Manuale Incubatore CO2 C 170 Geass
Manuale Incubatore CO2 C 170 Geass
C 170 (E3)
CO2 Incubator
with FPI-sensor system
and display controller RP1
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: [email protected] Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: [email protected]
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16
1. SAFETY .................................................................................................................. 5
1.1 Legal considerations ........................................................................................................................... 5
1.2 Structure of the safety instructions ...................................................................................................... 5
1.2.1 Signal word panel ..................................................................................................................... 5
1.2.2 Safety alert symbol ................................................................................................................... 6
1.2.3 Pictograms ................................................................................................................................ 6
1.2.4 Word message panel structure ................................................................................................. 7
1.3 Localization / position of safety labels at the chamber ....................................................................... 7
1.4 Type plate ........................................................................................................................................... 8
1.5 General safety instructions on installing and operating the chamber ................................................. 9
1.6 Precautions when working with CO2 gas .......................................................................................... 11
1.7 Precautions when handling gas cylinders ......................................................................................... 11
1.8 Intended use ..................................................................................................................................... 12
2. CHAMBER DESCRIPTION .................................................................................. 13
2.1 The CO2 measuring principle ............................................................................................................ 13
2.2 Chamber overview ............................................................................................................................ 14
2.3 Connection panel at the rear of the chamber.................................................................................... 16
2.4 Chamber doors ................................................................................................................................. 16
3. COMPLETENESS OF DELIVERY, TRANSPORTATION, STORAGE, AND
INSTALLATION.................................................................................................... 17
3.1 Unpacking, and checking equipment and completeness of delivery ................................................ 17
3.2 Guidelines for safe lifting and transportation..................................................................................... 18
3.3 Storage .............................................................................................................................................. 18
3.4 Location of installation and ambient conditions ................................................................................ 18
4. INSTALLATION AND CONNECTIONS ............................................................... 21
4.1 Shelf holder and shelves ................................................................................................................... 21
4.2 CO2 sensor ........................................................................................................................................ 22
4.2.1 Connecting the CO2 sensor .................................................................................................... 22
4.2.2 General notes ......................................................................................................................... 22
4.3 Water pan .......................................................................................................................................... 23
4.4 Gas connection ................................................................................................................................. 24
4.4.1 Connecting the CO2 gas cylinder ............................................................................................ 25
4.4.2 Connecting the gas hose to the chamber rear ....................................................................... 26
4.4.3 Gas cylinder connection kit (option)........................................................................................ 27
4.5 Electrical connection ......................................................................................................................... 28
5. START UP ............................................................................................................ 29
6. FUNCTION OVERVIEW OF THE RP1 CONTROLLER ....................................... 29
6.1 Preset factory parameters ................................................................................................................. 30
6.2 Performance after turning on the chamber ....................................................................................... 30
7. CONTROLLER RP1 OPERATING MODES ......................................................... 31
7.1 Selecting and setting the operating functions ................................................................................... 32
7.2 Operating mode HAND ..................................................................................................................... 32
7.3 Operating mode USER: Advanced settings ...................................................................................... 34
7.4 Operating mode LOCK: Locking/unlocking of the operating functions’ settings by operating mode
HAND ................................................................................................................................................ 38
7.5 Performance during and after power failures and shut down ........................................................... 39
1. Safety
This operating manual is part of the components of delivery. Always keep it handy for reference. The
device should only be operated by laboratory personnel especially trained for this purpose and familiar
with all precautionary measures required for working in a laboratory. Observe the national regulations on
minimum age of laboratory personnel. To avoid injury and damage observe the safety instructions in the
operating manual.
WARNING
Failure to observe the safety instructions.
Serious injuries and chamber damage.
Observe the safety instructions in this operating manual
Carefully read the complete operating instructions for the chamber.
Depending on the seriousness and probability of serious consequences, potential dangers are identified
with a signal word, the corresponding safety color, and if appropriate, the safety alert symbol.
DANGER
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious
(irreversible) injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor
(reversible) injury
CAUTION
Indicates a potentially hazardous situation, which, if not avoided, may result in damage to the product
and/or its functions or to property in its proximity.
1.2.3 Pictograms
Warning signs
Mandatory regulation Read operating Disconnect the power Lift with several persons
instructions plug
Observe all other notes and information not necessarily emphasized in the same way, in order to avoid
disruptions that could result in direct or indirect injury or property damage.
Hot surface
Replace safety labels that are no longer legible. Contact BINDER Service for these replacements.
CAUTION
Danger of overheating.
Damage to the chamber.
∅ Do NOT install the chamber in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
DANGER
Explosion hazard.
Danger of death.
∅ Do NOT operate the chamber in potentially explosive areas.
∅ KEEP explosive dust or air-solvent mixtures AWAY from the chamber.
DANGER
Explosion hazard.
Danger of death.
∅ Do NOT introduce any substance into the CO2 incubator which is combustible or explo-
sive at working temperature.
∅ NO explosive dust or air-solvent mixture in the inner chamber.
Any solvent contained in the charging material must not be explosive or inflammable. I.e., irrespective of
the solvent concentration in the steam room, NO explosive mixture with air must form. The temperature
inside the chamber must lie below the flash point or below the sublimation point of the charging material.
Familiarize yourself with the physical and chemical properties of the charging material, as well as the
contained moisture constituent and its behavior with the addition of heat energy and humidity.
Familiarize yourself with any potential health risks caused by the charging material, the contained mois-
ture constituent or by reaction products that may arise during the temperature process. Take adequate
measures to exclude such risks prior to putting the CO2 incubator into operation.
DANGER
Electrical hazard.
Danger of death.
∅ The chamber must NOT become wet during operation or maintenance.
The chambers were produced in accordance with VDE regulations and were routinely tested in accord-
ance to VDE 0411-1 (IEC 61010-1).
During and shortly after operation, the temperature of the inner surfaces almost equals the set-point.
CAUTION
The glass door, the glass door handle and the inner chamber will become hot during
operation.
Danger of burning.
∅ Do NOT touch the glass door, the glass door handle, the inner surfaces or the charging
material during operation.
WARNING
Stability hazard.
Danger of injury.
Damage to the chamber and the charging material.
Housing cover breakaway.
∅ Do NOT climb on the lower housing cover.
∅ Do NOT load the lower housing cover with heavy objects while the chamber door is
open.
WARNING
High concentration of CO2 (> 4 Vol.-%).
Risk of death by suffocation.
Danger of poisoning.
∅ Do NOT set up chambers in non-ventilated recesses
Ensure technical ventilation measures
Observe the relevant regulations for handling CO2.
Secure the gas cylinders against falling and other mechanical damage.
WARNING
Safety valve tearing off.
Sudden release of the stored pressure energy.
Risk of injury.
Secure gas cylinders against falling (chaining).
Transport gas cylinders with a cylinder cart.
The valve of the gas cylinder always must be closed before screwing on or unscrewing the gas hose.
WARNING
Opening the cylinder valve when the cylinder is not connected.
Sudden release of the stored pressure energy.
Risk of injury.
Close the gas cylinder valve before connecting or removing the gas hose.
After connecting the gas cylinder, check all gas connections for leaks (e.g. with leak spray or
diluted soap solution).
Observing the instructions in this operating manual and conducting regular maintenance
work (chap. 16) is part of the intended use.
Due to the special demands of the Medical Device Directive (MDD), these ovens are not
qualified for sterilization of medical devices as defined by the directive 93/42/EWG.
Always use proper protective equipment (clothing, gloves, safety glasses, etc.).
Always follow good hygienic practices according to GLP/SOP protocols.
Each individual operating the chamber is responsible for his or her own safety
Any component of the charging material must NOT be able to release toxic gases.
DANGER
Explosion or implosion hazard.
Danger of poisoning.
Danger of death.
∅ Do NOT introduce any substance combustible or explosive at working temperature into
the chamber, in particular no energy sources such as batteries or lithium-ion batteries.
∅ NO explosive dust or air-solvent mixture in the inner chamber.
∅ Do NOT introduce any substance which could lead to release of toxic gases.
In case of foreseeable use of the device there is no risk for the user through the integration of the cham-
ber into systems or by special environmental or operating conditions in the sense of EN 61010-1:2010.
For this, the intended use of the chamber and all its connections must be observed.
(1)
(2)
(3)
(1) Triangle instrument panel with RP1 controller for temperature and CO2 and
(2) Door handle
(3) Main power switch
Delay time for the temperature and CO2 tolerance range alarm :
After closing the outer door, the tolerance range alarm is turned off for a programmable delay
time. This prevents alarms being constantly triggered during the unstable operating phase
after opening the outer door.
Remove any protective lamination sheet on the inner metal surfaces prior to commissioning.
CAUTION
Sliding or tilting of the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
∅ Do NOT lift or transport the chamber using the door handle or the door.
Lift the chamber from the pallet at the four lower corners with the aid of four people.
If you need to return the chamber, please use the original packing and observe the guidelines for safe
lifting and transportation (chap. 3.2).
For disposal of the transport packing, see chap. 17.1.
Note on second-hand chambers (Ex-Demo-Units):
Second-hand chambers are chambers that were used for a short time for tests or exhibitions. They are
thoroughly tested before resale. BINDER ensures that the chamber is technically sound and will work
flawlessly.
Second-hand chambers are marked with a sticker on the chamber door. Please remove the sticker before
commissioning the chamber.
CAUTION
Sliding or tilting the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
Transport the chamber in its original packaging only.
Secure the CO2 incubator with transport straps for transport.
∅ Do NOT lift or transport the chamber using the door handle or the door.
Lift the chamber at the four lower corners with the aid of 4 people and place it on a
rolling pallet.
Move the chamber to the desired location and lift it from the rolling pallet with the aid of
four people.
• Permissible ambient temperature range for transport: -10 °C / 14°F to +60 °C / 140°F.
You can order transport packing and rolling pallets for transportation purposes from BINDER Service.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 17.2).
• Permissible ambient temperature range for storage: -10 °C / 14°F to +60 °C / 140°F.
• Permissible ambient humidity: max. 70% r.H., non-condensing
When after storage in a cold location you transfer the chamber to its warmer installation site, condensa-
tion may form in the inner chamber, on the housing or in the sensor compartment of the CO2 measure-
ment. Before start-up, wait at least one hour until the CO2 incubator has attained ambient temperature
and is completely dry.
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user there is no risk of temporary overvoltages in the sense of EN 61010-1:2010.
Do not install or operate the chamber in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
∅ Do NOT operate the chamber in potentially explosive areas.
∅ KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
Ambient conditions
• Permissible ambient temperature range for operation: +18 °C / 64.4 °F up to +30 °C / 86 °F. At elevat-
ed ambient temperature values, fluctuations in temperature can occur.
• Ideal ambient temperature: by at least 7 °C / 12.6 °F below the intended working temperature. E.g.,
working temperature 37 °C / 98.6 °F – resulting permitted ambient temperature 30 °C / 86°F and lower
In the event of working temperatures of less than 7 °C / 12.6 °F above the ambient temperature, the set
point can be exceeded.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +22 °C +/- 3 °C / 71.6 °F ± 5.4 °F to which the specified technical data relate. For
other ambient conditions, deviations from the indicated data are possible.
WARNING
High concentration of CO2 (> 4 Vol.-%).
Danger of death by suffocation.
Danger of poisoning.
∅ Do NOT set up chambers in non-ventilated recesses.
Ensure technical ventilation measures.
Observe the relevant regulations for handling CO2.
Observe the occupational exposure limit OEL for CO2 set by the national authorities (formerly maxi-
mum permitted workplace concentration). Check compliance when operating all chambers located in the
room.
• OEL for Germany: 5000 ml/m3 (ppm) = 0,5 Vol.-%
• CO2 lost with each door opening: about 16.4 g, i.e. 0.0084 cubic meters / 0.296 cubic feet (under nor-
mal pressure)
• CO2 lost during 12h at 5 vol.-% without door opening: approx. < 2 g, i.e. 0.001 cubic meters / 0.035
cubic feet (under normal pressure 1013 mbar / 14.7 psi)
An example of how to evaluate laboratory volume and air change rate:
Question: Is an air change rate of 1/h sufficient for a lab with a volume of 100 cubic meters / 3,531.5
cubic feet with 10 CO2 incubators, opened 4 times per hour?
Calculation: CO2 concentration = CO2 lost by door opening, multiplied by 10 chambers, multiplied by 4
door openings per hour, divided by lab volume
0.0084 cubic meters x 10 x 4 div. 100 cubic meters = 0.00336, i.e. 0.336% or 3360 ppm.
0.296 cubic feet x 10 x 4 div. 3,531.5 cubic feet = 0.00336, i.e. 0.336% or 3360 ppm.
Result: The maximum permissible value of 5000 ppm is not exceeded under these operation conditions.
Even when CO2 or systems operated with CO2 are handled carefully and appropriately, a residual risk
remains, which can lead to life-threatening situations under certain circumstances. Therefore we strongly
recommend continuous monitoring of CO2 concentration in the ambient air of the CO2 incubator. It must
be ensured permanently that the maximum permissible occupational exposure limit OEL for CO2 (0.5 vol -
% CO2 for Germany) is not exceeded.
If just one of the three supplied shelves shall be used in one of the three lower positions, we
recommend inserting a supplementary shelf as high as possible into the shelf holder for stabi-
lizing purpose.
Turn off the chamber. Open the door of the inner chamber and plug the CO2 sensor (4) into the perma-
nently installed holding tube located in the upper part of the rear of the inner chamber.
Figure 10: Plugged-in CO2 sensor (right) and gas mixing head for CO2
The sensor must click in correctly and sit tightly in the connection socket.
CAUTION
Connecting or removing the CO2 sensor during operation.
Damage to the CO2 sensor.
Connect or remove the CO2 sensor only with the chamber turned off.
Connect or remove the CO2 sensor without rotating and only when the incubator is turned off. Remove
the CO2 sensor before removing or replacing its filter cap. The PTFE filter of the CO2 sensor prevents dirt
and humidity from intruding into the measuring cell. It is available as a spare part. Replace it whenever it
is damaged or soiled.
The accuracy of the indicated values of CO2 depends on the ambient air pressure (approx.
0.08 vol.-% per 10 mbar / 0.15 psi). In order to compensate this effect when measuring the
CO2 concentration, the altitude of the installation site above sea level can be entered into the
controller (chap. 8.1).
CAUTION
Excess temperature.
Damage to the CO2 sensor.
∅ Do NOT autoclave the CO2 Sensor.
∅ Do NOT expose the CO2 sensor to hot-air sterilization.
CAUTION
Different CO2 sensor.
Invalid calibration.
∅ Do NOT change the CO2 sensor head.
Note down the serial number of the CO2 sensor.
CAUTION
Shocks of the CO2 sensor.
Damage to the CO2 sensor.
∅ Avoid strong shocks of the CO2 sensor (by putting it down hard, or dropping).
Figure 11: Position of the water pan centrally between the front cams (arrows)
Secure the gas cylinders against falling and other mechanical damage.
WARNING
Safety valve tearing off.
Sudden release of the stored pressure energy.
Risk of injury.
Secure gas cylinders against falling (chaining).
Transport gas cylinders with a cylinder cart.
The valve of the gas cylinder always must be closed before screwing on or unscrewing the gas hose.
WARNING
Opening the cylinder valve when the cylinder is not connected.
Sudden release of the stored pressure energy.
Risk of injury.
Close the gas cylinder valve before connecting or removing the gas hose.
After connecting the gas cylinder, check all gas connections for leaks (e.g. with leak spray or
diluted soap solution).
Carbon dioxide (CO2) in high concentrations is hazardous to health. It is colorless and almost odorless
and therefore practically imperceptible. Vent out any CO2 gas that may escape via good room ventilation
or a suitable connection to an exhaust system. We recommend installing a CO2 warning system.
WARNING
High concentration of CO2 (> 4 Vol.-%).
Danger of death by suffocation.
Danger of poisoning.
∅ Do NOT set up chambers in non-ventilated recesses.
Ensure technical ventilation measures.
Observe the relevant regulations for handling CO2.
The CO2 gas necessary for operation must have a technical grade of 99.5 %.
The gas connections must be established by qualified personnel who are trained in handling
the respective gases and familiar with the required safety measures.
A gas supply pressure above 2.5 bar / 36 psi will result in chamber damage.
Use a pressure reducer and make sure to avoid any excessive outlet pressure when connecting the
gas hose to the incubator.
The real outlet pressure of gas cylinders, sets of gas cylinders or central gas supplies am on the sec-
ond manometer must not exceed 2.5 bar / 36 psi.
CAUTION
Excessive outlet pressure > 2.5 bar / 36 psi.
Damage to the chamber.
∅ The outlet pressure must NOT exceed the indicated value of 2.5 bar / 36 psi.
Before connecting, check the outlet pressure on the pressure reducer of the cylin-
der.
Adjust the outlet pressure to 2.0 bar / 29 psi above the ambient pressure.
Observe the correct outlet pressure also when replacing the gas cylinders.
Establishing the connection to the incubator
Connect the supplied gas hose (internal diameter 6 mm / 0.24 inches) to the pressure reducer of the
gas cylinders or central gas supply and secure the connection with the supplied hose clamp.
Connect the pre-assembled hose nozzle of the gas hose to the quick acting closure socket (14) DN 6
on the chamber rear, as described in chap. 4.4.2.
Leak test
After connecting the gas cylinder, check all gas connections for leaks (e.g. with leak spray or diluted soap
solution).
Conversion table for gas inlet pressures, bar – psi, see chap. 19.6.
The gas hose, which will be used to establish the connection to a gas cylinder, is already attached to the
hose nozzle and secured by a hose clamp. Plug the hose nozzle into the corresponding quick acting clo-
sure socket (a) located at the rear of the chamber. This quick acting closure socket is closed by a rubber
cover (b).
Only use the supplied hose nozzle to connect to the quick acting closure socket.
Otherwise, the quick acting closure socket may leak, and/or it may become impossible to con-
nect the original hose nozzle. In this case, please contact BINDER Service.
(a)
(b)
(c)
Now fit the hose nozzle (c) in the quick acting closure socket. To remove the connection, pull the hose
nozzle off the quick acting closure socket.
The gas cylinder connection kit for CO2 (Art. no. 8012-0014) includes the
following parts for connecting a gas cylinder to the CO2 incubator:
• Pressure reducer with manometers for cylinder pressure (high pressure
gauge) and outlet pressure (low pressure gauge)
• 5 m pressure hose with pre-assembled hose nozzle for quick acting clo-
sure socket
• 1 hose clamp to connect the gas hose to the pressure reducer
The pressure reducer is also available as an individual accessory.
Instructions 7001-0249 included with the connection kit describe connecting and setting
the gas cylinder connection kit.
CAUTION
Excessive outlet pressure > 2.5 bar / 36 psi.
Damage to the chamber.
∅ The outlet pressure must NOT exceed the indicated value of 2.5 bar / 36 psi.
Before connecting, check the outlet pressure on the pressure reducer of the cylinder.
Adjust the outlet pressure to 2.0 bar / 29 psi above the ambient pressure.
The gas connections must be established by qualified personnel who are trained in handling
the respective gases and familiar with the required safety measures.
• The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the lat-
est technology. The protective conductors of the socket and plug must be compatible!
• Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (centrally located at the bottom of the left-hand side of the
chamber, see chap. 1.4)
• When connecting, please observe the regulations specified by the local electricity supply company
and as well as the VDE directives (for Germany). We recommend the use of a residual current circuit
breaker.
• Pollution degree (acc. to IEC 61010-1): 2
• Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Controller buttons
Button Function
Reduce value
Increase value
CAUTION
Danger of samples being destroyed.
Charge the chamber only after equilibration of temperature and CO2.
Check that the interior of the chamber is empty, including the trays and the water pan. If you do not know
what the chamber was last used for, for hygiene purposes you should clean and disinfect or sterilize the
interior (chap. 15).
Set the controller to the desired temperature and CO2 set points that are to be used to operate the cham-
ber.
The yellow LED indicates that the heating is active, and the yellow LED indicates that the CO2
valve is open.
Only insert samples into the chamber when it has reached its stable operating state:
Temperature: Equilibration time is approx. 2 hours.
CO2: After approx. 5 minutes, the CO2 concentration equilibrates automatically to the preset value.
If there is no accordance between the actual and set value shown in the display, proper opera-
tion of the chamber is not guaranteed.
Instructions when observing leakage of condensate from the CO2 sensor system and CO2 fluctua-
tion:
In the case of CO2 fluctuations and concomitant condensate leakage from the injection and suction noz-
zle of the CO2 sensor compartment after start-up or when taking the chamber back into service, the
chamber should be left to dry open for at least an hour running at 37 °C with the water pan empty. This
will result in flushing the sensor compartment and tubing from humid air. Removing the condensate from
the CO2 sensor system will ensure a turbulence-free CO2 measurement.
In operating modes HAND and USER, access the next operating function by pressing “MODE”.
Pressing “MODE” will confirm the changed value of the current operating function, i.e., the chamber will
from now on regulate at the new value.
If you do not want to accept the modification of a value, wait 30 seconds without pressing any
button. The controller reverts to Normal Display automatically without accepting the modifica-
tion.
Use “▲” and “▼” to change the value of the selected operating function or to select between defined
values.
To change numerical values by tenths, press “▲” and “▼” for a very short time. To change
them by whole numbers, press “▲” and “▼” for longer.
In some operating functions, you can read but not change the value (for service and maintenance pur-
poses).
To go back to the Normal display, press “MODE” repeatedly.
If a button for 30 seconds is not pressed after access to an operating function or changing a value, the
controller toggles back to Normal display automatically. If the “MODE” button is not pressed after chang-
ing a value in the displayed operating function, the value reverts to its previous value.
Safety controller set point when set point type is “Limit” (Lit)
Limit value, i.e., maximum permitted absolute temperature value. When
0.0 °C to 60.0 °C exceeded, the safety controller triggers an alarm. For technical reasons,
the limit value should exceed the temperature set point by at least 2 °C.
Typical value: 39.0 °C / 102.2°F.
Or
Safety controller set point when set point type is “Offset” (OFS)
Offset value, i.e. maximum over temperature above the selected set point
0.1 K to 10 K
When exceeded, the safety controller triggers an alarm. For technical
reasons, the offset value should be at least 2 °C (factory setting).
The operating function “At.S” in operating mode USER (chap. 7.3) specifies the safety control-
ler set point type “Limit” (Lit) or “Offset” (OFS).
• With set point type “Limit” (Lit), operating function “AL.S” (safety controller set point as limit
value) is displayed in operating modes HAND and USER. You can set a limit value.
• With set point type “Offset” (OFS), operating function “Ot.S” (safety controller set point as
offset value) is displayed in operating modes HAND and USER. You can set an offset val-
ue.
With set point type “Limit”, set the safety controller each time the set point for the temperature
is changed. Set the safety controller set point approx. 2 °C above the controller temperature
set point.
Factory setting: set point type “Offset” with safety controller set point 2 °C.
• In Normal display, press “MODE” and “EXIT” simultaneously for 3 seconds to access the options for
selecting the chamber's operating modes.
• Press “▲”until the value “OP.U” (operating mode USER) appears in the upper display.
• Press “MODE”. The display to enter the password appears.
The operating mode USER is password-protected by a number. The password is preset to "1" in factory.
You can change the password in operating function “PA.U”.
Inhibit time. This operating function is not required to operate the cham-
----- ber.
Its value should always be zero.
OFF, or -199,9 °C
Temperature adjustment serves to adjust the temperature sensor.
to +300 °C
OFF, or
-10.0 vol.-% to CO2 final value adjustment serves to adjust the CO2 sensor.
110.0 vol.-%
The operating function “At.S” specifies the safety controller set point type “Limit” (Lit) or “Off-
set” (OFS).
• With set point type “Limit” (Lit), operating function “AL.S” (safety controller set point as limit
value) is displayed in operating modes HAND and USER. You can set a limit value .
• With set point type “Offset” (OFS), operating function “Ot.S” (safety controller set point as
offset value) is displayed in operating modes HAND and USER. You can set an offset val-
ue.
To permanently unlock the settings, set operating function “PA.H” to “0” in HAND mode (chap.
7.2).
7.5 Performance during and after power failures and shut down
In the event of power failure, all controller functions are taken out of operation. The gas inlet valves are
closed so that no gas can escape into the ambient air. The zero-voltage relay alarm output (13) (chap.
10.3) is switched to alarm position indicating the alarm for the whole duration of the power failure.
After the power returns or after turning on the chamber by hand, the incubator regulates the temperature
and CO2 to the last entered set points.
If the chamber was in sterilization mode, the process is cancelled and the chamber continues normal
operating mode with the original set points previously entered .
All settings and set point values remain in the memory during power failure after power off.
Normal Display
Password request
Password entry
(factory setting: “1”).
4x
Entry of altitude
Entry of the current altitude [km] above sea level with arrow keys.
Unit of altitude above sea level for entry and displayed value: kilometer [km]
Correlation feet [ft] to kilometer [km]: see chap. 19.5.
Normal Display
When setting a lower temperature set point, in order to save time, we recommend cooling
down the chamber by turning it off and opening both doors of the chamber.
When setting a lower CO2 set point, the CO2 gas must be able to escape first. Open both
doors of the chamber for this purpose.
If the CO2 concentration is indicated as “-0.4”, the CO2 sensor is defective. Please contact
BINDER Service.
WARNING
High concentration of CO2 (> 4 Vol.-%).
Danger of death by suffocation.
Danger of poisoning.
∅ Do NOT set up chambers in non-ventilated recesses.
Ensure technical ventilation measures.
Observe the relevant regulations for handling CO2.
If CO2 is released, leave the area und inform the security service or fire department.
Regularly check the safety controller setting for set point type “Limit” or “Offset”.
Set the safety controller set point by approx. 2 °C above the desired temperature set point.
Factory setting: set point type “Offset” with safety controller set point 2 °C.
Select the safety controller set point type in operating mode USER (chap. 7.3). The operating function
“At.S” specifies the safety controller set point type “Limit” (Lit) or “Offset” (OFS).
• With set point type “Limit” (Lit), operating function “AL.S” (safety controller set point as limit value) is
displayed in operating modes HAND and USER. You can set a limit value.
• With set point type “Offset” (OFS), operating function “Ot.S” (safety controller set point as offset value)
is displayed in operating modes HAND and USER. You can set an offset value.
Normal Display
Password request
Password entry
(factory setting: “1”).
You can check and set the safety controller set point in operating modes HAND (chap. 7.2) or USER
(chap. 7.3).
• With set point type “Limit” (Lit), operating function “AL.S” (safety controller set point as limit value) is
displayed in operating modes HAND and USER. You can set a limit value.
• With set point type “Offset” (OFS), operating function “Ot.S” (safety controller set point as offset value)
is displayed in operating modes HAND and USER. You can set an offset value.
The settings of the safety controller are inactive during sterilization (chap. 15.4). They become
functional again following abortion of the sterilization and/or the restart of the chamber at the
main power switch.
Normal Display
Password request
Password entry
(factory setting: “1”).
10 x
Depending on the selected set point type (chap. 9.2.1) a different display appears:
Entry of limit value or offset value
With set point type Limit: With set point type Offset:
Enter a limit value Enter an offset value
or
Entry of limit value using the arrow keys. Entry of offset value using the arrow keys.
Display Event
Door open
If more than one alarm signal is sent simultaneously, they are displayed in a cycle, except for alarms 995
to 999. These take priority over all other operational displays and alarm signals of the controller.
Except for the tolerance range alarms, all alarms are displayed immediately when the fault occurs. The
tolerance range alarms are suppressed for a selected time (delay times set for temperature alarm and for
CO2 alarm) after opening the chamber door or turning on the chamber.
• Remove the cause of the alarm or wait until the chamber compensates for the reason of the error.
• The visual alarm disappears, when the cause of the fault has been remedied or the monitored operat-
ing function returns within its tolerance limits.
• Press the “EXIT” button to turn off the audible alarm signals.
Pin 1: Pole
ALARM Pin 2: Break relay
Pin 3: Make contact
DANGER
Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.
∅ Do NOT exceed the maximum switching load of 24V AC/DC – 2.5 Amp.
∅ Do NOT connect any devices with a higher loading capacity.
The alarm message on the controller display remains displayed during transmission of an alarm via the
zero-voltage relay outputs.
As soon as the cause of the alarm is identified and resolved, the alarm transmission via the zero-voltage
relay outputs resets automatically together with the alarm message on the controller display.
In case of a power failure, transmission of the alarm via zero-voltage relay outputs remains active for the
duration of the power failure. Afterwards, contact 1 will close automatically with contact 3.
Connection to an external monitoring system
To ensure short-circuit-proof alarm monitoring that will trigger the alarm when the chamber is connected
to an external alarm monitor, connect the external alarm monitoring system to the chamber via the con-
tacts 1 and 3.
No alarm signal is emitted during the temperature delay time after the outer door is closed
• Visual display
LED Upper display, alternating
Temperature Alarm code
Flashes
Flashes
flashes
• Audible alarm: buzzer (intermittent sound, beginning 5 minutes after the door is opened)
• Switching the zero-voltage relay alarm output (beginning 5 minutes after the door is opened)
Actions:
• Close the outer door.
• Use the “EXIT” button to turn off the buzzer even when the door is open.
• The alarm message is cancelled.
• The zero-voltage relay alarm output switches off.
CAUTION
Excessive outlet pressure > 2.5 bar / 36 psi.
Damage to the chamber.
∅ The outlet pressure must NOT exceed the indicated value of 2.5 bar / 36 psi.
Before connecting, check the outlet pressure on the pressure reducer of the cylinder.
Adjust the cylinder outlet pressure to 2.0 bar / 29 psi above the ambient pressure.
The recovery times of the gas concentrations inside the chamber following the door being opened, which
are indicated in the technical data (chap. 19.4), refer to a connection pressure of 2.0 bar / 29 psi. De-
creasing supply pressure down to the alarm point of 0.3 bar / 4.4 psi results in longer recovery times.
Check the pressure displays of your gas supply. If very short recovery times are required or the door is
opened frequently, replace the gas cylinders promptly when the pressure decreases below 2.0 bar / 29
psi.
Actions:
• Turn off the chamber.
• If necessary, clean, disinfect and sterilize the chamber.
• Please contact BINDER Service.
Flashes
Flashes
Flashes
Flashes
Actions:
• Turn off the chamber with the main power switch.
• Disconnect the power plug.
• Wait for 1 minute.
• Replace the power plug.
• Turn on the chamber with the main power switch.
• If the fault shows again, the chamber could be faulty. Please contact BINDER Service.
12.1.1 Measuring the CO2 concentration indirectly via the pH of the cell medium
By using the indirect determination of CO2 concentration via the pH-value of the nutrient, it is possible to
check the CO2 concentration inside the chamber. This is a simple method to test the correct CO2 concen-
tration without any special CO2 measuring equipment. You need only use an accurate pH indicator or a
pH-measuring electrode, which are standard equipment in cell culture laboratories.
This method is not suitable for calibrating the BINDER FPI sensor system.
This method is based on the acid base equilibrium of the buffer system in the culture media.
NaHCO3 buffers the common media. From the pH value of the medium, it is possible to conclude its CO2
concentration. Figure 23 shows the relationship between CO2 concentration in vol.-% and the pH of dif-
ferent NaHCO3 buffered media.
Recommended procedure:
• Incubate an empty sample with medium for 1/2 day under the same conditions as the cells. You can
perform the incubation in a cell culture bottle or in a 50 ml Falcon tube with open lid.
• After gassing, remove the empty sample from the incubator and within 5 minutes measure the pH-
value with a glass electrode.
During the measurement, the medium should have the least possible surface contact with the ambient
air, so that the CO2 can evaporate only slightly. A significant downward movement will happen only af-
ter 5 minutes, permitting sufficient time for measurement.
• In addition, you can of course use pH-test strips (pH range 6 to 8, non-bleeding).
Vol.% CO2
18
Trade names of common media:
3,70 g/l NaHCO3
16 NaHCO3 [g/l]
14 DMEM 3.70
2,20 g/l NaHCO3 BME 2.20
12
MEM 2.20
10 Medium 199 2.20
1,20 g/l NaHCO3 Mc Coy 2.20
8
F10 1.20
6 F12 1.20
0,85 g/l NaHCO3
4
This is a common do-it-yourself test for many users. A chemical color reaction in a glass tube shows the
CO2 concentration. A standardized volume of air from inside the incubator has to be suctioned through
this glass tube to get a quantitative test result. Therefore, use a special hand pump with a standardized
suction volume.
Procedure (example):
(1) Break off both ends of the glass tube or remove the plugs.
(2) Pin the end with the higher end of the scale to the adapter of the hand pump that belongs to that
test system.
(3) Pin the other end through the silicone access port of the inner chamber door.
(4) Take one sample volume out of the inner chamber volume by pressing the pump fully together and
releasing it afterwards.
(5) The standardized volume is suctioned through the glass tube and the chemical indicator changes
its color beginning from the side pinned into the chamber in the direction of the hand pump.
(6) The more CO2 is inside the chamber the more the chemical reaction will cause a color change of
the chemical reactor.
(7) You can read the CO2 concentration by the scale directly printed on the glass tube or a delivered
reference-reading rule.
(8) The result must then be corrected to the current ambient pressure. The required formula is printed
on the instruction sheet of such systems.
All the necessary equipment must be supplied by one manufacturer only and in a defined test system.
Note: These test systems are not very accurate. A typical accuracy is around 10% of the full-scale value.
These test systems are not suitable for calibrating the BINDER FPI sensor system.
Figure 24: Example of chemical indicator Figure 25: Example of a hand pump (foreground) and elec-
tubes trical pump (background)
The easiest way of measuring the CO2 concentration is by electronic sensor systems. BINDER offers the
portable measuring device model CTM 01 that was specifically designed to measure temperature and
CO2 concentration inside CO2 incubators. The CTM 01 is suitable both for reference measurements in
certified laboratories, and for service purposes. Please contact the BINDER INDIVIDUAL team.
13. Options
13.1 Silicone access ports 30 mm / 1.18 in, closable with 2 silicone plugs
(8012-0558 rear, 8012-0559 left, 8012-0560 right) (option)
(c)
(a)
(b)
Figure 26: Positions of the optional silicon access ports left (a), rear (b), and right (c)
WARNING
High concentration of CO2 (> 4 Vol.-%).
Risk of death by suffocation.
Danger of poisoning.
Tightly close each access port with two plugs during operation.
The mounting instructions 7001-0147 delivered with the base on castors describe its
installation (Art. No. 9051-0024).
The mounting instructions 7001-0145 delivered with the stacking adapter describe its
installation (Art. No. 9051-0025 for 2 C 170, Art. No. 9051-0027 for C 170 on top of CB
150/160).
DANGER
Electrical hazard.
Danger of death.
∅ Do NOT spill water or cleaning agents over the inner and outer surfaces.
Before cleaning, turn off the chamber at the main power switch (2) and
disconnect the power plug.
Completely dry the appliance before turning it on again.
15.1 Cleaning
Disconnect the CO2 incubator from the power supply before cleaning. Disconnect the power plug.
The interior of the chamber must be kept clean. Thoroughly remove any residues of the charg-
ing material.
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Alcohol-based solutions
Do not immerse the CO2 sensor into the solution.
CO2 sensor Disinfection with alcohol or an alcohol-based surface disinfectant without corro-
sive effect, free from acid or halides. We recommend using the disinfectant spray
Art. No. 1002-0022.
Silicone door gasket: Alcohol-based solutions or neutral cleaning agent Art. No. 1002-0016.
Zinc coated hinge
Standard commercial cleaning detergents free from acid or halides. Do NOT use
parts, rear chamber
a neutral cleaning agent on zinc coated surfaces.
wall
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or
the charging material. If there is doubt regarding the suitability of cleaning products, please contact
BINDER service.
We recommend using the neutral cleaning agent Art. No. 1002-0016 for a thorough cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
CAUTION
Danger of corrosion.
Damage to the chamber.
∅ Do NOT use acidic or chlorine cleaning detergents.
∅ Do NOT use the neutral cleaning agent on other kind of surfaces e.g., the zinc coated
hinge parts or the rear chamber wall.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
With every cleaning method, always use adequate personal safety controls.
Following cleaning, leave the chamber door open or remove the access port plugs.
The neutral cleaning agent may cause health problems in contact with skin and if ingested.
Follow the operating instructions and safety hints labeled on the bottle of the neutral clean-
ing agent.
Recommended precautions: To protect the eyes use sealed protective goggles. Suitable protective
gloves with full contact: butyl or nitrile rubber, penetration time >480 minutes.
CAUTION
Contact with skin, ingestion.
Skin and eye damage due to chemical burns.
∅ Do not ingest. Keep away from food and beverages.
∅ Do NOT empty into drains.
Wear protective gloves and goggles.
Avoid skin contact.
Following use of the neutral cleaning agent and prior to hot-air sterilization, remove any agent
residues by using a moistened towel in order to avoid formation of permanent residues.
For chemical disinfection, we recommend the disinfectant spray Art. No. 1002-0022.
Any corrosive damage that may arise following use of other disinfectants is excluded from
liability by BINDER GmbH.
In case of contamination of the interior by biologically or chemically hazardous material, there are two
possible procedures depending on the type of contamination and charging material.
1. Spray the inner chamber with an appropriate disinfectant.
Before start-up, the chamber must be absolutely dry and ventilated, as explosive gases may form dur-
ing the decontamination process.
2. You can sterilize the shelves and the shelf holder in a sterilizer or autoclave.
In case of eye contact, the disinfectant spray may cause eye damage due to chemical
burns. Follow the operating instructions and safety hints labeled on the bottle of the disin-
fectant spray.
VORSICHT
Eye contact.
Eye damage due to chemical burns.
∅ Do NOT empty into drains.
Wear protective goggles.
Following frequent use of the disinfectant spray and prior to hot-air sterilization, remove any
agent remainder by using the neutral cleaning agent and then a moistened towel to avoid for-
mation of permanent residues.
After using the disinfectant spray, allow the incubator to dry thoroughly, and aerate it suffi-
ciently.
CAUTION
Excess temperature.
Immersion of sensor into liquids.
Shocks of the CO2 sensor.
Damage to the CO2 sensor.
∅ Do NOT immerse the CO2 sensor into liquids.
∅ Do NOT expose the CO2 sensor to autoclaving.
∅ Do NOT expose the CO2 sensor to hot-air sterilization.
Avoid strong shocks of the CO2 sensor (by putting it down hard, or dropping).
CAUTION
Connecting or removing the CO2 sensor during operation.
Damage to the CO2 sensor.
Connect or remove the CO2 sensor only with the chamber turned off.
Recommended procedure:
• Turn off the chamber
• Pull out the sensor
• Spray the sensor head with alcohol or wipe it clean with a soaked cloth. Observe the reaction time of
the disinfectant used.
• Before reinserting the CO2 sensor, it must be completely dry.
• The filter in the front of the sensor only needs replacing when damaged or dirty.
The CO2 sensor head was especially adjusted for the specific chamber. To avoid confusion, an adhesive
label with a serial number is adhered to the sensor head. When exchanging the sensor, repeat CO2 ad-
justment.
CAUTION
Different CO2 sensor.
Invalid calibration.
∅ Do NOT change the CO2 sensor head.
Write down the serial number of the CO2 sensor.
The very first sterilization after operation may cause an odor. This is not a quality defect. We
recommend ventilating well the room during sterilization.
15.4.1 Overview
37 °C
approx. 2
4 hours approx. 6 hours
hours
Sterilization cycle of approx. 12 hours
The chamber can perform an automatically controlled hot-air sterilization cycle. This procedure will take
approx. 12 hours and consists of the following steps:
• Heating up phase: The incubator heats up the inner chamber as fast as possible to the sterilization
set-point temperature
• Holding phase: Constant sterilization set-point temperature
The sterilization set-point temperature is pre-set in factory to 190 °C / 374 °F. When this temperature
is reached, the holding phase begins. The duration of the holding phase is in total 4 hours. This en-
sures that 180 °C / 356 °F is maintained on all internal surfaces for at least 30 minutes.
• Cooling down phase: until 37 °C / 98.6°F is reached.
When 37 °C / 98.6 °F ± 1 ° is reached, the display "End" in the lower controller display (Figure 32 page
67) indicates the end of the sterilization procedure.
The precise duration of the entire sterilization cycle depends on the ambient temperature at the installa-
tion site and can thus vary. At an ambient temperature of 25 °C / 77 °F, the total duration is approx. 12
hours. You can shorten the cooling-down time by aborting the sterilization cycle (chap. 15.5) during the
cooling-down phase, i.e. no sooner than after 6 hours.
During sterilization, the CO2 valve is closed and the CO2 controller turns off entirely.
Before carrying out the first hot-air sterilization, remove any protective lamination sheet from
the inner metal surfaces.
The safety controller settings are inactive during sterilization. They become functional again
following abortion of the sterilization and/or restart of the chamber at the main power switch.
CAUTION
Excess temperature.
Damage to the CO2 sensor.
∅ Do NOT expose the CO2 sensor to hot-air sterilization.
The CO2 sensor head was especially adjusted for your specific chamber. To avoid confusion, an ad-
hesive label with a serial number is adhered to the sensor head. When exchanging the sensor, repeat
CO2 control adjustment.
CAUTION
Different CO2 sensor.
Invalid calibration.
∅ Do NOT change the CO2 sensor head.
Write to down the serial number of the CO2 sensor.
WARNING
Danger of implosion.
Damage to the chamber.
Empty the water pan before starting the hot-air sterilization.
• Press “▲”.
The upper display toggles from “OFF” to “ON”.
• Press “MODE” to start the sterilization cycle.
The upper display toggles between the current interior temperature in °C and “StE”.
The indication “ - - - “ in the lower display shows that the CO2 sensor is disconnected.
CAUTION
Interruption of the temperature reaction time.
Ineffective sterilization.
∅ Do NOT open the chamber doors during sterilization.
• The sterilization cycle ends after approx. 12 hours, when the lower display shows “End”.
CAUTION
Interruption of temperature reaction time.
Ineffective sterilization.
Repeat the sterilization.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
∅ Do NOT touch the glass door and inner surfaces for approx. 7 hours after aborting steri-
lization.
• When the inner chamber has cooled down to a value below 60 °C / 140 °F, turn off the chamber and
plug in the CO2 sensor.
Three events which can abort the automatic sterilization cycle prematurely:
• A manual abortion (chap. 15.5.1)
• Opening the outer door (chap. 15.5.2)
• Turning off the power to the chamber or a brief power failure (chap. 15.5.3)
For safety reasons, the sterilization cycle is automatically aborted if you open the outer chamber door.
• Do not open the glass door and close the outer door immediately.
• After an abortion caused by opening the outer door, the chamber reverts to its standard operational
status. The indication “ - - - “ in the lower display shows that the CO2 sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the chamber will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
• If necessary, repeat the sterilization.
• When the inner chamber has cooled down to a value below 60 °C / 140 °F, turn off the chamber and
plug in the CO2 sensor.
15.5.3 Aborting the hot-air sterilization using the main power switch
For safety reasons, the sterilization cycle is aborted if the chamber is turned off or if there is a power fail-
ure.
Normally you should not terminate sterilization in this way.
• Do not open the chamber doors until the interior temperature has dropped to 37 °C / 98.6 °F.
• After turning on again the chamber, it reverts to its standard operational status. The indication “ - - - “
in the lower display shows that the CO2 sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the chamber will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
• If necessary, repeat the sterilization.
• When the inner chamber has cooled down to a value below 60 °C / 140 °F, turn off the chamber and
plug in the CO2 sensor.
DANGER
Electrical hazard.
Danger of death.
∅ The chamber must NOT become wet during operation or maintenance work.
∅ Do NOT remove the chamber’s rear panel.
Before conducting maintenance work, turn off the chamber at the main power switch (2)
and disconnect the power plug
Ensure all maintenance work is conducted by licensed electricians or experts author-
ized by BINDER.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.
The CO2 sensor was especially adjusted for the specific chamber. When exchanging the sensor, you
must repeat the CO2 adjustment.
We recommend taking out a maintenance agreement. Please consult BINDER Service.
BINDER telephone hotline: +49 (0) 7462 2005 555
BINDER fax hotline: +49 (0) 7462 2005 93555
BINDER e-mail hotline: [email protected]
BINDER service hotline USA: +1 866 885 9794 or +1 631 224 4340 x3 (toll-free in the USA)
BINDER service hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
BINDER service hotline Russia and CIS +7 495 988 15 16
BINDER Internet website http://www.binder-world.com
BINDER address BINDER GmbH, post office box 102, D-78502 Tuttlingen
International customers, please contact your local BINDER distributor.
For security reasons we cannot accept a chamber delivery if it does not carry an authorization
number.
Solid wood
Wood recycling
(IPPC standard)
If recycling is not possible, all packing parts can also be disposed of with normal waste.
• Turn off the main power switch (2) and disconnect the chamber from the power supply (pull the power
plug).
• Turn off the CO2 supply. Remove the gas connection.
• Let the inner chamber sufficiently cool down before removing any parts.
• The water pan must not remain filled while the incubator is out of operation. Otherwise condensation
may occur on the inner surfaces, as well as in the injection and suction nozzle of the CO2 sensor com-
partment. If condensation formation has occurred, drops of condensate would leak from the openings
of injection and suction nozzle of the CO2 sensor. In this case, clean and dry the incubator running at
37 °C / 98.6 °F with doors open for at least one hour before loading it with samples. BINDER recom-
mends performing a hot air sterilization of the chamber before commissioning.
• Disassembling the shelf holder takes place inside the chamber. Remove the shelves. Then fold both
lateral parts to the middle and remove them.
You cannot take out the shelf holder of the incubator entirely together with the shelves.
CAUTION
Violation against existing law.
∅ Do NOT dispose of BINDER devices at public collecting points.
Have the device disposed of professionally at a recycling company which is certified
according to the German national law for electrical and electronic equipment (Elektro-
und Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl. I p. 1739).
or
Instruct BINDER Service to dispose of the device. The general terms of payment and
delivery of BINDER GmbH apply, which were valid at the time of purchasing the cham-
ber.
Prior to handing the chamber over to a recycling company, it is the user’s responsibility that it
is free from toxic, infectious or radioactive substances.
• Prior to disposal, clean all introduced or residual toxic substances from the chamber.
• Prior to disposal, disinfect the chamber from all sources of infection. Be aware that sources
of infection may also be located outside the inner chamber.
• If you cannot safely remove all toxic substances and sources of infection from the chamber,
dispose of it as “special” waste according to national law.
• Fill out the contamination clearance certificate (chap. 22) and enclose it with the chamber.
WARNING
Contamination of the device with toxic, infectious or radioactive substances.
Danger of intoxication.
Danger of infection.
∅ NEVER take a chamber contaminated with toxic substances or sources of infection for
recycling according to Directive 2012/19/EU.
Prior to disposal, remove all toxic substances and sources of infection from the cham-
ber.
A chamber from which all toxic substances or sources of infection cannot be safely
removed must be considered as “special” waste according to national law. Dispose of it
accordingly.
17.4 Disposal of the chamber in the member states of the EU except for the Fed-
eral Republic of Germany
According to Annex I of Directive 2012/19/EU of the European Parliament and of the Council on waste
electrical and electronic equipment (WEEE), BINDER devices are classified as “monitoring and control
instruments” (category 9) only intended for professional use“. They must not be disposed of at public col-
lecting points.
The chambers bear the symbol for the marking of electrical and electronic equipment man-
ufactured / placed on the market in the EC after 13 August 2005 and be disposed of in
separate collection according to the Directive 2012/19/EU on waste electrical and electron-
ic equipment (WEEE). WEEE marking: crossed-out wheeled bin with solid bar under.
At the end of the device’s service life, notify the distributor who sold you the device, who
will take back and dispose of the chamber according to the Directive 2012/19/EU on waste
electrical and electronic equipment (WEEE).
CAUTION
Violation against existing law.
∅ Do NOT dispose of BINDER devices at public collecting points.
Have the device disposed of professionally at a recycling company that is certified ac-
cording to conversion of the Directive 2012/19/EU into national law.
or
Instruct the distributor who sold you the device to dispose of it. The agreements apply
that were agreed with the distributor when purchasing the chamber (e.g. his general
terms of payment and delivery).
If your distributor is not able to take back and dispose of the chamber, please contact
BINDER service.
Prior to handing the chamber over to a recycling company, it is the user’s responsibility that it
is free from toxic, infectious or radioactive substances.
• Prior to disposal, clean all introduced or residual toxic substances from the chamber.
• Prior to disposal, disinfect the chamber from all sources of infection. Be aware that sources
of infection may also be located outside the inner chamber.
• If you cannot safely remove all sources of infection and toxic substances from the chamber,
dispose of it as “special” waste according to national law.
• Fill out the contamination clearance certificate (chap. 22) and enclose it with the chamber.
WARNING
Contamination of the device with toxic, infectious or radioactive substances.
Danger of intoxication.
Danger of infection.
∅ NEVER take a chamber contaminated with toxic substances or sources of infection for
recycling according to Directive 2012/19/EU.
Prior to disposal, remove all toxic substances and sources of infection from the cham-
ber.
A chamber from which all toxic substances or sources of infection cannot be safely
removed must be considered as “special” waste according to national law. Dispose of it
accordingly.
CAUTION
Alteration of the environment.
For final decommissioning and disposal of the CO2 incubator, please contact BINDER
Service.
Follow the statutory regulations for appropriate, environmentally friendly disposal.
The main board of the CO2 incubator includes a lithium cell. Please dispose of it according to national
regulations.
c b
Figure 33: Determination of the useable volume
A
Exterior dimensions
Width, net mm / inch 740 / 29.13
Height including feet mm / inch 870 / 34.25
Depth, net mm / inch 715 / 28.15
Depth including door handle, I-triangle, and
mm / inch 795 / 31.30
connections
Wall clearance rear (minimum) mm / inch 100 / 3.94
Wall clearance side (minimum) mm / inch 50 / 1.97
Doors
Number of doors 1
Number of inner glass doors 1
Interior dimensions
Width mm / inch 560 / 22.05
Height mm / inch 600 / 23.62
Depth mm / inch 500 / 19.69
Interior volume l / cu.ft. 168 / 5.93
Shelves
Number of shelves, series 3
Number of shelves, max. 8
mm x mm 538 x 447
Size of shelf, width x depth
inch x inch 21.18 x 17.60
Weight
Weight (empty) kg / lbs 110 / 242.5
Temperature data
Temperature range, 7 °C / 12.6 °F above ambient,
°C / °F 50 / 122
up to
Temperature fluctuation ≤ +/- K 0.1
Temperature uniformity (variation) at 37 °C/ 98.6 °F +/- K 0.3
Recovery time after door open for 30 sec
minutes 6
at 37 °C / 98.6°F
Humidity data
Humidity range % r.H. 90 to 95
CO2 data
CO2 range vol.-% CO2 0 to 20
Setting accuracy vol.-% CO2 0.1
Recovery time after door open for 30 sec
minutes 7
at 5 vol.-% CO2
Drift-free CO2 infrared absorption measurement
CO2 measurement
system
Connection hose nozzle DN6 for CO2
Connection to the chamber for hose with internal di- mm / inch 6 / 0.24
ameter
To operate the chamber, use only original BINDER accessories or accessories / components
from third-party suppliers authorized by BINDER. The user is responsible for any risk arising
from using unauthorized accessories.
Regular equipment
Microprocessor display controller with 2-channel technology for temperature and CO2
CO2 infra-red absorption measuring system
Fan-assisted air jacket system
Hot-air sterilization at 180 °C / 356°F
Gas mixing head
Weldless deep-drawn inner chamber made of stainless steel V2A (German material no. 1.4301,
US equivalent AISI 304), polished
Electronic error auto-diagnosis system with zero-voltage relay output
Temperature safety device class 3.1 acc. to DIN 12880:2007
Tightly closing inner glass door
3 perforated shelves, stainless steel (German material no. 1.4016, US equivalent AISI 430)
C 170-UL: cUL test mark
Options / accessories
Perforated shelf, stainless steel
Door hinged left (option only available when ordering the chamber, no retrofitting)
Silicone access ports, closable with 2 silicone plugs 30 mm / 1.18 in , rear, left or right side
Analog outputs 4-20mA for temperature and CO2, with DIN socket 6-poles, DIN plug included
Base on castors
Stacking adapter for direct, thermally decoupled stacking
Gas cylinder connection kit
Pressure reducer
Cleaning kit (neutral cleaning agent, disinfection spray and lint-free disposable wipes, protective gloves
and safety goggles)
Calibration of temperature and CO2 including certificate
Spatial temperature measurement including certificate
Spatial temperature measurement acc. to DIN 12880:2007 including certificate
Qualification folder
BINDER GmbH is responsible for the safety features of the chamber only, provided skilled
electricians or qualified personnel authorized by BINDER perform all maintenance and repair,
and if components relating to chamber safety are replaced in the event of failure with original
spare parts. The user is responsible for any risks arising from using unauthorized accesso-
ries/components.
• A completely filled out form must be transmitted via Fax (+49 (0) 7462 2005 93555) or by letter in ad-
vance, so that this information is available before the equipment/component part arrives. A second
copy of this form must accompany the equipment/component part. In addition, the carrier should be
notified.
Eine vollständig ausgefüllte Kopie dieses Formblattes soll per Telefax (Nr. +49 (0) 7462 2005 93555) oder Brief
vorab an uns gesandt werden, so dass die Information vorliegt, bevor das Gerät/Bauteil eintrifft. Eine weitere
Kopie soll dem Gerät/Bauteil beigefügt sein. Ggf. ist auch die Spedition zu informieren.
• Incomplete information or non-conformity with this procedure will inevitably lead to substantial delays
in processing. Please understand the reason for this measure, which lies outside our area of influence
and will help us to speed up this procedure.
Unvollständige Angaben oder Nichteinhalten dieses Ablaufs führen zwangsläufig zu beträchtlichen Verzöger-
ungen in der Abwicklung. Bitte haben Sie Verständnis für Maßnahmen, die außerhalb unserer Einflussmöglich-
keiten liegen und helfen Sie mit, den Ablauf beschleunigen.
3. Details about utilized substances / biological substances / Einzelheiten über die eingesetzten
Substanzen/biologische Materialien:
b) ____________________________________________________________________________
c) ____________________________________________________________________________
3.2 Safety measures required for handling these substances / Vorsichtsmaßnahmen beim Um-
gang mit diesen Stoffen:
a) ____________________________________________________________________________
b) ____________________________________________________________________________
c) ____________________________________________________________________________
b) ____________________________________________________________________________
c) ____________________________________________________________________________
d) ____________________________________________________________________________
3.4 Other important information that must be taken into account / Weitere zu beachtende und
wichtige Informationen:
a) ____________________________________________________________________________
b) ____________________________________________________________________________
c) ____________________________________________________________________________
4. Declaration on the risk of these substances (please checkmark the applicable items) /
Erklärung zur Gefährlichkeit der Stoffe (bitte Zutreffendes ankreuzen) :
4.1 For non toxic, non radioactive, biologically harmless materials / für nicht giftige, nicht radi-
oaktive, biologisch ungefährliche Stoffe:
We hereby guarantee that the above-mentioned unit / component part… / Wir versichern, dass o.g.
Gerät/Bauteil...
Has not been exposed to or contains any toxic or otherwise hazardous substances / weder giftige noch
sonstige gefährliche Stoffe enthält oder solche anhaften.
That eventually generated reaction products are non-toxic and also do not represent a hazard / auch
evtl. entstandene Reaktionsprodukte weder giftig sind noch sonst eine Gefährdung darstellen.
Eventual residues of hazardous substances have been removed / evtl. Rückstände von Gefahrstoffen
entfernt wurden.
4.2 For toxic, radioactive, biologically harmful or hazardous substances, or any other hazard-
ous materials / für giftige, radioaktive, biologisch bedenkliche bzw. gefährliche Stoffe oder
anderweitig gefährliche Stoffe.
We hereby guarantee that … / Wir versichern, dass …
The hazardous substances, which have come into contact with the above-mentioned equip-
ment/component part, have been completely listed under item 3.1 and that all information in this re-
gard is complete / die gefährlichen Stoffe, die mit dem o.g. Gerät/Bauteil in Kontakt kamen, in 3.1 aufgelistet
sind und alle Angaben vollständig sind.
That the unit /component part has not been in contact with radioactivity / das Gerät/Bauteil nicht mit Ra-
dioaktivität in Berührung kam
Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.)
__________________________________________________________________________________
___________________________________________________________________________________
Name: _______________________________________________________________________
Position/Title: _______________________________________________________________________
Equipment that is returned to the factory for repair must be accompanied by a completely filled
out contamination clearance certificate. For service and maintenance on site, you must submit
such a contamination clearance certificate to the service technician before the start of any
work. No repair or maintenance of the equipment is possible, without a properly filled out con-
tamination clearance certificate.
Please fill:
Reason for return request Duplicate order
Duplicate shipment
Demo Page one completed by sales
Power Plug / Voltage 115V / 230 V / 208 V / 240V
Size does not fit space
Transport Damage Shock watch tripped? (pictures)
Other (specify below)
_____________________
Is there a replacement PO? Yes No
If yes -> PO #
If yes -> Date PO placed
To protect the health of our employees and the safety at the workplace, we require that this form is com-
pleted by the user for all products and parts that are returned to us. (Distributors or Service Organizations
cannot sign this form)
NO RMA number will be issued without a completed form. Products or parts returned to our
NY warehouse without a RMA number will be refused at the dock.
A second copy of the completed form must be attached to the outside of the shipping box.
1. Unit/ component part / type:
2. Serial No.
3. List any exposure to hazardous liquids, gasses or substances and radioactive material
3.1 List with MSDS sheets attached where available or needed
(if there is not enough space available below, please attach a page):
a) ____________________________________________________________________________
b) ____________________________________________________________________________
c) ____________________________________________________________________________
3.2 Safety measures required for handling the list under 3.1
a) ____________________________________________________________________________
b) ____________________________________________________________________________
c) ____________________________________________________________________________
3.3 Measures to be taken in case of skin contact or release into the atmosphere:
a) ____________________________________________________________________________
b) ____________________________________________________________________________
c) ____________________________________________________________________________
d) ____________________________________________________________________________
a) ____________________________________________________________________________
b) ____________________________________________________________________________
c) ____________________________________________________________________________
Name: _______________________________________________________________________
Position: _______________________________________________________________________
Company: _______________________________________________________________________
Address: _______________________________________________________________________
Phone #: _______________________________________________________________________
Email: _______________________________________________________________________
Date: _______________________________________________________________________
Signature: _______________________________________________________________________