Specification Sheet: Specifications TC10B
Specification Sheet: Specifications TC10B
Specification sheet
Specifications TC10B
Tamson Circulator 10 Litres
General Safety
The TC10B can be used both as circulator or bath. The bath conforms to CE regulation. Further it is
The bath fluid is heated via a microprocessor equipped with multiple safety features in order to
controlled heating element. When working below guarantee safe around the clock operation. A
ambient temperature cooling fluid can be circulated mechanical safety thermostat protects the bath against
through the cooling to lower the bath temperature. high temperatures and will automatically reset itself
after switching the bath off and on again. The safety
The bath is equipped with fluid level detection. The thermostat can be set between ambient and +300°C.
microprocessor controlled heating with auto tune In case of over-temperature, electronics and motor are
function guarantees precise temperature control. switched-off. The pump motor is protected against
mechanical overload via a resettable fuse.
Span
The bath temperature can be set to a maximum of Temperature readout
200C (392F) in steps of 0.1degree. The overall Standard available in C, on request in F.
system accuracy is 0.02°C.
Pump
Accessories A pump circulates the bath content or provides
Top lid is included. external circulation. The pump pressure is 300 mBar
and maximum flow is 7 litres per minute with no
Options counter pressure.
P/N 03T0017 Leveling platform.
P/N 02T3025 RS232 communication.