6490LCMS QQQ Presentation
6490LCMS QQQ Presentation
the World’s
World s Most LC/MS Division
Sensitive and Agilent Labs
Agilent Technologies
Triple Quad MS Santa Clara, CA
New 6490 - World’s Most Sensitive LC/QQQ
iFunnel Technology
¾ Agilent Jet Stream
¾ Hexabore capillary
¾ Dual-stage ion funnel
2
NiFunnel
E W I F UTechnology
N N E L T E C Hcaptures
N O L O G Y 10X more ions
3
Agilent Jet Stream Ion Generation
N b li i gas
Nebulizing
Enhanced efficiency nebulizer
S
Super-heated
h d sheath
h h gas
Nozzle voltage
g
High
Hi h LLow
Pressure Pressure
Stage 1 Stage 2
Existing
RF Ion
Guide
Stage 2
Stage 1
St 1-3 Torr
8-12 Torr
Stage 1 offset breaks up the high pressure gas exiting the Hexabore Capillary
8
How Does an Ion Funnel Work?
The RF Voltage RF Drive
focuses the ions
to the center.
Unfocused
Ions and Gas Focused
Enter Ions Out
The DC Voltage
accelerates the
ions to the exit. DC Drive
How is an Ion Funnel Constructed?
Gold Plated
Previously, with many metal plates…. Rim
… resulting
l i iin a llarge capacitive
i i lload
d and
d
relatively large RF power electronics.
7x 11x
7x
6490 14x
7x
7x 5x 10x
5x
10x
113
200 400 600 800 1000 1200 1400 200 400 600 800 1000 1200 1400
m/z m/z
6490 with iFunnel vs 6460 in positive ion mode
103
x10 x103
6490
6490
1
6460
6460
0 0.1 0.2 0.3 0.4 0.5 0.6 0 0.2 0.4 0.6 0.8 1.0 1.2
A
Acquisition
i iti Time
Ti (min)
( i ) Acquisition Time (min)
12
New 6490 Increases Signal but also Reduces Noise
6460 6490
S/N = 1,157:1 S/N = 11,640:1
where noise = 1xRMS where noise = 1xRMS
6490 with iFunnel vs 6460 in negative ion mode
14
New iFunnel Technology Revolutionizes
Atmospheric
p Sampling
p g
RF & DC
Voltages Set
Confinement and
Axial Acceleration
Smaller
((3.5mm)) Larger
(5 mm)
Exit Entrance
Aperture
Focuses Aperture
I
Ions iinto
t Accepts more
MS2 Ions
New 6490 LC Triple with iFunnel Technology
“Unrivaled
Unrivaled Quantitative Excellence”
Excellence
6490
10000
9000
SNR 1 pg reserrpine
8000
7000
6000
5000
4000
6460
3000
2000
6430
6410
1000
0
200 zeptomoles
p
4.5 verapamil
Injected on-column
4.4
4.3
42
4.2
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
Acquisition Time (min)
Unprecedented Six logs of Linearity with 6490
100 Attogram to 100 Picogram varapmil
arapmil on col mn
on-column
10000000
Y = 65283 x – 0.390412
1000000 R2 = 0.9976
0 9976
100000
nse
ea Respon
10000
1000
Are
100
10
1
0.1 1 10 100 1000 10000 100000
Femtograms
e tog a s
50 pg/mL Fluticasone in plasma with 6460
25 Femtogram on-column
Area RSD = 23.7% (n=6)
Mean accuracy = 104
Fluticasone propionate
F
x10 1 *1.534 min.
S O
5.3 CH3
CH3 O
5.2 HO
O
CH3 CH3
5.1
5 F H
4.9 O
F
4.8
4.7
4.6
4.5
4.4 MRM unit / unit resolution
4.3 (0.7 0.7
(0 7 amu / 0 7 amu)
4.2
0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2 2.1 2.2 2.3 2.4
4.8 4.8 F
4.7
47 4.7
47
4.6 4.6
4.5 4.5
4.4 4.4
43
4.3 43
4.3
4.2 4.2
0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2 2.1 2.2 2.3 2.4 2.5
Acquisition
q Time [[min]]
Page 24
The Value of Enhanced Resolution on 6490
Fluticasone propionate (inhalant) in human plasma 2 5 fg on-column)
2.5 on column)
0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 2.2 2.4
Simetryn S/N = 9
Secbumeton S/N = 51
Prometon S/N = 40
Atrazine S/N = 28
Peptide MRM on 6460 vs new 6490 Triple Quad
1
x10 1 femtomole on 6460
4.5
LVNEVTEFAK
4.4 575.5 Æ 937.5
4.3
4.2
1
x10
10
100 attomole on 6490
4.5
4.4
4.3
4.2
Triggered MRM – What Problem Does tMRM Solve?
T t d Quant
Targeted Q t + Compound
C d Identification
Id tifi ti
DDA –scan
switch to PIS
tMRM for up to 8 Confirmatory Ion Transitions
- when primary compound MRM triggers confirmatory MRMs
Faster compound ID at higher sensitivity with tMRM
Plot counts versus m/z for product ions
“Composite MS/MS spectra for compound b”
Product %MRM
Ion Scan y
Sensitivity
Time
QTRAP
~ 100 msec ~ 10%
5500
countts
Page 31
The New 6490 Triple Quad System