U2 Troubleshooting
U2 Troubleshooting
TROUBLESHOOTING CHART
Err 0 CPU PCB RAM failure. (Evident at turn on.) 1. Turn system off/on.
2. Replace CPU PCB.
Err 1 CPU PCB ROM failure. CRC check of EPROM. 1. Turn system off/on.
(Evident at turn on.) 2. Replace EPROM U9.
3. Replace CPU PCB.
Err 2 CPU operation code failure. (Evident at turn on.) 1. Turn system off/on.
2. Replace EPROM U9.
3. Replace CPU PCB.
EEPROM FAILURE CPU PCB EEPROM CRC check failure. 1. Turn system off/on.
2. Replace CPU PCB.
85V Err 85V power supply out of tolerance 1. Turn system off/on.
2. Replace power supply assy.
U/S WDT Err U/S Watchdog Time Out 1. Release footpedal, and press U/S key, or
(Power to handpiece is disabled.) depress footpedal from position 0 to
position 1 to clear error message. Retune
handpiece and re-engage U/S power.
2. Replace U/S Driver PCB.
KeySwtch Err • Key is depressed for too long. Press and release key quickly.
U/S Err U/S output out of tolerance. 1. Push US key and try again.
2. Connect known good handpiece.
3. Replace U/S Driver PCB.
Table 4-3. TROUBLESHOOTING CHART - This chart can be used to help diagnose a problem with the Unversal II system.
Look for the condition or error message in the left column that describes the problem you observe, then look
to the right for the possible cause and suggested correction.
906-3940-501 4.7
Universal II
Gnd Flt Err U/S return path has been interrupted. 1. Connect known good handpiece.
2. Replace U/S Driver PCB.
3. Replace cable W2.
Pwr Sply Err Power supply voltages out of tolerance. 1. Perform power supply checks and adjust-
ments.
2. Replace power supply.
Irr Sol Err Indicates the irrigation solenoid or feedback 1. If solenoid does not move with depression
circuit has faulted. of ftsw:
• reseat solenoid connector.
• T-shoot 24 VDC.
• replace solenoid.
2. If solenoid does move:
• replace Irr Switch PCB.
• replace Front Panel PCB.
• replace cable W3.
Ftsw Conn Err • System does not recognize ftsw. Check footswitch connection or reseat connector.
Replace footswitch.
Vacuum display jumps to 400 mmHg. • Transducer feedback out of calibration. Perform vacuum calibration.
Comm Err • Communications cable between Universal II Securely connect communications cable
(when used as fragmentation and Universal Microtome Vitrectomy System between Universal II and UMVS.
system) (UMVS) not connected.
• Faulty printed circuit board. Replace/Reseat Front Panel PCB and /or CPU.
4.8 906-3940-501