0% found this document useful (0 votes)
19 views

Chol BS200

The document provides parameters for analyzing cholesterol levels using an enzymatic-spectrophotometric method on a BS-200 instrument. It details reagent preparation instructions, test parameters including reaction times and units of measurement, calibration procedures, and reference ranges for serum cholesterol analysis.

Uploaded by

michel
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
19 views

Chol BS200

The document provides parameters for analyzing cholesterol levels using an enzymatic-spectrophotometric method on a BS-200 instrument. It details reagent preparation instructions, test parameters including reaction times and units of measurement, calibration procedures, and reference ranges for serum cholesterol analysis.

Uploaded by

michel
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

CHOLESTEROL

Enzymatic-spectrophotometric
CHOLESTEROL OXIDASE/PEROXIDASE
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.

Instrument: BS-200 ( = Accent 200)


Reagent preparation
Reagent is ready to be used.

Parameters

Test name Chol R1 300µl


Test nº R2
Full name Cholesterol Sample Volume 3µl
Standard nº R1 Blank ----
Analy.Type End-point Mixed read Blank ----
Pri.Wave 510 Linearity range 0.3 - 1000
Secon.Wave - Linearity limit
Trend Ascending Substrate Limit
Reac time 0 - 18 Factor
Incuba Time □ prozone check q1 q2 q3 q4
Unit mg / dL
Precision 1 Pc Abs

Reference
N.o Gender Sample type Age Low.Lim Up.Lim
Serum 0 200

(*) values entered by user

Calibration
Rule
Rule One point Linear Calibrators
Sensivity Name Concentration
Replicates 3
Interval 99 days Std 1 (*)
Difference
Limit ----
SD 0
Blank response ----
Erros limit ----
(*)enter the
Coefficient 0 Standard value
Vers 0902