Paperpdnoise
Paperpdnoise
5 pC pd signal
ppd
k
0
0 20 40 60 80 100
k
PAG_Info2.doc/12.06.2005/PO/KF page 1 of 3
Presco AG
No. 2
June 2005
3. Calibrating in case of external noise or interference.
Supposing now, there is an external interference of 4 pC (could be random or a high frequency
sine or other). Without the calibration signal an oscilloscope would show :
4 pC noise signal
yn 0
k
4
0 20 40 60 80 100
k
If now a 5 pC calibration signal is applied, the PD detector will "see" the calibration impulse
superimposed on the noise signal :
Overlayed noise and signal:
10
6
sm
k
markk
4
0
0 20 40 60 80 100
k
As one can immediately see, the calibration impulse "dances" on the interference signal and the
instrument "sees" a higher calibration impulse than effectively injected (8.5 pC in our case). If
now the calibration is made, the sensitivity of the instrument is reduced because it is set to a
value "calibration magnitude + interference level". If the system is then used after such a wrong
calibration, the PD measurement will be wrong and it also shows a back ground noise level
lower that it is in reality. If after such a wrong calibration the back ground noise level is
measured, the PD detector will show
5
q gdisp
4 q gdisp 2.353
8.5
These 2.35 pC are wrong, it is a "embelished" back ground noise level leading to wrong PD
measurements.
PAG_Info2.doc page 2 of 3
Presco AG
No. 2
June 2005
4. Recommendations
Use a "honest" PD detector with a correct filter characteristic according the specific
application and without "dead zone" and good linearity.
Improve your PD system regarding grounding (earthing), screeneing, filtering. Our
experienced engineers can help.
Always calibrate your PD system with a high calibration impulse (e.g. 100 pC or even
1'000 pC). The higher the injected pulse for calibration is, the lower the influence of noise
on the calibration is.
With the good linearity of our instruments, after a calibration with a high value you can
easily check the performance of the instrument with the calibrator. Then check the real
background noise level the following way :
Calibrate with 100 pC (normally with gain 10)
Switch off the calibrator
Increase gain to 1000
Read back ground noise level.
5. Presco's PD Detectors
All our PD Detectors are supplied with Coupling Device, Battery Calibrator, Cable between
Coupling Device and PD Detector plus a Test Certificate granting traceability to international
standards
diagnosis of high voltage insulation systems
diagnosis of high voltage insulation systems
fulfils standard IEC 60270
fulfils standard IEC 60270
real analogue processing
selectable broad band filters
direct observation of PD activities by analogue
tuneable narrow band filter
indicator
combined analogue and digital display
good linearity and reproducibility of measurements
interference gating with up to 6 individually
Auto calibration adjustable gates
Auto ranging
Real analogue or digital processing
built-in RS232 interface
Auto calibration
one cable connection between coupling device and
Auto ranging
instrument
Phase resolved pattern in scope mode or fingerprint
including Battery Calibrator and Coupling Device mode
Options : different kinds of multiplexers, remote
one cable connection between coupling device and
control software instrument
low cost
including Battery Calibrator and Coupling Device
Options : different kinds of multiplexers and remote
control software, digital signal processing software
PAG_Info2.doc page 3 of 3