GE PC2IP IIB Manual3 2015521111351
GE PC2IP IIB Manual3 2015521111351
- ~ ARTISAN®
~I TECHNOLOGY GROUP
with experienced engineers and technicians on staff.
Note: When performing this procedure, it is recommended to work with the BEP2 lying on its side
(as shown in Figure 8-78 on page 8-91), to provide easier access.
NOTE: For clarification, the letters appearing in square brackets in the following steps correspond to the letters
shown in the illustration of the BEP2 Power Supply and cable connectors in Figure 8-82 below.
H G
Cable Connections
A Motherboard (beside CPU)
Not used
E B Motherboard (beside ECG)
C BEP rear
D Floppy Disk
F E PC-VIC
F Hard Disk
G CD-RW
I H MOD
C B
D I ECG
A
Figure 8-82 BEP2 Power Supply with Cable Loom and Connectors
1.) Route the cable loom down into the cavity of the BEP casing (placing the cable labelled [E] in Figure
8-82 at the front of the loom) and arrange each of the cables in the approximate position. Carefully
feed the cable connectors underneath the data cables.
2.) Holding the new BEP2 power supply in one hand, connect the 4-pin external connector [A] to the
motherboard (beside the CPU).
3.) Slide the power supply into the BEP power supply location (see Figure 8-83 below) and push the
metal stabilizing bracket (located on the side of the power supply) downwards to hold the power
supply in position.
Ground Cable
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
4.) Re-connect the mains power cable [B] to the motherboard (beside the ECG location).
5.) Re-connect the mains power cable (4-pin flat connector [C] - Part No. 2252981) to the rear of the
BEP, and fasten with cable ties.
6.) Re-connect the power cable to the Floppy Disk (labelled [D] in Figure 8-82); refer also to
Figure 8-83, above.
7.) Slide the PC-VIC & hard disk out sufficiently to be able to re-connect the power cable (labelled [E]
in Figure 8-82) to the PC-VIC; refer also to Figure 8-83, above.
8.) Re-connect the data cable to the PC-VIC (see Figure 8-83, above).
9.) Re-connect the power cable to the Hard Disk (labelled [F] in Figure 8-82); refer also to Figure 8-83,
above.
10.)Return the Ground cable and screw that fastens the PC-VIC in position, as shown in Figure 8-80
on page 8-92.
11.)Re-connect the power cables to the media drives (and CD-RW [G]) and MOD [H] .
12.)Carefully place the ECG Module in the front of the BEP, sliding it into position and re-connecting
the power cable [I] to the rear of the ECG Module. (Refer to the BEP2 - ECG Module Installation
Procedure on page 8 - 86).
13.)Gently slide the power supply up into the final position in the BEP casing.
14.)Check that all cable connections are secured firmly and fasten with additional cable ties, as
necessary.
NOTE: Make sure that all data cables, memory boards, etc. are securely in position and that nothing has
erroneously become disconnected during the power supply removal procedure.
15.)Stand the BEP2 upright.
16.)At the rear of the BEP2, return and loosely tighten the four screws that fasten the power supply in
position on the BEP, as shown in Figure 8-77 on page 8-91.
17.)Return the BEP front cover as described in the BEP2 - Front Cover Installation Procedure on page
8 - 68.
18.)Return the one screw previously removed from the top of the side cover, as shown in Figure 8-76
on page 8-90.
19.)Fully tighten the four screws at the rear of the BEP (refer to step 16) to secure the power supply
firmly in position.
20.)Return the BEP2 side covers, as described in the BEP2 - Side Cover Installation Procedure on page
8 - 66.
21) Carefully return the BEP2 to the original position on the machine chassis, and proceed as follows:
a.) Re-connect all the cables on the rear of the BEP (refer to the step-by-step instructions provided
in Back End Processor Installation Procedure on page 8 - 60).
b.) Return the screws of the lower left and right BEP brackets see Figure 8-54 on page 8-59.
a.) Return the two screws of upper BEP bracket (Figure 8-54).
22.)Replace the Vivid™ 3 rear cover - refer to the Rear Cover Installation Procedure on page 8 - 7.
23) Remove the Vivid™ 3 left and right side covers - refer to the Side Covers Installation Procedure on
page 8 - 4.
24) Re-connect the mains AC power cable for the mains power wall outlet.
25) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
26) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-5-17-1 Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the Hard Disk
replacement procedure.
8-5-17-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
6.) Disconnect all the cables at the rear of the hard disk.
7.) Remove the hard disk.
Note: Before installing a new hard disk, make sure the jumper settings are as shown below:
1) Place the new hard disk in position in the hard disk module holder.
2.) Reconnect all the cables at the rear of the hard disk.
3) Fasten each of the four screws (two on each side), as shown in Figure 8-85 on page 8-97.
4) Carefully slide the module holder into position in the unit.
5) Return and fasten the screw that holds the hard disk module in position (as shown in Figure 8-74
on page 8-88
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Note: When a new hard disk arrives from the manufacturer, usually it is partitioned and ready for
software installation.
Refer to Software Loading on page 8 - 121 for software installation instructions .
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 8-6
Lower Section Components Replacement
8-6-1 AC Distribution Box Replacement Procedure
8-6-1-1 Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the AC
Distribution Box replacement procedure.
8-6-1-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Primary Secondary
Connector Connector
5) Gently pull the AC Distribution Box towards you, pulling it out through the rear of the machine.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-6-2-1 Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the AC Input
Box replacement procedure.
8-6-2-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-6-3-1 Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Allen keys as
indicated in the keyboard or monitor cable replacement procedure.
8-6-3-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-6-4-1 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-6-5-1 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Securing nut
10) Remove the cable end stopper from the release handle.
11) Remove the gas spring release cable. Before removing the cable, pay attention to its path inside
the machine, and the position of each of its parts.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Note: It is recommended to clean the filter before returning it to its original location, as described
in Chapter 10 - Periodic Maintenance.
14) Verify that the gas spring release handle is functioning correctly.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
- Removal of the old-type (plastic) Up/Down Handle (Part No. 2260300) - see Up/Down Handle
Removal - Procedure 1, below.
- Removal of the new-type (metal) Up/Down Handle with support (Part No. 237002-2) - see
Up/Down Handle Removal Procedure 2 on page 8 - 107.
- Installation of the Up/Down Handle with support (Part No. 237002-2) - see Up/Down Handle
Replacement Procedure on page 8 - 108.
8-6-6-1 Tools
Use the appropriate the Phillips-type screw drivers, as indicated in the up/down handle removal
procedure.
8-6-6-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
NOTE: This describes the procedure for removing the old-type (plastic) Up/Down Handle
(Part No. 2260300).
1) Raise the control console to its maximum height. Remove the bottom keyboard cover, as described
in the Bottom Keyboard Cover Removal Procedure on page 8 - 14.
2) Loosen and remove the nut on the M6 screw (see Figure 8-90 below) that secures the up/down
handle in position.
Up/down Handle
M6 Screw
M6 Nut
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
3.) Extract the M6 screw from the up/down handle by sliding it out.
4.) Release the gas spring cable from the broken up/down handle - see Figure 8-91 on page 8-107.
5.) Remove the broken up/down handle.
Pic to be Replaced
NOTE: This describes the procedure for removing the new-type (metal) Up/Down Handle (Part No. 237002-2).
1) Remove the bottom keyboard cover, as described in the Bottom Keyboard Cover Removal
Procedure on page 8 - 14.
2) Remove the right and left probe holders, as described in the Right and Left Probe Holders Removal
Procedure on page 8 - 23.
3) Remove the front handle from the console, as described in the Front Handle Removal Procedure
on page 8 - 24.
4.) Release the gas spring cable by loosening the securing nut (shown in Figure 8-92 below). Slide the
cable out of the retaining bracket and slide the end of the cable out through the slot on the handle
assembly (see below, and refer to the detail in Figure 8-94 on page 8-109).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
5.) Remove the four screws (two short and two long) that secure the handle mounting brackets to the
chassis (see Figure 8-92, above) and remove the handle assembly. Refer also to Figure 8-95 on
page 8-109.
6.) Loosen and remove the M6 nut from the handle assembly axis screw - refer to Figure 8-93 on
page 8-108.
7.) Extract the axis screw from the handle assembly and remove the defective Up/Down handle.
NOTE: This describes the procedure for installing the new-type (metal) Up/Down Handle (Part No. 237002-2).
1) Insert the M6 Axis screw through the screw hole in the first handle mounting bracket, through the
two holes in the new metal Up/Down handle, then through the screw hole in the second bracket and
into M6 nut, as shown in Figure 8-93 below.
M6 Nut
Handle Mounting Bracket
Figure 8-93 Fixing the Up/Down Handle to the Handle Mounting Brackets
2.) Tighten the M6 nut fully to secure the handle mounting brackets to the handle assembly.
3.) Insert the end of the gas spring cable through the slot on the gas spring cable retaining bracket and
into the grooved hole on the Up/Down handle. Slide the end of the cable into the groove, carefully
aligning the groove on the cable with that of the groove in the hole (see Figure 8-94 below).
Tighten the securing nut against the gas spring cable retaining bracket, to hold the cable securely
in position.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Grooved Hole
shown in detail
(insert cable here)
4) Mount the handle mounting brackets onto the chassis using one long and one short screw for each
bracket, as shown in Figure 8-95. Tighten the screws to secure the brackets firmly in position.
Short Screw
Long Screw
Short Screw
Long Screw
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
5) Test the Up/Down handle to make sure the mechanism is operating properly and that the handle is
fastened securely on the chassis.
6.) Return the front handle to the console and secure in position, as described in the Front Handle
Installation Procedure on page 8 - 24
7) Return the right and left probe holders and secure in position, as described in the Right and Left
Probe Holders Installation Procedure on page 8 - 23.
8) Return the bottom keyboard cover and secure in position, as described in the Bottom Keyboard
Cover Installation Procedure on page 8 - 15.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-6-7-1 Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter, the appropriate open-ring
wrenches and a Hex M8x80mm screw as indicated in the gas spring replacement procedure.
8-6-7-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Screw
Screw
6) Remove the securing screw from the cable holder located on the gas spring rod.
Note: In the following steps, all the gas spring components indicated by letters can be seen in
Figure 8-97 and Figure 8-98 on page 8-112.
7) Turn the cable holder (B) to free the cable end sleeve (C) then remove the cable end stopper (D)
from the trigger arm (E).
DANGER: Do not activate the trigger arm as this will cause the expansion of the gas spring and may
cause damage and/or injury!
8) Remove the retaining washer (K) from the pin (L) at the lower end of the gas spring.
9) Insert the Hex M8x80mm screw in the appropriate hole to the left of the compartment at its base,
and then tighten the screw until the pin (L) at the lower end of the gas spring is loose enough to be
removed, and remove the pin.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
B
A
C
F
J
E
K
D
L
A
C
D
E
10) Remove the retaining washer (K) from the pin (L) at the lower end of the gas spring.
11) Insert the Hex M8x80mm screw in the appropriate hole to the left of the compartment at its base,
and then tighten the screw until the pin (L) at the lower end of the gas spring is loose enough to be
removed, and remove the pin.
Note: In the following step, the gas spring components indicated by letters can be seen in Figure
8-99 on page 8-113.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
N M
Note: All the gas spring components indicated by letters in the following steps can be seen in
Figure 8-99, above and Figure 8-97 and Figure 8-98 on page 8-112.
DANGER: Do not activate the trigger arm as this will cause the expansion of the gas spring and
may cause damage and/or injury!
Note: It is recommended to clean the filter before returning it to its original location, as described
in Chapter 10 - Periodic Maintenance.
13) Verify that the gas spring release handle is functioning correctly.
Chapter 8 - Replacement Procedures 8-113
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
When replacing the earlier type of front wheel (Part No. 2269437) with newer improved type (Part No.
2392789) make sure you replace BOTH front wheels. When ordering the new-type Front wheels, you
will receive a new transverse arm which should also be replaced (between the two wheels). Refer to
these instructions:
• Front Wheel Removal - Procedure 1 on page 8 - 114
• Front Wheel Installation - Procedure 2 on page 8 - 118.
If replacing the newer improved type (Part No. 2392789) with the same type of wheels, refer to the
following instructions:
• Front Wheel Removal - Procedure 2 on page 8 - 116
• Front Wheel Installation - Procedure 2 on page 8 - 118.
8-6-8-1 Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter, the appropriate Hex wrenches
and the appropriate Allen keys as indicated in the front wheel replacement procedure.
8-6-8-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Note: Before you begin, make sure you are working on a flat surface with the wheels facing the front
and the brake in the left (locked) position, preventing movement.
The machine will remain stable, even after one wheel has been removed. However, for a safer
working environment, it is recommended to support the unit with wooden blocks or a jack.
WARNING: Do not lean on the machine, or leave it unattended without support, if a wheel has
been removed.
1) Remove the Vivid™ 3 front cover, as described in the Front Cover Removal Procedure on page 8 - 5.
2) Using a flat screwdriver, push up and release the locking spring washer located in the center of the
transverse arm (wheel position bar), as shown in Figure 8-100 on page 8-115. Lift the wheel
position bar upwards (without completely removing it).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Figure 8-100 Releasing the Spring Washer from the Center of the Transverse Arm
3) Remove the four bolts (three wrench bolts and one Allen bolt) that secure the front wheel to the
chassis (see Figure 8-100 below) and remove the wheel.
Securing Bolts
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
WARNING: When performing this procedure, the machine must be safely supported on two
wooden blocks (or on a jack capable of supporting the weight of the machine). Do not lean on
the machine, or leave it unattended without support, when the wheels have been removed.
Always remove and replace one wheel at a time, before attempting to remove the second wheel.
Note: Before you begin, make sure you are working on a flat surface with the wheels facing the front
and the brake in the left (locked) position, preventing movement.
1) Remove the Vivid™ 3 front cover, as described in the Front Cover Removal Procedure on page 8 - 5.
2) Remove the locking washer, located on the black wheel position bar, and lift the wheel position bar.
3) Using a flat screwdriver, push up and release the locking spring washer located in the center of the
transverse arm (wheel position bar), as shown in Figure 8-100 on page 8-115.
4.) Release the locking spring washer from the underside of each end of the transverse arm, as shown
in Figure 8-102 below.
Figure 8-102 Releasing the Spring Washer from the End of the Transverse Arm
5.) Lifting the transverse arm upwards, unhook it from the central locking pin located above the brake
pedal and remove it completely. Refer to Figure 8-103 on page 8-117.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Transverse Arm
(Wheel Position Bar)
Securing Bolts
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Note: The machine will remain stable, even after one wheel has been removed. However, for a safer
working environment, it is recommended to support the unit with wooden blocks or a jack.
1) Install the new wheel (Part No. 2269437) on the machine chassis in the same position as that of the
previously-removed wheel. Make sure it is correctly positioned, parallel to the other front wheel.
2.) Lift the transverse arm (wheel position bar) and secure it into position by re-attaching the spring
washer to the central locking pin (refer to Figure 8-100 on page 8-115).
3.) Return the four bolts (three wrench bolts and one Allen bolt - removed in step 3 on page 8-115) and
make sure that wheel is secured firmly to the chassis (see Figure 8-101 on page 8-115).
4.) Return the Vivid™ 3 front cover, as described in the Front Cover Installation Procedure on page 8 - 6.
WARNING: When performing this procedure, the machine must be safely supported on two
wooden blocks (or on a jack capable of supporting the weight of the machine). Do not lean on
the machine, or leave it unattended without support, when the wheels have been removed.
Always remove and replace one wheel at a time, before attempting to remove the second wheel.
1) Having removed one wheel from the machine chassis (as described in Front Wheel Removal -
Procedure 2 on page 8 - 116), install one of the new wheels (Part No. 2392789) in its place,
ensuring that the wheel is correctly positioned.
Install the four bolts (three wrench bolts and one Allen bolt removed in step 6 on page 8-117 [refer
to Figure 8-104 on page 8-117]), making sure that the wheel is secured firmly to the chassis.
2) Having removed the second wheel from the machine chassis (as described in Front Wheel
Removal - Procedure 2 on page 8 - 116), install the other new wheel in its place, ensuring that the
new wheel is correctly positioned, parallel to the other new front wheel.
Install the four bolts (three wrench bolts and one Allen bolt removed in step 6 on page 8-117 [refer
to Figure 8-104 on page 8-117]), making sure that each wheel is secured firmly to the chassis.
3.) Install the new transverse arm (wheel position bar) by hooking it onto the central locking pin located
above the brake pedal and securing it into position (attach the spring washer [supplied] to the
central pin - refer to Figure 8-103 on page 8-117 and Figure 8-100 on page 8-115).
4.) Lift the transverse arm upwards and attach a new locking spring washer (supplied) to the underside
of each end of the transverse arm. Refer to step 4 on page 8-116 and Figure 8-102 on page 8-116.
5.) Return the Vivid™ 3 front cover, as described in the Front Cover Installation Procedure on page 8 - 6.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: This section describes the installation and removal procedure of the earlier-type of rear wheel
(Part No. 2269436) and the newer improved type (Part No. 2389102). The installation procedure is the
same for each type of rear wheel, as described in the procedure below.
NOTE: IMPORTANT - When replacing the earlier type (Part No. 2269436) with newer improved type (Part No.
2389102) make sure you replace BOTH rear wheels.
8-6-9-1 Tools
Use the appropriate flat and Phillips-type screw drivers and combination spanner as indicated in the
rear wheel replacement procedure.
8-6-9-2 Preparation
Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
Note: The machine will remain stable, even after one wheel has been removed. However, for a safer
working environment, it is recommended to support the unit with wooden blocks or a jack.
WARNING: When BOTH rear wheels are being removed, the machine must be safely supported
on two wooden blocks (or on a jack capable of supporting the weight of the machine). Always
remove and replace one wheel at a time, before attempting to remove the second wheel.
1) Remove the Vivid™ 3 side covers and the rear cover, as described in the Side Covers Removal
Procedure on page 8 - 4 and the Rear Cover Removal Procedure on page 8 - 7.
2) Position the machine so that the wheel that is to be changed is not in contact with the ground, while
ensuring that the other three wheels remain on the ground. If replacing both wheels, support the
machine on two wooden blocks (or a suitable jack).
3) Loosen the shaft screw (see Figure 8-105 on page 8-120) and remove the defective wheel from the
machine chassis.
4) Proceed to the Rear Wheel Installation Procedure on page 8 - 120 to install the new rear wheel in
place of the one removed.
5.) If replacing both wheels, repeat step 3 above on the other wheel, then proceed to the Rear Wheel
Installation Procedure on page 8 - 120 to install the second new rear wheel and complete the rear
wheel installation procedure.
WARNING: Do not lean on the machine, or leave it unattended without support, if a wheel has
been removed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
WARNING: Do not lean on the machine, or leave it unattended without support, if a wheel has
been removed.
WARNING: When BOTH rear wheels are being replaced, the machine must be safely supported
on two wooden blocks (or on a jack capable of supporting the weight of the machine). Always
remove and replace one wheel at a time, before attempting to remove the second wheel.
1) Having removed one rear wheel, position the new wheel on the chassis (in the same position as
that of the wheel removed).
2) Tighten the shaft screw (see Figure 8-105 below), making sure the wheel is firmly secured.
Shaft Screw
3.) If replacing both wheels, when the second rear wheel has been removed, repeat step 1 and step 2
above to install the second new rear wheel.
4) Return the Vivid™ 3 rear, left and right covers, as described in the Rear Cover Installation
Procedure on page 8 - 7 and the Side Covers Installation Procedure on page 8 - 4.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 8-7
Software Loading
8-7-1 Software Installation/Upgrade Procedure
Note: It is recommended to back-up the archive and the settings as described in the Vivid™ 3 Pro/
Vivid™ 3 Expert User Manual, before commencing software installation (or whenever
upgrading the software).
WARNING: Do not attempt to install software that was not designated for your Vivid™ 3 unit. Attempting
to install un-approved software will cause irreversible damage to hardware and software!
Normally the Vivid™ 3 system is supplied from the manufacturer with the software already installed. In
some cases, it is necessary to re-install the software (for example, after replacing the Back-end
Processor) by following the software installation procedure shown in Table 8-5 below.
Table 8-5 Software Installation
1 Press the <Eject> button of the CD drive and remove The CD drive tray ejects.
any media if present.
2 Insert the CD provided with the software kit, and press The CD drive tray will load the software CD.
the <Eject> button again.
3 Hold down the <ON/OFF> button for 5 seconds to The shutdown options are displayed on-screen.
shutdown the system.
5 Hold down the <ON/OFF> button for three seconds. The system powers ON and when system boot-up from the
CD is complete, the following message is displayed
prompting you to select the required option:
8 Follow the on-screen instructions, clicking OK (or The machine performs 4 reboots before completing the
Cancel) when prompted. installation. The installation process will take between
20 and 50 minutes - DO NOT ATTEMPT TO STOP THE
INSTALLATION BEFORE COMPLETION!
When the installation process is complete, the system will
be in regular display mode (Probe Select Screen).
Note: If this is a new software version, an adhesive Rating
Plate is supplied with the software CD. Adhere the plate in
the appropriate position on the metal cover under the
control console.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 8-8
Peripherals
8-8-1 B/W Video Printer Replacement Procedure
NOTE: This section describes the installation and removal procedure for the Sony UP-895MDW B/W Video
Printer.
8-8-1-1 Tools
Use the appropriate flat and Phillips-type screw drivers.
8-8-1-2 Manpower
One person - 30 mins.
8-8-1-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panel cover as shown in the Connector Panels Cover Removal Procedure
on page 8 - 8.
3.) Unpack the B/W Video Printer.
4.) Check that the B/W Video Printer AC voltage corresponds to that of your local voltage.
5.) Verify that all the DIP Switches on the rear panel of the B/W Video Printer are in the Down position
as shown in Figure 8-107 on page 8-124.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Figure 8-107 B/W Video Printer, DIP Switches on the Rear Panel
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
2.) Connect the Remote cable, the B/W Video IN cable, and the Power cable, as shown in Figure 8-
110 below.
Power cable
From: Right Connectors Panel, AC outlet
To: AC IN
NOTE: For shelf installation instructions refer to Color Printer installation in the Color Video Printer
Installation - Procedure1 section, page 8-157.
6.) Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation
Procedure on page 8 - 8.
7) Reconnect the mains AC power cable from the wall outlet.
8) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
9) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
10.)Select the activation key under System Configuration, VCR/ECG Tab, A/B Swap.
By default, Print B key will be B&W Video printer and Print A key will be Color Video printer. For
more information on switching printer settings, see VCR/ECG Tab on page 3 - 45.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Slide Switch
Settings on the
Paper Tray
Figure 8-111 B/W Video Printer, Slide Switches Location and Settings on the Paper Tray
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-8-2-1 Tools
No special tools needed.
8-8-2-2 Manpower
One person - 30 mins.
8-8-2-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Video Cassette
Recorder (VCR)
Figure 8-113 Video Cassette Recorder (VCR) Installed in the Vivid™ 3 Ultrasound Unit
Configuration with UP2950 Printer
Video Cassette
Recorder (VCR)
Figure 8-114 Video Cassette Recorder (VCR) Installed in the Vivid™ 3 Ultrasound Unit
(Configuration with UP21MD Color Printer and Shelf)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
2.) Connect one end of the VCR power cable (P/N: 2269430) to the power outlet on the Vivid 3 rear
panel and the other end to the AC power socket on the VCR rear panel - see Figure 8-115 on
page 8-130.
3.) Using the S-video cable (P/N: 2266743), connect one end to the Video OUT socket on the VCR
rear panel (refer to Figure 8-115) and the other to the Video IN socket on the Vivid 3 rear panel.
Audio IN Socket
AC Power Socket
6.) Press Config then, from the System Configuration window, activate the VCR / ECG tab.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
7.) Under the VCR Options list, choose Mitsubishi from the VCR model drop-down list (Figure 8-117).
8.) From the Video Mode drop-down list, choose PAL or NTSC (according to video system in your
region), then press OK to apply the changes.
9.) Proceed to the VCR Setup procedure, described below.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: The Mitsubishi VCR can be configured using the Menus on the front panel (Figure 8-118) as follows:
Jog Button
Enter Button
1.) Press the VCR Play B. button on the Vivid™ 3 control console to display the VCR menus on the
Vivid™ 3 monitor.
2.) Using the Jog button on the VCR front panel (Figure 8-118), select the required VCR menu
language (Figure 8-119), then press the Enter button to continue to the next menu.
3.) Select the <Menu1> settings, according to the configuration shown in Figure 8-120 below , then
press the Enter button to continue to the next menu.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
4.) Select the <Menu2> settings, according to the configuration shown in Figure 8-121 below , then
press the Enter button to continue to the next menu.
6.) Select the <RS-232C> settings, according to the configuration shown in Figure 8-123 below, then
press the Enter button to continue to the next menu.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
The SVO-9500MDP2 (PAL) and SVO-9500MD2 (NTSC) are alike, except for the Color Signal system.
8-8-3-1 Tools
No special tools needed.
8-8-3-2 Manpower
One person - 30 mins.
8-8-3-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
5.) Verify that the switches on the VCR front panel are as shown in Figure 8-127 below.
If necessary, adjust accordingly.
Not important.
Monitor Mix (Determines what is sent to MONITOR OUT jack.
Has no effect on AUDIO OUT).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
6.) Verify that the switches on the VCR rear panel are as shown in Figure 8-128 below. If necessary,
adjust accordingly.
1 2 3 4 5 6
OFF
OFF ON 4800
ON OFF 9500
ON ON 10200
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Video Cassette
Recorder (VCR)
Figure 8-129 Video Cassette Recorder (VCR) Installed in the Vivid™ 3 Ultrasound Unit
Configuration with UP2950 Printer
Video Cassette
Recorder (VCR)
Figure 8-130 Video Cassette Recorder (VCR) Installed in the Vivid™ 3 Ultrasound Unit
(Configuration with UP21MD Color Printer and Shelf)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
2.) Connect the RS-232C cable, the Audio and S-Video IN and OUT cables, and the Power cord, as
shown in Figure 8-131.
.
S-VIDEO IN cable
AUDIO OUT cable From: Left Connectors
From: Left Connectors Panel: Panel,
IN-R Connector (Red) S-Video OUT
IN-L Connector (Black) To: VCR S-Video IN
To: VCR Audio Output:
BLACK: CH1 RS-232 cable
From: Left Connectors Panel,
RED: CH2
Connector RS-232(1)
BLACK
BLACK
RED
To: VCR RS-232C Input
RED
AUDIO AUDIO
IN OUT
VIDEO IN
CH1
CH1
CH2 S S
VIDEO CH2 VIDEO
AC IN
Power cord
From: Left Connectors
Panel AC OUT
To: VCR AC IN
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: For shelf installation instructions refer to the Color Video Printer Installation - Procedure1 on
page 8 - 157.
6.) Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation
Procedure on page 8 - 8.
7) Reconnect the mains AC power cable from the wall outlet.
8) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
9) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
10.)Under System Configuration, VCR/ECG Tab, select the appropriate VCR mode options. For more
information, see VCR/ECG Tab on page 3 - 45.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
The SR-S388E (PAL) and SR-S365U (NTSC) models are alike, except for the Color Signal system.
Note: When installing the JVC VCR, use the cables supplied with the VCR and not those supplied
with the Vivid 3 system.
8-8-4-1 Tools
No special tools needed.
8-8-4-2 Manpower
One person - 30 mins.
8-8-4-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panel cover as shown in the Connector Panels Cover Removal Procedure
on page 8 - 8.
3.) Unpack the VCR and place it on a table (or firm surface).
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
5.) Proceed to the JVC VCR Setup Procedures on the next page.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Y/C TBC
RS-232C Hi Fi Select
1) Open the flip-down switch cover on the lower left side of the JVC front panel (by pulling downwards),
as seen in Figure 8-133, above.
2.) Check that the switch settings for each of the following switches is correct (as shown below).
If necessary, adjust accordingly:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: If you have an external monitor, you can connect it to the Monitor Out - Video socket on the VCR rear
panel (see Figure 8-134 below), which will enable you to view the different menus as they are being
edited for the BIOS setup. For best image quality during recording and playback, this is the
recommended setup procedure.
Connect External
Monitor Here
(if available)
1.) Connect the AC power cable to the AC Input socket on the VCR rear panel (upper left corner) and
plug the other end of the cable into the mains power wall outlet.
2.) Turn ON power to the VCR.
3.) If an external monitor is connected to the VCR, make sure power to the monitor is also ON.
4.) Press the Menu button on the JVC Front Panel - see Figure 8-135 below.
Menu Button
Select Button
Set Button
The LCD Display panel (refer to Figure 8-133 on page 8-142) becomes illuminated.
NOTE: If an external monitor is connected (see Note above), the Menu Switch Setting Screen is displayed (on
the external monitor). While making changes to the Menu Switch settings, these may be monitored in
this display.
If an external monitor is not connected, the Menu # and currently-selected value is only visible in the
LCD Display on the JVC VCR front panel.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
5.) Use the Select button (see Figure 8-135) to select MEMORY SW and then press the Set button.
6.) Using the Select and Set buttons, scroll through the various Menus and set the values for each of
the relevant Menus, in accordance with those shown in Table 8-6 below.
NOTE: Work in the same order as that given in Table 8-6 and do not change any settings other than those
listed.
NOTE: In Table 8-6, square brackets [ ] indicate factory presets. The numbers in parenthesis ( ) are seen on
the LCD display.
7.) When all the above-mentioned settings are correct, press the Menu button, select CLOCK ADJUST
and press the Set button. Using the Select and Set buttons, adjust the settings to the correct time.
When the correct time is visible in the LCD display, press the Menu button, to quit CLOCK ADJUST.
8.) Turn OFF power to the VCR. Remove the AC power cable from the mains power wall outlet and
disconnect the cable from the VCR rear panel.
If an external monitor was used for the BIOS setup procedure, turn OFF power to the external
monitor and disconnect the power cable (connected to the VCR rear panel Monitor Out - Video
socket - refer to Figure 8-134 on page 8-143).
NOTE: All settings are now stored in the VCR memory and the JVC VCR is ready for installation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
RS-232C
Video Input
Video Output
5.) Connect one end of the cable extender supplied with the VCR (see Figure 8-137 below) to the
AC power cable (already attached to the VCR).
NOTE: When connecting the VCR AC Power cable, always use the cable extender supplied with the VCR.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
6.) Connect the other end of the cable extender to the cable connector on the Vivid™ 3, as shown in
Figure 8-138 below (refer also to Figure 8-137, above).
7.) Carefully slide the VCR into the side compartment, and at the same time, route all the cables
through the gap between the compartment base and the right side of the compartment (as shown
in Figure 8-139 below) while gently pulling them in a downwards direction.
8.) Continue sliding the VCR all the way back into the final position with one hand, while gently pulling
the cables in a downwards direction. The VCR and attached cables are now positioned as shown
in Figure 8-140 below.
cables routed
VCR in position through gap
in compartment
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
9.) Check that the VCR fits properly in position and does not extend beyond the edge of the side
compartment - see Figure 8-141 below.
JVC VCR
NOTE: In order to eliminate the risk of the JVC VCR moving or falling out of the compartment whenever the
Vivid 3 ultrasound unit is moved, a removable L-shaped aluminium stopper (Part No. 2401662) is fitted
to the floor of the side compartment, as described below. This serves to secure the VCR firmly in
position.
10.)Make sure the surface of the side compartment is thoroughly clean and dry.
11.)Check that the outer sides of the L-shaped aluminium VCR stopper strip are clean and dry.
12.) Cut two pieces of self-adhesive velcro strip (supplied with the VCR), each approximately 8 cm (3 in)
in length.
13.)Holding one the velcro strips, remove the protective sheet from the self-adhesive side (hook
section) and carefully apply the strip to the outer side of the aluminium VCR stopper, placing it close
to the edge as shown in Figure 8-142 below. Press down to secure firmly in position.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
15.)Separate the upper (loop) section from each of the velcro strips and repeat step 13 to apply each
section to the floor of the side compartment, parallel to the lower edge of the VCR as shown in
Figure 8-143 below. Press down firmly to ensure the two sections are securely attached.
16.)Holding the aluminium VCR stopper with the velcro hook side facing downwards, fix it in position in
front of the VCR, taking care to properly align the velcro strips. Press down firmly to ensure the
stopper is securely fastened - refer to Figure 8-144 below.
17.)Bundle all the VCR cables together into a single loom and bind together with cable ties.
18.)With the control console still at the maximum height, route the cable loom down to the left rear
connectors panel (at the rear of the Vivid 3 unit).
19.)Fasten the loom to the Vivid™ 3 bottom cover using the square adhesive cable fastener (supplied).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
20.)Connect the RS-232C cable, the Audio and S-Video IN and OUT cables, and the Power cable to
the appropriate sockets on the Vivid™ 3 rear connectors panel - see Figure 8-145.
RS-232C
AUDIO IN
AUDIO OUT
21.)Return the console bottom metal cover (upper section) (removed in step 3). Make sure the VCR
cable loom is properly aligned to pass through the slot in the cover (as shown in Figure 8-146
below) and that there is no kinking or pinching of the cables.
Cable Loom
(routed through slot)
Figure 8-146 Routing Cable Loom through Slot in Console Bottom Cover (upper section)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
22.)Screw the cover firmly in position as described in the Control Console Bottom Cover (Upper
Section) Installation Procedure on page 8-19.
23.)Re-install the Connectors Panel Cover as described in the Connector Panels Cover Installation
Procedure on page 8-8.
24) Reconnect the mains AC power cable from the wall outlet.
25) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
26) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
27.)Under System Configuration, VCR/ECG Tab, select the JVC VCR option. For more information, see
VCR/ECG Tab on page 3-45.
28.)Insert a blank video casette tape into the VCR and perform a test recording to check correct
operation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: IMPORTANT - Before installation of either Panasonic VCR model, it is necessary to order the
Panasonic VCR RS-232 Adaptor (Part No. 066E8200).
8-8-5-1 Tools
No special tools needed.
8-8-5-2 Manpower
One person - 30 mins.
8-8-5-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
5.) Verify that the switches on the VCR front switch panel are as shown in Figure 8-148 on page 8-152.
If necessary, adjust accordingly.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Video Cassette
Recorder (VCR)
Figure 8-149 Panasonic Video Cassette Recorder (VCR) Installed in the Vivid™ 3 Ultrasound Unit
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
3.) Connect the Panasonic VCR RS-323 adaptor (Part No. 066E8200) to the RS-232C cable, and then
to the RS-323 socket on the rear of the VCR, as shown in Figure 8-131 below.
4.) Connect the Audio and S-Video IN and OUT cables, and the Power cable (also shown in Figure 8-131).
S-VIDEO IN cable
From: Internal IO Module, Connector A11
To: S-Video Input
BLA C K
RED
RS-232C Adaptor
Ground:
From the system chassis
BLACK
RED
Power cable
From: Distribution box
To: AC IN
AUDIO OUT cable
From: Internal IO Module: RED: Connector A7
BLACK: Connector A6
S-VIDEO OUT cable
To: S-Video Output: RED: CH1
From: Internal IO Module, Connector A10
BLACK: CH2
To: S-Video Output
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: For shelf installation instructions, see Color Video Printer Installation - Procedure1 on page 8 - 157.
8.) Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation
Procedure on page 8 - 8.
9) Reconnect the mains AC power cable from the wall outlet.
10) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
11) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
12.)Under System Configuration, VCR/ECG Tab, select the Panasonic VCR option, as shown in Figure
8-151 below. For more information, see VCR/ECG Tab on page 3 - 45.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-8-6 Sony UP 2950 MD & 2800P Color Video Printer Replacement Procedure
NOTE: This section describes the installation and removal procedures of the Sony UP-2800P and UP/2950MD
Color Video Printers.
The Sony UP-2950MD Color Video Printer has a NTSC/PAL Selector on the rear panel while the Sony
UP-2800P Color Video Printer is PAL only.
• For units that are equipped with a previously-installed B/W printer and/or VCR
(see Color Video Printer Installation - Procedure1 on page 8 - 157, and Color Video Printer Removal
- Procedure 1 on page 8 - 161).
• For units that are not equipped with a previously-installed B/W printer and/or VCR
(refer to Color Video Printer Installation - Procedure 2 on page 8 - 162, and Color Video Printer
Removal - Procedure 2 on page 8 - 162).
8-8-6-1 Tools
Use the appropriate flat and Phillips-type screw drivers.
8-8-6-2 Manpower
One person - 30 mins.
8-8-6-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Unpack the Color Video Printer.
4.) Check that the Color Video Printer AC voltage corresponds to that of your local voltage.
5.) Set the NTSC/PAL selector switch to NTSC (this applies only to Model UP-2950MD) as shown in
Figure 8-158 on page 8-160.
6.) Install the Color Video Printer accessories, as described in the Operating Instructions supplied with
the printer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: The Installation Procedure is for units that are equipped with a previously-installed B/W printer and/or
VCR. The color printer option includes a shelf with fastening belt, on which the printer is placed.
NOTE: If a B/W Printer and VCR are not installed, then the Color Video Printer can be placed on the top surface
of the Vivid™ 3 ultrasound unit (below the keyboard) and secured with the belt.
Note: Make sure that all the cables are fitted through the slot in the far end of the Metal Bottom
Cover, as shown in Figure 8-153 below.
Figure 8-153 Color Printer Cables through Metal Bottom Cover Slot
8) Remove the right and left screws located at the front of the lower section top cover, as shown in
Figure 8-154 on page 8-158.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
9) Place the shelf on the top surface, with its two legs facing the front, and the stoppers (located on
top of the shelf) facing the far end. Fasten the shelf to the top surface, using the screws removed
in step 8, as shown in Figure 8-155.
10) Route the left end of the top surface belt through the narrow left slot in the shelf (closest to the front),
working from below the shelf, upwards.
11) Route the right end of the top surface belt through the narrow right slot in the shelf (closest to the
front), working from below the shelf, upwards.
12) Route the two sections of the additional belt (supplied with the Color Printer Shelf), through the
second pair of (outer) right and left slots, as shown in Figure 8-155, above.
13) Return the B/W printer and VCR to their original positions, between the left and right sections of the
belt.
14) Lock the belt buckles to tighten the shelf to the top surface. Verify that the B/W printer and VCR are
secure.
8-158 Section 8-8 - Peripherals
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Figure 8-156 Color Video Printer and Shelf Installed on the Vivid™ 3
17.)Connect the Ground cable (hanging out of the Bottom Metal Cover) to the Ground screw at the rear
of the Color Video Printer, as shown in Figure 8-157 below
Ground Screw
Ground Cable
18) Connect the RGB and SYNC cables, Color Printer Remote cable, and AC power cable (hanging out
of the Bottom Metal Cover slot - see Figure 8-153 on page 8-157), to the Color Video Printer, as
shown in the diagram in Figure 8-158 on page 8-160.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Model
UP-2950MD
only
Power cord:
To: Color Video Printer
AC IN
19.)Position the Color Video Printer against the rubber stoppers at the rear of the shelf.
20.)Connect the buckles and tighten the belt. Verify that the Color Video Printer is secure.
21.)Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation
Procedure on page 8 - 8.
22) Reconnect the mains AC power cable from the wall outlet.
23) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
24) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
25.)Select the activation key under System Configuration, VCR/ECG Tab, A/B Swap.
By default, Print B key will be B&W Video printer and Print A key will be Color Video printer. For
more information on switching printer settings, see VCR/ECG Tab on page 3 - 45.
See also the Color Video Printer Settings section, page 8-163.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: This procedure (Procedure 1) is for units that are equipped with a previously-installed B&W printer and
a VCR. The color printer option includes a metal shelf and fastening belt on which the color printer is
placed.
1) Unlock the belt buckles located on either side of the color printer.
Belt Buckle
2) Move the color printer forwards, to gain access to the rear panel.
3) Open the lower middle screw from the color printer’s rear panel and remove the Ground cable lug.
4) Unplug the AC cable.
5) Unplug the BNC connectors R-G-B and SYNC from the color printer.
6) Remove the color printer from the top shelf.
Note: Perform the following steps only if the metal shelf must be removed.
7) Remove the right and left screws located in the legs on either side of the shelf, and remove the shelf.
Note: Do not perform the following step if the metal shelf will later be replaced.
8) Return and fasten the two screws (removed in step 7) to either side of the top cover.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: This procedure (Installation Procedure 2) is for units that are not equipped with a previously-installed
B&W printer and VCR. It describes all steps required to install a color printer. The color printer is placed
on the top cover of the unit’s lower section and the additional shelf and belt supplied with the color
printer are not required.
1) Unlock the belt buckles located on the lower section of the unit’s top cover, and move the belt parts
to the sides.
2) Place the color printer on the lower section of the top cover.
3) Connect the cable’s BNC connectors R-G-B and SYNC to the color printer R-G-B and SYNC Input
connectors.
4) Plug the color printer remote cable to the color printer Remote 2 Input.
5) Plug the AC cable to the color printer AC IN socket.
6) Screw the Ground cable lug to the lower middle screw on the color printer’s rear panel.
7) Set the NTSC/PAL switch in accordance with your local color system.
8) Lock the belt buckles over the color printer, and tighten the belt by pulling the remaining length of
the belt to fasten the printer securely in position.
9) Reconnect the mains AC power cable from the wall outlet.
10) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
11) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
12.)Select the activation key under System Configuration, VCR/ECG Tab, A/B Swap.
By default, Print B key will be B&W Video printer and Print A key will be Color Video printer. For
more information on switching printer settings, see VCR/ECG Tab on page 3 - 45.
See also the Color Video Printer Settings section, page 8-163.
NOTE: This procedure (Removal Procedure 2) is for units that are not equipped with a previously-installed
B&W printer and a VCR.
1) Unlock the belt buckles located on either side of the color printer, as shown in Figure 8-159 on
page 8-161.
2) Move the color printer forward to access its rear panel.
3) Open the lower middle screw on the color printer’s rear panel and remove the Ground cable lug.
4) Unplug the AC cable.
5) Unplug the cable's BNC connectors R-G-B and SYNC from the color printer RGB and SYNC Input
connectors.
6) Remove the color printer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: The settings for the Sony UP-2800P and UP2950MD Color Video Printers are regulated using the Video
Printer Control Panel, as described below.
8-8-6-8-1 Preparations
1.) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
2) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
3.) Open the front door of the Color Video Printer.
4.) Press MENU.
The Preset menu is displayed (see Figure 8-160 below).
There are five setup categories: COLOR (c), LAYOUT (l), PRINTER (p), INPUT (i) and OUTPUT
(o). Each setup category can be selected from the Top menu of each category.
Figure 8-160 Color Video Printer Control Panel and Printer Window Display
1.) Select COL using the or button (see Figure 8-160, above). The COLOR ADJUST top menu
is displayed on the Printer Window display (Figure 8-161 below).
Figure 8-161 Color Video Printer Window Display (COLOR ADJUST top menu displayed)
2.) In the COLOR ADJUST menu, adjust the required parameters (according to those shown in
Table 8-8 on page 8-164) by selecting the parameters one-by-one, using the or button; for
each selected parameter, adjust the value using the or button.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
LIGHT 0 +2
3.) Press EXEC on the Printer Control Panel to save the Color settings.
1.) Return to the COLOR ADJUST top menu (Figure 8-161 on page 8-163)), using the button.
2.) Select LAY by pressing the or button. The LAYOUT SETUP top menu is displayed (Figure
8-162 below).
Figure 8-162 Color Video Printer Window Display (LAYOUT SETUP top menu displayed)
3.) In the LAYOUT SETUP menu, select WINDOW by pressing the or button.
4.) Press the button to display the WINDOW SETUP menu (Figure 8-163 below).
Figure 8-163 Color Video Printer Window Display (WINDOW SETUP top menu displayed)
5.) In the WINDOW SETUP menu, adjust the required parameters (in accordance with those shown in
Table 8-8 below) by selecting the parameters one-by-one using the or button; for each
selected parameter, adjust the value using the or button.
:
Table 8-8 WINDOW SETUP Parameters
V START -8 -8
1.) Return to the WINDOW SETUP top menu using the button (Figure 8-163, above).
2.) Press the button to display the LAYOUT SETUP menu (Figure 8-162).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
3.) Select PRN by using the or button. The PRINTER SETUP top menu is displayed:
Figure 8-164 Color Video Printer Window Display (PRINTER SETUP top menu displayed)
4.) In the PRINTER SETUP menu, select FUNCTION using the or button.
Figure 8-165 Color Video Printer Window Display (FUNCTION SETUP top menu displayed)
6.) In the FUNCTION SETUP menu, adjust the parameters in accordance with those shown in Table 8-9
below, by selecting the parameters one-by-one using the or button; for each selected
parameter, adjust the value using the or button.
1.) Return to the PRINTER SETUP top menu using the button (Figure 8-164, above).
2.) Select IN by using the or button. The INPUT SETUP top menu is displayed:
Figure 8-166 Color Video Printer Window Display (INPUT SETUP top menu displayed)
3.) In the INPUT SETUP menu, select INPUT SEL using the button.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
1.) Return to the INPUT SETUP top menu using the button (Figure 8-166).
2.) Select OUT by using the or button. The OUTPUT SETUP top menu is displayed:
Figure 8-167 Color Video Printer Window Display (OUTPUT SETUP top menu displayed)
3.) In the OUTPUT SETUP menu, adjust the parameters in accordance with those shown in Table 8-
10 below, by selecting the parameters one-by-one using the or button; for each selected
parameter, adjust the value using the or button.
SOURCE EE
4.) Press EXEC on the control panel of the printer to save the settings.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
The Sony UP-21MD Color Video Printer has an NTSC/PAL Selector on the rear panel.
• For units that are equipped with a previously-installed B/W printer and/or VCR
(see Color Video Printer Installation - Procedure1 on page 8 - 157, and Color Video Printer Removal
- Procedure 1 on page 8 - 161).
• For units that are not equipped with a previously-installed B/W printer and/or VCR
(refer to Color Video Printer Installation - Procedure 2 on page 8 - 162, and Color Video Printer
Removal - Procedure 2 on page 8 - 162).
8-8-7-1 Tools
Use the appropriate flat and Phillips-type screw drivers.
8-8-7-2 Manpower
One person - 30 mins.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-8-7-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Unpack the Color Video Printer.
4.) Set the NTSC/PAL selector switch to PAL or NTSC as shown in Figure 8-166 on page 8-165.
5.) Install the Color Video Printer accessories, as described in the Operating Instructions supplied with
the printer.
NOTE: The Installation Procedure is for units that are equipped with a previously-installed B/W printer and/or
VCR. The color printer option includes a shelf with fastening belt, on which the printer is placed.
NOTE: If a B/W Printer and VCR are not installed, then the Color Video Printer can be placed on the top surface
of the Vivid™ 3 ultrasound unit (below the keyboard) and secured with the belt.
Note: Make sure that all the cables are fitted through the slot in the far end of the Metal Bottom
Cover, as shown in Figure 8-169 below.
Figure 8-169 Color Printer Cables through Metal Bottom Cover Slot
8-168 Section 8-8 - Peripherals
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8.) Remove the right and left screws located at the front of the lower section top cover, as shown in
Figure 8-170.
9) Place the protective metal cover on top of the printer shelf (either on the right or left side, as
preferred by the customer) and fasten securely into position to protect the connectors, using the
two screws provided - see Figure 8-171 below:
Screw Screw
Metal Cover
Figure 8-171 Printer Shelf with Protective Metal Cover Secured in Position
10) Place the shelf on the top surface, with its two legs facing the front, and the stoppers (located on
top of the shelf) facing the far end. Fasten the shelf to the top surface, using the screws removed
in step 8, as shown in Figure 8-172 below.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
11.)Route the end of one of the top surface securing belts down through the narrow left slot in the shelf
(closest to the front), then thread it below the shelf, backwards and upwards through the slot at the
rear of the shelf. Leave the opposite end of the belt (with the plastic buckle [female end] attached)
dangling towards the front, as shown in Figure 8-171 and Figure 8-172 on page 8-170.
12) Repeat step 11 to thread the second belt in position through the slot on the right.
13) Attach the male end of the two plastic buckles to each of the belt ends protruding from the rear right
and left slots.
14) Return the color printer and VCR to their original positions, between the left and right sections of the belt.
15) Lock the belt buckles to tighten the shelf to the top surface. Verify that the printer and VCR are secure.
16.)Re-connect the previously-disconnected Color Printer and VCR cables.
17) Place the UP-21MD Color Video Printer on top of the shelf, as shown in Figure 8-173 below.
UP-21MD
Color Video Printer
Figure 8-173 Sony UP-21MD Color Video Printer and Shelf Installed on the Vivid™ 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
18.)Connect the Ground cable (hanging out of the Bottom Metal Cover) to the Ground screw at the rear
of the UP-21MD Color Video Printer, as shown in Figure 1-17 below
Ground Cable
Ground Screw
19) Connect the RGB and SYNC cables, Color Printer Remote cable, and AC power cable (hanging out
of the Bottom Metal Cover slot - see Figure 8-169 on page 8-168), to the UP-21MD Color Video
Printer, as shown in the diagram in Figure 8-175 on page 8-172.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Power cord:
To: UP-21MD Color
Video Printer AC IN
NTSC/PAL
selector
REMOTE cable:
To: UP-21MD Color Video
Printer REMOTE 2 Connector
20.)Position the Color Video Printer against the rubber stoppers at the rear of the shelf, and against the
metal protective cover on the side (making sure that the connectors are not exposed).
21.)Connect the two plastic buckles and tighten the belts. Verify that the UP-21MD Color Video Printer
is secure.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
22.)Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation
Procedure on page 8 - 8.
23) Reconnect the mains AC power cable from the wall outlet.
24) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
25) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
26.)Select the activation key under System Configuration, VCR/ECG Tab, A/B Swap.
By default, Print B key will be B&W Video printer and Print A key will be Color Video printer. For
more information on switching printer settings, see VCR/ECG Tab on page 3 - 45.
NOTE: This procedure (Procedure 1) is for units that are equipped with a previously-installed B&W printer and a
VCR. The color printer option includes a metal shelf and fastening belt on which the color printer is placed.
Belt Buckles
2.) Move the color printer sideways, to gain access to the rear panel.
3.) Open the lower middle screw from the color printer’s rear panel and remove the Ground cable lug.
4.) Unplug the AC cable.
5.) Unplug the BNC connectors R-G-B and SYNC from the color printer.
6.) Remove the color printer from the top shelf.
Note: Perform the following steps only if the metal shelf must be removed.
7.) Remove the right and left screws located in the legs on either side of the shelf, and remove the shelf.
Note: Do not perform the following step if the metal shelf will later be replaced.
8.) Return and fasten the two screws (removed in step 7) to either side of the top cover.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: This procedure (Installation Procedure 2) is for units that are not equipped with a previously-installed
B&W printer and VCR. It describes all steps required to install a color printer. The color printer is placed
on the top cover of the unit’s lower section and the additional shelf and belt supplied with the color
printer are not required.
1) Unlock the belt buckles located on the lower section of the unit’s top cover, and move the belt parts
to the sides.
2) Place the color printer on the lower section of the top cover.
3) Connect the cable’s BNC connectors R-G-B and SYNC to the color printer R-G-B and SYNC Input
connectors.
4) Plug the color printer remote cable to the color printer Remote 2 Input.
5) Plug the AC cable to the color printer AC IN socket.
6) Screw the Ground cable lug to the lower middle screw on the color printer’s rear panel.
7) Set the NTSC/PAL switch in accordance with your local color system.
8) Lock the belt buckles over the color printer, and tighten each belt by pulling the remaining length of
the belt to fasten the printer securely in position.
9) Reconnect the mains AC power cable from the wall outlet.
10) Turn ON the main circuit breaker located on rear of the Vivid™ 3.
11) Turn ON the Vivid™ 3 by pressing the ON/OFF button on the Control Console.
12.)Select the activation key under System Configuration, VCR/ECG Tab, A/B Swap.
By default, Print B key will be B&W Video printer and Print A key will be Color Video printer. For
more information on switching printer settings, see VCR/ECG Tab on page 3 - 45.
NOTE: This procedure (Removal Procedure 2) is for units that are not equipped with a previously-installed
B&W printer and a VCR.
1) Unlock the belt buckles located on either side of the color printer, as shown in Figure 8-176 on
page 8-173.
2) Move the color printer sideways to access its rear panel.
3) Open the lower middle screw on the color printer’s rear panel and remove the Ground cable lug.
4) Unplug the AC cable.
5) Unplug the cable's BNC connectors R-G-B and SYNC from the color printer RGB and SYNC Input
connectors.
6) Remove the color printer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
The HP6540/3 Printer Kit (P/N: 2415435) contains the following items:
8-8-8-1 Tools
Use the appropriate flat and Phillips-type screw drivers.
8-8-8-2 Manpower
One person - 15 mins.
8-8-8-3 Preparations
1) Shut down the Vivid™ 3 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Carefully unpack the HP 6540/3 USB Printer.
4.) Check that the HP 6540/3 USB Printer AC voltage corresponds to that of your local voltage.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Transformer
To Vivid 3 rear connectors panel
To Power-IN Socket
USB Socket
WARNING: To avoid electrical hazards, the HP 6540/3 USB Deskjet Color Printer must be
connected to the AC Supply of the Vivid™ 3 Ultrasound Unit and never directly to the local AC
wall outlet.
3.) Use the USB cable (shipped with the printer) to connect the Vivid 3 system to the USB socket on
the HP6540/3 rear panel (see Figure 8-178).
4.) Turn ON power to the system and the printer and boot up the system in User mode.
5.) Wait until the application finishes loading and then press <Alt> +<Tab> simultaneously.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
6.) When the Found New Hardware Wizard opens, click Cancel, see Figure 8-179 below.
7.) Insert the HP6540/3 driver installation CD (P/N:2415381) provided with the printer into the CD-ROM
drive.
8.) Browse the CD (Drive H:) and double-click on the InstallPrinter.bat file (Figure 8-180)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
10.)When prompted, verify that Local Printer is selected. If not, activate the Local printer radio button
as shown in Figure 8-183.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
11.)Click Next.
12.)From the displayed list of available ports, choose one of the available USB ports, as shown in
Figure 8-184.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
14.)From the list of Printers, highlight the HP Deskjet 6500 Series - see Figure 8-185.
15.)Click Next.
16.)When prompted, choose Yes if you want this printer to be the default printer - see Figure 8-186.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
18.)When the following message appears, verify that Do not share this printer is selected, as shown
in Figure 8-187.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
The Printer Properties Dialog opens with the Advanced Tab activated, as shown in Figure 8-193.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
26.)Set up the printing defaults (Paper size, Paper type etc....) to your liking, then click OK.
The following message is displayed:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
29.)Confirm that HP Deskjet 6500 Series has been added to Currently Installed Printers list, then
click OK.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8-8-8-6 HP 5652 Deskjet Color Printer Installation Procedure - Vivid 3 BT00 Systems
NOTE: Since Vivid3 BT00 systems are based on the Windows NT operating system, they do not support
the Plug & Play feature.
Therefore, the HP990c driver (already installed on the system) can also be used for the
HP 5652 Deskjet Color Printer, without requiring the installation of any additional printer drivers. All
that is necessary is to set the HP990c as the default printer, as described in the procedure below.
1.) Place the HP 5652 Deskjet Color Printer on the Vivid 3 bottom shelf allowing sufficient room to
access the cables at the rear of the printer, or on a stable surface located not far from the
Vivid 3 scanner.
2.) Connect one end of the Printer power cable to the Vivid 3 rear connectors panel and the other end
(via the transformer [supplied]) to the Low Voltage Power-IN socket on the rear of the printer, as
shown in Figure 8-178 below.
Parallel Port
Transformer
To Power-IN Socket
To Vivid 3 rear connectors panel
WARNING: To avoid electrical hazards , the 5652 HP Deskjet Color Printer must be connected to
the AC Supply of the Vivid™ 3 Ultrasound Unit and never directly to the local AC wall outlet.
3.) Turn ON power to the system and the printer and boot up the system in User mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
4.) Press the Config button on the control console, then from the System Configuration window,
activate the System tab (Figure 8-198).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
6.) Highlight hp deskjet 990c series as shown above, then click on the Set as default button.
7.) Click on the Printing settings button to access the Printer Properties dialog box:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
8.) Set up the printing defaults (Paper size, Paper type etc....) to your liking, then click OK.
The System Configuration window is re-displayed (refer to Figure 8-198 on page 8-188).
9.) Click OK to close the window.
8-8-8-7 HP 5652 Deskjet Color Printer Removal Procedure - Vivid 3 BT00 Systems
1) Turn OFF power to the HP 5652 printer.
2) Unplug the AC power cable from the Power IN socket at the rear of the printer.
3) Unplug the parallel cable from the parallel port at the rear of the printer.
4) Remove the HP 5652 printer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Chapter 9
Renewal Parts
Section 9-1
Overview
9-1-1 Purpose of Chapter 9
This chapter gives you an overview of replacement parts for the Vivid™ 3 ultrasound unit.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 9-2
List of Abbreviations
• Assy - Assembly
• BEP - Back End Processor
• Ctrl - Control
• FEP - Front End Processor
• FRU 1 - Replacement part available in parts hub
• FRU 2 - Replacement part available from the manufacturer (lead time involved)
• LCD - Liquid Crystal Display
• Int - Internal
• I/O - Input/Output
• PWA - Printed Wire Assembly
• Recv - Receive
• XFRMR - Transformer
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 9-3
Renewal Parts Lists and Diagrams
9-3-1 Mechanical Hardware Parts
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Part 2258600 - Cover Right Pro Part 2258602 - Cover Top (Lower Section)
Part 2261572 - Cover Top (Upper Section) Part 2261573 - Cover Bottom Rear (Upper Section)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Part 2261576 - Probe Holder Left Part 2261577 - Probe Holder Right
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Part 2389102 - Caster Wheel (Rear) Part 2392789 - Caster Wheel (Front) with brake
Part 2253013 - Rear Connector Panel Right Part 2253008 - Rear Connector Panel Left
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
P/N Part Name FRU Refer to
BT03
2258594 Cover Rear 2 Figure 9-1 on page 9-3
2370002-2 Up/Down Handle (plastic, with support) 2 Figure 9-1 on page 9-3
2261573 Cover Bottom Rear (Upper Section) 2 Figure 9-2 on page 9-4
2261574 Cover Bottom Front (Upper Section) 1 Figure 9-2 on page 9-4
2392789 Wheel (Castor) - Front (with brake) 1 Figure 9-4 on page 9-6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
P/N Part Name FRU Refer to
BT03
2253013 Rear Connector Panel Right (with cables) 2 Figure 9-4 on page 9-6
2253008 Rear Connector Panel Left (with cables) 2 Figure 9-4 on page 9-6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
P/N Part Name FRU Refer to
BT03
2252994 Input AC Box and Circuit Breaker 1 Figure 9-5 on page 9-9
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Part 2253033-2 - PCB Front Board and TR4 Assembly (FB, 2Con.,V7)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Part 2253033-3 - PCB Front Board and TR4 Assembly (FB, 3Con.,V7)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
P/N Part Name FRU Refer to
BT03
2253030-2 Front End Crate (including back plan MB) 2 Figure 9-6 on page 9-11
2295971 FAN 92x92 Crate Fan 12V 1 Figure 9-6 on page 9-11
2270068 Low Voltage Power Supply (LVPS) 1 Figure 9-6 on page 9-11
1 a a
FC200507 PCB Radio Frequency Interface (RFI) Figure 9-8 on page 9-13
FA200764 PCB Front End Controller (FEC) 1 Figure 9-7 on page 9-12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Part 2336022-2 Samsung 15" Monitor (591-S) for Pro Part 2306835-6 Monitor for Expert 17"
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
BT03
P/N Part Name FRU Refer to
2340279 PC Video Encoder Control Assembly (PC-VIC) 1 Figure 9-15 on page 9-18
2256476 ECG Box Assembly (Patient I/O Module) 1 Figure 9-12 on page 9-15
2316279 PCB Plug & Scan Card 1 b b Figure 9-13 on page 9-16
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
BT03
P/N Part Name FRU Refer to
1 e e
FC200656 PCB PC2IP2 Card Figure 9-15 on page 9-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
9-3-5 Cables
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
J7 J8
BRN
J7 J8
~
/
J7/8
BLU
J7/8/
0
J7 J8
PHERI.-1/2 RED
PHERI.-1/1 RED
PHERI.-2/2 BLACK
PHERI.-2/1 BLACK
INTER./2 ORANGE
INTER./1 ORANGE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Part 2292972 Network Back End Rear Panel to Isolation Box (option)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
BT03
P/N Part Name FRU Refer to
2252966 Cable: Input Box / AC Distribution Box 2 Figure 9-21 on page 9-21
2252972 Cable: Fan (AC Dist. Box / Front End) 2 Figure 9-22 on page 9-22
2252976 Cable: Keyboard/ VIC External Power 2 Figure 9-22 on page 9-22
2252981 Cable: Keyboard/ VIC Internal Power/ Back-End 2 Figure 9-23 on page 9-23
2252978 Cable: Trackball / Back End Matrix Panel 2 Figure 9-23 on page 9-23
2252979 Cable: ECG - Back End Internal 2 Figure 9-23 on page 9-23
2252980 Cable: Patient I/O int. ECG - ECG/Phono 2 Figure 9-23 on page 9-23
2252982 Cable: ECG Int. Footswitch / Keyboard Control 2 Figure 9-23 on page 9-23
2253009 Cable: Com1: Back End / Rear Panel Left 2 Figure 9-24 on page 9-24
2253010 Cable: Ground / Rear Panel Left 2 Figure 9-24 on page 9-24
2253011 Cable: VIC / Rear Panel Left Col. Pr. 2 Figure 9-24 on page 9-24
2371429 Cable: Video / BEP rear (Frame Grabber) 1 a a Figure 9-21 on page 9-21
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
BT03
P/N Part Name FRU Refer to
2253014 Cable: AC Dist. / Rear Panel Right 2 Figure 9-24 on page 9-24
2253016 Cable: Parallel Port / Rear Panel 2 Figure 9-25 on page 9-25
2253019 Cable: Ground/ Rear Panel Right 2 Figure 9-25 on page 9-25
2253079 Cable: Rear Panel Left / B/W Printer 2 Figure 9-25 on page 9-25
2253080 Cable: B/W Printer / Rear Panel Left (Print Trig) 2 Figure 9-25 on page 9-25
2256477 Cable: ECG BEP Cable USA - External 1 Figure 9-25 on page 9-25
2256478 Cable: ECG BEP Cable Europe - External 1 Figure 9-25 on page 9-25
2256685 Cable: External ECG (with Echo Stress Option) 2 Figure 9-25 on page 9-25
2258801 Cable: Internal Audio / Keyboard Control Back End 2 Figure 9-26 on page 9-26
2258803 Cable: External Audio Back End / Rear Panel 2 Figure 9-26 on page 9-26
2258805 Cable: ON/OFF Motherboard / Keyboard Controller 2 Figure 9-26 on page 9-26
066E8200 RS-232 Adaptor - for Panasonic VCR 1 Figure 9-27 on page 9-27
2266742 Cable: RS232 VCR / Rear Panel - for Sony VCR 2 Figure 9-27 on page 9-27
2266744 Cable: Video OUT / Rear Panel 2 Figure 9-27 on page 9-27
2266745 Cable: Audio OUT / Rear Panel 2 Figure 9-27 on page 9-27
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Pro BT03
BT03
P/N Part Name FRU Refer to
2276874 Cable: PC2IP (Data Front End to Back End) 2 Figure 9-28 on page 9-28
2283903 Cable: Network Rear Panel Right to Isolation Box 1 Figure 9-29 on page 9-29
2300857 Cable: Power Peripherals and Printer 2 Figure 9-29 on page 9-29
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
9-3-6 Software
Pro BT03
P/N Part Name FRU
BT03
S/W Kit - contains s/w for systems:
2415386 1
Vivid™ 3, Vivid™ 4, and Vivid-i™
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
9-3-7 Probes
Pro BT03
BT03
P/N Part Name FRU
H4000P
Probe: 7S Sector 1
H40422LB
H40422LA
Probe: 5S Sector 1
H4901RA
H45521DX Probe: 6T OR 1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 9-4
Peripherals
BT03
Pro BT03
P/N Part Name FRU
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 9-5
Cabling Block Diagrams
SYSTEM MONITOR
17” 2306835-6
15” 2336022-2
COLOR
2269429 PRINTER
Speakers
2269309 R/G/B/ Sync
MONITOR
BRACKET
2269431
2269432
Track Color Printer Trig.
K.B Assy Ball K/B A/N K/B Matrix From Ac
2253043 On/off
From Ac System
System
2252978
BACK END 2315842-7
2252977
Mouse(PS2)
FOOT
PEDAL K.B (PS2)
006E007 ECG assy 2269422
Pci K/B. & Audio Ctrl 2253027-5
2252282 2252982 2252975
K/B Ctrl
2252979
Soft
S/D
2307289 &12V
2300847
FANS contactors
2253004
2253002
2256479
2253001
2295791 X 5
LAMBDA
PS
12VDC
Temp 2252999
Sens.
AC DIST. BOX
PHASED Beam
(64) (64) Former FC 200507
/LINEAR 2253038-3
PRO FA200900
ARRY
2253033-2
PROBES EXPERT
2253033-3 2252966
FA200114
TXPS
TR4-1
PENCIL C.B.
TR4-16
PROBE
AC
INPUT
Data Bus Data Bus MLA Bus BOX L.F.
2252994
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
SYSTEM MONITOR
17“ 2306835-6
15” 2336022-2
COLOR
2269429 PRINTER
Speaker
R/G/B/ Sync
2269309 MONITOR
BRACKET
2269431
2269432
Track Color Printer Trig.
K.B Assy Ball K/B A/N K/B Matrix From Ac
2392754 On/off
From Ac System
System
2252978
BEP 2315842-7 Basic
2252977
Mouse(PS2)
FOOT
PEDAL K.B (PS2)
066E0007 ECG assy 2256476
Pci K/B. & Audio Ctrl 2253027-5
2252982 2252975
2252282
K/B Ctrl
2252979
2300847
FANS contactors
2253004
2253002
2256479
2253001
2295971 X 5
LAMBDA
PS
12VDC
AC DIST. BOX
Front BF64
Mux Fec2 Fec Rft Image 2276875 2253000
ISOL.
Board Beam
(64) Port TRAN.
LVPS
2270068
TR4-1
PENCIL C.B.
TR4-16
PROBE
AC
INPUT
L.F.
BOX
Pipe link 2252994
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Chapter 10
Periodic Maintenance
Section 10-1
Overview
10-1-1 Purpose of Chapter 10
This chapter describes Periodic Maintenance (PM) procedures for the Vivid™ 3 ultrasound scanner and
its peripherals. These PM procedures are designed to protect your investment in the Vivid™ 3 system,
and to ensure constant operation at maximum efficiency.
After delivery and installation of the Vivid™ 3 ultrasound unit, your service representative will contact
you to schedule the first of several routine service visits. In addition to routine maintenance by trained
Field Service Engineers (FSEs), it is recommended that the procedures described in this chapter be
observed as part of your own internal routine maintenance program.
Table 10-1 Contents in Chapter 10
10-1-2 Warnings
DANGER: There are several places on the backplane, the AC distribution box, and the
DC distribution box that could be dangerous. Be sure to disconnect the system power plug and
to open the main circuit breaker before you remove any parts. Proceed with caution whenever
power is ON and covers are removed.
DANGER: Do not pull out or insert circuit boards while mains power to the system is ON.
CAUTION: Practice good ESD prevention. Wear an anti-static strap when handling electronic
parts and when disconnecting/connecting cables.
CAUTION: Do NOT operate the unit unless all board covers and frame panels are securely in
place, to ensure optimal system performance and cooling.
When covers are removed, EMI may be present.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-2
Why Perform Periodic Maintenance Procedures?
10-2-1 Keeping Records
It is good business practice that ultrasound facilities maintain records of periodic and corrective
maintenance. The Ultrasound Periodic Maintenance Inspection Certificate provides the customer with
documented confirmation that the Vivid™ 3 ultrasound scanner is regularly maintained on a periodic
basis.
A copy of the Ultrasound Periodic Maintenance Inspection Certificate should be kept in the same room
as the Vivid™ 3 ultrasound scanner, or nearby.
Quality Control testing of the system must be conducted routinely. The same tests are performed
regularly during each period, so that changes can be monitored over time and effective corrective action
taken, if required.
The results of Quality Control testing, corrective action, and the effects of corrective action, must be
documented and maintained on site.
Your GE Medical Systems Service Representative can help you with establishing, performing and
maintaining records for a Quality Assurance program.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-3
Periodic Maintenance Schedule
10-3-1 How Often Should PM Procedures be Performed?
The Periodic Maintenance Schedule (provided in Table 10-2 on page 10-4) specifies how often the
Vivid™ 3 ultrasound scanner should be serviced, and outlines items requiring special attention. Annual
periodic maintenance inspections are recommended. Each PM inspection requires approximately
2 hours for completion, excluding time needed for corrective action.
NOTE: It is the customer’s responsibility to ensure the periodic maintenance procedures are performed on the
Vivid™ 3 ultrasound scanner as recommended in the Periodic Maintenance Schedule, in order to retain
the high levels of safety, dependability, and system performance.
Your GE Medical Systems Service Representative has an in-depth knowledge of your Vivid™ 3
ultrasound scanning system and can best provide competent, efficient service. Please contact us for
further information, and to schedule routine maintenance servicing by our qualified professional FSEs.
The service procedures and recommended frequencies listed in the Periodic Maintenance Schedule
assumes use of the Vivid™ 3 scanner for an average patient load (10-12 patients per day). However,
this is not applicable if the system is used as a primary mobile unit which is transported between
diagnostic facilities.
NOTE: If conditions exist whereby typical usage and patient load is exceeded, it is strongly recommended to
increase the periodic maintenance frequencies.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Surgical Probe
As prescribed in Probe Manual.
Current Leakage Checks
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
DANGER: To avoid electrical shock hazard, do not remove the panels or covers from the
machine. This servicing must be performed by qualified service personnel only. Failure to
comply, could result in serious injury!
DANGER: If any defects are observed or any system malfunctions occur, DO NOT operate the
equipment. Inform a qualified service engineer immediately.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-4
Tools Required
10-4-1 Special Tools, Supplies and Equipment
46–194427P278 120V
Anti Static Vacuum Cleaner
46–194427P279 230V
Safety Analyzer 46–285652G1 DALE 600 KIT (or equivalent) for electrical tests
E7010GG 60 minutes
SVHS VCR Cassette
E7010GF 120 minutes
SVHS VCR Head Cleaner See VCR user manual for requirements
B/W Printer Cleaning Sheet See printer user manual for requirements
Color Printer Cleaning Sheet See printer user manual for requirements
Disposable Gloves
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-5
System Periodic Maintenance
10-5-1 Preliminary Checks
The preliminary checks take approximately 15 minutes to perform. Refer to the Vivid™ 3 Pro/Vivid™ 3
Expert User Manual whenever necessary.
St
ep Item Description
1 Ask & Listen Ask the customer if they have any problems or questions about the equipment.
Fill in the appropriate details at the top of the Vivid™ 3 Preventative Maintenance Inspection
2 Paperwork
Certificate (see page 10-30). Note all probes and system options.
Turn the system power ON and verify that all fans and peripherals turn on. Watch the displays
3 Power-up
during power up to verify that no warning or error messages are displayed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: Refer to Chapter 4 - Functional Checks, for additional details about the functional checks described in
this section.
Applicable Software Verify the basic operation of all optional modes such as Multi-Image, 3D, Contrast,
5 a Harmonics, Cine, Stress Echo,... etc.
Options Check the basic system controls that affect each options operation.
Perform an Element Test on each probe to verify that all the probe elements and system
6 Probe Elements channels are functional.
7 Perform the Automatic Test to verify that all boards are functioning according to
System Diagnostics
specifications.
8 Perform the Keyboard Test Procedure (external and alphanumeric) to verify that all
Keyboard Test
keyboard controls are OK.
Scan a gray scale phantom and use the measurement controls to verify distance and area
calculation accuracy.
10 Measurements
Refer to Chapter 18 of the Vivid™ 4 User Manual, for measurement accuracy
specifications.
a. Some software may be considered as standard, depending upon system model configuration
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
1 VCR Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
2 B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3 Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4 DICOM Verify that DICOM is functioning properly. Send an image to a DICOM device.
1 Unplug Cord Disconnect the mains cable from the wall and system.
2 Mains Cable Inspect the mains cable and its connectors for any damage.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals,
3 Terminals
and that no strands may cause a short circuit.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
10-5-4 Cleaning
1 External On a weekly basis, moisten a soft, non-abrasive folded cloth with a mild, general purpose,
Panels non-abrasive soap and water solution or general purpose disinfectant. Wipe down the top,
front, back and both sides of the unit. Do not spray any liquid directly onto the unit!
2 Monitor On a weekly basis, apply glass cleaner to a soft, non-abrasive folded cloth. Gently wipe the
monitor face.
3 Console Frequent and diligent cleaning of the Vivid™ 3 ultrasound unit reduces the risk of spreading
infection from person to person, and also helps to maintain a clean working environment.
Prior to cleaning, turn OFF the power to the system.
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Be careful not to get the cloth too wet so that moisture does not enter the console.
5 Probe Holder Clean the probe holders with warm water and a damp cloth to remove all traces of gel
(soaking may be required to remove excess gel).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: It is recommended that the air filter cleaning procedure is performed once every 3 months (quarterly).
However, the required frequency for air filter cleaning will vary in accordance with environmental
conditions.
Remove the air filter located at the front of the scanner, above the probe connector sockets.
1 Air Filter Clean the air filter as follows: shake it in an area away from the system, wash with a mild
soapy solution, rinse, and air dry.
NOTE: For convenience (or if the air filter is excessively dirty), replacement filters are available. Refer
to Chapter 9 - Renewal Parts for the air filter replacement part number.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
2 Scratches and Dents Inspect the console for dents, scratches or cracks. Use touch-up paint, as required.
Clean the monitor with a soft cloth dampened with mild detergent and water.
3 Monitor Repeat using only water, and wipe with a dry cloth. Inspect the monitor for
scratches and raster burns.
4 Probe Holders Clean the probe holders with warm water and a damp cloth to remove all traces of gel.
Verify ease of control console (Operator I/O Panel) movement in all acceptable
6 Control Console Movement
directions. Ensure that it latches in position, as required.
Check all wheels and castors for wear and verify operation of the foot brake to stop
7 Wheels and Brakes the unit from moving, and release mechanism. Check all wheel locks and wheel
swivel locks for proper operation.
Check that the FE crate cooling fans, BE fan and peripheral fans are operating and
8 Fans
clean.
Clean the drive head and media with the vendor-supplied cleaning kit. Advise the
9 user to repeat this often, to prevent future problems. MOD disks must be stored
MOD
away from dust and cigarette smoke. Do not use alcohol or benzene to clean the
MOD cartridge.
Clean the drive head and media with the vendor-supplied cleaning kit. Advise the
10 CD Drive user to repeat this often, to prevent future problems. CDs must be stored away from
dust and cigarette smoke. Do not use alcohol or benzene to clean the CD drive.
Check the rear panel connectors for bent pins, loose connections and loose or
11 Rear Panel Connectors
missing hardware. Verify that the labeling is in good condition.
Check the AC board connectors and the associated cabling for good connection
12 AC Unit
and proper insulation. Verify that the connections are secured.
Check the BEP rear connector for bent pins. Screw all the cable connectors tightly
13 BEP Rear Connector
to the connector sockets on the panel.
Check all internal cable harnesses and connectors for wear, and secure connector
14 Cables and Connectors seating. Pay special attention to footswitch assembly and probe strain or bend
reliefs.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation or
other deterioration, and verify continuity. Tighten the clamps that secure the power
15 Power Cord
cord to the unit and the outlet plug to the cord. Replace the power cord and clamp,
as required.
Check to ensure that all EMI shielding, internal covers, panels and screws are in
16 Shielding and Covers
place. Missing covers and hardware can cause EMI/RFI problems during scanning.
18 External Microphones Check for proper operation of any external microphones by recording an audio test.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-6
Probe Maintenance
WARNING: Any evidence of wear on a probe indicates that it must not be used. Improper
handling may easily damage ultrasound probes. See the Vivid™ 3 Pro/Vivid™ 3 Expert User
Manual and also refer to the probe manufacturer’s handling instructions, for more details.
Failure to follow these precautions can result in serious injury and equipment damage. Failure
to properly handle or maintain a probe may also void its warranty.
Check there are no traces of gel on the probe holders. If any gel present, clean the probe
1 Probe Holder holders with warm water and a damp cloth to remove all traces of gel (soaking may be required
to remove excess gel).
Check all the probes for wear and tear on the lens, cable and connector. Look for bent pins on
the connector and in the connector socket on the unit. Remove any dust from inside the
connector sockets.
2 Probes Verify that the EMI fingers around the probe connector socket housing are intact.
Check the probe locking mechanism and probe switch.
Perform a visual check of transducer pins and connector sockets before connecting a probe.
Verify that the system properly recognizes all probes.
Note: For detailed information on handling endocavity probes, refer to the appropriate supplementary
instructions for each probe.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
WARNING: Any evidence of wear on a probe indicates that it must not be used. Ultrasound
probes can be easily damaged by improper handling. See the Vivid™ 3 Pro/Vivid™ 3 Expert User
Manual and also refer to the probe manufacturer’s handling instructions, for more details.
Failure to follow these precautions can result in serious injury and equipment damage. Failure
to properly handle or maintain a probe may also void its warranty.
TEE and Interoperative probes often have special usage considerations; always refer to the individual
probe manufacturers’ handling instructions/user manual.
CAUTION: To help protect yourself from blood-borne diseases when cleaning and handling
probes, wear approved, non-allergic disposable gloves.
Note: IMPORTANT - For probe cleaning and disinfecting instructions, refer to the individual probe
Users Manual (or care instructions supplied with the probe).
See also the information provided in Chapter 13 of the Vivid™ 3 Pro/Vivid™ 3 Expert User
Manual.
Note: Failure to follow the prescribed cleaning/disinfection procedures will void the probe’s warranty.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
WARNING: Always disinfect a defective probe before returning it to the manufacturer. Be sure
to tag the probe as being disinfected.
CAUTION: To help protect yourself from blood-borne diseases when cleaning and handling
probes, wear approved, non-allergic disposable gloves.
Equipment being returned must be properly clean and free of blood and other potentially infectious
contaminants.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstances should a part or
equipment be shipped before being visibly clean and properly disinfected.
The purpose of the regulation is to protect employees in the transportation industry, as well as the
persons who will receive and/or open the package.
NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or dripping
with human blood that are now caked with dried blood; or which were used or intended for use in patient
care” are “regulated medical waste” for transportation purposes and must be transported as a
hazardous material.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-7
Electrical Safety Tests
10-7-1 Safety Test Overview
The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC
60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of
cord-connected, electrically operated, patient care equipment. If additional information is needed, refer
to the NFPA 99 (for USA) and IEC 60601-1 documents.
WARNING: THE USER MUST ENSURE THAT THE SAFETY INSPECTIONS ARE
PERFORMED AT LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF
THE PATIENT SAFETY STANDARD IEC-EN 60601-1. ONLY TRAINED PERSONS ARE
ALLOWED TO PERFORM THE ABOVE-MENTIONED SAFETY INSPECTIONS.
CAUTION: To avoid the risk of electrical shock, the unit under test must not be connected to
other electrical equipment. Remove all interconnecting cables and wires. The operator and
patient must not come into contact with the unit while the tests are being performed.
CAUTION: Possible risk of infection. Do not handle soiled or contaminated probes and other
components that have been in patient contact. Follow appropriate cleaning and disinfecting
procedures before handling the equipment.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
a. Mains Applied is the Sink Leakage Test. In this test, mains (supply) voltage is supplied to the part or equipment to
determine the amount of current that will pass (sink) to Ground if a patient comes into contact with the part.
Table 10-14 Type CF Applied Part Current Leakage Limits - Surgical Probes and ECG Connections
a. Mains Applied is the Sink Leakage Test. In this test, mains (supply) voltage is supplied to the part or equipment to
determine the amount of current that will pass (sink) to Ground if a patient comes into contact with the part.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
NOTE: No outlet tester can detect the condition in which the Neutral (grounded supply) conductor and the
Grounding (protective earth) conductor have been reversed. If later tests indicate high current leakages,
this should be suspected as a possible cause and the outlet wiring should be visually inspected.
Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
The Ground wire resistance should be less than 0.2 Ohms. Refer to the procedure in the IEC 601-1.1.
CONSOLE
OHMMETER
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
10-7-5-1 Definition
This test measures the current that would flow in a grounded person who touched accessible metal
parts of the bedside station, if the Ground wire should break. The test verifies the isolation of the power
line from the chassis.
The meter is connected from accessible metal parts of the case to Ground. Measurements should be
made with the unit ON and OFF, with the power line polarity Normal and Reversed. Record the highest
reading.
CAUTION: To avoid damaging the unit, never switch the polarity and the neutral status when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the Polarity switch
and/or the Neutral switch.
Figure 10-3 Set Up for Chassis Source Current Leakage, IEC 601-1 Clause 19
Continuous Current Leakages and Patient, Auxiliary Currents
When using the Microguard or a similar test instrument, its power plug may be inserted into the wall
outlet and the equipment under test be plugged into the receptacle on the panel of the meter. This
places the meter in the grounding conductor and the current flowing from the case to Ground will be
indicated in any of the current ranges. The maximum allowable limit for chassis source leakage is shown
in Table 10-12 on page 10-18.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Table 10-15 Typical Data Sheet for Chassis Source Current Leakage Test
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
10-7-6-1 Definition
This test measures the current which would flow to Ground from any of the isolated ECG leads. The
meter simulates a patient who is connected to the monitoring equipment and is grounded by touching
some other grounded surface.
CAUTION: To avoid damaging the unit, never switch the polarity and the neutral status when the unit
is powered ON. Be sure to turn the unit power OFF before switching them, using the Polarity switch
and/or the Neutral switch.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
CAUTION: Line voltage is applied to the ECG leads during this test. To avoid the risk of possible
electric shock, the system being tested must not be touched by anyone (patient or operator),
while the ISO TEST switch is depressed.
NOTE: It is not necessary to test each lead individually, or power condition combinations, as required
in previous tests.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Table 10-16 Maximum Allowance Limit for ECG Lead Leakage Test to Ground
Maximum
Allowance Limit
Table 10-17 Maximum Allowance Limit for ECG Lead Isolation Sink Test
AC Power Maximum
Source Allowance Limit
115V 20uA
Patient Lead Isolation Current Test
220/240V 5mA
ON NORM CLOSED
ON REVERSE CLOSED
ON NORM OPEN
ON REVERSE OPEN
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
10-7-9-1 Definition
This test measures the current that would flow to Ground (from any of the probes) through a patient who
is being scanned, and who becomes grounded by touching some other grounded surface.
CONSOLE
POWER N (WHITE)
OUTLET
G (GREEN)
MOMENTARY
SWITCH LEAKAGE TEST
METER
NOTE: Each probe will have a certain amount of leakage current, depending on its design. Small variations in
probe current leakages are normal from probe to probe. Other variations will result from differences in
line voltage and test lead placement.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
CAUTION: When power to the unit is ON, never switch the Polarity and the status of Neutral . Be
sure to turn OFF power to the unit before switching them, using the POLARITY switch and/or the
NEUTRAL switch. Failure to comply with this warning may cause damage to the unit!
Table 10-19 Typical Data Sheet for Transducer Source Current Leakage Test
Transducer Tested:
Tester Tester
Power GROUND
POLARITY or NEUTRAL
Unit Power Switch Switch Measurement
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-8
Excessive Current Leakage
10-8-1 Possible Causes of Excessive Current Leakage
Tighten all Grounds. Ensure star washers are under all Ground studs.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or
owner to correct any deviations. As a work around, check the other outlets to see if they could be used
instead.
NOTE: No outlet tester can detect the condition where the white neutral wire and the green grounding
wire are reversed. If later tests indicate high current leakages, this should be suspected as a
possible cause and the outlet wiring should be visually inspected.
NOTE: Each probe will have some amount of leakage, dependent on its design. Small variations in
probe current leakages are normal from probe to probe. Other variations will result from
differences in line voltage and test lead placement. The maximum allowable leakage current for
body surface contact probe differs from inter-cavity probe. Be sure to enter the correct probe
type in the appropriate space on the check list.
If excessive leakage current is slot dependent, inspect the system connector for bent pins, poor
connections, and Ground continuity.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
Section 10-9
PM and Safety Inspection Certificates
*Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array, or other
Measurement Accuracy
GE-approved Peripherals
Comments:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
DIRECTION 2300164-100, REVISION 7 VIVID™ 3 PRO/VIVID™ 3 SERVICE MANUAL
PROBES
Probe 1:
Probe 2:
Probe 3:
Probe 4:
Probe 5:
Probe 6:
Probe 7:
Probe 8:
Probe 9:
Final Check: All system covers are in place. The Vivid™ 3 system scans with all probes, as expected.
Accepted by:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GE MEDICAL SYSTEMS
GE Medical Systems-Americas: Fax 414.544.3384
P.O. Box 414; Milwaukee, Wisconsin 53201-0414, U.S.A.
GE Medical Systems-Europe: Fax 33.1.40.93.33.33
Paris, France
GE Medical Systems-Asia: Fax 65.291.7006
Singapore
GE Vingmed Ultrasound: Fax: +47 3302 1350
34 Horten, Norway
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Artisan Technology Group is an independent supplier of quality pre-owned equipment
Artisan Scientific Corporation dba Artisan Technology Group is not an affiliate, representative, or authorized distributor for any manufacturer listed herein.
We're here to make your life easier. How can we help you today?
(217) 352-9330 I [email protected] I artisantg .com