ABL 80 Specification
ABL 80 Specification
Specifications
Measured parameters Derived parameters
Type Parameter Units Measuring range cHCO3- (P)
cBase(B)
pH pH 6.00 – 8.00
cBase(B,ox)
Blood Gas pCO2 mmHg 0.0 – 150.0
cBase(Ecf)
kPa 0.00 – 20.00
cBase(Ecf,ox)
pO2 mmHg 0 – 760
cHCO3-(P,st)
kPa 0.0 – 101.3
ctCO2(P)
Electrolytes cCa2+ mmol/L 0.00 – 5.00
ctCO2(B)
mEq/L 0.00 – 10.00 cCa2+(7.40)
mg/dL 0.00 – 20.00 Anion Gap (K+)
cCl- mmol/L 0 – 250 Anion Gap
mEq/L 0 – 250 ctO2
cK+ mmol/L 0.00 – 20.00 sO2
mEq/L 0.00 – 20.00 ctHb
cNa+ mmol/L 0 – 210 pO2(A)
mEq/L 0 – 210 pO2(a/A)
Lactate cLac mmol/L 0.0 – 40.0 pO2(A–a)
mg/dL 0 – 360 RI
Hematocrit Hct % 0 – 85
The Measuring range for a parameter is the range within which the analyzer is physically capable of measuring. The measuring
range corresponds to the “range of indication” as defined in the “International vocabulary of basic and general terms in the
metrology” (VIM).
Sensor cassette
Sample volume ~ 70 µL
Measuring time ~ 100 – 115 sec (~ 125 sec with lactate)
Cycle time ~ 100 – 115 sec (~ 125 sec with lactate)
Startup time ~ 5 min (~ 12 - 15 min with lactate)
Shelf life 120 days (90 days with lactate)
Storage temperature 5 – 25 °C / 41 – 77 °F (2 – 8 °C / 36 – 46 °F with lactate)
Versions
Tests per day 0.4 0.8 1.7 3.3 5 6.7 10 20 40
Sensor cassettes without lactate
Patient tests 25 50 100 200 300 300 300 600
In-use lifetime (days) 60 60 60 60 60 30 15 15
Sensor cassettes with lactate
Patient tests 25 50 100 200 300 300 600
In-use lifetime (days) 30 30 30 30 30 15 15
Solution pack
Solution 1 Solution 2
Fluidic cycles 450 110
In-use lifetime Dependent on number of patient and QC samples and frequency of calibration.
Up to 60 days maximum.
Shelf life 150 days
Storage temperature 2– 25 °C / 36 – 77 °F
Calibration data
Details Default interval Interval options Duration
Automatic : 1-point cal With measurement - -
Automatic : 2-point cal 8 hours Every 2, 4 or 8 hours 2 min.
or manual
Hardware
Computer specifications Interfaces
Microsoft Windows®XP Embedded operating system 8.4” color TFT-LCD, resolution 800 × 600 SVGA Touch screen
Minimum 1 GB hard drive Barcode reader
ETX single board CPU Serial line RS232
Minimum 512 MB EDO-RAM RJ45 Ethernet port
2 USB 1.1
PS2 keyboard
Software
Correlation correction Security and QA features
Standard correlation mode: Seven programmable user-access levels
For whole blood: all parameters available User ID and access-level verification
Other fluids mode: For all parameters except Hct Automatic lockout of parameter that fails QC or option to
Hemodilution mode: For the Hct parameter only inactivate individual sensors for failed calibration
QC statistics and on-board Levey-Jennings plots
Data capacity Air-in-sample detection
Mandatory input fields
Patient results : 500
Manual QC results : 500
2-point cal. results : 500 Communication
Event records : 1500
Security records : 1500 HIS/LIS communication
User IDs : 1000 High-level protocols:
ASTM (E1394-97)
Printer display options ASTM 6xx
HL7 (Version 2.2/2.5)
Autoprint (on/off) Low-level serial protocols:
Select derived parameters ASTM (E1381-95)
Custom report layout Low-level network protocols:
QC ranges with results TCP/IP
Select input variables Radiometer IT solution
Reference ranges with results Interface via Ethernet adapter
Analyzer name (user-defined)
Edit log
Additional information
Dimensions Other
Width 22 cm 9 in Operating environment 12 – 28 °C / 54 – 82 °F
Height 40 cm 16 in Altitude 2290 m/7513 feet above sea level
Depth 28 cm 11 in Power 100 – 240 VAC, 50/60 Hz, 130 VA
Weight 8.5 kg 19 lbs Thermostat control 37.0 °C ± 0.2 within 10 sec