Balance Linearity
Balance Linearity
If you use a laboratory balance in your day-to-day work, you may have
wondered whether it is accurate. Here are four tests that you can use to confirm
your instrument's accuracy. In this article you will learn to make these tests
yourself.
Cornerload errors are those errors associated with different positions on the
weighing pan of the object being weighed. A given object should produce the
same reading regardless of its position on the weighing pan.
Span refers to the difference between the weight reading of a given mass
standard, and the actual value of that standard. This measurement is often done
at or near full capacity. This is the test (or adjustment) that most people thing of
when the refer to calibration of their instrument. For many users, it is the only
test or adjustment they will choose to make routinely.
The following pages contain specific instructions for performing these tests.
IES Corporation
Repair of Laboratory Balances
(800) 541-0852 www labbalancerepair.com
Preliminaries
Handling test weights
The weights used to test lab balances are precision devices and need to be handled
accordingly. When handling weights, avoid direct hand contact with weights by using clean
gloves. Avoid sliding weights across any surface, especially across the stainless steel
weighing pan of the balance under test.
Environment
In order to pass any test of reproducibility an instrument must be operating in an acceptable
environment. A poor environment will degrade the results of a standard deviation (SD) test
and falsely suggest that the performance is substandard. There are several aspects of the
environment which impact the performance of a lab balance.
Temperature
The accuracy and overall performance of any lab balance is affected by the room temperature.
For best stability and performance the room temperature should be regulated to within one
degree Fahrenheit without interruption. The instrument should remain with power ON
continuously.
Air Drafts
In the cases of measurements with resolution of .001 gram and less, the force exerted by
moving air is readily detectable. A shroud or enclosure around the weighing pan will shield
the pan from these effects. Avoid plastic materials for draft shields.
Static Electricity
Static electricity exerts a mechanical force which is readily detectable by analytical and
microbalances. An example of static electricity exerting a mechanical force would be lint
sticking to clothing. Static will be a problem when it exists on the object being weighed, on
the person using the balance, on draft shields, or on weighing vessels. Sources of static are
carpets, Vibrum shoe soles, plastic draft shields, plastic weighing vessels, and melamine
(Formica) table tops. Low ambient humidity exacerbates static problems.
You can test for a static problem easily. On an analytical balance place a metal enclosure (a
coffee can works well) over the weighing pan, so that the pan is enclosed by the can but NOT
touched by it. If the weight readings stabilize with the can in place, then static may be the
cause of the instability. Notice that the coffee can provides an effective draft shield too.
IES Corporation
Repair of Laboratory Balances
(800) 541-0852 www labbalancerepair.com
Reproducibility Testing
Reproducibility testing entails repeatedly weighing a given object, recording the results, and
analyzing those results. Select a test weight equal to, or nearly equal to, the weighing capacity
of the instrument. Utilize the chart titled "Reproducibility Test Chart."
IES Corporation
Repair of Laboratory Balances
(800) 541-0852 www labbalancerepair.com
Cornerload
Cornerload testing verifies that the instrument delivers the same weight reading, regardless of
where on the weighing pan the object being weighed is placed. Cornerload performance
specifications are often not advertised. Typical tolerances are shown below.
Cornerload Chart
Front Right Rear Left
IES Corporation
Repair of Laboratory Balances
(800) 541-0852 www labbalancerepair.com
Linearity Testing
Linearity Chart
0%– 50% 50% - 100%
Special Note:
A common error in linearity (accuracy) testing is to simply place test weights
on the weighing pan and observe the difference between the indicated weight
and the nominal value of the test weight. This process fails to account for the
fact that test weights are imperfect and that the difference between the nominal
value and the actual weight might be significant. This is especially true with
analytical balances, where the balance may be more accurate than any standard
test weight. The above procedure nullifies this problem by comparing the
weight readings of the same object, both with and without a preload. The
accuracy of the test weight is thus immaterial.
IES Corporation
Repair of Laboratory Balances
(800) 541-0852 www labbalancerepair.com
Span
Span refers to the adjustment of the sensitivity of the instrument across the full weighing
range. Span differs from the previous performance parameters in that it is readily adjustable,
whereas cornerload, linearity, and repeatability generally are not. Span adjustment of
instruments is different from instrument to instrument. Generally, the adjustment procedure is
described in the user's manual which comes with the instrument. Follow the instructions in
your user's guide.
Many (not all) instruments now include internal calibration weights, so calibration is as easy
as pushing a single button. The user may reasonably ask, "How do I know the internal
calibration weight is correct ?" The answer is that the only way to know is to have an external
standard for comparison. The advantage of an external standard, whether it's used to adjust
span, or just to confirm the internal weight, is that it is completely external to the instrument,
and can therefore be compared to external standards. By that process, the overall calibration
of the instrument is matched to international standards and all other weighing instruments.
Standardized test weights are made to various levels of accuracy. ASTM class 1 is the most
accurate weight class commonly available. Most weights in ASTM class 1 are accurate to one
part in 400,000. Since many lab balances are considerably more accurate (resolution of
greater than one part in a million), one might reasonably wonder how a standard weight can
be used to test or adjust analytical balances. The answer would be that the standard weight can
itself be calibrated to a level of accuracy that exceeds even the most accurate lab balances.
The characterization of the weight will determine the actual value of the weight to a much
higher degree of precision than required by the ASTM class 1 standard. Such a weight can be
used to verify the accuracy of an internal calibration weight. Simple "traceability" of an
external weight is insufficient. A calibration certificate for a class 1 weight must specify the
actual weight value with deviation from nominal. A calibration service which supplies such a
certificate should be registered (via A2LA, or similar) to provide that level of accuracy. The
registration confirms via third party testing that the calibration certificate issuer is capable of
such accuracy and is audited to be delivering that accuracy reliably.
After placing a test weight on an operating instrument and finding that the displayed weight
value does not exactly match the nominal value of the test weight, many users have concluded
that the instrument is miscalibrated. However, that conclusion is by no means certain unless
the test weight has been calibrated and its correction from nominal value is known. The
calibration weights internal to high quality lab balances are more accurate than commonly
available test weights. In the absence of a calibration certificate for a specific weight, users
should presume that internal weights are more accurate than external test weights.
IES Corporation
Repair of Laboratory Balances
(800) 541-0852 www labbalancerepair.com