage to the required
value.
The output circuit of the generator should also be able to
work into the input impedance of the unit under test to preserve
the waveshape of the test signal. Most generators are equipped
131
with an output selector switch which permits setting up the gen
erator for externa] loads ranging from a few ohms (voice coil
impedances) to several hundred thousand ohms (high-impedance
amplifier input). Cathode-follower amplifiers, employed in various
stages of some audio amplifiers, generally have a very high input
impedance and an output impedance of a few thousand ohms at
best. Amplifiers used in broadcast and professional audio work
frequently utilize transformer input designed to match 250- and
500-ohm audio lines. Crystal microphones and crystal pickups
usually connect to amplifiers having input impedances from 50,000
ohms to 1 megohm. Magnetic pickups for tape and disc recordings
VTVM
Fig. 803. Test setup for the measurement of very low a.c. voltages. With this
technique the approximate voltage gain of a preamplifier can be measured easily.
132
Frequency-response checks
It is general practice to plot five curves on an amplifier which
contains tone-control circuits. Individual curves are plotted with
tone controls set for flat response and for maximum and minimum
bass and treble response. The test for flat response can be made
first. Try this procedure:
1. Set up the v.t.v.m., amplifier and audio-signal generator as
shown in Fig. 804.
2. Set the amplifier volume control to a position which cor
responds to the power output level at which the test is to be made.
3. With a signal of the desired voltage applied to the amplifier
input, set the tone controls so that they have a minimum effect on
VTYM
Fig. 804. Technique for checking the frequency response of an audio amplifier.
the frequency response of the amplifier. On some amplifiers, the
“fiat” positions for these controls are indicated. In other cases, the
points of minimum effect will have to be determined experimen
tally.
4. Tune the generator to a frequency just below the estimated
bass limit of the amplifier. With the v.t.v.m. connected across the
speaker voice coil, record the a.c. voltage reading.
5. Tune the signal generator to the next highest frequency at
which the measurement is to be made. Record the v.t.v.m. reading.
Below 200 c.p.s., a reading can be taken every 20 cycles. The inter
val can be increased to 1,000 cycles when readings begin to level
off. The interval can be reduced to 100 c.p.s. on the high end when
readings start to decrease. Keep the audio-generator output con
stant for all test frequencies.
6. Tune the generator to a frequency in the treble range
(3,000 c.p.s. is a good starting point). Vary the bass control and
note the effect on the output voltage. Similarly vary the treble
control. If the treble control produces a large voltage change, lower
the frequency of the audio generator. Alternately vary the bass
and treble controls and adjust the frequency of the generator until
a frequency is found at which both bass and treble controls have
minimum effect on the output voltage. This frequency can be used
133
as a reference frequency for plotting a curve of output voltage vs.
frequency. The effect of varying the bass control should be plotted
from the lower limit of the amplifier up to the reference level,
while the effect of varying the treble control should be plotted
from the reference frequency to the upper frequency limit of the
amplifier.
Assume that 1,000 c.p.s. is found to be the frequency at which
Fig. 805. Typical curves obtained for various settings of the audio-amplifier
controls.
the bass and treble controls have the least effect on the output
voltage. Set the bass control to minimum and take output voltage
readings with the frequency varied from 30 to 1,000 cycles. Set the
bass control to maximum and repeat the output voltage readings
as the frequency is varied. The same process is followed in deter
mining the effect of the treble control except that the output
voltage reading need not be taken below 1,000 cycles.
From these voltage readings, curves such as those shown in
Fig. 805 may be plotted. The curves are plotted in decibel units
so that comparison of the results can be seen. These curves indi
cate the frequency response of the amplifier with minimum and
maximum bass and treble. The flat line corresponds to the re
sponse obtained at the flat setting of these controls.
134
generally used when measurements are being made. Gliding-tone
records are useful when it is desired to check for peaks but may
also be used for frequency-response measurements by appropriate
timing. These normally commence at a high frequency, maintain
constant velocity characteristics down to the crossover frequency,
then follow the recording characteristic down to the final fre
quency (approximately at the rate of 6 db/octave). In some cases
the high-frequency response follows the recording characteristic.
The test is made by playing the record on the phonograph and
monitoring the voltage output of the amplifier with the v.t.v.m.
Fig. 806. Test method for determining the input and output imped
ances of an audio amplifier. The same procedure can be used for
impedance measurements of circuits other than audio amplifiers.
in the record during the recording process. The usual crossover
frequencies (the frequency below which the bass is attenuated)
are 250, 300, 350 and 500 c.p.s. Various record manufacturers use
different crossover frequencies and consequently they have differ
ent recording curves.
Equalizer networks are designed to match a specific recording
curve. A properly designed equalizer in an ideal amplifier will
have an output which is the inverse of the recording curve. In
other words, if the recording attenuates frequencies below 500
c.p.s. at a certain rate, the equalized amplifier must boost these
frequencies at the same rate. The curves for these units are plotted
in the same manner as described earlier.
135
from the audio generator. This method, shown in Fig. 807, is
accurate if the impedance is resistive; approximate if it is reactive.
Since most amplifiers have inputs which are mostly resistive,
it is sufficient to check the input impedance with a low-frequency
signal. A frequency of 50 c.p.s. may be used when checking a high-
fidelity amplifier. For checking the input impedance of a public
address system or an amplifier with limited low-frequency re
sponse, a test frequency of 100 c.p.s. may be used. The circuit
shown in Fig. 806 is used to determine both high and low input
impedances. For high-impedance circuits use the high-impedance
output of the generator, for low-impedance circuits use the low-
impedance output. Ri is used for input impedance measurements,
Ro for output impedance measurements.
Set the frequency of the generator to either 50 or 100 c.p.s.,
depending on the quality of the equipment under test. Connect
the generator to the amplifier; set the output control so that there
is a convenient reading on the a.c. scale of the v.t.v.m. Keep the
signal level low so that the amplifier is not overloaded. Connect
the potentiometer (Rt) into the circuit and vary its resistance
until the voltage indicated on the v.t.v.m. is one-half the original
value. The resistance of the potentiometer is then equal to the
input impedance.
Output impedances are measured by the same technique except
that the potentiometer (Ro) is shunted across the output of the
amplifier. Since the output impedance changes as the frequency is
varied, it is advisable to check the output impedance over a wide
band of frequencies.
Decibel measurements
To study the relationship between the output voltages measured
with the v.t.v.m. in audio applications and the frequency response
of an amplifier, it is helpful to plot a curve showing the a.c. volt
age output in decibels against frequency. The conversion table
shown in Fig. 807 offers a convenient means of changing voltage
readings to decibels. Unless the same reference level and load
impedance is used, however, the figures obtained from the chart
will be relative values.
136
value is defined as the number of decibels above or below a zero
reference level of 1 milliwatt in 600 ohms at 1,000 c.p.s. Zero dbm,
therefore, would indicate a power level of 1 milliwatt; 10 dbm,
10 milliwatts; and 20 dbm, 100 milliwatts.
The graph makes possible rapid conversion of r.m.s. voltages
to corresponding dbm values. Associated power levels can be read
along the top of the graph. If the r.m.s. voltage is measured across
a resistive load other than 600 ohms, the correction factor given
here must be added algebraically to the dbm values read from the
graph in Fig. 807. For resistive loads not given in Table 7, the
following formula should be used for determining the correction
factor:
Correction factor = 10 log 600
R
where R is the load in ohms. If R is greater than 600 ohms, the
correction factor is negative.
•DBM is the increment to be added algebraically to the dbm value read from
the chart in Fig. 807.
137
that it corresponds to I milliwatt. Repeat this process, but this
time stop at the diagonal line and then move to the left side of the
graph. The corresponding dbm value is zero. Similarly, 2 volts
corresponds to 8 dbm and slightly less than 7 milliwatts.
The decibel may be defined in a number of ways, including the
following equations:
p E k
db = 10 log — db = 20 log =^. db = 20 log -2.
Px Ej Ij
In these equations Pi is the input power in -watts, P2 is output
power in watts, Ej is the input voltage in volts, E2 the output volt
age in volts, Ii is the input current in amperes, and I2 is output
current in amperes.
Standard reference levels
In making any series of measurements, it is necessary to employ
a standard specified set of conditions. If this standard is to be
I
L73VRMS <500*.
__1___
Fig. 808. This standard
is 6 mw across 500 ohms.
expressed in terms of voltage and power, the resistance of the load
used must be specified. Two different reference levels are com
monly used. These standards are shown in Fig. 808 and Fig. 809.
.7WRMS <600xk
I____________JPWR* I MW
139
scale are marked plus and units to the left are marked minus.
The standard load and zero-reference level employed in calibrat
ing these meters is stamped on the meter face plate or is given in
the instruction manual.
While the chart shown in Fig. 807 is extremely useful, a similar,
but more direct approach can be used. Thus, if extensive work is
to be done with a constant load resistance, power level, or voltage
value, the technician can easily modify his v.t.v.m. to provide
readings of decibel values directly from the face of the meter. This
procedure is described in detail in Chapter 10.
140
chapte
he radio amateur who has not yet discovered the value of the
T v.t.v.m. around the ham shack has overlooked a helpful and
versatile accessory to his hobby. The many advantages which make
the v.t.v.m. such a useful instrument in radio, TV, audio trouble
shooting, alignment, and adjustment, make it an extremely desir
able service aid in the ham shack. The v.t.v.m. can be put to good
work in troubleshooting communications receivers, frequency
meters, variable-frequency and crystal oscillators, exciter units,
r.f. amplifiers, power supplies of all sizes, modulators and speech
amplifiers, and many other types of specialized equipment used by
the amateur.
The basic troubleshooting methods described in earlier chap
ters can be applied equally well to amateur equipment. Many
of the parts used in amateur apparatus are the same as or similar
to the parts used in radio and television receivers. Capacitors, coils,
transformers, resistors, and other components used by amateurs
are subject to the same types of defects and breakdowns that occur
in broadcast receivers. Therefore, troubleshooting methods used
in localizing faulty operation in a section or stage in amateur
equipment are similar to those employed in repairing radio and
TV sets. The procedures for testing individual components are
the same as those discussed in Chapter 6.
In most amateur applications the v.t.v.m. is even more useful
because of its flexibility and accuracy. The communications re
ceiver, for example, utilizes many of the conventional circuits
141
found in better-quality broadcast receivers and also employs addi
tional circuits, such as beat-frequency oscillators, crystal filters,
and voltage-regulated power supplies. Because the communica
tions receiver is designed for stable performance and has a high
degree of sensitivity and selectivity, adjustment, maintenance and
alignment are obviously important. Good test procedures and a
v.t.v.m. are, therefore, especially necessary.
Some new and interesting applications, most of which involve
familiar procedures, can be found in servicing the amateur trans
mitter. In oscillator and exciter stages, for example, proper operat
ing voltages, good regulation, and correct driving voltages are
essential for frequency stability and a clean output signal. The
v.t.v.m. is especially useful in checking and adjusting such stages.
In checking receiving, transmitting and special-purpose equip
ment, the v.t.v.m. is valuable because of its high sensitivity, low-
loading characteristics, and adaptability to special-purpose probes
and cables.
Special considerations
1. Remember that there is danger inherent in testing amateur
equipment which operates at hazardous voltages. Therefore, the
amateur should thoroughly familiarize himself with the equip
ment under test before working on it, bearing in mind that high
voltages may appear at unexpected points in defective equipment.
2. It is good practice to remove power before connecting test
leads to high-voltage points. If this is impractical, be especially
careful to avoid accidental contact with equipment racks and other
objects which can provide a ground. Working with one hand in
your pocket and standing on a properly insulated floor lessens the
danger of shock.
3. Filter capacitors may store a charge large enough to be
hazardous. Therefore, discharge filter capacitors before attaching
test leads-.
142
4. Remember that leads with broken insulation provide the
additional hazard of high voltages appearing at exposed points
along the leads. Check test leads for frayed or broken insulation
before working with them. This is most important when making
measurements with the high-voltage probe.
5. To lessen the danger of accidental shock, disconnect the test
leads immediately after the test is completed.
6. Remember that the risk of severe shock is only one of the
possible hazards. Even a minor shock can place the operator in
hazard of more serious risk such as a bad fall or contact with a
source of higher voltage.
7. The experienced amateur continuously guards against in
jury and does not work on hazardous circuits unless another person
is available to assist in case of accident.
8. Do not attempt to make measurements in circuits contain
ing high-level r.f. signals. If an accessory r.f. probe is used, make
sure the d.c. and r.f. voltages do not exceed the input rating of
the probe. It is generally impossible to measure the r.f. signal of
exciter and power stages with the v.t.v.m. because of the relatively
large r.f. signal voltages.
9. Take special care to ground the v.t.v.m. thoroughly when
making any high-frequency measurements. Remember that the
presence of stray r.f. fields can cause some weird effects and upset
normal a.c. and d.c. voltage measurements.
10. Make sure the range and function switches on the v.t.v.m
are set correctly before connecting the v.t.v.m. to the circuit to be
tested. Failure to observe this precaution can result in damage to
the v.t.v.m.
143
For example, in crystal oscillators, the plate and screen voltages
must be adjusted properly to assure operation, not only within the
tube’s ratings, but also to prevent excessive grid-circuit current
from damaging the crystal. In both crystal and variable-frequency
oscillators, improper d.c. voltages can cause frequency shift or
chirp under keying conditions. In variable frequency oscillator
circuits, excessive frequency drift may result from improper
operating voltages.
Two basic types of oscillators employed in amateur transmitters
are shown in Fig. 901. Both the crystal oscillator shown in Fig.
901-a and the variable-frequency oscillator in Fig. 901-b have many
variations in their circuit design. If the amateur is familiar with
the requirements for making measurements in both the types of
oscillators shown here, however, he should be able to make meas
urements without difficulty in nearly every type of oscillator.
The crystal oscillator of Fig. 901-a provides output on a fre
quency fixed by the crystal in the control-grid circuit. The supply
voltage from the power supply is fed to the B plus and B minus
terminals. The correct value of screen voltage is obtained by drop
ping the higher B plus voltage through the screen-grid resistor.
In tetrode and pentode oscillators, the screen voltage exercises a
great influence on performance of the stage, and this voltage
should be measured with the v.t.v.m. set up for d.c. voltage meas
urements. If the oscillator is keyed, the plate and screen voltages
should be measured under both key-up and key-down conditions
because, unless a voltage-regulated power supply is used, the key
down (load) voltage will be lower than the key-up voltage.
If the oscillator must deliver much power, the plate voltage
should be measured with the v.t.v.m. at point D rather than at any
point between D and the plate pin (G) because the plate circuit
will contain a relatively strong r.f. signal which will disrupt the
voltage reading. The capacitor from point D to ground serves to
keep r.f. energy from getting into the supply line and also permits
the measurement to be made. Because there is no dropping resistor
in this plate circuit, the d.c. voltage measured at D should be the
same as that at the plate. In crystal-oscillator stages, as in doubler
and amplifier stages, the tuning of the plate-tank circuit will affect
plate and screen voltages. Off-resonance tuning will cause plate
current to increase, resulting in a change in the d.c. voltages at
the tube pins.
It is not usually possible to measure the r.f. signal voltages at
points E and F in power oscillators because the peak values of these
144
voltages normally exceed the maximum allowable input voltages
of the r.f. probes used with v.t.v.m.’s. Crystal-diode type r.f. probes
generally have a much lower signal-handling rating than have the
vacuum-tube probes. Before r.f. measurements are attempted,
therefore, the amateur should be sure that the v.t.v.m. and the r.f.
probe can handle the estimated signal voltage without damage to
the measurement equipment.
The screen voltage should be measured with the d.c. probe at
point C, which is at r.f. ground because of the bypass capacitor.
Fig. 90I-a,-b. Crystal oscillator (a) and variable-frequency oscillator (b) can be
checked at the various test points shown in the circuits.
Current measurements
If the v.t.v.m. has facilities for measuring direct current, the
peter can be inserted at the points marked “X” in the plate and
screen leads to read the plate and screen currents. The amount of
current flow will depend upon the tuning of the plate tank circuit
and also upon the load connected to point F. Remember that both
145
these leads are above ground and caution should be exercised
when connecting the v.t.v.m. into the circuit. For additional con
siderations on current measurements, see the section on direct
current measurements in Chapter 5.
A variable-frequency oscillator is shown in Fig. 901-b. The
operating frequency of this oscillator is determined by the tuning
of the split-stator capacitor in the grid circuit. Supply voltages, grid
bias, and output-signal voltages are measured with the v.t.v.m.
with the same procedures used in checking the crystal-oscillator
stage shown in Fig. 901-a. In the variable-frequency oscillator
stage, however, tuning of the grid capacitor may produce changes
in the output signal, in the bias and operating voltages. Because
these oscillators are usually constructed to cover a specified range
of frequencies, the grid tank circuit should be designed to give
uniform output throughout the tuning range. If essentially flat
(uniform) output cannot be obtained, the circuit should be ad
justed to give peak output near the center of the range. The
v.t.v.m. is especially important in checking these variable-fre
quency oscillators because they are more susceptible to power
supply voltage changes and keying. The v.t.v.m. is extremely
valuable in debugging such stages. Plate and screen currents can
be read at points marked “X”.
4 J
Fig. 902-a,-b. Two different types of r.f. amplifier circuits (buffer or voltage
amplifier). The drawing at the left (a) and that at the right (b) differ in the way
in which they are coupled to the driver and also in the arrangement of the
plate-tank circuit.
mum specified value when the key is up. A voltage divider and
bleeder, such as that shown in Fig. 902-a, is recommended for
c.w. stages. The series-dropping resistor arrangement used in the
screen circuit of Fig. 902-b is usually satisfactory for phone opera
tion because excitation is applied continuously to the stage. In
either circuit arrangement, the v.t.v.m. should be used to measure
the screen voltage under key-up and key-down conditions. The
screen bypass capacitor puts the screen circuit at r.f. ground poten
tial, permitting d.c.-voltage measurements at this point to be made
with the v.t.v.m.
147
— which means that no plate current flows when excitation is
removed (key up). Some final amplifiers used in c.w\ transmitters
are operated under class ABi, ABo or class B conditions, however,
and a small amount of plate current may flow under key-up condi
tions. Usually, however, bias is adjusted to a cut-off or below
cut-off value recommended by the tube manufacturer. For class-C
phone amplifier operation, the bias is usually increased to approxi
mately two and a half times the amount required for plate-current
cut-off.
Grid current can be measured in both of the stages shown by
inserting the meter at points marked “X” in the grid circuits.
These points should not contain r.f., which can disrupt the meter
reading. Total tube current is checked at points “X” in the
cathode circuits. The proper test points for screen and plate cur
rent (C and D) are also shown. See Table 8 for complete data on
oscillator and amplifier measurements.
TABLE 8
Key Letter V.T.V.M.
in Fig. 901 Setting
and Fig. 902 To Measure (Volts) Probe Type
A Grid bias -D.C. D.c. isolating probe
B Cathode bias 4-D.C.
ft
C D.c. screen voltage 4-D.C.
D Plate supply voltage 4-D.C.
E R.f. signal voltage ±D.C.* R.f. probe
F R.f. output signal voltage ±D.C* R.f. probe
G D.c. plate voltage 4-D.C. D.c. isolating probe
♦Polarity of v.t.v.m. setting depends on the polarity of the output voltage from the
r.f. (high-frequency) probe.
Several different methods of obtaining grid bias are used in
transmitter stages, some of which are shown in Fig. 903. In all
the arrangements shown (except Fig. 903-f) the total tube bias
voltage is measured with the v.t.v.m. by connecting the ground
cable of the v.t.v.m. to the chassis (because the cathodes are
grounded) and measuring the d.c. operating bias (Eop) on the
—DC volts ranges of the v.t.v.m. Bias should be measured under
both key-up and key-down conditions because the bias voltage may
change appreciably with removal and application of the excitation.
In Fig. 903-a, the operating bias voltage, Eop, depends on the excita
tion and is developed by grid current flow through the resistor, RI.
Fixed or protective bias (Epr) is used in Figs. 903-b to 903-e
inclusive. The operating bias in Fig. 903-c and Fig. 903-e is the
total of the fixed bias and the developed grid-leak bias. Both
protective and operating bias are provided by the arrangement
148
shown in Fig. 903-f although this method should not be employed
in c.w. stages.
In the circuits shown in Fig. 903, bias under key-up conditions
can be measured by connecting the ground cable of the v.t.v.m.
to the chassis and connecting the d.c. probe to the control grid pin
of the tube. Not all grid-circuit arrangements contain an r.f. choke
or r.f. bypass capacitor and the accuracy of measurement of the
grid bias at the tube pin under key-down or phone conditions will
e
Fig. 903-a to f. Bias arrangements for transmitter tubes. Circuit a uses
grid-leak bias while the battery in circuit b could represent a bias (power
supply) pack. Circuit c uses a combination of the types shown in a and b.
In circuit d, RI is the bleeder across the bias pack. Circuit e has the advan
tage of using a gas-tube regulator across the bias pack output. Bias is
developed across the cathode resistor, RI, in circuit f.
depend upon the driving power and the stability of the v.t.v.m.
under r.f. conditions.
Bias in the circuit shown in Fig. 903-f should be measured by
connecting the ground cable of the v.t.v.m. to the chassis and
connecting the d.c. probe to the cathode pin of the tube.
Preliminary checks
In setting up an r.f. amplifier for the first time, the v.t.v.m. can
be used to make several precautionary measurements before plate
and screen voltages are applied. In checking out a high-power
amplifier such as that shown in Fig. 904 the following procedure
should be used:
1. Apply filament and grid bias voltages. Remove excitation.
2. Set up the v.t.v.m. for a.c. voltage measurements. Check the
filament voltage of each tube directly at the tube filament pins (xx)
by connecting the ground cable of the vxv.m. to one pin and the
a.c. probe to the other pin of the same tube socket. The filament
voltage should be within at least ±5% of the recommended fila
ment voltage for the tube type.
3. Similarly check the filament voltage of the other amplifier
tube (yy). The voltage should be the same as that measured on
the first tube. If the voltage measured at either filament is off more
than ±5%, measure the line voltage on the primary side of the
filament transformers. Use the v.t.v.m. set to read a.c. volts for
this check. Make sure the range switch is in the proper position.
Too low or too high a line voltage may be the cause of improper
filament voltages. If the line voltage is correct, the voltage drop
150
in the filament leads may be responsible, in which case a heavier
conductor should be used, or one of the filament transformers
may need to be replaced because of improper turns ratio. In no
event should the high-power tubes be operated outside their fila
ment-voltage ratings when B— voltages are applied.
4. Set up the v.t.v.m. for minus d.c. voltage measurements.
Connect the ground cable to the amplifier chassis. The d.c. bias
supplied by the separate bias supply should be checked at the
control grid pins of both amplifier tubes by connecting the d.c.
probe to these points. This is a wise precaution, especially in
equipment which is not relay protected again grid-bias failure.
If equally correct voltages are measured, the operator may be
sure that wiring is correct and that the bias supply is functioning
properly.
The separate bias-supply leads feeding the two halves of the
grid circuit are provided as a means of checking the balance of
the push-pull circuit. Grid-bias voltage measurements (taken from
control grid to chassis) in the amplifier circuit should not be
attempted when excitation is applied because of the large circulat
ing r.f. current.
5. Adjust the grid-tank tuning and the coupling to obtain the
required amount of grid-current. If the v.t.v.m. is equipped with
a current-measuring circuit, the instrument can be used as a mil
liammeter. Open the bias leads at point A, apply excitation, and
read the amount of direct current on the meter. Repeat at point B.
151
Both readings should be the same. If not, then RI and R2 should
be adjusted until both readings are identical. We are assuming,
of course, that the coupling from the driver stage is such that
both VI and V2 receive equal amounts of excitation. Bias due to
excitation or drive is developed across RI and R2. The r.f. chokes
(RFC) and capacitors Cl and C2 will keep r.f. out of the v.t.v.m.
After obtaining the correct value of grid drive, remove excitation.
6. The amplifier should now be checked for plate current
cut-off or, if the amplifier is to be run with a small amount of
static plate current flowing, the amplifier can be checked for the
desired static plate-current flow.
With bias applied and excitation removed, apply plate and
screen voltages. Note the plate-current flow, if any, and adjust the
bias voltage from the bias supply, if necessary, to give the required
cut-off or static current. Remove the high voltage.
7. The plate and screen circuits can now be checked with
excitation applied. In circuits which employ high-perveance tubes,
special care should be taken to prevent excessive plate and screen
current flow while tuning adjustments are made. Unless dial set
tings of plate-tank resonant points are first established by means
of a grid-dip oscillator, a considerably reduced plate voltage should
be employed for tuning operations. The value of this voltage
depends upon the plate characteristics and permissible dissipation
of the tube type employed and can best be determined from the
manufacturers’ tube data for the type involved. The plate-supply
voltage should be measured at point C and the screen-supply
voltage should be measured at point D. The radio-frequency choke
(RFC) and capacitor C5 will keep r.f. out of the v.t.v.m. when
making a measurement of the supply voltage for the plate at
point C. Similarly, the radio-frequency choke and capacitors C3
and C4 keep point D at r.f. ground potential. Resistors R3 and R4
act as a voltage divider to put the proper amount of d.c. voltage
on the screens. If, when making d.c. voltage measurements of
screen potential you find that the screen voltage is too low, then
the tie point at D should be moved further up on R4 (toward
point C). Similarly, if the screen voltage is too high, then D should
be moved down on the voltage divider toward ground.
153
meat of the neutralizing capacitors or loops in the amplifier will
eliminate the r.f. from the plate-tank circuit. While these adjust
ments are made, however, it is necessary to have some additional
means of indicating the presence of this energy because the regular
metering circuits will usually not indicate the presence of the r.f.
A typical neutralizing setup employing a v.t.v.m. as an r.f.
indicator is shown in Fig. 905. The v.t.v.m. should be set to its
154
The tuned circuit consists simply of a coil and capacitor which
can be adjusted to the operating frequency. The coil is tapped
about a third of the length along its turns. The set to
read low voltage dx. is connected to the tap through the r.f.
(rectifying) probe and the circuit is tuned for maximum reading
on the v.tv.m. The readings obtained will be relative because the
field strength will vary considerably with the power input to the
equipment and the orientation of the antenna. A small length of
wire may be added to the tuned circuit to increase the r.f. pickup.
155
907 and 908. Regulator tubes can be obtained for 75, 90, 105 and
150 volts. Such tubes can be connected in series, as shown in
Fig. 908, to provide two or more regulated voltages. The same
supply can furnish an unregulated voltage by tapping off at a point
prior to that at which regulation takes place.
The v.t.v.m. can be used in setting the adjustment for current
through the tubes. A starting voltage approximately 30 percent
higher than the operating voltage is required to fire the tubes.
(Voltage regulation will not be obtained unless the gas tube fires.)
The value of the current limiting resistor R must be between that
156
chapter
A
miscellaneous applications
158
250 volts and the measured ripple voltage is 1.25 volts, the per
centage of ripple is 0.5. The percentage of ripple is usually greater
under load than without load because the a.c. component remains
relatively unchanged when the d.c. output voltage is dropped by
the load current. In general, the amount of ripple which can be
tolerated in audio amplifiers and radio receivers is less than 1
percent.
Power-supply regulation can be determined simply by measur
ing the d.c. output voltage under load (Eminimom) and no-load con
ditions (Emaximum). Percentage of regulation is equal to:
Ejnaximum Endnlmum
-------------------------- X 100
PWR TRANS
voltage reading. In making this test, do not shunt the cell with a
load resistor. Make the measurement using an isolating probe
and with the v.t.v.m. set to read on a low-voltage d.c. scale.
Checking transformers
The average radio junk box usually contains one or more mul
tiple-winding transformers for which no data are available.
Because it is not always possible to pair up leads by means of their
color codings, the v.t.v.m. can be used to determine which leads
are connected internally and what voltage ratings they have.
The windings of power transformers should be paired up first
by means of resistance measurements. Filament windings will have
a very low resistance, usually a few ohms or less, while the plate
winding has the highest resistance, ordinarily 100 ohms or more.
The resistance of the primary winding will vary, although it is
more than that of the filament windings and less than that of the
plate winding. The primary resistance ranges from about 1 to 5
ohms. Center taps of filament and plate windings should also be
determined.
160
The turns ratio of the transformer can be determined only by
voltage measurements; resistance measurements cannot provide an
accurate indication. After the related leads have been identified,
the transformer turns ratio can be found by applying an a.c.
voltage to the primary winding of the transformer and measuring
the a.c. output voltage with the v.t.v.m. connected, in turn, across
each winding. If the primary winding could not be clearly identi
fied by resistance measurements, only a low test voltage should be
applied to determine the turns ratio. A filament voltage from
another transformer or an audio-oscillator signal can be used.
The setup is shown in Fig. 1004. Using only enough test voltage
Checking potentiometers
The overall resistance, taper, value of tone-control taps, and
general condition of potentiometers can be determined easily
with the v.t.v.m. Few of these controls used in radio and tele
vision receivers and audio equipment exceed a maximum value
of 10 megohms. Consequently, the controls can be checked with
considerable accuracy with the v.t.v.m., which has a center-scale
value of 10 megohms on its highest resistance range.
Fortunately for the service technician, potentiometers have a
reputation for becoming noisy. This is not always due to the
design of the control but can be caused by rough handling, insuffi
cient power rating, poor choice of type for the circuit involved,
high humidity, defective circuitry, or wear.
The best test for noisy operation, of course, is to try the poten
tiometer in an operating circuit. A good approximation of the
condition of the control may be had by connecting the v.t.v.m.
(set to read ohms) across the arm and one end of the control and
varying the position of the shaft. If the control is in good condi
tion, the resistance should change smoothly as the shaft is rotated.
A spotty control will cause the meter pointer to flicker or jump in
spots as the shaft is turned.
If the control is used in high-gain circuits especially susceptible
to noise, however, such as audio preamplifier stages, the true con
dition can be determined only by installing the pot and listening
for noise in the output. Some of these circuits are extremely
sensitive and only a new control of good design and high quality
will give noise-free performance. Unless a spotty control can be
162
repaired, it should be discarded immediately to prevent its finding
its way back to the stock drawer.
Spotty controls often can be repaired by applying a few drops
of anti-noise solution to the base of the shaft and turning the shaft
back and forth several times. In a pinch, cigarette-lighter fluid
will also do the trick. In cases where a new control fails to correct
the noise it is wise to check the circuitry in the stage and make
sure that a leaky capacitor or other defective component is not
causing trouble in the control circuit.
Potentiometers are manufactured with a variety of tapers.
Basically, the taper indicates the rate of change of resistance per
unit of rotation throughout the range of the control. The taper
of a potentiometer may be determined by plotting resistance
readings against degrees of rotation on a graph. A small dial, cali
brated from zero to 300 degrees, can be placed around the shaft,
and a pointer knob used to indicate the exact amount of rotation.
In most radio repair work, the value of the potentiometer is not
critical. When replacing controls, however, care should be taken
when cutting the shaft to proper length that excess force does not
cause the shaft to loosen. Rough handling can make a new control
noisy.
When replacing these controls in TV and FM receivers, careless
handling of adjacent leads should also be avoided. This is es
pecially true in small FM receivers where improper lead dress can
cause a bad case of hum. In sets where wiring is compact, it is a
good idea to make a pencil sketch of adjacent leads, if no wiring
schematic is available, to make sure leads are dressed in their
original positions.
Calibrating an oscilloscope
The v.t.v.m. can be used to calibrate the vertical-deflection
amplifier of an oscilloscope. A voltage-calibrated oscilloscope is
extremely valuable in troubleshooting TV receivers and other
electronic equipment because calibration makes possible simultan
eous waveshape display and voltage measurement on the screen
of the oscilloscope.
The maximum deflection sensitivity of an oscilloscope is ex
pressed as the number of volts required at the scope input termin
als to produce one inch of vertical deflection on the oscilloscope
screen when the scope controls are set for maximum gain. For
example, if the input attenuator on the scope is set to its minimum
attenuation (maximum gain) position and 1 volt applied to the
163
input terminals produces a waveform one inch high on the screen,
the scope is said to have a vertical sensitivity of one volt per inch.
If scope amplification is linear, five volts applied at the input
should produce full-screen deflection on a five-inch cathode ray
tube. It is possible, therefore, to calibrate or scale the screen of
the cathode-ray tube directly in terms of volts. The v.t.v.m. is used
in the initial calibration. Once calibrated, the scope may be used
in voltage measurements without requiring the addition of the
v.t.v.m.
The test setup required consists of the oscilloscope, v.t.v.m., and
an audio oscillator. Before proceeding with the calibration, how
ever, it is necessary that some of the characteristics of the scope
be understood or established. The following are important:
1. Some scopes do not provide linear deflection at all points
on the tube screen. Nonlinearity is usually greatest near the edges
of the screen. Therefore, it may not be possible to obtain linear
deflection over more than two-thirds or three-fourths of the screen
height. In these cases, the calibration should be confined to the
center section of the screen or it will be necessary to plot a non
linear voltage scale. In the example which follows, both linear
and nonlinear scales will be plotted.
2. The frequency response range of the oscilloscope vertical
amplifier should be established. The response of the instrument
will not be the same at all frequencies, and the range over which
amplification is uniform should be determined. Unless this range
is established, inaccurate voltage measurements may be obtained.
For example, if scope response is flat from 30 cps to 500 kc a one-
volt input signal of any frequency within that range should pro
duce the same amount of vertical deflection on the screen. A one-
volt signal of 1.5 me, however, might produce only one quarter the
amount of vertical deflection on the screen — an error of 75%.
The frequency of the calibrating voltage, therefore, should lie
somewhere within the flat response range of the scope.
3. The attenuation ratio of the vertical-input attenuator
should be established by experiment. It is desirable that these posi
tions on the attenuator switch have a fixed attenuation ratio
between each other. If the ratio is a factor of ten, it is possible to
use one base scale for all the attenuator positions. If the attenuator
positions have a different ratio, a separate scale must be laid out on
the scope screen for each attenuator-switch position.
The test setup for calibrating the scope is shown in Fig. 1006.
The output from the audio oscillator is fed to the vertical-input
164
of the scope and also to the v.t.v.m. The vxv.m. is set up for
a.c. voltage measurements and the direct probe of the v.t.v.m. is
used. The scope may be calibrated in either r.m.s. or peak-to-peak
voltage values, although peak-to-peak values are more useful in
troubleshooting TV receivers and other equipment where com
plex waveforms are encountered. If the v.t.v.m. does not have
peak-to-peak voltage scales, it will be necessary to multiply the
- 40 -5
- 20 -
-0 —0 -0
A B c
167
ments can be increased by modification of the instrument as shown
in Fig. 1009. Parts required include two banana-plug jacks and a
well-insulated s.p.d.t. switch having low leakage. The switch and
jacks should be mounted on the front panel and connected as
shown by the heavy lines in Fig. 1009. The switch is inserted in
series with the divider network.
168
transformerless radio and TV receivers. A typical voltage-doubler
circuit employing two of these rectifiers is shown in Fig. 1010.
The v.t.v.m. can be used to check operating voltages as follows:
1. Set up the vxv.m. for a.c.-voltage measurements. With
power removed from the supply, connect the ground lead to the
B-minus bus and connect the a.c. probe to point A. Apply power.
The v.t.v.m. will read the line voltage.
2. Move the probe to point B. RI is a surge-current limiting
resistor of low value. Little or no drop in the voltage reading
should be noticed unless the resistor has increased in value. If
resistance is increased, look for a leaky or shorted filter capacitor
(C2 or C3) or faulty rectifier (CR1 or CR2).
3. Move the a.c. probe to the output of the rectifier section and
measure the ripple content at the input to the filter section (point
C). Using the d.c. probe, measure the d.c. voltage. Under no-load
conditions, the d.c. voltage will be approximately 2 to 214 times
the a.c. voltage (r.m.s.) measured at point B.
4. Measure the d.c. output voltage at point D. The a.c. output
ripple voltage should also be measured at this point.
In some supply circuits, a filter resistor (R2) may be used in
place of choke, LI. If the resistance of the choke or resistor is
known, the amount of d.c. load current may be determined by
Ohm’s law. Subtract the voltage measured at point D from the
voltage at point C and divide the difference by the resistance of
LI or R2.
When replacing defective rectifiers in either half-wave or volt
age-doubler circuits, make sure to observe correct polarity when
installing the new unit. If rectifier connections are reversed and
power is applied, the rectifier may be damaged quickly. Although
selenium rectifiers are best checked by means of current measure
ments in a specially designed tester, the front-to-back resistance
ratio as measured with the v.t.v.m. will usually indicate a shorted
or damaged unit.
169
Checking TV antennas and transmission lines
TV antennas and twin-line lead-ins can be checked easily with
the v.t.v.m. for continuity, shorts, and intermittent conditions in
test applications like those shown in Figs. 101 La and -b. In
installations which employ a continuous system (folded dipole
antenna), overall continuity can be checked at the TV receiver by
disconnecting the transmission line at the set and measuring the
continuity as shown in Fig. 101 La. The resistance of a properly
Fig. 1011-a, -b. Technique for testing transmission lines and TV antennas.
Disconnect the line from the receiver terminals before making the check.
170
this same method for checking cables used with test instruments.
Such cables, subjected to considerable flexing while in use, can
develop shorts, particularly where the cable is connected to the
probe.
Fig. 1012. The v.t.v.m. and a generator can be used to measure the
impedance of a coil or a capacitor, or combination of resistance, induc
tance and capacitance.
function of the v.t.v.m. and will give a close approximation of the
reactance or impedance at audio and low radio frequencies.
The setup requires use of a calibrated potentiometer or adjust
able resistor. The component under test is connected in series
with the resistance, and the junction point of the two components
is connected to ground of the v.t.v.m. A low-frequency test signal,
from 400 to 1,000 c.p.s., is applied across the two components.
The s.p.d.t. switch is supplied to permit reading the individual
voltage drops across the resistor and the component under test.
Repeated readings are taken while the resistor is adjusted until
the voltage drops (Ex and E2) across both the calibrated resistor
and the unit under test are the same. Equal voltage drops across the
units indicate that the reactance or impedance of the component
under test is equal to the resistance of the calibrating resistor.
If a calibrated resistor is not available, a good quality potentio
meter may be used. The potentiometer is adjusted as described
above for equal voltage drops. The potentiometer is then removed
from the circuit and its resistance checked with the v.t.v.m. The
setting of the pot should not be changed before the resistance
measurement is made. Power transformers may be checked using
171
a 60-c.p.s. signal from a filament transformer. In these applications,
a signal generator is not required. The setup for checking the
impedance of the primary of a power transformer is shown in
Fig. 1013. Be careful not to short any secondary windings of the
transformer under test. Each of the windings of the power trans
former can be checked in turn. Several different test potentio
meters (R) may be required if the impedances to be measured
have a wide range. Thus, you may need one test potentiometer for
175
figures in parentheses are used for power measurements, the proper
decibel values will be found quickly, without requiring calcula
tions and without extracting additional figures from a decibel con
version table.
Transforming these measurements, which are given with respect
to 1.5, into measurements referred to some other value is simple.
If, for example, the measurements are to be obtained with respect
to 1 volt, first perform the conversion in the usual way, with
respect to 1.5.
Now operate on the 10-volt range. Referring to the 0-12 volt
range, 10 volts in linear measurement corresponds to 16.5 db,
still with respect to 1.5. To find the final set of values, take each
of the decibel values of the measurements (with respect to 1.5) and
find the difference between them (subtract 16.5 db from each
value). The answer will be in decibels, referred to 10 volts.
Or, for example, if the power readings are to be given with
respect to 1 watt, the readings are first converted to decibels by
using the values in parentheses. Since the 0-3 range shows that
1 watt corresponds to —1.75 db, the difference is found between
—1.75 db and other power readings. Each value in the measure
ment has 1.75 db added to it. In this way, the final power measure
ment is found in decibels, with respect to 1 watt.
~ RFSGGEN
Fig. 1016. Setup for making a rough check of the gain of an r.f. amplifier stage.
Although the block diagram shows a TP receiver, the same technique can be
used for r.f. amplifiers in AM and FM sets.
177
The preamplifier provides an additional gain factor of approxi
mately 100. The lowest full-scale sensitivity, therefore, is 0.01
times the full-scale value of the lowest v.tv.m. range. For example,
the lowest range on the v.t.v.m. shown here has a full-scale value
of 1.5 volts. With addition of the preamplifier, the full-scale sensi
tivity is increased to 0.015 volt, or 15 millivolts. The frequency
response of the modified unit is within ±1 db from 25 to 30,000
Construction
The metal case for the preamplifier may be purchased or con
structed from a small sheet of aluminum. The layout is shown in
Fig. 1018. The tube socket mounts on the small bracket. Compo
nents should be connected and soldered to the socket before the
bracket is installed in the case. Keep all signal leads short to pre
vent hum pickup and feedback. (Miniature input and output
capacitors are used for this purpose.) Supply voltage leads from
178
the v.t.v.m. should be made long enough to permit the v.t.v.m. to
be taken out of the case without requiring disconnection of the
preamplifier. The complete circuit appears in Fig. 1019. The R-C
fro VTVM B+
——
x............ - FILTER v-------- -
ADDED IN VTVM
—
Fig. 1019. Diagram of the v.t.v.m. preamp
lifier. The circuit is that of a resistance-
coupled amplifier.
A small notch is cut in the v.t.v.m. case (lower left-hand comer)
for the voltage supply-lead grommet. Two holes are drilled in the
side of the case to permit attachment of the preamplifier with self
179
tapping screws. An underchassis view of the v.tv.m. preamplifier
showing connections to the v.t.v.m. is illustrated in Fig. 1020.
The B plus in the v.tv.m. shown here is 50 volts. Other v.t.v.m.’s
may employ higher voltages. If so, the frequency response of the
preamplifier may be extended because a greater amount of feed
back can be used. In any event, the B-plus voltage supplied to the
preamplifier should be limited to 75 volts by using a larger filter
resistance in place of the 10,000-ohm resistor shown in Fig. 1019.
Voltages higher than 75 volts might impose too heavy a load on
the v.t.v.m. voltage divider and cause unstable v.t.v.m. operation.
Fig. 1020. Underchassis view of the preamplifier shows connections to the v.t.v.m.
Calibration
Basic calibration of the preamplifier consists of applying a
0 31-volt signal to the input and adjusting the feedback control so
the v.tv.m. reads full scale on its lowest range. The 1-megohm
range control should be set to maximum during this adjustment.
The preamplifier may be calibrated by means of an accurate audio
180.
voltmeter or by using the simple calibration setups shown in
Figs. 1021 and 1022. When using these test setups, procedure is
as follows:
1. Set up the v.t.v.m. for a.c. measurements and measure the
line voltage.
2. Connect the ground cable to the bottom end of the 2,000-
ohm pot and connect the a.c. probe to the wiper arm. Adjust the
control to give a reading of 0.01 (one-hundredth) the value of the
line voltage. For example, if the line voltage is 115 volts, adjust
182
If the meter is used without an external shunt connected across
the binding posts, the maximum current limit will be equal to the
amount of current required for full-scale deflection. The sensitiv
ity of the meter (amount of current needed to move the pointer
needle to right-hand maximum) is usually given in the schematic
diagram or parts list for the v.t.v.m. Current readings should be
taken from a linear voltage scale.
The value of the external meter shunt will be determined by
Fig. 1023-a, -b. The meter in the bridge circuit (a) can be connected to
a double-pole, double-throw switch (b) for making current measurements.
The meter can be provided with external shunts for increasing its cur-
rent-measuring range.
the amount of current you wish to measure. The larger the cur
rent, the lower will be the value of the meter shunt resistance.
The shunt resistance can easily be calculated from the formula:
Shunt resistance = x meter resistance
shunt current
The resistance of the meter in the v.t.v.m. can be supplied by the
manufacturer, but often appears in the circuit diagram for the
instrument. As an example, suppose that your meter has a resis
tance of 50 ohms and a sensitivity of 1,000 ohms-per-volt. This
means that a current of 1 milliampere will produce full-scale
deflection. Let us also suppose that you wish to measure a current
of 25 milliamperes.
To protect the meter movement, it would be advisable to have
a shunt whose resistance value would permit the meter needle
to come to rest at about center scale. This can easily be done if
you estimate the shunt to carry twice the required amount of
current. Thus, in this example we would calculate the value of a
50-milliampere shunt (2 X 25 milliamperes). Substituting these
values in our formula we would have: Shunt resistance =
(.001/.05) X 50 = 1 ohm. The maximum current you could
measure with this shunt would be 51 milliamperes, since under
such conditions 50 milliamperes would flow through the shunt,
and 1 milliampere would go through the meter, producing full
183
scale deflection. If you wanted to measure a larger amount of
current you would have to calculate and supply another shunt
resistor.
A conversion chart utilizing a straight line can be plotted, show
ing voltage in terms of current. If another current meter is avail-
OIRECTHaeE
184
checking oscillator injection voltage, and in receiver and converter
alignment.
185
the d.c. probe and the ground lead. The total value of this net
work, which is usually within ±1% of the rated value, can be
determined by adding the values of all the series resistors, includ
ing the isolating resistor in the probe tip. The v.t.v.m. should be
disconnected from the power line when used in these applications.
186
schematic diagram of Fig. 1027 may be of interest. The instrument
utilizes only a few readily available components, a single twin
section tube (6SN7-GT) and an inexpensive 500-microampere
meter movement. Designed for use on a.c. and d.c. voltage meas
urements, the v.t.v.m. can be compactly constructed to suit the
needs of the builder. The instrument can be used as an auxiliary
device, as a portable v.t.v.m. in the service chest, or as an emer
gency unit in the event that the regular v.t.v.m. is in for repairs.
Designed as a compromise between the extremes of expense and
impracticability, this simplified v.t.v.m. will perform most of the
Construction details
The unit can be mounted in any small aluminum or steel case
(or in a shielded box). These can be purchased from any parts
distributor. A small shelf or chassis is required as a mount for the
tube socket, selenium rectifier and terminal strips. The v.t.v.m.
could have been made still more compact by using a transformer
less power supply, but this is not recommended because of fre
quent applications involving transformerless and a.c.-d.c. receivers.
Any small power transformer delivering about 120 volts at 40 ma
(or more) is suitable. This voltage, furnished to the rectifier and
187
filter will provide a relatively low and stable d.c. output voltage.
The transformer should also have a filament winding for the
v.t.v.m. bridge tube. Although a 6SN7-GT is shown in the circuit,
a miniature type 12AU7 can be substituted.
A variety of meter movements having different basic sensitivities
can be used. The circuit shown will work well with a movement
of 500-microampere sensitivity, although a meter having a greater
sensitivity (such as a 200-microampere type) should provide better
performance because it will permit operation over a smaller por
tion of the tube’s characteristic curve and make for increased
linearity and accuracy.
The voltage divider network, consisting of four half-watt carbon
Circuit operation
Operation centers around the vacuum-tube bridge circuit, which
is similar to those discussed in Chapter 2. The balance control
provides a means of adjusting the two halves of the bridge so that
188
the voltages appearing at the cathodes are equal with no signal
applied. A change in control grid voltage at pin 4 will upset the
balance of the bridge, causing current to flow through the meter
movement. The amount of meter-pointer deflection depends upon
the amount of current flow. The 6SN7-GT has sufficiently straight
characteristic curves to produce linear voltage readings over the
instrument’s basic operating range from — 1.5 volts to -j-1.5 volts,
as measured at grid pin 4.
Test probe
The d.c.-isolating probe can be constructed from any suitable
plastic tubing and an insulated or coaxial lead. A 1.2-megohm
isolating resistor should be connected in series with the probe tip
to reduce the capacitance loading of the cable and instrument on
Calibration
Before attempting calibration, the builder should disassemble
the meter and remove the face plate. A new face plate containing
189
blank scales drawn in ink or pencil can be pasted over the old
scales. Calibration consists simply of making d.c. voltage readings
with another v.t.v.m. of known accuracy and transcribing the read
ings to the points indicated on the blank scale when the home-
constructed instrument is connected to the same voltage source.
The calibration control should be adjusted to provide for most
accurate readings at the full-scale points.
Checking approximate stage gain in audio amplifiers
Remove or disable one of the i.f.-amplifier tubes so that external
signals cannot reach the audio. Set up the v.t.v.m. to measure a.c.
voltage on a low range (see Fig. 1029), and adjust the signal gen
erator to deliver an audio output signal.
L Connect the output cable of the signal generator to the grid
of the output tube, shown as point 1 in Fig. 1029. Connect the
v.t.v.m. to the same point. Adjust the output of the signal genera
tor to produce 1.0 r.m.s. volt of signal at the control grid. Shift
the v.t.v.m. probe to the plate of the stage and measure the audio
signal voltage at point 2. The approximate voltage gain of the
stage is equal to the numerical value of the signal voltage at the
plate. For example, with 1.0 volt of signal at the grid and 15 volts
of signal at the plate, the approximate gain is equal to 15/1 or 15.
In a.c.-d.c. radio receivers and in some small a.c. receivers, con
siderable hum, up to 10 or 15 volts, may be present at the plate of
the output tube. To minimize the error that this can cause, it is
necessary to measure the hum voltage separately at the plate and
subtract the hum voltage from the previous reading. The hum
voltage should be measured without application of an audio signal.
When checking the gain of an audio-output stage, it is essential
to have a speaker connected to the output transformer because the
gain of the output stage without a load is much higher than
normal.
M-
Fig. 1031-a, -b, -c. The v.t.v.m. can be used to check the output of phono
graph cartridges. Use a constant amplitude test record for making these
measurements.
of the most common types are shown in Fig. 1031. In all three
circuits, the v.t.v.m. is set up for a.c. measurements on a low
a.c. scale and the a.c. probe connected to test points A. The v.t.v.m.
ground lead connects to the" ground side of the potentiometer.
In Fig. 1031-a, a crystal cartridge is connected to the arm of the
input volume control. A maximum voltage reading will be ob
tained when the arm is at the top (maximum volume position) of
the resistor. In this position, the signal is fed directly to the control
grid of the amplifier.
Two input circuits for magnetic cartridges are shown in
Fig. 1031-b and Fig. 1031-c. A matching transformer, which steps
up the output voltage from the magnetic pickup is used in
Fig. 1031-c. In all three of these circuits, an input signal from a
193
phonograph record is required. If a banded-tone test record is
available, the frequency response of the cartridge can be deter
mined.
In making these tests of pickup units, the most convenient
point at which to clip the test lead is the control grid of the pre
amplifier tube (point A in Fig. 1031). Another test spot is the lead
connected to the moving arm of the potentiometer. Do not clip
the probe directly to the pickup unit itself or to the wire leading
from the pickup, since this could upset the weight and balance of
the tone arm. The lead could damage the test record. You can also
use this method for checking the flatness of output of the record.
Although we have given various miscellaneous applications in
this chapter, the ideas suggested here are not a complete list by
any means. The value of the v.t.v.m. is limited only by the re
sourcefulness and ingenuity of the user. In the hands of a skilled
technician this instrument can do many jobs over and beyond
those originally intended by the manufacturer of the v.t.v.m.
194
troubleshooting the vtv.m.
195
bile, the v.t.v.m. should be inspected and serviced at frequent
intervals to insure continued reliable operation. There are a num
ber of routine checks and adjustments which can be made periodi
cally to keep the v.t.v.m. in top condition. Such checks and adjust
ments are classified as preventive maintenance.
Preventive maintenance
Improper use of the v.t.v.m. is a major cause of failure. Misuse
can result in costly repairs. Some of the most frequent mistakes
made by the technician include improper setting of the function
control or the application of too much voltage to the input cir
cuits. For example, many v.t.v.m.’s are returned to their manufac
turer for repairs because the low-value precision resistors in the
ohms network have been damaged or burned out due to the fact
that the technician has set up the v.t.v.m. for ohms measurement
and applied a high d.c. or a.c. voltage to the instrument. Since the
ohms network uses precision resistors having a very low wattage
rating, misuse can easily result in damage. Immediate repair of
such damage is not always possible because the resistors are usually
made specially for the instrument in which they are used. If re
placement parts cannot be obtained locally, the technician must
obtain these from the instrument manufacturer.
Repairs of this nature are often expensive and require two to
three weeks “down” time for the instrument. A little additional
care and double checking by the technician when he is using the
instrument can prevent this type of trouble.
Improper connection of probes and cables, and failure to pro
vide transformer isolation of other equipment from the power
line in a.c.-d.c. or transformerless applications can also cause
similar costly repairs. Damage resulting from such misapplication
can be prevented by use of an isolation transformer, as described
on page 76.
Probes and cables are a frequent source of trouble because they
receive a great deal of physical handling. The constant flexing and
bending of cables causes the wires to break or mechanical connec
tions to loosen. Detachable probes may be lost or damaged if a
heavy chassis is set down on them or if they are dropped to the
floor. This frequent source of trouble can be eliminated by clamp
ing the probe to the instrument case when the probe is not being
used. Insulation of probes and cables can also be damaged if it
comes into contact with a hot soldering iron. Out-of-the way place
ment of the soldering-iron stand can prevent this.
196
Rough handling or dropping of the instrument is a frequent
source of trouble and invariably results in damage to the meter
movement. Meter bearings and coil mountings are relatively deli
cate and, while the v.tv.m. is a sturdy instrument, it should be
remembered that the movement can be damaged by heavy jars or
shocks. The shop v.t.v.m. should be firmly mounted in a suitable
place on the bench or shelf where it cannot be moved or pulled
onto the floor when cables are handled. Even the best constructed
meter movements cannot be expected to withstand a four-foot
drop.
While the preventive measures described here appear to be
obvious, they are, nevertheless, frequent sources of trouble. These
measures are concerned primarily with the everyday use of the
v.t.v.m. Below are listed some of the instrument checks and peri
odic servicing tips which can extend the trouble-free life of the
instrument.
Ohms battery
Failure of the ohms-measurement function can be prevented by
periodic inspection or replacement of the ohms battery. The con
dition of the battery may be checked as follows:
I. Set function switch to “ohms.”
2. Set range control to “R X 1” position.
3. Rotate the ohms adjust control to bring the pointer to
full-scale deflection. If it is impossible to adjust the pointer to full
scale, the battery should be replaced.
4. Short the ohms cable to the ground cable for about 10 sec
onds.
5. Open the short circuit and observe the scale indication. Any
appreciable deviation of the pointer from full-scale deflection
indicates a weak battery.
Exhausted batteries should not be left in the instrument. If
inspection at any time shows that the battery is leaking or corrod
ed, it should be replaced at once and any chemical deposits or
corrosion on the battery contacts or inside the case should be
removed. When replacing the ohms battery in a holder with spring
contacts, the contacts should be sanded clean and the tension
adjusted to insure solid contact and as low a contact resistance as
possible. In some instruments the internal ohms connections are
soldered directly to the battery terminals. It is important, also
that correct polarity be observed. In some v.t.v.m.’s, the positive
197
contact is grounded. In other instruments, the negative contact is
grounded.
General troubleshooting
Of necessity a v.t.v.m. contains electronic and mechanically
operated components. Mechanical parts, such as function and
range switches, are in constant use, often require repair or re
placement.
Switches
The apparently complicated construction of switches used in
the v.t.v.m. is a big source of difficulty to many technicians. Indeed,
the switches and their connections are the most difficult aspect of
troubleshooting the v.t.v.m. If the switches could be removed, the
v.t.v.m. would be an extremely simple instrument. With a little
patience and careful circuit tracing, however, the technician
should be able to follow through and understand the switching
circuits without difficulty.
Three symbolic methods of drawing rotary switches are shown in
Fig. 1101. These types are widely used for the function and range
switches in schematic diagrams of v.t.v.m.’s. These drawings show
how contacts, permanent connections, and wiper arms are sym
bolized. In Fig. 1101-b the numbered terminals are used to con
nect to seven different points of a resistance network in a v.t.v.m.
which has seven measurement ranges. Contact 1 corresponds to the
first or left-hand position of the range control. The unnumbered
terminal is the wiper contact. In nearly all v.t.v.m. schematics, the
switches are shown in their maximum counterclockwise positions.
A five-position switch of similar construction is shown symboli
cally in Fig. 1101-c. In this illustration, the wiper arms are shown
in position 1 (the first range position). As the switch is rotated to
the right, the wiper arm will move up to positions 2, 3, 4, and 5.
Ganging of the switch sections is shown by the dotted line. Ganged
switch sections may also be identified by symbol numbers, such
as SI-a and Sl-b.
As is necessary in repairing and maintaining TV tuners, the
v.t.v.m. switches should be checked and attended to periodically
to keep them in good working order. This is an important preven
tive measure because switch replacements are costly and are best
done by the factory. Panel nuts on switch shafts should be kept
tight to prevent movement of the switch assembly, broken leads
198
and shorted components. Because many of the precision resistors
used in the v.t.v.m. are mounted directly on the switch assemblies,
care should be taken when soldering to prevent overheating or
mechanically scraping the resistors. Excessive heating may perma
nently change the value of the resistors and result in inaccurate
meter readings. A pair of long-nose pliers may be clamped to the
lead between the resistor being soldered and the point of connec
tion at the switch to dissipate as much heat as possible. Care should
also be taken to prevent twisting or loosening of the switch con-
5ECTÎON SECTION
Sl-a S-b
05 o5
04 o4
JeO-WIFERARM
03 o3
SWITCH CONTACT 02 02
PERMANENT CONNHITION
WÉR ARMS GANGED
A £
Fig. 1101-a, -b, -c. Rotary switches used in v.t.v.m’s can be repre
sented in these three different ways in circuit diagrams.
tacts and to prevent solder from getting down into the switch.
Contacts should be cleaned periodically with a good contact
cleaner.
Replacement parts
Whenever possible parts should be obtained from the instru
ment manufacturer to insure exact replacement. Individual com
ponents for meter movements are often not stocked, however, and
the meter movement must be repaired by a shop which specializes
in such work.
When replacing selenium rectifiers in the v.t.v.m. or when
replacing crystal diodes in high-frequency probes, be sure to
observe polarity. Incorrect installation of these parts may result in
damage or improper circuit operation.
Tube replacement
Most v.t.v.m.’s are supplied with tubes which have been specially
aged at the factory before installation in the instrument. In some
instruments it is necessary to select tubes. Tubes are picked or
199
aged to insure a low value of gas and contact-potential current.
While the tube replacement requirements may differ somewhat
between instruments of different manufacturers, it is usually desir
able to age replacement tubes for bridge circuit and a.c.-signal
rectifier applications.
Proper selection and aging of the bridge tube is usually required
in v.t.v.m.’s which utilize bridge circuits of the types described in
Chapter 2. Unless the tube has suitable operating characteristics
and a low value of gas current, it may cause off-zero deflection of
the meter pointer when the range switch is changed from the low
est d.c.-voltage range to the higher ranges. This condition requires
that the operator rezero the meter pointer whenever ranges are
changed. Often, a tube will prove unsuitable for installation, even
after a lengthy aging period. The acceptability of a tube for bridge
circuit use will depend upon the design of the v.t.v.m. and the
manufacturer’s recommendations. In most cases, however, a change
of one or more divisions in the meter-zero setting as the range
switch is changed indicates that the tube requires more aging or
that the tube is unacceptable for use in the v.t.v.m. If a number
of new tubes are available, they should be installed in the instru
ment, one at a time, before aging, to determine which tube has the
lowest value of gas. The tube which produces the least amount
of off-zero deflection as the range is changed is the best tube to
select for aging.
Single or twin diodes used in the a.c.-signal rectifying section
may or may not require selection or aging before installation. As
requirements for replacement will depend upon circuit design
and the particular v.t.v.m. involved, the manufacturer’s recom
mendations should be followed. In general, however, the tube
selected should provide a minimum amount of off-zero deflection
of the meter pointer when the function switch is changed from
d.c. volts or ohms to a.c. volts as well as a minimum amount of
deflection as the range switch is rotated. Internal adjustments are
provided on many instruments, however, to compensate for differ
ences between various signal-rectifier tubes, hence tube selection
and aging may not be critical.
A new tube may be aged by installing it in the v.t.v.m. and run
ning the instrument for a number of hours or days. An alternate
and ordinarily quicker method of tube aging makes use of a simple
external setup such as that shown in Figure 1102. This setup may
be used for either triodes or diodes. In the case of triodes, the grids
and cathodes are tied to one side of the a.c. line and the plates are
200
tied to the other side of the line. Filaments should be operated at
a normal filament voltage supplied by a separate transformer.
Diodes can be aged in the same way, except that no grid connec
tions are required.
The aging period required is from 24 to 100 hours, depending
upon the type of tube and instrument involved. Aging of tubes by
this method usually requires a shorter length of time than when
the tube is aged in the instrument.
Whenever a new tube is installed in the v.t.v.m., the instrument
should be checked for calibration and accuracy. In some instru-
Measurement accuracy
The degree of accuracy or inaccuracy with which a v.t.v.m. will
provide a voltage reading is usually expressed in percent of full-
scale reading. A typical figure is ±3% for d.c.-voltage measure
ments. The error is introduced primarily by two factors: (1) error
in the meter movement and (2) variance in the values of the pre
cision resistors used in the divider networks. These errors may be
cumulative. For example, if the meter movement causes a reading
which is low by two percent and the resistor network introduces
an additional error of minus one percent, the v.t.v.m. will indicate
a value which is three percent lower than the actual value of the
test voltage. It is also possible for the tolerance of the meter move
ment to be off in one direction and for the error in resistance value
to be off in the opposite direction so that the total measurement
error is only 1%. The error can be high or low.
Before the technician decides that his v.t.v.m. is providing scale
indications which are out of tolerance with the manufacturer’s
201
specifications, however, he should make sure that the standard
which he is using is accurate. It is not sufficient to make a com
parison with the readings of other voltmeters. The v.t.v.m. read
ings should be checked against those provided by a laboratory
standard or against a calibrating voltage which is carefully regu
lated.
When it is suspected that changed-value divider resistors are
causing incorrect readings, these components should be checked
on an accurately calibrated resistance bridge. Because many resis
tors used in commercial v.t.v.m.’s are designed to be accurate to
three and four numerical places, conventional resistance-measur
ing meters cannot be depended upon to measure their values
accurately.
Factory-fresh flashlight cells are a fair substitute for a voltage
standard for checking measurement accuracy on d.c.-voltage
ranges. When new, a flashlight cell should provide a voltage of
1.55 volts under the high-resistance load of the v.t.v.m. Two of the
cells can be connected in series to make a three-volt test source.
Fig. 1103. This block diagram represents the circuit arrangement in many
v.t.v.m’s.
203
tracing out the basic circuits on the schematic diagram provided
in the instruction manual. Representative block diagrams of two
v.t.v.m.’s are shown in Fig. 1103 and Fig. 1104. Different commer
cial instruments will present numerous variations in design from
the layouts shown here, but the circuit switching and relationships
of the voltage-dividers and electron-tube circuits are similar.
Fig. 1103 shows a simplified and straightforward layout common
to many commercial v.tv.m.’s. This circuit employs three different
voltage dividers for d.c. volts, a.c. volts, and ohms measurements.
Analysis of the diagram will show that only one of these divider
circuits is operative at any one time. When the function switch
is set to plus or minus d.c. volts, for example, the ohms and a.c.
networks are disconnected from the circuit. The bridge circuit,
power supply, and indicating meter, however, are employed on all
three functions.
The technician should have little difficulty in localizing trouble
in an instrument of this design. For example, if the v.t.v.m. fails
to indicate on any function, the trouble must lie in the three sec
tions common to all functions — the bridge circuit, power supply,
or meter. Faulty operation on ohms measurements only would pin
point the source of trouble to the ohms divider network, the ohms
probe and cable, or the associated circuitry. The preliminary
checks described here are important to localizing and detecting
the source of trouble.
When troubleshooting the defective v.t.v.m., the technician
should not overlook the probes and cables as a source of trouble.
These accessories should be inspected carefully during the prelim
inary tests and moved and flexed about during each step of the
inspection to make sure no intermittent shorts or opens are re
sponsible.
A v.t.v.m. of somewhat different design is shown in block-dia
gram form in Fig. 1104. This arrangement is similar to the design
employed in the RCA WV-97A Senior VoltOhmyst and other
commercial instruments. As is common with many v.t.v.m.’s, a
single input circuit is used for both a.c. and d.c.-voltage measure
ments and a single test cable is used for both functions. The input
is switched inside the v.t.v.m. to channel the test voltage to the
proper circuit.
The probe used with this type of v.t.v.m. may be a switch type,
such as that described in Chapter 4, to permit removal or addition
of an isolating resistor for d.c.-voltage measurements. The instru
ment of Fig. 1104 employs a common voltage divider for both a.c.
204
and d.c. voltage measurements. On d.c. volts, the input signal is
fed directly to the voltage divider and then to the bridge circuit.
On a.c. volts, the signal is first rectified by the a.c.-signal rectifier
and then fed to the voltage divider as a d.c. voltage. A defect in the
a.c.-d.c. voltage-divider network will affect operation of the v.t.v.m.
on both these voltage functions. Improper operation on a.c.-voltage
measurements only would point to the a.c.-signal rectifier section
as a source of trouble.
The a.c. signal section incorporates a special high-voltage divider
network between the a.c.-voltage terminal and the signal rectifier
section. This network usually consists of two or three precision
Fig. 1104. Block diagram of the various circuits of a v.t.v.m. This instrument
uses a contact-potential compensating circuit for a.c. voltage measurements.
205
A complete schematic diagram for a v.t.v.m. which employs this
arrangement is shown in Fig. 1105.
206
Fig. 1105) is available for the measurement of high frequencies
up to 100 me.
By carefully tracing the switching arrangements used in the 303,
200HA/650a APPROX
Fig. 1107. A.c. voltage section of the v.t.v.m. The diode rectifies the a.c., passes
it on to the triode in the form of a d.c. voltage.
208
tion of the 6AL5 must be overcome. This is accomplished in the
303 by using one diode section of the 6AL5 in a bridge arrange
ment arid balancing the circuit by means of R31. A.c. calibration
is effected by means of R32. Both R31 and R32 are adjusted at the
factory.
Fig. 1108 shows the circuit arrangement used when the selector
switch is set to ohms. A 1.5-volt battery supplies the test voltage
for ohms measurements and is sufficient to cause full-scale deflec
tion of the meter pointer when the test leads are open. Exact
full-scale setting of the meter pointer is effected by adjustment of
R35, the front-panel ohms-adjust control. When the resistance
Fig. 1109. Complete circuit diagram of the vacuum-tube voltmeter described in the text. The troubleshooting procedure for this instru
ment can be applied to any v.t.v.m.
is shown in Fig. 1109. This instrument will measure resistance
values up to 1,000 megohms, d.c. and a.c. (r.m.s.) voltages up to
1,500 in seven ranges, and direct current from 10 microamperes
to 15 amperes in nine ranges. A simplified block diagram is shown
in Fig. 1110. Examination of the block diagram shows that separ
ate resistance dividers are used for each of the major functions
Fig. 1110. Complete block diagram of the v.t.v.m. illustrated in Fig. 501. This
instrument has provisions for measuring direct current, up to a maximum of
15 amperes.
and that the vacuum-tube bridge circuit is not used when the
v.t.v.m. is employed to measure current. In block-diagram form,
the instrument appears in its simplest form and the relationships
between the different circuits and functions is readily apparent.
A front-panel view of the Master VoltOhmyst is illustrated in
Fig. 501. Two interior views are shown in Figs. 1111 and 1112.
To gain some experience with circuit tracing and to acquaint
ourselves with some of the problems and techniques encountered
in troubleshooting any defective v.t.v.m., let us anticipate a variety
of possible troubles and, from examination of the block and sche
211
matic diagrams, list the possible causes. Because v.t.v.m. s have
so much in common in their design and operation, we should be
able to apply much of the following discussion to other instru
ments.
(Note: At this point, the technician might wish to list a number
of operating defects and, with the aid of the schematic diagram,
list for himself some of the possible causes. When the list has been
completed, it can be checked against the possible causes listed
below.) Components and tubes, described in the list of v.t.v.m.
troubles on this and following pages, are illustrated in Figs. 1109,
1111, and 1112.
General Troubles
Symptom
1. Instrument fails to operate on all functions except current;
tubes do not light.
Possible Causes
a. Line cord broken or not making contact at outlet.
b. Switch section S2F defective.
c. T1 defective. Check continuity of windings and resistance to
ground.
d. Fuse Fl open.
Symptom
2. Instrument fails to operate on all functions except current;
tubes light, zero adj does not work.
Possible Causes
a. Broken lead in ground cable or J2 defective.
b. Incorrect operating voltages on bridge tube. Measure volt
ages with another voltmeter.
c. V2 or C3 defective.
d. Open or short circuit in V3 bridge circuit. Check out with
ohmmeter. Check R26, C5, T1 secondary (plate winding),
and meter leads.
e. Switch section S2B, S2C, or S2D defective.
Symptom
3. Meter pointer sticks.
Possible Causes
a. Cracked or broken jewel bearing in meter.
212
b. Scale plate of meter is loose and is striking meter pointer
(remove meter from case and tighten screws).
Symptom
4. Meter pointer bangs hard left or right, depending upon set
ting Of FUNCTION Switch.
Possible Causes
a. Open or short circuit in bridge circuit.
Fig. 1111. View of the interior of a v.t.v.m. Voltage and current cali
bration adjustments are easily available, can easily be made with a
screwdriver. The battery (used for ohms measurements) is clamped
into place, but a new unit can be readily substituted.
Symptom
5. Intermittent operation on all functions except current.
Possible Causes
a. Loose or broken connection in ground cable. (Check by flex
ing cable and observing meter readings.)
b. Wiper contact on S2B, S2C, or S2D defective.
c. Loose or broken connection in bridge circuit wiring. With
213
power applied, check wiring, connections, and components
with insulated probe.
d. V2 or V3 defective.
Ohms-Function Troubles
Symptom
1. Instrument fails to operate on ohms; works normally on a.c.-
and D.c.-voltage measurements.
Possible Causes
a. Battery exhausted or not making contact.
b. Loose or broken connection in ohms cable. Check by flexing
cable and observing meter readings.
c. Defective wiper contact on switch section SID.
d. Open circuit in resistor network, or burned out resistor
around switch section SID (ohms-divider section). Note: This
network is made up of resistors which are added in series as
the switch is rotated. Therefore, one faulty resistor may cause
improper operation on one or more ranges.
e. Ohms jack, JI, or its attaching lead is open or shorted to other
wiring or to instrument case. Check resistance to case and
adjacent wiring.
f. Loose or broken “ohms” contact on switch section S2C (ter
minal 12) or S2D (terminal 6).
g. Cable connector not making contact.
Symptoms
2. Ohms adj control fails to give infinity setting on resistance
scale.
Possible Causes
a. Ohms battery exhausted.
b. Ohms adjustment potentiometer, RI 6, defective.
c. V3 defective.
Symptoms
3. Readings inaccurate on some or all resistance ranges.
Possible Causes
a. One or more resistors in network around switch section SID
have changed value.
b. Excessive leakage in “ohms” circuit. Check switch wafers on
214
SID and S2B for dirt or damage. Note: High humidity may
cause leakage and inaccurate readings on high ohms scales.
Bake out inside of instrument with low-wattage light bulb,
c. Poor contact from ground jack to ground bus of chassis,
d. Shorted wiring in resistor network around SID.
e. Faulty or high-resistance connections in “ohms” circuit.
R2
Fig. 1112. Underchassis view of a v.t.v.m. Note the way in which parts
are mounted on the wafer switches. Components with code numbers
can be identified by reference to the circuit diagram in Fig. 1109 on
page 210.
Symptoms
4. Meter pointer moves off infinity mark when range switch is
changed.
Possible Causes
a. Replace V3 with 12AU7 having a low value of gas current.
Note: A small deviation should be considered acceptable.
b. Excessive leakage in “ohms” circuit, possibly caused by high
humidity. Check by baking out for several hours with light
bulb. Do not overheat and damage capacitors.
215
A.C.-Voltage Measurement Troubles
Symptoms
1. Instrument fails to operate on any A.c.-voltage range; works
normally on ohms and ‘d.c. volts.
Possible Causes
a. VI defective.
b. A.c. volts contact on S2A, S2B, S2C or S2D defective.
c. Faulty wiper contact on SIA.
d. Open or short circuit in circuitry associated with VI. Check
out wiring and components with ohmmeter.
e. Cl or C3 open, C6 shorted, or R34 open.
Symptoms
2. Meter pointer moves off zero when ranges are changed.
Possible Causes
a. A.c. zero potentiometer out of adjustment. If adjustment
fails to correct trouble, then do steps b and c.
b. Interchange VI and V2 or replace one or both tubes.
c. Resistor network around SIE defective. Check continuity and
values with ohmmeter.
Symptoms
3. A.c.-voltage readings inaccurate on some or all a.c. ranges;
performance on ohms and D.c.-voltage ranges is normal.
Possible Causes
a. Resistor network around switch SIC defective. Check values
with ohmmeter.
b. Defective contacts or wipers or excessive leakage in SIC.
Check for loose or dirty contacts.
c. VI defective.
d. C6 leaky.
e. R34 changed in value.
Note: If instrument is inaccurate on 500- and 1,500-volt
ranges only, check RIO, Rll, R12, and C2. If inaccurate on
1.5-volt range only, replace VI.
D.C.-Voltage Measurement Troubles
Symptoms
1. Instrument fails to operate on any range of +volts or
—volts; works normally on ohms and a.c. volts.
216
Possible Causes
a. D.c. (isolating) probe is defective. Try applying low d.c.-
voltage directly to direct-probe tip.
b. D.c.-voltage contacts on switch sections S2A, S2B, S2C, or
S2D defective. Check for loose or broken contacts.
Symptoms
2. Instrument fails to operate on any range of ¿-volts but nor
mal on —volts or vice versa; works normally on ohms and
A.C. VOLTS.
Possible Causes
a. Potentiometer RI5 or RI3 defective. Check out with ohm
meter. Check R15 when ¿-volts is defective; check R13 when
—volts is defective.
b. Defective contacts on switch sections S2A, S2B, S2C, or S2D.
Check switches for loose or broken contacts.
Symptoms
3. Voltage readings inaccurate on ¿-volts, —volts, or both; a.c.
voltage and resistance readings are correct.
Possible Causes
a. RI3 and RI5 out of adjustment.
b. Isolating resistor in d.c. (isolating) probe has changed value.
Measure resistance on ohms function. Note: If one or more
of the low-value resistors in the network around SIB have
changed value, it is possible that the meter reading may be
inaccurate on the 500- and 1,500-volt ranges but the inaccur
acy may not be apparent on the lower ranges.
Symptoms
4. D.c.-voltage readings incorrect; resistance readings correct.
Possible Causes
a. Resistance in network around SIB has changed value. Check
out with ohmmeter and replace defective resistors.
b. Switch contacts on SIB or S2A defective. Check for loose or
broken contacts or foreign material.
Symptoms
5. Instrument fails to function on D.c.-voltage measurements;
works normally on ohms function.
217
Possible Causes
a. Check d.c. (isolating) probe and cable and connectors.
b. SIB network open.
Symptoms
6. -pD.c. and a.c. voltage readings incorrect; —d.c. normal.
Possible Causes
a. Recalibrate RI5.
b. RI5 defective.
Current-Measurement Troubles
Symptoms
1. Instrument fails to operate on any current range. A.c. and
D.c. volts and ohms functions normal.
Possible Causes
a. Switch contact on S2C (terminal 2) or S2D (terminal 8)
defective.
Symptoms
2. Instrument fails to operate on -|-ma current range; .+ I-5a
and +15a ranges normal.
Possible Causes
a. Faulty wiper contact on SIF.
b. Check +ma terminal for broken lead.
Symptoms
3. Instrument operates only on low-current ranges, or no opera
tion on any current range; a.c. and d.c. volts and ohms
functions normal.
Possible Causes
a. Resistor network RI through R9 defective. Check out with
ohmmeter.
b. Switch contact on S1F, S2C, or S2D defective.
218
index
219
Protective--------------------------------- 148 Contrast, Excessive Picture------------ 91
Supply, External---------------------- 92 Contrast, Insufficient Picture --------- 91
Breakdown of V.T.V.M.-------------------- 195 Control:
Bridge: Circuit----------------------------------------- 65
Balancing the------------------------------- 19 Function-------------------------------- 65, 198
Circuit----------- ------------------------16, 17 Ohms Adjust------------------------------ 65
Circuit, Cathode-Metered --------- 17 Range------------------------------------ 65, 198
Circuit, Open----------------- 212 Selector-------------------------------------- 65
Circuit Operation, Cathode- Zero-Adjust--------------------------- 19, 43
Metered ------------------- 20 Controls, Bass and Treble--------------- 134
Circuit, Plate-Metered ---------------- 17 Controls, Front-Panel-------- ------------- 65
Wheatstone-------------------------------- 167 Conversion Scales, DB-------------------- 174
Buffer Amplifiers----------------------------- 147 Conversion Table for Sine Waves — 26
Burnout, Meter--------------------------------- 127 Crossover Frequencies---------------------- 135
Burnout-Resistant Meter Circuit — 11 Crystal:
Buzz in Sound - ------------------------------------91 Diode Rectifier Probes —49, 51, 52
Oscillator __ ___________________ 144
Pickup, Checking ------- -------------- 193
Cable, Checking Coaxial------------------ 170 Curing Hum------------------------------------- 129
Cable Plug Tips------------------------------ 185 Current:
Cables 45, 46 Gas, in V.T.V.M. Tube - ------------ 200
Cables, Failure of--------------------------- 196 Grid____________________________ 148
Calibrating an Oscilloscope------------ 163 Measurements___ _______________ 145
Calibrating Voltage Output of Measurements, Meter Burnout in 182
Generators . ------------ ---------------- 157 Measurements, Modifying
Calibration of V.T.V.M. Preamplifier 180 V.T.V.M. for----------------------- 182
Cap, Meter —----------------------------------- 202 Measurements, Troubles in_____ 218
Capacitance, Excessive in Probe------ 186 Measuring Circuit . -------- 34
Capacitor Leakage -------------------------- 113 Readings with the V.T.V.M.____ 183
Capacitors, Checking----------------------- 83 Cut Rotor Plates ------------------------------- 118
Capacitors, Leaky Coupling------------ 105
Cartridge, Checking Voltage Output DB:
of---- - -------------- -------- 193 Conversion Scales------------------- -- 174
Cathode Bias, Measuring---------------- 148 Measurements ------------------------- 136
Cathode Follower Impedance - --------- 132 Readings with the V.T.V.M.___ 174
Cathode-Metered Bridge Circuit Scales 175
Operation----------------------------------- 20 DBM 137
Causes of Hum-------------------------------- 129 DBM vs. R.M.S. Voltages 136
Cells, Checking Bias------------------------ 160 D.C.:
Characteristics, Instrument------------ 7 Input to V.T.V.M. 21
Checking: Isolating Probe, Making a 189
Battery Voltages----------——- 159 Plate Voltage, Measuring 148
Bias Cells--------- —------------------- 160 Probe ................ ....-------------- 48
Coaxial Cable--------------------------- 170 Screen Voltage, Measuring___ 148
Current-------------------------------------- 182 Voltage-Divider Network 20
Defective Picture Tubes---------- 99 Voltage Measurement Troubles . 216
Frequency Response------------------ 133 Voltage Measurements ___ 47, 70
Gassy Tubes------------------------------ 162 Voltage Measuring Section of
Impedance--------------------------------- 171 the V.T.V.M. 207
Inductance-------------------------------- 172 Decibel:
Local Oscillator------------------------ 111 Measurements 136
Ohms Battery--------------------------- 197 Readings with the V.T.V.M........ — 174
Phonograph Equalizers------------ 135 Scales ........ —.................... 175
Pickups - ------------------------------------- 193 Defective Selenium Rectifier,
Potentiometers —--------------------- 162 Replacing_______ ________ ______ 169
Power Supplies-------------------------- 112 Deflection Circuits, Troubleshooting 100
Power Transformers---------------- 172 Deflection, Off-Zero __ _ ___ 200
Precision Resistors-------------------- 168 Deflection Sensitivity of Scope_____ 163
Q of Coils--------------------------------- 173 Demodulator Probe________ -________ 49
Reactance------ --------------------------- 171 Detail, Soft Picture _______ —.___ 101
R.F. Amplifier Gain------------------ 176 Detector:
Selenium Rectifier Circuits ------- 168 Alignment of Ratio 122
Stage Gain in Audio Amplifiers 190 Balanced Ratio 122
Transformers _--------------------------- 160 Circuit. Ratio_________ _______ „__ 121
Transmission Lines--------------------170 Discriminator 121
TV Antennas-------------------------- — 170 Probes 49
Voltage Output of Phonograph Ratio 122
Cartridges------ ---------------------- 193 Unbalanced Ratio 122
Voltage Regulator Circuits_____155 Dial Out of Calibration----------- .------- 115
Circuit: Direct Current Measurements __ 33, 72
Arrangement of V.T.V.M. ---------- 203 Direct Probes 46
Control 65 Discriminator Detectors 121
Loading ---------------------------,--------- 8 Divider Probes 56
Coaxial Cable, Checking 170 Divisions, Scale 39
Coils, Checking_______________________ 89 Doubler, Full-Wave Voltage 113
Coils, Checking Q of 173 Dress, Lead 163
Combined Probe ------------------------- 49, 50 Dress, Poor Lead 129
Commercial V.T.V.M/s, Analysis of 206 Dummy Load 131
Component Failures..................103, 107
Construction of Preamplifier for the
V.T.V.M. 179 Eliminating Hum — -------------- ------ 128
Contact Potential----- _-------- -------30, 205 Equalizers, Checking Phonograph 135
Contact Potential, Correcting for Error, Tracking w _76
Effects of 190 Experimental V.T.V.M______ .____— 186
220
Factor, Multiplier 38 Instrument Characteristics —_____ — 7
Failures, Component__________ ___ 107 Intercarrier TV, Test Points in____ 93
Field-Strength Meter____________ _ 154 Intermittent:
Filament Wiring Tests_____________ 104 Receiver Test Points 115
Filtering, Inadequate Power Supply 129 Sound 114
Fixed Bias_________________________ _ 148 Trouble _____________________ ___ 94
Flutter, Excessive Airplane 91 Intermittents __ _____________ __103, 114
FM: Intermittents in Antenna Systems _ 170
Alignment ._____________________ __ 103 Isolating:
Receiver Circuit 120 Effect of Probe 47
Receiver, Test Points in ______ 120 Probes 47, 184
Receivers 119 Resistor 47, 184
Repair------- --------------- ----------- 103 Isolation Transformer, Need for____ 109
Frequencies, Crossover_________ 135
Frequency Response: Kinks for the V.T.V.M. 184
Checks 133
Measurements 130
Of Scope 164 Lead Dress 163
Of V.T.V.M. __________________26, 27 Lead Dress, Poor 129
Vs. Output Voltage 136 Leakage:
Frequency Test Records 134 In Power Output Stage,
Front-Panel Controls 65 Heater-to-Cathode 129
Front-to-Back Resistance of Paths, Transformer 89
Selenium Rectifier 169 Resistance Test _90
Full-Wave Supply, Test Points in .. 114 Leaky Coupling Capacitors------------- 105
Full-Wave Voltage Doubler 113 Levels, Standard Reference------------139
Function Control-------------------------65, 198 Lever Adjustment, Meter 202
Limitations of the V.T.V.M.------- 62, 63
Gain, Checking R.F.Amplifier--------- 176 Linear Deflection of Scope 164
Gain, Voltage------------------------------------ 132 Linear Scales 41
Ganged Switches --------------------- 198 Load:
Gas Current in V.T.V.M.Tubes____ _ 200 Dummy 131
Gas-Tube Regulator------- ----------------- ' 156 Impedance, Voice-Coil--------- '— 131
Gassy Tubes - — ---------------------------- 105 Transformer Output 127
Gassy Tubes, Checking .. _— 162 Loading Effect, Meter 8
General Troubles in the V.T.V.M. _ 212 Local Oscillator, Checking........... — 111
Generators, Calibrating Voltage Localizing Trouble in the V.T.V.M. - 202
Output of---------------------------- 157 Low A.C. Voltage Scales --------- 42
Gliding Tone Records------------------------- 134 Low Circuit Loading in Voltage
Grid Bias, Checking 147 Measurements 8
Grid, Bias, Measuring----------------------- 148
Grid Current 148 Magnetic Pickups, Checking------------ 193
Grid-Leak Bias----- ---------------------------- 148 Maintaining V.T.V.M. Switches ------ 198
Grounding, Improper------------------------- 130 Maintenance of the V.T.V.M. —_— 196
Measurement:
Half-Wave Supply, Test Points in — 111 A Pr*TIT'S! (*V 201
Ham Shack. V.T.V.M. in the 141 Of Low A.C. Voltages, Test
Heater-to-Cathode Leakage 113, 129 Setup for ----------------------- 132
Heater Wiring Tests----------------------- 104 Of Power-Supply Ripple and
Height, Insufficient ------------------------ 101 Regulation ------------------- -------- 158
High-Fidelity Amplifier Circuit------ 124 Tips ..................................................... 74
High-Frequency Probes ------------- 49 Troubles, Current------------------- 218
High-Power Amplifiers, Adjustment Troubles, D.C. Voltage------------- 216
of _________________ 149 Measurements:
High Voltage: A.C. Voltage - --------------------- 25, 71
Circuit, TV______________ _______ 101 In Audio Amplifiers------------------ 125
Measuring —------------------------------- 152 Current ----------------------------------- 145
Precautions --------------- --------------- 142 D.C- Voltage________________ 47, 70
Probes ------ -------- - -------------------- 56 Decibel --------------------------------- 136
Rectifying Probe for A.C. Direct Current --------------------- 33, 72
Measurements-------------- ------- 186 Frequency Response ------------------ 130
Hints for the V.T.V.M.-------------------- 184 How to Take ------------------ 70
Hiss .................................. .............................. 115 Modifying the V.T.V.M. for
Horizontal Hold, Critical----------- — 91 Current ---------------- --------- 182
Horizontal Pulling---------------------------- 91 Of Screen Voltages in
Hum: Transmitters ——--------------- 147
Eliminating ___-------------------------- 128 Ohms---------------------- —— 71, 74
Excessive -------------------------------- 112 Voltage__________ ____ —75, 110
In Output Tube ------------------------ 190 Measuring:
Remedies for----------------------------- 129 Circuits. A.C.------------ —------------ 28
High Voltage--------------- —--------- 152
I.F. Response Curve------------------------ 97 Inductance------------------------- — 172
Impedance, Checking -------------------- 171 Power Output--------------------------- 192
Impedance of Power Transformers, 172
Checking - --------- :-----------------172 Unknown Impedance --------- 171
Improper Grounding -------- —— 130 Mechanical Zero Setting of the
Improper Use of the V.T.V.M. -- ------ 196 Meter---------------------- 202
Inadequate Shielding —------------- — 130 Meter:
Inductance, Measuring--------------------- 172 Burnout —--------------------------- 127
Infinity ........-.................................. Burnout in Current
42
Input and Output Impedances of Measurements--------------------- 182
Amplifier ____________ ____ _—— 135 Cap — ---------- .—.----------------- 202
Input Resistance of the V.T.V.M. — 21 Circuit, Burnout-Resistant —— 11
Input to V.T.V.M., D.C---------------- ---- 21 Cracked Bearing in-------------------- 212
221
Field-Strength 154 Picture Tube, Checking a Defective 99
Lever Adjustment----------------------- 202 Placements of Parts, Poor--------------- 130
Loading Effect -9 Plate Feed, Series and Parallel —— 146
Mechanical Zero Setting of------- 202 Plate Supply Voltage, Measuring — 148
Resistance---------------- 9 Plotting a Bandpass Curve -------------- 95
Screw Plug----------------------------------202 Plug Tips -------- 185
Shunts 183 Polarity of Ohms Battery----------------- 197
Misalignment___________________ 103, 115 Potential, Contact ------------------ 30, 205
Miscellaneous Applications--------------- 157 Potential, Contact, Correcting for
Misuse of the V.T.V.M___ 196 Effects of------------------------------ 190
Modifying the V.T.V.M. for Potentiometers:
Current Measurements 182, 183 Checking------------------------------------- 162
Modifying the V.T.V.M. Scales for Noisy _162
DB Readings 175 Taper of--------------------------------------- 163
Monitoring a TV Set 95 Power:
Mounting Stand for the V.T.V.M. _ 185 Input to Amplifier----------------------- 153
Multiplier: Output, Measuring---------------------- 192
Factor------------------------------------- 38 Output Stage, Heater-to-
Probes---------------- 56 Cathode Leakage in------------ 129
Resistor 57, 58 Power Supplies:
Stages in Transmitter---------------- 146 Checking ---------------------------------- 112
Multipliers------------------------------------------- 67 Checking Selenium Rectifier------ 168
Filtering Inadequate in-------------- 129
Neutralizing Indicator-----------------------153 Regulation of----------------------- 156, 159
Neutralizing Triodes------------------------ 154 Ripple in --------------------------- 158
Noisy Potentiometers----------------------- 162 Testing Voltage Doubler 169
Nonlinear Scales-------------------------41, 174 Tests of ---------------------------------- 88
Null Indication 73 Voltage Regulated----------------------- 156
Null Indications, Zero-Center---------167 Power Transformers, Checking_____ 172
Preamplifier for the V.T.V.M.--------- 177
Off-Zero Deflection--------------------------- 200 Preamplifier Voltage Gain,
Ohms: Measurement of ---- 132
Adjust Control----------------------------- 65 Precision Resistors, Checking Value
Battery--------------------------------------- 197 of ________________ 168
Battery Polarity-------------------------- 197 Precision Resistor, Using the
Battery, Replacing--------------------- 197 V.T.V.M. as a ............„.................185
Function Troubles---------------------- 214 Preventive Maintenance of the
Measurements 71, 74 V.T.V.M. 196
Scale 41 Probes:
Open Tubes 104 A.C. 49
Operating Bias--------------------------------- 148 Combined ------------------------------- 49, 50
Oscillator: Crystal Diode------------------------------- 49
Bias 145 Damage to --------------------------------- 196
Checking Local--------------------------- 111 D.C. 48
Crystal --- ---------------------------------144 Demodulator 49
Operating Voltages--------------------- 143 Design of 50
Stages, Checking ....--------------------- 143 Detector --------------------------------------- 49
Variable-Frequency 144 Direct ------------------------------------------ 46
Oscillators, Calibrating Voltage Divider ---------------------------------------- 56
Output of 157 Failure of ---------------------------------- 196
Oscillators, Testing of____ ___________ 144 High-Frequency ------------------------- 49
Oscilloscope: High-Voltage 56
Calibration of 163 Isolating ---------------------------- 47, 184
Frequency Response of 164 Isolating Effect of --------------------- 47
Voltmeter Scales for 166 Making D.C. Isolating 189
Output: Making Rectifying--------- ------------ 189
Measuring Power 192 Multiplier 56
Stage, Heater-to-Cathode Peak-to-Peak 55
Leakage in 129 Rectifying ------------ .--------------------- 49
Transformer Load 127 Resistance of_____ ._____________ 21
Tube Hum 190 R.F 49
Voltage vs. Frequency Response 136 Storage of ---------------------------------- 185
Output and Input Impedances, Switch Type-------- ----------------------- 49
Audio Amplifier 135 Vacuum Tube ------------------ 49, 53, 54
Overloading 91 Protective Bias ---------------------- 148
Push-Pull Transmitter Power
Parallel Plate Feed 146 Amplifier 151
Parasitics 146
Parts Placement, Poor______________ 130 Q Measurements 172
Parts, Replacement of 199
Peak-to-Peak: Range Control 65, 198
Probes 55 Range Switch-------- ----------------------- 38, 40
Rectifier ____ -________ ._________ 33 Ratio Detector:
Rectifier Circuit____„___________ 29 Balanced 122
Scales 42 Circuit 121
Value of Sine Wave 25 Unbalanced 122
Peak Value of Sine Wave__ „______ 25 Reactance, Checking_____ ._________ 171
Percentage of Regulation ___________ 159 APAiVAT* *
Phonograph Cartridges, Checking Alignment, AM _________ .____ 116, 117
Voltage Output of ________ 193 Circuit, FM 120
Phonograph Equalizer, Checking ___ 135 FM .................................... ................... 119
Pickups, Checking _______ __________ 193 Servicing the Shorted ........____ .... 191
Picture I.F. Amplifiers, Trouble Records:
shooting ------------------ 94 Banded Tone 134
222
Gliding Tone_______ ________ ____ 134 Selenium Rectifier Circuits,
Test Frequency _____________ 134 Checking _ _ 168
Rectifier: Selenium Rectifier, Replacing
Circuits, Checking Selenium ___ 168 Defective................. 169
Circuits, Peak-to-Peak_________ 29 Sensitivity of the V.T.V.M.__________ 10
Peak-to-Peak ___ 33 Series Plate Feed ___________ 146
Probes, Crystal-Diode . 51, 52 Servicing Audio Amplifiers ________ — 123
Replacing Defective Selenium __ 169 Servicing the ShortedReceiver_____ 191
Rectifying: Setup for Aging Tubes ____________ 201
Probe, Making a_______________ 189 Setup for AM Alignment, Test__117, 118
Probes . __ _ __________________ 49 Shielding ______ 78
Probes for High-Voltage A.C. Shielding, Inadequate ____________ 130
Measurements_____________ 186 Shock Hazard ____________________ 142
Reference Levels, Standard ______ 139 Shorted Receiver, Servicing the___ 191
Regulation, Percentage of___________ 159 Shunts, Meter_______________________ 183
Regulation, Power-Supply _____ 156, 158 Signal Tracing---------------------------------- 107
Regulator, Gas-Tube ________________ 156 Signal Tracing Audio Amplifiers__ 123
Repairs, AM_________________________ 103 Sine Waves, Conversion Table for__ 26
Repairs, FM--------------------------------------- 103 Slip on Probes_________ 46
Replacement of Tubes in the Slotted Rotor Plates ______________ 118
V.T.V.M. _________________ 199 Snow in Picture ---------——---------- 91
Replacement Parts for the Soft Picture Detail__________________ 101
V.T.V.M.___________________ 199 Sound I.F. Amplifiers, Trouble
Resistance: shooting ---------------- 94
Matching, Modifying the Sound, Intermittent_________________ 114
V.T.V.M. for 167 Spare Probes, Storage of-----------------185
Measurement Network _________ 23 Split-Sound TV, TestPoints in ______ 93
Measuring Circuit of the Stage Gain in Audio Amplifiers,
V.T.V.M.____________________ 209 Checking_ 190
Of Meter_____________ 9 Standard Reference Levels __________139
Of Transformer Windings_______ 160 Storage of Spare Probes ___ 185
Of V.T.V.M. Input_____ 21 Subdivision Marks on Scales------------- 39
Probe___ _____ 21 Substitution Tests, Capacitor------------ 86
Readings Over Wide Ranges___ 12 -Switch, Range------------------------ 38, 40
Resistor: Switch-Type Probe-------------------- 49
Isolating-------------------------------- 47, 184 Switches:
Multiplier_________________________ 57 Defective -------------------------- ----- — 212
V.T.V.M. as a Precision_________ 185 Ganged---------------------------- :---------- 198
Resistors, Checking_______________ 87 Rotary . _ — -------- ---------- 199
Response: Troubleshooting V.T.V.M.------ 198
Curve, I.F._______________________ 97 Switching Arrangements in the
Measurements, Frequency_______ 130 V.T.V.M. . _____ 207
Of V.T.V.M., Frequency__________ 27 Sync Circuits, Troubleshooting - ------- 100
R.F.: Sync Noise Immunity. Poor--------------- 91
Alignment----------------------------------- 115
Amplifier Circuits, Transmitter 147 Taper of Potentiometers------------------ 163
Amplifier Gain, Checking________ 176 Test Points:
Amplifier, Testing-------------------- 150 AM Receiver ----------------------------- 110
Output Signal Voltage, FM Receiver----------------------------- 120
Measuring----------------------- 148 Half-Wave Supply ----------------------111
Probes --------------------------------------- 49 Intermittent Receiver---------------- 115
Signal Voltage, Measuring---------- 148 Transformer Full-Wave Supply 114
Voltages, Measuring------------------ 52 Voltage-Doubler Supply------------- 169
Ripple, Power-Supply - ------------------ 158
Test Probe, Making a --------- 189
R.M.S. Value of Sine Wave - --------------25 Test Records, Frequency------------------ 134
R.M.S. Voltages vs. DBM -------------- 136 Test Setup for AM Alignment 117, 118
Rotary Switches in the V.T.V.M. _ 199 Test Setup for Measurement of
Rotor Plates, Cut and Slotted-----------118 Low A.C. Voltages---------- — 132
Testing:
Safety--------- 142 Amplifier Input and Output
Scales: Impedances ------------------------- 135
A.C., R.M.S.________ ___________ 42 Battery Voltages ----------------------- 159
A.C. Voltage_____________________42 Oscillators 144
DB Conversion__________________ 174 TV High-Voltage Circuits---------- 101
Division of ----------------------- 39 Time, Transit ------------------------------ — 51
For an Oscilloscope,Voltmeter _ 166 Tips, Plug----------------------------------------- 185
How to Read ------ 39 Tracing, Signal--------------------------------- 107
Linear and Nonlinear-------------------- 41 Tracking Error------- 76
Low A.C. Voltage----- w------------- 42 Transformer:
Modifying for DB Readings------ 175 Impedance Matching ------------ 161
Ohms --------- 41 Isolation, Need for--------------------- 109
Peak-to-Peak ----------------------------- 42 Load, Output ----------------------------- J27
Sharing of —-------- 38 Turns Ratio----------------------- —r 161
Subdivisions of___________________ 39 Transformers, Checking----------— 89, 160
Zero-Center ------ 43 Transit Time--------- - -------- --------------- 51
Scope (see also Oscilloscope): Transmission Lines, Checking--------- 170
Calibration of Transmitter:
163
Deflection Sensitivity------------------ 163 Amateur —----------------- —-------- —- 141
Frequency Response of--------------- 164 Amplifier Stages-------------------- 146
Linear Deflection of ——------ 164 Harmonics --------- —---------------146
Screen Voltage Measurements in Multiplier Stages — ------------------ - 146
Transmitter- 147 Trap Alignment----------------------------------8 J
Screw Plug, Meter ------ 202 Treble Control-------------------------------- J|4
Selector Control-------------------------------- 65 Triode Neutralization----------------------- 154
223
Trouble in the V.T.V.M., Analyzing Field-Strength Meter-------------------155
and Localizing 202 General Troubles in the------------ 212
Troubles, Tube----------------------------------- 80 Hints and Kinks for-------------------- 184
Troubleshooting: Improper Use of-------------------------- 196
Audio Amplifiers------------------------ 123 Localizing Trouble in the 202
Defective Components------------------ S3 Maintaining Switches in the — 198
Deflection Circuits ----------------------- 100 Measurement Accuracy of 201
FM Receivers------------------------------- 119 Meter, Setting of the-------------------202
Picture I.F. Amplifiers -------------- 94 Miscellaneous Applications _____ 157
Sound I.F. Amplifiers-------------------- 94 Mounting Stand for-------------------- 185
Switches in the V.T.V.M.198 Preventive Maintenance of_____ 196
Sync Circuits 100 Replacement Parts for----------------- 199
Tuners ------- 92 Rotary Switches in---------------------- 199
TV 79 Switching Arrangements in the 207
Video Amplifiers ------------- 98 Troubleshooting ----------------------- 195
Vacuum-Tube Voltmeters--------- 195 Troubleshooting Switches in the 198
Tube: Tube Replacement in the 199
Failures_____________________ 103, 104 Voltage Standard------------------------- 202
Heaters, Open----------------- 114 V.T.V.M. Defects:
Hum in Output--------------------------- 190 A.C. Voltage Range Defective _ 216
Replacement 199 A.C. Voltage Readings
Troubles -------------- 80 Inaccurate .. 216, 218
Tubes: A.C. Zero Potentiometer Out of
Aging for the V.T.V.M.------------- 200 Adjustment 216
Checking Gassy-------------------------- 162 Analysis of ---------------------------- -- 203
Gassy------------------------------------------- 105 Broken Connection in Ohms
Open ------------------------------------------- 104 Cable — -------------------214
Setup for Aging-------------------------- 201 Broken Ground Cable _ --------- 212
Tuner Operation, Improper -________ 93 Cable Connector Not Making
Tuners, Troubleshooting-------------------- 92 Contact —_______ ________ 214
Turns Ratio of Transformers —------- 161 Cracked Jewel Bearing in Meter 212
TV: Current Range Does Not Work 218
Antennas, Checking —.---------------- 170 D.C. Probe Defective------------------= 217
High-Voltage Circuit------- ---------- 101 D.C. Voltage Readings
Troubleshooting--------------------------- 79 Incorrect ----------- 217, 218
Defective Wiper Contact _____ 214
Unbalanced Ratio Detector 122 Excessive Leakage in Ohms
Unstable Picture 91 Circuit --------- 214
Exhausted Battery 214
Vacuum-Tube Probes 49, 53, 54 Fails to Operate ------------------------- 212
Vacuum-Tube Rectifying Probe--------- 51 Faulty Resistors---------------------------214
Variable-Frequency Oscillator --------- 144 Faulty Wiper Contact 216
Video Amplifier, Troubleshooting Fuse Open ........................................ * 212
in the 98 Inaccurate Reading .......................... 214
Voice-Coil Impedance Matching — 161 Instrument Does Not Read D.C.
Voice-Coil Load Impedance------------ 131 Voltages ----------- 217
Voltage: Instrument Falls to Work on
Checking Battery 159 Ohms —..... ---- -------------------- 214
D.C. Plate -____________________ 148 Intermittent Operation --------------- 213
D.C. Screen______________________ 148 Isolating Probe, Defective--------- 217
Divider Network, D.C. 20 Loose Connection in Ohms Cable 214
Doubler, Full-Wave------------------- 113 Loose or Dirty Contacts_____ . 216
Doubler Power Supply, Testing -169 Meter Pointer Off Scale ------ 213, 215
Gain 132 Meter Pointer Off Zero-------------- 216
Measurement of Screen------------ 147 Meter Pointer Sticks ----------------- 212
Measurement Troubles, A.C. — 216 Ohms Adjustment Control Fails
Measurements 75, 110 to Give Infinity Setting.. 214
Measurements, Low-Circuit Ohms Adjustment Potentiometer
Loading in 8 Defective --------- 214
Measuring High 152 Ohms Jack Shorted -------------------- 214
Output of Generators and Open Bridge Circuit ------------------ 212
Oscillators, Calibrating______ 157 Open Resistor Network 214
Plate Supply ------------------------------- 148 Probe Isolating Resistor Changes
Regulated Supply 156 Value -................... —------------- 217
Regulator Circuits, Gas Tube .... 155 Transformer Defective __ -_____ 212
R.F. Output Signal 148 Tubes Do Not Light --------- 212
R.M.S. vs. DBM 136 Zero Adjust Does Not Work — 212
Sources for Ohms Measurements 23
Standard for the V.T.V.M._____ 202
Voltmeter Scales for the Wheatstone Bridge --------------------------- 167
Oscilloscope 166 Whistling Between Stations-------------- 115
V.T.V.M.: Weak Signals .. - --------- 115
A.C. Voltage Section of 207 Wide-Range Frequency Response
Aging Tubes for 200 in A.C. Measurements______ 10
Analysis of Commercial------------ 206 Width, Insufficient _________________ - 101
Analyzing Trouble in the---------- 202
As a Precision Resistor - ------------ 185 Zero-Adjust Control —_______ 19, 43, 65
Circuit Arrangement of 203 Zero-Center Null Indication------- —— 167
Complete Circuit of 206 Zero-Center Scales . -------------- -------- 43
D.C. Voltage Measuring Section 207 Zero Setting of the Meter,
Experimental 186 Mechanical ------- .--- 202
224