Metra 18C and Metrawin 90 Modular Calibration System, CP
Metra 18C and Metrawin 90 Modular Calibration System, CP
GOSSEN-METRAWATT GMBH
METRAHit 18C
Rugged, Reliable EMC Design Modular Expansion to Calibration System (see also page 6)
The calibrator is shielded from ambient influences and is pro- The METRAHit 18C not only serves as a calibration standard. As
tected from rough handling (jolts and impacts) with a sealed, a professional calibration tool, it can be readily linked to a PC at
impact resistant housing and a protective rubber case which is any time as well, from which frequently used calibration data and
provided as standard equipment. Additional protection is pro- sequences can be uploaded. Measuring point specific data can
vided by the optional ever-ready case. The internal ferromagnetic be loaded for on-site calibration tasks, and device specific data
shield and electronic attenuators protect against interference for laboratory applications.
caused by electrical and magnetic leakage fields, as well as elec- The calibration of measuring transducers or transmitters for pro-
trostatic discharge. cess engineering applications can be easily automated through
the additional implementation of the METRAHit 18S multimeter
Automatic Blocking System (ABS) and METRAwin90 software, which allows for a complete com-
Our proven mechanical blocking system prevents incorrect con- parison of set-points and actual values, as well as the creation of
nection of measurement cables after the desired sensor function a test record with a PC and a printer.
has been selected at the rotary switch. The METRAHit BD232 interface adapter, the METRAHit SI232
memory adapter and METRAwin90 software are available as
accessories for connection to a PC.
Data Storage
Calibration sequences and programs can be stored and saved to Test Certificate
the integrated memory with the keypad and can be queried either
forward or in reverse. A retraceable calibration certificate is included with every
The calibrator can be linked to a PC with the optional METRAHit 18C calibrator at no additional cost.
METRAHit BD232 plug-in interface adapter. Individual values,
intervals and ramps which have been created and stored to a Automatic Reference Junction
data file with the help of METRAwin90 software can be Thermoelectric voltages can be simulated for the ten most com-
uploaded to the calibrator and saved to non-volatile memory. mon types of sensors. Thermoelectric voltage can be read out
with reference to an internal (0 °C) or an external reference junc-
Easy Operation tion. The temperature for the external reference junction can be
All calibration and simulation functions are controlled by means of adjusted manually at the calibrator or the PC.
the central rotary switch and four keys. In this way, the device to be calibrated does not have to be con-
Stored values and sequences can be selected, one after the nected to the calibrator with a compensation cable: a copper
other, with the ↑ and ↓ keys. conductor is sufficient.
2 GOSSEN-METRAWATT GMBH
METRAHit 18C
Ramp Temperature Across Resistance Thermometer
This output mode provides for the continuous read-out of step- RTD Source Range Inherent Deviation
less calibration values between the selected minimum and maxi- DIN/IEC 751 [°C] ± [% + K]
mum values of the device to be calibrated. Pt100 – 180 ... 850 0.1 + 0.25
Ramp delay for rising and falling ramps can be adjusted within a Pt1000 – 180 ... 250 0.1 + 0.1
range of 1 s to 60 min.
Ni100 – 60 ... 180 0.1 + 0.25
Memory Ni1000 – 60 ... 180 0.1 + 0.1
Values which have been manually stored to non-volatile memory Resolution 0.1 °C or 0.1 °F
(10 values per measuring range or sensor type) can be queried in Sensor Current 0.05 ... 0.1 ... 1.0 ... 1.4 mA
steps. Complete calibration sequences for devices or systems
can be uploaded from a PC and read out either manually or auto- Temperature Across Thermocouple
matically.
Thermoelectric Voltage Types Range Inherent Deviation
Source [°C] ext. ref. junction
DIN/IEC 584 ± [% + µV]*
Fe-CuNi type J – 200 ... 1200 0.1 + 15
Applicable Regulations and Standards Fe-CuNi type L – 200 ... 900 0.1 + 15
Cu-CuNi type T – 250 ... 400 0.1 + 15
IEC 61010-1 Safety requirements for electrical equipment for measure-
Cu-CuNi type U – 200 ... 600 0.1 + 15
DIN EN 61010 Part 1 ment, control and laboratory use
VDE 0411-1 NiCr-CuNi type K – 250 ... 1350 0.1 + 15
DIN 43 751 Digital measuring instruments NiCr-CuNi type E – 250 ... 1000 0.1 + 15
Generic standard for interference emission; Pt10Rh-Pt type S – 50 ... 1750 0.1 + 15
DIN EN 50081 Part 1
residential, business and light industry Pt13Rh-Pt type R – 50 ... 1750 0.1 + 15
Generic standard for interference immunity; Pt30Rh-Pt6Rh type B + 50 ... 1800 0.1 + 15
DIN EN 50082 Part 1
residential, business and light industry
NiCr-NiSi type N – 240 ... 1300 0.1 + 15
VDI/VDE 3540 Reliability of measuring, control and regulating devices
* With a fixed reference temperature of 0°C and thermoelectric voltage
DIN EN 60529 Test instruments and test procedures
VDE 0470 Part 1 – Protection provided by enclosures (IP code)
Industrial platinum resistance thermometers Resolution 1°C or 1°F
DIN IEC 751
and platinum resistance elements Reference Junction internal: inherent deviation: 2°C
external: entry: – 50 ... 100 °C
Frequency
Calibration Ranges and Inherent Deviation Frequency (square-wave pulse) Resolution [Hz] Inherent Deviation ± %
0.01 ... 999.99 Hz 0.01*) 0.05
Inherent deviation data are related to a reference temperature of *)
A pulse rate of 16 µs reduces resolution within the higher frequency range. The
23 °C ±3 K. The % values make reference to the respectively precise frequency is displayed.
selected set-point.
Pulse Train continuous or as pulse package,
Direct Current adjustable from 1 to 99,999
I Source/Sink Range Load Inherent Deviation Pulse Amplitude 0 ... 15 V
[mA] [Ω] ±[% + mA] Keying Ratio 50%
0 ... 20.000
Fixed Ranges ... 750 0.05 + 2
4 ... 20.000 Time Functions, Interval and Ramp
Variable Ranges 0 ... 24.000 ... 750 0.05 + 2 Inherent Deviation approx. ±1% of set-point ±0.1 ms
Overload limited by PTC thermistor to max. 42 V
Input Voltage 4 ... 27 V in “sink” mode
GOSSEN-METRAWATT GMBH 3
METRAHit 18C
Mechanical Design
Protection IP 50 per VDE 0470 / EN 60529
IP 20 at terminals and power pack
connector sockets
Dimensions 84 mm x 185 mm x 35 mm
Weight 0.4 kg with batteries
Ambient Conditions
Operating Temp. –10 °C ... +50 °C
Storage Temperature –25 °C ... +70 °C
Relative Humidity max. 75%, no condensation allowed Data Interface
Climatic Category 2z/0/50/70/75% in compliance with
Type RS232C per DIN 19241
VDI/VDE 3540
Data Transmission optical with infrared light through housing
Elevation to 2000 m
Baud Rate 9600 bits/s
Power Supply
METRAHit 18C Standard Equipment
Battery Operation 3 alkali manganese mignon cells
1.5 V/2.2 Ah, type IEC LR6 1 METRAHit 18C calibrator with 3 IEC LR6 batteries
Service Life approx. 15 h without current drain for Uout 1 KS17 cable set (yellow), consisting of 2 measurement
≤ 1.5 V (at 2.2 Ah pre cell) cables (1 black, 1 yellow) with angle plugs and test probes
approx. 10 h without current drain for Uout 1 Operating instructions
> 1.5 V 1 GH18 protective rubber case
approx. 12 h for I source 20 mA/≤ 200 Ω
approx. 6 h for I source 20 mA/> 200 to
750 Ω
4 GOSSEN-METRAWATT GMBH
METRAwin90
METRAwin90 Calibration Software (optional) Calibration “Test” Program in Accordance with ISO9001, 4.11
A calibration sequence can be pre-programmed in the TEST
METRAwin90 calibration software is shipped with the CP1 and mode which can be run in selectable, individual steps, or fully
the CP2 calibrator pack (see page 7). This allows for convenient automatically (compare: applications example on page 6).
configuration of the calibrator, as well as analysis of calibration The calibration values and a freely definable comment are read
sequences with the METRAHit 18C calibration system via PC. out after each calibration step. After the output parameters of the
The following settings can be configured: device to be calibrated (e.g. a transducer) have been logged and
evaluated with the multimeter, the measurement results are dis-
Direct Entry at the Calibrator played in tabular form. Test sequence results can be saved to a
The calibrator is controlled by the PC, and preset values are data file or printed out as a calibration certificate. All requirements
uploaded from the PC directly to the calibrator. in accordance with ISO9001 Abs.11.4 are fulfilled.
The following operating modes are provided with the indicated
calibration ranges:
• DIRECT A freely selectable, fixed numeric calibration value
can be entered.
• INTERVAL Continuous read-out of calibration values from a
calibration interval which can be subdivided into
intermediate steps as desired.
• RAMP Continuous, stepless read-out of calibration val-
ues from a calibration ramp with definable rise and
dwell periods. The ramp function may also be
operated continuously.
GOSSEN-METRAWATT GMBH 5
METRAHit 18C and METRAwin90
Modular Calibration System, CP
Isolation Amplifier
21.978 mA 109.89 °C
Temperature
Measuring Transducer
t
0 ... 100 °C 4 ... 20 mA
I/U
6 GOSSEN-METRAWATT GMBH
METRAHit 18C and METRAwin90
Modular Calibration System, CP
Standard Equipment included with CP2 Calibration System METRAHit 18S Technical Data
1 METRAHit 18C calibrator The METRAHit 18S is used within the CP2 calibration system for
including cable set and operating instructions the recording of actual values from the device to be calibrated
1 METRAHit 18S multimeter with 4 3/4 places and a 30,999 digit display. The METRAHit 18S
including cable set and operating instructions includes all of the functions of a precision multimeter in addition to
1 METRAHit BD232 interface adapter accurate logging of standard signals ranging from 0 ... 20/24 mA
and 0 ... 300 mV/10 V/30 V.
1 METRAHit SI232 memory adapter
1 F840 ever-ready case
1 RS-232 bus cable, 2 m, 9-pole – 9-pole Excerpt from List of Relevant Measuring Ranges
1 METRAwin10/METRAHit software package including for the Use of the METRAHit 18S in Calibrations Systems:
3½" program floppy disc and installation instructions
1 METRAwin90 software package including 3½" program Measurement Measuring Range Resolution Inherent Deviation
Function
floppy disc and installation instructions
1 NiMHc set (battery charger and 6 NiMH rechargeable batteries) 300.00 mV 10 µV 0.5 + 30 D (> 500 D)
3.0000 V 100 µV 0.3 + 30 D (> 300 D)
1 KS32 cable set for METRAHit 18C/S consisting of: V 30.000 V 1 mV 0.3 + 30 D (> 300 D)
4 measurement cables (2 black, 1 red, 1 yellow) ... ... ...
with angle plugs and straight plugs ... ... ...
4 plug-on alligator clips (2 black, 1 red, 1 yellow) mA 30.000 mA 1 µA 0.5 + 30 D (> 300 D)
4 plug-on adapters, 2 mm (2 black, 1 red, 1 yellow) ... ... ...
GOSSEN-METRAWATT GMBH 7
® ®
METRAHit 18C and METRAwin 90
Modular Calibration System, CP
GOSSEN-METRAWATT GMBH
Thomas-Mann-Str. 16-20
D-90471 Nuremberg, Germany
Telephone +49 911 8602-0
Telefax +49 911 8602-669
http://www.gmc-instruments.com
e-mail: [email protected]