Manual Multi EA 5100 en
Manual Multi EA 5100 en
multi EA 5100
C/N/S/X Analyzer
Manufacturer Analytik Jena GmbH
Konrad-Zuse-Straße 1
07745 Jena / Germany
Telephone: +49 3641 77 70
Fax: +49 3641 77 9279
E-mail: [email protected]
For a proper and safe use of this product follow the instructions.
Keep the operating manual for future reference.
Edition A (01/2021)
Table of contents
1 About this user manual .................................................................................................................................................. 7
2 Intended use.................................................................................................................................................................... 8
3 Security ............................................................................................................................................................................ 10
3.1 Safety labeling on the device ........................................................................................................................... 10
3.2 Requirements for the operating personnel..................................................................................................... 11
3.3 Safety instructions, transport and commissioning ......................................................................................... 11
3.4 Safety instructions: during operation .............................................................................................................. 12
3.4.1 General safety instructions during operation.................................................................................................. 12
3.4.2 Safety instructions – protection against explosion and fire........................................................................... 12
3.4.3 Safety instructions: electrical equipment ........................................................................................................ 13
3.4.4 Handling of auxiliary and operating materials ............................................................................................... 13
3.4.5 Decontamination after soiling ......................................................................................................................... 14
3.4.6 Behavior during overpressure faults (0206, gas pressure fault) ................................................................... 14
3.5 Behavior during emergencies .......................................................................................................................... 14
3.6 Safety instructions – maintenance and repair ................................................................................................ 15
6 Operation ........................................................................................................................................................................ 36
6.1 General notes for measuring operations......................................................................................................... 36
6.2 Selecting the measurement method ............................................................................................................... 37
6.3 Switching on the basic module and the modules ........................................................................................... 38
6.4 Switching off the basic module and the modules........................................................................................... 39
6.5 Recommissioning after emergency shutdown (Cl module 5100) ................................................................. 40
3
Table of contents multi EA 5100
13 Troubleshooting.............................................................................................................................................................. 98
13.1 General notes.................................................................................................................................................... 98
13.2 Remedying software message faults............................................................................................................... 98
13.3 Initializing the basic module and the system components ............................................................................104
4
multi EA 5100 Table of contents
5
Table of contents multi EA 5100
14.11.1 Replacing the sulfuric acid and cleaning the sulfuric acid container. ............................................................147
14.11.2 Maintenance of the measuring cell .................................................................................................................149
14.11.3 Maintenance and storage of the electrodes ...................................................................................................150
14.12 Maintenance of the S module 5100 basic and S module 5100 MPO sulfur detector ................................152
14.12.1 Replacing the UV lamp .....................................................................................................................................152
14.12.2 Replacing the chemical ozone decomposer ....................................................................................................154
14.13 Maintenance of the coulometric sulfur detector ............................................................................................155
14.13.1 Replacing the absorber.....................................................................................................................................155
14.13.2 Replacing the electrolyte solution ...................................................................................................................157
14.14 Maintenance of the TOC detector ...................................................................................................................158
14.14.1 Replacing the water traps ................................................................................................................................159
14.14.2 Replacing the halogen trap..............................................................................................................................159
14.14.3 Regenerating the TIC reactor...........................................................................................................................161
14.14.4 Cleaning the TIC reactor ...................................................................................................................................161
14.14.5 Replacing the pump hose of the condensate pump .......................................................................................163
14.14.6 Cleaning the condensation coil........................................................................................................................164
14.14.7 Replacing the catalyst in the TOC combustion tube .......................................................................................165
6
multi EA 5100 About this user manual
Conventions Instructions for actions which occur in chronological order are numbered and combined
in action units.
Warnings are indicated by a warning triangle and a signal word. The type, source and
consequences of the hazard are stated together with notes on preventing the hazard.
Elements of the control and analysis program are indicated as follows:
¡ Program terms are in bold (e.g., the System menu).
¡ Buttons are indicated by square brackets (e.g., [OK]).
¡ Menu items are separated by vertical lines (e.g., System | Device).
Symbols and signal words used The user manual uses the following symbols and signal words to indicate hazards or in-
in this manual structions. The warnings are always placed before an action.
WARNING
Indicates a potentially hazardous situation which can cause death or very serious (possi-
bly permanent) injury.
CAUTION
Indicates a potentially hazardous situation which can cause slight or minor injuries.
NOTICE
Provides information on potential material or environmental damage.
7
Intended use multi EA 5100
2 Intended use
The multi EA 5100 is a modular analyzer for determining the sulfur, nitrogen, chlorine
and carbon content in solid, liquid, paste-like, viscous and gaseous samples. Sample di-
gestion is achieved via pyrolysis and subsequent thermal oxidation of the achieved result
of the pyrolysis. The various element contents are determined in accordance with na-
tional and international standards by subsequent selective detection systems.
Depending on the functionality, the basic module includes at least one detection module
and one sampling module. The basic module and any connected modules are controlled
by the multiWin control and analysis software, and the analysis of the measurement
data is performed in this as well.
The analyzer may only be used for the analyses listed in the user manual. Only this spec-
ified use is regarded as the intended use, ensuring the safety of the user and the device.
Suitable sample types for the The analyzer can be used for the following sample types:
multi EA 5100 ¡ Organic solids, e.g., wax, polymer
¡ Organic liquids, e.g., fuels, solvents
¡ High-viscosity organic liquids, gels and paste-like samples, e.g., crude oil, asphalt,
bitumen, tar
¡ Organic gases and gas mixtures, e.g., natural gas, methane
¡ Compressed liquefied gases, e.g., LPG, NPG, butane, propane, propylene
¡ Carbon modifications, e.g., coal, elemental carbon, soot, coke
¡ TC/TOC/TIC/NPOC in water analyses, e.g., waste water
¡ EC/OC determination in particulate emissions, e.g., air quality control, three-way
catalytic converter research
¡ AOX and EOX for water, sludge and soil via the column or batch method
Unsuitable samples The following sample types may not be analyzed with the multi EA 5100:
¡ Highly flammable organic compounds – risk of explosion!
¡ Self-reactive materials and explosives, e.g., peroxides, epoxides, azides
¡ Organic and inorganic phosphorous compounds, e.g., phosphoric acid ester
¡ Silico-organic and reactive or unstable inorganic silicium compounds, e.g., silane
¡ Metal organic compounds, e.g., nickel carbonyl
¡ Corrosive or highly-reactive substances, inorganic acids, HF, Cl2, H2SO4
¡ Materials with high inorganic or organic salt content, in particular Na+ und K+ ions,
e.g., soil, fertilizers, feed, construction material
¡ Inorganic compounds, e.g., minerals, inorganic halogens
¡ Samples requiring more than 1100 °C for complete oxidation or digestion, e.g., met-
als, alloys, N2
¡ Pure elements, e.g., sulfur, nitrogen
¡ Sample with element contents exceeding the permissible operating range.
¡ TOC, TIC or TC analyses of solids, e.g., soil, sediment
Alkali salts and alkaline earth salts lead to crystallization of all quartz glass components
This process is accelerated by increased salt concentrations and combustion tempera-
tures, i.e. reducing the service life of the quartz glass components.
Gases used The device must only be used with argon and oxygen of the required quality as a carrier
gas.
Sum parameters Attaching the corresponding detectors and sampling modules allows the multi EA 5100
analyzer to determine the following parameters as sum parameters.
8
multi EA 5100 Intended use
Parameter Description
TS Total sulfur
Organically bound sulfur is detected. Inorganic sulfur compounds are
only digested partially or not at all.
TN Total nitrogen
Organically bound nitrogen is detected. Inorganic nitrogen compounds
and molecular nitrogen (N2) are only digested partially or not at all.
TC Total carbon
The total organic carbon and elementary carbon and contained CO2 is
detected. Inorganic carbon compounds are only digested partially or not
at all.
TX Total chlorine
The organically bound chlorine contained in the sample is detected.
Bromide and iodine compounds are only detected partially. Fluorine
compounds are not detected at all. The result is output as defined as to-
tal chlorine. Inorganic halogen compounds are only digested partially or
not at all.
EOX Extractable organically bound halogens
The EOX parameter is the sum of organically bound halogens (chlorine,
bromide, iodine) extracted under defined conditions from a water sam-
ple or from solids (sludge and sediment) using an organic solvent. Fluo-
rine is not detected.
AOX Adsorbable organically bound halogens
The AOX parameter is the sum of organically bound halogens (chlorine,
bromide, iodine) adsorbed under defined conditions (pH < 2 with
HNO3) from a water sample or a solid sample (sludge and sediment) in
active carbon using the vibration or column method. Fluorine is not de-
tected.
EC/OC Elementary carbon/organic carbon
The content of elementary carbon of a sample is detected directly after
the thermal desorption of the organic carbon. The amount of organic
carbon is then determined sequentially after thermal oxidation of the
remaining sample.
The following parameters can be determined from aqueous samples with the multi EA
5100:
Parameter Explanation
TC Total carbon
TIC Total inorganic carbon
The total inorganic carbon from carbonates, hydrocarbonates and dis-
solved carbon dioxide is determined.
TOC Total organic carbon
The total organic carbon content is determined mathematically by cal-
culating the difference: TOC = TC - TIC. Determination of the TC and TIC
content of the sample is required for this.
NPOC Non purgeable organic carbon
The total non-purgeable organic carbon content of a sample is de-
tected. Highly volatile organic compounds are not or only partially de-
tected.
9
Security multi EA 5100
3 Security
For your own safety and to ensure error-free and safe operation of the device, please
read this chapter carefully before commissioning.
Observe all safety instructions listed in this user manual and all messages and infor-
mation displayed on the monitor by the control and analysis software.
Warning and mandatory action labels have been attached to the device and must al-
ways be observed.
Damaged or missing warning and mandatory action labels can cause incorrect ac-
tions leading to personal injury or material damage. The labels must not be removed.
Damaged warning and mandatory action labels must be replaced immediately!
The following warning and mandatory action labels have been attached to the de-
vice:
Warning sym- Meaning Comment
bol
Warning against dan- In the device interior on the electronics
gerous electrical volt- cover
age Warning against electrical voltage. Be-
fore opening the electronics the device
must be disconnected from the mains.
Warning about hot In the device interior on the combus-
surface tion furnace
On the top cover of the combustion
furnace
On the heated gas transfer line from
the Cl module 5100
Touching the furnaces or the ends of
the gas transfer lines during operation
or shortly after shutdown of the device
can result in burn injuries.
Warning against cor- On the "high sensitive" measuring cell
rosive substances of the Cl module 5100
Exercise caution when handling the
electrolyte solution. It contains highly-
concentrated acetic acid.
On the sulfuric acid container and the
safety attachment of the Cl module
5100
Exercise caution when handling con-
centrated sulfuric acid
Warning against UV In the S module 5100 (basic, MPO)
radiation Switch off the detection module before
performing maintenance on the UV
lamp.
10
multi EA 5100 Security
The device must only be operated by qualified specialist personnel instructed in the
use of the device. This instruction also include teaching the contents of this user
manual and of the user manuals of the connected system components. We recom-
mend training by qualified employees of Analytik Jena or its representatives.
In addition to the safety instructions in this user manual, the general applicable
safety and accident prevention regulations of the respective country the device is op-
erated in must be observed and adhered to. The operator must ensure the latest ver-
sion of these regulations.
The user manual must be accessible to the operating and service personnel.
Incorrect installation can create serious hazards. This may result in electric shock and
explosion if the gases are not connected correctly.
¡ Only the Analytik Jena GmbH customer service or specialist personnel trained
and authorized by them is allowed to install and commission the device and its
system components.
¡ Unauthorized assembly and installation is not permitted.
Insufficiently secured components pose a risk of injury.
¡ During transport, secure the device components as specified in these operating
instructions.
¡ Loose parts must be removed from the system components and packed sepa-
rately.
11
Security multi EA 5100
To prevent health damage, the following must be observed when moving the device
in the laboratory (lifting and carrying):
¡ For safety reasons, two persons are required to transport the device who must
hold the unit by either side of the equipment.
¡ The device does not have any carrying handles. Therefore the device must be
gripped firmly with both hands at the lower end.
¡ Risk of damage to health due to improper decontamination! Perform a profes-
sional and documented decontamination of the device before returning it to An-
alytik Jena. The decontamination report is available from Service when register-
ing the return. Service must refuse acceptance of contaminated devices. The
sender may be liable for damage caused by inadequate decontamination of the
device.
The operator must make sure that the device and its safety equipment is in sound
condition each time before starting up the device. This applies in particular after
each modification or extension of the device or its repair.
Observe the following:
¡ The device may only be operated if all items of protective equipment (e.g. covers
in front of electronic components) are in place, properly installed and fully oper-
ational.
¡ The sound condition of the protection and safety equipment must be checked
regularly. Any defects must be corrected as soon as they occur.
¡ Protective and safety equipment must never be removed, modified or switched
off during operation.
¡ Free access to the power switch on the back of the enclosure must be ensured
during operation.
¡ The ventilation equipment on the device must be in good working condition.
Covered ventilation grilles or slots etc. may cause the device to break down or
may cause damage to it.
¡ The furnace operates at temperatures of 700 to 1100 °C. Hot components must
not be touched during or directly after the operation of the device.
¡ Caution when handling glass components. Risk of broken glass and therefore risk
of injury!
¡ Keep all combustible materials away from the analyzer.
¡ When operating the sampling modules (Autoinjector, MMS and ABD), there is a
crushing risk for hands and fingers. Maintain a safe distance.
¡ When handling the Autoinjector, there is a danger of stabbing yourself in the
hand or finger with the syringe. Exercise caution when handling the syringe.
¡ When operating the multi-matrix autosampler with temperature control func-
tion (MMS-T or Multi Matrix Sampler with Liquids kit TMP), the syringe holder
and the sample tablet achieve temperatures of up to 80 °C during heating. These
hot components present a risk of burn injuries.
12
multi EA 5100 Security
Smoking or handling open flames are prohibited in the room in which the device is
operated!
Lethal voltages may occur in the device! Contact with live components may cause
death, serious injury or painful electrical shock.
¡ The power plug must be connected to a proper power outlet to ensure that the
device meets protection class I (ground connector). The device may only be con-
nected to power sources whose nominal voltage is the same as that on the rating
plate of the equipment. Do not replace the removable power cable of the device
with a power cable that does not meet the specifications (with no protective
ground conductor). Extensions of the supply cable are not permitted!
¡ The basic module and the system components may only be connected to the
mains when they are switched off.
¡ Electrical connection cables between the basic module and the system compo-
nents may only be connected or disconnected when the device is switched off.
¡ Before opening the device, the device must be switched off at the main switch
and the power plug must be disconnected from the power outlet! The only ex-
ception is for instructions that explicitly indicate that the doors of the basic mod-
ule or of a detection module must be opened during operation. This is the case,
for example, during the end point routine of the Cl module 5100 or when
searching for a gas leak in the measuring gas connection between the basic
module and the Cl module 5100.
¡ Work on the electronics may only be carried out by the customer service of Ana-
lytik Jena and specially authorized technicians.
The operator is responsible for the selection of substances used in the process as well
as for their safe handling. This is particularly important for radioactive, infectious,
poisonous, corrosive, combustible, explosive and otherwise dangerous substances.
When handling hazardous substances, the locally applicable safety instructions and
instructions in the safety data sheets from the manufacturers of the auxiliary and op-
erating materials must be complied with.
Hazardous substances are used during operation of the following detection modules:
Detection module Hazardous substance Use
Cl module 5100 Concentrated sulfuric Drying agent in the sul-
acid furic acid container
Acetic acid Preparing the electrolyte
Concentrated nitric acid solution
Methanol
Thymol
S module 5100 coulo- Acetic acid Preparing the electrolyte
metric solution
TOC module 5100 40 % orthophosphoric Reagent in the TIC reac-
acid tor
0.2 mol/l hydrochloric Reagent for NPOC de-
acid termination
13
Security multi EA 5100
Acetic acid vapors causing strong irritation of the airways can occur in the measuring
cell of the Cl module 5100. Connect the measuring cell with the exhaust hose and
connect the detection module to the laboratory exhaust unit.
Extreme caution is required during system overpressure! Incorrect operation can en-
danger the operating personnel and damage the analysis system. If an overpressure
fault is detected, a warning message will appear in the multiWin program.
Observe the following:
¡ Never switch off a device subject to overpressure!
¡ Do not introduce any samples.
¡ Do not close the program.
¡ Do not switch off the modules.
¡ Do not switch off the gas supply.
¡ N/S/C branch: Wait until the overpressure in the system has dropped to normal
pressure. The routine for reducing the overpressure in the N/S/C branch runs au-
tomatically.
¡ Chlorine branch: For devices with an operational Cl module 5100, ventilation
must be performed manually (→ "Recommissioning after emergency shutdown
(Cl module 5100)" 40).
¡ After relieving the overpressure remove the obstruction in the gas path.
14
multi EA 5100 Security
¡ If there is no immediate risk of injury, switch off the device and the connected
system components immediately in hazardous situations or in the event of an
accident and/or disconnect the power plugs from the power outlets.
¡ Close the oxygen gas supply as soon as possible after switching off the equip-
ment!
¡ When closing the argon gas supply, a return of the acid used can occur in the Cl
module 5100. For this reason, the measuring gas lines to this module must be
disconnected before closing the argon gas supply.
¡ When attaching the Cl module 5100: There is a risk of injury from sulfuric acid!
During an emergency shutdown the vacuum may draw sulfuric acid into the
transfer line and the auto-protection valve assembly during the cooling down pe-
riod. After an emergency shutdown, check that the transfer line and the auto-
protection valve assembly are not contaminated with sulfuric acid (→ "Recom-
missioning after emergency shutdown (Cl module 5100)" 40) before recom-
missioning.
¡ When closing the gas supply, a return of the slightly acidic electrolyte solution
and the phosphoric acid in the TIC reactor can occur in the S module 5100 coulo-
metric and the TOC module 5100. Check the measuring gas hoses for contami-
nation before recommissioning. Observe the safety instructions for handling
acidic solutions during cleaning.
15
Function and design multi EA 5100
The multi EA 5100 analysis system is modular and can be adapted to each measuring
tasks by combining it with different detectors and sampling modules. The analysis sys-
tem is used to determine the sulfur, nitrogen, chlorine and carbon content in solid, liq-
uid, paste-like, viscous and gaseous samples. It can additionally analyze the AOX, EOX,
EC/OC or TOC, NPOC and TIC sum parameters.
The core of the analyzer system is the multi EA 5100 basic module, which performs
sample gas digestion and measuring gas drying. (As an exception, the measuring gas is
dried in the detection module for chlorine determination.)
The analysis system is controlled and the measurement data is evaluated via the multi-
Win control and analysis software.
A self-check system (SCS) is integrated into the analysis system. The SCS consists of a
combination of hardware components and software functions that automatically ensure
proper function of the entire analysis system. Depending on the system setup, the SCS
checks the important device safety and analysis quality parameters (e.g., gas flow, tem-
peratures, pressures, system integrity, baseline stability, signal drift, cooling time, flame
value, etc.) multiple times per second
The basic module is equipped with a double furnace. This innovative technology enables
operation of the combustion furnace in vertical or horizontal mode. Switching between
the two operation modes is simple, and is performed by the user.
Vertical operation mode In vertical operation mode, the sample aliquot is injected directly into the multi-purpose
combustion tube via the injection port. The sampling module required is a multi-matrix
autosampler, an auto-injector, a GSS module or an LPG module.
Advantages of vertical operation mode:
¡ For gases, LPG, liquid samples with normal viscosity
¡ Best for N, A and Cl trace and ultratrace analyses
¡ Fast analysis
¡ Small bench space required
Horizontal operation mode In horizontal operation, solids and liquids are transferred to the multi-purpose combus-
tion furnace on boats with Automatic Boat Drive (ABD). Sample feeding can be auto-
mated with an autosampler in combination with the ABD. Liquids can alternatively be
transferred into the horizontal combustion furnace via the auto-injector.
Gases and LPG are injected directly via the injection port of the sample port of the ABD.
Advantages of horizontal operation mode:
¡ For gases, LPG, solids, liquid samples independent of viscosity
¡ Best for volatile liquids
16
multi EA 5100 Function and design
Sample feed
Liquid/LPG/gaseous
Sample feed
Solid/liquid/LPG/gaseous
TS, TN, TC, TX and EOX Sample digestion to determine TS, TN, TC, TX and EOX can be performed in vertical and
horizontal operation mode.
Digestion is a two-stage process at 1000 to 1100 °C via pyrolysis with subsequent ther-
mal oxidation. In the first phase, the sample components are pyrolyzed in an argon flow,
and the generated pyrolysis gases are incinerated in an oxygen flow. Next, the remain-
ing pyrolysis products are re-incinerated in a pure oxygen flow during the second phase.
This digestion is summed up via the following equations:
R*-H + O2 → CO2 + H2O
R*-N + O2 → NOx + CO2 + H2O
R*-S + O2 → SO2 + CO2 + H2O
R*-X** + O2 → HX** + CO2 + H2O
R* - carbonic substance
X** - F-, Cl-, Br-, I-
TOC Digestion for TOC determination is performed in vertical operation mode via thermal ox-
idation in an oxygen flow at 700 °C. The aqueous samples are injected directly into the
TOC combustion tube via the injection port.
TIC Digestion for TIC determination is performed in the TOC module via wet chemical oxida-
tion with phosphoric acid. The aqueous samples are injected into the TIC reactor manu-
ally.
AOX Digestion for AOX determination is performed in the horizontal operation mode. The
loaded active carbon is combusted in the oxygen flow at at least 950 °C and forms hy-
drogen halogen, carbon dioxide and water. The loaded active carbon is transferred to
the multi-purpose combustion tube with or without a quartz container in the quartz
glass boat with ABD.
17
Function and design multi EA 5100
EC/OC Digestion for EC/OC determination is performed in horizontal operation mode. Digestion
is performed in two phases. In the first process phase, the OC content adsorbed on the
filter samples is thermally desorbed in a pure argon flow. The gaseous products are then
re-incinerated in oxygen. In the second process phase, the remaining EC proportion in
pure oxygen is converted completely to CO2.
The filter samples are transferred to the combustion furnace via the ABD and a special
feed program.
After leaving the combustion tube, the mixture is dried before it is transferred to the de-
tectors.
Parameter Method
TS, TN, TC, EC/OC Membrane dryer (in the basic module)
TX, AOX, EOX Concentrated sulfuric acid (in the Cl module 5100)
TOC, NPOC, TIC Condensation through Peltier cooling (in the TOC module
5100)
Main components of the basic The multi EA 5100 basic module contains the following main components:
module ¡ Electronics/internal device control
¡ Gas supply
¡ Combustion system
¡ Measuring gas transfer
All basic module components that must be operated or maintained by the user can be
accessed via the 2 doors on the front side or via the removable side panels.
Electrical connection and gas connections) and the interfaces for connecting the system
components are on the rear of the device.
Double furnace device ¡ Vertical and horizontal operating mode possible
¡ Opening with maintenance flaps at the right-hand side panel
Vertical operation ¡ Furnace in vertical installation position
¡ Right side panel closed
18
multi EA 5100 Function and design
7 1
6
2
3
5
1
4
2
3
19
Function and design multi EA 5100
Internal device control The control electronics are found on the rear of the basic module behind the panel when
viewed from the front. The control electronics provides the power supply and control of
the individual components and the communication with the control PC and other con-
nected system modules.
Operational LED
Power switch, ports, connec- The power switch and the interfaces for connecting the system modules and for con-
tions necting the control PC can be found on the rear of the device.
The control PC can be connected via a USB port. The interfaces for connecting the sam-
pling modules and for connecting the detectors are RS 232 interfaces.
20
multi EA 5100 Function and design
1
2
21
Function and design multi EA 5100
11
10 9 8 7 6 5 4
Figure 7 Interfaces for detector modules and sampling modules
1 "External" connection 2 Autoinjector connection
3 "Flame" flame sensor connection 4 S-coulometer connection
5 S-UVF connection 6 CI-coulometer connection
7 N-CLD connection 8 C-NDIR connection
9 Sampler connection (RS-232 bus) 10 Service connection
11 Control PC connection
Interfaces in the device The electrical connections for the combustion furnace, the flame sensor and the temper-
ature sensor can be found on the rear interior side of the device. The connections are
only accessible in the vertical installation position of the combustion furnace.
1 2 3
Figure 8 Connection for the sensors and the combustion furnace in the device
1 Temperature sensor 2 Combustion furnace
3 flame sensor
The connections for the auto-protect valve assembly and the heated transfer line (only if
the CI module is connected) can be found in the frame behind the door. The toggle
switch for opening and closing the pneumatic seal in the auto-protection valve assembly
is also located there.
22
multi EA 5100 Function and design
1 2 3
Figure 9 Connections for the auto-protection valve assembly and transfer line
1 Toggle switch for opening and clos- 2 Transfer line heating connection
ing the pneumatic seal in the auto-
protection valve assembly
3 Valve assembly connection
Hose diagrams The connection between the individual components is made with labeled hoses. The
numbers circled in the hose diagram match the labels on the hoses in the multi EA
5100.
23
24
Function and design
9 Filter
Filter 1 2 16
N/S/C Exhaust
outlet
Ar O2 ABD
7 15 Gas box
MFM
main
bypass
inlet
seal
dryer
Pump
6 3
8 12 11 4
Cl system tightness
test
Rotary bearing
Combustion furnace
10
Rotary bearing
13
5
Membrane dryer
multi EA 5100
to the N/S/C detector
9 Filter 1 2
multi EA 5100
N/S/C Exhaust
outlet
Ar O2 ABD 3
main
7 15
4
MFM inlet
Pump
11
6
seal
13
Membrane dryer
Cl system tightness
Gas box
12
8
bypass
5
Combustion furnace
Rotary bearing
Rotary bearing
14
Coupling
10
to the Cl detector
25
Function and design
Function and design multi EA 5100
Gas connections at the rear of The gas connections are located on the rear of the device. The gas supplies for oxygen
the device and argon must be connected to the "IN O2" or "IN Ar" connection via the included con-
nection hoses (AD 6 mm, ID 4 mm.
5 1
Gas connections on the gas box The two process gases, argon and oxygen, are controlled via the gas box in the basic
module. The gas box is located on the left side of the device.
26
multi EA 5100 Function and design
2
1
27
Function and design multi EA 5100
Gas supply control The composition of the gas mixture for optimum gas digestion is controlled via the flow
management system (FMS).
1
3
2
Figure 16 Multi-purpose combustion tube
1 Oxygen supply connection 2 Argon supply connection
3 Flame sensor connection 4 Screw cap with septum (only for vertical opera-
tion mode and operation with an auto-injector)
28
multi EA 5100 Function and design
4.2.1 Auto-injector
There are two types of auto-injector. The classic auto-injector is suitable for vertical and
horizontal operation mode, while the Autoinjector AI-EA is only used in vertical opera-
tion mode.
The auto-injectors are suitable for the following applications:
¡ In vertical operation mode for non-viscous liquids and colorless EOX extracts
¡ In horizontal operation mode only for volatile liquids
The auto-injectors are unsuitable for the following sample types:
¡ Viscous liquids and their solutions
¡ Solids suspended in solution
¡ Colored EOX extracts
¡ Water analyses (TC/NPOC determination)
29
Installation and commissioning multi EA 5100
Ambient conditions The climate conditions for the installation location are listed in the technical data (→
"Technical data for the multi EA 5100 " 172). If required, ensure that the room is
temperature-controlled.
Laboratory conditions The device is only approved for indoor use. The installation location should have the
characteristics of a chemical laboratory. It must meet the following conditions:
¡ Atmosphere free from hydrocarbons, halogens, sulfur compounds and nitrogen ox-
ides
¡ Atmosphere with low dust levels
¡ No vibrations
¡ No smoking in the operating room of the device
Installation location require- The requirements for the installation location of the device are as follows:
ments ¡ No caustic vapor in the immediate vicinity of the device and its system components.
These could corrode the device connections and modules.
¡ Free from draft air; do not install the device close to windows or doors
¡ Away from electromagnetic sources of interference
¡ No direct sunlight and away from radiant heaters
¡ The front door and air vents must not be obstructed by other equipment or furnish-
ings
¡ Maintain a safety distance of at least 20 cm from the rear of the device to other de-
vices or walls.
WARNING
Danger due to electrical voltage
¡ Only connect the device to a properly grounded socket which complies with the volt-
age indicated on the device’s rating plate.
¡ Do not use an adapter in the feeder.
30
multi EA 5100 Installation and commissioning
The operator is responsible for the gas supply and the corresponding connections and
pressure reducers.
The connection hoses with outer diameter (AD) 6 mm and inner diameter (ID) 4 mm
are provided with the device. Their lengths are 2 m. If other lengths are preferred,
please contact the Analytik Jena GmbH customer service department. The required gases
and their qualities are listed in the technical data (→ "Technical data for the multi EA
5100 " 172).
The space required for the modular analysis system depends on all the components that
make up the measuring station. The analysis station always includes:
¡ Basic module
¡ 1 sampling module (to the right of the basic device or on top of it)
¡ 1 detection module (to the left of the basic device)
Several modules may also be placed in a row.
Component Width x Height x Depth Mass [kg] Arrangement
[mm]
Basic module 510 x 470 x 550 mm 25 kg
multi EA 5100
Detection modules
N module 5100 300 x 500 x 550 mm 13 kg As last detector in a
row
S module 5100 300 x 470 x 550 mm 13 kg Left/right of other
basic detectors
S module 5100
MPO
Cl module 5100 300 x 470 x 530 mm 12 kg To the immediate left
of the basic module
S module 5100 300 x 470 x 530 mm 11 kg Left of the Cl module
coulometric 5100
C module 5100 300 x 470 x 530 mm 12 kg Left/right of other
detectors
TOC module 5100 300 x 470 x 530 mm 12 kg To the immediate left
of the basic module
or to the left of the Cl
module 5100
Sampling modules
ABD 520 x 210 x 500 mm 12 kg To the right of the
basic module
MBD 500 x 80 x 80 mm 0.35 kg To the right of the
basic module
Multi Matrix Sam- ca. 510 x 500 x 410 ca. 9.5 kg On the basic module
pler mm or on the ABD*
31
Installation and commissioning multi EA 5100
Figure 18Space required for the basic device and modules (vertical operation)
32
multi EA 5100 Installation and commissioning
Figure 19Space required for the basic device and modules (horizontal operation)
33
Installation and commissioning multi EA 5100
WARNING
Danger due to incorrect commissioning
¡ The analysis system may only be set up, mounted and installed by the Analytik Jena
GmbH customer service department.
¡ Any unauthorized access of the device can endanger the user and the operational
safety of the equipment and limits or completely voids any warranty claims.
NOTICE
Retain the transport packaging
Return transport for maintenance must be in the original packaging. This alone prevents
transport damage.
The basic module, the sampling modules and the detectors are unpacked and assembled
by the Analytik Jena GmbH customer service department or personnel trained and au-
thorized by them.
Check when unpacking the device for completeness and soundness of the delivery in ac-
cordance with the packing list included.
The customer service department tests the function of the analysis system after installa-
tion and documents the test.
Installing the basic module } Carefully remove the basic module and its components from the transport packaging.
Do not damage the transport packaging!
} Place the basic module at its intended location.
} Ensure adequate space for the additional system components (sampling modules,
detectors).
} Install the combustion furnace (→ "Removing and installing the combustion furnace"
141).
} Install the auto-protection valve assembly (→ "Maintenance of the auto-protection
valve assembly" 131).
} Install the combustion tube in the combustion furnace (→ "Maintenance of the
multi-purpose combustion tube" 126).
} Place the additional system modules on their intended locations and connect them.
CAUTION
Danger of short circuit!
¡ Only connect the basic module and other system components when they are pow-
ered down and switched off.
¡ Before connecting the power supply cable, set the power switch of the rear of the
device to "0".
¡ Only use the IEC connection cable included in the scope of delivery for the connec-
tion to the power supply (VDE label, 1.5 m long). Extensions of the supply cable are
not permitted!
34
multi EA 5100 Installation and commissioning
NOTICE
Settled condensation and temperature differences can damage individual com-
ponents of the basic module during commissioning.
¡ If there is any temperature difference between the storage location and the operat-
ing location, allow the analyzer system to acclimatize for at least one hour in the
room it will operate in.
Connecting power and gases } Connect the power cable to the connection of the rear of the basic module.
} Connect the power plug to a grounded power outlet.
} Connect the supplied connection hoses to the pressure reducers of the gas supply
and the the O2 and Ar gas connections on the rear of the device (media connections
of the rear of the device).
} Set the inlet pressure on the pressure reducers (600 kPa (6 bar)).
} Connect the computer and connect it to the basic module via the supplied USB cable.
} Connect further system components (detectors, sampling modules) to the basic
module.
ü The basic module is now ready for operation and can be switched on.
3
2
35
Operation multi EA 5100
6 Operation
Only use the correct sampling modules for the corresponding matrix and installation po-
sition of the furnace for sampling.
Furnace installation Sample type Sampling
position
Vertical: Liquids ¡ Autoinjector
¡ Multi Matrix Sampler
MMS
Gas, non-pressurized ¡ GSS module
Gas, pressurized ¡ GSS/LPG combi module
¡ GSS module with GSS
adapterbox
LPG ¡ LPG module 2.0
¡ GSS/LPG combi module
Horizontal: Solids ¡ Automatic Boat Drive
ABD
¡ ABD with MMS
Liquids ¡ ABD with manual dosing
syringe
¡ ABD with MMS
¡ Autoinjector
Gas, non-pressurized ¡ GSS module
Gas, pressurized ¡ GSS/LPG combi module
¡ GSS module with GSS
adapterbox
LPG ¡ LPG module 2.0
¡ GSS/LPG combi module
Observe the following during analysis:
¡ To add samples of flammable substances in horizontal operation, always use ABD
with a flame sensor in automatic or automatic plus mode or an Autoinjector.
¡ To add samples of flammable substances, do not use the ABD in parameter mode,
or only use it with program parameters designed and tested for it (risk of soot!).
¡ Do not exceed the maximum permissible sample amounts (→ "Technical data for
the multi EA 5100 " 172).
¡ Adjust the dosing rate to the sample matrix and observe the maximum dosing
speeds (→ "Technical data for the multi EA 5100 " 172).
¡ Standard solutions with organic solvents can change composition rapidly due to the
volatility. Therefore, ensure that the clear space above the liquid in the sample cup
is small when preparing and storing samples. Store the solutions in the refrigerator.
The boiling points of the materials used should also not differ by more than 50 °C.
¡ Adjust the sample volume to the expected concentration to remain within the mea-
suring range of the detector.
¡ Begin the analysis with a standard solution and determine the daily factor. If the
value measured for the standard solution differs by more than 20 % from the target
value, repeat the measurement. If required, look for the source of the fault. If re-
quired, re-calibrate the analysis system.
36
multi EA 5100 Operation
¡ For examining very low element contents, examining a blind value is recommended
before the daily factor. The blind value measurement cleans the analysis system.
Select the proper measurement method for each sample with the following table.
Some parameters can only be analyzed with horizontally or vertically installed furnaces
¡ Horizontal: AOX, EC/OC
¡ Vertical: TOC, TIC, NPOC (in water)
For other samples, the recommended measurement method depends on the nature of
the sample.
Sample Furnace installation po- Comment
sition and sampling
TS, TN, TX, TC in:
Organic solids, e.g., wax, Horizontal for ABD with flame sensor
polymer ABD For trace analysis:
High-viscosity organic liq- ABD with MMS Use 1 quartz glass boat
uids, gels and paste-like for all samples
samples, e.g., crude oil, as- MMS with temperature-
phalt, bitumen, tar control unit in heating
Bio-Diesel mode for high-viscosity
Oils such as crude oil, paraf- liquids
fin oil and vegetable oil ABD without MMS for
Volatile liquids such as pe- non-homogeneous sam-
troleum ether, methanol, ples
naphtha
Liquids or solids with high
element content
(> 100 mg/l)
Organic liquids with normal Horizontal for ABD with flame sensor
viscosity, such as fuels, sol- Autoinjector
vents
ABD
ABD with Multi Matrix
Sampler
Vertical for For trace analysis: must
Autoinjector and Au- be vertical
toinjector AI-EA
MMS
EOX Horizontal for ABD with flame sensor
ABD Sample (colorless, col-
ABD with MMS ored) in n-hexane or pe-
troleum ether
Vertical for Only colorless samples in
MMS n-hexane
Autoinjector
37
Operation multi EA 5100
38
multi EA 5100 Operation
NOTICE
Observe the run-in period
In the Status analyzer window, components that are not yet ready for operation are dis-
played in red. The heating time for 1050 °C is approx. 30 min. Measurement is not pos-
sible during the run-in period.
¡ If the analyzer is still not ready after approx. 30 min, perform a fault search (→
"Troubleshooting" 98).
NOTICE
Risk of overheating
If the basic module is switched off too early, the electronics can overheat and become
damaged due to lack of cooling.
¡ Only switch off the basic module after a cooling time of 1 h.
39
Operation multi EA 5100
WARNING
Chemical burns due to concentrated sulfuric acid
If a Cl module 5100 is connected to the basic module, sulfuric acid can still be in the gas
transfer line and the auto-protection valve assembly after an emergency shutdown.
¡ Wear protective clothing when working on the sulfuric acid container.
¡ Exercise particular caution when checking the gas transfer line and the auto-protec-
tion valve assembly.
¡ Observe all specifications in the safety datasheet.
CAUTION
Risk of burns from the hot furnace and from the gas transfer line
¡ Allow the device to cool before recommissioning.
Observe the following when recommissioning the basic module and Cl module 5100.
Vent the analysis system after an overpressure fault (fault message"206 – gas pressure
fault") as described in the following:
} Carefully disconnect the measuring gas hose from the measurement cell.
} Carefully disconnect the gas transfer line from the sulfuric acid container and remove
the sulfuric acid container from the module.
} Disconnect the gas transfer line of the auto-protection valve assembly in the basic
module. Remove the heating cable plug from its socket.
} Carefully remove the gas transfer line and check it for contamination with sulfuric
acid.
} If necessary, clean the gas transfer line:
– Rinse the gas transfer line with distilled water and then with ethanol.
– Dry the gas transfer line (e.g. by blowing it through with an inert gas).
} Wait for the pressure in the system to decrease. Then switch off the basic module.
Shut off the gas supply.
} Open the seal for the auto-protection valve assembly in the basic module. To do this,
flip the toggle switch up. Remove the valve assembly plug from its socket.
} Carefully remove the auto-protection valve assembly from the basic module and
check it for contamination with sulfuric acid.
} If necessary, clean and dry the assembly. Replace the filter.
If cleaning is not possible or if the auto-protection valve assembly is damaged, it
must be replaced prior to recommissioning.
} Reinstall the auto-protection valve assembly in the basic module. Connect the as-
sembly via the cable. Ensure that a filter has been inserted in the auto-protect valve
assembly.
} Refill the sulfuric acid container with sulfuric acid and insert it in the detection mod-
ule. Connect the measuring gas hose with the sulfuric acid container.
} Refit the gas transfer line:
40
multi EA 5100 Operation
– Connect the gas transfer line with the auto-protection valve assembly. Plug the
heating cable plug into its socket.
– Route the gas transfer line through the wall of the basic module top the detection
module. Connect the gas transfer line to the sulfuric acid container.
ü The basic module and the detection module can be switched back on again.
41
Nitrogen analysis with the N module 5100 multi EA 5100
Expansion of the basic module with the detection module allows the determination of
the nitrogen content in solids, liquids, and gases via chemiluminescence.
Organic nitrogen compounds can be determined as a TN sum parameter with the analy-
sis system. Inorganic nitrogen compounds are only detected if they can be digested in
the combustion furnace. Pure nitrogen cannot be analyzed.
The chemiluminescence of the reaction between nitrogen monoxide (NO) and ozone
(O3) is used for the determination. This reaction creates nitrogen dioxide in its excited
state (NO2*) for a short time. During the transition to the base state, the nitrogen diox-
ide emits electromagnetic radiation in the visible light spectrum. The light emitted is
proportional to the NO2* concentration. This allows measurement of the concentration
via light. NO is the only substance involved in the reaction, which means that this
method is very selective and not influenced by any other constituents of the measuring
gas.
NO + O3 → NO2* + O2
NO* → NO2 + hv
The measuring gas is created during incineration of organic nitrogen compounds in the
basic module. It contains a mixture of NO and NO2, generally known as NOx.
R-N + O2 → NOx + CO2 + H2O
R: Hydrocarbon residue, NOx: Mixture of NO and NO2
The measuring gas is passed through a converter which renders the NO2 portion usable
for the reaction and hence for detection purposes. The converter reduces NO2 to NO.
The ozone required for the reaction is created in the device from the supplied oxygen
(O2). Excess O3 is removed in the ozone decomposer after the reaction. The toxic gas is
not released into the ambient air.
7.1.2 Design
The detection module is used to determine the nitrogen content via chemiluminescence.
All components required for determination are mounted inside the sealed housing.
42
multi EA 5100 Nitrogen analysis with the N module 5100
7.1.3 Connection
The equipment door at the front is closed tightly and cannot be opened. An LED is in-
stalled on the door. The LED flashes during the run-in period of the detection module
and lights up permanently when the module is operational.
The following connections can be found on the rear of the detection module:
¡ Main switch, power connection, device fuses
¡ Media connections for gases and waste
¡ Interface for connection to a basic module
¡ Service interface with programming button.
A diagram at the center on the rear of the device explains the different connections.
43
Nitrogen analysis with the N module 5100 multi EA 5100
9
8
7
6 1
44
multi EA 5100 Nitrogen analysis with the N module 5100
7.2 Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
NOTICE
The detection module is equipped with a pump that can disrupt the function of other
optical detection modules or lead to serious errors.
¡ Always connect the detection module as the last in any series of the detectors.
Figure 23Absorber
1 Hose 6 connection 2 Holding clamp
3 Absorber
} Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
45
Nitrogen analysis with the N module 5100 multi EA 5100
46
multi EA 5100 Nitrogen analysis with the N module 5100
7.3 Operation
CAUTION
Risk of respiratory problems due to leaking ozone
If the gas hoses have not been properly connected to the ozone generator, ozone may
leak out of the detection module.
¡ If you can smell ozone, switch off the device and check the connection of the gas
hoses on the ozone generator.
47
Chlorine analysis with the Cl module 5100 multi EA 5100
Expansion of the basic module with the detection module allows determination of the
chlorine content in solid, liquid, paste-like, viscous and gaseous samples. When doing
so, the device will determine the bromine and iodine content as proportions of the total
chlorine of the sample as well. Fluorine is not detected.
Organic halogen compounds can be determined as TX, AOX and EOX sum parameters
with the analysis system. Inorganic halogen compounds are only determined if they can
be digested via the combustion furnace. Halogen hydracids or pure halogens cannot be
analyzed.
Organic halogen compounds are incinerated into hydrogen halogen, carbon dioxide and
water in the basic module.
The measuring gas flow is routed to the detection module via a transfer line and dried
there. The hydrogen halogens (HCl, HBr, Hl) are determined there via microcoulometric
titration. Hydrogen flouride (HF) is not determined.
In the first step, hydrogen chloride HCl* dissolves in the electrolyte, and dissociates to
hydrogen and chloride ions (H+, Cl-) In the measuring cell, the chloride ions react with
silver ions created via electrolysis to form silver chloride (AgCl). To achieve the most
complete reaction to AgCl, titration is carried out in a strong acetic acid electrolyte. The
solubility product of AgCl is reduced in the acetic acid electrolyte.
R-X + O2 → HX + CO2 + H2O
HX* → H+ + X-
Ag → Ag+ + e-
Ag+ + X- → AgX
R: Hydrocarbon residue, X: Cl, Br, I, * HBr und HI are determined as proportions of AgBr
and AgI.
The end point of the titration is displayed potentiometrically. In accordance with Fara-
day's law, the amount of chloride ions can be calculated from the amount of charge ex-
pended to create the silver ions.
8.1.2 Design
48
multi EA 5100 Chlorine analysis with the Cl module 5100
4 2
The "sensitive" measuring cell The "sensitive" measuring cell is used for chlorine contents between 1 to 100 µg.
The measuring cell consists of the electrode space, which holds the electrolyte solution,
and the stirrer block in the detection module. A generator anode in the form of a stable
silver plate (silver circle) is located on the bottom of the electrode space. The magnetic
stirring rod runs above the anode.
The cell is sealed airtight with a lid and three knurled head screws. The lid has two
openings:
¡ The opening marked "electrode" is for the amperometric combined electrode.
¡ The unmarked opening is used for direct injection into the measuring cell or to con-
nect to the exhaust.
49
Chlorine analysis with the Cl module 5100 multi EA 5100
2
1
2
1
50
multi EA 5100 Chlorine analysis with the Cl module 5100
1 5
2 3 4
Figure 28Combined electrode
1 Connector plug 2 Generator cathode (Pt)
3 Indicator electrodes (Ag) 4 Gas inlet to measuring cell
5 Measuring gas connection
(hose 20)
The "high concentration" mea- The measuring cell has the same function as the "sensitive" measuring cell, but has a
suring cell higher electrolyte volume. It is suitable for a chlorine content of 10 to 1000 µg, and is
recommended in particular for TX determination of highly contaminated waste and of
polymer samples with high PVC content.
Here, as well, the maintenance-free combined electrode is used.
1 2 3 4 5 6
51
Chlorine analysis with the Cl module 5100 multi EA 5100
The "high sensitive" measuring This measuring cell is used for very low chlorine content (0.01 to 10 µg). In particular, it
cell is recommended for use in EOX determination or for gaseous sample.
The measuring cell consists of the base element with a magnetic stirring rod and lid. The
positions of the required components are marked inside the lid of the measuring cell.
The central borehole is intended for an adapter that is connected to the exhaust hose. It
is used as an exhaust for the acetic acid vapors.
The small unmarked opening is used for direct injection into the cell. This opening is
closed with a plug when measuring gas is routed into the measuring cell from the com-
bustion furnace.
3 8
4
2
7
5
1
6
52
multi EA 5100 Chlorine analysis with the Cl module 5100
6
7
1 2
5 3
Figure 31Lid of the "high sensitive" measuring cell equipped with all electrodes
1 Platinum electrode 2 Silver electrode
3 Gas inlet tube with Teflon screw 4 Opening for direct injection, with
connection (for measuring gas) plug
5 Sensor electrode 6 Reference electrode
7 Adapter for exhaust connection
Sensor and reference elec- The sensor and reference electrode are always inserted into the measuring cell together.
trodes
The sensor electrode has a chlorinated sensor pin which is inserted into the measuring
cell, and a gold contact. The sensor pin is touch-sensitive. Scratch protection is applied
to it for storage.
The reference electrode is supplied dry, with no bridging electrolyte. The same acetic
acid electrolyte solution used for the measuring cell is used as a bridging electrolyte. The
bridging electrolyte is applied via the filling opening. The reference electrode is ready for
measurement after a run-in period of approx. 4 h in the measuring cell.
The filling opening of the reference electrode must be open during operation. For short-
term storage, close the filling opening and place the dry protective cap on the electrode.
The openings in the lid of the measuring cell are designed to align the sensor pin of the
sensor electrode toward the diaphragm of the reference electrode.
A pre-amplifier is applied to the reference electrode. Its connection cables connect the
reference electrode with the sensor electrode and with the connection on the detection
module.
53
Chlorine analysis with the Cl module 5100 multi EA 5100
1
5
2
3
Generator electrodes A pair of generator electrodes generate the silver ions needed for chemical precipitation.
The pair of electrodes consist of a platinum cathode with a salt bridge and a silver an-
ode.
54
multi EA 5100 Chlorine analysis with the Cl module 5100
3
1
55
Chlorine analysis with the Cl module 5100 multi EA 5100
8.1.3 Connection
An LED is installed on the front of the detection module. The LED lights up when the
module is switched on.
The power switch, the device fuse and the power connection are located on the rear of
the module. The RS 232 interface for connection to the basic module ("CL-Coul" inter-
face) is also located on the rear of the module. The gas outlet for connection to the ex-
haust hose of the laboratory exhaust system is located at the bottom left of the rear
panel.
5
4
3
56
multi EA 5100 Chlorine analysis with the Cl module 5100
1 2
5 4 3
Measuring gas transfer A heated gas transfer line is used to transfer the measuring gas. It connects the combus-
tion system of the basic module with the sulfuric acid container in the detection module.
The hose of the gas transfer line is connected to the sulfuric acid container connection
with a banjo bolt with a conical nipple. The other end of the gas transfer line is con-
nected to the auto-protection valve assembly in the basic device.
7 1
5 2
4
4 3
Figure 38Connection of the heated gas transfer lines to the sulfuric acid container
1 Safety attachment 2 Sulfuric acid container
3 Hose for measuring gas infeed 4 Connector
5 Banjo bolt 6 Heated gas transfer line
7 Measuring gas transfer to the mea-
suring cell with PTFE connector
The measuring gas is routed into the measuring cell via the safety attachment of the
sulfuric acid container.
¡ The measuring gas hose is connected directly to the combined electrode with the
"sensitive" and "high concentration" measuring cells.
57
Chlorine analysis with the Cl module 5100 multi EA 5100
¡ A cut glass tube is used to feed gas into the "high sensitive" measuring cell.
58
multi EA 5100 Chlorine analysis with the Cl module 5100
8.2 Installation
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
} Place the detection module to the immediate left of the basic module.
} Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
} Connect the detection module with the basic module via the interface:
CL-Coul interface on the rear of the detection module
CL-Coul interface on the rear of the basic module
} Connect the "waste" outlet with the exhaust hose. Connect the hose with the exhaust
system, or route the hose into a vent.
59
Chlorine analysis with the Cl module 5100 multi EA 5100
7 1
5 2
4
4 3
Figure 41Connection of the heated gas transfer lines to the sulfuric acid container
1 Safety attachment 2 Sulfuric acid container
3 Hose for measuring gas infeed 4 Connector
5 Banjo bolt 6 Heated gas transfer line
7 Measuring gas transfer to the mea-
suring cell with PTFE connector
} Insert the sulfuric acid container in the holders.
} Route the gas transfer line through the top right opening of the detection module.
} Connect the gas transfer line to the sulfuric acid container:
– Insert the thin hose (3) into the sulfuric acid container.
– Connect the connector (4) to the sulfuric acid container.
– Connect the gas transfer line via the banjo bolt (5) on the connector.
NOTICE! Ensure the proper fit of the conical nipples!
} Place the safety attachment on the sulfuric acid container and attach the safety clip.
} Connect hose 20 with the safety attachment. Connect the hose to the measuring cell
later.
} Connect the gas transfer line in the basic module:
– Route the free end of the gas transfer line through the top left opening on the
basic module.
– Connect the gas transfer line to the auto-protection valve assembly.
– Connect the heating cable of the gas transfer line to its socket.
60
multi EA 5100 Chlorine analysis with the Cl module 5100
Inserting measuring cells The "sensitive" and "high concentration" measuring cells are inserted into the detection
module as follows:
} Insert the measuring cell with magnetic stirring rod and lid into the de-
tection module.
} Fill the measuring cell with electrolyte solution.
} Insert the combined electrode into the marked opening of the measuring
cell.
} Fasten hose 20 from the sulfuric acid container to the connection of the
combined electrode with the PTFE connector (1).
1 2 } Connect the olive (2) with the exhaust hose (hose 21) in the detection
module.
} Connect the combined electrode and the measuring cell to the inside of
1 2 the rear panel of the detection module:
Combined electrode connection (1)
Measuring cell connection (2)
5 4 3 Do not use connections (3) to (5).
61
Chlorine analysis with the Cl module 5100 multi EA 5100
The "high sensitive" measuring cell is inserted into the detection module as follows:
NOTICE
Risk of damage to the sensor electrode
The sensor pin and the gold contact of the sensor electrodes are sensitive to touch.
¡ Apply scratch protection to the sensor pin for storage.
¡ Rinse the sensor pin with ultrapure water before use or to clean it. Do not touch it
again after this. Do not dry the pin or wipe it dry!
¡ Before use or to clean it, wipe the gold contact with a cloth and some ethanol. Do
not touch it again after this.
} Insert the measuring cell with magnetic stirring rod and lid into the de-
6 tection module.
7
} Fill the measuring cell with electrolyte solution.
1 2 } Insert the components in the following openings:
Platinum electrode with salt bridge (1): "Pt" opening
Silver electrode (2): "Ag" opening
Gas inlet hose (1): "Inlet" opening
3 Sensor electrode (5): "Sens" opening
5
Reference electrode (6): large "ref" opening
Adapter (7): Center "outlet" opening
The openings on the lid align the sensor electrode and the reference
4
electrode toward each other.
} Close the opening for direct injection (4, "test") with a plug.
} Connect the short hose with the T piece to the adapter (7). Connect the
exhaust hose in the detection module (hose 21) with a limb of the T
piece.
} Place the pre-amplifier on the reference electrode. Connect the pre-am-
plifier with the sensor electrode.
} Connect hose 20 from the sulfuric acid container to the gas inlet tube via
the PTFE connector.
NOTICE! The conical nipples of the PTFE connector must be in the
proper position on the hose. Otherwise, gas leaks may ensue.
} Connect the electrode on the inside of the rear panel of the detection
1 2 module:
Platinum electrode connection (3)
Silver electrode connection (4)
5 4 3 Sensor and reference electrode connection (5)
Do not use connections (1) and (2).
62
multi EA 5100 Chlorine analysis with the Cl module 5100
8.3 Operation
The measuring cells are similar in function. The "high concentration" measuring cell
works with a larger electrolyte solution volume.
WARNING
Risk of chemical burns
100 % acetic acid (glacial acetic acid), concentrated nitric acid and thymol can cause se-
vere chemical burns. Methanol is a toxic, highly flammable material.
¡ Wear protective clothing when preparing the electrolyte solution. Work under an ex-
tractor.
¡ Observe all instructions and specifications in the safety data sheets.
ð Reagents required: 200 ml acetic acid 100 % (glacial acetic acid), 4 ml concentrated
nitric acid, 4 g gelatin, 1.0 g thymol, 0.3 g thymol blue, 500 ml methanol
} Solution A:
Fill 500 ml of water into a 1000-ml volumetric flask, add 4 ml of HNO3 (conc.), care-
fully add 200 ml of acidic acid and top up with water to the marking.
} Solution B1:
Mix 4 g of gelatin in a beaker with 400 ml of water, allow to swell for 3 hours and
then dissolve whilst heating to 35 to 45 °C.
The excess gelatin will sediment at the bottom of the beaker. Only use the clear su-
pernatant. Filter the solution, if necessary.
} Solution B2:
Dissolve 1.0 g of thymol and 0.3 g of thymol blue in a beaker with 500 ml of
methanol.
} Solution B:
After solution B1 has cooled down to 18 to 22 °C, slowly add it to solution B2 while
stirring, transfer into a 1000-ml volumetric flask and top up with water to the mark-
ing.
} Solution C, finished electrolyte:
Pipette 8 ml of solution B in a 100 ml measuring cylinder and fill with solution A to
100 ml, or
pipette 40 ml of solution B in a 500 ml measuring cylinder and fill with solution A to
500 ml.
ü The electrolyte solution is finished.
Storage and shelf life of the electrolyte solutions:
¡ When stored at 4 ±3 °C in well-sealed bottles, solutions A and B can be kept for ap-
prox. 6 months.
63
Chlorine analysis with the Cl module 5100 multi EA 5100
¡ The finished electrolyte solution (solution C) can be kept for approx. 30 days if kept
in a well-sealed glass container at 20 to 25 °C.
Performing the end point rou- An end point routine is necessary after every electrolyte change. The end point routine is
tine used to adjust the electrolyte to the optimal operating range of the measuring cell. The
operating point of the measuring cell is: 1500 to 5000 Counts .
} Start the end point routine via the System |End point routine menu item.
} Remove the olive from the lid of the measuring cell. Dose the HCl solution directly
into the measuring cell when instructed to by the software.
"sensitive" measuring cell: 200 µl of 0.01 N HCl
"high concentration" measuring cell: 200 µl of 0.1 N HCl
} Directly after dosing, activate the end point routine by clicking on [OK].
} The End point routine status is displayed in the Status analyzer window during this
routine. After the end point routine, the Stand-by titration and the current indicator
value is displayed.
ü The system is ready for measurements.
The determined operation point of the combined electrode is displayed in the System |
Component test menu item in the Chlorine tab.
Protecting the combined elec- Observe the following to protect the combined electrode against unnecessary wear:
trode ¡ Always add fresh electrolyte to the measuring cell before the end point routine.
¡ Do not carry out the end point routine several times in a row.
Preparation of the electrolyte Measurements with the "high sensitive" measuring cell require an electrolyte solution.
solution The electrolyte solution is also used as a bridge electrolyte for the reference electrode.
WARNING
Risk of chemical burns
100 % acetic acid (glacial acetic acid) can cause severe chemical burns. Gases may de-
velop during shaking.
¡ Wear protective clothing when preparing the electrolyte solution. Work under an ex-
tractor.
¡ Observe all instructions and specifications in the safety data sheet.
ð Reagents required: 800 ml acetic acid 100 % (glacial acetic acid), 2.7 g sodium ac-
etate p.a. (CH3COONa), anhydrous
} Dilute 2.7 g of sodium acetate in 200 ml of ultrapure water in a 1-liter volumetric
flask.
} Carefully add 800 ml of glacial acetic acid. Keep moving the flask while pouring the
acid. Carefully shake the mixture.
NOTICE! Do not exceed the specified amounts of water and glacial acetic acid. Do
not fill the flask up to the 1-liter mark (volume contraction).
Performing the end point rou- The end point routine is used to set the electrolyte to the optimum operating point of
tine the sensor electrode within the operating range of the titration cell.
¡ Operating range: 1000 to 10000 Counts
¡ Optimal operation range: 3000 Counts
64
multi EA 5100 Chlorine analysis with the Cl module 5100
Automatic end point routine As soon as the indicator value is outside the operating range of the titration cell, the sys-
tem automatically triggers an end point routine. This routine may also be triggered be-
tween two measurements of a multiple determination. The End point routine status is
displayed in the Status analyzer window during this routine.
Indicator value is higher than the operating range:
¡ Silver ions are automatically produced. The electrolyte is set to the optimum operat-
ing range of 3000 Counts.
Indicator value is lower than the operating range:
¡ The software prompts the user to add the following solution to the measuring cell:
100 µl hydrochloric acid (HCl, 10 mg/l HCl)
¡ If the indicator value rises above the operating range again, silver ions will be pro-
duced automatically. The electrolyte is set to the optimum operating range.
When working with a MMS, the addition of chloride ions to the measuring cells can be
performed automatically if the indicator value goes below the operating range while an
analysis sequence is in progress. For this, the user must have prepared a suitable organic
chlorine solution and placed it in the intended position (110) on the sample rack.
After the end point routine, the measuring cell requires approx 15 min to achieve a sta-
ble cell potential. During this time, a negative drift can occur with indicator values below
3000 Counts.
Manual end point routine For indicator values within the operating range, the end point routine can be started
manually via the System |End point routine menu item.
After the end point routine, the current indicator value is displayed in the Status ana-
lyzer window. If the current indicator value is within the operating range and the drift is
stable, the system is ready for measurements
} Place the measuring cell with electrodes and electrolyte solution in the detection
module and connect it to electricity.
} Switch on the basic module and the detection module.
ü The devices boot up. The status LED on the front of the basic module light up in
green after approx. 30 s.
ü The LED on the front of the detection module flashes during the run-in period.
Depending on the detector, the run-in period can take up to 30 min. After this,
the LED will light up continuously. Starting a measurement is only then possible.
} Open the gas supply and set the required gas pressure.
} Switch on the PC.
} Start the control and analysis software and login with your username and password.
} Initialize the analysis system by clicking on [Initialize analyzer].
ü The initialization and automatic detection of all connected components will be
carried out.
} Ready the samples.
} Activate a pre-existing method via the Method |Method - activate menu item.
} Alternatively: Create a new method in the Method |Method - new menu. Select the
measurement parameter in the method. Release and activate the method.
} Select Start |Start - Analysis in the menu.
} Select an analysis group or create a new one and confirm via [OK].
} Create an analysis sequence.
65
Chlorine analysis with the Cl module 5100 multi EA 5100
CAUTION
Risk of respiratory problems due to leaking acetic acid vapors
The electrolyte solution of the "high sensitive" measuring cells contains high concentra-
tions of acetic acid.
¡ Ensure that the exhaust hose is connected to the measuring cell.
Check if the exhaust hose is connected to the "waste" outlet on the rear of the detec-
tion module, and that it is connected to the exhaust system.
¡ Before measurement operations, always close the front door of the detection mod-
ule and switch on the laboratory exhaust system.
WARNING
Risk of chemical burns
Concentrated sulfuric acid is used in the detection module as a drying agent. The con-
centrated acid can lead to severe chemical burns.
The 100 % acetic acid (glacial acetic acid), nitric acid and thymol used to create the elec-
trolyte solution can lead to severe chemical burns.
¡ Wear protective clothing when working with these hazardous substances.
¡ Observe all instructions and specifications in the safety data sheets.
¡ Fill the sulfuric acid container with fresh concentrated sulfuric acid every day (→
"Replacing the sulfuric acid and cleaning the sulfuric acid container." 147).
The "sensitive" and "high con- Fill the measuring cells with fresh electrolyte solution daily:
centration" measuring cells ¡ "sensitive" measuring cell: 15 to 20 ml
¡ "high concentration" measuring cell: 120 ml
The "high sensitive" measuring Fill the measuring cell daily with electrolyte solution: to approx. 65 ml
cell
Change the electrolyte solution:
¡ Once per week
¡ If analytical problems occur
¡ If crystalline deposits form
Open the refill opening of the reference electrode during operation.
66
multi EA 5100 Sulfur analysis with the S module 5100 (basic, MPO)
Expansion of the basic module with the detection module allows determination of the
sulfur content in solid, liquid, paste-like, viscous and gaseous samples via UV floures-
cence.
The measuring gas is created during incineration of organic sulfur compounds in the ba-
sic module. It contains sulfur dioxide (SO2).
R-S + O2 → SO2 + CO2 + H2O
R: Hydrocarbon residue
The detection is performed using the UV fluorescence method. Sulfur dioxide (SO2) ex-
cited via UV light emits a characteristic flourescence (220 to 420 nm). This flourescence
is measured. The SO2 concentration is determined from the changes to the flourescent
intensity.
9.1.2 Design
The detection module allows determinations of the sulfur content via UV flourescence.
All components required for determination are mounted inside the sealed housing.
67
Sulfur analysis with the S module 5100 (basic, MPO) multi EA 5100
The detection module is available with and without the MPO option. The MPO option is
not suited for multi-element methods that determine multiple elements simultaneously.
It can be switched on and off as needed via the control and analysis software.
Working with a method with the active MPO requires calibration performed with the
MPO active. Otherwise the measurement results will be too low. Conversely, use of cali-
bration with an active MPO for a method with no MPO can lead to false measurement
results that are too high.
9.1.3 Connection
The equipment door at the front is closed tightly and cannot be opened. An LED is in-
stalled on the door. The LED flashes during the run-in period of the detection module
and lights up permanently when the module is operational.
The device switch for switching the module on and off is located on the top right of the
rear of the device (when viewed from the front). The device fuse and power connection
are located beneath it.
Communication with the basic module is performed via a 9 pin interface cable. The in-
terface is labeled "S-UVF".
The hose for the measuring gas coming from the basic module is connected to the "sam-
ple in" gas inlet. The gas outlet is labeled "sample out".
The "Service" interface and the programming button are only required for service pur-
poses.
8
7
6
5 S-UVF
4
1
68
multi EA 5100 Sulfur analysis with the S module 5100 (basic, MPO)
9.2 Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
} Place the detection module to the left of the basic module. In the event of a series of
detection modules: place the detection module to the left or right of the others.
} Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
} Connect the detection module with the basic module via the interface:
S-UVF interface on the rear of the detection module
S-UVF interface on the rear of the basic module
} Connect the measuring gas hose of the basic module to the "sample in" gas inlet on
the rear of the module.
} Leave the "sample out" outlet unconnected or connect it to the measuring gas inlet of
the next detection module.
} For detection modules with MPO technology: Install the chemical ozone decomposer
on the rear of the module:
– Fasten the two holding clamps with the supplied screws.
– First press the chemical ozone decomposer into the upper holding clamp and
then into the lower one.
– Connect the hose from the "waste (MPO)" outlet to the ozone decomposer. Do not
pull the hose out of the device!
ü Connection of the detection module is complete.
69
Sulfur analysis with the S module 5100 (basic, MPO) multi EA 5100
9.3 Operation
CAUTION
Risk of respiratory problems due to leaking ozone
If the gas hoses have not been properly connected to the ozone generator, ozone may
leak out of the detection module.
¡ If you can smell ozone, switch off the device and check the connection of the gas
hoses on the ozone generator.
70
multi EA 5100 Sulfur analysis with the S module 5100 coulometric
Expansion of the basic module with the detection module allows the determination of
the sulfur content in solids, liquids, and gases via microcoulometric titration.
The organic sulfur compounds are incinerated into a mixture of sulfur dioxide (SO2) and
sulfur trioxide (SO3) in the basic module. Both oxides are created in a fixed relation. Car-
bon dioxide and water is also created during this incineration.
The amount of SO2 is proportional to the total sulfur amount in the sample.
The measuring gas flow is first dried and then routed to the detection module via a
transfer line. In the measuring cell, the sulfur oxides dissolve in the electrolyte and react
with iodine. This decreases the cell potential.
R-S + O2 → SO2 + SO3 + CO2 + H2O
2 H2O + SO2 + I2 → H2SO4 + 2 HI
R: Hydrocarbon residue
After a specified accumulation time dependent of the sulfur content of the sample, titra-
tion begins. The iodine ions are oxidized back to iodine at the anode. This increases the
cell potential. The end point of the iodometric titration has been reached when the
measuring cell has again reached its original potential.
With titration and end point routine: The electrode reactions consist of an anode reac-
tion (+) and a cathode reaction (-).
Anode (+): 2 I- → I2 + 2 e-
Cathode (-): 2 H+ + 2 e- → H2
10.1.2 Design
71
Sulfur analysis with the S module 5100 coulometric multi EA 5100
8 1
2
6
5 4
72
multi EA 5100 Sulfur analysis with the S module 5100 coulometric
7 3
6
4
Measuring cell The measuring cell is equipped with electrodes for generation and indication. The elec-
trodes are color-coded:
¡ Generation: Yellow anode, red cathode
¡ Indication: Black indicator electrodes
The measuring gas is routed into the measuring cell via the gas inlet tube. The solution
for the end point routine, the sodium sulfate solution (Na2SO3) and the solutions for the
cell measurements are dosed, for example, via the port for manual dosing.
The measuring cell must be filled with approx. 100 ml electrolyte solution (approx. to
the height of the manual dosing port).
73
Sulfur analysis with the S module 5100 coulometric multi EA 5100
7
1
6
3
4
Electrodes A diaphragm is located between the electrodes for generation (anode and cathode). The
diaphragm is only permeable for sulfations. This means that the diaphragm prevent
false analysis results due to undesired compounds.
The electrodes for indication are platinum electrodes.
74
multi EA 5100 Sulfur analysis with the S module 5100 coulometric
6
1
5 3
4
Absorber To clean the measuring gas, two absorbers are installed in the detection module. The
absorbers remove elements from the measuring gas that could disrupt the analysis.
The NOx absorber removes nitrogen oxides (NOx) from the measuring gas. High nitro-
gen oxide content has an influence on the analysis and leads to false results. The filling
of the absorber is normally green. If the color changes to yellow or light brown, the fill-
ing must be replaced.
The HX absorber removes hydrogen halogens (HX with X = Cl, Br, I) from the measuring
gas. Hydrogen halogens disrupt the analysis due to cross-sensitivities. The absorber con-
tains silver wool. The silver wool must be replaced if the color changes from metallic sil-
ver to dark gray.
75
Sulfur analysis with the S module 5100 coulometric multi EA 5100
2 1
Hose diagram Labeled hoses connect the measuring cell with the other components in the detection
module. The numbers in the hose diagram correspond to the labels on the hoses.
71
72
NOx absorber
HX absorber
Measuring cell
73
Mixer
76
multi EA 5100 Sulfur analysis with the S module 5100 coulometric
10.1.3 Connection
The electrical connection and the interface to the basic module can be found on the rear
of the detection module.
The device switch for switching the detection module on and off is located on the top
right of the rear of the device (when viewed from the front). The device fuse and power
connection are located beneath it.
Communication with the basic module is performed via a 9 pin interface cable. The in-
terface on the rear of the device is labeled "S-Coul".
10.2 Installation
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
77
Sulfur analysis with the S module 5100 coulometric multi EA 5100
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
NOTICE
The detection module cannot be operated together with optical detectors.
¡ Do not connect the detection module together with an optical detector.
78
multi EA 5100 Sulfur analysis with the S module 5100 coulometric
8 1
2
6
5 4
79
Sulfur analysis with the S module 5100 coulometric multi EA 5100
10.3 Operation
WARNING
Risk of chemical burns
100 % acetic acid (glacial acetic acid) can cause severe chemical burns. Gases may de-
velop during shaking.
¡ Wear protective clothing when preparing the electrolyte solution. Work under an ex-
tractor.
¡ Observe all instructions and specifications in the safety data sheet.
Performing the end point rou- A sodium sulfite solution (1000 mg/l) is required for the end point routine:
tine
ð Reagents required: 393.9 mg sodium sulfite
} Fill a 100 ml volumetric flask with the specified quantity of sodium sulfite.
} Fill the volumetric flask with ultrapure water up to the calibration mark. Shake the
flask until the sodium sulfite has completely dissolved.
Storage and handling of the sodium sulfite solution:
¡ The solution is sensitive to atmospheric oxygen and may thus be stored in a refriger-
ator for no longer than 1 month.
¡ For the end point routine, only 2 – 10 μl of the master solution are required.
¡ For better metering, diluted solutions can be used.
For the end point routine, the cell potential is set to the optimum operating range of the
measuring cell.
¡ Operating range: 110 to 160 mV
¡ Optimal operation range: 120 mV
80
multi EA 5100 Sulfur analysis with the S module 5100 coulometric
As soon as the cell potential goes outside of the operating range, the software will auto-
matically start an end point routine.
Cell potential is lower than the operating range:
¡ Automatic generation start (iodine generation)
Cell potential is higher than the operating range:
¡ Add sodium sulfite solution when the software instructs.
¡ Titration runs automatically until the cell potential sinks to 120 mV.
An end point routine is also automatically started when electrolyte is replaced.
When working with a MMS, the addition of chloride ions to the measuring cells can be
performed automatically if the indicator value exceeds the operating range while an
analysis sequence is in progress. For this, the user must have prepared a suitable organic
sulfur solution and placed it in the intended position (110) on the sample rack.
Checking the measuring cell Only check the measuring cell if a defect of the detection module is suspected.
A sodium sulfate solution (1000 mg/l) is required for the check of the measuring cell.
ð Reagents required: 1.5482 mg sodium thiosulfate pentahydrate
} Weigh the indicated quantity of sodium thiosulfate into a 100 ml volumetric flask.
} Fill the volumetric flask with ultrapure water up to the calibration mark. Shake the
solution until the salt has dissolved completed.
The solution can be kept for approx. 1 month in a tightly-sealed container.
Standard solutions can be prepared by diluting the solution.
100 μl of the diluted solutions contain the following TS amounts:
¡ 10 mg/l standard: 1 μg S absolute
¡ 100 mg/l standard: 10 μg S absolute
After the function test of the measuring cell with the sodium thiosulfate solution, the
electrolyte must be replaced.
} Place the measuring cell with electrodes and electrolyte solution in the detection
module and connect it to electricity.
} Switch on the basic module and the detection module.
ü The devices boot up. The status LED on the front of the basic module light up in
green after approx. 30 s.
ü The LED on the front of the detection module flashes during the run-in period.
Depending on the detector, the run-in period can take up to 30 min. After this,
the LED will light up continuously. Starting a measurement is only then possible.
} Open the gas supply and set the required gas pressure.
} Switch on the PC.
} Start the control and analysis software and login with your username and password.
} Initialize the analysis system by clicking on [Initialize analyzer].
ü The initialization and automatic detection of all connected components will be
carried out.
} Ready the samples.
} Activate a pre-existing method via the Method |Method - activate menu item.
} Alternatively: Create a new method in the Method |Method - new menu. Select the
measurement parameter in the method. Release and activate the method.
81
Sulfur analysis with the S module 5100 coulometric multi EA 5100
82
multi EA 5100 Carbon analysis with the C module 5100
Expansion of the basic module with the detection module allows determination of the
carbon content in solid, liquid, paste-like, viscous and gaseous samples.
The detection module contains a wide-range NDIR detector. The carbon content in or-
ganic compounds can be determined as TC and EC/OC sum parameters with the detec-
tion module.
Carbon dioxide (CO2) and water (H2O) is created during thermal oxidation of the sam-
ples in the basic module. The gas mixture is dried and transferred to the NDIR detector.
R + O2 → CO2 + H2O
R: Hydrocarbon
The radiation sensor is CO2-sensitive. The double bond between carbon (C) and oxy-
gen(O) has a specific absorption range in the infrared wavelength band.
If a ray of light is projected through a cuvette arrangement, the CO2 content in the mea-
suring gas absorbs a proportion of the total radiation proportional to the CO2 concentra-
tion.
11.1.2 Design
The detection module is used to determine the carbon content in solids, liquids and
gases by measuring the IR IR absorption. All components required for determination are
mounted inside the sealed housing.
The device door is closed tightly and cannot be opened. An LED is installed on the front
of the detection module. The LED flashes during the run-in period and lights up perma-
nently when the module is operational.
83
Carbon analysis with the C module 5100 multi EA 5100
11.1.3 Connection
1
2
3
84
multi EA 5100 Carbon analysis with the C module 5100
11.2 Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
} Place the detection module to the left of the basic module. In the event of a series of
detection modules: place the detection module to the left or right of the others.
} Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
} Connect the detection module with the basic module via the interface:
"C-NDIR" interface on the rear of the detection module
"C-NDIR" on the rear of the basic module.
} Connect the measuring gas hose of the basic module to the "sample in" gas inlet on
the rear of the module.
} Leave the "sample out" outlet unconnected or connect it to the measuring gas inlet of
the next detection module.
ü Connection of the detection module is complete.
85
Carbon analysis with the C module 5100 multi EA 5100
86
multi EA 5100 Carbon analysis with the TOC module 5100
The detection module contains a wide-range NDIR detector. Expansion of the basic
module with the detection module allow determination of the following sum parame-
ters:
Sum parameters Samples Basic module configuration
TC Organic liquids, solids and Vertical/horizontal opera-
gases tion
EC/OC Elemental and organically Horizontal operation
bound carbon from par- with:
ticulate emissions (fine ¡ Special combustion
particulate matter, diesel tube for EC/OC deter-
engine exhaust, smoke mination
gas
TC, TOC, NPOC, TIC Water samples Vertical operation with:
¡ TOC combustion tube
¡ Condensation coil
Carbon dioxide (CO2) and water (H2O) is created during thermal oxidation of the sam-
ples in the basic module. The gas mixture is dried and transferred to the NDIR detector.
R + O2 → CO2 + H2O
R: Hydrocarbon
The radiation sensor used in the TOC module is CO2-sensitive. The double bond between
carbon (C) and oxygen(O) has a specific absorption range in the infrared wavelength
band.
If a ray of light is projected through a cuvette arrangement, the CO2 content in the mea-
suring gas absorbs a proportion of the total radiation proportional to the CO2 concentra-
tion.
A proportion of the water sample is dosed manually into the TIC reactor to determine
the inorganically bound carbon (TIC) content. The sample reacts with phosphoric acid in
the TIC reactor. This creates CO2, which is determined via the NDIR detector.
The detection module determines the CO2 concentration multiple times per second and
generates an integral over time from the signals. The integral is proportional to the car-
bon concentration in the sample.
12.1.2 Design
All components of the module to be operated or serviced by the user can be access via
the door on the front of the module.
The module consists of the following main components:
¡ TIC condenser unit (with TIC reactor, gas-liquid separator, measuring gas dryer)
¡ Condensate pump
¡ Halogen trap and water traps for drying and cleaning the measuring gas.
¡ NDIR detector (in the rear part of the detector)
¡ Indicator and control elements, connections
87
Carbon analysis with the TOC module 5100 multi EA 5100
7
1
2
6
3
4
88
multi EA 5100 Carbon analysis with the TOC module 5100
TOC combustion tube The TOC combustion tube (reactor) is used to determine the TC, TOC and NPOC param-
eters in water samples. The combustion tube is made of quartz glass and is filled with a
catalyst and additives. If the effectiveness of the catalyst decreases, the combustion tube
has to be filled again.
89
Carbon analysis with the TOC module 5100 multi EA 5100
Screw the screw cap with the septum on the top opening of the combustion tube. Con-
nect the condensation coil with the aid of a fork type clamp to the spherical joint on the
lateral outlet.
The oxygen gas supply (hose 3 of the basic module) is connected via a FAST connector
at the lateral outlet directly under the screw cap. The tube holder is used to fasten the
TOC combustion tube in the furnace.
Sampling Water samples are dosed directly into the TOC combustion tube via the injection port
with microliter syringes. A syringe with scale is used for manual dosing. For automatic
sampling via the autosampler, special microliter syringes are used. The syringes have
special dimensions and therefore have no scale. They are unsuitable for manual opera-
tion for this reason. The syringes have a gas connection for analyses in NPOC mode. The
injection volume is: 50 to 500 µl. Optimum measurement results are achieved when 50
to 100 % of the volume of the microliter syringe is used.
The injection port is equipped with temperature-resistant septums with a high penetra-
tion tolerance.
A proportion of the sample is dosed directly into the TIC reactor of the TOC module with
a microliter syringe with a scale to determine the inorganic carbon content (TIC). Only
manual dosing is permitted for this.
Measuring gas drying and For the analysis of water samples, the basic module is equipped with a condensation coil
cleaning made of glass. The condensation coil is connected to the TOC combustion tube via the
spherical joint. Connect hose 80 to the other end of the condensation coil via the the
FAST connector.
The measuring gas cools quickly in the condensation coil and the water vapor con-
denses. The mixture of measuring gas and water is transferred to the TIC reactor in the
detection module via hose 80.
90
multi EA 5100 Carbon analysis with the TOC module 5100
2
1
91
Carbon analysis with the TOC module 5100 multi EA 5100
82 81
84
83 80 85
NDIR
NDIR measuring
gas inlet
86
12.1.3 Connection
The device switch for switching the detection module on and off is located on the top
right of the rear of the device (when viewed from the front). The device fuse and power
connection are located beneath it.
92
multi EA 5100 Carbon analysis with the TOC module 5100
1
2
3
NPOC The connection for the NPOC purging gas is located on the rear of the basic module and
is labeled with "out ABD". The purge hose is connected to the connector via a FAST con-
nector. The purge hose is fastened to the holder on the rear of the detection module
with the provided connector. From there, a hose with an outer diameter (AD) of 1.6 mm
continues to the holder on the sampler. For manual sample preparation, the hose can be
immersed directly in the sample.
TC, EC/OC The detection module can be used to determine TC and EC/OC.
TC determination can be performed in both the multi-purpose combustion tube (stan-
dard tube) and the special EC/OC combustion tube.
93
Carbon analysis with the TOC module 5100 multi EA 5100
12.2 Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
} Place the detection module to the immediate left of the basic module.
} Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
} Connect the detection module to the basic module via the two interfaces:
– 9-pin interface cable
"C-NDIR" interface on the rear of the detection module
"C-NDIR" interface on the rear of the basic module
– 25-pin interface cable
"external (in) interface on the rear of the detection module
"External" interface on the rear of the basic module
} Install the halogen trap and the water traps in the detection module as shown in the
illustration. Connect both with hose 81.
} Install the TIC reactor. Connect the TIC reactor to the water traps. Connect the TIC re-
actor to the condensate pump via hose 86.
} Connect the measuring gas hose (hose 80) from the basic module to the lateral out-
let of the TIC reactor.
Hose 80 is connected to the condensation coil outlet in the basic module later.
} Place the waste canister to the left of the detection module.
} Connect the waste hose to the "waste" outlet on the rear of the detection module.
Route the waste hose into the waste canister.
} Leave the "sample out" outlet unconnected or connect it to the measuring gas inlet of
the next detection module.
94
multi EA 5100 Carbon analysis with the TOC module 5100
7
1
2
6
3
4
Preparing the basic module } Fill the TOC combustion tube as shown in the illustration.
} Insert the TOC combustion tube into the combustion furnace of the basic module.
Connect the oxygen supply (hose 3). Fasten the combustion tube with the tube
holder.
} Install the condensation coil in the basic module. Connect the condensation coil with
the TOC combustion tube with the spherical joint. Secure the spherical joint connec-
tion with a clamp.
} Connect the condensation coil to the TIC reactor in the detection module with hose
80. Route the hose through the side openings of the modules for this.
95
Carbon analysis with the TOC module 5100 multi EA 5100
1
2
3
For NPOC determination } Prepare the detection module and the basic module as previously described.
} Connect the NPOC purge hose (hose 16) to the "out ABD" connection on the basic
module.
} Connect the purge hose with a hose with an outer diameter (AD) of 1.6 mm. Fasten
the hose to the rear on the TOC module.
} Route the hose to the holder on the sampler. Alternatively, use the hose for manual
purging of the acidified samples.
For TC, EC/OC determination The TOC module can be configured for TC determination with the basic module and the
multi-purpose combustion tube or the special combustion tube for EC/OC determina-
tion. For EC/OC determination, the special combustion tube for EC/OC determination
must be used. The measuring gas is routed directly from the combustion tube into the
NDIR detector via the "sample in" outlet.
} Remove hose 82 from the "sample in" measuring gas inlet on the rear of the detec-
tion module.
} Connect the "sample OUT N/S/C" outlet on the basic module with the "sample in"
measuring gas inlet with hose 9.
12.3 Operation
96
multi EA 5100 Carbon analysis with the TOC module 5100
} Start the control and analysis software and login with your username and password.
} Initialize the analysis system by clicking on [Initialize analyzer].
ü The initialization and automatic detection of all connected components will be
carried out.
} Ready the samples.
} Activate a pre-existing method via the Method |Method - activate menu item.
} Alternatively: Create a new method in the Method |Method - new menu. Select the
measurement parameter in the method. Release and activate the method.
} Select Start |Start - Analysis in the menu.
} Select an analysis group or create a new one and confirm via [OK].
} Create an analysis sequence.
} Enter the sample ID for all sample in the Name field.
} Release all sequence lines
} Confirm the entries with [OK].
} Click the [Start Measurement] button.
ü The prepared analysis sequence is processed.
For manual sampling, follow the instructions in the software.
97
Troubleshooting multi EA 5100
13 Troubleshooting
For fault analysis, log files can be recorded. Log file recording should be activated after
consultation with Analytik Jena GmbH customer service for specific faults.
The save location of the log files can be defined via the Extras |Configuration menu
item in the Configuration |Error analysis window.
CAUTION
¡ If faults cannot be remedied by the customer, the Analytik Jena GmbH service de-
partment must always be informed. This also applies for the repeated occurrence of
individual faults.
¡ Send the correspond files to the service department via email for fault diagnosis
(address on inside of the front cover).
Communication problems between the hardware and the software can often be reme-
died by a basic initialization of the measuring system (→ "Initializing the basic module
and the system components" 104).
NOTICE
Communication fault due to wrong USB cable
¡ Use the cable supplied by Analytik Jena GmbH.
¡ Extensions are not permitted for the USB connection!
98
multi EA 5100 Troubleshooting
99
Troubleshooting multi EA 5100
100
multi EA 5100 Troubleshooting
101
Troubleshooting multi EA 5100
102
multi EA 5100 Troubleshooting
103
Troubleshooting multi EA 5100
Performing the basic initializa- Basic initialization is always performed if the Device - edit window was opened and ex-
tion ited via [OK]:
} Select the Device |Device - edit menu item.
} Make any necessary changes and exit the Device - edit window via [OK].
} Click on [Initialize analyzer] in the main window.
ü The system is initialized and the method last used is activated. If the initialization
was successful, the [Start Measurement], [Activate method] and possibly the
[Start calibration] buttons are displayed in the main window.
Standard initialization Click on the [Initialize analyzer] button in the main window. Alternatively, select the Sys-
tem |Initialize menu item.
104
multi EA 5100 Troubleshooting
13.4.1 Overview
In the Status analyzer window, information on the device status or information on indi-
vidual modules is displayed.
Status analyzer multi EA 5100 1
TN liquid vertical (1) - liquid 2
Rack : 112 - Syringe : 50 µL 3
C-NDIR
N-CLD-5100 OK
0.15 4
S-UVFD-5100
Cl-POT
MFC 1 300
MFC 2 0
MFC 3 100
5
105
Troubleshooting multi EA 5100
13.4.2 Method
The following can be displayed in the top line of the Status analyzer window:
Display Description
TN(1) - liquid Example of a method: Name(version) – state
Possible states:
¡ Liquid
¡ Solid
¡ GSS
¡ LPG
¡ AOX, AOX solid
¡ EOX liquid, EOX solid
No method displayed (dis- Device not ready for measurement, no method active:
play empty) Activate method
106
multi EA 5100 Troubleshooting
13.4.4 Detectors
107
Troubleshooting multi EA 5100
End point routine required Indicator value outside the operating range of the
titration cell:
¡ Indicator value greater than 10000: End point
routine starts automatically
¡ Indicator value less than 1000: Start an end point
routine via the System |End point routine menu
item and follow the instructions
Cell temperature exceeds Current cell temperature does not correspond to cell
range temperature configured in the method:
¡ Wait until the desired cell temperature is
achieved
¡ Access the values via the System |Component
test |Chlorine menu item.
108
multi EA 5100 Troubleshooting
109
Troubleshooting multi EA 5100
110
multi EA 5100 Troubleshooting
111
Troubleshooting multi EA 5100
112
multi EA 5100 Troubleshooting
113
Troubleshooting multi EA 5100
114
multi EA 5100 Troubleshooting
115
Troubleshooting multi EA 5100
13.11 Device faults on the S module 5100 basic and S module 5100 MPO
These affect: Detection with the S module 5100 basic and the S module 5100 MPO
Error Possible cause Remedy
Scattering measurements ¡ UV lamp defective ¡ Replace lamp, see (→
"Replacing the UV
lamp" 152)
¡ Sample unsuitable for ¡ Use the horizontal
vertical operation operation mode
mode (droplet forma-
tion)
¡ In vertical operation ¡ Check the position of
mode: Quartz wool the quartz wool, ad-
not present or at the just as necessary
wrong position in the
combustion tube
¡ Sample evaporates ¡ Use a cooled sampler.
before dosing
116
multi EA 5100 Troubleshooting
117
Troubleshooting multi EA 5100
118
multi EA 5100 Troubleshooting
119
Troubleshooting multi EA 5100
120
multi EA 5100 Troubleshooting
13.18 Analytical problems during TC, EC/OC, TOC, NPOC and TIC determination
121
Troubleshooting multi EA 5100
122
multi EA 5100 Troubleshooting
123
Maintenance and care multi EA 5100
124
multi EA 5100 Maintenance and care
125
Maintenance and care multi EA 5100
CAUTION
Risk of injury from falling components
The user can be injured if the combustion tube falls down during maintenance.
¡ Exercise extreme caution when performing maintenance on the combustion tube.
126
multi EA 5100 Maintenance and care
CAUTION
There is a risk of burns at the combustion furnace and combustion tube!
¡ Only perform removal when cold. Allow the device to cool sufficiently.
¡ Wear the heat-proof gloves included in the delivery when handling hot components.
These gloves are suitable for temperatures of up to 200 °C.
Maintenance of the combustion tube is always performed when the furnace is is the ver-
tical installation position. Remove the combustion tube as follows:
} Exit multiWin.
} Switch off the basic module via the power switch and shut off the gas
supply.
} Remove the upper cover of the basic module.
} Open the device door. Flip the toggle switch for the pneumatic seal up.
ü The auto-protection valve assembly is opened.
} Remove hose 3 and hose 4 from the FAST connectors on the combustion
tube.
} Horizontal operating mode:
4 Carefully grip the flame sensor (FS) by the blue ring and pull it off the
FS combustion tube.
The connection to the combustion tube is very fragile!
} Carefully remove the combustion tube from the furnace.
3
} Check the combustion tube for crystallization, cracks and burst spots.
} If the vertical operation mode is in use: Check the condition and position
of the quartz wool plug
127
Maintenance and care multi EA 5100
ð Remove the combustion tube from the combustion furnace (→ "Removing the com-
bustion tube" 127).
} Remove the screw cap and septum, if present. Remove the 2 angled FAST connectors
from the combustion tube.
} For nitrogen and sulfur determination in vertical operation mode: Remove the quartz
wool plug from the combustion tube with a long hook.
Wear protective clothing when replacing the quartz wool (labcoat, protective gloves
and glasses) Wear a respiratory mask or work underneath an exhaust vent because
quartz wool dust irritates the respiratory tracts.
} Clean the inside of the combustion tube with a suitable solvent and a cotton swab or
bottle brush. Rinse with distilled water if the solvent is water-soluble. Otherwise,
rinse with ethanol.
} Dry the combustion tube (e.g., by blowing it through with an inert gas).
} Deposits from incomplete incineration, e.g., soot or solid pyrolysis residue, can also
be removed by firing the combustion tube in a muffle furnace at 750–900 °C or with
a suitable burner flame, e.g., propane.
} For nitrogen and sulfur determination in vertical operation mode: Insert a new
quartz wool plug in the combustion tube (→ "Inserting the quartz wool plug" 128).
ü The combustion tube is clean and can be reinserted.
For nitrogen and sulfur determination in vertical operation mode: Insert a quartz wool
plug in the combustion tube.
Missing quartz wool leads to sooting of the analysis system. Sample with high salt con-
tent form ash and solid oxides during incineration that can build up in the quartz wool.
The quartz wool must then be replaced. The quartz wool plug is not needed for opera-
tion in horizontal mode.
CAUTION
Skin and respiratory system irritation due to quartz wool
Quartz wool tends to form dust. Irritation can occur after breathing in or skin contact
with this dust.
¡ Avoid the formation of dust when working with quartz wool.
¡ Wear protective clothing and gloves.
¡ Work under an extractor or wear a respiratory mask.
NOTICE
Risk of device damage
¡ Only use the pure quartz wool supplied by Analytik Jena GmbH. Contaminated
quartz wool may damage the combustion tube and clog the filter.
¡ Ensure correct positioning of the quartz wool plug. When the plug is not in correct
position the sample does not vaporize evenly.
128
multi EA 5100 Maintenance and care
} Insert the quartz wool plug into the inner tube of the combustion tube
with a clean glass rod.
} Push the quartz wool plug into the tube until the positioning mandrel is
at the center of the plug.
The plug must not seal the bottom slot in the inner tube.
The plug should cover the entire cross section of the inner tube.
} After replacing the quartz wool: Clean the analysis system by performing
at least 3 measurements with a pure solvent (e.g., isooctane, toluene, xy-
lene).
WARNING
Risk of explosion and sooting due to improper connection of the gases to the
combustion tube.
¡ The connections for argon and oxygen on the combustion tube must not be mixed
up!
CAUTION
Risk of burns from hot components and possible damage to the seal of the auto-
protection valve assembly.
¡ Allow the combustion tube to cool before cleaning after incineration.
¡ Allow the combustion furnace to cool before inserting the combustion tube.
NOTICE
Alkaline salts (hand perspiration) cause crystallization in the quartz glass when heating
the combustion furnace. which reduces the service life of the combustion tube.
¡ Wear protective gloves when inserting the combustion tube and do not touch the
tube with bare hands.
¡ Wipe the outside of the combustion tube off with ethanol and a paper towel before
insertion.
129
Maintenance and care multi EA 5100
FS
130
multi EA 5100 Maintenance and care
CAUTION
Risk of injury from falling components
The user can be injured if the assembly falls down during maintenance.
¡ Exercise extreme caution during maintenance of the auto-protection valve assem-
bly.
CAUTION
Risk of burns from the hot furnace and the gas transfer line.
¡ Switch off the device and allow it to cool before performing maintenance.
131
Maintenance and care multi EA 5100
} Hold the auto-protection valve assembly with the left hand and pull at
the knob of the clamping mount with the right hand to open the lock. Re-
move the auto-protection valve assembly from the combustion furnace.
} The installation of the auto-protection valve assembly is in reverse order.
132
multi EA 5100 Maintenance and care
} Remove the housing with the pneumatic seal from the auto-protection
valve assembly.
} Unscrew the connection bushing for hose 11 from the seal housing.
133
Maintenance and care multi EA 5100
} Place the second PTFE disk onto the intermediate ring above the filter.
} Place the pneumatic seal onto the auto-protection valve assembly and
attach it with 4 screws.
ü The auto-protection valve assembly is ready for operation again.
CAUTION
Risk of burns at the hot furnace
¡ Switch off the device and allow it to cool before performing maintenance.
NOTICE
Risk of damage due to pinching or twisting
The sensitive membrane for water vapor exchange in the membrane dryer can be dam-
aged by squeezing or twisting.
¡ Do not pinch the membrane when installing the new membrane dryer.
¡ Do not twist the sensitive connections.
134
multi EA 5100 Maintenance and care
For better clarity, some work steps are displayed with the side panels removed. How-
ever, for the installation and removal of the membrane dryer it is not necessary to re-
move the side panels.
} Move the combustion furnace into the horizontal position.
} Detach the connection of the membrane dryer at the auto-protection
valve assembly. Slightly loosen the knurled screw (1) and pull the con-
nection (2) out toward the bottom.
1
2
} Carefully place the new membrane dryer in 2 strap loops, insert and
clamp it in the holder.
} The gas connection at the top end must point to the right and the gas
connection at the bottom end to the left.
NOTICE! The connections must not be pinched or twisted.
} Reinstall the holder with the new membrane dryer in reverse order.
ü The membrane dryer is installed and ready for operation.
135
Maintenance and care multi EA 5100
Check the hose connections regularly for leaks. Remove and replace faulty hoses and
hose connections. Check the system for leaks after maintenance (→ "Checking the sys-
tem for leaks" 139).
When replacing Fingertight screw connections, observe the following:
¡ Only use straight cut, round and unpinched hose ends for the connection.
¡ Slide the conical nipple with the conical side towards the banjo bolt onto the hose.
¡ The conical nipple and hose end must be flush.
2 3
1
In vertical operation mode, the septum on the injection port of the combustion tube
must be replaced if it is worn out, as this leads to leaks in the system.
Figure 67Replacing the septum on the injection port of the combustion tube
} Open the device door. Flip the toggle switch of the pneumatic seal up to open the
auto-protection valve assembly.
} Remove the cover from the top of the basic module.
} Unscrew the screw cap of the combustion tube.
} Check that the quartz wool plug is still in the proper position in the inner tube of the
combustion tube.
136
multi EA 5100 Maintenance and care
} Insert the septum and screw the screw cap back onto the combustion tube.
ü Replacing the septum on the combustion tube is now complete.
The check valves must be replaced if the gas flow can no longer be set to the target
value (observe the message in the software) and possible leaks in the system have been
excluded. The check valves are located in the valve block on the gas box on the left side
of the device.
} Switch off the basic module and disconnect the power plug from the
power outlet.
} Switch off the gas supply at the shut-off valve.
} Remove the protective ground conductor on the left side panel. Loosen
the 4 screws on the left side panel and lift off the side panel.
} Pull hoses 3 and 4 out of their connections on the valve block (see ar-
rows).
} Unscrew the screw on the valve block with a 2.5 mm hexagon socket
wrench.
} Remove the top part of the valve block and take the check valves out of
the "main" and "inlet" connections.
137
Maintenance and care multi EA 5100
} Replace the sealing rings of the check valves in the upper and lower part
of the valve block.
} Insert new check valves.
} Assemble the valve block and screw on the top part.
} Connect hose 3 to the "main" connection and hose 4 to the "inlet" connec-
tion.
} Connect the protective ground conductor to the side panel and close the
side panel.
} Open the gas supply at the shut-off valve.
} Connect the power cable on the basic module and switch on the module
via the power switch.
ü The basic module is now ready for operation again.
The "Ar" and "O2" gas inlets on the rear of the basic module are equipped with particle fil-
ters. The particle filters and check valves must be replaced if the process gases can no
longer be set to their target values (observe the message in the software) and possible
leaks in the system have been excluded.
} Switch off the basic module and disconnect the power plug from the
power outlet.
} Switch off the gas supply at the shut-off valve.
} When using the ABD: Disconnect the ABD from the basic module and
move the sampling module slightly to access the rear of the device.
} Pull the gas hose out of its connection on the rear of the basic module.
To do so, press down the red ring and pull the gas hose out of the con-
nection.
} Unscrew the gas connections with a 13 mm open-ended wrench.
138
multi EA 5100 Maintenance and care
MFC 0 ml/min (idle state) Oxygen for post-combustion phase, hose 4, gas
2 inlet on the combustion tube, value is set in the
method
MFC 100 to 200 ml/min Pyrolysis gas (argon), hose 4, gas inlet on the
3 (example) combustion tube, value is set in the method
The system tightness for N/S/C methods is monitored automatically. If the system is
leaking, the Gas leak message will appear in the Status analyzer window and the MFC 1
display is marked in red. Starting a measurement is not possible.
WARNING
Risk of chemical burns
Concentrated sulfuric acid is used in the detection module as a drying agent. The con-
centrated acid can lead to severe chemical burns.
¡ Wear protective clothing when working with these hazardous substances.
¡ Completely empty the sulfuric acid container prior to the system tightness check.
¡ Observe all notes and specifications in the safety data sheet.
NOTICE
Risk of destruction of the internal MFM due to corrosive gas
¡ To check the system tightness, only use the flow monitoring set included in the
scope of delivery.
The check of the system tightness is not performed automatically, but instead manually
using the hose set (flow monitoring set) included in the scope of delivery.
139
Maintenance and care multi EA 5100
} First, completely empty the sulfuric acid from the sulfuric acid container
of the Cl module 5100 (→ "Replacing the sulfuric acid and cleaning the
sulfuric acid container." 147). Then clean, dry and re-install the con-
tainer and all adjacent components (safety attachment, gas transfer
hose, connector).
} Measuring cell "high sensitive" (image):
– Remove the gas inlet hose from the measuring cell. Remove the elec-
trolyte from the gas inlet hose. Clean and dry the outside and inside
of the tube and the connector.
– Connect the gas inlet hose to hose 5 of the flow monitoring set.
} Measuring cells "sensitive" and "high concentration":
– Disconnect hose 20 from the combined electrode and connect it to
hose 5 of the flow monitoring set.
} Remove the "MFM in" screw connection on the cover of the control elec-
tronics in the basic module and connect the flow monitoring set (hose 2)
(see arrow).
} Read the current gas flow in the Component test |Flow System |Compo-
nent test menu item) window.
The target flow is the sum total of the measured inlet flows (main + inlet
+ argon bypass). For methods using gas samplers, the auxiliary gas flow
of the gas sampler must be added.
} If the displayed flow differs from the target flow by more than ± 15 ml/
min, search for and remedy the possible causes. Contact the customer
service department if this is not successful.
} After the flow measurements, remove the set and reconnect hose 5 to
the MFM inlet to ensure a complete measuring gas path for N/S/C meth-
ods.
} Refill the sulfuric acid container with sulfuric acid.
Alternatively, the system tightness can also be checked at the transfer line. In this case,
it is not necessary to drain the sulfuric acid, but the check does not cover the entire gas
path. As leaks usually occur on the basic module, this check is a simple and quick alter-
native.
} Open the front doors of the basic module and the Cl module 5100.
} Unscrew the heated transfer line from the connector of the sulfuric acid
container.
} Connect the Fingertight connector of the transfer line to the hose of the
flow monitoring set.
} Then proceed as described before.
140
multi EA 5100 Maintenance and care
System tightness for the gas path from the inlet of the basic module to the outlet of the
TOC module 5100 is not regulated automatically. Use the hose set included in the scope
of delivery and proceed as follows:
ð The basic module and the detection module are switched on and connected.
ð The carrier gas supply is open.
ð The multiWin control and analysis software has been started.
ð A method for TOC determination is active (see software manual).
} Assemble the flow monitoring set (→ "System tightness for Cl methods" 139).
} Connect hose 5 of the flow monitoring set to the "sample out" outlet on the rear of
the detection module.
} Open the doors of the basic module.
} If necessary, disconnect the connection on the "MFM in" connection. The connection
is located on the cover plate of the control electronics on the right in the basic mod-
ule (→ "System tightness for Cl methods" 139).
} Connect the other end of the hose set with the "MFM in" inlet on the basic module.
} Read the current gas flow in the System |Component test menu under Device |Con-
trol flow.
} If the displayed flow differs from the target flow by more than ± 5 ml/min, search for
and remedy the possible causes. Contact the customer service department if this is
not successful.
ü The target flow is displayed in the Status analyzer window:
Target Description
MFC 1 200 ml/min Primary oxygen (hose 3) in the basic module,
value cannot be changed in the method
MFC 2 0 ml/min MFC 2 and MFC 3 are in standby during TOC op-
MFC 3 eration.
WARNING
Risk of electrical shock
¡ Before installing the combustion furnace, switch off the basic device via the power
switch and disconnect the power plug from the power outlet!
CAUTION
Risk of burns at the hot furnace
¡ Switch off the device and allow it to cool before performing maintenance.
141
Maintenance and care multi EA 5100
CAUTION
Risk of injury from falling components
The user can be injured if the combustion furnace falls down during installation/re-
moval.
¡ Exercise extreme caution when installing/removing the combustion furnace.
} Remove the hoses from their holders on the combustion furnace (see ar-
row).
} Remove the protective grounding from the base plate of the combustion
furnace.
142
multi EA 5100 Maintenance and care
} Detach hose 14 (see arrow). Press the ring at the plug connector into the
connector and pull off the hose.
} If necessary, remove the connections on the membrane dryer (for S/N/C
methods).
} Carefully lift the combustion furnace of the basic module.
} The installation of the combustion furnace container is performed in the
reverse order.
WARNING
Risk of electric shock
High voltages are present in the interior of the device, which can lead to electric shock if
contacted.
¡ Before opening: Switch off the device via the power switch.
¡ Disconnect the power cable from the socket.
CAUTION
Risk of burns from the thermal ozone decomposer!
¡ Only perform maintenance in the inside of the device when cold/allow the device to
cool sufficiently.
CAUTION
Risk of respiratory problems due to leaking ozone
If the gas hoses are not properly connected to the ozone generator, ozone can leak out
of the detection module.
¡ Ensure proper hose connection.
¡ Check that the gas connections are sealed with indicator paper after maintenance.
143
Maintenance and care multi EA 5100
144
multi EA 5100 Maintenance and care
Checking the function of the Check the function of the detection module after maintenance via a control measure-
detection module ment.
} Perform a purge measurement with a solvent, e.g., isooctane.
} Analyze a standard solution (5 mg/l TNb) Compare the curve and area with earlier
measurements.
} For resuming measuring operations: Determine the daily factor to test calibration. If
the daily factor is outside of the tolerance range, the analysis system must be re-cali-
brated.
ð Replace the absorber if the analysis baseline has been permanently raised. Replace
the entire absorber (replacement part).
} Remove the screw connection from the absorber. Do not pull the hose out of the de-
vice!
} Pull the absorber out of the holding clamps.
} Press the new absorber into the holding clamps. Reconnect the hose.
ü The detection module is ready for measurement again.
Figure 69Absorber
1 Hose 6 connection 2 Holding clamp
3 Absorber
145
Maintenance and care multi EA 5100
WARNING
Risk of electric shock
High voltages are present in the interior of the device, which can lead to electric shock if
contacted.
¡ Before opening: Switch off the device via the power switch.
¡ Disconnect the power cable from the socket.
CAUTION
Risk of burns from the thermal ozone decomposer!
¡ Only perform maintenance in the inside of the device when cold/allow the device to
cool sufficiently.
ð Replace the entire chemical ozone decomposer annually. Optionally, have the de-
composer replaced by the customer service department.
} Switch off the detection module via the power switch.
} Remove the left side panel. Remove the 4 screws to do this. Remove the protective
grounding cable and remove the side panel.
} Disconnect the following hose connections:
Disconnect hose 25 from the T piece.
Unscrew hose 24 from the bottom of the ozone decomposers.
} Remove the ozone decomposer with its filter and hose 25 from the holding clamps.
Recommendation: Remove it from the top clamp first.
} Install the new filter in the reverse order.
ü The detection module is ready for measurement.
146
multi EA 5100 Maintenance and care
14.11.1 Replacing the sulfuric acid and cleaning the sulfuric acid container.
WARNING
Risk of chemical burns
Concentrated sulfuric acid can cause severe chemical burns!
¡ Before replacing the sulfuric acid: Switch off the gas supply via the software. Risk of
splashing if the gas supply is running.
¡ Wear protective clothing when working on the sulfuric acid container.
¡ Observe all instructions and specifications in the safety data sheet.
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
147
Maintenance and care multi EA 5100
The sulfuric acid absorbs the water that is produced during the combustion process. If
the acid content falls below 85 % the sulfuric acid can no longer sufficiently dry the reac-
tion gas. Chlorine values that are too low are then measured.
7 1
5 2
4
4 3
Figure 71Connection of the heated gas transfer lines to the sulfuric acid container
1 Safety attachment 2 Sulfuric acid container
3 Hose for measuring gas infeed 4 Connector
5 Banjo bolt 6 Heated gas transfer line
7 Measuring gas transfer to the mea-
suring cell with PTFE connector
ð Replace the sulfuric acid daily. High sample throughput may require more frequent
replacement.
} Exit the multiWin software and switch off the analysis system. Switch of the detec-
tion module via the switch on the rear of the device.
} Allow the heated gas transfer line to cool or wear heat-proof gloves when replacing
the sulfuric acid.
CAUTION! Risk of burns at the ends of the heated gas transfer line! The ends can
achieve temperatures of up to 100 °C during operation.
} Remove the banjo bolt from the connector, separating the heated gas transfer line
from the sulfuric acid container.
} Disconnect the PTFE connection and remove hose 20 from the safety attachment.
} Carefully pull the sulfuric acid container with its remaining components out of the
holding clamps toward the top and remove it from the module.
A large beaker is suitable for safe transport and storage (e.g., 500 ml).
} For the "high sensitive" measuring cell: Remove the gas inlet tube from the detection
module with PTFE screw joint and hose 20.
} Remove the safety attachment from the sulfuric acid container.
NOTICE! The bases of the PTFE screw joints remain on the safety attachment, the
hose and the gas inlet tube.
148
multi EA 5100 Maintenance and care
} Unscrew the connectors for the heated gas transfer line from the sulfuric acid con-
tainer. Remove the thin hose from the container.
CAUTION! Sulfuric acid residue can still be present on the hose.
} Empty the sulfuric acid via the top opening. Dispose of the sulfuric acid.
} Rinse the sulfuric acid container and the safety attachment multiple times with ultra-
pure water and then rinse with ethanol or methanol:
} Rinse hose 20 including the PTFE screw joint with ultrapure water and then with
ethanol or methanol.
} Dry the cleaned components, e.g., by blowing it through with an inert gas.
} Place the sulfuric acid container down somewhere safe and fill it with 20 ml of con-
centrated sulfuric acid.
} Install the filled sulfuric acid container in the detection module in reverse order (→
"Installation" 59).
NOTICE! Ensure the proper fit of the conical nipples when connecting the gas trans-
fer line and the PTFE connectors.
ü The detection module is ready for operation again.
WARNING
Risk of chemical burns
The electrolyte solution contains high concentrations of acetic acid.
¡ Wear protective clothing when replacing the electrolyte solution.
¡ Observe all instructions and specifications in the safety data sheet.
ð For the "sensitive" and "high concentration" measuring cells: Replace the electrolyte
solution daily.
ð For the "high sensitive" measuring cell: Refill the electrolyte solution daily.
Only replace the electrolyte solution in the event of analytical problem and if crys-
talline deposits form.
} For replacement of the electrolyte: Empty the measuring cell. Dispose of the elec-
trolyte solution.
} Rinse the empty measuring cell and the magnetic stirrer first with ultrapure water,
then with ethanol.
} Carefully wipe the measuring cell and the magnetic stirrer with tissue to remove any
possible silver chloride residue.
} Fill the measuring cell with fresh electrolyte solution:
"high sensitive" measuring cell: 65 ml
"sensitive" measuring cell: 15 to 20 ml
"high concentration" measuring cell: 120 ml
ü The measuring cell is ready for operation again.
Also observe the following:
¡ If the detection module is put out of operation for multiple days, clean the measur-
ing cell and store it in a dry place.
¡ Check the casing of the magnetic stirring rod for cracks regularly. If metal ions from
the stirring rod come into contact with the electrolyte solution, they can disrupt the
analysis.
149
Maintenance and care multi EA 5100
¡ Due to risk of short-circuits: Prevent fluids from entering the stirring/cooling block
and the plug-in contacts.
Combined electrode
NOTICE
Risk of possible destruction of the electrode due to cleaning, abrasive or polish-
ing agents.
The combined electrode consists of ceramic materials which is mechanically sensitive es-
pecially near the fused electrodes.
¡ Only rinse the combined electrode with ethanol and ultrapure water to clean it.
When handled incorrectly, the electrical connection of the combined electrode can
break.
¡ Carefully remove the electrode from the lid of the measuring cell.
¡ Grip the electrode from above and pull it straight out of the lid.
¡ Do not pull on or jerk the side connection sleeve for the electrical connections. The
connections in the sleeve will break otherwise (not visible from outside)!
150
multi EA 5100 Maintenance and care
¡ Do not touch the measuring cell and the electrode during operation (during a mea-
surement or an end point routine). The measurement result will be falsified other-
wise.
¡ The generator anode is located at the bottom of the measuring cell in the form of
the stable silver plate (silver circle). The silver electrode wears out with increasing
use. If necessary, the entire measuring cell must be replaced.
Sensor electrode
NOTICE
Risk of damage to the sensor electrode
The sensor pin and the gold contact of the sensor electrodes are sensitive to touch.
¡ Apply scratch protection to the sensor pin for storage.
¡ Rinse the sensor pin with ultrapure water before use or to clean it. Do not touch it
again after this. Do not dry the pin or wipe it dry!
¡ Before use or to clean it, wipe the gold contact with a cloth and some ethanol. Do
not touch it again after this.
Reference electrode
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
¡ Check the condition and filling level of the bridge electrolyte weekly. The bridge
electrolyte must be clear and free of deposits or other particles.
¡ If necessary: Refill or replace the electrolyte solution.
Storing the reference electrode:
¡ The reference electrode can be stored in the measuring cell for several days if it is
sufficiently filled with electrolyte solution.
¡ For storage of less than 1 month: Close the refill opening. Apply the empty protec-
tive cap to the electrode.
¡ For storage longer than 1 month: Empty the bridge electrolyte completely via the re-
fill opening with a syringe or disposable pipette. Close the refill opening. Apply the
empty protective cap to the electrode.
Putting the reference electrode back into operation after longer storage:
¡ Rinse the inside of the electrode with approx. 2 ml electrolyte solution.
¡ Fill the electrode with electrolyte solution up to the refill opening.
¡ Fill the measuring cell with electrolyte solution. Insert the electrode in the measur-
ing cell and allow the solution to stir for several hours in the detection module. Do
not connect the electrode to its electrical connection for this.
151
Maintenance and care multi EA 5100
Platinum electrode The platinum electrode is maintenance-free. The salt bridge of the platinum electrode
has a diaphragm. The electrolyte solution must not be allowed to crystallize inside the
diaphragm, otherwise it can clog. Remove the salt bridge and rinse with sufficient dis-
tilled water for longer storage intervals.
The silver electrode in the "high Wipe the silver surface with cellulose after use. Otherwise, the electrode is mainte-
sensitive" measuring cell nance-free. The silver electrode can show discoloration from longer use.
14.12 Maintenance of the S module 5100 basic and S module 5100 MPO sulfur
detector
WARNING
Risk of electric shock
High voltages are present in the interior of the device, which can lead to electric shock if
contacted.
¡ Before opening: Switch off the device via the power switch.
¡ Disconnect the power cable from the socket.
WARNING
Risk due to UV radiation
The UV lamp emits UV radiation that can damage eyes and skin.
¡ Before opening the detection module: Switch off the device via the power switch.
CAUTION
Risk of burns
The UV is hot directly after operation.
¡ Allow the lamp to cool before maintenance.
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
152
multi EA 5100 Maintenance and care
NOTICE
Contamination reduces the effectiveness of the UV lamp.
¡ Do not touch the glass of the new lamp. In particular, protect the radiation emission
point made of quartz glass.
¡ If the glass is touched, wipe it with a lint-free cloth and pure alcohol.
153
Maintenance and care multi EA 5100
} Place the lamp in the proper position when inserting: The pin on the
holder must fit into the groove of the lamp body.
} Fasten the new lamp with the 2 screws.
} Plug the connection plug back into the socket all the way.
} Fasten the side panel again.
ü The detection module is ready for operation again.
154
multi EA 5100 Maintenance and care
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
CAUTION
Skin and respiratory system irritation due to quartz wool
Quartz wool tends to form dust. Irritation can occur after breathing in or skin contact
with this dust.
¡ Avoid the formation of dust when working with quartz wool.
¡ Wear protective clothing and gloves.
¡ Work under an extractor or wear a respiratory mask.
ð Check the absorber once per week. Replace the filling if necessary.
ð Replace the filling of the NOx absorber if the color has turned from light green to
yellow or light brown.
ð Replace the filling of the HX absorber if the color has turned from metallic silver to
dark gray.
} Remove the hose connection of the absorber tube.
} Remove the absorber tube from the clamps.
} Disconnect the FAST connector from the tube on one side. Remove the quartz wool
plug.
} Remove the use filling from the tube.
} Fill fresh absorber material (silver wool for the HX absorber, ammonium iron(II) sul-
fate for the NOx absorber) into the tube. Reinsert the quartz wool plug. Connect the
FAST connector.
} Carefully place the absorber tube back in the clamps.
} Connect the hoses to the absorber tube.
ü The detection module is ready for operation again.
155
Maintenance and care multi EA 5100
2 1
156
multi EA 5100 Maintenance and care
8 1
2
6
5 4
157
Maintenance and care multi EA 5100
} Fill the measuring cell with approx. 100 ml of fresh electrolyte solution (approx up to
the manual dosing port). For creation of the electrolyte solution, see (→ "Preparing
the measuring cell" 80).
} Carefully place the stirring rod back in the measuring cell.
} Reinsert the electrodes in the measuring cell. Connect the electrodes to the "Genera-
tion" and "Indication" connections.
} Place the measuring cell in the holder on the magnetic stirrer.
} Insert the gas inlet tube into the measuring cell. Connect the gas inlet tube with the
HX absorber via hose 72.
} Activate the magnetic stirrer (set to stage 3 approx.).
NOTICE! The magnetic stirring rod can damage the electrodes if the rotation fre-
quency is set too high. Operate the rotary switch with care.
} Wait for approx. 5 min for the new electrolyte to collect in the salt bridge of the gen-
erator electrodes and for the measuring cell to establish equilibrium. Before starting
any measurement, the end point routine must be performed.
ü The detection module is ready for operation again.
NOTICE
Risk of gas leaks
System tightness for the gas path from the inlet of the basic module to the outlet of the
detection module is not regulated automatically.
¡ Always check the system tightness after performing maintenance on the detection
module (→ "System tightness for TOC methods" 141).
158
multi EA 5100 Maintenance and care
81
1
2
3
4
CAUTION
Skin and respiratory system irritation due to quartz wool
Quartz wool tends to form dust. Irritation can occur after breathing in or skin contact
with this dust.
¡ Avoid the formation of dust when working with quartz wool.
¡ Wear protective clothing and gloves.
¡ Work under an extractor or wear a respiratory mask.
159
Maintenance and care multi EA 5100
NOTICE
Risk of device damage due to aggressive combustion products
When the copper wool is used up, aggressive combustion products can damage the opti-
cal and electronic components of the detection module.
¡ Replace the complete filling of the halogen trap as soon as half of the copper wool
has turned black.
82 81
160
multi EA 5100 Maintenance and care
WARNING
Risk of chemical burns
The TIC reactor is regenerated and cleaned with 40 % phosphoric acid. Phosphoric acid
can irritate eyes, skin and mucous membranes.
¡ Wear protective clothing when handling the concentrated acid.
¡ Observe all notes and specifications in the safety data sheet.
NOTICE
Risk of leaks
A cannula that is too large will damage the septum on the septum port.
¡ Only use cannula with an outer diameter of 0.63 mm for the septum port.
ð For TIC determination or TIC determination in differential mode: The TIC reactor
must be regenerated daily. How often this must be done is dependent on the TIC
content of the samples. For high TIC content, regenerate the TIC reactor more than
once daily.
ð Regeneration of the TIC reactor is also required after longer times of inactivity.
ð If only the TC or NPOC mode is used, regeneration of the TIC reactor is not required.
} Select the System |Component test menu.
} Select Regeneration TIC reactor from the list field in the Device tab.
} Click the Regeneration TIC reactor button.
} After prompt via software: Add 40 % phosphoric acid to the TIC reactor via the sep-
tum port with the supplied 5 ml syringe.
ü The TIC reactor is drained off and purged.
WARNING
Risk of chemical burns
The TIC reactor is regenerated and cleaned with 40 % phosphoric acid. Phosphoric acid
can irritate eyes, skin and mucous membranes.
¡ Wear protective clothing when handling the concentrated acid.
¡ Observe all notes and specifications in the safety data sheet.
161
Maintenance and care multi EA 5100
ð Check the TIC reactor for deposits and cracks quarterly. Cleaning is only required if
TIC samples are no longer properly purged.
} Undo the connection between the TIC reactor and the water traps.
} Remove the 2 knurled screws on the cover of the cooling block. Remove the cover.
} Remove the FAST connector with hose 80 from the side outlet of the TIC reactor.
} Remove the waste hose (hose 86) to the condensate pump from the connection on
the bottom of the TIC reactor.
} Remove the TIC reactor from the detection module and check it for deposits and
cracks.
} Rinse the TIC reactor with ultrapure water.
} Reinstall the TIC reactor in the detection module is reverse order.
} Check the system for leaks.
ü The detection module is ready for operation again.
7
1
2
6
3
4
162
multi EA 5100 Maintenance and care
WARNING
Risk of chemical burns
Residue of 40 % phosphoric acid can still be found in the pump hose. Phosphoric acid
can irritate eyes, skin and mucous membranes.
¡ Wear protective clothing when handling the concentrated acid.
¡ Observe all notes and specifications in the safety data sheet.
ð Check the condensate pump for leaks every 3 months. If any liquid is escaping from
the pump hose, replace the pump hose.
ð When the pump body and the roller carrier are heavily damaged, they must be re-
placed. Inform customer service for this.
} Press the bracket on the condensate pump to the left.
} Remove hoses 85 and 86 from the pump's connections.
} Remove the guide piece with the pump hose from the pump body.
} Check the pump hose and the connections on excessive wear and cracks.
} Wipe the pump body and roller carrier with ultrapure water.
} Check the pump body and roller carrier for wear.
} Push the intact or new pump hose back into the guide piece.
During installation, the hose clamps must be turned downwards.
} Push the hose guide into the groove on the guide piece.
} Position the guide piece with the hose around the pump body again. Press the guide
piece down with one hand to do this. Move the bracket to the right until it engages
with the other hand.
} Press hoses 85 and 86 back onto the metal connections of the pump hose.
} Check the system for leaks.
ü The detection module is ready for operation again.
163
Maintenance and care multi EA 5100
CAUTION
Risk of burns at the hot furnace
¡ Switch off the device and allow it to cool before performing maintenance.
164
multi EA 5100 Maintenance and care
CAUTION
Risk of burns at the hot furnace
¡ Switch off the device and allow it to cool before performing maintenance.
CAUTION
Skin and respiratory system irritation due to quartz wool
Quartz wool tends to form dust. Irritation can occur after breathing in or skin contact
with this dust.
¡ Avoid the formation of dust when working with quartz wool.
¡ Wear protective clothing and gloves.
¡ Work under an extractor or wear a respiratory mask.
ð If the catalyst loses effectiveness, the combustion tube must be refilled. A check has
to be performed after the maintenance interval has elapsed (max. 1500 injections).
A software message will indicate when the maintenance interval has elapsed.
ð The service life of the catalyst depends largely on the samples. On average, approx.
1500 injections can be performed, more are sometimes possible. The service life can
be lower with especially loaded samples, especially high salt content.
} Switch off the basic modules via the power switch and allow the device to cool.
} Switch off the gas supply and disconnect the power plug from the power outlet.
} Open the doors of the basic module and remove the top cover.
} Remove the FAST connector with hose 3 from the combustion tube.
} Unscrew the knurled head screw at the fork clamp and remove the fork clamp con-
necting the outlet of the combustion tube to the condensation coil.
} Disconnect the spherical joint. The condensation coil remains in the basic module.
} Remove the tube holder for holding the combustion tube.
} Carefully pull the combustion tube out of the combustion furnace toward the top.
} Remove the screw cap with septum from the combustion tube.
} Remove the used catalyst filling.
} Check the combustion tube for heavy crystallization, cracks and burst spots. Only re-
use intact combustion tubes.
} Rinse the combustion tube with ultrapure water and allow it to dry.
ü The TOC combustion tube is clean.
NOTICE
Risk of devitrification of the quartz glass due to sweat
Alkaline salts in sweat from hands leads to devitrification of quartz glass when heated.
Devitrification reduces the service life of the combustion tube.
¡ Only touch the cleaned combustion tube when wearing gloves.
165
Maintenance and care multi EA 5100
} Insert approx. 500 mg of quartz glass wool in the combustion tube. Carefully press
the glass wool down to a height of approx. 1 cm with a glass rod. Do not press it
down too tightly.
} Carefully pour 16 g of platinum catalyst onto the quartz wool (fill height approx
4 cm).
} Cover the catalyst completely with approx 250 mg of quartz glass wool. Carefully
tamp down the quartz wool.
} Fill approx. 10 g of ground quartz glass into the combustion tube (fill height approx.
2 cm).
} Cover the ground quartz glass with a piece of high temperature fiber mat (HT mat)
(layer height approx 1 cm).
} Close the filled TOC combustion tube with septum and screw cap and reinstall it in
the furnace in reverse order.
} Check the analysis system for leaks.
ü The analysis system is ready for operation again.
1
2
3
NOTICE
The catalyst may emit gas during first heating, this can be seen as mist formation in the
TIC reactor.
¡ Allow the catalyst to normalize during first heating for approx. 30 min until mist no
longer forms. Disconnect the gas path between the TIC reactor and the water traps
for this.
166
multi EA 5100 Transport and storage
15.1 Transport
When transporting the device, observe the safety instructions in the "Safety instructions"
section.
Avoid the following during transport:
¡ Impact and vibration
Risk of damage due to shock, impact or vibration!
¡ Large temperature fluctuations
Risk of condensation!
CAUTION
Risk of injury during transport
Dropping the device poses a risk of injury and damage to the device.
¡ Proceed carefully when moving and transporting the device. Two persons are re-
quired to lift and carry the device.
¡ Grip the device firmly at the bottom with both hands and lift it simultaneously.
Observe the following when moving the device within the laboratory:
¡ Insufficiently secured components pose a risk of injury!
Before moving the device, remove all loose parts and disconnect all connections
from the device.
¡ For safety reasons, two persons are required to transport the device, one person on
each side of the device.
¡ As the device does not have carrying handles, grip the device firmly with both hands
at the lower end. Lift the device simultaneously.
¡ Observe the guide values and adhere to the legally mandated limits for lifting and
carrying loads without auxiliary means.
¡ Observe the installation conditions at the new location.
15.3 Storage
NOTICE
Risk of device damage due to environmental conditions
Environmental influences and condensation can destroy individual components of the
device.
¡ Only store the device in air-conditioned rooms.
¡ Ensure that the atmosphere is free of dust and corrosive vapors.
167
Transport and storage multi EA 5100
If the device is not installed immediately after delivery or not required for longer periods,
it should be stored in its original packaging. A suitable desiccant should be added to the
equipment to prevent damage from moisture.
The requirements for the climatic conditions of the storage location can be found in the
specifications.
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
168
multi EA 5100 Transport and storage
WARNING
Risk of chemical burns due to acid and electrolyte solution return
Concentrated sulfuric acid is used in the Cl module 5100, phosphoric acid in the TOC
module 5100 and a slightly acidic electrolyte solution in the S module 5100 coulomet-
ric.
A vacuum can be created in the analysis system when the combustion furnace is cooling.
This vacuum can lead to acid being drawn all the way into the auto-protection valve as-
sembly via connection hoses and lines.
¡ For the Cl module 5100: Only switch off the basic module and the gas supply when
the analysis system has cooled. The argon safety bypass on the auto-protection
valve assembly prevents a vacuum from being created during cooling for the chlo-
rine analysis branch. Alternatively: Remove the hose connection between the basic
module and the detection module before cooling.
¡ For the TOC module 5100 and the S module 5100 coulometric: Before shutting
down the basic module via the software, disconnection the hose connection to the
detection module.
} Switch off the detection module via the power switch. Disconnect the power plug
from the power socket.
} Shut off the gas supply.
} Disconnect all connections on the rear of the detection module.
} Seal the open gas connections with the ends of a short, fitting piece of hose to pre-
vent contamination during transport.
} Remove loose components such as the absorber from the rear of the detection mod-
ule and package them individually.
} If the detection module can be opened via the front door, remove all moving compo-
nents and package them individually. Observe the notes for individual detection
modules for this.
} Carefully package the detection module and the accessories (cables, glass compo-
nents, hoses, clamps) in the original packaging.
} Add a desiccant to the package to prevent moisture damage.
WARNING
Risk of chemical burns
The concentrated sulfuric acid used as a drying agent and the acetic acid electrolyte so-
lution can cause severe chemical burns.
¡ Wear protective clothing when working on the sulfuric acid container and the mea-
suring cell.
¡ Observe all instructions and specifications in the safety data sheets.
} Disconnect the electrodes (and the measuring cell) from the electrical connections on
the inside of the rear panel of the detection module.
} Disconnect the measuring gas hose from the gas inlet tube/from the combined elec-
trode. Ensure that the seals of the PTFE screw connection are not lost during trans-
port.
169
Transport and storage multi EA 5100
} Disconnect the measuring cell from the exhaust hose and remove the measuring cell.
Empty the measuring cell.
} Disconnect the gas transfer line from the sulfuric acid container.
} Remove the sulfuric acid container from the detection module. Empty and rinse the
container (→ "Replacing the sulfuric acid and cleaning the sulfuric acid container."
147).
} Put all electrodes into their original packaging. Observe the instructions on mainte-
nance and care of the electrodes (→ "Maintenance and storage of the electrodes"
150).
} Clean the measuring cell with distilled water and ethanol. Carefully wipe out/off the
measuring cell and the magnetic stirring rod with tissue.
} Fasten the exhaust hose in the detection module, e.g., with adhesive strips.
} Disconnect the electrode cables from the "Generation" and "Indication" connections.
} Disconnect the hose connections to the HX and NOx absorbers and remove both ab-
sorbers from the module.
} Remove the measuring cell from the module.
} Remove the electrodes and the gas inlet tube from the measuring cell.
} Empty the measuring cell, remove the magnetic stirring rod and rinse both with ul-
trapure water.
} Package the glass components and the electrodes in their original packaging.
WARNING
Risk of chemical burns
The TIC reactor can contain 40 % phosphoric acid residue. Phosphoric acid can irritate
eyes, skin and mucous membranes.
¡ Wear protective clothing when emptying and cleaning the TIC reactor.
¡ Observe all notes and specifications in the safety data sheet.
} Remove the halogen trap and the water traps from the detection module.
} Remove the TIC reactor from the detection module and rinse it out.
} In the basic module: Remove the condensation coil and the TOC combustion tube af-
ter the furnace has cooled.
CAUTION! Risk of burns around the hot furnace!
} Package all components and connection hoses in their original packaging.
170
multi EA 5100 Disposal
16 Disposal
When the respective service life has expired, the multi EA 5100 basic module, the detec-
tors and sampling modules must be disposed of as electronic waste in accordance with
the applicable regulations.
Waste water containing acids and sample material is generated during device operation.
Dispose of the neutralized waste in accordance with the legal requirements.
Cl module 5100 electrodes The metals used for the electrodes (platinum, silver) must not be allowed to contami-
nate the sewage systems, the surface or ground water or the soil. Dispose of the elec-
trodes in accordance with applicable regulations on hazardous waste.
The chemical ozone decom- The chemical ozone decomposer contains metal oxides. The absorber is filled with active
poser on the absorber of the carbon and soda lime. The used cartridges must be disposed of in accordance with local
N module 5100 or S module regulations.
5100 MPO
TOC module 5100 The TOC combustion tube contains a platinum catalyst. Dispose of the used catalyst in
consumables accordance with local regulations. Analytik Jena GmbH will accept the special catalyst
back for disposal. Please contact the customer service department (see inside front
cover).
The halogen trap contains copper. Contact the responsible institution (authority or
waste disposal company). There you will receive the information regarding recycling or
disposal.
171
Specifications multi EA 5100
17 Specifications
Sampling (vertical) TS, TN, TX, TC Direct injection of liquids into the multi-purpose combus-
tion tube via injection port with septum
Direct injection of gases with a special, long injection nee-
dle into the multi-purpose combustion tube via injection
port with septum
EOX Direct injection of an extract into the multi-purpose com-
bustion tube via injection port with septum
TC, NPOC Direct injection of aqueous samples into the TOC combus-
tion tube via injection port with septum
TIC Direct injection of aqueous samples into the TIC reactor
via injection port with septum
Sampling (horizontal) TS, TN, TX, TC Injection of liquids via injection port with septum (ABD) in
quartz glass boats or direct transfer of solid samples in
quartz glass boats into the multi-purpose combustion tube
Direct injection of liquids via injection port with septum
into the multi-purpose combustion tube
Injection of gaseous samples with a special, flexible injec-
tion cannula via injection port with septum (ABD) into the
multi-purpose combustion tube
AOX Transfer of the loaded active carbon with a quartz con-
tainer (column method, max. 18 x 6 mm columns) in the
quartz glass boat into the multi-purpose combustion tube
172
multi EA 5100 Specifications
Sample volume TS, TN, TX, TC, Liquids 1 to 100 µl (horizontally with ABD)
TOC 1 to 500 µl (vertically with MMS or TOC
module 5100 and direct manual dosing)
Solids 0.001 to 110 mg
Non-pressurized 1 to 100 ml
gases
Pressurized gases 1 to 20 ml (with GSS/LPG combi mod-
ule)
1 to 100 ml (with GSS module and GSS
adapterbox)
LPG 1 to 50 µl
EOX (extract) 10 to 100 µl
TOC (aqueous samples) 10 to 500 µl
173
Specifications multi EA 5100
174
multi EA 5100 Specifications
175
Specifications multi EA 5100
Procedural data Analytical parameters AOX, EOX, TX, TCl, TOX, TIX
Detection principle Micro-coulometric end point titration
(argentometry)
Sample feed ¡ In carrier gas stream (from basic
module)
¡ Direct injection of aqueous samples
and HCl for test purposes
(into the measuring cell)
Temperature control for measuring cell Integrated cooling
176
multi EA 5100 Specifications
177
Specifications multi EA 5100
178
multi EA 5100 Specifications
Procedural data Analytical parameters Total carbon TC, TIC, TOC, NPOC, EC/OC
Sample feed for TIC analysis Direct injection in the TIC reactor
Detection principle NDIR (non-dispersive infrared absorp-
tion
spectrometry)
Measuring range (C in aqueous samples) 0.2 to 10000 mg/l C
Measuring range (C in organic samples) 0.1 to 10000 mg/l C
Measuring range (C absolute) 0 to 500 mg C
179
Specifications multi EA 5100
Protection class and protection The device is protection class I. The housing is protection type IP 20.
type
Device safety The device complies with the following safety standards
¡ EN 61010-1
¡ EN 61010-2-081
¡ EN 61010-2-010
EMC compatibility The device has been checked for transient emissions and noise immunity.
It meets the requirements for transient emissions according to
¡ EN 61326-1 (EN 55011 group 1, class B)
The device meets the requirements for noise immunity according to
¡ EN 61326-1 (requirements for use in a basic environment)
Environmental compatibility The device has been tested for environmental compatibility and meets the requirements
according to
¡ ISO 9022-2
¡ ISO 9022-3
Guidelines for China The device contains substances subject to regulation (according to the directive GB/T
26572-2011). Analytik Jena guarantees that, if the device is used as intended, these
substances will not leak within the next 25 years and therefore will not pose a threat to
the environment or health within this time period.
180
multi EA 5100 Table of figures
Table of figures
Figure 1 Operating modes of the multi EA 5100 ............................................................................................................ 17
Figure 2 Front view in vertical operation mode ................................................................................................................ 19
Figure 3 Basic module in horizontal operation mode....................................................................................................... 19
Figure 4 Basic module with a sampling module and a detection module....................................................................... 20
Figure 5 Interfaces on the rear of the device .................................................................................................................... 21
Figure 6 Power connection, power switch......................................................................................................................... 21
Figure 7 Interfaces for detector modules and sampling modules.................................................................................... 22
Figure 8 Connection for the sensors and the combustion furnace in the device ............................................................ 22
Figure 9 Connections for the auto-protection valve assembly and transfer line ............................................................ 23
Figure 10 Hose diagram for horizontal operation .............................................................................................................. 24
Figure 11 Hose diagram for vertical operation ................................................................................................................... 25
Figure 12 Gas connections at the rear of the device........................................................................................................... 26
Figure 13 Gas connections on the gas box .......................................................................................................................... 27
Figure 14 Connections on the combustion tube ................................................................................................................. 27
Figure 15 Combustion furnace in the vertical and horizontal operating mode ................................................................ 28
Figure 16 Multi-purpose combustion tube.......................................................................................................................... 28
Figure 17 Membrane dryer................................................................................................................................................... 29
Figure 18 Space required for the basic device and modules (vertical operation) ............................................................. 32
Figure 19 Space required for the basic device and modules (horizontal operation) ........................................................ 33
Figure 20 Media connections of the rear of the device ...................................................................................................... 35
Figure 21 Basic module with detection module and sampling module............................................................................. 43
Figure 22 Rear of the nitrogen detector .............................................................................................................................. 44
Figure 23 Absorber ............................................................................................................................................................... 45
Figure 24 Connecting detection modules to the basic module .......................................................................................... 46
Figure 25 Design of the chlorine detector (without measuring cell) ................................................................................. 49
Figure 26 Measuring cell "sensitive" with lid........................................................................................................................ 50
Figure 27 Equipped measuring cell...................................................................................................................................... 50
Figure 28 Combined electrode ............................................................................................................................................. 51
Figure 29 Measuring cell "high concentration" with lid....................................................................................................... 51
Figure 30 Measuring cell "high sensitive" with lid ............................................................................................................... 52
Figure 31 Lid of the "high sensitive" measuring cell equipped with all electrodes ............................................................ 53
Figure 32 Reference electrode and sensor electrode.......................................................................................................... 54
Figure 33 Exhaust connection.............................................................................................................................................. 54
Figure 34 Platinum electrode with salt bridge .................................................................................................................... 55
Figure 35 Silver electrode ..................................................................................................................................................... 55
Figure 36 Rear of the chlorine detector............................................................................................................................... 56
Figure 37 Measuring cell connections ................................................................................................................................. 57
Figure 38 Connection of the heated gas transfer lines to the sulfuric acid container ...................................................... 57
Figure 39 Gas infeed tube with PTFE connector................................................................................................................. 58
Figure 40 Gas infeed to the "sensitive" measuring cell ....................................................................................................... 58
181
Table of figures multi EA 5100
Figure 41 Connection of the heated gas transfer lines to the sulfuric acid container ...................................................... 60
Figure 42 Connection of the gas transfer line in the basic module ................................................................................... 61
Figure 43 Basic module with detection module and sampling module............................................................................. 67
Figure 44 Rear of the sulfur detector................................................................................................................................... 68
Figure 45 Chemical ozone decomposer ............................................................................................................................... 69
Figure 46 Coulometric sulfur detector with measuring cell (without door) ...................................................................... 72
Figure 47 Coulometric sulfur detectors without measuring cell......................................................................................... 73
Figure 48 Coulometric measuring cell ................................................................................................................................. 74
Figure 49 Generator and indicator electrodes, gas inlet tube ............................................................................................ 75
Figure 50 NOx absorber and HX absorber........................................................................................................................... 76
Figure 51 Hose diagram ....................................................................................................................................................... 76
Figure 52 Rear of the coulometric sulfur detector .............................................................................................................. 77
Figure 53 Coulometric sulfur detector with measuring cell (without door) ...................................................................... 79
Figure 54 Basic module with detection module and sampling module............................................................................. 83
Figure 55 Rear of the carbon detector................................................................................................................................. 84
Figure 56 TOC detector, door opened.................................................................................................................................. 88
Figure 57 Components in the basic module ........................................................................................................................ 89
Figure 58 TOC combustion tube (unfilled) .......................................................................................................................... 89
Figure 59 Tube holder for the TOC combustion tube ......................................................................................................... 90
Figure 60 Condensation coil ................................................................................................................................................. 91
Figure 61 Hose diagram of the TOC detector...................................................................................................................... 92
Figure 62 Rear of the TOC detector ..................................................................................................................................... 93
Figure 63 TOC detector, door opened.................................................................................................................................. 95
Figure 64 TOC combustion tube........................................................................................................................................... 96
Figure 65 Status analyzer window ......................................................................................................................................105
Figure 66 Replacing the Fingertight connections ...............................................................................................................136
Figure 67 Replacing the septum on the injection port of the combustion tube ...............................................................136
Figure 68 Replacing the ozone generator ...........................................................................................................................144
Figure 69 Absorber ...............................................................................................................................................................145
Figure 70 Replacing the chemical ozone decomposer........................................................................................................147
Figure 71 Connection of the heated gas transfer lines to the sulfuric acid container ......................................................148
Figure 72 Correct handling of the combined electrode ......................................................................................................150
Figure 73 Replacing the chemical ozone decomposer........................................................................................................154
Figure 74 NOx absorber and HX absorber...........................................................................................................................156
Figure 75 Coulometric sulfur detector with measuring cell (without door) ......................................................................157
Figure 76 Replace the water traps .......................................................................................................................................159
Figure 77 Replacing the halogen trap .................................................................................................................................160
Figure 78 TOC detector, door opened..................................................................................................................................162
Figure 79 Installing the pump hose in the guide piece.......................................................................................................163
Figure 80 Components in the basic module ........................................................................................................................164
Figure 81 TOC combustion tube...........................................................................................................................................166
182