Analytical Methods
Analytical Methods
March 2021
2. Specifications
Outline of the Presentation
2. Specifications
4. Analytical Methods
8
Analytical Method Development……ctd
Validation data,
degradation products,
process impurities,
excipients,
23
Content and format of analytical procedures
• .
24
Content and format of analytical procedures cont…
• The following is a list of information that should typically
be included in a description of an analytical procedure.
Principle
Sampling and Preparation of Samples
Equipment and Equipment Parameters
Reagents
Preparation of Standards
Procedure
System Suitability Testing
Calculations
Reporting of Results
Qualitative and quantitative limits
25
Compendial methods-Verification
All methods are appropriately validated as specified under
Validation of Compendial Procedures (1225).
A validated method may be used to test a new formulation only
after confirming that the new formulation does not interfere with
the accuracy, linearity, or precision of the method (i.e.
Verification).
It may not be assumed that a validated method could correctly
measure the active ingredient in a formulation that is different
from the one used in establishing the original validity of the
method
Therefore, assay method for APIs may not be required?!
Compendial methods….ctd
Compendial methods should be verified for its proper
performance especially for dosage forms.
Nature of Matrices,
Intended applications,
Detector wavelength
flow rate
Specificity Precision
Accuracy LOD/LOQ
Robustness
Linearity
Pre-validation Works
32
Table: Test parameters with acceptance criteria
Precision -
Repeatability - + - +
Interm. precision - + - +
Specificity + + + +
Detection limit - - + -
Quantitation limit - + - -
Linearity - + - +
Range - + - +
Robustness - + - +
34
1.Specificity
• The ability to assess unequivocally the analyte in the
presence of components that may be expected to be
present, such as impurities, degradation products, and
matrix components
Specificity Applied
• Identification
– to ensure the identity of the analyte
• Purity Test
– accurate statement of the content of impurities of an
analyte (related substances, residual solvents, etc.)
• Assay
– an exact result which allows an accurate statement on the
content of potency of the analyte in a sample
36
Specificity Cont..
– Assay by titration
– Identification by UV absorbance
• When the criteria are not met, this indicates that the
method is not sufficiently developed
38
Samples for Specificity Testing
45
Linearity, practical example
Calibration Levels Concentration (micgr/mL) Response (Peak Area) Response factor
1 5 301568 60313.6
2 10 567413 56741.3
3 20 1126639 56332.0
4 40 2248305 56207.6
5 60 3421719 57028.7
6 80 4544013 56800.2
7 100 5702036 57020.4
8 120 6743868 56198.9
9 140 7885679 56326.3
10 160 8846651 55291.6
11 180 10002721 55570.7
12 200 10915064 54575.3
13 240 13286836 55361.8
14 280 15360306 54858.2
46
15 320 19269779 56024.1
16 360 21170824 58807.8
1.8
Linearity, practical example (Excluding
the reds)
Chart Title
18000000
12000000
Axis Title
10000000
Series1
8000000
Linear (Series1)
6000000
4000000
2000000
0
0 50 100 150 200 250 300
Axis Title 47
Linearity, practical example (Including
the reds)
Chart Title
25000000
y = 57901x - 165414
R² = 0.9965
20000000
15000000
Axis Title
Series1
10000000 Linear (Series1)
5000000
0
0 50 100 150 200 250 300 350 400
Axis Title 48
3. Accuracy
49
Accuracy…….
• For a drug substance, the common method of determining
accuracy is to apply the analytical procedure to the drug
substance and to quantitate it against a reference standard of
known purity.
• For the drug product, accuracy is usually determined by
application of the analytical procedure to synthetic mixtures of
the drug product components or placebo dosage form to
which known quantities of drug substance of known purity
have been added
• Experimental Considerations.
Typically, known amounts of related substances and the drug
substance in placebo are spiked to prepare an accuracy sample
of known concentration of related substance and drug product
Accuracy Cont…
Acceptance criteria
≥ 10 98 –102%
1 -10 90 –110%
0.1 – 1 80 – 120%
< 0.1 75 – 125%
51
Table Recovery of D1 and D2 three concentration levels 50 %, 100 % and 150 % of
the analytical concentration in triplicate preparations at each level
Levels Average peak area spiked Average peak area of Amount add (mg) Amount recovered (mg) Percent
sample standard recovered recovered
D1 D2 D1 D2 D1 D2 D1 D2 D1 D2
50 % - 1 3585970.0 7113609.0 1333420.3 2353278.9 2.0015 2.0027 2.0067 2.0349 100.3 101.6
50 % - 2 3615666.0 7162589.0 1351445.1 2377594.0 2.0015 2.0027 2.0344 2.0552 101.6 102.6
50 % - 3 3642196.5 7170121.5 1354633.1 2335796.7 2.0015 2.0027 2.0393 2.0203 101.9 100.9
100 % - 1 6252466.0 12056729.5 4011587.6 7321064.3 6.0044 6.0080 6.1114 6.1811 101.8 102.9
100 % - 2 6207407.5 12050038.5 3919844.1 7215713.7 6.0044 6.0080 5.9708 6.0932 99.4 101.4
100 % - 3 6207264.0 11942704.0 3954714.3 7182373.9 6.0044 6.0080 6.0243 6.0653 100.3 101.0
150 % - 1 9349575.0 17734261.0 7132039.0 13047925.7 11.0080 11.0147 10.8910 10.9608 99.0 99.5
150 % - 2 9509192.5 17917261.0 7256642.8 13156930.9 11.0080 11.0147 11.0850 11.0518 100.7 100.3
150 % - 3 9404765.5 17781983.0 7163887.1 13046317.8 11.0080 11.0147 10.9428 10.9595 99.4 99.5
Average 100.5 101.1
SD 1.1 1.2
RSD 1.1 1.2
Minimum 99.0 99.5
Maximum 101.9 102.9
4. Precision
• The precision of an analytical procedure expresses the
closeness of agreement (degree of scatter) between a series
of measurements obtained from multiple sampling of the
same homogeneous sample under the prescribed conditions.
• Determination
– Assay individual samples of a homogeneous preparation
– Calculate Standard Deviation or Relative Standard
Deviation
• Precision is expressed as RSD
53
Precision Cont…
Level of precision
A. Repeatability
– Agreement within a short period of the same analyst and
instrumentation
– System and method precision
– n=6, 100% level
B. Intermediate precision
– Agreement in results intra-laboratory but from different
days, analysts and equipment (as appropriate)
C. Reproducibility
– Agreement in results between laboratories (as in a
collaboration study)…..(If C is done, B is not required) 54
Precision Cont…
>10.0% ≤ 2%
1.0 up to 10.0% ≤ 5%
0.1 up to 1.0% ≤ 10%
< 0.1% ≤ 20%
55
:
Table: System precision data of performed at the analytical
concentration (n=6) of the standard solutions
Intermediate Precision
• FDA recommends a minimum of 2 different days with
different analysts
• Acceptance criteria
– Perform F-test
– results between analysts should not be statistically
different
– Typically about 2x repeatability (2 x RSD), or pooled RSD
and Absolute mean difference
58
Table Intermediate Precision data of method validation performed by multiple analyses
(n = 6) of the same homogenous sample by two analysts on two different days, using
independently prepared reagents and sample preparations, and on two different
equipment setups
Reproducibility
60
5. Range
61
Range Cont…
62
6. Limit of Detection / Quantitation
LOD
LOQ
• Instrumental/Chromatographic
• Non-instrumental Methods
65
Detection Limit/ Quantitation Limit
66
7. System Suitability
77
• Two approaches
B. Multi-variable
78
Peak Column Temperature in Mobile Phase flow rate Detector Wavelength in Injection
Characteri 0C in mL per minute (nm) Volume in µL
stics (25) (1.0) (220) (20)
Peak pH of the mobile Ion pair con. of the Organic con. of the Triethylamine con.
Characte phase Mobile Phase Mobile Phase of the Mobile Phase
ristics (7.0) (30) (51) (5)
6.8 7.2 27 mM 33 mM 50 % 52 % 4 mL 6 mL
82