MM5E Service Manual Ver 20.0
MM5E Service Manual Ver 20.0
Pre-installation
GW10078855.811.01.09.02
Installation
Installation Instructions
GW10078855.812.01.09.02
MULTIMOBIL 5E
Service Instructions
Service Instructions
GW10078855.898.01.15.02
Update Instructions
Update Instructions
GW10078855.896.01.07.02
The reproduction,
transmission or use of this
document or its contents is
not permitted without
express written authority.
Offenders will be liable for
damages.
GW10078855.802.01.20.02
Multimobil 5E
CV GW
Pre-installation
The reproduction
transmission or use of
this document or its
contents is not
permitted without
express written
authority. Offenders will
be liable for damages.
GW10078855.811.01.09.02
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are directed to contact one of the local offices of
Siemens or one of its affiliates before attempting installation or service procedures.
GW10078855.811.01.09.02 Page 2 of 20
INDEX
INDEX 3
1 PRE-INSTALLATION 4
1.1 TECHNICAL SPECIFICATIONS 4
1.2 INFORMATION ON ROOM PLANNING 6
1.3 PACKAGING 7
1.4 UNLOADING 8
1.5 UNPACKING THE MM5E 8
1.6 UNPACKING THE MONITOR TROLLEY 17
1.7 INSTALLATION OF THE MONITORS 19
1.8 INSTALLATION OF THE THERMAL PRINTER (OPTIONAL) 19
GW10078855.811.01.09.02 Page 3 of 20
1 Pre-installation
1.1 Technical Specifications
Sr.No Parameter Value
1. Mains Voltage 190 – 240VAC
2. Line frequency 50 Hz , ± 2 Hz
3. Mains Fuses 16A delayed fuse, provided in
eqiupment
4. Line impedance 1 W (max) @190 – 240V
5. Length of power cable 5.5 meters
6. Power Output- Radiography 2.5 kW @ 190-240VAC input
Nominal Electric power at
100kV and 100mSec.
7. Wave Shape Multipulse DC - Ripple 5kV max
8. KVp Range 40 – 100kV in 20 steps
9. KV steps 40, 42, 44, 46, 48, 50, 52, 55, 57, 60,
63, 66, 70, 73, 77, 81, 85, 90, 96,
100.
10. KV Accuracy £ ± 10 %
11. mA-range 13 – 63 mA @ 190-240V input
12. mAs Steps: 0.32, 0.4, 0.5, 0.64, 0.8, 1.0, 1.25,
1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.4, 8, 10,
12.5, 16,20, 25, 32, 40, 50, 64, 80,
100, 125, 160, 200
13. mAs Range 0.32 – 200mAs at 40kV
0.32 – 160mAs at 42 - 50kV
0.32 – 125mAs at 52 - 63kV
0.32 – 100mAs at 66 - 77kV
0.32 – 80 mAs at 81 - 100kV
14. mAs Accuracy £ ±10 % for mAs £ 20 mAs
£ ±5 % for mAs > 20 mAs
15. Exposure Time 20mSec - 5 Sec in 24 steps @ 190-
240V
16. Fluoroscopy voltage 40-110 kV
17. Fluoroscopy Modes ADR(Automatic dosage regulation):
-Anti-isowatt And Contrast
ADR Stop (Manual)
18. Fluoroscopy mA range Anti-isowatt 0.5 to 5 mA
Contrast 0.5 to 7mA
GW10078855.811.01.09.02 Page 4 of 20
GW10078855.811.01.09.02 Page 6 of 20
1.3 Packaging
The MULTIMOBIL 5E unit is delivered in two boxes.
One box consists of the Monitor trolley.
Second box contains the MM5E unit along with monitors, memory module and
accessories.
The trolley mounted on the wooden base and is fixed using plastic straps.
The C-arm unit is clamped using brackets on the wooden base.
The two monitors and the memory module are packed in a corrugated box
and placed on the wooden platform.
Second box consists of standard accessories i.e. Clips and user manual.
GW10078855.811.01.09.02 Page 7 of 20
1.4 Unloading
The unloading of the unit from the truck can be carried out using any of the
following ways depending on the site conditions:
1) Using a forklift.
2) Using an overhead travelling crane.
Once the unit is removed from the transport, place it on the ground on the
base.
Check for the tilt indicator (figure shown below). If the tilt indicator shows red
follow the instructions mentioned on the tilt indicator.
TILT INDICATOR
GW10078855.811.01.09.02 Page 8 of 20
1. Cut the straps and remove the ramp placed on top of the box
Ramp
Strap
GW10078855.811.01.09.02 Page 9 of 20
3. Remove the split pin and unlock the toggle clamps to remove the side walls
GW10078855.811.01.09.02 Page 10 of 20
4. Remove the side walls and then the ESD bag covering the unit
5. Cut the straps holding the accessories box, Monitor boxes and the memory module box. Remove the accessories box, Monitor boxes and the
memory module box from the base of the palette
GW10078855.811.01.09.02 Page 11 of 20
6. Cut the straps and remove the foam covers 7. Cut straps holding the single tank and the C-arm.
from the single tank and image intensifier Remove the supports by unscrewing the bolts.
Single
tank
C-Arm
support
support
Bolts
GW10078855.811.01.09.02 Page 12 of 20
8. Remove the metal clamps holding the MM5E to the palette by unscrewing the bolts.
Bolt
Side wheel support Bolt Metal clamps Rear support
GW10078855.811.01.09.02 Page 13 of 20
9. Release the brake and 10. Remove the supports near the
move the C-arm the position side wheels by sliding them out
as shown in the picture
Side wheel supports
11. With the help of lever and side wheel support, lift the unit
at the rear end as shown and slide out the rear support
Rear support
GW10078855.811.01.09.02 Page 14 of 20
11. Using the rear support and lever, lift the unit at the 12. Mount the ramp and roll
front end near the front wheel and remove the front down the unit from the base
support. palette
Front support
Ramp
GW10078855.811.01.09.02 Page 15 of 20
Observe the following steps for transporting the unit to the OT.
Unlock the C-arm and turn it to the position shown in the step No.9 and lock it.
This is the suitable position for moving the unit from the platform. This is also
the parking position.
Roll all the cables and hang it on the cable holder provided on the housing.
Release the brakes at the wheels and move the unit to the OT. Use the
steering for ease of maneuverability
GW10078855.811.01.09.02 Page 16 of 20
GW10078855.811.01.09.02 Page 17 of 20
4. Remove the ESD bag, corner supports and then cut the straps holding 5. Lift the trolley from the base and place it on the
the trolley to the base. floor. Follow the procedure for mounting the
monitors on the trolley
Trolley Base
GW10078855.811.01.09.02 Page 18 of 20
To install SONY make thermal printer in MM5E system following points has to be
ensured to have the better quality image on the monitor by keeping input
impedance of Printer to high impedance stage.
1) Connect the printer using “BNC cable provided with the printer”. Connect the
BNC cable to a BNC coupler and connect it to memory module input and other
end to the printer input.
3) Press the “MENU” button once and MENU selection assembly comes out and
Selected MENU is identified by “ * ”
GW10078855.811.01.09.02
5) Press the “D” button till you get 75 ohm on the display and select “OFF” using
R
Button
6) Finally press the MENU button and MENU selection assembly will comes to its
Original position.
GW10078855.811.01.09.02 Page 20 of 20
Installation Instructions
The reproduction
transmission or use of
this document or its
contents is not
permitted without
express written
authority. Offenders will
be liable for damages.
GW10078855.812.01.09.02
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are directed to contact one of the local offices of
Siemens or one of its affiliates before attempting installation or service procedures.
GW10078855.812.01.09.02 Page 2 of 11
INDEX
1 INSTALLATION 4
GW10078855.812.01.09.02 Page 3 of 11
1 Installation
1.1 Transport Position of Unit
After unpacking the unit as explained in pre installation, bring the unit to its
transport position as explained below.
NOTE: While transporting the unit on the slope ensure that the slope
should not exceed +5°. If the slope is more than 5°, there is a
risk of toppling.
The unit must be checked for the status of its safety measures at least every
year or any time desired by the user, not exceeding one year.
However, we recommend in the interest of the safety of the patient, the user,
and other personnel to have a yearly drill on operational safety.
GW10078855.812.01.09.02 Page 4 of 11
Before commencing the operation, the user must convince himself / herself
regarding all safety aspects and their proper functioning. He/she should also
ascertain that all displays and indicators are functioning as described.
The radiation indication lamp should light only when the exposure is released.
In case the lamp lights in any other condition, the unit should be immediately
switched off and SIEMENS SERVICE DEPARTMENT should be informed.
GW10078855.812.01.09.02 Page 5 of 11
In case of fault if the kV/mA tries to build beyond the safe limit of X-ray tube
the protection circuit blocks the HT.
In case of fault if the Filament current tries to increase beyond the safe limit
the protection circuit blocks filament boosting.
Maximum permissible resistance between the Earthing lead terminal in the
control and on the HT transformer shall not exceed 0.1 ohms.
1.2.4 Radiation Protection
Do not put any part of your body in the direct line of
radiation.
WARNING Keep maximum possible distances from the object
being radiographed. For this make the best use of 5
mtr of recoilable exposure release switch.
In the room, wear protective clothing (lead Apron).
Monitor radiation received using the personal TLD-
badges or pen-dosimeter.
GW10078855.812.01.09.02 Page 6 of 11
GW10078855.812.01.09.02 Page 7 of 11
GW10078855.812.01.09.02 Page 8 of 11
1.3.2.3 Switching ON
Press switch on the top panel. The displays will light up.
The Display will be as follows for the set-up time. (8 seconds approx)
kV mAs kV mA min
After setup and initial hardware checks the unit becomes ready and the
display
kVis as shown below.
kV mAs kV mA min
Fluoroscopy Mode is selected by pressing the Mode Selection switch and can
be initiated by pressing the footswitch. In fluoroscopy mode, the display is as
follows.
kV mAs kV mA min
Radiography mode can be selected by mode selection switch. In radiography
mode the display is as follows.
kV mAs kV mA min
The kV & mAs display and settings will now respond to the corresponding
switches on the panel.
In case of any errors while setting up / hardware check / stand-by operation,
the corresponding Error code will be displayed on the same displays.
The display will indicate Cod E99 in case
the unit while setting up/ after set-up
detects that last reset of the unit was due
to watch-dog and not hardware RESET.
GW10078855.812.01.09.02 Page 9 of 11
1.3.3 X - ray
1.3.3.1 Fluoroscopy
Switch ON the unit. Enter into the Fluoroscopy Mode, press foot switch check
Fluoroscopy is possible at the same time radiation Indication, Amber coloured
LED on the top panel and the lamp on the monitor trolley, will light up. The
Image will appear on the Monitor in ‘Live’ Mode, by releasing the foot switch,
the Last Image on the Monitor will be frozen (LIH function).
1.3.3.2 Exposure
Set the exposure parameters as 60kV, 10mAs.
Pull the recoilable 2 step Exposure Release Switch and release an exposure.
The Amber coloured LED on the top panel and the lamps on the monitor
trolley will light up for duration equal to exposure time.
An audible indication by a muffled sound from single tank as well as from the
buzzer will confirm the working of the inverter and the exposure.
GW10078855.812.01.09.02 Page 10 of 11
GW10078855.812.01.09.02 Page 11 of 11
The reproduction
transmission or use of
this document or its
contents is not
permitted without
express written
authority. Offenders will
be liable for damages.
Restricted GW10078855.844.01.10.02
Chapter Page Revision
All All 10
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are directed to contact one of the local offices of
Siemens or one of its affiliates before attempting installation or service procedures.
Restricted GW10078855.844.01.10.02
INDEX
1 CIRCUIT DIAGRAMS 4
2 WIRING DIAGRAMS 19
Page 3 of 33
GW10078855.844.01.10.02
1 Circuit Diagrams
Index
MULTIMOBIL 5E
Page 4 of 33
GW1008855.844.01.10.02
Page 6 of 33
GW1008855.844.01.10.02
Page 8 of 33
GW1008855.844.01.10.02
Page 10 of 33
GW1008855.844.01.10.02
Page 12 of 33
GW1008855.844.01.10.02
Page 14 of 33
GW1008855.844.01.10.02
Page 16 of 33
GW1008855.844.01.10.02
Page 18 of 33
GW1008855.844.01.10.02
1 Wiring Diagrams
GW10078855.844.01.10.02 Page 19 of 33
Page 20 of 33
GW1008855.844.01.10.02
Page 22 of 33
GW1008855.844.01.10.02
GW10078855.844.01.10.02 Page 23 of 33
Note: The connections to the Master Card also Applicable for D915A
Page 24 of 33
GW1008855.844.01.10.02
Note: The connections to the Master Card also Applicable for D915A
GW10078855.844.01.10.02 Page 25 of 33
Page 26 of 33
GW1008855.844.01.10.02
GW10078855.844.01.10.02 Page 27 of 33
Page 28 of 33
GW1008855.844.01.10.02
GW10078855.844.01.10.02 Page 29 of 33
Page 30 of 33
GW1008855.844.01.10.02
GW10078855.844.01.10.02 Page 31 of 33
Page 32 of 33
GW1008855.844.01.10.02
GW10078855.844.01.10.02 Page 33 of 33
The reproduction
transmission or use of
this document or its
contents is not
permitted without
express written
authority. Offenders will
be liable for damages.
Restricted GW10078855.850.01.08.02
Chapter Page Revision
All All 08
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are directed to contact one of the local offices of
Siemens or one of its affiliates before attempting installation or service procedures.
Restricted GW10078855.850.01.08.02
INDEX
1 CIRCUIT DESCRIPTION.................................................................................................................. 5
1.1 MASTER CARD D915 ......................................................................................................................... 5
1.1.1 Power Supply .......................................................................................................................... 5
1.1.2 RESET Generation .................................................................................................................. 5
1.1.3 Crystal Operation .................................................................................................................... 5
1.1.4 Microcontroller ....................................................................................................................... 5
1.1.5 Octal Latch.............................................................................................................................. 5
1.1.6 GAL ........................................................................................................................................ 6
1.1.7 EPROM ................................................................................................................................... 6
1.1.8 NVRAM ................................................................................................................................... 6
1.1.9 Keyboard................................................................................................................................. 6
1.1.10 Display ............................................................................................................................... 6
1.1.11 Serial Communication ......................................................................................................... 6
1.1.12 KVsoll and mAsoll ............................................................................................................... 6
1.1.13 Line voltage monitoring ...................................................................................................... 7
1.1.14 KV Feedback....................................................................................................................... 7
1.1.15 IH Feedback ....................................................................................................................... 7
1.1.16 mA Feedback ...................................................................................................................... 7
1.1.17 KV Regulation..................................................................................................................... 8
1.1.18 Filament Regulation ............................................................................................................ 8
1.1.19 mAs Integration................................................................................................................... 8
1.1.20 Control of the Firing Pulse frequencies REG and CAL......................................................... 8
1.1.21 Generation of actual Firing Signals ..................................................................................... 9
1.1.22 Main Inverter Short Circuit Feedback SCM ......................................................................... 9
1.1.23 kV max, mA max Feedback ....................................................................................................... 9
1.2 MASTER CARD D915A .................................................................................................................... 10
1.2.1 Power Supply ........................................................................................................................ 10
1.2.2 RESET Generation ................................................................................................................ 10
1.2.3 Crystal Operation .................................................................................................................. 10
1.2.4 FPGA .................................................................................................................................... 10
1.2.5 Multiplexer ............................................................................................................................ 10
1.2.6 NVRAM ................................................................................................................................. 10
1.2.7 Keyboard............................................................................................................................... 11
1.2.8 Display .................................................................................................................................. 11
1.2.9 Line voltage monitoring ......................................................................................................... 11
1.2.10 kV Feedback ..................................................................................................................... 11
1.2.11 IH Feedback ..................................................................................................................... 11
1.2.12 mA Feedback .................................................................................................................... 12
1.2.13 kV Regulation.................................................................................................................... 12
1.2.14 Filament Regulation .......................................................................................................... 12
1.2.15 mAs Integration................................................................................................................. 12
1.2.16 Main Inverter Firing Frequency ........................................................................................ 13
1.2.17 Filament Regulator Firing Frequency................................................................................ 13
........................................................................................................................................................... 14
1.2.18 Main Inverter Short Circuit Feedback SCM ....................................................................... 14
1.2.19 FLASH .............................................................................................................................. 14
1.3 INVERTER ASSEMBLY D64 ............................................................................................................... 14
1.3.1 IGBT’s Inverter Assembly ...................................................................................................... 14
1.3.2 Firing .................................................................................................................................... 14
1.3.3 Inverter ................................................................................................................................. 15
1.3.4 SCM ...................................................................................................................................... 15
1.4 FILAMENT & II INTERFACE PCB D602A........................................................................................... 15
1.4.1 ON/OFF Circuit .................................................................................................................... 15
Page 3 of 23
GW10078855.850.01.08.02
Page 4 of 23
GW10078855.850.01.08.02
1.1.4 Microcontroller
The micro controller J31 outputs demultiplexed lower order address bus and the
data bus on port ‘P0’ and higher order address bus on port ‘P2’.
The control signal RD , WR along with A8-A15 are provided to J28 (GAL) to
generate the chip select signals for accessing various peripherals.
The status signal PSEN is used for reading external program stored in EPROM J32.
The EA ‘LOW’ allows accessing the external program memory.
Different signals for controlling various operations of the entire system are provided
by the Micro controller on its ports P1, P3, P4, and P5.
The analog inputs AN0 - AN6 are used for monitoring analog signals.
The pins TxD and RxD are used for serial communication between the processor
and the external device such as a Computer for servicing.
The Service switch ST provided at pin P4.5 is used for operating the Unit in the
service mode.
Page 5 of 23
GW10078855.850.01.08.02
At the active ALE signal, the lower order address (A0-A7) is latched.
1.1.6 GAL
Since Memory mapping technique is used, the chip select signals for various
peripheral devices are generated by the Gated Array Logic IC J28. The inputs are
the address lines (A8 - A15), and the control signals RD and WR . The output lines
are CS1 to CS8 .
These output chip select lines are generated on the basis of the logic implemented
in the program stored in the GAL.
1.1.7 EPROM
The entire program is stored in a single EPROM J32.
1.1.8 NVRAM
The 2 Kbytes NVRAM stores the variables during the operation of the unit.
A0 - A10 and D0 - D7 are the address and the data lines provided to the RAM. The
first 2 Kbytes of entire memory space 0000H to 07FFH is used as RAM.
1.1.9 Keyboard
The keys are accepted through X12 from D601.
The keys kV+, kV -, mAs+, mAs-, ON, OFF are accepted on the data bus through
J38 and J37 at the address 9000H.
1.1.10 Display
The data bus, the control signal WR and the chip select signal CS2 are provided to
the Display Driver J24 on D601 through X1. Writing the control word and the string
of 8 byte data at the addresses 8000H and 8800H respectively displays the kV and
mAs values and lights up function related LED’s on D915.
Page 6 of 23
GW10078855.850.01.08.02
1.1.14 KV Feedback
The actual kVp across the tube is stepped down and fed back to the Master card
D915 at pins 1, of X 8 as PKV with respect to the common terminal at pin 3.
PKV is passed through the voltage follower J5 and then fed to the Differential
Amplifier formed by J5 to produce kVist used for kV regulation.
(KVist ratio is 1V = 20 kV)
As no differential configuration is used, NKV part is ignored by grounding it.
KVist is fed back to AN7of J31 for monitoring purpose.
1.1.15 IH Feedback
The filament current feedback is taken from the Interface card D602A through
X11.1.
J7 provides a gain of 5 to Ih.
The AC filament current Iheiz is converted in DC value Ih by RMS-DC converter J6.
J9 provides an appropriate gain to Ih (that can be adjusted by preset P5) so that the
ratio 1V = 1.13A holds true. Ih is fed back to J31 at AN2 for monitoring purpose and
the same is used for Ih regulation.
1.1.16 mA Feedback
The tube current (JR) is fed back at pin 5 of X 8 on D915 & is applied to the
Differential Amplifier formed by J4.
The output JR represents the actual tube current mAist which is fed back to J31 at
AN1 through J9 for monitoring purpose. JR is also used for tube current regulation.
Page 7 of 23
GW10078855.850.01.08.02
1.1.17 KV Regulation
The kVsoll and kVist values are fed to the P-I controller formed by IC J10.
The difference between the kVist and kVsoll value is amplified to produce an
appropriate error output, so as to compensate the error. The zener V12 limits the
maximum output voltage of the error amplifier to 10V.
The error amplifier output is fed to PLL IC J3, which is used as voltage to frequency
converter. The J10 output voltage produces an equivalent frequency REG at J3
output. For 10V, REG (the maximum firing frequency of the Main Inverter) is approx.
16.5 kHz. J3 is enabled by an active low INH2 signal provided by the microcontroller
J31 at P4.1 through J23. During HT, J3 is enabled through INH2 and J3 results in a
firing frequency of Main Inverter IGBT’s, to result in actual kV in line with the
required value.
Page 9 of 23
GW10078855.850.01.08.02
1.1.23.3 IH Maximum
This error is similar to JIM. This is monitored during preparation state and gives an
interrupt if the filament current exceeds the maximum permissible current that is 3.6
Amperes.
1.2.4 FPGA
· The supply of +3.3V, 2.5V and 1.2V is supplied to the FPGA-J20.
· The control signal clock and reset are provided for functioning of the FPGA.
· All the digital input, output and in/out signal are used to accessing various
peripherals of the Master Card D915A.
1.2.5 Multiplexer
· The multiplexer is used to supply 2.5V to the kV, JR and IH ADC during self
diagnosis
1.2.6 NVRAM
· The 256 Kbyte NVRAM, J31 stores the variables during the operation of the unit.
Page 10 of 23
GW10078855.850.01.08.02
1.2.10 kV Feedback
· The actual kVist(measurable at KV test point on D915A) across the tube is stepped
down and fed back to the Master card D915A at pins 1of X8 as PKV(measurable at
P_KV test point on D915A) with respect to the common terminal at pin 3.
· PKV is passed through the voltage follower and then fed to the Differential
Amplifier to produce kVist used for kV regulation.
(kVist ratio is 1V= 20 kV)
· Appropriate gain to kVist that can be adjusted by pot R9.
· As no differential configuration is used, NKV(measurable at N_KV test point on
D915A) part is ignored by grounding it.
1.2.11 IH Feedback
· The filament current feedback is taken from the Interface card D506 through X11.1.
· The AC filament current I(measurable at I test point on D915A) is converted in DC
value IH(measurable at IH test point on D915A) by RMS-DC converter.
· Gain is provided to IH (that can be adjusted by pot R72) so that the ratio 1V = 1A
holds true.
· IH is fed back to FPGA through ADC for monitoring purpose and the same is used
for IH regulation.
Page 11 of 23
GW10078855.850.01.08.02
1.2.12 mA Feedback
· The tube current JR(measurable at JR test point on D915A) is fed back at pin 5 of X8
on D915A & is applied to the Differential Amplifier.
· Appropriate gain to JR that can be adjusted by pot R43.
· The output JR represents the actual tube current mAist which is fed back to FPGA
through ADC for monitoring purpose and tube current regulation.
1.2.13 kV Regulation
· Depending on the kV value selected, it generates kVsoll value internally in FPGA.
When exposure is triggered, the kVsoll and kVist values are fed to the P-I controller
module in the FPGA. The difference between the kVist and kVsoll value is used to
produce an appropriate error output, so as to compensate the error. The error output
produces an equivalent frequency of inverter firing pulses M_INV1 and M_INV2.
This is given as the input to mains inverter which generates kV.
Page 12 of 23
GW10078855.850.01.08.02
3.3V
M_INV1
3.3V
M_INV2
16.5kHz
·
· The Main inverter firing pulses M_INV1 and M_INV2 are out of phase. These
signals are optically isolated by optocouplers and fed to the Main Inverter Driver
(D62 or D64) through X20.14 and X20.18.
1.2.17 Filament Regulator Firing Frequency
· The Filament inverter firing pulses F_INV1(measurable at F_INV1 test point on
D915A) and F_INV2(measurable at F_INV2 test point on D915A) have a maximum
frequency between them as 1.25 kHz. The pulse width of the pulses is 220µsec.
· The Filament inverter firing pulses F_INV1 and F_INV2 are out of phase. These
signals are inverted and fed to the Filament regulator firing circuit through X11.13
and X11.16.
Page 13 of 23
GW10078855.850.01.08.02
3.3V
F_INV1
3.3V
F_INV2
1.25kHz
1.2.19 FLASH
· The shorting link SW4 must be shorted for proper functioning of the FLASH.
Igbt Inverter Assembly houses IGBT’s configured as bridge inverter, along with
Inverter PCB D64 consisting of associated firing circuit, supply monitoring and short
circuit sensing circuit.
1.3.2 Firing
With selection of single step exposure release switch for Radiography or
Fluoroscopy foot switch and or Hand held Fluoro switch for x-ray generation, the
microcontroller prepares for the exposure loading the kV and mA values required for
the exposure into the DAC. This control circuit residing on the Master Card
D915/D915A then generates the firing pulses for the inverter which are available at
the connector X6 of the inverter PCB through interface PCB D602A.
Page 14 of 23
GW10078855.850.01.08.02
1.3.3 Inverter
U/V is the primary winding of the HT trafo which is in series with external resonating
capacitor C5 and external inductor L2.
With Disp1 pulse the current path is completed through DC(+), 3-1 of IG1, A3, U/V,
C5, L2, A2, 1-2 of IG2, DC(-). After 40 m Sec the gate pulse is turned off and the
IGBT’s are turned off. The resonating capacitor C5 which is now charged to a
voltage greater than the bank voltage discharges via the free-wheeling diodes in
anti-parallel to the IGBT’s.
With Disp2 pulse the current path is completed through DC(+), 3-1 of IG2, A2, L2,
C4, U/V, A3, 1-2 of IG1, DC(-). After 40 m Sec the gate pulse is turned off and the
IGBT’s are turned off. The resonating capacitor C4 which is again charged to a
voltage greater than the bank voltage discharges via the free-wheeling diodes. The
firing pulses have a maximum frequency of 16.5 kHz ± 0.3 kHz.
The presence of DC voltage is indicated by LED V1 on D64.
1.3.4 SCM
VCE of the IGBT’s is monitored continuously by the IGBT drivers. In case of short in
any IGBT’s VCE will increase resulting in generating of active low signal on ERR pin
of the driver. ERR pin is normally held high by a pull-up resistor in D602A. The ERR
pin can go low if the supply voltage falls below 13V instead of 15V. When this signal
is sensed by the microcontroller, firing pulses are inhibited and Error Code 11 is
displayed, and the IGBT’s are turned off.
Page 15 of 23
GW10078855.850.01.08.02
K2 gets completed and causes the Relay K2 to energise. This causes the contact
K2 to close and provide the ac voltage to the SMPS-1, PS-1, PS-2 (40 watts & 60
watts) and II supply input and Contactor Coil connections. As the coil K2 energises
the contact K2 (6, 8) is established and hence the unit remains ON even if the “ON”
switch is released.
When the switch OFF is pressed the Relay K1 coil is energized and this causes the
Normally Closed contact K1 (4, 6) of relay K1 to open thereby opening the circuit
connections for relay K2. The Unit thus is turned off.
The Relay US (K3) and LN (K6) coils are energized only when corresponding signal
from the Master Card D915/D915A is obtained. This is achieved through opto-
couplers J3 and J2 respectively. The relay ON signal from the Master Card
D915/D915A intern drives optocoupler J3 / J2 which in turn K3 / K6 relay. The VH
relay (k4) is used to switch over between 24V and 15V supply for large and small
focus respectively. It is energized through J7 optocoupler.
Page 17 of 23
GW10078855.850.01.08.02
The up/down movement controls are generated in this PCB and the interface to the
limit switches to limit the up and down extreme positions.
The signal available from the camera as an ABC (Automatic brightness control) is
feed to the input of the buffer amplifier constructed by using IC TLO82.The ABC
signal varies from -4 V to +7 V, this signal needs a voltage level conversion to be
applied as an ADR (Automatic Dose Regulation) to the port of master card. The
second stage is a summing amplifier used as a voltage level shifter by applying a
constant negative voltage at the summing input. This voltage can be adjusted by the
potentiometer. Refer Adjustment instructions in service instruction.
The output voltage is available as an ADR and varies as 0V to 5V DC, which is
proportional signal to the brightness signal from Camera.
1.5 Radiography & Fluoroscopy Display PCB D601
This PCB consists of
· ON / OFF switch.
· Parameter selection keys (kV & mAs)
· Service mode selection key(rear of the pcb)
· Seven segment display for Radiography (kV and mAs values).
· Control keys for the vertical movement of the motorized column
· Seven segment display for fluoroscopy kV and mA and cumulative fluoroscopy
time values
· Mode selection for standby/ fluoroscopy/radiography mode.
· Buzzer reset keys.
· Contrast mode key.
· Non ADR (Manual mode) key.
The control signals generated in this PCB by activation of the relevant keys are
sending to the Master card D915/D915A through the connecting cables.
2 Unit On D915/D915A
3 kV + kV increment D915/D915A
4 kV - kV decrement D915/D915A
5 mAs + mAs increment D915/D915A
6 mAs - mAs decrement D915/D915A
7 DL_Serv Service mode select D915/D915A
10 Up and Down Movement of motorised D602A
key column
Page 18 of 23
GW10078855.850.01.08.02
Page 19 of 23
GW10078855.850.01.08.02
Page 20 of 23
GW10078855.850.01.08.02
Page 21 of 23
GW10078855.850.01.08.02
2 Wiring Description
2.1 Mains Distribution
When the unit is switched ON the mains supply is directed through fuses U1,
U2.The Neutral is directly wired to the N input of the line filter whereas the phase is
directed to the P input of the line filter through the Emergency switch, and mains
voltage of 240V appears across X1.11 and X1.12 on D602A. At the same time 240
V also appears between i.e. at the primary input of transformer T1.The 24V from the
secondary of T1 appears between points X4.6 and X4.5 on D602A.This 24V AC is
rectified and regulated to give constant DC voltage of 24V on the D602A PCB.This
24V is used for operation of relays on the respective PCBs.The unit is switched ON
by pressing the switch on top panel; the circuit for the relay coil K2 and CS is
energised , on energizing K2 the mains voltage is available at the following points:
§ X1.8 – Input mains for power supply SMPS1.
§ X4.1 – Input mains for power supply PS1 and PS2 and II SUPPLY.
§ CS44 – Power supply for monitor trolley.
§ X1.6 – Input supply for DL contactor through normally open contact of relay K5.
After the unit is turned ON capacitor banks C1, C2 starts charging through charging
resistors R1//R2 (47R//47R),NO contact of CS(since CS is picked),NC contact of LN
relay and rectifier V1.
After 8 Secs of Switch ON the US relay picks up –
§ NC contacts, open disconnecting charging resistors R1&R2.
§ NO contact closes connecting mains voltage through NC contact of LN, V1, and
capacitor bank and across D64+ and D64-.
The line voltage is monitored by the microcontroller, if it falls below 200V AC the LN
relay picks up through optocoupler J2 on D602A
§ And opening the NC contact and disconnecting the mains voltage to the rectifier
input V1.
§ Causing NO contact to close and routing 0-220V AC from primary winding of
mains transformer T1 through 260V tapping to V1, capacitor bank and across
D64+ and D64-.
Page 22 of 23
GW10078855.850.01.08.02
When the unit is switched OFF by pressing the switch on D601, the relay coil
K1 on D602A is energised causing NC contacts of K1 to open thereby
disconnecting K2 relay. The DC voltage across the capacitor bank discharges
through K8 NC Contact and Resistor R2 and R3.
Page 23 of 23
GW10078855.850.01.08.02
Service Instructions
The reproduction
transmission or use of
this document or its
contents is not
permitted without
express written
authority. Offenders will
be liable for damages.
RESTRICTED GW10078855.898.01.15.02
Chapter Page Revision
All All 15
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are directed to contact one of the local offices of
Siemens or one of its affiliates before attempting installation or service procedures.
INDEX
INDEX .............................................................................3
1 SAFETY .....................................................................6
1.1 SAFETY PRECAUTIONS ................................................................................................................................................................. 6
1.2 OVERLOAD PROTECTION .............................................................................................................................................................. 6
1.3 RADIATION PROTECTION .............................................................................................................................................................. 7
1.4 MEASURING WITH OSCILLOSCOPE ................................................................................................................................................ 7
1.5 HANDLING PRECAUTIONS ............................................................................................................................................................. 7
1.6 MECHANICAL SAFETY ................................................................................................................................................................... 7
1.7 MAINTENANCE.............................................................................................................................................................................. 8
1.7.1 Product Disposal ........................................................................................................................ 9
1.7.2 Radioprotection Material ............................................................................................................ 9
1.7.3 Transformer Oil .......................................................................................................................... 9
1.7.4 Plastics....................................................................................................................................... 9
1.7.5 Electrolytic Capacitors ............................................................................................................... 9
2 SERVICING ............................................................. 10
2.1 TOOLS AND MEASURING INSTRUMENTS REQUIRED ...................................................................................................................... 10
2.2 REPLACING DAMAGED OR MISSING SCREWS ............................................................................................................................... 10
2.3 CLEANING AND DISINFECTING .................................................................................................................................................... 10
2.3.1 Cleaning the Unit ..................................................................................................................... 10
2.3.2 Disinfecting .............................................................................................................................. 11
2.4 PROTECTING THE SYSTEM FROM FLUIDS..................................................................................................................................... 11
2.5 ROUTINE C HECKS ...................................................................................................................................................................... 12
2.5.1 Daily checks ............................................................................................................................. 12
2.5.2 Monthly checks ......................................................................................................................... 13
2.5.3 Performance Check (Every Six months)..................................................................................... 13
2.6 CALIBRATION CHECK.................................................................................................................................................................. 13
2.6.1 Control Voltages....................................................................................................................... 13
2.6.2 Signals ..................................................................................................................................... 14
2.6.3 Potentiometer ........................................................................................................................... 15
2.6.4 Exposure .................................................................................................................................. 15
2.6.5 Timing ...................................................................................................................................... 15
2.6.6 kVp Measurement ..................................................................................................................... 16
2.6.7 mA / mAs Measurement............................................................................................................. 16
2.6.8 Tuning ...................................................................................................................................... 17
2.7 YEARLY CHECKS ........................................................................................................................................................................ 17
2.8 PARTS SUBJECT TO W EAR ......................................................................................................................................................... 17
2.9 CHECKS PRIOR TO SWITCH ON AFTER PROLONGED SHELF LIFE .................................................................................................. 18
2.9.2 Service programs available ....................................................................................................... 19
2.10 CHECKS PRIOR TO TRANSPORT ............................................................................................................................................. 22
GW10078855.898.01.15.02
PAGE 3 OF 141
6 MONITORS: ............................................................. 46
7 MEMORY MODULE OPERATIONS ........................ 49
7.1 USER INTERFACE: ...................................................................................................................................................................... 51
9 APPENDIX I ............................................................. 66
9.1 SPECIFIC ERROR CODE HANDLING............................................................................................................................................. 66
9.1.1 Unit not turning ON .................................................................................................................. 66
9.1.2 NO X-RAY ................................................................................................................................ 68
9.1.3 NO STANDBY .......................................................................................................................... 70
9.2 INITIALIZATION ERROR CODE ...................................................................................................................................................... 71
9.2.1 CODE 90 (EPROM CHECKSUM FAILURE)............................................................................ 71
9.2.2 CODE 96 (KV SOLL FAILURE) ............................................................................................... 71
9.2.3 CODE 97 (mA FAILURE)......................................................................................................... 71
9.2.4 CODE 99 (LAST RESET BY WATCH DOG TIMER) ................................................................. 71
9.3 STANDBY ERROR CODES............................................................................................................................................................ 73
9.3.1 CODE 02: +15 V Supply Error ................................................................................................. 73
9.3.2 CODE 03 & 04 : Iheiz < Istby & Iheiz > Istb ............................................................................ 74
9.3.3 CODE 05 kVist <>0 ................................................................................................................. 78
9.3.4 CODE 06: JR <> 0 .................................................................................................................. 79
9.3.5 CODE 33 : Main Inverter Short Circuit .................................................................................... 80
9.4 EXPOSURE ERROR CODES ................................................................................................................................................. 82
9.4.1 CODE 11 : Main Inverter Short Circuit .................................................................................... 82
9.4.2 CODE 12 : kVist > kVmax ........................................................................................................ 83
9.4.3 CODE 13 : Iheiz > Imax OR JR > Jrmax ................................................................................. 85
9.4.4 CodE 23 : JR > Jr max ............................................................................................................. 87
9.4.5 CODE 14 : kVist < kVsoll ......................................................................................................... 88
GW10078855.898.01.15.02
Page 4 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
INDEX
9.4.6 CODE 15 : JR < JRS ................................................................................................................ 90
9.4.7 CODE 17: Exposure terminated by Backup Timer ..................................................................... 94
9.4.8 CODE 18: Premature termination of Exposure ......................................................................... 96
9.4.9 CODE 21 : Iheiz > Iheiz maximum............................................................................................ 97
9.4.10 CODE 22: Maximum Preparation Time ................................................................................ 99
9.4.11 Interface PCB D602A......................................................................................................... 102
9.4.12 Master Card D915A ........................................................................................................... 103
GW10078855.898.01.15.02
PAGE 5 OF 141
1 Safety
1.1 Safety Precautions
In case of fault if the kVmA tries to build beyond the safe limit of X-ray tube the
protection circuit blocks the HT.
In case of fault if the Filament current tries to increase beyond the safe limit the
protection circuit blocks filament boosting.
Maximum permissible resistance between the Earthing lead terminal in the control
and on the HT transformer shall not exceed 0.1 ohms.
GW10078855.898.01.15.02
Page 6 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SAFETY
GW10078855.898.01.15.02
PAGE 7 OF 141
While moving various parts of the unit with hand, use of handgrips and handles is
recommended. The figure shows location where there is a danger of crushing for
patient or user due to parts moving relative to each other.
The system should be moved only manually, by the handles provided for this
purpose. Whenever exceptions must be made, note the points of potential injury
between the movable system parts and their guide openings. The system areas
marked in the drawings indicate points of crushing or impact hazards for the
patient or the operator.
Points of injury for operator Points of injury for patients.
1.7 Maintenance
All parts, modules of this product must be tested, inspected, for performance and
safety aspects every 12 months to ensure that the product functions properly and is
safe for patients, operating personnel and other persons.
Every 12 months, trained technical personnel should inspect and if necessary
replace system components of the unit, which may result in hazard due to excessive
wear and tear.
If more frequent inspections and maintenance are required by federal or local
regulations, ensure compliance with them.
GW10078855.898.01.15.02
Page 8 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SAFETY
GW10078855.898.01.15.02
PAGE 9 OF 141
2 Servicing
2.1 Tools and measuring instruments required
· Service tools.
· Digital Multimeter
· 2 channel storage oscilloscope
· mAs meter
· Megger
· Radiation detector
· Mechanical tool kit containing set of spanners, screwdrivers, Allen keys etc.
· FTP-2000
GW10078855.898.01.15.02
Page 10 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SERVICING
GW10078855.898.01.15.02
PAGE 11 OF 141
GW10078855.898.01.15.02
Page 12 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SERVICING
a) Cover single tank with multifold lead apron and switch on the fluoroscopy in
ADR mode. The kV and mA should reach to the maximum (110KV, 5mA).
b) Remove all objects from the x- ray beam path and switch on the Fluoroscopy
in a ADR mode. The kV and mA should reach to the minimum.
c) Check the image intensifier camera functions.
d) Last Image Hold.
e) ADR function.
f) Check the memory module functions
g) Memory Selection
h) Image Transfer (Memory write / Overwrite)
i) Image rotation clockwise.
j) Image rotation Counter Clockwise.
k) Image negation
l) Image contrast
m) Image averaging
GW10078855.898.01.15.02
PAGE 13 OF 141
2.6.2 Signals
Signal Measuring Point Range Adjustme Remarks
nt
KVsoll D915.TP.KVS 2.0 ± 0.1 V for D915.P7 Applicable for
During Exposure 60kV D915
KVist kV test point on D915A. 3.0± 0.1V for Pot R9 on Applicable for
During Exposure 60kV D915A D915A
Expos Capture the KV (TP KV) For e.g. @ 60 D915. P1 Applicable for
ure and the Ma(TP JR) kV, 10mAs D915
time waveform on the Time 238 – 262
oscilloscope and measure msec
the exposure time. (Refer to the
Exposure charts
section for
various
combinations of
kV & mAs for
units with 190-
240V.
GW10078855.898.01.15.02
Page 14 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SERVICING
2.6.3 Potentiometer
Potentiometers on D915A
R9 To adjust the tube voltage(kV)
R43 To adjust the tube current(JR)
R72 To adjust standby filament current (IH)
2.6.4 Exposure
2.6.5 Timing
GW10078855.898.01.15.02
PAGE 15 OF 141
2.6.6.1 KVp-meter
Ø Switch the unit ON.
Ø Place the kV sensor over the single tank and connect the cable to the meter.
Ø Set the kVp meter at 12pulse and 55-85kV ranges.
Ø Set exposure parameters as 60kV 20mAs.
Ø Release an exposure.
Ø The readings should be within ±5% of the selected value.
Ø Repeat with meter setting 77-150kV and exposure parameter setting as 90kV
20mAs.
2.6.6.2 Electrical
Ø Connect oscilloscope at D915.TP.kV/ (kV Test point on D915A).
Ø Set the exposure parameters as 60kV, 2.5mAs.
Ø Release an exposure.
Ø The recorded waveform should be within ± 5% of the set value. (1V Þ 20 kV)
Ø Repeat with 90kV 1.6mAs.
2.6.7.1 mAs-meter
Ø Remove mAs link on D602A PCB.
Ø Insert mAs-meter leads in the banana sockets mA+ and mA- on D602A.
Ø Set meter at DC mA, 200mAs and in the mAs mode.
Ø Set exposure parameters as 60kV 20mAs.
Ø Release an exposure.
Ø The reading should be within ± 5% of the selected value.
Ø Repeat for exposure parameters 90kV 16mAs.
Ø The actual tube current can be measured by setting the meter in mA mode.
Ø Set the exposure parameters as 60kV 50mAs.
Ø Release an exposure.
Ø During the exposure, the reading should be within ± 5% of the mA as per the
exposure chart.
GW10078855.898.01.15.02
Page 16 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SERVICING
Ø
2.6.7.2 Electrical
Ø Connect oscilloscope at D915.TP.JR/ (JR test point D915A).
Ø Set exposure parameters 60kV 2.5mAs.
Ø Release an exposure.
Ø The recorded waveform should be within ±5% of the mA as per the exposure
chart. (1VÞ20mA)
2.6.8 Tuning
It is necessary to re-tune the unit in case of undershoot or overshoot observed in the
mA wave shape.
GW10078855.898.01.15.02
PAGE 17 OF 141
7. Wheels
8. Stickers and symbols
9. Bearings, Rollers etc
10. Sprockets
11. Handles and knobs.
12. False covers
13. Threaded spindle and Threaded Bush
kV mAs kV mA min
3. Observe the gradual build up of DC voltage across the capacitors in the
inverter module.
4. Switch off the unit after the capacitor voltage reaches 150V DC.
5. Switch on the unit again.
6. Keep it on till the capacitor voltage reaches 250V
7. Repeat the above procedure for 5 times.
8. Switch on the unit and after 9 Secs the US contactor picks up
9. The voltage across the capacitors should be around 300-350V DC
GW10078855.898.01.15.02
Page 18 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SERVICING
2.9.2.1 Program 1
Not Used
2.9.2.2 Program 2
· To check the total number of exposures taken, enter the service mode and select
program 2.
- The display will be as below.
XXX XXX
- XXX XXX corresponds to the total number of exposures taken on the unit.
2.9.2.3 Program 3
· To check the type of error codes that occurred in the unit (maximum 20 errors), enter
the service mode and select program 3.
- The display will be
AA XX
- AA displays the serial number of the errors that have occurred, in chronological
order (max. 20) & XX displays the corresponding codes occurred on the unit
2.9.2.4 Program 4
· To erase all the error codes that have occurred in the unit, enter the service mode and
select program 4.
- The display will be
ErA Err
· Keep the mAs + key pressed for 4 seconds till the display becomes
0 Err
2.9.2.5 Program 5
· To store the Last Value of kV & mAs display, enter the service mode and select
program 5.
- The display will be
LAS VAL
- By entering into this program the last value displayed before turning OFF the unit
will be displayed. The same value is displayed next time when the unit is turned
ON.
2.9.2.6 Program 6
· To set the maximum kV & mAs possible with the unit, enter the service mode and
select program 6.
- The display will be as below
100 200
GW10078855.898.01.15.02
PAGE 19 OF 141
- 100 indicate the maximum kV available & 200 indicates the maximum mAs
available on the Unit.
· Set the required limit for kV and mAs using the kV /mAs increment and decrement
switches.
NOTE: In case the display values do not vary in standby by using the kV/mAs keys then enter
program 6 and set the kV & mAs to maximum using the kV /mAs increment and decrement
switches.
2.9.2.7 Program 7
(Applicable only for Microcontroller based Master Card D915)
· To adjust the REG frequency, enter the service mode and select Program 7.
- The display indicates.
Adj F
· Press the exposure hand switch and monitor REG signal on Digital storage
Oscilloscope at REG Test Point on D915.
- The display indicates
F On
- And max. Inverter firing frequency will be displayed on Oscilloscope.
· Adjust REG for 16.5 kHz ±0.3 kHz by using P4 on D915 card.
15V
REG
16.5kHz
2.9.2.8 Program 7
(Applicable only for FPGA based Master Card D915A)
· Generation of main inverter firing pulses, enter the service mode and select Program 7.
- The display indicates.
PUL 1
· Press the exposure hand switch and monitor M_INV1 signal on Digital storage
Oscilloscope by connecting the osc probe at M_INV1 Test Point on D915A.
- The display on the Oscilloscope at M_INV1 Test Point on D915A should show a
single inverter firing pulse.
GW10078855.898.01.15.02
Page 20 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
SERVICING
3.3V
M_INV1
20µsec
-
· Press the exposure hand switch and monitor M_INV2 signal on Digital storage
Oscilloscope by connecting the osc probe at M_INV2 Test Point on D915A.
- The display on the Oscilloscope at M_INV2 Test Point on D915A should show a
single inverter firing pulse.
3.3V
M_INV2
20µsec
Program 8
(Applicable only for FPGA based Master Card D915A)
- Not used
2.9.2.9 Program 9
(Applicable only for FPGA based Master Card D915A)
- Not used
2.9.2.10 Program 10
(Applicable only for FPGA based Master Card D915A)
· Using the mAs increment and mAs decrement keys select PR. 10.
· Press the DL_SERV switch on the backside of the Top panel.
· The display should read ‘5EV xxx’ (xx represents the software code version).
5EV xxx
GW10078855.898.01.15.02
PAGE 21 OF 141
· Press the DL_SERV switch again and the display should go back to Pr 10.
GW10078855.898.01.15.02
Page 22 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
EXPOSURE CHART
3 Exposure Chart
For units with 230V input
GW10078855.898.01.15.02
PAGE 23 OF 141
4 Error Codes
Initialisation Error Codes
EPROM Checksum Failure The EPROM Checksum is stored at 7FFEh and 7FFFh as a 16-bit word.
During self-Diagnostics the software calculates the checksum of the EPROM
and compares with the stored checksum.
96 kVsoll Failure During Self-Diagnostics the software outputs 7Fh to the D/A converter (B).
The 2.5V at the output of the D/A converter is Read by the µC through
Analog Port 4. The value read should be less than 73h and greater than 86h.
(I.e. between 2.25V and 2.62V)
97 JRS Failure During Self-Diagnostics the software outputs 7Fh to the D/A converter (A).
The 2.5V at the output of the D/A converter is Read by the µC through
Analog Port 3. The value read should be less than 73h and greater than 86h.
(I.e. between 2.25V and 2.62V)
99 Last Reset by watchdog The built-in Watchdog timer (WDT) is reset by the software every 25 msec. If
Timer due to some failure the software doesn’t reset the WDT, the WDT in turn will
reset the µC after 65 msec.
Standby Error Codes
02 +15V Supply Error The +15V supply from SMPS is polled by the µC through Analog port 0. The
+15V supply should be between +12V to +18V.
03 Iheiz < Istby The Filament Standby current is 1.5A. Iheiz read by the µC through Analog
port 2 should be less than 1.25A. i.e. 1.25V (Iheiz ratio : 1V = 1A)
04 Iheiz > Istby The maximum value of Standby current permitted is 1.8 A.
05 kVist <> 0 kVist is read by the µC through Analog port 7. During Standby the value of
kVist read should be Zero. (kVist ratio : 1 V = 20kV)
06 JR <> 0 JR is read by the µC through Analog port 1. During Standby the value of JR
read should be Zero. (JR ratio : 1 V = 20mA)
33 Main Inverter Short Circuit In standby mode if the Main Inverter Driver (Cable) is disconnected this error
gets activated.
Exposure Error Codes
11 Main Inverter Short Circuit This error is displayed when the drivers of the Main Inverter detect short
circuit.
12 kVist > kVmax The PkV is monitored for Max.115kV. If the actual value of kV is greater than
this i.e. if the voltage exceeds 3.83V error is displayed.
13 Iheiz > Imax OR JR > Maximum value of Iheiz above which error will be displayed is 4.0 A (4.0V).
Jrmax Maximum value of mAist is 71mA above which error is displayed.
14 kVist < kVsoll kVist is continuously polled during exposure. The value of kVist should not be
less than 85% of kVsoll
15 JR < JRS JR is continuously polled during exposure. The value of JR should not be
less than 50% of JRS.`
17 Backup Timer This error is displayed if regular means of terminating exposure fails and the
backup timer terminates exposure. The backup timer value is set to 2 * mSec
+100 mSec
18 Premature Exposure Exposure Release Switch is released before the mAs Integrator terminate the
Termination exposure.
21 Iheiz> Iheiz max This error is displayed if Iheiz exceeds the Iheiz max value.
22 Maximum Preparation This error is displayed if the exposure switch is in half pressed state for 15 sec
Time due to hardware failure.
51 kVist < kVsoll (Fluoro) Same as Code 14. This Error is displayed only during Fluoroscopy.
52 JR < JRS (Fluoro) Same as Code 15. This Error is displayed only during Fluoroscopy.
GW10078855.898.01.15.02
Page 24 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
ERROR CODES
For Master Card D915A:
Code Description Explanation
Initialization Codes
96 kVsoll Failure During Self Diagnostics, 2.5V supplied to the ADC. The read value should be greater
than 7Ah and less than 86h. (i.e. between 2.39V and 2.62V)
97 JRS Failure During Self Diagnostics, 2.5V supplied to the ADC. The read value should be greater
than 7Ah and less than 86h. (i.e. between 2.39V and 2.62V)
Standby Codes
02 +15V Supply Error The +15V supply from SMPS is polled by the µC through Analog port 0. The +15V
supply should be between +12V to +18V.
03 Iheiz < Istby The Filament Standby current is 1.5A. Iheiz should be greater than 1.25A. i.e. 1.25V
(Iheiz ratio : 1V = 1A)
04 Iheiz > Istby The maximum value of Standby current permitted is 1.8 A.
05 kVist <> 0 During Standby the value of kVist read should be Zero. (kVist ratio : 1 V = 20kV)
06 JR <> 0 During Standby the value of JR read should be Zero. (JR ratio : 1 V = 20mA)
33 Main Inverter Short Circuit In standby mode if the Main Inverter Driver (Cable) is disconnected this error gets
activated.
Exposure Codes
11 Main Inverter Short Circuit This error is displayed when the drivers of the Main Inverter detect short circuit.
12 kVist > kVmax The PkV is monitored for Max.115kV. If the actual value of kV is greater than this i.e.
if the voltage cross 3.83V error is displayed.
13 Iheiz > Imax Maximum value of Iheiz above which error will be displayed is 4.0 A (4.0V).
14 kVist < kVsoll kVist is continuously polled during exposure. The value of kVist should be greater
than 85% of kVsoll
15 JR < JRS JR is continuously polled during exposure. The value of JR should be greater than 50%
of JRS.
17 Backup Timer This error is displayed if regular means of terminating exposure fails and the backup
timer terminates exposure.
The backup timer value is set to 2 * mSec +100
18 Premature Exposure Exposure Release Switch is released before the mAs Integrator terminate the exposure.
Termination
21 Iheiz> Iheiz max This error is displayed if Iheiz exceeds the Iheiz max value.
22 Maximum Preparation Time This error is displayed if exposure mode is not entered by the system after exposure
switch selection for more than 15 sec because of hardware failure.
23 JR > Jrmax Maximum value of mAist above which CODE is displayed is 71mA. (3.59V)
51 kVist < kVsoll (Fluoro) Same as Code 14. This Error is displayed only during Fluoroscopy.
52 JR < JRS (Fluoro) Same as Code 15. This Error is displayed only during Fluoroscopy.
GW10078855.898.01.15.02
PAGE 25 OF 141
Troubleshooting
Troubleshooting In case of errors, which Impair radiography or fluoroscopy, perform
the following checks. The components can be identified as shown in the picture.
L2 V1
C4
C5
SMPS1 D64
PS1
D602A
PS2 L3
CAMERA
SMPS (PS3)
GW10078855.898.01.15.02
Page 26 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
TROUBLESHOOTING
U1
U2
CS, DL, US
Contactor
T1
C1
M1 C3
DL
TERMINAL BLOCK
GW10078855.898.01.15.02
PAGE 27 OF 141
GW10078855.898.01.15.02
Page 28 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
TROUBLESHOOTING
LED’s on D915A
D31 Green +3.3V on board
D35 Green +1.5V on board
D30 Green +2.5V on board
D34 Green +5V on board
D40 Green +15V on board
D41 Green -15V on board
GW10078855.898.01.15.02
PAGE 29 OF 141
-NC- = No Connection
GW10078855.898.01.15.02
Page 30 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
TROUBLESHOOTING
Ø Measure the CAL frequency at the CAL TP on D915(For D915A: Measure the
F_INV1 and F_INV2 frequency at the test points on D915A.)
Ø CAL freq. should be within 1.25 to 1.5kHz (For D915A : The frequency
between F_INV1 and F_INV2 should be within 1.25 to 1.5kHz)
Ø If not,tune P3 to set CAL within the above range. (For D915A: If not, tune Pot
R72 to set F_INV1 and F_INV2 within the above range.)
GW10078855.898.01.15.02
PAGE 31 OF 141
3.3V
F_INV1
3.3V
F_INV2
1.25kHz
GW10078855.898.01.15.02
Page 32 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
TROUBLESHOOTING
1V= 30kV at kVsoll.(Note: this not applicable for D915A)
Ø The readings should be within ±5% of the selected value.
Ø Repeat the measurement with the following parameters
1. 40kV 10mAs
2. 81kV 16mAs
3. 100kV 20mAs
GW10078855.898.01.15.02
PAGE 33 OF 141
Ø The mAs values should be within ±5% for values > 20 mAs and ±10 % for
mAs £ 20 mAs.
Ø If the values are not within the above range, tune P1 on D915 to adjust the
mAs values.
After resistance adjustment, Use FTP-2000 pattern to correct the ADR value
precisely, Place FTP-2000 on image intensifier as shown in picture, Select the
Fluoroscopy ADR mode In Anti-Isowatt mode, adjust the potentiometer on D602A
PCB till kV, mA displays 62-63 kV, 1.7-1.8 mA respectively.
GW10078855.898.01.15.02
Page 34 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
TROUBLESHOOTING
GW10078855.898.01.15.02
Page 36 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
CAMERA CONTROL
5 CAMERA CONTROL
5.1 INTRODUCTION
The Image intensifier camera (IIC) consists of a PCB and lens assembly. At the
heart of the IIC system, is the video processor. The video processor acquires
images from the Image Intensifier interface and performs operations on the acquired
image. The power subsystem provides regulated power to the video processor and
the other components. Clocks needed for the various digital components are derived
from the clock generation subsystem. The X-ray generator interface enables
connection with the X-ray generator. The configuration interface is used to set the
various adjustable parameters of the IIC.
5.2 RATINGS
GW10078855.898.01.15.02
PAGE 37 OF 141
Environmental
Temperature Humidity Air pressure
Operation 10 to 40 °C 30 – 85% RH 70 – 106 kPa
max. (not
condensing)
Storage & Transportation -15 to 45 °C 10 – 90%max.(not 50 – 106 kPa
condensing)
LIH function
When the LIH ON (X-ray OFF) signal is input, writing of the video signal into frame
memory is halted and the signal in the memory is played back continuously until the
LIH OFF (X-ray ON) arrives to obtain static images. It is possible to switch between
frame read-out and field read-out. When the LIH is OFF, the memory image is
updated and the LIH image is cleared.
GW10078855.898.01.15.02
Page 38 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
CAMERA CONTROL
Note: During the replacement of image intensifier camera (IIC) (as per PII 5609-
008_EUA_01S_00), ensure the proper orientation of the DSUB connector. Refer
below picture
Female 15 DSUB
connector from
Generator
Male 15 DSUB
Ensure the
connector from
orientation of IIC
DSUB
GW10078855.898.01.15.02
PAGE 39 OF 141
2. check and adjust the camera parameters displayed on seven segment by using
the rotary switch and INC/DEC switch
GW10078855.898.01.15.02
Page 40 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
CAMERA CONTROL
Increment
switch Decrement
switch
Seven segment
for displaying Rotary switch
camera for position
parameters selection
Default value
3. Check as per the inspection sheet supplied with the product whether the system
sensitivity has decreased as compared with that at the time of installation or not. . If
the sensitivity has decreased, check whether there is dust on the CMOS surface. If
there is dust on the CMOS surface, wipe off the dust with a clean soft cloth. If there
is no dust on the CMOS surface, the cause is due to the decrease of the II
brightness. Compensate for low brightness by widening the lens aperture.
However, do not open the aperture wider than F/2, to prevent peripheral resolution
from being decreased.
GW10078855.898.01.15.02
PAGE 41 OF 141
Note: For any unit between serial number 03336 to 03568, if CRT monitor flicker is
observed i.e. an intermittent horizontal flickering pattern (it will be different from text
flicker and only one horizontal line of the circular image will flicker) is seen then the
normal trouble shooting steps to set vertical linearity and vertical image size need to
be followed as shown in the monitor manual. In case the same is not solved, the
value of R121 potentiometer in camera card should be set to 110ohms as shown
below:
GW10078855.898.01.15.02
Page 42 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
CAMERA CONTROL
4. Use allenkey and open all 4 SOCKET HEAD SCREW M3X30 and remove PWR
and CFG PCB from IIC.
6. Use Multimeter and set R121 potentiometer resistor value to 110 Ohm
between pin 2 and pin3.
GW10078855.898.01.15.02
PAGE 43 OF 141
R121 Potentiometer
7. After Adjustment Place the threaded POWER PCB STANDOFF (along with
sticked Press pan washer) on the IIC PCB.
8. Place the PWR and CFG PCB on the all four POWER PCB STANDOFF
such that the mounting holes on the PWRCFG PCB align with the mounting
holes on the threaded POWER PCB STANDOFF and Complete the
GW10078855.898.01.15.02
Page 44 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
CAMERA CONTROL
assembly by mounting the final fastening screw/washer on the PWRCFG
PCB and ensure that the assembly is tight enough.
GW10078855.898.01.15.02
PAGE 45 OF 141
6 Monitors:
M10082944-17” TFT PC GRADE MONITOR
Refer PII 5609-007_EUA_01S_01 while upgrading earlier MM5E with M08645918-
Monitor 15'' to M10082944-17” TFT PC GRADE MONITOR
Monitors parameter setting:
TFT monitor supply with a default setting, few below parameter needs to adjust
during every monitor replacement.
1. Set the parameters on the monitors as mentioned above.
a. Image stetting: Image pixel : 85, phase: 60
b. Position: Horizontal: 100, Vertical : 32
2. Exchange both the VGA cables on the memory module under ON
condition.
GW10078855.898.01.15.02
Page 46 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MONITORS:
Fig. 3 Fig. 4
3. Connect the output of the adapter and BNC cable to both the monitors as
shown in Fig. 5.
Fig. 5
GW10078855.898.01.15.02
PAGE 47 OF 141
Fig. 6 Fig. 7
5. Connect BNC cable from monitor to memory module video OUT1 and video
OUT2 respectively & change memory module toggle switch to CCIR position
i.e to BNC configuration as shown Fig. 8.
Fig. 8
M4 Ornamental screws
GW10078855.898.01.15.02
Page 48 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MEMORY MODULE OPERATIONS
Powering up:
When system is switched ON XC-V boots up and ‘Blinking mirror LED’ indicates the
booting operation which may take up to 30 seconds. Subsequently images are
loaded from internal non-volatile storage, and message “Loading Image” appears on
the monitor.
Note: If the system fails to boot up, a continuous beep is heard.
After successful booting, Monitor1 shows a bar pattern. The multi gray bar pattern is
used to do any brightness and contrast adjustments on the monitor.
Monitor2 shows any loaded image from the non-volatile memory present when the
unit was stitched off last.
The image number (“01” to “25” or “REF”) is displayed on the top monitor2.
Video Monitors:
XC-V supports 2 monitors as describe above.
LIH monitor (Monitor1) is used to display Live/ LIH – captured images. All images
saved by foot switch operation (in LIH mode) are saved to this monitor.
Monitor-2 is used to display any of the 25 (100 optional) images from the built in
image memory of the unit.
Image selection on Monitor-2: Under normal operation of the memory unit, the rotary
dial can be rotated (left/ right) to select images on the Monitor-2 and the image
number is shown on top of the screen. (Image numbers 1-25 [0r 100] & REF is
displayed on top of the monitor)
GW10078855.898.01.15.02
PAGE 49 OF 141
On power up, all these (25/100) images are automatically restored from the non
volatile memory. Also, after power failures, the images are restored automatically
without any corruption.
(In pulse mode this permanent memory storage operation has to be initiated
manually)
In USB option:
GW10078855.898.01.15.02
Page 50 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MEMORY MODULE OPERATIONS
In LAN option:
For transfer of images to / from a PC over LAN, the XC- IMAGER s/w application
should RUN on the PC, and XC-V should be connected to the PC.
Refer s/w reference guide for details.
Recall key
XC-V
Read key
USB/LAN
Store key
ROTARY DIAL:
A rotary dial is present on the right side of the front panel. This is used as a flexible
input device to change images on the REF monitor, adjust rotation, or edit
parameters in the system configuration.
The Rotary Dial is similar to the tuning control knob on a Radio, and can be rotated
clockwise or anti-clockwise. The Dial could also be depressed (CLICKed) to perform
GW10078855.898.01.15.02
PAGE 51 OF 141
USB connector:
A USB2.0 full speed connector is present on the front panel to connect pen drives for
storing / reading captured images.
OPERATION:
1. MENU KEY:
Menu key is used to select the operating mode. The MODE is cycled as LIH-> LIVE->
PULSE-> CONFIG and LIH again.
Present mode is displayed in Monitor1.
LIVE-Mode:
The live image is shown when X-ray is ON, but image is not captured when the
footswitch is released.
This mode can be used to compare the live image against an image captured earlier
(on the LIH Monitor)
Configure-Mode:
This is to set the different configuration of the system such as File back up mode, IP
address setting for LAN, Date & Time setting and blanking circle position setting.
To enter this mode, select CONFIG by MENU key and CLICK rotary dial. (as shown
below)
CONFIG
AVG: 04
ROTN: 000
Click to config
X-Ray OFF
GW10078855.898.01.15.02
Page 52 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MEMORY MODULE OPERATIONS
LIH Monitor
Turning the Rotary dial cycles through sub menus FILE -> IP -> TIME -> CENTRE->
and FILE again. Click rotary dial to enter the required sub menu.
CONFIG
AVG: 04
ROTN: 000
USB
X-Ray OFF
This mode selects the default back up mode of the XC-V. (USB pen drive or LAN).
When FILE is displayed, Click rotary dial to enter this mode as shown above. Rotate
rotary dial to select file back up setting as USB or LAN.
Press menu key / click rotary dial to exit this mode.
(Not applicable for XC-VBM).
IP
AVG: 04
ROTN: 000
192.168.002.101
X-Ray OFF
From CONFIG mode, select IP and click encoder to enter this sub-menu.
IP address consists of four numerical address fields. (192, 168, 002, 107 in the above
image).
These address fields start blinking indicating user EDIT mode.
After editing all fields, the caption “IP saved. Restart” will be displayed for 3 seconds
(as shown below). The unit has to be restarted to effect the changes by removing and
re-applying power.
GW10078855.898.01.15.02
PAGE 53 OF 141
IP
AVG: 04
ROTN: 000
IP saved, Restart
X-Ray OFF
IP
AVG: 04
ROTN: 000
2008.01.01-01:01:01
X-Ray OFF
Select Time sub menu from config. Mode and change the time as in the previous
procedure. (YYYY.MM.DD – HH:MM:SS, YYYY- Year, MM-Month, DD-Date, HH-
Hour in 24hr. format, MM-minutes, SS-Seconds)
From CONFIG mode, select CENTRE and click encoder to enter this sub-menu.
CENTRE
AVG: 04
ROTN: 000
X = 126, Y = 25
X-Ray OFF
GW10078855.898.01.15.02
Page 54 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MEMORY MODULE OPERATIONS
(PRESS MENU to abort this operation if required.)
To save the settings Press SAVE key.
The unit has to be restarted by removing and reapplying power (RESTART message
is displayed on the screen).
2. AVG Key:
This key sets the number of video frames averaged for noise reduction (1, 2, 4, 8 or
16). Successive depression of this key changes the value. (default 4)
3. ROTATE key:
This key enables the Image rotation operation. Rotation is possible on the LIH
monitor only. The LED above this key lit when this mode is selected.
The default rotation angle step is 20 degrees, Clicking the encoder dial, changes the
step angle to 2 (or 1 depending on model) degrees (and back to 20 degrees).
4. MIRROR Key:
This is to view the Mirror view of the image displayed on the LIH monitor. The LED
above this key is lit when this mode is selected.
Mirroring is applied only on real time live images or images captured directly on the
LIH monitor.
Mirror is not applied for any images transferred from Monitor-2 (by RECALL key) or
by READ operation from USB / LAN.
5. NEG Key:
This (NEGATIVE) key is used to show gray scale negative of images displayed on
the monitors. (Press the NEG key again to display the normal image.)
6. CNTRST Key:
Image is captured with 256 levels (values 0-255) of grey scales (brightness levels).
Contrast processing operation involves stretching a portion of the Grey scale (say
between 30 & 80) to full brightness range so that the detailed information in this
range is amplified and displayed.
Press “Contrast” key to enter this mode, when following screen appears with a
Contrast Slider.
The shaded region of SLIDE BAR shows the portion of the image being stretched /
enhanced. (Extreme left side of the bar indicated image brightness of 0 (black) and
right most end indicates image brightness of255 (white).
GW10078855.898.01.15.02
PAGE 55 OF 141
This shaded region can be expanded / reduced OR moved within the intensity region
of image/ slider bar.
7. SAVE key:
This key is used to save the image on LIH Monitor to the current memory location on
the Monitor-2 (image numbers 01 to 25 (100) or REF image). For XC-VPC and above
models the image is also transferred to the permanent image storage.
8. RECALL Key:
The RECALL key is used to transfer the image displayed on Monitor-2 to LIH Monitor
(applicable only when X-ray is OFF / Foot switch is released).
This key can be used to selectively bring images from monitor-2 to LIH monitor and
then to STORE them to USB/LAN.
9. REF Key:
10. READ and STORE Keys ( Image store & Retrieval operations)
USB-pen drive:
Note: For this the File option in CONFIG. Sub menu should be = USB.
(images in the pen-drive are sequentially numbered (3 digit FILE-No) with a Time
stamp (Yr.Mo.Dt. Hr.Min. Sec). This image number can be used as a reference
number, to READ the images from the XC-V itself)
GW10078855.898.01.15.02
Page 56 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MEMORY MODULE OPERATIONS
3. Message “STORE USB” is displayed, and image on the LIH-Monitor is saved
to next available sequential FILE Number.
(STORE Key – LED also lights during this period).
Note: do not remove the pen drive during this period.
4. After successful operation, the image FILE – No. and time stamp is
Displayed) as below.
LIH
AVG: 04
ROTN: 000
(After inserting / re-inserting the pen drive wait for 15 seconds before doing next
operation)
6. insert USB – pen drive into the front panel – USB connector (after inserting a
drive, wait for 15 seconds before doing a read/ store operation).
7. Press “READ” key. Following screen appears, with a blinking Image File-No.
(003 in following figure)
LIH
AVG: 04
ROTN: 000
File : 003
X-Ray OFF
GW10078855.898.01.15.02
PAGE 57 OF 141
8. User may EDIT the FILE-No, using Rotary Dial. Rotate Dial to change the
first digit, press encoder to move to the next etc.
After editing 3rd digit, click dial again to start Read operation.
Note: do not remove the pen drive during this period
LIH
AVG: 04
ROTN: 000
X-Ray OFF
Store LAN
X-Ray OFF
GW10078855.898.01.15.02
Page 58 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MEMORY MODULE OPERATIONS
If LAN is not connected, or s/w application is not running on PC or IP address is
different, following message is displayed.
LIH
AVG: 04
ROTN: 000
Waiting LAN
X-Ray OFF
LIH
AVG: 04
ROTN: 000
READ LAN
X-Ray OFF
(if LAN connection is not established before doing the STORE / READ operations
“ Waiting LAN” appears on the monitor)
For a dedicated LAN connection (X-Ray controller and a single computer) ‘cross
wired’ cable should be used as shown in fig. If it is to be connected by using a
Hub / Switch / Router a standard LAN cable can be used.
GW10078855.898.01.15.02
PAGE 59 OF 141
GW10078855.898.01.15.02
Page 60 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MECHANICAL CHECKS AND ADJUSTMENTS
8.1.2 BEARINGS
Following bearings will need periodic adjustments.
IV. Floating bearings and delring roller fitted on the Trag arm.
V. Main Eccentric bearing pin assembly and Eccentric guide bearing pin
assembly fitted on the Trag arm holder.
GW10078855.898.01.15.02
PAGE 61 OF 141
8.1.3 V- BELT
The V-Belt is used for the transfer of power from the motor pulley to the main pulley.
If the tension in the belt is not sufficient or if it is loose, there will be loss of power
due to slip, resulting in noisy up-down motion, wobbling of the belt, rubbing sound,
jerks etc, which can cause inconvenience to the doctor during operation. Hence the
belt has to be maintained in a tight position.
1. Lift the unit from the rear side and place some wooden support below the rear
wheels so that there can be an access for tightening the belt.
2. Now, remove the housing cover.
3. Loosen the screws of the motor from bottom side of the trolley base.
4. Pull the motor towards the backside so that the belt is under tension. At this point
tighten the screws of the motor.
5. Place the unit back on the floor.
6. Refix the covers.
7. In case the motor fails during use of equipment and the C-Arm needs to be
moved it can be done by moving the V-Belt below the housing trolley base.
8.1.4 CHAIN
Roll-on chain is used in the steering mechanism. The chain is routed through the
sprockets of two rear wheels, 4 idler sprockets, and the drive shaft. The chain is in
two halves and there are two adjustable nuts provided to adjust the tension in the
chain.
When the chain is loose, there is a tendency that the chain can slip off from the
sprockets, hence proper tension should be maintained in the chain.
For adjusting the tension in the chain, following are the steps
a) Lift the unit up from the rear side and place a wooden platform below both the
rear wheels.
b) Locate the adjustable nuts in the chain.
GW10078855.898.01.15.02
Page 62 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
MECHANICAL CHECKS AND ADJUSTMENTS
c) Loosen the check nut (M5) used to lock the movement of the screws.
d) Tighten the adjustable nut in the clockwise/anticlockwise direction as
required.
e) Lock the screws with the check nut so that the adjustable nut does not turn.
f) Place the unit on the floor and operate the steering lever to check the
movement.
8.2.2 Bearings
Bearings, generally doesn’t get worn out under clinical conditions. Only possibility
that cannot be ruled out is the wearing out of balls after a period of five to eight
years. This may cause noise in the movement and misalignment of movement.
Hence, it is the duty of the service engineer to clean and grease the bearings
regularly.
GW10078855.898.01.15.02
PAGE 65 OF 141
9 Appendix I
9.1 Specific Error Code Handling
Note: Refer D602 as D602A PCB
9.1.1 Unit not turning ON
GW10078855.898.01.15.02
Page 66 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 67 OF 141
9.1.2 NO X-RAY
NO X-RAY Contd….
GW10078855.898.01.15.02
Page 68 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 69 OF 141
9.1.3 NO STANDBY
GW10078855.898.01.15.02
Page 70 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
D915
GW10078855.898.01.15.02
PAGE 71 OF 141
D915
D915
D915
GW10078855.898.01.15.02
Page 72 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
D915/D915A
D915/D915A
GW10078855.898.01.15.02
PAGE 73 OF 141
9.3.2 CODE 03 & 04 : Iheiz < Istby & Iheiz > Istb
For Microcontroller based Master Card D915:
GW10078855.898.01.15.02
Page 74 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 75 OF 141
and F_INV2=1.25kHz
to 1.5kHz on D915A
GW10078855.898.01.15.02
Page 76 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 77 OF 141
GW10078855.898.01.15.02
Page 78 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
No
GW10078855.898.01.15.02
PAGE 79 OF 141
GW10078855.898.01.15.02
Page 80 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 81 OF 141
GW10078855.898.01.15.02
Page 82 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
IS KVM PULSE NO
GOING HIGH ?
A
YES
IS R=20kOhm IS R=20kOhm
between pins(X8.1 NO between pins(X2.1 YES Replace cable
& X8.2) and (x8.1 & X2.2) on D602A X8
& X8.3) ?
No
YES
Change D512
STOP
GW10078855.898.01.15.02
PAGE 83 OF 141
START
NO
YES
Replace D915A
Switch OFF the Unit
IS R=20kOhm IS R=20kOhm
between pins(X8.1 NO between pins(X2.1 YES Replace cable
& X8.2) and (x8.1 & X2.2) on D602A X8
& X8.3) ?
No
YES
STOP
GW10078855.898.01.15.02
Page 84 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
For
GW10078855.898.01.15.02
PAGE 85 OF 141
YES
YES
YES
NO
NO
YES
GW10078855.898.01.15.02
Page 86 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
YES
YES
NO
Is R=50Ω at pin
X8.4 & X8.5 and
X8.5&X8.6? at
Radiography
YES
GW10078855.898.01.15.02
PAGE 87 OF 141
GW10078855.898.01.15.02
Page 88 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 89 OF 141
GW10078855.898.01.15.02
Page 90 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 91 OF 141
GW10078855.898.01.15.02
Page 92 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 93 OF 141
GW10078855.898.01.15.02
Page 94 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 95 OF 141
GW10078855.898.01.15.02
Page 96 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 97 OF 141
GW10078855.898.01.15.02
Page 98 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
APPENDIX I
GW10078855.898.01.15.02
PAGE 99 OF 141
START
TRIGGER A RADIOGRAPHY
No
IS CODE 22
DISPLAYED?
Yes
CHANGE D915A
STOP
GW10078855.898.01.15.02
Page 100 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
GW10078855.898.01.15.02
PAGE 101 OF 141
GW10078855.898.01.15.02
Page 102 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
GW10078855.898.01.15.02
PAGE 103 OF 141
GW10078855.898.01.15.02
Page 104 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
10 Replacement of Parts
10.1 Replacing the single tank
To replace the single tank, remove it from the C arm as follows:
Ø Bring the C-arm in the horizontal position such that the axis of the x-ray
tube and II are in the horizontal plane fully extended. Lock the brake of the
trag arm so that the C arm does not move. Place a wooden block or any
packing material of appropriate size below the base tray of the C-arm so
that on removing the single there is no damage to the base tray as well as
personnel injury as the unit may fall behind due to balancing problem.
Ø Remove the side cover and top cover by removing the ornamental screw.
Ø Disconnect the Single Tank connections by removing connections from
D512 and the earthing point.
Ø Remove the temperature sensor connection.
Ø Loosen the Socket headed screw (3 nos at bottom of single tank mounting
plate) used to anchor the single tank to the C-arm using Allen key.
Ø Lift the single tank off vertically and place it on a soft base.
Ø Detach the collimator by loosening 4x alignment studs
Ø Mount the new single tank with the collimator
Ø Make the connections in D512 and temperature sensor connections.
Ø Check alignment of X-ray beam with II and adjust if necessary.
Ø Mount the collimator with the help of 4X alignment studs.
Ø Do collimator centring with the help of 4X alignment studs, so as to align
the X-ray beam over the 9” II system.
GW10078855.898.01.15.02
PAGE 105 OF 141
GW10078855.898.01.15.02
Page 106 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
Counter Weight
GW10078855.898.01.15.02
PAGE 107 OF 141
· Keep the C-Arm in the position shown below such that the II and the single tank are in
one level.
· Ensure all brake movements of the C-Am are locked and then remove the II mounted on
the c-arm.
To remove the II refer following steps:
Rubber stopper
False cap
2. Remove the II cap by removing the ornamental screw (2x) and the CSK screws on the
sides.
Ornamental
screw
CSK screw
3. Disconnect the cables from II and the camera.
GW10078855.898.01.15.02
Page 108 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
5. Dismantle the II by removing the hardware (4x). One person should be holding the II
carefully to ensure it does not fall off (weight of II is approximately 25 Kg). Keep the
II carefully in a safe place.
Soc. Hd.
screw
Rubber
stopper
Hardware (4x)
6. Remove the II mounting bracket by unscrewing the soc. hd. screws (4x).
GW10078855.898.01.15.02
PAGE 109 OF 141
II mounting
bracket
· Remove the single tank cover by unscrewing the ornamental screw (4x).
GW10078855.898.01.15.02
Page 110 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
Disconnect these
wires. Refer wiring
diagram for
connections
GW10078855.898.01.15.02
PAGE 111 OF 141
· Remove the fixing plate C-arm cable by removing the 3 screws provided.
Screw
· Carefully pull out the single tank cable from the C-arm without damaging the
connectors.
· Remove the single tank cable from the fixing plate for C-arm cable by unscrewing the
nut on the metal gland.
· Hold the nut on the metal gland with a spanner and rotate the fixing plate for c-arm in
anticlockwise direction, this will loosen the nut.
Spanner
Nut
GW10078855.898.01.15.02
Page 112 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
Slide out the fixing plate for c-arm cable and the nut from the single tank cable.
· Remove the single tank cable anchored through the Rubber nozzle between Cable holder
fitted on the trag arm & Cable holder plate-upper.
· To disconnect the cable connections at the housing end. Refer the following steps.
2. Remove the cap below the handle and place it in a safe place.
Screw
Steering handle
GW10078855.898.01.15.02
PAGE 113 OF 141
3. Remove the Relief cover from the top panel by unscrewing the ornamental screws
(4x).
Relief Cover
4. Unwind the mains cable & foot switch cable from the cable winder and Place them
aside.
Rear cover
5. Remove the front cover LH, front cover RH, Rear cover and the front cover plate from
the unit by unscrewing the ornamental screws.
Ornamental screw
GW10078855.898.01.15.02
Page 114 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
6. Cut the cable ties and disconnect the cable from the terminal block, earthing terminal,
DL contactor, inverter card D64 and D602A PCB.
D602A
PCB
Single tank
cable
Inverter card
D64
Terminal block
Contactor
Earthing
terminal
GW10078855.898.01.15.02
PAGE 115 OF 141
7. Disconnect the cable from the display PCB mounted on the top panel.
8. Remove the top panel by removing the M4(6x) nuts on either side.
Earthing wire
9. Disconnect the wires connected to the safety switch mounted on the top panel and also
the earthing wire from the top panel
Safety switch
Safety switch
cable
GW10078855.898.01.15.02
Page 116 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
10. Loosen the nut on the metal gland of the single tank cable from the top panel side and
pull out the cable carefully without damaging the other components.
Top panel
GW10078855.898.01.15.02
PAGE 117 OF 141
Note:
· Kindly note connection of PTR connector before disconnecting or Refer
wiring diagram during connections.
Green PTR
Connector
Metal gland
· Carefully slide the fixing plate for c-arm cable and the nut through the single tank cable.
GW10078855.898.01.15.02
Page 118 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
· Mount the fixing plate to the single tank cable by holding the nut on the metal gland
firmly with a spanner and rotate the fixing plate in clockwise direction.
Note:
· Ensure that the fixing plate for c-arm and the nut is properly tightened.
· When tightening take care that the metal gland does not rotate as this will
damage the cable due to over tightening.
Spanner
Nut
· Carefully pass the single tank cable through the C-arm without damaging the connectors.
Note:
· Ensure the cables are routed through the c-arm in the correct direction.
GW10078855.898.01.15.02
PAGE 119 OF 141
· Route the II interface cable through the c-arm such that it is out at the II end.
II interface
cable
· Route the single tank cables through the c-arm such that the cables come out at the single
tank end and do the necessary connections. Refer wiring diagram for the connections.
Note:
· Ensure that the contact washer is placed after every earth wire while
connecting the ground wire to the single tank.
GW10078855.898.01.15.02
Page 120 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
· Mount the II mounting bracket to the c-arm by tightening the soc. hd. screw.
II mounting
bracket
· Mount the II on to the II mounting bracket and tighten with soc hd screw M5X16 (4x)
and spring washer A5 (4x). One person should hold the II carefully until the hardware’s
are tightened.
Note:
· New soc hd screw M5X16 (4x) and spring washer A5 (4x) has to be
used for mounting the II.
· Do not reuse the removed hardware for mounting the II.
GW10078855.898.01.15.02
PAGE 121 OF 141
Connect the 24V power supply cable to the HT power supply of the II.
GW10078855.898.01.15.02
Page 122 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
False cap
Screw
GW10078855.898.01.15.02
PAGE 123 OF 141
· Support the hanging single tank cable by anchoring it through rubber nozzle between
cable holder fitted on the trag arm & cable holder plate-upper.
· The clamping should be done such that sufficient length of cable is available for free
movement of the C-arm.
· The approx. length of the loop to be 1.35m (from metal gland to rubber nozzle).
Note:
· Move the c-arm and check if the loop left for the single tank cable is correct. If
not adjust the rubber holder nozzle by loosening the cable holder plate-upper.
1.35m
GW10078855.898.01.15.02
Page 124 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
· Remove the nut on the metal gland from the top panel side of the single tank cable and
pass the cable through the opening provided on the top panel.
Top panel
GW10078855.898.01.15.02
PAGE 125 OF 141
· Connect the safety switch wire and the earth wire to the top panel.
· Fix the top panel to the housing.
Safety switch
Safety switch
cable
Earthing wire
GW10078855.898.01.15.02
Page 126 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
· Connect the single tank cable wires to the terminal block, earthing terminal, DL
contactor, IGBT card D64 and D602A PCB. Refer wiring diagram for connections.
D602A
PCB
Single tank
cable
Inverter card
D64
Terminal block
Contactor
Earthing
terminal
GW10078855.898.01.15.02
PAGE 127 OF 141
· Route the single tank cable carefully and properly through the housing.
· Mount the counter weight lead plates on the C-arm. Take care that the II interface cable
is not damaged.
Counter Weight
Angle Scale
Screw
Steering handle
GW10078855.898.01.15.02
Page 128 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
Ø Move the rubber buffer washer backward in the direction shown to expose the
screws for the mounting of the cap as shown in fig 1.
Cap
Fig. 1
Ø Remove the M4 screws (3x) of the cap and keep the Cap bush and the cap aside.
Refer fig 2.
Fig. 2
GW10078855.898.01.15.02
PAGE 129 OF 141
Ø Remove the special grub screw of the adjustable handle and then remove the entire
adjustable handle out of the assembly. Refer fig 3.
Adjustable Handle
Fig. 3
Ø Remove the entire brake assembly out by removing the outer circlip,
stopper washer and the brake assembly lock washer. Refer fig 4.
Fig. 4
GW10078855.898.01.15.02
Page 130 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
Ø Unscrew the hardware of the cable holder and remove the single tank cable
out of this holder. Refer fig 5
Fig. 5
Ø Remove inner circlip and slide the shaft out of the square tube by holding
the entire C arm assembly carefully. Keep the C arm assembly aside. Refer
fig 6
Fig. 6
Note: Keep the C arm assembly safely on the ground by providing sufficient cushioning
as shown to ensure no damage to the components.
GW10078855.898.01.15.02
PAGE 131 OF 141
Ø Loosen all screws and replace the shaft assembly as shown in fig 7 and
tighten the bolts back with a torque of 10.2 Nm.
Ø Follow the steps back again for reassembly. Check the angulation
movement after the assembly is done to ensure that the parts are properly
assembled.
Fig. 7
Rubber stopper
False cap
GW10078855.898.01.15.02
Page 132 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
2. Remove the II cap by unscrewing the ornamental screw (2x) and the CSK
screws(2x).
Ornamental
screw
CSK screw
GW10078855.898.01.15.02
PAGE 133 OF 141
5. Dismantle the II by removing the hardware (4x) and Keep the II carefully in a
safe place.
Note: One person should be holding the II carefully to ensure it does not fall
off (weight of II is approximately 25 Kg) during dismantling and remounting
Rubber
stopper
Hardware (4x)
Rubber
stopper
Hardware (4x)
GW10078855.898.01.15.02
Page 134 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
7. Remove Grid (10077920) and grid plate (10076450) from Sirecon II and
assemble it to the II RING FOR 9" II FABRICATED - MM5E (10079298)
(use the same CASSETTE HOLDER MOUNTING PIN - MM5E (10079118)
hardware to mount it on Toshiba I.I.)
GW10078855.898.01.15.02
PAGE 135 OF 141
10. Adjust the camera to the center of the II and check the image quality by
placing the resolution pattern.
11. Mount the II top cover by passing the GND wire and camera cable through
the groove provided to the II top cover.
Rubber stopper
False cap
GW10078855.898.01.15.02
Page 136 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
2. Remove the II cap by unscrewing the screw (2x) under false caps
False caps
GW10078855.898.01.15.02
PAGE 137 OF 141
5. Dismantle the II by removing the hardware (4x) and Keep the II carefully in a
safe place.
Note: One person should be holding the II carefully to ensure it does not fall
off (weight of II is approximately 25 Kg) during dismantling and remounting
Rubber
stopper
Hardware (4x)
Rubber
stopper
Hardware (4x)
GW10078855.898.01.15.02
Page 138 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
7. Remove II RING and grid plate from failure Toshiba II and assemble it to the
replaced new Toshiba II by using the same CASSETTE HOLDER
MOUNTING PIN - MM5E (10079118) hardware to mount it on Toshiba I.I.
GW10078855.898.01.15.02
PAGE 139 OF 141
9. Adjust the camera to the center of the II and check the image quality by
placing the resolution pattern.
10. Mount the II top cover by passing the GND wire and camera cable through
the groove provided to the II top cover.
GW10078855.898.01.15.02
Page 140 of 141
Copyright Ó Siemens Healthcare Pvt. Ltd. All rights reserved. Restricted
Service Instructions
Multimobil 5E
10079118 10079252
Fig A/3.7
- Slide the 10079252 CASSETTE HOLDER COMPLETE (1X) into the step given
in mounting pin & lock the 10079252 CASSETTE HOLDER COMPLETE by
pushing it through the curve profiles given on cassette holder clamps.See Fig
A/3.7.
GW10078855.898.01.15.02
PAGE 141 OF 141
Update Instructions
The reproduction
transmission or use of
this document or its
contents is not
permitted without
express written
authority. Offenders will
be liable for damages.
Unrestricted GW10078855.896.01.07.02
Chapter Page Revision
All All 07
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are directed to contact one of the local offices of
Siemens or one of its affiliates before attempting installation or service procedures.
Unrestricted GW10078855.896.01.07.02
INDEX
INDEX
INDEX 3
1 UPDATE 4
GW10078855.896.01.07.02 Page 3 of 6
1 Update
Refer below product improvement information (PII) for changes carried out in MM5E and refer
PII document for:
1. Reason for improvement,
2. Changes carried out,
3. New part introduced,
4. Instructions for upgrading the earlier products
5. Parts to be ordered.
Intensifier (10083300)
2. II camera module (10080556) fitment is not back
compatible to the new Sirecon 23-1 ISX Image
Intensifier. Hence Al mounting plate on which the II
camera module is mounted has been modified to fit on
the new II. The new number for II camera module with
the modified Al plate is 10083304
3. II RING FOR 9" II (POWDER COATED) - MM5E
(10079117) diameter does not match with Sirecon 23-1
ISX Image Intensifier, therefore II RING FOR 9" II
(POWDER COATED) -MM5E (10079117) is replaced
by I.I. GRID MOUNTING RING (10083312) Sirecon 23-
1 ISX
4. Image Intensifier does not have an II cap so I.I. CAP
(LEAD PASTED) (10083315) is added.
5. Hardware (Screw and washer) is updated for camera
(10083304) and I.I. CAP (LEAD PASTED) (10083315)
6. Spacer for Sirecon II cap(10083324) is added to mount
II cap on Sirecon image intensifier
GW10078855.896.01.07.02 Page 5 of 6
(J6) & (J3) on the PCB and on the cable are from the
same manufacturer for better coupling /compatibility. It
was also ensured that the correct crimps were used for
the housing.
18 PII 5609-018_EUA_01S_00 In Multimobil 5E (M10078855) SMPS PS2
(M10083294) and PS3 (10080361) have been replaced
with SMPS LPS65 (M10077520)
19 PII 5609-019_EUA_01S_00 As a DTC measure one 18000µF capacitor (C3) is
eliminated from MM5E inverter module and the line
resistance is changed from 1.2 Ω to 1 Ω.
20 PII 5609-020_EUA_01S_00 1. Toshiba E5804-P3A Image Intensifier (10082045) is
implemented in place of Siemens Sirecon 23-1 ISX
Image Intensifier (10083300)
2. With this change II Camera - Sirecon 23-1 ISX
(10083304) will have to be replaced with II camera
module (10080556). (The difference in both the camera
assembly is with the thickness of the Al plate on which it
is mounted.)
3. With this change existing I.I. GRID MOUNTING RING
(10083312) will have to be replaced with II RING FOR
9" II FABRICATED - MM5E(10079298)
4. I.I. CAP (LEAD PASTED) (10083315) and Spacer for
Sirecon II cap (10083324) are not required with this
change as E5804-P3A Image Intensifier (10082045),
comes with II cap.
GW10078855.896.01.07.02 Page 6 of 6