Leica SP8
Leica SP8
Jordan Briscoe Kyle Marchuk Taylor Shagam John Eichorst Austin Edwards
January 2020
Contents
1 Introduction 2
2 Hardware Start-Up 2
6 Imaging Parameters 8
6.1 XY Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
6.2 Z-Stack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
6.3 Time Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
6.4 Tiling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
6.5 Mark and Find . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
6.6 Navigator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
8 LIGHTNING 11
1
1 Introduction
The Leica SP8 is an upright laser scanning confocal system configured for large volume (cleared tissue) multiplexed
imaging. The SP8 comes equipped with an AOBS tunable detection pathway (3 PMTs and 2 HyD detectors able
to detect from 380nm-800nm), a white light laser (470nm-670nm), and the Lighting deconvolution module. The
HyD detectors offer a “true black” background and are able to gate on fluorescence lifetime (allowing for the
in-line reduction auto-fluorescence).
The system also contains spectral scanning, three-dimensional (z) acquisition and time-lapse capabilities. A
programmable stage allows automated return-to-site and multiple-field stitching.
2 Hardware Start-Up
1. Turn on the three green switches on the right side of table, turn emission key to on position. See Figure: 1.
2
2. Turn on lamp box (top box in the middle of the table) This lamp is an LED that can be switched on and
off without cool down. See Figure: 2.
Figure 2
Figure 3
4. Turn white light emission key to on position (box all the way to the left). See Figure: 4.
Figure 4
3
must be manually changed in the software as well as on the body of the microscope. In order to update
objective changes, we need to go into the Leica LAS X Hardware. See Figure: 5.
Figure 5
Figure 6
Figure 7
4
6. Double click on the objective to select it for use. If the objective has been properly selected, it will appear
in the window at the top of the screen. See Figure: 8
Figure 8
Figure 9
9. Open the LAS X software. Choose the start up configuration options. The resonant scanner can only be
turned on and off in this window. See Figure: 10
5
Figure 10
10. Initialize the stage. THE STAGE MUST BE IN THE UP POSITION FOR INITIALIZATION OR IT
WILL HIT THE BODY OF THE SCOPE AND KNOCK IT OUT OF ALIGNMENT.
Figure 11
11. Mount the objective. If using the regular sized 16x, 20x, or 63x, you will need to add the silver extension
tube. If using the 20x water dipping, 16x motCorr, or 25x motCorr, you can mount them right to the
scope. See Figure ...
12. Mount your sample to a stage insert and place it on the stage. Screwing the insert down is not necessary
to have a stable image. Please note: you are adding your sample to the z-peizo so please be gentle.
13. Add the proper immersion liquid (water or oil) to your slide.
14. Ensure the z stepper motor is at its upper limit by using the 3-axis controller. See Figure ...
15. Raise the stage into its upper position. Please get prior BIDC approval to use any custom or non-Leica
stage inserts as the dimensions are extremely important to the safe usage of the microscope!
6
5. Click the IL-Shutter button to open and close the shutter.
6. Using the fine adjust only, bring your sample into focus.
7. Once the sample can be seen through the eyes, close the shutter and go to the LAS X software for imaging
setup.
Figure 12
2. In the Dye Assistant window, click the ’. . . ’ button in order to select the fluorescent markers that are
present in the sample. See Figure: 13
Figure 13
3. Choose which detector you want to pair up with the fluorescent marker in the ’PMT or HyD’ tab. Note:
the HyD detectors are meant for detecting markers that are dimmer or are more sensitive to higher laser
powers. If the marker is too bright and a HyD detector is used it can damage the detector. For this reason,
the detector has a safety shut off and can only be turned on when the fluorophore intensity is low enough.
This can be achieved by lowering the gain and laser power.
4. After choosing the fluorescent markers and the detectors, the Dye Assistant will provide a list of imaging
sequence options. The top option on the list will generally be the fastest sequence but will have the most
amount of bleed through. Whereas, the last option on the list will be the slowest option but will have the
least amount of bleed through. Users have the ability to edit their sequence choice (laser and detector
settings) within the Dye Assistant or after the settings have been applied.
None Sequential: All channels will be acquired simultaneously. Fastest option, but it limited to 5 channels.
Will likely have the most Crosstalk of the suggested settings.
7
Line Sequential: Switches Sequences each line of the image scanned. For there to be any speed in this
setting, the collection Bandpass will be set at the same width.
Frame Sequential: Switches Sequences at the end of each image plane. No limitation on the width of the
Bandpass. Switching sequences does take time.
Stack Sequential: Switches Sequences at the end of each Stack. Considerable faster than Frame Se-
quential and no limitation on Bandpass.
5. Once a sequence in the dye assist has been selected, the program will ask you if you want to turn on the
lasers needed for imaging. Select yes.
3. Turn the Power State to ON for the lasers you will use. See Figure...
4. Add laser lines by pressing the ’+’ button. See Figure...
Note: to add the 405 nm laser press the ’405’ button.
5. Turn on the PMTs you will use for each laser line and adjust the bandpass window.
Note: the minimum setting for the PMT bandpass window should be set at least 10 nm above the excitation
wavelength.
6. In the Sequence menu press the ’+’ to add the next sequence to your acquisition.
7. Assign PMTs, bandpass windows, and lasers (check and uncheck laser lines) to this sequence.
7. Click the Start button to collect collect and save an image of all sequences.
6 Imaging Parameters
After setting up the laser and gain acquisition settings for the sequences, the user can set up the rest of sample
acquisition parameters. Here is an overview of the most common methods and settings used for acquiring 3D
volumes.
8
6.1 XY Tab
In general, used for setting up the XY imaging parameters.
Format: The size of the image in pixels. Suggested size is 1024x1024 for most applications.
Speed: Image acquisition speed. Suggested speed is 600 Hz. Speed faster than 600 Hz will require increasing
the Zoom of the image.
Bidirectional X: Data is collected on both movement directions of the X galvo. Some acquisition parameters
are tied to this setting.
Zoom Factor: Digital Zoom of the image. Increasing the magnification of the objective will give a better
image and optical section capability.
Zoom In:
Image Size: Describes the size of the image in microns. This is dependent on the Objective, Format, and
Zoom Factor.
Pixel Size: Describes the size of the pixel in microns.
Optical Section:
Pixel Dwell Time: Length of data collection for each pixel.
Frame Rate: Time required to collect a single sequence at a given image plane.
Line Average: Number of times each line of the image will be acquired and averaged to produce the image.
Used to reduce noise.
Line Accu: Number of times each line is accumulated (summed) to produce and image. Can be used on very
dim samples.
Frame Average: Number of times each frame of the image will be acquired and averaged to produce the
image. Used to reduce noise.
Frame Accu: Number of times each frame is accumulated (summed) to produce and image. Can be used on
very dim samples.
Auto Gain:
Rotation:
Pinhole: Used to adjust the pinhole size. Increasing the pinhole produces a brighter signal at the expense of
your optical section capability.
6.2 Z-Stack
Used for creating the volumes of a 3D stacked image.
Begin: Using the Z Position knob, move the image plane to the ’Top’ of your sample and press this button
to assign this position.
End: Using the Z Position knob, move the image plane to the ’Bottom’ of your sample and press this button
to assign this position.
Down Arrow and Lines Icon:
Trash Can Icon: Deletes your position assignments.
Stack Direction (Z): Scanning direction for the volume acquisition. Typically want to move the stage in the
Up direction to scan from the Top of your sample to the Bottom.
9
z-Galvo: Uses the z-piezo actuator for movement. Fast, but limited to 500 microns of travel range.
00000000: Uses the z-stepper motor for movement. Slower, but has 13 mm of travel range.
Number of Steps: Within the given volume, determines z-step size by number of image planes.
Z-Step Size: Within the given volume, determines the number of steps by step size.
System Optimized: Determined by the objective; the smallest step size before over-sampling the volume.
Z-Compensation: Is used to associate laser power and gain with a z-position. Typically used to compensate
for scattering in a thick sample.
1. Check which options you’d like to associate with z-position. Excitation Gain is laser intensity. Emis-
sion Gain is the PMT gain.
2. Navigate to an imaging plane within your volume and adjust the laser intensity and PMT gain to your
satisfaction.
3. Move to other plane and repeat the process.
4. Delete any positions you do not need, or may have loaded when you started.
5. Select your next Sequence and repeat the process.
Galvo Flow: When On, the z-stage does not stop at each individual step. Instead it continuously moved
from the Begin to End point. Useful when needing super fast acquisitions.
Travel Range: Travel range of the selected z stage.
6.4 Tiling
The Tiling Module will create a 2D grid for acquisition based upon saved stage positions.
1. Locate an initial corner position (upper right, lower left, etc) for your grid either through the eyepiece or
while Live.
6. To decrease the grid size, add points within the current grid, and delete the unecessary point by selecting
that point in the drop down menu and pressing the Trash Can Icon.
7. Turn Merge Images, Auto Stitching, and Linear Blending to ON to automatically create a stitched
imaged at the end of acquisition.
8. To Use Focus Map a focus map will need to be made in the Navigator (See Below).
10
6.5 Mark and Find
The Mark and Find Module is a way to quickly save points and take a single-viewing-area image or volume.
1. Identify a position of interest while moving the stage in Live mode.
2. Add the position of interesting by pressing the Add Position button.
3. To remove an unwanted position select the position from the Positions drop down menu and press the
Trash Can Icon.
4. Select Same stack for all if you want to use the same imaging plane or Z-Stack setup for each position.
5. To define separate volumes for each position:
(a) Select your first position from the dropdown.
(b) Define your volume in the Z-Stack panel.
(c) Press Redefine Stack
(d) Select your next position and repeat the process.
6.6 Navigator
The Navigator Module is a large area explore tool that is capable of setting up multiple ROIs of different sizes
for acquisition.
Moving the Stage: Double-click on the viewing area and the stage will move and take an image.
Live: Press the Live button to continuously image at that stage position.
Fast Live: Press the Fast Live button to continuously image at that stage position in a lower resolution (faster
than Live).
8 LIGHTNING
9 Other Imaging Tips
9.1 Project Naming and Saving
It is important to adopt a project naming and saving convention to avoid future headaches when trying to analyze
the data.
11
9.2 USB Control Panel
The USB Control Panel (Figure ... ) can be customized to include controls you may need. For example; the xy
stage position can be controlled similarly to the z-piezo. To add additional features:
1. Press the USB Panel shortcut button (Figure...).
2. Select Scan Field Rotation and change it to x Position (Stage).
8. Turn off lamp box (top box in the middle of the table).
9. Turn off three green switches on the right side of table then turn key to off.
10. Ensure the area is clean and tidy for the next user.