M24486EN
M24486EN
Model: AP-338
Version: 1.0
IMPORTANT: Many factors (e.g. environment gas residue, smoke & various smell
in human mouth, etc…) may affect the Blood Alcohol Concentration (BAC) level, so
the Readings are For Reference Only.
Neither the reseller nor manufacturer accept any liability arising from the use of this
unit.
DO NOT drink before you drive!
www.aap.com.hk
Contents
Page
Measuring instrument……………………………………………………………………. 2
Packing List ...................................................................................................................... 3
Basic Scientific Principles…………………………………………………....................... 4
Maintenance…………………………………………………………………………….… 4
Intended Use………………………………………………………………………….…... 4
General Conditions…………………………………………………..……………….…… 5
Power & Charging………………………………………………………………………... 5
Operation instructions…………………………………………………………….…...….. 5
Active Test……………………………………………………………..……………....…. 5-7
Quick Test……………………………………………………….…………….….……….. 7
Test record…………………………………………………………………………………. 8
Sensor Test…………………………………………….…....……………………………… 8
System Setting…………………………………………………….…………………...…… 9
GPS..………………………………………………………………………………………… 10
Clock and Torch…………………………………………………………………………… 10
Specification…………………………………………………………………...………….... 11
Calibration and Service……………………………….…………………….…...…...…..... 11
Maintenance
Repair/maintenance of the AP-338 should only be carried out by trained technician from authorized professional
breath alcohol tester Service Center.
Only authentic spare parts can be used for maintenance.
Intended Use
* Don’t put this product in high temperature, high humidity or high dust environment.
* Don’t knock this product, especially do not press the TFT-LCD by force.
* Don’t dismantle the machine and use the alcohol, thinner or benzene to clean the surface of the product and its
components.
* Don’t use in the area prohibiting using of wireless devices.
* Don’t use in the flammable and explosive environment.
* Please use the touch pen when doing the touch screen calibration
* Before use, please fully charge up the Handset using the provided accessory charger.
General Conditions
- Use a new mouthpiece for each test or each person being tested.
- Make sure no alcohol and solvent vapors (disinfectants) in the ambient air.
- Do not allow a high alcohol concentration to reach the sensor. For instance, from drinking alcohol, eating
alcohol containing foods or rinsing the mouth with alcohol mouth washer just before measurement.
- Keep away from the antenna of mobile phones or transmitters.
Power ON/OFF
1.Press the “Power Key” for 2 seconds to turn ON the machine
then enter into the main interface.
2.Press and hold the “Power Key” for about 2 seconds to turn OFF the machine.
3.If no key input for 10 minutes, the machine will auto power off.
3 minutes for enter screen saver. (it can set in Advanced setting)
Operation instructions
1. Active Test: Under this test mode, the device will initialize the sensor
condition. In some case, it need take few seconds to 2 minutes.
1.1 Insert a new mouthpiece in CORRECT orientation into the inner tube chamber inlet..
1.2 Press the “ ” icon on the main menu, The LCD will shows “Tester Ready, Please Blow Continuously”,
user can start the alcohol test.
Evenly blow into the mouth piece for about 3-6 seconds with a continuous “Beep” sound announcement. (Active test)
The blowing time is related to the blowing pressure.
1.3 If the breath sample flow rate is too soft or blowing duration is too short, “Blow Abort or too Short” will be shown
on the display. The machine will return to the main menu automatically. Repeat 1.1 & 1.2 to test again.
Active test:
1.4 If the breath sample is taken successfully, the display will show “Sampling…Please wait”
The machine will analyze the breath sample and the test result will then be shown on the
display after analyzing.
1.5 Press “Retest” icon to start another test or press “Save” to input the relevant information.
Press “Return” to return back into the main menu;
1.6 If the breath sample is taken not successfully. Then “blow time-out” or “blow abort” will show as following.
User need to re-test again.
After inputted the relevant information, the Print or Retest interface will display on the LCD.
Press “Print” key to print out the test result or Press “Retest” to start another test or Press “Return” key to return back.
2. Quick Start: - Under this mode, the device allow to test without initialize
sensor condition. Eg: temperature & residual alcohol.
Ready-blow sampling
4.3. Delete: to delete the test record, need input user password
4.4. Exit: return to previous menu.
4.5 Delete all record: need input user password.
4. Sensor Test:
Caution: The calibration is already been done before shipping, user alters the setting parameters will override
all calibration data. Only FiT authorized technician is allowed to modify the calibration data.
① “CAL TEST”: This is the test for verification the accuracy for Lab Certification / Verification only.
② “CAL Flow”: calibration breath blowing flow/ pressure.
Please blow to the device use 15 Liter/minute flow for 6 seconds,
with device has Long “beep….!” Sound. A “beep beep” sound
when finish. Screen shows flow data, press “Save” to save the
date. Or press “Restart” to try again.
Or press “return” icon to exit the flow calibration menu.
④ “SET PASSWORD”: default user password is empty. User can set user own password
If forget password, you make need to send back to service center to re-set it.
⑤ “Set Unit”
Set the unit to Permillage (‰BAC), Percentage (%BAC) and mg/L,
mg/ml, mg/100ml and g/l. as picture.
For advanced user, it may set the BrAC ratio: 2100, 2000, 2200, 2300.
⑥ “USB CONNECT PC” : user can download the records and photos to the PC computer
with a PC software and USB cable. Read the PC download manual for more details.
⑦ “Advanced Setting”
Advanced Setting: For Authorized service center only
(Super-user password is required)
Only allowed to perform this function by authorized service center
please enter into the menu, to start the GPS postion detection.
After successful locating, then make the alcohol test with GPS
record. See picture below for your reference.
Remark:
1. The GPS Module will take about 5~20 minutes while the first time to use this function.
2. In the spacious places GPS position will take less time than in the places with dense buildings.
7. Clock: User can set the device data and time under this manual.
1. click the upper left corner setting icon to enter date and time setting interface. After change, press “OK” to
save the setting. Press “back” to exit without save.
2. Return to main menu by click the return icon at the upper right corner
8. System information
9. Torch: ON/OFF the build in LED in front of the device. Check the driver at the
roadside, it is helpful.
7. Mouthpiece: Easy to replace a new mouthpiece by pressing a key, very convenient and hygiene.
13. Long Term Storage condition: Temperature: -20 Degree C to +55 Degree C.. Humidity: 30 to 90% RH
14. Operating Temperature: From -20℃ to +45 ℃, ( 5℃ to +40 ℃ for optimum operation)
15. Memories: Min. 150,000 memories with full information:- Drive name, vehicle no., inspector name, test
result with time stamp, date and time, location, instrument no., Calibration date, test no. etc…
16. Measurement precision: Standard deviation +/–0.025 mg/L or +/–5 % of the measurement at calibration point
Drift: typically < 0.8 % of the measurement value/month
17. Alert level setting: Set “drunk driving” and “driving while intoxicated” by a supervisor password.
12. How often should I re-calibrate the tester or change a new sensor module?
The tester can remain accurate for 1 year of use. The more frequent the use, the more the unit will need to be
calibrated. According to DIN31051, we recommend recalibration every 12 months (or earlier) basis to ensure
maximum accuracy.
NB* It is important to observe the recommended 15 minute pause without consuming any alcohol, cigarettes or
anything else before testing. Failure to do this can shorten the life of the sensor. In such cases, the sensor can be
permanently damaged or may need more frequent calibration. This user manual should be read carefully for
information on caring for the unit sensor.
Each test change new mouth piece, for best performance, wait for 3 minutes between each test.
Many factors (e.g. environment gas residue, smoke & various smell in human
mouth, etc…) may affect the BAC level, so the readings are For Reference Only.
Neither the reseller nor manufacturer accepts any liability arising from the use
of the unit. Don’t drink before you drive!
Made in China.