This document describes how to load user programs onto different Memoskop systems. It provides instructions for loading programs onto Memoskop C-E, C-E100, C, C-Sub, and Siremobil Iso-C 3D systems. The document also lists the standard user programs for organ programs on Siremobil Compact and Iso-C systems.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
62 views

User Programs For Memoskop C-E, C-E100, C and Sub

This document describes how to load user programs onto different Memoskop systems. It provides instructions for loading programs onto Memoskop C-E, C-E100, C, C-Sub, and Siremobil Iso-C 3D systems. The document also lists the standard user programs for organ programs on Siremobil Compact and Iso-C systems.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

User Programs for Memoskop C-E, C-E100, C and Sub

This document describes how to load the Memoskop User Programs to the Memoskop.

Note: Floppy disk #3 contains the versions of the Memoskop User Programs, which were current in June 2004.
Please look, if there is a newer version of the Memoskop User Program delivered with the system. A newer
version has a larger number in the MEM file name.

Table of Contents: Installation of the User Programs for Memoskop C-E


Installation of the User Programs for Memoskop C, C-E 100 and C-Sub
Installation of the User Programs for Siremobil Iso-C 3D
Organ programs Siremobil Compact and Iso-C

Installation of the User Programs for Memoskop C-E

• Switch on the unit.


• Put the disk #3 "Memoskop User Programs" into the service PC drive A.
• Start the Memoskop Service Program and enter the password.
• Select the service PC serial interface.
• Click on "Get from Unit" in the Main menu. The currently stored parameters in the Memoskop are transferred
to the service PC. Wait for the transfer.
• Select the "File" menu.
• Select the "Load without technical Data" menu point.
• Select the file (A:\Userprograms_V1.mem) in the Data window.
Note:
For US a separate file “A:\Userprograms_V1_US.mem” exists.
For Lithostar Modularis a separate file “A:\Userprograms_V1_Litho_MUP.mem” exists.
• Click on "Open".

Because of some problems during transfer of the predefined data in the Memoskop software VC00A, you have
to delete the User Program PFC2, PFC3, DR3 and DR4 before transfer to the Memoskop. After successful
transfer you have to program it again with the Memoskop keyboard by hand (User Setup).

• Select the "Mode" menu in the Memoskop Service Program.


• Select the operating modes in turn (PFC, DR).
• Select the program to be erased in the respective menu points – PFC2 / PFC3 / DR3 / DR4.
• Select "Delete This Program" in the respective windows.
• After selecting all of the programs to be deleted click on "Put to Memoskop" in the main menu.
The parameters are transferred to the Memoskop and saved. Wait until the status Configuration load
done" is shown.
• Exit the Memoskop Service Program by selecting the "File" and "Exit" menus.
• Switch the unit off and then on again. Wait for start-up.

• Program the deleted User Programs PFC2 / PFC3 / DR3 / DR4 with the Memoskop keyboard. For
detail see below list of Organ Programs for Siremobil Compact and Iso-C.

• Select the individual operating modes on the system and check the number of available user programs by
pressing several times. The user program is shown and can be read in the top left corner of monitor A.

Readme.doc/.pdf VA00A.6 Page 1 of 5


User Programs for Memoskop C-E, C-E100, C and Sub

Installation of the User Programs for Memoskop C, C-E 100 and C-Sub

• Switch on the unit.


• Put the disk #3 "Memoskop User Programs" into the service PC drive A.
• Start the Memoskop Service Program and enter the password.
• Select the service PC serial interface.
• Click on "Get from Unit" in the Main menu. The currently stored parameters in the Memoskop are transferred
to the service PC. Wait for the transfer.
• Select the "File" menu.
• Select the "Load without technical Data" menu point.
• Select the file (A:\Userprograms_V1.mem) in the Data window.
Note:
For US a separate file “A:\US_Userprograms_V1.mem” exists.
For Lithostar Modularis a separate file “A:\Userprograms_V1_Litho_MUP.mem” exists.
• Click on "Open".

Because of some problems during transfer of the predefined datas in the Memoskop software VC02A and
VC02B, you have to delete the User Program DR3 and DR4 before transfer to the Memoskop. After successful
transfer you have to program it again with the Memoskop keyboard by hand (User Setup).

• Select the "Mode" menu in the Memoskop Service Program.


• Select the operating mode “DR”).
• Select the program to be erased in the respective menu points – DR3 / DR4.
• Select "Delete This Program" in the respective windows.

To optimize the User Programs for Subtraction and Roadmap you have to change some parameters at the K-
Factor Menu in the Memoskop Service Program.

• Select the 'K-Factor' menu.


• Select 'Subtraction Noise red Phase A' to the value 16.
• Select 'Subtraction Noise red PhaseB' to the value 8.
• Select 'Duration of phase A' to the value 1,5 s
• Select 'Roadmap Noise red PhaseA' to the value 16.
• Select 'Roadmap Noise red PhaseB' to the value 8.
• Do not change 'Roadmap Noise red PhaseC' ( MD2)
• Select 'Roadmap Duration of Phase A' to the value 1,5 s

• After all these changes click on "Put to Memoskop" in the main menu.
The parameters are transferred to the Memoskop and saved. Wait until the status "Configuration load done"
is shown.

• Exit the Memoskop Service Program by selecting the "File" and "Exit" menus.
• Switch the unit off and then on again. Wait for start-up.

• Program the deleted User Programs DR3 / DR4 with the Memoskop keyboard. For detail see below list of
Organ Programs for Siremobil Compact and Iso-C.

• Select the individual operating modes on the system and check the number of available user programs by
pressing several times. The user program is shown and can be read in the top left corner of monitor A.

Readme.doc/.pdf VA00A.6 Page 2 of 5


User Programs for Memoskop C-E, C-E100, C and Sub

Installation of the User Programs for Siremobil Iso-C 3D

• Switch on the unit.


• Put the disk #3 "Memoskop User Programs" into the service PC drive A.
• Start the Memoskop Service Program and enter the password.
• Select the service PC serial interface.
• Click on "Get from Unit" in the Main menu. The currently stored parameters in the Memoskop are transferred
to the service PC. Wait for the transfer.
• Select the "File" menu.
• Select the "Load without technical Data" menu point.
• Select the file (A:\Userprograms_V1.mem) in the Data window.
Note:
For US a separate file “A:\US_Userprograms_V1.mem” exists.
For Lithostar Modularis a separate file “A:\Userprograms_V1_Litho_MUP.mem” exists.
• Click on "Open".

To optimize the User Programs for Subtraction and Roadmap you have to change some parameters at the K-
Factor Menu in the Memoskop Service Program.

• Select the 'K-Factor' menu.


• Select 'Subtraction Noise red Phase A' to the value 16.
• Select 'Subtraction Noise red PhaseB' to the value 8.
• Select 'Duration of phase A' to the value 1,5 s
• Select 'Roadmap Noise red PhaseA' to the value 16.
• Select 'Roadmap Noise red PhaseB' to the value 8.
• Do not change 'Roadmap Noise red PhaseC' (MD2)
• Select 'Roadmap Duration of Phase A' to the value 1,5 s

• After all these changes click on "Put to Memoskop" in the main menu.
The parameters are transferred to the Memoskop. Wait until the status "Configuration load done" is shown.
• Exit the Memoskop Service Program by selecting the "File" and "Exit" menus.
• Switch the unit off and then on again. Wait for start-up.

• Select the individual operating modes on the system and check the number of available user programs by
pressing several times. The user program is shown and can be read in the top left corner of monitor A.

Please mention, these User Programs are not used in 3D-Mode.

Readme.doc/.pdf VA00A.6 Page 3 of 5


User Programs for Memoskop C-E, C-E100, C and Sub

Organ programs Siremobil Compact and Iso-C


Continuos fluoroscopy 1 1 **) 1 ***) 2 3 4 5
1. Program name Standard Standard Litho/ Pelvis High Low/ Gastro/
ESWL Contrast/ Paediatric Thorax
Spine
2. Dose rate Mid *) Mid Mid Mid High Mid Mid
3. Sirematic normal S2 HC1 HC2 S2 HC2 S1 S2
4. Sirematic push HC2 HC2 HC2 HC2 IOD S2 IOD
5. Noise reduction - low 4 MD1 4 4 4 2 2
6. Noise reduction - high MD2 MD2 8 8 8 8 8
7. Auto LIH disk transfer No No No No No No No
8. Disk transfer rate 0 0 0 0 0 0 0

Pulsed fluoroscopy 1 1 **) 1 ***) 2 3 4 5


1. Program name Puls Standard Litho/ Puls Puls High Puls Puls
Standard ESWL Pelvis Contrast/ Low/ Gastro/
Spine Paediatric Thorax
2. Dose rate Mid *) Mid High Mid High Mid Mid
3. Sirematic normal S2 HC1 HC2 S2 HC2 S1 S2
4. Sirematic push HC2 HC2 HC2 HC2 IOD S2 IOD
5. Noise reduction - low 4 4 4 4 4 4 4
6. Dose reduction - low High High High High High High High
7. Noise reduction - high 8 8 8 8 8 8 8
8. Dose reduction – high Mid High High High Mid High Mid
9. Auto transfer each image No No No No No No No
8. Auto LIH disk transfer No No No No No No No

_______________________

*) for US: Dose rate "High"


**) Factory default
***) Lithostar Modularis

Readme.doc/.pdf VA00A.6 Page 4 of 5


User Programs for Memoskop C-E, C-E100, C and Sub

DR 1 1 **) 1 ***) 2 3 4 5
1. Program name DR Standard Litho/ DR DR High DR DR
Standard ESWL Pelvis Contrast/ Low/ Gastro/
Spine Paediatric Thorax
2. Dose rate Mid High High Mid High Mid High
3. Noise reduction - low 8 8 4 4 8 8 4
4. Noise reduction - high 16 16 16 8 16 16 8
5. Auto disk transfer Yes Yes No Yes Yes Yes Yes

Subtraction 1 1 **) 1 ***) 2 3 4 5


1. Programm name Sub Standard Litho/ Sub Sub Sub Sub
Standard ESWL Standard Extremity Nitrogen Back-
2B/s 4B/s ground
2. Subtraction Max Max Max Max Max Min Max
3. Dose rate High High High High Mid High High
4. Landmark No No 10% No No No 20%
5. Videosignal docu. output Pos Pos Pos Pos Pos Pos Pos
6. Image display Pos Pos Pos Pos Pos Pos Pos
Phase B1:
7. Duration of the phase B1 0 0 0 0 0 0 0
8. Disk transfer rate 0 0 0 0 0 0 0
Phase B2:
9. Disk transfer rate 2 3 2 4 2 2 2

Road map **) ***)


1. Programm name Road Map Standard Litho/
ESWL
2. Subtraction Max Max Max
3. Dose rate High High High
4. Landmark No No No
5. Videosignal docu. output Pos Pos Pos
6. Image display Pos Pos Pos

_______________________

*) for US: Dose rate "High"


**) Factory default
***) Lithostar Modularis

Readme.doc/.pdf VA00A.6 Page 5 of 5