Nmlsop For Health Centre Level Laboratory - Jan-2024
Nmlsop For Health Centre Level Laboratory - Jan-2024
TANZANIA
MINISTRY OF HEALTH
NATIONAL MEDICAL
STANDARD OPERATING
PROCEDURES FOR HEALTH
CENTRE LEVEL
LABORATORY
Version 01:2024
i
JANUARY 2024
NATIONAL MEDICAL STANDARD OPERATING PROCEDURES FOR HEALTH CENTRE LEVEL
LABORATORY
PUBLISHED IN 2024
©MINISTRY OF HEALTH,
PO BOX 743,
EMAIL: [email protected]
WEBSITE: www.moh.go.tz
ISBN: 978-9912-9833-0-4
Any section(s) or clause(s) within this document, can be used by the Government of
Tanzania Departments, implementers and stakeholders, provided that, the Ministry of
Health is acknowledged
ii
DISCLAIMER
This document is provided on the basis that it is for the use of MoH and PO-RALG
only. MoH will not be bound to discuss, explain or reply to queries raised by any
agency/organisation other than the intended recipients of this document. MoH
disclaims all liability to any third party who may place reliance on this document and
therefore does not assume responsibility for any loss or damage suffered by any such
third party in reliance thereon.
The images and related drawings used in this document are intended solely as a
guiding support and should be considered as purely indicative, not restrictive of the
expected item characteristics nor an indication of the client’s choice and/or preference.
The views expressed in this manual are those of the individual contributors. Comments
on the usefulness of this manual, and suggestions for improvements in future editions
will be welcome, and should be addressed to the contact therein. This communication
may contain information that is proprietary, confidential, or exempt from disclosure. If
you are not the intended recipient, please note that any other dissemination,
distribution, use or copying of this communication is strictly prohibited. Anyone who
receives this document in a message in error should notify the sender immediately by
telephone or by return e-mail and delete it from their computer.
Table of Contents
ABBREVIATIONS AND ACRONYMS ................................................................................................. I
FOREWORD ................................................................................................................................VI
ACKNOWLEDGEMENTS ..............................................................................................................VII
SCOPE ........................................................................................................................................... 9
PRIMARY BENEFICIARIES .................................................................................................................... 9
SECONDARY BENEFICIARIES ................................................................................................................ 9
Abbreviations Acronyms
AIDS Acquired Immunodeficiency Syndrome
AJLM African Journal for Laboratory Medicine
AMMP-1 Adult Morbidity and Mortality Project Phase 1
AMR Antimicrobial Resistance
ANC Antenatal Care
APECSA Association of Pathologists of East, Central and Southern
Africa
APHL Association of Public Health Laboratories
APT Association of Pathologists of Tanzania
ASCP American Society for Clinical Pathology
ASLM African Society for Laboratory Medicine
ASM American Society for Microbiology
BEmONC Basic Emergency Obstetric and Neonatal Care
BMC Bugando Medical Centre
BRM Biorisk Management
BSC Biological Safety Cabinet
BSc Bachelor of Science
BSL Biosafety Level
BUQ Bottom-Up Quantification
CCM Chama Cha Mapinduzi (Ruling Party in the United
Republic of Tanzania)
CD4 Cluster of Differentiation 4
CDC Centers for Disease Control and Prevention
CDs Cluster of Differentiations
CEmONC Comprehensive Emergency Obstetric and Neonatal Care
CEPD Continuing Education and Professional Development
CHF Community Health Financing
CHMT Council Health Management Team
CHSB Council Health Service Board
CHWs Community Health Workers
CLS Community Laboratory Services
CLSI Clinical and Laboratory Standards Institute
COPECSA College of Pathologists, East, Central and Southern Africa
CPD Continuing Professional Development
CPL Central Pathology Laboratory
CTRL Central Tuberculosis Reference Laboratories
DDHCTSU Director, Diagnostic and Health Care Technical Services
Unit
DED District Executive Director
DHs District Hospitals
i
Abbreviations Acronyms
DLTs District Laboratory Technologists
DMO District Medical Officer
DO Data Officer
DP Development Partner
DQA Data Quality Assurance
DTS Dried Tube Specimen
EAC East African Community
EAPHLN East African Public Health Laboratory Network
ECSA East, Central, and Southern African
eLIS electronic Laboratory Information System
eLMIS Electronic Logistics Management Information System
EOC Emergency Operations Centre
EQA External Quality Assessment
FBO Faith Based Organisations
FYDP Five Years’ Development Plan
GCLA Government Chemistry Laboratory Agency
GHSA Global Health Security Agenda
GOT Government of Tanzania
HCRF Health Commodity Revolving Fund
HCTS Health Care Technical Services
HIV Human Immunodeficiency Virus
HLI Health Links Initiative
HLPC Health Laboratory Practitioners’ Council
HMIS Health Management Information Systems
HMTs Health Management Teams
HSSP IV Health Sector Strategic Plan IV
HSSP V Health Sector Strategic Plan V
I-TECH International Training and Education Centre for Health
IATA International Air Transport Association
ICAP International Center for AIDS Care and Treatment
Programs
IDSR Integrated Disease Surveillance and Response
IHR International Health Regulations
IMPACT Information Mobilise for Performance Analysis and
Continuous Transformation
IMTU International Medical and Technological University
ISBN International Standard Book Number
ISO International Organization for Standardization
IT Information Technology
KCMC Kilimanjaro Christian Medical Centre
KIU Kampala International University
KPI Key Performance Indicator
LEMM Laboratory Equipment Management Module
LIO Laboratory Information Officer
LIS Laboratory Information System
LMIS Logistic Management Information System
ii
Abbreviations Acronyms
LO Logistic Officer
LQA Laboratory Quality Assurance
MDG Millennium Development Goals
MeLSAT Medical Laboratory Scientists Association of Tanzania
MMAM Mpango wa Maendeleo ya Afya ya Msingi
MNCH Mother and Neonatal Child Health
MoH Ministry of Health
MOU Memorandum of Understanding
MS Marketing Surveillance
MSc Master of Science
MSD Medical Stores Department
MSD Medical Stores Department
MTBDR Mycobacterium Tuberculosis Drug Resistance
MTEF Medium Term Expenditure Framework
MUHAS Muhimbili University of Health and Allied Sciences
NACP National AIDS Control Programme
NACTE National Council for Technical Education
NBTS National Blood Transfusion Services
NCDs N0n-Communicable Diseases
NGO Non-Governmental Organisation
NHLS National Health Laboratory Services
NHLSP National Health Laboratory Strategic Plan
NIMR National Institute for Medical Research
NMCP National Malaria Control Programme
NMLSSP III National Medical Laboratory Services Strategic Plan III
NPHL National Public Health Laboratory
NPHLA National Public Health Laboratory Agency
NRL National Reference Laboratory
NSCLQS National Sub-Committee on Laboratory Quality System
NSGHLS National Standard Guidelines for Health Laboratory
Services
NSGRP National Strategy for Growth and Reduction of Poverty
NSLQMS National Sub-committee for Laboratory Quality
Management Systems
NTDs Neglected Tropical Diseases
NTLP National Tuberculosis and Leprosy Programme
PEPFAR President’s Emergency Plan for AIDS Relief
PHDR Poverty and Human Development Report
PHIA Population-based HIV Impact Assessment
PHLB Private Health Laboratory Board
PHLS Public Health Laboratory Services
PLHIV People Living with HIV and AIDS
PO-RALG President’s Office-Regional Administration and Local
Government
POC Point of Care
PPM Planned Preventive Maintenance
iii
Abbreviations Acronyms
PPP Public Private Partnership
PT Proficiency Testing
QA Quality Assurance
QC Quality Control
QMS Quality Management System
QO Quality Officer
RAS Regional Administrative Secretary
RBF Results-based Financing
RFM Result Framework Matrix
RHMT Regional Health Management Team
RLQA Regional Laboratory Quality Assurance
RLTs Regional Laboratory Technologists
RMO Regional Medical Officer
RRH Region Referral Hospital
RTQII Rapid Testing Quality Improvement Initiative
SADC Southern African Development Community
SADCAS Southern African Development Community Accreditation
System
SCM Supply Chain Management
SD Standard Deviation
SDG Sustainable Development Goals
SLIPTA Stepwise Laboratory Improvement Process Towards
Accreditation
SLMTA Strengthening Laboratory Management Toward
Accreditation
SO Safety Officer
SWOC Strength Weakness Opportunities and Challenges
TA Technical Assistance
TB Tuberculosis
TCU Tanzania Commission for Universities
THPS Tanzania Health Promotion Support
TIKA Tiba kwa Kadi
TMDA Tanzania Medicine and Medical Devices Authority
TOR Terms of Reference
TOT Trainer Of Trainee
TSPAS Tanzania Service Provision Assessment Survey
URT United Republic of Tanzania
WHO World Health Organization
ZACDS Zonal Advisory Committee on Diagnostic Services
iv
TERMS AND DEFINITIONS
For the purposes of this NMLSOP document, these terms and definitions will apply:
v
FOREWORD
This is the first National Medical Laboratory Standard Operating Procedure
(NMLSOP).
The document outlines the Standard Operating Procedures that the laboratory tests
are to be performed in various facilities during the period of one year’s towards
strengthening the provision of quality medical laboratory services. The plan was
developed through consultative process involving various stakeholders within the
Ministry of Health and other Ministry of the government of Tanzania.
This plan describes the background of medical laboratory sciences, as well as the
context in which it was established in the country. It also features the current roles,
functions and structure of the organisation. Furthermore, contents of the plan include;
stakeholders’ analysis, strengths, weaknesses, opportunities, and challenges that face
medical laboratory services. In identifying these elements, SWOC and Political,
Economic, Social, Technological, Environmental and Legal (PESTEL) analysis were
used.
The NMLSOP have been developed in line with the other medical laboratory’s technical
procedures. This NMLSOP is presented in seven (7) chapters set in a path to
overcome the weaknesses and challenges as well as proving a tool to take advantage
of organisation’s strength in exploring the existing opportunities. The SOPs include
purpose, scope, responsibilities, principle and many steps. This NMLSOP is expected
to be reviewed annually from January 2024 to January 2025.
vi
ACKNOWLEDGEMENTS
The National Medical Laboratory Standard Operating Procedure (NMLSOP) is a
product of dedicated efforts and contributions of various stakeholders. The MoH
acknowledges the contribution of PO-RALG, Development and Implementing Partners,
Non-Government Organisations (NGOs), Institutions or facilities (National Hospitals,
Zonal Hospitals, Specialised Hospitals, Regional Referral Hospitals, District Hospitals,
Health Centers, Dispensaries), Regulatory Agencies (Health Laboratory Practitioner’s
Council and Private Health Laboratories Board), Programmes (NASHCoP, NTLP,
NMCP, NBTS) and Individuals towards improving the quality of health care service
delivery through improved medical laboratory services.
In particular, the MoH would like to thank the Global Fund (GF) and Centres for
Disease Control and Prevention (CDC) for their financial and technical support through
the consultancy in development of this strategic plan. Special appreciations are
extended to technical experts and individuals for their active participation and
constructive inputs and comments provided in reviewing this guideline (ANNEX 01).
Our appreciations go to the Director of diagnostic and Health Care Technical Services
Unit (DDS) Dr Alex S. Magesa for steering up the whole process; Acting Head of
Laboratory Services (Ag. HLS) Mr. Reuben S. Mkala for field coordination; and to all
laboratory technical team or individuals who played a pivotal role in developing this
NMLSOP by participating in the consultation workshops, meetings, providing relevant
information and offering their expert opinion when consulted. Last but not the least;
appreciations go to David Ocheng, who facilitated the whole
Prof.Tumaini J. Nagu
CHIEF MEDICAL OFFICER
vii
EXECUTIVE SUMMARY
In developing this National Medical Laboratory Standard Operating Procedure
(NMLSOP) for the period 2024-2025 consideration has been made on a number of
National and International guiding and operational guidelines. National Standard for
Medical Laboratories (2017); Health Care Technology Policy Guideline (2004);
Operational Plan for the National Laboratory System to Support HIV and AIDS Care
and Treatment (2005); National Laboratory Quality Assurance Framework to Support
Health Care Interventions (2007); Standard Guidelines for the Facilities and
Operations of Forensic Bureau Laboratory (2008); International Health Regulations
(IHR) of 2005; Global Health Security Agenda (GHSA) 2024 Framework; and the
Ruling Party Election Manifesto 2020.
The main objective of this SOPs is to enable the laboratories at all levels to effectively
and efficiently carry out their core functions of laboratory tests as stipulated by their
mandate, strategically coordinate and allocate the available resources by prioritising
functions with effective impact in line with the National Health Policy (2007) and the
Sustainable Development Goals (SDGs).
There are four chapters in this document. Chapter one provides; a brief introduction
and background information, justification and objectives for the NMLSOP.
This National Medical Laboratory Standard Operating Procedure has four strategic
directions:
For each objective above, strategies, targets, activities and performance indicators
were derived. The NMLSOP Matrix specifically describes the sequence of objectives,
strategies, targets, activities and performance indicators are considered the guiding
framework for the development and implementation of annual operational plans.
Page 8 of 245
Scope
The scope of this National Medical Laboratory Standard Operating Procedure
(NMLSOP) is to provide a step-by-step instruction for performing laboratory tests.
Primary Beneficiaries
The primary beneficiaries are Laboratory Practitioners’ who perform examination of
biological samples in the laboratories and non-laboratory testers performed specialised
laboratory tests at point-of-care testing sites.
Secondary Beneficiaries
The Secondary users of this document will include but not be limited to:
Policy makers, medical care managers and administrators, medical devices regulatory
authorities, medical laboratories, intervention programmes, Regional Health
Management Teams and Council Health Management Teams, zonal medical
equipment workshops, biomedical engineers, healthcare technical services and end-
users engaged in strengthening the quality of medical diagnostic services in the
country.
Page 9 of 245
CHAPTER 1: SAMPLE COLLECTION
1.0 GENERAL CONSIDERATIONS
The collection of samples for laboratory tests from patients consists of following steps:
10
xiii. The most appropriate samples for isolation of viral, chlamydial or rickettsial
agents depend on the nature of the illness.
xiv.The material should be collected as early as possible in the acute phase of
the disease, because these agents tend to disappear relatively rapidly after
the onset of the symptoms.
xv. Vesicle fluid is preferably collected in a syringe or capillary pipette and
immediately diluted in an equal volume of skimmed milk or tissue culture
medium.
xvi.All samples for viral culture should be frozen and stored at -70°C until culture
is initiated.
1.1.3 Dispatch of Samples from Reception to the Laboratory Sections
i. Match the containers and respective request forms, number them and enter in the
dispatch register/computer.
ii. Verify while handing over/taking away to respective department of the laboratory.
iii. Notify the concerned department about urgent and special tests.
1.1.4 Sample Transportation
i. Exterior of the container should not be soiled/contaminated with the samples.
ii. Sufficient absorbent materials must be used to pack the sample, so that it absorbs
the spilled liquid in case of leakage/breakage during transit to reference/referral
laboratory.
iii. Sample containers must be leak proof and unbreakable. Plastic containers are
preferred.
iv. Samples must be promptly delivered to the laboratory for valid results.
v. If applicable, appropriate transport media should be used.
vi. Samples are to be refrigerated or incubated at 37°C, as the case may be, if there is
a delay in transport of samples to laboratory.
vii. An appropriately filled request form should always accompany all samples to guide
the pathologist/ health laboratory practitioner in selection of suitable media or
appropriate technique.
1.2 COLLECTION OF BLOOD SAMPLES
Consider the following recommended order of veins during blood drawing;
11
iv. Wherever applicable, paired samples are to be collected during acute and
convalescent phases of illness in certain viral and other infections to
document a diagnostic rise in antibody titre.
v. Protect blood samples from extremes of heat and cold during transport.
vi. Whole blood is to be stored at 2-8°C.
vii. Serum can be frozen at -20°C or lower temperature for long term storage.
1.2.2 Blood sample for culture
i. Make sure you have the appropriate media for blood culture, as the media
may vary depending upon the type of pathogen suspected.
ii. Wash the hands with soap and water and wear sterile gloves.
iii. Withdraw the blood following the procedure described on the procedure for
collection of venous blood. Change needle before injecting the blood into the
culture bottle.
iv. Thoroughly clean the rubber bung of the culture bottle with iodine solution
and inject an amount of blood equal to 10% of the volume of medium (for 30
ml medium 3 ml blood and for 50 ml medium, 5 ml blood is needed).
v. After the needle has been removed, the site should be cleaned with 70%
alcohol/spirit swab again.
vi. Don’t store the containers and caps separately.
vii. Blood obtained for culture of suspected anaerobes should not be exposed
to air in any way.
1.2.3 Venous blood
i. Welcome and greet the patient, introduce
yourself.
ii. Make the patient sit comfortably on the
phlebotomy chair. iii. Identify the patient by
asking his/her particulars and compare them
with the request form.
iii. Check the request form for the requested investigations and Inform the
patient about the samples to be collected.
iv. Where possible, ask the patient to remove any tight - fitting sleeved clothing,
or to roll up loose sleeves.
v. Lable the containers before obtaining sample.
vi. Select syringe of appropriate size so that the quantity of blood required can
be obtained in single prick. If multiple or high volume of samples is required,
use a butterfly needle or a canula.
vii. Select appropriate vein (preferably antecubital) from forearm. Cleanse the
skin over the venepuncture site in a circle approximately 5 cm in diameter
with 70% alcohol/spirit swab and allow to air dry, do not blow.
viii. If the sample is to be collected for blood culture then skin is to be thoroughly
sterilised, following the procedure as follows:
a. Starting in the centre of a circle
12
b. apply 2% iodine (or povidone-iodine) in everwidening circles until the
entire chosen area has been saturated with iodine.
c. Allow the iodine to dry on the skin for at least 1 min.
d. Completely remove the iodine with 70% alcohol/spirit swab following the
pattern of application.
viii. Apply a tourniquet tight enough to obstruct venous flow only and relocate
the vein to be punctured but don’t touch the proposed site of needle entry
or the needle itself. Ask the patient to clench the fist to make the veins
prominent. If the vein is not visible, palpate it with fingers. In case the veins
of forearm are not visible/palpable, other sites such as dorsum of the hand
may be selected.
ix. With bevel up, insert the correct needle size on the veins at an angle
between 15 - 30o then draw the blood into the appropriate collection tubes,
(make sure the patient arm in a downward position to prevent reflux).
x. Mix immediately after drawing each tube that contain an additive by gently
inverting the tube 8 - 10 times. Do not mix vigorously in order to avoid
haemolysis.
xi. Release the torniquet from the patient, Withdraw the needle and apply
pressure to the puncture site using dry cotton balls/gauze pad. Do not
withdraw the piston too forcefully as it can collapse the vein and it may cause
frothing/ haemolysis of the sample.
xii. Apply pressure with thumb on antiseptic swab at puncture site for 2 - 4 min
till the blood ooze stops.
xiii. If syringe was used, safely remove the needle from the syringe before
distribution.
xiv.The blood from syringe is distributed to appropriate, labelled containers.
NOTE:
In case of multiple blood sample collection, consider the following recommended order
of draw;
13
vi. Wipe first drop of blood with a clean dry gauze or cotton wool,
vii. Collect the sample with the second drop of blood using appropriate sample
collection devices such as blood capillary tube,
viii. Apply pressure with a clean dry gauze pad or cotton wool until bleeding stop,
ix. Transfer the collected blood sample into appropriate sample container or
testing devices.
1.2.5 Procedure for Performing Neonate Capillary Blood Collection
i. Use the most medial or lateral portions of the planter surface of the heel
Limit the depth of the puncture wound by using an automated lancet.
ii. Only consider using the whole plantar surface of the foot (using automated
lancets of 2.2mm in length or less) for neonates over 33 weeks’ gestation if
they are having multiple/frequent heel pricks
iii. Position the neonate: ensure the foot is lower than the body.
iv. Choose a puncture site – do not use a previous puncture site.
a. Clean the heel site (i.e. gauze and water) if the foot appears unclean (e.g.
faecal material).
b. Encircle the foot with the palm of the hand and the index finger.
c. Make a quick puncture with the automated lancet device
d. Wipe off the first drop of blood with a gauze swab
e. Collect the sample with the second drop of blood using any of the
collection devices such as slides or rapid test
f. Apply pressure with a clean dry gauze pad until bleeding stop
1.3 COLLECTION OF URINE SAMPLES
Urine samples are collected for routine and culture examinations to diagnose urinary
tract infections (UTIs), both lower UTIs (cystitis-infection of the bladder) and upper UTIs
(pyelonephritis-infection of the kidney). Unlike most other cultures, colony counts are
done on urine samples to determine the number of organisms present in the sample.
Generally, >100,000 organisms/ml of a single isolate indicate an individual has a UTI.
However, mixed UTIs do occur and some individuals with UTIs will have counts lower
than 100,000 organisms/ml.
Most organisms that cause UTIs are normal enteric flora, including E. coli, Proteus,
Klebsiella, Enterobacter, and Enterococcus species. In young, otherwise healthy,
sexually active females, Staphylococcus saprophyticus can be found to be the cause
of UTI.
UTIs caused by Proteus species can be complicated by the formation of urinary calculi
or stones. The large amount of urease that Proteus produces can alkalinize the urine.
14
If there are minerals such as phosphates or carbonates in the urine, the alkaline pH
can cause them to precipitate out and form stones.
E. Post-prandial
It is collected after meal (usually after 2 hours). It is good for glucose and protein
estimation. Urine sugar testing now has limited diagnostic or prognostic value.
F. Timed sample
It is a combination of all voiding over a length of time. Two-hour sample is good for
urobilinogen and 24-hour sample is good for quantitative urinary components
estimation. Timed urine samples are collected in dynamic function tests.
G. Foley catheter
Disinfect a portion of the catheter with alcohol, puncturing the tubing directly with a
sterile syringe and needle and aspirate the urine. Place urine in a sterile container, it
should never be collected from drainage bag.
H. Suprapubic urine
Urine sample collected by suprapubic aspiration and cystoscopy.
15
Urine sample is often collected by patient him/herself. Therefore, the patient needs to
be properly instructed to have correct sample collection. An uncontaminated
midstream urine (MSU) sample is the best and following methods are to be used for
its collection:
A. Females
i. Wash the genital area thoroughly with - Clean water (may be omitted for
urine Routine Examination).
ii. With two fingers of one hand, hold the outer folds of vagina (labia) apart.
With the other hand, rinse the area from the front to the back with running
tap water.
iii. Start urination so that the stream of urine should flow without touching the
skin. After a few moments, place a sterile container under the stream of
urine. Remove it from the urine stream the moment required amount of urine
is collected.
iv. Secure and tighten the cap on the container.
B. Males
i. Wash the genital, area thoroughly with Clean water (may be omitted for urine
Routine Examination).
ii. Start urination and after a few moments, place a sterile container under the
stream of urine. Collect the required amount of urine and remove the
container from urine stream.
iii. Secure and tighten the cap.
C. Infants, uncooperative and debilitated patients
i. Plastic bags may be attached after careful and thorough washing of genital
area.
ii. The bags should be watched so that they can be removed immediately after
patient has passed the urine.
iii. If the patient has not voided urine within 30 min the collecting bag is
removed.
iv. Patient needs to be re-scrubbed and a new collection device is to be
attached.
D. Urine collection for Mycobacterium tuberculosis
i. Three consecutive early morning samples (>90 ml each) collected in sterile
container are superior to 24h collection.
ii. Boric acid (1.6%) is used as preservative in case of 24h urine collection in
exceptional situations e.g., when patient cannot report daily for sampling.
iii. Suprapubic aspiration in ward by a doctor is preferred in catheterised
patients.
16
1.4 FAECAL SAMPLES COLLECTION PROCEDURE
Faecal samples are collected for routine and culture examinations to find the causative
agent of infectious diarrhoea. Rectal swabs are often helpful in identifying the cause of
acute bacterial diarrhoea when stool sample cannot be collected readily.
It is important to remember that there are many causes of diarrhoea other than
infectious agents such as metabolic disorders, certain drugs, and food intolerances
(allergies). In a routine stool culture, the following organisms are often isolated:
Salmonella species, Shigella species and Vibrio cholera. Viruses and parasites are
also common causes of diarrheal disease.
1.4.1 Procedure for Collection of faecal samples
i. Faeces should be passed directly into a clean, waxed cardboard container
that is fitted with a tight cover.
ii. Avoid contact with residual soap/detergent, disinfectant or urine in the
bedpan.
iii. Faeces obtained are transferred to another clean, wide mouthed and screw
capped container. The sample should include any pus, blood, mucus or
formed elements that may have passed with stool.
iv. Sample (~1 ml) is added to 10 ml sterile alkaline peptone water in suspected
cholera cases.
v. If viral infection is suspected, faeces are extracted with sterile buffered
saline. Faeces (~1 ml) are mixed with 9 ml sterile buffered saline, allowed to
sediment for 30 min (or centrifuged). The supernatant is transferred to a
sterile container, frozen and kept below -40°C until processed.
1.5 COLLECTION OF BODY FLUIDS
The primary body fluid that are collected for routine and culture examinations includes;
cerebrospinal fluid (CSF), joint fluid, pleural fluid, and peritoneal (ascites) fluid.
17
CSF is normally collected from sub-arachnoid space of spinal cord at lumber level by
puncture with a long needle. A physician in the ward under strict aseptic conditions
performs the procedure.
18
The three most common causes of septic arthritis are: Staphylococcus aureus,
Neisseria gonorrhoea, and Coagulase-negative staphylococci in patients with joint
replacement prosthetics
19
iv. Make a smear of the discharge on a microscope slide by gently rolling the swab
on the slide. This will avoid damaging pus cells which contain the bacteria.
Note: Very few pus cells may be present if the patient has recently passed urine.
Allow 2–4 hours after urination before collecting a samples.
i. When culture is indicated (see previous test),
ii. Collect a sample of pus on a sterile cotton-wool swab.
iii. If possible, before inserting the swab in a container of Amies transport medium,
inoculate a plate of culture medium.
iv. Label the samples and deliver them to the laboratory as soon as possible.
v. Isolation of N. gonorrhoeae from urine
Note:
• A rectal swab is also required from homosexual patients. A selective medium is
required to isolate N. gonorrhoea from a rectal sample.
• In acute urethritis, it is often possible to detect N. gonorrhoea in pus cells passed in
urine, especially the first voided urine of the day (centrifuged to sediment the pus
cells).
1.5.4 Cervical swabs and possible pathogens
A. From non-puerperal women:
Neisseria gonorrhoea, Chlamydia trachomatis (serovars D-K), Streptococcus
pyogenes, herpes simplex virus.
B. From women with puerperal sepsis or septic abortion:
Streptococcus pyogenes, other beta-haemolytic streptococci, Staphylococcus
aureus, Enterococcus species, anaerobic cocci, Clostridium perfringens,
Bacteroides, Proteus, Escherichia coli and other coliforms, Listeria
monocytogenes.
1.5.5 Collection of cervical samples from female patients
i. A samples collected from the endocervical canal is recommended for the
isolation of N. gonorrhoeae by culture. Use a sterile vaginal speculum to
examine the cervix and collect the samples.
ii. Moisten the speculum with sterile warm water, and insert it into the vagina.
iii. Cleanse the cervix using a swab moistened with sterile physiological
saline.
iv. Pass a sterile cotton-wool swab 20–30 mm into the endocervical canal and
gently rotate the swab against the endocervical wall to obtain a samples.
v. When gonorrhoea is suspected, before inserting the swab in Amies
transport medium, if possible inoculate a plate of culture medium. Label
20
the sampless and deliver to the laboratory as soon as possible. Inoculated
culture plates must be incubated within 30 minutes.
1.5.6 Vaginal Swabs
Vaginal discharge may be due to infection of the vagina or infection of the cervix or
uterus. Pathogens causing vaginal infections include Trichomonas vaginalis, Candida
species, and Gardnerella vaginalis with anaerobes.
1.5.7 Collection of vaginal discharge
To detect T. vaginalis, C. albicans and G. vaginalis
Two preparations are required:
i. Wearing protective rubber gloves, cleanse around the ulcer (chancre) using
a swab moistened with physiological saline. Remove any scab which may
be present.
ii. Gently squeeze the lesion to obtain serous fluid. Collect a drop on a clean
cover glass and invert it on a microscope slide.
iii. Immediately deliver the preparation to the laboratory for examination by
darkfield microscopy
21
1.6 WOUND SAMPLES TO INCLUDE PUS, ABSCESS, TISSUE, ETC.
Indicated for primarily to diagnose skin and soft tissue infections (SSTIs). SSTIs may
be caused by a variety of organisms; different organisms depending on how the wound
or injury occurred. Fungi, parasites, and viruses are also important causes of certain
types of SSTIs.
22
ix. When tuberculosis is suspected: Aspirate a sample of the pus and transfer it to a sterile
container.
x. When the tissue is deeply ulcerated and necrotic (full of dead cells): Aspirate a sample
of infected material from the side wall of the ulcer using a sterile needle and syringe.
Transfer to a sterile container.
xi. Fluid from pustules, buboes, and blisters: Aspirate a samples using a sterile needle and
syringe. Transfer to a sterile container.
xii. Serous fluid from skin ulcers, papilloma, or papules, that may contain Treponema:
Collect a drop of the exudate directly on a clean cover glass and invert it on a clean
slide. Immediately deliver the samples to the laboratory for examination by dark-field
microscopy.
Caution: Samples from patients with suspected plague or anthrax are highly
infectious. Label such samples HIGH RISK and handle them with care. In a health
centre for dispatch to a microbiology laboratory Collect the samples using a sterile
cotton-wool swab.
Insert it in a container of Amies transport medium, breaking off the swab stick to allow
the bottle top to be replaced tightly.
In a hospital with a microbiology laboratory
• Using a sterile technique, aspirate or collect from a drainage tube up to 5 ml of
pus.
• Transfer to a leak-proof sterile container. When the material is aspirated fluid
from a pustule, transfer the fluid to a sterile, leak-proof container. Stopper, and
seal in a leak-proof plastic or metal container.
Note: It is not possible to transport exudate from a suspected treponemal ulcer
because the Treponema remain motile for only a short time.
Make a smear of the material on a clean slide (for Gram staining) and allow to air-dry
in a safe place. Heat-fix the smear.
Caution: Do not make a smear for transporting when the samples is from a
patient with suspected anthrax.
Send the samples with a completed request form to reach the microbiology laboratory
within 6 hours.
1.8 COLLECTION OF THROAT AND NASAL SWABS
Throat cultures are performed to diagnose streptococcal pharyngitis (infection with
Streptococcus pyogenes (Group A streptococci). The most common causes of
pharyngitis (sore throat) are viruses, which cause over 75- 80% of all cases. Of the
23
bacteria that cause pharyngitis, Group A streptococci is the major cause and therefore,
with few exceptions the only bacteria that is reported from a throat culture is
Streptococcus pyogenes.
Exceptions include:
Corynebacterium diphtheria which in areas of the world where vaccination is prevalent,
is a rare cause of pharyngitis
Neisseria gonorrhoea can cause pharyngitis, however many pharyngeal infections with
N. gonorrhoea are mild or asymptomatic.
If the physician suspects either of these two organisms, he/she must let the lab know
because the isolation of either requires special culture techniques.
1.5.10 Throat Swabs
Throat swab cultures are to be taken under direct vision with good light.
Areas of exudation, membrane formation, any inflammation or if not seen then tonsillar
crypts are the sites of choice.
1.5.11 Nasal swabs
Nasopharyngeal swabs are better taken by treating physician/surgeon himself. For
recovery of viral agents, washings are collected after gargles with nutrient broth by the
patient.
Nasal Sample for Mycobacterium leprae
The nasal sample for M. leprae can be taken as follows:
Nasal swab
i. Make the patient sit with his head bent backwards but facing the light.
ii. Insert and repeatedly rotate the swab into one of the nasal cavities, against
upper part of the nasal septum.
iii. Make 2-3 evenly spread smears.
iv. Air dry the slides, wrap in a paper and send to the laboratory.
Nasal washings and nasal blow
i. Make the patient sit.
ii. Place a few drops of sterile saline in the nose.
iii. After 3 min, ask the patient to blow hard his nose on a small sheet of plastic or
cellophane. (This plastic or cellophane can be given to the patient to take it
home and ask him to blow hard onto the sheet, the following morning, soon after
waking and before washing.
iv. The patient can bring it directly to the laboratory).
v. Transfer some of the mucus pieces from the washing to a slide with a clean
wooden stick and make thin smear.
24
vi. Air dry slide and send it to the testing area
Collection of Nasopharyngeal Swabs
Assemble equipment for Nasopharyngeal swab collection and PPEs for prevention of
infections.
i. Lable the VTM with required information and fill the register and all necessary
forms with all necessary information.
ii. Remove the swab from the package. Do not touch the soft end with your hand
or anything else.
iii. Insert the nasopharyngeal swab into the nasopharnx region.
iv. Leave in place for a few seconds.
v. Slowly remove swab while slightly rotating it.
vi. Break the applicator’s stick end and put tip of swab into ATM containing vial
VTM
COLLECTION OF OROPHRYANGEAL SWABS
Assemble equipment for orophryangeal throat swab collection and prevention of
infections.
i. Label the VTM with required information and fill the register and all necessary
forms with all necessary and correct information.
ii. Remove the swab from the package. Do not touch the soft end with your hand
or anything else.
iii. Have the patient open his/her mouth wide.
iv. Insert swab in the area of tonsils.
v. Use tongue depressor if the patient is not able to resist gagging and closing the
mouth while the swab touches the back of the throat near the tonsils.
vi. Rotate swab to obtain adequate sample
vii. Break the applicator’s stick end and put tip of swab into vial containing VTM
25
i. Clean the skin at the smear site with swab and let it air dry. ii. Pinch the skin
firmly between your thumb and forefinger
ii. Make an incision in the skin about 5mm long and 2mm deep. Keep on pinching
to make sure the cut remains blood less iv. Turn the scalpel 90o and hold it at
right angle to the cut
iii. Scrap inside the cut once or twice with side of the scalpel to collect tissue fluid
and pulp (there should be no blood in the sample as this may interfere with
staining and reading of the slide
iv. Stop pinching the skin and absorb any bleed with dry cotton swab
v. Spread the material scrapped from the incision on to the slide, ensure you
spread it evenly with the flat of the scalpel making 8mm diameter
vi. Clearly label your slide with patient ID similar to that appearing on patient
request form
1.6.2 Sample from active lesion
i. Rub the scalpel with cotton wool drenched with alcohol.
ii. Pass the blade through the flame of the spirit burner for 3-4 seconds and let it
to cool without touching anything
iii. Select the most active looking lesion (active means lesion that are raised and
reddish in colour)
1.7 SPUTUM SAMPLES
Sputum cultures are performed to diagnose infections such as pneumonia and
pulmonary tuberculosis that is caused by Mycobacterium tuberculosis. Bacteria
associated with Community-Acquired Pneumonia (CAP):
26
1.6.3 Sputum Collection Procedure
i. Label the sample container on the body of the container, not on the lid and fill
out the sputum examination request form.
ii. Instruct the patient and demonstrate how she or he can produce and collect
good sputum;
a. Access to a well ventilated place (outside the laboratory working area),
b. Mouth wash (rinsing with water),
c. Breathe in deeply 2-3 times, and breathe out hard each time.
d. Cough deeply from the chest and collect the sputum into the container.
e. Opening and closing the sputum container so as there are no leaks or
smearing on the exterior of the container.
f. Hand wash steps
iii. Emphasize the need for the patient to supply the most useful sample, the
normally thick, yellowish (sometimes blood-streaked), purulent material brought
up from the lungs after a deep, productive cough.
iv. Emphasize that saliva produced by spitting is not sputum. However, if the only
sample the patient can produce is salivary, do submit it to the laboratory as it
can still yield useful information.
v. Encourage the patient to bring the collected sample back to the unit as quickly
as possible.
vi. For M. tuberculosis culture, a series of three fresh, early morning samples (5-
10 ml) are collected and kept in the refrigerator. If amount is less, the patient is
advised to collect 24 h sputum or until 50 ml is obtained.
vii. M.tuberculosis can be recovered from the gastric contents in infants, debilitated
patients and those who are unable to cooperate in the collection of sputum. This
can be obtained by gastric aspiration performed as an indoor procedure.
viii. Gastric washings are better collected early in the morning, in fasting state.
These are neutralised soon after collection by N/10 NaOH.
27
CHAPTER TWO: PARASITOLOGY
2.1. PROCEDURE FOR MALARIA RAPID TEST (MRDT)
2.1.1 Purpose
This procedure provides instructions for In vitro qualitative screening test for detection
of malaria parasites (P. falciparum, P. vivax, P. ovale and P. malariae) in whole blood
2.2.2 Scope
This procedure applies to all Health Laboratory Practitioners who works in parasitology
section on performing MRDT test.
2.2.2.1 Responsible
2.2.3 Principle
2.2.5 Equipment
Timer
2.2.6 Materials
28
2.2.7 Storage and Stability
MRDT sample should be tested within 1hour after collection if not possible stored at 2-
8oc for 7 days.
2.2.8 Safety
i. All personal protective equipment (PPE) must be worn when performing this
procedure.
ii. All samples must be regarded as potentially infectious.
iii. Refer to National infection prevention and control Guidelines (IPC)
2.2.9 Calibration
Not Applicable
Process known positive and negative blood sample daily before performing patient
samples.
i. Allow all kit components and specimen to room temperature prior to testing
ii. Remove the test device from the foil pouch; place it on a flat, dry surface.
iii.Label the test device with the patient identification number/name
iv. Transfer 5µl of whole blood collected in an inverted cup/special capillary
provided into sample well by touching sample pad. (see manufacture
instruction)
v. Add 4 drops of assay diluents into the squire assay diluents well. (See
manufacture instruction)
vi. Read the result within 15 min. (Do not interpret after 30 min) and read result.
(See manufacture instruction).
Not Applicable
Negative
i. The presence of one colour band (“C” Control line) within the result window
indicates a negative result. Positive
ii. “C” and “P. falciparum the presence of two coloured bands (“Pf” test line and
“C” control line) within the result window no matter which band appears first,
indicate P.F positive result. Positive
29
iii. “C”, “P.f” and “Pan” the presence of three colored band (“Pf”, Pan “Test line and
“C” Control line) within the result window no matter which band appears first,
indicate P.F positive or mixed infection of P.F and P.V or P.M or P.O Invalid
result
iv. If no coloured band appear, at control line “c” within stipulated time then the
result is invalid.
Test kit cannot detect malaria antigen if parasites are less than 100. The test is limited
to detect HRP2, an antigen to Malaria Plasmodium species that may persist after
treatment or passed P. falciparum infection.
2.2.17 References
Manufacturer inserts
2.2.1 Purpose
2.2.2 Scope
2.2.3 Responsible
30
2.2.4 Principle
Giemsa stain is a Romanowsky stain contains methylene blue which is basic stain and
eosin is acidic stain, the malaria parasites has DNA in the nucleus which is basic in
nature and RNA in the cytoplasm is acidic in nature, during staining the reactions takes
place whereby acidic part of a parasite will pick-up basic part of the stain and the basic
part of the parasite will pick-up the acidic part of the stain, that’s why the nucleus of the
parasite will show red color and the cytoplasm will stain bluish. Malaria parasites are
identified by microscopic examination of thick or thin blood films stained with Giemsa.
Thick blood films are used for detecting parasites, thin blood films are used for more
detailed morphological examination and for determining parasite species. Thick blood
films consist of several layers of blood cells, so that a large volume of blood is
examined. The thick film staining technique ruptures red cells, leaving white cells and
parasites intact. The thin film staining technique preserves the morphology of blood
cells and malaria parasites.
2.2.6 Equipment
Hot plate, Tally counters or differential counter, Microscope, Timer, Weighing scale,
and PH Meter
2.2.7 Maintenance
2.2.8 Materials
31
2.2.9 Storage and Stability
2.2.10 Safety
2.2.10 Calibration
Calibration of PH meter and weighing scale should be done once per year.
Maintenance of microscopes should be done as planned.
QC slides should be used to check the quality and performance of the Giemsa stain,
Microscope and Laboratory personnel. Malaria-positive and negative blood should be
used to prepare QC thick and thin films. Before examining patient slides, the QC slides
should be checked first, If the QC slides are satisfactory, the patient slides can be
examined. Slides must be selected regularly for cross-checking, either by sending
them to a crosschecking center or during routine supportive supervisory visits.
i. Prepare both thick and thin film on the same microscope glass slide.
ii. Label the slides with the unique patient ID including the date of examination.
iii. Put Slide card for making thin and thick blood films, showing size of blood drops
and area of slide to cover for a thin film and thick film
iv. Put the microscope glass slide on the slide card for thick and thin film
v. Pipette 2µl of blood and pour it on the smallest circle on the slide card
vi. Place the spreader in front of the 2µl drop of blood at a 30 o- 45o angle. Use a
clean microscope slide with a smooth edge as a spreader.
vii. Pull back the spreader and hold until the blood evenly spreads along the width
viii. Push the slide forward in a smooth continuous motion
ix. Avoid hesitation or a jerky motion when spreading the blood
x. Pipette 6µl of blood and pour on the large circle on the slide card.
xi. Using another slide, spread the drop of blood within the confined area to make
a thick film.
xii. Put the slide films on the slide rack and allow it to dry.
xiii. Fix the thin smear for one second by either spraying or dipping into absolute
methanol.
xiv.Air dry the slides on a slide rack with the fixed thin film facing down
32
i. Measure and dissolve 3.8g of Giemsa powder in 250ml of methanol
ii. Measure 250ml of glycerol and add to the solution above
iii. Ripen the stock solution by placing in direct sunlight for about 1 week or place
in water bath of temperature 56oC and shake at interval.
iv. Filter the stock solution before storing in a cool dry place with label and date of
preparation
i. Arrange the slides on a staining rack with the sample side facing up
ii. With an aid of a disposable pipette, flood the films with 10% Giemsa solution
and leave to stain for 15 minutes.
iii. Decant the Giemsa and wash in buffered water at pH 7.2
iv. Clean the back of each slide with cotton wool or gauze
v. Air dry the slides
33
2.2.14 Biological Reference Intervals Not Applicable
• Typical Malaria parasites have the following features on the Giemsa stained
films;
• Purple red chromatin dot, Blue cytoplasm, Brown-black/yellowish green pigment
and Distinct morphology
2.2.15.1 Interpretation
In stained blood films, trophozoites appear as red stained chromatin dots with blue
staining cytoplasm. If doubtful on parasites seen, search for definite ones. Do not make
a diagnosis on the basis of structures that resemble rings or chromatin dots alone.
Structures that may be confused with malaria parasites are platelets, portions or other
red cell inclusion bodies
2.2.15.2 Reporting
If malaria parasites are seen, report species identified and count e.g. “Plasmodium
species seen 50/200WBC”
The following are the Panic/Critical values for Malaria microscope test results. In case
of the patient result falling inside the indicated values, call the Doctor or relevant ward
and record the details of the conversation on the Panic Result Book
• Poor storage of reagents or using the reagents after expiry date may cause false
results.
• Difficulty in distinguishing young ring-stage parasites
34
2.2.19 Supporting Documents Not Applicable
2.2.20 References
2.3.1 Purpose
The purpose of this procedure is to provide step by step instructions for performing
macroscopic and microscopic examination of urine sediment samples
2.3.2 Scope
2.3.3 Responsible
2.3.4 Principle
The urine is visualized with naked eyes to determine its appearance (turbidity and
colour).
35
Terminal urine sample collected at any time of the day for demonstrating ova of
Schistosoma haematobium
First voided urine sample in the morning is used to demonstrate Trichomonas vaginalis
in males
In infants and babies, a random urine sample collected as early as possible is used for
all types of urine investigation.
2.3.6 Equipment
2.3.7 Materials
Gloves, Cover slips, Glass slides, Test tubes, Centrifuge tubes, Gauze, Grease pencil,
Lens paper, Waste containers, Urine container, Laboratory coat, Marker pen
Process urine sample within 1hour of collection, if not possible refrigerate at 2-8oC
immediately and test within 12 hours.
2.3.9 Safety
2.3.10 Calibration
2.3.12 Procedure
36
2.3.12.2 Microscopic examination
i. Label the centrifuge tubes with the laboratory unit number using a grease pencil
and arrange into a rack.
ii. If the urine sample is to be tested both for urinalysis and culture, mix well by
inverting three times and aliquot the sample into a centrifuge tube and label
exactly as the original sample
iii. Transfer the centrifuge tubes containing urine sample into the centrifuge.
iv. Make sure the centrifuge tubes are balanced well.
ii. Centrifuge the urines at 3000rpm for 5 minutes.
iii. Remove the centrifuge tubes from the centrifuge
iv. Pour off the supernatant into the sink with running water and leave the urine
deposit
v. Transfer the rack containing arranged centrifuge test tubes to the working area
vi. Arrange the marked slides according to the number of urine sample
vii. Re-suspend the urine deposit by tapping the bottom of the centrifuge tube.
viii. Transfer a drop of the deposit onto a clean glass slide.
ix. Put on a cover slip
x. Place the slide on the microscope stage and reduce the iris diaphragm
xi. Examine the slide systematically using x 10 objective and count the cells, casts,
crystals, amorphous phosphates and parasites (these structures are more likely
to be seen around the edges of the cover slip.
xii. Swing the x40 objective into position
xiii. Open the iris diaphragm very slightly to allow just enough light to provide a
contrast of cells, casts, crystal and amorphous phosphates against the bright
back ground (these structures are more likely to be seen around the edges of
the cover slip).
xiv.structures cannot be seen if a very bright light is used
xv. Put the used glass slide and cover slip in a container with disinfectant.
Macroscopy
Color Straw - Dark yellow
Appearance Clear - Hazy
Microscopy
Red blood cells: 0–2/hpf.
Casts 0-4/hpf
Bacteria Negative
A. Macroscopic
37
Report the Turbidity and color of the urine e.g. Clear yellowish, Blood stained etc.
B. Microscopic
• Report a range of the actual highest number of WBC (pus cells), RBC and
Parasites per high power field as counted under x40 objective (e.g. 0 – 2
RBC/HPF, 3 – 5 WBC/HPF, 10 – 12 S. haematobium ova/HPF) etc.
• For crystals, Epithelial cells, yeasts or casts, report as +, ++ or +++.
o Where + indicates 1 to 9 observed findings per field;
o ++ indicates 10 to 100 observed findings per field;
o +++ indicates above 100 observed findings per field.
• If no cells or parasites are seen report as “No parasites or cells seen.”
C. Results interpretation
38
2.4 PROCEDURE FOR STOOL ROUTINE EXAMINATION
2.4.1 Purpose
2.4.2 Scope
2.4.2.1 Responsible
2.4.3 Principle
Normal saline retains the morphology of the organism in its natural shape and color,
free helminthes eggs from debris.
Iodine - stains the internal structure of cyst to brown/yellow color so to allow the study
of cyst morphology. It also kills the organisms to allow internal structure easily seen.
Stool collected in clean plastic screw capped container which has spoon like. Include
macroscopic worms or worm segments as well as bloody and mucoid portions of the
sample.
2.4.5 Equipment
2.4.5.1 Maintenance
39
2.4.5.2 Materials
Normal saline, 1% Iodine, 10% formalin, Wooden applicator stick, Grease pencil,
Gloves, Microscopic slides, Cover slip, Marker and Stool container
Stool sample should be at the laboratory within two hours after collection. If a liquid or
soft stool sample can’t be examined within 30 minutes of passage, place it in a
preservative; if a formed stool sample can’t be examined immediately, refrigerate it or
place it in preservative.
2.4.6 Safety
2.4.7 Calibration
Quality control should be performed daily before processing patient samples by using
known positive and negative samples
2.4.8.1 Procedure
A.Macroscopic Observations
Observe the stool sample for the following; Color of the samples, Consistency,
Presence of blood, mucus, and, or, pus. Whether the samples contain worms.
40
iv. If stool is formed, the portion should include the inside and outside parts of the
sample
iv. For mucoid or watery stool, mix the entire contents before picking a portion to
mix with saline.
v. Cover the preparations with a cover slip.
vi. Start with 10x objective to systematically, examine the entire saline preparation.
vii. Use the 40x objective to assist in the detection and identification of parasitic
elements (eggs, ova, cyst etc). Always examine several microscope fields with this
objective before reporting ‘No parasites seen’.
viii. Use the iodine preparation to assist in the identification of cysts.
ix. Report the findings on Stool analysis worksheet.
x. Discard the used microscope slide on the container with disinfectant.
Not Applicable
2.4.10.1 Interpretation
2.4.10.1 Reporting
41
i. Colour of the samples.
ii. Consistency, i.e. whether formed, semiformed, unformed, watery.
iii. Presence of blood, mucus, and, or, pus. If blood is present note whether this is mixed
in the faeces. If only on the surface this indicates rectal or anal bleeding.
iv. Presence of worms, e.g. A. lumbricoides (large roundworm), E. vermicularis
(threadworm) or tapeworm segments, e.g. T. solium, T. saginata.
B.Microscopic Findings – Report the Following
• Report presence of any ova, trophozoites or cysts seen, specifying the species,
e.g. “Entamoeba histolytica trophozoites seen”.
• If no parasites are seen report as “ No ova or cysts seen”.
2.4.10.5 References
2.5.1 Purpose
42
2.5.2 Scope
2.5.3 Responsible
2.5.4 Principle
A. Giemsa stain
Giemsa stain is a Romanowsky stain contains methylene blue which is basic stain and
eosin is acidic stain, the parasites has DNA in the nucleus which is basic in nature and
RNA in the cytoplasm is acidic in nature, during staining the reactions takes place
whereby acidic part of a parasite will pick-up basic part of the stain and the basic part
of the parasite will pick-up the acid part of the stain, that’s why the nucleus of the
parasite will show red color and the cytoplasm will stain bluish.
B. Wet preparation
Whole blood collected in EDTA tube should be collected around midnight (22:00 –
04:00 for W. Bancrofti and 10:00 – 15:00 for L. Loa), as this is the time when parasite
is present in the blood for microfilaria worms.
2.5.6 Equipment
2.5.7 Materials
43
• buffer solution,
• Cover slips,
• Glass slide,
• Rack,
• Slide drying rack,
The Giemsa stock stain should be stored in a dark bottle and take precautions to avoid
moisture entering the stain.
2.5.9 Safety
2.5.10 Calibration
Positive and negative control samples are processed once a day in the morning before
patient samples.
44
vi. Examine the entire preparation microscopically for motile microfilariae using the
10× objective with the condenser iris closed sufficiently to give good contrast.
vii. Count the number of microfilariae in the entire preparation. Divide the number
counted by 10 to give the approximate number of microfilariae per ml of blood
(mf/ml).
viii. If unable to identify the species with certainty, continue with step 2 (examination
with 10% Giemsa stain under oil immersion objective).
i. Remove the cover glass and add a small drop of plasma, serum, or albumin
solution.
ii. Mix and spread thinly. Allow the preparation to dry completely.
iii. The addition of albumin, plasma, or serum (known to be microfilaria-free) will
help to prevent the preparation from being washed from the slide during
staining.
iv. Fix with absolute methanol or ethanol for 2–3 minutes.
v. Flood the films with 10% Giemsa solution and leave to stain for 15 minutes.
vi. Decant the Giemsa and wash in buffered water at pH 7.2 or distilled water
vii. Clean the back of each slide with cotton wool or gauze
viii. Place a drop of immersion on the thick film and place the slide on the
microscope stage
ix. Swing the x10 objective into position and bring the film into focus.
x. Swing the x100 oil immersion objective into position and focus.
xi. Examine the thick film systematically starting from the top left hand corner and
move from field to field.
A. Microfilaria
Species are identified by noting the arrangement of the nuclei towards the end of the
tail and the presence of sheath. W. bancrofti and Loa loa have a sheath; M. perstans
does not have a sheath. In Loa loa and M. perstans, the nuclei reach the tail tip and
the tail tip is rounded. In W. bancrofti, the tail tip is pointed.
B. Trypanosomes
Trypanosomes have elongated, flat, narrow bodies, often curved. In wet preparations,
they move rapidly by means of an undulating membrane and flagellum. In stained
preparations, the kinetoplast, a dark staining round body from which the flagellum
originates, is seen.
45
2.5.15 Limitation of the Procedure and Sources of Errors
2.5.18 References
46
CHAPTER THREE: BLOOD TRANSFUSION
3.1 PROCEDURE FOR ABO AND RHESUS BLOOD GROUPING
3.1.1 Purpose
This procedure provides instructions for performing ABO and Rhesus (D) blood
grouping using tube method.
3.1.2 Scope
This procedure is used in Blood Transfusion unit when performing ABO & Rhesus (D)
blood group typing for donors and patients.
3.1.3 Responsible
The head of Blood Transfusion and competent medical laboratory personnel are
responsible for ensuring this procedure is effectively implemented and maintained.
3.1.4 Principle
The D antigen on the red cells surface reacts which known D antibodies (anti-D) to
form agglutination which determines the Rhesus group of an individual, either as Rh
(D) Positive (agglutination) or Rh (D) Negative (no agglutination).
2-3mls of EDTA sample, 2-3mls of clotted sample from plain tube centrifuge the sample
at 3500rpm (RCF) for 5 minutes.
3.1.6 Equipment
47
3.1.7 Materials
Reagent Consumables
Anti –A Gloves
3.1.9 Safety
3.1.10 Calibration
Perform Quality Controls once in a week, when new antisera received and when new
control cells are prepared.
• Commercial anti-sera are quality controlled by reacting them with known cell
Suspension of A, B, O, Rh (D) Positive and Rh (D) Negative.
• Arrange 6 tubes labelled A, B, AB, O and D positive and D negative.
• Put one drop of 2-5% cell suspension of A, B, AB, O Rh (D) Positive and O
Rh (D) Negative to the corresponding tubes above
• Add one drop Anti-A, Anti-B, Anti-AB and Anti-D into corresponding tubes.
48
3.1.11.2 Controlling of O sensitized cells
cells Cells
Anti A + - + -
Anti B - + + -
Anti AB + + + -
Anti D +
LISS +++
AHG + -
Saline + to ++
Distilled Haemolysis
water
Key:
+ Agglutination
- No agglutination
49
3.1.12 Procedure Steps
A. Forward/Cell Grouping
i. For each patient/donor label 3 test tubes as Anti-A tube (A), Anti-B tube (B),
and anti-D tube (D).
ii. Arrange the labelled test tubes in the test tube rack
iii. Add one volume of anti A into tube A
iv. Add one volume of anti-B into tube B
v. Add one volume of anti-D into tube D
vi. To each of the above tubes add one drop of 2 - 5% cell suspension and mix
well.
vii. Centrifuge the three tubes at 1000rpm for one minute
viii. Examine the contents of the tubes for the evidence of agglutination.
ix. Read, interpret, and record the test results.
B. Backward/Serum Grouping
50
vi. If agglutination or haemolysis is observed at this stage, report result as Rh
(D) Positive and the procedure ends here.
vii. If there is still no agglutination, proceed as follows:
Not Applicable
- Means No Agglutination
Results report should include the ABO Type and Rhesus D reaction results, e.g. Blood
group A Rh (D) Positive, or Blood Group A Rh (D) Negative
51
3.1.16 Limitations of the Procedure and Source of Error
3.1.18 References
3.2.1 Purpose
3.2.2 Scope
3.2.3 Responsible
Trained qualified and competent certified registered medical personnel and other
authorised medical personnel.
3.2.4 Principle
A blood droplet is allowed to fall into copper sulphate solution of a specific gravity 1.053
and the movement of droplet is observed, If the specific gravity is higher than solution,
the drop will sink within 15 sec or else it will remain suspended for some time.
52
3.2.5 Sample Requirements
Capillary Blood
3.2.6 Equipment
Not applicable
3.2.7 Materials
• Reagents: 70% Ethyl alcohol, Copper Sulphate solution with specific gravity
of 1.053
• Consumables: Sterile swabs, Picker, Sharp container, Capillary tubes,
Gloves, Universal bottle, Waste containers and Timer
vi. Store and keep stock solution at room temperature in a tightly capped
brown glass bottle.
Copper Sulphate working solution stored at room temperature in brown bottle in three
months
53
3.2.11 Safety
3.2.12 Calibration
Not applicable
54
ix. Using the index finger and thumb, squeeze and hold the donor's·fingertip at
the upper joint tightly and prick
x. Wipe off the first drop of blood once with a dry swab
xi. Draw blood into the Capillary Tube
xii. Gently press the pricked fingertip to draw blood Note; Do not Squeeze the
pricked finger because it may introduce tissue fluid, dilute blood and give
false low Haemoglobin
xiii. Hold Capillary tube at approximately 60 degrees and let the blood flow into
the capillary tube up to not less than 3/4 capacity
xiv.Avoid air entering the capillary tube by ensuring smooth flow and
undisrupted flow of blood into the capillary tube
xv. Close the upper tip of capillary tube by placing your finger tip
xvi.Give clean dry swab to donor and instruct them to press finger with thumb
till bleeding stops.
xvii. Release a drop of Blood to Copper Sulphate Solution
xviii. Hold capillary tube at least 1 cm above the surface of copper sulphate
solution
xix.Release your finger tip to allow one drop of blood freely into labeled
container of fresh daily prepared and quality control copper sulphate
solution.
Not applicable
Not applicable
3.2.20 References
55
3.3 PROCEDURE FOR COMPATIBILITY TESTING
3.3.1 Purpose
This procedure provides instructions for performing compatibility testing, which is used
to select blood and blood components that will not cause harm to the recipient (patient)
3.3.2 Scope
3.3.3 Responsible
The head of Blood Transfusion and competent medical laboratory personnel are
responsible for ensuring this procedure is effectively implemented and maintained.
3.3.4 Principle
This method is based on immunophenotyping principle. The known red cell antigens
from donor are mixed with unknown antibodies from the recipient (patient) to detect if
there is any incompatibility caused by ABO, Rhesus and/or other blood groups
antibodies.
3.3.6 Equipment
3.3.7 Materials
Reagent Consumables
Incomplete anti-D (IgG) Test tubes, Test tube rack, grease pencil,
56
3.3.8 Storage and Stability
Venous blood must be used within 3 hours at room temperature then after should be
refrigerated at 2-8˚C. Blood Donor units should be stored at 1-6˚C
3.3.9 Safety
3.3.10 Calibration
Quality Controls for Antihuman globulin should be done on a daily basis before
performing patient samples..Internal quality control should be prepared from patient
samples or known organisation.
i. Select a unit of blood from the blood bank storage with the same group as the
recipient. If there is no unit with the same blood group as the recipient, blood
group O packed red cells can be used as universal donor. Also recipients with
AB blood group are considered as universal recipients of packed red cells.
ii. Label a tube for each donor red cell suspension being tested with the patient’s
serum.
iii. Add two drops of patient’s serum or plasma to each tube.
iv. Add one drop of donorcells suspension into appropriate test tubes
v. Rinse the Pasteur pipette 5 times during transferring of cells and serum to avoid
contamination
vi. Mix the contents of the tube(s) and centrifuge at 1000rpm for 60 seconds.
vii. Gently re-suspend the cell buttons and observe for haemolysis or agglutination.
viii. Read, interpret, and record test results. ix. If compatible, proceed with Indirect
Antiglobulin Technique
57
3.3.12.2 Interpretation and reporting:
Note: In emergency cases where the blood unit is required immediately, perform
Immediate Spin Saline Technique. Issue the blood unit if it is compatible then
proceed with Indirect Antiglobulin test. If there is any incompatibility,
immediately call the ward to stop the transfusion and re-call the blood unit.
Not applicable
Results should be reported in such a way that will indicate the recipient’s blood group
and the donor’s blood unit number to which the donor’s bloods is compatible or not
compatible.
58
3.3.18 References
Pam S. Helekar, D.P. Blackall et.al. American Association of Blood Bank 15 Edition,
1985. (method 3.1) and (method 3.2.1)
59
CHAPTER FOUR: HAEMATOLOGY
4.1 PROCEDURE FOR SICKLING SCREENING TEST
4.1.1 Purpose
The purpose of this procedure is to provide instruction for performing a screening test
to determine abnormal type of Haemoglobin called Haemoglobin S in blood.
4.1.2 Scope
This procedure is used by all trained laboratory staff while performing sickling test.in
the Laboratory
4.1.3 Responsible
Qualified, trained and competent Medical Laboratory Scientist, Technologist and
Technician are responsible for doing this test procedure.
Section heads are responsible for ensuring the effective implementation and
competency assessment for this procedure
4.1.4 Principle
When a drop of blood is sealed between a cover slip and a slide, the decline in oxygen
tension due to oxidative processes in the blood cells leads to sickling. In this method
added with blood drop chemical reducing agents such as sodium met bisulphite. This
rapidly reduces oxyhemoglobin to reduced haemoglobin, and then this will be
accelerating sickling.
4.1.6 Equipment
Light Microscope, and Hot plate
4.1.7 Materials
Freshly prepared 2% Sodium Metabisulphite (Diluting 0.2gm in 10ml of distilled water).
Vaseline/paraffin wax, Pipette, Cover glass, Glass slide and Applicator stick
60
4.1.9 Safety
Adhere to safety precautions as stated in the Safety manual/IPC guideline All personal
protective equipment (PPE) must be worn when performing this procedure.
All samples must be regarded as potentially infections.
4.1.10 Calibration
Not Applicable
61
4.1.17 Supporting Documents
Sample collection manual
4.1.18 References
Monica Cheesbrough Handwrite
4.2 PROCEDURE FOR URIT-12 HEMOGLOBIN METER
4.2.1 Purpose
This procedure is used to describe step by step on how to operate the URIT-12
Haemoglobin Meter using human whole blood sample in the laboratory.
4.2.2 Scope
This procedure is applied in testing hemoglobin parameter using human whole blood
sample in the haematology department/section in the laboratory
4.2.3 Responsible
A trained and competent laboratory scientist, laboratory technologists and assistant
laboratory technologists are responsible for performing this procedure.
The head of section or assigned personnel will be responsible for ensuring that this
procedure is effectively implemented.
4.2.4 Principle
The URIT-12 Hemoglobin Meter utilizes optical reflectance for determination of the
total hemoglobin. A drop of whole blood is applied to the test spot on the strip, blood
immediately disperses within the membrane contacting the reagent, then reaction
product could absorb spectrum in the range of 500nm – 600nm. The meter’s optical
detector automatically measures the change in membrane reflectance. The intensity
of reflectance is inversely proportional to the hemoglobin concentration. The meter
calculates and displays the total hemoglobin concentration in gram/decilitre (g/dL) in
12 seconds based on mathematical conversion.
4.2.6 Equipment
URIT-12 Hemoglobin Meter
4.2.7 Materials
The materials required in this procedure are Clean gloves, Laboratory coats,
Micropipettes, Cuvetes,Sharp container, Pricker and Alcohol swab
62
4.2.8 Storage and Stability
Anticoagulated blood is stable up to 72 hours at 2-8 ⸰c
4.2.9 Safety
• Personnel Protective Equipment must be worn at all times
• All samples must be treated as potentially infectious.
• Adhere to safety precautions as stated in the Safety manual/IPC guidline
4.2.10 Calibration
Not Applicable
63
4.2.15 Limitation of the Procedure and Sources of Errors
Only whole blood or EDTA anticoagulated blood should be used.
4.2.17 References
URIT – 12 Hemoglobin meter Operation Manual
4.3 PROCEDURE FOR DETERMINATION OF HAEMOGLOBIN LEVEL
USING HEMOCHROMAX PLUS
4.3.1 Purpose
The procedure provides instructions to laboratory staff on operation of Hemochroma
Plus Machine.
4.3.2 Scope
This procedure is applicable when performing quantification of Haemoglobin (Hb)
concentration in Hospital Laboratory.
4.3.3 Responsible
Qualified, trained and competent Medical Laboratory Scientist, Technologist and
Technician are responsible for doing this test procedure. Section heads are
responsible for ensuring the effective implementation of this procedure.
4.3.4 Principle
The hemochroma PLUS analyser utilizes a dual wavelength LED light sources by
which the haemoglobin absorbance is detected and converted into an electrical signal.
The signal is direct proportional to the amount of haemoglobin present in the blood
sample.
4.3.6 Equipment
Hemochroma PLUS machine
4.3.7 Materials
Hemochromax Plus Micro calibrator cuvette, Calibrator ID chip cuvettes, , Gauze, 70%
methylated spirit and Blood lancet or prickers
64
Anticoagulated blood is stable up to 3 days at 2-8 ⸰c
4.3.9 Safety.
• Adhere to safety precautions as stated in the Safety manual/IPC guideline
• All personnel must be worn protective equipment (PPE) when performing this
procedure.
• All samples must be regarded as potentially infections.
4.3.10 Calibration
It should be done when the machine is not working properly or when provides a doubt
result.
4.3.12 Procedure
i. Establish good relationship with the patient
ii. Make sure the patient is sitting comfortably
iii. Lightly massage to stimulate circulation. Only use the middle or ring finger.
The patient should not wear the ring on that finger iv. Pres lightly and draw finger-prick
blood into a micro cuvette by bringing the cuvette in contact with the blood drop on the
fingertip and puncture the side to a depth of cuvette.
v. Remove any excessive blood from the outside of the cuvette. vi. Insert the cuvette
containing blood sample to the Hb machine
vii. Wait until displaying of the test results and record the findings into the system
and register book
viii. Pull the cuvette holder out to its loading position and discard the used
microcuvette in sharp box
4.3.13 Biological Reference Interval
Infant 14.0 - 22.0g/dl, Children 11.1 - 14.1g/dl, Adult male 13.0 -17.0g/dl, Adult female
12.0-15.0g/dl
65
4.3.15 Limitation of the Procedure and Sources of Errors
i. Only whole blood should be used
ii. Air bubbles in the optical eye caused by inadequate filling of the cuvettes may
lead into false results
4.3.16 Performance Characteristics
Refer to method verification report
4.3.18 References
• Monica Cheesbrough. District laboratory practice in tropical countries, Part 2.
2000.
• Hemochromax PLUS package insert
• Hemochromax PLUS user manual
4.4 PROCEDURE FOR DETREMANATION OF HEMOGLOBIN LEVEL
USING HEMOCUE 201+ MACHINE
4.4.1 Purpose
This procedure provides instructions for the performance of Haemoglobin Estimation
using Hemocue 201+ machine.
4.4.2 Scope
This procedure applies to all competent laboratory staffs during determination of
haemoglobin level by using Hemocue 201+ machine.
4.4.3 Responsible
Qualified, trained and competent health laboratory practitioners in the laboratory are
responsible for implementation of this procedure.
4.4.4 Principle
The reaction in the microcuvette is a modified azidemethemoglobin reaction. The
erythrocytes are haemolysis to release the haemoglobin. The haemoglobin is
converted to methoglobin and the combined with azide to form azide methoglobin. The
measurement takes place in the analyser in which the transmittance is measured the
absorbance and haemoglobin level is calculated. The absorbance is directly
proportional the haemoglobin concentration
66
Capillary whole blood sample or Anti-Coagulated Whole blood (collected in EDTA
anticoagulant)
4.4.6 Equipment
Hemocue 201+ machine
4.4.7 Materials
Hemocue Hb 201+ microcuvette, Lancet/Pricker for capillary sample, Pipette or any
other transfer device for venous sample or control materials
4.4.9 Safety
• Personnel Protective Equipment must be worn at all times
• All samples must be treated as potentially infectious.
4.4.10 Calibration
Not Applicable
Note: The microcuvette should be filled within 3 minutes after it has been taken out of its
package.
67
Infant 17.0-22.0g/dl, Children 11.0-13.0g/dl, Adult male 13.0 -17.0g/dl, Adult female
12.0-15.0g/dl
4.4.18 References:
Monica Cheesbrough. District laboratory practice in tropical countries, Part 2. 2000.
Hemocue Hb 201+ cuvette kits insert and use manual.
4.5 PROCEDURE FOR CD 4 COUNT TEST BY USING BD FACS PRESTO
4.5.1 Purpose
The purpose of this procedure is to provide detailed information on how to analyse and
detect CD4 T Cell enumeration on blood sample by using BD FACS presto.
4.5.2 Scope
This procedure applicable in haematology to analyse and detect CD4 T Cell by using
BD FACS presto.
4.5.3 Responsible
Qualified, trained and competent Medical Laboratory Technician and Technologist are
responsible for doing this test procedure. Section heads are responsible for ensuring
the effective implementation and competency assessment for this procedure.
4.5.4 Principle
68
The BD FACS Presto™ cartridge the CD4/%CD4/Hb cartridge contains dried
fluorochrome-conjugated antibody reagents. When blood reacts with the reagents, the
antibodies in the reagent bind to the surface antigens on the BD FACS Presto
Cartridge: lymphocytes and monocytes. After the incubation period, the cells are
analysed on the BD FACS Presto Near-Patient CD4 Counter (the instrument). The
software identifies the cell populations of interest and calculates CD4 absolute counts,
CD4 percentages of lymphocytes, and haemoglobin concentration. The system
measures total haemoglobin by spectrophotometric method, using absorbance at an
isobestic point for oxyhemoglobin and deoxy-hemoglobin, with Correction for scatter.
4.5.6 Equipment
BD FACS Presto
4.5.7 Materials
Reagent Consumables
BD FACS presto Blue and yellow tips
cartridge BD FACS presto print out paper
BD disposable 100 µl Pipette
4.5.8 Storage and Stability
• Do not refrigerate whole blood SAMPLE before sample preparation.
• Do not use previously fixed and stored samples.
4.5.9 Safety
• Blood samples may contain infectious agents that are hazardous to your health.
Observe Standard Universal precautions.
• Ensure the instrument and environment you working are kept clean and free
from infectious substance such as human blood to avoid contamination.
• Spills should be immediately disinfected with 0.5% Sodium Hypochlorite
Solution.
4.5.10 Calibration Not Applicable.
4.5.11 Quality Control
BD Facs presto has internal electronic self-test for IQC or the use of known higher and
low CD4 counts made in-house can also be applied
69
ii. Face the inlet port up. iii. Invert the tube 10 times to mix the contents well. iv.
Use the pipette to obtain the sample.
v. Gently squeeze the bulb on the pipette to form a drop of blood on the tip of the
pipette. vi. Carefully dispense the sample into the inlet port. Hold the cartridge by its
ridges only.
vii. Make sure the blood reaches the top of the inlet port. If necessary, gently
squeeze the bulb on the pipette to dispense more blood.
viii. Make sure the cartridge is level, with the barcode side up, at all times. Make
sure that blood appears in the part of the channel not covered by the channel protector,
next to the containment zone.
ix. Discard the pipette into a biohazard us waste container.
x. Close the cartridge cap securely and Set the on-board timer. xi. Place the
cartridge, barcode side up, on the work station
xii. Press the Run Test tab. xiii. Press Patient ID.
xiv. Enter the patient’s ID and press Accept. xv. The Confirmed Patient ID
screen opens. xvi. Press Accept and Insert the cartridge:
xvii. Select your Operator ID and press Accept. Then cartridge door on the
instrument opens will open.
xviii. Note: If possible, complete the following two steps within 30 seconds. xix.
Remove the channel protector from the cartridge. xx. Hold the cap with the
channel facing upwards.
xxi. Insert the prepared cartridge into the cartridge door. The cartridge door closes.
xxii. Press Accept to eject the cartridge within 30 seconds.
4.5.13 Biological Reference Interval
Analyte Gender Reference interval SI UNIT
Absolute CD 4 count Male 462 – 1306 cells/uL
Female 440 – 1602 cells/uL
%CD4 of lymphocytes Male 29 – 54 %
Female 32 – 55 %
HAEMAGLOBIN Male 13.5 – 18.0 g/dL
Female 12.0 – 16.0 g/dL
4.5.14 Interpretation and Reporting of Results
Interpretation of results
• The results are displayed on the screen and print automatically.
Reporting of Results
• Report the obtained results as displayed on the screen.
Critical values
• If the CD4 count is low (below 200cells/ l for adult and below 450 cells/ l for
children) the result will be regarded as critical and communicate with clinician
4.5.15 Limitations of the procedure and sources of error
70
• Use the cartridge only with the BD FACS Presto instrument.
• The use of expired cartilage may result into false results
• Improper filling of the test device may not give the proper results
4.5.16 Performance Characteristics
Refer to Method verification report
4.5.18 References
Becton Dickinson BD FACS presto Operating Manual
4.6.2 Scope
This procedure is to be performed at point of care or any health facility to detect the
clotting time
4.6.3 Responsible
The Section heads are responsible section is responsible for ensuring the effective
implementation and maintenance of this procedure. Qualified, competent and
registered Medical Laboratory practitioners are responsible for implementing this test
procedure.
4.6.4 Principle
The presence of activator augments the contact activation phase of coagulation, which
stimulates the intrinsic coagulation pathway. Clotting time can be performed manually,
whereby the operator measures the time interval from when blood is injected into the
test tube to when clot is seen along the sides of the tube.
4.6.6 Equipment
Stop watch, water bath and thermometer
4.6.7 Materials
71
Disposable gloves, Laboratory coat, 70% alcohol, Masks, and Sterile lancets
4.6.9 Safety
• Adhere to safety precautions as stated in the facility Safety manual
• All personal protective equipment (PPE) must be worn when performing this
procedure.
• All samples must be regarded as potentially infections.
• Refer to National infection prevention and control Guidelines for health waste
management and safety practice.
4.6.10 Calibration
• Perform equipment calibration of the Stop watch, water bath and thermometer
as per auxiliary equipment calibration schedule
72
iii. This may be normal in patients taking anticoagulant therapy.
iv. This is usually normal when the intrinsic and common pathways are present in
an amount not exceeding 1% of the normal plasma level.
4.6.16 Performance Characteristics
Refer the method verification report of the procedure.
4.6.18 References
A manual of laboratory diagnostic tests. Edition 7, Lipipicontt William and Wilkins, by
Frances Talaska Fishbach, RN, BSN, MS, and MarshallaBrnet 11, RN, BSN, MS,
Ph.D.
4.7.2 Scope
This procedure is used in Haematology section when performing bleeding time
4.7.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
implementing this test procedure.
The Head Haematology is responsible for ensuring the effective implementation and
maintenance of this procedure.
4.7.4 Principle
Bleeding time is a medical test done to assess platelet function of a patient. It involves
cutting the underside of the subject's forearm, in an area where there is no hair or
visible veins. The cut is of a standardized width and depth, and is done quickly by an
automatic device. A blood pressure cuff is used above the wound, to maintain venous
pressure at a special value. The time it takes for bleeding to stop (i.e. the time it takes
for a platelet plug to form) is measured. Cessation of bleeding can be determined by
blotting away the blood every several seconds until the site looks 'glassy'.
73
4.7.6 Equipment
Timer, Thermometer, Light Microscope, Sphygmomanometer
4.7.7 Materials
Marker pen, Examination Gloves, Filter paper, 70% alcohol swabs and Sterile lancets
4.7.9 Safety
• Adhere to safety precautions as stated in the Safety manual/IPC guideline
• All personal protective equipment (PPE) must be worn when performing this
procedure.
• All samples must be regarded as potentially infections.
4.7.10 Calibration
Perform calibration of equipment (Timer and Thermometer) as per calibration schedule
74
NOTE; If the bleeding time is greater than 10 minutes, stop the test and apply pressure to the
wound. Report the results as greater than 10 minutes.
Reporting of results
Report results as: Bleeding time (in minutes)
Critical value
Findings greater than 10 minutes.
4.7.18 References
4.8 PROCEDUIRE FOR FULL BLOOD COUNT BY USING URIT BH – 40P
HAEMATOLOGY ANALYSER.
4.8.1 Purpose
This procedure provides instruction on how to operate urit BH-40P haematology
analyser in determining full blood count
4.8.2 Scope
This procedure applies in the Haematology section when performing FBC (Full blood
count) analysis using URIT BH - 40P Haematology Analyser.
4.8.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
implementing this procedure. The Head Haematology is responsible for ensuring the
effective implementation and maintenance of this procedure.
4.8.4 Principle
The URIT BH – 40P Automated Haematology Analyser is a multi-parameter, it can
display 21 parameters and 3 histograms. Analyser adopts electrical impedance
method for WBC, RBC and PLT test and colorimetric method for HGB test. The
electrical impedance method is based on non-conductivity of blood cells. When the
blood cells in diluents pass through the ruby aperture, resistance will change, based
75
on that we can get the counting and volume of blood cells. The colorimetric methods
to measure and calculate HGB. Add lyse into the diluents sample, and then RBC will
be dissolve and release haemoglobin. Then the haemoglobin combines with lyse to
form cyanohemoglobin. Measure the transmission light intensity of this compound in a
sample cup through the monochromatic light of 540 nm wavelength and then compare
it with the result in blank state to get the haemoglobin concentration (blank state refers
to the state that only has diluents in sample cup).
4.8.6 Equipment
• URIT BH-40 Automated Haematology Analyser
• Perform start up, maintenance, troubleshooting and shut down the URIT BH-
40 Automated Haematology analyser as per manufacturer’s instrument instructions
4.8.7 Materials
Diluent, Lyse, Probe detergent, Set of Controls kit, Marker pen, Examination Gloves
Capillary tube, Plane test tube, Thermal paper, and Gloves
76
• At main menu screen, click “Drain”; the diluents will be dispensed into the tube.
• Remove the tube, add 20µl of the blood sample to the tube, and gently shake
the tube to make them well mixed.
• After that present the well mixed sample under the aspiration probe; make sure
the probe touches bottom slightly.
• Press RUN key on the front panel and remove the sample after hearing beep
sound. The result will be available after analysis is performed. ii. Whole Blood
Mode
• Gently shake the tube to well mix the blood sample, then present the sample
tube beneath the probe, make sure the probe touches tube bottom slightly.
• Press RUN key and remove the sample after hearing beep sound. The results
will be available after analysis is performed.
iii. Ant coagulated Peripheral Blood Mode
• Gently shake the tube to well mix the blood sample, then present the sample
tube beneath the probe, make sure the probe touches tube bottom slightly.
• Press RUN key and remove the sample after hearing beep sound. The results
will be available after analysis is performed.
4.8.13 Biological Reference Interval
See annex 1.
4.8.18 References
77
URIT BH- 40P Operation manual
4.9 PROCEDURE FOR FULL BLOOD COUNT USING OF ABX PENTRA 80
HAEMATOLOGY ANALYSER
4.9.1 Purpose
This procedure provides instructions for operation and maintenance of ABX Pentra 80
analyzer for Full blood picture
4.9.2 Scope
This procedure applies to all Full Blood Count tests done on ABX Pentra 80 analyser
in the haematology section
4.9.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
implementing this test procedure.
The Head Haematology is responsible for ensuring the effective implementation and
maintenance of this procedure.
4.9.4 Principle
The ABX Pentra 80 is an automated haematology analyser used for counting and
differentiating the cellular components in whole blood using electrical impedance,
cytochemical staining, light scatter and spectrophotometer.
The principle behind cell counting is based on disruption of electric current as particles
pass through an orifice. An electric current applies on both sides of this orifice. Cells
do not conduct electric current, therefore their passage through the orifice leads to a
change of the electric current established between both electrodes. These electric
current differences are registered and increment a counter at every cell passage.
78
haemoglobin reducer (potassium cyanide) and then the haemoglobin measurement
follows Drabkin method with a 540nm reading. The integration of luminous intensity
transmitted is evaluated according to the BEER-LAMBERT formula. An enzymatic
liquid (ABX CLEANER) ensures the system cleanliness between every analysis and
prevents carryover between samples.
4.9.9 Safety
• Adhere to safety precautions as stated in the Safety manual/IPC guideline
• All personal protective equipment (PPE) must be worn when performing this
procedure.
• All samples must be regarded as potentially infections.
4.9.10 Calibration
Calibrate the ABX Pentra 80 analyser under the following conditions:
• Change of software
• Major component replacement
4.9.11 Quality Control
Run control QC materials (Low, Normal and High) before patient samples
79
v. Place the sample on the tube holder of analyser.
vi. Press the door of analyser inside to run the sample.
vii. The patient results will be printed automatically.
4.9.13 Biological Reference Interval See annex 1.
4.9.14 Interpretation And Reporting Of Results
• Normal: if the results are within acceptable range.
• Abnormal (pathological): if the results are out of acceptable range
• Results are automatically printed from the ABX Pentra 80 and then review by section
head.
• Communicate the Critical value with clinicians
Test Critical Value
Haemoglobin Less than 5 g/dl (50 g/L)
White Blood Count Less than 2.0 or greater than 18.0 cells/mm3
Platelet Count Less than 50,000/ mm3 or greater than
800,000/mm 3
80
Refer to the method verification report of this procedure.
4.9.18 References
ABX PENTRA 80 operator’s manual
4.10 PROCEDURE FOR FULL BLLOD COUNT USING SINNOWA HB - 7021
4.10.1 Purpose
This procedure provides instructions for operation and maintenance of SINNOWA HB
– 7021 Hematology Analyser.
4.10.2 Scope
This procedure applies to all Full Blood Count tests done on SINNOWA HB – 7021
analyser in the haematology section
4.10.3 Responsible
• Qualified, trained and competent health laboratory practitioners are responsible
for implementing this test procedure.
• The Haematology section head is responsible for ensuring the effective
implementation and maintenance of this procedure.
4.10.4 Principle
The instrument adopts the method of impedance to measure and count cells. The
conductivity liquid (mainly diluents) provides constant current source to the electrode
thus they circuit can form a steady impedance circulation. When cells pass through the
aperture, the conductive liquid is replaced by cells. Change of circuit resistance
produces electrical pulse. The amplitude varies when the cells of different size pass
through the aperture. Consequently the number and volume of cells pass through the
aperture can be calculated based on the amplitude.
4.10.5 Sample Requirements
2 to 4mls of whole blood collected in K3 EDTA tube
4.10.6 Equipment
• SINNOWA HB – 7021 Haematology Analyzer, Refrigerator
• Perform start up, maintenance, troubleshooting and shut down the SINNOWA HB –
7021 Haematology analyser as per manufacturer’s instrument instruction
4.10.7 Materials
Reagents consumables
81
SINNOWA Diluent Reagent, Laboratory coat, Biohazard waste container.
SINNOWA Lyse Reagent, 0.5% Sodium hypochlorite solution, Distilled
SINNOWA Detergent, SINNOWA water, Protective gloves, Methanol, Printing
Probe Detergent, paper
4.10.8 Storage and Stability
• Control kit is stable until expiry date and should be kept at 2 oC – 8oC
• Blood sample be kept at temperature between 2oC- 8oC for 7 days.
• Store reagents as per manufacturer instructions
4.10.9 Safety
• Adhere to safety precautions as stated in the Safety manual/IPC guideline
• All personal protective equipment (PPE) must be worn when performing this
procedure.
• All samples must be regarded as potentially infections.
4.10.10 Calibration
The SINNOWA HB – 7021Hematology Analyser System requires commercial
calibrator material or assayed whole blood for calibration.
Calibrate the SINNOWA HB – 7021Hematology Analyser under the following
conditions:
• Change of software
• Major component replacement
4.10.11 Quality Control • Run the three levels of control LOW, NORMAL and
HIGH” for “SINNOWA HB – 7021 Hematology Analyzer to ensure quality of results
Perform Quality Control:-
i. Before analyzing the samples
ii. After replacement of the reagents
iii. After maintenance component replacement, or a field service action
iv. If there is any doubt in accuracy of the test results
v. After a reagent lot number change
vi. After a software change
vii. Following calibration
viii. According to your laboratory’s quality control program
ix. According to manufacturer requirements
4.10.12 Procedural Steps
i. Turn on the power on rear panel and indicator shows red light
ii. The instrument initializes test program
iii. Diluents, lyses and detergent will be sucked and tubing system cleaned
iv. If initialization is finished the display shows all parameters WBC, RBC and PLT
82
v. Shift to select ID and then press OK
vi. Select the test selection from the drop down menu
vii. Gently mix the sample
viii. Open the sample tube and place it under the Open Mode Probe
ix. Raise the tube until the end of the probe is deeply immersed in the sample.
x. Press the Touch Plate to activate aspiration
xi. Remove the tube when the beep sounds and replace the cap.
xii. When the cycle is finished, the results post to the Data log and are displayed in
the Run View
xiii. Print the results
4.10.13 Biological Reference Interval See annex 1.
4.10.14 Interpretation and Reporting of Results
• Normal: if the results are within acceptable range.
• Abnormal (pathological): if the results are out of acceptable range
• Results are automatically printed from the SINNOWA HB – 7021 and then review by
section head.
• Communicate the Critical value with clinicians
Test Critical Value
Haemoglobin Less than 5 g/dl (50 g/L)
White Blood Count Less than 2.0 or greater than 18.0 cells/mm3
Platelet Count Less than 50,000/ mm3 or greater than
800,000/mm 3
4.10.18 References
• SINNOWA HB – 7021 Hematology Analyser user manual.
83
This procedure provides instructions for operation and maintenance of BHA 3000
Haematology Analyser.
4.11.2 Scope
This procedure applies in haematology section for performing Full Blood Count tests
done on BHA 3000 analyser
4.11.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
implementing this test procedure. The Haematology section head is responsible for
ensuring the effective implementation and maintenance of this procedure.
4.11.4 Principle
BHA -3000 Automatic Hematology Analyser provides a 3 part differential blood count
uses an electrical impedance to count red blood cells, platelets and volume
distributions, and uses calorimetry to measure the haemoglobin and relevant
parameters will be enumerated.
This system uses electrical impedance method to count red blood cells, platelets and
white blood cells. When the absorbed quantitative sample is diluted by a quantitative
conductive solution, it is sent to the detection unit of the instrument. The detection unit
has a detection aperture, with a pair of positive and negative electrodes on both sides
of the aperture, which is connected with a constant current power supply. Due to the
bad conductor characteristic of these cells, when the cells in the diluted sample pass
through the detection aperture under a constant negative pressure, the DC resistance
between the electrodes will change, thus forming a pulse change at both ends of the
electrode in portions to the size of the cell volume. When cells continuously pass
through the aperture, a series of electrical pulses are generated at both ends of the
electrodes. The number of pulse is equal to the number of cells passing through the
aperture. And the pulse amplitude is proportion to the cell volume. Amplify the collected
electrical pulses, and calculates the number of electric pulse in the red blood/platelets
channel and WBC channel. The pulses are then classified according to different
channel voltage threshold, hence determines the cell volume distribution
4.11.6 Equipment
84
BHA 3000 Haematology Analyzer, Refrigerator
4.11.7 Materials
DIL-3 Diluent and HR-1 Lyse, Controls (Normal level (N), Low- level (L) and High- level
(H) Laboratory coat, Biohazard waste container. 0.5% Sodium hypochlorite solution,
Distilled water, Protective gloves, Printing paper
• Change of software
• Major component replacement
4.11.11 Quality Control
Run the three levels of control LOW, NORMAL and HIGH” for BHA 3000 Haematology
Analyzer to ensure quality of results
Perform Quality Control:
85
iv. If initialization is finished the display shows all parameters WBC,RBC and PLT
v. Shift to select ID and then press OK
vi. Select the test selection from the drop down menu
vii. Gently mix the sample
viii. Open the sample tube and place it under the Open Mode Probe
ix. Raise the tube until the end of the probe is deeply immersed in the sample.
x. Press the press Plate to activate aspiration
xi. Remove the tube when the beep sounds and replace the cap.
xii. When the cycle is finished, the results post to the Data log and are displayed in the Run
View
xiii. Print the results
4.11.13 Biological Reference Interval See annex 1.
4.11.14 Interpretation and Reporting of Results
• Normal: if the results are within acceptable range.
• Abnormal (pathological): if the results are out of acceptable range
• Communicate the following Critical value with clinicians
Test Critical Value
Haemoglobin Less than 5 g/dl (50 g/L)
White Blood Count Less than 2.0 or greater than 18.0 cells/mm3
Platelet Count Less than 50,000/ mm3 or greater than
800,000/mm 3
4.11.18 References
BHA 3000 Haematology Analyser user manual.
4.12 PROCEDURE FOR PERFORMING FULL BLOOD COUNT BY USING MS4
HAEMATOLOGY ANALYSER
4.12.1 Purpose
86
This procedure provides instructions for operation and maintenance of MS4 analyser
for Full blood count
4.12.2 Scope
This procedure applies to all Full blood count tests done on MS4 analyser in the
Laboratory hematology section
4.12.3 Responsible
• Qualified and trained Health Laboratory Practitioners are responsible for
implementing this test procedure.
• Section head is responsible for ensuring the effective implementation and
maintenance of this procedure.
4.12.4 Principle
The MS4-S is an automated haematology analyser used for counting and
differentiating the cellular components in whole blood using electrical impedance,
cytochemical staining, light scatter and spectrophotometer.
The principle behind cell counting is based on disruption of electric current as particles
pass through an orifice. An electric current applies on both sides of this orifice. Cells
do not conduct electric current, therefore their passage through the orifice leads to a
change of the electric current established between both electrodes. This electric
current difference is registered and increment a counter at every cell passage.
A chemical agent is used to separate erythrocyte and leukocyte populations, because
of size overlapping and quantities discrepancies. This chemical agent contained in
Lysis (ACTI-DIFF) pops the cytoplasmic membrane of the red cells. Erythrocyte
population disappears leaving the leukocytes. A Haemoglobin preservative is added in
lysing agent to measure Haemoglobin scaled down in a 540nm photometric tank at the
end of the counting. The Haemoglobin measurement is made from the first dilution.
The lysing agent has a powerful Haemoglobin reducer (potassium cyanide) and then
the Haemoglobin measurement follows Drabkin method with a 540nm reading. The
integration of luminous intensity transmitted is evaluated according to the BEER-
LAMBERT formula.
An enzymatic liquid (TRANSFLUX) ensures the system cleanliness between every
analysis and prevents carryover between samples.
87
2 to 4mls of whole blood collected in K3 EDTA purple top color tube
4.12.6 Equipment
MS4-S Hematology Analyser, and roller mixer
4.12.7 Materials
MS4-S analyser reagent pack, Lab coat, Biohazard waste container, 5-6% Sodium
hypochlorite, Distilled water, Protective gloves and Methanol
4.12.9 Safety
i. Treat all samples as potentially infectious.
ii. Keep hands away from the sample carrier when the analysis begins
iii. Some components inside the MS4s have sharp edges or angular corners, therefore
operate with caution to avoid cuts to the hands
iv. When performing maintenance procedure, take similar precautions as you would take
when handling patient samples
4.12.10 Calibration
All auxiliary equipment should be calibrated annually
88
ix. The QC results will be printed automatically
x. Plot the results of the QC on the Levy-Jennings chart and verify that the QC
has passed before running patients’ samples.
4.12.12 Procedural Steps
Running of Patient Samples
i. Press Analysis key in the machine
ii. Press RIGHT arrow
iii. Select the gender female or male by using UP and DOWN arrows (↑) or (↓)
iv. Press the ENTER key (using key board) or ( ) on the machine to validate
v. Write the name of the patient (Full Name)
vi. Mix well the sample
vii. Place the sample on the tube holder of the machine.
viii. Press ENTER on keyboard or ( ) on the machine to start running the sample.
ix. The patient results will be printed automatically.
4.12.13 Biological Reference Intervals
See annex 1.
Critical value/results should be recorded and immediately communicated to the
requesting clinician.
89
Method verification should be carried out and its report will be referred to fulfil this
requirement.
Laboratory quality policy manual, Laboratory safety policy manual and Laboratory
sample collection manual
4.12.18 References
90
CHAPTER 5: CLINICAL CHEMISTRY AND IMMUNOLOGY
5.1 PROCEDURE FOR BLOOD GLUCOSE BY USING ACCU-CHECK
GLUOCOMETER
5.1.1 Purpose
This procedure provides instructions for use of blood glucose test strips for
determination of Blood Glucose level. Blood glucose is measured mainly in the
diagnosis and maNot applicablegement of diabetes mellitus.
5.1.2 Scope
This procedure will be used for Blood Glucose testing in the laboratory and at point
of care testing sites (POCT).
5.1.3 Responsible
Qualified and competent registered and licensed Health laboratory practititioners and
Trained health care providers respectively are responsible for doing this test
procedure.
The head of clinical chemistry section is responsible for ensuring the effective
implementation of this procedure.
5.1.4 Principle
The test strips contain a capillary that sucks up a reproducible amount of blood.
Glucose in the blood reacts with an enzyme electrode containing Glucose oxidase
(or Glucose dehydrogeNot applicablese).The enzyme is reoxidized with an excess of
a mediator ferricyanide ion, a ferrocene derivative or osmium bipyridyl complex. The
mediator in turn is reoxidised by reaction at the electrode, which generates an
electrical current. The total charge passing through the electrode is proportioNot
applicablel to the amount of glucose in the blood that has reacted with enzyme.
91
ii. Post-prandial samples: This describes blood collected after a meal has
been taken. The sample is usually taken as a 2 hour post - prandial
samples.
iii. Random samples: This refers to a blood sample collected at any time,
regardless of food intake.
5.1.6 Equipment Glucometer Maintenance
Conduct maintenance as required by manufacturer instructions
5.1.7 Materials
Glucose test strips, Lancets, Cotton wool or gauze or alcohol swab, Sharp box or
Container, waste bin, Disposable gloves, Laboratory coat
5.1.9 Safety
i. All samples must be considered as potentially infectious and must be handled
and examined with care. ii. All person applicable protective equipment (PPE) should
be worn when performing procedure
iii. Adhere to safety precautions as stated in the Safety manual
Refer to National infection prevention and control Guidelines for health waste
management and safety practice.
5.1.10 Calibration
Equipment should be calibrated as per schedule.
92
v. Wear gloves clean the patient’s finger using the alcohol swab and allows it to
dry.
vi. Take one strip from the container. Close cap tightly and make sure the yellow
color in the round window on the back of the test strip matches the yellow
color above 0 mg/dL on the container. If it looks green do not use it.
vii. Insert the test strip, with the orange pad facing up, until it will go no further
into the meter. Do not bend the test strip. (The arrow heads are almost no
longer visible when the test strip is inserted correctly.)
viii. Make sure the code on the meter matches the code on the test strip container.
ix. When you see the flashing blood drop, hold the lancet device against the side
of patient fingertip and press the release button.
x. Gently squeeze patient fingertip to get a drop of blood. xi. Once the the blood
drop appears on the screen, you have 2 minutes to apply the drop of blood.
xii. Touch the blood drop to the center of the square orange pad. Do not bend the
test strip.
xiii. An hourglass symbol appears on the screen, and then the test result appears.
xiv. To remove the lancet, take off the lancet device cap and point the lancet end
away.
xv. Slide out the rejector to discharge the lancet into an appropriate container for
sharp objects.
xvi. Applying Blood with Test Strip outside of the Meter. xvii. Take the test strip
out of the meter
xviii. Touch the center of the square orange pad to the drop of blood. Do not bend the
test strip. xix. Within 20 seconds, insert the test strip, with the orange pad to the drop
of blood.
Critical value
93
Fasting blood glucose <2mmol/L
>20mmol/L
5.2.2 Scope
This procedure for Gluco-plus device will be used for blood glucose chemistry testing
in health facility in Tanzania
5.2.3 Responsible
Trained, qualified and competent laboratory registered practitioners are responsible
for performing this procedure.
The head of section for chemistry is responsible for ensuring the effective
implementation and competency assessment for this procedure .
94
5.2.7 Materials
Test strips and Lancing device
95
2 Press S button until you hear a beep sound. Press off button to save the
strip code.
3 Month number will blink use S and M button to select the required month
4 Press the off button to save month and date number will blink and use S
and M button to select the required date.
5 The time hours will blink use S and M to set time .
Performing a Quality Control test
1 Perform weekly and as per meter (machine) protocol.
2 Control used is GLUCOPLUS Selected Control solution.
3 Prepare and Apply Control solution.
4 Touch the sample tip of the test strip to the control drop. Verify check
window filled.
5 Meter display will count down from 5 to 1 and then display result.
6 Compare the control solution results with correct Control range printed on
test strips vial. (Example; 6.2 – 8.2 mmol/L). If the result are not within
control range repeat the control solution test.
Testing Blood
1 Use lancing device to puncture site, normally fingertip.
2 Insert a test strip into the strip port to turn on meter.
3 Touch the blood sample to the sample tip at the end of the test strip.
4 The meter display will count down from 5 to 1 and then display result.
5 If the “HI” or “LO” message appears on the meter, the result is above 33.3
or below 1.1. Repeat test to verify.
Error Codes
E-1 Problem with Meter – Do not use the meter
E-2 Meter or strip Problem – Repeat the test with New strip.
E-3 Meter was not ready – Repeat the test with a new strip. Apply SAMPLE
or CONTROL appears on the display.
E-4 Strip Problem – Repeat test with a new strip.
E-5 Strip problem or Sample too small – Repeat the test with a new strip
and new sample.
HI.E Temperature too high—repeat test in a cooler area.
LO.E Temperature too low—repeat test in a warmer area.
E-6 Battery LOW — replace battery soon.
5.2.13 Biological Reference Interval
Fasting Blood glucose (mmol/L) Random Blood Glucose
Blood/Plasma: 3.9 – 5.6 mmol/L (70 - 100 mg/dl) ≤6.9 (125 mg/dl)
5.2.14 Interpretation and Reporting of Results vii. Results are displayed in either
mg/dl or mmol/liter depending on which unit of measurement is selected.
Report the value in the agreed SI unit.
viii. If the result is lower than 10mg/dL (0.6mm/L) “Lo” is displayed instead of a
result.
96
ix. “Lo” may indicate that your blood is very low.
x. If the result is higher than 600mg/dL (33.3mmol/L), “Hi” is displayed instead
of a result.
xi. Fasting blood glucose between 5.6 – 6.9 mmol/L (100 – 125 mg/dl) indicates
high risk to diabtes. xii. Two separate test results of 7.0 mmol/L (126 mg/dl)
or higher indicate diagnosis of diabetes.
Critical value
Fasting blood glucose <2mmol/L
>20mmol/L
5.2.18 References
1. User manual for Gluco-plus
2. Manufacturers package insert
5.3 PROCEDURE FOR URIT 50 (URINE CHEMISTRY ANALYZER)
5.3.1 Purpose
This procedure is provides description for performing urine biochemistry by using
URIT 50 semi-automated Urine analyser.
5.3.2 Scope
This procedure is applied for testing Urine sample at the health facilities in Tanzania.
5.3.3 Responsible
A trained and competent health laboratory practitioners are responsible for
performing this procedure. The head of section of biochemistry and parasitology is
responsible of ensuring the implementation of this procedure.
5.3.4 Principle
The analyser measures change of the reflectance of reagent strips pads, A detector
integrated in the system is composed of light source and a light receiver, the light
from which goes through in the spherical integrator and reflect to the reagent pads
97
on strip. The absorbance (reflectance) varies according to the color of reagent pads,
the darker of the reagent pads higher the absorbance is and less light is reflected.
Conversely the lighter the reagent pad is the lower the absorbance is and more light
is reflected degree of color developed is direct proportion to the concentration of
analyte in urine” .
5.3.6 Equipment
Urit-50
5.3.7 Materials
Disposable gloves, Laboratory coats, Urine container, Waste container, Marker pen,
Urine strips from urit G10, G11 or G14, Gauze
5.3.9 Safety
Samples and control materials at this section should be treated as infectious material
and should be handled careful.
Note: Return calibrator strip into its container and discard other used materials according to
standard operating procedures
98
i. Deep the reagent strips of G series into urine sample and put it on the dry gauze
to remove excess urine on the back of the strip.
ii. Put the sample strip on the strip holder till D sound
iii. Read the patient result on the machine
iv. Record result to the register
5.3.13 Biological Reference Intervals See annex 3.
5.3.14 Interpretation And Reporting Of Results
Refer to the insert which present on the reagents strips of G series G10, G11, G14.
Report result according to insert present on the reagent strip bottle
5.4.2 Scope
This procedure is used in Clinical chemistry section of the user when performing urine
using the URIT-560 analyzer
5.4.3 Responsible
The section head of Clinical Chemistry is responsible for ensuring this procedure is
effectively implemented and maintained
5.4.4 Principle
The analyzer measures changes in reflectance of the reagent strips pads. A detector
integrated in the system is composed of a light source and a light receiver, the light
from which goes through spherical integrator and reflect at the reagent pads on the
strips. The absorbance varies according to the color of the reagent pads. The darker
the reagent pads is the higher the absorbance is and less light is reflected.
Conversely, the lighter the reagent pad is the lower the absorbance is, and more light
99
is reflected; ie. The degree of color development is proportion to the concentration of
analyte in urine.
The reflected light goes in to the optical-electronic detector system, which transforms
the optical into electrical. The strength of the electricity correlates which reflectance.
Then the electrical cables will be processed by CPU after being transformed by I/V
converter, and the test results can be printed out by printer.
5.4.6 Equipment
URIT-560, Printer
5.4.7 Materials
Urine strips (URIT G Series), Printer thermal paper, Urine container, Gloves
5.4.9 Safety
i. Personnel Protective Equipment must be worn at all times
ii. Samples must be treated as potentially infectious.
5.4.10 Calibration
Perform calibration as per manufacturer instructions
100
Bilirubin BIL Negative
Ketones KET Negative
Specific gravity S.G 1.003-1.029
Occult blood BLD Negative
Ph Ph 4.5 - 7.8
Protein PRO Negative
Nitrite NIT Negative
Leukocytes LEU Negative
5.4.14 Interpretation And Reporting Of Results
Results are automatically printed from the machine. Attach results printout with report
form
5.4.18 References
• URIT-560 operator’s manual
• Standard guard line for Health Laboratory 2007 Edition.
5.5 ROCEDURE FOR PERFORMING URINE BIOCHEMISTRY BY USING
CYBOW READER 300
5.5.1 Purpose
This Standard Operating Procedure (SOP) is aimed to describe step by step on how
to operate the CYBOW™ READER 300 semi-automated Urine analyser using Urine
sample at the health facilities in Tanzania
5.5.2 Scope
This procedure applies to all staff who works in parasitology section on performing
urinalysis test.
5.5.3 Responsible
101
A trained, qualified and competent laboratory registered practitioner are responsible
for performing this procedure. The head of sections is responsible for ensuring the
implementation of this procedure.
5.5.4 Principle
The CYBOW reader 300 are reflectance photometer. The strip is illuminated by white
light, and the reflected light from the strip is detected by the sensor. The RGB signal
is digitized, and this digitized image is interpreted by the processor. The intelligent
image analyser SW locates the strip and the pads, and based on this colour data the
parameter values are determined. The results including the date and the time of
measurement, sequence number and the ID are stored printed out by the internal
printer.
5.5.6 Equipment
Perfom the CYBOWTM READER 300 procedure for start-up, maintenance,
troubleshooting and shut down the urine analyser as per manufacturer’s instrument
instruction.
5.5.7 Materials/REAGENTS
Gloves, Marker pen, Laboratory coat, Waste container, Gauze, Urine container and
CYBOW strip.
5.5.9 Safety
• Adhere to safety precautions as stated in the Safety manual
• All personal protective equipment (PPE) must be worn when performing this
procedure.
• All samples must be regarded as potentially infections.
• Front cover of machine should be covered during operation to avoid sample
contamination.
• Used only power cord specified for CYBOWTM READER 300
• Avoid excessive dust, wet/damp condition and provide proper ventilation.
102
• Do not wipe the body clean with benzene, thinner, gasoline. This may
discolour
5.5.10 Calibration
The calibration of instrument should be done prior to first time use and then the 2 nd
calibration process is recommended in every 4 weeks with calibration strip provided
in the package.
103
• Select the quick mode by press direction key(◄) and press Enter
button
• Put the strips (incubation is done) once strip loading plate
continuously.
• Once the last strip is placed on the loading plate, press Enter button.
• Test result is shown on the LCD and automatically saved in the
memory
5.5.13 Biological Reference Intervals See annex 3.
5.5.14 Interpretation And Reporting Of Results
Chemical urinalysis
Report the reading when the immersed strip is compared to the colours on the strips
container
Macroscopic examination
Report whether the urine is Clear, Slightly cloudy, Cloudy or turbid. Report the colour
of the urine which will range from Light yellow, Yellow, Amber, Red to Brown
Microscopic examination of urine sediment
• White Blood Cells/Pus Cells/ leucocytes.
Report the average number of cells per High Power Field, example 2-5
WBCs/HPF
• Red Blood Cells
Report the average number of cells per High Power Field, example 2-5
RBCs/HPF
• Casts
Identify the type of cast and report as number per High Power Field
• Crystals
Identify the type of crystals and report their presence • Epithelial cells
Identify the type whether squamous, transitional or renal epithelial cells,
quantify them and report per high power field. Otherwise report the presence
of epithelial cells if can’t be identified.
• Trichomonas Vaginalis
Report as “seen” or “not seen”
• Yeast
Report as “seen” or “not seen”
• Spermatozoa
Report for males and not for females
104
If any delay happen put the urine sample in the refrigerator to avoid bacterial growth.
Do not use urine dipsticks beyond expiry date
5.5.18 References
Manufacturer’s Package Insert in multistrips kit
Cheesbrough, Monica Health Laboratory Manual for Tropical Countries
Graff’s Text Book for Urinalysis and body fluids, Second edition, Lillian A. Mundt and
Kristy Shanahan
5.6 PROCEDURE FOR DETERMINATION OF ALT BY USING DIRUI-DR 7000
CHEMISTRY ANALYZER
5.6.1 Purpose
This procedure provides instructions for determining Alanine Aminotransferase
(ALT) using the DIRUI-DR7000 Analyzer
5.6.2 Scope
This procedure is used in Clinical chemistry section for analysing ALT using the
DIRUI-DR7000 clinical chemistry Analyzer.
105
5.6.6 Equipment
DIRUI DR7000 Semi Automated Chemistry Analyzer, Centrifuge machine
5.6.7 Materials
Reagent Kits, Calibrator/Stsndard, Control Kits (LEVEL I and II), Disposable gloves
Laboratory coat, Sample Cups, Reaction Wells, Transfer Pipettes
5.6.9 Safety
i. Adhere to safety precautions as stated in the facility laboratory Safety
manual/IPC guidline
ii. All personal protective equipment (PPE) must be worn when performing this
procedure.
iii. All samples must be regarded as potentially infections.
5.6.10 Calibration
i. Perfom equipment calibration when; - ii. There is a change in the reagent lot
number. iii. If the QC result falls outside the acceptable ranges. iv. The machine
blinks on the QC-calibration indicating that the calibration is expired.
v. There is a change in the system software
vi. System maintenance/ component replacement procedure is performed
5.6.11 Quality control
QC shluld be performed before patient samples, after calibration of reagent
106
Result reporting
Once all of the results are accepted or validated, a final report will automatically be
printed out.
5.6.18 References
DIRUI-DR7000 Chemistry analyser user manual
5.7 PROCEDURE FOR DETERMINATION OF AST BY USING DIRUI-DR7000
CHEMISTRY ANALYZER
5.7.1 Purpose
This procedure provides instructions for determining ASAT using the DIRUI-DR7000
Analyzer
5.7.2 Scope
This procedure is used in Clinical chemistry section of the user when performing
blood analysis using the DIRUI-DR7000 clinical chemistry Analyzer
5.7.3 Responsible
The section head of Clinical Chemistry is responsible for ensuring this procedure is
effectively implemented and maintained.
5.7.4 Principle
Kinetic method for the determination of Aspartat-Aminotransferase (AST) activity
according to the recommendations of the Expert Panel of the International Federation
of Clinical Chemistry (IFCC). Without pyridoxalphosphate activation. AST is
measured by the reagent rate analysis by the coupled reaction with Malate
dehydrogenase (MDH) to reduce NADH (measured at a wavelength of 340nm) to
NAD+. The rate of decrease in absorbance at 340 nm due to NADH depletion is
proportional to the AST activity in the sample.
107
Serum or plasma, sample free from hemolysis and not contaminated
5.7.6 Equipment
DIRUI DR7000 Semi Automated Chemisry Analyzer•
Cleaning and Maintenance
i. Use a large amount of distilled water to rinse the tubing by click the rinse
interface.
ii. And drain the liquid from the tubing if necessary. iii. Remove the waste liquid
bottle from the back of the analyzer.
iv. Keep the instrument vertical during move and transport.
v. Try best to avoid vibration. vi. And check and debug the instrument before use.
5.7.7 Materials
Reagent Kits, Calibrator/Standard,Control Kits (LEVEL I and II), Supplies, Disposable
gloves, Laboratory coat, Sample Cups, Reaction Wells and Transfer Pipettes
5.7.10 Calibration
It is suggested to use supplementary calibrator as instructed. When lot number is
changed or QC is invalid, calibration shall be conducted again. Procedures for
reagent, Calibration, QC and Sample preparation
108
5.7.13 Biological Reference Intervals.
See annex 4.
5.8.3 Responsible
All qualified, trained and competent laboratory scientist, laboratory technologists and
assistant laboratory technologists are responsible for performing this procedure. The
head of section of biochemistry is responsible of ensuring the implementation of this
procedure.
5.8.4 Principle
The principle of the instrument is based on the phenomenon of different wave band
absorbance from substance, which is in line with Lambert-Bill Law.
(The greater the concentration of the sample, the more light is absorbed, the less
light is transmitted, and the darker of the color)
109
5.8.5 Sample Requirements
The 2 - 4ml of whole blood collected in plain tube (red top) for serum or EDTA (purple
top) for plasma or in heparinized tube free from hemolysis. Cerebral spinal fluid (CSF)
when required.
5.8.6 Equipment
SA-30 semi- automated chemistry analyser, Centrifuge
5.8.7 Materials
Micropipettes and tips, Marker pen, Thermal paper and PPE
5.8.10 Calibration
Quality Control procedure is the same as analytic procedure of unknown sample
(Use Randox Calibrator (pipette 500µl of reagent to tube and add 25µl calibrator mix
gently and incubate at 370C for 8-10 minutes for end point test method, no need of
extra incubation for kinetics method)
110
vi. Press PUSH button to aspirate distilled water, to calibrate AD value. The
AD value should be 45000 to 60000.
vii. Click “Continue” to test reagent blank. viii. Select “YES” to aspirate
reagent blank to test reagent blank absorbance.
ix. Press PUSH button to aspirate reagent blank to test reagent blank
absorbance.
x. Click “Continue” to test STD
Select “NO”, device will use last factor and perform the sample test directly.
xi.
Select “YES”, device will aspirate standard to test STD xii. Press
PUSH button to aspirate standard, then device will test standard
absorbance and calculate factor automatically
xiii. Click “Continue” to next to test the sample directly or perform control test.
xiv. Press PUSH button to aspirate sample or control, then device will test
the sample or control and display test result automatically
5.8.13 Biological Reference Intervals See annex 4.
5.8.14 Interpretation And Reporting Of Results
Interpretation
i. If the result of the particular parameter lies within the established
reference range, it means that the patient has normal particular
parameter.
ii. If the result of the particular parameter lies below or above the
established reference range, it means that the patient has abnormal
particular parameter and requires intervention as per the clinical history
and the laboratory findings. Reporting results
Report the results as they are displayed on the screen of the machine
111
5.8.16 Performance Characteristics Refer to method verification report
5.8.17 Supporting Documents
Sample collection manual, Safety manual, Quality manual
5.8.18 References
SA-30 semi automated chemistry analyser user manual
5.9 PROCEDURE FOR OPERATING CLINDIAG FA 200 CHEMISTRY TEST
5.9.1 Purpose
This procedure provides instructions for determining basic Chemistry tests using
Clindiag FA 200 analyzer.
5.9.2 Scope
This procedure is used at Clinical chemistry section for processing basic chemistry
tests using Clindiag FA 200 analyzer.
5.9.3 Responsible
Qualified and trained Health Laboratory Personnel are responsible for doing this test
procedure.
5.9.4 Principle
The test principle of the biochemistry analyser is mainly based on the lambert Beer
law.
The reagents and the samples to be tested are mixed at a certain proportion. The
mixture is placed in a calorimetric dish at a certain temperature for incubation, its
absorption of light of specific wavelength is continuously measured, and finally the
concentration of the measured substance is automatically calculated according to the
value of absorbance (change).The procedure uses Lambert Beers Law which state
that the absorptive capacity of a dissolved substance is direct proportional to its
concentration to the solution.
5.9.6 Equipment
Clindiag FA200 Chemistry analyser.
5.9.7 Materials
Reagent kits, cleaning solution, HD- high efficiency cleaning agent, Sample cups,
112
Calibrators and controls, Gloves, Laboratory coat, A4 paper, Micropipette,
Micropipette tips
5.9.9 Safety
i. Adhere to safety precautions as stated in the Safety manual. ii. All personal
protective equipment (PPE) must be worn when performing this procedure.
iii. All samples must be regarded as potentially infections. iv. Refer to National
infection prevention and control Guidelines for health care.
v. Avoid any contact between hands and eyes and nose during sample collection
and testing.
vi. Do not use kit beyond the expiration date. vii. Do not reuse the test device.
viii. All spills should be wiped thoroughly using 1% sodium hypochlorite
solution.
5.9.10 Calibration
Machine should be calibrated when
113
vi. Select the sample cup number
vii. Click OK
viii. Click Test on left side ix. Select test sample
x. Then click start test
xi. When sample processing is complete, Select the results from the menu bar,
patient result review will display, select name of patient on the left side.
xii. All the undone tests will be shown on the results as NA with a reason on the
right side. xiii. Re-running patient sample, repeat the patient order with no
change in a sample cup number
5.9.18 References
Refer to equipment instruction manual Clindiag FA 200
114
5.10 PROCEDURE FOR DETERMINATION OF IMMUNOASSAYS BY USING
GETEIN 1100 IMMUNOANALYSER
5.10.1 Purpose
This procedure provides description on determination of immunoassays by using
Getein 1100 Immunofluorescence Quantitative analyser.
5.10.2 Scope
This procedure is used for processing and analysis of immunol assay tests in the
biochemistry department/section at the hospital laboratory.
5.10.3 Responsible
A qualified,trained and competent laboratory scientist, laboratory technologists and
assistant laboratory technologists are responsible for performing this procedure. The
head of section of biochemistry is responsible of ensuring the implementation of this
procedure.
5.10.4 Principle
The detection element scans the binding area and converts the optical signal to
electrical signal. The voltage variation between the test line and background has a
linear relationship with the antigen concentration which can be used to calculate the
concentration. In conclusion the antigen concentration in whole blood, plasma,
serum, urine can be calculated quantitatively according to optical signal of the test
line.
5.10.6 Equipment
Getein1100 Immunofluorescence Analyzer, stop watch
5.10.7 Materials
Disposable gloves, Micropipettes and its tips, Containers for waste segregation,
Marker pen, Getein test card
115
Calibrator, controls and test kit devices should be stored as per manufactures
instruction
5.10.9 Safety
i. Samples and control materials should be treated as infectious material ii. Worn
PPE all of the time while working
5.10.10 Calibration
Calibration of the assay should be peformed as per manufacturer instruction.
116
5.10.17 Supporting Documents
• Sample collection manual
• Material data sheet or reagent manual
• Safety manual
5.10.18 References
Getein 1100 user manual
5.11.2 Scope
This FIA8000 is an analyzer that used to measure biomarkers in human whole blood,
serum, plasma or urine samples.
5.11.3 Responsible
Qualified and competent registered Health Laboratory practititioners are responsible
for doing this test procedure. The head of section of chemistry is responsible for
ensuring the effective implementation and competency assessment for this
procedure.
5.11.4 Principle
The combination of the antigensin the sample, the gold-labelantibodyin the colloidal
gold pad or nitrocellulose membrane, and theantibody pre-coated on the test linecan
form a purplish red streak on the test line. The colour intensity of the test lineis
proportionate to the quantity of antigens detectedin the sample. The analysersystem
can obtain the photo-electric signal intensity of the complexby scanning the test line
with a photo-electric component. Then the voltage difference between the voltage of
the test line and the background is obtained. The voltagedifference has a linear
relationship with the antigen concentrationwhichcan be used to calculate the antigen
concentration. The relationship has been established and varying from the measured
parameter. In conclusion, the antigen concentration in whole blood, plasma, serum,
urine can be calculated quantitatively in one-step according to the colour intensity of
the test line.
117
5.11.6 Equipment
FIA 8000 Quantitative Immunoassay Analyzer
5.11.7 Materials
Test kit, Power source, Printing paper, QC Card, QC SD, Gloves
5.11.9 Safety
i. Decontaminate working surfaces twice daily, in the morning and afternoon ii.
Adhere to safety precautions as stated in the Safety manual
iii. All personal protective equipment (PPE) must be worn when performing this
procedure.
iv. All samples must be regarded as potentially infections.
v. Avoid any contact between hands and eyes and nose during sample collection
and testing.
vi. All spills should be wiped thoroughly using 1% sodium hypochlorite solution
vii. Decontaminate the biohazpus waste before disposal.
5.11.10 Calibration
Perfom equipment calibration when; -
i. There is a change in the reagent lot number.
ii. If the QC result falls outside the acceptable ranges.
iii. The machine blinks on the QC-calibration indicating that the calibration is
expired.
iv. There is a change in the system software
v. System maintenance/ component replacement procedure is performed
5.11.11 Quality Control
Quality Control (QC) card should be run before processing patient sample for each
day
118
viii. Open the test card slot.
ix. Carefully insert the test card into the slot, making sure that the arrows on the
test card are pointing in the same direction as the arrows on the analyzer.
x. Close the test card slot.
xi. Add the sample.
xii. Follow the on-screen instructions to add the sample to the test card. Be
careful not to overfill the sample well.
xiii. Touch the screen to start the test.
xiv. The analyzer will automatically begin to process the sample.
xv. Read the results.
xvi. The analyzer will automatically read the results and display them on the
screen.
xvii. The results may be displayed in a variety of formats, such as quantitative
results, qualitative results, or graphs.
119
5.11.15 Limitation of the Procedure and Sources of Errors
Haemolyzed sample should not be used since the color changes caused by
haemolysis may result to wrong results. Refer to package insert for interfering
substances for specific test
5.12.2 Scope
This procedure is used in Clinical chemistry section for processing HbA1c,Lipid panel
and C-Reactive Protein.
5.12.3 Responsible
A trained, qualified and competent laboratory registered practitioners are responsible
for performing this procedure. The head of section for chemistry is responsible for
ensuring the effective implementation and competency assessment for this
procedure .
5.12.4 Principle
A Test Cartridge with patient sample or control is placed in the cartridge chamber of
the Analyzer. By manually closing the lid, the Test Cartridge is transported into the
analysis compartment of the Analyzer. Test and lot-specific information is obtained
from the barcode label (Figure 2). When the Test Cartridge enters the Analyzer, the
integrated camera reads the barcode. The calibration data for the actual lot are read,
which then initiates the processing of the Test Cartridge. The sample and reagents
are automatically transferred between the wells. An integrated camera monitors the
entire process. Light-emitting diodes (LEDs) illuminate the reaction area, which can
be either a colored membrane or a reaction well. The camera detects the reflected or
transmitted light, which is converted to a test result and displayed on the touch
120
screen. When the user accepts the result, the lid covering the cartridge chamber
opens automatically and the used Test Cartridge can be removed and discarded. The
Analyzer is then ready for the next run.
5.12.6 Equipment
Alere Affinion AS100
5.12.7 Materials
Test cartridge, Calibrators, Controls, PPEs, Pipette tips 100 -200ul, Pipette tips 100,
1000ul, Sample container rack, waste bin
121
5.12.12 Procedural Steps Analyzing a patient/control sample
1 Touch to enter the patient sample mode.
Touch to enter the control mode. A “C” in the upper left corner
indicates that the Analyzer is in the control mode. The lid opens
automatically.
If the lid is left open from the previous run and “Insert Cartridge” is
displayed, this step is omitted and you can start with step 2.
2 Insert the Test Cartridge with the barcode label facing left.
Be sure that the Test Cartridge is correctly placed in the cartridge
chamber.
3 Close the lid manually. The Analyzer will start processing the Test
Cartridge.
The processing time depends on the test in use.
6 Remove the used Test Cartridge from the cartridge chamber and
discard it in a suitable waste container. Insert a new Test Cartridge or
close the lid manually.
Keep the lid closed to protect the cartridge chamber when the
Analyzer is not in use.
122
5.12.15 Limitation of the Procedure and Sources of Error
i. Icteric samples that appears with yellow colour of the serum or plasma due to
bilirubin accumulation.
ii. Samples tested after 24 hours may give unreliable results Avoid using lipemic
samples
5.12.16 Performance Characteristics Refer the method verification reports .
5.12.17 Supporting Document
i. Sample Collection Manual, Safety Manual, Quality Manual
5.12.18 References
ALERE AFFINION AS100analyzer user manual
Manufacturer package insert
123
CHAPTER SIX: SEROLOGY
6.1 PROCEDURE FOR SYPHILIS ANTIBODIES RAPID TEST
6.1.1 Purpose
This procedure provides instructions for Qualitative detection of antibodies of all
isotopes against Treponema pallidum
6.1.2 Scope
The procedure is used in all Laboratory areas for screening syphilis infection
6.1.3 Responsible
Qualified and trained Medical Laboratory Technicians, Technologists and Scientist
are responsible for implementing this test procedure.
The Head serology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.1.4 Principle
The syphilis Ab Rapid test strip (Serum/plasma) is a lateral flow chromatographic
immunoassay based on the Principle of the double antigens–sandwich technique, In
this test syphilis recombinant antigen is immobilized in the test line region of the strip
test device, After sample is added to the sample well of the device it react with syphilis
recombinant antigen coated particles in the test. This mixture migrates
chromatographically along the length of the test strip and interacts with immobilized
syphilis antigens.
If the sample contains syphilis antibodies a coloured line will appear in the test line
region indicating positive results. If the sample does not contain syphilis antibodies a
coloured line will not appear in the region, indicating a negative result
6.1.6 Equipment
Centrifuge, Timer, Micropipette
Maintenance
Maintenance of the equipment should be performed as per schedule
6.1.7 Materials
124
Reagents Consumables
Syphilis Ab Rapid Test Strips kit Marker pen
Known Positive control, Examination Gloves
Known Negative control
125
Negative - Only one coloured band appears on the control(C) region. No
apparent band on the test (T) region
Positive - In addition to a pink coloured control (C) band, a distinct pink coloured
band will also appear in the tests (T) region
Invalid – a total absence of colour in both regions or no coloured line appears
on the control (C) region is an indication of procedure error or the tests
deterioration. Repeat the test with a new kit. Reporting of results
Not applicable
6.1.18 References
Manufacturer Kit insert for syphilis
6.2 PROCEDURE FOR (HIV) TESTING BY USING BIOLINETM HIV 1/2 3.0 TEST
6.2.1 Purpose
The purpose of this procedure is to describe the method of performing Bio line
HIV1/HIV-2 rapid test assay.
6.2.2 Scope
This procedure is applicable to all HIV-1/HIV-2 rapid test using Bio line HIV-1/HIV-2
6.2.3 Responsible
Qualified, registered, licenced and trained Medical personnel are responsible for
implementing this test procedure.
126
The Head of Serology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.2.4 Principle
SD Bioline HIV-1/HIV-2 is a rapid HIV qualitative immune-chromatographic assay
used to detect antibodies to HIV in human blood, serum or plasma as the sample had
been added to sample pad. As the sample migrates through the conjugate pad, it
reconstitutes and mixes with the selenium colloid-antigen conjugate. This mixture
continues to migrate through the solid phase to the immobilized recombinant antigens
and synthetic peptides at the sample window site. If antibodies to HIV-1 and/or HIV-
2 are present in the sample, the antibodies bind to the antigen-selenium colloid and
to the antigen at the client window, forming a red line at the client window site. If
antibodies to HIV-1 and/or HIV-2 are absent, the antigen selenium colloid flow past
the client window and no red line is formed at the client window site.
6.2.6 Equipment
Timer, Centrifuge, Micropipette, Refrigerator.
Maintenance
Maintenance of equipment should be performed as per schedule
6.2.7 Materials
Abbott Bioline TM HIV 1/2 3.0 Test/kit, Assay diluent, Disposable gloves, Laboratory
coat
6.2.9 Safety
127
Adhere to safety precautions as stated in the facility Safety manual /IPC guideline
All personal protective equipment (PPE) must be worn when performing this
procedure.
All samples must be regarded as potentially infectious.
6.2.10 Calibration
Perform calibration of equipment as per calibration schedule
128
Not applicable
6.2.18 References
Package insert Abbott Bioline TM HIV-1/2 3.0
129
The purpose of this procedure is to describe the method of testing HIV-1and HIV-2
using Trinity Biotech Uni-Gold test assay.
6.3.2 Scope
This procedure is applicable in all sites that perform Trinity Biotech Uni-Gold HIV test
6.3.3 Responsible
Qualified, registered, licenced and trained Medical personnel are responsible for
implementing this test procedure.
The Head serology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.3.4 Principle
Recombinant proteins representing the immune-dominant regions of the envelope
proteins of HIV-1 and HIV-2, glycoprotein gp-41, gp120 (HIV-1) and glycoprotein
gp36 (HIV-2) respectively are immobilized at the test region of the nitrocellulose strip.
These proteins are also linked to colloidal gold and impregnated below the test region
of the device. A narrow band of the nitrocellulose membrane is also sensitized as a
control region. Antibodies to HIV-1 and HIV-2 react with the colloidal gold linked
antigens. The antibody protein-colloidal gold complex moves chromatographically
along the membrane to the test and control regions of the test device.
6.3.6 Equipment
Stop watch, Micropipette, Centrifuge, refrigerator
6.3.7 Materials
Uni-GoldTM HIV test-kit, Disposable gloves, Laboratory coat, 70% alcohol
6.3.9 Safety
130
• Adhere to safety precautions as stated in the Facility Safety manual/ IPC
guideline
• All personal protective equipment (PPE) must be worn when performing this
procedure.
• All samples must be regarded as potentially infections.
6.3.10 Calibration
Perform equipment calibration as per Schedule
A pink/red line of the intensity adjacent to the letter ‘C’ (control). But no
pink/red line adjacent to ‘T’ (test) this indicates a Non-Reactive result. Invalid
results
131
No pink/red line appears in the device window adjacent to the letter “C”
(control) irrespective of weather or not a pink /red line appears in the device
window adjacent to “T” (test). This is an INVALID result that cannot be
interpreted. An invalid result must be repeated
Results reporting
• Reactive test results: Report as HIV TEST POSITIVE
• Non-reactive test results-repeat 1st and 2nd tests following the national
HIV testing algorithm
• If the test results are still discordant report - INCONCLUSIVE then
inform the patient for retesting after 14 days
• If after 14 days, the test is still discordant report - INCONCLUSIVE
Collect fresh venous blood sample, refer the sample for ELISA testing
Critical Values Not Applicable
6.3.15 Limitation of the Procedure and Sources of Error
i. Avoid hemolyzed sample and beware of lipemic samples when
interpreting results.
ii. The BIOLINE HIV-1/2 test is designed to detect antibodies to HIV-1 and
HIV-2 in human serum, plasma and whole blood. Other body fluids or
pooled samples may not give accurate results.
iii. Intensity of the patient bar does not necessarily correlate to the titer of
antibody in the sample. iv. A negative result with BIOLINE HIV-1/2 does
not exclude the possibility of an infection with HIV.
6.3.16 Performance Characteristics
Refer to the method verification report
6.3.18 References
Manufacture package insert (Trinity Biotech Uni-Gold HIV)
6.4.2 Scope
The procedure is used in the serology section in detection of pregNot applicablency
6.4.3 Responsible
132
Qualified,trained and competent Medical Laboratory Technicians, Technologists and
scientist are responsible for implementing this test procedure.
The Head Microbiology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.4.4 Principle
The Human Chorionic GoNot applicabledotropin One Step PregNot applicablency
Test Strip (urine) is rapid chromatographic immunoassay for the qualitative detection
of Human Chorionic GoNot applicabledotropin in urine to aid in early detection of
pregNot applicablency. The test uses two lines to indicate results. The test line is pre
coated with a monocloNot applicablel Human Chorionic GoNot applicabledotropin
antibody to selectively detect elevated level of Human Chorionic GoNot
applicabledotropin. The control line is pre coated with goat anti-mouse IgG antibody.
The test also includes a burgundy coloured conjugate paid containing another
monocloNot applicablel HCG antibody conjugated with colloidal gold. The assay is
conducted by immersing the test strip in a urine specimen and observing the
formation of coloured lines .The specimen migrate via capillary action along the
membrane to reach with coloured conjugate. Positive specimen reacts with the
specific antibody HCG coloured conjugate to form a coloured line at the test line
region of the membrane. Absence of this coloured line suggest a negative results
6.4.6 Equipment
Stop watch
Refrigerator
6.4.7 Materials
Reagents Consumables
HCG test kit Disposable gloves, Laboratory coat
133
6.4.9 Safety
i. Adhere to safety precautions as stated in the Facility Safety manual/ IPC
guideline. ii. All personal protective equipment (PPE) must be worn when
performing this procedure.
iii. All samples must be regarded as potentially infections.
6.4.10 Calibration
Perform equipment calibration as per schedule
134
ii. Very dilute urine specimen as indicated by low specific gravity may not
contain representative level of hCG. If pregnancy is still suspected the first
morning urine specimen should be collect 48 hours later and tested.
iii. first trimester pregnancies terminate for natural reasons, a test result that
is weakly positive should be confirmed by retesting with first morning urine
specimen collected 48 hours later.
iv. This test may produce false positive results. A number of conditions other
than pregnancy including trophoblastic disease and certain non
trophoblastic neoplasms including testicular tumours, prostate cancer,
breast cancer and lung cancer, causes elevated level of hCG. Therefore
the presence of hCG in the urine should not be used for the diagnosis
pregnancy unless this condition has been ruled out.
v. This test may produce false negative results. False negative results may
occur when the levels of hCG are below the levels of sensitivity level of
the test. When pregnancy is still suspected a first morning urine specimen
should be collected 48 hours later and tested. In case pregnancy is
suspected and the test continue to produce negative result see a physician
for further diagnosis.
6.4.16 Performance Characteristics
Refer to the method verification report of this procedure
6.4.18 References
HCG package insert
6.5.2 Scope
This procedure is applicable to all site perform hepatitis C Virus Antibody rapid tests
6.5.3 Responsible
It is the responsibility of the Head of serology Section to ensure effectively
implemented by all personnel working in the serology section.
6.5.4 Principle
The HCV Ab Rapid test Strip is a lateral flow chromatographic immunoassay based
on the Principle of the double antigen- sandwich technique. The test strip consists of
135
1) a burgundy colored conjugate pad containing HCV antigens conjugated with
colloidal gold (HCV Ag conjugates) and rabbit IgG-gold conjugates, 2) a nitrocellulose
membrane strip contain a test band (T band) and a control band (C band). The T
band is pre-coated with non- conjugated HCV antigens, and the C band is pre-coated
with goat anti-rabbit IgG. When an adequate of test specimen is dispensed into the
sample well of the strip, the specimen migrates by capillary action across the strip.
The antibodies: either the IgG, the IgM, or the IgA, to HCV if present in the specimen
will bind to the HCV Ag conjugates. The immunocomplex is then captured on the
membrane by the pre coated HCV antigens, forming a burgundy colored T band,
indicating a HCV Ab positive test result. Absence of the T band suggests a negative
result. The test contains an internal control (C band) which should exhibit a burgundy
colored band of the immunocomplex of goat anti rabbit IgG / rabbit IgG-gold
conjugates regardless the presence of any antibodies to HCV. Otherwise, the test
result is invalid and the specimen must be retested with another device.
6.5.6 Equipment
Timer, Centrifuge and Refrigerator
6.5.7 Materials
Reagents Consumables
HCV Ab test kit (Test strips, disposable Disposable gloves,
dropper and HCV Ab buffer) Laboratory coat
Micropipette
6.5.8 Storage and stability
a. The test device in the sealed pouch can be stored at 2-400C or as instructed
by manufacturer up to the expiration date. The test device must remain in the
sealed pouch until use. DO NOT FREEZE
b. Store whole blood at 2-80c for up 3 days
c. Serum and plasma maybe stored at 2-80c for up 7 days, for long term storage,
serum and plasma specimens should be kept at-200c or below.
6.5.9 Safety
i. Adhere to safety precautions as stated in the Facility Safety manual/ IPC
guideline
136
ii. All personal protective equipment (PPE) must be worn when performing
this procedure.
iii. All samples must be regarded as potentially infections.
6.5.10 Calibration
Listed equipment will be calibrated as per calibration schedule.
137
Invalid; No line appears in the control region. Under no circumstances should a
positive sample be identified until the control line forms in the viewing area if the
control line does not form, the test result is inconclusive and the assay should be
repeated
Reporting of results
Reactive test result - HCV Ab rapid test positive
Non - reactive test results - HCV Ab rapid test Negative
6.5.15 Limitations Of The Procedure And Source Of Error
i. The HCV Ab Rapid Test cassette (whole blood, serum, plasma) is for in
vitro diagnostic use only. This test should be used for the detection of
antibodies to HCV in whole blood, serum or plasma sample.
ii. The HCV Ab Rapid Test cassette (whole blood, serum, plasma) will only
indicate the presence of antibodies to HCV in the sample and should not
be used as the sole criteria for the diagnosis of hepatitis C viral infection.
iii. A negative result can occur if the quantity of the antibodies to HCV
present in the specimen is below the detection limits of the assay, or the
antibodies that are detected are not present during the stage of disease in
which a sample is collected.
6.5.16 Performance Characteristics
Refer to method verification report of this procedure
6.5.18 References
HCV Ab Package inserts
Kit manufacturer paper insert
6.6.2 Scope
This procedure will be used by all staff and students perform CrAg test
6.6.3 Responsible
138
It is the responsibility of the Head of serology Section to ensure effectively
implemented and maintained.
6.6.4 Principle
The CrAg Lateral Flow Assay is a dipstick sandwich immune-chromatographic assay.
Specimens and specimen diluent are added into an appropriate reservoir, such as a
test tube, and lateral flow device is placed into the reservoir. The test uses specimen
wicking to capture gold- conjugated, anti-Crag monocloNot applicablel antibodies and
gold conjugated control antibodies deposited on the test membrane. If Crag is
present in the specimen, then it binds to the gold conjugated, anti-CrAg. The gold
labelled antibody antigen complex continues to pick up the membrane where it will
interact with the test line, which has immobilized ant Crag monocloNot applicablel
antibodies. The gold labelled antibody-antigen complex forms a sandwich at the test
line causing a visible line o form. With proper flow and reagent reactivity, the wicking
of any specimen, positive or negative, will cause the gold- conjugated control
antibody to move to the control line. Immobilized antibodies at the control line will
bind to the gold conjugated control antibody and form a visible control line. Positive
test results create two lines (test and control). Negative test results from only one line
(control). If control line fails to develop then the test is invalid.
6.6.6 Equipment
Timer, Centrifuge and Refrigerator
6.6.7 Materials
Reagents Consumables
CrAg test kit (CrAg LF Test strip, LF Disposable gloves,
Specimen diluents, LF titration diluents, Laboratory coat
CrAg positive control) Micropipette
6.6.8 Storage And Stability
139
i. If a delay is encountered in sample processing, store at 2-8ºc for up to
72 hours is permissible.
ii. CSF, plasma, serum may be stored for longer period at -20ºc provided
they are not repeatedly thawed and refrozen
iii. Whole blood in transit should be maintained at 2-8ºc not -20ºc.
6.6.9 Safety
i. Adhere to safety precautions as stated in the Facility Safety manual/ IPC
guideline
ii. All personal protective equipment (PPE) must be worn when performing this
procedure.
iii. All samples must be regarded as potentially infections.
6.6.10 Calibration
Perform Equipment calibration as scheduled.
140
No visible band at all or there is visible band only in the test region and not in the
control region, Repeat the procedure.
Reporting of results
For the positive result report as Cryptococcal antigen rapid test Positive
For the Negative result report as Cryptococcal antigen rapid test Negative
6.6.15 Limitations Of The Procedure And Source Of Error
i. The assay Performance Characteristics have not been established for
matrices other than serum, plasma, whole blood and CSF
ii. Depending on the disease and organism prevalence, testing should not be
performed as screening procedure for general population. The predictive
value of a positive or negative serologic result depends on the pre-test
likelihood of cryptococcal disease being present. Testing should only be done
when clinical evidence suggests the diagnosis of cryptococcal disease
iii. Testing hemolyzed serum samples could lead false negatives due to the high
background color on the strip
iv. This assay was not evaluated for potential interference related to specimen
pre-treatment with 2-mercatoethanol or with specimens including the
following substances: Vaginal cream, caffeine, ascorbic acid, intraconazole,
amphotericin B, acetaminophen, or acetylsalicylic acid
6.6.16 Performance Characteristics
Refer to method verification report of this procedure
6.6.18 References
6.7.2 Scope
This procedure is applicable in all sites that perform Hepatitis B Virus surface antigen
in whole blood ,serum or plasma sample qualitatively.
6.7.3 Responsible
It is the responsibility of the Head of serology section and all laboratory personnel to
ensure effective implementation of this procedure.
6.7.4 Principle
141
HBsAg is an antibody sandwich immunoassay. Colloidal gold conjugated
monoclonal antibody reactive to HBsAg is dry-immobilized onto a nitrocellulose
membrane strip. When the sample is added, it migrates by capillary diffusion through
the strip rehydrating the gold conjugate. If present, HBsAg will bind with the gold
conjugate antibody to form particles. These particles will continue to migrate along
the strip until the test zone (T) where they are captured by ant-HBs antibody
immobilized there and a visible red line appears. If there is no HBsAg in sample, no
red line will appear in the T zone. The gold conjugate will continue to migrate alone
until is captured in the control zone (C) by immobilized goat, ant-mouse IgG antibody
aggregating a red line, to serve as an internal process control, a control band should
always be seen after test is completed. Absence of a colored control line in the control
region is an indication of an invalid result.
6.7.6 Equipment
i. Timer
ii. Refrigerator
iii. Centrifuge
6.7.7 Materials
Reagents Consumables
142
iii. All samples must be regarded as potentially infections.
6.7.10 Calibration
Perform Equipment calibration as scheduled.
143
6.7.15 Limitations Of Procedure And Source Of Error
i. HBsAg Rapid Test Kit detects HBsAg in human serum or plasma and
is only screening test. All reactive samples should be confirmed by
supplemental assays like PCR OR ELISA.
ii. A non -reactive result does not exclude the possibility of exposure or
infection with Hepatitis B virus.
iii. Patients with auto-immune liver diseases may show falsely reactive
results.
iv. This test is standardized to work best when the test procedure
mentioned in the package insert is strictly followed. Any deviation from
the test procedure may lead to erroneous results.
6.7.16 Performance Characteristics
Refer to method verification report of this procedure
6.7.18 References
HBsAg Package insert kit
6.8.2 Scope
This procedure is to be performed at point of care or any health facility
6.8.3 Responsible
The Head of Serology is responsible for ensuring the effective implementation and
maintenance of this procedure
Qualified, competent and registered Medical Laboratory practitioners are responsible
for implementing this test procedure.
6.8.4 Principle
It is a rapid chromatographic immunoassay for qualitative detection of specific
antigens to SARS-CoV-2. When the liquid sample is dropped on the sample pad, the
antigen in the sample forms an immunocomplex with the antibody labelled with
colloidal gold. Its complex moves along with the liquid sample, and makes a contact
144
with the antibody immobilized on the membrane, followed by forming an
immunocomplex with the immobilized antibody, resulting in generation of a coloured
red purple line. Appearance of red purple line on the membrane indicates the
presence of antigen in the sample. Since the liquid of the sample migrates through
the membrane very fast, it makes it possible to detect the presence or absence of
antigen within 15 minutes.
6.8.6 Equipment
Stop watch, Micropipette/Supplied capillary
6.8.7 Materials
Reagent Consumables
Test devices Buffer Disposable gloves
Laboratory coat
70% alcohol
Mask
6.8.8 Storage And Stability
Store Covid 19 rapid kit devoice 2-30OC Protected from sunlight and should not be
frozen
6.8.9 Safety
i. Adhere to safety precautions as stated in the facility Safety manual
ii. All personal protective equipment (PPE) must be worn when performing
this procedure.
iii. All samples must be regarded as potentially infections. iv. Refer to
National infection prevention and control Guidelines for health waste
management and safety practice.
6.8.10 Calibration
Perform equipment calibration as per schedule
145
i. Peel off the aluminium foil seal from the sample processing tube containing
the extraction buffer.
ii. After the sample collection, plunge the swab up and down in the sample
processing tube for at least 15 seconds, taking care not to spill the contents
out of the tube.
iii. Remove the swab while pinching wall of the tube with the swab and rotating
the swab, to extract the liquid from the swab. iv. Firmly attach the dropper lid
to the top of the sample processing tube.
v. Remove the test cassette from the sealed pouch. vi. Sample adding: Reverse the
sample processing tube, holding the tube upright, and slowly add 3-4 drops to the
sample ole (S) of the test cassette then start the timer.
vii. Timing observation: judge the result 15 minutes after sample adding; do not
observe the results after 30 minutes later.
viii. After the test, put the medical wastes into the biosafety bag.
6.8.13 Biological Reference Interval
Not Applicable
6.8.18 References
146
Manufacture package insert kit for SARS COV-2
6.9.2 Scope
This procedure is used in serology section when performing rapid Dengue antibody
rapid tests.
6.9.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
implementing this test procedure.
The Head serology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.9.4 Principle
Dengue test utilizes immunochromatography whereby mouse anti- human IgM and
human IgG antibodies are immobilized on the nitrocellulose membrane respectively,
as two individual test lines (IgM line and IgG line) in the test window of the test device.
The IgG line in the test window is closer to the sample well and followed by IgM line.
As the test sample flows through the membrane within the test device, the coloured-
dengue specific recombinant antigen-colloidal gold conjugate complexes with
specific antibodies (IgM and or IgG) of dengue virus if present in the sample. This
complex moves further on the membrane to the test region where it is captured by
the anti-human IgM and or human IgG antibodies coated on the membrane leading
to formation of a coated band, which indicates a positive test results. Absence of the
coloured band in the test window indicates a negative test result. A built in control
line will always appear in the test window when the test has performed properly
regardless of the presence or absence of anti-Dengue virus antibodies in the sample.
Dengue NSI antigen test is a solid phase immunochromatographic assay. As the test
sample flows through the membrane within the test device and mobilize the gold anti-
NSI conjugate that it is coated on the conjugate pad if NSI it is present then the result
it is the formation of coloured band of the test (T)line region
147
6.9.5 Sample Requirements
Plasma/whole blood, Serum (2-5ml)
NOTE 1: To obtain serum, Centrifuge blood collected in plain red top tube at
3000rpm/RCF per 3 minutes
NOTE 2: To obtain Plasma, Centrifuge blood collected in EDTA tube at
3000rpm/RCF per 3 minutes
6.9.6 Equipment
Timer ,Centrifuge ,refrigerator and Thermometer
6.9.7 Materials
Reagents consumables
• Dengue IgG/IgM antibody and NSI • Marker pen
antigen Cassette • Examination Gloves
• Buffer • Gauze
• Transfer pipette for dengue NSI
• Capillary pipette for
Dengue IgG/IgM
• Known Positive control,
• Known Negative control
6.9.8 Storage and Stability
i. The test kit should be stored at 15-25ºC in the sealed pouch for the
duration of the shelf-life (refer to manufacturer instruction)
ii. If the samples are not to be tested they should be refrigerated
immediately at 4 - 8 ºC
iii. If storage periods > 5 days the sample should be frozen at -200C
6.9.9 Safety
i. Adhere to safety precautions as stated in the Safety manual/IPC
guideline
ii. All personal protective equipment (PPE) must be worn when performing
this procedure.
iii. All samples must be regarded as potentially infections.
6.9.10 Calibration
Perform calibration of equipment (Timer ,Centrifuge
,refrigerator and Thermometer)as per calibration schedule
148
6.9.12 Procedural Steps
i. Bring the samples and the test components to room temperature if
refrigerated or frozen. Mix the sample well prior to assay
ii. Remove the test strip from the foil pouch and use it as soon as possible
iii. Use transfer pipette to transfer sample by depressing the bulb of the pipette
iv. If capillary used withdraw 5ul of sample, the black bar near the opening end
of pipette indicates the required 5ul of sample.
v. Drop the sample in the corner pointed by S1»
vi. Hold the pipette in a vertical position over the left “S” sample well on the
device
vii. Transfer 2 drops of sample into well
viii. Dispense 2 drops of sample buffer to the right ”S” sample well
ix. Read result at the end of 20 minutes
6.9.13 Biological Reference Interval
Not Applicable
6.9.14 Interpretation and Reporting of Results
Interpretation of results
NEGATIVE
• If only the “C” line is developed, the test indicates that no detectable
antibodies to dengue are present in the specimen
POSITIVE
• Lines showed to control and NSI –NS1 POSITIVE during window period
• Lines showed to control and IGM-IgM Positive Chronic dengue
• Lines Showed to control and IgG- IgG Positive during early infections BUT
has been treated
INVALID
If NO line is developed at “C”, the assay is invalid regardless of colour development
on the “T” line. Repeat the test Reporting of results
Report results as: Dengue - Negative or Dengue – Positive
Critical value
Positive findings
149
Refer manufacture kit insert for specificity and sensitivity. Also Method verification of
this procedure should be done and that the report should be referred to verify
compliance to this requirement. for Dengue rapid test procedure
6.10.2 Scope
This procedure is used in serology section when performing rapid Plague rapid
tests(F1RDT)
6.10.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
implementing this test procedure.
The Head serology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.10.4 Principle
F1RDT detect pathogen‐specific antigens (the F1 capsular antigen, which is part of
the outer surface of Yersinia pestis, the bacteria causing plague) in a small quantity
of different body fluids through lateral flow immunochromatography. The test is
simple to perform and provides a result within 15 minutes. It can be performed in the
pus contained in the buboes (swellings), or in the sputum (mucous coughed up from
the respiratory tract) of people with suspected pneumonic plague.
6.10.6 Equipment
Centrifuge, timer, thermometer
150
6.10.7 Materials
Reagents Consumables
F1RDT test kit • Marker pen
• Examination Gloves
• Gauze
6.10.9 Safety
i. Adhere to safety precautions as stated in the Safety manual/IPC
guideline
ii. All personal protective equipment (PPE) must be worn when performing
this procedure.
iii. All samples must be regarded as potentially infections.
6.10.10 Calibration
Perform calibration of equipment (Timer ,Centrifuge and Thermometer)as per
calibration schedule
151
6.10.17 Supporting Documents
Sample collection manual
6.10.18 References
Chanteau S, Rahalison L, Ratsitorahina M, Mahafaly, Rasolomaharo M, Boisier P, et
al. Early diagnosis of bubonic plague using F1 antigen capture ELISA assay and
rapid immunogold dipstick. International Journal of Medical
Microbiology 2000;290(3):279-83.
6.11.2 Scope
This procedure is used in serology section to all rapid H. pylori antigen tests.
6.11.3 Responsible
Qualified, trained, and competent health laboratory practitioners are responsible for
implementing this test procedure.
The Head serology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.11.4 Principle
This is a ready to use test that is based on the homogeneous membrane system
technology with latex microspores to perform the test, an aliquot of diluent sample is
added to sample well of the test cassette. The sample flows through a label pad
containing Pylori antibody coupled to red –colored colloidal gold. In the presence of
antigens, they bind to the antibody coated on the colloidal gold particles to form
antigen-antibody-gold complexes. These complex moves on the nitrocellulose
membrane by capillary action towards the test line region on which Pylori specific to
the antibody on the membrane in the form of a line. A second red control line will
always appear in the results windows to indicate that the test has been correctly
performed and the test device functions properly. If pylori is not present or lower than
the detection limit of the test, only the control line will be visible. If the control line
does not develop, the test in invalid.
152
6.11.5 Sample Requirements Stool sample.
6.11.6 Equipment
Timer
6.11.7 Materials
Reagents consumables
Test cassette dilution buffer Marker pen
-Known Positive control, Examination Gloves
-Known Negative control
153
viii. Do not read result after 15 minutes
6.11.13 Biological Reference Interval
Not Applicable
6.11.18 References
Manufacturer Kit insert for H pylori
6.12 PROCEDURE FOR HELICOBACTER PYLORI ANTIBODY RAPID TEST
6.12.1 Purpose
154
Rapid chromatographic immunoassay for the qualitative detection of antibodies (IgG)
anti-Helicobacter pylori (H. pylori) in human serum or plasma. It is used as an aid in
the diagnosis of infection with H. pylori
6.12.2 Scope
This procedure is used in serology section when performing rapid H. pylori antibody
tests.
6.12.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
implementing this test procedure.
The Head serology is responsible for ensuring the effective implementation and
maintenance of this procedure.
6.12.4 Principle
The H. Pylori rapid Test device is a lateral flow chromatographic assay based on the
Principle of the antibody – sandwich technique. The membrane is pre coated with
H.pylori Antigen on the test line region of the test. During testing H. pylori Antibodies
in the serum or plasma sample reacts with coated antigen and migrates upward on
the membrane chromatography by capillary action to the membrane and generate a
colored line. The presence of this colored line in the test region indicates a positive
result, while its absence indicates a negative result
6.12.6 Equipment
Timer, Centrifuge, refrigerator and Thermometer
6.12.7 Materials
Reagents consumables
Test cassette/strips buffer Marker pen
-Known Positive control, Examination Gloves
-Known Negative control
155
6.12.8 Storage and Stability
III. Store unopened test device at 2- 30 ̊C. If stored at 2-8 C̊ , ensure that
the test device is brought to room temperature before opening. Do not
freeze the kit or expose the kit over 30 C ̊ . (refer to manufacturer
instruction)
IV. Store sample at 2-8 C ̊ for up to 72 hours.
6.12.9 Safety vii. Adhere to safety precautions as stated in the Safety manual/IPC
guideline
viii. All personal protective equipment (PPE) must be worn when performing
this procedure.
ix. All samples must be regarded as potentially infections.
6.12.10 Calibration
Perform calibration of equipment (Timer, Centrifuge,
refrigerator and Thermometer)as per calibration schedule
156
14.2 Reporting of results
Report results as: H pylori - Negative or H pylori – Positive
14.3 Critical value Not applicable
6.12.15 Limitation of the Procedure and Sources of Error
• The test is for in vitro diagnostic use only
• The test should not be used as the sole criteria for the diagnosis of pylori
infection since it only indicates the presence of antibodies in the sample
6.12.16 Performance Characteristics
Refer manufacture kit insert for specificity and sensitivity. Also Method verification of
this procedure should be done and that the report should be referred to verify
compliance to this requirement. for H. pylori procedure
6.12.18 References
Manufacturer Kit insert for H pylori
6.13 PROCEDURE FOR BRUCELLA ANTIBODY DETECTION
6.13.1 Purpose
This procedure provides instructions for Qualitative detection of antibodies of all
isotopes against Brucella species.
6.13.2 Scope
The procedure is used in the serology section in the diagnosis of brucellosis.
6.13.3 Responsible
Qualified, registered and competent health laboratory practitioners are responsible
for implementing this test procedure. Section heads are responsible for ensuring the
effective implementation and maintenance of this procedure.
6.13.4 Principle
This test is based on antigen /antibody reaction. The smooth, attenuated stained
Eurocell antigen suspensions are mixed with the patient’s serum. Specific antibodies
to Brucella antigens if present in the patient serum will react with the antigen
suspensions to produce an agglutination reaction. No agglutination indicates the
absence of specific antibodies to Brucella antigens.
157
Serum sample is prefarable for this procedure. Allow blood to clot and Centrifuge the
sample at 3000rpm for 5 minutes
6.13.6 Equipment
Centrifuge, Pipettes, Shaker and Stop watch
6.13.7 Materials
Stained Eurocell-A/ Eurocell -M Antigen suspensions, Slide Test, 70%
alcohol, Known Positive control, Known Negative control, Marker pen,
Examination Gloves.
6.13.9 Safety
i. Adhere to safety precautions as stated in the facility Safety manual
ii. All personal protective equipment (PPE) must be worn when performing this
procedure.
iii. All samples must be regarded as potentially infections. iv. Refer to National
infection prevention and control Guidelines for health care services in
Tanzania
6.13.10 Calibration
Calibration of Centrifuge should be done as per schedule.
Maintenance
Maintenance of Centrifuge should be done as planned.
158
separate mixing sticks. viii. Gently rock the slide back and forth/observe for
agglutination macroscopically for one minute
6.13.13 Biological Reference Interval
Not applicable
6.13.18 References
Monica Cheesbrough: District Laboratory Practice in Tropical Countries, Vol 1,
Tropical Health Technology, 1998.
URIT 560 Urine Analyzer– User Manual for Urine Chemistry Analyzer
Practical Laboratory Manual-Jane Carter and Orgenes Lema
6.14 PROCEDURE FOR SALMONELLA TYPHI ANTIBODIES QUANTIFICATION
METHOD
6.14.1 Purpose
This procedure provides instructions for in vitro detection and quantitative estimation
of specific antibodies to salmonella present in Human serum
6.14.2 Scope
The procedure is used in serology section when performing widal test by slide
method.
6.14.3 Responsible
159
Qualified full registered and competent laboratory practitioners are responsible for
implementing this test procedure.
The Head of serology section is responsible for ensuring the effective implementation
and maintenance of this procedure.
6.14.4 Principle
This test is based on the Principle of direct agglutination reaction. The smooth
suspension of killed salmonella bacilli carries homologous O and H antigens. When
patient serum (containing antibodies to S typhi and S paratyphi) is incubated with
respective antigens, visible agglutination occurs. Arising titre of antibodies is
indicative of Enteric fever
6.14.6 Equipment
Timer, Centrifuge, Pipettes, Water bath
6.14.7 Materials
Stained salmonella antigen set, Stained salmonella antigen S typhi “O”, Stained
salmonella antigen S typhi “H”, Disposable gloves, pipettes, White Tile or slide,
Sample rack, Test tube rack, 5%Sodium hypochlorite to wipe and disinfect the spills
and Marker Pen
6.14.9 Safety
i. Decontaminate working surfaces twice daily, in the morning and
afternoon
ii.
iii. Adhere to safety precautions as stated in the facility Safety manual
iv. All personal protective equipment (PPE) must be worn when
performing this procedure.
160
v. All samples must be regarded as potentially infections. vi. Refer to
National infection prevention and control Guidelines for health care
services in Tanzania,
6.14.10 Calibration
Centrifuge should be calibrated as per schedule.
1.1. Maintenance
Maintenance of Centrifuge should be done as planed
161
Granular agglutination in case of “O” and flocculating agglutination in case of “H”
indicate positive reaction Quantitative slide/white tile test
A diagnostic titre of 1:80 suggest positive results
6.14.18 References
Reagent package insert for widal test
162
6.15 PROCEDURE FOR CHORELA RAPID DIAGNOSTIC TEST
6.15.1 Purpose
This procedure provides instructions for performing rapid Bioline Cholera Ag O1/
0139 test
6.15.2 Scope
The procedure is used for performing rapid Bioline Cholera Ag O1/ 0139 test in
Microbiology section.
6.15.3 Responsible
Qualified, trained and competent health laboratory practitioners are responsible for
performing this test
6.15.4 Principle
Cholera antigen test contains a membrane strip which is precoated with mouse
monoclonal anti- vibrio cholera O1 antibody on test line O1 region and with the
device. These line in result window are not visible before applying any specimen. The
control line is used for procedural control. Control line should always appear if the
test procedure is performed properly and the test reagents of control line are working.
A purple test line will be visible in the result window if vibrio cholera O1 and/or O139
antigens are present in the specimen.
When a specimen is added to the test, vibrio cholera O1 and O139 antigens in the
specimen react with colloidal gold-labelled V. Cholerae O1 and V. Cholerae O139
specific antibodies and form a complex of antigen-antibody colloidal gold conjugates.
As this complex migrates along the length of the result window by capillary action,
the complex is captured by the mouse monoclonal V. cholerae 01 antibody on test
line 01 (O1) and mouse monoclonal V. cholerae 0139 antibody in the test line 0139
(0139) across the result window and generates a colored line. In the absence of V.
cholerae 01 and 0139 antigen in specimens, a complex is not formed and no colored
test line appears in the result window of test device.
6.15.6 Equipment
Refrigerator
6.15.7 Materials
163
Disposable gloves, Timer, Biohazard bag, Test device, Sample collection tube with
extraction buffer, Sample collection swab for solid stool samples and Sample
collection dropper for liquid stool samples
164
b) Draw liquid specimens up to the fill line (about 300ul) using disposable
dropper.
c) Immediately transfer liquid specimen into the specimen collection tube.
d) Discard the dropper and then assemble filter cap on the specimen collection
tube.
6.15.13 Testing Procedure.
a) Bring the test device and sample to reach a temperature between 15-30⁰C
for at least 30 minutes in case they were refrigerated.
b) Remove the test device form the foil pouch and place it on a flat, dry surface.
c) Shake the collection tube thoroughly to ensure proper mixing of the sample
with extraction buffer.
d) Loosen the nozzle cap of the specimen collection tube.
e) Hold the collection tube vertically and dispense 3 drops(70ul) into the
specimen well of test device.
f) Wait a minimum of 15 minutes then read results. Do not read test results
after 15 minutes; reading results after 15 minutes can yield false results.
6.15.14 Biological Reference Intervals
Not applicable
165
6.15.16 Limitations Of The Procedure And Sources Of Error
a) A negative result does not exclude the possibility of V. cholerae 01 and/or
0139 infection in a patient. Failure to detect V. cholerae 01 and/or 0139 may
be a result of factors such as collection of specimens at an improper time in
the disease when few bacteria are present and improper sampling or handling
of the specimen.
b) A positive result does not preclude the presence of other enteric pathogens.
While the relationship between cholera and gastroenteritis is well established,
concurrent infection with other microbial pathogens is possible. Additional
microbiological tests should be performed in parallel with Bioline Cholera Ag
O1/ 0139 test kit in order to exclude other possible causes of the illness.
c) Test results should be interpreted in conjunction with information available
from epidemiological studies, clinical symptoms of the patient and other
diagnostic procedures.
6.15.17 Performance Characteristics
Refer to manufacture user manual
6.15.19 References
Refer to manufacture user manual
166
CHAPTER 7: BACTERIOLOGY AND MYCOLOGY
7.1 POTASSIUM HYDROXIDE WET MOUNT PREPARATION
7.1.1 Purpose
This procedure provides instructions for examination of wet preparations that is
mainly used to examine samples and cultures for motile bacteria, C.S.F for
capsulated yeast cells and for fungi.
7.1.2 Scope
This procedure applies to the microbiology section and health laboratory practitioners
in the laboratory settings.
7.1.3 Responsible
Competent Health Laboratory Practitionersare responsible for implementing this test
procedure.
The Head of Microbiology/who asned is responsible for ensuring the effective
implementation and maintenance of this procedure.
7.1.4 Principle
A KOH preparation is used for Samples such as skin scrapings, nail, infected hairs,
or for other Samples such as sputum to clear out background debris that may be
confused with fungal elements. KOH dissolves proteinaceous tissues, including
keratin, and renders them transparent. This enables fungi to be visualized more
easily. The use of KOH is not necessary for Samples such as CSF where background
debris is minimal.
167
7.1.8 Storage and Stability
Potassium hydroxide, 10 or 20% solution is stable for one year at room temperature
7.1.9 Safety
Observe standard safety precautions when handling Samples. Refer to Safety
Manual
Discard all used materials (e.g., sticks, pipettes, disposable forceps) in a bucket
containing bleach.
Note: Keratinous (skin, nails) Samples should be left at room temperature for 20 to 30 minutes
to allow digestion and “clearing” of the keratin, after clearing, press coverslip gently to make
a thin mount.
v. Examine the slide microscopically on low power and confirm observations with
high power. Observe for hyphae, conidia, budding yeasts, spherules or
sclerotic bodies, etc. Consult photomicrographs in appropriate references
when necessary, to identify fungal structure.
168
Artifacts, such as fibers, may be distinguished from hyphae from the lack of septa,
tapering ends, and size differences.
Reporting of Results
Report the type of fungal structure seen. Do not report quantity.
Examples: “Fungal elements seen (septate hyphae)”
“Fungal elements seen (yeast cells and pseudohyphae)”
Report negative results as: “No fungal elements seen”
Report as: “Fungal elements seen (Malassezia spp.)”
169
Guidelines on Standard Operating Procedures for Laboratory Diagnosis of
HIV/AIDS Opportunistic Infections in Patients. WHO Regional Office for
Southeast Asia. New Delhi. June 2001
Bailey and Scott’s Diagnostic Microbiology. 13th edition. Mosby, Inc. St. Louis,
Missouri, USA. 2013.
7.2.2 Scope
This procedure is to be used for detection of Acid Fast Bacilli in sputum or other body
fluid samples in the Laboratory
7.2.3 Responsible
The section heads and technical staffs are responsible for implementing this
procedure.
7.2.4 Principle
This procedure is used to stain mycobacterium tuberculosis and mycobacterium
leprae. These bacteria are also called acid fast bacilli. They stain with carbolfuschin,
which is a red dye. They retain the dye when treated with acid, which is because of
the presence of mycolic acid in their cell wall.
7.2.6 Equipment
Centrifuge (when necessary), Bunsen burner/spirit lamp, Light Microscope and
Biosafety cabinet
7.2.7 Materials
ZN stains reagents: 1% Carbol Fuchsin stain, 20% Sulphuric acid and 0.125%
Methylene blue. Acid Fast Bacilli Positive and negative Control slides for zn.
Personal protective gears, Gauze, Glass slides, Staining racks, Drying rack, Clean
Glass slides, Applicator stick, Micropipette or Pasture
pipettes,sputum container,timer
170
7.2.8 Storage and Stability
Sputum samples should be proccesed within 3days if stored at room temperature
and if not possible store it at 2 to 8°C for 10 days.
7.2.9 Safety
Decontaminate working surfaces as recommended by IPC Guidelines.
All personal protective equipment (PPE) must be worn when performing this
procedure.
All samples must be regarded as potentially infections.
Refer to National infection prevention and control Guidelines for health waste
management and safety practice.
7.2.10 Calibration
Centrifuge and safety cabinety should be calibrated as per schedule
171
xiii. Using forceps take the slide from rack, drain off water and stand the slide on
the edge to air dry on the drying rack.
xiv. Examine the smear microscopically first with the 40x objective to see the
distribution of material, then with the oil immersion objective to look for Acid
Fast Bacilli . Open fully the condenser iris when using the oil immersion lens.
After examining a positive smear, the oil immersion objective must be wiped
clean.
Note: The stained smear should show a light blue colour from methylene blue. If the smear is
dark blue, it usually indicates that the smear is too thick.
172
7.2.17 Supporting Document
Patients Results Register - TB 05
Internal Quality Control review form
7.2.18 References
Clinical Microbiology Procedures Handbook. American Society for Microbiology.
Washington D.C., USA, 2nd edition, 2007.
Monica cheesbrough (2005). District Laboratory Practice in Tropical countries.
Cambridge University Press, New York, USA, 2nd edition, 2005.
WHO, (2003). Mannual of basic techniques for a health laboratory. Geneva. 2 nd
edition, 2003.
Manual of Clinical Microbiology. American Society for Microbiology (ASM),
Washington D.C., USA. 9th edition, 2007.
7.3 PROCEDURE FOR AURAMINE O PHENOL STAINING
7.3.1 Purpose
The Auramine O staining technique applies to identification of Acid Fast Bacilli in
patient sample using fluorescence Microscopy which increase the rate of detection
compared to the light microscopy.
7.3.2 Scope
This procedure is to be used for detection of Acid Fast Bacilli in sputum or other body
fluid samples in the Laboratory by using Auramine O reagents.
7.3.3 Responsible
Competent Health Laboratory Practitioners are responsible for implementing this test
procedure.
The Head Microbiology is responsible for ensuring the effective implementation and
maintenance of this procedure
7.3.4 Principle
The fluorochrome dye, Auramine-Rhodamine, forms a complex with mycolic acids
contained in the acid fast cell wall of organisms which resist decolorization by
acidalcohol. The counterstain, Methylene blue 0.3%, renders tissue and its debris
non fluorescent, thus reducing the possibility of artifacts. The cells visualized under
ultraviolet light appear bright yellow or reddish orange.
7.3.6 Equipment
173
Biosafety cabinet, Fluorescent microscopes, Timer, Bunsen burner/ Spirit lamp
7.3.7 Materials
Auramine –O phenol 0.1%, Hydrochloric acid in Alcohol 0.5%, Methylene blue 0.3%,
Microscopic slides, Slide holding box, Gloves, Gauze/Cotton wool, Beaker, Forceps
Applicator sticks/Disposable, Pencil, Dry rack, Slide racks, wide mouth sputum
container for routine samples or Falcon tube 50ml for referral samples.
7.3.9 Safety
Decontaminate working surfaces as recommended by IPC Guidelines.
All personal protective equipment (PPE) must be worn when performing this
procedure.
All samples must be regarded as potentially infections.
Refer to National infection prevention and control Guidelines for health waste
management and safety practice.
7.3.10 Calibration
Centrifuge and safety cabinety should be calibrated as per schedule
7.3.12 Procedure
Film Preparation
i. Label the slides properly using a unique laboratory number
ii. Place the labeled slides, the samples and the applicator stick /Pasteur pipette
in the Biological safety cabinet
iii. Match each slide with the corresponding sputum or sample container. iv.
Proceed to smearing, taking the labeled slides and opening containers one by
one, do the smearing from the center of the slide, outwards making small coil
– like movements (A thin smear ,allow to air dry).
v. Select a small portion of purulent or muco-purulent material with the
applicator stick for a direct sputum smear and then transfer it to the slide, if a
stick is used, break it in two pieces and used ragged ends for dissecting
sputum and for smearing.
174
vi. Spread the material carefully over the area equal to about 2×1 cm using
repeated coil like movements, without touching the margins of the slides and
should be in the middle.
vii. Make the smear as even as possible by continuing this process until no thick
parts remain. Remove excess material with the second stick and discard in
the biohazard bag.
viii. The thickness of the smear should be such that a newspaper can bared bye
read through the dried smear held about 10cm above it (Translucent).
ix. Warm the slides on the slide warmer in the biological safety cabinet to dry
and fix the smear for at least 30minutes.
x. Re-fix the smears by passing a flame under the slides before staining.
xi. If the slide warmer is not functioning, air dry the smear then fix them by
passing a flame under the smear ,the smear should face upward ,do not over
heat ,or else Acid Fast Bacilli staining will be poor.
Fluorescent Microscopy Staining Procedure
Place the slides on staining rack. IT IS A MUST to keep distance of at least 1cm
between every slide. Otherwise there is a possibility that acid fast bacilli may cross
contaminate the negative smears due to over flooding or splashes from positive
smear to a negative smear
i. Cover the smear completely with filtered 0.1% auramine solution Do not
heat
ii. Leave for 15 minutes iii. Wash the slides well with distilled water or
running water iv. Pour the acid alcohol solution over the slides.
v. Allow to act for 2- 3 minutes.
vi. Gently rinse each slide with distilled or running water. vii. Repeat
decolourization if macroscopically visible stains are still visible.
viii. Flood smear with 0.5% methylene blue counter solution for 1 minute .Time is
critical because counterstaining for longer time may quench the acid fast
bacilli
ix. Gently wash off counterstain with distilled or running water.
x. Stand the slides on edge to drain and air dry on the slide rack away from
direct sun light.
Examination
i. Keep stained smears in the dark (box or folder) till reading, and read as soon
as possible (within 24hours) since fluorescence fades quickly out of the box.
ii. Use 20 x objectives for scanning and 40x confirmation; scan the stained smear
systematically from one side to another side
iii. One length has to be scanned before reporting a Negative. iv. Acid-fast
bacilli appear bright yellow against the dark background material.
v. Store the slides in a slide box according to the study, following the laboratory
Number as they will be needed for EQA.
175
Note: Acid fast bacilli appear bright yellow against dark background, report as possible for
Acid Fast Bacilli if at least one acid-fast bacillus was seen in a well stained smear, even if you
think they might be other mycobacterium other than tubercle bacilli. Tubercle bacilli are quite
variable in shapes from very short fragments to elongated types. They may be uniformly
stained or with one or many gaps, or even granular. They occur singly or in small groups
(coded), and rarely in large clumps. The typical appearance of bacillli are usually rather long
and slender, straight or slightly curved rods.
Critical values
Presence of Acid Fast Bacilli
7.3.18 References
176
International union against tubercle and lung diseases. The public health service
national tuberculosis Reference laboratory and the national laboratory Network
,Paris 1998
Smithwick R.W laboratory Manual for acid fast microscopy .US Department of
Health ,Education and Welfare,CDC,1979
Angra P, Becx-Bleumink M,Glipin C,et al,Ziehl Neelsen staining ,strong red on
week blue, or weak red under strong blue Int J Tuberic Lung Dis 2007:11:1160-
1
7.4.2 Scope
This procedure will be used by all laboratory personnel perform gram staining to
identify bacteria
7.4.3 Responsible
It is the responsibility of the Head of Microbiology Section to ensure effectively
implemented and maintained .
7.4.4 Principle
Gram stain based on the ability of bacteria cell wall. When the bacteria are stained
with primary stain (crystal violet) and fixed by the mordant (Iodine), Gram Positive
bacteria retain the primary stain when decolorized by ethanol (alcohol) because the
cell walls of gram positive bacteria have THICK layer of protein-sugar components
called peptidoglycan and low lipid contents. Upon decolourization Gram positive
bacteria causes thick cell wall to dehydrate and shrink, which closes the pores in the
cell wall and prevent the stain from existing the cell, therefore ethanol cannot remove
crystal Violet-Iodine complex that is bound to the thick layer which peptidoglycan and
appears BLUE. While Gram Negative bacteria are decolorized by ethanol. For the
gram negative bacteria cell wall takes up the crystal violet-iodine complex but due to
the thin layer of peptidoglycan and thick outer layer with form of lipids, crystal violate-
iodine complex gets washed off, when they are exposed to decolorizer dissolves the
lipids in the cell walls which allow the crystal violet-iodine complex to lead out of the
cells then when again stained with counterstain (safranin) they pick up the safranin
and appears red in color.
177
7.4.5 Sample Requirement
Fresh collected Pus, urine sediment, CSF, sputum, other body fluids
7.4.6 Equipment
Microscope, Timer, Bunsen burner/hot plate
7.4.7 Materials
Reagents Consumables
• Crystal violet as primary stain
• Lugol’s iodine as mordant • Gloves
• 10% Acid/Acetone as • Gauze
decolourizer • Waste bins
• Grass slides
• Neutral red/safranin as counter
• Applicator stick
stain
7.4.8 Storage and Stability
i. Samples are stable at 2-80C for 7 daysdays
ii. store reagentas instructed by manufacturer
7.4.9 Safety
i. Adhere to safety precautions as stated in the Facility Safety manual/ IPC
guideline
ii. All personal protective equipment (PPE) must be worn when performing this
procedure.
iii. All samples must be regarded as potentially infections.
7.4.10 Calibration
Perform Equipment calibration as scheduled.
178
viii. Rinse off iodine gently with flowing tap water ix. Decolorize by letting the
reagent flow over the smear while the slide is held at an angle or tilt slide.
x. Adjust decolourisations time to thickness of smear
xi. Remove excess decolorizer with gentle flow of tap water
xii. Flood with neutral red for 30 seconds
xiii. Remove excess neutral red with gentle flow of tap water
xiv. Drain slide and air dry in an upright position
Reading smears
Place the slide on the microscope and ensure that the smear is facing upward
179
CHAPTER EIGHT: MOLECULAR BIOLOGY
8.1 DIAGNOSIS OF MTB/RIF TESTING USING A TRUENAT MACHINE
8.1.1 Purpose
The purpose of SOP is to describe the stepwise procedure for the rapid detection of
Mycobacteria tuberculosis (MTB) and detection of rifampicin resistance using
Truenat MTB/RIF test for use with the Truenat Dx system is a semi-quantitative
realtime PCR in-vitro diagnostic test for: The detection of MTB DNA in sputum
samples, detection of rifampicin resistance-associated mutations of the rpoB gene.
The purpose of SOP is to describe the stepwise procedure for the rapid detection of
Mycobacteria tuberculosis (MTB) and detection of rifampicin resistance using
Truenat. MTB/RIF test for use with the TrueNat Dx system is a semi-quantitative real-
time PCR in-vitro diagnostic test for; detection of MTB DNA in sputum samples and
detection of rifampicin resistance-associated mutations of the rpoB gene.
8.1.2 Scope
This SOP describes the use of the Truenat™ MTB Plus assay, a chip-based
RealTime Polymerase Chain Reaction (PCR) test, for the semi-quantitative, detection
and diagnosis of Mycobacterium tuberculosis complex bacteria (MTBC) in human
sputum samples
8.1.3 Responsible
Qualified and competent health laboratory practitioners are responsible for
implementing this test procedure.
The section head is responsible for ensuring the effective implementation and
maintenance of this procedure.
8.1.4 Principle
The TrueNat assays utilize chip-based real-time micro-polymerase chain reaction
(PCR) for detection of TB and RIF-resistance from Deoxyribonucleic Acid (DNA) that
is extracted from sputum sample within an hour.
Truenat™ MTB Plus works on the principle of Real-Time Polymerase Chain Reaction
where the sputum sample is first liquefied and lysed thereafter the extracted DNA
from the sample is then amplified by the Truelab Real-Time microPCR analyser. The
purified DNA is then dispensed into the reaction well of the Truenat™ MTB Plus chip
and the test is started. Description of the apparatus
180
Trueprep AUTO Sample Prep device Truelab™ Real Time micro PCR Analyser
Figure1. Truenat Dx System hardware components
8.1.5 Sample Requirements
Sputum sample type, collected as either Spot or morning sputum sample
8.1.6 Equipment
True prep AUTO Sample Prep device, Truelab Real Time micro PCR Analyser and
Refrigerator
8.1.7 Materials
Extraction of Amplification of Materials required but Reagents and
DNA purified not provided Solutions
DNA
• Trueprep • Truenat™ • Liquefaction
AUTO v2 • Truelab Duo Positive buffer
Universal Real Time Control • Lysis buffer
Cartridge Quantitative Kit - Panel I • Conc. bleach
Based micro PCR • Powder free and 70%
Sample Analyzer disposable alcohol
Prep • Truenat™ MTB gloves
Device Plus micro PCR • Two waste
• Trueprep chip disposal
AUTO • Truelab™ micro containers, with
MTB PCR printer lids, containing
Sample • Truepet SPA bleach
Pre- fixed volume (6 solutions
treatment µl) Precision • Timer
Pack micropipette • Two waste
• Trueprep bags
• • DNase and
AUTO v2 • Micro tube
RNase-free
Universal Stand
pipette tips with
Cartridge • . Cartridge
filter barrier
Based Holder
Sample • Two cryovials
Prep Kit racks
Note: Connect a new reagent pack to the Trueprep Auto v2 device by inserting the Plug-in
Connector into the slot provided.
181
8.1.8 Storage and Stability store at room temperature for 3 days
Store at 2°C -8°C
8.1.9 Safety
• Treat all biological samples, including used cartridges, as if capable
of transmitting infectious agents. Because it is often impossible to
know which might be infectious, all biological samples should be
treated with universal precautions.
• Wear protective disposable gloves, laboratory coats and eye
protection when handling samples and reagents.
• Wash hands thoroughly after handling samples and test reagents.
Follow safety procedures for working with chemicals and handling
biological samples, (See safety manual)
• Dispose used cartridges according to infectious waste material
disposal guidelines.
8.1.10 Calibration
Calibration is perfumed as per schedule
182
iii. Add 2 drops of liquefaction buffer to the sputum sample
iv. Close the cap and swirl gently to mix
v. Incubate for 10 minutes at room temperature. If sample is not pipetteable after
10 minutes, incubate for another 5 minutes with swirling at 2-minute intervals
vi. Transfer 0.5 ml of liquefied sputum sample into the lysis buffer bottle using a 1 ml
transfer pipette
vii. Add 2 drops of liquefaction buffer into the lysis buffer bottle, swirl gently to
mix and incubate for 3-5 minutes
viii. Remove the cartridge from the pouch, label it and place it on the cartridge
stand. Take out the elute collection tube (ECT) and label it appropriately.
Keep it aside for later use. Keep the elute transfer pipette in the pouch for
later use.
ix. Transfer the entire contents of the lysis buffer tube to the sample chamber
(black cap) of the cartridge using 3 ml transfer pipette
x. Switch “on” the Trueprep® AUTO v2 device. Press “eject” button to open and
gently pull out the cartridge holder
xi. Place the cartridge in the tray, and gently push to close the cartridge holder.
Press “start.”
The device will beep at the end of the DNA extraction process (20 minutes),
and the cartridge holder will eject automatically.
xiii. Gently pull out the cartridge holder, remove the cartridge, and place it on the
cartridge stand.
xiv. Carefully pierce the elute chamber with the provided elute transfer pipette, and
transfer the entire elute into the ECT. Discard the transfer pipette and cartridge
C. Running a PCR TB Test
i. Switch “on” the Truelab microPCR analyser by pressing the red button in the
back right corner for 2 seconds. LED will glow in Green. Wait for 30-50
seconds for “boot-up screen” to appear followed by “home screen.”
ii. Select USER ID, enter password. Press “Sign in” to Log in
iii. Select test profile “MTB” or “MTB Plus”. To confirm selection tap “PROCEED”
and enter patient details (referred by, patient ID, gender, patient name & age)
iv. Select sample type (sputum).
v. Open a TRUENAT™ MTB Plus chip pouch *Pull out the orange desiccant
pouch and confirm that it is orange in colour.
vi. Gently take out the chip without touching white well portion and place it on the
chip tray by aligning it in the slot provided
vii. Press “START TEST” on the screen. Chip tray opens. “Please Load Sample” will
appear. (Don’t press “YES” until chip loading is complete.
viii. Open the master mix tube, discard the stopper and place the tube in the micro
tube stand. *Check for white cake at the bottom of the micro tube.
183
ix. Attach the 6ul micro tip provided in the pouch to the single push pipette.
x. Transfer 6ul of the elute from ECT into the master mix tube
xi. Allow the master mix to stand for 30 SECONDS to get a clear solution. *Do
not mix by tapping, shaking or reverse pipette. *Do not discard the pipette tip.
xii. Transfer the elute from the master mix tube to the white reaction well of the
chip (Figure 16). *Avoid spillage of the clear solution outside the white reaction
well. *Discard the pipette tip and master mix tube.
xiii. Click “YES” on the device screen to start the test. The PCR will be completed
in 35 minutes.
xiv. Tap the “Open/Close Tray” button to eject the chip tray and discard the used
chip immediately after the reaction.
xv. If MTB is detected test the same elute for RIF resistance using the Truenat
MTB RIF Dx chip as a follow-on test. The test takes about 55 minutes.
D. Running a RIF-Resistance Test
i. If MTB is detected in a sample, Run a RIF resistance test.
iii. Use a portion of the same DNA eluate to test for RIF resistance using a
Truenat MTB-RIF Dx chip.
iv. Start by returning to Step 3 in the PCR TB test process and repeat for
RIFresistance by Selecting “MTB RIF” as the test type in the Truelab micro–
PCR Analyser.
v. RIF-resistance testing takes an additional 60 minute.
8.1.13 Biological Reference Interval Not applicable
8.1.14 Interpretation and Reporting of Results
Interpretation of results
At the end of the test run, the result screen will display;
NOTE: Invalid samples have to be repeated with fresh samples from the sample
preparation stage.
IPC will co-amplify in most positive cases also, in some samples having a high
target load, the IPC may not amplify, however, the test run is still considered valid
Sample with MTB DETECTED should be tested for MTB RIF
Reporting of results
Click “VIEW RESULTS” on the menu bar. The View Results window appears.
184
Optional: Press “Print” to print the result page using Truelab® microPCR printer.
Critical value
MTB Detected RIF Resistance Detected
185
This procedure provides instructions for performing sample which is suspect with
MTB/Rif
8.2.2 Scope
This procedure is used for detection of the Mycobacterium tuberculosis complex
bacteria and their rifampicin susceptibility using the Gene Xpert MTB/Rif system in
microbiology section in the Laboratory.
8.2.3 Responsible
Qualified, trained, Competent and Registered health laboratory personnel is
responsible for ensuring the effective implementation for this procedure.
8.2.4 Principle
The Gene Xpert MTB/RIF system is a fully automated nested real-time PCR system,
which detects MTB complex DNA in smear positive and negative sputum samples
and other body fluid i.e. pleural fluid, ascetic fluid CSF and Pus. It simultaneously
identifies mutations in the rpoB gene, which are associated with rifampicin
resistance.
8.2.7 Equipment
• Gene Xpert machine, Microscope, Personal Protective Equipment such as N95
respirator and Timer
186
i. Decontaminate working surfaces twice daily, in the morning and afternoon
ii. Adhere to safety precautions as stated in the Safety manual
iii. All personal protective equipment (PPE) must be worn when performing
this procedure.
iv. All samples must be regarded as potentially infections.
v. Avoid any contact between hands and eyes and nose during sample
collection and testing.
vi. Do not use kit beyond the expiration date which appears on the package
label or device in a damaged pouch.
vii. All spills should be wiped thoroughly using 1% sodium hypochlorite
solution.
8.2.10 Calibration
Calibarate auxillary equipment used in this operating this procedure once a year and
kep record available is performed once per year.
187
• Using the sterile transfer pipette provided, aspirate the liquefied
sample into the transfer pipette until the meniscus is above the
minimum mark (= 2ml). Open the cartridge lid. Transfer sample into
the open port of the Xpert MTB/RIF cartridge
NOTE: It is crucial that no bubbles are created when transferring the sample
into the cartridge as this can lead to an error (no result).
Dispense slowly to minimize the risk of aerosol formation.
Close the cartridge lid. Make sure the lid snaps firmly into place.
Note: Remaining liquefied sample may be kept for up to 12 hrs at 2-8oC
should repeat testing be required.
8.2.14 Start the test on the GeneXpert instrument
Note: Before start processing the sample, check that the Gene Xpert instrument is functioning
and the modules are available.
I. Turn on the computer, and then turn on the Gene Xpert instrument.
II. On the Windows desktop, double-click the Gene Xpert shortcut icon.
III. Log on to the Gene Xpert System software using your user name and
password.
IV. Click on “CHECK STATUS” and check if modules are available. If not
proceed to “Troubleshooting” in User’s manual.
V. In the GeneXpertDx System window, click “CREATE TEST”. The Scan
Cartridge Barcode dialog box appears.
VI. Scan the barcode on the Xpert MTB/RIF cartridge.
VII. The Create Test window appears.
VIII. Using the barcode information, the software automatically fills the boxes for
the following fields: Select Assay, Reagent Lot ID, Cartridge SN, and
Expiration Date.
IX. In the Sample ID box (ID=Patient names), scan or type the sample ID (e.g.
NEW X_2016, HIV X_2016 or KID X_2016). Make sure you type the correct
sample ID. The sample ID is associated with the test results and is shown in
the “View Results” window and all the reports X. Choose module.
XI. Click “Start Test”.
XII. In the dialog box that appears, type your password.
XIII. Open the instrument module door with the blinking green light and load the
cartridge.
XIV. Close the door.
XV. The test starts and the green light stops blinking.
XVI. Wait until the system releases the door lock at the end of the run, then open
the module door and remove the cartridge.
XVII. Dispose of used cartridges in the appropriate sample waste containers
according to your institution’s standard practices
8.2.15 Biological Reference interval Not Applicable
188
Interpretation and Reporting of Results
• MTB DETECTED, Rif Resistance DETECTED –RR
• MTB DETECTED, Rif Resistnce INDETEMINATE –TI
• MTB DETECTED Rif Not DETECTED – T
• MTB NOT DETECTED – N
• This is DNA based test, meant for New TB suspect, ensure don’t enrol
‘follow-up’ patient.
The results are interpreted by the GeneXpert Dx system from measured
fluorescent signals and embedded calculation algorithms and will be
displayed in the “View Results” window. Lower Ct values represent a
higher starting concentration of DNA template; higher Ct values
represent a lower concentration of DNA template.
Critical values MTB DETECTED, Rif Resistance DETECTED –RR
8.2.16 Limitation of the Procedure and Sources of Error
• Perform the test and validate results as per this SOP and details of test package
insert. Reliable results depend on proper sample collection, handling, and
storage. A positive test result does not necessarily indicate the presence of
viable organisms. It is however, presumptive for the presence of MTB and
rifampicin resistance. The results might be affected by antecedent or
concurrent anti-TB drug therapy.
8.2.19 References
1. Clinical Microbiology Procedures Handbook. American Society for
Microbiology. Washington D.C., USA, 2nd edition, 2007.
2. Monica cheesbrough (2005). District Laboratory Practice in Tropical countries.
Cambridge University Press, New York, USA, 2nd edition, 2005.
3. WHO, (2003). Mannual of basic techniques for a health laboratory. Geneva.
2nd edition, 2003.
4. Manual of Clinical Microbiology. American Society for Microbiology (ASM),
Washington D.C., USA. 9th edition, 2007.
5. Manual for the Laboratory Identification and Antimicrobial Susceptibility
Testing of Bacterial Pathogens of Public Health Importance in the Developing
World. U.S. Centers for Disease Control and Prevention (CDC), Atlanta,
Georgia, U.S.A, and World Health Organization (WHO) Geneva Switzerland.
2003.
189
6. International Union against Tuberculosis and Lung Disease. The Public Health
Service National Tuberculosis Reference Laboratory and the National
Laboratory Network. Paris; 1998.
8.3 DETERMINATION OF HIV -1 VIRAL LOAD BY USING GENEXPERT
SYSTEM
8.3.1 Purpose
This SOP outlines the steps for qualitative in vitro diagnostic HVL test by using
automated GeneXpert system.
8.3.2 Scope
The Xpert HIV-1 VL assay is an in vitro reverse transcriptase polymerase chain
reaction (RT-PCR) assay for the detection and quantification of Human
Immunodeficiency Virus type 1 (HIV-1) RNA in human plasma from HIV-1 infected
individuals, using the automated GeneXpert Instrument Systems. The assay can
quantify HIV-1 RNA over the range of 40 to 10,000,000copies/mL. The Xpert HIV-1
VL assay is validated for quantification of RNA from HIV-1 Group M (subtypes A, B,
C, D, F, G, H, J, K, CRF01_AE, CRF02_AG, and CRF03_AB), Group N, and Group
O.
The Xpert HIV-1 VL assay is intended for use in conjunction with clinical presentation
and other laboratory markers for disease prognosis and for use as an aid in assessing
viral response to antiretroviral treatment as measured by changes in plasma HIV-1
RNA levels.
The Xpert HIV-1 VL assay is not intended to be used as a donor screening test for
HIV-1 or as a diagnostic test to confirm the presence of HIV-1 infection
8.3.3 Responsible
The Qualified, competent and registered health laboratory practitioners are
responsible to carry out this procedure.
The head of section is responsible for ensuring the effective implementation and
competency assessment for this procedure
190
running tests and viewing the results. The systems require single-use disposable
GeneXpert cartridges that contain the RT-PCR reagents and carry out the sample
extraction and RT-PCR processes. Because the cartridges are self-contained,
crosscontamination between samples is minimized. For a full description of the
systems, refer to the appropriate GeneXpert Dx Operator Manual or GeneXpert
Infinity Operator Manual.
The HIV-1 VL assay includes reagents for the detection of HIV-1 RNA in specimens
and two internal controls used for quantitation of HIV-1 RNA. The internal controls
are also used to monitor the presence of inhibitor(s) in the RT and PCR reactions.
The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the
cartridge, probe integrity, and dye stability
Principle of operation
Each GeneXpert Dx module processes one sample. You insert the sample and
applicable reagents into a GeneXpert cartridge, create a test, load the cartridge into
an available instrument module, and then start the test. During the test, the system
performs the following steps:
i. Moves the sample and reagents into different chambers in the
cartridge for sample preparation.
ii. Hydrates the reagent beads. iii. Performs probe checks to
ensure that the sample preparation is successful (only if the assay
definition requires this step).
iv. Moves the sample and reagent mixture into the reaction tube.
v. Starts the PCR cycles and real-time detection.
8.3.5 Sample Requirement.
• DBS collected as per SOP for collection of DBS.
• Anticoagulated whole blood(WB) B in sterile tubes using EDTA (lavender
top) as the anticoagulant as per the manufacturer’s intructions for use.
• A minimum of 100 μL of WB is required for the HIV-1 Qualitative assay.
8.3.6 Equipment
Perform equipment Start up, Maintenance, trouble shoot and shut down refer
manufacturer instructions.
Biosafety cabinet, Data computer connected to LIS (Optional), Printer (optional),
GeneXpert instrument and GeneXpert Software should be available.
8.3.7 Materials
Reagent kit content Extra supplies
191
The HIV-1 Qual assay kit contains sufficient reagents to • Serviette/Wipe
process 10 specimens or quality control samples. • Bleach
The kit contains the following: • 70 % alcohol or
HIV-1 Qual assay Cartridges with Integrated Reaction Tubes methylated
10 spirit
• Bead 1, Bead 2, and Bead 3 (freeze-dried) 1 of each • Labelling
per cartridge marker
• Lysis Reagent (Guanidinium Thiocyanate) 1.4 mL per • Optional:
cartridge sterile pipettes
• Rinse Reagent 0.5 mL per cartridge for sample
• Elution Reagent 2.5 mL per cartridge processing
• Lab coat,
• Binding Reagent 2.4 mL per cartridge
• Non -
• Proteinase K Reagent 0.48 mL per cartridge
powdered
gloves
192
ii. Wear protective disposable gloves, laboratory coats, and eye protection
when handling specimens and reagents. Wash hands thoroughly after
handling specimens and test reagents.
iii. Follow safety procedures for working with chemicals and handling
biological samples.
iv. When processing more than one sample at a time, open only one
cartridge; add sample and close the cartridge before processing the next
sample. Change gloves between samples.
v. Do not substitute HIV-1 Qual assay reagents with other reagents.
vi. Do not open the HIV-1 Qual assay cartridge lid except when adding the
Sample Reagent and WB or the Sample Reagent treated DBS sample.
vii. Do not use a cartridge if it appears wet or if the lid seal appears to have
been broken.
viii. Do not shake the cartridge. Shaking or dropping the cartridge after
opening the cartridge lid may yield invalid results.
ix. Do not use a cartridge that has a damaged reaction tube.
x. Each single-use HIV-1 Qual assay cartridge is used to process one
specimen. Do not reuse spent cartridges. xi. The single-use
disposable pipette is used to transfer one specimen. Do not reuse spent
disposable pipettes.
xii. In the event of contamination of the work area or equipment with specimen
or control materials, disinfect the area with a 1:10 bleach solution and then
70% ethanol. Wipe work surfaces dry completely before proceeding.
8.3.10 Calibration
There is no need to calibrate the GeneXpert Dx instrument. Cepheid performs all of
the necessary calibrations before you receive the system. However, Cepheid
recommends that the instrument be recalibrated after 1 year of use, based on the
initial installation date (or based on the previous calibration for subsequent years) or
at 2000 tests per instrument module, whichever comes first.
193
used to indicate sample-inhibitor contamination. The endogenous control is taken
from the specimen sample.
Each test includes a Sample Volume Adequacy (SVA), a Sample Processing Control
(SPC) and Probe Check Control (PCC).
Sample Volume Adequacy (SVA): Ensures that the sample was correctly added to
the cartridge. The SVA verifies that the correct volume of sample has been added in
the sample chamber. The SVA passes if it meets the validated acceptance criteria. If
the SVA does not pass, an ERROR 2096 will display if there is no sample or an
ERROR 2097 if there is not enough sample. The system will prevent the user from
resuming the test.
Sample Processing Control (SPC): Ensures that the sample was correctly
processed. The SPC is an Armoured RNA in the form of a dry bead that is included
in each cartridge to verify adequate processing of the sample virus. The SPC verifies
that lysis of HIV-1 has occurred if the organism is present and verifies that the
specimen processing is adequate. Additionally, this control detects
specimenassociated inhibition of the RT-PCR reaction. The SPC should be positive
in a negative sample and can be negative or positive in a positive sample. The SPC
passes if it meets the validated acceptance criteria.
In addition to the controls, the GeneXpert Dx instrument performs a probe check
during the first stage of the test.
Probe Check Control (PCC): Before the start of the PCR reaction, the GeneXpert
Instrument System measures the fluorescence signal from the probes to monitor
bead rehydration, reaction tube filling, probe integrity, and dye stability. The PCC
passes if it meets the validated acceptance criteria.
External Controls: Internal quality controls should be done weekly by using known
HIV-1 DETECTED and HIV -1 NOT DETECTED as the same as the routine EID/DBS
samples.
8.3.12 Procedural steps
Follow the actions described step by step to do each specific task
194
Preparing of cartridge(s)
EDTA anticoagulated Whole Blood
i. Wear protective disposable gloves. ii. Disinfect the work area by 0.5
% bleach solution followed by 70 % alcohol
iii. Wear protective disposable powder free gloves.
iv. Label the Sample Reagent vial with the specimen identification.
v. Inspect the test cartridge for damage. If damaged, do not use. vi.
Open the cartridge lid.
vii. Use the 1 mL transfer pipette provided to transfer 750 μL of the sample
reagent into the sample chamber of the cartridge.
viii. Allow the Sample Reagent to adjust to room temperature and mix the
bottle by inverting before transferring to the cartridge. Transfer exactly
750 μL into the sample chamber of the cartridge.
ix. Mix the Whole Blood sample by inverting the vial (EDTA or lavendertop
tube) at least seven times. Immediately transfer 100 μL using the
micropipette provided by squeezing the upper bulb and then releasing
to aspirate the blood.
x. Squeeze again to dispense the blood into the sample chamber of the
cartridge where it will mix with the Sample Reagent already in the
sample chamber. Alternatively, use an automatic pipette to dispense
the blood into the sample chamber of the cartridge. Do NOT pour the
specimen into the chamber!
195
Figure 2. HIV-1 Qual Assay 100 μL Transfer Micropipette
❑ DBS sample
i. Wear protective disposable gloves.
ii. Disinfect the working area. iii. Wear protective disposable
powder free gloves. iv. Before starting, remove the vial containing
the Sample Reagent from the kit and, if it was refrigerated, allow to adjust
to room temperature. If the vial has not been stored in an upright position,
make sure the buffer is settled in the bottom by giving the vial a firm
shake.
v. Turn on Thermo Mixer to heat to 56 °C.
vi. Label the Sample Reagent vial with the specimen identification.
vii. Using sterilized scissors, excise one entire DBS from the filter
paper card for each specimen. Follow the delineated lines when
excising the DBS. If perforated circles are used, use clean and
sterile pipette tips to detach the DBS.
viii. Unscrew the lid on the vial containing the Sample Reagent and
place one DBS in the vial. Ensure that the DBS is fully submerged
in the Sample Reagent buffer.
196
ix.
Place the vial with the DBS in a Thermo Mixer and incubate for 15
minutes at 56 °C while rotating at 500 rpm.
x. Inspect the test cartridge for damage. If damaged, do not use. xi.
Open the cartridge lid
xii. Use the 1 mL transfer pipette provided to transfer all the liquid from
the lysed DBS specimen into the sample chamber of the cartridge.
Ensure the pipette is filled above the third mark on the transfer
pipette. Avoid suction of the DBS with the pipette. Do NOT pour the
specimen into the chamber!
xiii. Close the cartridge lid, ready to start the test.
Notes! Change gloves between specimen, and each new procedure.
Starting the Test
i. In the GeneXpert System window, click Create Test. The scan
Cartridge Barcode dialog box appears.
ii. Scan the barcode on the HIV-1 Qual assay cartridge. iii. Using the
barcode information, the software automatically fills the boxes for the
following fields: Select Assay, Reagent Lot ID, Cartridge SN, and
Expiration Date.
iv. Type the Patient ID, make sure the Patient ID is typed correctly.
v. Type in the Sample ID. Make sure the Sample ID is typed correctly.
vi. On the Notes field, enter the words KATAVI RRHL to indicate a
testing laboratory name on the patient report.
vii. Open the instrument module door with the blinking green light and load
the cartridge.
viii. Click Start Test (Gene Xpert Dx). Enter you’re your user name and
password, if requested.
ix. Close the door.
x. The test starts and the green light stops blinking. When the test is
finished, the light turns off. xi. Wait until the system releases the
door lock before opening the module door and removing the cartridge.
Result viewing, and printing
i. In the Gene Xpert Dx System window, Click the View Results icon to view
results. This view result window appears.
ii. If the software reports “Error”, Invalid, or No result, repeat the test using
new DBS circle.
iii. Should the test again show Error, Invalid, or No result, proceed to
troubleshooting manual to exclude technical problems before requesting a
new specimen.
iv. Upon completion of the test, click the Report button of the View Results
window to view and/or generate a PDF report file.
197
v. Report should be done in the Laboratory HEID register and a tick on the
respective result blank space on the HEID request form. Report as; HIV -
1 DETECTED*.
HIV -1 NOT DETECTED
vi. Report “please submit a new sample” if the system repeatedly did not
produce a result and you have excluded and/or fixed a technical problem.
vii. HIV -1 DETECTED is a critical result that needs immediate action
including but not limited to; retesting using new DBS circle with a new
cartridge, and result notification to the respective referring health facility.
8.3.13 Biological Reference Intervals Not Applicable
8.3.14 15. Result interpretation and Reporting of Results.
• The results are interpreted automatically by the GeneXpert Instrument System
from measured fluorescent signals and embedded calculation algorithms and
are clearly shown in the View Results window. Possible results are shown in
Table below:
Result Interpretation
HIV-1 The HIV-1 target nucleic acids are detected.
DETECTED • The HIV-1 target nucleic acids have a Ct within the
See Figure 1. valid range.
• SPC: NA (not applicable); SPC is ignored because
the HIV-1 target amplification occurred.
• Probe Check: PASS; all probe check results pass.
HIV-1 NOT The HIV-1 target nucleic acids are not detected. SPC meets
DETECTED acceptance criteria.
See Figure 2. • SPC: PASS; SPC has a Ct within the valid range.
• Probe Check: PASS; all probe check results pass.
INVALID Presence or absence of the HIV-1 target nucleic acids
cannot be determined. Repeat test with new sample and
cartridge.
• SPC: FAIL; SPC Ct is not within valid range.
• Probe Check: PASS; all probe check results pass
ERROR Presence or absence of HIV-1 target nucleic acids cannot be
determined. Repeat test with new sample and cartridge.
• HIV-1: NO RESULT
• SPC: NO RESULT
• Probe Check: FAIL*; all or one of the probe check
results fail.
*If the probe check passed, the error is caused by the
maximum pressure limit exceeding the acceptable range or
by a system component failure.
198
NO RESULT Presence or absence of HIV-1 target nucleic acids cannot be
determined. Repeat test with new sample and cartridge.
A NO RESULT indicates that insufficient data were collected.
For example, the operator stopped a test that was in
progress.
• HIV-1: NO RESULT
• SPC: NO RESULT
• Probe Check: NA (not applicable).
199
Rare mutations within the target region of the HIV-1 Qual assay may affect primer
and/or probe binding resulting in failure to detect the virus.
A negative test result does not preclude HIV-1 infection. Results from the HIV-1 Qual
assay should be interpreted in conjunction with clinical presentation and other
laboratory markers.
8.3.16 Performance Characteristics
Refer to the manufacture package insert and verification report for detailed
information on Performance Characteristics of the testing procedure
8.4.2 Scope
The HIV-1 Qual assay, is a qualitative in vitro diagnostic test designed to detect
Human Immunodeficiency Virus Type 1 (HIV-1) total nucleic acids on the automated
Gene Xpert Systems using human whole blood (WB) and dried blood spot (DBS)
specimens from individuals suspected of HIV-1 infection. The HIV-1 Qualitative assay
is intended to aid in the diagnosis of HIV-1 infection in conjunction with clinical
presentation and other laboratory markers. The assay is intended to be used by
laboratory professionals or specifically-trained healthcare workers. The assay is not
intended to be used as a blood donor screening test for HIV-1.
8.4.3 Responsible
200
Section head is responsible effective implementation of this procedure. Only
competent laboratory staffs should carry out this procedure. It is the responsibility of
each staff to read, understand and implement this procedure.
8.4.4 Principle
Principle of the Procedure
The GeneXpert (GX) Instrument Systems automate and integrate sample
preparation, nucleic acid extraction and amplification, and detection of the target
sequence in simple or complex samples using real time reverse transcription PCR
(RT-PCR). The systems consist of an instrument, personal computer, and preloaded
software for performing tests and viewing the results. The systems require the use of
single-use disposable GeneXpert cartridges that hold the RT-PCR reagents and host
the RT-PCR processes. Because the cartridges are self-contained,
crosscontamination between samples is minimized. The HIV-1 Qual assay includes
reagents for the detection of HIV-1 total nucleic acids in specimens as well as an
internal control to ensure adequate processing of the target and to monitor the
presence of inhibitor(s) in the RT and PCR reactions. The Probe Check Control
(PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity,
and dye stability.
201
EDTA Anticoagulated WB in sterile tubes using EDTA (lavender top) as the
anticoagulant as per the manufacturer’s instructions for use. A minimum of 100 μL of
WB is required for the HIV-1 Qualitative assay.
8.4.7 Equipment
Start up, Maintainance, trouble shoot and shud down refer manaufacturer instructions
Thermo Mixer C for incubation with smart block, Biosafety cabinet, Data computer
connected to LIS (Optional) and Printer (optional)
8.4.8 Materials
Materials (Reagents and consumables) used to perform the test.
Reagent kit content Extra consumables
The HIV-1 Qual assay kit contains sufficient reagents to • DBS Collection
process 10 specimens or quality control samples. Kit (Filter paper
The kit contains the following: cards, e.g.,
HIV-1 Qual assay Cartridges with Integrated Reaction Whatman 903,
Tubes 10 Munktell or equivalent,
• Bead 1, Bead 2, and Bead 3 (freeze-dried) 1 of each lancets, desiccants,
per cartridge plastic
• Lysis Reagent (Guanidinium Thiocyanate) 1.4 mL per • sealable bags,
cartridge and swabs)
• Rinse Reagent 0.5 mL per cartridge • Scissors,
• Elution Reagent 2.5 mL per cartridge sterile
(recommended for
• Binding Reagent 2.4 mL per cartridge
excising DBS from filter
• Proteinase K Reagent 0.48 mL per cartridge
paper if not using a
HIV-1 Qual assay Sample Reagent Set (Sample Reagent)
perforated DBS card)
10
• Sterile pipette
• Lysis Reagent (Guanidinium Thiocyanate) 1.0 mL per
tips
vial Disposable 1 mL Transfer Pipettes 1 bag of 10 per kit
• Serviette/Wipe
Disposable 100 μL Transfer Micropipettes 1 bag of 10 per • Bleach
kit CD 1 per kit
• 70 % alcohol or
• Assay Definition Files (ADF) methylated
• Instructions to import ADF into GeneXpert software spirit
• Instructions for Use (Package Insert) • Distilled water
• Labelling
marker
• Optional:
sterile pipettes
for sample
processing
202
• Lab coat,
• Non - powdered
gloves
8.4.10 Safety
i. Treat all biological specimens, including used cartridges, as if capable of
transmitting infectious agents.
ii. Wear protective disposable gloves, laboratory coats, and eye protection
when handling specimens and reagents. Wash hands thoroughly after
handling specimens and test reagents.
iii. Follow safety procedures for working with chemicals and handling biological
samples.
iv. When processing more than one sample at a time, open only one cartridge;
add sample and close the cartridge before processing the next sample.
Change gloves between samples.
v. Do not substitute HIV-1 Qual assay reagents with other reagents.
vi. Do not open the HIV-1 Qual assay cartridge lid except when adding the
Sample Reagent and WB or the Sample Reagent treated DBS sample.
vii. Do not use a cartridge if it appears wet or if the lid seal appears to have been
broken.
viii. Do not shake the cartridge. Shaking or dropping the cartridge after opening
the cartridge lid may yield invalid results.
ix. Do not use a cartridge that has a damaged reaction tube.
x. Each single-use HIV-1 Qual assay cartridge is used to process one
specimen. Do not reuse spent cartridges. xi. The single-use disposable
pipette is used to transfer one specimen. Do not reuse spent disposable
pipettes.
xii. In the event of contamination of the work area or equipment with specimen or
control materials, disinfect the area with a 1:10 bleach solution and then 70%
ethanol. Wipe work surfaces dry completely before proceeding.
8.4.11 Calibration
You do not need to calibrate the GeneXpert Dx instrument. Cepheid performs all of
the necessary calibrations before you receive the system. However, Cepheid
203
recommends that the instrument be recalibrated after 1 year of use, based on the
initial installation date (or based on the previous calibration for subsequent years) or
at 2000 tests per instrument module, whichever comes first.
204
bead rehydration, reaction tube filling, probe integrity, and dye stability. The PCC
passes if it meets the validated acceptance criteria.
External Controls: Internal quality controls should be done weekly by using known
HIV-1 DETECTED and HIV -1 NOT DETECTED as the same as the routine EID/DBS
samples.
8.4.13 Procedural steps
Follow the actions described step by step to do each specific task
Note: Before start processing the specimen, check that the GeneXpert instrument is
functioning and the modules are available.
205
Figure 6. HIV-1 Qual Assay 1 mL Transfer Pipette
206
v. Turn on Thermo Mixer to heat to 56 °C.
vi. Label the Sample Reagent vial with the specimen identification. vii. Using
sterilized scissors, excise one entire DBS from the filter paper card for each
specimen. Follow the delineated lines when excising the DBS. If perforated circles
are used, use clean and sterile pipette tips to detach the DBS.
viii. Unscrew the lid on the vial containing the Sample Reagent and place one DBS
in the vial. Ensure that the DBS is fully submerged in the Sample Reagent buffer.
ix. Place the vial with the DBS in a Thermo Mixer and incubate for 15 minutes at
56 °C while rotating at 500 rpm.
x. Inspect the test cartridge for damage. If damaged, do not use.
xi. Open the cartridge lid
xii. Use the 1 mL transfer pipette provided to transfer all the liquid from the lysed
DBS specimen into the sample chamber of the cartridge. Ensure the pipette
is filled above the third mark on the transfer pipette. Avoid suction of the DBS
with the pipette. Do NOT pour the specimen into the chamber!
xiii. Close the cartridge lid, ready to start the test.
207
The results are interpreted automatically by the GeneXpert Instrument System from
measured fluorescent signals and embedded calculation algorithms and are clearly
shown in the View Results window. Possible results are shown in Table below:
Result Interpretation
HIV-1 DETECTED The HIV-1 target nucleic acids are detected.
See Figure 1. • The HIV-1 target nucleic acids have a Ct within the valid
range.
• SPC: NA (not applicable); SPC is ignored because the
HIV-1 target amplification occurred.
• Probe Check: PASS; all probe check results pass.
HIV-1 NOT The HIV-1 target nucleic acids are not detected. SPC meets
DETECTED acceptance criteria.
See Figure 2. • SPC: PASS; SPC has a Ct within the valid range.
• Probe Check: PASS; all probe check results pass.
INVALID Presence or absence of the HIV-1 target nucleic acids cannot be
determined. Repeat test with new sample and cartridge.
• SPC: FAIL; SPC Ct is not within valid range.
• Probe Check: PASS; all probe check results pass
ERROR Presence or absence of HIV-1 target nucleic acids cannot be
determined. Repeat test with new sample and cartridge.
• HIV-1: NO RESULT
• SPC: NO RESULT
• Probe Check: FAIL*; all or one of the probe check results
fail. *If the probe check passed, the error is caused by the
maximum pressure limit exceeding the acceptable range or by a
system component failure.
NO RESULT Presence or absence of HIV-1 target nucleic acids cannot be
determined. Repeat test with new sample and cartridge.
A NO RESULT indicates that insufficient data were collected. For
example, the operator stopped a test that was in progress.
• HIV-1: NO RESULT
• SPC: NO RESULT
• Probe Check: NA (not applicable).
208
8.4.16 Limitation of the Procedure and Sources of Error.
Good laboratory practices and changing gloves between handling specimens are
recommended to avoid contamination of reagents.
Rare mutations within the target region of the HIV-1 Qual assay may affect primer
and/or probe binding resulting in failure to detect the virus.
A negative test result does not preclude HIV-1 infection. Results from the HIV-1 Qual
assay should be interpreted in conjunction with clinical presentation and other
laboratory markers
209
8.4.19 References
Ministry of Health, Community Development, Gender, Elderly and Children, Standard
Operating Procedures for qualitative HIV-1 HEID testing using GeneXpert
Xpert HIV-1 Qual -1 Assay Package Insert 308-3048 Rev J
GeneXpert Dx System. Operator Manual
210
CHAPTER NINE: ANATOMICAL PATHOLOGY
9.1 PROCEDURE FOR MORTUARY SERVICES
9.1.1 Purpose
This procedure provides instructions for providing mortuary services including
autopsy practice, embalming as well as safety of personnel, visitors and community.
9.1.2 Scope
This document provides guidelines for the mortuary staff and administration
recommended standards for mortuary facilities in settings and for communication
between staff involved in autopsy procedures or autopsy related processes.
Body storage
A body cold store having a capacity appropriate for the mortuary workload should be
maintained at a temperature of about 2- 6 o C.
If long term storage is required, the body should have maintained at approximately
20 o C.
Labelling procedures should be established so that body identification is made easy.
9.1.3 Responsible
Qualified and trained mortuary personnel are responsible for implementing this
procedure.
The Head of Laboratory is responsible for ensuring the effective implementation and
maintenance of this procedure.
9.1.7 Materials
10% formalin, Gloves, Leak proof bag, gloves, water resistant gown/plastic apron
over water repellent gown, and surgical masks, goggles or face shield, shoe covers
9.1.9 Safety
211
- PPE shall be used to prevent skin and mucous membrane contact with blood
and other body fluid. These may include the use of gloves, N95 masks,
protective eye wear, face shields, shoe covers, plastic aprons/gowns hair
bornets, cut resistant gloves and laboratory coats.
Surgical or post mortem gloves must be worn by all personnel involved in the
autopsy procedure Hand washing:
- Hand and other skin surface should be washed with soaps and water
immediately after contact with blood or other body fluid. Hand shall be washed
each time with running water and soap.
- Sufficient and appropriate disinfectant should be 0.5% chlorine solution for
routine mortuary work, embalming and post-mortem, then removed and rinsed
with distilled water before being dried and stored.
- Appropriate vaccination and follow up of immunity status should be offered to
all mortuary staff and record should be maintained including any refusal of an
offer of immunization.
- The Mortuary personnel is responsible to ensuring that he/she complies with
policies guarding personal safety, handling of bodies and bodily fluids as well
as the safe operation of Mortuary equipment.
- Put bio hazardous waste in a red biohazard bag
- Put the red bag in the bio hazardous waste box
- Tape box closed
- Put taped box in bio hazardous waste pick up location
- The Medical Attendants must wear complete PPE when handling hazardous
waste.
Autopsy Instruments:
- All instruments are to be cleaned and disinfected between examinations,
instruments should be dried and laid out on a non-metal surface
- Autopsy Tables and Garbage Disposals: Autopsy tables should be cleaned
and disinfected between autopsies or at the end of the day
- Following the autopsy, the disposals and drain on each autopsy table should
be cleaned.
9.1.10 Calibration
Calibration for refrigerators should be done as per schedule
212
9.1.13 In the ward you should observe the following before:
If the dead body is well labelled
> If the mortuary forms have been filled in and signed by the nurse
>Make sure you sign the death book before you take the body
Note: Tags have to be attached to the big toes /or around the circumference of the
ankle
Receiving dead body from outside the facility brought by police, relatives or
good Samaritans.
After receiving a dead body from the police or relatives from home the following
procedures should be followed
i. First get the information, document the police/transport ID/telephone number
ii. Prepare the trolley to receive the dead body
iii. Check the body appearance and attached properties, if it’s a police case after
documentations put the dead body directly to the cooling unit and wait for
post-mortem exam to be done.
iv. Provide the mortuary identity card to the police, if will be later on provided to
the relative
v. If it’s a home cases after documentation inform the relatives about the other
process of payments.
vi. After payments request the nearby relative to attend the embalming
procedure then keep the dead body in the cooling unit after following the
identification procedure and make sure it is well labelled.
vii. Make sure you label the form to indicate which unit the dead body has been
put in.
9.1.15 Releasing the died body from mortuary to relatives
Before giving out dead bodies following procedure should be followed;
i. For the dead body from the facility make sure the relative comes with the
discharge summary, burial certificate and the proper receipt.
ii. Check if the information has been filled well in the mortuary register book.
iii. When relative have accomplished all the process and have all the form
needed, assist them to prepare the dead body and keep it in a good condition
213
iv. Allow them to leave after finishing all the process.
v. For out - patients make sure the relatives come with the burial certificate/or
letter from local government if they have it and proper receipts.
vi. When they have accomplished all process, assist them to prepare the body in
the washer rooms and make sure equipment, water and sanitation is well
controlled.
9.1.16 Handling personal property and clothing
Personal property that arrives with decedent remains shall be processed as follows;
i. Decedent’s personnel property will be entered into register book upon
admission.
ii. Clothing should be described by colours and items
iii. Personal effect such as jewellery, watch as well as money should only
be removed in the presence of relatives/police. iv. Personal effects
such as jewellery should be described in non-valuable terms e.g. yellow
metal or coloured stone.
9.1.17 Embalming procedure
It consists of arterial infusion of embalming fluid (10% Formalin). Since most bodies
are not disposed within 72hrs after death, hence the importance of a mortuary
establishing this service.
i. Arrange all your equipment that are needed for embalming. ii.
Prepare the formalin solution 10% concentration
iii. Put on the proper protective equipment
iv. Prepare the body to be embalmed
v. Start the embalming procedure
vi. Make sure that body has a label and tag
vii. Clean the area and the equipment that have been used.
9.1.18 Autopsy procedure
i. Autopsy has to be complete routinely.
ii. A complete autopsy is defined to include a detailed external examination as
entire body and an internal examination to include the removal and dissection
of all thoraco –abdominal and neck organs, opening the head with the removal
and examination of the brain
iii. A complete autopsy does not require histological examination
iv. A patient autopsy is defined as an examination that surface any part of the
defined complete autopsy e.g. not opening any of the body cavities or not
examining organs.
v. An external examination is defined as a detailed description of the decedents
remains including scars, surgical incisions, medical devices and tattoos.
9.1.19 Pre-autopsy procedures
214
Prior to autopsy the medical attendant will set up the autopsy work room according
to the case examination status including preparing tables for body dissection,
preparing instruments, preparing sample containers and collection tubes, preparing
paper work for daily case load and taking radiographs.
Autopsy work room should be set up with the following instruments supplies, in
certain cases it will be necessary to equip the autopsy work room with specialized
instruments or additional supplies.
215
pleural, incising the psoas muscles, assisting with preparation of sex kits and
DNA cards. Measure the length of a died.
9.1.22 Post Autopsy Procedures
i. Put organs in viscera bag ii. Replace body organs
iii. Close the thoracic, abdominal and cranial cavities with sutures
iv. Clean the body and replace in body bag
v. Indicate completion of examination by inciting a ONE on body bag
vi. Return body to refrigerated storage
vii. Put specimen in designated area depending on processing instructions;
a) Toxicology samples are put in toxicology refrigerator in the yellow
tray labelled” toxicology”
b) Histology sections are put in the yellow tray on top of toxicology
refrigerator labelled “histology”
c) Microbiology sample are put in the yellow tray on top of toxicology
refrigerator labelled “micro”
viii. Thoroughly clean and disinfect autopsy and dissection tables, sinks, drains,
instruments, dry erase boards floor area
ix. Between examinations all instruments surfaces should be cleaned with a 10%
bleach solution.
9.1.23 Body storage and organization
i. Body should be stored in clean, closed body bags with no leakage of fluids
on rack or tray
ii. Place body in empty compartment head toward the wall
iii. Tag body storage compartment with decedent name and case number
iv. Log body on body part(s) into box and cart inventory sheet
v. Maintain an accurate box or cart inventory sheet. The inventory sheet should
be updated appropriate daily
vi. Empty trays to the cleaned and disinfected and kept in box
vii. All trays should be cleaned and disinfected following a release or transfer
9.1.24 Receiving and releasing remains
Receiving Remains;
i. Identify ID bracelet with Police to tag
ii. Obtain appropriate signatures on transport notification form
iii. Take photography using the camera
iv. Obtain decedent height and weight
v. Record personal property (clothing)
vi. Complete intake of body into the logbook
vii. Complete intake of body in FACTS
viii. Log body on box and cart sheet ix. Place body in refrigerated
storage
Releasing Remains:
216
i. Confirm in FACTS that the body is ready to be released and that the receipt
of remain seen has been properly completed by communications unit.
ii. Print and sign copy of the Receipt of remains form.
iii. Sign body out of release logbook
iv. Obtain funeral home representatives signatures, number and initials where
appropriate
v. Have to tag and receipt of remains form witnesses
vi. Clean and disinfect tray before retaining it to cold box
9.1.25 Sample Storage and Retrieval
All stock specimens retained in clean and dry bottle in 10% buffered formalin label
write the Post Mortem Examination Number, name, age, sex, address and
anatomical site
9.1.28 References
1. Standard guidelines for the facilities and operation of mortuaries in Tanzania
2008
2. Hutchins GM. Practice guidelines for autopsy performance. Archives of
pathology and laboratory medicine, 1994, 118(1):19-25
3. Advices ED, Sims KL. Enhancing autopsy performance and reporting. A
system for a 5-day completion time. Archives of pathology and laboratory
medicine,1996,120(3):249-53.
217
Annex 1: List of Subject Matter Experts who developed these SOPs
218
30. Idrissa Hassan Laboratory Service PORALG, Dodoma
Coordinator
31. Innocent A. Chinguile Laboratory Scientist & Training, Dodoma MoH,
Epidemiologist
32. Jackson Luvamunda Quality Officer Sinza HC, Dar es Salaam
33. Jackson Zengo Deputy National DHCTSU, Dodoma MoH,
Laboratory Quality Officer
34. Jacob Lusekelo National Laboratory DHCTSU, Dodoma MoH,
Quality Officer
35. James Mziray Laboratory Service PO-RALG, Dodoma
Coordinator
36. Joakim Chacha LS-HSEEP DHCTSU, MoH, Dodoma
37. Joyce Samson Laboratory Quality Officer Sekou-Toure RRH,
Mwanza
38. Kadogosa Saimon Laboratory Technologist Singida RRH, Singida
39. Lanja Kahemela Laboratory Quality Officer MNH, Mloganzila, Dar es
Salaam
40. Maige Ngeleja DSCMC MoH, Dodoma
41. Mary F. Mtui Registrar HLPC, MoH, Dodoma
42. Mayala Lushina Laboratory Quality Officer MNH, Upanga, Dar es
Salaam
43. Michael Mazoya Laboratory Scientist NBTS, Dar es Salaam
44. Mzelifa Daudi Microbiologist University of Dodoma
(UDOM), Dodoma
45. Pius R. Tarimo Laboratory Scientist & Kilimanjaro Christian
Epidemiologist Medical University
College (KCMUCo),
Moshi
46. Rajabu Muninge Laboratory Scientist Maranatha Hospital,
Mbeya
47. Rashid Nassoro Laboratory Quality Officer Morogoro RRH, Morogoro
48. Reuben Abednego Laboratory Scientist NPHL, Dar es Salaam
49. Reuben Lema Deputy Laboratory Morogoro Morogoro RRH,
Manager
50. Reuben S. Mkala Ag. Head of Laboratory DHCTSU, Dodoma MoH,
Service
51. Richard Kinyaha Laboratory Scientist Kibong’oto IDH,
Kilimanjaro
52. Robert Makala Regional Laboratory RAS - Manyara
Technologist (RLT)
53. Sabra Rashid Laboratory Scientist MoH, Unguja-Zanzibar
54. Said Lunemhya Laboratory Scientist Benjamin Mkapa ZRH,
Dodoma
55. Scolastica S. Laboratory Quality Officer Mirembe Hospital,
Manyama Dodoma
56. Susu Jeremiah Susu Laboratory Quality Officer Benjamin Mkapa ZRH,
Dodoma
57. Tamaly Lutufyo Laboratory Manager CCBRT, Dar es Salaam
219
58. Ulfat Rahim Laboratory Technologist TMJ, Dar es Salaam
59. William S Mollel Laboratory Technologist MOLLEL LABORATORY,
Dodoma
60. Yahya Mnung’a MeLSAT-President MeLSAT, Dar es Salaam
61. Yohana Shirima Laboratory Technologist NBTS Northern zone,
Moshi
62. Yulitha Barnabas Laboratory Manager Dodoma RRH, Dodoma
63. Zacharia Omary Laboratory Manager Sekou-Toure RRH,
Mwanza
220
Annex 2: Biological Reference Intervals for Full Blood Count
221
Annex 3: Biological Reference Intervals for Coagulation Profile
222
Annex 4: Biological Reference Intervals for Urine Biochemistry
223
Annex 5: Biological Reference Intervals for Clinical Chemistry and
Immunoassays
224
Test Name Normal range SI Unit
Sodium ( Na ) 136 - 145 mmol/L
Potassium ( K ) 3.5 – 5.1 mmol/L
Chloride (Cl ) 98 – 107 mmol/L
Amylase Total 25- 125 U/L
Male 30 – 200 U/L
Female 29- 168 U/L
Creatine Kinase ( CK )
Male child 30 – 200 U/L
Female child 29-168 U/L
Lactate dehydrogenase 125 – 220 IU/L
(LDH)
Lipase 13- 60 U/L
Male 2.1 – 2.55 mmol/L
Famale 2.1-2.55 mmol/L
Calcium
Male child 2.2 – 2.7 mmol/L
Female child 2.2 – 2.7 mmol/L
Glucose 3.3 - 6.1 mmol/L
Male 5.5 – 25.8 µmol/L
Female 4.5 - 25.8 µmol/L
Iron
Male child 5.5 – 25.8 µmol/L
Female child 4.5 – 25.8 µmol/L
% Saturation (Iron 20 - 50 %
saturation)
Male 0.21 - 0.42 mmol/L
Female 0.15 - 0.35 mmol/L
Uric Acid
Male child 0.21 - 0.42 mmol/L
Female child 0.15 - 0.35 mmol/L
Phosphorus 0.74 - 1.52 mmol/L
Sodium 24hrs Urine 27 – 287 mmol/24 hours
Potassium 24hrs Urine 25 – 125 mmol/24hrs
Male 1.74 - 3.64 g/L
Female 1.8 - 3.82 g/L
Transferrin
Male child 1.86 – 3.88 g/L
Female child 1.86 - 3.88 g/L
Alpha Feto Protein 0.0 - 1.09 ng/ml
13.8-17.5 Female pg/ml
High Sensitive Troponin
28.9-39.2 Male pg/ml
Vitamin B12 187-883 pg/ml
Ferritin 10 - 250 ng/ml
Folate 3.72 - 50.4 ng/ml
PSA 0.0 - 4.0 ng/ml
TSH 0.49 - 4.67 IU/ml
T4 0.47 - 4.67 ng/L
T3 1.45 - 3.48 pg/ml
CK-MB 0.0 - 6 %
225
Test Name Normal range SI Unit
Tacrolimus 3 – 20 ng/ml
BNP 0-142 pg/ml
Cyclosporine 30.0-1500 ng/ml
CEA 0-5 ng/ml
CA-125 0-35 IU/mL
Less than 5 for non pregnant mIu/mL
B-HCG
25 for early pregnancy mIu/mL
Vitamin D 0 - 160 ng/ml
Immunoglobulin G 5.40-18-22 Male g/l
5.52-16.31 Female g/l
1-12 months <15 IU/mL
1-5 years <60 IU/mL
Immunoglobulin E 6-9 years <90 IU/mL
10-15 years <200 IU/mL
Adults <100 IU/mL
Male 63 – 645 mg/dl
Female 65 – 517 mg/dl
Immunoglobulin A
Male child 21- 291 mg/dl
Female child 21 - 281 mg/dl
Male 0.22-2.40 g/l
Female g/l
Immunoglobulin M
0.33-2.93 g/l
Either 0.22-2.93 g/l
D - Dimer 0 .0 – 198 ng/L
CRP 0.0-5.0 mg/L
Follicular phase 21-251 pg/ML
Midcycle phase 38-649 pg/ML
Estradiol Lueal phase 21-312 pg/ML
Postmenopausal female 1028 pg/ML
Male 11-44 pg/ML
Male ng/ml
Prolactin 3.46-19.40 ng/ml
Female5.18-26.53 ng/ml
Male 4.94-32.01 nmol/L
Testosterone
Female 0.38-1.97 nmol/L
Follicular phase 0.1-0.3 ng/ml
Luteal phase 1.2-15.9 ng/ml
Postmenopausal 0.1-0.2 ng/ml
Progesteron First trimester 2.8-147.3 ng/ml
Second trimester 22.5-95.3 ng/ml
Third trimester 27.9-242.5 ng/ml
Male 0.1-0.2 ng/ml
Male 1.14-8.75 ng/ml
LH Follicular phase 2.39-6.60 ng/ml
Midcycle peak-9.06-74.24 ng/ml
226
Test Name Normal range SI Unit
Luteal phase 0.909.33 ng/ml
Postmenopausal ng/ml
10.39 - 64.57 ng/ml
Male 0.95 - 11.95 mIu/mL
Follicular phase 3.03 - 8.08 mIu/mL
Midcycle peak 2.55 - 16.69 mIu/mL
FSH
Luteal phase 1.38 - 5.47 mIu/mL
Postmenopausal mIu/mL
26.72 - 133.41 mIu/mL
Male 0.66 - 1.07 mmol/L
Female 0.66 - 1.07 mmol/L
Magnesium
Male child 0.70 - 0.86 mmol/L
Female 0.70 - 0.86 mmol/L
ADA 0 - 15 U/L
Glycated haemoglobin
4-6 %
(HBA1C)
227
Annex 6: Critical or Panic Values that call for Immediate Actions
228
Annex 7: Charts for Biochemical Identifications of Common Enterobacteriaceae and other Enteric Organisms
MAC
Organism TSI Oxidase H2S Gas Motility Indole Urea Citrate Haemolysis Comment
Reaction
Serratia Red pigment at
NLF K/A or A/A - - + + - - + -
mercesens room temp on MHA
Grow with
Proteus mirabilis NLF K/A - + + + - + +(weak) - swarming xters on
BA
A/A or Grow with
Proteus vulgaris NLF K/A - + + + + + +/- - swarming xters on
BA
Citrate pos (non-
Salmonella sp NLF K/A - + + + - - + -
typhoid salmonella)
Salmonella typhi NLF K/A - Wk+ + + - - - - Black ppt on
SSA&XLD
Shigella sonnei NLF K/A - - - - - - - - No black ppt on
SSA/XLD
Other Shigella sp NLF K/A - - - - - - -
Vibrio cholerae NLF A/A - + + - - - - + String test-positive
Vibrio NLF K/A + - - + + - + +
parahaemolyticus
Green pigmentation
P.aeruginosa NLF K/NC + - - + - + + +/-
on MHA
Acinetobacter spp NLF NC - - - - - - - - Coccoide rods
Morganella NLF K/A - - + + + + - -
morganii
Providencia spp NLF K/A - - + + + -
Yersinia NLF K/A - - - +(25/-35°C) +/- +/- - -
enterocolitica
Edwardsiella NLF K/A - + + + + - - -
tarda
Grows with
E.coli LF A/A - - + + + - - +/- precipitate bile salt
on MCA
229
MAC
Organism TSI Oxidase H2S Gas Motility Indole Urea Citrate Haemolysis Comment
Reaction
Enterobacter Often resistant to
aerogenes LF A/A - - + + - - + - Ampicillin and
cephalosporin
Grow with very
Klebsiella
LF A/A - - + - - + + - mucoid
pneumoniae
colonies
Klebsiella oxytoca LF A/A - - + - + + + -
Citrobacter Late LF A/A or K/A - + + + - +/- + -
freundii
230
NOTE PAD
231
NOTE PAD
232
NOTE PAD
233
FOR FURTHER INFORMATION CONTACT:
PERMANENT SECRETARY
MINISTRY OF HEALTH,
MAGUFULI GOVERNMENT CITY, AFYA ROAD/STREET, MTUMBA,
PO BOX 743,
40478 DODOMA, TANZANIA.
LANDLINE: +255 (0)26 232 3267
EMAIL: [email protected]
WEBSITE: www.moh.go.tz