Operating Manual NOVOTEST T-UD3 1
Operating Manual NOVOTEST T-UD3 1
HARDNESS TESTER
NOVOTEST Т
Operating Manual
2014
Development and manufacturing of non-
destructive testing units and system
CONTENTS
1. Introduction 2
2. Appointment 2
3. Specifications 2
4. Packing list 5
5. Principle of operation 5
6. Preparations for operation 7
7. Basics of operation 8
8. Operation 9
9. Hardness measurement 19
10. Measurement modes 30
11. Calibration 35
12. Technical maintenance, special conditions of operation 46
13. Precautions and trouble shouting 48
14. Manufacturer’s guarantee and service maintenance 49
1. Introduction
In doing this you will be able to take full advantage of the function
range of the instrument. At the same time, you will also avoid errors
and wrong operation which in turn would cause incorrect test results
and thus could lead to injury and damage.
2. Appointment
3. Specifications
3.7. To save battery charge in the menu there are settings of device
auto switch off and brightness of the display.
4. Packing list
5. Principle of operation
The Vickers diamond is fixed to the tip of a round metal rod. This
metal rod is excited, to its resonant frequency of approx. 78 kHz,
into longitudinal oscillations. When the Vickers diamond contacts the
sample surface, the resonant frequency will change. This change
happens in relation to the size of the indent area from the Vickers
diamond. The size, in turn, is a measure for the hardness of the
tested material. Resonant frequencies can be measured very
accurately. This is why the UCI method is suited to make the
evaluation of Vickers indents, and thus of the complete test
procedure, so much easier and quicker.
There are also two additional advantages:
- the measurement is made under load. (No impairment of the
measurement due to elastic resilience);
- the hardness measurement is based on the area of the indent and
not on the length of the indent diagonals.
The measurement is thus less affected by surface roughness; even
gunmetal-finished surfaces can be measured.
Concerning the UCI method, the measurement value is also
dependent on the Young’s modulus of the material.
7. Basics of operation
7.1. Keys
Materials switch
Hardness scale
Modes switch switch
Navigation keys
8. Operation
Move through the menu using the navigation keys, for entering
press key .
8.3.1. Measuring
8.3.2. Calibration
8.3.3. Archive
8.3.4. Settings
Auto Off: setting of the automatic switch off device when it is not in
use.
Flash: enable / disable the cameras flash (for version with camera).
Tolerance, %: This parameter is used only for Smart mode. Set the
% value adjusts the range of deviations of measurements to be
included in the calculation of the average for the series in Smart
mode. Detailed description of the Smart mode, see paragraph 10.4.
8.3.6. Information
In this menu you can view information about the manufacturer and
offices around the world.
9. Hardness measurement
9.1. Using of the UCI method
The design of UCI Probe is shown in Figure 1a. Probe has a special
removable collapsible nozzle with the puck (Figure 1b). The puck
provides the convenience of positioning the probe relative to the
testing object and a clip for the measurements
One side of the puck is flat to using probe on flat surfaces. Another
side has grooves for using a probe on cylindrical surfaces.
It is marked slots designed for ease of measurement of hardness on
cylindrical products of various diameters. Probe with demounted
nozzle is usually using to measure hardness in difficult places, such
as narrow or deep groove.
The probe may be removed from the nozzle is used for measuring
the hardness in the narrow and hard to reach places.
Figure 2
Presence of user
calibration
Mode
Then select the scale and material hardness for which there is a
proper calibration (how to calibrate device see paragraph 11).
For select the hardness scale press and chose the scale that
For select the material press and chose the material that will
During the measurement you can select other hardness scale. The
displayed measurement reading will be converted according with the
new hardness scale if it is calibrated.
Install the probe puck on the sample surface, keeping it in the tough
skirt as shown in Figure 3a. By clicking on the skirt thrust both hands
to bring the diamond tip of the probe perpendiculary to the sample
surface to the touch (Figure 3b). Slowly (in about 0.5 seconds) by
pressing with a force of (5 or 1 kg depending on probe type) thrust
skirt, push the diamond tip into the metal surface, preventing
swinging (Figure 3c). After the sound signal, remove the probe from
the tested object.
a b c
Figure 3
NOTE! For correct operation of the UCI probe try not to exceed the
pressing force which is correspond to the probe (1 or 5 kg).
You can get the value of the current measurement or the average
for the series of measurements, depending on the settings in
Settings menu Current or Average:
Keyboard appears:
Move the cursor over the keyboard with navigation keys and select
Angle of the
measurements
Place the rebound probe vertically onto the test surface and press it
slightly against the surface. Charge the impact body by a charging
mechanism pressing the probe enclosure until the clicks, as shown
below:
After pressing the shutter release button and the impact body hitting
in the area of measurement sounds a beep and the display shows
the value of measured hardness.
Filled with the color readings indicate that the device has fixed the
general sequence and Smart mode is active.
Mode is active only for Leeb probe and shows the voltage produced
during the impact and rebound of impact body. M is the maximum
value of the conditional code number which corresponds to the
received signal.
11. Calibration
11.1. Main calibration
To calibrate the probe you will need 3 samples of material with a
known hardness. Hardness range should be wider than the
hardness of the materials which will be measured (The values
should be the maximum or more, minimum or less and average).
The device goes to the table, each cell of the table corresponds to
the certain scale of calibration for the certain material:
Steel
Alloy Steel
Stainless Steel
Cast Iron
Nodular Iron
Aluminums
Brass
Bronze
Copper
User material 1
User material 2
All calibrations of the device can be calibrated for any materials and
any scales, and are divided in this way just for practicality.
Press the navigation keys to choice cell for calibration, for example
HRC for Steel:
The device, making the measurement, gets the nominal codes, the
purpose of calibration - is to find the correlation between the nominal
code and hardness values (construction of the relation function).
For ending setting of first value press again , then the button
Number of
measurements
Current value
of the code
To check the correctness of the got values, move the cursor to the
X:000 and take a few measurements on one of the sample.
Hardness tester NOVOTEST T 39
Development and manufacturing of non-
destructive testing units and system
About the stored user calibration will symbolize the next state of the
cell:
12.5. Storage
12.5.1. Hardness tester shall be kept in the carry case, the probe
and the batterys shall be disconnected.
12.5.2. If hardness tester is kept in the carry case for than 14 days,
the battery shall be taken out from its compartment in the
information processing unit.
12.5.3. It is recommended to keep hardness testers in closed
premises with the relative humidity not more than 80%, there shall
be no mold, paints, acids, chemical agents and other chemicals, the
evaporation of which my give a harmful effect. Sharp fluctuations of
temperature and humidity which can result in dew formation are not
allowed.
12.6. Transportation
12.6.1. Hardness tester transportation in the carry case shall be only
in closed vehicles, where the possibility of mechanical damage or
atmospheric precipitation is excluded.
12.6.2. The way packed in carry cases hardness testers are located
inside the vehicle shall exclude.