Lung Lab Responses Weebly
Lung Lab Responses Weebly
2. How much of the PTV is covered entirely by the 100% isodose line?
The DVH for this AP/PA plan shows that only 7.8% of the PTV is receiving the Rx dose of
60Gy (Figure 1.2).
3. In your own words, summarize two advantages of using a parallel opposed plan?
(Review Khan, 5th ed., 11.5.A, Parallel Opposed Fields)
The use of parallel opposed fields in RT is the simplest technique for combining beams,
which results in favorable reproducibility of the treatment setup. This is advantageous
because reducing setup/treatment complexity helps minimize errors that may occur and can
reduce the time a patient spends on the table. Another benefit of parallel opposed beams is
the improved homogeneity of the dose distribution surrounding the target because there are
two beams contributing dose opposite one another. Field weighting can also be applied to
these beams to adjust the contribution from each, further improving target coverage.
PTV
7.82%
6000cGy
2. How much of the PTV is covered entirely by the 100% isodose line?
There is now 18.8% of the PTV being covered by the prescribed dose of 60Gy (Figure 2.2).
PTV
18.8%
6000cGy
Plan 3: 6MV – 5 Field (equal weighting)
Figure 3.1. Plan 3 – 6 MV, 5 Field
2. In your own words, summarize why beam energy is an important consideration for lung
treatments? (Review Khan, 5th ed., 12.5.B3, Lung Tissue)
Treating tumors in the lungs can create challenges in radiation therapy due to the Compton
effect being the predominant interaction in RT and the vast difference in tissue density
between the lung (air), tumor, and tissues surrounding the lungs. The electron density of a
material is directly related to its beam attenuation ability, so less density results in less beam
attenuation and a loss of lateral electronic equilibrium. This will occur as a treatment beam
travels through lung tissue. Then, when the beam eventually reaches a material of higher
density (i.e. tumor), a re-buildup region occurs at that interface, creating a second dmax and a
significantly lower dose delivered to the periphery of the tumor. As beam energy is increased
(and field size decreases), the loss of lateral electronic equilibrium will escalate, worsening
this phenomenon and decreasing target coverage. This is why lower energy beams are
typically preferred when treating tumors within the lungs.
Figure 3.2. Plan 3 – DVH
PTV
20.5%
6000cGy
PTV
30.4%
6000cGy
Plan 5: 6MV – 5 Field (unequal weighting with wedges)
Figure 5.1. Plan 5 – 6 MV 5-Field with wedges
1. Embed a screen capture of the beams-eye view (BEV) for each field that you used a
wedge.
See Figure 5.2 below.
Figure 5.2. Plan 5 – 6 MV, BEV of AP field (a), PA field (b), and LLat field (c) with wedge orientation
2. List the wedge(s) used and the orientation in relation to the patient and describe its
purpose. (ie. Did it push dose where it was lacking or move a hotspot?)
For the AP and the PA beams, I utilized 30° enhanced dynamic wedges (EDW) with the heel
toward the inferior of the patient and the toe at the superior. My reasoning for this was to
increase dose to the superior aspect of the PTV as I had noticed in my previous plans that
coverage was lacking the most in that area. The 95% isodose line did showed improved
coverage to this area, decreasing its distance by half when compared to Plan 4.
For the left lateral beam (LLat), I added a 45° EDW oriented with the heel toward the
anterior part of the patient’s body and the toe toward the posterior. I chose to use this wedge
as a compensator with the heel side acting as an additional attenuator in the portion of the
beam’s path encountering less tissue. This helped create a more homogenous dose
distribution along the lateral aspect of the PTV.
3. Describe how your PTV coverage changed (relating to the 100% isodose line) with your
final wedge choice(s).
The prescription dose coverage of the PTV decreased with the addition of wedges, from
30.4% PTV coverage in Plan 4 down to 21.1% in this plan (Figure 5.3). However, I chose to
continue with this arrangement because I was able to increase the dose that 95% of the PTV
was receiving, from 5437.2cGy (Plan 4) vs 5601.5cGy (Plan 5), the hotspot was decreased
from 105.2% (Plan 4) to 103.1% (Plan 5), and I liked the dose distribution that was provided
by the wedges as well (Figure 5.1).
PTV
21.1%
6000cGy
Plan 6: 6MV – 5 Field (plan normalized - 100% Rx to 95% PTV)
Figure 6.1. Plan 6 – 6 MV 5-Field with normalization
95%
PTV
6000cGy
Plan 7: 6MV – Final 3D Lung Treatment Plan
Figure 7.1. Plan 7 – 6 MV Final Plan with normal structure contours
3. Where is the region of maximum dose (“hot spot”), what is it, and is this outcome
clinically acceptable?
The final maximum dose was located centrally and slightly posteriorly within both the PTV
and ITV contours (Figure 7.2). The hotspot was 106.8% of the Rx dose, or 6410cGy, which
would be considered clinically acceptable because it is less than the traditionally allowed
maximum dose of 110%.
4. Embed a screen cap of your final plan’s isodose distribution in the axial, sagittal and
coronal views.
See Figure 7.1 above.
5. Include a final screen capture of your DVH and embed it within this assignment. Make
it big enough to see (use a full page if needed). Be sure to provide clear labels on the
DVH of each structure versus including a legend. *Tip: Import the screen capture into
the Paint program and add labels. See example in Canvas.
See Figure 7.3.
Figure 7.3. Plan 7 – Final DVH with all structures
ITV
PTV
Primary
Bronchus
Lt Lung
Trachea
6. Use the table below to list typical OAR, critical planning objectives, and the achieved
outcome. Please provide a reference for your planning objectives.
Figure 7.4. Combination of RTOG 0623, 06172, and Timmerman3 Dose Constraints
Planning Objective
Organ at Risk (OAR) Desired Planning Objective Met?
Outcome
Spinal Canal (Cord) D0.03cc < 4500cGy 842.14cGy Y
Esophagus D0.03cc < 6300cGy 1714.97cGy Y
Esophagus Mean Dose < 3400cGy 424.9cGy Y
Heart V60Gy < 33% 0% Y
Heart V45Gy < 67% 0.05% Y
Heart V40Gy < 100% 0.11% Y
Heart Mean Dose < 4000cGy 289.2cGy Y
Total Lung – ITV* V20Gy < 37% 22.7% Y
Total Lung – ITV* Mean Dose < 2000cGy 1171.2cGy Y
Trachea3 V44Gy < 5cc 0.00cc Y
Trachea3 D0.035 < 6000cGy 1585.3cGy Y
Primary Bronchus3 V44Gy < 5cc 8.1cc N
Primary Bronchus3 D0.035 < 6000cGy 6237.3cGy N
*RTOG 0623 defines constraints for Total Lung – GTV.
Extra Plan: 6MV – Lung DCA
My preceptor sometimes prefers to use dynamic conformal arcs (DCAs) for a treatment instead
of static beams, especially in palliative pain cases, for a reduced treatment time. For this reason, I
wanted to attempt a plan with the DCA technique as well, which can be seen below (Figure 8.1).
Figure 8.2. Comparison of Plan 7 (left) and Plan 8 (right) with the hotspots displayed
Figure 8.3. Plan 8 – DVH Lung DCA
ITV
PTV
Primary
Bronchus
Lt Lung
Trachea
It was interesting to see the differences between Plan 7 and Plan 8 (Figure 8.2). Both plans met
required constraints and would be considered acceptable for treatment. However, Plan 7 was less
hot and had better sparing for all contoured OAR, except for the heart. This was not surprising
seeing as the DCA plan delivers beam around the patient’s entire circumference without blocking
for normal structures when possible. The OAR that received the highest increase in dose from
Plan 8 included the spinal canal, trachea, and esophagus (Figure 8.3).
References:
1. Gibbons JP. Khan’s the Physics of Radiation Therapy. 6th ed. Wolters Kluwer; 2020:222-230.