-DRS Mobile Pro 320

Technical Specifications

Option:

Electrical Specifications
Mains connection: 115 V / 230 V ±10 % - 50/60 Hz ± 5 Hz, standard monophase with ground
conductor.

Absorbed current 10 A

Line compensation Automatic

Line resistance <1 Ω @ 115 / 230 Vac

Standard socket 16 A @ 230 V

Power supply cable 2.9 m, retractable

Mechanical Specifications

Changes in product illustrations or changes which serve the technical progress, as well as mistakes are reserved.
Fixed Column Option: Telescopic Column

W x L x H (in transport position) 576 mm x 1168 mm x 1855 mm 576 mm x 1200 mm x 1421 mm

Transport handle height 900 mm 900 mm

Focus-ground distance (Z-axis) 726 mm ÷ 2013 mm 660 mm ÷ 1354 mm


Publisher: EXAMION GmbH, Erich-Herion-Str. 37, 70736 Fellbach, Germany.
Max. height 2242 mm 2242 mm

Arm extension (X-axis) 360 mm 360 mm

Front range 710 mm ÷ 1070 mm 710 mm ÷ 1070 mm

Lateral range 517 mm ÷ 877 mm 536 mm ÷ 896 mm

Rotation of the monobloc • Around the arm axis (α swivel): ± 180°


• Around the column axis (β swivel): ± 320° from transport position
• Around its axis (γ swivel): 127° (90° forward, 37° backward)
Movement Motorized, speed proportional to the force impressed on the transport handle.

Speed forward (transport position) 0 ÷ 1,4 m/s

Speed forward in working position 0 ÷ 0,7 m/s


Version: 04/ 2021

Speed backwards 0 ÷ 0,7 m/s

Page 1 of 4 www.examion.com
Max. superable inclination 12° (21 %)

Diameter of the wheels • Front: 125 mm, antistatic, damped


• Rear: 300 mm, antistatic
Weight 412 kg, batteries included

Environmental Conditions OperatingTemperature:


• 10° C to 40° C (50° F to 104° F)
• Humidity: 30 % to 75 % (non-condensing)
• Atmospheric pressure: 700 hPa to 1060 hPa

Transport / Storage
• Temperature: -25° C to 70° C (-4° F to 122° F)
• Humidity: 10 % to 90 % (non-condensing)
• Atmospheric pressure: 500 hPa to 1060 hPa

Batteries
Battery type VRLA (Valve regulated lead acid batteries)

Battery pack 12 batteries x 12 V, 18 Ah, Pb, 144 Vdc

Vacuum nominal voltage 144 Vdc

Load voltage (worst case) 100 V @200 A

Battery charger
Power supply 115 Vac / 230 Vac ±10 % single-phase, 50 / 60 Hz

Absorption by mains 10 A max

Charge indicator Yes

Changes in product illustrations or changes which serve the technical progress, as well as mistakes are reserved.
Charge time 5h

Safeties Overtemperature, overvoltage, charge timeout

X-ray Specifications
Max power of the generator 32 kW (320 mA @ 100 kV)

Inverter frequency in high voltage 40 kHz Publisher: EXAMION GmbH, Erich-Herion-Str. 37, 70736 Fellbach, Germany.

kV variation range 40 ÷ 125 kV in steps of 1 kV

mA variation range 70 ÷ 320 mA automatically associated to kV

mAs variation range 0,1 ÷ 320 mAs with increases of 12,5 %

Exposure times range 0,001 ÷ 3 sec according to set mAs

Continuous thermal dissipation 60 W

mAs range according to kV


Small focus 40 ÷ 100 kV; 0,1 ÷ 200 mAs
Version: 04/ 2021

Large focus • 40 ÷ 100 kV; 0,63 ÷ 320 mAs


• 101 ÷ 125; 0,63 ÷ 250 mAs

Page 2 of 4 www.examion.com
X-ray group
Monoblock • Nominal power (100 kV – 320 mA) 0.1 s: 32 kW
• Max voltage at the tube: 125 kV
• Max current at the tube: 350 mA
• Min. inherent filtration @ 75 kV: 1.4 mm Al
• Thermal capacity available: 600 kJ
• Continuous thermal dissipation: 60 W
• Weight: 19.0 kg

X-ray tube • Rotation speed: 3000 min-¹


• Nominal High voltage: 130 kVp
• Nominal focus dim. (IEC 60336): 0,8 mm small focus, 1,3 mm large focus
• Nominal anodic power (IEC 60613): 16 kW small focus, 32 kW large focus
• Thermal capacity of the anode: 80 kJ (107 kHU)
• Max continuous anode dissipation: 300 W
• Min. inherent filtration (IEC 522): 0,7 mm Al

Collimator • Collimation: Manual with internal light source, multilayer, squared field.
• Coverage of the field at 100cm FFD (SID): min 0 x 0 cm, max 43 cm x 43 cm
• Lighting source: Clusters of high-brightness LED power.
• Focal distance measurement: Retractable tape measure (max extension 3 m)
• Rotation: ± 120°
• Weight: 5.5 kg
• Inherent filtration @ 75 kV: 2 mm Al

Detector Integration
Charging Charging via cable

Detector

Changes in product illustrations or changes which serve the technical progress, as well as mistakes are reserved.
Selectable Detectors X-DR L WiFi X-DR L WiFi Premium

Pixel Size 150 µm 100 µm

Pixel Matrix 2304 x 2800 3500 x 4300

Weight 3.4 kg 2,8 kg

Protection Class IPX1 IP56

Uniform Load 150 kg 300 kg Publisher: EXAMION GmbH, Erich-Herion-Str. 37, 70736 Fellbach, Germany.

Local Load 100 kg (4 cm diameter) 150 kg (4 cm diameter)

Panel PC
CPU Intel® CoreTM i7

RAM 4 GB

Mass Storage 60 GB SSD, 460 GB HDD

Network Gigabit LAN RJ45, WLAN IEEE 802.11 ac / a / b / g / n

Display 19", 1280 x 1024, multi-touch screen


Version: 04/ 2021

Operating System Windows 10 IoT

Page 3 of 4 www.examion.com
Page 4 of 4
Software
Configurations

• DICOM Store SCU

EXAMION X-AQS PACS


• DICOM Worklist SCU

• Integrated Mini-PACS
EXAMION X-AQS DICOM Modality SCU

www.examion.com
Version: 04/ 2021
Changes in product illustrations or changes which serve the technical progress, as well as mistakes are reserved.
Publisher: EXAMION GmbH, Erich-Herion-Str. 37, 70736 Fellbach, Germany.