0% found this document useful (0 votes)
93 views19 pages

Fugro Suhaimi Scope of Accreditation

This document provides the scope of accreditation for Fugro Suhaimi Limited's Calibration Laboratory in Dammam, Saudi Arabia. The laboratory is accredited to perform calibrations for various equipment across several parameters including acoustical, chemical, dimensional, and others. Ranges and measurement uncertainties are provided for different equipment types and parameters.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
93 views19 pages

Fugro Suhaimi Scope of Accreditation

This document provides the scope of accreditation for Fugro Suhaimi Limited's Calibration Laboratory in Dammam, Saudi Arabia. The laboratory is accredited to perform calibrations for various equipment across several parameters including acoustical, chemical, dimensional, and others. Ranges and measurement uncertainties are provided for different equipment types and parameters.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 19

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

FUGRO SUHAIMI LIMITED


Calibration Laboratory Dammam
King Abdulaziz Port Road
PO Box 2165.
DAMMAM, Kingdom of Saudi Arabia 31451
Ziyad Alrabeh Phone: 00966 13 857 4200

CALIBRATION

Valid until: November 30, 2024 Certificate Number: 3512.06

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 7:

I. Acoustical

Parameter/Equipment Range CMC2 () Comments

Sound Level Calibrators 114 dB, 1 kHz 0.99 dB Casella model CEL-
110/2

Sound Level Meters, 114 dB, 1 kHz 0.64 dB Casella model CEL-
Fixed Point 110/2

II. Chemical

Parameter/Equipment Range CMC2, 8 () Comments

pH Measuring 4 pH 0.22 pH pH buffer solutions


Equipment, Fixed 7 pH 0.24 pH
Points 10 pH 0.24 pH

TDS Measuring 1000 µg/ml 13 parts in 106 Reference solution


Equipment

(A2LA Cert. No. 3512.06) Revised 3/13/2023 Page 1 of 18


Parameter/Equipment Range CMC2, 4, 8 () Comments

Conductivity, Liquid 500 µmhos/cm


At 20 ⁰C 8.5 µmhos/cm Conductivity
At 25 ⁰C 11 µmhos/cm solutions

1413 mhos/cm
At 20 ⁰C 11 mhos/cm
At 25 ⁰C 13 mhos/cm

XRF Analyzer Cu (0.0077 to 93.95) % 5.4 % Standard reference


Sn (0.00051 to 0.270) % 0.2 % material
Pb (0.0018 to 0.33) % 0.2 %
Zn (0.111 to 0.656) % 0.22 %
Ni (0.012 to 4.17) % 0.4 %
P (0.001 to 0.318) % 0.2 %
Fe (0.060 to 2.87) % 0.22 %
Si (0.0065 to 0.35) % 0.26 %
Mn (0.001 to 8.33) % 0.41 %
As (0.001 to 0.44) % 0.2 %
Al (5.27 to 8.42) % 0.4 %
C (0.0026 to 1.18) % 0.6 %
S (0.0093 to 0.020) % 0.2 %
Cr (1.076 to 16.36) % 2.0 %
Mo (0.00077 to 1.318) % 0.4 %
Al 0.131 % 0.2 %
Co (0.0027 to 0.0048) % 0.4 %
N (0.0026 to 0.0225) % 0.4 %
Ti 0.0024 % 0.3 %
As 0.0036 % 0.01 %
Nb 0.0146 % 0.01 %
V 0.00015 % 0.01 %

Fixed-Gas Aanalyzer

H2S; 0-200 ppm 1.44 ppm Fixed Reference


CO; 0-10 ppm 0.79 ppm Gases
CH4; 0-200 % 1.17 %
O 0-25% 0.71 %

Multi-Gas Analyzer

(H2S; CO; CH4; O) ( Up to 200 ppm; Up to (5.29 % / ppm) Mixed Reference


1,000 ppm; Up to100 %; Gases
Up to 30 %)

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 2 of 18


III. Dimensional

Parameter/Equipment Range CMC2, 6 () Comments

Calipers (All Types)


Outside, Step, Depth Up to 300 mm 1.5 µm Mitutoyo gage block
set Grade 1
Inside Up to 300 mm 1.8 µm Step Gage

Inside, Outside and Up to 300 mm 1.4 µm Mitutoyo inside


Depth Micrometers micrometer checker,
gauge blocks
Anvil Flatness 25 mm 1.4 µm
Optical Flat

Theodolite, Total
Station & Auto Level

Level 0⁰ 2.0⁰ Optical collimator

Distance 34.6 m 2.5 mm Laser Distance Meter

Dial and Digital Up to 25 mm 2.1 µm Kudale dial calibration


Indicators tester

Measuring Tape/Steel Up to 1000 mm 1 mm Kudale model EL302-L


Rule scale & tape
calibrator

Height Gauge Up to 300 mm 2.2 µm Mitutoyo gage block


set Grade 1, Surface
Plate

Thickness Gauge Up to 60 mm 0.89 µm Mitutoyo gage block


set Grade 1

Coating Thickness Up to 2900 µm 2.8 µm Elcometer model


Gauge – Ferrous T99022255-2C
thickness foils set

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 3 of 18


Parameter/Equipment Range CMC2 () Comments

CBR Equipment –
ASTM D1883
Spacer Up to 180 mm 0.13 mm Caliper
Weights Up to 5 kg 0.02 kg Balance

Compactor Marshall or
Proctor – ASTM
D6926/D698/D1557
Drop Up to 458 mm 0.07 mm Tape Measure –
Weights Up to 4.55 kg 0.03 kg Balance
Hammer Diameter Up to 101 mm 0.14 mm Caliper

Conical Mold and


Tamper –
ASTM C128
Weight Up to 340 g 1.2 g Balance
Diameters Up to 90 mm 0.07 mm Caliper
Height Up to 75 mm 0.02 mm Depth Gage

BS 812
Flakiness Gauge – Openings Up to 63 mm 0.006 mm Caliper

LA Abrasion Machine –
ASTM C131 / C535
Inside Diameter Up to 711 mm 0.02 mm Measuring Tape
Inside Length Up to 508 mm 0.09 mm Measuring Tape
Average Dia of Charge Up to 47 mm 0.11 mm Caliper
Average Mass of Charge Up to 445 g 0.70 g Balance
Revolutions per Min. Up to 33 rpm 1.3 rpm Tacho
meter

Liquid Limit Device and


Groover –
ASTM D4318
Cup Radius (52 to 56) mm 1.2 mm Caliper
Dimensions Up to 150 mm 0.11 mm Caliper
Rate Up to 2 rps 0.24 rps Tachometer
rps = revolutions per
second

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 4 of 18


Parameter/Equipment Range CMC2 () Comments

Molds CBR –
ASTM D 1883
Dimensions Up to 200 mm 0.1 mm Caliper
Volume Up to 3000 ml 0.5 ml Balance

ASTM C470
Molds Cylinder – Up to 300 mm 0.45 mm Caliper
Dimensions

ASTM D 6926
Molds Marshall – Up to 100 mm 0.07 mm Caliper
Dimensions

ASTM C109
Molds Mortar Cube – Up to 50 mm 0.47 mm Caliper
Dimensions

ASTM D698/D1557
Molds Proctor – Volume Up to 3000 ml 2.3 ml Balance

ASTM D698/D1557
Proctor Straightedge – Up to 3 mm 0.13 mm Caliper
Beveled Thickness

ASTM D854, ASTM


C128, ASTM D2041
Pycnometer – Gravimetric Up to 250 mL 0.15 mL Balance
Up to 500 mL 0.17 mL
Up to 4000 mL 2.5 mL

Slump Cone and Rod –


ASTM C143
Diameters Up to 200 mm 0.12 mm Caliper
Heights/Lengths Up to 300 mm 0.13 mm Measuring Tape

Sieves (2 to 100) mm 0.5 mm Calipers

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 5 of 18


IV. Electrical – DC/Low Frequency

Parameter/Equipment Range CMC2, 4, 5, 8 () Comments

DC Voltage – Measure Up to 200 mV 14 µV/V + 0.09 µV Fluke 8508A


(0 to 2) V 9.0 µV/V + 2.3 µV
(0 to 20) V 9.0 µV/V + 16 µV
(0 to 200) V 12 µV/V + 38 µV
(0 to 1000) V 14 µV/V + 0.8 µV

DC Voltage – Generate (0 to 329.9999) mV 23 µV/V + 1.2 µV Fluke 5522A


(0.32 to 3.299999) V 11 µV/V + 2.8 µV
(3.29 to 32.99999) V 12 µV/V + 30 µV
(30 to 329.9999) V 18 µV/V + 0.24 mV
(100 to 1020) V 18 µV/V + 2.4 mV

DC High Voltage – (0 to 20) kV 0.042 V Fluke 87V


Measure³ multimeter/high
(0 to 40) kV 1.6 V voltage probe

DC Current – Generate 329.999 µA 0.014 % + 0.02 µA Fluke 5522A


3.29999 mA 93 µA/A + 0.06 µA
32.9999 mA 93 µA/A + 0.31 µA
329.999 mA 93 µA/A + 3.1 µA
1.09999 A 0.019 % + 48 µA
(1.1 to 2.99999) A 0.035 % + 48 µA
(0 to 10.9999) A 0.044 % + 0.6 mA
(11 to 20.5) A 0.093 % + 170 mA

DC Current – Measure (0 to 200) µA 44 µA/A + 0.004 µA Fluke 8508A


(0 to 2) mA 27 µA/A + 0.007 µA
(0 to 20) mA 29 µA/A + 0.04 µA
(0 to 200) mA 68 µA/A + 0.82 µA
(0 to 2) A 0.035 % + 0.03 µA
(0 to 20) A 0.052 % + 0.51 µA

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 6 of 18


Parameter/Equipment Range CMC2, 4, 5, 8 () Comments

DC Resistance – Generate (1 to 10.9999) Ω 41 µΩ/Ω + 120 µΩ Fluke 5522A


(11 to 32.9999) Ω 29 µΩ/Ω + 160 µΩ
(33 to 109.9999) Ω 25 µΩ/Ω + 190 µΩ
(110 to 329.9999) Ω 25 µΩ/Ω + 0.72 mΩ
330 Ω to 1.099 999 kΩ 25 µΩ/Ω + 1.4 mΩ
(1.1 to 3.299 999) kΩ 25 µΩ/Ω + 2.3 mΩ
(3.3 to 10.999 99) kΩ 25 µΩ/Ω + 14 mΩ
(11 to 32.999 99) kΩ 25 µΩ/Ω + 23 mΩ
(33 to 109.9999) kΩ 25 µΩ/Ω + 0.12 Ω
(110 to 329.999 99) kΩ 29 µΩ/Ω + 0.12 Ω
330 kΩ to 1.099 999 MΩ 29 µΩ/Ω + 1.2 Ω
(1.1 to 3.2999 99) MΩ 46 µΩ/Ω + 1.6 Ω
(3.3 to 10.999 99) MΩ 0.013 % + 0.01 kΩ
(11 to 32.999 99) MΩ 0.024 % + 0.02 kΩ
(33 to 109.9999) MΩ 0.047 % + 0.02 kΩ
(110 to 329.9999) MΩ 0.29 % + 0.16 kΩ
(330 to 1100) MΩ 1.4 % + 1.2 MΩ

Fixed Points (1 to 10) MΩ 0.8 % Decade Megohm


(>10 to 100) MΩ 0.8 % Box/ Vaiseshika
(>100 to 1000) MΩ 0.8 % 8400-HV
(>1 to 10) GΩ 1.2 %
(>10 to 100) GΩ 1.8 %
500 GΩ 2.2 %
1000 GΩ 4.0 %

DC Resistance – Measure (0 to 2) Ω 57 µΩ/Ω + 4 µΩ Fluke 8508A


(0 to 20) Ω 25 µΩ/Ω + 9 µΩ
(0 to 200) Ω 14 µΩ/Ω + 0.2 mΩ
(0 to 2) kΩ 12 µΩ/Ω + 0.6 mΩ
(0 to 20) kΩ 11 µΩ/Ω + 0.01 Ω
(0 to 200) kΩ 12 µΩ/Ω + 0.1 Ω
(0 to 2) MΩ 20 µΩ/Ω + 1.66 Ω
(0 to 20) MΩ 30 µΩ/Ω + 0.17 kΩ
(0 to 200) MΩ 0.030 % + 0.009 MΩ
(0 to 2) GΩ 0.2 % + 1 MΩ

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 7 of 18


Parameter/Range Frequency CMC2, 4, 5, 8 () Comments

AC Voltage – Generate

(1.0 to 32.999) mV (10 to 45) Hz 0.070 % + 7.1 µV Fluke 5522A


(0.045 to 10) kHz 0.014 % + 7.1 µV
(10 to 20) kHz 0.019 % + 7.2 µV
(20 to 50) kHz 0.093 % + 7.2 µV
(50 to 100) kHz 0.35 % + 15 µV
(100 to 500) kHz 0.70 % + 58 µV

(33 to 329.999) mV (10 to 45) Hz 0.029 % + 15 µV


(0.045 to 10) kHz 0.017 % + 15 µV
(10 to 20) kHz 0.018 % + 16 µV
(20 to 50) kHz 0.035 % + 17 µV
(50 to 100) kHz 0.070 % + 40 µV
(100 to 500) kHz 0.19 % + 83 µV

(0.33 to 3.29999) V (10 to 45) Hz 0.029 % + 60 µV


(0.045 to 10) kHz 0.017 % + 71 µV
(10 to 20) kHz 0.019 % + 71 µV
(20 to 50) kHz 0.029 % + 68 µV
(50 to 100) kHz 0.064 % + 160 µV
(100 to 500) kHz 0.24 % + 0.7 mV

(3.3 to 32.9999) V (10 to 45) Hz 0.029 % + 0.8 mV


(0.045 to 10) kHz 0.015 % + 0.7 mV
(10 to 20) kHz 0.026 % + 0.8 mV
(20 to 50) kHz 0.035 % + 0.8 mV
(50 to 100) kHz 0.087 % + 21 mV

(33 to 329.999) V (0.045 to 1) kHz 0.018 % + 29 mV


(1 to 10) kHz 0.019 % + 72 mV
(10 to 20) kHz 0.026 % + 72 mV
(20 to 50) kHz 0.028 % + 94 mV
(50 to 100) kHz 0.19 % + 59 mV

(330 to 1020) V (0.045 to 1) kHz 0.029 % + 18 mV


(1 to 5) kHz 0.024 % + 18 mV
(5 to 10) kHz 0.029 % + 25 mV

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 8 of 18


Parameter/Range Frequency CMC2 ,4, 5, 8 () Comments

AC Voltage – Measure

(0 to 200) mV 10 Hz to 20 kHz 0.025 % + 0.004 mV Fluke 8508A


(0 to 2) V 0.013 % + 0.05 mV
(0 to 20) V 0.014 % + 0.35 mV
(0 to 200) V 0.014 % + 4.0 mV
(0 to 1000) V 0.018 % + 24 mV

AC Current – Generate

(29.00 to 329.99) µA (10 to 20) Hz 2.8 µA/A + 0.85 µA Fluke 5522A


(20 to 45) Hz 0.1 µA/A + 0.12 µA
(0.045 to 1) kHz 0.1 µA/A + 0.12 µA
(1 to 5) kHz 0.3 µA/A + 0.2 µA
(5 to 10) kHz 0.7 µA/A + 0.23 µA
(10 to 30) kHz 1.4 µA/A + 0.5 µA

(0.33 to 3.29999) mA (10 to 20) Hz 0.2 µA/A + 0.27 µA


(20 to 45) Hz 0.1 µA/A + 0.27 µA
(0.045 to 1) kHz 0.1 µA/A + 0.2 µA
(1 to 5) kHz 0.2 µA/A + 0.24 µA
(5 to 10) kHz 0.5 µA/A + 0.35 µA
(10 to 30) kHz 0.9 µA/A + 0.69 µA

(3.3 to 32.9999) mA (10 to 20) Hz 0.2 µA/A + 2.3 µA


(20 to 45) Hz 0.1 µA/A + 17 µA
(0.045 to 1) kHz 0.04 µA/A + 2.3 µA
(1 to 5) kHz 0.1 µA/A + 2.3 µA
(5 to 10) kHz 0.2 µA/A + 3.5 µA
(10 to 30) kHz 0.4 µA/A + 4.6 µA

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 9 of 18


Parameter/Range Frequency CMC2, 4, 5, 8 () Comments

AC Current – Generate
(cont)

(33 to 329.999) mA (10 to 20) Hz 0.2 µA/A + 160 µA Fluke 5522A


(20 to 45) Hz 0.1 µA/A + 160 µA
(0.045 to 1) kHz 0.04 µA/A + 29 µA
(1 to 5) kHz 0.1 µA/A + 61 µA
(5 to 10) kHz 0.2 µA/A + 0.12 mA
(10 to 30) kHz 0.4 µA/A + 0.23 mA

(0.33 to 1.099 99) A (10 to 45) Hz 0.2 µA/A + 0.56 mA


(0.045 to 1) kHz 0.04 µA/A + 0.23 mA
(1 to 5) kHz 0.6 µA/A + 1.2 mA
(5 to 10) kHz 2.3 µA/A + 5.8 mA
(1.1 to 2.999 99) A (10 to 45) Hz 0.2 µA/A + 0.56 mA
(0.045 to 1) kHz 0.1 µA/A + 0.56 mA
(1 to 5) kHz 0.6 µA/A + 1.3 mA
(5 to 10) kHz 2.3 µA/A + 5.7 mA

(3 to 10.9999) A (45 to 100) Hz 0.06 µA/A + 5.9 mA


(0.1 to 1) kHz 0.1 µA/A + 2.8 mA
(1 to 5) kHz 2.9 µA/A + 5.9 mA

(11 to 20.5) A (45 to 100) Hz 0.12 µA/A + 7.9 mA


(0.2 to 1) kHz 0.15 µA/A + 7.9 mA
(1 to 5) kHz 2.9 µA/A + 12 mA

Capacitance – Generate

(220.0 to 399.9) pF 10 Hz to 10 kHz 0.60 % + 15 pF Fluke 5522A


(0.4 to 1.0999) nF 10 Hz to 10 kHz 0.60 % + 0.63 nF
(1.1 to 3.2999) nF 10 Hz to 3 kHz 0.60 % + 0.58 nF
(3.3 to 10.9999) nF 10 Hz to 1 kHz 0.30 % + 0.63 nF
(11 to 32.9999) nF 10 Hz to 1 kHz 0.30 % + 0.77 nF
(33 to 109.999) nF 10 Hz to 1 kHz 0.30 % + 2.8 nF
(110 to 329.999) nF 10 Hz to 1 kHz 0.30 % + 7.5 nF
(0.33 to 1.099 99) µF (10 to 600) Hz 0.30 % + 27 nF
(1.1 to 3.299 99) µF (10 to 300) Hz 0.30 % + 50 nF
(3.3 to 10.9999) µF (10 to 150) Hz 0.30 % + 250 nF
(11 to 32.9999) µF (10 to 120) Hz 0.50 % + 750 nF
(33 to 109.999) µF (10 to 80) Hz 0.50 % + 2.8 µF
(110 to 329.999) µF (0 to 50) Hz 0.50 % + 2.8 µF
(0.33 to 1.09999) mF (0 to 20) Hz 0.50 % + 3.0 µF
(1.1 to 3.299 99) mF (0 to 6) Hz 0.52 % + 3 µF
(3.3 to 10.9999) mF (0 to 2) Hz 0.52 % + 10 µF
(11 to 32.9999) mF (0 to 0.6) Hz 0.90 % + 100 µF

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 10 of 18


Parameter/Equipment Range CMC2 () Comments

Temperature Simulation –
Measuring Equipment³

Type K (-196 to -100) °C 0.65 °C Fluke 744


(-100 to -25) °C 0.80 °C multifunction
(-25 to 120) °C 0.94 °C calibrator, Type K TC
(120 to 1000) °C 1.25 °C
(1000 to 1372) °C 2.65 °C

Electrical Calibration of
Thermocouple Devices –

Type B (600 to 800) °C 0.55 °C Fluke 5522A


(800 to 1000) °C 0.45 °C
(1000 to 1550) °C 0.41 °C
(1550 to 1820) °C 0.44 °C

Type E (-250 to -100) °C 0.59 °C


(-100 to -25) °C 0.22 °C
(-25 to 250) °C 0.20 °C
(350 to 650) °C 0.22 °C
(650 to 1000) °C 0.27 °C

Type J (-210 to -100) °C 0.33 °C


(-100 to -30) °C 0.22 °C
(-30 to 150) °C 0.20 °C
(150 to 760) °C 0.23 °C
(760 to 1200) °C 0.29 °C

Type K (-200 to -100) °C 0.40 °C


(-100 to -25) °C 0.24 °C
(-25 to 120) °C 0.22 °C
(120 to 1000) °C 0.32 °C
(1000 to 1372) °C 0.48 °C

Type N (-200 to -100) °C 0.48 °C


(-100 to -25) °C 0.28 °C
(-25 to 120) °C 0.25 °C
(120 to 410) °C 0.24 °C
(410 to 1300) °C 0.33 °C

Type R (0 to 250) °C 0.69 °C


(250 to 400) °C 0.46 °C
(400 to 1000) °C 0.44 °C
(1000 to 1767) °C 0.51 °C

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 11 of 18


Parameter/Equipment Range CMC2 () Comments

Electrical Calibration of
Thermocouple Devices –
(cont’d)

Type S (0 to 250) °C 0.58 °C Fluke 5522A


(250 to 400) °C 0.47 °C
(400 to 1000) °C 0.48 °C
(1000 to 1767) °C 0.58 °C

Type T (-250 to -150) °C 0.74 °C


(-150 to 0) °C 0.30 °C
(0 to 120) °C 0.22 °C
(120 to 400) °C 0.20 °C

Type U (-200 to 0) °C 0.56 °C


(0 to 600) °C 0.27 °C

Electrical Calibration of
RTD Devices –

Pt 385, 100 Ω (-200 to -80) °C 0.061 °C Fluke 5522A


(-80 to 0) °C 0.061 °C
(0 to 100) °C 0.083 °C
(100 to 300) °C 0.11 °C
(300 to 400) °C 0.12 °C
(400 to 630) °C 0.14 °C
(630 to 800) °C 0.27 °C

Pt 3926, 100 Ω (-200 to -80) °C 0.05 °C


(-80 to 0) °C 0.05 °C
(0 to 100) °C 0.07 °C
(100 to 300) °C 0.09 °C
(300 to 400) °C 0.12 °C
(400 to 630) °C 0.15 °C

Pt 3916, 100 Ω (-200 to -190) °C 0.25 °C


(-190 to -80) °C 0.04 °C
(-80 to 0) °C 0.05 °C
(0 to 100) °C 0.06 °C
(100 to 260) °C 0.07 °C
(260 to 300) °C 0.08 °C
(300 to 400) °C 0.09 °C
(400 to 600) °C 0.10 °C
(600 to 630) °C 0.23 °C

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 12 of 18


Parameter/Equipment Range CMC2 () Comments

Electrical Calibration of
RTD Devices – (cont’d)

Pt 385, 200 Ω (-200 to -80) °C 0.050 °C Fluke 5522A


(-80 to 0) °C 0.050 °C
(0 to 100) °C 0.050 °C
(100 to 260) °C 0.061 °C
(260 to 300) °C 0.14 °C
(300 to 400) °C 0.16 °C
(400 to 600) °C 0.17 °C
(600 to 630) °C 0.19 °C

Pt 385, 500 Ω (-200 to -80) °C 0.050 °C


(-80 to 0) °C 0.061 °C
(0 to 100) °C 0.061 °C
(100 to 260) °C 0.072 °C
(260 to 300) °C 0.094 °C
(300 to 400) °C 0.094 °C
(400 to 600) °C 0.11 °C
(600 to 630) °C 0.13 °C

Pt 385, 1000 Ω (-200 to -80) °C 0.040 °C


(-80 to 0) °C 0.040 °C
(0 to 100) °C 0.050 °C
(100 to 260) °C 0.061 °C
(260 to 300) °C 0.10 °C
(300 to 400) °C 0.10 °C
(400 to 600) °C 0.10 °C
(600 to 630) °C 0.23 °C

Cu 427, 10 Ω (-100 to 260) °C 0.3 °C

V. Fluid Quantities

Parameter/Equipment Range CMC2, 8 () Comments

Volumetric Beakers and Up to 200 ml 0.5 µl Analytical balance,


Glassware (200 to 700) ml 2.0 µl thermometer
(700 to 2000) ml 23 µl

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 13 of 18


VI. Mechanical

Parameter/Equipment Range CMC2, 8 () Comments

Force,3 Compression – (5 to 100) kN 0.51 kN Proving ring


Measuring Equipment

Universal Testing (50 to 3000) kN 0.52 kN Matest load cell &


Machine³ (Compression) digital indicator

Torque – Measuring (41 to 410) lbfꞏft 2.3 lbfꞏft Torque tester Kudale
Equipment Instruments model
(0 to 2000) lbfꞏft 4.0 lbfꞏft (0 to 5500) lbf.in torque
tester Proto model 6190

Concrete Air Meter Up to 10 % of air in 2.6 % of air in ASTM C231


(Pressure Method) concrete concrete Balance

Indirect Verification
with a Force
Verification of Brinell
Hardness Machines –
HBW 10/3000/15 212 HBW 2.4 HBW Indirect verification
method per ASTM E 10
Hardness Blocks

Indirect Verification of HRBW Indirect verification per


Rockwell Superficial Low 1.6 HRBW ASTM E 18
Hardness Machines3 – Medium 1.3 HRBW Hardness Blocks
High 1.3 HRBW
HRC
Low 1.2 HRC
Medium 1.2 HRC

Indirect Verification of 274 HV 3.5 HV Indirect verification per


Microindentation ASTM E384 & ASTM
Hardness Machines3 – E92
Vickers Hardness Blocks

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 14 of 18


Parameter/Equipment Range CMC2,8 () Comments

Impact Testing Machines – ASTM E 23


Metals3
Level
Distance Between Anvils ≈ 40.00 mm 0.12 mm
Caliper
Striker Centered on Anvils ≈ 0.406 mm 0.16 mm
Striking Bit Parallel to Anvils Up to 0.127 mm/mm 0.17 mm
Machine Level Up to 0.076 mm/mm 0.11 mm/mm
Free Swing Zero Check Up to 10 lbfꞏft 1.1 lbfꞏft

Schmidt Hammer 80 HRC 2.8 HRC ASTM C805


Hardness Anvil

Scales & Balances³ 20 mg to 50 g 2.60 mg OIML Class F2


200 mg to 5 kg 12 mg OIML Class E2
>10 g to 10 kg 21 mg OIML Class E2
>10 kg to 20 kg 1.2 g OIML Class M1
>20 kg to 25 kg 2.3 g OIML Class M1

Asphalt & Concrete Batch Up to 600 kg 29 g Scales


Plants³

Pressure – Measuring Equipment

Hydraulic Gage Pressure


(10 to 5000) psi 2.9 psi DH Budenberg
(5000 to 8 000) psi 4.5 psi model 380D deadweight
tester, high pressure
(7 500 to 30 000) psi 590 psi comparator, and gauge

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 15 of 18


Parameter/Equipment Range CMC2, 8 () Comments

Mass – Measure OIML R111-1


Standard Mass Set, Mass Comparators

Fixed Points 20 kg 90 mg OIML Class F2 and Less


10 kg 89 mg OIML Class M1 and Less
5 kg 88 mg OIML Class M2 and Less
2 kg 0.11 g OIML Class M3
1 kg 0.16 g OIML Class M3
200 g 1.0 mg OIML Class M1 and Less
100 g 1.7 mg OIML Class M2 and Less
50 g 1.2 mg OIML Class M2 and Less
20 g 0.39 mg OIML Class M1 and Less
10 g 0.19 mg OIML Class F2 and Less
5g 0.39 mg OIML Class M1 and Less
2g 0.20 mg OIML Class M1 and Less
1g 0.15 mg OIML Class M1 and Less

VII. Thermodynamics

Parameter/Equipment Range CMC2, 8 () Comments

Temperature – Measuring (-30 to 140) °C 0.14 °C Dry block calibrator


Equipment (140 to 200) °C 0.07 °C (Fast Cal Low & Druck
DBC 650)

(200 to 500) °C 0.07 °C Dry block calibrator


(500 to 600) °C 0.08 °C (Druck DBC 650)

(650 to 950) ⁰C 1.17 °C Portable furnace


(950 to 1100) ⁰C 1.45 °C

Curing Room/Cabinet –

Temperature (10 to 25) °C 0.4 °C ASTM C511


Humidity (10 to 75) % RH 3.6 % RH Hygrometer

Curing Tank –
ASTM C511
Temperature (10 to 25) °C 0.13 °C Thermometer

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 16 of 18


Parameter/Equipment Range CMC2, 8 () Comments

Temperature – Oven (20 to 270) °C 0.46 °C ASTM D 2216 and


ASTM E 145 Section 4
Thermometer

Temperature &
Humidity – Measuring
Equipment

Temperature (5 to 45) ºC 0.72 ⁰C Temperature &


humidity calibration
Humidity (20 to 90) % RH 2.0 % RH INDEECON calibrator

VIII. Frequency

Parameter/Equipment Range CMC2, 8 () Comments

Rotation Frequency –

Contact – Measure Up to 1000 rpm 2.0 rpm Tachometer


(1000 to 15 000) rpm 4.1 rpm

Photo – Measure Up to 1000 rpm 3.0 rpm


(1000 to 90 000) rpm 6.1 rpm

Digital Tachometers Up to 100 000 rpm 2.0 rpm Fluke 5522A with
LED

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 17 of 18


IX. Welding Machine Calibration

Parameter/Equipment Range CMC2 () Comments

Welding Machine Calibration

DC Voltage – Measure (10 to 80) V 97 mV Welding machine


calibrator SMP 1337,
AC Voltage – Measure (10 to 80) V 95 mV Current Clamp
60 Hz

DC Current – Measure (10 to 600) A 0.98 A

AC Current – Measure (10 to 600) A 1.0 A


60 Hz

___________________________________
1
This laboratory offers commercial calibration service and field calibration service.
2
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a
laboratory can achieve within its scope of accreditation when performing more or less routine calibrations
of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and
Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of
confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific
calibration performed by the laboratory may be greater than the CMC due to the behavior of the
customer’s device and to influences from the circumstances of the specific calibration.
3
Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4
In the statement of CMC, the value is defined as the percentage of reading, unless otherwise noted.
5
CMCs are expressed as either a specific value that covers the full range or as a combination of a
percent/fraction of the reading/output plus a range specification.
6
In the statement of CMC, L is the numerical value of the nominal length of the device measured in
inches.
7
This scope meets A2LA’s P112 Flexible Scope Policy.
8
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges
indicated for the listed measurement parameter.

(A2LA Cert. No. 3512.06) Revised 03/13/2023 Page 18 of 18


Accredited Laboratory
A2LA has accredited

FUGRO SUHAIMI LIMITED


Dammam, SAUDI ARABIA
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017
General requirements for the competence of testing and calibration laboratories. This laboratory also meets R205 – Specific
Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical competence for a
defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 13th day of February 2023.

_______________________
Mr. Trace McInturff, Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 3512.06
Valid to November 30, 2024

For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.