Apnea Monitor Based On Bluetooth With An
Apnea Monitor Based On Bluetooth With An
Abstract— Apnea monitor is a device that is used to give a warning if there is stop breathing. Stop breathing while sleeping is one
form of obstructive sleep apnea. This cessation of breath cannot be underestimated, this is related to the main risk factors for health
implications and increased cardiovascular disease and sudden death. The purpose of this study is to design an apnea monitor with the
Android interface. This device allows the users to get how many times sleep apnea happens while sleeping and got data to analysis
before continuing with a more expensive and advanced sleep test. This device used a flex sensor to detect the respiration rate, the sensor
placed on the abdomen or belly so it can measure expand and deflate while breathing. The microcontroller uses an Arduino chip called
AT-Mega328. Bluetooth HC-05 used to send respiration data to Android, MIT app inventor used for the android programmer, and on
the android, there are plotting of respiration value and when the device detected apnea so the android also gives a warning to the user.
Based on the results of testing and measurement then compare with another device, the results of the average% error were 3.61%. This
apnea monitor design is portable but there are needs some improvement by using another sensor for detected respiration and using a
module other than Bluetooth.
This study uses a flex sensor which will be placed on the Start
patient's abdomen to detect the patient's breathing. Using
ATMega328 as a microcontroller, and Bluetooth to send data
to android.. Initialization
2) Experiment
No
In this study the researchers measured subject’s respiration
who were randomly selected and compared the results of
respiration with standard. Bluetooth
Connected?
B. The Diagram Block
In this research, flexible sensors were used to identify the Yes
swelling and deflection of the abdomen of the patient during
breathing as seen in Figure 1. The output sensor that is still in Received Data
the form of resistance will enter the analog signal conditioning from
circuit so that it can be processed by the microcontroller.
While the breathing detected, the microcontroller will send Plotting of Respiration Value
data for respiration value counters to android via Bluetooth.
When sleep apnea occurs, the module will activate the alarm
on the device and also on android. On the Android display No
there is also a patient's respiration value in the form of plotting
the graph of respiration value. Apnea?
C. The Flowchart
Bluetooth HC-05
The Arduino program is based on the flowchart in Fig. 2.
After initializing the microcontroller will detect changes in input
Fig. 1. The diagram block of the apnea monitor voltage produced by the flex sensor. Furthermore, this data
Alarm will be sent via Bluetooth to Android.
Start Handphone
The Android program flowchart can be seen in Figure
Plotting of respiration rate Fig.3, where after initializing, the operator needs to connect
Initialization Android to Bluetooth from the device. When breathing occurs,
the value of the patient's respiration will appear, and when its
Fig. 2. The flowchart of the arduino program
one minute there will be an update to plot the graph of
Detecting Respiration respiration value on the android display.
by Flex Sensor
D. The Analog Circuit
The most important part of the design is the analog signal
Sending Data via
Blueooth to Android
conditioning circuit, as shown in figure 4 which is a voltage
divider circuit, the difference amplifier circuit and filter in
Figure 5. This circuit serves to process analog signals so that
Stop the output is ready to be processed by a microcontroller.
1) Voltage Divider
The circuit shown in Figure 4 is a voltage divider circuit,
where this circuit will experience a change in output voltage
when the flex sensor changes the curve caused by the
breathing process of the patient.
III. RESULTS
In this study, a trial has been carried out on the tool
module by measuring breathing in patients directly and taking
measurements to determine the accuracy of the module
Fig. 4. Voltage divider
Serial.println
(nafasmanual); beat=0; }
}}
//===========pembacaan nilai
respirasi=========== if (flex>hold){
beat=1; detik=0; kirim=0;
digitalWrite (buzzer,LOW);
digitalWrite (ledPernapasan,
LOW); }
if
(beat==1
){
kirim=0;
digitalWrite (ledPernapasan, HIGH);
if
(waktumonostabil>delayMonostabil){
(a) (b)
Fig. 10. Sensor placement to measured the respiration (a) left (b) right
side
10 10.20 0.20
s1 10.08 0.08
0
10 10.16 0.16
10 10.20 0.20
Average
Number Device
(RR) STDEV Ua Error
Standard 11.2 0.45 0.20
1
Design 10.8 0.84 0.37 3.57%
Standard 19.6 0.89 0.40
Fig. 18. Respiration signal on the left of umblical region (Subject 4) 2 5.10%
Design 18.6 2.07 0.93
Standard 12.6 0.89 0.40
3 4.76%
Design 12.0 1.00 0.45
Standard 17.2 1.30 0.58
4 6.98%
Design 16.0 2.35 1.05
Standard 13.0 2.00 0.89
5 3.08%
Design 12.6 1.95 0.87
Fig. 19. Respiration signal on the right of umblical region (Subject 5)
6 Standard 18.8 2.17 0.97 2.13%
Design 18.4 2.88 1.29 [7] S. Sunit, R. Pallab, J. L. Mathew, M. Jayashree, and Dhanalakshmi,
Standard 15.0 2.00 0.89 “Apnea in New Born,” Indian J. Pediatr., vol. 75, no. 1, p. 57, 2008.
7
Design 14.8 2.17 0.97 1.33%
Standard 23.4 2.30 1.03
8
Design 22.0 2.55 1.14 5.98%
Standard 18.8 1.48 0.66
9
Design 18.2 1.92 0.86 3,.19%
Standard 19.0 1.41 0.63
10
Design 19.0 1.41 0.63 0.00%
The measurement results from the patient's Respiration
Rate compared to the standard. The results are shown in table
2.
IV. DISCUSSION
The apnea monitor design has been examined and tested
completely in this study. Based on the result after measuring
respiration signal from 10 subject, as seen above in the table
can be conclude that the average of amplitude from the right
side is
2.52 Volt and the average from the left side is 1.812, so the
right side is better to place the sensor.
Furthermore, based on the results after compared
measurement using standard there are differences in the
measurement results. The biggest error value is 6.98% and the
smallest is 0.00%. The average error is 3.61%. The uncertainty
value is obtained because there are still other factors that
influence the measurement results, such as the location of the
flex sensor at the time of measurement, and respiration from
the respondent is not always stable. So the value of uncertainty
must also be calculated. The biggest value is 1.29 and the
smallest value is 0.20.
V. CONCLUSION
This study has demonstrated the development of the apnea
monitor to monitoring the respiration of the patient’s while
sleeping. This study was built based on Bluetooth with
android interface. After testing the design, there are needs to
improvement by using other sensor for detected respiration
and using module other than Bluetooth.
REFERENCES
[1] Kementerian Kesehatan, “Keputusan Menteri Kesehatan Republik
Indonesia Nomor 118/MENKES/SK/IV/2014,” pp. 33–34, 2014.
[2] N. Maske and A. Gaikwad, “Monitoring Of Obstructive Sleep Apnea
Using Mobile,” Int. J. Ind. Electron. Electr. Eng. ISSN, vol. 4, no. 4, pp.
71–75, 2016.
[3] K. Sneha, “Detection of Sleep Apnea using Pressure Sensor,” Int. Res. J.
Eng. Technol., vol. 03, no. 10, pp. 709–712, 2016.
[4] Dipti Patil, V. M. Wadhai, S. Gujar, K. Surana, P. Devkate, and S.
Waghmare, “APNEA Detection on Smart Phone,” Int. J. Comput. Appl.,
vol. 59, no. 7, pp. 15–19, 2012.
[5] R. M. Taylor-Gjevre, “Obstructive sleep apnea, hypoxia and
inflammatory arthritis: How may they be linked?,” Int. J. Clin.
Rheumtol., vol. 10, no. 4, pp. 219–222, 2015.
[6] G. J. Rossini, “O bstructive S leep A pnea A nd O besity,” J. Lancaster
Gen. Hosp., vol. 4, no. 4, pp. 139–142, 2009.