3590 Series

The document provides instructions for calibrating different types of weighing systems connected to AF03GT indicators, including a single analogue weighbridge, multiple analogue weighbridges, digital cell weighbridges, wireless weighing systems, and dual scale systems. The calibration process involves connecting the weighing system to the indicator, checking the signal stability, selecting the number and type of load cells, setting the scale parameters, and performing the calibration by loading known weights.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
38 views32 pages

3590 Series

The document provides instructions for calibrating different types of weighing systems connected to AF03GT indicators, including a single analogue weighbridge, multiple analogue weighbridges, digital cell weighbridges, wireless weighing systems, and dual scale systems. The calibration process involves connecting the weighing system to the indicator, checking the signal stability, selecting the number and type of load cells, setting the scale parameters, and performing the calibration by loading known weights.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 32

AF03GT

Program for weighbridges

CALIBRATION
MANUALE
METHODD’USO
- v7 ITALIANO
ENGLISH

For 3590 series touch screen indicators

www.diniargeo.com
Index

Calibration of a channel connected analogue weighbridge  4

Analogue calibration with equalisation  8

Setting and calibration of digital cell weighbridges  12

Calibration of a WWS wireless weighing system  16

Calibration of a dual digital weighbridge system  20

Calibration of a dual scale system consisting of wireless WWS platforms  26

AF03GT 3
TECH_MAN_CAL_ENG_AF03GT_v7
Calibration of a channel connected analogue weighbridge

PREMISE

CASE 1: An indicator connected to a weighbridge.

Load cells
Weighbridge
Junction box

CASE 2: An indicator connected to a scale composed of several weighbridges in series (up to 48 1000 Ohm cells).

BOOT

Procedure:

A. With switched-off indicator and power supply unplugged, connect the weighbridge to the indicator.
SENSOR
RS232
COM3 SIG-
GND

J2 J3
+5V

SIG+ J3
485 EXC-
-
-

EXC+ SEN + / REF + EXC + SEN + / REF - EXC -


ON J2
RS232 RS232 CH4 CH3 CH2 CH1
COM1 COM2
SEN + /
SEN + /
EARTH

REF +

EXC+
+Vdc
CTS2

REF -
GND

SIG+
GND

EXC-
A(+)

SIG-
RX1
TX1

TX2

B(-)
IN2

RX

13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

AF03GT 4
TECH_MAN_CAL_ENG_AF03GT_v7
B. Turn on the indicator, enter the setup menu and check that the signal coming from the weighbridge is stable and increases with
weight increase.

1. 2. 3.
Technical set-up
...
Diagnostics
... kg
Channel
SCALE 1
15000 W1
A
f=7
d=74
ADC min MAX
... ... 1 344673 308333 308333

Diagnostic Weight test


... ...
... ...
... ...
Esc Esc Esc ADC-mV

C. Select the number of weighbridges and each weighbridge load cell type.
1. 2.
Calibration Scale selection
... ...
... ...
... Number of scales =1
Scale selection Type of load cell = Analogue
... Number of channels (Scale 1) =1
... ...
Esc Esc

D. Set scale metric parameters (capacity / division, etc.)

1. 2. 3.
Calibration Scale configuration Parameters
... ... Number of decimals
... ... Unit of measure
... Parameters Range number
Scale 1 ... Range 1 division
... ... Capacity / Range 1
... ... Range 2 division
Esc Esc Esc

Example of dual scale weighbridge configuration: Number of decimals = 0; Unit of measure = kg; Range number = 2;
Range 1 division = 10; Capacity / Range 1 = 30000; Range 2 division = 20; Capacity / Range 2 =60000.

E. Perform calibration.
1. 2. 3.
Calibration Scale configuration Scale configuration
... ... Calibration

... ...
Insert the number of calibration
... ... points other than the zero.

Scale 1 Calibration Calibrazione OK


Cancel
... ...
... ...
Esc Esc Esc

AF03GT 5
TECH_MAN_CAL_ENG_AF03GT_v7
4. 5. 6.
Calibration Scale configuration Scale configuration
Calibration points Calibration Calibration

1 Zero acquisition.
1 ~ 8
Unload the platform and press Zero acquisition underway
1 2 3 / Esc the OK button. Please wait...

4 5Bilancia
6 1 * Calibrazione OK
Cancel Calibrazione
ADC=308261
7 8 9 - BkSp
30%
C 0 . + OK

Esc Esc Esc

7. 8. 9. Load the weighbridge and distribute


Calibration Scale configuration the mass evenly.
Calibration Point 1 weight

30000
0 ~ 999999
Acquisition of point 1.
Insert the weight value. 1 2 3 / Esc

Bilancia 1
Cancel OK 4 Calibrazione
5 6 *

7 8 9 - BkSp
kg kg kg kg kg
C 0 . + OK

Esc Esc

10. 11. 12.


Scale configuration Calibration Scale configuration
Calibration Calibration Calibration

Acquisition of point 1. Acquisition of point 1 underway. Calibration terminated


Load the platform with 30000kg Please wait... successfully.
and press the OK button. Punti / div = 84
ADC=308261
Calibrazione OK
Cancel Bilancia 1 Calibrazione
OK

30%

Esc Esc Esc

AF03GT 6
TECH_MAN_CAL_ENG_AF03GT_v7
F. Repeat the procedure from step C for calibration of the second weighbridge (if present).

G. Perform backup
1. 2. 3.
Technical set-up Backup & Restore Backup & Restore
Configuration backup
... ...
... ...
Backup all settings?
... ...
Backup & Restore Backup of the configuration Calibrazione Yes
No
... ...
... ...
Esc Esc Esc

i It is possible to enter a protection password to restore the stored parameters.

i Saving has completed correctly only if the asterisk is present: Backup of the configuration (*)

H. Receive calibration with Dinitools.

I. Exit setup and save.

1. 2.
Technical set-up Technical set-up
Technical set-up
...
... Setup changed.
Save changes before restart the
... indicator?

... Cancel ...No Yes


...
...
Esc Esc

AF03GT 7
TECH_MAN_CAL_ENG_AF03GT_v7
Analogue calibration with equalisation

PREMISE

CASE 1: The weighbridge load cells are connected to the 4 indicator channels.

ch1 ch4
Load cells
Weighbridge
ch2 ch3

CASE 2: Scale composed of various weighbridges in sequence, each connected to a channel of the indicator.

Up to 48
ch1 ch2 ch3 ch4 1000 Ohm cells.

i Each weighbridge shall be individually equalised via junction box.

Procedure:

A. With switched-off indicator and power supply unplugged, connect the scale to the indicator.

B. Turn on the terminal, enter the setup menu and check that the signal coming from the scale is stable and increases with weight
increase.

1. 2. 3.
Technical set-up
...
Diagnostics
... kg
Channel
SCALE 1

ADC
15000
min
W1
A
f=7
d=74
MAX
... ... 1 344673 308333 308333

... ...

Diagnostic Weight test


... ...
... ...
Esc Esc Esc ADC-mV

AF03GT 8
TECH_MAN_CAL_ENG_AF03GT_v7
C. Select the number of scales and load cell type.

1. 2.
Calibration Scale selection
... ...
... ...
... Number of scales =1
Scale selection Type of load cell = Analogue

... Number of channels (Scale 1) = 2 ... 4


... ...
Esc Esc

D. Set scale metric parameters (capacity/division) etc.

1. 2. 3.
Calibration Scale configuration Parameters

... ... Number of decimals


... ... Unit of measure
... Parameters Range number
Scale 1 ... Range 1 division
Scale 2 ... Capacity / Range 1
... ... Range 2 division
Esc Esc Esc

E. Perform the guided equalisation procedure.

Scale configuration
...

...
Equalisation
...
...
...
Esc

For CASE 1: equalisation aims to optimise the angles of the scale.

kg
kg kg

kg kg kg = 1/8 Max (kg)

For CASE 2: equalisation aims to make each weighbridge the same, in order for the weight measured on each individual one
to be the same.

kg

kg kg kg kg
kg = 1/8 Max (kg)

AF03GT 9
TECH_MAN_CAL_ENG_AF03GT_v7
F.
1. 2. 3.
Calibration Scale configuration Scale configuration
Calibration

Insert the number of calibration


points other than the zero.

Scale 1 Calibration Calibrazione OK


Cancel

Esc Esc Esc

4. 5. 6.
Calibration Scale configuration Scale configuration
Calibration points Calibration Calibration

1 Zero acquisition.
1 ~ 8
Unload the platform and press Zero acquisition underway
1 2 3 / Esc the OK button. Please wait...

4 5Bilancia
6 1 * Calibrazione OK
Cancel Calibrazione
ADC=308261
7 8 9 - BkSp
30%
C 0 . + OK

Esc Esc Esc

7. 8. 9. Load the weighbridge and distribute


Calibration Scale configuration the mass evenly.
Calibration Point 1 weight

30000
0 ~ 999999
Acquisition of point 1.
Insert the weight value. 1 2 3 / Esc

Bilancia 1
Cancel OK 4 Calibrazione
5 6 *

7 8 9 - BkSp
kg kg kg kg kg
C 0 . + OK

Esc Esc

10. 10. 11.


Scale configuration Calibration Scale configuration
Calibration Calibration Calibration

Acquisition of point 1. Acquisition of point 1 underway. Calibration terminated


Load the platform with 30000kg Please wait... successfully.
and press the OK button. Punti / div = 84
ADC=308261
Calibrazione OK
Cancel Bilancia 1 Calibrazione
OK

30%

Esc Esc Esc

i Perform the procedure for each channel.

AF03GT 10
TECH_MAN_CAL_ENG_AF03GT_v7
G. Perform backup
1. 2. 3.
Technical set-up Backup & Restore Backup & Restore
Configuration backup

Backup all settings?

Backup & Restore Backup of the configuration Calibrazione Yes


No

Esc Esc Esc

i It is possible to enter a protection password to restore the stored parameters.

i Saving has completed correctly only if the asterisk is present: Backup of the configuration (*)

H. Receive calibration with Dinitools.

I. Exit setup and save.

Technical set-up Technical set-up


Technical set-up

Setup changed.
Save changes before restart the
indicator?

... Cancel ...No Yes

Esc Esc

AF03GT 11
TECH_MAN_CAL_ENG_AF03GT_v7
Setting and calibration of digital cell weighbridges

PREMISE
CASE 2:

BOOT
RS485 RS485
Load cells
Weighbridge
Junction box

Procedure:

A. With switched-off indicator and power supply unplugged, connect the scale to the weighing terminal 3590.
SENSOR

ChipRS232
485
COM3 SIG-
GND
+5V

SIG+ J3
485 EXC-
-
-

EXC+ J2
ON
RS232 RS232 CH4 CH3 CH2 CH1
COM1 COM2 Dip switch to enable termination resistor

i
SEN + /
SEN + /
EARTH

REF +

EXC+
+Vdc
CTS2

REF -
GND

SIG+
GND

EXC-
A(+)

SIG-
RX1
TX1

TX2

B(-)
IN2

For digital cells operation installation of


RX

chip 485 is required.


13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

A(+) B(-)

B. Select the number of scales and load cell type.


1. 2.
Calibration Scale selection

Number of scales =1
Type of load cell = RCPTD, RCD, SP/CS, CB50X-DL, RC3D
Scale selection RL5426/RL5416, C16i, ZSFD, DGX, WWS
Number of cells = 1 - 16

Esc Esc

C. Set digital cell serial number.

Premise
Number the load cells for easier identification during maintenance, for example:

1 2 3 4 2 4 6 8 10 12 14 16

5 6 7 8 1 3 5 7 9 11 13 15

AF03GT 12
TECH_MAN_CAL_ENG_AF03GT_v7
1. 2.
Scale selection Serial number configuration
... ...
... ...
= Enter for each load cell its serial number,
Serial number configuration Cell n
displayed on the casing of the cell.
... ...
... ...
... Address assignment
Esc Esc

D. Perform the load cell automatic numbering procedure:


1.
Serial number configuration
...
...
Cell n
...
...
Address assignment = The indicator assigns to each load cell its number, which will identify it in each
programming or diagnostics menu.
Esc
If the operation does not succeed, check connection between cells and
indicator.

i If the operation succeeds, set the dip switch to “ON” to enable termination resistor.

E. Set scale metric parameters (capacity/division) etc.

1. 2. 3.
Calibration Scale configuration Parameters

... Number of decimals


... Unit of measure
Parameters Range number
Scale 1 ... Range 1 division
... Capacity / Range 1
... Range 2 division
Esc Esc Esc

F. Perform equalisation following the indicator’s guided procedure.

1.
Scale configuration Scale configuration
Equalisation

Excute load cells equalization?


Equalisation

No Yes

Esc Esc

AF03GT 13
TECH_MAN_CAL_ENG_AF03GT_v7
Example of a 4 cell weighbridge: equalisation aims to Example of an 8 cell weighbridge: equalisation aims to
optimise each cell of the scale. optimise each cell of the scale

kg kg
kg kg kg kg kg kg

kg kg
kg = 1/8 Max (kg) kg kg kg kg
kg = 1/8 Max (kg)

G. Perform calibration.

1. 2. 3.
Calibration Scale configuration Scale configuration
Calibration

Insert the number of calibration


points other than the zero.

Scale 1 Calibration Cancel OK

Esc Esc Esc

4. 5. 6.
Calibration Scale configuration Scale configuration
Calibration points Calibration Calibration

1 Zero acquisition.
1 ~ 8
Unload the platform and press Zero acquisition underway
1 2 3 / Esc the OK button. Please wait...

4 5 6 * Cancel OK ADC=308261
7 8 9 - BkSp
30%
C 0 . + OK

Esc Esc Esc

7. 8. 9. Load the weighbridge and distribute


Calibration Scale configuration the mass evenly.
Calibration Point 1 weight

30000
0 ~ 999999
Acquisition of point 1.
Insert the weight value. 1 2 3 / Esc

Cancel OK 4 5 6 *

7 8 9 - BkSp
kg kg kg kg kg
C 0 . + OK

Esc Esc

AF03GT 14
TECH_MAN_CAL_ENG_AF03GT_v7
10. 11. 12.
Scale configuration Calibration Scale configuration
Calibration Calibration Calibration

Acquisition of point 1. Acquisition of point 1 underway. Calibration terminated


Load the platform with 30000kg Please wait... successfully.
and press the OK button. Punti / div = 84
ADC=308261
Calibrazione OK
Cancel Bilancia 1 Calibrazione
OK

30%

Esc Esc Esc

i Perform the procedure for each scale.

H. Perform backup
1. 2. 3.
Technical set-up Backup & Restore Backup & Restore
Configuration backup

Backup all settings?

Backup & Restore Backup of the configuration Calibrazione Yes


No

Esc Esc Esc

i It is possible to enter a protection password to restore the stored parameters.

i Saving has completed correctly only if the asterisk is present: Backup of the configuration (*)

J. Receive calibration with Dinitools.

K. Exit setup and save.

Technical set-up Technical set-up


Technical set-up

Setup changed.
Save changes before restart the
indicator?

... Cancel ...No Yes

Esc Esc

AF03GT 15
TECH_MAN_CAL_ENG_AF03GT_v7
Calibration of a WWS wireless weighing system

PREMISE

System diagram:

i This configuration lets


you manage up to 16
platforms.

CAL

J1

DISPLAY
POWER

Vin
1

GND
2

+BAT
3

GND
4

OUTPUT

+Vaux
5

OUT
6

COM
Procedure : OUT1
7

OUTPUT

OUT2
8

A. OUT3
9

Calibrate and set metric parameters of each WWS platform. Check that each platform is configured on the same
INPUT

OUT4
10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28

ON

radio channel (channel “0”)


DISPLAY

IN
B. COM
With indicator switched-off and disconnected from the electrical grid, connect the radio module to the AUX port.
IN1 TX
RX
COM1
RS232

IN2 -
+5V
TX1 -
GND
RX1 CPU Board Radio module
GND
SENSOR

MODULO
COM2
RS232

CTS2
TX2 RADIO
RX
GND 1 TX RX 1
2 RX TX 2
485

+Vdc
COM3
RS232

3 3
A(+) 4 5÷12V 5÷12V 4
BOOT

5 5
B(-) 6 GND GND 6
ON

EARTH
CH4 CH3 CH2

SIG+

AF03GT SIG- 16
SEN + /
REF + TECH_MAN_CAL_ENG_AF03GT_v7
SEN + /
C. Set the instrument’s radio channel

1. 2. 3.
Technical setup Serial ports Radio Frequency interface
Input texts Serial ports function mode Communication port
Serial ports Printer port configuration
Printout Pc port configuration
Ext. keyboard - Barcode reader Auxiliary port configuration
Remote scale Radio Frequency interface
Digital outputs
Esc Esc Esc

4. 5. 6.
Communication port Radio Frequency interface Radio Frequency interface
Channel
Disabled Communication port Communication port
PC port Channel Channel
0
0 ~ 7

Printer port 1 2 3 / Esc

Auxiliary port 4 5 6 *

7 8 9 - BkSp

C 0 . + OK

Cancel OK Esc Esc

D. Select the number of WWS which make up the system.

1. 2.
Calibration Scale selection

Number of scales =1
Scale selection Type of load cell = WWS
Number of cells (Scale 1) = Corresponds to the number of WWS platforms
which make up the system, set from 1 - 16.

Esc Esc

E. Establish communication between indicator and WWS platform.

1. 2. Set the gravity value of the area of use.


WWS configuration WWS configuration
Remote WWS configuration Gravity value

WWS configuration acquisition 9.80390


9.75001 ~ 9.84999

1 2 3 / Esc
= Acquires metric parameters from
4 5 6 *
the WWS and modifies them on
the indicator. 7 8 9 - BkSp
If the operation does not suc-
C 0 . + OK
ceed, check connection with the
Esc Esc
platforms.

AF03GT 17
TECH_MAN_CAL_ENG_AF03GT_v7
i This operation overwrites the data in the memory.

i
If WWS platforms are approved for use in relation with third parties:
• total system capacity must be equal to the sum of each individual WWS platform’s capacity.
• system division must be equal to the sum of each individual WWS platform’s division rounded up to the highest one.
• set these parameters to be able to approve the system.
E.g.: in a system consisting of 8 WWS platforms with “Max.=15,000kg” and “e=5kg”, maximum capacity will be
“Max.=120,000kg” and the division “e=50kg”.

If the platforms are not approved:


• total system capacity must be equal to the sum of each individual WWS platform’s capacity.
• system division equals division of the single WWS platform.
E.g.: in a system consisting of 8 WWS platforms with “Max.=15,000kg” and “e=5kg”, maximum capacity will be
“Max.=120,000kg” and the division “e=5kg”.

The system’s total capacity and division can be modified to create specific internal use systems (see step F).

F. Set indicator metric parameters (if required).

1. 2. 3.
Calibration Scale configuration Parameters

... Number of decimals


... Unit of measure
Parameters Range number
Scale 1 ... Range 1 division
... Capacity / Range 1
... Range 2 division
Esc Esc Esc

G. Perform backup
1. 2. 3.
Technical set-up Backup & Restore Backup & Restore
Configuration backup

Backup all settings?

Backup & Restore Backup of the configuration Calibrazione Yes


No

Esc Esc Esc

AF03GT 18
TECH_MAN_CAL_ENG_AF03GT_v7
i It is possible to enter a protection password to restore the stored parameters.

i Saving has completed correctly only if the asterisk is present Backup of the configuration (*)

H. Receive calibration with Dinitools.

I. Exit setup and save.

Technical set-up Technical set-up


Technical set-up

Setup changed.
Save changes before restart the
indicator?

... Cancel ...No Yes

Esc Esc

AF03GT 19
TECH_MAN_CAL_ENG_AF03GT_v7
Calibration of a dual digital weighbridge system

PREMISE

System diagram:

Weighbridge

RS485

Load Junction
cells box

i It is possible to install weighing systems with more than one weighbridge based on the number of cells of each
(Max. 16 digital cells, max. 8 CCI AD cells).

With this configuration the weight of each platform is displayed separately.

GROSS Weight “A”, Weight “B”


or Sum “A” + “B”
kg
A B (enabling “A+B” function).
kg kg See page 23
PRESET NOTES
CUSTOMER SUPPLIER
MATERIAL
VEHICLE
FREE PLATE
0 0
1. WEIGH 2. WEIGH PRINT >>

AF03GT 20
TECH_MAN_CAL_ENG_AF03GT_v7
Procedure:

A. With switched-off indicator and power supply unplugged, connect the scale to the weighing terminal 3590.
SENSOR

Chip 485
RS232
COM3 SIG-
GND
+5V

SIG+ J3
485 EXC-
-
-

EXC+ J2
ON
RS232 RS232 CH4 CH3 CH2 CH1
COM1 COM2 Dip switch to enable termination resistor.

SEN + /
SEN + /
EARTH

REF +

EXC+
+Vdc
CTS2

REF -
GND

SIG+
GND

EXC-
A(+)

SIG-
RX1
TX1

TX2

B(-)
IN2

RX

For digital cells operation


13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 installation of chip 485 is required.

A(+) B(-)

B. Select the number of scales and load cell type

1. 2.
Calibration Scale selection

Number of scales =2
Scale selection Type of load cell = RCPTD, RCD, SP/CS, CB50X-DL, RC3D
RL5426/RL5416, C16i, ZSFD, DGX, WWS
Number of cells (Scale 1) = 1 - 16
Number of cells (Scale 2) = 1 - 16
Esc Esc

i Total of selectable cells in the dual scale configuration is “16”.

AF03GT 21
TECH_MAN_CAL_ENG_AF03GT_v7
C. Set digital cell serial number.

Premise
Number the load cells for easier identification during maintenance, for example:

2 4 6 8 10 12 14 16

1 3 5 7 9 11 13 15

1. 2.
Scale selection Serial number configuration
... ...
... ...
= Enter for each load cell its
Serial number configuration Cell n
serial number, displayed on the
... ... casing of the cell.

... ...
... Address assignment
Esc Esc

D. Perform the load cell automatic numbering procedure:


1.
Serial number configuration
...
...

Cell n
...
...
Address assignment = The indicator assigns to each load cell its number, which will identify it in each
programming or diagnostics menu
Esc
If the operation does not succeed, check connection between cells and indicator.

i If the operation does not succeed, set the dip switch to “ON” to enable termination resistor.

E. Set scale metric parameters (capacity/division) etc.

1. 2. 3.
Calibration Scale configuration Parameters

... Number of decimals


... Unit of measure
Parameters Range number
Scale 1 ... Range 1 division
Scale 2 ... Capacity / Range 1
... Range 2 division
Esc Esc Esc

Example of dual scale weighbridge configuration: Number of decimals = 0; Unit of measure = kg; Range number = 2;
Range 1 division = 10; Capacity / Range 1 = 30000; Range 2 division = 20; Capacity / Range 2 =60000.

AF03GT 22
TECH_MAN_CAL_ENG_AF03GT_v7
F. Perform equalisation following the indicator’s guided procedure.

1.
Scale configuration Scale configuration
Equalisation

Excute load cells equalization?


Equalisation

No Yes

Esc Esc

For CASE 1: equalisation aims to optimise the angles of the scale.

kg
kg kg

kg kg kg = 1/8 Max (kg)

For CASE 2: equalisation aims to make each weighbridge the same, in order for the weight measured on each individual one
to be the same.

kg

kg kg kg kg
kg = 1/8 Max (kg)

G. Perform calibration.

1. 2. 3.
Calibration Scale configuration Scale configuration
Calibration

Insert the number of calibration


points other than the zero.

Scale 1 Calibration Calibrazione OK


Cancel
Scale 2

Esc Esc Esc

4. 5. 6.
Calibration Scale configuration Scale configuration
Calibration points Calibration Calibration

1 Zero acquisition.
1 ~ 8
Unload the platform and press Zero acquisition underway
1 2 3 / Esc the OK button. Please wait...

4 5Bilancia
6 1 * Calibrazione OK
Cancel Calibrazione
ADC=308261
7 8 9 - BkSp
30%
C 0 . + OK

Esc Esc Esc

AF03GT 23
TECH_MAN_CAL_ENG_AF03GT_v7
7. 8. 9. Load the weighbridge and distribute
Calibration Scale configuration the mass evenly.
Calibration Point 1 weight

30000
0 ~ 999999
Acquisition of point 1.
Insert the weight value. 1 2 3 / Esc

Bilancia 1
Cancel OK 4 Calibrazione
5 6 *

7 8 9 - BkSp

kg kg kg kg kg
C 0 . + OK

Esc Esc

10. 11. 12.


Scale configuration Calibration Scale configuration
Calibration Calibration Calibration

Acquisition of point 1. Acquisition of point 1 underway. Calibration terminated


Load the platform with 30000kg Please wait... successfully.
and press the OK button. Punti / div = 84
ADC=308261
Calibrazione OK
Cancel Bilancia 1 Calibrazione
OK

30%

Esc Esc Esc

i Perform the procedure for each scale.

How to display the sum of “A”+”B”


With this configuration it is possible to display simultaneously the weights of the two platforms and their sum.

1. 2. 3.
Technical set-up AF03 functions A+B function mode

Disabled

Enabled

AF03 functions
Operation A+B

Esc Esc Cancel OK

AF03GT 24
TECH_MAN_CAL_ENG_AF03GT_v7
H. Perform backup
1. 2. 3.
Technical set-up Backup & Restore Backup & Restore
Configuration backup

Backup all settings?

Backup & Restore Backup of the configuration Calibrazione Yes


No

Esc Esc Esc

i It is possible to enter a protection password to restore the stored parameters.

i Saving has completed correctly only if the asterisk is present: Backup of the configuration (*)

I. Receive calibration with Dinitools.

J. Exit setup and save.

Technical set-up Technical set-up


Technical set-up

Setup changed.
Save changes before restart the
indicator?

... Cancel ...No Yes

Esc Esc

AF03GT 25
TECH_MAN_CAL_ENG_AF03GT_v7
Calibration of a dual scale system consisting of wireless WWS platforms

PREMISE

System diagram:

i This configuration lets


you manage up to
16 platforms.

Scale 1
Traction unit

Bilancia 2
Trailer

Trailer Traction unit

With this configuration the weight of each platform is displayed separately.

GROSS Weight “A”, Weight “B” or


Sum “A” + “B” (enabling
kg
A B “A+B” function).
kg kg See page 23
PRESET NOTES
CUSTOMER SUPPLIER
MATERIAL
VEHICLE
FREE PLATE
0 0
1. WEIGH 2. WEIGH PRINT >>

AF03GT 26
TECH_MAN_CAL_ENG_AF03GT_v7
GND

OUTPUT
+Vaux

5
Procedure: OUT

6
COM
OUT1

7
A.

OUTPUT
Calibrate and set metric parameters of each WWS platform. Check that each platform is configured on the same radio channel
OUT2

8
(channel “0”)
OUT3

9
B. With indicator switched-off and disconnected from the electrical grid, connect the radio module to the AUX port.

INPUT
OUT4

10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

ON

DISPLAY
IN
COM
IN1 TX
RX

COM1
RS232
IN2 -
+5V
TX1 -
GND
RX1 CPU Board Radio module
GND

SENSOR
MODULO
COM2
RS232

CTS2
TX2 RADIO
RX
GND 1 TX RX 1
2 RX TX 2
485

+Vdc
COM3
RS232

3 3
A(+) 4 5÷12V 5÷12V 4

BOOT
5 5
B(-) 6 GND GND 6
ON

EARTH
CH4 CH3 CH2 CH1

SIG+
SIG-
SEN + /
REF +
SEN + /
REF -
C. EXC+the platforms for weighing the traction unit from those for weighing the trailer.
Separate and identify
This division shall EXC-
be followed for each weighing procedure.
EXC-
SIG-
EXC+
SIG+

i
J2

J3

Traction unit
= Scale 1
10 8 6 4 2 WWS group
9 7 5 3 1

Traile
= Scale 2
WWS group

Trailer WWS group Traction unit WWS group

D. Select the number of scales and number of WWS platforms for each scale.

1. 2.
Calibration Scale selection

Number of scales = 2 (“scale” means a group of WWS)


Scale selection Type of load cell = WWS
Number of cells (Scale 1) = 1 - 16 Enter the number of cells of the
traction unit group and of the
Number of cells (Scale 2) = 1 - 16
trailer group
Esc Esc

i The total of selectable WWS platforms in the dual scale configuration is “16”.

AF03GT 27
TECH_MAN_CAL_ENG_AF03GT_v7
E. Establish communication between indicator and WWS platform.
1. 2. Set the gravity value of the area of use.

WWS configuration WWS configuration


Gravity value
Remote WWS configuration
WWS configuration acquisition
9.80390
9.75001 ~ 9.84999

1 2 3 / Esc

= Acquires metric parameters from 4 5 6 *


the WWS and modifies them on the
7 8 9 - BkSp
indicator.
If the operation does not succeed, C 0 . + OK

Esc check connection with the platforms. Esc

i This operation overwrites the data in the memory.

i
If WWS platforms are approved for use in relation with third parties:
• total system capacity must be equal to the sum of each individual WWS platform’s capacity.
• system division must be equal to the sum of each individual WWS platform’s division rounded up to the highest one.
• set these parameters to be able to approve the system.
E.g.: in a system consisting of 8 WWS platforms with “Max.=15,000kg” and “e=5kg”, maximum capacity will be
“Max.=120,000kg” and the division “e=50kg”.

If the platforms are not approved:


• total system capacity must be equal to the sum of each individual WWS platform’s capacity.
• system division equals division of the single WWS platform.
E.g.: in a system consisting of 8 WWS platforms with “Max.=15,000kg” and “e=5kg”, maximum capacity will be
“Max.=120,000kg” and the division “e=5kg”.

The system’s total capacity and division can be modified to create specific internal use systems (see step F).

F. Set indicator metric parameters (if required).

1. 2. 3.
Calibration Scale configuration Parameters

... Number of decimals


... Unit of measure
Parameters Range number
Scale 1 ... Division of range 1
Scale 2 ... Capacity / Range 1
...
Esc Esc Esc

AF03GT 28
TECH_MAN_CAL_ENG_AF03GT_v7
How to display the sum of “A”+”B”
With this configuration it is possible to display simultaneously the weights of the two platforms and their sum.

1. 2. 3.
Technical set-up AF03 functions A+B function mode

Disabled

Enabled

AF03 functions
A+B function mode

Esc Esc Cancel OK

G. Perform backup
1. 2. 3.
Technical set-up Backup & Restore Backup & Restore
Configuration backup
... ...
... ...
Backup all settings?
... ...
Backup & Restore Backup of the configuration Calibrazione Yes
No
... ...
... ...
Esc Esc Esc

i It is possible to enter a protection password to restore the stored parameters.

i Saving has completed correctly only if the asterisk is present Backup of the configuration (*)

H. Receive calibration with Dinitools.

I. Exit setup and save.

Technical set-up Technical set-up


Technical set-up

Setup changed.
Save changes before restart the
indicator?

... Cancel ...No Yes

Esc Esc

AF03GT 29
TECH_MAN_CAL_ENG_AF03GT_v7
Notes

This publication, or portions thereof, may not be duplicated without written permission from the Manufacturer. All information herein
is based on the data available at the time of publication. The Manufacturer reserves the right to make changes to its products at
any time without notice and without incurring any penalty. We therefore recommend that you always check for any updates.
The individual in charge of operating the scale must ensure that all safety regulations in force in the country of use are applied, ensuring that the appliance is used in
accordance with the purpose it is intended for and to avoid any danger for the user.
The Manufacturer declines any liability arising from any weighing operation errors.

AF03GT 30
TECH_MAN_CAL_ENG_AF03GT_v7
HEAD OFFICE
Via Della Fisica, 20

www.diniargeo.com
Via Dell’Elettronica, 15
SERVICE ASSISTANCE

41042 Spezzano di Fiorano, Modena - Italy


41042 Spezzano di Fiorano, Modena - Italy
Tel. +39 0536 843418 - Fax +39 0536 843521

Tel. +39 0536 921784 - Fax +39 0536 926654


Authorised service centre stamp
The information in this document is approximate and can be subject to variations without prior notice by Dini Argeo, with respect of the norms in force. The official
TECH_MAN_CAL_ENG_AF03GT_v7
technical data is available in the updated version on the www.diniargeo.com website or by contacting the Dini Argeo Customer Service.
Rev. 28.01.2019 - 07.10.xx