Keg 500
Keg 500
KEG-500
OPERATION MANUAL
KOENG Co., Ltd. 1302, DIGITAL EMPIRE, 1130, BEOMAN-RO, GEUMCHEON-GU, SEOUL, KOREA
Tel: +82-2-838-6300 / Fax: +82-2-839-4700 / [email protected]
To use this analyzer with safety and effectiveness, please read this
manual before usage.
1. This manual should be referred with regard to the product design and
safety assurance of the configuration.
5. Please don't use the printer until the printing paper is prepared.
- The contents of this manual may be changed without notice for the
purpose of functional improvement.
3/63
4/63
5/63
(1) A damage or fault, which occurs due to mishandling in use and negligence in maintenance
and keeping.
(2) A damage or fault, which originates from a change or reconstruction.
(3) When the parts and consuming materials that naturally wear out become damaged
and needed to be exchanged.
(4) A damage or fault, which originates from an environmental factor such as a fire, earthquake,
storm and flood, or other natural disasters.
(5) A damage or fault, which occurs as the designated pure part is not used.
(6) A mistaken guarantee claim procedure
(for example, absence of type or serial number).
When you issue a claim against this product in accordance with above provision, please contact
the agent where you purchased:
(1) When you get a fault: Please verify the product by referring Chapter 8. fault diagnosis and
troubleshooting.
(2) Still in bad operation: Please request a repair to the agent where you purchased.
(3) Repair after guaranteed period: After the guaranteed period, the repair cost will be charged.
(4) Repair within guaranteed period: The guaranteed period is for 12 months from the purchase
and within this period the product will be repaired in accordance with our quality guarantee
provision.
(5) For details or questions on the after service, please contact our local agent.
(6) When you contact our local agent, please inform us your product type, serial number, date of
purchasing, and fault conditions in detail.
6/63
This analyzer is a equipment to measure the gas emission density of an automobile enabling to
diagnose the automobile status and its preventive maintenance so that it can provide a function
to prevent the air pollution in advance.
① This analyzer should be manipulated only by the trained personnel who well understood the
usage.
② A spot check or periodical check should be executed in accordance with this manual prior to
a measurement.
③ The probe should be installed in a place where not affected by a wind.
④ If there occurs a unusual case during operation, please stop the operation and contact us for
its inspection and verification.
(2) Warning
① This analyzer is designed for the use of AC110V only or AC220V only. Please verify the
power sources prior to the usage.
② The probe is so hot by emitted gas. Therefore, you need to pay high attention not to be
burned during when you insert or remove.
③ During the analysis, please do not stay long in the place to which the gas is being emitted.
④ The emitted gas normally includes the CO, which may induce a fatal damage to a human
body, so that it must be used only in well-ventilated place.
7/63
3-1 Specification
KEG – 500
CO(carbon monoxide),HC(Hexane), CO2(carbon dioxide),
Measuring item
O2(Oxygen), Lambda(air surplus rate),
AFR, NOX (optional)
Measuring CO, HC, CO2 : NDIR Method
method O2, NOx : Electrochemical Cell
Measuring range 0.00 ~ 9.99% 0 ~ 9999 ppm
Resolution CO 0.01% HC 1 ppm
Display 4 digit 7segment LED 4 or 5 digit 7segment LED
Measuring range 0.0 ~ 20.0% 0.00 ~ 25.00 %
Resolution CO2 0.1% O2 0.01 %
Display 4 digit 7segment LED 4 digit 7segment LED
Measuring range 0 ~ 2.000 0.0 ~ 99.0
Resolution Lambda 0.001 AFR 0.1
Display 4 digit 7segment LED 4 digit 7segment LED
Measuring range 0-5000 ppm
NOx
Resolution 1 ppm
(Option)
Display 4 digit 7segment LED
Repeatability Less than ±2% FS
Response time Within 10 seconds (more than 90%)
Warming up time About 2 ~ 8 minutes
Sample collecting
4 ~ 6 L/min
quantity
Power 220V AC or 110V AC ±10% 50 / 60Hz
Power consumption About 50 W
Operation
0℃ ~ 40℃
temperature
Dimensions 420 (W) × 298 (D) × 180 (H) mm
Weight About 6.9 kg
8/63
This manual is explaining mainly HOW TO USE 4GAS analyzer. 5GAS analyzer is just
added with NO sensor on 4GAS analyzer. So, for 5GAS analyzer, use this same
manual. You also can use the same manual for motorcycle which emits high density of
HC. Measuring range of the HC is 0 - 20,000 ppm and display window is specially
custom-made in order to indicate 5 digits segments, and for CO, CO2, O2, the display
window is indicating 4 digit segments as in general.
Other features such as the sequence and a series of procedures including GAS
calibration is same as those for 4GAS analyzer. So, you can use the same manuals as
well.
9/63
10/63
3-3-9 ▼
KEY
ZERO
3-3-10 ▲
KEY
PURGE
ENT
3-3-11 KEY
MEAS
ESC
3-3-12 KEY
STAND-BY
3-3-13 Internal printer case: The print paper comes out this case.
11/63
3-4-1 CAL(Calibration)
This is a gas inflow gate used in time of standard gas adjustment.
3-4-2 INLET
This is a measuring inlet where one end of the probe mounted into this inlet in time of gas
emission measurement.
12/63
3-4-8 OPTION
This is a terminal to be directly connected with a normal PC printer.
13/63
14/63
PART
NO. PART NAME Q'TY
NUMBER
1. PROBE QG0-M026 1PC
** As for the standard gas, we recommend you to buy it from local dealers.
15/63
The electrochemical method measures the gas density by using the quantity of electron which
produced in the time of oxidation and reducing reaction of the gas.
16/63
17/63
5-1 Installation
5-1-1 Fit end of the probe hose into the measuring probe and the other end of the hose into the
gas inlet in the rear of the analyzer. If the fitting condition is not good and the air comes
into flow from outside, it may yield an incorrect measured value.
Therefore, please carefully verify the fitting condition before use.
5-1-2 Turn off the power switch and then connect the power cable to the power socket located
in the rear of the analyzer
18/63
5-1-4 Verify again the connection status of the analyzer and then turn on the power switch.
5-1-5 Connect the ground cable to GROUND on the rear of the analyzer.
19/63
▶ This analyzer is operated with AC110V only or AC220V only. Please verify the power source
before use. (Note that you can not change power source which is fixed upon supplying this
analyzer.)
▶ The following factors should be considered to properly locate the analyzer body:
━ To be setup in a place of no direct light rays, humidity, vibration and abrupt temperature
change.
▶ The analyzer should not be moved during the operation. It may cause an
incorrectness in the measured value.
▶ The probe should be located in a place where not affected by the wind.
▶ The probe may be so hot during the operation so that it should be carefully handled not to be
burned in insertion or removal.
▶ It needs to be noted that you should not stay in the gas emitting place during the operation.
▶ The analyzer should be used in a well-ventilated place only because the emitting gas
contains the CO and it can cause a fatal damage to a human body.
▶ This analyzer is Type-approved in accordance with the related government laws so that it is
extremely prohibited to disassemble, change, or rebuild the analyzer and also if you
disassemble the NDIR Module, you may not be guaranteed for the repair.
▶ NDIR module is the core part of the analyzer. Disassembly or assembly work is
possible only by the manufacturer's technician. In any circumstance, you or anyone
else should not attempt to disassemble it. If disassembled or re-assembled, proper
A/S is not going to be possible.
20/63
6-1. Warming- Up
① Switch the power [ON]. KEG-500 will perform the initialization process for 10 seconds,
displaying the following.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
② KEG-500 will display current set date and time for about five seconds.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
21/63
the order of display window verification, communication, internal sensor, and memory
as follows. If the items diagnosed are normal, KEG-500 will display the [PASS]
message.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
22/63
% ppm
CO2 O2
% %
AFR
Lambda
NOX
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
23/63
⑤ The warming-up will be performed while the counter-value decreases by 1 between 120 and
480. ( Perform the warming-up for 2 through 8 minutes according to the neighboring
Activate the pump one minute before ending the warming-up and clear up the inside of
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
24/63
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
Processing
Processing
Processing
Processing
Processing
Automatic
25/63
The standby means when KEG-500 is in a non-measurement state for a long period of time. In
this state, the power is switched [ON] while the suction pump is switched [OFF].
ESC ENT
HOLD SELECT
ZERO PURGE STAND-BY MEAS
PRINT
Press the ESC key after the measurement or the purge function. And the suction pump will
be deactivated.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
If the window displays as shown above, rdy mode(ready mode) means the stand-by state and
the sign of “gASo” means the selected fuel to calculate the Lambda value.
KEG-500 has the power-saving software in it. It operates in a measurement mode and then
turns into a standby mode after 10 minutes as the pump automatically switches itself OFF.
It is also programmed to operate in a purge mode for just 10 minutes and turns to be in a
In this manner, in short term KEG-500 will elongate life cycle of a filter exchange period and in
long term, will elongate the life cycle of pump and other parts as well.
26/63
27/63
The [PURGE] mode indicates that the gas remaining in the sample-cell in KEG-500 is being
PURGE mode of KEG-500 cleans the probe for about 120 seconds and internal sample cell for
20 seconds.
ESC ENT
HOLD SELECT
ZERO PURGE STAND-BY MEAS
PRINT
① Press PURGE key and the following will be displayed. The probe is cleaned for
two minutes.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
② After the probe clearance, zero calibration is automatically performed as follows for
28/63
% ppm
CO2 O2
% %
AFR
Lambda
NOX
★Press the PURGE key one more time during the PURGE operation, and the PURGE
After continuously using the exhaust gas of a vehicle for a long time, there may be lots of
remaining gas in the probe and sample-cell. If the HC value indicates over 10 ppm after the
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
This preparation process is for the next accurate measurement and for the protection of
sample-cell in the equipment. Therefore, after the continuous operation for a long time, please
29/63
Zero key.
ESC ENT
HOLD SELECT
ZERO PURGE STAND-BY MEAS
PRINT
① Press the ZERO key. The counter values will be decreased by 1 from 20 as shown in the
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
30/63
6-5. Measurement
① Place the probe in the clean air to perform the [ Zero calibration ].
③ Push the probe deep into the exhaust outlet of the vehicle and measure exhaust gas by
ESC ENT
HOLD SELECT
ZERO PURGE STAND-BY MEAS
PRINT
※ The measurement operates for 10 minutes and the pump is automatically stopped by the
activation of the power-saving mode. Press MEAS key again to measure the exhaust gas
for more than 10 minutes.
③ Pull the probe out of the exhaust outlet of the vehicle. Then clean the inside of KEG-500 with
the clean air by pressing the PURGE key until the measurement values drop to 0.
④ If all the measurements fall close to 0, press ESC key to maintain KEG-500 in a stand-
31/63
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
6-6. Hold
This function stops measurement screen temporarily.
ESC ENT
HOLD SELECT
ZERO PURGE STAND-BY MEAS
PRINT
6-7. Print
32/63
Hold mode, and the car number registration screen is displayed as shown in the Figure above
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
④ Press the ▲ key, and the blinking figure increases. After adjusting the desired figure,
press the ▼ key and the figure in the next digit blinks. After registering the car number
by using the ▼ and ▲ keys, press the PRINT key. Then the print starts.
2012/12/15
PM 01:30
----------------------------------------------------------
33/63
Key Contents
Use the key to perform the leak test, the PEP value indication, the program
version indication, and the standard gas calibration. Press the [SELECT] key
Use this key to end the selection mode and change into a measurement
ESC
mode.
Use this key to select the present mode indicated, or to execute the number
ENT
substituted.
A function that looks for any possible air leakage from the sample-cell in order to indicate the
result.
ESC ENT
HOLD SELECT
ZERO PURGE STAND-BY MEAS
PRINT
① Press the SELECT key once in the stand-by mode to select the Leak Test mode.
CO HC
% ppm
CO2 O2
34/63
AFR
Lambda
NOX
② If the message ‘Leak test’ is indicated on the indication window as shown above, mount the
③Press ENT key. The pump operates and the count values decrease by 1 from 20
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
④ If the test turns out normal without any leak after 20 seconds, a message is indicated
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
35/63
% ppm
CO2 O2
% %
AFR
Lambda
NOX
⑤ KEG-500 returns to the Standby mode automatically unless other keys are pressed
36/63
This test measures HC element accumulated in sample extract probe line and inside this
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
② Press ENT key. Perform the zero calibration for 20 seconds as shown below.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
③ As the pump starts operating, the counter value is decreased by one from 30 as shown in the
following and the remaining HC test is performed for 30 seconds. During the test, HC density
is displayed.
37/63
% ppm
CO2 O2
% %
AFR
Lambda
NOX
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
④ If HC value is below 20ppm for 30 seconds, it means normal, and displays ‘PASS’.
If HC value is over 20ppm for 30 seconds, give another try from above ③.
If the result of the second try is over 20ppm, ‘FAIL’ sign is displayed.
⑤ Clean inside this analyzer by pressing the PURGE key when “Fail” is indicated because
remaining gas leaves from the end of the probe to inside this analyzer. But if “Fail“ is
indicated again after the purge mode process, please exchange various expendables filters
to have a remaining HC test again
38/63
This is the function that selects fuel for testing vehicle. It is used to calculate air surplus rate
(Lambda) and AFR. This analyzer can select the fuel such Gasoline, LPG, CNG, and Alcohol.
GASOLIN LPG CNG ALCOHOL
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
② Selected fuel is indicated as in the above. Use ▼ , ▲ keys until it indicates the fuel to
select.
④ For Example, If the “LPG” fuel is selected, the window indicates as shown in the below.
⑤ The selected fuel display will remain as set until the power supply is switched off in the
standby mode. It is programmed to display gASo (gasoline), the most commonly used fuel,
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
39/63
This is the function that selects ratio between hydrogen and carbon (Hcv) and ratio between
oxygen and carbon (Ocv) for the fuel. This function is used to calculate Lambda value which is
defined by the ratio of actual AFR and theoretical AFR through Brettschneider formula.
Following table is for normal Hcv and Ocv, but there are cases to apply other values depending
on formulation of fuels and country to use. Once Hcv and Ocv values are setup, they are stored
in memory and are maintained even after rebooting.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
!
CAUTION
The value of Hcv and Ocv is not possible to change. If it requires to change the value of Hcv
and Ocv due to the difference of molecular structure of the fuel, please contact those whom you
purchased the analyzer from.
40/63
Propane Equivalency Factor (PEF) indication function indicates the figure to convert from the
propane gas into the N-Hexane. The value is the unique constant of relevant measurement cell
and should not be modified or adjusted at random. Multiple the propane density by the PEF
values to convert into the N-Hexane density.
① Press SELECT key for five times. Then, the PEF values are indicated as the
following.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
Press the SELECT key six times in the Standby mode. The program and internal sensor
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
41/63
① Press SELECT key seven times in standby mode. Then current time is displayed as
shown below.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
② The values of date and time are set in the order of Year→Month→Day→
AM/PM→Hour→Minute, and are increased or moved with the ▼ and ▲ keys.
Upon completing the setup, press the ENT key to store it.
The range of setup value is shown in the Table below.
Setup Contents
Year Increase or decrease between three years before and after the set up year.
42/63
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
③ Select YES for NOX display mode, NO for non-NOX display mode, then press ENT
※ “A” is attached as shown below in the AFR(Air/fuel rate) display mode and disappears
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
43/63
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
ppm
This analyzer is supplied after accurate calibration with the standard gas (the gas authenticated
by the government-authorized organization). However, it is recommended to adjust the gas at
any time (at least once in every six months) to maintain the accuracy of this analyzer.
This analyzer is designed to prevent any the faulty calibration due to mishandling. Try to fully
understand these manuals and operate KEG-500 according to the specified procedure, for it has
a great influence on accuracy of a standard gas calibration.
① The gas used for the standard gas contains poisonous compositions and causes a fatal
damage to human body. Perform the calibration process in a well-ventilated place to
prevent any accident caused by the gas poisoning.
② If the mixed gas that has no test record or that is expired has been on use, the accuracy of
this analyzer can be seriously deteriorated. Therefore, it is recommended to use the
standard gas authenticated by the authorized organization.
③ In the NDIR type analyzer, the change of the pressure and the flow may influence the
measurement values in its principle of measurement. So, the standard gas with its pressure
and flow out of the specified range may lower the accuracy of the equipment.
④ The standard gas calibration process is complicated and may affect measurement values. It
is recommended to request the service center for the process, if possible.
44/63
Items Contents
The standard gas with the attached test record from the authorized
Standard gas
organization
Density range of the
C3H8 : 250∼3000 ppm, CO: 2.00∼6.00 %, CO2 : 10∼15 % N2 balance
standard gas
Recommended
density of the C3H8 : 2000 ppm, CO : 5.00 %, CO2 : 14 % N2 balance
standard gas
Pressure supplied 2GAS : 0.5 kgf/㎠±0.1
(Secondary 4GAS : 0.5 kgf/㎠±0.1
pressure) 5GAS : 0.7 kgf/㎠±0.1
Flow supplied 4∼5 ℓ/min
Use the device with the flow-meter, the flow adjustment valve and the
Pressure controller pressure gauge indicating both the first pressure and the secondary
pressure.
① Connect the standard gas cylinder to the CAL (standard gas suction nipple) in the rear of this
analyzer main frame. Be sure to close all the gas valves of the gas tank by turning them in
the [Close] direction to cut off the gas. The standard gas is only to be injected under the
② Unwind the pressure controller counterclockwise to cut off the pressure, and open the valve
of the standard gas cylinder counterclockwise. If the first pressure falls to 10kgf/cm2 or below,
④ Close the flow control valve and turn the control adjuster clockwise little by little to adjust the
45/63
using the pressure controller because the secondary pressure may change during the
[Figure 1]
46/63
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
ESC ENT
HOLD SELECT
ZERO PURGE STAND-BY MEAS
PRINT
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
④ Press ENT key in the above state. The [Zero calibration] starts automatically for
47/63
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
⑤ When the [Zero calibration] is completed, the values will be indicated on the
indication window as shown and the value in the first digit will blink.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
⑥ The ▲ key makes the blinking number increased. After the setup of the required
number, press ▼ key to move the blinking part to the next field. Use ▼ ,▲
keys to setup to standard gas value of gas cylinder, then press ENT key. The each value
of CO, CO2 and O2 is entered the one displayed on the gas cylinder, but the HC value is
entered the N-Hexane value which is explained in 7-6. For example, HC value = the value is
displayed on the gas cylinder (propane HC value) x PEF value.
If the HC value of the gas cylinder is 1002 ppm, multiply it by 0.5 of PEF value and the
outcome value of 501 ppm is gas calibration.
※ Adjustment of the NO value is just to input the value indicated on cylinder, then it will be
adjusted. When you do this process, you should be very careful and perform it in the area
where it has a good ventilation since NO gas is very toxic and dangerous.
48/63
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
⑦ Open the valve of the standard gas cylinder to spill the standard gas in a constant pressure.
If the measurement values are in a stable state, press both the ENT key and SELECT
key at the same time to start the gas calibration.
CO HC
% ppm
CO2 O2
% %
AFR
Lambda
NOX
⑧ If the gas calibration is successfully completed, the message [PASS] will be indicated on
each concerned window. If the gas calibration is successfully completed, the message
[PASS] will be indicated on each concerned window. When ENT key is pressed, the
calibrated values indicate on each concerned window. When ENT key is pressed one
more time, this analyzer cleans the calibration line for about 120 seconds and performs the
[Zero calibration] and returns to the standby mode.
⑨ If the gas calibration fails, the message [FAIL] is indicated on each concerned window.
Press the ENT key. KEG-500 will perform the [Zero calibration] and will return to
the standby mode.
49/63
% ppm
CO2 O2
% %
AFR
Lambda
NOX
* If the [FAIL] is displayed, return to the standby mode and press the PURGE key to
clean the sample-cell with a clean air for two or three minutes. Then perform the gas
calibration again.
50/63
51/63
9-1-1 Loosen five screws (four in the side and one in the rear) by using the (+) screwdriver.
9-1-2 After lifting a little the upper cover case of analyzer, pull it back for
dismounting.
52/63
9-2-1 If the probe filter contains lots of moisture or is polluted through the visible
verification, replace it.
9-2-2 How to replace the probe filter is as follows; Loosen B counterclockwise for
separation. Then replace the probe filter with new one and assemble it again
in the reverse order of disassembly.
9-3-1 Loosen the main filter housing (A) and (B) counterclockwise.
9-3-2 Replace the first main filter and the second main filter with new ones.
9-3-3 Assemble them in the reverse order of disassembly.
53/63
When replacing the printing paper, make sure the correct side, as the printing paper has
pintable side.
9-4-2. Insert the paper as shown on drawing 2. If it inserted oppositely with wrong side of paper,
it won't be printed, since only one side is printable.
9-4-3. When the rubber stick of printer falls off or when replacing the rubber stick, place it with
correct direction to fit to the gear.
CAUTION : When rubber stick and printer parts is moistened or blemished, major parts of
the printer could be damaged. When moistened, dry it before use.
When blemished, clean it before use.
9-5-1 Prt Set in the SELECT is to adjust the density of the printing letter. It is possible to adjust
the density from 0 to 399. To set the density, press the ▼ and ▲ keys in Prt Set.
Upon completing the setup, press the ENT key to store it.
9-5-2 Normal density is around 100 and depending on the printing material or how much dark
and light printing required, it could be adjusted. (At the time of delivery from factory, it is
set at 100).
54/63
CAUTION 1: Set Mode is generally used in winter time or very cold area since in cold
temperature the printing is usually produced light. Unless it produces very light,
it is better to keep density value at 100 in set mode.
CAUTION 2: Printing paper is thermal paper which is sensitive to the heat and bright light
such as sunlight, and when it is exposed to such circumstances, the condition of
the paper is turning to unprintable. So when keeping it in storage, store it in
black plastic bag to avoid the contact from the light.
9-6 O2 sensor
9-6-3 Split upper case of analyzer and dismount the O2 sensor connector.
9-6-5 After verifying that the rubber "O" ring is attached to new O2 sensor, mount it while
matching it with the screw thread.
55/63
9-7-1 If the dust filter contains lots of moisture or is polluted through the visible verification,
replace it.
9-7-3 After replacing the filter with new one, re-assemble it in the reverse order of disassembly.
9-8-3 After replacing the filter with new one, re-assemble it in the reverse order of disassembly.
56/63
9-9-1 If the analyzer power is not turned on because the fuse is gone, replace it.
9-9-3 Insert the (-) screwdriver or pin to “A” part for dismounting.
9-9-4 After replacing the fuse (250V 2A), re-assemble it in the reverse order of disassembly.
57/63
Product number
Product name Solid diagram Remarks
Specification: W×D×H(mm)
QG0-M066
Front panel case
420×170×30
QG0-M067
Rear panel case
417×167×20
QG0-M064
Upper cover case
420×140×270
QG0-M065
Bottom case
40×20×34
housing Ø 49×94
QG0-A017
Main filter
Ø 29×33
Main filter
QG0-A019
medium rubber
Ø 22×14
membrane
QG0-A013
Dust filter
Ø 35×81
58/63
Activated
QG0-A014
charcoal filter
Ø 35×81
(Zero filter)
QG0-A021
Probe filter
Ø 23×82
QG0-A006
Check valve
Ø 19×43
QG0-A007
Orifice
Ø 5×6
Printer QG4-M023
QG4-M062
Printer paper
Ø 48×57
QG4-M070
Rubber stick
Ø 8×64
59/63
QG0-A022
Probe handle
Ø 39×119
QG0-M026
Probe tip
Ø 8×345
Probe
QG0-M027
(stainless steel)
QG0-A015
Probe hose
Ø 6.5×7000
Solenoid QG0-M069
supporter 39×22×9
Solenoid QG0-A002
QG0-E017
Power supply
94×119×40
QG0-E021
Power socket
30×22×49
60/63
QG0-E022
Fuse
(250V 2A)
QG0-E019
Power cable
0.75SQMM×3C
QG0-E002
RS232 port
53×12×15
QG0-E009
Fan
80×25×80
QG0-A001
Pump
80×70×128
QG4-E015
O2 sensor
Ø 29×41
QG4-E016
NOx sensor
Ø 29×42
O2 sensor QG0-MO60
supporter 58×37×22
(Bench) 224×74×102
61/63
QG4-N019
Display PCB
152×18×123
QG4-M006
T-type fitting
Ø 5×31×21
OPERATION
MANUAL
supporter 40×20×34
62/63
Lambda is not the value measured from the vehicle exhaust gas, but is a constant
calculated based on the Bretschineider formula that measures the density of
CO, HC, CO2 and O2 and applies the percentage of H, C, and O of vehicle gasoline.
Bretschineider formula
Lambda is the value calculated based on the density of CO, HC, CO2, O2 of
gasoline gas and percentage of gasoline (C8H18) elements. Therefore,
measuring 4 gases accurately is prerequisite to the Lambda calculation.
63/63