Glass Test
Glass Test
Types of Glass
Properties
Tests
Specifications
- Glass containers must comply with identity and surface hydrolytic resistance
specifications to be classified as Type I, II, or III glass
- Type I or Type II containers for aqueous parenteral products must be tested for
extractable arsenic
- Glass containers must meet specifications for hydrolytic resistance, thermal
shock resistance, and chemical durability
Reference Materials
Test Selection
1. Autoclave
- Temperature control: 121 ± 1°C
- Equipment:
- Thermometer or calibrated thermocouple device
- Recorder
- Pressure gauge
- Vent cock
- Tray (sufficient capacity to hold containers above water level)
- Cleaning: Clean the autoclave and other apparatus thoroughly with Purified
Water before use
2. Mortar and Pestle
- Material: Hardened steel
- Specifications: Figure 1
4. Sieves
- US Sieve Nos. 25, 40, and 50
- Square-mesh stainless steel
5. Sieve Shaker
- Mechanical sieve-shaker or sieving machine
6. Hammer
- Tempered, magnetic steel
7. Magnet
- Permanent magnet
8. Weighing Bottles
- For weighing samples
9. Stoppers
- For weighing bottles
10. Metal Foil
- Aluminum or stainless steel
11. Hot Air Oven
- Temperature control: 140 ± 5°C
12. Balance
- Accuracy: 0.005 g
- Capacity: 500 g
13. Desiccator
- For drying samples
14. Ultrasonic Bath
- For cleaning and degassing samples
Reagents
- Preparation:
- Boil Purified Water vigorously for 5 minutes or more
- Allow to cool while protected from carbon dioxide absorption
- Alternative:
- Use Purified Water with a resistivity of at least 18 Mohm-cm
- Preparation:
- Dissolve 50 mg of methyl red in:
- 1.86 mL of 0.1 M sodium hydroxide
- 50 mL of ethanol (96%)
- Dilute with Purified Water to 100 mL
- Sensitivity Test:
- Mix:
- 100 mL of carbon dioxide-free water
- 0.05 mL of 0.02 M hydrochloric acid
- 0.1 mL of methyl red solution
- Result: Solution should be red
- Add NMT 0.1 mL of 0.02 M sodium hydroxide to change color to yellow
- Color change indicates pH change from 4.4 (red) to 6.0 (yellow)
.........
Glass Grains Test
Sample Preparation
Procedure
Equipment
Sample Requirements
Test Conditions
1. Temperature: 140°C
2. Time: 20 min
3. Acetone volume: 30 mL
4. Ultrasonic cleaning time: 1 min
5. Cooling: Desiccator
Method
1. Weigh 10.00 g of cleaned and dried grains into two separate conical flasks (test
solutions) and one conical flask (blank).
2. Pipet 50 mL of carbon dioxide-free Purified Water into each of the conical
flasks (test solutions and blank).
3. Distribute the grains evenly over the flat bases of the flasks by shaking
gently.
4. Close the flasks with neutral glass dishes or aluminum foil rinsed with Purified
Water or with inverted beakers so that the inner surfaces of the beakers fit snugly
down onto the top rims of the flasks.
5. Place all three flasks in the autoclave containing water at ambient temperature,
ensuring they are held above the level of the water in the vessel.
6. Insert the end of a calibrated thermometric device in a filled container through
a hole of approximately the diameter of the thermocouple and connect it to an
external measuring device. If the container is too small to insert a thermocouple,
apply a thermocouple in a suitable, similar container. Alternatively, use the
internal thermometer of the autoclave.
7. Close the autoclave door or lid securely but leave the vent-cock open.
8. Start automatic recording of the temperature versus time, and heat the autoclave
at a regular rate such that:
- Steam issues vigorously from the vent-cock after 20–30 min.
- Maintain a vigorous evolution of steam for a further 10 min.
- For autoclaves using a steam generator, it is not necessary to maintain the
temperature for 10 min at 100°.
9. Close the vent-cock, and raise the temperature from 100° to 121° at a rate of
1°/min within 20–22 min.
10. Maintain the temperature at 121 ± 1° for 30 ± 1 min from the time when the
holding temperature is reached.
11. Cool down to 100° at a rate of 0.5°/min, venting to prevent formation of a
vacuum, within 40–44 min.
12. Do not open the autoclave until it has cooled to 95°.
Titration
Note: Where necessary to obtain a sharp endpoint, decant the clear solution into a
separate 250-mL flask. Rinse the grains by swirling with three 15-mL portions of
carbon dioxide-free water, and add the washings to the main solution. Add 0.05 mL
of the Methyl red solution. Titrate, and calculate as before. In this case, also
add 45 mL of carbon dioxide-free Purified Water and 0.05 mL of Methyl red solution
to the blank solution.
Method:
1. Select six dry vials or bottles from the sample lot, or three if their capacity
exceeds 100 mL.
2. Remove any dirt or debris.
3. Weigh the empty containers with an accuracy of 0.1 g.
4. Place the containers on a horizontal surface.
5. Fill them with Purified Water to about the rim edge, avoiding overflow and the
introduction of air bubbles.
6. Adjust the liquid levels to the brimful line.
7. Weigh the filled containers to obtain the mass of the water expressed to:
- Two decimal places for containers having a nominal volume less than or equal
to 30 mL.
- One decimal place for containers having a nominal volume greater than 30 mL.
8. Calculate the mean value of the brimful capacity in mL.
9. Multiply it by 0.9 to get the filling volume for the particular container lot,
expressed to one decimal place.
Ampuls:
Titration
1. Carry out the titration within 1 h of the removal of the containers from the
autoclave.
2. Combine the liquids obtained from the containers, and mix.
3. Introduce the prescribed volume (see Table 5) into a conical flask.
4. Transfer the same volume of carbon dioxide-free water, to be used as a blank,
into a second similar flask.
5. Add to each flask 0.05 mL of Methyl red solution for each 25 mL of liquid.
6. Titrate the blank with 0.01 M hydrochloric acid.
7. Titrate the test solution with the same acid until the color of the resulting
solution is the same as that obtained for the blank.
8. Subtract the value found for the blank titration from that found for the test
solution, and express the results in mL of 0.01 M hydrochloric acid per 100 mL of
test solution.
9. Express titration values of less than 1.0 mL to two decimal places; express
titration values of greater than or equal to 1.0 mL to one decimal place.
Method:
1. Compare the results obtained from the Surface Etching Test to those obtained
from the Surface Glass Test.
2. For Type I glass containers, the values obtained are close to those found in the
Surface Glass Test.
3. For Type II glass containers, the values obtained greatly exceed those found in
the Surface Glass Test; and they are similar to, but not greater than, those
obtained for Type III glass containers of the same filling volume.
CAUTION: Hydrofluoric acid (HF) is extremely aggressive and can cause life-
threatening injuries. Even small quantities can cause severe burns and permanent
damage. Handle with extreme caution and wear appropriate protective equipment.