1741 08
1741 08
CALIBRATION
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 10:
I. Chemical
II. Dimensional
Linear Indicators3 – Dial & Up to 4 in (3.2 + 8.8L) µin + 0.6R Gage blocks
Test
Angle Indicators & 15°, 30°, 45°, 60°, 0.03° Angle block set
Protractors3 75°, 90°
Diameter /Radius/Fixture
Gages/Weld Gages3 –
Optical Comparators3 –
Surface Plates3 –
Grades AA, A, & B
Interim Verification of
Coordinate Measuring
Machines3 –
Length 3,8 –
DC Power3 – Generate
Capacitance3 –
Generate
AC Voltage3 –
Generate
AC Voltage3 – Measure
AC Current 3 – Measure
(150 to 1025) A
DC Volt Amplitude:
Rise Time:
Electrical Simulation of
Thermocouples3 –
Electrical Simulation of
Thermocouples3 –
Electrical Simulation of
RTDs3
V. Fluid Quantities
Viscosity Dip Cups3 (0 to 100) mm²/s 2.2 cSt Certified viscosity oil
(Kinematic Viscosity, Efflux (100 to 1000) mm²/s 2.2 cSt
Time)
(0.01 to 100) psig 0.07 % of full scale Fluke 754 w/ 700 series
(0.1 to 1000) psig 0.07 % of full scale modules
(5 to 10 000) psig 0.07 % of full scale
Vacuum 3 (0.01 to 28.5) in∙Hg 0.07 % of full scale Fluke 754 w/ 700PD6
Indirect Verification of
Brinell Hardness Testers at
Test Condition 3 –
Indirect Verification of
Rockwell Hardness Testers3 HRA Indirect verification per
(20 to 65) 0.82 HRA ASTM E18
(70 to 78) 0.81 HRA
(80 to 84) 0.81 HRA
HRBW
(40 to 59) 0.82 HRBW
(60 to 79) 0.82 HRBW
(80 to 100) 0.81 HRBW
HRC
(20 to 30) 0.84 HRC
(35 to 55) 0.84 HRC
(60 to 65) 0.81 HRC
HR15N
(70 to 77) 0.82 HR15N
(78 to 88) 0.82 HR15N
(90 to 92) 0.81 HR15N
HR30N
(42 to 50) 0.81 HR30N
(55 to 73) 0.81 HR30N
(77 to 82) 0.81 HR30N
HR45N
(20 to 31) 0.82 HR45N
(37 to 61) 0.83 HR45N
(66 to 72) 0.81 HR45N
15T:
Low 0.80 HRBW
Medium 0.80 HRBW
High 0.82 HRBW
30T:
Low 0.80 HRBW
Medium 0.80 HRBW
High 0.82 HRBW
Speed3 –
Optic/Non-Contact:
RPM (5 to 200 000) rpm 0.018 % Monarch PT200
Contact:
RPM (0.5 to 20 000) rpm 0.22 % Monarch PT200
Totalizer/Rate Meters (1 to 6561.7) rpm 0.22 %
Light Booths3 –
VIII. Thermodynamics
_______________________________________________________________
1 This laboratory offers commercial calibration, dimensional testing and field calibration services, where
noted.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a
fraction/percentage of the reading plus a fixed floor specification.
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General
requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of R205 –
Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical competence for
a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated
April 2017).
_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 1741.08
Valid to May 31, 2023
For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.