Tuv Sud Ivdr Technical Documentation Submission Requirements
Tuv Sud Ivdr Technical Documentation Submission Requirements
Documentation
Submission
Requirements
Depending on the classification and Conformity Assessment procedure chosen, a Notified Body assesses the Tech-
nical Documentation and whether it complies with the requirements described in the IVDR.
This document provides an overview on how to submit a Technical Documentation, and what documentation can be
expected. It does NOT give guidance on what the contents of the Technical Documentation should look like and how
regulatory requirements are fulfilled. It is the sole responsibility of the manufacturer to comply with all requirements
given in Annex II and Annex III of the Regulation.
By following this document, you help to perform an efficient and timely assessment of the Technical Documentation.
If this document or parts thereof is not followed, it will still be possible to conduct a successful assessment of the
Technical Documentation, however, the time needed, and hence the costs associated, could increase.
We hope this document provides you with a helpful tool in submitting your Technical Documentation. Please do not
hesitate to contact your personal client manager
To allow an efficient assessment of the Technical Documentation, the following chapter should be used to point us to the
sections in the Technical Documentation, where the relevant information can be found.
Please complete each section below, by either entering the file(s) where the relevant information can be found, or by
ticking the “Not Applicable”-box (n.a.). If a section is not applicable, please provide a reference to the location in the
01
Technical Documentation, where a rationale for non-applicability can be found. Please enter the filename, and the
document title and version/date, in the respective fields. If information is to be found at multiple locations within the
Technical Documentation, please indicate what information can be found where (be precise and specific).
n.a.
02
General device description:
n the description of the principle of the assay method or the principles of operation of the instrument;
n the description of the reactive ingredients of relevant components such as antibodies, antigens, nucleic acid
primers; and where applicable:
n the description of the specimen collection and transport materials provided with the device or descriptions of
specifications recommended for use;
n for instruments of automated assays: the description of the appropriate assay characteristics or dedicated assays;
n for automated assays: a description of the appropriate instrumentation characteristics or dedicated instrumentation;
n a description of any software to be used with the device
n.a.
03
1.8 Principle of action
What is the principle of action of the device. The description should be specific enough to allow the user to understand the
functioning of the device.
n.a.
If the device is to be connected to other device(s) in order to operate as intended, a description of this combination/
configuration including proof that it conforms to the general safety and performance requirements when connected to any
such device(s) having regard to the characteristics specified by the manufacturer.
Accessories provided separately need to have their own labelling, instruction for use, packaging and certification.
n.a.
1.10 Accessories / equipment required for use but not provided with the devices
A description of all accessories / equipment that is required for use, but not provided with the device. Must also be
mentioned in the IFU.
n.a.
04
2. Information supplied by the manufacturer
(Regulation (EU) 2017/746 (IVDR), Annex II Section 2)
The label or labels on the device and on its packaging, such as single unit packaging, sales packaging, transport packag-
ing in the case of specific management conditions, in the languages accepted in the Member States where the device
is envisaged to be sold; The instructions for use in the languages accepted in the Member States where the device is
envisaged to be sold.
n.a.
A summary on the design process with reference to the applied implemented documented procedure(s) and versions date
shall be included. For CDx, the design with regards to suitability of the device in relation to the medicinal product concerned.
n.a.
Flow Chart including operation steps, time points of in-process controls (monitoring) and final controls, reference of
manufacturing procedures (ID numbers sufficient for traceability). A summary of manufacturing processes allowing an
understanding of the critical process steps and utilities and process chemical required to product the device.
In case of sub-contracted (outsourced) processes:
n For non-critical component suppliers (e.g. bulk) identification of supplier only.
n For critical component suppliers (e.g. outsourced manufacturing of sterile device) overview of manufacturing
processes and corresponding control measures (e.g. references to verification and validation activities, copy of the
certificate shall be included).
n.a.
05
4. General Safety and Performance Requirements
(Regulation (EU) 2017/746 (IVDR), Annex II Section 4)
Fulfilment of all applicable General Safety and Performance Requirements must be shown. If this is done by a GSPR
checklist, please make sure: Non-applicable General Safety and Performance Requirements shall have a justification as
to why they are not applicable to the device.
Demonstration of conformity includes a precise identity of the controlled documents offering evidence of conformity
with harmonised standards, common specification or other method employed to demonstrate conformity with the
General Safety and Performance Requirements. A cross-reference to the location of such evidence is provided.
n.a.
All risks associated with these hazards occurring during intended use and during reasonable foreseeably misuse must be
estimated and evaluated.
Risk control measures related to use error conform to safety principles, taking account of the generally acknowledged state
of the art. Risks related to the ergonomic features of the device and the environment in which the device is intended to be
used must be reduced as far as possible.
The technical knowledge, experience, education, training and use environment must be considered.
The medical and physical conditions of intended users must be considered.
n.a.
06
6. Benefit-Risk Analysis and Risk Management
(Regulation (EU) 2017/746 (IVDR), Annex II Section 6)
Note: All verification and validation documents in the sections below shall at a minimum comprise a plan, including
acceptance criteria and a rationale for sample size, and a report that analyses and summarised the results.
Including description of study design, populations studies, method / mode of specimen characterisation, statistical methods.
The results should be summarised in the analytical performance report. It shall contain a conclusion if the Safety and
Performance requirements (including Common Technical Specifications / Common Specifications) concerning the
07
sensitivity and specificity are fulfilled. It shall demonstrate the analytical performance, taking into consideration the
state of the art (e.g. CE-marked reference test applied).
n.a.
If Clinical Performance Studies are performed, a Clinical Performance Study Plan and Report shall be provided. The
report shall include, where applicable:
Diagnostic sensitivity
n Diagnostic specificity
n Positive predictive value
n Negative predictive value
n Likelihood ratio
n Expected values in normal and affected populations
n.a.
08
6.6 Performance Evaluation Report
An overall Performance Evaluation Report shall summarise all performance evaluation results and shall contain at a
minimum, where applicable:
n Justification for the approach taken to gather clinical evidence.
n Literature search methodology / protocol / report of literature review (or a reference to it).
n Technology on which the device is based, the intended purpose of the device and any claims made about the
device’s performance and/or safety.
n Scientific validity and analytical / clinical performance.
n Clinical evidence demonstrating the state-of-the-art of the device.
n New conclusions derived from Post-Market Performance Follow-Up (PMPF).
n.a.
Real-time stability, accelerated stability, on-board stability, open vial stability, shipping stability.
n.a.
09
6.9 CHEMICAL, PHYSICAL AND BIOLOGICAL PROPERTIES
(REGULATION (EU) 2017/746 (IVDR), Annex II Section 6.5)
In the case of devices placed on the market in a sterile condition, a description of the methods used, including the
validation reports, with respect to packaging, sterilisation and maintenance of sterility. The validation report shall
address bioburden testing, and, if applicable, testing for sterilant residues.
n.a.
10
6.9.5 Constructional Safety
6.9.5.5 Packaging
A description of the packaging including a documentation the demonstrates that packaging will not adversely affect
the device characteristics and performances during the shelf life of the device. Validation of packaging with regards to
integrity, cleanliness, and sterilisation, where applicable.
n.a.
11
6.9.5.6 Devices with connection to other device(s)
Information how to obtain a validated and safe combination (including key performance characteristics).
A demonstration that the whole combination (including the connection system) is safe and does not impair the specified
performances. Information given on known restrictions to combinations. If device needs to be connected to other
equipment, description of combination given including proof that it conforms to the general safety and performance
requirements.
n.a.
12
IV. POST-MARKET SURVEILLANCE
(REGULATION (EU) 2017/746 (IVDR), Annex III)
13
1.1 Periodic Safety Update Report (PSUR) (only class C and D devices)
The results and conclusions of the analyses of the gathered post-market surveillance data according to Annex III
together with a rationale and description of any preventive and corrective actions taken shall be gathered in a PSUR,
all PSURs are part of the Technical Documentation.
Throughout the lifetime of the device concerned this report shall set out:
n the conclusion of the benefit risk determination;
n the main findings of the Post Market Performance Follow-up Report and
n the volume of sales of devices and an estimate of the population that use the device
n and, where practicable, the usage frequency of the device.
Manufacturers of class C and D devices shall update the report at least annually and it shall be part of the technical
documentation as specified in Annexes II and III.
n.a.
It should include. at least items according to Annex XIII part B 5.2. Conclusions of the PMPF evaluation reports shall
be taken into account for the performance evaluation and in the risk management. Performance evaluation report
shall be updated as per the PMPF plan.
n.a.
14
Contact us to discuss your needs with
the fastest, most reliable global partner
www.tuvsud.com/mhs
[email protected]