Wave User Manual
Wave User Manual
User Manual
2 Wave User Manual
Version 3
Wave User Manual 3
Contents
1 INTRODUCTION 8
1.1 WHAT IS WAVE? 8
1.2 APPLICATION CONCEPT 8
Version 3
4 Wave User Manual
7 THE FILTERS 42
7.1 SETTING A FILTER 42
Version 3
Wave User Manual 5
11 SETTINGS 67
11.1 GENERAL 67
11.1.1 BRIGHTNESS 67
11.1.2 HIGH CONTRAST 68
11.1.3 DATE AND TIME 68
11.2 LANGUAGE 68
11.3 NETWORK 68
11.3.1 WIRED NETWORK CONNECTION 68
11.3.2 NETWORK CONNECTION VIA WI-FI 70
11.4 ABOUT 71
11.5 LEGAL 71
12 UPDATES 72
14 WAVE COMPANION 79
14.1 CONNECT WAVE TO THE WAVE COMPANION 79
14.2 VERSION COMPATIBILITY 79
15 QUALITY ASSURANCE 80
16 MAINTENANCE 81
16.1 TEMPERATURES 81
16.2 SCREEN 81
16.3 CONNECTORS 81
16.4 BATTERY AND POWER SUPPLY 81
16.5 USB-C CONNECTOR 82
16.6 TRAVELLING 82
17 ADDITIONAL RESOURCES 83
17.1 PROBES 83
17.2 ACCESSORIES 83
17.3 SONATEST WEBSITE 83
Version 3
6 Wave User Manual
18.1 WARRANTY 84
18.2 DISCLAIMER OF LIABILITY 84
19 TROUBLESHOOTING 85
19.1 BATTERY 85
19.2 USB-C CONNECTOR 85
19.3 SYSTEM STATUS 85
19.4 HARD SHUTDOWN 85
19.5 APPLICATION 85
20 GLOSSARY 87
Version 3
Wave User Manual 7
Copyright
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or
transmitted in any form or by any means, electronic, mechanical, photocopying, recording, or
otherwise, without the prior written permission of Sonatest Limited.
Sonatest Limited or its subsidiaries provide this manual “AS IS” without warranty of any kind, either
express or implied. Included but not limited to the implied warranties or conditions of merchantability
or fitness for a particular purpose. In NO events shall Sonatest Limited, its subsidiaries, its directors,
officers, employers, or agents be liable for indirect, special, incidental, or consequential damages,
even if Sonatest Limited or its subsidiaries have been advised of the possibility of such damages arising
from any defect or error in this manual or product.
Specifications and information contained in this manual are furnished for informational use only, and
are subject to change at any time without notice, and should not be construed as a commitment by
Sonatest Limited or its subsidiaries.
Version 3
8 Wave User Manual
1 Introduction
1.1 What is Wave?
Wave is a new generation, smart flaw detector. It has been designed to simplify the work of
technicians and engineers. It is based on simple, custom-made applications that can be tailored to
meet the specific requirements of different inspection procedures.
Wave also brings new tools to make the interpretation of A-Scans easier. A scan plan and complete 2-
D view can be used to represent the part being inspected. A smart ray tracer then displays the
ultrasonic beam path travelling within the piece of material, precisely showing the echoes, echo
amplitude and gate positions.
As well, Wave allows using all the main sizing methods, including DAC and DGS to name a few.
Applications are easily created using the Wave Companion Software. Once created, these new apps
are transferred and installed on a Wave unit. They are then immediately available to the Wave user.
A custom app will greatly simplify the instrument usage by building a much simplified environment.
This allows streamlining the user’s task, as well as lowering the risk of misusing the instrument.
Version 3
Wave User Manual 9
● Wave instrument;
● Smart Li-on battery pack;
● AC/DC battery charger;
● Ethernet to USB-C adapter;
● Ethernet cable;
● Quick Start Guide booklet;
● Couplant bottle.
Version 3
10 Wave User Manual
Touch
screen
LED
Power button
Figure 2 – Wave with the touch screen, power button, and LED indicator
The unit does not present any physical button apart from the power button. Operating the unit is done
using the integrated touch panel. Using the latest projective capacitive (P-CAP) technology, the touch
panel allows multi-touch as well as drag and pinch gestures. Using working gloves is not a problem as
the touch screen controller has been programmed to adapt to different circumstances. Moreover,
water or couplant will not interfere or affect the behaviour of finger detection.
NOTE: Disabling the touch screen can be useful for cleaning the screen.
Version 3
Wave User Manual 11
IMPORTANT: To turn the unit off, you must use the on-screen power-off button
appearing at the bottom right of the Home screen. See Section 4 for more
details.
Version 3
12 Wave User Manual
When using the instrument in pulse-echo mode, only the TX/RX output is used. For pitch and catch,
both TX/RX and RX must be hooked up to the probe.
TX/RX
RX
Expansion
GPIO
Version 3
Wave User Manual 13
NOTE: The UT and GPIO connectors, when not used, should always be hidden
by their protective covers. Though these connectors are rugged and rated IP67,
they can eventually become damaged by the accumulation of dirt, which may
cause some contacts to fail.
Version 3
14 Wave User Manual
USB-C
door
Wave’s USB-C port is hence used to recharge the instrument’s battery. The AC-DC power supply
provided with the unit allows fast charging. Typically, a full recharge of a dead battery to 100% charge
level will take just over 3 hours even if the unit is turned on during recharge.
USB-C hubs can be inserted between the power supply and the instrument, allowing extra connections
(like being hooked-up to a wired network) at the same time as the instrument’s battery is being
recharged.
Version 3
Wave User Manual 15
Battery
compartmen
t
NOTE: Always turn off the instrument before removing the battery. Doing
differently may cause undesirable effects.
Version 3
16 Wave User Manual
Version 3
Wave User Manual 17
Settings
Applications
Software
Update
Power-off
button
To launch an app, simply tap it to bring the Configuration Detail screen that displays a short
description of the app as well as a list of configuration files associated to this specific app. Moreover,
a section to the right lists the main parameters of the highlighted configuration, helping to select the
right configuration file prior to launching the app.
Main parameter
details of the
List of selected
configuration configuration
files
NOTE: From the Configuration Detail screen, you can also delete the selected
configuration file by using the Delete button. This requires a long press, as
indicated by the long press icon.
Version 3
18 Wave User Manual
Version 3
Wave User Manual 19
● Set parameters
● Select which measurements to display
● Move gates
● Change the A-Scan view to your preference
● Zoom in an A-Scan to view finer details
● Capture screens (screenshots)
Measurement bar
On-screen action
buttons
Menu items
Tool menu
Parameters
associated to Signal
menu item,
Version 3
20 Wave User Manual
NOTE: You may have noticed a small hand and finger in the upper right corner
of some on-screen buttons. This means that a long press will trigger a specific
action. For example, on the Gain button, this will set the amplitude of the A-
Scan under the selected gate to a predetermined FSH, usually 80%.
Version 3
Wave User Manual 21
Numeric keypad
Pre-set values
Increment values
For example, in the above figure, simply select a new value for Velocity from the list of pre-set values.
You can also change the Velocity value by increments of ±1 or ±10. Another option would be to make
the on-screen keypad appear in order to then enter a specific value.
NOTE: The A-Scan grid can be aligned with the main tick mark of the horizontal
ruler. On the other hand, the grid can be fixed, displaying the traditional 10
vertical lines.
Version 3
22 Wave User Manual
View the following video for more details about the envelope
View the following video for more details about grid types and overlays.
Click on Signal menu item and then on Ref. A-Scan to activate it.
The alternate way of zooming in an A-Scan is simply by using the same zoom gesture you use on your
smartphone, that is by using 2 fingers and spreading them apart. It is then possible to zoom out by
simply pinching your fingers or use the Reset Zoom button.
When zoomed, you can pan through the entire range of the A-Scan by using the usual pan action, that
is dragging the screen with one finger.
View the following video for more details about zooming the A-Scan.
Version 3
Wave User Manual 23
NOTE: Some parameters can only take two (2) different values. When changing
the value of such binary type parameters, hitting the on-screen button will
immediately toggle the parameter value. The Ruler Mode parameter is one of
these parameters as it can only take one of two values: Distance or Time.
View the following video for more details about the A-Scan rectification
Button Action
Gain Change gain, from 0dB to 110dB
Version 3
24 Wave User Manual
Reset Zoom Resets A-Scan to normal (“zoom out”) mode. The button is enabled (non-
greyed out) when the A-Scan is zoomed. The A-Scan will be zoomed after
zooming in on a gate (by double tapping the gate) or after zooming in part of
the A-Scan.
Freeze Freezes the A-Scan. Once frozen, all gates remain active and can be moved
around. All measurements show live values according to new gate positions
and different measurements can be selected. Also note that when frozen, the
Screen Capture
The reference level 80% can be set to any FSH percentage by using the Ref. Amplitude parameter
found in the Cal menu item.
View the following video for more details auto 80% FSH
5.7 Gates
Wave supports four (4) gates. All can be used simultaneously and all offering all measurements. Please
note that gate to gate measurements are possible between Gate 1 and Gate 2 (G1-G2), and between
Gate 3 and Gate 4 (G3-G4).
Version 3
Wave User Manual 25
To enable a gate, simply select the gate from the menu items. Note that Gate X stands for Gate 1,
Gate 2, Gate 3, or Gate 4. Tap GX Enabled to change its value to Yes.
NOTE: Gates are easily differentiated by their unique colour. Gate 1 is red; Gate
2 is blue; Gate 3 is yellow; and Gate 4 is pink.
While on A-Scan view mode, user can set a gate by directly clicking on the gate of interest. Once the
user clicks on the gate, it will appear a double arrow on each corner of the gate and a cross in the
centre. At the same time, it will enable the corresponding gate on the Menu items. User can set the
starting point and the length of the gate by clicking on the arrows, respectively. By clicking on the
centre cross, user can set the height and the position (X-axis) of the gate.
While on Scan Plan view mode, user can set a gate by directly clicking on the gate of interest. Once
the user clicks on the gate, Menu will change to the corresponding gate, and the user can set gate
parameters (gate enabling, gate start, gate width, gate level and gate polarity) according to his needs.
Note that even if a signal peak amplitude is lower than a gate, the gate will display the FSH% if this is
the measurement selected. The exception being for a signal below 5% FSH.
You can touch a gate and move it around by dragging it to a new position. When you do so, the gate
is “selected” and appears bold. Also, the proper gate menu item is then automatically selected,
allowing to easily change gate parameters.
Version 3
26 Wave User Manual
Alternatively, you can also tap a gate. After tapping, the gate will appear with handles.
The left arrow, when dragged left or right, will change the starting point of the gate and its length. The
right arrow will allow changing the gate length without changing the gate starting point. Finally, the
cross-arrow handle is used to move the gate with useful constraints. Move up or down without
changing the gate position, or move right and left, without changing its level.
NOTE: When a gate is selected, the upper right corner of the A-Scan view always
shows the gate level in FSH percentage
NOTE: When moving a gate with your fingers, i.e. when you drag it, the value
of the gate level (FSH) is displayed live. This value is rounded to the nearest 5%.
Use the Gate Level parameter if you want to set a gate to any specific level.
● Absolute: Considers all signal for evaluation (available only with non-rectified signal)
● Positive: Considers only the positive signal for evaluation (available with non-rectified or
rectified or positive only signal)
● Negative: Considers only the negative signal for evaluation (available only with non-
rectified or negative only signal)
Option Meaning
Gx Enabled Unable or disable a gate (Yes or No)
Gx Start It sets the starting point of a gate (in mm or µs).
Gx Width It sets the width of a gate (in mm or inches).
Gx Level It sets the level of the gate (in %FSH).
Gx Polarity It sets the polarity of a gate (positive or negative).
Version 3
Wave User Manual 27
For a gate adaptive triggering set to F-F, it has the same described behaviour. However, but the flank
is the new source of position.
In echo to echo mode, this prevents Gate 2 to trigger the same peak as Gate 1. Consequently, a G2-
G1 measurements would have been set to 0 mm.
Version 3
28 Wave User Manual
5.8 Alarms
Alarm 1 Alarm 2 Alarm 3
Alarm 4
There are four alarms, simply named Alarm 1 to Alarm 4. They all appear at the top of the Wave
display. If an alarm is not active, it will not be visible. If it is enabled but not triggered, it will appear
dimmed, like a turned-off LED. When enabled and triggered, the alarm will appear bright, like a turned-
on LED.
In the above table, we can see the status of each of the four (4) alarms.
Table 9 - Alarms
NOTE: Touching the alarms at the top bar of the Wave is a direct short cut to
the Alarms menu.
The above figure shows the default alarm colours. It is possible to override the default source of each
alarm. That is, you can decide that Alarm 1 will be triggered by Gate 2. This is accomplished in the
Version 3
Wave User Manual 29
Alarms menu item. Moreover, the alarm LED will take the colour of its source gate. So, as an example,
if Alarm 1 is triggered by Gate 2, Alarm 1 will appear blue (instead of its default red colour).
When an alarm is triggered by a complex condition, it will appear differently, as shown below.
In the example above, we had G1 or G2 as the source of Alarm 1, and G1 and G2 as the source of
Alarm 2.
As an example, the above figure shows measurements that can be displayed when only
Gate 1 is enabled. The choice of measurements is then dynamic, making it easier to choose from
depending on the application running on the instrument and also depending on the number of gates
enabled.
Version 3
30 Wave User Manual
Gate to gate measurements are grouped in pairs. That is, GX-GY measurements are offered between
Gate 1 and Gate 2 or between Gate 3 and Gate 4. The example below shows this.
Version 3
Wave User Manual 31
The following table shows the various measurement choices for different gates.
Table 10 - Measurement choices versus gates
NOTE: The Gate Start and Gate Stop positions in the table above can be
expressed in a number of different ways. The usual way of representing theses
position is based on the sound path. It might also be useful to express them
relative to the depth, the true depth or the surface distance. This allows to
better understand which part of the piece being expected is covered by the gate
Flank: The first positive flank to cross the gate (the first value higher than the gate threshold that
crosses the gate).
First Peak: The highest value between the first two gate crossings (full positive flank and full negative
flank), with the first gate crossing being the flank.
The following figure shows the aforementioned definitions in graphical form. The peak corresponds
to B. The flank is represented by A. The first peak is B.
Version 3
32 Wave User Manual
There is also a difference between the depth and true depth. Depth is considered as the depth without
taking into account multiple half skips.
True depth is considered as the depth in which multiple haft skips are taken into account. True depth
is less or equal to the thickness of the part.
Here are visualisations of true depth measurements for different parts that available on Wave:
Version 3
Wave User Manual 33
Curved part
Figure 27 – Visual representation of true depth measurement
Version 3
34 Wave User Manual
GX First Peak Sound Path GX^1⇘ Sound path distance measured at the
first peak
GX First Peak Depth GX^1⇓ Depth at the first peak position
GX First Peak True Depth GX^1⇓⊻ True depth at the first peak position
(takes skips into account)
GX First Peak Surface Distance GX^1⇒ Distance from the exit point of the
probe to the position of first peak
relative to the surface
GX First Peak Surface Distance GX^1⇒-X Distance from front of the probe to
Minus X-offset the position of first peak relative to
the surface
GX First Peak %Ref GX^1%REF Amplitude referred to the curve
(DAC, TCG and DGS) or to the
reference amplitude (Ref. Amplitude)
GX First Peak dB Ref GX^1dBREF Amplitude referred to the reference
dB. This is the difference between
the Ref dB and the actual dB
amplitude.
GX Flank Sound Path GX/⇘ Sound path distance measured at
flank
GX Flank Depth GX/⇓ Depth at flank position
GX Flank True Depth GX/⇓⊻ True depth at flank position (takes
skips into account)
GX Flank Surface Distance GX/⇒ Distance from the exit point of the
probe to position of flank relative to
the surface
GX Flank Surface Distance GX/⇒-X Distance from front of the probe to
minus X-offset position of flank relative to the
surface
GX Threshold --- GX threshold This is the gate level
GX Start Sound Path GX⇘start Gate start position relative to sound
path
GX Start Depth GX⇓start Gate start position relative to depth
Version 3
Wave User Manual 35
Version 3
36 Wave User Manual
Version 3
Wave User Manual 37
Surface distance
Depth
Say that the red dot in the above figure represents the position of a flaw, as it would be indicated in
the measurement of a gate. The red dot is positioned over the horizontal line where the probe lies. As
depth is usually represented as a positive number, referencing the red dot’s position requires using a
negative number.
Hence, we need to make sure we have a unique reference system for positioning, both in terms of
surface distance and in depth.
Version 3
38 Wave User Manual
Toolbar menu
Version 3
Wave User Manual 39
When saving a configuration, you will be presented with the following screen, allowing you to enter a
filename under which the file content will be saved.
Version 3
40 Wave User Manual
Probe model
available by
clicking here
IMPORTANT: It is not possible to save a Generic probe file. On the other hand,
you can save these parameters by saving the configuration as it is.
By touching the Probe Model menu parameter, the Probe Selection screen appears.
The left columns include fields (Manufacturer, Frequency, Type, Angle, DGS Only, Show Recent) to
help you search for a probe. The table on the right shows the list of available probes according to the
Version 3
Wave User Manual 41
fields chosen. In the probe list, it is possible to sort them by clicking Model, Manufacturer, Frequency,
Angle or Size.
Once a probe is loaded, it is possible to modify the Probe Model, the Angle, the Wedge Velocity, the
Probe Zero, the Frequency, the Probe Width and the Probe Type. Probe Height will always remain
greyed and X-Offset is only available for angular probes.
NOTE: DGS probes are identified with a small DGS logo next to the
manufacturer's name. DGS probes are the only probes that can be used to make
a DGS sizing method. Once a DGS probe is loaded all the menu parameters are
greyed out except Probe Model and Probe Zero.
View the following video for instructions on how to choose a probe in the Probe
Selection
Version 3
42 Wave User Manual
7 The Filters
Wave offers a set of 28 filters. Each of these is centred around the most usual transducer frequencies,
namely 2.25MHz, 4MHz, 5MHz and 7.5MHz.
Following is the list of these filters with their centre frequency, as well as their low and high pass
frequencies.
Table 13 - Filters
Filter Name Centre Frequency High Pass Freq. Low Pass Freq.
(MHz) (MHz) (MHz)
0.5 MHz Narrow 0.5 0.38 0.63
0.5 MHz Wide 0.5 0.25 0.75
1 MHz Narrow 1 0.75 1.25
1 MHz Wide 1 0.5 1.5
1.25 MHz Narrow 1.25 0.94 1.56
1.25 MHz Wide 1.25 0.63 1.88
2 MHz Narrow 2 1.5 2.5
2 MHz Wide 2 1 3
2.25 MHz Narrow 2.25 1.69 2.81
2.25 MHz Wide 2.25 1.13 3.38
3.5 MHz Narrow 3.5 2.63 4.38
3.5 MHz Wide 3.5 1.75 5.25
4 MHz Narrow 4 3 5
4 MHz Wide 4 2 6
4.5 MHz Narrow 4.5 3.38 5.63
4.5 MHz Wide 4.5 2.25 6.75
5 MHz Narrow 5 3.75 6.25
5 MHz Wide 5 2.5 7.5
7.5 MHz Narrow 7.5 5.63 9.38
7.5 MHz Wide 7.5 3.75 11.3
10 MHz Narrow 10 7.5 12.5
10 MHz Wide 10 5 15
15 MHz Narrow 15 11.3 18.8
20 MHz Narrow 20 15 25
20 MHz Wide 20 10 30
Broadband Low 2.25 to 5 1 6
Broadband High 7.5 to 10 2 15
Version 3
Wave User Manual 43
Version 3
44 Wave User Manual
Flat T-joint
NOTE: The part type choice influences the weld types. Thus, not all weld types
will be available depending on the chosen part.
Version 3
Wave User Manual 45
NOTE: For the Curved type, the Inside Diameter is automatically calculated. It
is therefore impossible for the user to modify it. It is a read-only parameter.
That is why it is always greyed out.
Version 3
46 Wave User Manual
Double V Single U
Double U Single J
Version 3
Wave User Manual 47
The availability of these welds depends on the type of part chosen beforehand. Table 15 illustrates
these possibilities. A checkbox indicates that the weld is visible for a given part.
Table 15 - Part types vs welds
Table 16 shows the dimensions according to the type of weld chosen. An X indicates that the
parameter is visible for a given weld.
Table 16 - Weld types and their dimensions
Version 3
48 Wave User Manual
Single V X X X X X
Double V X X X X X X X X
Single U X X X X
Double U X X X X X X X X X
Single J X X X X X X
Double J X X X X X X X X X X
Bevel Groove X X X X X
Double Bevel X X X X X X X X
Single Fillet X X
Weld
Double Fillet X X X X
Welds
NOTE: For Single U and Double U welds, the root opening (R) is equal to zero.
Since the parameter cannot be modified by the user, it is not visible.
● R: Root opening
● α: Bevel angle α
● β: Bevel angle β
● f: Root face
● S1: Weld depth of the bevel #1
● S2: Weld depth of the bevel #2
● r1: Groove radius #1
● r2: Groove radius #2
● L1: Length of the left-side fillet weld
● L2: Length of the right-side fillet weld
● Lw1: Effective throat of the left-side fillet weld
● Lw2: Effective throat of the right-side fillet weld
8.3 Wizard
The goal of the wizard is to quickly and simply guide the user to define the part to be inspected and
the weld, if any. Here is a summary of the steps to follow:
View the following video for how to define a Flat part with a Single V weld.
View the following video for how to define a T-joint with Fillet Welds.
Version 3
Wave User Manual 49
Part view
The Part View always includes vertical and horizontal rulers. The units of the rules are in SI or imperial
system depending on the user preferences (see section 5.9.5)
Version 3
50 Wave User Manual
Grid Type None, Plain, By default, the dimensions are 10x10mm (1”x1”), but can
Cross change as the user zooms in or out. Grid lines may help user
on sizing and distance evaluation.
Weld Visible Yes, No This feature allows the user to see the weld overlay. It helps
the user to be more confident with defects location and
evaluation
Signal Visible Yes, No This feature allows the user to have the A-scan displayed over
the ray tracer, in real time. It helps the user to be more
confident with defects location and evaluation.
Distance Visible Yes, No It allows the user to show or hide the distance.
NOTE: Distances located on the left side of the vertical reference point (Y axis
origin) are set by default as negative values. Distances located on the right side
of the vertical reference point (Y axis origin) are set by default as positive values.
View the following video for how to move a probe into the Scan Plan View.
Version 3
Wave User Manual 51
The Auto Cal automates this process so that only two readings are required, one on the thin sample
and one on the thick sample. The Wave then sets the Velocity and Probe Zero. This Auto Cal operation
must be repeated any time the test material is changed (velocity) or the transducer is changed (zero).
10.2 DAC
One of the many ways to perform echo evaluation is by means of the distance-amplitude correction
(DAC). DAC curves use a series of reference echoes from a reference block having artificial flaws. Due
to sound attenuation on materials and beam spread while using angle probes, echo heights of same-
sized reflectors are different at distinct depths (sound path).
Couple your probe with your standard block and try to maximise the echo from the first reflector.
Depending on your range size and the sound path associated with your reflector, the refracted echo
can be located outside of the reference window. You can move the reference window
(forward/backward) through the A-Scan until it encompasses the echo; simply click on the touch
panel. You can also enlarge or reduce the window width by directly clicking on the window and then
adjusting it according to your needs.
Version 3
52 Wave User Manual
At this point, you do not need to be aware of echo amplitude (gain level). You just need to make sure
that the first echo corresponds to your reference reflector to maximise its amplitude. When the first
echo is recorded, Wave automatically sets the Gain at 80% FSH by default. You can change the
reference amplitude to other values with the Ref. Amplitude button and according to your
requirements. The gain is locked and set as the reference gain. To record an echo, you should click on
Add Point. Once the reference gain is set and the first echo is recorded, users can randomly record
the other points to create a DAC curve. Users can also delete or replace another point as much as
necessary if they consider that the point was not properly taken. To delete or replace a point on a DAC
curve, move the reference window over the point and then click on the Delete Selected Point option.
If an echo amplitude falls below 20% FSH, click Split DAC Enabled. Once this option is switched on,
continue recording points. Wave will automatically set the gain so that echo amplitude will reach the
level of amplitude set by users.
Users can add up to 16 points to a DAC curve and also have four Split DACs, depending on their needs.
Version 3
Wave User Manual 53
Option Meaning
Transfer Loss The transfer loss is the gain to compensate for the difference of
attenuation between a reference standard and the part to be inspected.
DAC Trigger It sets the trigger of the alarm when DAC curve is on (DAC, G1, 6dB, 0dB,
-2dB, -6dB, -10dB, -12dB and -14dB). Options are directly dependent on
DAC sub curves.
Gain Mode Absolute or relative
Split DAC Enabled Turns on/off Split DAC option
DAC Subcurves -2/-6/-10, -6/-14, JIS and Custom
Figure 39 – Gain mode set to Relative and 6 dB of scanning gain has been added
In the figure above, 6 dB of scanning gain has been added. All the DAC curves have also been increased
by 6 dB. Please noted that all relative amplitude measurements have not changed ( +0 dB ).
The figure below is based on the same previous example; however, the Gain Mode is set to Absolute
instead. The A-scan is indeed +6 dB higher, while the DAC curves have not moved. Please note that
corrected (C) measurements provide a compensated amplitude reading. It is the advised assessment
method to use for DAC amplitude measurements. Uncorrected (U) measurements are not
compensated with respect to the DAC calibration.
Version 3
54 Wave User Manual
Figure 40 – Gain Mode is set to Absolute and 6 dB of scanning gain has been added
Version 3
Wave User Manual 55
Corrected measurements, such as the screenshot above, is the recommended way to use DAC
measurements. Regardless of the Gain Mode, the scanning gain is 1 dB stronger than the current
reference gain (42.7 dB+ 1.0 dB = 43.7 dB). The calibration has not been changed, so the corrected
readings are also 100.0% FSH and 0.0 dB. The uncorrected readings are 112.0% FSH and 1.0 dB. This
would mean a larger reflector. It also breaks the current calibration because a higher A-scan echo from
the same reflector should give the same DAC gain difference.
10.3 TCG
Another alternative to perform echo evaluation is by the time correction gain (TCG). TCG uses a series
of reference echoes from a calibration block with artificial flaws. This sizing method is a way to
compensate material attenuation and beam characteristics so that reflectors of equal size at distinct
depths into a piece will provide the same amplitude response on A-Scan.
Couple your probe with your standard block and try to maximise the echo from the first reflector.
Depending on your range size and the sound path associated to your reflector, the refracted echo can
be located outside of the reference window. You can move the reference window (forward/backward)
through the A-Scan until it encompasses the echo by just clicking on the touch panel. You can also
enlarge or reduce the window width by directly clicking on the window and then adjusting it according
to your needs.
At this point, you do not need to be aware of echo amplitude (Gain level). You just need to make sure
that the first echo corresponds to your reference reflector to maximise its amplitude. When the first
echo is recorded, Wave automatically sets the Gain at 80% FSH by default. You can change the refence
amplitude to other values on Ref. Amplitude button based on your requirements. The gain is locked;
it is set as the reference gain. To record an echo, you should click on Add Point. Once the reference
Version 3
56 Wave User Manual
gain is set and the first echo is recorded, users can randomly record the other points to create a TCG
curve. Users can also delete or replace another point as many times as necessary if they consider the
point was not properly taken. To delete or replace a point on your TCG curve, you need to move your
reference window over the point and then click on the Delete Selected Point option.
Option Meaning
Lock Points It allows users to lock or unlock (yes or no) the points on TCG curve
Range It allows users to set the range according to their needs.
Delete Selected Point It allows users to delete selected points on the TCG curve.
Version 3
Wave User Manual 57
10.4 DGS
Distance-gain-size diagram is a theoretical model for discontinuity sizing evaluation. This model allows
technicians to correlate discontinuity size between natural flaws and theoretical reflectors in terms of
distance compensation, and without requiring a large range of reference standards in order to
establish an evaluation curve.
DGS flaw sizing shows the sound response from “equivalent reflectors” of different sizes at distinct
depths, by performing mathematical predictions in terms of probe effective diameter, frequency and
near-field length.
When performing DGS evaluation, keep in mind that this defect sizing method shows some limitations
and it may lead to no definitive conclusions in some cases. As far as discussed above, it is a predictive
method and it tends to work better on standard configurations, circular probe elements and narrow
band.
NOTE: To perform DGS evaluation, click the Probe menu and choose a proper
DGS probe from the probe list. Otherwise, the DGS sizing option will remain
greyed out on Wave.
Version 3
58 Wave User Manual
Couple your probe with your standard block and try to maximise the echo from the first reflector.
Depending on your range size and the sound path associated to your reflector, the refracted echo can
be located outside of the reference window. You can move the reference window (forward/backward)
through the A-Scan until it encompasses the echo by just clicking on the touch panel. You can also
enlarge or reduce the window width by directly clicking on the window and then adjusting it according
to your needs.
Press Calibrate once the echo is correctly located in the calibration window. Wave will automatically
set the gain as the Reference Gain. Wave will then apply it to finalise the calibration procedure.
Version 3
Wave User Manual 59
Once a DGS probe is selected, the user may set some additional parameters.
Table 20 – Additional parameters for DGS sizing options
Option Meaning
Ref. Attenuation Attenuation finds the attenuation losses at the calibration step.
Material Att. Material attenuation of the test piece. Once the DGS calibration is
done, it applies the attenuation effect on all DGS curves.
Ref. Type Reference echo can be a side-drilled hole, flat bottom hole or the infinite
reflector size, also known as the backwall echo.
Ref. Type Reference echo can be a side-drilled hole, flat bottom hole or the infinite
reflector size, also known as the backwall echo.
Ref. Size Reference size during the calibration, its type is defined in the Ref. Type
tile
Ref. Block Velocity The velocity of the reference block
View the following video for how to create a DGS curve
Option Meaning
Transfer Loss The Transfer Loss is the gain to compensate for the difference of
attenuation between a reference standard and the part to be inspected.
DGS Trigger It sets the trigger of the alarm when DGS curve is on (DGS or G1).
Gain Mode Absolute or relative
ERS Equivalent Reflector Size
DGS Subcurves Ref, ERS or dB
10.5 AWS
Users can evaluate discontinuities according to AWS D1.1 (American Welding Society). This evaluation
is based on signal amplitude comparison. The echo amplitude from a discontinuity is compared to the
echo amplitude of a known reflector. It is defined as an indication rating according to the code. Sound
attenuation in test part is also taken into account for evaluation.
Version 3
60 Wave User Manual
At this point, you do not need to be aware of echo amplitude (Gain level). You just need to make sure
that the echo corresponds to your reference reflector and maximise its amplitude. When the echo is
recorded, Wave automatically sets the Gain at 80% FSH by default. You can change the refence
amplitude to other values on Ref. Amplitude button based on your requirements. Since the reference
level is taken (dB level), it is no longer possible to make adjustments on the reference gain; however
users can still set the range. For calibration, echoes must reach the gate at a proper amplitude level.
Version 3
Wave User Manual 61
Option Meaning
Transfer Loss The Transfer Loss is the gain (dB) to compensate for the difference of
attenuation between a reference standard and the part to be inspected.
AWS C Rounded It allows users to choose the rounding method for the C parameter (0.1 or
1 dB precision).
The AWS C rounded parameter is a material attenuation factor. It is expressed in terms of dB. It relates
the sound attenuation. It can assume positive or negative values depending on sound path length (in
inches). These values can be rounded to the unit or decimal according to customer needs.
10.6 API
Users are able to perform discontinuity evaluation according to American Petroleum Institute (API)
Recommended Practice for Ultrasonic Evaluation of Pipe Imperfections by using the Amplitude
Distance Differential Method (ADDM), also known as API Recommended Practice 5UE.
This evaluation method is a combination of amplitude comparison and a 6 dB drop in echo amplitude
to determine the radial depth of a discontinuity. It assumes that radial depth of a discontinuity affects
either the amplitude of the received echo signal and the differential time of flight of the incident
ultrasonic wave as it passes through the discontinuity.
Version 3
62 Wave User Manual
Option Meaning
Transfer Loss The transfer loss is the gain to compensate the difference of attenuation
between a reference standard and the part to be inspected
10.7 T-log
The method to log thickness data, the T-log (also known as Thickness logging), enables users to log
several thickness measurements in a table. These values are colour-coded based on different
thresholds (Min Threshold, Max Threshold, Tolerance).
The T-log is built using a wizard. The first step enables users to define the grid. The second step allows
users to record thickness measurements.
Version 3
Wave User Manual 63
The following table describes the parameters with respect to data acquisition.
Table 25 – Data acquisition parameters
As indicated in Table 22, the measurement that will be logged in the grid is the first one in the
measurement bar from the left-hand side.
Version 3
64 Wave User Manual
Measurement that
will be saved in the
grid
No logged valued
Version 3
Wave User Manual 65
below. The symbol is added to the measurement field at the top of the screen in order to highlight
to the user that the T-Min function is enabled.
This function allows users to set the minimum thickness value as a reference while taking
measurements, depending on the values previously established for minimum threshold, maximum
threshold, and tolerance (mm / in /%).
Once the T-Min function is enabled, a new button is also enabled, the Reset T-Min. This button allows
users to reset a new reference value for the minimum thickness, starting from the active cell. Once
activated, all measures taken after enabling this function will be taken as a reference for the minimum
thickness, which represents the new value established.
This function can be very interesting for users when they perform measurements with significant
variations in thickness. It allows you to perform all the necessary measurements without the need to
save them in separate files.
To access the files, users must connect to the device through the Wave Companion software. Once
connected, they will be able to select the desired files and transfer them to their computers. View the
figure below for more details.
Version 3
66 Wave User Manual
Figure 50 – Companion interface that shows stored CSV files in the instrument
The headers and the measurement relative to each file will be present at the top of the file's data
table. Values will be shown in the table as a function of measurements taken by users and the table
size (columns x lines). View the details in the figures below.
Version 3
Wave User Manual 67
11 Settings
Wave settings are available by clicking the gear wheel located in the top right corner of the Home
screen.
11.1 General
The General includes three sub-menus, which include the brightness, contrast, and date and time.
11.1.1 Brightness
You can change the brightness of the screen by moving your finger on the slider.
Version 3
68 Wave User Manual
Figure 55 – Normal contrast (black background) versus high contrast (white contrast)
11.2 Language
You can choose the language of the instrument from the list of available languages. The changes are
done automatically. In short, the instrument does not need to be rebooted.
11.3 Network
The device can be connected to a network either by a wire of by Wi-Fi.
Figure 56 – Network
Version 3
Wave User Manual 69
No matter how Wave is connected to a network (wire or Wi-Fi), an IP address will be assigned to the
device. In the case of a wire connection, the IP address is visible in the Network section, under the
Instrument Name.
Figure 58 – The IP address of a Wave unit when it is connected to a network via a wire
In addition, an icon, which represents a wired network connection, will appear at the top upper right-
hand corner, next to Time.
Version 3
70 Wave User Manual
The available networks are displayed. Some icons may appear next to a Wi-Fi network:
Private network
Public network
Version 3
Wave User Manual 71
By clicking on a private network, a window will open in order to enter the password.
NOTE: If the Wave is connected to a network by a wire and Wi-Fi at the same
time, the wire network will take precedence over the Wi-Fi network.
11.4 About
This section gives you some basic information about your Wave, like the date of the Last Calibration,
the Software Version, the Firmware Version and the Serial Number.
11.5 Legal
Wave was designed to be ultrasonic inspection equipment. It has been extensively tested before
leaving the manufacturer. However, operating software may experience untoward issues.
The instrument must be used only for industrial applications and material testing. Any other use than
that prescribed in this manual is strictly prohibited.
If you face any operating issues, please contact Sonatest’s customer service or our nearest
representative in your home country.
Version 3
72 Wave User Manual
12 Updates
IMPORTANT: Wave must be connected to the Internet to be able to update.
The Software Update is available by clicking on the circular arrows displayed on the Home screen.
Software Update
The Software Update screen allows the user to update the device via the Internet. If an update is
available, the user will be able to click on the Update button. Release notes are available for each
update.
Version 3
Wave User Manual 73
○ 4 MHz wide
Version 3
74 Wave User Manual
○ Pulse-Echo ● DGS ○ 1
○ 2
○ Through ● API
Transmission ○ 3
● AWS
● Voltage ● DAC Curve 1
● DAC to TCG
○ 100 V ● DAC Curve 2
● T-Log
○ 150 V ● DAC Curve 3
Sizing Options
○ 200 V ● AWS C Rounded
● Transfer Loss
○ 0.1
○ 250 V
● DAC Trigger ○ 1
○ 300 V
○ DAC ● ERS
○ 350 V
○ G1 ● Gain Mode
○ 400 V
○ 0 dB ○ Relative
○ 450V
● Gain Mode ○ Absolute
○ 500V
○ Relative ● DGS Trigger
● Pulse Type
○ Absolute ○ DGS
○ Spike
● Split DAC Enabled ○ G1
○ Square ● DGS Sub curves
○ Yes
● Frequency ○ Ref.
○ No
● PRF ○ dB
● TCG Visible
○ ERS
○ Yes
● Nb of DGS Curves
○ No
○ 0
● DAC Sub curves
○ 1
○ DAC ○ 2
○ JIS ○ 3
○ -2/-6/-10 ● DGS Curve 1
○ -6/-14 ● DGS Curve 2
○ Custom ● DGS Curve 3
● ERS Curve 1
● ERS Curve 2
● ERS Curve 3
Version 3
Wave User Manual 75
Part Probe
Part menu item is a wizard in fact. See section 8 for ● Probe Model
more details.
○ Probe Selection
● Angle
● X-offset
● Wedge Velocity
● Probe Zero
● Frequency
● Probe Width
● Probe Height
● Probe Type
○ Mono Circular
○ Mono Rectangular
○ Dual Crystal
Version 3
76 Wave User Manual
Version 3
Wave User Manual 77
○ None ○ None
○ 0 ○ 0
○ 1 ○ 1
○ 2 ○ 2
○ 3 ○ 3
● Alarm 2 GPO ● Alarm 4 GPO
○ None ○ None
○ 0 ○ 0
○ 1 ○ 1
○ 2 ○ 2
○ 3 ○ 3
Version 3
78 Wave User Manual
Version 3
Wave User Manual 79
14 Wave Companion
Wave Companion is the desktop application that accompanies Wave. It offers an environment in
which dedicated applications can be designed and uploaded to Wave units as well as a way to
download files.
NOTE: For more information about Wave Companion, please refer to the Wave
Companion User Guide.
• Wave Companion can manage several applications that are based on the current version or
older versions. However, these applications should not be corrupted.
• Wave Companion will transfer an application if its version is not as recent as the version of
Wave. Once exception is version 1.0.X of Wave Companion.
• When an application based on an old version is transferred to Wave, a clock icon will appear
on the application. This helps users distinguish old from current applications.
• When Wave is updated, default applications (Default imperial, T-joint, Default, DGS, DAC and
DASH) are also updated. For all other applications, users must update them in Wave
Companion or leave them as is.
Version 3
80 Wave User Manual
15 Quality Assurance
Wave was designed to meet and to surpass several international standards. All Wave units are tested
for quality assurance before delivery to customers.
The following list shows the standards that Wave has been tested for:
• EN 12668
• EN 61326-1
• FCC Part 15
• ICES 003
• CISPR11
• IEC 61000-4
Version 3
Wave User Manual 81
16 Maintenance
Wave requires basically no maintenance, only a few procedures for cleaning.
Do not leave the equipment close to electromagnetic sources because it may lead to electronic
interference. Likewise, keep it away from carpets that can provide electrostatic shocks.
To clean the instrument and its accessories use a soft cloth and ordinary liquid soap. Try to clean Wave
as often as possible to avoid any accumulation of dust, oil, grease, or couplant. Special attention needs
to be given to the battery door. Make sure the battery door gasket is properly clean and free of
particles.
IMPORTANT: Do not use any methyl alcohol or solvents for dye penetrant to
clean the equipment. It may damage some parts or brittle plastic components.
16.1 Temperatures
Wave’s operating temperatures (when the fan is working) are from -10°C to 45°C (14°F to 113°F).
16.2 Screen
For the Wave screen, we recommend the following advice:
● Do not use an abrasive cloth or brush to clean the screen.
● Do not use a sharp object to point functions on the screens.
● Do not leave or store the unit where the screen will be directly exposed to sunlight for a long
period of time.
16.3 Connectors
We strongly recommend users to protect all connectors to avoid contact with dust, oil, grease, and
water. All connectors come with a rubber cover which just needs to be put back in the right place
when the connector is not in use.
This procedure will avoid major damages to connectors and further issues related to pin corrosion.
Do not use the battery if it has been severely impacted or dropped, or if the battery has been exposed
to heavy pressure. Some damage may not be visible on the outside of the case, thus the cells inside
the battery may rupture or catch fire.
Do not expose the battery to seawater, rain, snow, drop in water or any other liquids. Do not charge
or use a wet battery. If the battery gets wet, make sure to wipe it dry before using.
Version 3
82 Wave User Manual
If the equipment is going to be stored for more than 30 continuous days with no utilisation, we
recommend fully recharging the battery, no matter of the battery level. This procedure will avoid
premature mortality of the battery.
The approximate level of remaining battery life is shown on the top right corner display by the battery
icon and the percentage level. When the battery indicator is down to one-quarter full, charge the
battery as soon as possible.
NOTE: The instrument automatically shuts off when the battery is too weak for
reliable operations. However, current settings are saved and then restored when the
instrument is turned on. For long work journeys on remote locations or if you are
unable to have access to main power supply, always carry a spare battery.
When the AC adapter is connected to the instrument, the battery icon indicates a percentage of full
charge of the battery. If the AC adapter is not connected to the instrument when removing the battery
to install a spare, the instrument will automatically turn off. However, if the adapter is connected, the
instrument will remain on.
16.6 Travelling
When travelling, Wave should be placed in its transportation case to avoid any major damages and
mechanical impacts.
Version 3
Wave User Manual 83
17 Additional Resources
17.1 Probes
Before operating the Wave, connect it to a probe. Probes can be connected to sockets on the top of
the instrument. Make sure to use appropriate cables and to properly connect them to the probes.
Otherwise, it may lead to considerable power losses or to echo waveform perturbation.
Wave can either work with single or double element probes. Single element probes need to be
connected to the proper socket as indicated on the following picture.
TX/RX
RX
Expansion
GPIO
17.2 Accessories
IMPORTANT: It is recommended that the instrument should be switched off, or at
least the pulser stopped, for the probe installation. If the Wave is not turned off, you
may cause some damage to the electronic circuits.
For options availability and quotations, please get in touch with one of our distributors.
Version 3
84 Wave User Manual
The Sonatest warranty only covers equipment that has been used in a proper manner, as described in
this document and that has not been subjected to excessive abuse, attempted unauthorised repairs,
or modification.
Inspect materials thoroughly on reception for evidence of external or internal damage that might have
occurred during shipment. Immediately notify the carrier making the delivery of any damage, since
the carrier is normally liable for damage during shipment. Retain packing materials, waybills, and other
shipping documentation needed in order to file a damage claim. After notifying the carrier, contact
Sonatest Limited for assistance with the damage claim and equipment replacement if necessary.
Neither the seller nor the manufacturer shall be liable either in contract or in tort for any direct or
indirect loss or damage (whether for loss of profit or otherwise), costs, expenses or other claims for
consequential or indirect compensation whatsoever (and whether caused by the negligence of the
company, its employees or agents or otherwise).
Version 3
Wave User Manual 85
19 Troubleshooting
19.1 Battery
Wave does not start when press I press the power-on button.
● Make sure the battery is fully charged and connected to the Wave.
● If the battery is flat or not fully charged, connect the charger which came with the unit.
● Remove and reinsert the battery then try to turning Wave on again.
● If the above does not work send it back for repairs.
My battery life is very short despite having charged it for 3-4 hours.
● When you received the unit for the first time, it is strongly recommended to complete a full
charge and discharge cycle.
● If possible, try charging the battery with another certified charger.
● If you do not have a spare charger, try with another battery. If the battery life is longer, the
battery is faulty.
● Have a look at the Sonatest website (sonatest.com) to see all the adapters compatible with
Wave.
● Yes. However Sonatest does not guarantee the compatibility of the adapter.
• Make sure you did not click the Freeze button or double-clicked the power-on to lock the
screen.
• If nothing is working with Wave, you can do a hard shutdown by long pressing and holding the
power-on button for 10 seconds. This action should be considered as a last resort.
19.5 Application
I have transferred an application and it is not working.
• Make sure you have not transferred a corrupted application. A corrupted application features
a logo with a crossed-out circle. You cannot open this application. You must replace the
application with one that is not corrupted or simply deleted the corrupted application.
Version 3
86 Wave User Manual
Version 3
Wave User Manual 87
20 Glossary
α (degree): Bevel angle α
First peak: The highest value between the first two gate crossing (full positive flank and full negative
flank), the first gate crossing being the flank.
Flank: The first positive flank to cross the gate (the first value higher than the gate threshold that
crosses the gate).
Probe zero (μs): Time which the sound spends in the wedge.
Reject type: Signal rejection criteria. It removes noise from the A-Scan.
Version 3
88 Wave User Manual
Transfer loss (dB): dB loss due to surface conditions, poor coupling etc.
X-Offset (mm or in): Distance between the front of the probe and the exit point.
Version 3
Wave User Manual 89
Version 3