www.IetLtd.

com Proudly serving laboratories worldwide since 1979

CALL 001.847.913.0777 for Refurbished & Certified Lab Equipment

API 5000™ LC/MS/MS System


The API 5000™ LC/MS/MS System is today's most sensitive triple quadrupole mass
spectrometer for small molecule analysis.

• Highest sensitivity triple quadrupole mass spectrometer for small molecule


analysis.
• Lowest limits of detection available for DMPK and ADMET studies.
• Robust, high-throughput platform for all stages of pharmaceutical development.
• Unequalled accuracy and precision for quantitative analysis.

Reach the Lowest Limit of Detection


New innovations in sensitivity like QJet™ ion guide technology make it possible to reach
the lowest limits of detection. Combining QJet ion guide technology with the proven
TURBO V™ ion source and a powerful new generation of Analyst® software results in a
robust, high throughput platform ideal for all phases of pharmaceutical development.

Eliminate Cross-Contamination
Innovative QJet ion guide technology dramatically improves sensitivity by capturing and
focusing more ions generated by the efficient TURBO V ion source. And the TURBO V
source virtually eliminates cross-contamination, even with high-concentration samples.

Flexible Options Increase Throughput


The optional DuoSpray™ source provides an ideal tool for faster method development
and increased throughput. This source has both TurboIonSpray® ion source and APCI
inlets allowing optimal computer-controlled ionization parameters for each compound in
an LC run. Another option, the PhotoSpray® source for atmospheric pressure
photoionization increases your analysis possibilities.

Unattended High Throughput Operation


The API 5000™ system continues the tradition of Applied Biosystems/MDS Sciex
rugged, reliable triple quadrupole systems. Advanced automation features enable
continuous duty operation in regulated high throughput laboratories.

Expert Results - Even for Non-Experts


The API 5000™ system makes it easy for you to get you the answers you need - even if
you're just getting into mass spectrometry. From automated methods development to
quick, simple routine maintenance, it is not only the most powerful triple quad available,
it's the easiest to use too.

System Set-Up Specifications


Gas SuppliesGas Supplies: 100 psi high purity nitrogen, 100 psi zero grade air
Performance/Throughput
Mass Accuracy (with internal
0.01% over the entire mass range
reference)
Gas Supplies: 100 psi high purity nitrogen, 100 psi
Gas 1/Gas 2
zero grade air
Dynamic Range 1 cps to 4 x 106 cps (pulse counting)
Mass Range m/z 5 – 1250
Physical Specifications
Width 130 cm (53 in)
Depth 50 cm (20 in)
Weight141 kg (310.8 lb)
Height 52 cm (21.5 in)
Computer Requirements (PC)
Operating SystemWindows 2000/XP
CD-ROM CVC RW (CD RW capable)
Memory 1 GB RAM
CPU 3.2 GHz
Hard Disk 200GB
Display 19" Viewable Color monitor recommended
Power Requirements
2 separate 207 V minimum to 242 V maximum (3 plug recepticles
Voltage
minimum)
Circuit
15 A grounded single phase circuit
Current

www.IetLtd.com Proudly serving laboratories worldwide since 1979