Revision History:

Version Date Created By Details


1.0 2020/12/04 Brian Cao

Ref: ZN-GE-0001 (Example) Version: 1.0 (Example) page 1 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

Contents
1. Backup ECG Data & Export Setting......................................................................... 3

1.1 Backup ECG Data .............................................................................................. 3

1.2 Export Setting ..................................................................................................... 4

2. Preparation for Upgrade ............................................................................................ 5

2.1 Hardware and Software needed: ........................................................................ 5

2.2 USB Formatting ................................................................................................. 5

2.2 Upgrade File Preparation ................................................................................... 6

3. Upgrade Procedure .................................................................................................... 7

3.1 Plug-in the Flash Disk ........................................................................................ 7

3.2 Upgrade Procedure for SE-601 Series, SE-12 Series and SE-1201................... 7

3.3 Upgrade Procedure for SE-18 .......................................................................... 8

3.4 Upgrade Procedure for SE-301 ........................................................................ 8

3.5 Update Program ............................................................................................... 9

4 Self Check and Restore Preset Data ........................................................................... 10

4.1 Check the Device Info ...................................................................................... 10

4.2 Import the Backup Data for SE-601, SE-12 Series and SE-1201. ................... 11

4.3 Import the Backup Data for SE-18................................................................... 12

Ref: ZN-GE-0001 (Example) Version: 1.0 page 2 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

1. Backup ECG Data & Export Setting


1.1 Backup ECG Data
Export Setting & Backup Internal ECG Data in advance is necessary when it comes to
software upgrading, in case of booting failure or system collapse.
 Prepare a vacant flash disk whose storage is no more than 8G, format it into FAT32.
 Plug the flash disk into the USB port, enter the “File” interface of ECG machine,
and press “Export All” to export internal data for backup.
 For SE-18, please enter “Archives>>Export” to export internal data for backup

Figure 1. Back up ECG Data for SE-12 Series, SE-601 Series and SE-1201

Figure 2. Back up ECG Data for SE-18.

Ref: ZN-GE-0001 (Example) Version: 1.0 page 3 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

1.2 Export Setting


 Back to Menu and enter “Maintenance” option.
 Choose “Export Settings” option to back up the setting files into flash disk, which
should be restore when you accomplish the whole upgrade procedure.
 For SE-18, please choose “Setup>>Maintenance>>Basic Setup>>Export Setup” to
proceed the backup.

Figure 3. Export Settings for SE-12 Series, SE-601 Series and SE-1201

Figure 4. Export Settings for SE-18

Ref: ZN-GE-0001 (Example) Version: 1.0 page 4 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

2. Preparation for Upgrade

2.1 Hardware and Software needed:

Flash Disk X1 Upgrade File


(Recommend: 1GB to 8GB)

Note: The Upgrade File is provided by certificates engineers from EDAN service team, file can be
varied according to different requirements.

2.2 USB Formatting


 Please format the USB Flash Disk to FAT32 format in advance and make sure there
is no additional file in the disk.

Figure 5. USB formatting

Ref: ZN-GE-0001 (Example) Version: 1.0 page 5 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

2.2 Upgrade File Preparation


 Please extract the files package, copy the extracted files to the root directory of USB
Flash Disk.

Note: Please place all upgrade files directly in the root directory of the USB flash drive, instead of
placing them in the folder in the root directory.

 Make sure no other irrelevant files in the Flash Disk.

Figure 6. Extract the files package

Note: The Upgrade File is provided by certificates engineers from EDAN service team, file can be
varied according to different requirements.

Ref: ZN-GE-0001 (Example) Version: 1.0 page 6 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

3. Upgrade Procedure

3.1 Plug-in the Flash Disk


 Plug the Flash Disk with software files into the USB port of ECG machine.

Figure 7. Insert the USB disk to the ECG machine.

WARNING: Only qualified service personnel is eligible to perform software upgrade.

3.2 Upgrade Procedure for SE-601 Series, SE-12 Series and SE-1201
 Enter the Menu of ECG machine
 Press “F1>>Password: 999999” and get access to the Advanced Setup Menu.
 Choose “Upgrade>>Upgrade All” to proceed the upgrade procedure.

Figure 8. Upgrade Procedure for SE-601 Series, SE-12 Series and SE-1201

Ref: ZN-GE-0001 (Example) Version: 1.0 page 7 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

3.3 Upgrade Procedure for SE-18


 Enter the Menu of ECG machine
 Choose “Maintenance>>Advanced Setup>>Password: 999999” and get access to
the Advanced Setup Menu.
 Choose “Upgrade>>Upgrade All” to proceed the upgrade procedure.

Figure 9. Upgrade Procedure for SE-18

3.4 Upgrade Procedure for SE-301


 Enter the Menu of ECG machine
 Press “Advanced Setup>>Password: 999999” and get access to the Advanced Setup
Menu.
 Choose “Upgrade>>Upgrade All” to proceed the upgrade procedure.

Figure 10. Upgrade Procedure for SE-301

Ref: ZN-GE-0001 (Example) Version: 1.0 page 8 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

3.5 Update Program


 The Upgrade procedure will be initiated automatically, please wait patiently and
make sure the power supply is stable.
 Wait until process done. Then the ECG will reboot itself automatically, which
indicates the upgrade has been successfully done.

Figure 11. Update Process OK

Ref: ZN-GE-0001 (Example) Version: 1.0 page 9 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

4 Self Check and Restore Preset Data

4.1 Check the Device Info


 If the upgrade is successful, the device will boot up normally.
 Please enter the Advanced Setup Menu of ECG machines respectively, choose
“Device Info” and check there current software version.
 If the software version hasn’t been upgraded to expectant version, perform the
upgrade procedures once again.

Figure 12. Device Info

 Re-plug the Flash Disk with backup files to the ECG machine that has been
upgraded.

Figure 13. Re-plug the Flash Disk with backup files.

Ref: ZN-GE-0001 (Example) Version: 1.0 page 10 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

4.2 Import the Backup Data for SE-601, SE-12 Series and SE-1201.
 Enter the “File” interface of ECG machine, and press “Import” to import the backup
data.

Figure 14. Import the Backup Data for SE-601, SE-12 Series and SE-1201

 Back to Menu and enter “Maintenance” option. Choose “Import Settings” option to
restore the backup setting files from Flash Disk, should be restore when you
accomplish the whole upgrade procedure.

Figure 15. Import Settings

Ref: ZN-GE-0001 (Example) Version: 1.0 page 11 of 12


深圳市理邦精密仪器股份有限公司
EDAN Instruments, Inc.
Document Name: Instructions for ECG Machine Software Upgrade Date: Dec 7th 2020

4.3 Import the Backup Data for SE-18


 For SE-18, please enter “Archives>>Import” to import the backup ECG data.

Figure 16. Import the Backup Data for SE-18

 For SE-18, please choose “Setup>>Maintenance>>Basic Setup>>Import Setup” to


proceed the restore.

Figure 17. Import Settings