0% found this document useful (0 votes)
134 views36 pages

ES-3830/3831 EC-3831S2/M/M2/F/F2/L EC-3830S2/M/M2/F/F2/L: Pentax Video Gi Scopes

Uploaded by

gdebesa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
134 views36 pages

ES-3830/3831 EC-3831S2/M/M2/F/F2/L EC-3830S2/M/M2/F/F2/L: Pentax Video Gi Scopes

Uploaded by

gdebesa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 36

OPERATION MANUAL

PENTAX VIDEO GI SCOPES


ES-3830/3831
EC-3830S2/M/M2/F/F2/L
EC-3831S2/M/M2/F/F2/L
EC-3832L
EC-3430M/F/L
EC-3830TM/TF/TL
EC-3832TL
EC-3830MZ/FZ/LZ
EC-3430MZ/FZ/LZ

This manual describes the recommended procedures for inspecting and preparing
the equipment prior to its use. For the cleaning and maintenance of the equipment after
its use, please refer to the separate maintenance/reprocessing manual.
Intended Use
The Video Lower G.I. Endoscopes are intended to provide optical visualization (via a video monitor) of, and therapeutic access to,
the Lower Gastrointestinal Tract. The Lower Gastrointestinal Tract includes, but is not restricted to, the organs, tissues, and
subsystems: Large Bowel.
These instruments are introduced per rectally when indications consistent with the requirement for the procedure are observed in
Adult and Pediatric patient populations.
These video endoscopes contained in this manual can only be used with Pentax Video Processors, Models EPM-3300 and EPM-
3500.

Notes
Read this manual before operating, and save this book for future reference. Failure to read and thoroughly understand the
information presented in this manual, as well as those developed for ancillary endoscopic equipment and accessories, may result
in serious injury including infection by cross contamination to the patient and/or user. Furthermore, failure to follow the instructions
in this manual or the companion Reprocessing/Maintenance Manual may result in damage to, and/or malfunction of, the
equipment.
It is the responsibility of each medical facility to ensure that only well educated and appropriately trained personnel, who are
competent and knowledgeable about the endoscopic equipment, antimicrobial agents/processes and hospital infection control
protocol be involved in the use of these medical devices. Known risks and/or potential injuries associated with flexible endoscopic
procedures include, but are not limited to, the following: perforation, infection, hemorrhage, burns and electric shock.
This manual does not describe how an actual procedure is to be performed, nor does it attempt to teach the beginner the proper
technique or any medical aspects regarding the use of the equipment.
The text contained in this manual is common for various types/models of Pentax endoscopes and users must carefully follow only
those sections and instructions pertaining to the specific instrument models appearing on the front cover. For the cleaning and
maintenance after its use, please refer to the separate “Reprocessing/Maintenance Manual”.

Sterility Statement
The instruments identified in this instructional booklet are reusable medical devices. Since they are packaged non-sterile, they
must be high-level disinfected or sterilized BEFORE initial use. Prior to each subsequent procedure, they must be subjected to an
appropriate cleaning and either high-level disinfection or sterilization process.
Refer to the companion Pentax Reprocessing/Maintenance Manual describing in detail the recommended instructions on the
care, cleaning, disinfection and sterilization of these endoscopes.

Conventions
Throughout this manual, the following conventions will be used to indicate a potentially hazardous situation which, if not avoided;

WARNING : could result in death or serious injury.

CAUTION : may result in minor or moderate injury or property-damage.

NOTE : may result in property-damage. Also, advises owner/operator about important information on the use of this equipment.

If you have any questions regarding any of the information in this manual or concerns pertaining to the safety and/or use of this
equipment, please contact your local Pentax representative.

Prescription Statement
Federal (U.S.A) law restricts this device to sale by or on the order of a physician or other appropriately licensed medical
professional.
TABLE OF CONTENTS
1. NOMENCLATURE AND FUNCTION ..................................................................................... 1
1-1. VIDEO ENDOSCOPE................................................................................................................... 1
1-2. ACCESSORIES............................................................................................................................. 3
1-3. VIDEO PROCESSOR ................................................................................................................... 4

2. PREPARATION AND INSPECTION FOR USE ..................................................................... 5


2-1. INSPECTION OF THE VIDEO PROCESSOR ............................................................................ 5
2-2. INSPECTION OF ENDOSCOPE.................................................................................................. 6
2-3. PREPARATION JUST BEFORE INSERTION OF ENDOSCOPE............................................. 18

3. DIRECTIONS FOR USE ........................................................................................................ 19


3-1. PRETREATMENT ........................................................................................................................ 19
3-2. INSERTION AND WITHDRAWAL............................................................................................ 19
3-3. BIOPSY ......................................................................................................................................... 21
3-4 CHOLANGIOPANCREATOGRAPHY (ERCP) ........................ DUODENOSCOPES ONLY 22
3-5 BILIARY DRAINAGE (ERBD).................................................. DUODENOSCOPES ONLY 22
3-6. LASER............................................................................. EXCLUDING DUODENOSCOPES 23
3-7. ELECTRO-SURGERY.................................................................................................................. 25

4. CARE AFTER USE ................................................................................................................ 26


LEAKAGE TESTER INSTRUCTIONS ................................................................................... 27
SPECIFICATIONS

WARNING:
Instrument repairs should only be performed by an authorized Pentax service facility. Pentax assumes
no liability for any patient/user injury, instrument damage or malfunction, or REPROCESSING FAILURE
due to repairs made by unauthorized personnel.

WARNING:
Never drop this equipment or subject it to severe impact as it could compromise the functionality and/or
safety of the unit. Should this equipment to be mishandled or dropped, do not use it. Return it to an
authorized Pentax service facility for inspection or repair.
1. NOMENCLATURE AND FUNCTION
1-1. VIDEO ENDOSCOPE
AIR/WATER FEEDING VALVE
(Part # OF-B132 for scopes with a single/common channel)
(Part # OF-B121 for all other scopes)
Covering of hole in the top of the valve delivers
pressurized air. Covering of the hole and fully
depressing the valve delivers pressurized water.

SUCTION CONTROL VALVE (Part # OF-B120)


Depress to remove fluids or air through
the instrument channel.
RUBBER INLET SEAL
“F” (FREEZE) BUTTON Allows passage of accessories while
Push to freeze an image. preventing escape of fluids and air.
INSTRUMENT
“C” (COPY) BUTTON CHANNEL INLET
Push to activate the hardcopy system that For introduction of
was selected between “FILE” and “HARD COPY”. biopsy forceps

U
CONTROL BODY and other accessories.

L
MODEL DESIGNATION F

F
WATER JET CHECK VALVE ADAPTER (OF-B119)
(Only for models with a water jet channel system)
R
D

WATER JET CONECTOR CAP (OF-B118) RIGHT/LEFT DEFLECTION


(Only for models with a water jet channel system) CONTROL KNOB

FORWARD WATER JET CONNECTOR RIGHT/LEFT DEFLECTION LOCK


(Only for models with a water jet channel system) Functions similar to Up/Down lock
Allows connection of special irrigation tube (Part # OF-B113)
for pressurized source of a spray directed at the UP/DOWN DEFLECTION CONTROL KNOB
endoscopically visualized surface.
CANNULA/FORCEPS ELEVATOR
UP/DOWN DEFLECTION LOCK CONTROL KNOB
When this lever is in the “F” position, turned clockwise, (DUODENOSCOPES AND OTHER SCOPES
the bending section moves freely. When turned counterclockwise, WITH OBLIQUE OPTICS ONLY)
the bending section becomes progressively more stabilized. To guide and direct cannula or forceps.
MAGNIFICATION CONTROL KNOB
“V” (VCR) BUTTON
(EG/EC-3430Z/EC-3830Z SERIES ONLY)
Push to activate the VCR for recording live procedures. Pushing the knob forward magnifies an image

UMBILICAL CABLE

TWO-CHANNEL SCOPES ONLY


RUBBER INLET SEAL
Allows passage of accessories while YELLOW DOT IDENTIFIES
preventing escape of fluids and air. A SMALL CHANNEL

INSTRUMENT CHANNEL SELECTOR (OF-B122)


Alignment of the indicator to the prescribed colored
U
L

positions allows the user the choice of suction capability


through either channel (2.8mm or 3.8mm) or
F

simultaneous suction through both channels.


R
D

GREEN DOT IDENTIFIES A LARGE CHANNEL

–1–
DISTAL END
(Refer to the inside rear cover of this manual)

BENDING SECTION

INSERTION TUBE

FEED BACK TERMINAL


Certain manufacturers’ electrosurgical units, such VENTING CONNECTOR
as older Olympus models, require the connection Accepts “RED” ETO GAS Sterilization Venting cap.
of a scope feedback cord (S-cord) to this Also accepts Leakage Tester.
feedback terminal.
AIR/WATER PORT ETO GAS STERILIZATION VENTING CAP RED OF-C5
To connect feeding tube from water Provides venting of endoscope interior to equalize
bottle assembly. internal and external pressures. This cap must be
removed before immersion.
NOTE: See important separate section
regarding the use of this cap!

PVE CONNECTOR
Can be rotated within PVE SOAKING CAP
a 180 range OE-C2
This cap must be securely
attached before immersion.

ELECTRICAL CONTACTS

SUCTION NIPPLE
For attachment to
external suction LIGHT GUIDE
source Transmits light from
light source to distal
end of endoscope.

WARNING: Note:
Immediately after use, the metal light guide prong Ensure that the soaking cap has been securely
and the electrical contacts/pins of the endoscope attached (by properly rotating it) to prevent the cap
may be HOT. To avoid burns, do not touch these from coming off during reprocessing. Failure to
areas immediately after use. For safer handling after securely attach the soaking cap can result in scope
a procedure, grasp the PVE connector housing. damage.

–2–
1-2. ACCESSORIES
1) Biopsy Forceps

ENHANCED FLEXIBLE GRIP


PORTION FLEXIBLE SHAFT

CUPS HANDLE
Pink handle defines
autoclavable forceps.

2) Cleaning Brush for Instrument Channel

WHITE BRISTLE FLEXIBLE SHAFT HANDLE

(CS6030SN for VSB-SERIES)


(CS6021SN for all other scopes)

3) Cleaning Brush for A/W Channel (thin) (40/40K SCOPES)

WHITE BRISTLE FLEXIBLE SHAFT HANDLE

(CS3030S for VSB-3440/3440K)


(CS3025S for all other 40/40K series scopes)

4) Cleaning Brush for A/W Suction Valve Cylinder

( CS-C5S )

5) Cannula (DUODENOSCOPES ONLY)

JAPAN (TG1918S)

CAUTION:
Because of the effect that accessories used through the instrument channel of the endoscope can have on
the performance of the endoscope itself, it is strongly recommended that PENTAX accessories be used with
PENTAX endocopes. If a unique or highly specialized accessory is available from another source, please
contact PENTAX to arrange a test of its compatibility before using it through the PENTAX endoscope.

NOTE:
Maximum outer diameter of an endoscopic accessory instrument must be at least 0.2 mm less than the speci-
fied instrument channel diameter in Pentax endoscopes. Working length of an endoscopic accessory instru-
ment may be approximately 30 cm longer than the endoscope working length.

NOTE:
For patient contact endoscopic accessories, follow the specific and detailed instructions on use, care and
maintenance supplied with each product.

–3–
1-3. VIDEO PROCESSOR

CAUTION:
Please refer to the instruction supplied with the processor.

NOTE
Do not use the new Pentax OS-H4 water bottle cap with the older OS-H2 water container/bottle. Although the
cap may appear to fit onto the bottle, air may escape resulting in insufficient pressure and flow of air and water
during the endoscopic procedure. Both the Pentax water bottle cap and bottle (container) are identified by their
appropriate model designation. Ensure that an OS-H4 cap is used only with the OS-H4 water container/bottle.
Do not overtighten the bottle cap. Overtighting can cause the bottle cap to break.

1) FOR 30/40 SERIES SCOPES

NOTE:
For 30/40 series Video Endoscopes with serial numbers beginning E0**** or higher the Infra Red feature can-
not be used.
Leave the Color/IR switch set at ‘Color’ for those scopes.

EPM-3300 EPM-3500
BRIGHTNESS CONTROLS LAMP IGNITION SWITCH
DOCUMENTATION CONTROLS (main/auxiliary)
LAMP IGNITION SWITCH - Hardcopy, VCR etc.
(main/auxiliary) - Freeze BRIGHTNESS
CONTROLS
COLOR BALANCE
WATER BOTTLE BUTTONS

3500
EPM-

INTERFACE
SOCKET

POWER SWITCH
DOCUMENTATION
I : ON COLOR/IR CONTROLS
O: OFF SWITCH X-LUM
POWER SWITCH - Hardcopy, VCR etc.
KEYBOARD CONNECTOR SWITCH
X-LUM SWITCH AIR PUMP WATER BOTTLE I : ON - Freeze
SWITCH O: OFF INTERFACE
SOCKET AIR PUMP
COLOR BALANCE BUTTONS SWITCH

2) FOR 30K/40K SERIES SCOPES


EPK-700 EPK-1000

BRIGHTNESS CONTROLS DOCUMENTATION BRIGHTNESS


CONTROLS CONTROLS
- Hardcopy, VCR etc. AIR PUMP SWITCH
- Freeze

EPK-700
PENTAX E CONTRO
L EPK-1000
BALANC
COLOR
ESS
BRIGHTN COPE
+5
LAMP
AUTO
RED
SERVICE MAN BLUE
0 FREEZE

AVE
CENT -5
XLUM

POEWR
P
AIR PUM ON

ON OFF
OFF

POWER SWITCH POWER SWITCH


I : ON LAMP SWITCH I : ON
WATER BOTTLE O: OFF O: OFF
WATER BOTTLE
AIR PUMP SWITCH INTERFACE
X-LUM SWITCH SOCKET
INTERFACE SOCKET

–4–
2. PREPARATION AND INSPECTION FOR USE
Prior to use, the endoscope, video processor and endoscopic accessory instruments must be carefully inspected for cleanliness
and proper function to determine that they are appropriate for patient use:

NOTE:
PENTAX video endoscopes contained in this manual are only compatible with PENTAX video processors,
model EPM-3300 and EPM-3500 (for 30/40 scopes) or EPK-700 and EPK-1000 (for 30K/40K scopes).

2-1. INSPECTION OF THE VIDEO PROCESSOR


Please refer to the Owner’s Manual of the specific model of Pentax video processor for complete instructions.
1) Attach water bottle assembly, 2/3 filled with sterile water to the appropriate location on the left side of the video processor.

WARNING:
The addition of defoaming agents to the water supply is NOT recommended. Due to their nature, these sil-
icone based agents cling tenaciously to surfaces. Unless they are rinsed very thoroughly, a “barrier” could
be created which could reduce the effectiveness of the disinfection/sterilization process. Additionally,
repeated use of such defoamers could eventually lead to residual silicone build up resulting in equipment
malfunction such as clogged air and/or water channels.

2) Set the drain lever on the water bottle assembly to the


upright position labeled A/W (air/water).
EPM-3300 (for 30/40 scopes) 3) Plug the processor into a properly grounded receptacle with
the power switch in the OFF position.
• 4) Make sure that the Pentax PVE connector is aligned with the
interface socket on the front panel of the processor.
• 5) Connect the endoscope to the interface socket on the proces-
sor as illustrated. Rotate the lever of the interface socket
clockwise after insertion (EPK-700/EPK-1000).
• 6) Connect the air/water feeding tube from the water bottle
air/water feeding tube assembly to the air/water port on the side of the PVE con-
nector.
EPK-700 (for 30K/40K scopes)
7) Turn the processor and air pump to the “ON” position and
check for proper functioning.
EPK-700
8) Prior to each procedure, check the endoscope image quality
PENTAX CONTROL

LAMP

SERVICE
BRIGHTNE

AUTO
MAN
SS
+5
COLOR
BALANCE

BLUE
RED
COPE
displayed on the monitor. Confirm that the image quality,
0 FREEZE

XLUM
AVE
CENT -5

P
POEWR
color, automatic brightness (iris) functions are acceptable as
AIR PUM ON

ON OFF
OFF

per the instructions provided with the Pentax video processor.

NOTE: 30/40 SCOPES


air/water feeding tube Leave the Color/IR switch on the unit Front Panel
of the processor Model EPM-3300 set at ‘Color’ for
30/40 series Video Endoscopes with serial num-
bers E0**** or higher.

–5–
2-2. INSPECTION OF ENDOSCOPE

CAUTION:
If the endoscope is intended to be clinically used after testing of individual scope functions (suction, air/water
delivery, water jet, etc.) without further reprocessing,the following precaution should be exercised.
Use “fresh” distilled or sterile water during individual scope function tests to avoid recontamination of the previous-
ly reprocessed instrument by waterborne microorganisms. Tap water, especially that which may be left idle and
uncovered for a prolonged period of time, should not be used during any inspection/testing of the endoscope.

Before proceeding with inspection of individual functions, PENTAX Endoscopes should be tested for the integrity of their
water-tight design (example: tear in the instrument channel). This test is described in another section of this manual entitled:
“Leakage Tester Instructions.”

ST
• 1) Inspection of the Insertion Tube
TE 19.6
a) Check the entire surface of the insertion tube for abnor-
D
AN

;;;;;;;;
29.4

;;;;;;;;
GER

9.8

mal conditions such as dents, wrinkles or bite marks.


;;;;;;;;
LEKAGE TESTER
SHA-P2

;;;;;;;;
39.2
0

;;;;;;;;
kPa

;;;;;;;;
;;;;;;;;
Any indentation of the flexible shaft of the endoscopes
;;;;;;;;
;;;;;;;; can cause damage to the internal mechanisms of the
endoscopes.
b) Similarly, check the condition of the umbilical cable for
outward signs of damage such as buckling, crush marks,
Leakage Tester etc.

CAUTION:
To avoid further damage to the endoscope or the possibility of malfunction during a procedure, do not use an
endoscope with any abnormalities or outward signs of damage.

c) Make sure that the entire endoscope is clean and has been subjected to either a high-level disinfection or sterilization pro-
cess before each patient use.

WARNING:
All instruments must be reprocessed prior to first time use, after any repairs/service and before every
patient use. When utilizing chemo-thermal processes for reprocessing the endoscope, the instrument
should be allowed to return to room temperature prior to use and/or further handling.

NOTE:
The distal end of the endoscope must be protected against damage from impact. Never apply excess force
such as twisting, or severe bending to the flexible portion of the endoscope.

–6–
2) Inspection of Deflection Controls and Locks
a) Slowly manipulate the Up/Down and the Right/Left con-
RIGHT-LEFT • trol knobs to see that they function smoothly. Be certain
that a full and appropriate range of deflection is possible.
b) Engage the deflection locks to be certain that the position
of the deflected tip can be stabilized.
UP-DOWN

NOTE:
The deflection locks are of the friction type and the degree of friction is adjustable. The degree of locking fric-
tion depends on the rotational position of the lever.

NOTE: (SMALL BOWEL ENTEROSCOPE ONLY)


Due to the relatively long length of this particular instrument and the tortuous paths it must negotiate, some
“knob play” has been intentionally incorporated into the angulation control system of the enteroscope. This
“looseness” in the control knobs is most noticeable when the insertion tube is in a straight position.
During actual clinical use when the scope is looped, the operator should feel normal tension in the control system.

CAUTION:
ANY lack of smooth operation of the deflection controls may be an early indication of internal damage to
and/or part(s) failure within the endoscope’s angulation system. To avoid the possibility of further endoscope
damage or the potential for malfunction of the angulation system, do NOT use the endoscope if the angula-
tion mechanism does not operate properly.
Prior to use ensure that the deflection controls can rotate smoothly, that there is no grinding or excess fric-
tion within the angulation system and that the distal bending section bends freely and smoothly. NEVER
APPLY EXCESSIVE FORCE TO THE DEFLECTION CONTROLS!
When an endoscope exhibits excessive “knob play” or if angulation is lost in any direction, do NOT use the
instrument. Excessive “knob play” can be defined as rotating of the angulation control knob(s) in any one
direction for more than 30 degrees without any corresponding distal tip deflection. The examples above are
indications that service is requlred to avoid more serlous problems with the angulation control system,
including angle or pulley cable/wire breakage and/or the possibility of a “frozen” distal bending section.
A “frozen” bending section can make instrument extraction from a patient more difficult.

–7–
3) Inspection of Distal rubber tip (40/40K SCOPES EXCEPT SINGLE/COMMON-CHANNEL SCOPES)
a) It is imperative that a rubber distal tip (cap) be attached to the cream colored plastic endoscope distal end during clinical
use. This rubber tip incorporates the built-in common air/water nozzle for precise delivery of air and water onto the distal
objective lens.
b) Prior to use, ensure that a rubber tip is properly attached to the plastic distal end of the scope. For easier identification and
confirmation, the plastic scope tip is always cream colored and the distal rubber tip is black in color.

WARNING:
Prior to use, always ensure that a distal rubber tip in good condition is properly attached to the distal end
of the endoscope. A worn or damaged tip should be replaced with a new one which has already been sub-
jected to a high-level disinfection or sterilization procedure. For repeated use, always ensure that the tip
has already been reprocessed. A damaged, worn or missing tip could result in poor visibility during the
procedure due to the inability to properly deliver air and water to clear debris off the distal objective lens,
and/or difficulty in passing accessories through the scope tip. If the tip is not secured properly, a potential
exists for the tip to fall off during the endoscopic procedure. Additionally, if the tip is not attached, the
uncovered scope tip may scratch and traumatize patient tissue.

CAUTION :
If there is any doubt, replace the distal rubber tip with a new, fully reprocessed one for each procedure.

c) To attach or replace a rubber distal tip, proceed as follows


Tab
(ALL SCOPES EXCEPT DUODENOSCOPES)
➀ Align the tab(s) in the rubber tips with the notch(s) in
the plastic scope tip.
• ➁ Gently press the distal rubber cap down onto the scope
Groove tip until the rubber completely covers the circular
groove around the scope.
Notch

WARNING:
Failure to attach the distal rubber tip properly could result in poor visibility during the procedure (due to the
inability to properly clear debris off the distal objective lens) and/or in difficulty passing accessories through
the scope tip. Additionally, if the distal rubber tip is not attached properly, a possibility exists for the rubber
tip to fall off during the endoscopic procedure. It is imperative that the distal rubber tip be securely attached
to the tip of the endoscope.

• ➂ Gently depress the rubber air/water nozzle down so


that it is flush with the plastic distal scope tip.

CAUTION :
Make sure that there is no space between the rub-
WRONG CORRECT ber tip and plastic scope end.
If there is any clearance between rubber cap and
scope tip, air and water may not flow properly.

–8–
d) To attach or replace the distal rubber tip on DUODENOSCOPE, proceed as follows;
➀ Align the rubber air/water nozzle opening with the objective lens on the scope distal tip and slide the rubber cap onto
the scope.
➁ Press the rubber cap down until it seats properly onto the circular groove around the scope tip.

WARNING:
Failure to attach the distal rubber tip properly could result in poor visibility during the procedure (due to the
inability to properly clear debris off the distal objective lens) and/or in difficulty passing accessories through
the scope tip. Additionally, if the distal rubber tip is not attached properly, a possibility exists for the rubber
tip to fall off during the endoscopic procedure. It is imperative that the distal rubber tip be securely
attached to the tip of the endoscope.

➂ Gently depress the rubber air/water nozzle down so that it lies flush with the metallic distal scope tip.

CAUTION :
Make sure there is no space between the rubber tip and metallic scope end. If there is any clearance
between the rubber cap and scope tip, air and water may not flow properly.

e) (ALL SCOPES EXCEPT DUODENOSCOPES) To remove a rubber distal cap from a scope tip, gently hold the distal
rigid portion of the endoscope, just at the black epoxy, with one hand while holding the proximal tab area of the rubber
tip. Gently pull back the rubber tab and slip the rubber cap off the scope.
For DUODENOSCOPES, gently hold the distal rigid portion of the duodenoscope, with one hand while grasping the
very distal portion of the rubber cap with the other hand. Press the distal portion of the rubber cap down and pull off.

• CAUTION :
When removing the rubber cap, do not hold the scope
too tightly and do not grasp the bending section of the
scope. Applying too much pressure to the endo-
WRONG
scope’s bending section could damage the scope.

4) Inspection of Insulated Distal Cap (FOR ED-2330)


Since the ED-2330 incorporates a distal metal tip, an insulat-
CORRECT ed distal cap MUST be attached to the endoscope PRIOR to
use of any electrosurgical devices. See warning statement on
page 25 of this manual. For easier identification, the distal
rubber tip is black in color.

WARNING:
Prior to performing electrosurgery with the ED-2330, always ensure that an isolated distal cap in good con-
dition is properly attached to the distal end of the endoscope. A worn or damaged cap should be replaced
with a new one which has already been subjected to a high-level disinfection or sterilization procedure. For
repeated use, always ensure that the cap has already been reprocessed. NEVER use any electrosurgical
devices with the ED-2330 without the insulated distal cap, OE-A35, attached to the scope distal end.
Electrosurgery performed without this insulated distal cap could result in electrical shock or burn to the
patient or user. Additionally a damaged, worn or missing cap could result in poor visibility during the proce-
dure, and/or in difficulty passing accessories through the scope tip. If the cap is not secured properly, a
potential exists for the cap to fall off during the endoscopic procedure.

–9–
CAUTION :
If there is any doubt, replace the insulated distal cap with a new, fully reprocessed one for each procedure.

a) To attach the insulated distal cap on the ED-2330, moisten the cap with sterile water and align the cap with the scope dis-
tal tip and slide the rubber cap onto the scope and press the cap down until it seats properly.

CAUTION :
Make sure there is no space between the cap and metallic scope end.

b) To remove an insulated distal cap from a scope tip, gently hold the distal rigid portion of the duodenoscope, with one
hand while grasping the very distal portion of the cap with the other hand. Press the distal portion of the cap down and
pull off.

CAUTION :
When removing the cap, do not hold the scope too tightly and do not grasp the bending section of the
scope. Applying too much pressure to the endoscope’s bending section could damage the scope.

5) Inspection of Distal rubber hood (SCOPES WITH ZOOM FUNCTION)


For ideal magnified viewing of close-up surfaces, a rubber distal hood may be attached to the distal end of “zoom” endo-
scopes use.

WARNING:
Prior to use, ensure that a distal rubber hood in good condition is properly attached to the distal end of the
endoscope. A worn or damaged hood should be replaced with a new one which has already been subject-
ed to a high-level disinfection or sterilization procedure. For repeated use, always ensure that the hood
has already been reprocessed. Confirm that the rubber hood lost its elasticity and can still slide over and
hold onto the scope distal tip. If the hood is not secured properly, a potential exists for the hood to fall off
during the endoscopic procedure.

CAUTION :
If there is any doubt, replace the distal rubber hood with a new, fully reprocessed one for each procedure.

➀ To attach or replace a rubber distal hood, gently press the distal rubber hood down onto the scope tip until the hood com-
pletely covers the circular groove around the scope.
➁ To remove a rubber distal hood from a scope tip, gently hold the distal rigid portion of the endoscope, just at the black
epoxy, with one hand while holding the proximal tab area of the rubber hood. Gently pull back and slip the rubber hood
off the scope.

CAUTION :
When removing the rubber hood, do not hold the scope too tightly and do not grasp the bending section of
the scope. Applying too much pressure to the endoscope’s bending section could damage the scope.

– 10 –
OF-B121 6) Inspection of Air/Water Feeding
a) ALL SCOPES EXCEPT SINGLE/COMMON-CHANNEL SCOPES
Prior to use, the air/water feeding valve (OF-B121) should be inspected.
Remove the valve from the control body and make sure that the black
rubber check-valve (OB-B117) in good condition is properly attached to
Don’t let the skirt the bottom of the metal valve stem.
fold under
• b) ALL SCOPES EXCEPT SINGLE/COMMON-CHANNEL SCOPES
If the rubber check-valve is missing or not attached properly, correctly
reposition the check-valve by turning it several times on the air/water
OF-B117 valve stem. For proper positioning, there should be no clearance (gap)
between the rubber check valve and the metal air/water valve stem.

WARNING:
Prior to use, always ensure that a check-valve in good condition is properly attached to the air water valve.
A worn or damaged check-valve should be replaced with a new one which has already been subjected to a
high-level disinfection or sterilization procedure (Check valve set, model OF-B129 is optionally available). For
repeated use, always ensure that the check-valve has already been reprocessed. A damaged, worn or miss-
ing check-valve could create continous air flow or excessive air insufflation and result in potential patient injury
such as pneumatic perforation. If the check-valve is not attached properly, it could fall off during the endoscop-
ic procedure and result in potential injury to the patient.

CAUTION :
If there is any doubt, replace the check-valve with a new, fully reprocessed one for each procedure. A set of
20 pieces of the OF-B117 is available as the OF-B129.

c) ALL SCOPES EXCEPT SINGLE/COMMON CHANNEL SCOPES


To confirm that the check-valve is attached properly, visually inspect the check-valve and install the air/water valve into
the air/water cylinder on the control body. Proceed to test the air delivery function.
d) Connect the scope to the video processor. Turn air pump “ON” to desired pressure setting.
Place the scope distal tip into sterile water and confirm that no air bubbles exit the distal air nozzle.

WARNING:
If air bubbles are observed during the test, the rubber check-valve MUST BE REPLACED. Repeat the test pro-
cedure with a new check valve (OF-B117).

• e) To inspect air delivery, cover the hole at the top of the


air/water valve and confirm that air flows freely from the
air/water nozzle at the scope disital tip.
Action
• f) By depressing the air/water feeding valve, the water
delivery system is activated. Water should flow in a
steady stream from the air/water nozzle at the distal tip
of the endoscope. (This may take several seconds on the
Air Water initial attempt.) USE STERILE WATER ONLY.
Result –
Feeding Feeding g) Release the air/water valve to determine if the valve
freely returns to its OFF (neutral) position and delivery
of water (and air) ceases.

NOTE: SINGLE/COMMON-CHANNEL SCOPES


For the scopes with a single, common channel for air/water delivery and suction, all three functions occur
within a single channel, the following may be experienced when alternating between these functions:
• A slight delay in air delivery with subsequent residual water (fluid) flow may occur immediately after switch-
ing from water (or from suction) to air activation. This flow of fluid is due to the purging (emptying) of residual
fluid from the common internal channel.
• A slight delay in water delivery may occur if activated immediately after air delivery (or after suction) due to
filling of the empty common internal channel occupied by air.

– 11 –
OF-B121 OF-B132 • h) If the air/water valve (OF-B132 for single/common chan-
nel scopes and OF-B121 for all other scopes) does not
move smoothly, remove the valve and apply a small
amount of silicone oil lubricant onto all the O-rings.
Remove/wipe off excessive lubricant with a soft gauze.
Do not use excessive silicone oil.

O Ring : OF-B143
V Ring : OF-B138

NOTE:
Excessive silicone oil (lubricant) should be avoided to prevent occlusion of the internal air or water
channels/nozzles and potential impairment of the normally clear endoscopic image.

i) If air and/or water do not flow properly, NEVER attempt to clear the air or water nozzles with a needle or any other sharp
object. Instead, the following steps should be followed.
➀ Disconnect the endoscope from the video processor.
➁ Remove the air/water feeding valve.
➂ Using a cotton tipped applicator and alcohol, clean the valve recess (receptacle) in the control body thoroughly to remove
any debris. Do NOT attempt to insert the applicator into the small openings within the valve receptacle as the cotton or
applicator could become lodged within these openings and cause channel blockage.
➃ (30/30K SCOPES)
• Following the companion Reprocessing/Maintenance Manual for cleaning the air and water channel with detergent,
flush detergent through both the air and water channel.
Then flush the air several times to force any residual solution out of the channel.
• Remove the adapters and install the air/water feeding valve.
(Alternate) By leaving the air/water valve in the cylinder instead of the adapter, one may direct pressured fluid (or air)
independently to either channel to expel debris from and/or more forcefully flush solution into either the air or water
channel. This should not be attempted on a completely clogged/blocked air or water channel/nozzle.

NOTE:
If blockage of the line is encountered, avoid use of excessive pressure to prevent scope damage.

➄ (40/40K SCOPES EXCEPT SINGLE/COMMON-CHANNEL SCOPES)


• After attaching the air/water channel cleaning brush adapter (OF-B116) to the air/water valve receptacle and after
removing the distal rubber tip with nozzles from the endoscope distal tip, follow the companion Reprocessing/
Maintenance Manual for cleaning the air and water channels with the A/W channel cleaning brush.
• If prior to using the cleaning brush, the air or water channel was completely clogged, the channel should be reprocessed
using the channel cleaning adapters and exposure to an appropriate cleaning solution and high-level disinfectant.
• Remove the air/water channel cleaning brush adapter and reattach the distal rubber tip and air/water valve. Test for nor-
mal delivery of air and water.
➅ (SINGLE/COMMON CHANNEL-SCOPES)
• After attaching the air/water channel cleaning brush adapter (OF-B133) to the air/water valve receptacle, follow the com-
mon Reprocessing/Maintenance Manual for cleaning the air/water and instrument channel with the cleaning brush.
• If prior to using the cleaning brush, the air or water channel was completely clogged, the channel should be reprocessed using
the channel cleaning adapters and exposure to an appropriate cleaning solution and high-level disinfectant.
• Remove the brush adapter (OF-B133) and reattach the air/water valve. Test for normal delivery of air and water.
➆ Test for normal delivery of air and water. It may be necessary to repeat the above procedure if normal air and water deliv-
ery is still not available.

NOTE:
Do NOT apply excessive force in an attempt to unblock a clogged channel as the channel and/or brush could
become damaged. Whenever channel damage is suspected, the endoscope should be leak tested.
If repeated attempts to flush the air/water system are unsuccessful, do not attempt to use the endoscope on a
patient. Contact the PENTAX Service Department.

– 12 –
NOTE:
Prior to clinical use, it is important that the entire air channel system be dry. Failure to thoroughly dry the air
system could result in an unclear or blurry image caused by very fine droplets of moisture being swept over
and/or onto the objective lens at the distal end of the scope.

7) Inspection of Forward Water Jet (Only for models with a


water jet channel system)
a) Prior to use, the water jet check-valve adapter (OF-B119)
should be inspected. Open the water jet connector cap
(OF-B118). Turn the water jet check-valve adapter coun-
terclockwise and remove it from its cylinder.
OF-B119
Make sure that the black rubber check-valve (OF-B117)
is properly attached to the bottom of the water jet adapter
unit (OF-B119).
Don’t let the skirt
fold under
• b) If the rubber check valve is missing or not attached prop-
erly, correctly reposition the check valve by turning it
several times on the water jet adapter unit.
OF-B117 For proper positioning, there should be no clearance
(gap) between the rubber check-valve and the water jet
adapter stem.

NOTE:
The rubber check-valve OF-B117 is a reusable component and it as well as the OF-B119 check-valve adapter,
water jet cylinder, irrigation tube and water jet channel should be reprocessed after each use.

WARNING:
A worn or damaged check-valve should be replaced with a new one which has already been subjected to
a high-level disinfection or sterilization procedure (Check valve set, model OF-B129 is optionally avail-
able). For repeated use, always ensure that the check-valve has already been reprocessed. A damaged,
worn or missing check-valve could create a risk of cross contamination to the end user due to the potential
for reflux (spit-back) of patient fluids through an unsealed path. If the check-valve is not attached properly,
it could fall off during the endoscopic procedure and result in potential for end user cross-contamination.

CAUTION :
If there is any doubt as to the suitability for use of the rubber check-valve, replace with a fully reprocessed one prior
to each procedure. A set of 20 pieces of the OF-B117 check-valve are available in a package as model OF-B129.

forward water jet connector


• c) Attach the water jet check-valve adapter (OF-B119) and
water jet connector cap (OF-B118) opened and then irri-
water jet check
valve adapter (OF-B119)
gation tube (OF-B113) to the forward water jet connector
on the tip of the control head.
• d) Attach a syringe filled with water to this tube and flush
water through the tube. Water should flow in a steady
irrigation tube (OF-B113) and forceful stream from the water jet nozzle at the distal
water jet connector cap end of the scope (this may take several seconds on the
(OF-B118) initial attempt).
e) Use only sterile water in the forward water jet system.
f) If water does not flow properly, NEVER attempt to clear
the water jet nozzle with a needle or any other sharp
object. Instead, the following steps should be followed:
– 13 –
➀ (30/30K SCOPES)
• Attach a syringe filled with a compatible enzymatic detergent solution to the irrigation tube.
• Flush solution through the tube and nozzle. Soaking of the water jet channel with detergent solution should help dis-
solve and dislodge whatever is restricting the normal forceful stream from the water jet nozzle.
• Remove the syringe with solution and attach a syringe filled with air. Flush air through the irrigation tube and nozzle
several times to remove any residual solution from the tubing and nozzle.
➁ (40/40K SCOPES)
• Remove the irrigation tube (OF-B113) and water jet check-valve unit (OF-B119) from the proximal end of the control
body. Remove the distal rubber cap from the endoscope tip.
• Following the companion Reprocessing/Maintenance Manual, use the A/W channel cleaning brush to brush clean the
internal channel surfaces of the forward water jet system.
• If prior to using the cleaning brush, the forward water jet channel was completely occluded, the channel should be repro-
cessed using the water jet irrigation tube and exposure to an appropriate cleaning solution and high-level disinfectant.

NOTE:
Do NOT apply excessive force in an attempt to unblock a clogged channel as the channel and/or brush could
become damaged. Whenever channel damage is suspected, the endoscope should be leak tested.

CAUTION:
It may be necessary to repeat the above step several times to obtain proper water jet function.
If water does not flow properly, contact your local Pentax service facility.

WARNING:
Although the forward water jet system may not be clinically used during each procedure, it still MUST be
properly cleaned and subjected to the same disinfection/sterilization processes as other internal channels
of the endoscope.

CAUTION:
The forward water jet channel is made of stainless steel, fluorine-contained polymers and poly phenylene
oxide. When any fluids are used with these scopes, please read carefully and follow all instructions in the
manual supplied with the fluids for use and pay special attention to any reactions with the materials identi-
fied in the intended fluid path. Only the user can determine if the fluids are appropriate for patient use.

• NOTE:
The forward water jet system featured in Pentax
FORWARD WATER JET
water shoots straight endoscopes should not be confused with “auxiliary”
out to area of interest or manual water feed systems whose function is
simply to clean the distal objective lens. A true for-
ward water jet allows the endoscopist to direct a
AUXILIARY WATER SYSTEM forceful stream of water to clear blood, debris, etc.
water is directed across
scope tip to clean lens from a particular area of interest to improve visual-
ization.

– 14 –
8) Inspection of Instrument Channel Selection (TWO-CHANNEL SCOPES ONLY)
a) Check the condition of the instrument channel selector OF-B122. The selector knob should move smoothly when rotated
and click into position at the prescribed colored indicators. The instrument channel selector allows the user the choice of
suction capability through either channel, or simultaneous suction through both channels.

HANDLING OPERATION

Simultaneous Yellow Dot Alignment of the instrument channel selector to the combined
suction through Instrument yellow and green indicator allows simultaneous suction
Channel
both channels. Selector through both channels.

Green Dot NOTE: The instrument channel selector knob can be rotated
either clockwise or counterclockwise.

Yellow Dot
Suction through Alignment of the instrument channel selector knob to the sin-
a small channel gle yellow indicator allows suction through the small channel
only. only.

Suction through Alignment of the instrument channel selector knob to the sin-
a large channel gle green indicator allows suction through the large channel
only. only.
Green Dot

• b) Should the instrument channel selector not rotate


smoothly, it might require further cleaning.
➀ Remove the entire instrument channel selector mech-
anism from the endoscope.
➁ Using the large bristle of the specially designed clean-
SELECTOR ing brush, CS-C5S, insert the brush into the opening
CYLINDER of the instrument channel selector cylinder.
Thoroughly clean the surface areas.
• ➂ Next, using the smaller side of the cleaning brush
(CS-C5S) thoroughly clean the instrument channel
selector and rinse well.
• 9) a) Inspection of Cannula/Forceps Elevator (DUODENOSCOPES
AND OTHER SCOPES WITH OBLIQUE OPTICS)
This is the control that will guide and direct either the
cannula, biopsy forceps or other accessory during a pro-
cedure. To inspect, push elevator control knob forward
with thumb of the left hand. The elevator in the distal tip
should elevate in proportion to the distance the control
knob is moved. The motion of the elevator and the knob
should be smooth and easy without any “play” involved.

Cannula/Forceps
Elevator Control
Knob
DUODENOSCOPES

SCOPES WITH
OBLIQUE OPTICS

– 15 –
b) Inspection of Cannula (DUODENOSCOPES ONLY)
Make sure a cannula is clean and free from kinks, and the lumen is patent.

WARNING:
Cannulas and other accessories which enter the biliary tract should be sterile.

• 10)Inspection of magnification control (SCOPES WITH


ZOOM FUNCTION SERIES ONLY).
Push the magnification knob forward with thumb of the left
hand. The image should be magnified in proportion to the
distance the knob is moved. Push the magnification knob
backward. The image should be reduced to original size.
magnification control knob The motion of the knob should be smooth.
Position for viewing standard,
non-magnified image
• 11)Inspection of Suction Mechanism
a) Connect suction tubing from an external suction source
to the suction nipple located on the PVE Connector at the
end of the umbilical cable. Place the distal tip of the
endoscope in a basin of sterile water and depress the suc-
tion control valve. Water should be rapidly aspirated into
the suction system collection container.
Standard image Magnified image b) Release the suction control valve to determine if the
valve freely returns to its OFF position and the aspiration
DEPRESS of water ceases.

CAUTION :
If the instrument is to be used immediately after
SUCTION the inspection, use only “fresh” distilled or sterile
TUBE
water. To avoid recontamination of a previously
SUCTION
NIPPLE reprocessed endoscope, avoid use of idle/uncov-
ered tap water.

NOTE:
Rubber inlet seals in good condition must be on the
instrument channel inlet to prevent the loss of suc-
tion and a risk of cross contamination to the end
user due to the potential for reflux (spit-back) of
patient fluids. Worn seals will result in leakage and
should be replaced. To ensure maximum perfor-
OF-B120
mance of these sealing mechanisms, consider
replacing the rubber inlet seal with a new fully repro-
cessed one for each procedure.

• c) If the suction valve does not move smoothly, remove the


valve from the suction cylinder on the control body of
the endoscope. Apply a small amount of silicone oil
O-Ring Silicone oil
lubricant, OF-Z11, onto rubber part and rubber O-rings.
OF-B127
Remove/wipe off excess lubricant with a soft gauze. Do
not use excessive silicone oil.

– 16 –
CLOSE 12)Inspection of Biopsy Forceps and Instrument Channel
OPEN a) Make sure there are no kinks in the flexible shaft of the
biopsy forceps.
• b) The jaws of the forceps must be free of any residual
debris. Any debris must be cleaned from the forceps
before they are used. Use only sterile forceps.
c) The handle mechanism on the forceps should be operated
to open and close the jaws. This mechanism should oper-
ate freely.
d) Close and inspect the jaws of the forceps to make sure the
cups are in proper alignment. If the forceps has a spike, the
spike must be completely straight and fully within the cups.

WARNING:
The use of any forceps or accessory that shows any sign of damage or difficulty of operation must be
avoided. Any malfunction of a forceps or accessory during a patient procedure could result in serious
injury to the patient. Also, the use of damaged forceps or accessories may result in serious and costly
damage to the endoscope.

e) Any accessory should be slowly inserted through the instrument channel inlet with the endoscope in a straight position.
There should be no resistance encountered. If resistance is encountered, do not attempt to introduce the accessory further.
The instrument channel may be damaged and the endoscope should not be used. Contact the Pentax Service Department.

WARNING:
All patient contact accessories must be thoroughly cleaned and subjected to an appropriate high-level disin-
fection or sterilization process before being used for the first time and subsequently after each clinical use.

CAUTION:
The instrument, A/W and the water jet channel systems are made of stainless steel, poly phenylene oxide
and fluorine-contained polymers. When any fluids are used with this scope, please read carefully and follow
all instructions in the manual supplied with the fluids for use and pay special attention to any reactions with
the materials identified in the intended fluid path. Only the user can determine if the fluids are appropriate for
patient use.

13)Other Accessories/Devices (SMALL BOWEL ENTEROSCOPES)


Overtubes, stiffening devices, etc. have been used with the enteroscopes to assist in the introduction and advancement of the
insertion tube. These auxiliary accessories help prevent bowing (looping) of the flexible shaft as it passes through the G.I.
tract. Users are advised to exercise extra caution and care with these peripheral devices for the well being and safety of both
patient and instrument.

WARNING:
Depending upon the accessory, anatomy of a patient, and/or the degree of tip deflection, the user may
encounter resistance due to increased friction between the accessory and accessory/suction channel. The
user is cautioned to be prepared for a sudden exit of an accessory from the scope distal tip as the scope
position, anatomy, or tip deflection changes.

NOTE:
Accessories should always be inspected and checked with the particular endoscope prior to each procedure.

– 17 –
2-3. PREPARATION JUST BEFORE INSERTION OF ENDOSCOPE

WARNING:
Every endoscope should be properly disinfected or sterilized before being used for the first time. The
endoscope should have been properly cleaned, disinfected or sterilized after any previous use and after
being returned for any repairs/service.
Refer to the companion manual describing in detail Pentax reprocessing instructions.

WARNING:
Current infection control guidelines require that endoscopes and their patient contact accessories either be
sterilized or at the least be subjected to high-level disinfection. Accessories which ENTER STERILE TIS-
SUE or THE VASCULAR SYSTEM must be sterilized before patient use. It is recommended that any
accessory intended for use in the biliary tract be subjected to an appropriate sterilization process. Only the
user can determine if an instrument has undergone appropriate infection control procedures prior to each
clinical use.

1) Gently wipe the insertion tube with a dampened gauze.


2) Gently clean the objective lens with a cotton-tip applicator moistened with alcohol. A lens cleaner (anti-fogging agent) may
also be applied via gauze or other applicator.
3) Check the endoscopic image and confirm that it is of acceptable quality for clinical use. Refer also to the owner’s manual
supplied with the Pentax video processor for inspection of the image quality.
4) (SCOPES TO BE INTRODUCED TRANSORALLY)
Place a bite block on the insertion tube to protect the insertion tube before the scope is introduced.
5) Apply a medical grade water soluble lubricant to the insertion tube. Do not use petroleum based lubricants.

NOTE:
The objective lens must be kept free of the lubricant and excess lens cleaner.

WARNING:
Never drop this equipment or subject it to severe impact as it could compromise the functionality and/or
safety of the unit. Should this equipment to be mishandled or dropped, do not use it. Return it to an autho-
rized Pentax service facility for inspection or repair.

– 18 –
3. DIRECTIONS FOR USE

WARNING: This instrument should only be used by physicians who have thoroughly studied all the characteris-
tics of this instrument and who are familiar with the proper techniques of endoscopy. During the procedure, always
wear protective garments such as gloves, gowns, face masks, etc. to minimize the risk of cross contamination.

3-1. PRETREATMENT
1) The patient should be prepared in your normal endoscopy regimen.

NOTE: (SMALL BOWEL ENTEROSCOPE ONLY)


It has been suggested by various physicians that prior to the enteroscopic procedure, transfusing and hydrating the patient
with obscure bleeding will provide adequate blood volume to fill vascular channels and improve the diagnostic yield.

3-2. INSERTION AND WITHDRAWAL


1) (SCOPES WITH ZOOM FUNCTION)
Push the magnification control knob backwards for standard, non-magnified viewing.

WARNING: For safety reasons, always insert and advance the endoscope in the standard, non-magnified
mode. Magnified vision reduces the depth of the viewing field making it difficult to clearly see and identify anatomi-
cal landmarks. Do not advance the endoscope through the intestine in the magnified (zoom) mode.

• 2) Slowly insert the scope under direct vision.


3) (SCOPES TO BE INTRODUCED TRANSORALLY)
When the distal end of the scope is passed through the phar-
ynx, the patient should be gently biting down on the bite block
BITE BLOCK
to maintain the bite block’s position during the procedure.
4) Adjust the intensity of the video processor to obtain a bright-
ness level suitable for observation.

WARNING: The light emission from the endoscope could cause thermal injury. To minimize the risk, use only
the minimum amount of brightness and avoid close stationary viewing and unnecessary prolonged use.

5) The deflection controls should be used as needed to position the scope. The deflection of the tip should be done under direct
vision in a gentle and deliberate manner.

CAUTION: ANY lack of smooth operation of the deflection controls may be an early indication of internal dam-
age to and/or part(s) failure within the endoscope’s angulation system. To avoid the potential for malfunction of
the angulation system, do NOT use the endoscope if the angulation mechanism does not operate properly.
Ensure that the deflection controls can rotate smoothly, that there is no grinding or excess friction within the
angulation system and that the distal bending section bends freely and smoothly.
NEVER APPLY EXCESSIVE FORCE TO THE DEFLECTION CONTROLS!
If during a procedure angulation is lost in any direction such as when “cables snap” (broken pulley wire, broken
angle wire, etc.), do NOT continue to use the instrument and do NOT rotate the deflection controls. Should the
angulation system fail for any reason, stop the procedure, release the lock lever and carefully withdraw the
endoscope under direct visualization.
The examples above are indications that service is required to avoid more serious problems with the angulation
control system, including the possibility of a “frozen” distal bending section.
A “frozen” bending section can make instrument extraction from a patient more difficult.

– 19 –
NOTE: (SMALL BOWEL ENTEROSCOPE ONLY)
– The response of the distal tip deflection may be somewhat slower compared to a standard size gastroscope.
The distal bending section may appear sluggish when returning to its neutral position.
– A flexible overtube may be used to facilitate passage beyond the stomach/pylorus by preventing looping of the
scope shaft along the greater curvature of the stomach.
– To help advance the enteroscope, certain techniques such as palpation of the abdomen may be necessary.
Palpation may also be used to unpleat mucosal folds of the bowel to improve intraluminal visualization.

6) Insufflation can be controlled by the combined use of the air valve or CO2 valve to increase the amount of insufflation and
the suction control to decrease the level of insufflation. The user may elect to use CO2 delivery in lieu of air delivery by
using the CO2 gas adapter (OF-G11) in combination with a special CO2/water valve.

WARNING:
It must be recognized that variations in air flow (pressure and volume) for patient insufflation may exist from one
manufacturer’s equipment (processor and/or scope type) to another. It is, therefore, important to closely monitor the
patient at all times and to aspirate excessive air to prevent overinsufflation and potential pneumatic perforation.

CAUTION: (SINGLE/COMMON-CHANNEL SCOPES)


Because the air and water functions of these scopes share a common channel, alternating between each results in
a slight prolonging of each function prior to actual delivery of the activated feeding mechanism. Therefore, the
amount of air or water flow is more compared to a conventional gastroscope with standard independent channels.

7) Mucous or other debris may be aspirated via the instrument channel to improve visualization. Do not attempt to aspirate solid materials.

NOTE: SINGLE/COMMON-CHANNEL SCOPES


For the scopes with a single, common channel for air/water delivery and suction, all three functions occur
within a single channel, the following may be experienced when alternating between these functions:
• A slight delay in air delivery with subsequent residual water (fluid) flow may occur immediately after switch-
ing from water (or from suction) to air activation. This flow of fluid is due to the purging (emptying) of residual
fluid from the common internal channel.
• A slight delay in water delivery may occur if activated immediately after air delivery (or after suction) due to
filling of the empty common internal channel occupied by air.

8) The objective lens may be cleaned during the procedure by alternately using the air/water and suction control valves.

NOTE:
Should debris on the objective lens be difficult to clean, one can temporarily use the HIGH air pressure set-
ting on the processor and simultaneously press the air/water and suction control valves. Return air pressure
setting to original selection before proceeding.

9) (Only for models with a water jet channel system)


A Forward Water Jet Spray may be directed at the target area during the procedure as necessary. Use sterile water only.
10)Image capture, hard copy documentation, video recording, etc. may be carried out as necessary.
11)Before withdrawing the scope, trapped air should be suctioned to reduce patient discomfort.
12)When attempting to withdraw the scope, return the deflection locking levers to their free position. Always withdraw the
scope under direct visualization.
UP/DOWN DEFLECTION LOCK
FREE POSITION
(LOCK RELEASED) RIGHT/LEFT • WARNING:
DEFLECTION LOCK
FREE POSITION
If for any reason, the image is lost due to power shortage,
(LOCK RELEASED) lamp or processor failure, etc. the deflection locking levers
should be released, the scope tip should be straightened
to its neutral position, and the insertion tube should be
LOCK POSITION carefully and slowly withdrawn from the patient.
LOCK POSITION

– 20 –
3-3. BIOPSY

CAUTION:
For ALL types of endoscopic accessory instruments, always maintain a view of the accessory during
advancement, use and withdrawal of the device.

1) Insert the forceps through the slit in the rubber inlet seal. Be certain to hold the forceps handle in such a way to ensure that
the jaws of the forceps are in a fully closed position during insertion.

• NOTE:
When the cups are first passed through the inlet
5c
m seal, a temporary resistance will be encountered.
Hold the shaft tightly at about 5cm from the cups
and push it through.

NOTE:
During insertion, if the forceps are found hard to advance further due to resistance, decrease the deflection of
the bending section to a level suitable for smooth insertion and insert the forceps again.

CAUTION:
Never apply excessive pressure when introducing
any accessory since the instrument channel may
be damaged. Malfunction of the scope as well as
costly repairs may result.

• 2) When a portion of the cups of the forceps becomes visible in


the viewing field, carefully advance the forceps onto the tar-
get area.
3) Open the forceps cups and advance the forceps against the
target area. Carefully squeeze the forceps handle to close the
cups and obtain a specimen within the cups. Always main-
tain a view of accessory during advancement.
4) Withdraw the forceps slowly with the cups closed.

NOTE:
Because of the effect accessories used in the instrument channel of the endoscope can have on the perfor-
mance of the endoscope itself, it is strongly recommended that only PENTAX accessories be used with PEN-
TAX endoscopes. If a unique or highly specialized accessory is available from another source, please contact
PENTAX to arrange a test of its compatibility before using it through the PENTAX endoscope.

WARNING: (SMALL BOWEL ENTEROSCOPE ONLY)


Depending upon the accessory, anatomy of a patient, and/or the degree of tip deflection, the user may
encounter resistance due to increased friction between the accessory and accessory/suction channel.
The user is cautioned to be prepared for a sudden exit of an accessory from the scope distal tip as the
scope position, anatomy, or tip deflection changes.

– 21 –
3-4. CHOLANGIOPANCREATOGRAPHY (ERCP) DUODENOSCOPES ONLY
1) Insert the cannula into the biopsy/instrument channel
through the rubber inlet seal, there could be strong resistance
• from the inlet at first. Hold the cannula approximately
5~10cm from the distal tip and push it through the inlet. Use
repeated short strokes to advance the cannula.
2) Attach a luer lock syringe filled with contrast material to the
cannula. Inject, until air is eliminated from cannula. This
will maintain the integrity of the lumen and contrast media
while the cannula is in use.
3) Insert the tip of the cannula into the Ampulla of Vater.
4) Inject contrast material slowly into the duct under visualization.
5) Remove cannula slowly.

NOTE:
Should resistance in passing the cannula be encountered at the distal portion of the scope, gently pull back
the cannula, reduce the angle of the cannula elevator, then re-advance the cannula.

CAUTION:
If the cannula elevator is not deflected at all, the cannula may not be seen in the field of view since this is a
side viewing instrument. It is recommended that the elevator be slightly deflected so that the cannula exits
the distal scope tip and advanced slowly only under full view.

WARNING:
Accessories which ENTER STERILE TISSUE or THE VASCULAR SYSTEM must be sterile. It is recom-
mended that accessories intended for use in the biliary tract should be sterilized before patient use.

3-5. BILIARY DRAINAGE (ERBD) DUODENOSCOPES ONLY

NOTE:
Endoscopic Retrograde Biliary Drainage should be performed only by those physicians who are completely
familiar with endoscopy and ERBD procedure.
The following is meant solely for the safe passage of catheters and biliary prosthesis through the flexible
endoscope. It is not meant to be used as instructions for the procedure itself.

1) Pass the guidewire into the desired location and keep it in position for prosthetic implantation.
2) Thread prosthesis onto the guidewire, then using the pushing catheter, advance prosthesis through the scope and into position.
3) When the prosthetic device has been placed in desired position, withdraw the guidewire and pushing catheter.

NOTE:
There are several types of prosthetic devices available, be sure to review the specific instructions supplied
with each prosthesis.

– 22 –
3-6. LASER (EXCLUDING DUODENOSCOPES)
Laser equipment should only be used by physicians who have thoroughly studied all the characteristics of the equipment and
who are familiar with the proper techniques of endoscopic laser therapy. The user must carefully read and follow all instructions
in the operating manual supplied with the Laser equipment. The Laser equipment should be carefully and thoroughly inspected
and calibrated. Only the user can determine if the condition of the Laser equipment is suitable.

WARNING:
The Pentax endoscopes identified in this manual and with serial numbers beginning with A*****, B*****,
C***** or D***** are compatible with Nd:YAG laser only. Do not use these endoscopes with other types of
laser such as Diode, KTP, He-Cd, or Excimer Laser Systems. Pentax Video Endoscopes with serial num-
bers beginning E***** or higher are compatible with Diode Laser (wavelength 800 nm ~ 1000 nm) in
addition to YAG Laser. Never use those endoscopes with any types of laser other than Diode laser and
YAG laser. It could result serious injury to the patient.

CAUTION:
Pentax video duodenoscopes are not laser compatible, therefore never attempt to use laser equipment with
these instruments. Use of laser with non-laser compatible video scopes will create a total “white out” result-
ing in a momentary loss of image while the laser is activated.

CO2 GAS
• 1) The user has the option of using a non-explosive gas for
ADAPTER insufflation. Non-explosive gas from a pressure-regulated
OF-G11
and flow-rate controlled source can be connected to the pro-
vided or optionally available gas adapter, Model OF-G11, as
CO2 GAS
CYLINDER illustrated.

NOTE:
Never connect an unregulated source of gas to the
PROCESSOR LIGHT GUIDE PLUG PENTAX endoscope.

2) The gas adapter, which can be secured to the air/water port on the PVE connector, has a luer receptacle to accept tubing from
an external source of non-explosive gas. As long as the air/water feeding tube from a PENTAX water bottle assembly is con-
nected to the gas adapter and the air pump in the video processor is turned OFF, non-explosive gas can be delivered.

NOTE:
Set the pressure below 49kPa (0.5kg/cm2, 7.1PSI) and the flow rate at about 1 liter/min.

CAUTION:
Open the valve of the CO2 gas cylinder only AFTER turning off the pump switch of the processor. Failure to
do so will apply excessive pressure to the processor and can cause damage to the air pump.

3) Flow of gas from the nozzle at the distal end of the scope can be checked by placing the tip of the scope under water and cov-
ering the hole on the top of the air/water valve. The flow rate of gas should be no greater than the rate of air delivery when
the air feeding valve on the control head of the scope is covered.
4) The water delivery system is activated by depresing the air/water feeding valve.
5) The operator and assistant(s) should wear surgical gloves to avoid burns during use of Laser equipment.

– 23 –
NOTE:
It should be noted that as long as the valve of the CO2 gas cylinder is OPEN and the hole at the top of the A/W
feeding valve is NOT covered, CO2 gas will constantly be vented through the A/W valve into the room. To reduce
excessive CO2 concentrations, it is, therefore, recommended to close the CO2 gas cylinder valve, work in a well
ventilated room, and use air delivery whenever possible during examinations which are lengthy or in very con-
fined quarters. As an alternative, the optionally available gas/water valve, Model OF-B130, may be used in
place of the standards air/water valve. OF-B130 is a closed two-stage valve mechanism. Pressing the first
stage delivers CO2 gas and fully depressing the second stage activates water delivery.
When using the OF-B130 valve, there will be no venting of CO2 gas into the room.
Replace OF-B130 with the air/water valve OF-B121 after using the CO2 gas.

NOTE:
One may choose to leave the OF-G11 adapter attached to the endoscope during conventional air insufflation
using the standard air/water valve. However, the luer sideport of the OF-G11 must be capped.
Similarly, for normal water delivery, the air pump must be turned ON and the plastic luer lock cap must be
secured to the OF-G11 adapter.

6) The laser probe should be introduced through the endoscope in the same manner as described for biopsy forceps in section 3-3.
7) The position of the active portion of the laser probe should always be clearly visualized before Laser equipment is activated.
8) It should be recognized that a variety of factors can effect the quality of the video endoscope image during laser use.
Intensity of the aiming beam, high power setting of the laser, close distance of laser fiber to scope tip, excessive tissue burn-
ing, can each adversely influence image quality. To obtain optimum results, it is recommended that the power settings of the
aiming beam and laser be adjusted to minimal levels capable of achieving the desired clinical effect.
9) Follow standard hospital protocol regarding safe-use of lasers, including the wearing of safety eyewear.

WARNING:
Prior to activation of the laser, make sure that the laser fiber exits the distal channel opening of the endoscope.
Failure to do so can result in scope damage and possible ignition/burning of the endoscope.

NOTE:
It is normal for the aiming light to appear white in the video scope image.

SMEAR • 10)Should the distal tip of the endoscope be moved closer than
20mm from the irradiated tissue surface, the aiming beam
IRRADIATE
AREA may create a “smear” in the image as shown at left. If this
smear affect becomes too severe and distorts the visual field,
the intensity of the aiming light should be decreased.
• 11)When activating the laser at high power (about 100 W for
Yag Laser and 60W for Diode Laser.) and/or if the endo-
PROBE scope tip comes to within 10mm of the irradiated tissue,
flare may appear at the corners of the image as shown at left.
FLARE

WARNING:
Activation of the laser at high power settings
may cause patient injury or thermal damage of
the endoscopes tip.
Avoid use of high power.

PROBE

– 24 –
3-7. ELECTRO-SURGERY

WARNING:
Please refer to the operating manual provided with the electrosurgical unit. Electrosurgical systems may
be of the floating type (BF type, CF-type) or non-floating (B type). To avoid patient and user burn, use only
the floating type ESU (such as ERBOTOM ICC 200) /accessory systems. Do not use the non-floating (B
type) electrosurgical systems. The electrosurgical generator and any electrosurgical accessory should be
carefully and thoroughly inspected. Only the user can determine if the condition of the electro-surgical gen-
erator and the electro-surgical accessory are suitable.

WARNING: FOR ED-2330


Always attach the insulated distal cap model OE-A35, which has been subjected to a high-level disinfec-
tion or sterilization to the distal end of the ED-2330 prior to performing any electrosurgical procedures with
this endoscope. NEVER use any electrosurgical devices with the ED-2330 without the insulated distal cap,
OE-A35, attached to the scope distal end. Electrosurgery performed without this insulated cap could result
in electrical shock or burn to the patient or user.

CO2 GAS • 1) The user has the option of using a non-explosive gas for
ADAPTER
OF-G11
insufflation. Non-explosive gas from a pressure-regulated
and flow-rate controlled source can be connected to the pro-
CO2 GAS vided or optionally available gas adapter, Model OF-G11, as
CYLINDER
described for Laser in section 3-6.
2) The electrosurgical accessories should be introduced through
LIGHT the endoscope in the same manner as described for biopsy
SOURCE LIGHT GUIDE PLUG forceps in section 3-3.

WARNING:
To avoid patient and user burn, follow the instruction below before electrosurgical energy is delivered.
1) Use only the electrosurgical generator with the floating grounding type (BF or CF Type). Do not use the
non-floating (B type) electrosurgical systems.
2) Wear rubber gloves and face masks.
3) The position of the target area, the insulated distal portion of the electrosurgical accessory and the
active portion of the electrosurgical accessory, should be visible.
4) The active portion of the electrosurgical accessory should not touch the metallic distal portion of the
endoscope directly or via fluids.
5) The metallic portion of the endoscope should not touch the surrounding tissue directly or via fluids.
6) The active portion of the electrosurgical accessory should not touch the surrounding tissue directly or via fluids.
7) The head of any lesion such as polyp should not touch the surrounding tissue directly or via fluids.
8) Physicians and assisting personnel should avoid contact with the patient while high frequency energy is
delivered.
9) Electro-surgical energy should be delivered for as short a time period as necessary to accomplish the
desired clinical effect.
10) Select a high frequency output power setting suitable for the particular intended procedure in order to
avoid thermal invasion of the tissue or insufficient coagulation resulting in excessive bleeding.
11) To avoid the risk of thermal injury, use only insulated accessories.
Never use non-insulated devices while performing endoscopic electrosurgical procedures.
12) Never simultaneously use a diathermic device (e.g. snare) in one channel and another accessory (e.g. for-
ceps, retrieval device) through the other channel UNLESS both accessories are insulated. (TWO-CHANNEL
SCOPES)

– 25 –
4. CARE AFTER USE
The following steps summarize the basic recommended manual reprocessing procedures for the Pentax endoscopes identified on
the front cover of this operation manual. For more details, please refer to the separate companion manual for reprocessing/main-
taining Pentax endoscopes.

(1) Immediately after the procedure, clean/wipe all debris from the insertion tube with an enzymatic detergent solution.
(2) Alternate aspiration of the detergent and air several times through the biopsy/suction channel.
(3) Using the A/W valve, purge the air and water channels within the scope.
(4) Soak removable components (A/W valve, suction valve, etc.) in detergent solution.
(5) Attach the soaking cap (make sure that the ETO vent cap has been removed).
(6) Leak Test the scope PRIOR to immersing in any fluid.
(7) Immerse the entire scope in the enzymatic detergent solution.
(8) Clean all removable components in the detergent solution.
(9) Using the supplied cleaning brushes and/or brush guides, brush clean all applicable areas.
(10) Using the supplied channel cleaning adapters, flush detergent into each channel with a syringe.
(11) Remove the endoscope from the detergent.
(12) Rinse the scope and all internal channels with clean water to remove residual detergent and debris.
(13) Purge all internal channels and dry external scope surfaces.
(14) Immerse the scope and the scope components in the disinfectant.
(15) Using the supplied channel cleaning adapters, flush disinfectant into each channel with a syringe.
(16) Remove the endoscope from the disinfectant.
(17) Flush air to purge disinfectant from the channels.
(18) Rinse scope and channels with clean water. Purge channels.
(19) Use an alcohol rinse followed by forced air to thoroughly dry all surfaces.
(20) When utilizing chemo-thermal processes for reprocessing the endoscope, the instrument should be allowed to return to
room temperature prior to use and/or further handling.
(21) If sterilization is required, refer to the companion manual for reprocessing/maintaining Pentax endoscopes in addition to
any supplemental information identifying Pentax compatible reprocessing agents.

WARNING:
Instrument repairs should only be performed by an authorized Pentax service facility. Pentax assumes no
liability for any patient/user injury, instrument damage or malfunction, or REPROCESSING FAILURE due
to repairs made by unauthorized personnel.

WARNING:
Never drop this equipment or subject it to severe impact as it could compromise the functionality and/or
safety of the unit. Should this equipment to be mishandled or dropped, do not use it. Return it to an autho-
rized Pentax service facility for inspection or repair.

– 26 –
LEAKAGE TESTER INSTRUCTIONS
GREEN ZONE
• The PENTAX Leakage Tester allows for simple two (2) stage
TE
ST
19.6
RED ZONE testing of the watertight integrity of the PENTAX Endoscopes.
D

Air pressure is delivered by means of a hand operated bulb,


AN

29.4
AIR BLADDER
GER

9.8
LEKAGE TESTER
SHA-P2

eliminating the need for any electro-mechanical devices.


39.2
0
kPa

PRESSURE LEAKAGE
RELEASE TESTER
VALVE (Be certain CONNECTOR DRY TEST, STAGE I
HAND this connection
BULB is secure.)
BEFORE IMMERSION, PENTAX Endoscopes should be test-
ed for any major loss of integrity in their watertight construction
(example: major tear in the instrument channel). All valves,
inlet seals, distal rubber tips and other removeable components
should be detached from the scope prior to leak testing.
TES
T
• 1) Secure the Leakage Tester connector to the air vent on the
0.2

scope light guide. The PVE soaking cap should be securely


D
AN

0.1 0.3
GER

0 0.4
kg/cm2

attached to/over the electrical contacts. The Leakage Tester


PVE connector and the air vent on the scope light guide MUST be
Scope Connector Soaking Cap
dry before connecting. Proper connection will require align-
ment of the air vent pin and clockwise rotation of the
Leakage Tester connector.
• 2) Turn the gauge faceplate to ‘zero’ the pressure indicator.

ST
Faceplate
rotates.
• 3) Pressurize the scope by pumping the hand bulb until the
TE 19.6 indicator on the gauge is in the GREEN zone.
D
AN

29.4 DO NOT pressurize into the RED zone, as it may cause seri-
GER

9.8
LEKAGE TESTER
SHA-P2

39.2
ous damage to the scope.
0
kPa
NOTE:
During the leak test procedure, the insertion tube
of the scope should be flexed in various positions
and the distal bending section should be fully
angulated in all directions to confirm the absence
GREEN ZONE of a leak.

TE
ST 4) Observe the gauge pressure to determine if the indicator
19.6
D

remains in the GREEN zone. If the indicator drops from the


AN

29.4
GER

9.8
LEKAGE TESTER
RED ZONE GREEN zone rapidly, a major leak may be indicated.
SHA-P2

39.2
0
kPa NOTE:
Be certain that the pressure release valve on the
handle of the Leakage Tester has been tightened.
DO NOT IMMERSE the entire instrument if the
gauge indicator does not remain in the GREEN
zone. Instead, contact your PENTAX Service
Center.

– 27 –
WET TEST, STAGE II
After determining the absence of any major leak in Stage I test-
ing, PENTAX Endoscopes may be immersed to test for loss of
integrity in their watertight construction.

• CAUTION:
PVE
Soaking Cap BEFORE IMMERSING:
ETO Cap
a) The ‘Red’ ETO gas sterilization venting cap
must be taken OFF.
b) The PVE Soaking cap should be securely ON
the electrical contacts.

• 1) With the Leakage Tester securely attached to the scope and


the scope pressurized with the gauge indicator in the
;;;;;;
;;;;;;
;;;;;;
;;;;;; GREEN zone, and the PVE Soaking Cap securely attached
;;;;;;
;;;;;;
over the electrical contacts plug on the umbilical cable of the
scope, the entire scope with ALL valves and inlet seals
removed may be immersed in clean water.

NOTE:
Only the Leakage Tester connector and a small
portion of its tubing should be immersed. NEVER
immerse the entire Leakage Tester.

• 2) Observe the instrument carefully while fully angulating the


;;;;;;
;;;;;; distal tip of the scope. A few bubbles may occur initially
;;;;;;
;;;;;;
;;;;;;
;;;;;;
from recessed areas of the scope. This is normal. If a contin-
uous stream of bubbles is observed from the same spot, a
leak is indicated. Immediately remove the scope from the
water.
DO NOT use the scope.
3) After removing the scope from water, release the air pres-
sure within the scope by opening the pressure release valve
on the handle of the Leakage Tester. After the gauge indi-
cates ‘zero’, disconnect the Leakage Tester from the scope.

NOTE:
NEVER connect or disconnect the Leakage Tester
under water. This will cause leakage of water into
the scope and Leakage Tester.

4) If leakage was discovered in step (2), thoroughly dry the


instrument and contact your PENTAX Service Center.
5) If no leakage was discovered in step (2), you may proceed
with cleaning and disinfection of the scope as described in
the companion Reprocessing/Maintenance Manual.

– 28 –
MEMO

– 29 –
MEMO

– 30 –
SPECIFICATIONS
EC-3830M EC-3830F
ES-3830 EC-3830S2 EC-3830L
EC-3830M2 EC-3830F2
Field of View 120°
Depth of Field 5 ~ 100 mm
Up - Down 180°-180°
Tip Deflection
Right - Left 160°-160°
Rigid Distal Diameter ø13.4 mm
Insertion Tube Diameter ø12.8 mm
Diameter of Instrment Channel ø3.8 mm
Insertion Tube Working Length 700 mm 1,150 mm 1,300 mm 1,500 mm 1,700 mm
Total Length 1,027 mm 1,477 mm 1,627 mm 1,827 mm 2,027 mm

EC/ES-3830 SERIES NOTE: Specifications are subject to change without prior notice and without any
LIGHT GUIDE obligation on the part of the manufacturer.
OBJECTIVE LENS

AIR/WATER NOZZLE
INSTRUMENT Directs pressurized air
CHANNEL across objective lens to
remove moisture, also
provides distension of
cavity being examined
WATER JET NOZZLE or directs pressurized
Directs pressurized spray water across objective
at surface being viewed lens to clear debris
SPECIFICATIONS
EC-3430M EC-3430F EC-3430L
Field of View 120°
Depth of Field 5 ~ 100 mm
Up - Down 180°-180°
Tip Deflection
Right - Left 160°-160°
Rigid Distal Diameter ø11.5 mm
Insertion Tube Diameter ø11.7 mm
Diameter of Instrument Channel ø3.5 mm
Insertion Tube Working Length 1,300 mm 1,500 mm 1,700 mm
Total Length 1,627 mm 1,827 mm 2,027 mm
ES-3831 EC-3831S2 EC-3831M EC-3831F EC-3831L EC-3832L EC-3830TM EC-3830TF EC-3830TL EC-3832TL
EC-3831M2 EC-3831F2
Field of View 140° 120°
Depth of Field 3 ~ 100 mm 5 ~ 100 mm
Up - Down 180°-180°
Tip Deflection
Right - Left 160°-160°
Rigid Distal Diameter ø13.4 mm
Insertion Tube Diameter ø12.8 mm ø13.2 mm ø12.8 mm ø13.2 mm
Diameter of Instrment Channel ø3.8 mm ø3.7 mm, ø2.7 mm
Insertion Tube Working Length 700 mm 1,150 mm 1,300 mm 1,500 mm 1,700 mm 1,300 mm 1,500 mm 1,700 mm
Total Length 1,027 mm 1,477 mm 1,627 mm 1,827 mm 2,027 mm 1,636 mm 1,836 mm 2,036 mm
EC-3830MZ EC-3830FZ EC-3830LZ EC-3430MZ EC-3430FZ EC-3430LZ
Field of View 120° (standard) 40°(at maximum magnification)
Depth of Field 3-100 mm (standard) 2~3 mm (at maximum magnification)
Magnification Degree* Varies with monitor size. x105 (at maximum magnification with a 14” monitor)
Up - Down 180°-180°
Tip Deflection
Right - Left 160°-160°
Rigid Distal Diameter ø13.4 mm ø12.3 mm
Insertion Tube Diameter ø12.8 mm ø11.7 mm
Diameter of Instrment Channel ø3.8 mm ø2.8 mm
Insertion Tube Working Length 1,300 mm 1,500 mm 1,700 mm 1,300 mm 1,500 mm 1,700 mm
Total Length 1,620 mm 1,820 mm 2,020 mm 1,620 mm 1,820 mm 2,020 mm
Ambient temperature 10 ~ 40°C
Operating
Relative humidity 30 ~ 85%
environment
Air pressure 700 ~ 1060 hPa
Ambient temperature -20 ~ 60°C
Storage
Relative humidity 0 ~ 85%
environment
Air pressure 700 ~ 1060 hPa
Degree of protection against electric shock BF type

EC-3431 SERIES EC-3430 SERIES EC/ES-3831 SERIES


EC-3832L
OBJECTIVE LENS OBJECTIVE LENS
OBJECTIVE LENS LIGHT GUIDE
LIGHT GUIDE WATER NOZZLE
LIGHT GUIDE LIGHT GUIDE
LIGHT GUIDE INSTRUMENT CHANNEL Directs pressurized water across
AIR NOZZLE objective lens to clear debris

AIR/WATER
WATER NOZZLE NOZZLE AIR NOZZLE
Directs pressurized air across WATER JETNOZZLE
INSTRUMENT WATER JET INSTRUMENT objective lens to remove moisture, Directs pressurized spray at surface
CHANNEL WATER JET also provides distension of cavity being viewed
NOZZLE CHANNEL
NOZZLE being examined

EC-3830T SERIES
EC-3832TL EC-3830Z SERIES EC-3430Z SERIES

LIGHT GUIDE WATER OBJECTIVE


AIR/WATER LIGHT GUIDE OBJECTIVE LENS
NOZZLE LENS
NOZZLE
LIGHT GUIDE INSTRUMENT
OBJECTIVE CHANNEL AIR NOZZLE
LENS AIR NOTE: Specifications are subject to
NOZZLE
change without prior notice and
INSTRUMENT WATER JET INSTRUMENT
CHANNEL
WATER NOZZLE LIGHT GUIDE without any obligation on the part of
CHANNELS NOZZLE LIGHT GUIDE
the manufacturer.
NOTICE
These instruments are used with Class A Medical Equipment (specified EN55011) and are
intended for hospital or health care districts.
When used in clinical or residential areas near radio and TV receiver units, these instruments may
be subjected to radio interference.
To avoid and resolve adverse electromagnetic effects, do NOT operate these instruments near the
RF energy equipment.

HOYA Corporation
2-7-5 Naka-Ochiai, Shinjuku-ku, Tokyo 161-8525 Japan

HOYA Corporation PENTAX Life Care Tokyo Office


3-16-3 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 Japan
Tel: 03-5953-2592
Fax: 03-5953-2595

PENTAX Medical Company


A Division of PENTAX of America, inc.
102 Chestnut Ridge Road Montvale, New Jersey 07645-1856
Tel: 201-571-2300 Toll Free : (800) 431-5880
Fax: 201-391-4189

PENTAX Canada Inc.


1770 Argentia Road Mississauga, Ontario L5N 3S7
Tel: 905-286-5585
Fax: 905-286-5571
PENTAX Medical Singapore Pte. Ltd.
102E Pasir Panjang Road, #06-06 Citilink Warehouse Complex,
Singapore 118529
Tel: 65-6271-1669
Fax: 65-6271-1691
88030
● Specifications are subject to change without notice and without
any obligation on the part of the manufacturer. 2009. 11. 6217001 Z231 R15 printed in JAPAN

You might also like