EVM500041 Rev M Manual, User, EVolution 3e
EVM500041 Rev M Manual, User, EVolution 3e
eVolution 3e
®
International Version
Preface
Revision M of this user manual: Part Number EVM500041 (Ventilator, User Manual, International Version)
November-2022.
Illustrations in this manual apply to the eVolution 3e Series Ventilator and installed software / firmware
versions are displayed upon power up on the initial screen.
Inspiration, eVolution and CliniNet are registered trademarks of eVent Medical. Smart Sigh, Smart
Nebulizer, Virtual Report and MiniWeb are trademarks of eVent Medical. Other acronyms and products
referenced in this manual may be trademarked by separate companies.
Upon request, eVent Medical will provide certain technical, operation, sales and marketing information that
will assist clinicians and trained service professionals.
Table of Contents
TABLE OF CONTENTS..................................................................................................................................................................... 3
1. INTRODUCTION TO THE EVOLUTION 3E VENTILATOR ...................................................................................................... 12
1.1 INTENDED USE ....................................................................................................................................................................................................................... 12
1.2 CONTRAINDICATIONS: ........................................................................................................................................................................................................... 12
1.3 FREQUENT USE FUNCTIONS:.................................................................................................................................................................................................. 12
1.4 SAFETY ................................................................................................................................................................................................................................... 12
1.5 WARNINGS, CAUTIONS AND BASIC SAFETY INFORMATION ................................................................................................................................................... 13
1.5.1 Copyright Notice .............................................................................................................................................................. 18
1.5.2 Trademark Notice ............................................................................................................................................................ 18
1.5.3 Notice of Conformity, Compliance and Approvals ........................................................................................................... 18
1.5.4 Electromagnetic Compatibility ........................................................................................................................................ 18
1.5.5 CE Notified Body .............................................................................................................................................................. 19
1.5.6 Classification ................................................................................................................................................................... 19
1.5.7 Regulatory Notice ............................................................................................................................................................ 19
1.5.8 Transport ......................................................................................................................................................................... 19
1.5.9 Storage ............................................................................................................................................................................ 19
1.6 DEVICE LABELS AND SYMBOLS ............................................................................................................................................................................................... 19
1.7 RESPONSIBILITIES AND GUARANTEES .................................................................................................................................................................................... 23
1.8 CLINICAL SPECIFICATIONS ...................................................................................................................................................................................................... 23
1.8.1 Essential Performance Requirements .............................................................................................................................. 23
1.8.2 Ventilation Mode Types and Mode Selections ................................................................................................................ 24
1.8.3 Breath Type by Control Variables .................................................................................................................................... 24
1.8.4 Other Ventilation Mode Options ..................................................................................................................................... 25
1.8.5 Patient Types by Ideal Body Weight ................................................................................................................................ 25
1.8.6 Patient Startup Settings................................................................................................................................................... 25
1.8.7 Standard Mode Settings .................................................................................................................................................. 25
1.8.8 SPAP Mode Settings ........................................................................................................................................................ 31
1.8.9 Tube Compensation (optional software) ......................................................................................................................... 32
1.8.10 NCPAP+ Mode Settings*.................................................................................................................................................. 34
1.8.11 Hi FlowO2 (Optional Software) ....................................................................................................................................... 34
1.8.12 Apnea Backup Settings .................................................................................................................................................... 35
1.8.13 Configuration Screen Setting Options (Config) ................................................................................................................ 35
1.8.14 Monitored Data / Displayed Patient Values .................................................................................................................... 36
1.8.15 Graphics: Real-Time Curves ............................................................................................................................................. 44
1.8.16 Alarm Limit Settings ........................................................................................................................................................ 45
1.8.17 Alarms by Priority ............................................................................................................................................................ 46
1.8.18 SBT Mode Settings........................................................................................................................................................... 48
1.9 TECHNICAL SPECIFICATIONS .................................................................................................................................................................................................. 48
1.9.1 Power and Gas Supply ..................................................................................................................................................... 48
1.9.2 Environmental Data ........................................................................................................................................................ 49
1.9.3 Physical Data ................................................................................................................................................................... 49
1.9.4 Technical Data ................................................................................................................................................................. 49
1.9.5 Technical Data: Capnography Sensors ............................................................................................................................ 50
1.9.6 Compliance and Approvals .............................................................................................................................................. 50
1.9.7 Electromagnetic Compatibility Declaration ..................................................................................................................... 50
2 PREPARING THE VENTILATOR FOR USE ............................................................................................................................ 55
2.1 VENTILATOR SETUP................................................................................................................................................................................................................ 55
2.2 POWER SUPPLY ...................................................................................................................................................................................................................... 56
2.3 AC MAINS SUPPLY .................................................................................................................................................................................................................. 56
2.4 DC (EXTERNAL BATTERY) SUPPLY ........................................................................................................................................................................................... 57
2.5 DC (INTERNAL BATTERY) POWER SUPPLY .............................................................................................................................................................................. 57
2.6 ON AND OFF POWER SWITCH ................................................................................................................................................................................................ 58
2.7 GAS SUPPLY ........................................................................................................................................................................................................................... 58
2.8 HIGH PRESSURE BASED SYSTEM: CONNECTING TO OXYGEN AND AIR SUPPLY ...................................................................................................................... 59
2.9 OPTIONAL, TURBINE BASED SYSTEM: CONNECTING TO OXYGEN SUPPLY ............................................................................................................................. 59
Index of Figures
FIGURE 1: INSTALLING O2 SENSOR ........................................................................................................................................................ 55
FIGURE 2: CALIBRATE OXYGEN SENSOR SCREEN ............................................................................................................................... 56
FIGURE 3: POWER SUPPLY SOCKET....................................................................................................................................................... 57
FIGURE 4: DC, EXTERNAL BATTER SOCKET .......................................................................................................................................... 57
FIGURE 5: ON/OFF SWITCH ...................................................................................................................................................................... 58
FIGURE 6: HIGH PRESSURE O2 AND AIR INLET ..................................................................................................................................... 59
FIGURE 7: HIGH PRESSURE O2 INLET FIGURE 8: LOW PRESSURE O2 INLET ................................................................................ 59
FIGURE 9: COMMUNICATION PORTS, ETHERNET AND NURSE CALL ................................................................................................. 60
FIGURE 10: NURSE CALL CONNECTOR PIN ASSIGNMENTS ................................................................................................................ 61
FIGURE 11: BREATHING CIRCUIT ............................................................................................................................................................ 61
FIGURE 12: NCPAP PRONGS .................................................................................................................................................................... 62
FIGURE 13: EVOLUTION 3E EXHALATION VALVE SYSTEM ................................................................................................................... 63
FIGURE 14: FLEX ARM ASSEMBLY ........................................................................................................................................................... 64
FIGURE 15: ATTACHING THE IRMA MAINSTREAM CO2 SENSOR TO THE BREATHING CIRCUIT ..................................................... 65
FIGURE 16: ZERO CO2 SENSOR CALIBRATION SELECTION ................................................................................................................ 66
FIGURE 17: PERFORMING A ZERO CO2 SENSOR CALIBRATION ......................................................................................................... 67
FIGURE 18: ATTACHING THE ISA SIDESTREAM CO2 SENSOR TO THE VENTILATOR ....................................................................... 68
FIGURE 19: ATTACHING THE SIDESTREAM CO2SENSOR AND SAMPLE LINE ................................................................................... 70
FIGURE 20: EVOLUTION 3E FRONT PANEL ............................................................................................................................................. 71
FIGURE 21: STANDBY KEY OPTIONS ....................................................................................................................................................... 72
FIGURE 22: EVOLUTION POWER UP & PATIENT SETUP SCREEN ....................................................................................................... 73
FIGURE 23: GRAPHIC USER INTERFACE ................................................................................................................................................ 75
FIGURE 24: GRAPHIC USER INTERFACE, MODE SELECTION .............................................................................................................. 75
FIGURE 25: STANDBY NOT VENTILATING (SAFETY MESSAGE) ........................................................................................................... 81
FIGURE 26: CONTROL SETTING LIMIT (SOFT BOUNDARY) FEATURE ................................................................................................. 82
FIGURE 27: POWER UP SCREEN ............................................................................................................................................................. 84
FIGURE 28: PATIENT SELECTION SCREEN – STARTUP SCREEN ........................................................................................................ 84
FIGURE 29: NEW PATIENT STARTUP SCREEN ....................................................................................................................................... 85
FIGURE 30: IBW PATIENT STARTUP SCREEN ........................................................................................................................................ 86
FIGURE 31: PREVIOUS PATIENT STARTUP SCREEN ............................................................................................................................. 86
FIGURE 32: IDEAL BODY WEIGHT SCREEN ............................................................................................................................................ 87
FIGURE 33: PATIENT SETUP PAGE .......................................................................................................................................................... 87
FIGURE 34: HUMIDIFICATION TYPE SELECTION OPTION ..................................................................................................................... 87
FIGURE 35: PRE-PATIENT SYSTEM TESTS ............................................................................................................................................. 88
FIGURE 36: CALIBRATION OPTIONS ........................................................................................................................................................ 89
FIGURE 37: CURRENT SETTINGS SCREEN............................................................................................................................................. 90
FIGURE 38: PROPOSED SETTINGS SCREEN .......................................................................................................................................... 90
FIGURE 39: MODE TYPE / MODE SELECTIONS ...................................................................................................................................... 91
FIGURE 40: NEBULIZER SCREEN ............................................................................................................................................................. 96
FIGURE 41: SMART SIGH SECTION .......................................................................................................................................................... 97
FIGURE 42: TREND DATA SETTINGS ....................................................................................................................................................... 98
FIGURE 43: TUBE COMPENSATION SETTINGS ...................................................................................................................................... 98
FIGURE 44: TUBE COMPENSATION WAVEFORM ................................................................................................................................... 99
FIGURE 45: SBT SETTINGS SCREEN ..................................................................................................................................................... 100
FIGURE 46: SBT AUTOMATIC EVENT MARKERS .................................................................................................................................. 103
FIGURE 47: HOME TAB ............................................................................................................................................................................ 103
FIGURE 48: DATA TAB ............................................................................................................................................................................. 105
FIGURE 49: TRENDING WAVEFORM AND PARAMETER SELECTION SCREENS ............................................................................... 108
FIGURE 50: ALARMS SCREEN ................................................................................................................................................................ 110
FIGURE 51: EVENT LOG .......................................................................................................................................................................... 114
FIGURE 52: REAL-TIME GRAPHICS (CAPNOGRAMS) ........................................................................................................................... 132
FIGURE 53: REAL-TIME GRAPHICS (SBCO2CURVES) .......................................................................................................................... 133
FIGURE 54: CONFIG SCREEN ................................................................................................................................................................. 133
FIGURE 55: MONITORS............................................................................................................................................................................ 136
FIGURE 56: COMPLIANCE COMPENSATION ......................................................................................................................................... 136
FIGURE 57: O2 SENSOR – HIGH PRESSURE SYSTEM ......................................................................................................................... 137
FIGURE 58: LOW FLOW O2 – OPTIONAL, BLOWER BASED SYSTEM ................................................................................................. 137
FIGURE 59: AUDIO / VISUAL OPTIONS ................................................................................................................................................... 138
FIGURE 60: CALIBRATIONS..................................................................................................................................................................... 139
FIGURE 61: CALIBRATE EXHALATION FLOW SENSOR ZERO OFFSET .............................................................................................. 140
FIGURE 62: PRE-PATIENT TESTS SCREEN ........................................................................................................................................... 141
FIGURE 63: SYSTEM TEST FAILURE MESSAGE ................................................................................................................................... 141
FIGURE 64: CALIBRATE OXYGEN SENSOR SCREEN ........................................................................................................................... 142
FIGURE 65: CALIBRATE EXHALATION FLOW SENSOR SCREEN ........................................................................................................ 143
FIGURE 66: EXHALATION FLOW SENSOR ............................................................................................................................................. 145
FIGURE 67: EXHALATION VALVE SYSTEM ............................................................................................................................................ 145
FIGURE 68: EXHALATION VALVE SYSTEM COMPONENTS ................................................................................................................. 147
FIGURE 69: EXHALATION SYSTEM ASSEMBLY 1 ................................................................................................................................. 147
FIGURE 70: EXHALATION SYSTEM ASSEMBLY 2 ................................................................................................................................. 148
FIGURE 71: EXHALATION SYSTEM ASSEMBLY 3 ................................................................................................................................. 148
FIGURE 72: FAN FILTER AND COVER .................................................................................................................................................... 151
FIGURE 73: AIR INLET HEPA FILTER ...................................................................................................................................................... 151
FIGURE 74: O2 SENSOR .......................................................................................................................................................................... 152
FIGURE 75: INTERNAL BATTERIES (TURBINE UNIT) ............................................................................................................................ 152
FIGURE 74: INTERNAL BATTERY (HIGH PRESSURE UNIT) ................................................................................................................. 153
FIGURE 77: FUSE .................................................................................................................................................................................... 153
FIGURE 78: SELECTING A MONITORING PARAMETER ........................................................................................................................ 157
FIGURE 79: BASIC MONITORED PARAMETERS.................................................................................................................................... 158
FIGURE 80: MECHANICS MONITORED PARAMETERS ......................................................................................................................... 158
FIGURE 81: WEANING MONITORED PARAMETERS ............................................................................................................................. 159
FIGURE 82: FLOW AND PRESSURE WAVEFORMS CSTAT, RINSP & REXP ....................................................................................... 161
FIGURE 83: HOME SCREEN GRAPHICS DISPLAY ................................................................................................................................ 163
FIGURE 84: LOOP WINDOWS (LEFT PANEL) ......................................................................................................................................... 164
FIGURE 83: WAVEFORM CONFIGURATION ........................................................................................................................................... 164
FIGURE 84: FREEZING GRAPHICS ......................................................................................................................................................... 166
FIGURE 85: TREND DISPLAY .................................................................................................................................................................. 167
FIGURE 86: TREND CONFIGURATION.................................................................................................................................................... 167
FIGURE 87: SELECTING MANEUVERS ................................................................................................................................................... 168
FIGURE 88: SUCTION SUPPORT MANEUVER SELECTION .................................................................................................................. 168
FIGURE 91: SUCTION SUPPORT PREPARATION PHASE ..................................................................................................................... 169
FIGURE 90: SUCTION SUPPORT PATIENT DISCONNECTED PHASE .................................................................................................. 169
FIGURE 91: SUCTION SUPPORT POST OXYGENATION PHASE ......................................................................................................... 170
FIGURE 92: P0.1 MANEUVER .................................................................................................................................................................. 171
FIGURE 95: P0.1 MANEUVER MEASUREMENT GRAPH ........................................................................................................................ 171
FIGURE 96: PIMAX MANEUVER .............................................................................................................................................................. 173
FIGURE 97: INSPIRATORY HOLD MANEUVER ...................................................................................................................................... 174
FIGURE 98: EXPIRATORY HOLD MANEUVER........................................................................................................................................ 174
FIGURE 99: LUNG MODEL LOCATION .................................................................................................................................................... 175
FIGURE 100: LUNG COMPLIANCE IMAGES AND RANGES ................................................................................................................... 175
FIGURE 101: LUNG MODEL RESISTANCE IMAGES AND RANGES IN MANDATORY MODES ........................................................... 176
FIGURE 102: LUNG MODEL RESISTANCE IMAGES AND RANGES IN SPONTANEOUS MODES ....................................................... 176
FIGURE 103: WEANING TARGET TOOL LOCATION .............................................................................................................................. 177
FIGURE 104: WEANING TARGET TOOLS FOR DIFFERENT TYPE OF PATIENTS. .............................................................................. 177
FIGURE 105: V–CMV MODE ..................................................................................................................................................................... 179
FIGURE 106: V-SIMV MODE ..................................................................................................................................................................... 180
FIGURE 107: P-CMV MODE...................................................................................................................................................................... 182
FIGURE 108: P–SIMV MODE .................................................................................................................................................................... 183
FIGURE 109: SPONT MODE ..................................................................................................................................................................... 187
FIGURE 110: SPAP MODE........................................................................................................................................................................ 189
FIGURE 111: ADAPTIVE WEANING CRITERIA, VE ASSURED. ............................................................................................................. 193
FIGURE 112: CURRENT SETTINGS STANDARD CRITERIA .................................................................................................................. 193
FIGURE 113: CURRENT SETTINS, ADVANCE CRITERIA. ..................................................................................................................... 194
FIGURE 114: AUTO CONTROL WINDOWS ............................................................................................................................................. 196
FIGURE 115: AUTO CONTROL STATISTICS ........................................................................................................................................... 197
FIGURE 116: AUTO CONTROL SETTINGS.............................................................................................................................................. 198
FIGURE 117: HIGH PRESSURE SYSTEM, PNEUMATIC SCHEMATIC................................................................................................... 205
FIGURE 118: OPTIONAL, TURBINE BASED SYSTEM, PNEUMATIC SCHEMATIC ............................................................................... 206
Index of Tables
Revision History
Table 1: Revision History
Susan Gonzalez J Changed 180 lpm to 200 lpm in sections 1 and 10. 27 Feb 2020
Facundo Bermejo L Update section 1.8.5 and Tidal Volume and Target Tidal Volume and 19 April 2021
section 1.8.6.
Only qualified and properly trained personnel should attempt to use, service, or
maintain the eVolution ventilator. Read this User Manual carefully and keep
available for reference.
The device is not to be used in the presence of flammable anesthetics.
Before operating the eVolution, check the ventilator for proper operation by
performing the System Test, Alarms Test and Calibration procedures described in
this manual.
1.2 Contraindications:
Do not operate the ventilator in a magnetic resonance imaging (MRI) environment.
1.4 Safety
The symbols below draw your attention specifically to the remaining dangers associated with proper
use and to emphasize important technical requirements.
To ensure proper ventilator operation and avoid the possibility of physical injury, only qualified
medical personnel should attempt to set up the ventilator and administer treatment with the
ventilator.
In case of ventilator failure, the lack of immediate access to appropriate alternative means of
ventilation can result in patient death. An alternative source of ventilation, such as a self- inflating
manual bag resuscitator (as specified in ISO 10651-4 with mask) should always be available
when using the ventilator.
If any damage to the ventilator is apparent, its life-supporting function can no longer be
guaranteed. Stop using the ventilator immediately and use an alternate form of ventilation.
The ventilator system is not intended to be a comprehensive monitoring device and does not
activate alarms for all types of conditions. For a detailed understanding of ventilator operations,
be sure to thoroughly read this manual before attempting to use the ventilator system.
Do not add attachments or other components or sub-assemblies to the ventilator breathing
system can change the pressure gradient across the ventilator breathing system and that such
changes to the ventilator breathing system can adversely affect the ventilator performance
Nebulization or humidification can increase the resistance of breathing system filters and that
the operator needs to monitor the breathing system filter frequently for increased resistance and
blockage.
Ventilator accuracy can be affected by the gas added by use of an external nebulizer.
To prevent patient injury, do not make unauthorized modifications to the ventilator.
The audio alarm volume level is adjustable. The operator should set the volume at a level that
allows the operator to distinguish the audio alarm above background noise levels.
Preventive maintenance, cleaning and sterilization activities must be conducted in accordance
with the procedures and recommended intervals detailed in this manual to ensure prolonged
operation of the eVolution
Before each use, check the water traps on the back (gas inputs) for any residual water or
particles.
Do not replace any accessories or other parts of the eVolution while a patient is being
ventilated.
• Environment of Use
Do not position the ventilator next to anything that blocks or restricts the gas inlet or cooling air
circulation openings, gas exhaust port, fan intake, or alarm speaker, as this may:
Limit the air circulation around the ventilator, potentially causing over-heating;
Limit the ventilator's ability to exhaust patient exhaled gas leading to potential harm;
Limit the clinician’s ability to hear ventilator alarms.
To avoid injury, do not position the ventilator in a way that makes it difficult to disconnect the
patient.
To ensure proper operation, do not position the ventilator in a way that makes it difficult to
access the AC power plug.
Do not use the ventilator in a hyperbaric chamber. It has not been validated for use in this
environment.
Do not use the ventilator in the presence of strong magnetic fields. Doing so could cause a
ventilator malfunction.
Do not use the ventilator during radiotherapy (i.e. cancer treatment using ionizing radiation), as
doing so could cause a ventilator malfunction.
To avoid the risk of ventilator malfunction, operate the ventilator in an environment that meets
specifications.
Nitric oxide shall not be attached to the ventilator inlet gas source.
14 eVolution®3e Series User Manual (International Version) EVM500041, Rev M
Section 1 Introduction to the eVolution 3e Ventilator
Before activating any part of the ventilator, be sure to check the equipment for proper operation
and, if appropriate, run system test, oxygen sensor calibration and proximal flow sensor
calibration.
Lock the ventilator’s casters during use to avoid the possibility of extubation due to inadvertent
ventilator movement.
The eVolution Ventilator is not intended to be a comprehensive vital sign monitor for patients on
life-support equipment.
An alternative means of ventilation shall be available whenever the ventilator is in use. If a fault
is detected in the ventilator or its life-support functions are in doubt, disconnect the ventilator
from the patient and immediately start ventilation with such a device (for example, a resuscitation
bag). The ventilator must be removed from clinical use and serviced by eVent Medical authorized
service personnel.
It is recommended that additional independent monitoring devices be used during mechanical
ventilation. The operator of the ventilator must still maintain full responsibility for proper
ventilation and patient safety in all situations.
Check that the remote alarm system (nurse call) is activated before leaving the patient
unattended.
Do not put a vessel filled with a liquid on the ventilator. If a liquid enters the product, a fire and/or
electric shock may occur.
When using non-invasive ventilation (NIV), the patient’s actual exhaled volume may differ from
the exhaled volume reported by the ventilator due to leaks around the mask.
To reduce the risk of fire or explosion, do not place the eVolution ventilator in a combustible or
explosive environment (for example, around flammable anesthetics or other ignition sources).
Do not use it with any equipment contaminated with oil or grease.
Only medically pure oxygen should be used for ventilation. Do not use anesthetics and
potentially explosive gases. Ensure that oxygen supplies are completely oil-free.
To avoid any potential fire hazard, keep all matches, lighted cigarettes, and other sources of
ignition away from the device.
To minimize the risk of fire, do not use high-pressure gas hoses that are worn or contaminated
with combustible materials like grease or oil.
To reduce the risk of fire, use only breathing circuits intended for use in oxygen-enriched
environments. Do not use antistatic or electrically conductive tubing.
In case of fire, immediately secure the patient’s ventilatory needs, switch off the eVolution
ventilator, and disconnect it from its gas and electrical sources.
Replacement of ventilator batteries by inadequately trained personnel could result in an
unacceptable risk, such as excessive temperatures, fire, or explosion.
Class I ME Equipment: to avoid the risk of electric shock, this equipment must only be connected
to a supply main with protective earth
If the “Battery not Available” alarm is displayed at startup, take the eVolution ventilator to your
Biomedical service department to ensure that the internal battery is installed and or recharged
before use.
EVM500041 Rev M eVolution®3e Series User Manual (International Version) 15
Section 1 Introduction to the eVolution 3e Ventilator
If the “Battery not Available” alarm message is displayed on the user interface during use, you
must check the battery status as soon as possible to prevent ventilator inoperable in case of AC
mains power is lost. If no action is taken and AC mains power is lost, the battery will not work
and the ventilator will shut down with high pitch continuous buzzer alarm (Ventilator Inoperable).
The eVolution ventilator is not intended for use during transports.
To ensure battery backup power is available, do not use the ventilator if the internal battery is
not installed or if the internal battery has a charge of less than 50%.
The eVolution ventilator should always be plugged in to an AC power source during use and
storage to ensure the devices internal battery remains charged.
Before using the eVolution ventilator, power on the device, disconnect it from the AC power
outlet, and ensure the battery charge indicator positioned in the upper right corner of the screen
displays a charge of at least 50%.
If the ventilator has been stored for an extended period without being plugged into an AC power
source, verify the devices internal battery has at least a 50% charge and connect to an AC power
source before use.
Use only a 24VDC for DC Input.
To ensure proper servicing and to prevent possible physical injury, only qualified personnel
should attempt to service the ventilator.
Do not service the ventilator or replace parts during patient ventilation or when the unit is running.
To reduce the risk of electrical shock, disconnect electrical power from the ventilator before
servicing. Be aware that battery power remains even after the mains is disconnected. Be aware
that when the inspiration ventilator is off, some parts still carry high voltage.
Do not attempt service procedures other than those specified in the service
manual. Use replacement parts supplied by eVent Medical only.
Any attempt to modify the ventilator hardware or software without the express written approval
of eVent Medical automatically voids all warranties and liabilities.
To ensure the ventilator’s safe operation, always run the tests and calibrations prescribed in this
manual before using the ventilator on a patient. If the ventilator fails any tests, remove it from
clinical use immediately. Do not use the ventilator until necessary repairs are completed and all
tests have passed.
Maintenance should always be conducted in compliance with all relevant safety regulations.
Repairs, assembly and use must be conducted by trained personnel; and the ventilator must be
checked by trained personnel annually.
Orient the CO2 sensor and adapters so that they cannot fall on the patient.
Use appropriate sampling lines or airway adapters for the patient type.
Components that are too large for the patient can add dead space to the breathing circuit,
and components that are too small for the patient can cause excessive flow resistance.
To avoid occluding of the CO2 sensor, do not use the ISA (Sidestream) CO2 sensor with
metered-dose inhalers or nebulized medications.
Verify that the correct sensor is used for the patient type.
To ensure accurate CO2 sensor zeroing, place the sensor and adaptor in a well-ventilated place
in ambient air (21% O2 and 0% CO2) and avoid breathing near the sensor/adapter before or
during the zeroing procedure.
Incorrect probe zeroing will result in false gas readings.
Measurements can be affected by mobile and RF communications equipment. Ensure that the
CO2 sensor is used in the electromagnetic environment specified in this manual.
CO2 monitoring is intended only as an adjunct to patient assessment. It must be used in
conjunction with other assessments of clinical signs and symptoms.
Replace the sampling line if the input connector LED flashes red or the ventilator displays an
occlusion message.
Do not sterilize or immerse Nomoline Family sampling lines in liquid.
User only airway T-adapters with the sampling point in the center of the adapter.
Strong scavenging suction pressure might affect the sample flow.
Due to the risk of patient cross-infection, always use a bacteria filter on the exhaust port side if
sample gas is intended to be re-breathed.
No modification of the CO2 monitoring devices is allowed.
CO2 sensors are not designed for MRI environments. During MRI scanning, place the
ventilator outside the MRI suite
• Capnography Sensors:
The IRMA airway adapters are non-sterile devices. Do not autoclave the devices as this will
damage them. Refer to IRMA Mainstream Gas Analyzer and ISA Sidestream Gas Analyzer User
Guides for cleaning instructions.
Never sterilize or immerse the CO2 sensors in liquid.
Do not apply negative pressure to the Nomoline (i.e., by syringe) to remove condensed water.
Do not apply tension to the CO2 sensor cables.
The ISA “Plug-in and measure” sensors should be securely mounted in order to avoid the risk
of damage to the ISA.
• Electromagnetic Susceptibility
The eVolution ventilator complies with the EN/IEC 60601-1-2 EMC (Electro Magnetic
Compatibility) Collateral Standard, which include E-field immunity and ESD requirements.
However, even though the device is compliant at the levels of immunity specified in the EN/IEC
60601-1-2 standard, certain transmitting devices (e.g., cellular phones, walkie-talkies, cordless
phones, paging transmitters etc...) emit radio frequencies that could potentially influence or
disturb device operation if located in close proximity to the ventilator. Practitioners should be
aware that radio frequency emissions are additive, and that the ventilator must be located a
sufficient distance from transmitting devices to avoid interruption.
Do not operate these transmitting devices within the vicinity of the eVolution ventilator. Do not
use the eVolution ventilator in an environment with magnetic resonance imaging (MRI)
equipment.
• Disposal
The ventilator system complies with the requirements of EN/IEC 60601-1-2 (EMC Collateral
Standard) including the E-field susceptibility requirements at a level of 10 volts per meter, at
frequencies from 80 MHz to 2.7 GHz. However, even at this level of device immunity, certain
transmitting devices (cellular phones, walkie-talkies, cordless phones, paging transmitters, RFID
devices, etc.) emit radio frequencies that could interrupt ventilator operation if operated in a range
too close to the ventilator. Practitioners should be aware of possible radio frequency interference if
portable devices are operated in close proximity to the ventilator. For more detail information,
consult the service manual.
SGS
1.5.6 Classification
1.5.8 Transport
Handle with care at transport. Don’t stack during transport. eVolution must be shipped
securely packaged.
During intrahospital transport:
• Make sure the unit is securely attached to the cart
• Make sure all accessories are securely attached to the unit
• Make sure the battery is fully charged
• Avoid tipping the cart
1.5.9 Storage
Don’t stack during storage. When the eVolution is in storage, keep it connected to mains AC to
maintain full charge in the batteries. Make sure temperature is -20 – 70 ºC and humidity is 10-
95% during storage.
On/Off switch
% O2 Increase key
Warning and Caution Symbol used in the device user manual: Refer to the
user manual for Information regarding warnings and Cautions intended to
prevent potential damage to the patient, caregiver or device.
Earth Ground
Hot Surface
Do Not Twist
Max O2 200 l/min The Oxygen inlet port label is positioned next to the Oxygen inlet port.
2 – 6 bar (29 – 86 psi)
Max Air 200 l/min Air inlet port label is positioned next to the Air inlet port.
2 – 6 bar (29 – 86 psi)
GUI Display: Indicates that the Vti limit alarm is set to alarm when
Vti limit is violated.
GUI Display: Indicates that the Vti limit alarm has been set to silent.
• When on internal battery this icon will be displayed in the top right of
the user interface.
On battery backup
requirements are given by the following items which are specified in this section:
1.1.1.1. Ventilator tidal volume and airway pressure accuracy delivery;
1.1.1.2. High and Low oxygen alarms;
1.1.1.3. High and Low pressure alarms;
1.1.1.4. High and Low expiratory minute volume alarms;
1.1.1.5. Occlusion alarm;
1.1.1.6. Gas supply alarm (Air and O2);
1.1.1.7. Battery alarm management
CMV P–CMV
Assist Control Continuous Mandatory Ventilation
V–CMV
(CMV)
PRVC–CMV
SIMV P–SIMV
Synchronized Intermittent Mandatory Ventilation
V–SIMV
(SIMV)
PRVC–SIMV
Hi FlowO2 N/A
High Flow Oxygen Therapy
Non-Invasive Ventilation
(mode feature for enhanced leak NIV
compensation)
I-Time
All Patient Types 0.1 to 10 s
Accuracy:± 0.02 s
PEEP / CPAP
PEEP 0 to 50 cmH2O or
Adult: 0 to (100 cmH2O -
Pcontrol, Psupport)
Pediatric: 0 to (80 cmH2O
- Pcontrol, Psupport)
Infant: 0 to (60 cmH2O -
Pcontrol, Psupport)
whichever is less
Accuracy:± 10% or ± 2cmH2O whichever is greater
Pcontrol
1 to 100 cmH2O or
Pcontrol (100 cmH2O - PEEP),
whichever is less
Accuracy:± 10% or ± 2cmH2O whichever is greater
Psupport
0 to 100 cmH2O or
Psupport (100 cmH2O - PEEP),
whichever is less
Accuracy:± 10% or ± 2cmH2O whichever is greater
Trigger Type
Flow or Pressure
Leak Comp
On or Off
Automatic Leak Compensation
If ON and NIV is set to Off the ventilator will
0.0 to 25.0 l/min
compensate for leaks automatically from:
If ON and NIV is set to ON the ventilator will
compensate for leaks automatically from:
0.0 to 60.0 l/min
Note: Leak Compensation disabled when Tube
Compensation is “on”
Base Flow
NIV
On or Off
Setting Ranges
Pause
Inspiratory Plateau
Pause Adult: 0.0 to 2.0s (default: 0.0 s)
Pediatric: 0.0 to 1.0s (default: 0.0 s)
Infant: 0.0 to 1.1s (default: 0.0 s)
Accuracy:± 0.02 s
Oxygen
O2 21 to 100 %
% O2 Increase (button) 100% O2 for 3 min (adult/ped)
+20% for 3 min (neonatal)*
Accuracy (Delivery): ± (3 %) full scale
Suction Support (optional software)
Suction Support On/Off
(for complete details on Suction Support, see Section 9.3.1)
Rise Time 1 - 20
1 = Slow and 20 = Fast
Flow Pattern
Flow Pat. Decel (Decelerating), De50% (Decelerating 50%),
or Square
Exhalation Sensitivity
Esens 1 to 80 % of peak insp flow
Unlocked
Locked
Low PEEP
Plow 0 to 50 cmH2O or
0 to Phigh or
Adult: 0 to (100 - Psup low) cmH2O
Pediatric: 0 to (80 - Psup low)
cmH2O
Infant: 0 to (60 - Psup low) cmH2O
Plow cannot be set > Phigh
Accuracy: ± 10% or ± 2cmH2O whichever is greater
Accuracy:± 0.2 s
Cycle
(Low PEEP to High PEEP cycles per minute)
Adult 1 to 60 c/min
Pediatric 1 to 120 c/min
Infant 1 to 150 c/min
Accuracy:
0 to 100 b/min (± 1 b/min)
> 100 b/min (± 2%)
H : L Ratio
H:L 1:59.0 to 59.0:1
Accuracy: Derived from accuracies for Thigh and Tlow
Hi FlowO2 Settings
To access Hi FlowO2 under Mode Types you must be in Standby mode.
Pediatric and Adults Pediatric and Adults
(Default Settings: O2: 40%, Flow: 10 l/min)
Neonate Neonate
(Default Settings: O2: 40% Flow: 2l/min)
O2 Range: 21 to 100 %
Flow Range: 2 to 12 l/min
Humidification is recommended.
Monitors
5 or 8
Select Number of parameters
displayed in the Monitoring Bar
Compliance Comp.
Compliance Compensation On or Off
Low Flow O2
(Turbine models)
Low Flow O2; On or Off
O2 Sensor; On or Off
O2 Sensor
(High pressure models)
O2 Sensor; On or Off
Audio / Visual
Audio Level; 35-100%
Screen Clicks On or Screen
Clicks Off
LCD Brightness; 20-100%
Technical
Password Access To view or adjust the additional
Technical Settings enter the
access code is 2634 and press
OK or CANCEL to return to
previous screen.
Patient Height: cm or in
Keyboard: Standard or Generic
Alpha
Ptrach
Calculated peak tracheal -50 to 120 cmH20
pressure during a breath.
Available with the tube
compensation feature and
only calculated when tube
comp is on
0 to 120 cmH2O
Accuracy: ± 10% or ± 2 cmH2O whichever is greater
Pmin
Calculated peak end -50 to 120 cmH20
expiratory tracheal pressure
during a breath. Available
with the tube compensation
feature and only calculated
when tube comp is on.
Accuracy: ± 10% or ± 2 cmH2O whichever is greater
Vti 0 to 3000 ml
Inspired tidal volume
Accuracy: 0 to 200 ml: ± (10 ml + 5%)
201 to 3000 ml: ± 10%
Accuracy: + 10%
Accuracy: + 10%
Accuracy: + 10%
Accuracy: + 10%
Accuracy:
Accuracy:
O2 15 to 103 %
Monitored FiO2 Displayed as ---- if O2 sensor is
disabled.
Accuracy: ± 0.2 s
Accuracy: ± 0.2 s
39 eVolution 3e Series User Manual (International Version)
®
EVM500041, Rev M
Section 1 Introduction to the eVolution 3e Ventilator
Accuracy: ± 0.1 s
Ti/Ttot 1 to 99.9%
Respiratory Time Fraction
Spont% 1h 0 to 100 %
Percentage of spontaneous breaths for the Displayed as ----- if Auto Control is
last 1 hour set to Off and does not display if
Auto Control is Disabled.
Spont% 8h 0 to 100 %
Percentage of spontaneous breaths for the Displayed as ----- if Auto Control is
last 8 hours set to Off and does not display if
Auto Control is Disabled.
Accuracy: ± 10%
ΔP 0 to 120 cmH2O
Driving Pressure is the ratio of tidal volume
to (static) respiratory system compliance.
Calculation: ΔP = Plateau - PEEP
Accuracy: Derived from Plateau - PEEP
Power 0 to 150 Joules/min
The total mechanical power is defined as
total work/time.
Calculation: Power=RR x Vti x WOBtot
Accuracy: Derived from RR, Vti and WOBtot
C20/C 0 to 5
Overdistention index is the ratio of the
dynamic compliance during the last 20% of
Inspiration (C20) to the total dynamic
compliance (C).
Calculation: The slope of a line drawn
between 2 points of zero flow (the start and
the end of inspiration in this case) is
compared to the slope of a line drawn
during the last 20% of inspiration. If the C20
is less than the Cdyn, the ratio between
them is <1.0 indicating overdistension. A
ratio of greater than 1.0 indicates that the
PV loop can’t be used to assess over-
distension. This happens during PC
ventilation when the ventilator maintains a
constant inspiratory pressure.
Accuracy: Derived from Vt and Ppeak
Elastance 0.001 to 1.00 cmH2O/ml
Relationship between the difference of
pressure applied to the respiratory system
vs the tidal volume
Calculation: Elastance (E ) = (Ppeak –
PEEP) / (Vte)
Accuracy: Derived from Vt and Ppeak
SI -99.99 to 99.99
Stress Index provides a tool to monitor the
occurrence of tidal recruitment and over
distention. Stress Index is only available
during Volume modes with square
waveform.
Pressure-Volume Loop
P / V Loop Measured cmH2O and ml
Pressure displayed on x-axis and volume on the y-axis
Flow-Volume Loop
F / V Loop Measured l/min and ml
Flow is displayed on the y-axis and volume on the x-axis
Pressure-Flow Loop
Measured cmH2O and ml
P / F Loop
Flow is displayed on the x-axis and pressure on the y-
axis
1, 2 or 3 curves OR 1 to 2 loops can be selected to display. The eVolution ventilator provides a
user configurable auto–scale or manual scale feature for each graphic displayed. Graphs
displayed in the right panel can be frozen and provide touch and scroll cursor control with cursor
X and Y axes values displayed.
Standard
Standard
Adult 5 or Vte Low+5 to 3500 ml
Pediatric 2 or Vte Low+2 to 1000 ml
Neonate* 1 or Vte Low+1 to 300 ml
Vte Low
Adult 5 to 3490 ml or Vte High-5, or Off
Pediatric 2 to 995 ml or Vte High-2, or Off
Neonate* 1 to 298 ml or Vte High-1, or Off
Vti Limit
Adult 0 to 3500 ml or Off(default New Patients = 15ml/kg x IBW)
Pediatric 0 to 1000 ml or Off(default New Patients = 15ml/kg x IBW)
Neonate* 0 to 300 ml or Off(default New Patients = 15ml/kg x IBW)
55 mmHg
PetCO2 High
Range: 2 (or 1 > PetCO2Low) to 150 mmHg
25 mmHg
PetCO2 Low
Range: 1 to 149 mmHg or 1 < PetCO2High, or Off
Apnea Disconnection
High MV Low MV
High O2 Pressure Low O2 Pressure
High O2 Conc Low O2 Conc
High VT Low VT
TF-XXX
Battery Too Low
(see section 5.6.10.)
High O2% Imminent Shutdown
Low O2% Connect AC Now
High O2 Inlet Pressure High Pressure
Low Air Supply Low Pressure
SBT Time
15 to 120 min
(default = 30 min)
Accuracy:
±0.1 s
PEEP
0 to 30 cmH2O
(Default = Set Non-SBT mode PEEP unless changed by User.)
Accuracy:
± (2 cmH2O + 4%)
Psupport
0 to 30 cmH2O
(Default = IBW based default or Set Non-SBT mode Psupport unless changed by User.)
Accuracy:
± (2 cmH2O + 4%)
The ventilator is intended for use in the electromagnetic environment specified below.
The user of the ventilator should assure that it is used in such an environment.
5% Ut 5% Ut
(>95% dip in Ut ) (>95% dip in Ut )
for 5 sec for 5 sec
Power Frequency
(50/60 Hz) Power frequency magnetic fields should be
magnetic field. 30 A/m 30 A/m at levels characteristic of a typical location in
a typical hospital environment.
IEC 61000-4-8
Note: Ut is the a.c. mains voltage prior to application of the test level.
10 Vrms
10 Vrms
150 kHz to
80 kHz in
Radiated RF
IEC 61000-4-
3
10 V/m
10 V/m
80 MHz to
2.7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m) (b)
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey. (c)Should be
less than the compliance level in each frequency range. (d) Interference may occur in the vicinity of
equipment marked with the following symbol.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
(a) The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz
(b) The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range of 80
MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communication equipment could cause
interference if it is inadvertently brought in to the patient areas. For this reason, an additional factor of 10/3 is used in
calculating the recommended separation distances for transmitters in these ranges.
(c ) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the ventilator is used exceeds the applicable RF
compliance level above, the ventilator should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the ventilator.
(d) Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 1V/m.
The ventilator is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The use of the ventilator can help to prevent electro-magnetic disturbances by maintaining the
minimum distance between portable and mobile RF communications equipment prior (transmitters) and
the ventilator as recommended below, according to the maximum output of power of the communications
equipment.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: The ISM (industrial, scientific and medical) bands between 150 Hz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26..957MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz
NOTE 3: An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 KHz and 80 MHz and in the frequency range
80MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could
cause interference if it is advertently brought into patient areas.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagations affected by
absorption and reflection from structures, objects and people.
Unpack, install and calibrate O2 sensor before using the ventilator. The Oxygen sensor
calibration will permit the operator to perform a calibration of the integral O2 measurement
system prior to use and as required thereafter.
Figure 1: Installing O2
Sensor
Power cords must be secured using the retaining clip to prevent inadvertent
disconnection of the power cord from the ventilator.
Do not compromise the AC power cord grounding protection by using an
ungrounded extension lead.
Use only a 12VDC for DC Input.
Fuses must comply with IEC 60127-2/5, and power cords must comply with
IEC 245 Code 53 or IEC 227 Code 53.
To ensure battery backup power is available, do not use the ventilator if the internal battery
is not installed or if the internal battery has a charge of less than 50%.
The eVolution ventilator should always be plugged in to an AC power source during use
and storage to ensure the devices internal battery remains charged.
Before using the eVolution ventilator, power on the device, disconnect it from the AC power
outlet, and ensure the battery charge indicator positioned in the upper right corner of the
screen displays a charge of at least 50%.
If the ventilator has been stored for an extended period without being plugged into an AC
power source, verify the devices internal battery has at least a 50% charge and connect
to an AC power source before use.
If the “Battery not Available” alarm message is displayed at Startup, take the eVolution
ventilator to your Biomedical service department to ensure that the internal battery is
installed and or recharged before use.
If the “Battery not Available” alarm message is displayed on the user interface during use,
check the battery status as soon as possible. If no action is taken and AC mains power is
lost, the battery will not work and the ventilator will shut down with Device Alert high pitch
buzzer alarm.
Ventilator battery replacement shall be performed only by qualified and trained personnel.
In the event of a loss of AC power, the internal battery provides temporary backup power. With
a full charge, the internal battery should power ventilator operation for a minimum of 2 hours for
eVolution HP models and a minimum of 4 hours for the eVolution Turbine models. Actual
operating time is dependent on battery age, charge level, utilization frequency, and ventilator
settings.
To ensure battery backup power is available, do not use the ventilator if the internal battery
is not installed or if the internal battery has a charge of less than 50%.
The eVolution ventilator should always be plugged in to an AC power source during use
and storage to ensure the devices internal battery remains charged.
Before using the eVolution ventilator, power on the device, disconnect it from the AC power
outlet, and ensure the battery charge indicator positioned in the upper right corner of the
screen displays a charge of at least 50%.
If the ventilator has been stored for an extended period without being plugged into an AC
power source, verify the devices internal battery has at least a 50% charge and connect
to an AC power source before use.
If the “Battery not Available” alarm message is displayed at Startup, take the eVolution
ventilator to your Biomedical service department to ensure that the internal battery is
installed and or recharged before use.
If the “Battery not Available” alarm message is displayed on the user interface during use,
check the battery status as soon as possible. If no action is taken and AC mains power is
lost, the battery will not work and the ventilator will shut down with Device Alert high pitch
buzzer alarm.
Ventilator battery replacement shall be performed only by qualified and trained personnel.
The gas supply hose used for oxygen must be approved for at least 147 psi
(10 bar) pressure. The permissible gas supply operating pressure range is
29to 86 psi (2 to 6 bar) for High pressure O2 and 5 to 29 psi (0.3 to 2 bar) for
the Low flow O2 inlet.
Use medically pure oxygen only.
Oil impairs the individual ventilator components' ability to operate. Oxygen
supplies must be completely oil-free. Fault-free operation cannot otherwise be
guaranteed.
Warning: pure oxygen combined with oil is highly explosive!
Oxygen Air
connection connection nut
Supply pressure Supply pressure
29 to 86 psi 29 to 86 psi
(2 to 6 bar) (2 to 6 bar)
For additional details on CliniNet® Virtual Report™ please visit our website at
www.event-medical.com/products/clininet.
To order our free (30 day) trial software, please contact customer service at
[email protected] or contact our office directly.
1 Not in use
2 NC (Normal Closed)
3 NO (Normal Open)
4 Not in use
5 Common
6 Not in use
Figure 10: Nurse Call Connector Pin Assignments
Use the eVolution ventilator only with standard breathing circuits (single patient use or reusable)
that comply with international standards ASTM F1100-90, ISO 5356-1, EN1281-1 and EN12342.
The breathing circuit may include these components:
standard neonate/infant circuit with neonate / infant nasal prongs replacing the wye (Figure
10). Use of a proximal pressure line with the eVolution ventilators is not required when using
nasal prong interfaces. NCPAP+ mode is appropriate for use with most nasal prong
systems as shown below.
Attach the nasal prong interface to the circuit as required. Cap or close any open pressure
ports on the nasal prong interface as the eVolution NCPAP+ mode does not require use of a
proximal pressure line.
Tube
Tube
Mounting Holder
Flex Arm Holder
Bracket
Connection
Screw
2.13.1 Classifications
IRMA Mainstream Gas Analyzer:
• According to the degree of protection against harmful ingress of water or particulate
matter:- IP44
• According to sterility: The IRMA system contains no sterile parts.
• According to the mode of operation: CONTINUOUS OPERATION
• According to the degree of protection against electric shock: The IRMA probe is
classified as DEFIBRILLATION-PROOF TYPE BF APPLIED PART
Figure 15: Attaching the IRMA Mainstream CO2 Sensor to the Breathing Circuit
• To attach the IRMA sensor to the Adult or Infant airway adapter, snap the sensor
head onto the airway adapter as shown above in Figure 13. The sensor head clicks
into place when properly seated. The airway adapter will only connect to the sensor
in one orientation as shown above in Figure 13.
• Connect (interface) the airway adapter to the circuit:
o Connect the 15mm male connector to the Wye.
o Connect the 15mm female connector to the patient interface.
Do not place the airway adapter between the endotracheal tube and an elbow:
Doing so can allow secretions to block adapter windows, resulting in incorrect
CO2 sensor operation.
• Orient the airway adapter and sensor head/LED facing up as shown in Figure 13.
Avoid direct contact with the patient, and use insulation material if needed.
To keep secretions and moisture from pooling in the airway adapter windows,
position the adapter and probe head so that the IRMA green LED faces up.
• Verify that the sensor head and airway adapter are securely attached to the
breathing circuit.
• When the ventilator is on, a green LED on the CO2sensor indicates that the CO2
monitor is being powered and properly connected.
• Zero CO2Sensor Calibration:
o Snap in a new IRMA airway adapter into the IRMA probe, without connecting
the airway adapter to the patient circuit, enter Standby and then navigate to the
New or Previous Patient Start Up screen and press the Calibrations button.
o Select the Zero CO2 Sensor tab from the Calibrations screen as shown below in
Figure 14.
Zeroing the CO2 sensor should be performed only when an offset in gas
value is observed, or when an unspecified accuracy message is displayed.
o If the CO2 airway adaptor is positioned by mistake between the flow sensor and
circuit Wye, the Capnography waveforms and curves may display frequent
downward dips during exhalation due to the flow sensor continuous FiO2
compensation purge flow.
Reposition the CO2 airway adaptor between the breathing tube and the
flow sensor to resume normal Capnography performance.
ISA
Mounting Sensor
Allen Bracket
Wrench
Figure 18: Attaching the ISA Sidestream CO2 Sensor to the Ventilator
To reduce the risk of explosion, do not place the CO2 sensor in a combustible or
explosive environment (for example, around flammable anesthetics or other
ignition sources).
Position airway adapters with windows in a vertical, not a horizontal, position. This
helps keep patient secretions from pooling on the windows.
To prevent premature failure of the CO2 sensor, eVent Medical recommends that
you remove it from the circuit whenever an aerosolized medication is delivered.
This is due to the increased viscosity of the medication, which may contaminate
the airway adapter window.
• Connect a Nomoline Family sampling line to the ISA analyzer input connector. When
the ventilator is on, a green LED on the ISA CO2 sensor indicates that the sensor is
being powered and properly connected.
• Perform an initial pre-use check on the ISA sidestream CO2sensor:
o Block the sampling line for 10 seconds and verify that the CO2 monitor LED
flashes red indicating that the line is occluded.
o Attach the sample line to a sample port on bacteria filter and breathe into the
filter to verify that the ventilator displays valid CO2 waveforms and values.
• Attach the sample line to the sample port of an airway adapter or bacteria filter and
interface with the breathing circuit. Verify that the sampling line is securely attached.
o If the CO2 airway adaptor is positioned by mistake between the flow sensor and
circuit Wye, the Capnography waveforms and curves may display frequent
downward dips during exhalation due to the flow sensor continuous FiO2
compensation purge flow.
Reposition the CO2 airway adaptor between the breathing tube and the
flow sensor to resume normal Capnography performance.
• Replacement of the sampling line: The Nomoline sampling line is for single
patient use (non-reusable). The Nomoline sampling line should be replaced
every two weeks or whenever “Sampling Line Clogged” appears (whichever
comes first).
3 Operation
The evolution ventilator must be operated only by clinicians qualified to assess
the patient’s condition and treatment.
Touch screen or Knob turns to highlight fields and keys (current selections are highlighted), and pushing the knob
selects a field or confirms selections
o Patient Setup: Selecting this option will stop ventilation and load the patient setup
screen.
o Standby: Selecting this option will stop ventilation, place the device in standby and load
the main screen. Standby displays in the upper left corner below the date and time.
o Cancel: Selecting this option will close this screen, ventilation continues uninterrupted
and the previous screen will load.
Neonate Patients*
o Once pressed, 20% above the set FiO2 will be delivered
o An indicator appears in the status bar when the +20% O2 function is active.
After three minutes the delivered oxygen reverts to its original setting. Pressing 100% O2
before the interval is complete turns O2 ⬆ delivery off and the delivered oxygen reverts back
to its original setting. Oxygen related alarms are silenced during the 100% delivery phase. If
the oxygen gas source is disconnected or interrupted, the “Oxygen Supply” and “O2 Increase
Not Avail.” alarm will be triggered. If corrected these alarms automatically silence and latch.
In addition to its normal functionality, when the O2 ⬆Key is pressed a one point 100%
oxygen sensor calibration will be performed (Adult/Pedi patients only).
Alarm Pause/Silence: Pressing the Alarm Silence Key silence key silences the
audible portion of the currently active alarms for two minutes. If the alarm
condition is resolved, the visual indicator disappears. If the alarm condition has
not been resolved, the alarm remains latched on the display screen until the
condition is resolved. If a new alarm event occurs an alarm will sound and
display. Pressing the alarm silence key again will deactivate the current alarm
silence session.
(Exception for Battery Too Low, Imminent Shutdown and Connect AC Now)
Alarm Off: Pressing and holding the Alarm Silence Key for two seconds will
disable all alarms from sounding for two minutes. If a new alarm event occurs
no alarm will sound but all alarms will display. Pressing the alarm silence key
again will deactivate the Alarm Off session.
(Exception for Battery Too Low, Imminent Shutdown and Connect AC Now)
3.3 Calibration
Prior to installation, the eVolution ventilator should receive a full system and touch screen
calibration for higher precision. For additional information, please refer to the eVolution
Technical Service Manual. See index for keyword calibration to view specific calibration
criteria. See Section 6.1 for additional information on system calibrations.
The term Touch Screen refers to the front LCD screen of the eVolution, which is configured
with touch screen functionality.
The following information provides a general understanding of the touch screen functionality for
the eVolution ventilator.
• The eVolution when shipped will be configured with the latest software for that model.
• You must be familiar with using the touch screen and knobs to select, activate and
confirm parameters.
(For additional technical information, please refer to the eVolution Technical Service
manual)
The Patient Type (Adult, Pediatric or Infant) is visible from all user screens and is
Patient Type
displayed in the top left section of the screen
The current active ventilation mode is visible from all user screens and is displayed
three different ways depending on the following.
BOLD BLUE: During normal ventilation the Current Mode will be displayed in bold
blue. When the ventilator is in Standby, the word Standby is displayed bold blue in the
Current Mode Current Mode field.
When the selected mode is AUTO Control and the patient has transitions to the
corresponding spontaneous mode, the Current Mode (PRVC-CMV and PRVC-PS) will
be displayed in bold blue.
BOLD RED: During Apnea Backup ventilation the Current Mode will be displayed in bold red.
A square information icon is located to the left of the current mode display.
For Mandatory delivered breaths, this icon will display a White “M” letter with a Blue
background, indicating that the current breath being delivered is a Mandatory breath.
This icon will display a White “M” letter with a Grey background during the expiratory
Patient Effort /
phase and until a new delivered breath.
Trigger Indicator
For Spontaneous delivered breaths this square icon will display a White “S” letter with
a Blue background, indicating that the current breath being delivered is a Spontaneous
breath. This icon will display a White “S” letter with a Grey background during the
expiratory phase and until a new triggered breath.
The Status Bar is used to display criteria based indicator icons which inform the user of
specific conditions or events as follows.
Indicates the NIV control is set to ON. When NIV is active the
NIV ventilator will compensate for leaks up to 60.0 l/min
O2 Sensor Off Indicates the Oxygen sensor has been disabled (turned Off)
Indicates the Manual Inspiration Key has been pressed and a Manual
Manual Insp
breath has been delivered.
Status Bar
Nebu Indicates the Nebulizer has been turned on and is active.
The current date and time is visible from all user screens and is displayed in the top
Current Date and left section of the screen.
Time Date and Time settings are provided in the password protected Technical
section on the Config screen.
The following tabs are displayed for screen navigation and positioned below the
Status Bar on the Home screen as well as all other user screens.
Navigation Tabs (Home, Settings, Alarms, Config)
Touch-selecting one of these navigation tabs loads the screen and user options for
that tab.
Up to three simultaneous active alarm indicators can be displayed in the Active
Alarm Field and will be visible on Main screen as well as all other user screens.
Active alarms display on the alarm field according to priority from left to right.
Alarm Message
Once an alarm condition is corrected, the alarm message will latch (remain displayed
on the alarm bar) touch the alarm message to clear it. To clear all latched (inactive)
alarms, press and hold a latched alarm message for at least 2 seconds.
The monitoring bar is below the Alarm Display Field and is where the user defined
monitored parameters are displayed.
User can choose a monitored parameter for each monitor field.
Monitoring Bar By default 5 monitored parameters are displayed on the Monitoring Bar. The
user can choose to have five (5) or eight (8) monitored parameters displayed on
the Monitoring Bar.
The number of monitored parameters displayed on the Monitoring Bar can be
adjusted on the Config screen Graphics page under Audio/Visual Monitors.
The Left display panel is visible from all user screens and is positioned to the Left of
the Right display panel. From the Home screen, user can click on Left Panel toggle
buttons for Monitors or Maneuvers. When selecting Monitors tab, the user can choose
to display; Loops (P/F loop, P/V loop, F/V loop) or the Lung (Lung Model).
Each screens Left panel is used to display different but specific selections for
Left Display Panel that screen.
The last selected left panel will remain selected until changed or New Patient is
started.
Loop Display (Line Colors)
Mandatory breaths: (red = inspiration) and (light blue = exhalation)
Spontaneous breaths: (yellow = inspiration) and (light blue = exhalation)
The Right display panel is visible from all user screens and is positioned to the right
of the left display panel. From the Home screen, user can view configurable options
of 1, 2 or 3 waveforms, or 1, 2 or 3 loops.
User can use the right panel toggle buttons to choose the order of waveforms
Right Display Panel or loop display.
Waveform and Loop Display (Line Colors)
Mandatory breaths: (red = inspiration) and (light blue = exhalation)
Spontaneous breaths: (yellow = inspiration) and (light blue = exhalation)
The commonly adjusted settings specific to the current or selected proposed mode
Settings Bar are displayed on the settings bar and allow users quick access to change these
parameters.
To the right of the settings bar, the following action button options will be available
as described below.
START Button: At startup prior to mode activation from all user screens accept the
Current Settings screen, a grey START button will be displayed. Pressing the grey
START button will load the Current Setting Screen and the button will change to
green, indicating the mode can then be activated.
During ventilation, when selecting a new mode of ventilation from the Settings Tab,
the Proposed Settings screen loads and a green START button displays in the
START bottom right of the screen,
Note: There is a 2 minute timeout feature associated with non-activated mode
selections as follows:
A mode selection (during ventilation or startup) will remain in the selected state
for 2 minutes post user interaction. If this 120 second time frame elapses before
the next user interaction, the Settings screen will revert back to the Current or
default mode and settings.
Once START is pressed or upon navigation away from the Proposed Settings
screen, the green START button will automatically change to the Unlock icon.
UNLOCK / LOCK: Once a mode is activated the Unlock button will display in its place
indicating the screen is not locked. Pressing the Unlock button locks the screen and
LOCK / UNLOCK the Lock button displays indicating the screen is locked. To unlock the screen press
the Lock button, the screen will unlock and the Unlock button will again display.
Alarm Silence / Alarm Silence / Alarms Pause: The icons for Alarm Silence/Pause will display in
Alarms Pause the lower right hand corner of the screen.
4.5 Safety
4.5.1 Screen Display Defaults to Main Screen (2 minute rule)
During ventilation, the ventilator’s UI display will automatically change to the Home Screen
two minutes (120 seconds) post user interaction. This rule applies to all ventilation
screens/windows with the exception of those listed below.
Screens/Windows excluded from 2-minute rule
SCREENS (excluded) WINDOWS/MENUS (excluded)
Home All Home Screen windows
Monitoring All Monitoring Data windows
Engineering All Trend Data windows
During Standby: Prior to mode activation, the ventilator will be in the Standby mode. While
in Standby, the graphics panel on the Home screen will display the standby safety message
(Standby NOT Ventilating!) as shown below in Figure 24. While the ventilator is in Standby,
the ventilator’s UI display will automatically change to the Home screen two minutes (120
seconds) post user interaction. This ensures the Standby NOT Ventilating! messages
shown below is displayed to the user. This means while the ventilator is in Standby, if the user
navigates to any non-engineering screen, the screen will automatically change to the Home
screen two minutes (120 seconds) post user interaction.
is provided.
The Soft Boundary will be reached when one of the applicable controls is
selected and adjusted to a value of:
When the user selects one of the setting controls, the control will display in
its selected state with the following visual markers.
• A WHITE arrow will display on the controls green border
• Min value is at the lower-left corner
• Max value (Max Boundary) is at the lower-right corner
• A Soft Boundary is the point that the border changes from green color to light green color.
5 Ventilator Set Up
Only trained users should change ventilator configuration: changing the
configuration can have serious consequences for subsequent ventilation.
An alternate form of oxygen monitoring must be provided if the internal oxygen
sensor is turned off in the configuration screen.
• WELCOME SCREEN
• eVent Medical Logo
• eVolution Ventilator
Logo
• Bars of arch light up
one by one from small
to large and are timed
with boot-up sequence
• Indicated Self Test in
progress
For New Patient Startup, IBW is a required setting that must be entered before the Home
screen will load.
As shown below in Figure 30, pressing the ACCEPT button without establishing or manually
entering an IBW will result in the below message being displayed in the right panel above
the ACCEPT button. The IBW is used to establish initial ventilator and alarm settings and
therefore is required before the settings screen will load.
Previous patient:
Retains ventilator parameters at previous settings based on saved input. Ideal Body Weight
and Humidification setting can be verified and changed here if necessary. User may also
access and perform calibrations; Exhalation Flow Sensor Calibration, Systems Test and O2
Sensor Calibration at this time.
5.2.6 Calibrations
Takes user to the Calibration options screen where you can perform the following:
• Circuit Check
o Sensor Zeroing: zeroing of the pressure sensors
o Compliance: Calculation of tubing compliance.
o Safety Valve: Safety valve checking
o Insp. and Exp. Resistance: Calculation of tubing resistance.
• O2 Sensor: Calibrates the O2 Sensor
• EXH Flow Sensor: Perform complete calibration of the exh flow sensor
Setting Description
Mode Settings Panel Set mode CMV → P-CMV or V-CMV or PRVC-CMV; SIMV →P-SIMV
or V-SIMV or PRVC-SIMV; SPONT → PS or VS or SPAP
Current Settings Panel Set advanced parameter settings (parameters displayed dependent
on mode selected) includes Smart Nebulizer, Sigh, Tube Comp. and
Apnea Backup.
Basic Parameter Setting Bar Set basic parameter settings (parameters displayed are dependent
on mode selected)
Monitored Patient Parameters
Bar Display five (5) or eight (8) user-selectable monitoring values.
START
Initiates ventilation on user set parameters, switches to Home Screen
Adult/pediatric
patients
Neonatal patients
Once the Mode and Breath Type are selected, review and adjust all settings
(Current/Proposed, Apnea Backup and Alarm settings) as required. Only the setting
applicable to the selected mode will be displayed.
The user must select desired advanced parameter settings from the Advanced Parameter
panel, and the desired basic parameter settings from the basic parameter bar. Only active
settings are displayed. When ready, press the START button to start ventilation. If the selected
mode is not activated within two minutes (120 s) from the last user interaction, the ventilator
will revert back the previous settings.
Neonate:
1 to 60 l/min
Ptrig Pressure trigger sets the patient inspiratory pressure that Adult:
triggers a breath. Applies to all modes except NCPAP+. 0.2 to 20.0 cmH2O
NOTE: Adjusting Ptrig changes ventilator sensitivity to patient
effort. In case of auto-triggering, check the ventilator and Pediatric:
breathing circuit before increasing the Ptrig setting. 0.2 to 15.0 cmH2O
Neonate:
0.1 to 10.0 cmH2O
PEEP/CPAP PEEP (positive end expiratory pressure) and CPAP 0 to 50 cmH2O or
(continuous positive airway pressure), baseline pressures 0 to (90 cmH2O –
applied during the expiratory phase. Pcontrol, Psupport),
Applies to all breaths in all modes except SPAP whichever is less
Phigh Sets the value of the baseline pressure during Thigh. Applies 5 to 50 cmH2O or
to SPAP mode only. Plow to 50 cmH2O or
5 to (90 - Psup high)
Pause Sets the time (in seconds) that inspiration is paused after set 0 to 2 sec
tidal volume has been delivered. Applies to Volume Control
mandatory breaths only.
Plow Sets the value of the baseline pressure during Tlow. 0 to 50 cmH2O or
0 to Phigh or
Applies to SPAP mode only. 0 to (100 - Psup low)
Psup high A positive airway pressure setting in SPAP. Pressure (above 0 to 100cmH2Oor
Phigh) to be applied to a patient-triggered breath during the 0 to (100 - Phigh)
Thigh phase in SPAP. Applies to all Thigh cycles in SPAP whichever is less
mode. Phigh + Psup High will not exceed 90 cmH2O
Psup low A positive airway pressure setting in SPAP. Pressure (above 0 to 90 cmH2O or
Plow) to be applied to a patient-triggered breath during the 0 to (90 - Plow)
Tlow phase in SPAP. Applies to all Tlow cycles in SPAP whichever is less
mode. Plow + Psup Low will not exceed 90 cmH2O
Psupport Pressure (above PEEP/CPAP) to be applied to a patient 0 to 100cmH2O or
triggered breath during the inspiratory phase. 0 to (100 cmH2O –
Applies to spontaneous breaths in SIMV and PS modes. PEEP),
whichever is less
Rate Respiratory rate sets the number of mandatory breaths Adult:
delivered by the ventilator. 1 to 60 b/min
Applies to all CMV and SIMV modes.
Pediatric:
1 to 120 b/min
(Not available in PS, VS, SPAP)
Neonate:
1 to 150 b/min
Thigh Duration of Phigh airway pressure level. 0.1 to (60 - Tlow
Applies to all Phigh phases in SPAP setting)
Ti Set inspiratory time allowed to deliver a mandatory breath. 0.1 to 10.0 sec
Applies to P-CMV, P-SIMV, PRVC-CMV and PRVC-SIMV
breaths only.
Neonate:
2to 100 ml
+O2 Available only in Apnea Backup settings. Sets the FiO2 used 10-80 %
during Apnea Backup. FiO2 delivered equals Current FiO2
plus number chosen in +02 setting.
• The nebulizer is Volume and Flow compensated only, FiO2 is compensated +/- 10 %
during nebulizer use.
• Optional Blower Base System: The nebulizer function is not available when high
pressure O2 is not available. The nebulizer is Volume and Flow compensated only,
FiO2 is not compensated during nebulizer use.
During the nebulization time, the nebulizer valve opens synchronous to inspiration to
Smart Nebulizer feature will be disabled and the “Nebulizer Not Available” information
message will display to the user when:
o the High pressure oxygen supply is not available
o Orin V-Modes and VTV-Modes if
Vt setting – Vnebulizer + Vcompliance = < 100ml
o During nebulizer activation or use when delivered flow drops below 5 l/min.
o Check Events
o Change to specific Trend
When TC is activated, the ventilator automatically calculates the pressure drop across the
breathing tube. The ventilator then adjusts the airway pressure to deliver the set pressure (PC
or PS setting) to the distal (carinal) end of the breathing tube.
• TC: Off
• Tube Type: ETT
• Tube Diameter (new patient):
o IBW 0.2 – 5 kg* 3.0 mm (allowable range 2.0 – 6.0)
o IBW 5.1 – 10 kg 4.0 mm (allowable range 3.0 – 8.0)
o IBW 11 – 40 kg 5.0 mm (allowable range 4.0 – 8.0)
o IBW > 40 kg 7.5 mm (allowable range 5.0 – 10.0)
• TC Factor: 80%
A dotted line will be displayed 5 cmH2O below the High Pressure Alarm setting to represent
the pressure limit for TC breaths (based on alarm setting).
The SBT mode is a weaning tool provided for the clinician to evaluate the patient’s ability to
tolerate spontaneous breathing with minimal support in an effort to determine if the patient is
ready for weaning and or extubation according to clinician’s judgment.
o Clinicians should follow established scientific clinical guidelines for weaning criteria and
patient readiness for extubation.
o SBT Mode is not available for Neonate / Infant patients.
Setting Description
Patient Type The current Patient Type is displayed above the current mode
Current Mode The current ventilation mode is displayed
Setting Description
Monitoring Data Allows user to view 3 tabs of monitored data; Basic, Advanced and
Mechanics
Table 14: Monitoring Tab Descriptions
Setting Description
Trend Data Allows user to view 1, 2 or 3 trend graphs of any of 24selectabletrended
parameters as shown below.
Setting Description
Sets upper and lower limits automatically based on Setting and current measured
parameters
o For NCPAP+ mode, the Auto Set Alarm feature has no affect because all settable
alarms on the Alarms screen are disabled in NCPAP+ Mode.
Setting Description
Event Log Records the last one thousand (1,000) events. The Event Log is
accessed under the Alarms Tab. The Event Log stores the last one
thousand (1,000) events. Data can be filtered by Date, Type and
Event.
Table 21: Alarms Settings Descriptions
Alarms are always displayed in order of priority from left to right across the top of the screen.
Informational messages are only shown once per occurrence.
Once an alarm condition is corrected, the alarm message will latch (remain displayed) touch
the alarm message to clear it. To clear all latched (inactive) alarms, press and hold a latched
alarm message for at least 2 seconds.
To silence an active alarm for two minutes: Press and release the Alarm Silence
Key. A new alarm condition cancels the alarm silence. Pressing and holding the
Alarm Silence Key for 2 seconds silences all current and new alarms for 2 minutes
(Exception for Battery Too Low, Imminent Shutdown and Connect AC Now)
During an alarm silence, all current and new alarms are visually displayed with their
corresponding alarm priority message on the Alarm bar until resolved and canceled as
described above in Table 22.
Air Supply The air supply has been interrupted. Ventilation continues with
100% Oxygen, if
This alarm applies to the HP models only.
connected to an oxygen
high pressure gas source.
Oxygen Supply The oxygen supply has been interrupted. Ventilation continues with
air supply (21%).
Blower The Blower/Turbine gas supply has been Ventilation continues with
interrupted due to a blower failure. 100% Oxygen, if
connected to an oxygen
This alarm applies to the blower/turbine
high pressure gas source.
models only.
High Inh Pressure Patient system pressure has reached the Ventilator immediately
high pressure alarm setting. switches to exhalation.
Ventilation continues but
pressure is limited at the
alarm setting.
Low Inh Pressure Pressure cannot be attained in the patient Ventilation continues
breathing circuit, possible due to a leak or
disconnect.
Apnea The set Apnea alarm time has elapsed Device will look for
without a patient triggered or mandatory patient’s next inspiratory
breath being delivered as set on the Alarm effort. After two
Settings screen Apnea time setting. consecutive spontaneous
breaths the Apnea alarm
silences and the Apnea
alarm message will latch.
Note:
If apnea backup settings
are set to an Apnea
Backup mode (NOT
OFF), this alarm does not
display as the Apnea
Backup Active alarm will
display because the
Apnea Backup mode will
become active.
Low O2 Pressure The oxygen supply has been interrupted. Ventilation continues with
Blower supply (21%).
Possible causes: oxygen supply not
available while 100% O2 is being
administered or O2 calibration has been
started.
High O2 Pressure The oxygen inlet pressure exceeded max Ventilation continues with
level. Blower supply (21%).
Battery Too Low AC power disconnected. Critical battery Ventilation continues but
supply. will shutdown if not
connected to AC power.
Replace HEPA Filter The HEPA filter is dirty and should be Ventilation continues
replaced. (eVolution Turbine model only)
High Rate Respiratory rate is too high when compared with Ventilation continues
the high rate alarm limit.
Low Rate Respiratory rate is too low when compared with Ventilation continues
the low rate alarm limit.
Low PEEP Patient system PEEP pressure is too low Ventilation continues
compared with the set alarm limit for PEEP Low.
High PEEP Patient system PEEP pressure is too high Ventilation continues
compared with the set alarm limit for PEEP High.
Over Temperature The temperature inside the device is too high. Ventilation continues.
High Leak Rate Monitored value for Leak % is too high compared Ventilation continues
with the set alarm limit for Leak %.
Volume Not Delivered Upper limit of operational pressure for VTV Ventilation continues.
ventilation reached, further increase not possible
(Pop high = Ppeak alarm – 5mbar) Volume delivered is maybe
too low
Vti Limit Reached Delivered volume is being limited by Vti Limit Ventilation continues with
alarm settings delivered volume limited.
Check Pcontrol / Pmax Pcontrol and/or high-pressure alarm settings are Ventilation continues
incorrect i.e. Pcontrol > high-pressure alarm
setting. Pressure is limited at the
high-pressure alarm setting
until the settings are
corrected.
Check Psupport / Pmax Psupport and/or high-pressure alarm settings are Ventilation continues
incorrect i.e. Psupport > high-pressure alarm
setting. Pressure support is limited
at the high-pressure alarm
setting until the settings are
corrected.
Low ATM Pressure Barometric pressure lower than low limit of 9.5 psi Ventilation continues
High ATM Pressure Barometric pressure higher than high limit of 16 Ventilation continues
psi
Heater Faulty Heater temperature exceeds high limit Heater is turned off and
ventilation continues
The CO2 sensor measurements are out of range 1. Verify correct airway
Ck. CO2 Sensor
Possible cause: Software Error, Hardware Error, adapter installation.
Motor Error, Calibration lost, O2 sensor error 2. Perform CO2 sensor
Ventilation continues calibration.
3. Replace CO2 sensor.
CO2 sensor hardware issue, out of calibration 1. Verify correct sensor
CO2 Sensor Accuracy?
Possible cause: N20 outside range, O2 outside installation.
range, CO2 outside range, An agent is outside 2. Perform CO2 sensor
range, Temperature outside range, Pressure calibration.
outside range, Agent unreliable
Ventilation continues 3. Replace CO2 sensor.
Information
Description of Information Ventilator Response
Message
Battery Not Available The internal battery is not available. Ventilation Continues
Possible causes: battery flat, battery missing, faulty
cable connections.
Flow Sensor Error Exhalation flow sensor failure: Device switches over to a
hardware/disconnected/malfunctioning/reading out of pressure trigger mode.
range Exhaled volume
measurement switches to
Inspired volumes.
Information
Description of Information Ventilator Response
Message
connected.
Inverse I:ERatio Ventilation settings have been set to an inverse ratio Ventilation continues
(Exhalation time > Inhalation time)
100% O2 Not The O2↑key was pressed when no oxygen source is Ventilation continues
Available available
The internal Oxygen sensor is disconnected or 1. Check to make sure
O2 Sensor Not
depleted. the O2 sensor cable is
Available Displays each time at patient startup, if the O2 sensor
connected and
is depleted or disconnected. undamaged.
Ventilation Continues. 2. Replace O2 sensor
There is no signal from the CO2 sensor. This 1. Verify CO2 adaptor
No CO2 Sensor
message is not shown if the CO2 monitoring feature connection.
is disabled. Depending on the specific reason for the
loss of signal between the ventilator and the CO2 Connect and or replace
CO2 adaptor.
sensor, other related alarms may also be triggered.
Displays anytime the ventilator loses
signal/communication with an IRMA or ISA sensor.
configuration (calibrations, settings, etc) data structure does not match the stored CRC.
This check is performed at power-up only when loading the stored configurations and
settings.
o TF-08 Air Inhalation Sensor Failure
Software tests the Air Inhalation Flow Sensor during system startup to detect sensor
failures, and declares a technical failure. During ventilation the technical failure is
declared if Air Inhalation flow sensor failure is detected and O2 is not available.
o TF-10 Extremely Low Internal Battery
o Charge
Software monitors the remaining charge of the internal battery when the system is power
by it and declares a technical failure for extremely low charge (below 10%).
o TF-11 O2 Inhalation Sensors CRC Failure
Software tests the CRC integrity of the nonvolatile stored configurations of the Oxygen
Inhalation Flow Sensors during system startup, and declares a technical failure if the data
of the sensor is corrupted.
o TF-12 O2 Inhalation Sensor Failure
Software tests the Oxygen Inhalation Flow Sensors during system startup to detect sensor
failures, and declares a technical failure.
Soft key / membrane key Mechanical or SW fault, Device fully or partially EMC and ESD testing
failure ESD Damage, EMC inoperable
Interference Contact Biomed
LCD failure LCD failure, video controller LCD remains dark, no Main LED and Audible
failure, background lighting information for user. alarms
failure or end of life, ESD,
cable, driver failure. Unusable device Contact Biomed
User uncertain, incorrect Wrong language selected in User does not understand Language-specific manuals
operation configuration labeling and labeling. User training.
Information on correct
Missing particle filter in Possible contamination.
Service tech/user maintenance in service and
the gas input Device can become faulty.
user manual.
Information on correct
Dust and dirt may enter into
No fan filter User misuse maintenance in service and
device housing.
user manual.
Setting of conflicting Lack of user knowledge or Inappropriate patient Safe setting limits
ventilation parameters simple entry mistakes ventilation
Setting of conflicting Lack of user knowledge or Inappropriate patient Detailed setting instructions
ventilation parameters simple entry mistakes ventilation in User manual
Inappropriate setting of Lack of user knowledge or Inappropriate patient Safe setting limits
VTV parameters simple entry mistakes ventilation
Inappropriate setting of Lack of user knowledge or Inappropriate patient Detailed setting instructions
VTV parameters simple entry mistakes ventilation in VTV User manual.
5.7 Capnography
Capnography is a clinical tool to evaluating the adequacy of ventilation, as the obvious goal of
ventilation is to remove the CO2 produced by the body’s metabolic processes. But conventional,
time-based Capnography allows only qualitative and semi-quantitative, and oftentimes
misleading, measurements, thus volumetric Capnography has emerged as the preferred method
to assess the quality and quantity of ventilation.
To this end, the eVolution ventilator’s volumetric Capnography feature provides the following
information.
• CO2 Elimination (VCO2/min, VteCO2) measurements provide continuous feedback regarding
ventilation and perfusion and permit assessment of metabolic rate and treatment progress.
• End-tidal CO2 (PetCO2, PeCO2, FetCO2 and FeCO2) measurement provides a proper assessment of
arterial CO2.
• Dead Space (Vdana, Vdalv) and Alveolar Volume (Valv and Valv/min) measurements permit a
proper assessment of actual alveolar ventilation.
• Physiological dead space (Vd/Vt phy) measurement permits the assessment of total dead space to
tidal volume ventilation in support of the weaning process.
• Capnograms (PCO2+Time, FCO2+Time and VteCO2+Time) provide additional CO2 elimination
visualization and the combination capnogram and spirogram SBCO2 curves (PCO2–Vte and FCO2–
Vte) shape (SlopeCO2) permits assessment of COPD, asthma, and inefficient ventilation.
The eVolution ventilator is designed to work with the Masimo IRMA (mainstream) and ISA
(Sidestream) Capnography sensors and adapters. The eVolution’s Capnography software is
designed with plug-n-play functionality and will recognize which sensor (IRMA or ISA) is attached
and will automatically load all the Capnography related monitoring and trend data parameters,
graphics screens, event markers, settings and alarms specific to the type of Masimo CO2 adaptor
attached to the ventilator.
The IRMA (Mainstream) and ISA (Sidestream) sensors and accessories are optional
components available for purchase and are required for Capnography CO2 and VCO2
monitoring.
The eVolution can display measurements derived from the CO2 sensor as numeric values,
waveforms, trends, and loops. The Capnography waveforms are a valuable clinical tool that can
be used to assess a variety of pulmonary, perfusion and metabolic conditions as well as patient
airway integrity and proper endotracheal (ET) tube placement.
The eVolution’s Capnography features provides twelve (12) monitored parameters, a PaCO2
entry panel for calculation of alveolar volumes and Vd/Vt and a Capnograms screen which
displays PCO2, FCO2, VteCO2 waveforms as well as a SBCO2 panel to view PCO2/Vte and
FCO2/Vte curves.
The CO2 sensor can be easily transferred from one eVolution ventilator to another, even "on the
fly," during ventilation.
When the Masimo CO2 sensor is disconnected, the ventilator will display an information message
alarm “No CO2 Sensor” to alert the user and all the Capnography related features will be
automatically disabled. All information is retained, however, and will resume at previous settings
once the sensor connection is re-established.
Humidification HME, Humidifier, The triple action scroll allows one of three selection options:
Type None Select HME when using a heat-moisture exchanger between the
circuit wye and patient.
Select Humidifier when using any type of heated humidifier with or
without a heated circuit.
Select None when using the ventilator with a test lung and an
unheated, non-humidified circuit.
Ventilator software uses the selected humidity type to calculate an
accurate BTPS factor for volume ventilation compensation
Smart On/Off, Nebulizer can be manually turned On or Off and configured for set
Nebulizer™ On Time, nebulization time and at a user defined interval.
Interval On/Off,
Interval Time
Smart Sigh™ On/Off, Sigh can be manually turned On or Off and configured for sigh
Sigh Factor, amplitude, sigh interval, and sigh number for single or multiple
Sigh Frequency, sighs.
Sigh Number
Alarm Option Full/Basic User may change alarm options between Full and Basic.
Full: High and Low (Ve, VTe, RR, Ppeak, and PEEP or Pmean),
Leak, Apnea and Vti Limit. Under Standby, you may change the
Alarm Selection under User Configuration between PEEP and
Pmean.
Basic: High (Ppeak, Ve, RR), Low Ve, Apnea and Leak
Monitors 5, 8 User configures how many monitored parameters can be viewed
on the top Monitored Parameter Bar
Compliance On/Off User can turn compliance compensation On or Off.
Compensation Default setting is On
O2 Sensor On/Off User can turn the O2 Sensor On or Off.
Default setting is On
Low Flow O2 On/Off User can turn Low Flow Inlet O2 Pressure On or Off.
Available on the Note, when Low Flow O2 is Activated:
Blower Based Low O2 Pressure alarm will be activate if O2 inlet pressure <1.5 psi
System Only
When the Low Flow Inlet O2 feature is enabled, connect a low flow oxygen source to the eVolution’s low flow
O2 inlet port and adjust flow as required to achieve desired FiO2. The graph shown in the figure below
provides a general reference to determine a starting point for the low flow setting. To achieve FiO2 delivery
of 100%, the flow rate of the low flow gas source will need to be 1-2 l/min higher than the ventilators peak
flow rate. Low O2 Pressure Alarm will be activated if O2 inlet pressure <1.5 psi.
Graphs ranges and setting relationships based on minimum peak flow settings. Adjust flow as required to
achieve desired FiO2
NOTE: please turn the device off and on after changing the
language.
Patient Height: cm or in
Keyboard: Standard or Generic Alpha
5.8.1 Monitors
Select Monitors under the Config Tab. The Monitors selection screen allows you to;
o Select the number of monitored parameters to display in the monitoring bar above the
left and right panels as shown below.
5.8.3 O2 Sensor:
Select O2 Sensor under the Config Tab. The O2 Sensor screen allows you to;
o Turn the O2 Sensor On/Off
The System Test and Alarms Test should always be performed before using
the ventilator.
When you turn on the ventilator, the Startup screen appears. From the Startup screen, select
CALIBRATIONS. At the Calibrations screen, you can select a calibration to perform.
Calibration
Description
Procedure
Zero Flow sensor Performs a zeroing calibration of the exhalation flow sensor
System Test Checks patient circuit for leaks,calculates the circuit’s compliance factor
for volume compensation. and checks safety valve
This process is required every time the sensor is replaced, or there is drift.
The Calibrate Exhalation F - S Zero Offset screen allows the service technician to zero the
internal exhalation flow sensor. Information on the screen gives step by step instructions on
how to perform this task. This calibration is performed if an exhalation flow sensor is replaced,
an exhalation flow related repair is made, or if incorrect exhalation volumes are observed. A
Calibration Successful message is displayed when the calibration is complete. Press the
SAVE button to save the calibration information. A Save calibration data successfully
message appears when the date is saved. For additional information, see eVolution Technical
Service Manual.
2. When ready block the circuit wye and select ‘OK’ to begin the System test procedure.
The series of tests require the wye piece be closed or blocked.
3. On completion of the test, the device will report a pass as ‘OK’ accompanied by leakage
rate. A failure will be reported with a specific error number to prompt further
investigation.
Selecting Bypass allows the user to proceed to ventilation and the previous (last
successful) System tests tubing compliance will be used. This response is then stored as
a log entry.
Selecting Cancel takes the user back to the Patient Startup screen.
Make the desired changes and perform the System test again.
• When ready, select ‘OK’ in order to begin the System test procedure.
• Repeat procedure following the on screen instructions as defined above.
• Upon completion of the test, the device will report a pass as ‘OK’, a failure will be
reported with a specific error number to prompt further investigation.
The Calibrate Exhalation Flow Sensor screen allows the service technician or medical staff to
calibrate the internal exhalation flow sensor. Information on the screen gives step by step
instructions on how to perform this task. This calibration is performed if an exhalation flow is
replaced or if incorrect exhalation volumes are observed. A Calibration Successful message
is displayed when the calibration is complete.
After the completion of each calibration, the Patient Startup screen is updated with the status
for each, and a time and date stamp for each to indicate when it was completed.
Cleaning Hot Wire Flow Sensor: ONLY clean the expiratory hot wire flow sensor using an
Ethanol solution 70% pre-soaking solution:
o Prepare an Ethanol solution 70% pre-soaking solution or equivalent
o Immerse the part to be cleaned in the prepared solution for 2-5 minutes making sure
that Exhalation Flow Sensor is submerged in the solution, and agitate periodically.
o Remove the part from the solution after 2-5 minutes and rinse immediately by
immersing in at least 1 gallon of sterile distilled water at 20-30 degrees Celsius (68-86
degrees Fahrenheit). Leave the part in the rinsing bath for 1-2 minutes, agitating it
periodically to ensure thorough rinsing.
o Visually inspect the part after removing it from the rinse to ensure that no debris
remains on the part.
o Clean the flow sensor following two (2) disinfection cycles with Ethanol solution 70%
Replace the flow sensor if the sensor is still contaminated
o Let air dry in room air
7. Spare Parts
Description Part Number
8. Preventative Maintenance
Perform all cleaning, disinfection and sterilization carefully and according to all applicable
regulations to ensure the ventilator functions safely and effectively.
The cleaning section of this manual complies ISO 14937 and ISO 17664.
Perform these preventive maintenance procedures on the eVolution at the following intervals.
Check the Cooling Fan Filter regularly for blockages and contamination and
clean as required.
EVM500041, Rev M
151 eVolution®3e Series User Manual (International Version)
Section 8 Preventative Maintenance
8.5 Fuses
As Required: Replace Fuses
EVM500041, Rev M
153 eVolution®3e Series User Manual (International Version)
Section 8 Preventative Maintenance
Follow these instructions to test ventilator operation and the alarms identified below before
attaching the eVolution ventilator to a patient.
Oxygen supply/Low oxygen alarm tests: Change the Oxygen% setting to 100% and
disconnect the oxygen supply line. Verify that the Oxygen supply alarm sounds
immediately, and that the Low oxygen alarm sounds within a few breaths. Change the
Oxygen% setting back to 21%.
Air supply/Low air pressure alarm test: Disconnect or clamp-off the air supply line.
Verify that the air supply alarm sounds immediately, and that the Low Air Pressure alarm
sounds within a few breaths. Reconnect or remove the clamp allowing air pressure to
return to normal.
Continuous High pressure alarm test: Change the Ppeak High alarm to 20 cmH2O
and block the expiratory limb of the circuit. Verify that the High pressure alarm sounds
within two breaths, and that the alarm is continuous until you unblock the circuit. Unblock
the circuit and change the Ppeak High alarm back to 40 cmH2O.
Apnea backup active alarm test: Change the apnea interval setting to 10 seconds and
respiratory rate setting to 5 b/min. Verify that the Apnea backup active alarm sounds and
apnea ventilation begins within 20 seconds (an Apnea indicator appears in the upper
right corner of the screen during apnea ventilation). Change the respiratory rate back to
10 b/min.
Power source transition: Disconnect the mains power cord and verify that an audible
tone sounds immediately. Verify that the ventilator begins battery operation without
interrupting ventilation (a battery icon and estimated capacity are displayed in the lower
left corner of the screen during battery operation). Reconnect the mains power cord and
verify that the AC power indicator is on.
Battery low and battery flat alarm: If testing of the internal battery is desired, assure
that the battery is fully charged to greater than 95%. Disconnect the AC mains and
operate the ventilator on 21%. The ventilator should operate for approximately 2 hours
for eVolution HP models and 4 hours for eVolution Turbine models, or more depending
on ventilator settings. Assure that a medium priority low battery alarm sounds as the
ventilator discharges to a low value. Assure high priority battery too low alarm sounds
and cannot be silenced. Assure that a high priority Imminent shutdown and connect ac
now alarm sounds and cannot be silenced. The ventilator should continue to run until
the ventilator detects a “Battery Flat” condition. The ventilator should turn off at this time
and immediately emit a continuous high pitch Vent Inop alarm for a minimum of one
minute. (The Vent Inop alarm will actually continue alarming for much longer than 1
minute). The ventilator will use an external battery first if attached until it is depleted and
then switch to the internal battery.
High FIO2 alarm: If testing of the high FIO2 alarm is desired, remove the cover plate for
the O2 sensor. Disconnect the electrical connector from the O2 cell. Operate the
ventilator with test settings. The O2 value will raise the high value. Assure that the high
O2 alarm sounds within a few minutes.
8.7 Disposal
To dispose of the ventilator and its components and accessories, the user must:
• Deliver the device, free of charge and duty paid, to the manufacturer for disposal,
• Surrender the device to a licensed private or public collection company, or
• Properly recycle or dispose of the parts.
All disposals must be according to the applicable local regulations. Consult the responsible
authorities for guidance and requirements. Recycling and disposal procedure must not:
• Pose risk to human health.
EVM500041, Rev M
155 eVolution®3e Series User Manual (International Version)
Section 8 Preventative Maintenance
The galvanic oxygen sensor cell may contain lead and acid. Dispose of these
materials according to hospital procedures and local governing ordinances.
Risk of electrocution! Some of the device’s electronic components carry AC
mains voltage.
Always disconnect the mains plug and remove the internal battery before
opening the ventilator.
o Cstat
Static compliance (Cstat) measures the stiffness of the lung and is automatically
calculated and displayed for all mandatory breaths when the flow is less than 5 LPM
at end inspiration. During spontaneous SIMV breaths, Cstat will only be displayed if
an Inspiratory Hold is performed for 0.2 s or more. Pressure of the stable PIP – PEEP
or Pplateau – PEEP serves as the denominator and delivered tidal volume less
compliance compensation is the numerator. Cstat = Vt/(Stable PIP – PEEP if any) or
Vt/(Pplateau – PEEP if any)
To perform a manual static compliance measurement the ventilator should be in a
volume mode with a square wave flow pattern. The square waveform provides zero
flow when inspiratory pressure is reached and therefore a static measurement can
be taken. From the Home screen Left Panel loops window scroll until the maneuver
window is displayed. Press Insp. Hold. If there is an inspiratory plateau for greater
than 200 milliseconds and a flow of less than 5 liters per minute a Cstat value will be
measured.
It should be noted that if a patient is making respiratory effort during the maneuver it
may alter measured values and give erroneous measurement of lung compliance.
o Rinsp
Inspiratory Resistance (Ri) measure the airflow resistance of the airways, circuit
tubing and lungs during inspiration. Ri is automatically calculated during V-Modes
only if the Flow Pattern is set to Square with a Pause set to > 0.20 s.
o Rexp
Expiratory Resistance (Re) measure the airflow resistance of the airways, circuit
tubing and lungs during expiration. Re is automatically calculated during V-Modes
only if the Flow Pattern is set to Any with a Pause set to > 0.20 s.
o Auto PEEP
Auto PEEP measures end expiratory pressure exceeding the set value for PEEP,
also known as intrinsic PEEP or inadvertent PEEP.Auto PEEP = Total PEEP – Set
PEEP and is calculated after a user initiated Exp hold is performed. Hold times will
affect calculated value. When hold is released, the The PEEP sample
measurement just before terminating hold is used as total PEEP.
o Pplateau
Plateau Pressure is calculated in all modes after a user initiated Insp hold is
performed and in V-Modes if a Pause of > 0.1 s is set.
o RSBI
Rapid Shallow Breathing Index is derived from the measured breathing rate divided
by Insp tidal volume.
o ΔP
Driving Pressure is the ratio of tidal volume to (static) respiratory system compliance.
Calculation: ΔP = Plateau - PEEP
o P0.1
Occlusion Pressure at 100 ms. P0.1is also known as P100 and is the maximal slope
of the airway pressure drop (Max negative Inspiratory pressure) during the first 100
ms when the airway is occluded. P0.1 indicates the patient’s respiratory drive and
efforts.
o PiMax
Maximum negative inspiratory pressure against an occluded airway (cmH2O), an
indicator of inspiratory muscle strength. Not available for neonatal/infant patients.
o P0.1 / PiMax
Airway occlusion pressure (P0.1) divided by maximal inspiratory pressure (%).
o WOBimp
Work of Breathing Imposed (WOBimp) is a monitored and trended parameter defined
as the work performed by the patient to breathe through the ventilator’s valves,
breathing circuit, and humidifier. When a patient is actively initiating a breath the
airway pressure is decreased slightly to signal the ventilator to commence
inspiration. Imposed work of breathing is the work required by the patient to
overcome the resistance imposed by the breathing apparatus in order to initiate a
breath. As such, WOBimp is only measured during patient initiated breaths and it is
the change in airway pressure (Paw) integrated over tidal volume (Vti) normalized to
total inspiratory tidal volume.
o RCe
Expiratory Time Constant (RCe) is monitored and trended parameter defined as the
product of the total respiratory system compliance and the total expiratory resistance.
Where R is the respiratory resistance and C is respiratory system compliance. The
product of R and C has units of time and is called the time constant. During
mechanical ventilation, RCe is the product of resistance and compliance of the
patient's respiratory system and the additional resistive elements, such as the
endotracheal tube (ETT), filters, ventilator tubing and the ventilators expiratory valve.
Since ETT resistance is markedly flow-dependent, RCe is also flow-dependent.
RCe is determined using the calculated method which allows easy calculation of RCe
breath to breath in all modes and is proven to be an accurate approximation of actual
RCe. By analysis of the slope of the expiratory flow volume curve, this method
determines the expiratory time constant using the ratio of exhaled tidal volume to
peak expiratory flow at 75% of Expiration. Thus RCe is calculated breath by breath
as the relationship (ratio) between Vte and PFe at 75% of the maximum expiratory
volume.
Calculated Expiratory Time Constant: Vte / PFe = 1 Time Constant (s)
Controlled
Flow Inhalation Exhalation
Qendinsuff
Qplateau
Tpause
Tinhale Texhale
Resulting Tcycle
Pressure
Pendinsuff
Pplateau ΔP
PEEP
Figure 83: Flow and Pressure Waveforms Cstat, Rinsp & Rexp
The pressure-time waveform can help assess pressure settings or triggering thresholds.
• The flow-time waveform can be used to assess peak spontaneous flows during
spontaneous breathing before and after bronchodilator therapy.
• The flow-time curve can also be used to detect intrinsic PEEP (also known as occult or
auto PEEP), when expiratory flows do not reach baseline before the next breath. The
ventilator displays measured circuit PEEP above the PEEP setting, identifying intrinsic
PEEP.
• To determine if there is additional pressure in the lungs, select Exp. Hold during active
exhalation. This allows gas in the patient’s lungs to equilibrate with the circuit pressure,
and the ventilator displays the equilibrated total PEEP and Auto PEEP at the next
update: total measured PEEP - set PEEP = Auto (intrinsic) PEEP.
Flow Volume Loop Pressure Volume Loop Pressure Flow Loop Maneuvers Screen
Up to three trended parameters can be displayed at one time. The configuration of the trend
display is accessed in the Trend tab. Select the Trend tab and it will change the number of
waveforms from 1 to 3.
To change the trended parameter touch the heading of the graphic trend, a list of all the
trended parameters will be displayed. Select the parameter desired and the trend display
will be updated.
9.3. Maneuvers
To access respiratory maneuvers, from the Home screen, touch the left panel Maneuvers tab.
disconnection is not detected within 120s, the O2% will return to original set.
During the Post Oxygenation phase, the ventilator will recognize the breathing
circuit reconnection and will resume ventilation with Post Oxygenation period of 1
minute unless canceled by the user.
o Additional Information
STOP button: the STOP button will cancel the Suction Support any time is pressed
and normal ventilation will be resumed with original set FiO2.
Screen Navigation: during Preparation Phase and Patient Disconnected Phase,
you must STOP the procedure if you want to select different Tabs such as Home,
Settings, Alarms and Config. Only during the Post Oxygenation phase, the
navigation to another tab is allowed.
Information message: the “Suction Active” information message priority alarm will
display on the alarm bar.
Automatic Event Marker: each time the Suction Support maneuver is started, the
Event Log will update to display an Automatic Event Marker corresponding with the
exact time the maneuver was started.
• Once P2 is reached:
o the P0.1 maneuver is complete and the P0.1 monitored parameter will update to
display the measured P0.1 value.
o the inspiratory valve opens
o the current modes breath type will be delivered and normal ventilation will resume
o The P0.1 value for each P0.1 maneuver performed will be recorded in the Event
Log for review.
9.3.2.2 P0.1 Automatic Termination Criteria
• As a safety precaution the P0.1 maneuver once started will automatically terminate
(STOP) based on the following IBW specific time based criteria.
o This means if the P0.1 maneuver is started and the patient fails to generate the
negative pressure (P1) within the background Patient Type (IBW defined) specific
time based criteria, the P0.1 maneuver will automatically terminate (STOP) and
normal ventilation will resume.
In Spont modes, Insp Hold is limited to protect the patient. The maximum amount of time will
be 0.25s, unless the user turns this safety limit OFF in Clinical configuration screen.
The alveolar inspired Lung Model is located under the monitors tab. The second location of the Lung
Model is in the right panel. You will find the Lung Model symbol located as a tab on the left side
frame of the right panel. Once selected, the Lung Model will open up in the right panel. To close this,
select the tab again.
An Information button is located next to the Lung Model symbol. When selected, a brief
description of the Lung Model appears. This information is shown in a yellow font color.
When operating under SPONT mode, the following parameters are displayed:
• Cdyn
• RCe
• Vt/kgIBW
• RSBI
• Rate Spont
Note: if Cstat and Rinsp would like to be obtained, the user must use the inspiratory hold
Compliance
Resistance
Figure 102: Lung Model Resistance Images and Ranges in mandatory modes
Figure 103: Lung Model Resistance Images and Ranges in spontaneous modes
select the hexagon symbol . This tool is only available in SPONT modes. When the Weaning
Target Tool is selected, it will display the following parameters:
• Rate Spont
• Vt/kgIBW
• RSBI (adult and pediatric patients)
• MAP (neonatal patients)
• PS
• PEEP
• O2
The Weaning Target Tool displays the Height and IBW of the patient. Height and IBW are displayed as a
reminder to the clinician. This shows the patients inputted values. The age of the patient can be set as well.
By doing this, the ventilator will automatically set the Rate Spont setting according to a NIH scale, which is
based on Normal ranges of respiratory rate.
The yellow ball is the actual value of that parameter. Its position is related to the difference to the target
set by the user. By setting the inputted parameter, you are able to adjust the target value for your patient.
Since each patient is different, this tool allows for customization of each parameter for any patient. When
the yellow ball gets closer to the green target, it means the actual value is getting closer to the set target
value.
On the same ball the current monitored value is showed.
Depending on the type of patient the weaning target tool will show different goals. Below it is showed the
different Weaning Target tools
EVM500041, Rev M
177 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
The set High Pressure Alarm is the pressure limit of the breathing circuit. A
High Pressure Alarm will terminate inspiration and the total set Tidal Volume
will not be delivered. An unattended High Pressure Alarm may result in
patient hypoventilation.
The patient can trigger a mandatory breath during the exhalation phase trigger window.
EVM500041, Rev M
179 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
Mandator
During the spontaneous phase, peak inspiratory flow is the criterion for switching from
inspiration to exhalation. When inspiratory flow drops to the set exhalation sensitivity (Esens),
which is a percentage of peak flow, exhalation begins. Esens controls the desired exhalation
sensitivity. Example: Esens is set to 40%, if the peak inspiratory flow is 60 l/min, the inspiratory
phase will end when the peak flow decelerates to 24 l/min. The patient’s inspiratory effort will
determine peak flow during spontaneous breathing.
In SIMV, the rate setting determines the number of mandatory breaths per minute. The peak
flow setting applies to the mandatory breaths.
PS Tmax PS Tmax is the max time limit control setting for all modes s
with a pressure support (PS) setting or PS based breaths. PS
Tmax functions as the inspiration termination criteria for all PS
based breaths. PS breaths when delivered will terminate by
flow, pressure and time. Whichever termination criterion is
met first will cause the PS breath to cycle to exhalation.
NOTE: When Psupport is set to 0 cmH2O, the PS Tmax feature is
background disabled. This will prevent time based termination of the
patient’s breath and allow vital capacity (VC) measurements to be
obtained.
Leak Comp. Leak compensation (default setting is On) Up to l/min
25 l/min with NIV set to Off
Up to 60 l/min with NIV set to On
The ventilator automatically limits the setting range for Rate, I:E, Peak Flow and Plateau to ensure that:
Inspiration time is never less than 10 % of the breath.
Exhalation time is never less than 20 % of the breath.
Maximum peak flow is 200 l/min (3 l/second).
In Pressure Controlled Continuous Mandatory Ventilation (P-CMV) mode, the inspiratory and
expiratory breathing phases are time-controlled. The user sets the inspiration pressure
(Pcontrol) above the PEEP (baseline pressure), and the ventilator maintains Pcontrol during
inspiration. The patient can trigger a mandatory breath within the trigger window.
Increasing Pcontrol or Ti should result in an increase in delivered tidal volume depending on the
compliance and resistance of the patient circuit and patient. If changes occur to the compliance
and/or resistance of the patient circuit or patient this may affect the delivered tidal volume.
EVM500041, Rev M
183 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
PS Tmax PS Tmax is the max time limit control setting for all s
modes with a pressure support (PS) setting or PS based
breaths. PS Tmax functions as the inspiration
termination criteria for all PS based breaths. PS breaths
when delivered will terminate by flow, pressure and
time. Whichever termination criterion is met first will
cause the PS breath to cycle to exhalation.
NOTE: When Psupport is set to 0 cmH2O, the PS Tmax
feature is background disabled. This will prevent time based
termination of the patient’s breath and allow vital capacity (VC)
measurements to be obtained.
Leak Comp. Leak compensation (default setting is On) l/min
Up to 25 l/min with NIV set to Off
Up to 60 l/min with NIV set to On
VTV provides pressure based breaths (with decelerating flow pattern and square pressure
pattern) that target a set tidal volume by adjusting delivered pressure on a breath-to-breath
basis. VTV modes deliver pressure controlled or pressure-supported breaths, the pressure
varies automatically and as needed to achieve the user defined target tidal volume.
Delivered pressure levels do not vary by more than 3 cmH2O from breath to breath. The
ventilator starts by delivering a standard volume ventilation test breath with a decelerating flow
curve and short plateau. If the plateau pressure is stable, the ventilator calculates compliance
and establishes a pressure target for the first PRVC breath. The ventilator then delivers a
pressure control style breath with the Ti set at that starting pressure.
During VTV an automatic over volume alarm is set at 1.5 times the set
target tidal volume. To prevent excessive tidal volume delivery, the breath
will terminate when the inspired tidal volume alarm limit (Vti Lim) is reached
or when a volume of 1.5 x the set tidal volume (Vt) has been reached,
whichever value is lower.
EVM500041, Rev M
185 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
PS Tmax PS Tmax is the max time limit control setting for all modes s
with a pressure support (PS) setting or PS based breaths.
PS Tmax functions as the inspiration termination criteria
for all PS based breaths. PS breaths when delivered will
terminate by flow, pressure and time. Whichever
termination criterion is met first will cause the PS breath to
cycle to exhalation.
NOTE: When Psupport is set to 0 cmH2O, the PS Tmax feature
is background disabled. This will prevent time based termination
of the patient’s breath and allow vital capacity (VC)
measurements to be obtained.
Leak Comp. Leak compensation (default setting is On) l/min
Up to 25 l/min with NIV set to Off
Up to 60 l/min with NIV set to On
PS Tmax PS Tmax is the max time limit control setting for all modes s
with a pressure support (PS) setting or PS based breaths.
PS Tmax functions as the inspiration termination criteria
for all PS based breaths. PS breaths when delivered will
terminate by flow, pressure and time. Whichever
termination criterion is met first will cause the PS breath to
cycle to exhalation.
NOTE: When Psupport is set to 0 cmH2O, the PS Tmax feature
is background disabled. This will prevent time based termination
of the patient’s breath and allow vital capacity (VC)
measurements to be obtained.
Leak Comp. Leak compensation (default setting is On) l/min
Up to 25 l/min with NIV set to Off
Up to 60 l/min with NIV set to On
EVM500041, Rev M
187 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
EVM500041, Rev M
189 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
PS Tmax PS Tmax is the max time limit control setting for all modes with a s
pressure support (PS) setting or PS based breaths. PS Tmax
functions as the inspiration termination criteria for all PS based
breaths. PS breaths when delivered will terminate by flow,
pressure and time. Whichever termination criterion is met first will
cause the PS breath to cycle to exhalation.
NOTE: When Psupport is set to 0 cmH2O, the PS Tmax feature
is background disabled. This will prevent time based termination
of the patient’s breath and allow vital capacity (VC)
measurements to be obtained.
SPAP Type Varies the type of controls in SPAP
(Cycle + Time), (Cycle + Ratio) or (Time Only)
Leak Comp. Leak compensation (default setting is On) l/min
Up to 25 l/min with NIV set to Off
Up to 60 l/min with NIV set to On
The ventilator automatically limits the setting ranges to ensure that:
Insufflation time is never less than 100 ms.
Exhalation time is never less than 200 ms.
As with every adaptive ventilation system, there are inputs to the system, outputs of the system and a
feedback that helps to control the system reducing the difference to the target. In this case the system is
the ventilator.
EVM500041, Rev M
191 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
• Inspiratory pressure: Since Auto Control would change from PRVC-CMV to PRVC-PS and vice
versa if the patient needs more support, the level of pressure would change every breath trying
to get the lowest level of pressure that allows the volume target.
• Mandatory rate: When the patient is being ventilated with PRVC-CMV, the mandatory rate is
what the user selected. After CMV hold is elapsed and also ATT is OFF or Break time is
elapsed, then a trigger trial could begin (see below definition).
Trigger Trial:It is a defined as the process when the ventilator slowly reduces the mandatory support and
monitors the total breathing rate and the minute volume of the patient. If the patient can keep the total
breathing rate and the minute volume within a set range, then the ventilator would change modes.
The range of total breathing rate should be above “low Ve %” and below the high rate alarm. The range
of minute volume should also be above “low Ve %” multiplied by the Set Ve.
• Vt assured: When Vt assured is selected the ventilator would try to wean the patient keeping
the Vt at target. Vt is one of the settings of the mode.
• Ve assured: When Ve assured is selected, the ventilator would try to wean the patient keeping
the minute volume at target. User sets the Ve (minute volume) as part of the setting. Then
based on the set rate the ventilator would get the current Vt target.
Under current settings, the first setting on the right it is called “criteria”. There are two possible choices:
Standard and advance criteria. Both choices, Standard and Advance,both have a set of variables that
defines how Auto control works. The main difference is that Standard has fixed variables and Advance
allows the user to set different values for each parameter. A change on this setting would change the
behavior of the algorithm. This will allow the user to set the ventilator based on what the user believes is
the best for the patient. Based on this setting, the behavior of the ventilator will emulate the protocol that
the user would normally follow, however now it will work automatically.
The group of variables that will affect the behavior of the ventilator can be found under the window
called “Adaptive Weaning Criteria”. Following is an explanation of each variable:
showed; , then the max Ppeak acceptable would autoset at 5 cmH2O below the high Ppeak
limit alarm (see alarms section for more information)
• Low Ve% (%): Lowest Ve % acceptable during PS mode compared to set Ve at CMV mode.
• CMV hold (hours): This is the time the patient would stay at CMV. When this time has elapsed,
then the ventilator will begin a trigger trial (see below more information)
• Adaptive Trigger Trial (min): In case ATT is ON, the ventilator will monitor the activity of the
patient. In case the patient triggers breaths for 3 consecutive minutes in PRVC-CMV mode, then
the ventilator will begin to wean the patient. The trigger trials would be suspended for “Break time”
in case the weaning of the patient failed.
Standard Criteria:
To begin the Auto Control, Standard criteria (Ve assured or Vt assured), please select the mode and standard
criteria and then press Start.
By default, Auto control is using this algorithm. Patient initially will be ventilated with PRVC-CMV
and will stay for the first 30 minutes. User could press the trigger trial button and begin a trigger
trial at any moment.
During these 30 minutes, the ventilator will start trigger trial anytime when Adaptive Trigger Trial
(ATT) detects a patient breath. The “Break time” is set at 30 at 30 minutes
After 30 minutes the patient will go through a process called trigger trial (explained before). If the patient can
maintain the minute volume and total breathing rate within the set range, then the ventilator will switch from
PRVC-CMV to PRVC-PS.
In case the patient is not able to keep the ventilation targets within the set range, then the patient will stay at
PRVC-CMV. The ventilator will adapt the setting shown on the adaptive weaning criteria window, based on
the history of the patient. The ventilator will increase the time for the next trigger trial if the patient fails.
Advanced criteria:
The other option to use Auto control is the advance criteria. This would allow the user to change all the
Adaptive weaning criteria variables. By default the initial setting would be the same as the Standard critieria
offers.
Apnea: This is the time that the ventilator would take as “apnea” when the patient has no breath during this
period. This value is the same value that is set at the alarm limits. If the alarm limit is changed, this parameter
will also be changed and vice versa. If the patient has no breath during this period, then the patient would
switch from PRVC-PS to PRVC-CMV.
Ppeak: PRVC changes the level of pressure trying to get a targer tidal volume. If the monitored volume is
below the target volume, then the pressure will increase (see PRVC mode for more information). If the
monitored volume is above the target volume, then the inspiratory pressure will be reduced. So this mode
EVM500041, Rev M
193 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
regulates the inspiratory pressure to achieve a target volume. The range of pressure where PRVC modulates
the inspiratory pressure is given indireclty by the level of PEEP + 10 cmH2O and the level of Ppeak set at this
parameter. By default is set at 5 cmH2O below the high pressure alarm. User could change this value.
Low Ve% / Low vt%: In case the user selects Ve assured, then this parameter would be Low Ve %. In case
the user sets Vt assured then this parameter would be Low Vt%. This is the lowest Ve or Vt% acceptable
during PS mode compared to set Ve or vt at CMV mode.
During Ve assured, the ventilator would compare the monitored Ve with the target Ve. During Vt assured the
ventilator will compare the monitored exhalated Vt with the target Vt. In case the patient cannot keep the
target levels, then the ventilator will switch over PRVC-CMV.
CMV hold: Set by default at 0.5 hs (30 minutes). If CMV hold is increased then the amount of time that the
patient will stay at PRVC-CMV would increase.
ATT Break time: If the ATT Break is different than OFF, when the patient triggers the ventilator for 3
consecutive minutes, then a trigger trial would begin. If the ATT break time is increased, then after a trigger
trial is performed without success, then the patient would stay more time at PRVC-CMV until the next trigger
trial. (the amount of time set under ATT break time). The user could also set ATT break time at OFF. In this
case if the patient would stay at PRVC-CMV the time set at CMV hold. ATT is an exit of PRVC-CMV in case
the patient has trigger activity.
To begin the Auto Control, Advanced criteria (Ve assured or Vt assured), please select the mode and
Advanced criteria and then press Start.
Auto Control would begin ventilation with PRVC-CMV (see detailed information of this mode in previous
sections). The patient would stay in this mode initially for 30 minutes by default. The time the patient stays at
PRVC-CMV is set by CMV hold (under Adaptive weaning criteria). After this time elapses the ventilator would
begin a trigger trial. The mandatory rate would be slowly reduced allowing the patient to adapt to this change
of mandatory support. As the mandatory rate is reduced and the patient begins to trigger and keeps the Ve
or Vt within the setting (Low Ve% or Low Vt% setting) the ventilator would switch from PRVC-CMV to PRVC-
PS.
If the patient is not able to keep the target of Ve or Vt, then the patient would stay at PRVC-CMV. CMV hold
will set the time between trigger trials.
In case the user would like to do a trigger trial, there is a button called Trigger Trial. When this button is
pressed, even if the CMV hold has not elapsed, the patient would go through a trigger trial.
Another way to allow the ventilator to do a trigger trial automatically is using the Adaptive Trigger trial (see
Adaptive Weaning Criteria Adaptive T. T.). If ATT is ON (use set a break time), then the ventilator would
constantly observe if there is activity of the patient. Then if the patient triggers at least one breath for 3
consecutive minutes, then the ventilator would automatically begin a trigger trial. If patient is not able to keep
the targets, then patient would stay at PRVC-CMV. In order to avoid another trigger trial, ATT allows to set a
Break time. By default, ATT is set at 30 minutes. This means that if the patient fails the trigger trial, even if
Changing Criteria’s
It is possible for the user to change criteria from Standard to Advance and vice versa. In order to do that go to
“settings”, then “Auto Control”, then “Current Settings” and push the criteria button. By doing that the ventilator
would request confirmation opening a window as show on the below graphic.
If the user presses “Continue”, and then “Start”, then the ventilator will continue with the current mode of
ventilation using previous settings. If use press “ReStart” and then “start”, then ventilator will restart the whole
process again. In this last case the mode would be PRVC-CMV and the settings would be the default settings.
When patient is in Auto Control and being ventilated with PRVC-PS, the ventilator will analyze if the patient
meets conditions to proceed with a SBT. If those conditions are met, then the ventilator will activate the SBT
advisory message.
SBT Advisory Conditions are:
• FiO2<=50 %
• PEEP<= 8 cmH2O
• PS<= 10 cmH2O
• Monitored Ve >= set Ve (Ve assured)
• Monitored Vt >= set Vt (Vt assured)
• Set rate <+ total rate < High rate alarm
• Vt >= 5 ml/Kg IWB
“SBT advisory” message will appear with only one beep. It will disappear once user click off the green bar. It will
come back in 24 hours or when the Auto Control was restarted, whichever comes earlier
As it was explained before, the user would select Low vt% or Low Ve% (depending on the selection, Vt
assured or Ve assured). This is the lowest Ve % acceptable during PS mode compared to set Ve at CMV
mode. The yellow curve represents this parameter.
Not all the combinations of rate and volume would be safe for the patient. Therefore, a light blue rectangle is
showed as the patient’s safe zone called lung protective zone.
At the bottom of the screen the user would be able to see the amount of time the patient has been ventilated
in Spont mode or the time to elapse the CMV hold time A short explanation of all this representation would
Another window available when auto control is being used, Is the auto control statistics. Press the following
button to access it:
On this window there are 2 pies that show in grey the % of time on mandatory mode (PRVC-CMV) and in
yellow the % of time the patient was ventilated in spontaneous mode (PRVC-PS). The first pie shows the total
time % since the patient was in Auto control mode. The second pie would show the last trends, this is how
the patient is doing over the last period. The period that the user would like to see can be set. On the above
example it was set at 1hrs however it could be set up to 8 hs..
In case the ventilator is also used in other modes of ventilation, then a white portion would be shown.
Also, on the left of each pie the user will see the amount of mode changes and the amount of trigger trials the
patient had over that period.
At the bottom of the screen the user would be able to see the amount of time the patient has been ventilated
in Spont mode or the time to elapse the CMV hold time.
EVM500041, Rev M
197 eVolution®3e Series User Manual (International Version)
Section 10 Theory of Operation
. Trigger Trial Activates a trigger trial when user press it ON/OFF
A qualified caregiver must evaluate whether settings and alarms limits are
clinically appropriate.
If Flow trigger is selected, a breath will initiate when a patient effort decreases the bias flow in
the circuit equal to the value set by the flow sensitivity (Ftrig control). With Leak Compensation
On, bias flow is equal to the value of the current measured leak flow rate plus 4 l/min.
EVM500041, Rev M
199 eVolution®3e Series User Manual (International Version)
Section 11 Abbreviations
Term Definition
Ratio of time at high to low PEEP levels
H:L
(SPAP mode setting and monitored value)
l Liter
l/min Liters per minute
LCD Liquid crystal display
Leak % Calculated leak Percentage
Leak compensation is a setting that allows the ventilator to automatically
Leak Comp
compensate for leaks during ventilation.
Milibar
mbar
(1 mbar equals 1 hPa, which equals 1.016 cmH2O)
Milliliter
ml
(1 ml = 10 –3 l)
ms Millisecond
min Minute
NCPAP+ Nasal continuous positive airway pressure plus with rate and base flow
NIV Non-Invasive Ventilation
O2 Oxygen and or Delivery Concentration
+ O2% Apnea Backup Oxygen Setting
OI Oxygen Index
OID Object Identification Table
PIB Product Information Bulletin
P/V Loop Pressure + Volume Curve Graph
P0.1 Airway Occlusion Pressure Measurement (manual or maneuver)
P0.1/PiMax Airway Occlusion Pressure to maximal inspiratory pressure ratio
Pause Inspiratory Pause Time Setting
P–CMV Pressure Control Continuous Mandatory Ventilation
Term Definition
Control pressure, inspiratory pressure above PEEP delivered during a
Pcontrol
pressure-controlled breath (Pressure above PEEP)
PEEP Peak End Expiratory Pressure
PEEP Positive end expiratory pressure (Peak End Expiratory Pressure)
PF Peak Inspiratory Flow
PFe Peak Expiratory Flow
Phigh High PEEP level (SPAP mode setting)
The maximum negative inspiratory pressure that is achieved by the
PiMax
patient, during a PiMax maneuver.
Plow Low PEEP level (SPAP mode setting)
Pmean Mean Airway Pressure
P0.1 Airway occlusion pressure (P100)
VTV breaths are delivered within a defined operation pressure (P op)
Pop
range PopLow to PopHigh.
Ppeak Peak Inspiratory Pressure
Pplateau Plateau Pressure
Pplateau Plateau (inspiratory hold or pause) pressure
PRVC Pressure Regulated Volume Control
Pressure Regulated Volume Control Continuous Mandatory Ventilation
PRVC–CMV
(a VTV mode)
Pressure regulated volume control synchronized intermittent mandatory
PRVC–SIMV
ventilation, a VTV mode
PS Pressure Support (SPONT Mode)
Pressure per square inch
psi
(1 bar = 14.50 psi)
Term Definition
RS232 Serial interface
Rapid Shallow Breathing Index
RSBI
(Calculated respiratory rate divided by tidal volume)
s Second
SNMP Simple Network Management Protocol
SPAP Spontaneous Positive Airway Pressure
SPONT Spontaneous Ventilation (CPAP + PS)
Square Square Flow Pattern (Ramp)
STPD Standard temperature and pressure, dry
Tapnea Apnea Alarm Time Setting
Target Vt Target tidal volume (a VTV mode setting in PRVC and VS)
Te Expiratory Time
TF Technical fault
Thigh Time at high PEEP level (SPAP mode setting)
Ti Set or Monitored inspiration time
Ti/Ttot Calculated I–Time divided by cycle time
TIB Technical Information Bulletin
Tlow Time at low PEEP level (SPAP mode setting)
Torr is a unit of measurement for pressure
Torr
(1 Torr = 1 mmHg) and (760 Torr = 1 ATM)
TS Touch screen
UI User Interface
V Volt
VA Volt-ampere (watt)
VAC Volt, alternating current
V–CMV Volume Control Continuous Mandatory Ventilation
CO2 elimination. Net exhaled volume of CO2 per minute
VCO2/min
(Displayed if optional CO2 sensor is connected)
Term Definition
Vte Exhaled Tidal Volume
Vti Inspiratory Tidal Volume
VTV Volume Targeted Ventilation includes PRVC and VS modes.
W Watt
WBM Web based monitoring
Work of Breathing Imposed: WOBimp is a monitored parameter defined
WOBimp as the work performed by the patient to breathe through the ventilator’s
valves, breathing circuit, and humidifier.
μA Microampere
Table 48: Abbreviations
13. Index
Contact Information
Clinical Support..................................................................................
1 US Office Address ............................................................................ 2
US Office Fax ................................................................................... 2
Alarm Settings US Office Phone ............................................................................... 2
Standard............................................................................................. US Sales .............................................................................................
Alarms Control Settings
Auto Set Feature ................................................................................ Soft Boundary ....................................................................................
Reset Feature ............................................................................... 125 Display
Technical Faults................................................................................ 0 Freeze / Unfreeze ..............................................................................
Alarms Medium Priority Monitors ...................................................................................... 133
SBT High Rate ..................................................................................... Parameters .................................................................................. 133
Audible alarm Selecting Parameters .........................................................................
setting level .................................................................................. 134 eVolution Ventilator part numbers ...................................................... 2
Auto Control ........................................................................................... EXH Flow Sensor Calibration ............................................................ 142
Battery Exhalation Sensitivity (Esens) .................................................................
Status Bar Icon ................................................................................... Assembly ............................................................................. 146, 147
Breath Termination Exhalation valve cover
PS Tmax .............................................................................................. Manual cleaning .......................................................................... 145
Buzzer Exhalation Valve System .................................................................. 144
setting audible alarm level ........................................................... 134 EZ-Flow sensor
reusable ....................................................................................... 143
single patient use ........................................................................ 143
C
C20/C ......................................................................................................
F
Calibrations
Calibration Required Status Bar Icon ................................................. F/V Loop .................................................................................................
EXH Flow Sensor .......................................................................... 142 Graphics
New Patient ....................................................................................... Capnograms Screen ..................................................................... 127
O2 Sensor ............................................................................. 138, 141 Changing Scale...................................................................................
Pre-Patient Tests ................................................................ , 138, 139 Cleaning and sterilization ............................................................ 143
System Test .......................................................................... 138, 139 Humidification Type ...............................................................................
Zero Flow Sensor.......................................................................... 139
Capnograms Screen Graph Selection ................................................ 131
Left Panel SBCO2 Curves............................................................... 131 I
Right Panel Waveforms................................................................ 131
Capnography Maintenance .................................................................................... 138
Mainstream (IRMA) Sensor ................................................................ Alarm Tests ........................................................................................
Circuit compliance calculation .......................................................... 138 Manual Breath
Circuit sterilization ............................................................................ 142 Front Panel Key .................................................................................
Classification ........................................................................................... Mechanics
Cleaning and sterilization.................................................................. 142 Monitored Data .................................................................................
Clearing Alarms ................................................................................. 125 Mode description
CMV Auto Control ......................................................................................
Pressure ............................................................................................. Monitored Data
Communication Advanced ...........................................................................................
CliniNet Server ................................................................................... Monitoring Bar
Compliance Selecting Parameters .........................................................................
Cdyn ................................................................................................... Nebulizer setting .............................................................................. 133
Compliance Comp ............................................................................. 135 New Patient Startup ...............................................................................
Config. Screen ......................................................................................... Nurse call
Configuration Interface pin assignments .................................................................
Compliance Comp ........................................................................ 135
Trend Data .........................................................................................
Configuration Screen Settings........................................................... 133 O
Configuration Screens ............................................................... 132, 135
Configuring Graphic Display.................................................................... O2 ⬆ Key
T
P
Table of Contents ................................................................................. 3
P/V Loop ................................................................................................. Table of Tables .......................................................................................
Part Number ..................................................................................... 148 Technical Fault ..................................................................................... 0
1881 ............................................................................................. 148 TF-01 ................................................................................................ 0
EV200012 ..................................................................................... 148 TF-02 ................................................................................................ 0
EVL220008 ................................................................................... 148 TF-03 ................................................................................................ 0
EVL220047 ................................................................................... 148 TF-04 ................................................................................................ 0
EVL370017 ................................................................................... 148 TF-05 ................................................................................................ 0
EVL500501 ................................................................................... 148 TF-06 ................................................................................................ 0
F710616 ....................................................................................... 148 TF-07 ................................................................................................ 0
F910028 ....................................................................................... 148 TF-08 ............................................................................................ 125
F910037 ....................................................................................... 148 TF-10 ............................................................................................ 125
F910038 ....................................................................................... 148 TF-11 ............................................................................................ 125
MR 850 ......................................................................................... 148 TF-12 ............................................................................................ 125
MR370.......................................................................................... 148 Technical Specifications
RT 130 .......................................................................................... 148 EMC ...................................................................................................
RT 137 .......................................................................................... 148 Ventilation Modes
Part numbers .................................................................................... 148 Types & Selection ..............................................................................
Parts list ............................................................................................ 148 Ventilator
Patient Selection – New Settings ............................................................ Alarm Definitions ...............................................................................
Product Config Screen............................................................................... 132
Disposal .............................................................................................. Controls Adjustment .........................................................................
Product Specifications exterior cleaning .......................................................................... 145
Alarm High Priority............................................................................. GUI Navigation ..................................................................................
Regulated Volume Control Continuous Ventilator Modes
Mandatory Ventilation ( PRVC-CMV ) Mode ...................................... Activating...........................................................................................
Regulated Volume Control Synchronized Ventilator Settings
Intermittent Mandatory Ventilation ( PRVC-SIMV ) Mode ................ Apnea Backup ....................................................................................
Resistance
Expiratory (Rexp) ...............................................................................
Settings W
Base Flow ...........................................................................................
Settings Config. Screen (Technical) Warning
Password Access to User Config. ....................................................... CO2 monitor placement ....................................................................
Settings SBT Mode
Sigh
Status Bar Icon ...................................................................................
SIMV
Z
Pressure .............................................................................................
SPAP Mode Zero Flow Sensor .............................................................................. 139
Sales
www.event-medical.com