E118 STEAM STERILIZER INSTRUCTION MANUAL

Instruction Manual
E118
E118 STEAM STERILIZER INSTRUCTION MANUAL

Thank you for choosing our steam sterilizers.


Prior to operating this instrument, please read the operations manual carefully and follow
all installation instructions.

Caution: Be sure connect the exhaust tube and discharge container before start the
sterilizer.

Instruction Manual
STE-18-E

Ningbo Ican Machines Co., Ltd.


No. 77 Yunlin East Road, Gulin Town, Ningbo, China.
www.icanclave.com

Subject to technical changes


E118 STEAM STERILIZER INSTRUCTION MANUAL

Technical Specifications

Specifications
Model Number STE-18-E

Electrical supply 220-240 VAC 110-100 VAC

Chamber (mm, diameter/depth) Ø230×360 mm

Power (VA) 1100 VAC

Sterilization temperature (°C) 121 °C 134 °C


Overall dimensions (mm) 605 (width)x445(height)x400(depth)

Net weight (kg) 30 kg

121 °C 30 min.
Sterilization time
134 °C 10 min.

Packing content

Sterilizer 1

Rack 1

Trays holder 1

Trays 2

Exhaust tube 2

Dorr adjustment wrench 1

Description
E118 STEAM STERILIZER INSTRUCTION MANUAL

Installation
• Position the device on a plane surface with minimum capacity 60 Kg.
• The sterilizer should be placed on a level worktable.
• Improper water level in the chamber could cause a sterilizer malfunction.
• Leave at least 10 cm between the device rear part and the wall. The clearance required to open
the door is 40cm.
• Position the autoclave at such a height as to make it possible for the operator to check the whole
sterilization chamber and carry out the normal cleaning operations.
• The room where the device is installed must have sufficient ventilation.
• Do not install the autoclave near washing basins, taps, etc. where it is likely to be splashed.
• Do not install the sterilizer nearby a heat source.
• Do not lean on the door when it is opened.
• Do not place trays, papers, fluid containers, etc. on the sterilizer.
• Check the label on back panel of sterilizer to verify voltage rating for the unit. Failure to connect
the autoclave to an appropriate power supply could result in damage to the unit, and electrical
shock to personnel.
• Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit
only used for the sterilizer is recommended.

Setup
Open the door and remove all the inner contents for unpacking.
Connect the power cord to an outlet of the appropriate voltage. (See installation).
Connect the draining tube to the vent and the end of the draining tube is fixed into the discharge
container (preferably a metal container; if it is a plastic container, make sure that the container is filled
with a part of cold water to prevent high temperature deformation).
To power on, press the main switch, the machine turns on the LED.
Note: If the door is open, the display will show “do”; When the door is closed, the display will show “Ld”.

Preparation of sterilization materials


For the most effective sterilization and to preserve the sample, please follow below:
• Clean instruments immediately after use.
• Treat the instruments by ultrasonic cleaner.
• Residual chemicals left over after cleaning and disinfecting process may damage and corrode
parts of the autoclave, always rinse off the instruments using distilled water.
• Follow instrument manufacturer’s guidelines and recommendations for handling and cleaning
instruments prior to sterilization.
• Check the manufacturer’ s instructions as to the proper procedure for sterilizing of each item.
• Arrange the samples of different materials on different trays or with at least 3cm of space
between them.
• Clean and dry instruments thoroughly before placing them into tray.

Operation
Close the exhaust valve by turning it clockwise to the maximum position.
Fill with the chamber with distilled water (or deionized water).

Select the program temperature.


After selecting program, the materials to be sterilized can now be placed on the tray placed inside the
E118 STEAM STERILIZER INSTRUCTION MANUAL

chamber using the tray handle.


After the instruments are loaded, you may close the chamber door. The display will show the current
internal temperature.
Caution: You must turn the door handle to the maximum position.

Start the sterilization program.


After the Start button is pressed, the sterilizers will begin the cycle automatically.
Note:
The draining tube must be always connected to the vent and the end of the draining tube must be fixed
into the discharge container.

End of cycle
After the sterilization is complete, you will hear a beep and the screen will show “End”.
Then you can open the door to remove the sterilized items.
Note:
After the screen shows End, if the gauge is pointer different than 0 bar, do not open the door until it
down to zero-scale position.
Note:
After each cycle check the water level of the chamber.

Max. Load
Program Type Max. Load (kg)
per tray (kg)

121 °C / 134 °C Unwrapped solid material 2.4 1.2

Error codes

Code Description Proposed solution


E1 Heater overheating Check the water level in the chamber.
Check the Thermostat
E2 Chamber overheating Check temperature sensor connection.

E4 Heating overtime Somewhere is leaking. Check the door seal.


Heater malfunction.

During operation, the door has Confirm that the door handle has been closed to the maximum
E6 been opened or the door micro position.
switch is damaged. Check or replace the door micro switch

Maintenance

Frequency Maintenance Operation

Monthly Clean the door seal.

Clean the chamber, the trays and


Weekly the rack.

Every Year Replace the door seal.


E118 STEAM STERILIZER INSTRUCTION MANUAL

Clean Chamber, Trays, Door Seal, chamber filter and Tray Rack.
Remove the trays and tray rack from the chamber. Clean trays, rack and inside of chamber with mild
soap.
Rinse the trays, rack and inside of chamber with a smooth cloth and distilled water.
Examine gaskets for possible damage.
Clean gasket and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials / substances in chamber. Failure to comply
may result in damage to the chamber and/or other components.
Caution: To prevent burns, allow unit to cool before cleaning gaskets and mating surfaces.

Door adjustment
Under normal circumstances the chamber door does not require
adjustments. However, if the seal
fails (resulting in steam leaking from the front of the chamber), you may use
the spanner tool to
tighten the door seal.
Open the door and Insert the spanner tool in the gap beneath the plastic
cover; use the spanner to grip the adjusting nut.
Turn the nut counter clockwise. This will tighten the sealing plate.
Turn the nut until the sealing plate is tight. If the door knob is too tight, you
may also turn the nut clockwise to loosen it.

Replacement of the door seal

Open the chamber door. Remove the door


seal ring carefully by hand.
Clean the door seal ring carefully with a
smooth cloth with distilled water.
Moisten the new seal with medical
disinfectant or isopropyl alcohol.
Insert the new seal and press in sequence
as follows:

Caution: Please ensure the chamber and the


door have cooled prior to replacing the seal ring.