DBB-27 Technical Manual
DBB-27 Technical Manual
DBB-27
Technical manual ver.3.0
IMPORTANT
This manual contains the installation method,
setting and adjusting items of DBB-27 and
others. Before start of operation, read this
manual fully to understand proper operation
procedures and cautions, and always operate
the equipment in right and safe manner. Having
read this manual, keep it together with the
warranty sheet at a prominent place near the
equipment where all concerned personnel can
refer to it.
NIICKISO
® «©0
NIKKISO CO.,LTD.
IMPORTANT
This manual is effective for the DBB-27 mounted with the software of
Version 3.0.
Please confirm not only the model name, but also the software
version.
1. The model name is mentioned in the rating plate at the rear panel.
2. The software version is confirmed by the following operation on the
LCD of the monitor panel.
Touch the monitor displayed on the Technician menu bar in the LCD
and the pull down menu appears at the lower side of the monitor.
Touch the I/O data (to be displayed when input the password) to open
the I/O screen. At the upper left side, the software version is shown.
Chapter 2 Switches
3.1. UFC self test and conductivity self test .................................................................... 3-1
9.2. Micro-computer test ...........---- ........................................................................................ 3 _2
3.3. Blood monitor self test ...................................................................................................... 3-3
3.4. Coupling leak test ............................................................................................................. 3-4
This indicates the presence of a serious danger that will cause death or
serious injury if the contents are not followed.
This indicates the presence of a dangerous condition that may cause death or
serious injury if the contents are not followed.
This indicates the presence of a condition that may result in damage to the
aution
equipment (physical damage) or incorrect treatment results if the contents
are not followed.
This indicates important points that the operator should know when
Or* Notes
operating this equipment, and important points about conditions that the
operator should remain aware of.
Anyone operating this equipment must first thoroughly read and understand this
manual before starting the trial run adjustments and initial operation.
You are required to use the DBB-27 only within the scope of this
technical manual chapter 1.2 “ Items to be kept before the trial run
“ and the operating instruction manual chapter 1.7 “ Specifications of
DBB-27
Nikkiso can only be responsible for the safety, reliability and
performance of the DBB-27 when it is used in direct accordance with
the items detailed in [Chapter 1.2 Precautions that must be observed
before the trial run] in the Technical Manual (this manual) and
[Chapter 1.7 Specifications of DBB-27] in the Operating Manual.
aution or case
transport or he storage for 15 weeks or longer,
please strictly keep above “ Environmental conditions “.
2) Equipotential earth
In case DBB-27 to be used or installed at the medical room whose electric facility
has the earth resistance of greater than 10 Q and less than 100 Q at its earth
electrode or equipotential earth is necessary by other reason, equipotential earth
shall be made by using the terminal for equipotential earth on the rear of DBB-27.
As for the location of the terminal for equipotential earth, please refer to Operating
instruction manual chapter 2.1.4 “Power supply panel/Connector panel
Do not use defibrillators while the equipment is connected to a
patient.
[Electric shock from the defibrillator may be conducted to the
device via the patient, and cause electric shock to persons in the
vicinity.!
3. Spacing
DBB-27 must be installed in a place where the location to the patient, operation,
maintenance and control are at most rational level. To ensure safe operation of
DBB-27, installed it on the flat floor where no objects that are likely to fall down or
hinder operation exist around DBB-27. Also, to connect the utilities (power supply,
water supply and drain ) , install DBB-27 so that the power cable ( 3.5 m ) and drain
hose ( 3 m ) are not stretched.
Do not use this equipment within a controlled area for radiation
or MRI. Mobile phones, transceivers, and other electric
communication devices must not be brought into the room in
which this equipment is installed, or into a neighboring machine
room or preparation room.
[Electromagnetic interference may cause malfunction.]
3. Water supply
1) Water
Use only the water suitable for the hemodialysis to operate DBB-27. The water to be
used for DBB-27 must be the water treated by the reverse osmosis system, pure
water and softened water.
2) Conditions
© Water supply pressure: 70~300 kPa ( 0.7~3.0 bar )
If water supply pressure is higher than above value, lower that by using the
reducing valve.
5. Drain
The drain system shall be made so that it can process the drain at used flow rate
( standard : 500 ml/min. ) according to the local rules and regulations of each district. Be
careful of the reverse siphon and the air accumulation between drain connector and
drain port. For the case that disinfecting Equid with high temperature is flown through
the drain pipe, use the drain pipe made of materials durable of high temperature.
Concerning the drain connection, please refer to the technical manual chapter 1.3.1
“ Installation “.
EF Notes For details on using the Dialysate filter, see the [Dialysate Filter
Instruction Manual]
1) Turn levers at top and bottom of the holder to perpendicular direction, and push in
dialysate filter to the holder at the back side of DBB-27.
lever
holder
Dialysate holder (EF-02)
I
A -“ ■
lever
[Fig. 1-2 Attachment diagram of dialysate filter ( EF-02 ) ]
2) Firmly install the EF-02 at the upper and the lower ports to the holder and lock with
lever. ( See the diagram below)
Turn levers after the dialysate pushed in enough to touch two points at
aution top and bottom to the holder, (points shown as * 1 in the figure. If the
filter is locked without enough pressing, it could lead to leakage.
lever
holder
*1
lever
silicon hose
Tpipe (for</> 5 )
To EF-02 holder
aution Please refer to the technical manual chapter 5.2 Adjustment items
table, 17 Selection of use.
Concerning the details on the operation method, please refer to the Technical Manual
Chapter 2.2.3 “Technicians switch screen (Alarm, test ) Item 12 “ UFC self test key”.
The leak test of dialysate filter cannot be carried out when DBB-27 is in
aution
following cases at the pipe self check;
1. In the Treatment preparation mode, the bypass connector is not
placed in the bypass connector holder.
2. The mode is Gas purge, Connect, Disconnect, Start or Stop.
aution
1. After the dialysate filter is mounted or exchanged, do not fail to carry
out the CF leak test and confirm that any liquid leak does not exist.
2. If any abnormality from the CF leak test is observed, take necessary
action for restoration, then carry out the CF leak test again to
confirm that no problem exists. The treatment must, thereafter, be
started.
3. To prevent over-ultrafiltration, always check for leak from the
dialysate filter and piping during the treatment and at the daily
check and inspection. If any leak is observed from the dialysate filter
or the piping, stop the treatment and take necessary action
immediately such as exchange the filters or piping parts.
L_
7. Signal, cable
If necessary, the connection of signal, cables shall be made.
1) External input
Connect the external signal, which runs or stops the hydraulic unit or the
extracorporeal line, to the external input connector.
2) Monitor output 2 ( External output 1 )
Connect a cable to the Monitor output 2 connector to transmit the output signal of
the DBB-27 operating status to the external auxiliary equipment.
3) Monitor outputl ( External output 2 )
Connect a cable to the Monitor output 1 connector to transmit the signal of DBB-27
operating status to displaying plate and the external auxiliary equipment. (Contact
output : Max DC24V/120 mA)
4) Nurse call
Connect the Nurse Call cable ( optional accessory ) to the Nurse Call connector.
5) Blood pressure measuring manual switch
Connect the Blood Measuring manual switch cable ( optional accessory ) to the Blood
Measuring switch connector.
6) Peripheral equipment 1,2
Connect a cable to the peripheral equipment communication connector to transmit
computer data.
EF Notes Use "External input" or "Monitor output 1 (output 2)" for the
supply line hot rinse program.
For details, refer to the instruction manual for breakout box
DSB-05/DSB-06.
8. Irrigator stand
Set the irrigator stand to the stand holder.
aution( The water supply hose must be the one attached to DBB-27 or the one
J specified by Nikkiso ( ID 8mm ).
15. Drain
Connect the drain connector to the drain device. The height of the drain port must be
below 30cm from the floor face. Use the attached hose or appropriate one specified by
Nikkiso ( ID 8mm ). drain hose
air hole
below 30cm
I di ain duct
[Fig. 1-6 drain connection]
To prevent siphon phenomenon from the drain duct, make the air hole
to release the drain duct to the atmosphere. Let the drain hose not
touch the drain liquid. Do not touch carelessly the drain duct as hot
disinfectant may flow through it. When high temperature disinfecting
liquid is flown through the drain pipe, use the material durable to it.
16. Fuse i
Insert the battery fuses to the fuse holder and lock.
Be sure to carry on the insertion of fuses and locking of the fuse holder.
If these are not done surely, fuses may be broken causing contact failure
Notes
______ in f° r screen,
A s the
name m
please °d-
e suc as
? rese t toManual
refer tohInstruction be indicated
chapteron 7.2
the “status
Manualbar
rinse mode _________________________________________________
Mode screen opens.
5. Touch the |Rinse| key for 1.5 sec. or longer. The color of the |Rinse| key changes.
6. Having entered the Rinse mode, DBB-27 starts Rinse operation.
7. The Rinse process is performed for 10 min. or longer.
After power is turned on, make rinse at least for 5 minutes until the
equipment has filled with water. In order to prevent air from entering,
perform rinse without heating. Observe DBB'27 for leaks, strange
noises, or the presence of other abnormalities while rinsing.
8.Touch the Preset key for 1.5 sec. or longer. Key’s color changes and enters the Preset
mode.
9. Adjust the air detector voltage by reference to the Maintenance manual "5.15 Air
detector voltage adjustment".
11. Always check if a nut in hydraulic part were securely fastened after disinfection
performed at the end of installation.
To avoid cracking a nut due to much force, fasten manually without any
tools.
I , , ... <• .
aution To avoid a crack due to exceeded force, tighten it up with tools at 49N • cm
(5kgf7cm).
1.4. Transport
1.4.1. Indoor transport
1. Ordinary carrying
1) Release the caster break
2) Then, DBB-27 can be pushed or rotated to all direction.
( DBB-27 can be attached with transporting handle as option. )
2. Get over the small difference in level ( For example: Entrance to elevator )
To avoid being damaged or fallen down of DBB-27, move the equipment little by little and
get over the small difference in level slowly.
3. Get over the big difference in level ( For example: stairs )
When DBB-27 gets over the big difference in level, it must be lifted up. For this, at
least 4 persons are needed.
1) Remove the break of caster.
2) Incline DBB-27.
3) Transport DBB-27 lifting up.
Do not hold the monitor, external display lamp, or irrigator stand holder when
lifting.
4) Having placed DBB-27 on the floor, set it up by reverse procedures.
handle for transport
( optional accessory )
Chapter 2 Switches
This chapter contains the descriptions about the function of each switch. As an example,
the Technicians Screen is used. The switches described can be customized and displayed
on each screen. When they are customized to each screen, please refer to Chapter 4
“ Setting items “ of this Technical Manual.
(D UP RATE (t/h)
KEYS
■ The tag touched turns [blue] and selected switches are displayed on the screen.
■ When the key with shadow L-,-.;;J is operated ( ON/OFF ) from displayed switches,
touch it for 1.5 sec. or longer until the sound “ Pi —“ is heard.
■ When switch is in the state of ON, key color becomes any of green, orange, red, blue
or yellow. When switch is in the state of OFF, key color becomes grey.
■ When switch color is grey and the key letter is white, this key cannot be opetated.
1 5) (2 3
BPM
SINGLE BLD GIR.
DATA
NEEDLE DRAIN
RESET
NO It INITIAL
THP ZERO TUP
UF-FACT
CALC INITIAL
MEASURE
DIALYSAT
FLOW Na CONO.
SELECT CHECK
Notes UFC off does not mean that the UF rate is zero. At UFC off, UF pump
does not operate.
Do not operate the UFC off key when a patient is connected to this
equipment. [Doing so may cause an excess of UF.]
Warning Don’t touch the CONNECT OPEN AIR(PLATE) key in case of hollow
fiber dialyzer.
[It may cause excessive fluid removal or clogging of dialyzer.]
Don’t keep the |CONNECT OPEN AIR (PLATE)] key turned on more
than necessary, in case of Parallel-Plate dialyzer.
[It may cause excessive fluid removal.]
No Name Description
18 Na COND. This key is for extracting diasylate when Na concentrate priming and
CHECK Na concentrate infusion have been performed (step 1, or reflect the
infusion conductivity set at the start)
Execute the |Na COND. CHECK| key when the Na infusion profile is
[ON].
When performing Na infusion conductivity checks, with the dialyzer
coupEng set in the equipment, connect the concentrate connector to the
concentrate suction connector on the concentrate tank, and touch the
|Na COND. CHECK] key for more than 1.5 seconds to make it “Green”
(ON). ______________
The |Na COND. CHECK) key is only available during Preparation mode
and Preparation (washing) mode.
When in Preparation mode, the mode switches to washing mode.
When Na infusion conductivity check has completed, the mode
automatically switches to Preparation mode.
In the gas purge after Na infusion conductivity check, dialyzer priming
is possible with dialysate with the Na concentration infused (step 1, or
infusion conductivity set at the start is reflected).
If [Na infuse cond check func Use sei ] in setting Hospital 2 [27.3 Na
Infusion] is set to fo: Do not use], the |Na COND. CHECK| key is not
displayed on the screen.
(Sr* Notes
If the bypass connector is detached from the bypass connector holder,
the Na concentrate priming operation is interrupted.
Bfl S ffl Fs l HO
BLD II | VEN PRES | | OIBP | I|COUPLING||
OFFSET ZERO ZERO LEAK
MEASURE CALC CALC
J® 3
Notes The value of light receiving side voltage may not be normal if the air
remains inside the blood leak detector.
Notes When gas purge is performed during blood leak test, air may
cause blood leak detector test failure or blood leak detector dirt
alarm.
When blood leak test is performed manually after UFC self-test
or drain, air may cause blood leak detector test failure or blood
leak detector dirt alarm.
Perform gas purge for air removal prior to manual blood leak
test. i
No. Name Description
12 UFC SELF Key to perform self test of hydraulic circuit manually. The self test can
TEST be executed when treatment mode is HD and the mode is Preparation
or Stop. (6 : Balance test can be carried out in Gas Purge, Dialyzer
warming, Connect, Disconnect and Operation (Start) mode. )
Touch this key, then Ten key appears. Input the desired item by its
number (shown as below) and press the set button. The selected test
starts.
0 All test
1 Piping leak test
2:UF test
SUF leak test
4:Heater cut relay test
5:SV leak detector test
6 balance test
While UFC self test is being executed, the key lights up in green ( ON ).
TREAT
DATA
TIMER
no
/4\caution Per orm continuous battery charging (quick charging)
> unless the battery power is depleted after the battery is replaced
or is used by the blood pump.
[Continuously charging (quick charging) the battery when it is
fully charged causes depletion of battery power, which will cause
decreased battery operating time for the blood pump, or a
shortened battery lifespan.]
It will take approximately 48 hours to completely recharge the
battery when quick charge is off.
.....
I— ... ---
HO I-UF
DIS- DRAIN
■CON,
Key to drain the leftover of liquid inside the dialysate filter ( CF ) when it is exchanged.
During the draining, the key lights up in orange.
(1) Description of the operation screen.
Press the |CF replace] key on the Technicians switch screen (Hydro circuit -related ) to
open the CF replacement screen.
CF REPLACEMENT
RUN T I
RUNNING OF
TIME CONFIRM WATER
RESET REMOVAL
(h) ——J s—. I .... ........
Message
OF Replacement i s p o s s i b l e on P r e s e t ,
Treatment p r e p a r a t i o n and P r e p a r a t i o n modes
Press [CF WATER REMOVAL]key t o c a r r y out CF
replacement.
• In case that bypass coupler is dislocated from the bypass connector holder:
Message _____________________________________
Preparation (Air removal)
Connect dialyzer coupling to bypass connector
and return it to the bypass connector holder.
Step 2- Let the inside of the dialyzer hose be negative pressure.
To prevent leakage at removing the dialyzer coupler, make the inside of the
dialyzer hose to negative pressure. In the Message window, a message is
displayed.
Message
Please wait
® Remove the bypass connector out of the bypass connector holder to release the
dialyzer coupler. When the bypass connector is removed, the Confirm key turns to
flashing in green and gray. After releasing the dialyzer coupler, press the Confirm
key.
• In case that the bypass connector is dislocated from the bypass connector
holder:
Message
Under CF is draining (sidel)
Connect dialyzer coupling to bypass connector
and return it to the bypass connector holder.
• In case that the bypass coupler is dislocated from the bypass connector
holder:
Message
Connect dialyzer coupling to bypass connector
and return it to the bypass connector holder.
(3) CF replacement
After having replaced the CF, press the Confirm key.
Step 7: All stop ( Confirm the CF replacement and its completion )
Here, the removing and mounting of the CF is enabled. Following the
message indication, exchange the CF. The Confirm key flashes in green and
gray. After CF replacement is over, press the Confirm key to complete the
operation. Message is shown in the Message window.
• In case that the bypass coupler is mounted in the bypass connector holder.
Message
Replace a old filter
1 h YES NO
By pressing YES, the running time is reset.
3) Operation to complete the CF replacement
After the CF is replaced, filling up the liquid and rinsing of the CF are necessary. The
guiding message about the operation after CF replacement is displayed.
Message
Perform [Treatment preparation] to use for
the treatment.
Not for the use of the treatment, perform the
Cleaning program.
In the Preparation mode, by conducting the CF leak test, filling up the liquid and
rinsing of the CF can be performed. In the air removal process of the CF leak test,
filling up of liquid and rinsing is conducted for 5 minutes.
LCD
YELU
.UE
V
Col 1 Col 2
Preparation
Operat Operat
Stop
(HD) (I-UF)
® ■AS
|| n aboI = ■■■■■■■■
®
© Telop
SH zone KEY|
®
•Function key
No. Name Description
1 Key to return to Color scheme at shipment
Col 1 Col 2 Col 1 : light color (colors used for models before software
V2.1E)
Col 2 : dark color
2 This key switches between pages with sample colors.
Nurse Nurse
back c o l o r letter color
JWENUJJAf J
Technician Technician
back c o l o r letter color
[Function]
6 Key to set the color of menu bar letters and background.
[item to be set]
Nurse background color: Selective background color on menu bar at Nurse
mode.
Nurse letter color: Letter color on menu bar at Nurse mode.
Technicians background color: Selective background color on menu bar at
Technicians mode.
Technicians letter color: Letter color on menu bar at Technicians mode. _______
No key Item to be set
Disinfection Preparation
Back c o l o r
Operat Operat
Stop
(HD) (I-UF)
[Function]
Key to set the screen background color.
[Item to be set]
• Rinse • Disinfection’- Screen background color of Rinse • Disinfection, Preset and
Drain modes.
• Treatment preparation: Screen background color of Treatment preparation,
Preparation, Gas purge and Dialyzer warming modes.
• Operation ( HD ): Screen background color of HD operation mode.
• Operation ( I-UF ): Screen background color of I’UF operation mode.
• Stop: Screen background color of Stop, Connect and Disconnect modes. ________
No Key Item to be set
Honi l a b e l Letter c o l o r
[Function]
Key to set letter color of monitor label.
8
[Item to be set]
DIALV TEMP
•Setting key
No Key Item to be set
[Function]
Key to set background color and data letter color of the monitors which can be
operated by touching the data displaying area among monitors. The monitor
looking solid is the Settable monitor.
[Items to be set]
OIALY TEMP
etter
Background Color ofJ-*’ 3 / Color of Settable Monitor
settable Monitor (Color of numeric part)
[Function]
Key to set the background color and letter color of Multi- panel.
START TIME
12:59'
Background color of Multi-panel
•Setting key
No Key Item to be set
[Function]
Thia key is located on the right side of the screen, and sets the background color of the
switch area that includes the START and STOP key.
[Item to be set. __________
Background color of the switch area
STOP I
Background color of the switch area]
[Function]
This key sets background color and data font color of the monitors on which the data
display area cannot be touched to open the settings screen. Monitors with a solid
background color do not provide access to settings.
[Item to be set] ___
VEN PRESS (mmHg)] n a +a VEN PRESS M:
| Monitor data
letter color
12 IV*—
[Monitor background on]
Monitor background
color
70
[Monitor background off]
[Fig.2-16 Monitor background and font color outside the multi panel]
Data display area inside the multi panel
NURSING
START TIME
Monitor data letter color
[Fig.2-17 Monitor background and font color inside the multi panel]
' ' Monitors that do not provide access to settings are displayed with a
Notes solid background color or are surrounded by fixed color inside the multi
' panel so that they remain easily visible when the color scheme for the
monitors are changed.
•Setting key
No Key Item to be set
User 1 User 2
r— ------- -
Telop
User 3 Back c o l o r
13 [Function]
This is a key to set the color of the Telop background and letters.
[Item to be set]
User 1 : Letter color for User 1 Telop
User 2 : Letter color for User 2 Telop
User 3 : Letter color for User 3 Telop
Back color : Telop background color
No Key Item to be set
KEY ] Letter c o l o r
14
[Function]
Key to set letter color of the switch key
[Item to be set]
Font color of the switch key
11ICAUTION!11
All customized switch layout is initialized.
See technical manual for detailed switch layouts by type.
B« FLW (riL/dId)
MODE PRESET
[Type 1]
Based on the default settings, with the BUZZER MUTE, START, and STOP keys placed at
the top of the screen during all steps. _______________________________
Treatment Start/Stop Disconnect Rinse/Disinfect
preparation Connect mode ion mode
mode mode
mode
No.l Buzzer mute Buzzer mute Buzzer mute Buzzer mute Buzzer mute
No.2 Start Start Start Start _____Start _____
No.3 Stop Stop Stop Stop Stop
No.4 --- Preparation ... ... ...
No.5 IP rapid feed ... ... Drain IP rapid feed
No.6 ... Disconnect Disconnect Disconnect Weekly
program
[Type 2]
The START and STOP keys are fixed, and minimum numbers of keys are shown.
Treatment
Start/Stop Disconnect Rinse/Disinfect
preparation Connect mode
mode mode ion mode
mode
No.l ... ... ... ... ...
No.2 Start Start Start Start Start
No.3 Stop Stop Stop Stop Stop
No.4 ... ... ... ... ...
No.5 IP rapid feed ... ... Drain ...
No.6 ... Disconnect Disconnect Disconnect Weekly
program
[Type 3]
The START and STOP keys are placed at the top of the screen for all steps.
STEP keys are provided to make transitions between steps without switching the screen.
Treatment
Start/Stop Disconnect Rinse/Disinfect
preparation Connect mode
mode mode ion mode
mode
No.l Start Start Start Start Start
No.2 Stop Stop Stop Stop Stop
No.3 ... Preparation ... Preparation Preparation
No.4 ... Connect Connect ... ...
No.5 IP rapid feed IP rapid feed ... Drain ...
No.6 ... Disconnect Disconnect Disconnect Weekly
program
DBB-27 conducts the UFC self test and conductivity self test under the Treatment
Preparation mode to confirm the working of hydraulic parts and the conductivity-related
matters.
Also, confirming the working of micro-computer and blood pump and others by
conducting micro-computer test and blood monitor self test.
If any failure is resulted from each self test, it will be informed with alarm according to
the test on which failure is found out. As for details, please refer to the ITnstruction
Manual Chapter 10 Alarm and Action to be taken] .
The UFC self test and the conductivity self test can also get started manually by the
UFC self test| key or the [Conductivity self test| key in the Technicians switch ( alarm •
test ) screen.
1. Before the UFC self test
2. Under execution of the UFC self test ( first half part )
According to the start of the Treatment Preparation mode, the UFC self test ( first
half part ) gets automatically started. The contents of self test execution are displayed
as messages.
3. Under execution of the conductivity self test
When the UFC self test ( first half ) is over, the conductivity self test gets
automatically started. Under the conductivity self test in the Treatment Preparation
mode, SV close test is executed. The message of “Under execution of conductivity self
test” is displayed.
4. Under execution of the UFC self test ( last half )
When the conductivity self test is over, the UFC self test ( last half) gets automatically
started. The contents of self test execution are displayed as messages.
In the case that sodium infusion program is used, the sodium infusion
Notes
line needs to be filled up with sodium concentrate. After the mode
entered the Treatment Preparation, insert the sodium concentrate
suction nozzle ( purple ) to the tank and connect the concentrate
connector ( purple ) to the concentrate suction nozzle ( purple ).
CPU test
RAM test
ROM test
3 P RAM test
Data base test
2. When the blood monitor self test is over, message displays “blood monitor self test
passed”.
While the blood monitor self test is being carried, the blood pump
rotates reversely by approximately 1/8 revolution. This is not abnormal,
just for confirming its working.
Notes If coupling test fails in the Rinse mode, Rinse operation can be stopped
to avoid the Equid leakage which may occur while the coupling is
connected again to the bypass connector.
(Relevant setting: [Hydrocircuit activation at coupkng leak test failure]
in Hospital2 [43 Coupling leak test], initial value is [0=to continue the
Rinse operation])
• If [l=to stop the Rinse operation temporarily] is set, the Rinse
operation is stopped for 5 minutes after the couphng test failure
occurs, then resumed. However, if alarm state is released with the
Alarm reset key or others, Rinse operation can be resumed at any
time. When Rinse operation is stopped by above setting, message
displays “Rinse operation is stopped temporarily”.
• If [2=to stop the Rinse operation] is set, the Rinse operation is stopped
from the time when couphng test failure occurs to the time when the
alarm state is released. When Rinse operation is stopped by above
setting, message displays “Rinse operation is stopped.
In the case that the couphng leak test is executed at the Rinse mode of
the cleaning program, the time of cleaning program is not counted
while Rinse operation is stopped on account of the test execution and
test failure on the couphng leak test.
By turning the blood flow rate setter to the right , blood flow rate increases.
By turning the blood flow rate setter to the left , blood flow rate decreases.
NtXICISO
decrease increase
No Item Description
By touching |Return| key, ten-key
1 |Return|
Ti 7® screen closes.
Set ext
Name of| Displays name of item to change
2 ( Not displayed when detail screen
NEW set item| opens at the side of ten-key )
mmmH:: 4
5
olcl|
MAX
MIN
Displays the set value ( value before
changed )
$
By touching |CLR| key, the input value
aaacr s
6
7
CLR
and
is erased. ( For example, when input
error is made, touch [cLR| key )
aution] When input a new value, do not fail to confirm that values in the |new|
field and the [old] field are the same.
1. Touch the upper end of the bar graph and move the finger up and down to reach the
desired data.
2. The upper end of the bar graph moves according to the finger position.
3. When detail value is desired, input by ten-key.
0.0 0
| MONITOR I SETTING
Alarm p o i n t ~ ....................................................
2. Alarm p o i n t
2. 3 D i a l y s a t e pressure alarm y (
2, 4 TMP alarm
2. 5 Pressure l o s s alarm
2. 6 D i a l y s e r i n l e t b l o o d pressure alarm
8. 1 C o n d u c t i v i t y p r o f i l e alarm
8. 2 Na c o n d u c t i v i t y alarm
Alarm do i nt
2. 1 Venous pressure alarm __________________________,
HD • 1-UF Connect SN
Disconnect
F i x e d upper 400 mmHg 533 mmHg ■ 533 mmHg
[ Fig. 4-6 Select alarm point items screen and setting screen ]
flQSF* Notes There are Adjusting data as those other than Setting data. Adjusting
data are those adjusted proper to each DBB w27 and not changed except
at check and maintenance.
2. To change and confirm Hospital 1 data and Hospital 2 data, password input is
necessary. On treatment data come in two kinds. One is require password input and
the other is not.
On treatment Patient
data Data
Hospital 1
Data ( 16 persons )
Alarm
Alarm Alarm point
point data data etc.
point data
etc.
Hopital 2
Data
Notes Setting
does not.
data comes in 2 types. One requires password and the other
• ***’• Setting data within this frame have 2 kinds. One requires
: • • • • • ’ password to input and the other does not.
4. Patient data can be identical with Hospital 1 data. (Refer to ® on Fig. 4-7
Classification of the setting data)
When patient name is changed on the Patient setting screen, following screen will
appear;
Do you copy [HOSPITAL 1 ] data t o t h e data base
of t h i s p a t i e n t ?
| YES | | NO |
IlhSr* Notes
This functon is convenient in setting new patient data by clearing
previous patient data when patients move. Data for Hospital 1 should
be set to values appropriate for each facility.
Remarks
Notes
The table of setting items is structured basically as above arrangement,
however, some setting items may be arranged in different way as they
can not be structured as above. For this case, follow the display of the
table.
1. Dialysis condition
2. Alarm point
2.1. Venous pressure alarm
Patient connect • SN Set range
H D • I-UF
Blood collection
Fixed upper 300 400 400 -200-500 (mmHg) *1
Auto-set maximum 300 -200-500 (mmHg) *2
Auto-set width upper 50 0-400 (mmHg)
Auto-set width lower -30 -400-0 (mmHg)
Auto-set minimum 10 -200-500 (mmHg) *2
Fixed lower -50 -200 -50 -200-500 (mmHg) *1
*1 : Upper limit>Lower limit ( Upper limit can not be set lower than Lower Emit. Lower Emit can
not be set higher than Upper Emit, either.) Set Emit can not be set beyond the value of
Hospital 2 F9.1 Fixed alarm limit value J • F9.2 Fixed alarm Emit value ( Single Needle )j
*2 : Upper Emit>Lower Emit ( Upper Emit can not be set lower than Lower limit. Lower Emit can
not be set higher than Upper Emit, either.)
3. UF profile
Final
Step No. 1 2 3 4 5 6 7 8 9 10 position
Power
Initial Operation HD HD HD HD HD HD HD HD HD HD OFF
10
value Index 200 150 100 50 0 0 50 100 150 200
1 Operation
Set
value Index
Initial Operation HD HD HD HD HD HD HD HD HD HD
10 OFF
value Index 150 150 150 150 150 50 50 50 50 50
2 • Operation
Set
value Index
Initial Operation HD HD HD HD HD HD HD HD HD HD
10 OFF
value Index 50 50 50 50 50 150 150 150 150 150
3 Operation
Set
value Index
Initial Operation HD HD HD HD HD HD HD HD HD HD
10 OFF
value Index 200 200 160 160 120 120 80 80 40 40
4 Operation
Set
value Index
Initial Operation HD HD HD HD HD HD HD HD HD HD
value Index 200 200 160 160 120 120 80
10 OFF
80 40 40
5 Operation
Set
value Index
Initial Operation HD HD HD HD HD HD HD HD HD HD
value Index 40 40 80 80 120 120 160 160 200 200
10 OFF
6 Set Operation
value Index
Initial Operation HD HD HD HD HD HD HD HD HD HD
value Index 150 150 150 150 150 150 150 150 50
10 OFF
50
7 Set Operation
value Index
Initial Operation HD HD HD HD HD HD HD HD HD HD
value Index 150 150 150 150 150 150 150 150 50 50
10 OFF
8 Set Operation
value Index
Operation HD/ECUM OFF/O
Set range 1-10
Index 0-200 N*1
*1 : Among pattern 1~9, only pattern 1 can set the power ON.
_9 ______________________________________________
UF index number ( start ) 100 0-200
UF index number ( end ) 100 0-200
UF program course 1.0 0.0-2.0
Power OFF OFF/ON *1
*1 : Among pattern 1—9, only pattern 1 can set the power ON. ________
4. Conductivity profile
Switching
Power
position
Final
time
Step No. 1 2 3 4 5 6 7 8 9 10
Dialysate 14.0 14.0 14.0 14.0 14.0 13.5 13.5 13.5 13.5 13.5
Initial 30 OFF
°
value B liquid 2.50 2.50 2.50 2.50 2.50 2.50 2.50 2.50 2.50 2.50
1
Set Dialysate
value B liquid
Initial Dialysate 13.5 13.5 13.5 13.5 13.5 14.0 14.0 14.0 14.0 14.0
10 30 OFF
value B liquid 2.50 2.50 2.50 2.50 2.50 3.00 3.00 3.00 3.00 3.00
2
Set Dialysate
value B liquid
Initial Dialysate 14.0 14.0 13.5 13.5 13.5 13.5 13.5 13.5 14.0 14.0
10 30 OFF
value B liquid 3.00 3.00 3.00 3.00 3.00 2.50 2.50 2.50 2.50 2.50
3
Set Dialysate
value B liquid
Initial Dialysate 13.5 13.5 13.5 14.0 14.0 14.0 14.0 13.5 13.5 13.5
10 30 OFF
value B liquid 3.00 3.00 2.50 2.50 2.50 3.00 3.00 2.50 2.50 2.50
4
Set Dialysate
value B liquid
Initial Dialysate 13.5 13.5 14.0 14.0 13.5 13.5 14.0 14.0 13.5 13.5
10 30 OFF
value B liquid 3.00 2.50 2.50 3.00 3.00 3.00 3.00 2.50 2.50 3.00
5
Set Dialysate
value B liquid
Initial Dialysate 13.5 13.5 14.0 14.0 14.0 14.0 14.0 14.0 13.5 13.5
10 30 OFF
value B liquid 3.00 3.00 3.00 2.50 2.50 2.50 2.50 3.00 3.00 3.00
6
Set Dialysate
value B liquid
Initial Dialysate 14.0 14.0 13.5 13.5 13.5 13.5 13.5 13.5 14.0 14.0
10 30 OFF
value B liquid 2.50 2.50 3.00 3.00 3.00 3.00 3.00 3.00 3.00 3.00
7
Set Dialysate
value B liquid
Dialysate 12.5 15.5 (mS/cm) 1~
OFF
Set range 1~ 99 /ON
B liquid 1.50-7.00 (mS/cm) 10 (min *1
) *2
*1 : Among pattern 1~9, only pattern 1 can set the power ON.
*2. The switching time can only be set if [Select corres act of cond pro and UF pro] is selected to [No link operation] in
Hospital 1 [1.3 Dialysis condition (Dialysis related)].
Switching
position
Power
time
Final
Step No. 1 2 3 4 5 6 7 8 9 10
Initial Dialysate 13.5 14.0 13.5 14.0 13.5 14.0 13.5 14.0 13.5 14.0
10 30 OFF
value B liquid 3.00 3.00 3.00 3.00 3.00 3.00 3.00 3.00 2.50 2.50
8
Set Dialysate
value B liquid
Dialysate 12.5-15.5 (mS/cm) 1~
1- OFF
99 /ON
Set range 10 (min
B liquid 1.50-7.00 (mS/cm) *1
) *2
*1 : Among pattern 1~9, only pattern 1 can set the power ON.
*2- The switching time can only be set if [Select corres act of cond pro and UF pro] is selected to [No link operation] in.
Hospital 1 [1.3 Dialysis condition (Dialysis related)].
9
Start 13.5 12.5-15.5 (mS/cm)
Dialysate
End 15.0 12.5-15.5 (mS/cm)
conductivity
Course 2.0 0.0-2.0
4.2. Na profile
Na infusion
Step No. 1 2 3 4 5 6 7 8 9 10 switching Power
time
9 ___________________________________________
Start 0 0~50 (mEq/L) *1
Na Conduc 0~50 (mEq/L) *1
End 0
tivity
Course 1.0 0.0-2.0
Power OFF OFF/ON *2
*1 : Set range differs depend, upon setting of upper limit of Na infusion conductivity operating
range.
*2 : Among pattern 1 9, only pattern 1 can set the power ON.
5. Operation range
Blood flow operation range maximum 300 40~600 (mL/min)
Infusion pump operation range 10.0 0.0~10.0 (mL/h)
maximum _______________
UF rate operation range maximum 2.00 0.00-4.00 (L/h)
Dialysate temp, operation range 40.0 33.0-40.0 (°C) *1
maximum
Dialysate temp, operation range 33.0 33.0-40.0 (°C) *1
minimum
Na infusion operation range maximum 50 0-50 (mEq/L)
*1 : Upper limit>Lower limit ( Upper limit can not be set lower than Lower limit. Lower limit can
not be set higher than Upper limit, either.) __________ _ ___________________ __________
6. Blood pressure
Notes When maximum blood pressure lower limit alarm point is set to 0, it is
not monitored and the combined operation described in Hospital 1 [6.5
blood pressure (combined operation with maximum blood pressure
alarm)] is not allowed.
7. Alarm monitoring
9. Blood volume
9. 1. Blood volume 1
dBV drop alarm point 1 -20.0 -100-0 (%) *1
dBV drop alarm point 2 -40.0 -100-0 (%) *2
dBV changed rate alarm point -10.0 -50.0-0.0 (%/min) *3
:
*1 Alarm point 1 alarm point 2. Not monitored for 0 (%).
*2: Alarm point 1 alarm point 2. Not monitored for -100 (%).
*3: Not monitored for 0 (%/minute).
9. 2. Blood volume 2
Sei. of blood volume 1 Selection *1
*1 : 0=Not use l=Use
9. 4. Blood volume 4
dBV minimum UF rate 0.00 0.00~2.00(L/h)
dBV minimum UF prolonged time 5 0~30(min)
Pre weight 0.00 0.00— 300.00(kg) ______________
1. Blood pump
2. Priming setting
3. Dialyzer warming
Dialyzer warming time 2 l~60 (min)
Dialyzer warming after gas purge 0 Selection *1
Dialyzer warming stop time 10 l~60 (min)
Rinsing of dialyzer (Rinse extraction) 0.00 0.00~8.00(L)
Rinsing of dialyzer (Rinse rate) 1.20 0.00~4.00(L/h)
*1 : 0=Notused l=Used
4. Buzzer
5. LCD
6. Customize
VEN PRESS Cmmiic>; OIALY PRESS tone) UF VOLUME (0 up GOAL (D OF RATE (L/h
150
SET.
I CONDUCT
100 CONT
50
MEAS
11371 (mS/ciri)
AUTO MEAS
30 M INT OIALY FLOW
120 100 240
8. FLW (mLMn>
OPERAT
CALC
MODE
(HD) 120
Menu
Monitor A
Indicator
Switch
Monitor B
Multi-panel
j®* Notes When fCKBar graph screen J is selected from f Select basic screen during
treatment J of Hospital 2 F5.6 LCD ( screen )j , indicator is not displayed
even if “display indicator by customizing” is selected.
Others
Select display of [Dialyser warming] KEY 1 (Display)
Select display of [Priming assit] KEY 1 (Display) Selection*!
Select display of [Auto, priming] KEY 1 (Display)
*1 : 0=Not displayed l=Displayed
Selection *1
Selection*3
Selection*2
Selection*!
13
8 1 54
No.3 Total
UF volume Display Stop
conductivity
0 99 16
7 No Dialysate
No.4 Not
UF goal
display display flow rate
0 99
50 — No
No.5 Not —
UF rate display
display
— 57
No.6 — — — Dis-
—
connect
*1 : Refer to chapter 4 Setting items 4-34 TMonitors to be display able by customizing J
*2 : 0=Not display l=Display
*3 : Refer to chapter 4 Setting items 4-34 Switches to be displayable by customizing J
only)
Selection *2
Selection *1
13
8 1 54
NO.3 Total
UF volume Display conductivity
Stop
0 16
7 99
NO.4 Not Dialysate
UF goal No display
display flow rate
0
50 — 99
NO.5 Not
UF rate No display
display
— 57
NO.6 — — —
Disconnect
*1 : Refer to Chapter 4 Setting items 4-34 Monitors to be displayable by customizingj
*2 : 0=Not display l=Display
*3 : Refer to Chapter 4 Setting items 4'34 1Switches to be displayable by customizingj
50 — 56
NO.5 Not
UF rate Drain
display
—
— — —— — 57
NO.6 —
Disconnect
*1 : Refer to Chapter 4 Setting items 4:- 34 ("Monitors to be disp[ ayah e by customizingj
*2 : 0=Not display l=Display
*3 : Refer to Chapter 4 Setting items 4:”34 rSwitches to be disp] [ayah Le by customizingj
13
8 1 99
NO.3 Total No display
UF volume Display conductivity
0 16
7 20
NO.4 Not Dialysate
UF goal IP quick sw
display flow rate
0
50 — 99
NO.5 Not
UF rate No display
display
— — — 61 weekly
NO.6
timer
*1 : Refer to Chapter 4 Setting items 4-34 CM[onitors to be displayable by customizingj
*2 : 0=Not display l=Display
*3 : Refer to Chapter 4 Setting items 4-34 FS' witches to be displayable by customizingj
25.61 I 0| ALARM
RESET
START TIME ELAPSED TIME REMAIN TIME COMPLETE
TIME
PRESET
1 2 3 4
14 15 16
The 13
Dialysate
1st Total Dialysate
Bicarbonate temperatur
line conductivity flow rate
conductivity e
29 21
The Treatment 20 Treatment
35 Treatment
2nd 1
Start time r1 elapsed r* complete r
line remain time
o time .s .s time .2
•8 ■8 6 ■8
The rS Js
8 O 7 (D UF rate <D 99 <D
3rd CO CO CO LO
UF volume UF goal (monitor No display
line
only)
The
33 99 99 99
4th
IP total No display No display No display
line
*1 : Refer to Chapter 4 Setting items 4-34 TMonitors to be displayable by customizing J
1 2 3 4
The 9 99
11 99
1st Blood No display No display
Blood total
line pump
23 44
4 24
The Dialysate Na
TMP Water
2nd pressure conductivity
(monitor supply (monitor
line (monitor pressure
only) 4
only)
only) 5 r r 4
* d
30 d
Selection
d 3
1 o
Dialyzer 45
I
Dialyzer Dialysate
The Venous inlet blood CD pressure
CD
(D <72 co inlet blood
3rd pressure tn blood
loss and venous
line (monitor pressure mean
(monitor
only) (monitor pressure
only)
only)
31
The 32
Average 58 99
4th Volume
blood Delta BV No display
line stroke
pump flow
*1 : Refer to Chapter 4 Setting items 4-34 fMonitors to be displayable by customizingj
VEN PRESS (Mils) 01 ALY PRESS Mis) UF VOLUME (0 UF GOAL (D UF RATE (Uh)
92 -mi’ ;H0
IP START
NUR
START TIME ELAPSED TIME COMPLETE TIME REMAIN TIME Tzom CinO’
01 ALY TEMP
12:59 0:05 17:03 36J|
(<d)
IP TOTAL W n
| j CONDUCT
0.0 7
■ 13.7 |
&sM
I 01 ALY FLOW
CONDUCT SINGLE
UF
PROFILE PROFILE NEED | 500 |
, B. FLW (itl/min)
7201
[ Fig. 4-12 Nurse screen ]
1 2 3 4
21 20
35 29 Treatment Treatment
Selection *1
Selection *1
Selection *1
Selection *1
Selection *2
* *
PI 107
105 106 111
Key single
UF pro ’1 Na pro cond. pro
needle
o <D
co zn
*1 : Refer to Chapter 4 Setting items 4-34 ["Monitors to be displayable by customizing J
*2 : Refer to Chapter 4 Setting items 4-34 ["Switches to be displayable by customizing]
START
300-Vsib
STOP
f TEMP
MODE
OPERA!
CALC
VEN PRESS
START
PROGRESS
1?00
STOP
OF RATE
BLOOD FLOW
BLOOD TOTAL
IPRATL
IP TOTAL
VEN. PRESS 119
01 ALY FLOW
500 .conJ
I mode;
6.10. Screensaver
Press a tag to move to the setting screen directly.
Screens
UP
VOLUME
<SETTING> : O:D1B PRESS. , ,
Set monitor No. 1 to be displayed on
screensaver
Initial
<Setting
setting!
data>
7. IP rate setting
0.75 L/h
99! No display
YOIOUSPJXS.
92<*m
.
B.
||SET|
<4»* oimicp.
-79m, I W
* * “ “VJIHUAW
______i 8 0
[Fig4-15 Screensaver]
7. Weekly program
Day of the week Process Com pletion time
1 0 Selection *1 8:00 00:00-23:59 (h: min)
Sunday 2 0 Selection *1 14:00 00:00-23:59 (h: min)
3 0 Selection *1 20:00 00-00 23-59 (h- min)
1 0 Selection *1 8:00 00:00-23:59 (h: min)
Monday 2 0 Selection *1 14:00 00:00-23:59 (h: min)
3 0 Selection *1 20:00 00:00-23:59 (h: min)
1 0 Selection *1 8=00 00:00-23:59 (h: min)
Tuesday 2 0 Selection *1 14:00 00:00-23:59 (h: min)
3 0 Selection *1 20:00 00:00-23:59 (h: min)
1 0 Selection *1 8=00 00:00-23:59 (h: min)
Wednesday 2 0 Selection *1 14:00 00:00-23:59 (h: min)
3 0 Selection *1 20:00 00’00 23*59 (h- min)
1 0 Selection *1 8:00 00:00-23:59 (h: min)
Thursday 2 0 Selection *1 14:00 00:00-23:59 (h: min)
3 0 Selection *1 20:00 00-00 23-59 (h- min)
1 0 Selection *1 8:00 00:00-23:59 (h: min)
Friday 2 0 Selection *1 14:00 00:00-23:59 (h: min)
3 0 Selection *1 20:00 00:00-23:59 (h: min)
1 0 Selection *1 8=00 00=00-23:59 (h: min)
Saturday 2 0 Selection *1 14:00 00:00-23:59 (h: min)
3 0 Selection *1 20:00 00:00-23:59 (h: min)
*1 : 0= Not SET 1=CLEAN 1 2= CLEAN 2
3= CLEAN 3 4= CLEAN4 5= CLEAN 5
Hot rinse
Day of the week Select _____________Setting time
Sunday 0 Selection *1 0:00 00:00-23=59 (h: min)
Monday 0 Selection *1 0=00 00:00-23:59 (h: min)
Tuesday 0 Selection *1 0=00 00:00-23:59 (h: min)
Wednesday 0 i Selection *1 0:00 00’00 23’59 (hl min)
Thursday 0 Selection *1 (X00 00:00-23:59 (h: min)
Friday 0 Selection *1 0:00 00:00-23:59 (h: min)
Saturday 0 Selection *1 0-00 00-00 23-59 (h- min)
*1 : 0=None l=Perf.
8. Clock
Setting of clock ( year ) 00 00-99
Setting of clock 1.01 1.01-12.31
( month and the day )
Setting of clock 0 Selection *1
( the day of the week )
Setting of clock 0=00 0:00-23=59
( the hour & the minute )
*1 : 0=Sunday l=Monc Lay 2=Tuesday 3=Wednesday
4=Thursday 5=Frida ,y 6=Saturday
Settings for [8. Clock] cannot be changed during treatment (when one of
0 Notes the following conditions is fulfilled).
• Each mode of Connect, Treatment (start), Stop, Disconnect, or Drain
• After selecting the patient
• After confirming communication
Cleaning program 1
]L 2 3 4 5
Rinse Chemical Rinse — —
Operation
disinfection
Selection *1
30 | 40 40 J 0 0
Time
0-999 (min)
6 7 8 9 10
— — — — —
Operation
Selection *1
0 0 1 0 0 1 0
Time
0-999 (min)
Cleaning program 2
1 2 3_______ 4 5
Chemical
Operation
0-999 (min)
6 7 8 9 10
— — — — —
Operation
Selection *1
0 1 0 1 0 1 0 0
Time
0-999 (min)
*1 : Preset Dwell Rinse Hot rinse
Chemical disinfection Acid cleaning Heat citric acid
Note : Chemical disinfection, acid cleaning and heat citric acid are set in detail.
Set range of time is changeable depending upon former process.
( Details- Refer to the time to be possible to set for each process. )
Cleaning program 3
2 3 5
Rinse Hot rinse Rinse — —
Operation
Selection*!
9 25 10 0 0 1
1
Time
0-999 (min)
6 7 8 9 10
— — —- — —
Operation
Selection *1
0 1 0 1 0 1 0 1 0
Time
0-999 (min)
*1 : Preset Dwell Rinse Hot rinse
Chemical disinfection Acid cleaning Heat citric acid
Note : Chemical disinfection, acid cleaning and heat citric acid are set in detail.
Set range of time is changeable depending upon former process.
( Details- Refer to the time to be possible to set for each process. )
Cleaning program 4
2 3 4 5
Acid Chemical Rinse
Rinse
Operation
Rinse disinfection
cleaning
Selection *1
30 30 1 40 | 1 40 r
| 40 |
Time
0-999 (min)
6 7 ______ 8 9 10
Pre-set Rinse — — —
Operation
Selection *1
0 1 30 1 0 1 0 1 0
Time
0-999 (min)
*1 : Preset Dwell Rinse Hot rinse
Chemical disinfection Acid cleaning Heat citric acid
Note : Chemical disinfection, acid cleaning and heat citric acid are set in detail.
Set range of time is changeable depending upon former process. ( Details- Refer to the
time to be possible to set for each process. )
Cleaning program 5
1 2 3 5
Rinse — — — —
Operation
Selection *1
30 | 0 1 0 1 1 0 1 1 0 I
Time
0-999 (min)
6 7 8 9 10
— — — — —
Operation
Selection *1
0 1 0 1 o 0 0
Time
0-999 (min)
*1 : Preset Dwell Rinse Hot rinse
Chemical disinfection Acid cleaning Heat citric acid
Note : Chemical disinfection, acid cleaning and heat citric acid are set in detail.
Set range of time is changeable depending upon former process. ( Details: Refer to the
time to be possible to set for each process. )
Chemical Conductivity
Infusion Disinfetion Conductivity Min.time
disinfectant
volume temperature lower limit upper limit
port
OF option
Concentrate
nozzle Please refer to below table 1.
washing
1.0 6.0
Chemical 20 (min)
0.0-6.0 1.0-19.9
disinfection
(mS/cm) (mS/cm)
Acid cleaning 20 (min)
Acid
92.00 Equid 2.0
Heat citric side 35(min)
acid 80.00-92.00 0.0-6.0
Selection *1
(°C) (mS/cm)
*1 : Sodium hypochlorite side Acid liquid side (Although it is possibel to select between
either side, select the acid Equid side.)
NOTE- Selection is changeable by keeping touch on the key. The initial infusion value depends
upon the existence of Na infusion option.
Table 1
Infusion volume
Na infusion Not Not Not Not
Exist Exist Exist Exist
option exist exist exist exist
Not Not Not Not
CF option Exist Exist Exist Exist
exist exist exist exist
Concentrate
Not Not Not Not
nozzle Exist Exist Exist Exist
exist exist exist exist
washing
Chemical
27 22 25 20 29 23 26 21
disinfection
Acid cleaning 45 37 42 34 49 39 44 35
Heat citric 52 42
54 44 50 40 58 46
acid
Dwell 1 (min)
Hot rinse
Pre -rinse Hot rinse Rinse wait Post rinse
o 1 30 1 | io 1 0
Time
0-120 (min) 10-120 (min) 10-120 (min) 0-120 (min)
— 92.0 — —
Temperature
80.0~92.0°C
Note : When post-rinse is set to Omin, waiting is skippec >•
If the parts operating time exceeds the time set in [18. Replacement time of
Or Notes parts], you will be notified tha tthe replacement timing for the parts has
come in Preparation mode, after Treatment Preparation mode.
19. Ultrafiltration
19.2. Ultrafiltration 1
Selection of Treatment time count ( Stop ) 1 Selection*!
Sei. of treat, time count ( UF rate down ) 0 Selection *2
Sei. of UF pump after UF completion 0 Selection *3
UF rate ( UF rate down ) 0.00 0.00-2.00 (L/h)
UF rate limit to blood flow rate 30 0-100 (%)
Selection of UF rate auto-recalculation 0 Selection *4
Redemand permit of over UF rate range 10 1-100 (%)
*1 : 0=Count l=Not count
•When To (Count)J is selected, time is counted as treatment time even during FStop modeJ
and r Bypassing at Start mode J . «=> Treatment time is enlarged.
*2 : 0=Count at UF rate is lowering. l=Not count at UF rate is lowering.
• When 1 ( Not count ) is selected, time is not counted as treatment time even by operating the
UFP rate down key. Treatment time is not enlarged.
*3 : 0=UF pump rate lowering 1=UF pump running on 2=UF pump stops
*4 : 0=not recalculate l=Recalculate
• When r1( Recalculate )j is selected, UF rate is automatically recalculated to complete UF
in the treatment remaining time.
For example, When Fo ( Count ) J is selected from FTreatment time conditions, Count at
Stop modej , time is counted and passed as treatment time even in the state that UF pump
is not operated such as in Stop mode or in bypassing at Start mode.
Also, when Fo ( Count the time at UF rate lowering ) j is selected from FSelection of
treatment time count ( UF rate down ) j , time is counted and passed even at the state that
UF rate is being lowered by operating the UFP rate down key.
As above mentioned, when UFP is stopped or its rate is down by any reason, later it
becomes normalized and resumes operation, if F1 ( Recalculate ) J is selected from FSelect
auto-recalculation of UF rate J , UF rate is automatically recalculated to complete UF in the
treatment remaining time.
19.3. Ultrafiltration 2
UF pump cell abnormal alarm point 2.000 1.000-3.000 (V)
UF pump discharge down alarm 80 0-600
Select, of display/set. digit of UF goal 0 Selection* 1
Reset of UF rate after UF complete 0 Selection*2
Reset of UF data in Drain mode 1 Selection*3
Select UF calculation screen 0 Selection*4
Select display of [UF vol. reset] KEY 1 Selection* 5
*1 : 0=Maximum 9.99L 1= Maximum 39.99L
*2 : 0=Without UF rate reset l=With UF rate reset
• When 1 ( With UF rate reset ) is selected, UF rate is set to zero at UF completion.
• Once UF rate is reset to zero, UF rate can not be input durng information of UF completion.
*3 : 0=Without reset l=With reset
• When 0 ( Without reset ) is set, UF rate, UF target set and UF total are not reset in the
treatment data reset of drain mode.
• When 1 ( With reset ) is set, UF rate and UF target set are reset to Set 1 value.
• This set is valid for treatment data reset in the Treatment Preparation mode according to
1Treatment data reset delay timej of Hospital 2 135 Others 1J .
*4 : 0=Up/Down key type l=Ten-key type
*5 : 0=No display l=Display
19.4. Ultrafiltration 3
Sei dia remain time extend func by UF calc _________ 0 Selection*!
*1 : 0=No extension for remaining time l=Extension for remaining time
25. TMP
TMP zero calibration starting time 20 0-30 (min)
TMP zero calibration time 190 120-600 (sec)
TMP alarm prolonged time 2 0~60 (sec)
Changing of blood flow rate under TMP zero 1 Selection *1
calibration
Blood flow changed rate under TMP zero 70 50-100 (%)
calibration
UFC monitoring mode at HD —» I-UF 0 Selection*2
*1 : 0=No change l=Change
*2 : 0=Continue TMP monitoing (TMP auto-forecast monitoring or TMP auto-set monitoring)
l=dialysate pres monitoring
27. Conductivity
27.1. Conductivity 1
Upper limit _________ Lower limit
B conduc. auto-set alarm 5.0 0.0-9.9 (°/o) -5.0 -9.9-0.0 (%)
width
Total conductivity auto-set -5.0 -9.9-0.0 (%)
5.0 0.0-9.9 (%)
alarm width
27.2. Conductivity 2
Selection of external intensive piping 0 Selection* 1
B concentrate pump compensation 1 Selection *2
A concentrate pump compensation 1 Selection *2
Flushing daily, vol. after cond. alarm is
500 500-5000 (mL)
returned
*1 : 0=Tank l=Intensive piping
*2 : 0=No efficiency correction 1= Efficiency correction
27.3. Na infusion
Na priming function Use selection 0 Selection *1
Na infuse cond check func Use sei 0 Selection *1
Na infuse cond check func Infuse time 3 1 5 (min)
Op sei after Na infuse cond check 0 Selection *2
*1 : 0= Not use 1= Use
*2 : 0=Na pump stop l=Na pump operate (Step 1, or start infusion concentration)
30. Override
Setting of override time 120 1-120 (sec)
Limited override revolution value 7 1-7 (revolution)
32. OK Monitor
Selection *2
(sec)
Selection *1
BP rate 0 1 2 0 0
0—600
No blood circuit 1 2 0 0 (sec) *4
1 0-600 0
Detection of saline 2 30
(sec) *3
Clip-type air detector 1 0-600 0
2 0
power (sec)
* 1 : 0 = OK monitor buzzer OFF 1= OK monitor buzzer ON
*2 : O=Light off l=Light on 2= Flickeringl (1Hz) 3= Flickering2 (5Hz)
*3 : OK monitor does not report when set to 0
*4 : No continuous buzzer when set to 0
1 0 0-600
UF rate down 2 0 (min)
(sec) *4
UF rate too much 1 2 0 0
UF rate exceeds 1 0-600 0
2 0
operation range (sec)
UF rate exceeds 1
2 0
limited range
* 1 : 0 = OK monitor buzzer OFF 1= OK monitor buzzer ON
*2 : O=Light off l=Light on 2= Flickeringl (1Hz) 3= Flickering2 (5Hz)
*3 : OK monitor does not report when set to 0
*4 : No continuous buzzer when set to 0
Selection *2
Selection *1
IP rate 0 1 2 0 0-600 0 0-600
No syringe 1 2 0 (sec) 0 (sec) *3
IP bolus volume 0 1 2 0 0
* 1 : 0 = OK monitor buzzer 0. FF 1= OK monitor buzzer ON
*2 : O=Light off l=Light on 2= Flickeringl (1Hz) 3= Flickering2 (5Hz)
*3 : OK monitor does not repiort when set to 0
Treatment completion 1 2 15
1 0-600 0 0-600
Stop 2 0 (sec) *4
(sec)
Forgot to turn on 1 0—60 0
2 5
rinse/ disinfection (min)
0-600
Disconnect 1
(sec) *3
* 1 : 0 = OK monitor buzzer OFF 1= OK monitor buzzer ON
*2 : O=Light off l=Light on 2= Flickeringl (1Hz) 3= Flickering2 (5Hz)
*3 : OK monitor does not report when set to 0
*4 : No continuous buzzer when set to 0
0-600 0 0-600
pressure alarm 1 2 0
activation (sec) (sec) *3
Selection *1
Selection *2
(BV meter) (sec) (sec) *3
UF rate 1 2
(dBV UF rate down)
dBV reference area 0-480 0-600
0 1 60 0
deviance (sec) (sec) *3
* 1 : 0 = OK monitor without buzzer 1= OK monitor with buzzer
*2 : 0= Light off 1= Light on 2= Flickeringl (1Hz) 3= Flickering2 (5Hz)
*3 : No continuous buzzer when set to 0
1 0-600 0 0-600
2 0
plate dialyzer (sec) (sec) *3
Port cleaning
connection 1 2 0 0
Forgot return Na cond 0-600
conn (Drain mode) 1 3 1
(sec) *4
* 1 : 0 = OK monitor buzzer OFF 1= OK monitor buzzer ON
*2 : O=Light off l=Light on 2= Flickeringl (1Hz) 3= Flickering2 (5Hz)
*3 : No continuous buzzer when set to 0
*4 : OK monitor does not report when set to 0
Selection *2
Selection *1
No blood line at 0—600 0-600
1 2 10 0
clip-type air detector (sec) *3 (sec) *4
35. Others
35.1. Others 1
Selection of Monitor output 1 action 0 Selection *1
Reset Monitor output 1 1 Selection *2
Select action of external input signal 1 Selection *3
Treatment data reset prolonged time 300 0~600 (sec)
Reset time of treatment data at Treatment—>
1 Selection *4
Rinse / Disin
Select BM self TEST auto-execution 1 Selection *5
* 1 : 0= Output only at bypass disconnected l=Output always 2=Output only when
bypass is disconnected UF output 3=Always output, UF output
*2 : 0=No reset l=reset at buzzer stop
*3 : 0=No selection of action l=Stop of hydraulin unit 2=Stop of blood pump
3=Stop of hydraulic unit plus Blood Pump
*4 : 0=No reset l=With reset
*5 : 0=No execution l=Execution
35.2. Others 2
Alarm point of back pressure set. fail.
10 0-300
( H l : supply)
Alarm point of back pressure set. fail.
10 0-300
( H2: Drain )
Select the operation corresponding RO 0 Selection *1
RO stand-by time to correspond 10 0~30 (sec)
*1 : 0=Notlink 1=1fink
35.3. Others 3
Battery charge time 5 2-10 (sec)
Select air detector power ( Disconnect ) 0 Selection *1
Prep mode Flu spill prev func use sei 0 Selection *2
Prep mode Flu spill prev func op time 10 l~600(sec)
Concentrate nozzle liquid purge function 1 Selection *2
*1 : O=impossible l=possible to operate
*2 : 0=Not use l=Use
37. BPM
41. CF replacement
Preparatory running time at CF replacement 10 5-120 (sec)
Primary water drain time at CF replacement 10 5-120 (sec)
Secodary water drain time at CF replacement 30 5-120 (sec)
CP2 ( Booster Pump ) rate at CF replacement 1300 1000-2600 (rpm)
Fluid, filling time after CF replacement ______ 5 5~30 (min)
External lamp action 2 Selection*!
( During CF replacement ) ________
*1 : 0=Liglitoff l=Lighton 2=Flickering 1 (1Hz) 3=Flickering 2 (5Hz)
Selection *4
No. 2 Timer 2 0:00 0
Timer3 :
No. 3 *1 0 00 *2 *3 0
No. 4 Timer 4 0=00 0
No. 5 Timer 5 0:00 0
*1 : Input of timer names by key board ( 15 letters in ful . size )
*2 : 0 : 00 — 23: 59 (h ; min)
*3 : 0:00~23:59 (h: min)
As for Remaining time and Completion time, items s et later are effective.
Ineffective items are displayed as F ----- ; -----j ,
Remaining time is to be effective at initial.
*4 : 0=Buzzer 7=Melody 7 14=Melody 14
l=Melody 1 8=Melody 8 15=Melody 15
2=Melody 2 9=Melody 9 16=Melody 16
3=Melody 3 10=Melody 10 17=Melody 17
4=Melody 4 ll=Melody 11 18=Melody 18
5=Melody 5 12=Melody 12 19=Melody 19
6=Melody 6 13=Melody 13 20=Melody 20
Selection *4
No.12 Timer 12 (KOO 0
No. 13 Timer 13 *1 OW *2 *3 0
No. 14 Timer 14 0:00 0
No.15 Timer 15 0:00 0
*1~*4 : Please refer to 42.1 Timet set 1.
Selection *4
No.17 Timer 17 0:00 - - 0
No. 18 Timer 18 *1 0-00 *2 - - *3 0
No.19 Timer 19 0:00 — 0
No.20 Timer 20 0:00 - - 0
*1 *4 : Please refer to 42.1 Timer set 1.
42.5. Ti
Select number of timer to use 1 Selection *1
*1 : 0=Use one ____________1= Multiple use _____________________________________
1. Touch the [Standard Setting) displayed on Technicians Menu bar in the screen.
Below the [Standard Setting), a pull-down menu will appear.
echnicians
\ / Nurses
LU
VEN PRESS
| MODE II MONITOR
(Mils) DI ALY PRESS (-Us)
II
UF VOLUME
SETTING
(0
JI GRAPHIC 1 STANDARD SETTING
UF GOAL co
SELECTION
■e
3 O.oo0.00 REGISTRATION
SCHEDULE
4. Touch the desired blank name column. The Patient name setting key board will open.
[ Back |
aaa
aa a a a aaa a|T Fl I Clear
aaa a a a a a a a aa [ Space
aaFaamaa a F a aaa
aa a aa aa a a aa a
aaaaaa a aaa a a | SET |
[Fig. 4-17 Patient name setting keyboard screen (Numbers & Alphabet)]
Clear
Back
aaaaaaaaaaaaa I Space
aaaaaaaaaaaaa I I
aaaaaaaaaaaaa
aaaaaaaaaaaaa SET
aaaaaaaaaaaaa I I
By touching the applicable key on the patient name setting keyboard, the following
functions are performed.
Back
acoaaacotosQHacoizj
HHEOHQHQEZIQHQQCl
□□(IEQ0Q3I1EQQQ ( SET
No Name Description
A new The input name (character string) is displayed.
B old The entered name (character string) is displayed.
When the |Back| key is touched, one entered character is deleted.
1 Back (For example, if you make a mistake entering one character, touch the
|Back| key.)
When the |Clear| key is touched, all of the entered characters are deleted.
2 |Clear| (For example, if you make a mistake entering many characters, touch the
Clear key.)
You can enter spaces by touching the |Space| key.
3 |f3pac< (For example, you can insert a space between [First name] and
[Surname].)
Touch the [Set key to enter the inputted patient name.
4 Set When the [Set key is touched, the [Patient data] can be made the same as
[Hospital 1 data]. For details, refer to [Technical Manual Chapter 4.3
Setting data classification].
[Fig. 4'18 Keyboard screen for setting patient names (numbers and alphabet)]
MOOEL OOB-21
RINSE
Rins DISINFECTION IP
DJUSTMENT
__E8ESEL_ l E20iriLl (itiL/h)
2. Amplifier circuit
Adjusting items Initial value Adjust range Unit
2.1 Amplitude circuit ZERO (L) ........ ........ °C
2.2 Amplitude circuit SPAN (L) ........ ........ °C
2.3 Amplitude circuit ZERO (H) ........ ........ °C
2.4 Amplitude circuit SPAN (H) ........ ........ °C
3. Hydrocircuit sensor
Adjusting items Initial value Adjust range Unit
3.3 Dialysate pressure sensor ZERO -500 -600 —400 mmHg
3.4 Dialysate pressure sensor SPAN 500 400-600 mmHg
3.5 Water supply pressure sensor ZERO 0 ■100 -100 mmHg
3.6 Water supply pressure sensor SPAN 400 300-500 mmHg
4. Conductivity cell
Adjusting items Initial value Adjust range Unit
4.4 Cond. cell 2 constant 7.000 5.000—9.999 ........
6. Pump ( UFP / DP )
Adjusting items Initial value Adjust range Unit
6.1 UF pump discharge volume 0.776 0.740—0.803 mL/st
6.2 Dialysate flow rate (Measurement) 500 300-700 mL/min
6.3 Dialysate flow rate ( Theoretical ) ........ ........ ........
6.4 Pump cell voltage 0.00 0.00-4.00 L/h
8. Conductivity (Registration)
Adjusting items Initial value Adjust range Unit
8.1 Concentrate 1 ........ ........ ........
8.2 Concentrate 2 ........ ........ ........
8.3 Concentrate 3 ........ ........ ........
8.4 Concentrate 4 ........ ........ ........
8.5 Concentrate 5 ........ ........ ........
*1 : TUseJ • PNotuseJ