Seattle Pap Bubble CPAP
Seattle Pap Bubble CPAP
Oxymixer
Main flow outlet 0 – 16 l/min
Secondary flow outlet with bleed 0 – 3 l/min
Bleed flow ≤ 3 l/min at 3.4 bar (49.3 psi)
Maximum combined flow (all outlets) ≥ 30 l/min
Emergency flow (in the event of the air or oxygen supply failing) >15 l/min
Alarm signal in the event of loss of supply pressure Alarm on when the pressure difference of both gases is
0.9 – 1.8 bar (13 – 26.1 psi).
Alarm volume ≥80 dB ata distance of 30 cm
Adjustment range of oxygen concentration 21 – 100%
Gas supply pressure 3.2 – 6.5 bar (46.4 – 94.3 psi) air and oxygen with max. 0.7 bar
(10.1 psi) pressure difference between the two
Accuracy of the gas mixture (FIO2) ± 3 % oxygen
Connection types DISS outlets for gas mixture and NIST inlets for air and oxygen
Dimensions LxWxH 13 x 16.5 x 12.2 cm (5.1 x 6.5 x 4.8 inch)
Weight approx 1,600 g
Operating temperature +5 °C to +50 °C (41 to 122 °F)
Seattle PAP
Length of breathing hoses 1.8 m (70.86 in), including 40 cm
(15.74 in) hose extension
Material
Breathing hoses PP, EVA, ABS
Humidifier chamber PC, SBC, silicone, aluminum
Connector PP
Pressure relief valve PC, PE,PP, stainless steel
Bubble CPAP chamber PP, POM
O2 hose PVC (DEHP free)
Performance characteristics
Insp. resistance at 15 L/min <2 mbar* (or hPa or cmH2O)
Compliance at 60 mbar <1 mL/hPa (or mL/mbar)
Leakage at 60 mbar <50 mL/min
Electrical connection data 22 Ohm
EEP
hingga menghindari pasien lelah