10 - Ch-06 Test Selection 0324
10 - Ch-06 Test Selection 0324
Test selection
A. Outer tray
Layer 1: position # 1 - 20
Layer 2: position # 21 – 40
Figure 6.1-1
B. Inner tray
Figure 6.1-2
z Set a sample cup for “Pediatric” on the specific sample tube (diameter: 13mm,
length: 85mm) and place it in the Outer tray.
111
Chapter 6. Test selection
D. Sample positioning
Measurement starts in numeric order accordingly. The samples in the Inner tray are
measured prior to those in the outer tray. Tray selection varies depending on the
sample type.
When performing calibration at the same round, assign younger position # for the
calibrators than the position # for the normal samples.
112
Chapter 6. Test selection
Figure 6.2-1
1. Start
Specify the start position # in the Sample tray.
Select the position # (1 to 50) from drop-down menu.
Or specify the position directly from SID list.
For the samples with Barcode labels, do not specify position # and leave as
blank.
The position # 41 to 50 is not available for normal and emergency samples.
113
Chapter 6. Test selection
2. End
Specify the end position # to copy or delete. (This is effective only when ticking
on “C/D”.)
3. Sample Category
Specify the Sample Category from drop-down menu.
Sample Category Description
4. Template
Select the Template from drop-down menu.
5. Cup
Select “Normal” or “Pediatric” from drop-down menu.
6. Sample Barcode
Tick the “Sample Barcode” ON when applying barcode reader to enter the
sample ID.
7. C/D
Tick “C/D” ON when copying or deleting the registered order.
114
Chapter 6. Test selection
8. E
Tick “E” ON when specifying normal sample as an emergency sample.
9. SID
Enter the Sample ID.
12. PID
Enter the Patient ID. For the registered PID, select from drop-down menu.
When entering a new Patient ID, a registration screen for the PID appears.
14. Physician
Select Physician name from drop-down menu if any registered information.
15. Department
Select the Department from drop-down menu if any registered information.
17. Number
Specify measurement number for replicate measurement. (This is effective for
normal sample and emergency sample.)
20. Restart
Perform to restart measurement. (Effective only when adding a sample.)
115
Chapter 6. Test selection
23. Save
Saves the setting.
24. Cancel
Cancels the editing data.
25. Print
Printouts the registered orders.
26. Delete
Deletes the “Pos” information you select.
116
Chapter 6. Test selection
Figure 6.2-2
117
Chapter 6. Test selection
z When selecting “Pediatric” for sample cup, the Sample ID is assigned the
number with 8999 as first 4 digits.
When entering last 4 digits of the Sample ID, the first 4 digits will be added
automatically.
6. Select the Sample type (Serum, Urine, Plasma, or Others) from drop-down
menu.
7. Select the Sample volume (Low, Normal, or High) from drop-down menu.
For each volume of the sample, specify on the [Chemistry Parameters 1]
screen.
11. Enter the Physician, the Department, and the Sample Comment.
Directly enter or select from the list of the drop-down menu.
12. Specify the Number (number of measurement) for the selected method from
drop-down menu.
In the default setting, it is "01". The available range is “1 – 99”.
When selecting “2” or more at “No.”, auto rerun is not performed.
C. Copy procedure
3. Select the Start # as the beginning Position # for copying from drop-down menu.
4. Select the End # as the end Position # for copying from drop-down menu.
D. Delete procedure
2. When deleting the specific order, select the Pos” or SID on the SID list and click
the “Delete” icon.
119
Chapter 6. Test selection
Figure 6.2-3
120
Chapter 6. Test selection
z The SID# for Emergency sample is started with “99000” in 8 digits. The SID#
is classified by sample cup type.
z When selecting "Normal” as sample cup, “99000” as the first 5 digits is added
in the SID.
121
Chapter 6. Test selection
Figure 6.2-4
Under managing with barcode reader, the analyzer recognizes the necessary
information from barcode labels.
No order registration is required except placing necessary QC samples in the Sample
container.
For the QC samples set in Inner tray or without managing with barcode reader, follow
the procedure below.
5. Click the “Save” icon to confirm the order. The SID for the QC sample is listed
on the SID list.
6. For deleting, click the SID on the SID list, click “Delete” icon.
For the deletion confirmation, click the “Save” icon.
122
Chapter 6. Test selection
Figure 6.3-1
Under managing with barcode reader, the analyzer recognizes the necessary
information from barcode labels.
No order registration is required except placing necessary QC samples in the Sample
container.
For the calibrators set in Inner tray or without managing with barcode reader, follow the
procedure below.
3. Select the Sample type (Serum, Urine, Plasma, or Others) from drop-down
menu.
123
Chapter 6. Test selection
4. Specify the method. When ticking ON at the method name, its measurement
will be ordered.
(This is available to select only 1 method. When using a single calibrator to
prepare several calibration curves for the methods, select “Multi-Calibrator.”)
The method that registered on the [Calibration Registration] screen is available
to be ordered.
6. For deleting, click the SID on the SID list, click the “Delete” icon.
For the deletion confirmation, click the “Save” icon.
6.3.2 Multi-Calibrator
Using a single calibrator, prepare several calibration curves for the methods.
Measurement starts in “Pos” numerical order.
Multi-Calibrators can be set in both Inner and Outer trays. The multi-calibrators in the
Inner tray will be measured first.
Figure 6.3-2
Under managing with barcode reader, the analyzer recognizes the necessary
information from barcode labels.
No order registration is required except placing necessary multi-calibrators in the
Sample container.
124
Chapter 6. Test selection
For the multi-calibrators set in Inner tray or without managing with barcode reader,
follow the procedure below.
3. Select the Sample type (Serum, Urine, Plasma, or Others) from drop-down
menu.
7. Click the “Save” icon to confirm the order. The SID for the Multi-Calibrator
sample is listed on the SID list.
8. For deleting, click the SID on the SID list, click the “Delete” icon.
For the deletion confirmation, click the “Save” icon.
125
Chapter 6. Test selection
Figure 6.3-3
Under managing with barcode reader, the analyzer recognizes the necessary
information from barcode labels.
No order registration is required except placing necessary diluted calibration samples
in the Sample container.
For the diluted calibrators set in Inner tray or without managing with barcode reader,
follow the procedure below.
1. Specify the Start# for the Diluted Calibrator from drop-down menu.
3. Select the Sample type (Serum, Urine, Plasma, or Others) from drop-down
menu.
4. Specify the method. Tick the method ON for the measurement order.
5. Click the “Save” icon to confirm the order. The SID for the Diluted calibration
sample is listed on the SID list.
6. For deleting, click the SID on the SID list, click the “Delete” icon.
For the deletion confirmation, click the “Save” icon.
126
Chapter 6. Test selection
Figure 6.3-4
Under managing with barcode reader, the analyzer recognizes the necessary
information from barcode labels.
No order registration is required except placing necessary blank samples in the
Sample container.
For the blank samples set in Inner tray or without managing with barcode reader, follow
the procedure below.
3. Select the Sample type (Serum, Urine, Plasma, or Others) from drop-down
menu.
5. Click the “Save” icon to confirm the order. The SID for the blank sample is
listed on the SID list.
6. For deleting, click the SID on the SID list, click the “Delete” icon.
For the deletion confirmation, click the “Save” icon.
127
Chapter 6. Test selection
6.3.5 Mask
The Mask setting allows to set methods not performed. The methods indicate in a
measureable status that has analysis conditions, the calibration curves, and its
reagents in the Reagent container.
Figure 6.3-5
A. Mask selection/release
2. Select the Mask type (Normal, QC sample, or Calibrator) from drop-down menu.
4. By ticking the method ON to unselect the mask, the method becomes unmasked
and measurable status. Tick the checkmark again to remove the checkmark.
It becomes the mask status active.
128
Chapter 6. Test selection
6.3.6 Orderless
Orderless selection allows to perform measurement in all methods available for all
samples in the Sample container.
Prior to orderless measurement, specify the methods to be measured.
For the orderless measurement, refer to “7.1.5 Orderless measurement.”
Figure 6.3-6
2. Select the Sample type (Serum, Urine, Plasma, or Others) from drop-down
menu.
3. Select the Normal range from the pop-up list by clicking the drop-down menu.
It is “Human Auto” as the default setting.
Normal range is automatically selected from patient information such as age
and gender registered in the [Patient Information] screen and applied for
judgment.
When not specifying in age of the patient, “G2” is applied. When not specifying
in gender, “Male” is applied.
4. Specify the Number (number of measurement) for the selected method from
drop-down menu.
It is "01" as the default setting. The available range is “1 – 99”.
When selecting “2” or more, auto rerun is not performed.
z When some orders specified in the [Test Selection] screen, the orders are
measured prior to the orderless measurement.
129
Chapter 6. Test selection
Figure 6.3-7
1. Select the Start# from drop-down menu for the wash solution.
z SID list indicates the SID No. “96000201” as the Sample ID code.
1. Select the Start# from drop-down menu for the wash solution.
130
Chapter 6. Test selection
Figure 6.3-8
1. Select the Start# from drop-down menu for the ISE calibrator.
z SID list indicates the SID No. “96000001” as the Sample ID code.
1. Select the Start# from drop-down menu for the ISE calibrator.
131
Chapter 6. Test selection
z 500 µL of the ISE cleaning solution (CLEAN99K) is required for one ISE
cleaning procedure.
Figure 6.3-9
1. Select the Start# from drop-down menu for the ISE cleaning solution.
z SID list indicates the SID No. “96000201” as the Sample ID code.
1. Select the Start# from drop-down menu for the ISE cleaning solution.
132
Chapter 6. Test selection
Cleaning of the Na electrode requires ISE Etching. The ISE Etching needs to be
placed on the Sample container and specified its position.
With the sample barcode reader, the position of the ISE Etching can be read from the
barcode label.
Without the sample barcode reader, the position of the ISE Etching needs to be
confirmed.
For the ISE Etching, refer to “18.2.5 ISE electrodes (option).”
z 500 µL of the ISE Etching is required for one ISE cleaning 2 procedure.
Figure 6.3-10
1. Select the Start# from drop-down menu for the ISE Etching.
z SID list indicates the SID No. “96000301” as the Sample ID code.
1. Select the Start# from drop-down menu for the ISE Etching.
133
Chapter 6. Test selection
Perform ISE electrodes activation for stable results after replacement. The ISE
activator (serum or specific QC sample) needs to be placed on the Sample container
and specified its position.
With the sample barcode reader, the position of the ISE activator can be read from the
barcode label.
Without the sample barcode reader, the position of the ISE activator needs to be
specified.
For the ISE activation, refer to “C. Procedure on the ISE etching” of “18.2.5 ISE
electrodes (option).”
z 500 µL of the ISE activator is required for one ISE activation procedure.
Figure 6.3-11
1. Select the Start# from drop-down menu for the ISE activator.
z SID list indicates the SID No. “96000401” as the Sample ID code.
1. Select the Start# from drop-down menu for the ISE activator.
134
Chapter 6. Test selection
6.4 Template
It is available to select the Template that set the samples in the sample tray with its
specified methods as a set order.
The template is required to register in the [Auto Template] screen prior to selecting in
the [Test Selection] screen. For registering the template setting, refer to “14.8
Template setting.”
Figure 6.4-1
Figure 6.4-2
135
Chapter 6. Test selection
When selecting “On line Real time” at the Host Communication Mode on the [System
Setup 1] screen, automatically receives orders from host connection and no need the
order setting.
When selecting “On Line Batch” at the Host Communication Mode on the [System
Setup 1] screen, it needs to obtain orders from host communication.
When selecting “On Line Real time” at the Host Communication Mode on the [System
Setup 1] screen, it needs to register the position # and the SID to specify the position #
manually placed on the sample tray.
When selecting “On Line Batch” at the Host Communication Mode on the [System
Setup 1] screen, it needs to register the position # and the SID to specify the position #
manually placed on the sample tray. It also needs to obtain orders from host
communication.
A. Order setting procedure (in the “On Line Batch mode” with the Sample barcode
reader)
B. Order setting procedure (in the On Line Real time mode without the Sample
barcode reader)
136
Chapter 6. Test selection
C. Order setting procedure (in the On Line Batch mode without the Sample
barcode reader)
137