EEL101 Lab Manual E4 RL RC RLC Circuit Response
EEL101 Lab Manual E4 RL RC RLC Circuit Response
Aim: To study the transient response of a series RL, RC and RLC circuit and
understand the time constant concept with square and sinusoidal AC power supply.
Apparatus Required:
S.No. Instrument Range Quantity
1. Digital Storage Oscilloscope (DSO)
2. Function Generator/AC supply
3. Resistor
4. Inductor
5. Capacitor
6. Bread Board
7. Connecting Wires
8. CRO Probes
Theory:
1. Transient Response: When a circuit's state is changed, typically by applying or
removing a voltage or current source, the circuit response changes over time
before reaching a steady state. This change is known as the transient
response.
2. Time Constant (τ): It is a measure of time required for certain changes in
voltages and currents in RC and RL circuits. Generally, when the elapsed time
exceeds five-time constants (5τ) after switching has occurred, the currents
and voltages have reached their final value, which is also called steady-state
response.
a. For an RC circuit, the time constant tau is given by tau = RC, where R is
the resistance and C is the capacitance.
b. For an RL circuit, the time constant tau is given by tau = L/R, where L is
the inductance.
c. For an RLC circuit, the time constant depends on the damping factor
and is more complex to calculate.
2
Circuit Diagrams
3
Precautions:
1. Ensure all circuit connections are secure before applying power.
2. Do not exceed the voltage and current ratings of the components.
3. Be cautious when handling the power supply to avoid electric shock.
Procedure:
For RC circuit:
Make the connections as shown in Figure 1
1. Make sure that the switch is in off position and no power supply is provided to
the circuit from the function generator or AC supply.
2. Circuit Assembly: Connect a resistor and capacitor in series on a breadboard
or the test bench for 230V AC application.
3. Power Supply Connection: Attach the function generator or AC supply across
the RC series combination.
4. Switch the main switch on so that the function generator or AC supply will get
connected to the RC circuit.
5. For breadboard circuit: Apply the square wave of 2Vpp voltage and frequency
corresponding to the tau, five times tau and fifteen times tau through the
function generator to the circuit, where tau is the time constant of the RC
circuit.
For AC setup: Apply 0-230V, 50 Hz supply to the circuit.
6. Measurement Setup: Connect the oscilloscope (DSO) across the function
generator/AC supply and capacitor to measure voltage and the response.
7. Observe the transient response (exponentially rising) on DSO till the steady
state is achieved.
8. Now immediately press RUN/STOP Switch of DSO to hold the response shown
on the DSO screen.
9. Capture the response in any USB 2.0 type storage device.
For RL circuit:
Make the connections as shown in Figure 2
1. Make sure that the switch is in off position and no power supply is provided to
the circuit from the function generator or AC supply.
2. Circuit Assembly: Connect a resistor and inductor in series on a breadboard
the test bench for 230V AC application.
3. Power Supply Connection: Attach the function generator or AC supply across
the RL series combination.
4. Switch the main switch on so that the DC function generator or AC supply will
get connected to the RC circuit.
5. For breadboard circuit: Apply the square wave of 2Vpp voltage and frequency
4
corresponding to the tau, five times tau and fifteen times tau through the
function generator to the circuit, where tau is the time constant of the RL
circuit.
For AC setup: Apply 0-230V, 50 Hz supply to the circuit.
6. Connect DSO across the inductor.
7. Observe the transient response (firstly sudden increase in voltage and then
exponentially decaying on DSO. Now immediately press RUN/STOP Switch off
DSO to hold the response shown on the DSO screen.
8. Capture the response in any USB 2.0 storage device.