UT2 Written Insturction
UT2 Written Insturction
Procedure:
Prepared by:
Date:
1. Objective
2. Equipment
2.1 Immersion tank with associated bridge manipulator (type and make)
2.2 Ultrasonic Flaw Detector - Use Olympus Epoch 600 (serial#) or equivalent.
2.3 Immersion probe – Use 5MHZ, 1/2 inch diameter zero degree transducer
(serial#) or equivalent.
2.4 Calibration Blocks - A set of ASTM Distance amplitude blocks made of 7075
Aluminum. All blocks shall have a 5/64-inch flat bottom hole (serial #). Blocks
include 7075-5-0012, 7075-5-0025, 7075-5-0050, 7075-5-0075, 7075-5-0150, and
7075-5-300.
3. Equipment Checks
3.1 Inspect immersion tanks. Movement of bridge and manipulator arms should be
unobstructed.
3.2 Ensure the Ultrasonic Flaw detector calibration sticker is attached and dated
within the previous six (6) months. Ensure the calibration record indicates
acceptable evaluation of Horizontal limit and linearity also Vertical limit and
linearity within the last 6 months.
3.3 Ensure the immersion probe being used has a sound exit face and is not worn or
chipped.
3.4 Ensure cleanliness of all calibration blocks.
NOTE: Any variances in the mentioned above must be reported to the Supervisor
immediately. DO NOT begin inspection until all discrepancies are repaired.
4. Personnel
5.1 Turn on Ultrasonic flaw detector and allow unit to warm up for 3 minutes.
5.2 Place Lucite pads in the bottom of the immersion tank with the 7075-5-0025
Aluminum calibration block on top.
5.3 Lower probe to 3 inch above the Calibration block. This is your water path. (Use
3-inch spacer to achieve this if necessary)
5.4 Adjust the angle of the probe to 90 degrees in regards to the top surface of the
calibration block 7075-5-0025. Avoid the flat bottom hole.
5.5 Calibrate the instrument to a five (5) inch time base. (see below for screen
presentation example)
5.6 Align search probe above the 7075-5-0075 block and maintain a three (3) inch
water path for a maximum signal amplitude from flat bottom hole. Adjust the gain to
obtain eighty (80) % full screen height on signal.
5.7 Record decibel settings as being the reference of the Distance amplitude
correction (DAC) curve from the 7075-5-0075.
5.8 Complete the DAC by using the following blocks: 7075-5-0012, 7075-5-0025,
7075-5-0050, 7075-5-0150 and 7075-0300.
5.9 After DAC is complete, proceed to add +6db above reference db and commence
scanning.
6. Inspections
6.4 Record all indications signal found on marking tape along with the indication
number (e.g. Ind#1, Ind#2, Ind#3) for later assessment. Using the 6db drop method
size and record all indications found on marking tape and on a separate sheet of
paper. Record Length (y-axis), Width (x-axis) and depth.
6.5 Document the length, width and depth of each defects on the plan and side view
of specimen sketch.
7. Data Interpretation
Any indications whose amplitude equals of exceeds fifty (50)% of the reference level
shall be rejected
8. Reporting