Alphasense Pid-Ah2 Datasheet en 1
Alphasense Pid-Ah2 Datasheet en 1
Notes:
1. Do not obstruct Ø3.5 sensing area 5. Input voltage selector hole:
2. Seal betweenTop Ø6.2View Bottom View
and Ø9.0 (if different to atmosphere) a) When filled with solder the onboard Side View
regulator is disabled.
Notes: 3. Pin out details: A regulated supply of 3.2 - 3.6 V (typically 3.2 V) is then required.
Pin 1: + V supply (See note 5) b) When not filled with solder the onboard regulator is enabled.
1. Do not obstruct Ø3.5 sensing area 5. Input voltage selector hole:
Pin 2: Signal output A regulated or unregulated supply between 3.6 - 10 V is then required
2. Seal between PinØ6.2 and
3: 0V Ø9.0 (if different to atmosphere)
supply a)for
When filled with solder
IS applications, or uptheto onboard regulator
18 V for non-IS is disabled.which
applications A will be
3. Pin out details:
4. All dimensions ±0.1mm unless otherwise stated regulated supply of 3.2
internally regulated to 3.3V. - 3.6 V (typically 3.2 V) is then required.
- Pin 1: + V supply (See note 5) Nomally
b) Whenshipped
not filled with regulator
with solder enabled.regulator is enabled. A
the onboard
- Pin 2: Signal output regulated or unregulated supply between 3.6 - 10 V is then
PERFORMANCE
- Pin 3: 0 V supply (using 10.6 eV lamp 001-0019-04) required for IS applications, or up to 18 V for non-IS applications
Target gases
4. All dimensions ±0.1mm unless otherwise statedVOCs with ionisation potentials < 10.6
which willeV
be internally regulated to 3.3V.
Minimum detection level ppb isobutylene Normally shipped with regulator enabled. 1
Linear range ppm isobutylene 3% deviation 40
Overrange ppm isobutylene 40
Performance
Sensitivity Target gases
linear range VOCs with ionisation
mVpotentials < 10.6 eV
/ ppm Isobutylene > 25
(using 10.6 eV lamp 001-0019-04) Minimum detection level ppb isobutylene 1 5
Full stabilisation time minutes to 20 ppb
Warm up time Linear range
seconds ppm isobutylene 3% deviation
time to full operation 40 5
Offset voltage OverrangemV variable between
ppmdetectors
isobutylene 46 40
to 60
Response time (t90) Sensitivityseconds linear range mV /diffusion mode
ppm Isobutylene > 25< 3
Full stabilisation time minutes to 20 ppb 5
ELECTRICAL Warm-up time seconds time to full operation 5
Power consumption Offset voltage
Onboard regulatormV variable
enabled between detectors
(default): < 100 mW at 3.6 V, < 550 mW transient for46200ms
to 60
(at switch on) Response timeOnboard
(t90) regulatorseconds
disabled:
diffusion mode< 85 mW at 3.2 V, < 300 mW transient for 200ms<3
Supply voltage 3.2 to 3.6 VDC Ideally regulated ±0.01V (onboard regulator disabled)
Electrical 3.2 to 10 VDC (onboard
Power consumption Onboardregulator enabled)
regulator enabled (default): < 100 mW at 3.6 V, < 550 mW transient for 200ms
(maximum 10V forOnboard
(at switch on) IS approval, maximum
regulator disabled:18 V mW
< 85 for non-IS)
at 3.2 V, < 300 mW transient for 200ms
Output signal Offset voltage (minimum
Supply voltage 3.2 to 3.646
VDCmV) to Vmax
Ideally regulated ±0.01V (onboard regulator disabled)
(Vmax = Vsupply -0.15
3.6 toV10when regulator
VDC (onboard isdisabled,
regulator enabled)or 3.15 V when regulator is enabled)
(maximum 10V for IS approval, maximum 18 V for non-IS)
ENVIRONMENTAL Output signal Offset voltage (minimum 46 mV) to Vmax
Temperature range -40ºC to +55ºC (Intrinsically Safe);
(Vmax = Vsupply -40ºC
-0.15 V whentoregulator
+65ºC is(non-IS)
disabled, or 3.15 V when regulator is enabled)
Temperature dependence 0°C to 40°C 90% to 100% of signal at 20°C
Environmental -20°C
Temperature range 140% ofSafe);
-40°C to +55°C (Intrinsically signal at 20°C
-40°C to +65°C (non-IS)
Relative humidity range Non-condensing
Temperature dependence 0°C to 40°C of signal at 20°C 0 to 95%
Humidity sensitivity During operations: 0% to 75% rh transient
-20°C 140% of signal at 20°C
near zero
Relative humidity range Non-condensing 0 to 95%
KEY SPECIFICATIONS
Humidity sensitivity During operations: 0% to 75% rh transient near zero
Operating life 5 years (excluding replaceable lamp and electrode stack)
IS Approval
Key specification OperatingIECEx
life Ex ia IIC T4;5ATEX Ex ia II 1G
years (excluding -40ºC < Ta
replaceable < +55ºC
lamp (< 10VDC
and electrode stack)supply)
Onboard filter To
IS Approval remove liquids and particulates
IECEx Ex ia IIC T4; ATEX Ex ia II 1G -40 C < Ta < +55 C (< 10VDC supply)
o o
Alphasense Ltd, Sensor Technology House, 300 Avenue West, Skyline 120, Great Notley. CM77 7AA. UK
T: +44 (0)1376 556700 Telephone: +44 (0) 1376 556
E: [email protected] W:700 Fax: +44 (0) 1376 335 899For further
www.alphasense.com E-mail: [email protected]
information Website:
on the performance of this www.alphasense.com
sensor, on other sensors in the range or any other
subject, please contact Alphasense Ltd. or visit our website at “www.alphasense.com”.
PID-AH2 Perf
Performance Da
erformance ta
Data
Technical specifications Version 1.0
PID-AH2 Perf
Performance Da
erformance ta
Data
Figure 2 Linearity to Isobutylene at 3.6 V
0.6
Figure 2 shows reduced
0.5 sensitivity at higher
concentrations is a chemical/
tion
Figure
0.4 2 Linearity to Isobutylene at 3.6 V
physical effect and can be
0.6
0.3 Figure 1 shows reduced sensitivity at higher
Figure
corrected 2 inshows
softwarereduced
for a
(ppm) error (ppm)
0.5
0.2 concentrations
specific VOC. at is a chemical/physical
sensitivity higher effect and can
be correctedisinasoftware
concentrations chemical/for a specific VOC.
tion
0.4
0.1
Non-linearity
physical effect andcorrection
can be
Non-linearity correction
depends depends on the VOC
0.3
correctedoninthe VOC being
software for a
Linearity errorLinearity
0.0
0.2 0 5 10 15 20 25 30 35 40 45 50 being
measured. measured.
specific VOC.
-0.1
ica
0.1
-0.2 Non-linearity correction
0.0
-0.3 depends on the VOC being
0 5 10 15 20 25 30 35 40 45 50 measured.
-0.1
-0.4
ica
-0.2
-0.5
-0.3
Isobutylene (ppm)
Specif
-0.4
-0.5
Figure 3 Selecting the right lamp
600
Isobutylene (ppm)
Figure 3 compares the output
Specif
300 10.6 eV
lamps.
400
200
9.6 eV
200
0
1800 2000 2200 2400 2600
100 Time (s)
hnical
9.6 eV
Table
0 1: PID Replaceable Parts/Consumables List
1800 2000 2200 2400 2600
Minimum sensitivity Time
Minimum Custom er inform ation:
Lamp type Product code (s)range Lamp life
mV/ppm ppm isobutylene lit hours Part No Regulator Lam p Usage voltage Certified
9.6 eV 001-0030-00 0.25 8,000 TBD PID-AH2 Disabled HPPM 10.6 eV 3.2 to 3.6 Yes
Table 1: PID Replaceable Parts/Consumables List
PID-AH2 Enabled HPPM 10.6 eV 3.6 to 10 (10.1 to18) Yes (NO)
10.0 eV 001-0030-02 10 100 5,000
Minimum sensitivity Minimum range Lamp life PID-AH20 Disabled Custom
LLHS 10.6ereV
inform ation:
3.2 to 3.6 Yes
10.6
LampeVtype
(HPPM) 001-0019-04
Product code 25 40 5,000
mV/ppm ppm isobutylene lit hours PID-AH20
Part No Enabled
Regulator LLHSLam
10.6peV to 10 (10.1
3.6Usage to18)
voltage Yes (NO)
Certified
10.6 eV (LLHS) 001-0030-01 25 40 5,000
Tec
Electrode
Table 1 PID001-0018-01
Replaceable Parts/Consumables List PID-AH29
PID-AH2 Enabled
Customer
Enabled 9.6 10.6
HPPM
information
eV eV 3.6 10(10.1
3.6toto10 (10.1to18)
to18) Yes (NO)
Yes (NO)
10.0 eV stack 001-0030-02 10 100 5,000
PID-AH2X
PID-AH20 Disabled
Disabled 10.010.6
LLHS eV eV 3.2
3.2toto3.6
3.6 Yes
Yes
Stack removal
10.6 eV tool
(HPPM) 001-0020-00
001-0019-04 25 40 5,000
Minimum Part No eV eV Regulator to18) Lamp Usage voltage Certified
Minimum PID-AH2X
PID-AH20 Enabled
Enabled 10.0
LLHS 10.6 3.6 10(10.1
3.6toto10 (10.1to18) Yes (NO)
Yes (NO)
Lamp spring
10.6 eV (LLHS) Product
001-0023-00
001-0030-01 25 40 range
5,000 Lamp life
Tec
For further information on the performance of this sensor, on other sensors in the range or any PID-AH2X
other subject, please contact Alphasense
Enabled 10.0 eVLtd. 3.6 to 10 (10.1 to 18) Yes (NO)
Stack removal tool 001-0020-00
For Application Notes visit “www.alphasense.com”.
Lamp spring
In the 001-0023-00
interest of continued product improvement, we reserve the right to change design features and specifications without prior notification. The data contained in this
document is for guidance only. Alphasense Ltd accepts no liability for any consequential losses, injury or damage resulting from the use of this document or the information
Lamp cleaning kit 001-0024-00
contained within. ( ©ALPHASENSE LTD ) Doc. Ref. PID-AH2/AUG19
At the end of the product’s life, do not dispose of any electronic sensor, component or instrument in the domestic waste, but contact the instrument manufacturer, Alphasense or its distributor for disposal
instructions. NOTE: All sensors are tested at ambient environmental conditions, unless otherwise stated. As applications of use are outside our control, the information provided is given without
legal responsibility. Customers should test under their own conditions, to ensure that the sensors are suitable for their own requirements.
In the interest of continued product improvement, we reserve the right to change design features and specifications without prior notification. The data contained in this document is for guidance only.
Alphasense Ltd accepts no liability for any consequential losses, injury or damage resulting from the use of this document or the information contained within.(©ALPHASENSE LTD) Doc. Ref. PID-AH2/SEP22
T: +44 (0)1376 556700 E: [email protected] W: www.alphasense.com For further information on the performance of this sensor, on other sensors in the range or any other
subject, please contact Alphasense Ltd. or visit our website at “www.alphasense.com”.