Sensor Signal Conditioning (Part 3)
Sensor Signal Conditioning (Part 3)
f1 = 1/2pR1C1.
21 of 33 03/08/2021, 08:57
Sensor Signal Conditioning (part 3) http://www.industrial-electronics.com/DAQ/snsr-fnd_4c.html
FIG. 4.30 shows two ways to buffer and amplify the output of a charge output
transducer. Both require using an amplifier which has a very high input
impedance, such as the AD745. The AD745 provides both low voltage and low
current noise. This combina tion makes this device particularly suitable in
applications requiring very high charge sensitivity, such as capacitive
accelerometers and hydrophones.
The first circuit (left) in FIG. 4.30 uses the op amp in the inverting mode.
Amplification depends on the principle of conservation of charge at the inverting
input of the amplifier. The charge on capacitor CS is transferred to capacitor CF,
thus yielding an output voltage of ?Q/CF. The amplifier's input voltage noise will
appear at the output amplified by the AC noise gain of the circuit, 1 + CS/CF.
The second circuit (right) shown in FIG. 4.30 is simply a high impedance follower
with gain. Here the noise gain (1 + R2/R1) is the same as the gain from the
transducer to the output.
22 of 33 03/08/2021, 08:57
Sensor Signal Conditioning (part 3) http://www.industrial-electronics.com/DAQ/snsr-fnd_4c.html
FIG. 4.30: Balancing source impedances minimizes effects of bias currents and
reduces input noise.
Without the AC coupling capacitor C1, the amplifier will operate over a range of
0°C to +85°C. If the optional AC coupling capacitor C1 is used, the circuit will
operate over the entire -55°C to +125°C temperature range, but DC information is
lost.
Hydrophones
23 of 33 03/08/2021, 08:57
Sensor Signal Conditioning (part 3) http://www.industrial-electronics.com/DAQ/snsr-fnd_4c.html
Since the hydrophone signals of interest are inherently AC with wide dynamic
range, noise is the overriding concern among sonar system designers. The noise
floor of the hydrophone and the hydrophone preamplifier together limit the
sensitivity of the system and therefore the overall usefulness of the hydrophone.
Typical hydrophone bandwidths are in the 1 kHz to 10 kHz range. The AD743 and
AD745 op amps, with their low noise figures of 2.9 nV_/Hz and high input
impedance of 10^10 ohm (or 10 G ohm) are ideal for use as hydrophone
amplifiers.
The AD743 and AD745 are companion amplifiers with different levels of internal
compensation. The AD743 is internally compensated for unity gain stability. The
AD745, stable for noise gains of five or greater, has a much higher bandwidth and
slew rate. This makes the AD745 especially useful as a high-gain preamplifier
where it provides both high gain and wide bandwidth. The AD743 and AD745
also operate with extremely low levels of distortion: less than 0.0003% and
0.0002% (at 1 kHz), respectively.
The AD743 and AD745 op amps are the first monolithic JFET devices to offer the
24 of 33 03/08/2021, 08:57
Sensor Signal Conditioning (part 3) http://www.industrial-electronics.com/DAQ/snsr-fnd_4c.html
low input voltage noise comparable to a bipolar op amp without the high input
bias currents typically associated with bipolar op amps. FIG. 4.32 shows input
voltage noise versus input source resistance of the bias-current compensated
OP27 and the JFET-input AD745 op amps. Note that the noise levels of the
AD743 and the AD745 are identical. From this figure, it is clear that at high source
impedances, the low cur rent noise of the AD745 also provides lower overall
noise than a high performance bipolar op amp. It is also important to note that,
with the AD745, this noise reduction extends all the way down to low source
impedances. At high source impedances, the lower DC current errors of the
AD745 also reduce errors due to offset and drift as shown in FIG. 4.32.
FIG. 4.32: Effects of source resistance on noise and offset voltage for OP27
(bipolar) and AD745 (BiFET) op amps.
FIG. 4.33: A pH probe buffer amplifier with a gain of 20 using the AD795 precision
BiFET op amp.
25 of 33 03/08/2021, 08:57
Sensor Signal Conditioning (part 3) http://www.industrial-electronics.com/DAQ/snsr-fnd_4c.html
A typical pH probe requires a buffer amplifier to isolate its 10^6 to 109 ohm
source resis-tance from external circuitry. Such an amplifier is shown in FIG. 4.33.
The low input current of the AD795 allows the voltage error produced by the bias
current and electrode resistance to be minimal. The use of guard ing, shielding,
high insulation resistance standoffs, and other such standard picoamp meth ods
used to minimize leakage are all needed to maintain the accuracy of this circuit.
The slope of the pH probe transfer function, 50mV per pH unit at room
temperature, has an approximate +3500 ppm/°C temperature coefficient. The
buffer shown in FIG. 4.33 provides a gain of 20 and yields an output voltage equal
to 1 volt/pH unit. Temperature compensation is provided by resistor RT which is a
special tem perature compensation resistor, 1 k-ohm , 1%, +3500 ppm/°C,
#PT146 available from Precision Resistor Co., Inc. (Reference 18).
26 of 33 03/08/2021, 08:57