Smart Lab
Smart Lab
X-ray Diffractometer
Instruction Manual
This manual describes the correct use of the product as well the usage precautions to be observed. To
obtain full-expected performance from the product, thoroughly read this manual.
Also, store this manual at an easily accessible place so that you can promptly refer to it whenever
it is necessary.
The contents of this manual are subject to change without prior notice.
This manual described in it may not be disclosed to a third party or copied, in whole or in part, without
the written consent of Rigaku Corporation.
In no event will Rigaku Corporation be responsible for the results of the use of this manual.
As a rule, one set of the instruction manual has to be purchased for each product.
If there are any missing or incorrectly collated pages in the delivered instruction manual, please
contact the sales representative who you purchased the instrument from for instruction manual
replacement.
Microsoft, Windows, Windows XP, Windows Vista, Windows 7, Microsoft Internet Explorer, Microsoft Excel, and
Microsoft Word Operating System are either registered trademarks or trademarks of Microsoft
Corporation in the United States and/or other countries. The ™ and ® marks have been omitted.
Adobe, the Adobe logo, and Reader are either registered trademarks or trademarks of Adobe Systems
Incorporated in the United States and/or other countries. The ™ and ® marks have been omitted.
The names of actual companies and products mentioned herein may be the registered trademarks or
trademarks of their respective owners.
Before Using the Product
Read this manual cover to cover before attempting to use SmartLab. Carefully read the Guidelines for safe
use of the X-ray diffractometer described in the beginning of this manual.
Copyrights
2. No part of this manual may be disclosed to third parties. If the contents of this manual must be
disclosed to a third party, you must obtain our written approval before doing so for each
specific case.
3. No part of this manual may be cited without our permission. No part of this manual may be
translated or disclosed to a third party without permission. If you must cite or translate any
part of this manual, you must obtain our written approval before doing so for each specific
case.
4. In general, unless special agreements are reached, each product unit is provided with a copy of
this manual.
Liability
1. Rigaku shall not be held liable for any accidents caused by or resulting from any of the
following.
* Use of the product for a purpose other than the purpose intended
* End of product life
* Unauthorized modifications
* Inadequate maintenance by the user
* Natural phenomena, armed conflicts, civil disturbances
* Use or action in breach of instructions given in this manual
* Installation conditions failing to meet the recommended ambient parameters
* Consumables
2. Rigaku Corporation cannot be responsible for the results of using this manual to operate the
product or the effects of the results of such operation.
Relocation
Please contact Rigaku Corporation before attempting to move the product from the originally installed
location.
Trademarks and registered trademarks
Microsoft and Windows are trademarks or registered trademarks of Microsoft Corporation in the United
States and/or other countries.
Pentium is a registered trademark of Intel Corporation.
Other company names and product names are trademarks or registered trademarks of their respective
companies.
Note that this manual omits the TM (™) and R (®) symbols.
Contents
Overview ............................................................................................................1
Sample Plates..................................................................................................87
Height reference sample plate................................................................................87
Transmission SAXS sample plate (option: 2680J111)............................................87
Capillary holder (option: 2430C101).......................................................................88
Wafer sample plates and sample spacers..............................................................88
Installing and removing sample spacers..........................................................91
Installing and removing wafer sample plates...................................................92
Sample Holders...............................................................................................93
Glass sample holder...............................................................................................93
Aluminum sample holder ........................................................................................93
Transmission SAXS sample holder (option: 2680J111)..........................................93
Accessories.....................................................................................................95
2-bounce
monochromator
+
SmartLab 2-bounce
Parallel beam Soller slit Soller slit
system analyzer
Film thickness analysis
Soller slit Rocking curve
4-bounce
monochromator Reciprocal space mapping
2-bounce
Soller slit
analyzer
Particle-/Pore-size
SAXS Soller slit Soller slit
distribution analysis
4000
β α
Characteristic
特性X線
x-rays
Intensity
強度
2000
Continuous
x-rays
連続X線
0
0 0.5 1 1.5 2
Wavelength
波長(Å)(Å)
The wavelengths of characteristic x-rays depend on the type of target used. Typical x-ray diffractometry uses Kα
x-rays generated by several types of metal targets, as shown in the following table. Kα x-rays contain Kα1 x-rays
and Kα2 x-rays whose wavelengths are quite close. Although this does not pose serious problems for ordinary
measurement of powder samples for phase identification (ID) analysis, optimal results can be achieved by using
only Kα1 x-rays in certain cases when making measurements for crystal structure analysis with powder samples or
when performing precise measurements of thin film samples. In recent years, it has become possible to use just Kα1
x-rays by employing an incident optical system comprised of a multilayer mirror and a Ge or Si monochromator
crystal.
Attachment
XY attachment for automated XY mapping or an RxRy attachment for sample
tilt alignment to be conducted before in-plane or reciprocal space map (RSM)
measurement. A newly developed connector allows easy changing of
attachments.
X-ray generator
Even with a horizontal sample mount goniometer with a moving x-ray source,
the product can incorporate a state of the art high-intensity 9 kW rotating anode
x-ray generator. When combined with a multilayer mirror, this x-ray generator
produces a high intensity x-ray beam (approx. 6 to 7 times the intensity provided
by a sealed-tube system) equal to an 18-kW rotating anode x-ray generator,
while reducing power consumption by 50% due to the high-brightness focal spot.
The system also offers lower operating costs compared to earlier models. A 3 kW
sealed tube x-ray generator can also be used with the SmartLab system. The
specifications of each generator are summarized in the following table.
• CBO unit
To select the incident beam.
• Standard incident optics unit
To select Soller slit, 2-bounce monochromator (with Soller slit), or 4-bounce
monochromator.
• Standard incident slit box
Equipped with variable slit and length-limiting slit.
2.
Main unit
4. 3. 4. 3.
5. 5.
2. 1. 2. 1.
11.
10.
12.
6.
8. 9.
7.
Goniometer
1. 3.
5.
2.
6.
7.
4.
Goniometer
Theta_s arm
• Selection slit
Theta_d arm
• Kβ filter
→Receiving slit box # 1
Sample attachment
• Sample plate
Height reference sample plate,
4-inch sample plate, etc.
→Sample Plates
• Attachment
Standard, XY-4”φ, etc.
→Attachments
• Sample stage
Standard sample stage, β sample stage, etc.
→Sample Stage
X-ray tube
4. 3.
6.
2.
1.
5.
3.
1.
2.
1. Cooling water pipe Connect the x-ray generator cooling water pipe.
2. X-ray tube X-ray generating source.
3. Shutter Controls x-ray emissions.
Theta_s arm
2.
3.
4.
1.
Theta_s arm
The control cable of the optical device
1. Incident connector box
connects here.
2. Cross beam optics (CBO) unit Optics unit for switching optics.
Soller units, monochromators, etc. are
3. Incident optics unit
installed here
4. Incident slit box Variable slit on the incident side.
Theta_d arm
3. 4.
2.
5. 6.
1.
Theta_d arm
The control cables for the optical devices
1. Receiving connector box
connect here.
2. Receiving slit box # 1 Variable slit on receiving side.
Receiving optical devices (analyzer) are
3. Receiving optics unit # 1
installed here.
4. Receiving optics unit # 2 Soller slits, etc. are installed here
5. Receiving slit box # 2 Variable slit on receiving side.
6. Attenuator Adjusts x-ray intensity.
1.
2.
1. 2. 3.
x
∫ dI D = 1 − exp⎛ − 2μx ⎞
Gx = ∞0 ⎜ ⎟
∫0 DdI ⎝ sin θ ⎠
Substitute the values for angle θ of the incident x-ray beam and x-ray absorption coefficient μ(/cm) in the following
formula to obtain x cm, the effective thickness.
Diffracted x-ray
utilization rate Gx 50 75 90 95 99 99.5
(%)
Kx 0.69 1.39 2.30 3.00 4.61 6.91
The following table shows the results of calculations of the sample thicknesses x cm that give a diffracted x-ray
usage rate of 90% when Kx = 2.30 and θ = 90°.
2
密度ρρ
Density absorption coefficient μ/ρ2(cm
Mass質量吸収係数μ/ρ(cm Thickness for 90% utilization rate (mm)
90%利用率の厚さ(mm)
/g) /g)
Substance
物質 3
(g/cm ) MoKα CuKα CoKα CrKα MoKα CuKα CoKα CrKα
α‐Al2O3 3.97/2 3.34 31.1 48.0 97.6 1.74 0.187 0.121 0.060
Si 2.42/2 6.44 60.6 93.3 189 1.48 0.157 0.102 0.050
α‐SiO2 2.65/2 3.71 34.5 53.1 108 2.34 0.252 0.164 0.080
Powder
CaCO3 2.71/2 8.02 70.9 107 205 1.06 0.120 0.079 0.041
粉末
α‐Fe2O3 5.26/2 27.3 216 42.7 86.5 0.160 0.020 0.103 0.051
ZnO 5.60/2 44.7 50.7 78.2 159 0.092 0.081 0.053 0.026
Ag2O 7.22/2 24.1 204 300 548 0.132 0.016 0.011 0.0058
PbO I(正方)
(tetragonal) 9.53/2 112 216 311 540 0.022 0.011 0.0078 0.0089
Al 2.70 5.16 48.6 74.8 152 0.826 0.088 0.057 0.028
α‐Fe 7.86 38.5 308 52.8 108 0.038 0.0048 0.028 0.014
Metal
金属 SUS(18-8) 7.93 37.9 278 110 106 0.038 0.0052 0.013 0.014
Cu 8.92 50.9 52.9 81.6 166 0.025 0.024 0.016 0.0078
Pb 11.34 120 232 334 579 0.0085 0.0044 0.0030 0.0018
Polymer ポリエチレン
ポリマー
Polyethylene 0.93 0.59 4.0 6.1 12.5 21.0 3.09 2.02 0.989
Vinyl
塩化ビニールchloride 1.41 6.73 62.0 94.2 186.2 1.21 0.132 0.086 0.044
Single crystal Si 2.42 6.44 60.6 93.3 189 0.738 0.079 0.051 0.025
単結晶
substance α‐SiO
2 2.65 3.71 34.5 53.1 108 1.17 0.126 0.082 0.040
Turning On and Off SmartLab
Described below are the procedures for turning on and off SmartLab.
Turning on SmartLab
1 Make sure that the HV ENABLE key on the main panel is unlocked
(the key is in the horizontal position).
If it is locked (the vertical position), turn the key clockwise to unlock it.
Main panel
The OPEARATE lamp blinks while SmartLab is starting up and when an error occurs. For
more information, see Appendix Using the OPERATE Lamp.
2 The login window will appear. Enter the registered user name and
password in the Login name and Password boxes, and click the OK
button. (When launching the software for the very first time, enter
administrator in the Login name box and click the OK button without
entering a password.)
Startup
Click Package Measurement under the Task menu and click Startup on the
displayed Package Measurement flow bar.
Startup
When not using the water circulation pump, make sure the cooling water is flowing
properly. If the cooling water is not flowing or the amount, pressure of water are
incorrect, an alarm will be issued and remain active for three minutes after the
startup procedure is initiated. Adjust the water volume to shut off the alarm within
three minutes, then repeat the startup procedure. For information on how to adjust
the water volume, refer to the instruction manual of the water circulation pump.
For the startup procedure, refer to Section 2.1 of the “CBO Maintenance Package
Measurement” Help Topic of the online help section of the SmartLab Guidance software.
Shutdown
Click Package Measurement under the Task menu and click Shutdown in the
displayed Package Measurement flow bar.
Shutdown
CAUTION When not using the water circulation pump, wait three minutes following
shutdown (i.e., three minutes after the yellow lamp located at the upper
section of SmartLab goes out), then halt the supply of cooling water. (Halting
the cooling water supply in less than three minutes may result in damage to
SmartLab.)
For more information on the shutdown procedure, refer to Section 2.6 of the “CBO
Maintenance Package Measurement” Help Topic of the online help section of the SmartLab
Guidance software.
While SmartLab is not used, it is recommended to turn off only the x-ray, and do not turn
off the vacuum unit. In this case, the controller remains powered; therefore SmartLab can be
started from SmartLab Guidance. If SmartLab is unused for a while (e.g. during the New
Year Holidays), press the “O” OFF button to turn the power of SmartLab off.
1 Click the Exit command under the File menu to exit SmartLab
Guidance.
main panel
This procedure completely shuts down the SmartLab internal power supply.
3 The DOOR LOCK button will flash, and you will hear beeping.
The shutter cannot be opened when the door is unlocked. The following message
also appears when the door is unlocked. SmartLab Guidance cannot be operated
until the door is locked again.
3 When the door is properly locked, the DOOR LOCK button stops
flashing, and the beeping stops. The instrument is then ready for
operation.
LAMP button
1 Attach the CBO unit to the shutter. Press the CBO unit against the
surface in the back and secure the unit in place with the Allen wrench
provided.
2 Insert the cable into the CBO connector of the incident connector box.
3 Insert a selection slit into the slit box of the CBO unit.
Corresponding
Name of optics Abbreviation Illustration
CBO
CBO(Cu)
For para-focusing CBO(Cu/Co) BB
CBO-E
For para-focusing(Co)
(included in the CBO (Cu/Co) CBO(Cu/Co) BB Co
unit)
CBO(Cu)
For micro area measurements
CBO(Cu/Co) MA
(option: 2680G111)
CBO-E
The type of selection slit installed can be identified from SmartLab Guidance.
Parallel beam
φ0.3 mm PB03
(2431C302)
Parallel beam
φ0.1 mm PB01
(2431C301)
*1 : Switchable to the point source x-rays only using the sealed tube type
(2080B211/2080B212/2080B411).
The type of selection slit installed can be identified from SmartLab Guidance.
Above selection slits is used in combination with a collimator as a double pinhole
collimator.
1 Install the incident parallel slit adaptor to the CBO unit by sliding it
down from the top section of the CBO unit. Secure it in place with the
Allen wrench provided.
3 Install the parallel slit on the adaptor and secure it in place with the
Allen wrench provided
Divergence
Type Abbreviation Illustration
limit angle
Open
None Soller_Slit_open
(option: 2680D115)
Soller slit
2.5º Soller_Slit_2.5 deg
(option: 2680D114)
In-plane PSC
(Parallel Slit Collimator) 1.0º In-plane_PSC_1.0 deg
(option: 2680D113)
In-plane PSC
0.5º In-plane_PSC_0.5 deg
(option: 2680D112)
In-plane PSC
0.15º In-plane_PSC_0.15 deg
(option: 2680D111)
The type of installed incident parallel slit can be identified from SmartLab Guidance.
2-bounce monochromator
Diffraction
Crystal Abbreviation Illustration
plane
Ge
220 Ge(220)x2
(option: 2680B111)
Ge
400 Ge(400)x2
(option: 2680B112)
Open
None Soller_slit_open
(provided with the unit)
Soller slit
5.0º Soller_slit_5.0 deg
(option: 2680D213)
Soller slit
2.5º Soller_slit_2.5 deg
(option: 2680D212)
In-plane PSC
0.5º In-plane_PSC_0.5 deg
(option: 2680D211)
The type of parallel slit installed on the adaptor can be identified from SmartLab Guidance.
4-bounce monochromator
Diffraction
Crystal Abbreviation Illustration
plane
Ge
220 Ge(220)x4
(option: 2680B211)
Ge
440 Ge(440)x4
(option: 2680B212)
Converges the parallel beam in the longitudinal direction onto the sample. The
optical device is used for the measurements of micro amount of a sample and of
micro area.
Fine alignment of the CBO-f unit is performed from SmartLab Guidance.
CBO-f unit
1 Install the CBO-f unit to the theta_s arm from the top, and secure it in
place with the Allen wrench provided.
This slit box adjusts the slit width to control the beam divergence angle (for the
para-focusing method) or the beam width (for the parallel beam method). For
small-angle scattering measurements, this slit guards against parasitic scattering.
A length-limiting slit can also be inserted to limit the beam length along the
longitudinal axis.
1 Attach the standard incident slit box to the Zs axis. Secure it in place
with the Allen wrench provided.
2 Connect the cable to the SLIT connector on the incident connector box.
15 mm 15
10 mm 10
5 mm 5
2 mm 2
0.5 mm
micro area optics 0.5
(option: 2680G111)
The type of inserted length-limiting slit can be identified from SmartLab Guidance.
Some types of pinhole collimators can be installed for micro area measurements.
Kβ filter for Cu can also be installed (optional).
BB collimator holder
Used in point focus optics.
(2431D101)
Name Illustration
Absorber case
(2431H101)
Kβ filter for Cu
(2431H201)
Kβ filter for various wavelength can be supplied. Please contact the sales representative who
you purchased the instrument from.
1 Install the collimator holder to the Zs axis and secure it in place with
the Allen wrench provided.
2 Connect the cable to the SLIT connector on the incident connector box.
φ 0.2 mm
0.2
(2431E103)
φ 0.1 mm
0.1
(2431E102)
φ 0.05 mm
0.05
(2431E101)
CAUTION Do not use the pinhole collimator (option) with the XY-4” phi attachment.
When using the pinhole collimator with the XY-4” phi attachment, the end of
the pinhole collimator collides with the sample and/or the sample plate from
the translation of X and Y axes, and thus the pinhole collimator, the sample,
and the sample stage may become damaged.
The type of inserted pinhole collimator can be identified from SmartLab Guidance.
For para-focusing optics, this slit box controls the anti-scatter slit width to
reduce the level of background scatter from the sample. For parallel beam optics,
it functions as the first slit in a double-slit receiving system.
The slit width can be controlled from SmartLab Guidance. A Kβ filter can be
inserted to remove Kβ x-rays from characteristic x-rays.
1 Attach the standard receiving slit box # 1 to the rail of the theta_d arm.
Secure it in place with the Allen wrench provided.
The Kβ filter reduces the Kβ characteristic x-rays by approximately 90% while only
reducing the Ka characteristic x-rays by approximately 50%. Therefore almost all Kβ x-rays
can be removed. (Used for measurements based on the para-focusing method.)
4 The installation position can be read from the scale on the theta_d arm.
This adaptor is used to install a parallel slit analyzer (PSA). A parallel slit
analyzer is installed so that the plates inside are mounted horizontally thereby
providing angular resolution for the theta_d axis.
1 Attach the receiving optical device adaptor to the rail of the theta_d
arm. Secure it in place with the Allen wrench provided.
3 Attach a parallel slit analyzer to the adaptor. Secure it in place with the
Allen wrench provided.
Open 45 mm PSA_open
1.0º
45 mm PSA_1.0 deg
(option: 2680F113)
0.114º
90 mm PSA_0.114 deg
(option: 2680F111)
Open 90 mm PSA_open
The standard configuration of SmartLab cannot incorporate either the PSA 0.114
deg. or PSA open (90 mm) in conjunction with a receiving parallel slit. In order to
do this, the configuration must be modified by, moving the receiving parallel slit
adaptor, standard slit box # 2, attenuator, and detector (counter) to the right.
The type of parallel slit analyzer installed on the adaptor can be identified from SmartLab
Guidance.
The 2-bounce channel cut analyzer provides increased resolution on the 2-theta
axis by only allowing diffracted x-rays that satisfy the diffraction condition for
the analyzer crystal to enter the detector.
Fine alignment of the channel cut crystal can be performed from SmartLab
Guidance.
2-bounce analyzer
Ge
220 Ge(220)x2
(option: 2680C111)
Ge
400 Ge(400)x2
(option: 2680C112)
The type of crystal currently installed can be identified from SmartLab Guidance.
1 Attach the receiving parallel slit adaptor to the rail of the theta_d arm.
Secure it in place with the Allen wrench provided.
3 Attach a parallel slit to the adaptor. Secure it in place with the Allen
wrench provided.
Aperture
Type Abbreviation Illustration
angle
Soller slit
2.5º Soller_slit_2.5 deg
(option: 2680F114)
In-plane PSA
1.0º In-plane_PSA_1.0 deg
(option: 2680F113)
In-plane PSA
0.5º In-plane_PSA_0.5 deg
(use of PSA 0.5 deg.)
In-plane PSA
(use of PSA 0.114 deg.) 0.114º In-plane_PSA_0.114 deg
(option: 2680F111)
The type of parallel slit currently installed on the adaptor can be identified from SmartLab
Guidance.
The standard receiving slit box # 2 contains the variable receiving slit and
controls measurement resolution in the para-focusing and double-slit receiving
configurations.
If a double-slit receiving system is used, it will function as the second slit.
Slit width and vertical position can be controlled from SmartLab Guidance.
1 Attach the standard receiving slit box # 2 to the rail of the theta_d arm.
Secure it in place with the Allen wrench provided.
When using the para-focusing method, set the installation position to 300 mm.
Attenuator
The attenuator adjusts the intensity of the detected x-rays.
Standard attenuator
The system uses several types of attenuators on a rotating disc. Attenuators for
optics alignment are also built-in.
The type of attenuator can be switched from SmartLab Guidance. Attenuators
can be switched automatically for various measurements based on detected
intensity.
Standard attenuator
1 Attach the standard attenuator to the rail of the theta_d arm. Secure it
in place with the Allen wrench provided.
2 Connect the cable to the ATT connector on the receiving connector box.
The attenuator types used during actual measurements are as follows: Open, 1/70, 1/1000,
1/10000. The attenuator types used during alignments are as follows: 9 kWBB, 9 kWPB, 3
kWBB, 3 kWPB for alignment.
CALSA
2 Attach CALSA on therail of the theta_d arm and set the installation
position to 200 mm. The position can be read from the scale on the
theta_d arm.
4 Connect the ADPT1, ADPT2, and SLIT2 cable to the ADPT1, ADPT2,
and SLIT2 connector on the receiving connector box.
Install the parallel slit in a manner such that the stacking direction of the foils
should be the same as the direction of the theta_d axis scan (the direction of
axial divergence).
The type of the parallel slit installed on the adaptor can be identified from SmartLab
Guidance.
The slit, attenuator, and knife edge are prepared to be used with CALSA.
Insert the beam conditioner into CALSA to condition the X-ray beams.
The type of the beam conditioner inserted into CALSA can be identified from SmartLab
Guidance.
D/teX Ultra
Detector adaptor
Used to attach the detector.
Detector adaptor
1 Attach the detector adaptor to the rail of the theta_d arm. Secure it in
place with the Allen wrench provided.
Scintillation counter
The scintillation counter is attached to the detector adaptor. It detects x-rays
received by the receiving optics.
Scintillation counter
1 When detecting x-rays from the receiving optics directly (in absence of
the diffracted beam monochromator), install the scintillation counter
to the base with the Allen wrench provided.
2 Attach the base to the detector adaptor. Secure it in place with the
Allen wrench provided.
3 The installation position can be read from the scale on the theta_d arm.
Changeover dial
Diffracted beam monochromator unit using Cu for para-focusing optics only (Bent only) can
be supplied (2726H113).
2 Install the unit on the detector adaptor. Secure it in place with the
Allen wrench provided.
3 The unit installation position can be read from the scale on the theta_d
arm.
When using the monochromator in Bent mode with para-focusing optics, the unit
must be installed at a specific position for a specific target given in the table below.
When using the monochromator in Flat mode, the exact position is not important.
Type of target Installation position
Cu 351.5 mm
Co 359.8 mm
Fe 364.7 mm
Cr 376.1 mm
Use the RSm only when using the monochromator in Bent mode for the
para-focusing method. The RSm is not used when using the monochromator in Flat
mode.
D/teX Ultra
1 When detecting x-rays from the receiving optics directly (in absence of
the diffracted beam monochromator), install D/teX Ultra to the base
with the Allen wrench provided.
3 Install the base to the detector adaptor, and secure it in place with the
Allen wrench provided.
When the measurement is made in para-focusing optics, install the detector adaptor
at position 351 mm
Detector position
3 Install the unit on the detector adaptor. Secure it in place with the
Allen wrench provided.
When the measurement is made in para-focusing optics, install the detector adaptor
at position 351 mm.
Detector position
Sample stages can be installed to the sample height alignment unit. To install a
sample stage, join the stage with the rail of the sample height alignment unit as
shown in the following figure, and slide it from top down. Make sure that the stage
is placed horizontally and secure it in place with the Allen wrench provided.
The direct beam stopper (option: provided with 5741A300) and knife edge (option:
2431G101) can also be installed to the sample height alignment unit. They will
reduce the background intensity in high-speed measurements using D/teX Ultra.
β sample stage
Knife edge
Name Illustration
Al sample holder with base 0.5
for ASC
(2455E445)
When measurements are performed in the reflection geometry using a knife edge,
install the reflection knife edge (REF. KE) to the extended base for ASC as shown
below.
Table Accessories
Name Illustration
Mylar foil -
HTK16 CHC
1 Click Manual Control under the Control menu. Move the chi axis to 0°
and the phi axis to 0°.
3 Open the clasps for the right and left sides. Remove the attachment.
Pin
Clasps
Hook
CAUTION Do not use the pinhole collimator (option) with the XY-4” phi attachment.
When using the pinhole collimator with the XY-4” phi attachment, the end of
the pinhole collimator collides with the sample and/or the sample plate from
the translation of X and Y axes, and thus the pinhole collimator, the sample,
and the sample stage may become damaged.
・ Normally the height reference sample plate and transmission SAXS sample plate
should be used on top of the standard attachment. If using the height reference sample
plate or transmission SAXS sample plate with an attachment other than the standard
attachment, you must set each attachment axis to 0.
・ To install/remove each sample plate to/from an attachment, see Installing and removing
wafer sample plates.
CAUTION Placing a sample larger than the sample plate may damage the sample during
measurement.
Magnets supplied with the product are generally used to secure the sample in
place. For information on how to mount samples, see Wafer-shaped samples.
The sample holding magnets can be used for samples measuring up to approximately
3 mm in thickness.
The Z axis is used to align the sample surface with the center of the x-ray beam.
Since the thickness range for which the Z axis can be adjusted is 3 mm, you
must select the appropriate sample spacer based on the thickness of the sample
to be measured.
0~3 mm
0-3 mm
(option: 2680A406)
3~6 mm
3-6 mm
(option: 2680A405))
6~9 mm
6-9 mm
(option: 2680A404)
9~12 mm 9-12 mm
(option: 2680A403) 0-3 mm(blue stamped mark)
12~15 mm
12-15 mm
(option: 2640A402)
0-3 mm(blue stamped mark)
0~3 mm(*1)
15 mm~18 mm
15-18 mm
(option: 2680A401)
3-6 mm(blue stamped mark)
3~6 mm(*1)
*1: Sample thickness for using XY-4” φ attachment. Refer to the abbreviations in blue.
Name Illustration
Table List of possible combinations of attachment, sample spacer, and sample plate
Measurable
Attachment Sample spacer Sample plate
sample thickness
0-3 mm
3-6 mm 4-inch wafer sample plate 0-3 mm
6-9 mm 3-6 mm
6-inch wafer sample
9-12 mm 6-9 mm
plate(*1)
12-15 mm 9-12 mm
15-18 mm 8-inch wafer sample 12-15 mm
Standard attachment plate(*1)
RxRy attachment (black stamped 15-18 mm
XY-20 mm attachment mark)
4-inch wafer sample plate (for 18-21 mm) 18-21 mm
4-inch wafer sample plate
6-inch wafer sample plate(*1) 21-24 mm
8-inch wafer sample plate(*1)
0-3 mm
3-6 mm 4-inch area mapping sample 0-3 mm
(blue stamped plate 3-6 mm
XY-4"φ attachment mark)(*2)
4-inch XY mapping sample plate (for 6-9 mm) 6-9 mm
4-inch XY mapping sample plate 9-12 mm
*1: Both X and Y must be 0 mm when the sample is placed on the XY-20 mm attachment.
*2: When the XY-4” φ attachment is used, the thickness indicated by the blue stamp mark on the sample spacer is valid.
To remove the sample spacer from the attachment, apply pressure to the bar
indicated by the arrow in the figure below, then turn the sample spacer
counterclockwise.
To mount a wafer sample plate to a sample spacer, align the sections indicated
by the arrows in the figure below, then turn the sample plate clockwise to secure
it in place.
The wafer sample plate can be removed by turning it counterclockwise.
See Installing and removing sample spacers for information on direct mounting (removing)
a wafer sample plate to (from) an attachment.
For information on preparing samples, refer to Section 2.8 of the “Quick theta/2-theta
measurement (BB)” or “Quick theta/2-theta measurement (PB/PSA)” Help Topic of the
online help section of the SmartLab Guidance software.
Selection
選択スリット(BB)slit (BB)
X-ray
X 線源 source
(フィラメント)
Sample
試料
Optical
Opticalpath
path
光路AA(divergent
(divergentbeam)
A(発散ビーム) beam)
Optical path
光路 B(平行ビーム)
B (parallel beam)
Switching between para-focusing optics and parallel beam optics by changing selection slits
CBO-E keeps the CBO feature which makes the optical paths A and B intersect on the sample. The only difference
is that the shape of the multilayer mirror used in the CBO-E unit is ellipsoidal while the mirror shape used for CBO
is parabolic.
The ellipsoidal multilayer mirror used in CBO-E has been designed so that one focus of the ellipsoid is placed at the
focus of the x-ray source and the other focus of the ellipsoid falls on the detector plane. When measuring capillary
samples or samples on the transmission SAXS sample plate in the transmission geometry, the measurement can be
performed very efficiently if a 1D detector like D/teX Ultra is used in combination. The CBO-E unit also switches
between measurements in the reflection geometry in para-focusing optics and those in the transmission geometry in
convergent optics. In both cases, D/teX Ultra enables very efficient measurements.
Detector position
X-ray source
Sample
Name Illustration
Center slit
Type of Resolution
Applications
crystal (FWHM)
For applications requiring monochromatized Kα1
Ge(220)x2
0.008º< x-rays or for measurements of materials with
crystal
diffraction planes with d spacings of ca. 2 Å.
0.0022º For measurements of materials with diffraction
Ge(400)x2 planes with d spacings of ca. 1.4 Å.
crystal (when sample d
spacing is 1.41A)
Ge(220)x4 For applications requiring high resolution over a
0.0033º<
crystals broad 2-theta range.
Ge(440)x4 For applications requiring very high resolution over
0.0015º<
crystals a broad 2-theta range.
Slit collimation
スリットコリメーション Zoom
Ge(220)22結晶
Ge(220) crystal view
拡大
[任意単位] unit]
Ge(220)44結晶
Ge(220) crystal
強度[unspecified
Cu Kβ
Powder samples
Place powder samples in the indentation on the glass sample holder or aluminum
sample holder, then insert into the height reference sample plate.
Powder sample
Bulk samples
Use a 4-inchi wafer sample plate, 6-inch wafer sample plate, or 8-inch wafer
sample plate to measure bulk samples. Select the appropriate sample spacer and
sample plate based on sample thickness. For detailed information, refer to Wafer
sample plates and sample spacers.
Place the sample so that the sample surface to be measured is parallel to the
sample plate. A piece of clay may be used to secure the sample in place. If the
bulk sample is small enough to pass through the window on the provided
aluminum sample holder, you can secure the sample in the aluminum sample
holder and use the height reference sample plate.
Wafer-shaped samples
Use a 4-inch wafer sample plate, 6-inch wafer sample plate, or 8-inch wafer
sample plate to measure wafer-shaped samples.
Wafer-shaped sample
When placing or removing the sample, we recommend removing the sample plate and
sample spacer from the attachment. When removing the sample plate and sample spacer, we
recommend setting the chi axis to 0° and positioning the sample horizontally.
Place the sample at the center of the sample plate. Use the magnets provided to
hold the sample in place. The magnets can be used with samples that are up to 3
mm thick. For information on measuring samples thicker than 3 mm, see Bulk
samples.
CAUTION
Placing a sample larger than the sample plate may result in damage to the
sample during measurement.
Place an 8-inch wafer on the center of the 8-inch wafer sample plate; because the
space available for the magnets securing the sample in place is small. Be careful
while measuring of 8-inch wafers, because increasing the tilt axis (chi axis) can
cause the wafer to bend under its own weight.
For information on the startup procedure, refer to Section 2.1 of the CBO Maintenance
Package Measurement Help Topic of the online help section of the Smartlab Guidance
software.
Distance table
Limit table
3 Set the position of each optical device. To change the optical devices of
the receiving optics, inspect the installation positions of the standard
receiving slit box # 1, standard receiving slit box # 2, and detector
adaptor and enter the correct corresponding values.
For information on the procedures of each Package Measurements, refer to the Package
Measurement Help Topic of the online help section of the SmartLab Guidance software.
Optics maintenance
The characteristics of the detector will change over time. Replacement of the
detector is required when it reaches the end of its service life. However, you can
minimize performance degradation by making the appropriate adjustments.
We recommend performing HV/PHA adjustments and dead-time correction
measurement at periodic intervals.
・ HV/PHA adjustments
・ Dead-time correction measurement
To make these adjustments automatically, select Package Measurement from
the Task menu, then select Maintenance Package measurement under Utility.
For detailed information on measurement procedures, refer to the “CBO
Maintenance Package measurement” Help Topic of the online help section of the
SmartLab Guidance software.
Cooling water
Check the amount of cooling water when using the water circulation pump. If
the water level is low, add water. Periodic replacement of cooling water is also
recommended.
Target (2080A211/A411)
The target metal rotates at high speed to boost x-ray output. Pay close attention
and note any abnormal noise or vibrations, and if these occur please contact the
sales representative who you purchased the instrument from.
When contaminant (tungsten from the filament) from the electron gun (filament)
adheres to the target metal surface, you may observe characteristic x-rays from
the contamination, in addition to characteristic x-rays from the target metal. If
this happens, maintenance is required. Please contact the sales representative
who you purchased the instrument from.
Filament (2080A211/A411)
As electrons are discharged, the filament will become thinner, eventually
breaking the required circuit. Upon breaking, the filament will release a sudden
burst of electrons that can damage the target surface. To avoid this, we
recommend periodically changing the filament.
* If a problem other than those described above occurs or if the problem cannot be corrected by taking the countermeasure described,
please contact the sales representative who you purchased the instrument from.
The OPERATE lamp indicates the status of the SmartLab controller. The lamp is usually green, but will
change to red and begin flashing at set intervals in the event of an abnormality. To reset the generated
alarm, you must restart SmartLab. After shutting down SmartLab Guidance, press the “O” OFF button,
then press the “|” ON button to restart SmartLab.
After SmartLab restarts, the OPERATE lamp will be flashing in green for 10 to 20 seconds. After that,
when the lamp lights up in green, SmartLab is ready to use. If the OPERATE lamp does not turn green,
flashing red instead after startup, or if the lamp turns green but changes to a flashing red immediately
during use, please contact the sales representative who you purchased the instrument from. Before
contacting, please refer to following tables to identify the alarm being generated.
Continuously
Condition Color Pattern
On / flashing
Immediately 250 ms
Green Flashing
after Power ON
250 ms
100 ms
10 seconds after
Green Flashing
Power ON
100 ms
Able to be
controlled from Continuously
Green
SmartLab On
Guidance
Table Patterns of the OPERATE lamp flashing and alarm sounds when an alarm generated
Continuously
Alarm point Color Pattern
On / flashing
400 ms 50 ms
Cause Description
An error has occurred in the x-ray generator (XG). See # 4 in
XG alarm
Troubleshooting.
The cover of the attachment connection cable is open in the radiation
enclosure. Close the cover.
Attachment
connecting cover is
open
A communication error has occurred in the control board for the optical
Communication
devices in the incident and receiving optics. The cables for the incident and
error
receiving optics may be disconnected or broken.
An error has occurred in the encoder used to control the goniometer. The
Encoder error
encoder may not be adjusted correctly, or the detecting part may be stained.
The protection circuit is working as a result of motor electrical overload. See
Servo alarm
# 3 in Troubleshooting.
K absorption
edge
Wavelength (Å)
For efficient absorption of Kβ x-rays only, select an element as a Kβ filter with its absorption edge wavelength
located between the Kα x-ray and Kβ x-ray wavelengths. Those elements will have an atomic number one or two
less than that of the target element. The following table shows the target elements commonly used for x-ray
diffractometry and their corresponding filters.
Table Kβ filters
Target Substa of
Mass
Kα1 Kβ1 absorption
-nces Thickness per unit Kα1
edge
(mm) area transmissivity
(Å)
(g/cm2)
Cr 2.290 2.085 V 2.269 0.011 0.007 63
Fe 1.936 1.757 Mn 1.896 0.011 0.008 62
Co 1.789 1.621 Fe 1.743 0.012 0.009 61
Cu 1.541 1.392 Ni 1.488 0.015 0.013 55
Mo 0.7093 0.6323 Zr 0.689 0.081 0.053 43
Ag 0.5594 0.4970 Rh 0.534 0.062 0.077 41
LINE OF BUSINESS
X-ray Equipment for Science and Industry
X-ray Diffractometer and Attachments
X-ray Generators
X-ray Diffraction Cameras
X-ray Spectrometers
Piezo Goniometers
X-ray Stress Analyzers
Industrial X-ray Radiographic Equipment
Vacuum Rotary Feedthrough