±12V/0-24mAVoltage and current

source signal generator Manual


1.Technical Index
1.1 External power supply DC15-30V power 1w (not charging)/4W (charging);

1.2 3.7V lithium battery capacity 1100mAh, maximum charging current 0.6A

(optional);

1.3 External microUSB-5V current 0.2A (not charging)/0.8A (charging);

1.4 The voltage output is adjustable from ±12V, the accuracy after calibration

is 0.05V, and the maximum current is 20mA;

1.5 The current output is adjustable from O-24mA, the accuracy after

calibration is 0.05mA,The current sampling resistance is not more than 500

ohms;

1.6 The digital tube displays 4 digits, two decimal places;

1.7 Encoder knob 1 circle: 30 pulses;

2.Power Supply

3 power supply methods, as long as one is used, it can work, and it can be

connected at the same time;


Battery indicator status:
Battery green yellow Red Flashing Green
indicator does not
flash
Battery Electricity≥80% 80%≥Electricity≥40% 40%≥Electricity recharging。。。 Brimming
status
Complementary:Hold down“MODE”key 1 second,The digital tube flashes for 5 seconds to display
the actual battery voltage。

3. Dimensions
4. Wring Diagram
How to judge whether it is active or passive connection? Common methods are

as follows:

1 :See the internal schematic document of the device, I+ is 24V+,it is the

passive connection; I- is 24VG,it is the active connection;

2 :Use a multimeter to measure the voltage between I+ and I-, =0V is the

active connection, =24V is the active connection;


5. Voltage/Current switching

Press the MODE button, one key switch to adjust the voltage and current

mode;The indicator shows what state it is in,System operation corresponds to

which mode to operate;The original mode after switching,Will continue to

output,It’s just not in a state of adjustment;

6. System Opetion(Where“OK<confirm key>”is:Push the knob


diwn,Turn clockwise to “+<ADD>”,Anticlockwise
rotation”-<reduce>”):

6.1 In the normal operation screen, short press the confirm key to

save the output value ,After letting go,Screen

display"...",Successfully saved,After the next boot,The output

value is this saved value;When debugging equipment, We need to


randomly adjust the output value,Just don’t press the encoder knob,

After restarting, the original value will be saved;

6.2 Parameter settings:

6.2.1 In the normal operation screen,Press and hold the confirm

button for 2 seconds,Enter setting state ,Display”F001”(Parameter

number 001);

6.2.2 Turn the knob to change the parameter number (First entry F002

Need to enter the password first“+ - - + (Rotate the knob)”);

6.2.3 Parameter number interface,Press the confirm key to enter the

corresponding parameter value setting,Rotate the knob to modify the

parameters;

6.2.4 Parameter number interface,Press the confirm key to enter the

corresponding parameter value setting,Rotate the knob to modify the

parameters;

6.2.5 Rotate the knob directly to the last parameter number, press

the OK key, the setting is completed and enter the normal operation

screen;

6.2.6 If there is no operation on the parameter setting screen for

more than 10s, it will exit the setting state and enter the normal

operation screen;

6.3 For example:

6.3.1 Modify "F001" coarse and fine adjustment


adjustment;

Figure 1→2: Press and hold the knob for 2 seconds,Enter setting

state,Display “F001”;

Figure 2→3: Press "OK",Set the value of "F001"设,Turn the knob

to modify it;

Figure 3→4: Press "OK",Exit the setting state after saving,Enter

the normal;

6.3.2 Modify "F006" calibration value


Figure 1→2: Press and hold the knob for 2 seconds,Enter setting state,

Display“F001”

Figure 2→3: “+”1 Pulse,Display“- - - -”;

Figure 3→4: Enter password“+ - - +”,Display“= = = =”;

Figure 4→5: Press "OK", if the password is entered correctly, it will display

"F002"Or "F100", otherwise it will display "Err" to exit, there may be an error,

pleaseenter it again;

Figure 5→6: Rotate the knob "+" to display to“F006”;

Figure 6→7: Press "OK",Enter to set the parameter value of "F006",

Then modify this value to make the actual output equal to -10V or 4mA

Figure 7→ : Press "OK" to save, or automatically exit the setting state after 10s
of inactivity, and the parameters will not be saved;

6.4 In the voltage regulation mode, enter the voltage setting, parameter

description table:

Serial number Description Remark Default


0:Coarse adjustment,1:fine
Coarse adjustment or fine
F001 adjustment,2:Shortcut mode(Need to 0
adjustment
be set firstF100>0)
0:±12V,1:±5V,2:0-10V,3:2-10V,4:0-5V,
F002 Output mode 5:1-5V,6:0-3.3V,7:0-2.5V,8:0-1V, 0
9:-10-0V
0:Actual voltage,
F003 Display mode 1:percentage0-100.0, 2:50HZ, 0
3:0-1500
Coarse adjustment mode 1-50 plus and minus figures for each
F004 plus and minus value/each pulse, no decimal point concept (1-50) x 1
pulse 10
fine adjustment mode plus 1-50 plus and minus figures for each
F005 and minus value/each pulse, no decimal point concept (1-50) x 1
pulse 10
-999 -- +999 Internal reference only,
Output calibration value
F006
-10V Please be cautious when modifying
-999 -- +999 Internal reference
Output calibration value
F007 only,Please be cautious when
0V
modifying
-999 -- +999 Internal reference
Output calibration value
F008 only,Please be cautious when
10V
modifying
1-8: Level 1 is the darkest, level 8
F009 Digital tube brightness 1
is the brightest
0:Quick output mode is not used,
F100 Quick output points 0
2-9:Points
Range -12V to +12V How many points
F101...F109 9-point output value
are there how many values can be set
6.5 In the regulation current mode, Enter the current setting,parameter

description table:
Serial
Description Remark Default
number
0:Coarse adjustment,1:fine
Coarse adjustment or
F001 adjustment,2:Shortcut mode(Need 0
fine adjustment
to be set firstF100>0)
F002 Output mode 0:0-20mA, 1:4-20mA ,2:0-24mA 0
0: Actual current, 1: percentage
F003 Display mode 0
0-100.0, 2: 50HZ

Coarse adjustment mode 1-50 plus and minus figures for each
F004 plus and minus pulse, no decimal point concept 1
value/each pulse (1-50) x 10

fine adjustment mode 1-50 plus and minus figures for each
F005 plus and minus pulse, no decimal point concept 1
value/each pulse (1-50) x 10

-999 -- +999 Internal reference


Output calibration
F006 only,Please be cautious when
value 4mA
modifying
-999 -- +999 Internal reference
Output calibration
F007 only,Please be cautious when
value 12mA
modifying
-999 -- +999 Internal reference
Output calibration
F008 only,Please be cautious when
value 20mA
modifying
Digital tube 1-8: Level 1 is the darkest,
F009 1
brightness level 8 is the brightest
0:Quick output mode is not used,
F100 Quick output points 0
2-9:Points

Range 0-24mA How many points are


F101...F109 9-point output value
there how many values can be set

7 Current output test method: AIo GND


We often encounter people who say that there is no current when measuring

current. In fact, the ammeter is broken, the gear is wrong or the wrong hole is

inserted. There is a fuse inside the multimeter to measure the current. This is

easy to blow, so we use the method of voltage measurement. Verify the

transition of the signal generator as follows:

7.1 The simplest method: power on, directly measure the voltage between

AIo and GND when there is no load, =0V is a defective, otherwise 99% may be

no problem, because when there is voltage to measure current, it is equivalent

to short circuit and there will definitely be current ;

7.2 Connect an LED indicator, if the brightness changes with the adjustment,

then the signal generator current output is no problem; 7.3 Connect a 50-300
ohm resistor to AIo GND. If there is voltage and it meets "Voltage = Current X

Resistance", the voltage will change with the adjustment, then the signal

generator current output is no problem;

7.4 If there is a PLC/inverter, measure the resistance between I+ and I-. If it is

between 50-300 ohms, you can use it as a test resistance. Connect AIo to I+,

GND to I-, and measure the voltage. If there is voltage, and Satisfy "Voltage =

Current X Resistance", the voltage will change with the adjustment, then the

signal generator current output is no problem;

8 NOTE:
8.1 Please read this manual carefully before use;

8.2 Turn off the power before wiring

8.3 Exceeding the range shown in the technical indicators may cause the

instrument to work abnormally or even be damaged;