Silica Stamolys Ca 71 Si OM
Silica Stamolys Ca 71 Si OM
Stamolys CA71SI
Silicate Analyser
BA364C/07/en/06.04
51512332
Valid of:
Software version 5.4
Brief overview
AUTO MESSEN
3 30
38
163
15
M CE E K
110
Ø 27
15 9 Photometer +17V Photometer -17V Photometer 0V fx
Analog Out Kanal 1
Channel 1
Kanal 2
Schirm/ MP5 MP6 Channel 2
0/4-20mA
250 BK 65 COM
F4 M 0.2 A M 0.2 A Screen 28 22 16
ALARM 2
F5
ALARM 2
330
OPTION
Schirm
GN 64 0V BK 58 40 NC
160 Schirm Schirm
Screen
keine Probe
27 21 15
80 75 V4 70 Screen BN 63 -17 V YE 57 39 I2 + NO
Screen No sample
15 140 OR 79 74 V3 69
n.b.
WH 62 +17 V WH 56 Reserve 38 I2 - COM
26 20 14
SCHRITTMOTOR
n.c. 25 19 13
STÖRUNG
STEP MOTOR
ALARM 1
100
ALARM 1
Mess 2
ERROR
n.b. n.b.
BK 78 73 V2 68 YE 61 fx 55 37 31 NC
n.c. Meas 2 n.c.
60
24 18 12
n.b. n.b.
LEAK S.
+24V
LECKF.
RD 77 72 V 1 BU 67 60 54 36 I1 + 30 NO
n.c. n.c. 23 17 11
71 0 V BN 66 Status GN 53
A
YE 76 GY 59 0V 35 I1 - 29 0V RELAIS / Relay:
max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC
120
D
Klemmen führen auch bei ausgeschaltetem Gerät Spannung !!!
243
Ø 26
Terminals voltage to ground even when the unit is switched off !!!
200
F2 NETZ
AC Motor MAINS
7 L2
Ø 21
T 0.1A
165
M E K
125
AC MOTOR
N 3 L
47
6
RS 232
L1 Elektronik F1 N
5 2
555
524
Electronics
580
648
4 PE T 0.5A 1 PE
78
119
326
Ø 21
Ø 26
ENDRESS+HAUSER R
STAMOLYS CA 71
360 207
436
B C
1 2 3
12 4
M E K RS 232
11 5
A
130 154
E
130 120
A 121
143
B
141
170/171
142
S R1 R2 R3
P 121 122 B
P
135 135
F
C07-CA71XXx-00-08-00-xx-100.eps
Endress+Hauser
Stamolys CA71SI
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 4 8 Accessories . . . . . . . . . . . . . . . . . . . . . 40
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 8.1 Collecting vessel . . . . . . . . . . . . . . . . . . . . . . . . . . 40
1.2 Installation, commissioning and operation . . . . . . . . 4 8.2 Reagents, cleaner, standard solution . . . . . . . . . . . . 40
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 5 8.3 Hose cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 8.4 Additional accessories . . . . . . . . . . . . . . . . . . . . . . 40
1.5 Notes on safety icons and symbols . . . . . . . . . . . . . . 5
9 Trouble-shooting . . . . . . . . . . . . . . . . . 41
2 Identification . . . . . . . . . . . . . . . . . . . . 6 9.1 Trouble-shooting instructions . . . . . . . . . . . . . . . . . 41
2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6 9.2 System error messages . . . . . . . . . . . . . . . . . . . . . . 41
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 9.3 Process errors without messages . . . . . . . . . . . . . . 42
2.3 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 8 9.4 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9.5 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3 Installation . . . . . . . . . . . . . . . . . . . . . . 9 9.6 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6 Commissioning. . . . . . . . . . . . . . . . . . 32
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.2 Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7 Maintenance. . . . . . . . . . . . . . . . . . . . 34
7.1 Maintenance schedule . . . . . . . . . . . . . . . . . . . . . . 34
7.2 Replacing reagents . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.3 Replacing pump hoses . . . . . . . . . . . . . . . . . . . . . . 35
7.4 Checking the sample flow rate . . . . . . . . . . . . . . . . 37
7.5 Replacing valve hoses . . . . . . . . . . . . . . . . . . . . . . 37
7.6 Replacing the static mixer . . . . . . . . . . . . . . . . . . . 38
7.7 Replacing the photometer optical cell . . . . . . . . . . . 38
7.8 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.9 Placing out of service . . . . . . . . . . . . . . . . . . . . . . . 39
Endress+Hauser 3
Safety instructions Stamolys CA71SI
1 Safety instructions
4 Endress+Hauser
Stamolys CA71SI Safety instructions
1.4 Return
If the analyser has to be repaired, please return it cleaned to the sales centre responsible.
Please use the original packaging, if possible.
Please enclose the completed "Declaration of contamination" (copy the second last page of these
Operating Instructions) with the packaging and the transportation documents.
No repair without completed "Declaration of contamination"!
Warning!
# This symbol alerts you to hazards. They can cause serious damage to the instrument or to persons
if ignored.
Caution!
" This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.
! Note!
This symbol indicates important items of information.
Endress+Hauser 5
Identification Stamolys CA71SI
2 Identification
2.1.1 Nameplate
Check the order code on the nameplate (at the analyser) with the product structure (see below) and
your order.
Stamolys CA71
order code / Best.Nr.: CA71SI-A10A2A1
serial no. / Ser.-Nr: 3B90003C3AN1
C07-CA71SIx-18-08-00-xx-001.EPS
Measuring range
A Measuring range 1.0 ... 200 µg/l Si
B Measuring range 50 ... 5000 µg/l Si
Y Special version acc. to customer’s specification
Sample transfer
1 Sample transfer from one measuring point (one-channel version)
2 Sample transfer from two measuring points (two-channel version)
Power supply
0 Power supply 230 V AC / 50 Hz
1 Power supply 115 V AC / 60 Hz
Housing version
1 Without housing
2 With GFK housing
3 With stainless steel 1.4301 (AISI 304) housing
Communication
A 0/4 ... 20 mA, RS 232
Additional equipment
1 Quality certificate
2 Quality certificate + set of inactive reagents
3 Quality certificate + three sets of inactive reagents
6 Endress+Hauser
Stamolys CA71SI Identification
! Note!
Please, order reagents separately with analyser version CA71 XX-XXXXXX1.
With all other versions, inactive reagents are included in the scope of delivery. You have to mix the
reagents before using them. Please, read the mixing instructions attached to the reagents.
! Note!
Please, order reagents separately with analyser version CA71 XX-XXXXXX1.
With all other versions, inactive reagents are included in the scope of delivery. You have to mix the
reagents before using them. Please, read the mixing instructions attached to the reagents.
Endress+Hauser 7
Identification Stamolys CA71SI
2.3.1 4 approval
Declaration of conformity
The product meets the legal requirements of the harmonised European standards.
The manufacturer confirms compliance with the standards by affixing the 4 symbol.
8 Endress+Hauser
Stamolys CA71SI Installation
3 Installation
1 2 3
12 4
M E K RS 232
11 5
10 6
9 8 7
C07-CA71XXx-03-08-00-xx-001.eps
Endress+Hauser 9
Installation Stamolys CA71SI
15/0.59
163/6.42
110/4.33
Ø 27/1.06
15/0.59 9/0.35
250/9.84
330/12.99
160/6.30
15/0.59 140/5.51
100/3.94
60/2.36
A
120/4.72
Ø 26 / 1.02
243/9.57
200/7.87
165/6.50
Ø 21/0.83
125/4.92
47/1.85
580/22.83
648/25.51
555/21.85
524/20.63
78/3.07
119/4.69
Ø 21/0.83
326/12.83
Ø 26/1.02
360/14.17 207/8.15
436/17.17 mm / inch
C07-CA71xxx-06-08-00-en-001.eps
10 Endress+Hauser
Stamolys CA71SI Installation
38/1.5
250/9.84
160/6.30
110/
4.33
140/5.51
Ø 27/1.06
100/3.94
330/12.99
60/2.36
Ø 26/1.02
7-11/
120/
4.72
0.28 - 0.43
(M 6)
243/9.57
Ø 21/0.83
165/6.50
200/7.87
125/4.92
47/1.85
555/21.85
Ø 21/
648/25.51
0.83
78/3.07
119/4.69
Ø 26/1.02
326/12.83
Ø 16/0.63
360/14.17
436/17.17 Ø 27/1.06
465/18.31 120/4.72
500/19.69
mm / inch
C07-CA71xxx-06-08-00-en-002.eps
163 / 6.42
250/9.84
120 / 4.72
160/6.30
Ø 26/1.02
Ø 16/0.63
140/5.51
100/3.94
330 / 12.99
60/2.36
Ø 26/1.02
7-11/
120/
4.72
0.28 - 0.43
(M 6)
243/9.57
Ø 21/0.83
165/6.50
200/7.87
125/4.92
47/1.85
555/21.85
Ø 21/
648/25.51
0.83
78/3.07
119/4.69
Ø 26/1.02
326/12.83
Ø 16/0.63
360/14.17
436/17.17 Ø 27/1.06
465/18.31 120/4.72
500/19.69
mm / inch
C07-CA71xxx-06-08-00-en-020.eps
Endress+Hauser 11
Installation Stamolys CA71SI
381/15
351/13.82
19/0.75 197/7.76
192/7.56
Ø6/0.24
RS 232
540/21.26
562/22.13
175/6.89
325/12.80
9/0.35 8/0.31 mm/inch
C07-CA71xxx-06-08-00-en-003.eps
1 5
2 6
3 7
C07-CA71xxx-11-08-00-xx-001.eps C07-CA71xxx-06-08-02-en-003.eps
12 Endress+Hauser
Stamolys CA71SI Installation
One-channel version
Two-channel version
• Depending on the ordered version, one or two collecting vessels (without level measurement) are
included in the scope of delivery.
• Only one collecting vessel can be mounted at the housing. The second is to be placed nearby the
analyser.
1 2 3
V3 V1 V2 V4
7 6 5 4
C07-CA71xxx-00-08-00-xx-005.eps
Endress+Hauser 13
Installation Stamolys CA71SI
! Note!
The sample can be obtained as follows:
– Directly or after a reversible flow filter or a cross current filter by means of a small pump
(rating approx. 300 ml/min), suitable for clear media, e.g. in the discharge channel of a
sewage treatment plant
– From a sedimentation tank or after microfiltration; suitable for media containing flocculants,
e.g. in an activated sludge basin
– Sample conditioning using ultrafiltration for heavily soiled media,
e.g. from the primary settling tank
For questions regarding sample conditioning and its automation, please contact your Service
or your Sales Centre responsible.
8. Connect the tubes from canisters containing reagents, standard and cleaning agents to the
following nozzles:
Sample P
Reagent 1 SI1
Reagent 2 SI2
Reagent 3 SI3
Standard solution S
Cleaner R
! Note!
• The pressure of the hose cassette is set at the factory such that the sample and reagent are fed in
without bubbles.
• Only change the pressure if the factory setting does not meet your requirements. The setting is
changed by turning the adjustment screw using a 2.5 mm Allen key.
• The minimum flow rate in the sample line must be 14 ml/min.
14 Endress+Hauser
Stamolys CA71SI Installation
! Note!
Ensure that there is always sufficient sample
C07-CA71xxx-11-08-00-xx-006.eps
available for both analysers. Observe this when Fig. 11: Installation example
selecting maintenance intervals for CAT411 and
1 Sample from CAT430
CAT430.
2 Sample from CAT411
3 Collecting vessel
4 Collecting vessel overflow
5 Sampling lines
6 T-pieces
Endress+Hauser 15
Wiring Stamolys CA71SI
4 Wiring
" Caution!
Before swinging out the analyser frame, remove the hoses from the outlet pipe. Otherwise there is
danger of overflow. After swinging in the frame, reconnect the hoses.
! Note!
To reach the terminal strip, you must swing out the analyser frame.
1. Loosen the two lower Allen screws SW 6 by 3 to 4 turns (Fig. 12, pos. 1).
2. Unscrew the two upper Allen screws until the analyser frame swings out. Thus you can reach
the terminal strip (pos. 2).
M E K RS 232
1 2
C07-CA71xxx-04-08-00-xx-002.eps
0V fx Kanal 1 Kanal 2
Photometer +17V Photometer -17V Photometer Analog Out Channel 1
Schirm/ MP5 MP6 Channel 2
BK 65 0/4-20mA COM
F4 M 0.2 A M 0.2 A Screen 28 22 16
ALARM 2
F5
ALARM 2
OPTION
Schirm
GN 64 0V BK 58 40 NC
Screen 27 21 15
Schirm Schirm keine Probe
80 75 V4 70 Screen BN 63 -17 V YE 57 39 I2 + NO
Screen No sample 26 20 14
n.b.
OR 79 74 V3 69 WH 62 +17 V WH 56 Reserve 38 I2 - COM
SCHRITTMOTOR
n.c. 25 19 13
STÖRUNG
STEP MOTOR
ALARM 1
ALARM 1
Mess 2
ERROR
n.b. n.b.
BK 78 73 V2 68 YE 61 fx 55 37 31 NC
n.c. Meas 2 n.c. 24 18 12
n.b. n.b.
LEAK S.
+24V
LECKF.
RD 77 72 V 1 BU 67 60 54 36 I1 + 30 NO
n.c. n.c. 23 17 11
YE 76 71 0 V BN 66 GY 59 Status GN 53 0V 35 I1 - 29 0V RELAIS / Relay:
max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC
Terminals voltage to ground even when the unit is switched off !!!
F2 NETZ
AC Motor MAINS
7 L2 T 0.1A
AC MOTOR
6 N 3 L
L1 Elektronik F1 N
5 2
Electronics
4 PE T 0.5A 1 PE
C07-CA71xxxx-04-08-00-a2-001.eps
16 Endress+Hauser
Stamolys CA71SI Wiring
! Note!
The device does not have a mains switch. Therefore, it is advantageous to have a fused socket near
to the device.
L 3 3
Mains N 2 2
PE 1 1
COM 25 25
NO 23 23
COM 28 28
NO 26 26
COM – 13
NO – 11
COM – 16
NO – 14
COM 19 19
Fault NC 18 18
NO 17 17
COM 22 22
NO 20 20
+ 36 36
Analog output 1
– 35 35
0/4 ... 20 mA
Screen PE1 PE1
+ – 39
Analog output 2
– – 38
0/4 ... 20 mA
Screen – PE1
Input – 55
Channel switch-over
0V – 53
1) Brass screw with bolt top right in the connection compartment (marked with *)
Endress+Hauser 17
Wiring Stamolys CA71SI
! Note!
• Alarm values 1 and 2 do not need to be connected if the PLC sets its own alarm values at the
analog output.
• When using a sample conditioning system:
Connect terminals 57 and 53 on the analyser to the corresponding terminals on the sample
conditioning system. For the allocation of these terminals, please see the sample conditioning
system operating instructions.
• If there is a 24 V voltage at terminal 57, the analyser will not begin measurement (sample not
ready). To start measurement, the voltage must stay at 0 V for at least 5 seconds.
C07-CA71xxx-03-08-00-xx-001.eps
! Note!
With the two-channel version, place the cable cores of all cables (data cables to analog output 1 and
to analog output 2) through the interference suppressor.
18 Endress+Hauser
Stamolys CA71SI Wiring
Meas. 2 0 V = channel 1
Channel selection
24 V = channel 2
1) Alternative selection
Connection Terminal connection for Terminal connection for Terminal connection for
condition fulfilled condition not fulfilled power off
AV 1 A: 25 - 23 A: 25 - 24 25 - 24
R: 25 - 24 R: 25 - 23
AV 2 A: 28 - 26 A: 28 - 27 28 - 27
R: 28 - 27 R: 28 - 26
Fault A: 19 - 17 A: 19 - 18 19 - 18
R: 19 - 18 R: 19 - 17
Unassigned 22 - 20 22 - 21 22 - 21
16 - 14 16 - 15 16 - 15
13 - 11 13 - 12 13 - 12
Endress+Hauser 19
Wiring Stamolys CA71SI
Two-channel version
Connection Terminal connection for Terminal connection for Terminal connection for
condition fulfilled condition not fulfilled power off
AV 1 - 1 A: 25 - 23 A: 25 - 24 25 - 24
R: 25 - 24 R: 25 - 23
AV 1 - 2 A: 13 - 11 A: 13 - 12 13 - 12
R: 13 - 12 R: 13 - 11
AV 2 - 1 A: 28 - 26 A: 28 - 27 28 - 27
R: 28 - 27 R: 28 - 26
AV 2 - 2 A: 16 - 14 A: 16 - 15 16 - 15
R: 16 - 15 R: 16 - 14
Fault A: 19 - 17 A: 19 - 18 19 - 18
R: 19 - 18 R: 19 - 17
Channel ½ A: 22 - 20 A: 22 - 21 22 - 21
measurement end R: 22 - 21 R: 22 - 20
A = NO current configured
R = NC current configured
! Note!
Condition fulfilled means:
• AV 1: concentration > Alarm value 1
• AV 2: concentration > Alarm value 2
• Fault: error occurred
Contacts AV 1, AV 2 and fault are only affected during automatic operation.
3 TxD RxD 2
2 RxD TxD 3
8 CTS RTS 7
CTS 8
5 GND GND 5
The results (measured value+unit of measure+CR) are output in the "Data memory Measured
values" menu.
The calibration results (measured value+unit of measure+CR) are output in the "Data
memory-Calibration factors" menu.
! Note!
• A null modem cable is required (not a crossed one).
• The analyser does not have to be configured for the interface.
The following commands can be sent from the PC, in order to read out data:
• "D" = Data memory-Measured values
• "C" = Data memory-Calibration factors
• "S" = Setup (parameter entry, configuration...)
• "F" = Frequency (current)
20 Endress+Hauser
Stamolys CA71SI Wiring
Does the supply voltage correspond to the data on the nameplate? 230 V AC / 50 Hz 115 V AC / 60 Hz
Cable type properly disconnected? Guide power supply and signal lines
separate over the entire travel distance.
Separate cable channels are ideal.
Are power supply and signal lines connected correctly according to wiring
diagram?
Endress+Hauser 21
Operation Stamolys CA71SI
5 Operation
1 2
M CE E K RS 232
4 3
C07-CA71xxx-19-08-00-xx-005.eps
F – Select option
– Adopt value, forwards in the sub-menu (horizontal)
Retrieve error message
22 Endress+Hauser
Stamolys CA71SI Operation
! Note!
"Measuring" flashes when the analyser is ready for the next measurement but has not yet received
the enable signal from the sample collector or the sample conditioning unit.
Endress+Hauser 23
Operation Stamolys CA71SI
5.3.3 CONFIGURATION
! Note!
Some settings that can be made in this menu affect the defaults in the PARAMETER ENTRY menu.
In view of this, complete the CONFIGURATION menu first during initial start-up.
! Note!
In the following table and in the tables in the next chapter, example images can be found for each
option under "Display". In addition to the numerical values, the parameter is also displayed in some
options. This is not shown in the images. Moreover, individual numerical values in the images can
differ from the actual settings.
The actual factory settings can in any case be found in column 2 "Range of adjustment / Factory
settings" in bold.
Input 03.
Code number 03 Code-Nr. If an incorrect code is entered the program exits the
0
sub-menu.
Concentration c-Offset The offset specifies the zero shift of the calibration
0.00 ... 50.0 mg/l
offset + 0.00 mg/l function. (Change the sign with the J key.)
Dosing time for sample or standard (20 ... 120 s). During
20 ... 300 s this time, the entire system is flushed with sample or
Delay to sample 120 s (SI-A) Delay to sample standard, so when the reagent is added there is definitely
80 s
160 s (SI-B) only fresh sample in the mixer. If there is sufficient
sample available, select the highest possible value.
24 Endress+Hauser
Stamolys CA71SI Operation
Alarm val. 1-1 Setting for whether contact for alarm value 1, channel 1
Alarm value AV 1-1
norm. closed works as NO current or NC current contact.
Alarm val. 2-1 Setting for whether contact for alarm value 2, channel 1
Alarm value AV 2-1
norm. closed works as NO current or NC current contact.
NO current
NC current Two-channel version only!
Alarm value AV 1-2
! Note!
Changes will only be activated after a
Alarm val. 1-2
norm. closed
Setting for whether contact for alarm value 1, channel 2
works as NO current or NC current contact.
Reset (Power off/on)!
Two-channel version only!
Alarm value AV 2-2 Alarm val. 2-2 Setting for whether contact for alarm value 2, channel 2
norm. closed
works as NO current or NC current contact.
Frequency offset1
Calibrate offset yes / no Calibrate offs Pressing the J key starts a blind value measurement for
yes: K no: E
compensating the reagent's inherent colour.
1) Do not change the default setting. The frequency offset has to be 0 Hz. A determination is not necessary.
Endress+Hauser 25
Operation Stamolys CA71SI
Alarm value SI-A: 1 ... 200 µg/l / 100 µg/l Alarm val. 1-1 Concentration threshold value limit relay 1, channel 1
AV 1 - 1 SI-B: 100 ... 5000 µg/l / 2500 µg/l 2.50 mg/l (differential hysteresis 2% of alarm value).
Alarm value SI-A: 1 ... 200 µg/l / 200 µg/l Alarm val. 2-1 Concentration threshold value limit relay 2, channel 1
AV 2 - 1 SI-B: 100 ... 5000 µg/l / 5000 µg/l 1.25 mg/l (differential hysteresis 2% of alarm value).
26 Endress+Hauser
Stamolys CA71SI Operation
0 ... 720 h
Flush. interval Time between two flushings.
Flushing interval 0 h (SI-A)
48 h The "0 h" setting stops cleaning.
48 h (SI-B)
1) All channels set to 0 means, the channel selection is provided by an external device. All channels set to 1 means, alterning beginning with channel 1.
! Note!
• Always synchronise calibration and flushing time.
• Carry out flushing with standard cleaning solution approx. 3-4 hours before the next calibration.
• Flushing with special cleaning solution (e.g. hydrochloric acid) has a lasting effect on calibration.
Therefore, carry out this cleaning 3-4 hours after calibration.
5.3.5 LANGUAGE
The following languages are available:
• Deutsch
• English
• Français
• Suomi
• Polski
• Italiano.
! Note!
• This menu is a "Read-Only-Menu".
• You can find the individual error messages, their meaning and solutions to problems in chapter
"Trouble-shooting instructions".
• If there is at least one error message, the signal output is set to "fault".
• Causes of faults are requested for every measurement. If an error which occurred previously no
longer exists, it is automatically cancelled. If this should not happen automatically, error messages
can be deleted by quickly switching the analyser off and back on again.
Endress+Hauser 27
Operation Stamolys CA71SI
5.3.7 SERVICE
Navigate through the main menu with the W key to the "SERVICE" option and press the F key
to enter the menu.
! Note!
P1 and P2 are inactiv, when G is at go. If P1 or P2 is at
go, G is not available.
The following valve combinations are possible:
(applies to one-channel and two-channel version,
whereby where the latter is concerned selection is made
by positioning valve 3 between channel 1 and 2)
– V1: P, V2: S
Passage for the sample. This combination is
automatically reset on leaving the service menu.
– V1: S, V2: S
Passage for standard solution
– V1: S, V2: R
Passage for cleaning agent
28 Endress+Hauser
Stamolys CA71SI Operation
! Note!
Two menus, "DATA MEMORY 1" and "DATA MEMORY 2", are only for the two-channel
version. In the one-channel version there is only one menu "DATA MEMORY".
! Note!
To enter this menu select the CONFIGURATION menu, browse to the "Calibration factor" option
and press the J key.
Endress+Hauser 29
Operation Stamolys CA71SI
5.4 Calibration
! Note!
Compare these values to your own data.
After changes in the CONFIGURATION menu and in case of software updates, you can check and,
if necessary, change the calibration data in the sub-menu.
5.4.2 Interferences
30 Endress+Hauser
Stamolys CA71SI Operation
" Caution!
After the calibration has finished, the analyser automatically goes into measuring mode. You now
have to set the time of the 1st calibration back into the future in order to align the calibration and
rinsing times to each other. The rinse must be performed 3-4 hours before the next calibration.
Proceed as described above to change the setting for the time of the 1st calibration. After changing
to measuring mode, the analyser automatically begins measuring, rinsing and calibrating at the
defined times.
Endress+Hauser 31
Commissioning Stamolys CA71SI
6 Commissioning
6.2 Switch-on
! Note!
• If possible, let the analyser warm up in standby mode before commissioning ("Auto measuring"
display). The time can be defined via the "1st measurement" option in the PARAMETER ENTRY
menu.
• At the start of measurement with a cold analyser, the first measuring results will be errored. The
reaction is temperature-dependent and if the temperature is too low the pre-defined reaction time
is insufficient for a complete reaction. For this reason, never carry out calibration with a cold
analyser. Wait at least two hours before carrying out calibration.
When the analyser has been configured and calibrated, the measuring cycle starts automatically.
Entering parameters is no longer necessary.
To perform initial start-up or to readjust the device parameters, proceed as follows:
1. Plug-in the plug into a socket.
2. Press the H key until AUTO MEASURING is displayed.
3. Select the CONFIGURATION menu and program the individual options up to and including
the "Current date/time". With H you can return to the main menu.
4. Now complete the PARAMETER ENTRY and SERVICE menus.
With H you can return to the main menu.
5. Select CONFIGURATION again and use F to go to the "Calibrate offset" option.
6. Connect a vessel containing distilled water to the "Sample" connection and start the frequency
offset (J key). The registered value is displayed and saved.
7. Then reconnect the sampling line.
With H you can return to the main menu.
The analyser starts the "Calibration", "Measurement", and "Flushing" procedures automatically
(triggered by control signal or integrated timer) in accordance with the device parameters that you
have set (1st calibration, 1st measurement, 1st flushing times and the respective intervals control
the temporal procedure).
32 Endress+Hauser
Stamolys CA71SI Commissioning
The following summary shows procedures carried out in the device and the programmed intervals:
Function Duration [s] Range of adjustment
Flushing Pump cleaning solution Flush hold on: 2 PARAMETER ENTRY / "Flush hold on"
Allow to react 5
Pump cleaning solution Flush hold on: 2
! Note!
For the two-channel version, valve V3 must be set additionally for switch-over between channel 1
and channel 2.
8. Now proceed as with dry commissioning (from step 2).
Endress+Hauser 33
Maintenance Stamolys CA71SI
7 Maintenance
" Caution!
You must not carry out any procedures not listed in the following chapters, yourself.
This work must only be carried out by the service.
monthly – Flush sample line hose system with – see chapter "Replacing reagents"
pressurized water (disposable syringe), check – Remove the cassette of the
and replace sample pump.
reagents if necessary – Connect the one-way syringe instead of the
– Flush the sample tubing system with 12.5% sample inlet.
bleaching lye (sodium hypochloride) and – SERVICE:
reflush thoroughly with water V1: S, P1: g, P2: s, V2: S
# Warning!
Corrosive. Wear protective gloves and
Add solution to sampling connection
every 6 months – Replace pump hoses – see chapter "Replacing pump hoses"
– Replace valve hoses
! Note!
Whenever working on the reagent hoses, the hoses must be disconnected from the canisters, in
order to prevent contamination of the reagents.
34 Endress+Hauser
Stamolys CA71SI Maintenance
Endress+Hauser 35
Maintenance Stamolys CA71SI
1 Wheel head
1 2 2 Upper holder of the cassette
3 Lower holder of the cassette
4 Guidance at the pump hose
5 Nipple with guidance
5 3
C07-CA71xxx-00-08-00-xx-998.eps
Fig. 16:
" Caution!
Ensure that you connect the new pump hoses to the correct connections at theT-connector.
For the order numbers of the pump hoses see chapter "Trouble-shooting"/"Spare parts".
36 Endress+Hauser
Stamolys CA71SI Maintenance
1 2
C07-CA71XXx-03-08-00-xx-002.eps C07-CA71XXx-00-08-00-xx-997.eps
Fig. 17: Sample hose at pump outlet Fig. 18: Sample pump
1 Sample hose from pump outlet 3 Adjusting screw for the application force
2 T-piece
Check the sample flow rate after replacing the pump hoses:
1. Remove the sample hose from the T-piece (Fig. 17, pos. 2) and put it into a measuring cylinder.
2. Switch on the sample pump (SERVICE menu).
3. Measure the sample volume in a defined time interval.
4. If the determined flow rate is not 14 ± 0.5 ml/min, you must adjust the application force by
the adjusting screw (Fig. 18, pos. 3, Allen screw) at the hose cassette.
5. Measure resp. adjust until the flow rate is correct.
Endress+Hauser 37
Maintenance Stamolys CA71SI
1 2 3
V3 V1 V2 V4
7 6 5 4
C07-CA71xxx-00-08-00-xx-005.eps
" Caution!
Do not under any circumstances touch the optical window of the cell with your fingers!
Otherwise, traces of grease remain on the optical surfaces. This can lead to corrupted measured
values.
38 Endress+Hauser
Stamolys CA71SI Maintenance
7. Connect the cell to the hoses such that the sample is fed in from below.
8. Secure the hoses with the supplied cable connectors to stop them from slipping off the cell.
9. Reassemble the photometer and tighten the screws.
10. Attach the photometer housing and screw it down.
11. After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).
12. Carry out a calibration (PARAMETER ENTRY).
7.8 Cleaning
" Caution!
When cleaning, ensure that you do not damage the nameplate on the analyser. Do not use any
solvent-based cleaning agents.
To clean the analyser housing, proceed as follows:
• Stainless steel housing (stainless steel SS 1.4301 (AISI 304)):
with a lint-free cloth and Glittol RG 10.51
• GFR housing:
with a damp cloth or with tenside-based (alkaline) cleaning agent.
" Caution!
Before placing the instrument out of service, thoroughly rinse all of the lines of the measuring system
with clean water.
To place the analyser out of service, proceed as follows:
1. Remove the reagent and standard hoses from the canisters and immerse them in a tank
containing clean water.
2. Switch valve 1 to "Standard" and switch pumps 1 and 2 on for one minute (SERVICE menu).
3. Remove hoses from the water and allow the pumps to run until the hoses are completely dry.
4. If you are using a continuous sample supply,
disconnect sampling line.
5. Flush sampling hoses with clean water and then with compressed air, in order to completely
empty the hoses.
6. Remove the valve hoses from the valves.
7. Remove the load from the pump hoses by removing the hose cassette from the bracket below.
Endress+Hauser 39
Accessories Stamolys CA71SI
8 Accessories
! Note!
Please, note the disposal instructions in the safety data sheets of the reagents!
40 Endress+Hauser
Stamolys CA71SI Trouble-shooting
9 Trouble-shooting
If a calibration fails, you can enter a new calibration factor manually (CONFIGURATION menu,
"Calibration factor"). Cancel the error message by switching the analyser off and on briefly.
If the error occurs frequently, you will have to search for the cause.
Air bubbles in system Start calibration manually (PARAMETER ENTRY, "1st calibration", change the date accordingly,
start measurement) or enter a new calibration factor.
Incorrect concentration of standard Check the concentration in the laboratory. Adjust the standard accordingly (PARAMETER ENTRY,
Calibration failed "Calibration solution") or replace the standard.
Reagents contaminated or aged Simple check: Add about 10 to 100 ml of standard solution with about 1 ml of reagent in a beaker.
If it does not change colour after max. 10 min, replace the reagents.
Standard dosing defective Check the valves for contamination, obstructions (visual inspection). Replace valve hoses if
neccessary.
Optical cell dirty Not enough light intensity sent to – Flush with 12.5% bleaching lye
receiver, e.g. due to sedimented – When using a CAT430:
particles Check the filter.
Wrong photometer Wrong photometer Check the setting in the CONFIGURATION menu, "Photometer".
Leak at canisters or hoses Replace defective components and clean and dry the CA 70 or those components affected by the
Spillage error
leak.
No measuring signal Electrical connection Check all electrical connections and the tight fit of the fuses.
Endress+Hauser 41
Trouble-shooting Stamolys CA71SI
Reagents contaminated or Simple check: Add about 10-100 ml of standard solution with about 1 ml of reagent in a vessel. If it does
aged not change colour after max. 10 min, replace the reagents .
Measured values always
the same No sample, no reagents Make sure sample and reagents are supplied, check level monitor and clean if necessary
Incorrect concentration of Check the concentration in the laboratory. Adjust the standard accordingly
standard ("PARAMETER ENTRY", "Calibration solution") or replace the standard.
Reagents contaminated or Simple check: Add about 10-100 ml of standard solution with about 1 ml of reagent in a vessel. If it does
aged not change colour after max. 10 min, replace the reagents.
Reagent blank value too After replacing the reagents, carry out an offset calibration and then a calibration (CONFIGURATION,
high "Offset calibration")
Wrong dimension Check the setting in the CONFIGURATION menu, "Measuring unit".
Wrong optical cell Check the setting in the CONFIGURATION menu, "Photometer"
Sample suction time too Increase the suction time (CONFIGURATION, "Delay to sample")
short
Measured values Matrix effects Detect interfering substances (see Technical Information, "Measuring principle"),
inaccurate (substances that interfere possibly use sample conditioning
with the photometric
methods)
Standard is dosed to the Check the valves and valve settings. Replace valve hoses if necessary.
sample
Filter life too long Take counter sample at analyser inlet and check the concentration in the laboratory. If the analyser
measured values have no deviations, clean ultrafiltration modules or backflush filters more frequently.
Optical cell dirty Flush with 12.5% bleaching lye first and then with 5% hydrochloric acid
Counter sample in the Sample ageing Shorten time between sampling and analysis.
laboratory delivers
deviating measured
values
Wrong size of analog Check the setting (CONFIGURATION, "Analog output 1" " 2").
output
Measured value transfer
Incorrect measuring Adjust measuring range (PARAMETER ENTRY, "Measuring range")
errored
range
Background noise Check line for interference from strong sources of induction.
Analyser will not No power Check the electrical connection and ensure power supply.
switch on Fuse Replace fuse F1 (time-lag 0.5 A)
Analyser is running but Initialisation failed Switch the analyser off and after approx. 30 seconds back on.
display reading is garbled
or off
Leak sensor bypassed Interrupt contact between two leak sensors (pins 67-66)
Pumps fail to run
Fuse Check all fuses and replace if necessary.
Time of 1st measurement The date must be between 01.01.1996 and the current date.
"Measurement" display not reached
flashes
Interval not expired Change parameters.
42 Endress+Hauser
Stamolys CA71SI Trouble-shooting
Time of 1st calibration The date must be between 01.01.1996 and the current date.
not reached
Time of 1st flushing not The date must be between 01.01.1996 and the current date.
Flushing does not start reached
155 131 134 185 161 160 161 160 144 200
V3 V1 V2 V4 184 184
130 A 154
V4
V1 V2
H
121 170/171 R S R2 R3
R1
121
B
135 135 123 182
120
C
C07-CA71XXx-00-08-xx-111.eps
Fig. 20 shows the components of the analyser. Please, take the spare parts order numbers from the
following sections.
Endress+Hauser 43
Trouble-shooting Stamolys CA71SI
V3 V1 V2
130 A 154
V1 V2
H
121 170/171 R S R2 R3
R1
121
B
135 135 142 123
122 120
C
C07-CA71XXx-00-08-xx-102.eps
44 Endress+Hauser
Stamolys CA71SI Trouble-shooting
Fig. 21 shows the components of the analyser. Please, take the spare parts order numbers from the
following sections.
9.5 Return
If the device requires repair, please send it cleaned to the sales centre responsible. Please use the
original packaging, if possible.
Please enclose the completed "Declaration of contamination" (copy the second last page of these
Operating Instructions) with the packaging and the transportation documents.
No repair without completed "Declaration of contamination"!
9.6 Disposal
The device contains electronic components and must therefore be disposed of in accordance with
regulations on the disposal of electronic waste. Please observe local regulations.
Endress+Hauser 45
Technical data Stamolys CA71SI
10 Technical data
10.1 Input
10.2 Output
Signal on alarm Contacts: 2 limit contacts (per channel), 1 system alarm contact
optional: end of measurement (with two channel version display of channel
no. available)
Time between two measurements tmes = reaction time + sampling + waiting time + rinse again time + 2x
filling time + reagent refusal time (min. waiting time = 0 min)
Calibration interval 0 to 72 h
Sampling selectable from 20 to 300 s (standard = 160 s for SI-A, 120 s for SI-B)
46 Endress+Hauser
Stamolys CA71SI Technical data
10.5 Environment
Ingress protection IP 43
10.6 Process
Valves Tygon®
Endress+Hauser 47
Appendix Stamolys CA71SI
11 Appendix
AUTO
MEASURING F Calibration Measuring Flushing Waiting
PARAMETER
ENTRY F Measuring
range start
Measuring
range end
Alarm value
1
Alarm value
2
1st
Measurem.
Measuring
interval
1st
Calibration
Calibration
interval
CONFI-
GURATION F Code
number
Photometer
Default
settings
Measuring
unit
Calibration
factor
Concentrat.
offset
Dilution
Delay to
sample
LANGUAGE
F Deutsch English Français Suomi Polski Italiano
ERROR
DISPLAY F Error
messages
SERVICE
F Pumps
and valves
Frequency
DATA
MEMORY F Measured
values
Output Clear data
48 Endress+Hauser
Stamolys CA71SI Appendix
Analog Alarm value Alarm value Error act. Date/ Calibrate Frequency
output 1 2 contact time offset offset
Endress+Hauser 49
Appendix Stamolys CA71SI
Two-channel version
AUTO
MEASURING F Calibration Measuring Flushing Waiting
PARAMETER
ENTRY F Measuring
range start1
Measuring
range start2
Measuring
range end 1
Measuring
range end 2
Alarm value
1-1
Alarm value
2-1
Alarm value
1-2
Alarm value
2-2
CONFI-
GURATION F Code
number
Photometer
Default
settings
Measuring
unit
Calibration
factor
Concentr.
offset
Dilution
Delay to
sample
LANGUAGE
F Deutsch English Français Suomi Polski Italiano
ERROR
DISPLAY F Error
messages
SERVICE
F Pumps
and valves
Frequency
DATA
MEMORY 1 F Measured
values
Output Clear data
DATA
MEMORY 2 F Measured
values
Output Clear data
50 Endress+Hauser
Stamolys CA71SI Appendix
Analog Analog Alarm value Alarm value Alarm value Alarm value Error act. Date/ Calibrate Frequency
output 1 output 2 1-1 2-1 1-2 2-2 contact time offset offset
Endress+Hauser 51
Appendix Stamolys CA71SI
to Fax no:
52 Endress+Hauser
Stamolys CA71SI Appendix
to Fax no:
1) Please, take the photometer type and the order no. from chapter "Trouble-shooting/Spare parts" and write it down here!
Endress+Hauser 53
Appendix Stamolys CA71SI
Place:
Type:
Serial no. analyser:
Serial no. Photometer:
Software version:
Date:
Photometer Type:
Measuring unit:
Calibration factor:
Dilution:
Delay to sample: s
Frequency offset: Hz
AV 1: ❑ mg/l ❑ µg/l
AV 2: ❑ mg/l ❑ µg/l
1st measuring:
1st calibration:
Calibration interval: h
1st rinsing:
Rinse interval: h
Rinse time: s
54 Endress+Hauser
Stamolys CA71SI Appendix
Submenu
Error mask:
MB >:
MBE:
Rinse again:
Filling time:
Reaction time:
U/min:
K floating mean:
N: Points
Endress+Hauser 55
Appendix Stamolys CA71SI
Form
weekly
⇒ Check and note calibration factor
⇒ visible control (Soiling or blinding, pump tubings, reagent, sample inlet etc.)
done KW 1 KW 2 KW 3 KW 4 KW 5 KW 6 KW 7 KW 8 KW 9 KW 10 KW 11 KW 12
date
done KW 13 KW 14 KW 15 KW 16 KW 17 KW 18 KW 19 KW 20 KW 21 KW 22 KW 23 KW 24
date
done KW 25 KW 26 KW 27 KW 28 KW 29 KW 30 KW 31 KW 32 KW 33 KW 34 KW 35 KW 36
date
done KW 37 KW 38 KW 39 KW 40 KW 41 KW 42 KW 43 KW 44 KW 45 KW 46 KW 47 KW 48
date
done KW 49 KW 50 KW 51 KW 52 KW 53
date
2-weekly
⇒ Check concentration of the calibration factor in the laboratory
May be Concentration in the menue parameter entry chanching or new standard produce.
⇒ Flush sample tubing system with pressurised water (disposable syringe). Remove hose holder unit from the
pump.
Done KW 1 KW 3 KW 5 KW 7 KW 9 KW 11 KW 13 KW 15 KW 17 KW 19 KW 21 KW 23
Date
Done KW 25 KW 27 KW 29 KW 31 KW 33 KW 35 KW 37 KW 39 KW 41 KW 43 KW 45 KW 47
Date
Done KW 49 KW 51 KW 53
Date
C07-CA71XXx-16-08-00-en-006.pdf
56 Endress+Hauser
Stamolys CA71SI
Index
A I
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Hose cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Maintenance kit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 9–10, 13
Standard solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Analyser overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Interference suppressor . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
AUTO MEASURING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Interferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
C K
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Klemmenleiste erreichen . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Checking L
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 LANGUAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 M
Cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Collecting vessel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 32 Maintenance kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Dry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Wet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Menu
CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Auto Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Connection Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Sample line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Error display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Parameter entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
D Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
DATA STORAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 8 N
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Nameplate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
O
Operating matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
E Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 22
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
ERROR DISPLAY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Process errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
P
System errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 PARAMETER ENTRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Performance characteristics . . . . . . . . . . . . . . . . . . . . . . . . 46
F Photometer optical cell . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Placing out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
H Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Hose cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Product structure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Pump hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Pumps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Q
Quality certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Endress+Hauser 57
Stamolys CA71SI
R
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35, 40
Replacing
Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Photometer optical cell. . . . . . . . . . . . . . . . . . . . . . . . . 38
Pump hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Return. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 45
S
Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Sample flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Sample line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Screening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Serial interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
SERVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Signal connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Standard solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Switching contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46–47
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
V
Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
W
Wiring
Connection sticker . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
58 Endress+Hauser
Declaration of contamination
Dear customer,
Because of legal determinations and for the safety of our employees and operating equipment, we need this "Declaration of
contamination" with your signature before your order can be handled. Please, include the completely filled in declaration with the
device and the shipping documents in any case. Add also safety sheets and / or specific handling instructions if necessary.
SAFE
Company data
Company: Contact person:
Department:
Address: Phone:
Fax / e-mail:
Your order no.:
I hereby certify that the returned equipment has been cleaned and decontaminated acc. to good industrial practices and is in
compliance with all regulations. This equipment poses no health or safety risks due to contamination.
BA364C/07/en/06.04
Printed in Germany / FM+SGML 6.0 / DT
51512332