Instructions For ECG Machine Software Upgrade
Instructions For ECG Machine Software Upgrade
Contents
1. Backup ECG Data & Export Setting......................................................................... 3
3.2 Upgrade Procedure for SE-601 Series, SE-12 Series and SE-1201................... 7
4.2 Import the Backup Data for SE-601, SE-12 Series and SE-1201. ................... 11
Figure 1. Back up ECG Data for SE-12 Series, SE-601 Series and SE-1201
Figure 3. Export Settings for SE-12 Series, SE-601 Series and SE-1201
Note: The Upgrade File is provided by certificates engineers from EDAN service team, file can be
varied according to different requirements.
Note: Please place all upgrade files directly in the root directory of the USB flash drive, instead of
placing them in the folder in the root directory.
Note: The Upgrade File is provided by certificates engineers from EDAN service team, file can be
varied according to different requirements.
3. Upgrade Procedure
3.2 Upgrade Procedure for SE-601 Series, SE-12 Series and SE-1201
Enter the Menu of ECG machine
Press “F1>>Password: 999999” and get access to the Advanced Setup Menu.
Choose “Upgrade>>Upgrade All” to proceed the upgrade procedure.
Figure 8. Upgrade Procedure for SE-601 Series, SE-12 Series and SE-1201
Re-plug the Flash Disk with backup files to the ECG machine that has been
upgraded.
4.2 Import the Backup Data for SE-601, SE-12 Series and SE-1201.
Enter the “File” interface of ECG machine, and press “Import” to import the backup
data.
Figure 14. Import the Backup Data for SE-601, SE-12 Series and SE-1201
Back to Menu and enter “Maintenance” option. Choose “Import Settings” option to
restore the backup setting files from Flash Disk, should be restore when you
accomplish the whole upgrade procedure.