Abl90 Reference Manual English
Abl90 Reference Manual English
ABL90 FLEX
reference
manual
Reference manual
Note to the operators of the ABL90 FLEX analyzer
Instructions to Put the Note to the operators in the binder of your manual and replace the
operators corresponding old front and date of issue pages with the new pages in this update kit.
The contents of this document will be added to the manual the next time the manual is
updated.
3. Wet section
4. Electronics
6. User-defined corrections
9. Solutions
Index
Date of issue
System performance
The procedures described in this manual must be observed in order to ensure proper system
performance, and to avoid hazards.
Radiometer cannot provide or verify system performance characteristics if the system is not installed,
used and maintained in accordance with Radiometer procedures or if accessories not meeting the
specifications provided by Radiometer are used.
Radiometer warrants that the data media on which the software included in the system is furnished is
free from defects in material and workmanship under normal use for three (3) months from the date
of delivery as evidenced by a copy of invoice or receipt.
Confidentiality
The contents of this document shall not be reproduced or communicated to any third party without
the prior written consent of Radiometer.
Changes
This document is subject to change without notice and you are urged to contact Radiometer to verify
whether the document has been changed.
While every effort is made to ensure the correctness of the information provided in this document as
changed from time to time, Radiometer disclaims any liability for errors and omissions.
Radiometer, the Radiometer logo, ABL, AQT, TCM, RADIANCE, AQURE, PICO, CLINITUBES and QUALICHECK are trademarks of or
used under license by Radiometer Medical ApS.
Contents
Reference method.................................................................7-14
BiasPrim.ref .............................................................................7-14
BiasSec.ref and Repeatability – blood samples..............................7-14
Performance test results – FO2Hb .......................................................7-15
Reference method.................................................................7-15
BiasSec.ref and Repeatability – blood samples..............................7-15
Performance test results – FCOHb ......................................................7-15
Reference method.................................................................7-15
BiasPrim.ref .............................................................................7-15
BiasSec.ref and Repeatability – blood samples..............................7-16
Performance test results – FMetHb .....................................................7-16
Reference method.................................................................7-16
BiasPrim.ref .............................................................................7-16
BiasSec.ref and Repeatability – blood samples..............................7-16
Performance test results – FHHb ........................................................7-17
Reference method.................................................................7-17
BiasSec.ref and Repeatability – blood samples..............................7-17
Performance test results – FHbF .........................................................7-17
Reference method.................................................................7-17
BiasPrim.ref and Repeatability – blood samples.............................7-17
Performance test results – bilirubin.....................................................7-18
Reference method.................................................................7-18
BiasPrim.ref .............................................................................7-18
BiasSec.ref ..............................................................................7-18
External test results ..............................................................7-18
Interference tests ............................................................................7-20
pH/blood gas........................................................................7-20
Electrolytes ..........................................................................7-21
Metabolites ..........................................................................7-23
Oximetry parameters ............................................................7-25
FHbF sensitivity for pH changes ..............................................7-26
ctBil sensitivity for MCHC variations.........................................7-26
Anticoagulants (sampling)......................................................7-28
List of references ..................................................................7-29
8. Parameters ...................................................................................... 8-1
General information........................................................................... 8-2
The Deep Picture ................................................................... 8-2
Symbols ............................................................................... 8-3
Ranges and limits .................................................................. 8-4
Derived parameters................................................................ 8-4
Measured parameters ............................................................. 8-5
Contents ABL90 FLEX reference manual
Derived parameters..........................................................................8-15
General information ..............................................................8-15
Acid-base derived parameters.................................................8-15
Oximetry derived parameters .................................................8-16
Oxygen derived parameters....................................................8-16
Units and numerical format of derived parameters ................................8-19
Calculated versus estimated parameters ..................................8-19
Electrolyte parameters...........................................................8-19
Possible ranges and precision (number of decimals) ...................8-20
List of equations ..............................................................................8-23
Units and symbols.................................................................8-23
pH(T) ..................................................................................8-23
cH+(T).................................................................................8-23
pCO2(T)...............................................................................8-23
cHCO3–(P)............................................................................8-23
cBase(B) .............................................................................8-24
cBase(B,ox) .........................................................................8-24
cBase(Ecf) ...........................................................................8-24
cBase(Ecf,ox) .......................................................................8-24
cHCO3–(P,st) ........................................................................8-24
ctCO2(P) ..............................................................................8-24
ctCO2(B)..............................................................................8-25
pH(st) .................................................................................8-25
Hct .....................................................................................8-25
pO2(T).................................................................................8-25
pO2(A) ................................................................................8-26
pO2(A,T)..............................................................................8-26
pO2(a)/FO2(I) .......................................................................8-26
pO2(a,T)/ FO2(I) ...................................................................8-27
p50.....................................................................................8-27
p50(T) ................................................................................8-27
p50(st)................................................................................8-28
pO2(A-a)..............................................................................8-28
pO2(A-a,T) ...........................................................................8-28
pO2(a/A)..............................................................................8-28
pO2(a/A,T) ...........................................................................8-28
(or px).................................................................................8-29
ctO2 ....................................................................................8-29
–
ctO2(av ) ............................................................................8-29
BO2.....................................................................................8-29
ctO2(x) (or cx) ......................................................................8-30
Contents ABL90 FLEX reference manual
·
D O2 ....................................................................................8-30
·
Q t ......................................................................................8-30
·
VO2 .....................................................................................8-30
FShunt ................................................................................8-31
FShunt(T) ............................................................................8-32
RI.......................................................................................8-32
RI(T)...................................................................................8-32
Qx.......................................................................................8-33
sO2 .....................................................................................8-33
FO2Hb .................................................................................8-33
FHHb ..................................................................................8-34
V(B)....................................................................................8-34
Anion Gap,K+ .......................................................................8-34
Anion Gap............................................................................8-34
cCa2+(7.4) ...........................................................................8-34
mOsm .................................................................................8-34
FHbF ...................................................................................8-34
pO2(x,T) ..............................................................................8-35
VCO2/V(dry air) ....................................................................8-36
VO2/V(dry air) ......................................................................8-36
Oxyhemoglobin dissociation curve (ODC).............................................8-37
ODC equations .....................................................................8-37
The ODC reference position ....................................................8-37
The ODC displacement...........................................................8-38
The actual ODC position.........................................................8-39
Determining the actual displacement .......................................8-39
Coordinates on the ODC.........................................................8-41
Conversion of units ..........................................................................8-42
SI units ...............................................................................8-42
Temperature ........................................................................8-42
cK+, cNa+, cCl– .....................................................................8-42
cCa2+ ..................................................................................8-42
Pressure ..............................................................................8-42
ctHb ...................................................................................8-42
–
ctCO2, ctO2, ctO2(av ), BO2 ...................................................8-42
·
VO2 .....................................................................................8-42
cGlu....................................................................................8-43
cLac....................................................................................8-43
ctBil ....................................................................................8-43
ABL90 FLEX reference manual Contents
Index
Date of issue
1. Setup
1-2
ABL90 FLEX reference manual 1. Setup
Menu
Utilities
Setup
Calibration schedule
Replacement setup
Parameters
Units Printers
User-defined data items Printer setup
User-defined notes Automatic printing
Automatic archiving
Automatic backup Miscellaneous setup
1-3
1. Setup ABL90 FLEX reference manual
Analyzer security
General This program allows you to hand over the control of the operators and
security passwords to the RADIANCE system and to allow an anonymous use of the
analyzer and to define the logoff time of an operator.
To enter this program, press Menu > Utilities > Setup > Analyzer Security
> General Security.
To hand over the control of the operators and passwords to the RADIANCE
system, activate the check button next to "Enable centralized User
Management".
With this option enabled you can only view the operators, not add, edit or
remove any of them. All users defined on the ABL90 FLEX analyzer are deleted
and the list of users in the RADIANCE system is copied to the ABL90 FLEX
analyzer.
NOTICE: This option can only be enabled if RADIANCE communication is
enabled in the RADIANCE Connection Setup program.
To define, how the user should log on, use the up/down arrow buttons in the
"Authenticate" box to select the desired logon option. The following options are
available:
User ID/Password as primary
This option allows you to enter or scan a User name and password in the
Logon screen. By pressing the Log On BC button a Logon-barcode can be
scanned.
User ID/Password only
This option allows you to enter or scan the user name and password in the
Logon screen.
Logon-barcode as primary
This option allows you to enter or scan a Logon-barcode in the Logon screen.
By pressing the Extended Log On button a user name and password can be
scanned.
1-4
ABL90 FLEX reference manual 1. Setup
Logon-barcode only
This option allows you to enter or scan a Logon-barcode in the Logon screen.
To allow an anonymous use of the analyzer, i.e. use without logon, use the
up/down arrow buttons to select "Yes" in the "Anonymous use" box (default)
and select the desired access profile of the anonymous operator, in the "Access
profile for anonymous operator" box. The "Access profile for anonymous
operator" box only appears, when "Yes" is chosen in the "Anonymous use" box.
See Access profiles further in this section for information on how to define the
access profiles.
To set the time interval to elapse, before an operator is automatically logged off,
press the Logoff time button. Select the logoff time in minutes (from 0 to 60)
and seconds (from 0 to 50 in 10-second intervals). The default logoff time is
three minutes. Press Back to return to the General Security screen.
Operators and This program allows you to add, edit or remove operators and to assign an
passwords access profile to each operator.
NOTICE: If the Centralized User Management option is enabled in the General
Security screen, you cannot add, remove or edit the operators, but only view
the access profiles of the individual operators.
To enter the Operators and Passwords program, press Menu > Utilities >
Setup > Analyzer Security > Operators and Passwords.
When the analyzer is taken into use, the following default operators are
available:
1-5
1. Setup ABL90 FLEX reference manual
1-6
ABL90 FLEX reference manual 1. Setup
2. Touch and highlight the box to be changed. Enter the change. If you
change the passwords, confirm them again.
The password must be at least four characters long. The logon
barcode and the password can be identical.
3. Press Back.
If the password is not accepted, the Add New Operator screen
remains open and a message, telling you what was wrong, appears.
If the password is accepted, the Operators and Passwords screen
is displayed.
4. In the Operators and Passwords screen, select the desired access
profile of the new operator.
To remove an operator from the list, use the up/down arrow buttons in the
"Operator" box to highlight the operator and press Remove Operator.
1-7
1. Setup ABL90 FLEX reference manual
Access profiles This program allows you to define the permitted actions, the available menu
items and button shortcuts of an access profile.
To enter this program, press Menu > Utilities > Setup > Analyzer
Security > Access Profiles.
To define the permitted actions of an access profile, select the desired profile in
the "Profile names" box and activate the desired check buttons in the "Permitted
actions" box.
To deactivate an action, press the check buttons once again.
To define the available menu items and button shortcuts of an access profile do
the following:
Step Action
1. In the Access Profiles screen highlight the desired access profile in
the "Profile names" box and press Menus and Buttons.
Note that this button is grayed-out for the service technician profile.
2. Select the desired menu items in the "Menu Items in Quick Menu"
box.
Selected profile
is named on the
screen.
1-8
ABL90 FLEX reference manual 1. Setup
Open/close submenus
Enabling the “My Results” option will give the operator an easy access to all
Patient Results made by that operator, by displaying the Patient Result Log,
filtered on the operator name.
1-9
1. Setup ABL90 FLEX reference manual
Analysis setup
Program Press Menu> Utilities > Setup > Analysis setup and activate a button to
enter a program.
The following programs are available:
Syringe mode
Capillary mode
Patient reports
Reference ranges
Sample age evaluation setup
Reportable ranges
Sample pre-registration
Syringe/capillary modes
1-10
ABL90 FLEX reference manual 1. Setup
Selecting a
parameter
profile
Step Action
1. Press Parameters on the Syringe modes setup or Capillary
modes setup screen.
2. Select parameters for a given measuring mode by activating a
parameter check button (see Screen elements in section Software,
chapter 2 in the ABL90 FLEX operator's manual).
3. Activate the check button in the "Use dynamic parameters" box to
select parameters during a sample measurement.
Disabled A parameter is disabled, i.e. excluded from the Parameter profile screen and
versus the parameter bar, in General setup > Parameters and input.
deselected
A parameter that has been deselected for the given syringe or capillary mode
parameter
will be measured, but excluded from the displayed and printed patient report.
You can further select or deselect a parameter before or after a measurement
see chapter 4: Sample measurement in the ABL90 FLEX operator's manual.
1-11
1. Setup ABL90 FLEX reference manual
In this program you can enter your own reference ranges and critical limits for
all measured and calculated parameters. For each parameter, you can choose
whether or not to differentiate between the categories of sample type, sex and
age group.
NOTICE: Press the check button to activate a function; press the check button
again to deactivate it.
1-12
ABL90 FLEX reference manual 1. Setup
Setting age
group limits
0 days-1 month,
1-5 months,
5-8 months,
>8 months"
Step Action
1. Press the Age groups button on the Reference ranges and
critical limits screen.
2. Use the following to set/change the age groups:
Left/right arrows to choose the age group limit you want to
change (indicated by a blue circle with a white cross)
Up/down arrows to scroll through the list of possible age limits. As
the list is scrolled, the text on the age-group bar changes
accordingly.
3. Repeat step 2 for each limit to be changed.
4. Press Back when completed to return to the Reference ranges
and critical limits screen.
5. Activate the Age group check button in the Reference ranges
and critical limits screen and select the desired age group.
1-13
1. Setup ABL90 FLEX reference manual
Setting
reference and
critical limits
for each
parameter
Step Action
1. Highlight a parameter on the Reference ranges and critical
limits screen, using the up/down arrows.
2. Enter sample type, sex and age group, if required.
3. Press Edit to edit entries for a highlighted parameter.
4. If any entries are present and you cannot use any of them, press
Clear limits.
Then enter new critical and reference limits, using the keypad and
confirming each entry with Enter.
5. To change a value, touch and highlight it. Then enter the limit and
confirm with Enter.
6. Press Back to return to the Reference ranges and critical limits
screen.
7. Highlight a parameter in the "Parameter list" box to view the limits
on the Reference ranges and critical limits screen.
1-14
ABL90 FLEX reference manual 1. Setup
Reportable ranges
Step Action
1. Scroll to the desired parameter, using the up/down arrows or the
scroll bar.
2. Key in the desired lower limit and confirm with Enter on the
keypad.
3. Key in the upper limit and confirm with Enter on the keypad.
4. To change the reportable range to the default (primary) setting,
highlight the desired parameter and press Set default.
5. To change all parameters to the default values, press Set all
default.
Press Continue to change the reportable ranges of all parameters
to the default ones.
Press Cancel to keep the user-defined reportable ranges and return
to the previous screen.
6. Press Close to exit and confirm the selected settings.
1-15
1. Setup ABL90 FLEX reference manual
Program In this program you can create a number of new layouts for patient reports or
modify the existing ones.
Press Menu > Utilities > Setup > Analysis setup > Patient reports to
access the program.
Creating a
layout
Step Action
1. Press New to make a new layout (marked "New") or Copy to make
a copy of a highlighted layout.
2. Press the Keyboard button next to the "Name" box, type in a new
name for your layout and confirm with Enter to return to the
Patient report setup screen.
3. Enforce, if desired, the Radiometer default settings on the
highlighted layout by pressing -R- default. This will give you a
starting point for designing your own layout.
4. Edit your layout as described in Editing a layout further in this
section.
5. If desired, make the highlighted layout a default for your analyzer
by pressing Make default. It will be marked with () in the list of
layouts.
6. Make a test printout, if desired, of the highlighted layout (patient ID
items and selected parameter groups with the parameters/units for
each parameter group) by pressing Preview. This test print will be
labeled "Preview".
7. To delete a highlighted layout, press Delete.
Note that the Radiometer default layout cannot be deleted. The
button is disabled if only the Radiometer default layout is available.
1-16
ABL90 FLEX reference manual 1. Setup
1-17
1. Setup ABL90 FLEX reference manual
Default values
Changes to
List
Step Action
1. Highlight the desired item in the "Selected items" box with the
arrow buttons.
2. Set the default:
If the item has a value, press Keyboard, enter the value and
press Enter on the keyboard to confirm
If the item has a list of options, press List, highlight the option
using the arrow buttons and press Enter to confirm
3. Set or change other default values in a similar manner.
1-18
ABL90 FLEX reference manual 1. Setup
Editing patient
result layout
Step Action
1. Highlight a patient report layout on the Patient report setup
screen and press Edit patient results layout.
2. Add a highlighted item in the "Available items" box to the list of
and press .)
4. Select another parameter group along with parameters for this
group in the same manner.
5. Use layout commands:
<New group> (items following this command are placed at the
top of the next half of the screen)
<New Line> (a line is inserted between items)
<New Page> (items following this command appear on next
screen page) as desired
and press .
1-19
1. Setup ABL90 FLEX reference manual
This program allows you to select interpretation of the barcode and the patient
data that can be confirmed before and during a sample measurement.
Press Menu > Utilities > Setup > Analysis setup > Sample pre-
registration to access the program.
1-20
ABL90 FLEX reference manual 1. Setup
This program allows you to set up a maximum sample age for the individual
parameters to enable an automatical sample age evaluation.
Press Menu > Utilities > Setup > Analysis setup > Sample age evaluation
setup.
To enable the sample age evaluation of the individual parameters, do the
following:
Step Action
1. Press the check button in the "Enable sample age evaluation" box.
2. Select the maximum sample age in minutes for pH, using the arrow
buttons.
3. To enable the same number of minutes for all parameters, press the
check button next to "Same rule for all the parameters".
4. Press Close to return to the main screen.
1-21
1. Setup ABL90 FLEX reference manual
1-22
ABL90 FLEX reference manual 1. Setup
Step Action
1. Highlight a slot, using the arrow buttons.
2. Solutions from Radiometer (the QUALICHECK5+ control solution):
scan the barcode or press Keyboard to enter the barcode
information (see Barcode reader in section Hardware, chapter 2:
What is what in the ABL90 FLEX operator's manual)
Non-Radiometer control solutions: press Add Non-R-.
3. To delete a control solution, highlight the desired slot and press
Delete to cancel the operation.
A warning that this will irreversibly delete all statistical data related
to the selected slot appears. Press Delete to delete the control
solution.
1-23
1. Setup ABL90 FLEX reference manual
CAUTION – Changing QC
Changing a QC assigned to a slot will delete all current QC statistics
obtained on that slot. If you want a copy of the statistics for the last
QC month, create a WDC Report disk – see chapter 2: Disk functions
setup programs.
In this program you schedule QC measurements, both built-in and manual, for
your analyzer for all days of the week.
Navigation:
and
and
1-24
ABL90 FLEX reference manual 1. Setup
Step Action
1. Select the desired time and press Add to display the screen above.
2. Touch the "QC slot" box to activate it, if not already activated.
Select the desired slot/quality control solution, using the up/down
arrows in the box. Confirm with Select.
Built-in QC results are assigned to the slots A, B and C.
3. Highlight the "Week days" box using Field down and activate the
relevant check buttons to select the days of the week on which this
measurement should be performed.
4. Highlight the "Start time" box, using Field up, and key in the time
to perform a measurement and confirm with Enter on the keypad.
5. Highlight the "Repeat" box and select the interval with which the
measurement should be repeated, using the up/down arrows in the
box.
The QC schedule reminder "Lock analyzer when QC overdue"
(selected in Corrective actions – see further in this chapter) will
work on the basis of the setting selected in this box.
The symbols for the built-in or manual quality control will
automatically appear in the schedule.
6. Press OK to return to the Quality control schedule setup screen.
1-25
1. Setup ABL90 FLEX reference manual
By default the analyzer is set up to run Built-in QCs after replacement and
startup. To deactivate this function deactivate the check button next to the "Run
Built-in QCs after replacement and startup".
QC ranges
1-26
ABL90 FLEX reference manual 1. Setup
Step Action
1. Press Next slot to display the desired slot and then press Edit.
1-27
1. Setup ABL90 FLEX reference manual
Step Action
1. Display the desired slot, using Next slot.
2. Press Update all.
3. Press Continue to update the control ranges for all parameters
under the specified slot, or press Cancel to cancel updating.
NOTICE: Once the Fixed SD has been activated, you cannot update the control
ranges to limits that are narrower than those determined by the Fixed SD, for
both single-parameter and multiple-parameter updates.
QC input setup
In this program you can select the following for the Quality control
identification screen during a manual QC measurement:
Mandatory temperature entry by the operator
The default temperature always displayed (unless changed by the operator)
1-28
ABL90 FLEX reference manual 1. Setup
Step Action
1. Activate the Mandatory temperature check button
2. Or highlight the default temperature in the "Default temperature"
box, enter a default temperature on the keypad and confirm with
Enter.
NOTICES:
A will appear next to an empty temperature box on the
Quality control identification screen during each QC
measurement; otherwise the result cannot be retrieved.
The value in C or F is automatically entered on the Quality
control identification screen during measurement. The
temperature can be changed for a particular measurement but
will return to the default setting for future measurements.
QC statistics
Step Action
1. Key in the desired statistics factor (from 1.0 to 9.9) on the keypad
and confirm with Enter. The default value is 1.5.
2. Activate the check button in the “Built-in QC” field to automatically
make a printout of the QC statistics if the lot is changed.
3. Press Close to exit.
NOTICE: Statistics factor expands the control range to the statistics range (it is
the range within which QC results must fall in order to be included in the QC
statistics).
1-29
1. Setup ABL90 FLEX reference manual
In this program you can select Westgard Rules for all slots or for specific
parameters.
Step Action
1. Select the desired slot, using Next slot.
2. Press On/Off to activate the assigned Westgard Rules for the slot
or press this button again to deactivate them.
NOTICE: The Westgard Rules are a set of statistical rules. When applied to the
QC results, they can increase the probability of detecting an error in the
sampling procedure or in the analyzer itself, or they can help detect a shift or
trend in your QC results by comparing current measurement values of a QC
solution with previous values.
1-30
ABL90 FLEX reference manual 1. Setup
Step Action
1. Display the desired slot, using Next slot and press Edit.
2. Display the desired parameter using Next param or Prev param.
3. Activate the desired Westgard Rule(s) by pressing the
corresponding check button.
(All future quality control data for the given slot/parameter will be
evaluated according to the selected Westgard Rule(s).)
4. Select Westgard Rules for other parameters or levels in the same
manner.
5. Press Back to return to the Westgard Rules setup screen.
Step Action
1. Display the desired slot, using Next slot.
1-31
1. Setup ABL90 FLEX reference manual
2. Press Select All or Deselect all and verify the information on the
screen.
Press Continue. Changes are made and shown in the Westgard
Rules Setup.
Press Cancel. No changes are made.
Reference Westgard JO, Barry PLL. Cost effective quality control: managing the quality and
productivity of analytical processes. Washington: AACC Press, 1992.
RiLiBÄK ranges
The RiLiBÄK ranges program allows you to define a set of rules to control the
maximum deviation of any parameter from the assigned target value.
The assigned target values are given on the QC insert.
It is possible to define more than one rule for the individual parameters.
To activate or deactivate the RiLiBÄK rules, do the following:
Step Action
1. Press the On/Off button to activate/deactivate the assigned
RiLiBÄK rules.
1-32
ABL90 FLEX reference manual 1. Setup
2. Select the desired parameter from the parameter list shown in the
right side of the screen.
3.
Press until the first "Lower Limit" box is highlighted and enter
the desired lower limit.
4. Highlight the next box and select "<" or "<=".
5. Highlight the first "Upper Limit" box and select "<" or "<=".
6. Highlight the next "Upper Limit" box and enter the desired lower
limit.
7. To select the desired +/- range, press the desired radio button.
8. Enter the desired +/- range in the "Ranges" box.
9. Press Back to return to the RiLiBÄK ranges screen. The added
RiLiBÄK rule is now shown in the screen.
1-33
1. Setup ABL90 FLEX reference manual
2.
Use or to jump between the input boxes and edit the
desired values.
3. Press Back to return to the RiLiBÄK ranges screen.
1-34
ABL90 FLEX reference manual 1. Setup
Replacement setup
Program Press Menu > Utilities > Setup > Replacement setup to access the
Replacement setup and activate a button to enter a program.
The following programs are available:
Replacement schedule
User activities
Maintenance planning
Replacement warning
In this program you can schedule routine replacements along with the current
scheduled date and interval for replacement. The settings selected here are then
used in Replacement on the Analyzer status screen.
1-35
1. Setup ABL90 FLEX reference manual
Step Action
1. Highlight the replacement action to be scheduled and press Edit.
1-36
ABL90 FLEX reference manual 1. Setup
User activities
In this program you can formulate and schedule your own activities (e.g.
cleaning analyzer, replacing printer paper, etc.) along with the current
scheduled date and interval. The settings selected here are then used in the
Replacement status.
Adding a user
activity
Step Action
1. Press Add to display the Edit user activities schedule screen.
2. Press the Keyboard button and type a new activity. Confirm with
the Enter button on the keyboard.
3. Select the interval, using the up/down arrows in the "Interval" box.
4. Type in the "Next date", using the screen keypad. Confirm with
Enter.
5. Press Back to return to the User activities screen and repeat
steps 1-4 for each activity to be scheduled.
1-37
1. Setup ABL90 FLEX reference manual
1-38
ABL90 FLEX reference manual 1. Setup
Maintenance planning
In this program you can plan replacements during a week and the shift.
Step Action
1. Activate the check buttons for the days on which maintenance is to
be performed.
2. Select the time at the beginning or the end of the shift, using the
up/down arrows.
3. Press Close to confirm the settings.
1-39
1. Setup ABL90 FLEX reference manual
Replacement warnings
In this program you can set the time for a warning to appear before a
replacement. This will affect the status of the traffic light on the main screen.
Step Action
1. Select the number of remaining tests before the warning should be
given, using the up/down arrows.
2. Select the time before a replacement warning, using the up/down
arrows.
3. Select the expected measurements per day, using the up/down
arrows.
4. Press Close to confirm the settings.
1-40
ABL90 FLEX reference manual 1. Setup
Parameter setup
Disabling/enabling a parameter:
Step Action
1. Highlight a parameter on the screen, using the scroll facilities.
2. Press Enable/Disable to include/exclude the parameter from a
parameter profile and the parameter bar. Note that pH, pCO2 and
pO2 cannot be excluded.
1-41
1. Setup ABL90 FLEX reference manual
Locking/unlocking a parameter:
Step Action
1. Ensure that the analyzer is not connected to the RADIANCE system,
as parameters can be locked/unlocked from the RADIANCE system.
2. Highlight a parameter on the screen, using the scroll facilities.
3. Press Lock/Unlock. (This button is grayed-out if the analyzer is
connected to the RADIANCE system.)
4. To unlock a parameter, highlight it and press Lock/Unlock. The
traffic light on Analyzer status will change from YELLOW to a color
corresponding to the analyzer's overall status.
NOTICE: A locked parameter will show YELLOW on the parameter bar and will
change the overall analyzer status traffic light on the Analyzer status screen to
YELLOW. The parameter value will be absent from the printout; however, the
locked parameter will be calibrated.
1-42
ABL90 FLEX reference manual 1. Setup
Step Action
1. Highlight the desired parameter in the Parameter setup screen and
press Edit.
2. Activate (or deactivate) the following check buttons to select (or
deselect) the following functions:
Repression (repress parameter value in patient result in case of
any problems)
Out-of-range suppression for oximetry parameters or ctBil. When
activated, this function is applied to the oximetry/ctBil results
(including those obtained in the past) as follows:
o ctHb values lower than "0 g/dL", but inside the range of
indication will be shown as "0 g/dL"
o Oximetry parameter values (exclusive ctHb) inside the
range of indication, but lower than "0" or higher than
"100 %" will be shown as "0" or "100 %", respectively
o ctBil values lower than "0 µmol/L", but inside the range
of indication will be shown as "0 µmol/L".
3. Enter correction offset and correction slope. Confirm each entry with
Enter.
4. Press Back to return to the Parameter setup screen and repeat
steps 1-3 for another parameter, if desired.
Units setup
In this program you can select the unit for each parameter or group of
parameters.
1-43
1. Setup ABL90 FLEX reference manual
Step Action
1. Highlight a parameter or a group of parameters, using the arrow
buttons.
2. Select the unit, using the arrow buttons, and confirm with Select.
3. Change units for other parameters in a similar manner.
4. Press Close to return to the main screen.
In this program you can include other patient data in the Patient ID layout than
those already available there.
Step Action
1. Highlight an item on the User-defined patient data items screen
and press Edit.
1-44
ABL90 FLEX reference manual 1. Setup
2. Press the Keyboard button on the keypad and type in the new
name of up to 20 characters. Confirm with Enter on the keyboard.
3. Select the data type with the up/down arrows and press Select.
For "Text" entry, go to step 8
For "Numerical" entry, go to step 4
4. Highlight "Unit" and press the Keyboard button on the keypad.
Type in the new name of up to 20 characters and confirm with
Enter.
5. Highlight "Decimals" (if not already done) and the box with "0", "1",
"2","3" is displayed. Choose the number of decimals with the
up/down arrows and press Select to confirm.
6. Highlight "Max. value". Type in the value and confirm with Enter on
the keypad.
7. "Min. value" is now highlighted. Type in the value and confirm with
Enter on the keypad.
8. Press the check button to activate the "Use selection list" function.
To make a list:
Press Add
Type in the item on the displayed keyboard (up to 20 characters)
Confirm with Enter
Add as many items as you wish in the same manner.
9. Press Back to return to the User-defined patient data items
screen. The new entry will be included in the list
NOTICE: The check button in the "Use selection list" can be activated only if the
list contains two or more items.
Step Action
1. Press Menu > Utilities > Setup > Analysis setup > Patient
reports > Edit patient ID layout.
2. Follow the procedure described in Patient reports in this chapter.
Restore default:
To restore the default settings press Restore.
1-45
1. Setup ABL90 FLEX reference manual
User-defined notes
Press Utilities > Setup > General setup > Parameters and input > User-
defined notes.
Adding a Note:
Step Action
1. Activate one of the check buttons on the screen.
2. Press Add.
3. Type the text for the Note, using the screen keyboard. Confirm with
Enter to save the text and return to the previous screen.
Editing a Note:
Step Action
1. Highlight a Note in the "Notes" box, using the up/down arrows, and
press Edit.
2. Edit the text and confirm with Enter.
Deleting a Note:
Step Action
1. Highlight a Note in the "Notes" box.
2. Press Delete.
NOTICE: A list of Notes made for a given option will be marked with a pencil
icon on the relevant screen(s).
1-46
ABL90 FLEX reference manual 1. Setup
Analyzer settings
Program Press Menu > Utilities > Setup > General setup > Analyzer settings and
activate a button to enter a program.
The following programs are available:
Analyzer ID
Time/Date
Acoustic signal
Barometer
Language
Analyzer identification
Step Action
1. Touch and highlight the "Analyzer name" box if not already
highlighted.
2. Type in an identification name and/or number for the analyzer (up
to 32 characters), using the screen keypad or keyboard. Confirm
with Enter.
NOTICE: The installation number cannot be changed. Quote this number in any
technical inquiries you may have to Radiometer.
1-47
1. Setup ABL90 FLEX reference manual
Time/date setup
In this program you can change the current time and date setting.
Step Action
1. Highlight the "Time" box by touching it on the screen.
2. Key in the time on the screen keypad. Confirm with Enter.
Separators are automatically added between hours, minutes and
seconds.
3. Repeat steps 1-2 to set the date.
4. To revert to the previous settings, press Current.
1-48
ABL90 FLEX reference manual 1. Setup
In this program you can set up a short beep to sound after certain events.
Step Action
1. Activate the desired check button(s).
2. Select volume for the acoustic signal or activate the "Mute all
acoustic signals" check button.
1-49
1. Setup ABL90 FLEX reference manual
Barometer setup
In this program you can adjust the automatic barometer in accordance with the
reference barometer in your laboratory.
Step Action
1. Key in the desired pressure value on the keypad.
2. Confirm with Enter. The value will be shown in the "Measured
adjusted" box.
Maximum accepted correction is ±19 mmHg (i.e. the difference
between the "Measured unadjusted" and "Measured adjusted"
settings).
Barometer pressure limits are 450-800 mmHg, or 60.0-106.7 kPa,
or 450-800 Torr.
The units are selected in the Setup program: Units.
1-50
ABL90 FLEX reference manual 1. Setup
Languages
In this program you can select or change a language of your choice from the list
of languages on your analyzer.
NOTICE: It is only possible to select or change languages if they have been
installed. Not all listed languages may be available.
Step Action
1. In the “Select a language from the list” box, select the desired
language with the arrows and press Set language.
To choose a special regional setting, e.g. English (US), select the
desired regional setting in the “Regional language” box and press
Set regional settings.
2. Press Continue to restart the analyzer.
Press Cancel to continue operating the analyzer with the language
unchanged.
1-51
1. Setup ABL90 FLEX reference manual
Communications setup
Program Press Menu > Utilities > Setup > General setup > Communications and
activate a button to enter a program.
The following programs are available:
RADIANCE connection
LIS/HIS connection
Automatic data transmission
Automatic data request
Patient lookup setup
QA Portal connection
See Rear in section Hardware, chapter 2 in the ABL90 FLEX operator's manual
for the identification and location of the serial RS-232 interface connection
(COM) and the network (TCP/IP) RJ45 Ethernet connection.
In this program you can connect the analyzer to the RADIANCE system.
Step Action
1. Touch and highlight the "Server address" box. Type in the TCP/IP
address of your RADIANCE PC, using the screen keypad or
keyboard.
2. Touch and highlight the "Port" box. Type in the port number, using
the keyboard.
3. Touch and highlight the "Password" box. Type in your RADIANCE
password, using the keyboard.
1-52
ABL90 FLEX reference manual 1. Setup
Step Action
4. Press the check button in the "RADIANCE communication" box to
activate connection.
5. The "Connection status" box with the handshake indicates an
established connection to the RADIANCE system.
6. The Icon in the "Connection status" box indicates the state of the
RADIANCE connection. "Connected" indicates an established
connection to the RADIANCE system.
A RADIANCE icon in the Information bar will indicate the established
connection as well.
7. Clear the queue by activating the recycle bin icon. (The "Output
queue" box shows the number of data queued up for transfer. It will
be sent to the RADIANCE system.)
8. Press Close to exit.
In this program you can select the communication protocol for a connected
device.
The "Output queue" box shows the number of data queued up for transfer to
LIS/HIS. Clear, if necessary, the queue by activating the recycle bin icon.
1-53
1. Setup ABL90 FLEX reference manual
Step Action
1. Press Add.
1-54
ABL90 FLEX reference manual 1. Setup
Step Action
5. Connection specifications for serial low-level protocol:
Press Edit to display the Connection specifications screen.
Press Edit to enter the options screen and use the up/down arrows
in each box to select baud rate, Com port and port configuration.
Baud rate: 1200, 2400, 4800, 9600, 14400, 19200, 38400 –
default is 9600
Com Port: COM1, COM2 – default is COM1
Port configuration:
- Data bits: 5, 6, 7, 8, – default is 8
- Stop bits: 1, 1.5, 2 – default is 1
- Parity: None, Even, Odd – default is None
6. Connection specifications for network low-level protocol:
Press Edit to display the Connection Specifications screen.
Touch the screen to highlight the following boxes one after another:
Server Address
Com Port
Reconnect interval
Use the keypad/keyboard to enter the relevant information.
7. Connection specifications for POCTDML1A low-level protocol:
Touch the screen to highlight the following boxes one after another:
Server Address Port Reconnect Interval.
Use the keypad/keyboard to enter the relevant information.
1-55
1. Setup ABL90 FLEX reference manual
Step Action
1. Highlight a desired connection device on the screen, using the
up/down arrows.
2. Activate the relevant check button(s) to select the data to be sent
to the highlighted connection.
NOTICE: If the requested patient data (e.g. Patient Last Name) was received
after leaving the Patient identification screen, the patient report will be
transmitted without the data. To prevent this, select one of the patient data
items transferred from LIS/HIS as mandatory.
1-56
ABL90 FLEX reference manual 1. Setup
In this program you can select the conditions for requesting patient
demographics automatically from the connected RADIANCE system or from the
LIS/HIS computer system when entering patient ID, accession number or
sampler ID.
Step Action
1. Select a connected device in the "From connection" box, using the
up/down arrows.
2. Activate the relevant check button(s) to request patient
demographics when entering:
Patient ID
Accession number
Sampler ID
3. Press Close when completed.
NOTICE: If the requested patient data (e.g. Patient Last Name) was received
after leaving the Patient identification screen, the patient report will be stored
without the data in the Patient report log. The requested patient data will be
stored as a patient profile in the analyzer's database without, however, being
attached to any patient report.
1-57
1. Setup ABL90 FLEX reference manual
In this program you can select the data source from which to obtain the patient
information on the Patient identification screen.
Step Action
1. Select a data source from the established connections (local
database, RADIANCE or LIS/HIS connections).
2. Select the number of days you want each patient to be kept in the
list, using the up/down arrows in the box.
3. Press Close.
1-58
ABL90 FLEX reference manual 1. Setup
Step Action
1. Touch and highlight the "Server address" box.
Type in the TCP/IP address of your QA Portal, using the screen
keypad or keyboard.
2. Touch and highlight the "Port" box. Type in port number, using the
keypad.
3. Press the check button in the " QA Portal communication" box to
activate the connection.
The icon in the "Connection status" box indicates the state of the
QA Portal connection. "Connected" indicates an established
connection to the QA Portal.
4. Press Close to exit the screen.
1-59
1. Setup ABL90 FLEX reference manual
In this program you can select automatic archiving of the data logs by activating
the relevant check buttons.
1-60
ABL90 FLEX reference manual 1. Setup
Step Action
1. Activate the check button to select automatic archiving on the
analyzer's disk.
2. To select another destination, deactivate the check button in the
"Archive destination" box and press the drive icon that appears.
1-61
1. Setup ABL90 FLEX reference manual
In this program you can select automatic backup of all data and system files.
Step Action
1. Activate the check button.
2. Select time for automatic backup by highlighting the "Time" box and
typing in the time, using the screen keypad. Confirm with Enter.
3. Enter the interval between subsequent backups in the "Interval
(days)" box and type in the number of days, using the screen
keypad. Confirm with Enter.
4. Press the drive icon next to the "Destination" box to select
destination.
5. Highlight the drive or folder and press Expand/Collapse to open a
folder in a directory or within a folder.
Note that automatic backup can be selected for the internal disk or
the network.
When completed, the correct destination appears in the upper part
of the box.
5. Press Back to return to the Automatic backup setup screen.
6. Press Close to return to the main screen.
1-62
ABL90 FLEX reference manual 1. Setup
Printers
Program Press Menu > Utilities > Setup > General setup > Printers and activate a
button to enter the program.
The following programs are available:
Printer setup
Automatic printing
Printer setup
In this program you can set up other printers than the analyzer's printer for
making printouts.
Step Action
1. Highlight a printer from the list, using the up/down arrows.
2. Press Select/Deselect to select the highlighted printer for printing.
You can install any number of printers, but only up to 10 printers
can be selected at a time.
3. Activate the check button in the "Manual printing" box to display the
list of printers every time the Print button has been pressed.
If not activated, all selected printers will make a printout every time
the Print button is pressed.
4. Press Edit to display the keyboard to change the highlighted
printer's name, Type a name and confirm with Enter.
5. To install a new printer, press Install printer.
The Add Printer Wizard program appears. This function can be used
by a Radiometer service representative or a person with network
knowledge. To get the desired printer installed the analyzer will run
a restart.
1-63
1. Setup ABL90 FLEX reference manual
Automatic printing
In this program you can select automatic printout of patient, QC (both manually
and built-in) and calibration results plus Activity log messages.
Step Action
1. Activate the desired check buttons for automatic printout.
2. Select automatic printout of several copies (1-5) of patient results,
using the up/down arrows in the "Patient results print options" box.
3. Press User, Manager or Service in the "Message level" box to
select the level for the messages in the Activity log.
4. Press Close to return to the main screen.
1-64
ABL90 FLEX reference manual 1. Setup
Corrective actions
Program In this program you can select the following:
Corrective actions for the events listed in the "Conditions" box
Traffic light signal, if available, for an event
Analyzer action for the subsequent measurements
Step Action
1. Highlight the desired condition, using the up/down arrows in the box.
2. Select an action for this condition from the options in the
"Corrective action(s)" box – see the table below.
3. Select the desired traffic light signal (YELLOW or GREEN, if
available) for the specified event by pressing the traffic light in the
"Traffic light signal" box – see the table below.
4. Select corrective actions/traffic light signal for the other conditions
in a similar way.
Conditions and Conditions and corresponding corrective action options are as follows:
corrective
actions
Condition Corrective action Traffic light
Calibration error(s) Do not run scheduled built-in QC GREEN or
present YELLOW
Calibration schedule Message on next patient result GREEN or
reminder(s) YELLOW
QC error(s) present "?" on specific parameters YELLOW
QC schedule Message on next patient result GREEN or
reminder(s) YELLOW
Lock analyzer when QC overdue GREEN or
YELLOW
1-65
1. Setup ABL90 FLEX reference manual
NOTICE: Critical system messages will always result in a RED traffic light
signal.
The specified traffic light signal and the messages will continue to appear until
the condition no longer exists.
1-66
ABL90 FLEX reference manual 1. Setup
Miscellaneous setup
Program In this program you can select the following options (use the arrow buttons to
display the rest of the options):
1-67
1. Setup ABL90 FLEX reference manual
Option Function
Show parameter bar If disabled, the parameter bar will not appear on the
main screen.
Selecting HbF This option disables HbF correction for all levels, or enables it for all levels or for
correction HbF levels higher than 20 %.
To select the desired option, use the arrow buttons in the box.
Guidelines for selecting/deselecting HbF correction:
For neonatal Use "Enabled for all levels".
samples: It is important to enable HbF correction to obtain
correct results for ctBil, sO2, FO2Hb, FMetHb, FCOHb
and FHHb.
For adult samples: Use "Disabled" or "Enabled for levels > 20 %".
NOTICE: When an adult sample is measured with HbF correction "Enabled for
all levels" or "Enabled for levels > 20 %", it will slightly affect the measurement
of sO2, FO2Hb, FMetHb, FCOHb and FHHb, and will cause a marginal number of
adult samples reported with HbF present.
1-68
ABL90 FLEX reference manual 1. Setup
Analyzer A message, sent from the RADIANCE system to the connected analyzer and
messages displayed on the main screen, can be changed or deleted in this program.
Step Action
1. Press the Keyboard button, type the message (up to 40 characters
long) and confirm with Enter.
To delete the current message, press Delete on the keyboard, or
delete a message and type a new one, if desired.
2. Confirm the change with Enter to return to the Miscellaneous
setup screen.
1-69
1. Setup ABL90 FLEX reference manual
A B C D E F G H I
User X X X (X) X
Supervisor X X X X X X X X
Manager X X X X X X X X
Service techn. X X X X X X X X X
Guest X (X)
Custom 1 (X)
Custom 2 (X)
Custom 3 (X)
Remote X X X X X X X X
operator
Columns D, E, G and H are controlled via the Menu and button configuration
screen settings, not via the check buttons on the Access profiles screen.
(X) means restricted access to data logs:
User can view the logs, but there is no access to the archived data logs
Guest and Custom can view Patient results log and Quality control log
1-70
ABL90 FLEX reference manual 1. Setup
1-71
1. Setup ABL90 FLEX reference manual
cHCO3 (P,st)
1-72
ABL90 FLEX reference manual 1. Setup
1-73
1. Setup ABL90 FLEX reference manual
1-74
ABL90 FLEX reference manual 1. Setup
Vt Numerical L 2
Ve Numerical L 2
Peak flow Numerical L/min 1
Liter flow Numerical L/min 2
Ti Numerical seconds 1
PEEP Numerical cmH2O 1
Pressure support Numerical cmH2O 1
CPAP Numerical cmH2O 1
CMV Numerical Rate 1
SIMV Numerical Rate 1
Flow-by Numerical L/min 1
HFV Numerical Rate 1
I:E ratio Numerical None 2
Wave Numerical None None
ICD9 code Numerical None None
Oxygen device 1 Numerical None None
Oxygen device 2 Numerical None None
Diagnostic code Numerical None None
1-75
1. Setup ABL90 FLEX reference manual
1-76
ABL90 FLEX reference manual 1. Setup
1-77
1. Setup ABL90 FLEX reference manual
Print setup
With this program you can print out all or part of your analyzer setup.
Step Action
1. Press Menu > Utilities > Setup > Print analyzer setup.
2. All check buttons are activated.
Deactivate relevant check buttons to deselect those setups that you
do not wish to be printed out.
3. Press Print to start printing the selected setups or press Close to
return to the main screen.
1-78
ABL90 FLEX reference manual 1. Setup
Press Menu > Utilities > Disk functions > Restore default setup to access
the Restore radiometer default setup program.
You can restore the Radiometer default setup or a setup you have customized
(Customer setup) and saved.
Selecting or deselecting items in the setup – see Loading/restoring setup,
chapter 2: Disk functions setup programs.
Each setup group of settings is described in this section.
Parameters The following settings (i.e. screens and their data) will be restored in the
group Parameters group:
1-79
1. Setup ABL90 FLEX reference manual
General group The following settings (i.e. screens and their data) will be restored in the
General group:
1-80
ABL90 FLEX reference manual 1. Setup
Schedules, etc. The following ini files (i.e. screens and their data) will be restored in the
Schedules, etc. group:
1-81
1. Setup ABL90 FLEX reference manual
Interfacing facilities
Connecting a A mouse connected to the analyzer may be used to activate all the analyzer's
mouse screen functions instead of the operator touching the screen.
A standard PS/2 port mouse or a USB mouse is the sole item that is required for
connection to the analyzer.
Connecting a mouse:
Step Action
1. For a standard PS/2 port mouse only: Switch off the analyzer.
2. Connect the mouse to the mouse port at the rear of the analyzer.
3. For a standard PS/2 port mouse only: Switch on the analyzer. After
restart the mouse is ready for use.
A USB mouse can be used right after it has been connected.
Connecting an An external alphanumeric keyboard may be used to enter data instead of the
alpha-numeric on-screen keyboard. However, to select individual buttons on the analyzer
keyboard screen, you must use a mouse or must touch the screen.
An IBM enhanced personal computer keyboard or a USB keyboard is the sole
item that is required for connection to the analyzer. The keyboard layout must
correspond to the language version used by the analyzer.
Connecting a keyboard:
Step Action
1. For an IBM enhanced personal computer keyboard only: Shut down
the analyzer.
2. Connect the keyboard to the keyboard port at the rear of the
analyzer.
3. For an IBM enhanced personal computer keyboard only: Turn on the
analyzer. After restart the keyboard is ready for use.
A USB keyboard can be used right after it has been connected.
1-82
ABL90 FLEX reference manual 1. Setup
Step Action
1. Use a shielded data cable with an RJ45 connector to connect the
analyzer to a network.
2. The analyzer is first connected to the computer controlling the
information system via one of the following two interfaces:
A serial line (RS232 interface)
An Ethernet interface (TCP/IP)
3. Once the analyzer has been physically connected to the network,
one of two of the protocols stated below is used for communication
with the central computer.
ASTM
HL7
POCTDML1A
External An external barcode reader can be connected and used side by side with the
barcode reader built-in barcode reader – contact your Radiometer service representative.
1-83
1. Setup ABL90 FLEX reference manual
Sample counter
Purpose The sample counter lets you keep track of measurements, calibrations and QC.
Press Utilities > Sample counter to enter the program.
Description
Element Function
Parameter and List the parameters and how many times each has been
Count measured by the analyzer. Normally the count is the same
as the total number of measurements if the parameters have
not been excluded from the measurement(s).
Counters Shows the number of sample measurements, calibrations
and QC measurements made since the sample counter was
last reset ("User" column). The following is registered:
Activity Number of...
Total Completed sample/QC measurements/
calibrations only. Interrupted or aborted activities
are excluded.
Aborted Aborted sample/QC measurements/calibrations
due to sample errors, wet-section errors, etc. –
interrupted activities excluded.
User All completed sample/QC measurements/
calibrations performed by all operators since the
sample counter was last reset.
User counters Gives the date when the counters in the "User" column were
last reset last reset to zero.
Buttons Reset counters resets the counters in the "User" column
(on analyzers with no logon protection of the Setup
programs).
Print prints out information in Counters and in Parameter.
1-84
2. Disk functions setup programs
2-1
2. Disk functions setup programs ABL90 FLEX reference manual
General information
Disk functions To access the Disk functions programs, press Menu > Utilities > Disk
programs functions.
The following programs are available by pressing a corresponding button.
Button Function
WDC report To make a Worldwide DATACHECK report.
Backup all To make a backup of all data. Data is stored as a backup at a
data designated location.
Restore all To restore a backup of all data files to the analyzer's internal
data disk from a designated location.
Export data To export selected records from selected data logs.
logs
Import/ To import externally archived data logs.
Export
To export or delete archived data logs.
archives
Save setup To save the current setup of your analyzer.
Load setup To load a previously saved setup.
Restore To restore all or only some Radiometer default settings.
default setup
Definitions Setup data refers to information or files that configure the analyzer to operate
according to settings defined in the Setup programs.
All data refers to data in the analyzer's internal database, including but not
limited to data logs, setup and system files.
Data storage Information is stored on or retrieved from the internal disk, a network, a
options connected CD-drive (CD-RW, CD-R/RW) or a removable drive (USB mass
storage device).
Disk handling The CD-drive and removable drive (USB mass storage device) should be
rules handled according to the instructions on the packaging.
2-2
ABL90 FLEX reference manual 2. Disk functions setup programs
Step Action
1. Touch the "From:" box in the “Select period” box and set the dates
for the desired month, using the up/down arrows. The date in the
"To:" box will change automatically.
2. Highlight the desired drive or folder (another directory, removable
or externally connected CD-drive) by pressing the Disk drive
button on the screen, and touching it on the screen. Press
Expand/Collapse to open a folder in a directory or within a folder.
When completed, the correct destination appears in the upper part
of the box.
Press Back to return to the previous screen.
3. In the "File name" box (the WDC report screen), press the
Keyboard button to type the file name: you can change the four
characters "WDC_".
Confirm with Enter on the keyboard and return to the WDC report
screen.
4. Send the file to the selected destination by pressing Export data in
the "Export data" box.
Wait until the WDC report screen appears and remove the disk, if
any, with the WDC report.
NOTICES: "Could not create output file" appears if the destination is not
accessible.
"No statistical data found. WDC data not generated" appears
if no data is available for the selected month
2-3
2. Disk functions setup programs ABL90 FLEX reference manual
2-4
ABL90 FLEX reference manual 2. Disk functions setup programs
Step Action
1. Press Change destination to choose the destination.
2. Highlight the drive or folder by touching it on the screen. Press
Expand/Collapse to open a folder in a directory or within a folder.
When completed, the correct destination appears in the upper part
of the box.
If a removable drive is used, connect it to the USB port
Press Back to return to the previous screen.
3. On the Backup all data screen press Start to continue.
4. The backup process begins.
Network drive or internal disk: Backup continues without any
further action from the operator
Removable drive: Wait for the data to be prepared (see the timer
in the current task field located next to the status indicator in the
upper left corner of the screen) and press Start.
5. If the analyzer status shows a "Backup done" message, the process
is complete. Press Close to return to the main screen.
2-5
2. Disk functions setup programs ABL90 FLEX reference manual
Step Action
1. Press Change source to choose the source drive/directory.
2. Highlight the drive or folder by touching it on the screen. Press
Expand/Collapse to open a folder in a directory or within a folder.
When completed, the correct destination appears in the upper part
of the box.
If a removable drive is used, connect it to the USB port.
Press Back to return to the previous screen.
3. On the Restore all data screen press Start to continue.
(Or press Close to cancel and return to the main screen.)
4. The restore process begins.
Network: Restoring does not require any further action
Removable drive: Press Start
5. Complete restoring all data.
When restoring is complete, the analyzer shuts down and restarts
automatically, configured to the information obtained from the
backup file.
2-6
ABL90 FLEX reference manual 2. Disk functions setup programs
Step Action
1. Activate the check buttons next to the data logs to be exported.
2. Activate the calendar icon, the Choose date screen appears. Type
the "From:" date and confirm with Enter. Repeat the same for the
"To:" date.
Press Back to return to the Export data logs screen.
3. Activate the Disk drive button on the Export data logs screen.
Activate the desired drive by touching it on the screen.
Press Expand/Collapse to open a folder in a directory or within a
folder.
When completed, the correct destination appears in the upper part
of the box.
Press Back to return to the Export data logs screen.
4. On the Export data logs screen press Start. The Save Data Logs
screen, showing the data logs to be exported, the amount of saves
and the From-To export dates, appears. Press Start to begin the
export of data to the selected destination.
5. If the dates are different for each exported data log, repeat steps
2-5 for each data log.
2-7
2. Disk functions setup programs ABL90 FLEX reference manual
Importing/exporting archives
Purpose This function allows you to do the following:
Export (or delete) archived data logs stored onto any drive
Import externally archived data logs into the analyzer's archive directory from
any location
Importing an To import an archive, follow the procedure for exporting an archive, using the
archive right-hand section of the screen and Import.
Deleting an To delete an archive from a directory, highlight the desired archive and press
archive Delete.
2-8
ABL90 FLEX reference manual 2. Disk functions setup programs
Saving setup
Purpose You can copy your analyzer's current setup configuration onto a CD-RW,
removable drive, or network. It can be reloaded if the current setup is lost or
damaged or if the same setup configuration should be loaded on other analyzers
without performing all the Setup programs.
Step Action
1. Press Edit location to select destination.
2. Select the required location by touching it on the screen.
If a removable drive is used, connect it to the USB port
Press Expand/Collapse to open a folder in a directory or within a
folder.
When completed, the correct destination appears in the upper part
of the box.
3. Press Back to return to the Save setup screen.
4. On the Save setup screen press Start.
5. When saving is complete, press Close to return to the main screen.
6. Remove the removable drive, if any.
2-9
2. Disk functions setup programs ABL90 FLEX reference manual
Loading/restoring setup
Purpose You can reinstall a saved setup quickly and easily without performing the Setup
programs. If desired, only part of the setup can be loaded, e.g. operators.
Step Action
1. Press Select all to include all items from the list on the screen. Or
press Deselect all to exclude all items from the list.
2. To select single items, highlight the desired item, using the
up/down arrows.
2-10
3. Wet section
Introduction
Definition The wet section of the analyzer is where all samples and solutions are
transported for measurement, calibration, rinse and quality control.
All solutions for the ABL90 FLEX analyzer are contained in the solution pack.
Gas tanks are not necessary with the ABL90 FLEX analyzer, as gas is included in
the solution pack.
3-2
ABL90 FLEX reference manual 3. Wet section
3-3
3. Wet section ABL90 FLEX reference manual
Measuring processes
Introduction The following pages describe the process that occurs within the analyzer during
sample introduction, rinse, calibration and quality controls. The various types of
sampling modes are discussed separately.
All processes refer to the wet section diagram earlier in this chapter.
General information
Prior to When the analyzer is in the Ready mode prior to a measurement, the sensor
measurement cassette contains CAL 1 from the solution pack.
Heating The sensor cassette measurement chamber and the cuvette in the hemolyzer
unit of the optical system is thermostatted to 37 °C to ensure correct measuring
conditions.
Solutions All necessary solutions contained in the solution pack are introduced
automatically as required into the sensor cassette and oximetry module via the
flow selector and inlet.
Waste removal All waste liquids are transported to the waste pouch contained in the solution
pack. This includes blood sample waste.
3-4
ABL90 FLEX reference manual 3. Wet section
Patient samples
Measuring The following table describes the analytical process of a blood sample
process measurement with the ABL90 FLEX analyzer system.
Stage Description
1. The analyzer is ready to accept a patient sample.
"Ready" message is displayed
Traffic light is displaying a GREEN or YELLOW light
The desired parameters are available
2. At the Ready screen, the user lifts the inlet handle to the syringe or
capillary position. The sample (syringe or capillary tube) is pressed
against the inlet gasket and the inlet probe extends into the sample,
which is automatically aspirated. A 1-point calibration is performed
by sampling on the CAL1 (rinse) solution.
3. The sample is drawn into the sensor measuring chamber and the
oximetry module. This process is controlled by liquid sensors that
also check sample homogeneity with respect to air bubbles. A "?"
appears in case of an inhomogeneous sample.
In case of problems during the process or in case of insufficient
sample, the measuring process is aborted, as the validity of the
measuring result may be compromised.
4. When the aspiration is finished, close the inlet.
5. Measurement of the sample is performed as soon as the sample is
positioned in the measuring chambers. The measurement takes 35
seconds. Concurrent with sample analysis, the user enters patient
information as necessary.
6. When the measurement is complete, the results are computed and
then displayed on the screen and a rinse process starts. For further
information on the rinse process see page 3-6.
3-5
3. Wet section ABL90 FLEX reference manual
Rinse process
After a measurement is complete a rinse is performed. The rinse process is the
same, no matter what kind of measurement (patient measurement, QC and
calibration) is performed. The following table describes this rinse process.
Stage Description
1 After the measurement is complete the first part of the rinse is
performed with a mixture of solutions and air.
2. The next part of the rinse is performed with a mixture of solutions
and gas.
3. Thereafter the wet section is checked. The system is filled with gas
to equilibrate the measuring chambers.
4. The entire measuring path is filled with CAL 1 (rinse) solution. The
calibration status is reestablished and the device is now ready for a
new measurement.
3-6
ABL90 FLEX reference manual 3. Wet section
Calibration
The calibration can be divided into four kinds of calibrations:
pO2 calibration
pCO2, cGlu, cLac calibration
pH, cK+, cNa+, cCa2+, cCl– calibration
Oxi calibration
pO2 calibration pO2 is sensitivity calibrated on ambient air and status checked on CAL1. For
further information on the pO2 calibration see Calibration of the pO2 sensor in
section pO2 sensor in chapter 5 in this manual.
pO2 is sensitivity calibrated once a day and status checked with every
measurement.
pCO2, cGlu, pCO2, cGlu, cLac are sensitivity calibrated on CAL3 and status calibrated on
cLac CAL1. For further information on the pCO2, cGlu, cLac calibration, see
calibration Calibration of the pCO2 sensor in section pCO2 sensor and Calibration of the
metabolite sensors in section Metabolite sensors in chapter 5 in this manual.
pCO2, cGlu, cLac are sensitivity calibrated every four hour and status calibrated
with every measurement.
pH, cK+, cNa+, pH, cK+, cNa+, cCa2+, cCl– are sensitivity calibrated on CAL2 and status
cCa2+, cCl– calibrated on CAL1. For further information on the pH, cK+, cNa+, cCa2+, cCl–
calibration calibration, see Calibration of the pH and electrolyte sensors in section pH and
electrolyte sensors in chapter 5 in this manual.
pH, cK+, cNa+, cCa2+, cCl– are sensitivity calibrated once a day and status
calibrated with every measurement.
Oxi calibration ctHb and ctBil are sensitivity calibrated on S7770 ctHb Calibration Solution and
ctHb, ctBil and the oximetry parameters are status calibrated on a transparent
solution (CAL3) from the solution pack. For further information on the oxi
calibration, see Calibration of the optical system in section ctHb and derivates in
chapter 5 in this manual.
It is recommended that ctHb and ctBil are sensitivity calibrated (cuvette factor)
manually every three months by performing the tHb calibration. Also the
wavelength is calibrated. For further information on the ctHb calibration, see
section tHb calibration in chapter 6: Calibration in the ABL90 FLEX operator's
manual.
ctHb and the oximetry parameters are status calibrated every four hours and if
the temperature of the oximetry optical system changes to a temperature
outside drift limits.
3-7
3. Wet section ABL90 FLEX reference manual
Automatic QC
Measuring The solution pack contains three levels of QC solution. The analyzer is designed
process to run each level once every 24 hours. However, it is possible to set up a
schedule to run QC more often, if required, as described in section QC Schedule
in chapter 1.
The QC solutions that come from pouches in the solution pack enter the sample
path through the inlet as a normal blood sample. The only difference is the
position of the inlet that remains in the closed position.
Stage Description
1. When an automatic QC is scheduled to be run, it will postpone
measurements etc., unless the analyzer is busy measuring a blood
sample. In this case, the scheduled QC will be run after the analyzer
has completed the measurement.
2. The QC measuring procedure begins:
For QC level B only: A measurement of high oxygen is performed
on a gas from a pouch that is aspirated before the QC solution.
Measurement of the QC solution is performed as soon as it is
positioned in the measuring chambers.
3. The result is saved in the Quality Control log.
4. The result is compared with the defined control range, measuring
range and statistics range.
5. The absence of any markings next to a parameter indicates that a
parameter was measured without any fault.
Marking Explanation
? Error in the previous calibration, or analyzer
malfunction.
W A violated Westgard Rule.
R A violated RiLiBÄK rule.
Parameter value is outside the control range, but
inside the statistics range.
Only the values within the statistics range are
considered accepted and are included in the QC
statistics.
Parameter value is outside the statistics range and
is not included in the statistics.
Parameter value is outside the range of indication.
Measurement is not included in the statistics.
* Parameter values with user-defined corrections –
see section Parameters and input setup, chapter 1
for details
….. Parameter value could not be calculated, most likely
due to a system error or malfunction. These values
will for the most part be accompanied by a "?". To
obtain a possible explanation, press Message.
6. After the measurement is complete, it is followed by a rinse. For
further information on the rinse process see page 3-6.
3-8
ABL90 FLEX reference manual 3. Wet section
Manual QC samples
Measuring The following table describes the analytical process of a manual QC
process measurement using the manual QC option.
Stage Description
1. The analyzer is ready to accept a QC sample.
"Ready" message is displayed
Traffic light is GREEN or YELLOW.
The desired parameters are available
2. At the Ready screen, the user lifts the inlet handle to the syringe
position. Press Ampoule – QC.
3. The adapter is pressed against the inlet gasket and the inlet probe
extends into the sample, which is automatically aspirated.
NOTE: It is mandatory to use the adapter to minimize the risk of
possible glass pieces from the ampoule getting into the system of
the analyzer.
4. When the aspiration is finished, close the inlet. The screen is now
ready to accept QC information.
5. After the measurement is complete, it is followed by a rinse. For
further information on the rinse process see page 3-6.
3-9
3. Wet section ABL90 FLEX reference manual
3-10
4. Electronics
General information
General The electronics of the ABL90 FLEX analyzer can be subdivided into the following
information modules:
The user interface module which consists of a touch screen, a built-in barcode
scanner and an embedded computer module
An integrated thermal printer
Electronics for control of the wet section pump, valve, sensor cassette,
solution pack and flow selector
Interface to electronic chip for solution pack identification
Power supply unit
Inlet positioning
Sample mixer
4-2
ABL90 FLEX operator's manual 4. Electronics
Sensor Module The Sensor module includes the Wet Section Control, Sensor Interface,
Oximetry System, Selector Detector and Cassette/Instrument ID.
Wet Section Control:
The Wet Section Control PCB handles the measurement system and takes care
of data collection and actuator controls.
It includes a microcontroller circuit, motor drivers, oximetry circuit and
barometer.
It interfaces to User Interface Module, Sensor Interface PCB, Oximetry System
(spectrophotometer and hemolyzer) and other peripheral sensor module PCB´s.
Sensor Interface:
The Sensor interface PCB handles data collection from electro-chemical and
optical sensors.
It includes high-impedance amplifiers and integrated analog-to-digital
converters to acquire sensor signals and transmit those data to Wet Section
Control.
Selector Detector:
The Selector Detector PCB handles the position detection of the selector function
in the liquid cassette.
It communicates with the Wet Section Control through an I2C interface.
Cassette / Instrument ID:
The Cassette/Instrument ID PCB handles data collection for the instrument- and
liquid cassette.
It includes a unique instrument ID and a connector to collect data from the
liquid cassette ID chip.
Oximetry System:
The oximetry system consists of a hemolyzer with cuvette and a 138-
wavelength spectrophotometer with a measuring range of 467-672 nm. The
spectrophotometer is connected via an optical fiber to a combined hemolyzer
and measuring chamber.
Inlet The Inlet Position with guide plate handles the detection of inlet positions and
positioning signaling LED´s.
It includes Hall detectors and an input/output port that communicates with the
Wet Section Control through an I2C interface.
User Interface The User Interface Module includes the CPU Unit, display unit and barcode
Module reader.
4-3
4. Electronics ABL90 FLEX reference manual
CPU Unit:
The CPU Unit runs the operating system and application software.
It includes a compact ETX-PC module mounted on a baseboard that interfaces
signals to internal and external connectors.
It also includes a speaker, a fan and a solid-state disk (CF), in which all
operating system and software files are stored, along with system database
files.
Display Unit:
The display unit handles the Man Machine Interface.
It includes a 8,4” TFT display, resistive touch panel, and an Interface PCB for
LVDS signal, backlight and touch control.
Barcode Reader:
The barcode reader acts as input device for consumables, user and patient
barcodes.
It includes a laser scan engine, a proximity sensor, a buzzer, and a serial
interface.
Printer unit The 4” printer unit handles printouts of instrument and patient results.
The printer unit includes a 4” clamp shelf printer mechanism, a DC-DC converter
and a printer controller with USB interface.
Sample mixer The sample mixer detects and mixes blood samples in safePICO.
(for safePICO
It includes a mixing motor, detectors and a micro-controller that communicates
only)
with the Wet Section Control through an I2C interface.
4-4
5. Sensors and measuring technologies
Overview
General construction
Sensors In this manual, the term sensor refers to an individual sensor as part of the
sensing array within a sensor cassette. The electrical signal from each sensor is
measured by proprietary analog electronics contained within the analyzer unit.
The sensors are located on sensor boards in the sensor cassette.
Top sensor
board:
Ref K Na pCO 2 pH Cl Ca pO 2
Lac G lu
5-2
ABL90 FLEX reference manual 5. Sensors and optical system
Introduction There are four different measuring principles employed in the sensors in the
ABL90 FLEX analyzer.
Potentiometry: The potential of a sensor chain is recorded using a
voltmeter, and related to the concentration of the sample (the Nernst
equation). The potentiometric measuring principle is applied in the pH, pCO2,
K+, Na+, Ca2+ and Cl— sensors.
Amperometry: The magnitude of an electrical current flowing through a
sensor chain is proportional to the concentration of the substance being
oxidized or reduced at an electrode in the chain. The Amperometric measuring
principle is applied in the cGlu and cLac sensors.
Optical pO2: The optical system for pO2 is based on the ability of O2 to
reduce the intensity and time constant of the phosphorescence from a
phosphorescent dye that is in contact with the sample. This measuring
principle is applied in the pO2 sensor.
Spectrophotometry: Light passes through a cuvette containing a hemolyzed
blood sample. The specific wavelengths absorbed and their intensity
generates an absorption spectrum used to calculate oximetry parameters.
This measuring principle is used for measuring ctHb, sO2, FO2Hb, FCOHb,
FHHb, FMetHb, FHbF and ctBil.
The first three measuring principles are described under the sensors, where they
are applied. Spectrophotometry is described in the section titled ctHb and
derivates.
Activity vs. Strictly speaking, in potentiometry the potential of a sensor chain is related to
concentration the activity of a substance, and not its concentration.
The activity of a substance can be considered the "effective concentration" of a
species, taking non-ideality of the medium into account.
Activity and concentration are related by the following equation:
ax = cx
where:
ax = the activity of the species x
= the activity coefficient of species x under the measurement conditions
(for ideal systems = 1)
cx = the concentration of species x (mol/L)
Conversion of The analyzer automatically converts activities into concentrations. The term
activity to concentration is therefore used in explanations of the measuring principles for
concentration each of the sensors further on in this chapter.
5-3
5. Sensors and optical system ABL90 FLEX reference manual
Calibration
5-4
ABL90 FLEX reference manual 5. Sensors and optical system
General information
Definition Calibration is the process that relates the electrode signals during the calibration
sequence to the values of the calibrating solutions and air. Calibration enables
the electrode signals to be converted to the accurate values for an unknown
sample.
Calibration Calibration of all sensors is performed using air, CAL 1 (also used for rinse), CAL
solutions 2 and CAL 3 (see chapter 9, Solutions for more information on the solutions).
The calibration solutions contain known concentrations of the substrates to be
measured. These concentrations are vital in determining the measurement
accuracy of the analyzer.
The concentration of each substance in the calibration solutions is programmed
into the integrated smart chip of the solution pack. The information is
automatically read by the analyzer when a solution pack is installed in the
analyzer.
Traceability of The traceability certificate for the solution pack is found in chapter 9 of this
calibration manual.
solutions
5-5
5. Sensors and optical system ABL90 FLEX reference manual
Use The calibration line forms the basis of the scale used by the analyzer to convert
electrical measurements to concentrations.
Scale The calibration line now forms the scale used to convert the potential measured
at the pH sensor during sample analysis to an actual pH value.
5-6
ABL90 FLEX reference manual 5. Sensors and optical system
Sensitivity
Definition The electrode sensitivity illustrates the slope of the calibration line compared to
the slope of the theoretical electrode.
The sensitivity of the theoretical electrode is 100 % or 1.00.
Measured
potential 2-point calibration line
(mV) 96 Slope = 58.4 mV/pH
Sensitivity = 95 %
pH
6.8 7.3
Sensitivity The slope of the calibration line is described by the sensitivity value.
Status The calibration status values are, in general, defined as the sensor signals of
CAL 1, except for pO2, which is only calibrated in one point (pO2 status reflects
the cal check):
pH pCO2 pO2 cK+ cNa+ cCa2+ cCl– cGlu cLac
mV mV mmHg mV mV mV mV pA pA
Min. -50 -50 -20 150 150 200 -50 0 0
Max. 250 250 20 350 350 400 100 3000 3000
Drift Drift describes the variation in location of the calibration line between
consecutive calibrations. Typically, sensitivity drift is insignificant compared to
status drift. The analyzer automatically compensates for this drift by performing
a 1-point calibration with every measurement.
5-7
5. Sensors and optical system ABL90 FLEX reference manual
Measurement
Sample A blood sample gives a potential reading of 4.8 mV at the pH sensor. Reading
measurements off from the calibration line shown below, this potential corresponds to a pH of
7.35.
Corrections To compensate for deviations from ideal behavior (ex. dilution of sample with
residual rinse solution, and change in gas level by contact with the sample
path), a correction is applied, to give the final value.
The correction is typically a linear correction, and is described for each sensor
type in the following.
5-8
ABL90 FLEX reference manual 5. Sensors and optical system
Quality Management
Introduction This section describes the functionalities that the analyzer, apart from the built-
in quality checks, applies to ensure the measurement quality.
Analysis checks:
Analysis checks are performed in connection with analysis – be it patient sample
analysis, calibration or a QC measurement – consist of the following checks:
Status calibrations/checks
(to check the status of the sensors a status calibration is performed on all the
sensors except on the pO2, where is status check is performed. A detailed
description of the calibrations can be found further on in this chapter under
the individual sensor types.)
Sample integrity checks
(to check for sensor response stability, air bubbles, insufficient sample volume
and sample path obstructions. Blockages can indirect be revealed during these
checks.)
Temperature checks
(to check the sensor array, spectrophotometer and the temperature inside the
analyzer. Are performed continuously and with every measurement.)
Mechanical checks
(to check the pressure, solution pump and flow selector and to check for
leakages.)
5-9
5. Sensors and optical system ABL90 FLEX reference manual
Electronical checks
(to check the sensor impedance and leak current.)
Measurement preparation checks
(to ensure that the analyzer, after each activity, is ready for a new
measurement.)
Consumable lifetime checks
(To check the expiration date and lifetime of the sensor cassette. To check the
expiration date, lifetime and number of remaining tests of the solution pack.)
Software checks:
During the software check, the correct version software in the DMS is checked
against the correct version of the wet section. In case of software consistency
errors the analyzer does not allow measurements.
Communication checks:
During the communication check, the communication between the PC and
embedded systems is checked. In case of errors the analyzer automatically tries
to re-establish the connection.
Temperature checks:
During temperature checks, the temperatures of the sensors,
spectrophotometer, cuvette and barometer (internal temperature) are checked.
If a check of the continuous temperature monitoring fails the analyzer enters
the User intervention required mode that is automatically left again if the
temperature check is ok. After power on or replacements the analyzer waits for
the temperatures to be within limits. In case of temperature drift of the Oxi
spectrophotometer, an Oxi calibration is set pending and performed in
connection with a measurement, if no scheduled calibration has been
performed, thus compensating for the drift. The temperature is monitored and
logged during aspiration of sample, QC and Cal solutions.
Sensor checks:
During sensor response checks, the signal responses of the sensors are
checked. Stability errors are reported by an error message and a "?" next to the
parameter results.
During calibration checks, the calibration values of the sensors are checked to
ensure that the analyzer is ready for measurement. In case of calibration errors
(except severe fluid transport errors) the calibration is retried. The User
intervention required mode is not entered if no other severe errors are
encountered.
Mechanical checks:
During flow selector checks, the calibration/positioning of the flow selector in
the solution pack is checked. If the flow selector calibration fails, the activity is
stopped and retried. If the second calibration fails, the User intervention
required mode is entered.
During pump checks, the volume of the pump flow is checked. If the pump
calibration fails, the activity is stopped and retried. If the second calibration
fails, the User intervention required mode is entered.
Electronical checks:
During leak current checks, the leak current between the reference electrode
and the chassis is checked to detect liquid leaks in the solution pack. If the
check fails, the check is repeated. If the second check fails, the User
intervention required mode is entered.
During impedance checks, the impedance between each pH, cK+, cNa+, cCa2+,
cCl– sensor and the reference electrode is checked. The internal impedance of
5-10
ABL90 FLEX reference manual 5. Sensors and optical system
the pCO2 sensor is also checked. If the check fails, the check is repeated. If the
second check fails, the User intervention required mode is entered.
During liquid sensor checks, the inlet, sensor and oxi liquid sensors check the
liquid transport in the wet section. The calibration of the liquid sensors is also
checked. In case a calibration fails, a rinse is performed and the system
calibration repeated. If the second calibration fails the User intervention
required mode is entered. If the solution is inhomogeneous, the refill program
will perform three retries before aborting. If a rinse, as part of another program,
fails, a new rinse is tried and the User intervention required mode is entered if
the second attempt fails. During aspiration of a sample or internal solution, the
liquid sensors check the liquid transport (both the air segments and the liquid
are expected to trig off the sensors within certain time limits). In case of errors
the activity will be aborted.
Measurement preparation checks:
A rinse is performed after each activity. With the rinse, the temperature and
homogeneity of the rinse solution is checked. In case of rinse error, a rinse will
be performed. If this rinse fails too, the User intervention required mode is
entered.
Sample integrity checks:
The pO2 check is used to check for any blockage or leak in the flow path, and,
furthermore, the pO2 sensor is checked for air in front of the pO2 sensor. In case
of any air the pO2 parameter will be marked with a ? and a corresponding error
message is given. During pO2 checks, pressure tests are performed. In case the
tests fail, the activity is retried. If this activity fails, the User intervention
required mode is entered.
Consumables checks:
During consumables check, the sensor cassette and the solution pack lifetime
and expiration date are checked by inspecting the smart chip data of the
individual consumables. In case of chip data errors the User-intervention-
required mode is entered. If the consumables are used up or have expired, the
User-action-needed mode is entered. The user has the possibility to perform a
replacement.
Apart from the checks mentioned above a flow selector check, pressure test,
refill, pump calibration and a rinse are also performed during a solution pack
replacement. Furthermore, it is checked whether the Solution pack has been
used before. In case of errors, the User intervention required mode is entered.
The user has the possibility to perform a replacement.
During a sensor cassette replacement it is also checked whether the sensor
cassette has been used before and if the minimum/maximum conditioning time
has been met. Thereafter, a pressure test, liquid sensor check, pump
calibration, rinse and a calibration of all the sensors are performed. In case of
any errors the sensor cassette is considered to be unconditioned, and the
analyzer will perform an automatic conditioning and a new calibration that
prolong the startup time to 30 minutes.
By default each of the three built-in QC will be run after replacements and
startup. If this function has been deactivated it is recommended to perform QCs
after replacements and startup.
If the solution pack and the sensor cassette have been replaced at the same
time, all the above checks will be carried out. In case of errors the User-
intervention-required or the User-action-needed mode is entered. The user has
the possibility to perform a replacement.
Furthermore some system checks are also performed after the replacement of
the solution pack and sensor cassette.
5-11
5. Sensors and optical system ABL90 FLEX reference manual
After sensor replacement, the metabolite sensors have a significant drift in the
sensitivity. The analyzer automatically compensates for this by performing a
calibration, when needed, after every measurement that is used to calculate the
measured metabolite values.
5-12
ABL90 FLEX reference manual 5. Sensors and optical system
Reference electrode
5-13
5. Sensors and optical system ABL90 FLEX reference manual
Purpose The purpose of the reference electrode is to provide a stable, fixed potential,
against which other potential differences can be measured.
The potential at the reference electrode is not altered by the sample
composition.
Fixed potential A fixed potential is maintained at the reference electrode by the following
equilibrium reactions:
These reactions are possible because the electrode is made of an Ag rod coated
with AgCl to provide the Ag/Ag+ equlibrium in a solution with constant Cl
concentration and to determine the reference potential.
5-14
ABL90 FLEX reference manual 5. Sensors and optical system
Diagram
Parts and
functions Item Part Description/Function
1 Membrane Interface to the sample.
2 Electrolyte Acts as a salt-bridge solution that maintains
solution an electrical contact between the electrode
and the sample.
3 Electrode Provides the contact between the Electrolyte
solution and the electrical contact.
4 Electrical contact The point of electrical contact between the
electrode and the analyzer.
5 Housing Sensor cassette housing with integrated
reference electrode.
5-15
5. Sensors and optical system ABL90 FLEX reference manual
5-16
ABL90 FLEX reference manual 5. Sensors and optical system
Diagram The pH and electrolyte sensors are of solid-state design with a H+, K+, Na+ and
Ca2+ sensitive PVC membrane. The Cl– sensor is of solid-state design with a Cl–
sensitive epoxy membrane.
23 34
4
Parts and
description Item Part Description
1 Membrane Ion-selective membrane that is in direct
contact with the sample or calibration
solution and that is sensitive to a specific ion,
e.g. the H+ ions.
2 Solid-state The point of electrical and ionic contact with
contact the membrane.
3 Electrical contact The point of electrical contact between the
sensor and the analyzer.
4 Electrode base The structural platform on which the
electrode is formed.
5-17
5. Sensors and optical system ABL90 FLEX reference manual
Potentiometric The pH and electrolyte sensors are measured according to the potentiometric
measuring measuring principle, where the potential of an electrode chain recorded at a
principle voltmeter is related to the concentration of a substance via the Nernst equation.
Electrode chain The electrode chain (or electrical circuit) set up to measure pH/electrolytes is
illustrated in the following diagram:
2 3 4 5 6
Parts and The electrode chain describes an electrical circuit consisting of the following:
description
Item Part Function
1 Voltmeter Measures the voltage potential in the circuit.
2 Reference electrode Provides electrical connection to the voltmeter.
3 Liquid junction Point of contact between the reference sensor
and the sample.
4 Sample The unknown liquid being measured.
5 Membrane An ion-sensitive membrane, which is sensitive
to H+/electrolyte ions.
6 Solid-state contact Provides electrical connection to the voltmeter.
Electrode chain Every element in the electrode chain contributes a voltage to the total potential
potential drop through the chain. Thus:
When immersed in the appropriate electrolyte solution, both electrodes exhibit
separate potentials
The membrane junctions between the sample and electrolyte solutions also
exhibit separate potentials
The total potential across the electrode chain, therefore, is the sum of these
separate potentials, all but one of which are known and constant, as outlined in
the table on next page.
5-18
ABL90 FLEX reference manual 5. Sensors and optical system
Unknown The unknown potential difference across the ion-sensitive PVC membrane is the
potential difference between the measured total potential and the sum of the known
potentials:
Esample=Etotal Eref ELJ EE
Nernst Having measured the unknown potential (Esample), the potential difference across
equation the membrane in the sensor can be expressed by the Nernst equation:
RT
Esample E0 ln ax
nF
where:
Activity and As shown in the equation above, measuring the potential of each of the
concentration electrode chains gives a reading of the activity of the ions in the sample.
Activity expresses the "effective concentration" of a species and is explained in
more detail in the section General measuring principles earlier in this section.
The activity of the ions is automatically converted to a concentration value by
the analyzer.
The relationship between activity and concentration is explained in the section
General measuring principles at the beginning of this chapter.
5-19
5. Sensors and optical system ABL90 FLEX reference manual
Introduction The pH and electrolyte sensors are calibrated by determining the E0 and
sensitivity from 2-point calibrations. Slight variations in sensor performance
between calibrations are addressed by performing a measurement of CAL 1
within every sample measurement process.
2-point A 2-point calibration is performed at preset intervals using two solutions from
calibration the solution pack. The precise values for these solutions are contained in the
smart chip located on the solution pack.
Calibration The pH and electrolyte values for CAL 1 and CAL 2 are as follows (approximate
levels values):
Level
Substance Unit CAL 1 CAL 2
pH - 7.3 6.8
cNa+ mmol/L 150 70
+
cK mmol/L 4 10
–
cCl mmol/L 95 50
cCa2+ mmol/L 0.5 2.3
The solution pH and electrolyte values are known and contained in the solution
pack smart chip.
Calibration The sensitivity is calculated in the following way and expressed as the
percentage of the theoretical sensitivity, calculated from the sensor signal of the
two calibration solutions (mV) and the nominal calibration values:
pH:
mVcal2 mVcal1
S
61,5mV (pHcal2 pHcal1 )
Electrolyte sensors:
n(mVcal2 mVcal1 )
S
c
61,5mV log10 ( cal2 )
ccal1
5-20
ABL90 FLEX reference manual 5. Sensors and optical system
Measurement The pH value measured from the sample is calculated as follows, from the
sensor signal of the sample mVsample:
mVsample mVcal1
pH pHcal1
61,5mV S
c ccal1 10 61,5mV×S
cdisplayed k1 c k 2
The sensor response stability is defined as the standard deviation of the last 5
updatings of the response.
5-21
5. Sensors and optical system ABL90 FLEX reference manual
pCO2 sensor
5-22
ABL90 FLEX reference manual 5. Sensors and optical system
Diagram
Parts and
description Item Part Description
1 Silicone membrane A membrane separating the sample and the
electrolyte solution. Is only permeable to CO2.
2 Electrolyte solution A solution separating the silicone membrane and
the pH membrane, Ag/AgCl sensors. The
electrolyte solution is vented by CO2 whereby pH
is changed.
3 pH membrane H+ sensitive membrane.
4 Reference Ag/AgCl electrode
5 Solid-state contact The point of electrical contact between the pH
for the pH system membrane and the analyzer.
6 Solid-state contact The point of electrical contact between the
for the Ag/AgCl reference electrode and the analyzer.
system
7 Electrode base The structural platform on which the electrode is
formed.
5-23
5. Sensors and optical system ABL90 FLEX reference manual
Electrode chain The electrode chain (or electrical circuit) set up to measure pCO2 is illustrated in
the following diagram:
Parts and The electrode chain describes an electrical circuit consisting of the following:
description
Item Part Description
1 Voltmeter Measures the voltage potential in the
circuit.
2 pH electrode Provides electrical connection to the
voltmeter
3 Electrolyte solution Medium for connection
4 Internal reference electrode Provides electrical connection to the
(Ag/AgCl) voltmeter
Electrode chain The potential differences at all the junctions in the electrode chain are known
potential and constant, except that at the pH-sensitive membrane. (See the section pH
and electrolyte sensors for a full explanation.)
The potential difference at the pH-sensitive membrane depends on the pH of the
electrolyte solution, which in turn depends on the CO2 content of the sample.
This is explained in the measuring process below.
5-24
ABL90 FLEX reference manual 5. Sensors and optical system
Measuring The following is an account of the measuring process in the pCO2 sensor.
process
Part Function
Transport of CO2 CO2 from the sample permeates the membrane.
Dissolution of CO2 The CO2 dissolves in the electrolyte solution.
This produces carbonic acid:
H2O + CO2 H2CO3
Dissociation of Carbonic acid dissociates according to the following
carbonic acid equilibrium reaction:
H2CO3 H+ + HCO3–
pH change The release of H+ ions changes the H+ concentration, and
thus the pH of the inner buffer solution on one side of
the pH-sensitive membrane.
Measurement of The concentration gradient of H+ ions across the
potential membrane creates a potential difference across the
membrane.
This change in potential across the membrane is
measured by the voltmeter.
Relation of pH to The pH value is related to the partial pressure of CO2 in
pCO2 the sample by the following equation:
pH = pK a+ log
HCO
-
3
* pCO2
where:
pKa = log Ka, the equilibrium constant for the
dissociation of carbonic acid in water
= solubility coefficient for CO2 in water
The structure of the pCO2 sensor is similar to the pH
sensor, including the presence of a pH-sensitive
membrane. The major difference is in the internal
electrolyte solution present in the pCO2 sensor which
allows the dissolution and ultimate dissociation of
carbonic acid mentioned above.
If [cHCO3–] and in the electrolyte solution is constant
this results in the following:
pH = K - log pCO2
Where
K contains the equilibrium constant pKa, the
solubility coefficient and the concentration of
bicarbonate [cHCO3–].
5-25
5. Sensors and optical system ABL90 FLEX reference manual
Introduction The pCO2 sensor is calibrated by determining the sensitivity from 2-point
calibrations. Calibration measurements are performed on two levels of solution.
Slight variations in sensor performance between calibrations are addressed by
performing a measurement on CAL 1 within every sample measurement
process.
Calibration The ABL90 FLEX analyzer is equipped with a solution pack. This pack contains
levels precision-tonometered fluids. The tonometry calibration gas mixture is of a
known composition.
The partial pressure of CO2 (pCO2) and the solution pH values are known and
contained in the solution pack smart chip.
Sensitivity The sensitivity is calculated in the following way and expressed as the
percentage of the theoretical sensitivity, calculated from the sensor signal of the
two calibration solutions (mV) and the nominal calibration values:
mVcal2 mVcal1
S
pCO2 (cal2)
61,5mV log10 ( )
pCO2 (cal1)
5-26
ABL90 FLEX reference manual 5. Sensors and optical system
Measurement – pCO2
Measurement The pCO2 value measured from the sample is calculated as follows, from the
sensor signal of the sample mVsample:
Esample Ecal1
The sensor response stability is defined as the standard deviation of the last 5
updatings of the response.
5-27
5. Sensors and optical system ABL90 FLEX reference manual
pO2 sensor
5-28
ABL90 FLEX reference manual 5. Sensors and optical system
Optical system The optical system for pO2 is based on the ability of O2 to reduce the intensity
for pO2 and time constant of the phosphorescence from a phosphorescent dye that is in
contact with the sample.
The optical system for measuring pO2 is shown in the following diagram:
Measuring The green LED emits light, which is reflected by a dichroic mirror onto the pO2
sequence sensor. Due to the phosphorescence, red light is emitted back through the
dichroic mirror and onto a photo detector. The photo detector sends the
electrical signals, proportional to the light intensity, to the analog/digital
converter and the data processing unit. The calculation of the pO2 is performed.
Calculations The pO2 is calculated on the basis of the Stern-Volmer equation, which describes
the relationship between the phosphorescence intensity/time constant () and
the pO2 value in a sample:
pO2 ( ) k 0 1
5-29
5. Sensors and optical system ABL90 FLEX reference manual
Introduction The pO2 sensor is calibrated to determine its sensitivity by measuring one
calibration point during a sensitivity calibration process. Performance of the
sensor from calibration to calibration is checked and during sample or quality
control analysis any drift on sensitivity is checked. The pO2 sensor is calibrated
on ambient air.
Sensitivity The sensitivity is defined as the percentage of the measured pO2 on ambient air
compared to the reference value:
pO2 (meas)
S
pO2 (ref)
where pO2(ref) is the pO2 tension in ambient air saturated with water vapor:
where FO2 is the pO2 fraction in ambient air, and pH2O is the partial water vapor
pressure of saturated air at 37 oC, and p(amb) is the barometric pressure.
Status In connection with the sensitivity calibration performed on ambient air, also the
CAL1 (rinse) solution is measured to obtain a status. This status aims to check
the performed calibration. This is done by comparing the measured value of the
CAL1 (rinse) solution to the reference value of CAL1, given by the smart chip):
pO2 (status,cal) pO2 (CAL1,cal) pO2 (CAL1,ref)
5-30
ABL90 FLEX reference manual 5. Sensors and optical system
Measurement - pO2
On whole blood, pO2 is adjusted with the sensitivity value and the measured pO2
is therefore determined as follows:
pO2 (meas)
pO2 (sens,adjusted)
S
The measured value is applied a 2nd order blood correction, to compensate for
the varying buffer value of blood, as a function of pO2 tension. A second-order
correction is applied:
5-31
5. Sensors and optical system ABL90 FLEX reference manual
Metabolite sensors
5-32
ABL90 FLEX reference manual 5. Sensors and optical system
Basic The cGlu and cLac sensors are three-electrode sensors consisting of an internal
description silver/silver chloride reference electrode, a platinum auxiliary electrode, and a
platinum anode. The sensors are covered by a multi-layer membrane bound to
the sensor board.
The membrane consists of four layers:
The biocompatible layer
The outer membrane – permeable to cGlu/cLac
The enzyme layer
The inner membrane – permeable to H2O2
Diagram
Parts and
description Item Part Description
1 Biocompatible Biocompatible layer
layer
2 Outer membrane Outer membrane permeable to glucose –
diffusion control
3 Enzyme layer Contains glucose/lactate oxidase.
4 Inner membrane Cellulose acetate.
5 Reference Ag/AgCl electrode.
6 Anode Platinum electrode.
7 Cathode Platinum electrode.
8 Electrode base The structural platform on which the sensor is
formed.
Zero current The zero current is a small background current measured at the electrode when
no cGlu/cLac is present in a solution. As CAL 1 contains no cGlu/cLac, a baseline
representing the zero current, I0 as a function of time (I0 = f(t)), is obtained
from continuous measurements on CAL 1.
This I0 baseline is obtained as follows:
At the end of a rinse, with CAL 1 in the measuring chamber, the zero
current of the metabolite electrodes is measured periodically
The previous N (N = 8) measurements on the CAL 1 – before a calibration
or a sample measurement starts – are used to obtain a baseline
representing the time function of I0
5-33
5. Sensors and optical system ABL90 FLEX reference manual
Sensitivity The sensitivity of the cGlu and cLac sensors is calculated by measuring the
current from CAL 3 then subtracting the zero current as measured from CAL 1.
CAL 3 has a nominal glucose concentration of 10 mmol/L and a nominal lactate
concentration of 10 mmol/L. The precise values are specific to the individual lot
of the solution pack and are contained in the solution pack smart chip.
The current at the cGlu and cLac sensors with CAL 3 in the measuring chamber
is measured at regular intervals after the chamber is filled with solution. The
current, when signal stability is reached, is used to determine the sensitivity of
the cGlu or cLac sensor.
where I0 is the zero current estimated to the time of measurement from the 8
samples taken on CAL 1 (rinse).
Status is defined as I0.
5-34
ABL90 FLEX reference manual 5. Sensors and optical system
Measurement – metabolites
Measurement The glucose or lactate concentration in a sample is calculated from the following
equation, using the difference between the current in the sample and the
extrapolated zero current from the rinse solution:
Isample I0
c
S
The sensor response stability is defined as the standard deviation of the last 5
updatings of the response for CAL 1 (rinse).
For CAL 3, it is defined as the standard deviation of a linear regression for the
last 5 samples, normalized with the signal magnitude.
5-35
5. Sensors and optical system ABL90 FLEX reference manual
Amperometric The cGlu and cLac sensors are measured according to the amperometric
measuring measuring principle, in which the magnitude of an electrical current flowing
principle through an electrode chain is related to the concentration of a substance being
oxidized or reduced at an electrode in the chain.
Electrode chain The electrode chain set up to measure glucose/lactate is illustrated in the
following diagram1:
Parts and The electrode chain describes the electrical circuit consisting of the following:
functions
Item Part Description
1 Amperemeter Measures the current flowing through the circuit in
nanoamperes.
2 Cathode Negative electrode where a reduction reaction
occurs and electrons are consumed.
3 Membrane Allows the appropriate molecules to pass through
from the sample.
4 Sample Contacts the membrane.
5 Electrolyte Provides electrical contact between the anode and
cathode.
6 Anode Positive electrode where an oxidation reaction
occurs and electrons are released.
7 Applied voltage Applies the necessary potential for the reduction or
oxidation reaction under study.
1
Note that polarization voltage is applied between the anode and the reference
electrode (not shown). The current runs through the anode and cathode chain.
5-36
ABL90 FLEX reference manual 5. Sensors and optical system
Measuring A constant polarization voltage is applied to the electrode chain. The current
process through this chain is measured by an amperemeter.
Glucose or lactate molecules, in solution, are transported across the outer layer
of a multilayer membrane system. The enzymes glucose oxidase or lactate
oxidase, immobilized between the outer and inner layers, converts
glucose/lactate according to the following reactions:
Glucose: Glucose + H2O + O2 Gluconic Acid + H2O2
The oxygen for this reaction is supplied by the membrane system as well as by
the oxidation of H2O2 at the platinum anode.
The H2O2 produced by the enzyme reaction is transported across the inner
membrane to the platinum anode.
When a potential is applied to the electrode chain, the oxidation of H2O2
produces an electrical current proportional to the amount of H2O2, which in turn
is directly related to the amount of glucose/lactate.
H2O2 2H+ + O2 + 2e–
5-37
5. Sensors and optical system ABL90 FLEX reference manual
5-38
ABL90 FLEX reference manual 5. Sensors and optical system
General information
Measured The optical system of the ABL90 FLEX analyzer is designed to measure the
parameters following parameters:
Parameter Description
ctHb Concentration of total hemoglobin
sO2 Oxygen saturation
FO2Hb Fraction of oxyhemoglobin
FCOHb Fraction of carboxyhemoglobin
FHHb Fraction of deoxyhemoglobin
FMetHb Fraction of methemoglobin
FHbF Fraction of fetal hemoglobin
ctBil Concentration of total bilirubin (the sum of unconjugated
and conjugated bilirubin) in plasma
Spectrophotometer
Array of
photodiodes
Grating
Mirrors
Slit
Optical
Hemolyzer Hemolyzing
fiber
unit
cable
LED light
source
Cuvette
Sample
5-39
5. Sensors and optical system ABL90 FLEX reference manual
Measurement The method used in the analyzer's optical system is visible absorption
cycle spectroscopy. The measurement cycle consists of the following steps:
Step Description
1 The blood sample is transported to the cuvette positioned in the
hemolyzer unit. The temperature of the cuvette is regulated to 37 °C.
2 A back pressure is exerted on the sample. This one atmosphere
over-pressurization is maintained throughout the hemolyzation and
measurement to eliminate air bubbles in the sample and to enhance
the hemolyzation process.
3 The one-µL sample in the cuvette is ultrasonically hemolyzed at a
frequency of about 30 kHz. This hemolyzation process ruptures the
walls of the red blood cells, evenly mixing the content of the red
blood cells with the plasma and producing an optically clear
solution.
4 Light from a white LED is emitted to the cuvette and the light
transmitted through the cuvette is guided to the spectrophotometer via
an optical fiber.
5 The light passes through a slit that directs it towards an
arrangement of mirrors and a grating.
6 The grating separates the light into the colors of the rainbow and
the mirror focuses the light on a photodiode array.
7 The photodiode array has 256 diodes or pixels, one for each
wavelength, which convert the monochromatic light signals to
currents.
8 The currents and therefore the intensity of the light signals are
measured at each of the 256 diodes, which form the basis for the
absorption spectrum for a particular sample.
9 The spectrum is sent to the analyzer’s computer, where the
calculations of the oximetry parameter values are made.
The 256 channels are standardized into 138 selected wavelengths.
Lambert-Beer's Absorption spectroscopy is based on Lambert-Beer's law, which states that the
law measured absorbance for a single compound is directly proportional to the
concentration of the compound and the length of the light path through the
sample [2]:
Ay y cy l
where:
Ay = absorbance of compound y at wavelength
5-40
ABL90 FLEX reference manual 5. Sensors and optical system
Absorbance The absorbance (A) of a compound is defined as the logarithm of the ratio of the
light intensity before and after transmission through the compound.
In practice it is the logarithm of the ratio of the light intensity transmitted
through water to the light intensity transmitted through the compound.
I0
A =log
I
where:
Total For samples containing more than one optically active compound, the total
absorbance absorbance (Atotal) is the sum of the individual compounds’ absorbance, since
absorbance is an additive quantity.
For example, if a sample contains six compounds y1, y2, ….y6, the total
absorbance measured for that sample at wavelength 1 is:
1
Atotal Ay1 Ay1 Ay1 Ay1 Ay1 Ay61
1 2 3 4 5
l y1 cy1 y1 cy2 y1 cy3 y1 cy4 y1 cy5 y1 cy6
1 2 3 4 5 6
If there are Y compounds and measurements are taken at n wavelengths, a
general expression can be written for Atotal at the wavelength n:
Y
n
Atotal yn cy l
y 1
where:
n = the individual wavelengths.
Continuous n
Atotal can be depicted graphically as a function of wavelength, and if the
spectrum
differences between the wavelengths are small enough, a continuous spectrum
is produced.
5-41
5. Sensors and optical system ABL90 FLEX reference manual
Spectrum The figure below shows three spectra; pure O2Hb, pure HHb in a low
examples concentration, a spectrum of 92 % oxygenated hemoglobin obtained by adding
the spectra of O2Hb and HHb. The additivity of absorption and the continuity of
the spectra can be seen.
Absorption
480 500 520 540 560 580 600 620 640 660 680
Wavelength/nm
O2Hb (9.2 mmol/L)
HHb (0.8 mmol/L)
92 % oxygenated hemoglobin (i.e., 92 % O2Hb + 8 % HHb)
0.1
0.08
Absorba nce
0.06
0.04
0.02
0
470 520 570 620 670
nm
Determining In the spectrum taken of a sample, the absorption recorded at each wavelength
concentrations contains contributions from each of the compounds in the sample. The task then
is to determine the magnitude of that contribution and thereby the
concentration of each compound in the sample.
The concentrations are determined using the following equation:
138
cy K yn Atotal
n
n 1
where:
K yn = a constant specific to compound y at wavelength n.
5-42
ABL90 FLEX reference manual 5. Sensors and optical system
Matrix of The constants ( K yn ) are determined using Multivariate Data Analysis 2 where
constants
the spectra of the calibration compounds are considered together with the
reference values of the calibration compounds. The essential interfering
substances (intralipids and sulfhemoglobin) were also taken into account.
5-43
5. Sensors and optical system ABL90 FLEX reference manual
Zero point The zero point, Io, is the current (or intensity) measured by the photodiode
array on the transparent solution in the cuvette. During this "blank calibration"
the ctHb is calibrated to this zero point.
Cuvette path The cuvette path length (i.e. the length of the light path) is determined from
length Lambert-Beer’s Law by measuring the absorbance of the colored dye present in
the tHb Calibration Solution (S7770), which has a known equivalent hemoglobin
concentration.
tHb calibration It is recommended that a tHb calibration is performed every three months. See
frequency section tHb calibration in chapter 6: Calibration in the ABL90 FLEX operator's
manual for further information about the tHb calibration.
5-44
ABL90 FLEX reference manual 5. Sensors and optical system
HbF versus Fetal hemoglobin (HbF) does not have the same spectrum as adult hemoglobin
HbA (HbA) due to a slight variation in molecular structure. The presence of HbF in a
sample will interfere with the result if a correction is not performed.
It is thus important when measuring hemoglobin levels in premature neonates and
neonates aged 0-3 months, as well as adults suffering from e.g. thalassemia, to
take into account this difference [4], and the ABL90 FLEX analyzer automatically
corrects for HbF.
NOTICE: Hb types other than HbA and HbF interfere with hemoglobin
measurements and are not compensated for in the ABL90 FLEX analyzers.
The diagram below shows the transition from fetal hemoglobin to adult
hemoglobin [5].
Deviation of If the difference between the adult and fetal types of hemoglobin is not taken
results into account in measurements on samples containing HbF (e.g. from premature
neonates and neonates aged 0-3 months) then a deviation in the measurement
will arise.
The deviation is most important for measurements of oxygen saturation (sO2
and FO2Hb) and the fraction of carboxyhemoglobin (FCOHb), since inaccurate
measurements of these parameters can lead to incorrect diagnostic
interpretation of the results, and consequent risk of inappropriate treatment.
Detecting HbF The presence of HbF in a sample is detected from the difference spectrum
between fetal and adult oxyhemoglobin. From the size of the difference
spectrum the concentration of fetal oxyhemoglobin, cO2HbF, can be determined.
Correcting for The amount of cO2HbF exceeding a certain level indicates HbF interference. The
HbF analyzer automatically corrects for this interference by subtracting the
difference spectrum of fetal oxyhemoglobin from the measured spectrum.
Repressing Repressing the spectra of the likely interfering substances is done in two ways
spectra depending on the substance:
Either the substance is taken account of in the calculation of the matrix of
constants, K. This applies to Intralipids and Sulfhemoglobin.
Or the substance is detected, and the measured spectrum is corrected
accordingly. This applies to HbF.
5-45
5. Sensors and optical system ABL90 FLEX reference manual
Residual The measured spectrum is compared to a model spectrum calculated from the
spectrum determined concentrations. The difference between these two spectra is called
the residual spectrum. If this residual spectrum is too high, the oximetry module
parameters ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHHb, FHbF and ctBil will be
flagged with a warning.
In addition, a warning will accompany the results if any of the following
conditions exist:
ctHb < 0.1 mmol/L or ctHb > 25 mmol/L
FHb(deriv) < –2 % or FHb(deriv) > 102 %
where FHb(deriv) is defined as sO2, FO2Hb, FCOHb, FMetHb, FHHb
SHb < 2 % or SHb > 10%
Value of Turbidity < 0.5 % or > 5%
5-46
ABL90 FLEX reference manual 5. Sensors and optical system
FHbF cHbF
=
ctHb
where:
where:
5-47
5. Sensors and optical system ABL90 FLEX reference manual
5-48
ABL90 FLEX reference manual 5. Sensors and optical system
ctHb(sample)
ctHb(sample,corr)
Fcuv
where:
ctBil(sample)
ctBil(sample,corr)
Fcuv
Fcuv is the same as for tHb.
5-49
5. Sensors and optical system ABL90 FLEX reference manual
References
1. CLSI/NCCSL document C12-A, Clinical and Laboratory Standards Institute, 940 West
valley Road, Suite 1400, Wayne, PA 19087.
2. Ewing GW. Instrumental methods of chemical analysis. 5th ed. McGraw.Hill, 1985
3. Martens H. Multivariate calibration: quatitative interpretation of non-selective
chemical data: Dr. Techn. Thesuis. NTH Univ. of Trondheim, 1986.
4. Krzeminski A. Why correct for fetal hemoglobin in blood oximetry measurements?
Radiometer Publication Info. No. 1992-3. Copenhagen: Radiometer Medical A/S,
1992.
5. Huehns ER, Beanen GH. Developmental changes in human hemoglobins. Clin Dev Med
1971; 37: 175-203.
5-50
6. User-defined corrections
General information
Purpose of use User-defined corrections are most commonly implemented in situations where
the values measured for a particular parameter by two or more analyzers
deviate consistently from each other.
NOTICE: Since the performance of all ABL90 FLEX analyzers is tested as
described in chapter 7: Performance characteristics in this manual, and each
instrument is assumed to operate accurately and optimally, the unnecessary
correction of parameter values by the user can lead to inaccurate measurements
being reported.
Slope = 1
Offset
Measured
(uncorrected)
parameter value
0.0 (x axis)
Preparatory Prior to entering corrections for any parameter, the user must obtain the
action reference values for the chosen parameters, using the method accepted in
his/her laboratory.
6-2
ABL90 FLEX reference manual 6. User-defined corrections
Entering user- The slope/offset for each parameter are configured in General Setup >
defined Parameters and Input > Parameters. User-corrected values are marked with
corrections a "*" after the result.
NOTICE: The user-defined corrections will be applied to Quality Control
measurements unless the "Apply parameter corrections to QC" option was
deactivated in Miscellaneous setup as described in the section Miscellaneous
setup in chapter 1.
For detailed instructions on how to enter user-defined corrections, refer to the
section Parameters and input setup in chapter 1.
6-3
6. User-defined corrections ABL90 FLEX reference manual
6-4
ABL90 FLEX reference manual 6. User-defined corrections
6-5
6. User-defined corrections ABL90 FLEX reference manual
FO2Hb and The units for FO2Hb and FHHb are Fraction.
FHHb
After the user-defined corrections of the parameters sO2, FCOHb and FMetHb
have been carried out, FO2Hb and FHHb are automatically calculated using the
formulae stated below, since the sum of the fractions FCOHb, FMetHb, FO2Hb
and FHHb as defined must be equal to 1.0:
FO2Hb:
FO2Hb = (1 FCOHb FMetHb) sO2
FHHb:
FHHb = (1 FCOHb FMetHb) (1 sO2)
6-6
ABL90 FLEX reference manual 6. User-defined corrections
6-7
6. User-defined corrections ABL90 FLEX reference manual
Resetting The Radiometer default values for the electrolyte and metabolite parameters
corrections to must be reset manually by the user on the Parameter setup screen to 1.000
default values for each parameter.
6-8
7. Performance characteristics
General information
Performance specifications are achieved by comparison between the ABL90 FLEX
analyzer and the primary reference methods, and by comparison between the ABL90
FLEX analyzer and the ABL735 analyzer.
Performance specifications of the ABL90 FLEX analyzers are described, using the
following:
BiasPrim.ref = the mean difference between the ABL90 FLEX analyzer and the primary
reference methods.
BiasSec.ref = the mean difference between the ABL90 FLEX analyzer and the ABL735
analyzer.
Repeatability (imprecision estimate)
Reproducibility (imprecision estimate)
Total variation range.
7-2
ABL90 FLEX reference manual 7. Performance characteristics
Definition of terms
Bias The bias of a quantity is defined as the mean difference between the measured value
on a group of test instruments and the estimated true value (as assayed by the
reference method or certified standard reference material). BiasPrim.Ref is determined as
follows:
BiasSec.Ref is a relative bias between the ABL90 FLEX analyzer and the ABL735 analyzer
in macromode (C195 µL mode), and is determined as follows:
Secondary
Reference
Parameter Primary Reference Method Reference Reference
methods
Method
pH Capillary-type glass pH electrode ABL735 [Ref. 1,2]
with a saturated calomel reference
electrode and a liquid junction
saturated with KCl (BMS Mk2).
The calibration standards are
traceable to the Primary Reference
Standards for pH.
pCO2 Tonometry. N/A [Ref. 3]
The gases used for tonometry are
traceable to NIST-certified Standard
Reference Materials.
pO2 Tonometry. N/A [Ref. 3]
The gases used for tonometry are
traceable to NIST-certified Standard
Reference Materials.
cCa2+ Calcium transfer standards were ABL735 The
used. These are traceable to NIST standards
SRM 915 and SRM 956b and have were
an ionic strength of 160.0 mmol per produced as
kg of water and pH 7.40 at 37 °C, indicated in
using 1 mmol/L (37 °C) HEPES [Ref. 4]
buffer.
cCl- NIST-certified Standard Reference ABL735
Material SRM 909b (human serum)
and SRM 956b.
cK+ NIST-certified Standard Reference ABL735
Material SRM 909b (human serum)
and SRM 956b.
cNa+ NIST-certified Standard Reference ABL735
Material SRM 909b (human serum),
NIST 956b and Radiometer-specified
standard serum material (specified
using flame photometry).
7-3
7. Performance characteristics ABL90 FLEX reference manual
Secondary
Parameter Primary Reference Method Reference Reference
Method
cGlu Spectrophotometry, using the N/A [Ref. 5]
hexokinase (HK) method
recommended by NCCLS, measured
on serum.
cLac Spectrophotometry using a lactate N/A [Ref. 8]
dehydrogenase (LDH) method,
measured on serum.
ctHb HiCN method recommended by ABL735 [Ref. 6]
NCCLS.
sO2 Tonometry: ABL735
100%: whole blood is tonometered
with a gas mixture containing
94.4% O2 and 5.6% CO2. 0%:
whole blood is tonometered with a
gas mixture containing 94.4% N2
and 5.6% CO2 + dithionite.
FO2Hb Measured according to the following ABL735
relation:
FO2Hb = 1 - (FHHb + FCOHb +
FMetHb)
FHHb 0%: whole blood is tonometered ABL735
with a gas mixture containing
94.4% N2 and 5.6% CO2 +
dithionite.
FCOHb Gas chromatography: The ABL735
Standards are carbon monoxide
mixtures with atmospheric air,
whose purity is validated in
accordance with NIST SRM 1678 (50
ppm CO in N2)
FMetHb Spectrometry, modified Evelyn- ABL735 [Ref. 7]
Malloy method.
ctBil The reference method for total ABL735
bilirubin is a spectrophotometric
method (wet chemistry based on a
method from Bayer Healthcare,
Tarrytown USA).
The method is traceable to NIST
SRM916a Bilirubin.
FHbF The reference method is based on ABL735 [Ref. 15]
Cation Exchange HPLC.
General reference: [Ref. 10].
7-4
ABL90 FLEX reference manual 7. Performance characteristics
Coefficient of The coefficient of variation is reported as a percentage and calculated from the mean
variation (or measuring level) and standard deviation as follows:
(CV%)
Standard deviation
CV% = 100
Measuring level
Confidence Confidence interval provides a range of values estimated from a study group that is
intervals highly likely to include the true, but unknown, value ("confidence interval" applies to
the results of a statistical analysis). A 95% confidence interval means that there is only
a 5% chance that the true value is not included in the interval.
Repeatability/ Repeated measurements using one analyzer on samples assumed to be identical will
Reproducibility not necessarily yield identical results. The degree of variation in the results is a
measure of the precision of the analyzer.
The table on the next page describes the parameters used to characterize precision
obtained via the performance tests on the ABL90 FLEX analyzer. [Ref. 9]
Parameter Description
S0 Repeatability
This is a standard deviation obtained from repeated
measurements within a short interval of time using:
The same instrument and location
The same measurement procedure
Identical portions of the same sample
One operator per instrument
S0 for each level is pooled for all test instruments and test
days.
Sx Reproducibility is obtained from repeated measurements over
several days using:
Random instrument
Random sample
Random operators
Reproducibility for each level is pooled for all test instruments
and test days.
Total analytical TEA, total analytical error is a quality specification that sets a limit for both the random
error error (imprecision) and systematic error (inaccuracy) in a single measurement or single
test result. In Radiometer reference manual the following expression for total analytical
error is either expressed in an absolute number
TEA = (|Bias|+1.96×Sx)
or in %
TEA = (|Bias%|+1.96×CVx) %
The formula we are using for total error allowable works at 95% probability to allow for
5% error.
7-5
7. Performance characteristics ABL90 FLEX reference manual
Test conditions
Test conditions to determine biasPrim.ref, biasSec.ref, repeatability, reproducibility and total
variation for pH, pCO2, pO2, cCa2+, cCl–, cK+, cNa+, cGlu, cLac, ctHb, sO2, FO2Hb,
FCOHb, FMetHb, FHHb, ctBil and FHbF were as follows:
Item Description
Reference analyzers Five ABL735 analyzers with AutoCheck module were
used as a reference. The capillary mode was used
for pCO2 and pO2, and the syringe mode for all the
other parameters.
Primary/secondary As specified for each parameter earlier in this
reference methods chapter.
Analyzers and test 8-10 ABL90 FLEX analyzers were tested in syringe
modes and capillary mode.
Blood samples Heparinized blood samples from healthy, voluntary
donors.
The blood is prepared to obtain different
concentration levels of each measured parameter.
Blood measurements Measurements on every parameter are performed
on all analyzers, with 5 measurements on every
sample of each run, repeated for 3 days.
The measurements were performed by different
operators.
Solution pack All calibration solutions and gases used for the tests
are traceable to Primary Reference Standards.
Traceability certificates for the ABL90 FLEX
calibration solutions and gases are found at the end
of chapter 9: Solutions.
Experimental Ambient temperature: 22-25 °C
conditions
Relative humidity: 30-50 %.
Barometric pressure: 730 - 780 mmHg
7-6
ABL90 FLEX reference manual 7. Performance characteristics
Mode Volume
Syringe 65 µL
Capillary 65 µL
Number of The number of measurements during the test are listed below:
measurements
7-7
7. Performance characteristics ABL90 FLEX reference manual
Reference Capillary-type glass pH electrode with a saturated calomel reference electrode and a
method liquid junction saturated with KCl (BMS Mk2) [Ref. 1,2].
The calibration standards are traceable to the Primary Reference Standards for pH.
BiasPrim.ref pH BiasPrim.ref N
7.0 0.008 45
7.4 -0.004 45
7.6 0.004 45
N = number of measurements employed
7-8
ABL90 FLEX reference manual 7. Performance characteristics
7-9
7. Performance characteristics ABL90 FLEX reference manual
Reference NIST-certified Standard Reference Material SRM 909b (human serum) and Radiometer-
method specified standard serum material (specified using flame photometry).
BiasSec.ref
and
cCl– BiasSec.ref S0 SX CVX% TEA TEA
Repeatability
– blood (mmol/L) (%)
samples 80 0.8 0.1 0.9 1.1 2.56 3.2
100 0.8 0.1 1.1 1.1 2.96 3.0
120 1.1 0.1 1.5 1.3 4.04 3.4
140 0.8 0.1 1.6 1.1 3.94 2.8
150 0.7 0.1 1.7 1.1 4.03 2.7
7-10
ABL90 FLEX reference manual 7. Performance characteristics
Reference The calcium transfer standards were used. These are traceable to NIST SRM915 and
method have an ionic strength of 160.0 mmol per kg of water and pH 7.40 at 37 °C, using 1
mmol/L (37 °C) HEPES buffer.
The standards were produced as indicated in [4].
Bias ABL90 FLEX - Prim.ref = Bias ABL90 FLEX - ABL735 + Bias ABL735 – Prim.ref
7-11
7. Performance characteristics ABL90 FLEX reference manual
7-12
ABL90 FLEX reference manual 7. Performance characteristics
Setup:
Adult samples. HbF correction is not activated.
7-13
7. Performance characteristics ABL90 FLEX reference manual
Reference Tonometry:
method
100%: whole blood is tonometered with a gas mixture containing 94.4% O2 and 5.6%
CO2.
0%: whole blood is tonometered with a gas mixture containing 94.4% N2 and 5.6% CO2
+ dithionite.
7-14
ABL90 FLEX reference manual 7. Performance characteristics
Setup:
Adult samples. HbF correction is not activated.
Reference Gas chromatography: The Standards are carbon monoxide mixtures with atmospheric
method air, whose purity is validated in accordance with NIST SRM 1678 (50 ppm CO in N2).
Setup:
Adult samples. HbF correction is not activated.
7-15
7. Performance characteristics ABL90 FLEX reference manual
Setup:
Adult samples. HbF correction is not activated.
BiasSec.ref and
Repeatability –
blood samples ctHb FMetH BiasSec.ref S0 SX CVX% TEA TEA
(g/dL) b (%) (%)
15 0.0 -0.04 0.10 0.23 - 0.49 -
15 5.0 0.02 0.09 0.25 5.0 0.51 10.2
15 10.0 -0.04 0.12 0.33 3.3 0.69 6.9
15 20.0 -0.18 0.08 0.27 1.4 0.71 3.5
15 30.0 -0.26 0.08 0.34 1.1 0.93 3.1
15 50.0 -0.21 0.09 0.43 0.9 1.05 2.1
15 99.0 0.11 0.05 0.61 0.6 1.31 1.3
7-16
ABL90 FLEX reference manual 7. Performance characteristics
Reference 0%: whole blood is tonometered with a gas mixture containing 94.4% N2 and 5.6% CO2
method + dithionite.
Setup:
Adult samples. HbF correction is not activated.
Reference The reference method is based on Cation Exchange HPLC. The method is described in
method [Ref. 15]. The method is performed by the Hæmatology Laboratory at Herlev Hospital,
Denmark.
Setup:
Mixed adult and Fetal samples. HbF correction enabled for all levels.
7-17
7. Performance characteristics ABL90 FLEX reference manual
Reference The reference method for total bilirubin is a spectrofotometric method (wet chemistry
method based on a method from Bayer Healthcare, Tarrytown USA.
The method is calibrated using NIST SRM916a Bilirubin.
The method is performed by the Laboratory Unilabs AS., Denmark.
Setup:
HbF correction is not activated
NOTES:
a. Adult/fetal blood, pH = 7.4 0.1, normal MCHC and albumin variation. Spiked
with unconjugated bilirubin.
External test The purpose of the bilirubin external tests was to make a regression study of ABL90
results bilirubin against reference hospital analyzers on hospital neonatal blood samples.
A limited study was performed on hospital adult samples [Ref. 13].
For neonatal use: The bilirubin method has been evaluated on whole blood.
The allowed analytical error is 10% to satisfy average
clinical requirements for bilirubin measurement
[16,17,18,19,20]. For whole blood the analytical error is
slightly higher.
7-18
ABL90 FLEX reference manual 7. Performance characteristics
Wet Chemistry analyzer Roche Modular with Roche Calibrator was used as a reference
[Ref. 21]. Each external test site had two Modulars – a total of four. ctBil was measured
in mol/L.
400
y = 1.014x - 0.828
R2 = 0.985
350
300
N=175
250
Syx=11.6
ABL90
200
150
100
50
0
0 100 200 300 400
Modular
7-19
7. Performance characteristics ABL90 FLEX reference manual
50
40
30
20
10
Difference
0
0 100 200 300 400
-10
-20
-30
-40
-50
ctBil
Lines indicate “Mean” and +/- “15 umol or 10%”. Values are in µmol/L. Difference =
ABL90 – Modular.
Interference tests
The results from the interference tests are given as the deviation from the correct result
[Ref. 14].
pH/blood gas The following interference results are found for the pH and blood gas electrodes:
Interference on ...
Test pH pCO2 pO2
Substance Test matrix
Concentration mmHg
(level 7.4)
7-20
ABL90 FLEX reference manual 7. Performance characteristics
Electrolytes The following interference results are found for the electrolyte electrodes:
Interference on …....
+
cK cNa+ cCa2+ cCl-
(4 mM
Test con- (140 mM (1.25 mM (105 mM Test
level)
Substance centration level) level) level) matrix
Lithium (Li+) 3.2 mM <0.1 <1 <0.02 N/A Plasma
+
Potassium (K ) 12 mM N/A <1 <0.02 N/A Plasma
2+
Calcium (Ca ) 3.4 mM <0.1 1.2 N/A N/A Plasma
pH Aqueous/
-0.037
N/A N/A mM/pH N/A buffer
6.8 – 8 N/A N/A N/A <1 Plasma
-
Bromide (Br ) 37.5 mM N/A N/A N/A 76.6 Plasma
A "-" indicates that interference has not been measured on the respective parameter.
7-21
7. Performance characteristics ABL90 FLEX reference manual
Interference on …....
+
cK cNa+ cCa2+ cCl-
(4 mM
Test con- (140 mM (1.25 mM (105 mM Test
level)
Substance centration level) level) level) matrix
Acetylsalicylic acid 0.91 mM N/A N/A N/A <1 Plasma
7-22
ABL90 FLEX reference manual 7. Performance characteristics
Metabolites The following interference results are found for the metabolite electrodes:
Interference on
.......
Test Test
Substance Concentration Glu mM Lac mM matrix
Acetaminophen=paracetamol 2 mM < 0.1 < 0.1 Whole blood
Acetoacetate (Lithium
acetacetoacetate) 2 mM < 0.1 0.11 Whole blood
Citrate (sodium citrate 2H2O) 1 mM < 0.1 < 0.1 Whole blood
7-23
7. Performance characteristics ABL90 FLEX reference manual
Interference on
.......
Test Test
Substance Concentration Glu mM Lac mM matrix
2-deoxy Glucose 2.5 mM 2.25 N/A Whole blood
Unconjugated Bilirubin 0.2 g/L < 0.1 < 0.1 Whole blood
Conjugated Bilirubin 0.2 g/L < 0.1 < 0.1 Whole blood
Ascorbic acid 170 µmol/L < 0.1 < 0.1 Whole blood
7-24
ABL90 FLEX reference manual 7. Performance characteristics
Oximetry The substances against which the oximetry parameters (ctHb, sO2, FO2Hb, FCOHb,
parameters FMetHb, FHHb, FHbF) and ctBil were tested for interference are given in the table
below:
(SAT100 blood reference test sample: ctHb = 15 g/dL, sO2 = 100 %, FCOHb = 0.7%,
FMetHb = 0.5%, ctBil = 0, pH = 7.4. Parameter sensitivity from the influence on the
absorbance spectrum from various substances.)
7.4 Reference pH
Fluorescein 250 mg/L 1.34 -3.2 -9.6 -4.1 10.7 2.9 ND -1115
beta-carotene*) 3.7 µmol/L <|0.5| <|1%| <|1%| <|1%| <|1%| <|1%| <|20%| <|30|
Patent Blue V 10 mg/L <|0.5| <|1%| 2.2 <|1%| <|1%| <|1%| ND <|30|
Cardio Green 7 mg/L <|0.5| <|1%| <|1%| <|1%| <|1%| <|1%| <|20%| <|30|
Evans Blue 5 mg/L <|0.5| <|1%| <|1%| <|1%| <|1%| <|1%| <|20%| <|30|
2% (200
mg/dL) <|0.5| <|1%| <|1%| <|1%| <|1%| <|1%| <|20%| <|30|
Intralipid 5% (1000
mg/dL) <|0.5| <|1%| <|1%| <|1%| <|1%| <|1%| <|20%| <|30|
Bilirubin (unconj) 500 µmol/L <|0.5| <|1%| <|1%| <|1%| <|1%| <|1%| <|20%| 524
Bilirubin (conj) 400 µmol/L <|0.5| <|1%| <|1%| <|1%| <|1%| <|1%| <|20%| 377
0.4 g/L 0.53 <|1%| -4.0 <|1%| 3.4 <|1%| <|20%| <|30|
0.2 g/L <|0.5| <|1%| -2.4 <|1%| 1.8 <|1%| <|20%| <|30|
7-25
7. Performance characteristics ABL90 FLEX reference manual
0.4 g/L 0.71 -1.2 -5.2 <|1%| 3.2 1.2 <|20%| <|30|
0.2 g/L <|0.5| <|1%| -2.2 <|1%| 1.4 <|1%| <|20%| <|30|
∆FHbF =
− 0.51% /(nmol/L) × (cH + − 40 nmol/L)
pH ∆FHbF %
7.15 -15.8
7.25 -8.2
7.4 0
7.5 4.1
7.6 7.7
ctBil sensitivity MCHC (Mean Corpuscular Hemoglobin Concentration) is used to estimate hematocrit,
for MCHC Hct, which is used in the ctBil measurement. MCHC is an average Hb concentration in
variations the red blood cell (RBC). If the RBC volume decreases, MCHC increases. If an RBC has
iron deficit, MCHC decreases.
Hct is determined from ctHb as follows:
ctHb
Hct =
MCHC
7-26
ABL90 FLEX reference manual 7. Performance characteristics
If MCHC is defined as MCHC = 332 g/L MCHC, then the contribution to the relative
error on the ctBil measurement is as follows:
ctBil Hct MCHC
ctBil 1 Hct MCHC
If the reference value for Hct is known, it is possible to correct the displayed ctBil value,
using the following equation:
1 ctHb(displayed) 0.0301
ctBil(corrected) ctBil(displayed)
1 Hct(reference)
7-27
7. Performance characteristics ABL90 FLEX reference manual
Anticoagulants Anticoagulants containing sodium salts give erroneously high cNa+ results. Sodium
(sampling) fluoride with or without EDTA and oxalate (disodium) affects cGlu results. Sodium
fluoride gives erroneously high cNa+ and low cCa2+, cGlu and cLac results. Trisodium
citrate affects cNa+, cK+ and cGlu results.
Radiometer, therefore, recommends the use of heparin as the only anticoagulant.
7-28
ABL90 FLEX reference manual 7. Performance characteristics
List of
references 1. Kristensen HB, Salomon A, Kokholm G. International pH scales and certification of pH.
2. Definition of pH scales, standard reference values, measurement of pH and related
terminology (Recommendations 1994). Pure and Appl Chem 1985; 57, 3: 531-42.
3. Burnett RW, Covington AK, Maas AHJ, Müller-Plathe O et al. J Clin Chem Clin Biochem 1989;
27: 403-08.
4. IFCC reference methods and materials for measurement pH, gases and electrolytes in blood.
Scand J Clin Lab Invest 1993; 53, Suppl 214: 84-94.
5. Glucose. CLSI/NCCLS Publication RS1-A. Clinical and Laboratory Standards Institute, 940 West
Valley Road, Suite 1400, Wayne, PA 19087, 1989.
6. Reference and selected procedures for the quantitative determination of hemoglobin in blood.
Approved Standard (3rd edition), CLSI/NCCLS Publication H15-2A. Clinical and Laboratory
Standards Institute, 940 West Valley Road, Suite 1400, Wayne, PA 19087, 2000.
7. Evelyn K, Malloy H. Microdetermination of oxyhemoglobin, methemoglobin and sulfhemoglobin
in a single sample of blood. Biological Chem 1938; 126: 655-62.
8. Begmeyer. Methods of enzymatic analysis. 3rd ed., Verlag Chemie Deerfield Beach 1984; 6:
582-88.
9. VIM93: ISO, International Vocabulary of Basic and General Terms in Metrology, Geneva:
International Organization for Standardization; 1993.
10. Kristensen H.B. Traceability to the primary reference standards at Radiometer. Copenhagen:
Radiometer Medical ApS, 2004. Code 918-541.
11. CLSI Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guidelines,
EP5-A, Vol. 19, No. 2.
12. CLSI Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved
Guidelines, EP17-A, Vol. 24, No. 34.
13. CLSI Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline –
Second Edition, EP9-A2, Vol. 22, No. 17.
14. CLSI approved guideline for interference testing in clinical chemistry, EP7-A, Vol. 22, No. 27.
15. Tan GB, Aw TC, Dunstan RA & Lee SH, Evaluation of high performance liquid chromatography
for routine estimation of haemoglobins A2 and F. Journal of Clinical Pathology 46: 852-856.”
16. Fraser CG. The application of theoretical goals based on biological variation data in proficiency
testing. Arch Pathol Lab Med 1988; 112: 402-15.
17. Ehrmeyer SS, Laessig RH, Leinweber JE, Oryall JJ. 1990 Medicare/CLIA final rules for
proficiency testing: minimum intralaboratory performance characteristics (CV and bias)
needed to pass. Clin Chem 1990; 36, 10: 1736-40.
18. Fraser CG, Petersen PH, Ricos C, Haeckel R. Proposed quality specifications for the imprecision
and inaccuracy of analytical systems for clinical chemistry. Eur J CLin Chem Clin Biochem
1992; 30: 311-17.
19. Westgard JO, Seehafer JJ, Barry PL. Allowable imprecision for laboratory test based on clinical
and analytical test outcome criteria. Clin Chem 1994; 40, 10: 1909-14.
20. Vanderline RE, Goodwine J, Koch D, Scheer D, Steindel S, Cembrowski G. Guidelines for
providing quality stat laboratory services. 1987 Laboratory Quality Assurance Commitee.
21. Wahlefeld AW, Herz G, Bernt E. Modification of the Malloy-Evelyn method for a simple, reliable
determination of total bilirubin in serum. Scand J Clin Lab Invest 1972;29 Supplement 126:
Abstract 11:12.
7-29
7. Performance characteristics ABL90 FLEX reference manual
7-30
8. Parameters
General information
The Deep The Deep Picture developed by Radiometer [1] (visit our website www.deep-
Picture picture.com) expands traditional pH and blood gas analysis by evaluating the
capability of arterial blood to carry sufficient oxygen to tissues and to release it.
It simplifies interpretation by dividing the process into three stages:
Stage Description
Oxygen Oxygen uptake in the lungs indicates whether the pulmonary gas
uptake exchange is efficient enough to oxygenate arterial blood.
The uptake of oxygen in the lungs can be described by
parameters in combination, primarily the arterial oxygen tension
(pO2(a)), fraction of O2 in dry inspired air (FO2(I)) and shunt
· ·
fraction of perfused blood (Qs /Q t)
However, other parameters may also be used, such as the
difference in alveolar air and arterial blood oxygen tension
(pO2(A-a)).
Oxygen Oxygen transport reveals if arterial blood contains sufficient
transport oxygen.
The oxygen concentration of arterial blood (ctO2(a)) also termed
oxygen content is determined by the concentration of total
hemoglobin (ctHb(a)), the fraction of oxygenated hemoglobin
(FO2Hb(a)) and the arterial oxygen tension (pO2(a)).
Other parameters that should be known are the oxygen
saturation (sO2 (a)) and the fractions of dyshemoglobins
(FCOHb(a) and FMetHb(a)).
Oxygen Oxygen release describes the ability of arterial blood to release
release oxygen to the tissues.
The release of oxygen from capillaries to tissues is determined by
the oxygen tension gradient between the two. This release of
oxygen is also influenced by the hemoglobin-oxygen affinity,
which is indicated by the oxygen tension at 50 % saturation,
p50.
8-2
ABL90 FLEX reference manual 8. Parameters
Symbols The symbols for the parameters are based on the principles described by
Wandrup [2]. Each symbol consists of three parts, described below:
component
Example:
pO2(a)
property system
The parameters are listed by symbol in three groups: measured, input and
derived.
8-3
8. Parameters ABL90 FLEX reference manual
Critical limits are user-defined and can be entered into the analyzer software –
see section Reference ranges and critical limits in chapter 1.
If one or more default values have been used in the calculation, the result may
deviate significantly from the true value. The deviation on "estimated" oxygen
status parameters might become particularly significant if default values are
used instead of measured blood oximetry data.
In some cases, however, the default value is not accepted as the input for the
calculation. This is because the actual values of the missing parameter may
deviate significantly from the default value, thus making the estimation clinically
inappropriate. If sO2 cannot be measured due to severe errors, it will be
calculated.
8-4
ABL90 FLEX reference manual 8. Parameters
Measured Some of the listed parameters are measured, depending on the analyzer
parameters configuration. In these cases the equation given only applies if that parameter is
not directly measured by the analyzer.
Sample type Unless otherwise stated, a parameter will be calculated or estimated irrespective
of the choice on the Patient Identification screen: "Arterial", "Capillary",
"Venous", "Mixed venous" or "Not specified". Some parameters, however, are
defined for arterial samples only; they will be calculated only for sample types
entered as "Arterial" or "Capillary".
The symbol for system (blood (B) or plasma (P)) is not stated in the equations
unless it is important for the calculation.
Units The units given for each parameter refer to the units available on the analyzer for
that parameter.
Default values The default values are listed in Default values at the end of this chapter.
8-5
8. Parameters ABL90 FLEX reference manual
Measured parameters
General The following is the used:
information
m= male
f= female
Reference For adult arterial blood
range
Reference [10]
(unless otherwise specified)
8-6
ABL90 FLEX reference manual 8. Parameters
8-7
8. Parameters ABL90 FLEX reference manual
8-8
ABL90 FLEX reference manual 8. Parameters
8-9
8. Parameters ABL90 FLEX reference manual
8-10
ABL90 FLEX reference manual 8. Parameters
Age ctBil
24 hrs, premature 103-205 µmol/L
1.0-8.0 mg/dL
10-80 mg/L
24 hrs, full-term 34-103 µmol/L
2.0-6.0 mg/dL
20-60 mg/L
48 hrs, premature 103-205 µmol/L
6-12 mg/dL
60-120 mg/L
48 hrs 103-171 µmol/L
6-10 mg/dL
60-100 mg/L
3-5 days, premature 171-239 µmol/L
10-14 mg/dL
100-140 mg/L
3-5 days, full-term 68-137 µmol/L
4-8 mg/dL
40-80 mg/L
>1 month 3.4-17 µmol/L
0.2-1.0 mg/dL
2-10 mg/L
The following table shows the possible ranges and precision (number of
decimals) of the measured parameters.
These ranges can be narrowed by calculation ranges, reportable ranges, range
of indication, etc., but should be taken into consideration when external systems
are interfaced to the analyzer.
8-11
8. Parameters ABL90 FLEX reference manual
8-12
ABL90 FLEX reference manual 8. Parameters
Input parameters
Definition Input parameters are the parameters keyed in by the operator on the Patient
Identification screen or transferred from an interfaced database.
All input parameters are given in this section.
ctHb Is used if the analyzer version does not include the oximetry measuring system.
Definition Concentration of total hemoglobin in blood.
Input range /Unit g/dL: 0.0-33.0
g/L: 0-330
mmol/L: 0.0-20.5
Conversion ctHb (g/dL) = 1.61140 ctHb (mmol/L);
ctHb (g/L) = 16.1140 ctHb (mmol/L);
ctHb (mmol/L) = 0.62058 ctHb (g/dL) =
0.062058 ctHb (g/L)
8-13
8. Parameters ABL90 FLEX reference manual
8-14
ABL90 FLEX reference manual 8. Parameters
Derived parameters
General In the Type column the following symbols are used:
information
ms for measured parameters
dv for derived parameters
in for input parameters
8-15
8. Parameters ABL90 FLEX reference manual
8-16
ABL90 FLEX reference manual 8. Parameters
8-17
8. Parameters ABL90 FLEX reference manual
8-18
ABL90 FLEX reference manual 8. Parameters
Electrolyte The table below lists the electrolyte derived parameters for the analyzers.
parameters
8-19
8. Parameters ABL90 FLEX reference manual
Possible The following table shows the possible ranges and precision (number of
ranges and decimals) of the measured parameters. These ranges can be narrowed by
precision calculation ranges, reportable ranges, range of indication, etc., but should be
(number of taken into consideration when external systems are interfaced to the analyzer.
decimals)
Symbol Unit Numerical format within the following ranges:
Range Range
pH(T) - 4.000 11.000
+
cH (T) nmol/L -999999.0 199.9 200 9999999
pCO2(T) mmHg 0.0 99.9 100 750
kPa 0.00 9.99 10.0 100.0
–
cHCO3 (P) mmol/L 0.0 100.0
cBase(B) mmol/L -50.0 50.0
cBase(B,ox) mmol/L -100.0 100.0
cBase(Ecf) mmol/L -50.0 50.0
cBase(Ecf,ox) mmol/L -100.0 100.0
–
cHCO3 (P,st) mmol/L 0.0 150.0
ctCO2(P) mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
ctCO2(B) mmol/L 0.0 100.0
Vol % 0.0 224.1
mL/dL 0.0 224.1
pH(st) - 4.000 11.000
VCO2/V(dry % -10.0 110.0
air)
fraction -0.100 1.100
Hct % -10.0 110.0
fraction -0.100 1.100
pO2(T) mmHg 0.0 99.9 100 750
kPa 0.00 9.99 10.0 100.0
pO2(A) mmHg 0.0 750.1
kPa 0.00 100.00
pO2(A,T) mmHg 0.0 750.1
kPa 0.00 100.00
p50 mmHg 0.00 750.06
kPa 0.00 100.00
p50(T) mmHg 0.00 750.06
kPa 0.00 100.00
p50(st) mmHg 0.00 750.06
kPa 0.00 100.00
8-20
ABL90 FLEX reference manual 8. Parameters
·
V O2 mL/min 0 22414
mmol/min 0.0 1000.0
8-21
8. Parameters ABL90 FLEX reference manual
8-22
ABL90 FLEX reference manual 8. Parameters
List of equations
Units and All definitions and equations are based on SI units. If "T" for patient
symbols temperature is not stated, the calculation is based on a temperature of 37.0 °C.
The following SI units are used:
concentration in mmol/L
temperature in °C
pressure in kPa
fractions (not %)
cH+(T) Eq. 2:
cH+(T )=10
9 pH (T )
where
pK p 6.095
8-23
8. Parameters ABL90 FLEX reference manual
If ctHb is not measured or keyed in, the default value will be used.
If sO2 is not measured, it will be calculated from equation 39.
cBase(Ecf,ox) Eq. 8:
cBase(Ecf,ox) = cBase(Ecf) – 0.3062 × 3 ×(1 – sO2)
where
Eq. Description
9.1 a'= 4.04 10-3 4.25 104 ctHb
8-24
ABL90 FLEX reference manual 8. Parameters
pH pK
ctCO2(B) 9.286 103 pCO2 ctHb 1+10 Ery Ery
ctHb
ctCO2 (P) 1
21.0
where
Eq. Description
11.1 pHEry 7.19 0.77 pH- 7.40 0.035 1 sO2
Eq. Description
5.3 5.33 pH(Hb) pH
pH(st) pH log
pCO 2 log p CO2 (Hb) log(7.5006p CO )
2
8-25
8. Parameters ABL90 FLEX reference manual
where
Eq. Description See…
14.1 S = ODC(P,A,T) Eq. 47
14.2 S 1- FMetHb FCOHb Eq. 46.12
sO2,i(T )=
1-FCOHb-FMetHb
14.3 P Eq. 46.10
pO2,i (T )=
FCOHb
1
sO2,i(T ) 1 FCOHb FMetHb
Changes in the equations for pH(T) and ctO2 correspond to less than 0.5% of
pO2(T) in the reference range for pH, pCO2, pO2 and ctHb and T in the interval
32-42 °C, using FHbF = 0.5%.
pCO2 RQ1 FO2(I) RQ1 1
If FO2(I) and RQ are not keyed in, they are set to the default values.
The calculation requires entering the sample type as "Arterial" or "Capillary".
If FO2(I) and RQ are not keyed in, they are set to the default values.
The calculation requires entering the sample type as "Arterial" or "Capillary".
The calculation cannot be performed on the basis of the default FO2(I) value,
and the calculation requires entering the sample as "Arterial" or "Capillary".
8-26
ABL90 FLEX reference manual 8. Parameters
The calculation cannot be performed on the basis of the default FO2(I) value,
and the calculation requires entering the sample as "Arterial" or "Capillary".
where
Description See...
P = ODC(S,A,T) Eq. 47
A=a
T = 37.0 °C Eq. 46.13
where
Description See…
P = ODC(S,A,T) Eq. 47
pH
A a 1.04 T 37.0
(T )
pH
1.47 102 6.5 103 pH(37) 7.40
(T )
8-27
8. Parameters ABL90 FLEX reference manual
Description See…
S = 0.5 Eq. 46.11
A = a6 corresponds to pH = 7.40, pCO2 = 5.33 kPa, FCOHb = 0, Eq. 46.13
FMetHb = 0, FHbF = 0
T = 37.0 °C
8-28
ABL90 FLEX reference manual 8. Parameters
26.5 A=a
26.6 T = 37 °C
When ti = ctO2 – 2.3 mmol/L, then pO2,i = pO2(x), where ctO2 is determined as
described in equation 27.
pO2(x) cannot be calculated on the basis of a default ctHb value.
pO2(x) can only be calculated if the measured sO2(a) 0.97.
The calculation requires entering the sample type as "Arterial" or "Capillary".
– Eq. 28:
ctO2(av )
– –
ctO2(a v ) = ctO2(a) – ctO2(v )
where ctO2(a) and
–
ctO2(v ) are calculated from equation 27 for arterial and mixed
venous blood, respectively. The calculation requires two measurements and input
– –
of both pO2(v ) and sO2(v ).
Eq. 29 [7]:
8-29
8. Parameters ABL90 FLEX reference manual
where
· Eq. 31:
D O2
. .
D O2 ctO2 Q t
· ·
Q t is the cardiac output and is an input parameter for the calculation of D O2.
· ·
If Q t is not keyed in, D O2 will not be calculated.
The calculation requires entering the sample type as "Arterial" or "Capillary".
· Eq. 32:
Qt
.
. V O2
Qt
ctO2(a-v)
· ·
If V O2 is not keyed in, Q t will not be calculated.
· Eq. 33:
VO2
. .
V O2 Qt ctO2(a-v)
· ·
If Q t is not keyed in, VO2 will not be calculated.
8-30
ABL90 FLEX reference manual 8. Parameters
ctO2(c) ctO2(a)
FShunt=
ctO2(c) ctO2 (v)
and
Eq. Description
34.1 ctO2(A) ctO2(a)
FShunt
ctO2(A) ctO2(v)
34.2
1
ctO2(a) ctO2(v)
FShunt= 1
ctO2(A) ctO2 (a)
where
ctO2(c): total oxygen in pulmonary capillary blood
ctO2(a): total oxygen in arterial blood
ctO2(A): total oxygen in alveolar air. Oxygen tension = pO2(A).
–
ctO2(v ): total oxygen in mixed venous blood
34.3 ctO2 (a) 0.0105pO2 (a) ctHb 1 FCOHb FMetHb sO2 (a)
8-31
8. Parameters ABL90 FLEX reference manual
where
ctO2(a,T): total oxygen in arterial blood at patient temperature
ctO2(A,T): total oxygen in alveolar blood at patient temperature
–
ctO2(v ,T): total oxygen in mixed venous blood at patient temperature
If sO2(a) > 0.97, the default p50(st) (3.578 kPa) will be used to
determine the ODC.
35.11 – –
ctO2(v ,T) ctO2(v ) at 37 oC is calculated from
equation 27 for mixed venous blood values of pO2 and sO2.
If no mixed venous sample is measured, the FShunt(T) is
estimated assuming ctO2(a,T) –
–
ctO2(v ,T) = 2.3 mmol/L in equation 35.
RI Eq. 36:
pO2(A) pO2(a)
RI=
pO2(a)
8-32
ABL90 FLEX reference manual 8. Parameters
Qx Eq. 38 [8]:
The ODC is determined as described in equations 46-47 in the section
Oxyhemoglobin Dissociation Curve further in this chapter.
2.3
Qx
ctO2(a) ti
S 1 FMetHb FCOHb
sO2
1-FCOHb-FMetHb
where
Description See…
S = ODC(P,A,T)
pO2 FCOHb Eq. 46.9
P pO2
sO2 (1 FCOHb FMetHb)
A=a
T = 37.0 °C
8-33
8. Parameters ABL90 FLEX reference manual
Eq. Description
42.1 V (CO)
V (B)
21.84 FCOHb(2) FCOHb(1) ctHb
42.2 V(CO) = volume (in mL) of carbon monoxide injected according to the
procedure and the value keyed in
42.3 FCOHb(1) = fraction of COHb measured before the CO injection
42.4 FCOHb(2) = fraction of COHb measured after the CO injection
Eq. 46-47 See Oxyhemoglobin dissociation curve (ODC), further in this chapter.
8-34
ABL90 FLEX reference manual 8. Parameters
In the calculations the following are assumed: pH = 7.4, pCO2 = 5.33 kPa,
FCOHb = 0, FMetHb = 0, cDPG = 5 mmol/L, and temp = 37 °C.
Eq. Description See…
49.1 An estimate of FHbF is made: FHbFest = 0.8
49.2 pO2,est = ODC (sO2,A,T); Eq. 47
where the constant A depends on FHbF = FHbFest
49.3 sO2 (for fetal blood) = ODC (pO2,est, A,T); Eq.47
where FHbF = 1
49.4 cO2HbFest = sO2 (fetal blood) ceHb FHbFest
49.5 cO2HbFmeas. cO2HbFest
FHbFest
ceHb
50.8 pO2,i=pO2(x,T )
8-35
8. Parameters ABL90 FLEX reference manual
pO2
VO2 /V (dry air)
p(amb) 6.275
8-36
ABL90 FLEX reference manual 8. Parameters
y y o (x x o ) h tanh k o x x o
where ko = 0.5343
Eq. Description
46.1 x ln p
46.2 s
y ln
1- s
46.3 so
yo ln where so = 0.867
1-so
46.4 x0 = x00 + a + b = In(p00) + a + b where poo = 7 kPa.
The actual position of the ODC in the coordinate system (ln(s/(1–s)) vs ln(p))
used in the mathematical model, is expressed by equations 46.3 and 46.4.
The symbols "a" and "b" reflect the ODC displacement from the reference
position to its actual position in this coordinate system:
"a" describes the displacement at 37 C.
"b" the additional displacement due to the patient temperature difference from
37 C.
The ODC The reference position of the ODC was chosen to be the one that corresponds to
reference the default value for p50(st) = 3.578 kPa, which is traditionally considered the
position most likely value of p50 for adult humans under standard conditions, namely:
pH = 7.40
pCO2 = 5.33 kPa
FCOHb, FMetHb, FHbF = 0
cDPG = 5 mmol/L
8-37
8. Parameters ABL90 FLEX reference manual
The ODC The ODC displacement which is described by "a" and "b" in the coordinate
displacement system (ln(s/(1–s)) vs ln(p)), is given by the change in p50 from the default to
its actual value in a more common coordinate system (sO2, pO2).
Eq. Description
46.5 p
x xo ln ab
7
46.6 h ho a where ho = 3.5
46.7 b 0.055 (T T o ) To = 37 °C
pO2 pCO
M , to give eq. 46.9
cO2Hb cCOHb
46.10 p
pO2
FCOHb
1
sO2 1 FCOHb FMetHb
Eq. Description
46.11 cO2Hb cCOHb
s
cO2Hb cCOHb cHHb
or
sO2 1-FCOHb-FMetHb FCOHb
1 FMetHb
46.12 s 1-FMetHb FCOHb
sO2
1 FCOHb FMetHb
8-38
ABL90 FLEX reference manual 8. Parameters
The actual ODC The actual position of the ODC at 37 C for a given sample is, in principle,
position determined in two steps:
1. The calculation of the combined effect on the ODC position at 37 C of all
known causes for displacement (= ac in equation 46.13), and based on
this position.
2. The computation by a numerical method of the actual position of the
ODC curve by shifting it to pass through the known set of coordinates
(P0, S0).
Eq. Description
46.13 a = ac + a6
46.14 ac = a1 + a2 + a3 + a4 + a5
46.15 a1 = 0.88 (pH 7.40)
46.16 pCO2
a2 0.048 ln
5.33
8-39
8. Parameters ABL90 FLEX reference manual
Step Description
II: sO2 > 0.97 (or erroneous) and p50(st) is known.
Coordinates (P0, S0) are calculated from (p50(st), 0.5)
using equations 46.9 and 46.11.
Reference position of the ODC.
NOTICE: The curves are used only to illustrate the principles of the ODC
determination.
8-40
ABL90 FLEX reference manual 8. Parameters
8-41
8. Parameters ABL90 FLEX reference manual
Conversion of units
SI units The equations stated above are based on the SI-unit system. If parameters are
known in other units, they must be converted into a SI unit before entering the
equations. The result will be in an SI unit.
After the calculation the result may be converted to the desired unit. Conversion
of units may be performed, using the equations stated below:
Temperature 9 o
T F = (T C ) 32
5
5 o
T C = (T F 32)
9
Pressure
p 7.500638 × p(kPa)
p (mmHg) = =
(Torr)
ctHb 4
ctHb (g/dL) = 1.61140 ctHb (mmol/L)
ctHb (g/L) = 16.1140 ctHb (mmol/L) or
ctHb (mmol/L) = 0.62058 ctHb (g/dL)
ctHb (mmol/L) = 0.062058 ctHb (g/L)
· · ·
VO2 VO2 mmol/min = VO2/22.41 mL/min
8-42
ABL90 FLEX reference manual 8. Parameters
cGlu 22
cLac 22
8-43
8. Parameters ABL90 FLEX reference manual
Default values
Values The following default values are used in the analyzer, if other values are not
keyed in.
T = 37.0 °C
FO2(I) = 0.21 (21.0 %)
RQ = 0.86
ctHb = 9.3087 mmol/L, (15.00 g/dL or 150 g/L)
FCOHb = 0.004 (0.4 %)
FMetHb = 0.004 (0.4 %)
p50(st) = 3.578 kPa (26.84 mmHg)
In addition to the above default values, the analyzer uses the following default:
Ambient temperature = 25.0 C.
8-44
ABL90 FLEX reference manual 8. Parameters
References
1. The Deep Picture™, critical information from blood gas analysis. Copenhagen:
Radiometer Medical A/S, 1993: 1-14.
2. Wandrup JH. Physicochemical logic and simple symbol terminology of oxygen status.
Blood Gas News 1993; 2,1: 9-11.
3. Siggaard-Andersen O, Durst RA, Maas AHJ. Approved recommendation (1984) on
physicochemical quantities and units in clinical chemistry. J Clin Chem Clin Biochem
1987; 25: 369-91.
4. Siggaard-Andersen O. The acid-base status of the blood. 4th revised ed. Copenhagen:
Munksgaard, 1976.
5. Siggaard-Andersen O, Wimberley PD, Fogh-Andersen N, Gøthgen IH. Measured and
derived quantities with modern pH and blood gas equipment: calculation algorithms with
54 equations. Scand J Clin Lab Invest 1988; 48, Suppl 189: 7-15.
6. Burnett RW, Noonan DC. Calculations and correction factors used in determination of
blood pH and blood gases. Clin Chem 1974; 20,12: 1499-1506.
7. Wimberley PD, Siggaard-Andersen O, Fogh-Andersen N, Zijlstra WG, Severinghaus JW.
Hemoglobin oxygen saturation and related quantities: definitions, symbols and clinical
use. Scand J Clin Lab Invest 1990; 50: 455-59. Available as AS104.
8. Siggaard-Andersen O, Gøthgen IH, Wimberley PD, Fogh-Andersen N. The oxygen status
of the arterial blood revised: relevant oxygen parameters for monitoring the arterial
oxygen availability. Scand J Clin Lab Invest 1990; 50, Suppl 203: 17-28. Available as
AS108.
9. Wandrup JH. Oxygen uptake in the lungs. Blood Gas News 1992; 1,1: 3-5.
10. Tietz NW, Logan NM. Reference ranges. In: Tietz NW, ed. Fundamentals of clinical
chemistry. 3rd ed. Philadelphia: WB Saunders Company, 1987: 944-75.
11. Siggaard-Andersen O, Wimberley PD, Fogh-Andersen N, Gøthgen IH. Arterial oxygen
status determined with routine pH/blood gas equipment and multi-wavelength
hemoximetry: reference values, precision and accuracy. Scand J Clin Lab Invest 1990;
50, Suppl 203: 57-66. Available as AS106.
12. Siggaard-Andersen O, Thode J, Wandrup JH. The concentration of free calcium ions in
the blood plasma ionized calcium. In: Siggaard-Andersen O, ed. Proceedings of the IFCC
expert panel on pH and blood gases held at Herlev Hospital 1980. Copenhagen:
Radiometer Medical A/S, 1981: 163-90. Available as AS79.
13. Severinghaus JW. Blood gas calculator. J Appl Physiol 1966; 21,3: 1108-16. Available as
ST36.
14. Christiansen TF. An algorithm for calculating the concentration of the base excess of
blood. In: Siggaard-Andersen O, ed. Proceedings of the IFCC expert panel on pH and
blood gases held at Herlev Hospital 1980. Copenhagen: Radiometer Medical A/S, 1981:
77-81.
15. Kokholm G. Simultaneous measurements of blood pH, pCO2, pO2 and concentrations of
hemoglobin and its derivatives – a multicenter study. Scand J Clin Lab Invest 1990; 50,
Suppl 203: 75-86. Available as AS107.
16. Siggaard-Andersen O, Wimberley PD, Gøthgen IH, Siggaard-Andersen M. A mathematical
model of the hemoglobin-oxygen dissociation curve of human blood and of the oxygen
partial pressure as a function of temperature. Clin Chem 1984; 30: 1646-51.
17. Siggaard-Andersen O, Wimberley PD, Gøthgen IH, Fogh-Andersen N, Rasmussen JP.
Variability of the temperature coefficients for pH, pCO2 and pO2 in blood. Scand J Clin Lab
Invest 1988; 48, Suppl 189: 85-88.
18. Siggaard-Andersen O, Siggaard-Andersen M. The oxygen status algorithm: a computer
program for calculating and displaying pH and blood gas data. Scand J Clin Lab Invest
1990; 50, Suppl 203: 29-45.
19. Bartels H, Christoforides C, Hedley-Whyte J, Laasberg L. Solubility coefficients of gases.
In: Altman PL, Dittmer DS, eds. Respiration and circulation. Bethesda, Maryland: Fed
Amer Soc Exper Biol, 1971: 16-18.
20. Roughton FJW, Darling RC. The effect of carbon monoxide on the oxyhemoglobin
dissociation curve. Am J Physiol 1944; 141: 17-31.
8-45
8. Parameters ABL90 FLEX reference manual
21. Engquist A.. Fluids electrolytes nutrition. Copenhagen: Munksgaard, 1985: 56-68, 118.
22. Olesen H et al. A proposal for an IUPAC/IFCC recommendation, quantities and units in
clinical laboratory sciences. IUPAC/IFCC Stage 1, Draft 1, 1990: 1-361.
23. Kokholm G, Larsen E, Jensen ST, ChristiansenTF. 3rd ed. Blood gas measurements at
high altitudes. Copenhagen: Radiometer Medical A/S, 1991. Available as AS109.
24. Blood gas and pH analysis and related measurements; approved guideline. NCCLS (CLSI)
document C46-A2, Vol. 29 No. 8, 2009
25. Burton DR. Clinical physiology of acid-base and electrolyte disorders. 4th ed. New York:
McGraw-Hill, 1994
8-46
9. Solutions
General information ......................................................................... 9-2
Solutions ........................................................................................ 9-3
Certificate of traceability ................................................................... 9-5
9. Solutions ABL90 FLEX reference manual
General information
Introduction The ABL90 FLEX analyzer utilizes a solution pack for all calibrations, QC and
rinse procedures, and for the collection of waste fluids.
Lot Each solution pack has a lot number, which identifies the solution packs
assembled in one production lot.
In vitro All the solutions described in this chapter are for in vitro diagnostic use.
diagnostic use
Expiration date The expiration date of the solution pack is found on the solution pack barcode
label. A solution pack can be used on the analyzer for up to 30 days (or until no
more activities are left) but not after the expiration date. This means that if you
install a solution pack 5 days before the expiration date, it can only be used for
5 days.
Storage The solution pack storage temperature range is 2-25 °C. The storage altitude
range is sea level to 4000 meters. The barometer pressure should lie between
450-800 mmHg, or 60.0-106.7 kPa, or 450-800 Torr.
Material safety Material Safety Data Sheets (MSDS) for all solutions in the solution pack are
Data Sheets available from your Radiometer distributor.
9-2
ABL90 FLEX reference manual 9. Solutions
Solutions
Use The solutions contained in the pouches of the solution pack are used for either
calibration or quality control of all analytes. During sample analysis and quality
control measurements CAL 1 also acts as a rinse solution, removing the sample
from the sensor cassette measuring chamber.
Concentration
Substance Unit CAL 1 CAL 2 CAL 3
S1920 S1930 S1940
pH 7.30 6.8 NA
pCO2 mmHg 35 NA 80
pO2 mmHg 180 NA NA
cNa+ mmol/L 150 70 NA
+
cK mmol/L 4 10 NA
cCl– mmol/L 95 50 NA
2+
cCa mmol/L 0.5 2.3 NA
cGlu mmol/L 0 NA 10
(background)
cLac mmol/L 0 NA 10
(background)
ctHb g/dL NA NA 0
NOTICE: The actual analyte concentrations for each solution in a solution pack
lot are included in the smart chip contained in each solution pack. The values
are read into the analyzer when the solution pack is installed in an analyzer.
9-3
9. Solutions ABL90 FLEX reference manual
Concentration
Substance Unit S9030 S9040 S9050
(QC 1) (QC 2) (QC 3)
Solution Solution Gas pO2 Solution
(at 760
mmHg)
pH 7.2 6.8 NA 7.5
pCO2 mmHg 30 67 NA 15
pO2 mmHg 180 300 20
(42.07 %)
cNa+ mmol/L 140 118 NA 175
+
cK mmol/L 4 7 NA 1.8
–
cCl mmol/L 105 95 NA 125
cCa2+ mmol/L 0.8 1.65 NA 0.3
cGlu mmol/L 0 15 NA 7
cLac mmol/L 0 8 NA 4
ctHb mmol/L 0 8 NA 12
sO2 % 97 NA 70
FO2Hb % 92 NA 49
FCOHb % 3 NA 20
FMetHb % 2 NA 10
FHbF % 80 NA 50
ctBil µmol/L 0 300 NA 450
NOTICE: The actual analyte concentrations for each solution in a solution pack
lot are included in the smart chip contained in each solution pack. The values
are read into the analyzer when the solution pack is installed in an analyzer.
9-4
ABL90 FLEX reference manual 9. Solutions
Certificate of traceability
9-5
9. Solutions ABL90 FLEX reference manual
9-6
10. Messages
10-1
10. Messages ABL90 FLEX reference manual
Step Action
1. Open the Activity log.
2. Find the relevant error.
3. Highlight it by touching the screen.
4. Press Troubleshoot.
5. Follow the procedures given to remedy the error.
The following table describes possible errors and how to remedy them.
NOTICE: The list contains all possible errors and may thus list errors that will
not be relevant for all analyzer variants. Furthermore, operator actions are in
relation of the analyzer and may differ from local procedures in your institution.
In that case, follow local procedures.
10-2
ABL90 FLEX reference manual 10. Messages
* The ABL90 FLEX reference manual includes a quality control appendix for
manual QC.
10-3
10. Messages ABL90 FLEX reference manual
94 Value below The parameter value is below - Check for and remedy
reportable range the reportable range. other errors related to
the result, system
messages or calibration
status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Perform a
measurement on new
blood sample.
128 LIS/HIS: Failed The communication hardware - Check that the remote
to open was busy or the remote system is running,
connection system did not respond. correctly configured and
responding.
- Check communication
parameters, e.g. baud
rate, parity, IP address,
etc., as defined in
Communication Setup.
- Reboot the analyzer.
10-4
ABL90 FLEX reference manual 10. Messages
204 Westgard Rule The difference between two - Verify procedure and
(R.4s) violation consecutive measurements repeat measurement.
exceeds 4 SD. This may - Check Replacement
indicate an inconsistency in Status for pending
your procedure or an electrode replacements.
unstable analyzer. - Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
10-5
10. Messages ABL90 FLEX reference manual
211 Quality control One or more errors were - Check Quality Control
error(s) present registered during last QC Status for errors. View
measurement on one of the QC error messages and
installed QC levels. take required corrective
action.
10-6
ABL90 FLEX reference manual 10. Messages
290 Warning: SHb FSHb detected in the range - No action required. For
detected of 1-10 %. information only.
293 Oxi compensated OXI parameters have been - No action required. For
for HbF HbF compensated. Parameter information only.
FHbF may be shown or not
shown.
10-7
10. Messages ABL90 FLEX reference manual
375 Calibration status The status value is outside - Check for and remedy
out of limits the range for the given any system messages.
parameter. - Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
376 Calibration Drift The Drift 1 value exceeds the - Check for and remedy
1 out of range tolerance. any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
377 Calibration Drift The Drift 2 value exceeds the - Check for and remedy
2 out of range tolerance. any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
10-8
ABL90 FLEX reference manual 10. Messages
378 Calibration The sensitivity value is out of - Check for and remedy
sensitivity out of range for the given any system messages.
range parameter. - Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
484 Today is last day After the current day, quality - Print the QC statistics
in stat. month - control statistics obtained if a copy is required.
remember to over the month will be
print QC deleted and new statistics
statistics started.
487 A new statistical A new statistical month has - Make a WDC report
month has begun begun. disk.
- remember to
Removal condition:
export WDC data
- A WDC report disk has
been made.
10-9
10. Messages ABL90 FLEX reference manual
522 Calibration error One or more calibration - Check for and remedy
values are erroneous. any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration
523 Calibration drift Calibration drift exceeds - Check for and remedy
out of range defined limits. any System Messages.
- Perform any pending
replacements including
electrodes.
- Check that electrodes
are properly installed.
- Verify that proper
solutions and gases are
used.
- Perform the Electrode
Troubleshooting
procedure.
Removal condition:
- Calibration drift within
defined limits.
10-10
ABL90 FLEX reference manual 10. Messages
529 Inlet LS failed to Inlet liquid sensor failed to - Repeat the liquid
calibrate calibrate. sensor calibration.
- Contact Radiometer
service representative.
531 Sensors LS failed Liquid sensor near the sensor - Repeat the liquid
to calibrate cassette failed to calibrate. sensor calibration.
- Check solution pack
status and replace if
necessary.
- Contact Radiometer
service representative.
537 OXI LS failed to OXI module liquid sensor - Repeat the liquid
calibrate failed to calibrate. sensor calibration.
- Check solution pack
status and replace, if
necessary.
- Contact Radiometer
service representative.
581 OXI spectrum Spectrum deviates from the - Check the patient
mismatch expected blood or QC record for medication
spectrum. Measurement may containing possible
be unreliable. interfering substances.
- Start a calibration.
- Contact Radiometer
service representative.
10-11
10. Messages ABL90 FLEX reference manual
595 Liquid sensor One or more of the liquid - Repeat the liquid
calibration error sensors failed calibration. sensor calibration.
- Check solution pack
status and replace, if
necessary.
- Contact Radiometer
service representative.
606 Cal expired (pH) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful calibration.
reported as ".....".
608 Cal expired Too long time passed since - Perform a calibration.
(pCO2) the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 2-point
reported as ".....". calibration.
609 Cal expired (pO2) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 2-point
reported as ".....". calibration.
610 Cal expired (K) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 2-point
reported as ".....". calibration.
611 Cal expired (Na) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 2-point
reported as ".....". calibration.
612 Cal expired (Ca) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 2-point
reported as ".....". calibration.
10-12
ABL90 FLEX reference manual 10. Messages
613 Cal expired (Cl) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 2-point
reported as ".....". calibration.
614 Cal expired (Glu) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 1- or 2-
reported as ".....". point calibration.
615 Cal expired (Lac) Too long time passed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 1- or 2-
reported as “…..”. point calibration.
616 Cal expired (OXI) Too long time elapsed since - Perform a calibration.
the last successful calibration
Removal condition:
of the parameter. Parameter
measurement values are - Successful 1- or 2-
reported as ".....". point calibration.
643 ABL/DMS PC The connection from the DMS - Shut down and restart
disconnected PC to the wet section is lost. the analyzer.
from wet section - Contact Radiometer
service representative.
648 Calibration failed The last calibration was - Check solution pack
or not accepted aborted or not accepted. status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
- Check for and remedy
system messages.
- Repeat the calibration.
Removal condition:
- Successful calibration.
10-13
10. Messages ABL90 FLEX reference manual
682 OXI module not The OXI module is not - Shut down the
active responding due to an internal analyzer, using the
communication problem, or Temporary Shutdown
the software configuration function; then restart it.
does not match the analyzer - Contact Radiometer
type. service representative.
Removal condition:
- OXI module ready, or
software configured
without OXI module
support.
688 ctHb/ceHb too ctHb < 1 mmol/L, or ceHb < - If Oxi derivates are
low for OXI 0.75 mmol/L. If ctHb is too wanted, elevate tHb
calculation low, FHHb, FO2Hb, FCOHb and/or sO2.
and FMetHb are not
calculated. If ceHb = cHHb +
cO2Hb is too low, sO2 is not
calculated.
693 ABL not The analyzer was refused - Enter the correct
connected to connection to RADIANCE due password in the
RADIANCE - to incorrect password. analyzer's RADIANCE
incorrect Communication Setup.
password
Removal condition:
- RADIANCE connection
established or disabled.
10-14
ABL90 FLEX reference manual 10. Messages
10-15
10. Messages ABL90 FLEX reference manual
734 General WSM The data management - Wait a few minutes for
exception system establishes the connection to
connection to the analyzing establish.
unit, or the connection is - Restart the analyzer.
lost. - If the error persists,
contact Radiometer
service representative.
745 Low disk space Free disk space is low. - Move archive files to
another storage device.
Removal condition:
- Sufficient free hard
disk space.
770 Failed to restore The setup could not be - Download the setup
Custom Setup restored. data from another
floppy disk, hard disk or
network drive.
- Contact Radiometer
service representative if
the error persists.
10-16
ABL90 FLEX reference manual 10. Messages
787 Export of data Export of data logs was - No action required. For
logs started started by the user. information only.
799 User logged off User logged off. - No action required. For
information only.
800 Logon attempt User tried to log on but did - Provide a valid
failed not provide a valid password. password to log on.
10-17
10. Messages ABL90 FLEX reference manual
10-18
ABL90 FLEX reference manual 10. Messages
10-19
10. Messages ABL90 FLEX reference manual
832 pCO2 unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
833 pO2 unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
839 Glu unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
840 Lac unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
10-20
ABL90 FLEX reference manual 10. Messages
841 tHb unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
842 MetHb unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
843 COHb unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
844 HHb unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
845 O2Hb unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
846 sO2 unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
847 HbF unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
848 tBil unlocked The message is used in the - No action required. For
Activity Log to indicate that a information only.
previously locked parameter
has been unlocked.
855 Base Excess out Base Excess exceeds the +/- - For information only.
of range 30 mmol/L range. No analyzer error was
detected.
875 Sample aged The specified limit for sample - Draw and analyze new
age has been exceeded. sample.
10-21
10. Messages ABL90 FLEX reference manual
964 Leak current in Leak currents were detected - Replace solution pack.
relation to during system calibration and
- Contact Radiometer
solution pack may distort measuring
service representative.
detected results.
970 Replace solution This message is shown when - Replace solution pack.
pack the solution pack needs to be
replaced. The analyzer will
enter "User-intervention
required".
978 Flow selector Shown in the Activity Log - The analyzer will
calibration error when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
984 The analyzer Shown in the Activity Log - The analyzer will
could not when "User-intervention automatically enter
aspirate required" has been entered "User-intervention
homogeneous due to this reason. required". Follow the
calibration instructions shown on
solution the screen.
10-22
ABL90 FLEX reference manual 10. Messages
1002 pO2 dark data is Can be shown on a result if - Restart the analyzer.
out of range unable to calculate oxygen - Perform a calibration.
due to an unexpected system - Contact Radiometer
error. service representative.
1011 Oxi has no Blank Missing Blank Cal. Not - Perform a calibration.
Cal necessarily a hardware error. - Restart the analyzer.
- Contact Radiometer
service representative.
Removal conditioning:
- Successful Blank
calibration
1012 Oxi has no The system has not made a - Repeat the
sample spectrum sample measurement yet, or measurement.
there is a hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
10-23
10. Messages ABL90 FLEX reference manual
1014 Oxi Blank Cal. The spectrometer received - Check solution pack.
intensity too high too high light intensity during During Oxi Blank
Blank Cal. calibration, the cuvette
must be filled with
liquid.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
1019 Oxi Blank Cal. Peak value of Blank Cal. - Check solution pack.
outside limits spectrum intensity is outside The cuvette must be
acceptance limits. filled with liquid during
Blank calibration.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
10-24
ABL90 FLEX reference manual 10. Messages
1028 Oxi neon voltage Oxi hardware problem. - Restart the analyzer.
outside limits - Contact Radiometer
service representative.
1029 Oxi light source Oxi hardware problem. - Restart the analyzer.
voltage outside - Contact Radiometer
limits service representative.
1033 Oxi task was not Internal software problem. - Restart the analyzer.
finished - Contact Radiometer
service representative.
10-25
10. Messages ABL90 FLEX reference manual
1062 Pressure test A leak has been found in the - The analyzer will
pressure error solution transport. automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1063 Pressure test A leak has been found in the - The analyzer will
vacuum error solution transport. automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
10-26
ABL90 FLEX reference manual 10. Messages
10-27
10. Messages ABL90 FLEX reference manual
1090 No rinse solution Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1092 No cal 2 solution Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1093 No cal 3 solution Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1094 No QC1 solution Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1095 No QC2 solution Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1096 No QC3 solution Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
10-28
ABL90 FLEX reference manual 10. Messages
1099 Pump calibration Shown in the Activity Log - The analyzer will
error when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1100 Outlet LS not Shown in the Activity Log - The analyzer will
empty during when "User-intervention automatically enter
pump calibration required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1101 Outlet LS not full Shown in the Activity Log - The analyzer will
during pump when "User-intervention automatically enter
calibration required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1112 LS inlet not Shown in the Activity Log - The analyzer will
empty when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1113 LS sensors not Shown in the Activity Log - The analyzer will
empty when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1114 LS outlet not Shown in the Activity Log - The analyzer will
empty when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
10-29
10. Messages ABL90 FLEX reference manual
1121 The port did not This message is shown in the - Reinstall the sensor
open during Activity Log after a failed cassette.
sensor sensor cassette replacement. - Restart the analyzer.
replacement - Contact Radiometer
service representative.
1123 The sensor chip This message is shown in the - Reinstall the sensor
data could not be Activity Log after a failed cassette.
read or written sensor cassette replacement. - Restart the analyzer.
during - Contact Radiometer
replacement service representative.
1134 The chip This message is shown in the - Reinstall the solution
information for Activity Log after a failed pack.
the solution pack solution pack replacement. - Restart the analyzer.
cannot be read - Contact Radiometer
or written service representative.
1135 The solution pack This message is shown in the - Reinstall the solution
has been used Activity Log after a failed pack.
before solution pack replacement.
10-30
ABL90 FLEX reference manual 10. Messages
1140 The solution pack This message is shown in the - Reinstall the solution
has used the Activity Log after a failed pack.
maximum solution pack replacement.
number of
measurements at
installation
1142 The printer door Printer door open. - Ensure that the printer
is open. Printing paper is properly
not possible installed.
- Close the printer door.
1143 Internal printer is Printer hardware error. - Ensure that the printer
offline. Printing paper is properly
not possible installed.
- Close the printer door.
1144 Check that Printer hardware error. - Ensure that the printer
printer door is paper is properly
closed and that installed.
paper is present - Close the printer door.
1145 A printer error Printer hardware error. - Ensure that the printer
has occurred. paper is properly
Call service installed.
technician - Close the printer door.
1146 Printer paper This message is shown in the - No action required. For
replaced Activity Log after information only.
replacement of printer paper.
1147 Inlet opened Shown in the Activity Log - The analyzer will
during rinse when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1148 Inlet open during Shown in the Activity Log - The analyzer will
calibration when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1149 Inlet open during Shown in the Activity Log - The analyzer will
wet section when "User-intervention automatically enter
activity required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1150 Inlet closed This message is shown in the - No action required. For
without Activity Log when a information only.
aspirating measurement has been
sample cancelled due to inlet being
closed before aspiration
could be completed.
10-31
10. Messages ABL90 FLEX reference manual
1151 Inlet not closed: This message is shown in the - No action required. For
no sample Activity Log when a information only.
aspirated measurement has been
cancelled due to inlet being
closed too late.
1152 The solution pack This message is shown in the - Repeat replacement
chip data could Activity Log when a operation.
not be read or replacement of the sensor
written during cassette or solution pack has
replacement failed. The reason was that it
was impossible to
communicate with the chip
on the consumable.
1160 The top The top thermistor is not - Restart the analyzer
thermistor is not connected - If still present, replace
connected top thermistor
1161 The top The top thermistor short- - Restart the analyzer
thermistor short- circuited - If still present, replace
circuited top thermistor
1163 The sensor The sensor cassette - Restart the analyzer
cassette thermistor is not connected - If still present, replace
thermistor is not sensor cassette
connected
1165 Solution pack not Shown in the Activity Log - The analyzer will
properly installed when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1166 Solution pack Shown in the Activity Log - The analyzer will
expired when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1167 Sensor cassette Shown in the Activity Log - The analyzer will
not properly when "User-intervention automatically enter
installed required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
10-32
ABL90 FLEX reference manual 10. Messages
1168 Sensor cassette Shown in the Activity Log - The analyzer will
expired when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1169 Unable to pump Shown in the Activity Log - The analyzer will
solutions when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1170 Inlet has been Shown in the Activity Log - The analyzer will
open for too long when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1171 Inlet is missing Shown in the Activity Log - The analyzer will
or in unknown when "User-intervention automatically enter
state required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1172 Sensor cassette Shown in the Activity Log - The analyzer will
damaged when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1173 Solution pack Shown in the Activity Log - The analyzer will
damaged when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1174 Inlet opened Shown in the Activity Log - The analyzer will
while the when "User-intervention automatically enter
analyzer was required" has been entered "User-intervention
busy due to this reason. required". Follow the
instructions shown on
the screen.
10-33
10. Messages ABL90 FLEX reference manual
1176 A liquid sensor Shown in the Activity Log - The analyzer will
error was when "User-intervention automatically enter
detected required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1177 A flow selector Shown in the Activity Log - The analyzer will
error was when "User-intervention automatically enter
detected required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1180 An error occurred Shown in the Activity Log - The analyzer will
when trying to when "User-intervention automatically enter
communicate required" has been entered "User-intervention
with wet section due to this reason. required". Follow the
instructions shown on
the screen.
10-34
ABL90 FLEX reference manual 10. Messages
1184 Leak detected Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1185 Warning: Free The internal memory is low. - Restart the analyzer
memory is low
1189 FTC aborted, LS Shown in the Activity Log - The analyzer will
state change when "User-intervention automatically enter
error required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1190 Inlet open Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1193 QA Portal output Shown in the Activity Log - No action required. For
queue cleared when the QA Portal has been information only.
reset.
10-35
10. Messages ABL90 FLEX reference manual
1195 ABL not The analyzer was refused - Enter the correct
connected to QA connection to the QA Portal password in the
Portal - incorrect due to incorrect password. analyzer's QA Portal
password Communication Setup.
1199 FTC program has This message is found in the - No action required. For
been retried Activity Log when a information only.
measurement or calibration
activity has been retried due
to error.
1200 Solution pack Shown in the Activity Log - The analyzer will
empty when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1201 Solution pack Shown in the Activity Log - The analyzer will
lifetime expired when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1202 Expiration date Shown in the Activity Log - The analyzer will
reached when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
10-36
ABL90 FLEX reference manual 10. Messages
1217 No more tests Shown in the Activity Log - The analyzer will
left when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1218 Expiration date Shown in the Activity Log - The analyzer will
reached when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
10-37
10. Messages ABL90 FLEX reference manual
1223 Analyzer did not The analyzer DMS has not - Restart the analyzer.
connect at start- been able to establish - Contact Radiometer
up contact to the WS(M) at service representative.
start-up.
1225 The sample is The time between sampler - Either sampler draw
older than a day draw time and aspiration is time has been entered
larger than 1 day. incorrectly or time of
the analyzer is incorrect.
Change either to correct
the error.
1226 The sample age The time between sampler - Either sampler draw
is negative draw time and aspiration is time has been entered
less than zero. incorrectly or time of
the analyzer is incorrect.
Change either to correct
the error.
1234 Demonstration
software - not for
clinical purposes
10-38
ABL90 FLEX reference manual 10. Messages
10-39
10. Messages ABL90 FLEX reference manual
1295 Activity has been This message is shown in the - No action required.
repeated due to activity log when an activity
the following is repeated automatically. It
reason: lists the error and parameter
id that was the cause of the
repeat.
1296 Printer out of The printer is out of paper. A - Insert a new paper roll
paper new paper roll must be
inserted
10-40
ABL90 FLEX reference manual 10. Messages
1297 Printer is offline The printer is offline due to - Check the power
either a bad or missing connection
power / USB connection - Check the USB
connection
- Contact Radiometer
service representative.
1298 Printer lid open The printer lid is open - Close the printer lid
1306 Solution pack Shown in the Activity Log - The analyzer will
manually when "User-intervention automatically enter
removed required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1307 Disk space less The disk space on the - Delete some archives
than fifteen analyzer is low. to free up space on the
percent drive.
1308 Disk space less The disk space on the - Free disk space. E.g.
than one percent analyzer is less than 1 % deleting some archives
10-41
10. Messages ABL90 FLEX reference manual
1311 The analyzer chip It's not possible to read or - Contact Radiometer
data could not be write data to the analyzer service representative.
read or written chip
1312 Export data logs The export data log - Make sure the selected
failed operation has failed. export path exists.
- Make sure enough
space is available.
1313 Export data logs The export data log - No action required.
done operation has completed
successfully.
1315 Cal backlog error Cal backlog error (pH), - Perform rinse
(pH) leaping signals on rinse
10-42
ABL90 FLEX reference manual 10. Messages
1327 Analyzer locked User has locked the analyzer - No action required.
by user
10-43
10. Messages ABL90 FLEX reference manual
1341 Leak detected Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1342 Leak detected Shown in the Activity Log - The analyzer will
when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1343 Unable to pump Shown in the Activity Log - The analyzer will
solutions when "User-intervention automatically enter
required" has been entered "User-intervention
due to this reason. required". Follow the
instructions shown on
the screen.
1353 User Action The analyzer has entered - User should perform
Needed entered User Action Needed action shown on screen
1354 User Action The analyzer has exited User - No action required
Needed exited Action Needed
10-44
ABL90 FLEX reference manual 10. Messages
10-45
10. Messages ABL90 FLEX reference manual
10-46
I Appendix - Quality control
General information ............................................................................. I-2
Statistical parameters .......................................................................... I-3
Control ranges (for manual QC only) ...................................................... I-4
User control ranges (for manual QC only) ............................................... I-6
Statistics factor and statistics range ....................................................... I-9
Temperature corrections (for manual QC only)....................................... I-10
Westgard rules.................................................................................. I-12
Quality control evaluation ................................................................... I-15
Appendix – Quality control ABL90 FLEX reference manual
General information
I-2
ABL90 FLEX reference manual Appendix – Quality control
Statistical parameters
This section describes the terms and statistical parameters used in the topic
quality control:
Parameter Definition
Mean value, X The mean value is the average value as shown below:
X
X
n
where
X = single result
1 SD
(X X) 2
n 1
and can be illustrated on the normal distribution curve:
I-3
Appendix – Quality control ABL90 FLEX reference manual
About control The control ranges are the ranges within which the result of a quality control
ranges measurement should fall in order for the analyzer to be well functioning. Each
package of quality control ampoules is supplied with an insert with control ranges
for all parameters and analyzers which the given quality control system can
evaluate.
As the control ranges determine whether an analyzer is judged to be well
functioning or not, it is imperative that they are founded on well-established
reference methods and certified reference standards, and in-depth knowledge of
the correlation between an analyzer type and the quality control solution.
Blood gas analyzers are developed to analyze whole human blood, and different
types of analyzers will measure identical values (within the specifications). The
blood algorithm of the analyzers ensures this. When quality control solutions,
which are not blood, are measured, the same corrections will be applied to the
quality control result. Therefore, different types of analyzers will measure
differently, and control ranges must be established separately for each type of
analyzer.
The width of the control ranges is determined during the development phase by
measuring on a number of analyzers. The measurements are performed by
different persons, over several days, using different dispensers and with different
lot numbers of calibration solutions – all this to ensure that all natural variations
such as:
person-to-person day-to-day
analyzer-to-analyzer dispenser-to-dispenser
lot-to-lot variation from calibration solutions, etc.
Definitions The following terms are used by Radiometer in connection with control ranges:
I-4
ABL90 FLEX reference manual Appendix – Quality control
The inserts have control ranges for all analyzers for which the quality control
solution can be used. By using the data codes on the inserts, the information
about the type and level of the quality control solution and control ranges for all
parameters are transferred to the analyzer. The analyzer is then able to
recognize the quality control solutions.
I-5
Appendix – Quality control ABL90 FLEX reference manual
Introduction The specified control ranges in the inserts, which are established by Radiometer,
include all well-functioning analyzers, and the uncertainty budget will therefore
include a contribution derived from the analyzer-to-analyzer variation. This
means that results obtained on one specific analyzer should fall within control
ranges that are narrower than the insert control ranges.
Before establishing analyzer-specific or so-called user-defined control ranges,
you should ensure that your analyzer functions correctly and is properly
maintained. The procedure below should be followed. User control ranges must
be established and updated each time you start using a new lot of quality
control solutions.
Step Action
1. Perform at least 20 measurements as described above.
2. Enter the QC screen (press Menu > Utilities > Setup (if
necessary, log on) > QC Setup > QC Ranges).
Select the desired control solution by pressing Next Slot.
3. Press Edit to change the lower/upper limit in the "Lot to date range
(2SD)" column. Confirm each change with Enter on the keypad.
Use Next or Prev. Param. to change to another parameter.
4. Press Close to exit the program.
To get full benefit from the evaluation procedure, Radiometer recommends the
use of a statistics factor of 1.5 (default) to establish the statistics range (see
section Statistics factor and statistics range further in this appendix.
Another option is to manually correct the 2 SD control ranges in order to include
uncertainty contributions from:
chemical decomposition of the QUALICHECK5+ solution
inhomogeneity of QC lots
calibration solutions
I-6
ABL90 FLEX reference manual Appendix – Quality control
To correct the 2 SD control ranges for the above contributions, do the following:
Step Action
1. Find the mean ( X ) and the two times standard deviation (2 SD)
value in the Quality Control log (press: Data Logs > Quality
Control Log > Statistics) or calculate the values from the last 20
quality control measurements.
2. Find SDtotal in tables below.
3. Determine SDcorrected as follows:
2 SD 2 SDtotal
2 2
SDcorrected
Level 2
Parameter: ABL90 FLEX analyzer
pH 0.0054
pCO2 kPa 0.09
pO2 kPa 0.26
cK+ mmol/L 0.036
+
cNa mmol/L 0.6
I-7
Appendix – Quality control ABL90 FLEX reference manual
Level 3
Parameter: ABL90 FLEX analyzer
pH 0.0074
pCO2 kPa 0.07
pO2 kPa 0.44
+
cK mmol/L 0.036
+
cNa mmol/L 0.6
cCa2+ mmol/L 0.019
cCl mmol/L 1.08
cGlu mmol/L 0.4
cLac mmol/L 0.2
ctBil mol/L 4.1
ctHb g/dL 0.18
FHbF % 2.6
Other Hb derivatives % 0.19
Level 4:
Parameter: ABL90 FLEX analyzer
pH 0.0046
pCO2 kPa 0.26
pO2 kPa 0.73
+
cK mmol/L 0.055
cNa+ mmol/L 0.5
2+
cCa mmol/L 0.023
cCl mmol/L 0.95
cGlu mmol/L 0.1
cLac mmol/L 0.1
ctBil mol/L 2.2
I-8
ABL90 FLEX reference manual Appendix – Quality control
User control ranges (2 SD): The statistics range will correspond toX 3 SD.
Insert control ranges: The statistics range will correspond to 1.5
insert control range.
All results outside the statistics range will be excluded from the statistics and
marked accordingly.
Example The control range is: pH low = 6.986 and pH high = 7.016
To calculate the statistics range, do the following:
Step Action
1. Calculate the mean value: X = (6.986 + 7.016)/2 = 7.001.
2. Calculate the 2 SD: pH high X = 7.016 7.001 = 0.015.
3. Calculate the 3 SD: (0.015 3)/2 = 0.0225 = 0.023.
4. The statistics range will then be:
pH low = 7.001 0.023 = 6.978.
pH high = 7.001 + 0.023 = 7.024.
I-9
Appendix – Quality control ABL90 FLEX reference manual
Purpose Temperature correction is done to ensure that the quality control results reflect
the analyzer performance and are not influenced by fluctuations in the ambient
temperature.
Parameters The insert control ranges are determined at a reference temperature of 25 °C.
that require Deviations from this temperature will have an impact on the following
temperature parameters: pH, pCO2 and pO2.
correction
The reason for temperature correction is as follows:
To ensure that the QC result actually reflects the performance of the analyzer and
not just fluctuates because of temperature variations, it is important to keep the
ampoule at a stable and known temperature, so that variations can be corrected
for in the correct way.
Temperature Parameters
pH pCO2 pO2
> 25 C
< 25 C
Radiometer recommends that ampoules that have been stored in a cool place are
conditioned at a known room temperature for at least 5 hours before a
measurement, and we strongly advise not to keep ampoules on the top of an
analyzer as the temperature there can vary.
In the ABL90 FLEX analyzers the software will automatically temperature correct
the results on Radiometer QUALICHECK5+ solutions once the ambient
temperature has been entered.
Radiometer uses a reference temperature of 25 C.
I-10
ABL90 FLEX reference manual Appendix – Quality control
Temperature Range+ QUALICHECK solutions (usable for e.g. Calibration verification, see
corrections for section Calibration verification in Chapter 6: Calibration in the ABL90 FLEX
pH, pCO2 and operator's manual).
pO2
Parameter Equation for temperature correction
pH: pH(corr. to 25 C) = pH(meas.) A(t 25)
where A = a temperature constant. The values are given in the table below.
I-11
Appendix – Quality control ABL90 FLEX reference manual
Westgard rules
About The Westgard rules are a set of statistical rules that, when applied to the quality
Westgard rules control results, can aid the following:
Westgard rules are based on the calculation of the mean and standard deviation
(SD) of quality control measurement values for a particular parameter and a
specific device, through modification of control ranges. They are best expressed
in the form of plots.
Westgard rules are divided into two types:
Shows 3 SD ranges
Shows control ranges (2 SD)
Shows the mean value
The Westgard rules described in the following are selected for evaluation of
quality control measurement results. All six rules are applicable to manual QC.
Only four of the rules are applicable to built-in QC.
I-12
ABL90 FLEX reference manual Appendix – Quality control
Measurement value is
outside the mean 2 SD.
Measurement value is
outside the mean 3 SD.
Two consecutive
measurements are
outside the mean 2 SD
on the same side of the
mean.
I-13
Appendix – Quality control ABL90 FLEX reference manual
Four consecutive
measurements outside
the mean 1 SD on the
same side of the mean.
10 consecutive
measurements on the
same side of the mean.
I-14
ABL90 FLEX reference manual Appendix – Quality control
No. Question
1. Did you store the ampoules according to specifications?
2. Did you condition the ampoules according to the specifications?
3. Did you key in the correct ampoule conditioning temperature?
4. Did you shake the ampoule vigorously for 15 seconds before using
it?
5. Did you remember to hold the ampoule between your thumb and
index finger when you shook it? (This is done to avoid heating up
the ampoule contents and thus change its temperature.)
6. Did you analyze control solutions immediately after opening the
ampoule?
(Each QC ampoule must be used immediately after being opened,
for one measurement on one analyzer only, in order to ensure the
reliability of the measurement).
7. Did you use Radiometer QUALICHECK adapter?
8. If own user-defined limits were entered: Did you enter too narrow
control ranges?
9. If everything is OK – repeat the QC measurement.
I-15
Appendix – Quality control ABL90 FLEX reference manual
I-16
II Appendix -
Traceability to the primary standards at
Radiometer
Appendix – Traceability to the primary standards at Radiometer ABL90 FLEX reference manual
Introduction
The Metrology Section at Radiometer is responsible for establishing metrological
traceability for the parameters pH, pCO2, pO2, cK+, cNa+, cCa2+, cCl–, cGlu,
cLac, sO2, FCOHb, FMetHb, FHbF, Hct, ctBil and ctHb. This booklet documents
the traceability for each of these parameters.
II-2
ABL90 FLEX reference manual Appendix – Traceability to the primary standards at Radiometer
Traceability
pH The primary pH standards are traceable to the definitive method for pH. The
definitive method is based on a Hydrogen Electrode System. The primary pH
standards are obtained from the Danish primary laboratory for Electrochemistry
(DPLEC) at the Danish Institute of Fundamental Metrology (DFM). This primary
laboratory is accredited by Danish Accreditation (DANAK accreditation no. 255)
Certification is done in accordance with the method recommended by the
International Union of Pure and Applied Chemistry (IUPAC). The Hydrogen
Electrode System of DLPEC is validated by comparison with Standard Reference
Materials (SRMs) produced by the National Institute of Standards and
Technology (NIST). The primary standards are therefore also traceable to NIST.
The IUPAC-recommended method is described in Ref. 1.
The NIST SRMs used are: 186I/II, 185, 187, 191-192.
Using the primary pH standards, the secondary pH standards are certified in the
Metrology Section. These are normally of the same composition as the primary
buffers, tapped into 2-mL glass ampoules and heat sterilized. The secondary
buffers are stored at 5 °C. Measurements of the secondary buffers are done
using a glass electrode with a saturated calomel reference electrode and a liquid
junction of saturated KCl. The liquid junction is a vertical, cylindrical and open
liquid junction. Measurement of a secondary buffer is done using a primary
buffer together with a certified secondary buffer as standards for making a 2-
point calibration of the glass electrode arrangement.
pCO2 and pO2 The primary gases used are Standard Reference Materials (SRMs) produced by
NIST. The NIST SRMs used are: 1674b, 2625a, 2658a and 2659a. The NIST
SRM gases are used to validate primary gravimetric working gas standards,
certified by Scott Medical, Air Liquide or Air Products. The primary gravimetric
working gas standards are validated using a computer-controlled gas
chromatography system, introducing the NIST SRM gases as samples and
comparing the obtained results with the certified values.
The primary gravimetric working gas standards are used as standards in the gas
chromatography system, so that the composition of secondary working gas
standards can be determined.
By using the secondary working gas standards in a tonometer together with an
aqueous buffer solution, a solution with a known pCO2 and pO2 is produced. This
aqueous buffer solution is then used to determine the pCO2 and pO2 of
secondary working standards. These secondary working standards are aqueous
buffer solutions kept in 2-mL ampoules.
cK+ and cNa+ The primary working standards used are gravimetric standards produced from
KCl and NaCl Suprapur, produced by Merck. These primary working standards
are validated using Standard Reference Materials (SRMs) produced by NIST, so
that traceability to NIST is achieved. The NIST SRMs used are: 919a (NaCl) and
999 (KCl). Validation of the primary working standards is done using a flame
photometer together with the NIST SRMs.
The flame photometer method of validating the primary working standards is
described in Ref. 2.
The primary working standards are used to determine the sodium and
potassium concentrations of the secondary working standards. The
concentrations of the secondary working standards are measured using a flame
photometer. The determination of the sodium and potassium concentrations of
fluorocarbon-based secondary standards is done using ion-selective K and ion-
selective Na electrodes on the ABL735 analyzer. The determination takes place
using the primary working standards.
II-3
Appendix – Traceability to the primary standards at Radiometer ABL90 FLEX reference manual
cCa2+ The primary standards used are the so-called Ca2+ transfer standards, produced
from CaCO3 Urtitersubstanz®, produced by Merck. The transfer standards are
pH-stabilized to pH = 7.4, with 1 mmol/L HEPES and an ionic strength of 160.0
mmol per kg. Validation of the Ca2+ transfer standards is done using similar
standards produced from NIST SRM 915.
The transfer standards are used to determine the calcium concentrations of
secondary standards. These measurements take place using ion-selective Ca
electrodes on the ABL735 analyzer.
cCl– The primary working standards are gravimetric standards, prepared from KCl
Suprapur, produced by Merck. The primary working standards are validated by
making comparative titrations using similar standards prepared from NIST SRM
999 (KCl). The titrations are done using an AgNO3 solution as the titrant, and
potentiometric titration equipment.
The standardized AgNO3 solution is used as the titrant for the determination of
the chloride concentration of the secondary standards, using the potentiometric
titrator (TitraLab 900 from Radiometer Analytical, France).
cGlu The primary working standards are prepared from NIST SRM 917a (D-glucose).
These primary standards are used to determine the glucose concentration of
secondary standards. The measurements take place using the glucose reference
method, which is the hexokinase/glucose-6-phosphate dehydrogenase method
recommended by CLSI. This method is described in Ref. 3.
cLac No certified standard reference material for lactate is available at present. The
primary working standards are therefore prepared from a pure commercially
available material, namely the Lithium salt of L (+) Lactic Acid (Cat. No, L-2250)
supplied by the Sigma Chemical Company.
These primary standards are used to determine the lactate concentration of
secondary standards.
The measurements take place using a spectrophotometric method. The method
is based on a reaction of lactate, catalyzed by L-Lactate Dehydrogenase (LDH).
The reaction produces dihydronicotinamide (NADH), which is measured at 339
nm. The method is described in Ref. 4.
ctHb The primary standard used is an oxygenated whole-blood sample. The ctHb
value of this sample is determined by the use of the HiCN reference method.
This method is described in Ref. 5. The HiCN reference method is a
spectrophotometric method. The spectrophotometer used is calibrated using a
NIST SRM 930D filter. This method is further validated using the certified
reference material Hemoglobin-cyanide standard (product no. 3061) produced
by J.T Baker, Holland.
The primary standard is used to calibrate the ABL735 reference instruments.
II-4
ABL90 FLEX reference manual Appendix – Traceability to the primary standards at Radiometer
Saturation – The primary working standard used is a whole-blood sample, with the ctHb
sO2 = 100 % value adjusted to between 13 and 15 g %. The blood sample is tonometered
with 5.6 % CO2 - 94.4% O2, traceable to NIST SRM gases.
The primary standard is used to calibrate the ABL735 reference instruments.
Saturation – The primary working standard used is a whole-blood sample. The blood sample
sO2 = 0 % is centrifuged and the resultant blood concentrate is deoxygenated using Argon
and treated with a dithionite solution.
The primary working standard is used to calibrate the ABL735 reference
instruments.
FCOHb – The primary standards used are CO with atmospheric air mixtures, produced in
normal value a container of known volume. The CO used for making these gas mixtures has a
certified purity of 99.997 %. Validation of the mixing method is done by
comparison with NIST SRM 1678 (50 ppm CO in N2).
The produced mixtures are used as calibration standards in connection with a
gas chromatography method. The gas sample, injected into the gas
chromatograph, is the headspace of a blood sample which has been treated so
that all the bound CO is released from the hemoglobin. The analyzed result is
measured in % CO, and from this the FHbCO is calculated. The method is
described in Ref. 6.
The measured blood sample is used as secondary standard and is used to
calibrate the ABL735 reference instruments.
FCOHb – The primary working standard used is a whole-blood sample. The blood sample
100 % is tonometered with 100 % CO, with a certified purity of 99.997 % CO. The
primary working standard is used to calibrate the ABL735 reference
instruments.
FHbF The primary working standard is a whole-blood sample. The FHbF of this sample
is determined using the Cation Exchange HPLC reference method. The method is
described in [Ref. 9]. The method is performed by the Hæmatology Laboratory
at Herlev Hospital, Denmark.
II-5
Appendix – Traceability to the primary standards at Radiometer ABL90 FLEX reference manual
Radiometer measurements
The Hct measurement is based on conductivity measured in a sample and then
corrected for the presence of sodium ions. A Sigma 201 Micro hematocrit
centrifuge with RCF of 12620*g, which fulfills most of the CLSI requirements,
has been used for the test together with 75 mm Microhematocrit capillary tubes
with an inner diameter between 1.1 and 1.2 mm. The centrifugation time has
been 5 minutes.
The conductivity and sodium concentration has been measured on
approximately 1000 blood samples with a sodium concentration varying from 80
mmol/L to 180 mmol/L. Hct measurements have then been correlated to the Hct
measured by the reference method.
ctBil The primary working standard is a whole-blood sample. The total bilirubin is
determined on a serum sample prepared from this. The determination is
performed using a Hitachi 717 wet-chemistry analyzer, which uses the
Boehringer Mannheim reagency kit, DPD method, given in Ref. 11. The
reference instrument is calibrated using four levels of NIST SRM916a
unconjugated bilirubin standard material.
II-6
ABL90 FLEX reference manual Appendix – Traceability to the primary standards at Radiometer
References
1. Measurement of pH. Definition, standards, and procedures. (IUPAC Recommendations
2002). Pure and Appl Chem 2002; 74, 11: 2169-2200.
2. Standardization of sodium and potassium ion selective electrode systems to the flame
photometric method. NCCLS (CLSI) Publication C29-A2. Villenova, Pa.: NCCLS, 2000.
3. NCCLS (CLSI) Publication RS1-A, Villenova, Pa.: NCCLS .
4. Bergmeyer HU. Methods of enzymatic analysis. 3rd ed. Deerfield Beach: Verlag
Chemie, 1984; 6: 582-88.
5. Reference methods for the quantitative determination of hemoglobin in blood samples.
NCCLS (CLSI) Publication H15-A3. Villenova, Pa.: NCCLS, 2000
6. Collison HA, Rodkey FL, O'Neal JD. Determination of carbon monoxide in blood by gas
chromatography. Clin Chem 1968; 14, 2: 162-71.
7. Evelyn KA, Malloy HT. Microdetermination of oxyhemoglobin, methemoglobin, and
sulfhemoglobin in a single sample of blood. J Biol Chem 1938; 126: 655-62.
8. Kristoffersen K. An improved method for the estimation of small quantities of alkali-
resistant hemoglobin in blood. Scand J Clin Lab Invest 1961; 13: 402.
9. Tan GB, Aw TC, Dunstan RA & Lee SH, Evaluation of high performance liquid
chromatography for routine estimation of haemoglobins A2 and F. Journal of Clinical
Pathology 46: 852-856.
10. Procedure for determining packed cell volume by microhematocrit method. 2nd ed.
Approved standard. NCCLS (CLSI) Publication H7-A3. Villenova, Pa.: NCCLS, 2000.
11. Wahlfeld AW et al. Bile pigments: Technical aspects, modification of Malloy-Evelyn
method for a simple reliable determination of total bilirubin in serum. Scand J Clin Lab
Invest 1972; 29, Suppl 126: Hitachi Abstr. 11.12.
II-7
Appendix – Traceability to the primary standards at Radiometer ABL90 FLEX reference manual
II-8
Index
Date of issue
Publication: 201207
Edition: A
Code number: 995-433
Date of issue ABL90 FLEX reference manual