TI01001F/00/EN/28.

22-00 Products Solutions Services


71603692
2022-12-13

Technical Information
Levelflex FMP51, FMP52,
FMP54
Guided wave radar

Level and interface measurement in liquids

Application
• Rod, rope or coax probe
• Process connection: Starting 3/4" thread, flange or process connections for hygiene
requirements (Tri-Clamp, 11851)
• Process temperature: –196 to +450 °C (–320 to +842 °F)
• Process pressure: –1 to +400 bar (–14.5 to +5 800 psi)
• Maximum measuring range: Rod 10 m (33 ft); rope 45 m (148 ft); coax 6 m (20 ft)
• Accuracy: ±2 mm (±0.08 in)
• International explosion protection certificates; WHG; marine approval; steam
boiler approval; EN10204-3.1
• Linearity protocol (3-point, 5-point)

Your benefits
• Reliable measurement even for changing product and process conditions
• HistoROM data management for easy commissioning, maintenance and
diagnostics
• Highest reliability due to Multi-Echo Tracking
• Hardware and software developed according to IEC 61508 (up to SIL3)
• Seamless integration into control or asset management systems
• Intuitive user interface in national languages
• Bluetooth® wireless technology for commissioning, operation and maintenance via
free iOS / Android app SmartBlue
• Easy proof test for SIL and WHG
• Heartbeat Technology™
Levelflex FMP51, FMP52, FMP54

Table of contents

Important document information . . . . . . . . . . . . . . . 4 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . 72


Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Function and system design . . . . . . . . . . . . . . . . . . . 5 Process temperature range . . . . . . . . . . . . . . . . . . . . . . 73
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Process pressure range . . . . . . . . . . . . . . . . . . . . . . . . . 73
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Dielectric constant (DC) . . . . . . . . . . . . . . . . . . . . . . . . 74
Extension of the rope probes due to temperature . . . . . . . . 74
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Mechanical construction . . . . . . . . . . . . . . . . . . . . 75
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Blocking distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Probe length tolerances . . . . . . . . . . . . . . . . . . . . . . . . 83
Measuring frequency spectrum . . . . . . . . . . . . . . . . . . . 15 Surface roughness of AlloyC coated flanges . . . . . . . . . . . 83
Shortening probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Materials: GT18 housing (stainless steel, corrosion-
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
resistant) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Materials: GT19 housing (plastic) . . . . . . . . . . . . . . . . . . 85
Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Materials: GT20 housing (die-cast aluminum, powder-
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
coated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Materials: Process connection . . . . . . . . . . . . . . . . . . . . 87
Materials: Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Materials: Mounting bracket . . . . . . . . . . . . . . . . . . . . . 92
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Materials: Adapter and cable for remote sensor . . . . . . . . 93
Device plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Materials: weather protection cover . . . . . . . . . . . . . . . . 94
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . 34
Operation with remote display and operating module
Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
FHX50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Operation via Bluetooth® wireless technology . . . . . . . . . . 97
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . 35
Integration into the Tank Gauging System . . . . . . . . . . . 101
SupplyCare inventory management software . . . . . . . . . 102
Performance characteristics . . . . . . . . . . . . . . . . . . 36
Reference operating conditions . . . . . . . . . . . . . . . . . . . 36
Certificates and approvals . . . . . . . . . . . . . . . . . . 105
Reference accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
RoHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
RCM marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Influence of ambient temperature . . . . . . . . . . . . . . . . . 39
Ex-approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Influence of gas phase . . . . . . . . . . . . . . . . . . . . . . . . . 40
Dual seal according to ANSI/ISA 12.27.01 . . . . . . . . . . . 105
Gas phase compensation with external pressure sensor
Functional safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
(PROFIBUS PA / FOUNDATION Fieldbus) . . . . . . . . . . . . 40
Overfill protection . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Gas phase compensation using reference signal (option for
Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . 105
FMP54) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
AD2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
NACE MR 0175 / ISO 15156 . . . . . . . . . . . . . . . . . . . . 106
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 NACE MR 0103 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Mounting requirements . . . . . . . . . . . . . . . . . . . . . . . . 43 ASME B31.1 and B31.3 . . . . . . . . . . . . . . . . . . . . . . . 106
Pressure equipment with allowable pressure ≤
200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Pressure equipment with allowable pressure >
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 65
200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . 65
Steam boiler approval . . . . . . . . . . . . . . . . . . . . . . . . . 106
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Marine approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Radio approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Altitude according to IEC61010-1 Ed.3 . . . . . . . . . . . . . . 72
CRN approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Experience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Test, certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Cleaning the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

2 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Product documentation on paper . . . . . . . . . . . . . . . . . 110


Other standards and guidelines . . . . . . . . . . . . . . . . . . 111

Ordering information . . . . . . . . . . . . . . . . . . . . . . 112


Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . 112
3-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . 113
5-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . 114
Customized parameterization . . . . . . . . . . . . . . . . . . . 116
Labeling (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . 116

Application packages . . . . . . . . . . . . . . . . . . . . . . 116


Heartbeat Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . 116
Heartbeat Verification . . . . . . . . . . . . . . . . . . . . . . . . 117
Heartbeat Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . 117

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . 118
Communication-specific accessories . . . . . . . . . . . . . . . 129
Service-specific accessories . . . . . . . . . . . . . . . . . . . . . 130
System components . . . . . . . . . . . . . . . . . . . . . . . . . . 130

Supplementary documentation . . . . . . . . . . . . . . 130


Brief Operating Instructions (KA) . . . . . . . . . . . . . . . . . 130
Operating Instructions (BA) . . . . . . . . . . . . . . . . . . . . . 130
Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . . 130
Functional Safety Manual (FY/SD) . . . . . . . . . . . . . . . . 130

Endress+Hauser 3
Levelflex FMP51, FMP52, FMP54

Important document information


Symbols Safety symbols
DANGER
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or
fatal injury.
WARNING
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or
fatal injury.
CAUTION
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or
medium injury.
NOTICE
This symbol contains information on procedures and other facts which do not result in personal
injury.

Electrical symbols

Direct current

Alternating current

Direct current and alternating current

Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system.

Protective earth (PE)


Ground terminals that must be connected to ground prior to establishing any other connections.
The ground terminals are located on the inside and outside of the device.
• Inner ground terminal; protective earth is connected to the mains supply.
• Outer ground terminal; device is connected to the plant grounding system.

Symbols for certain types of information and graphics


Permitted
Procedures, processes or actions that are permitted
Forbidden
Procedures, processes or actions that are forbidden
Tip
Indicates additional information

Reference to documentation

Reference to graphic

Notice or individual step to be observed


1. , 2. , 3.
Series of steps

Result of a step
1, 2, 3, ...
Item numbers
A, B, C, ...
Views
Temperature resistance of the connection cables
Specifies the minimum value of the temperature resistance of the connection cables

4 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Function and system design


Measuring principle General principles
The Levelflex is a "downward-looking" measuring system that functions according to the time-of-
flight method (ToF). The distance from the reference point to the product surface is measured. High-
frequency pulses are injected to a probe and led along the probe. The pulses are reflected by the
product surface, received by the electronic evaluation unit and converted into level information. This
method is also known as TDR (time domain reflectometry).

20 mA
100%
D
E
LN F

4 mA
0%

A0011360

1 Parameters for level measurement with guided wave radar


LN Probe length
D Distance
L Level
R Reference point of measurement
E Empty calibration (= zero)
F Full calibration (= span)

If the εr value is less than 7 in the case of rope probes, measurement is not possible in the area
of the tensioning weight (0 to 250 mm (0 to 9.84 in) from the probe end), (lower blocking
distance).
The reference point R of the measurement is located at the process connection.

Endress+Hauser 5
Levelflex FMP51, FMP52, FMP54

Dielectric constant
The dielectric constant (DC) of the medium directly affects the degree of reflection of the high-
frequency pulses. In the case of large DC values, such as with water or ammonia, there is strong
pulse reflection while, in the case of low DC values, such as with hydrocarbons, pulse reflection is
weak.

Input
The reflected pulses are transmitted from the probe to the electronics. Here, a microprocessor
evaluates the signals and identifies the level echo which was caused by the reflection of the high-
frequency pulses at the product surface. This clear signal detection system benefits from over 30
years of experience with pulse time-of-flight procedures that have gone into the development of the
PulseMaster® software.
The distance D to the product surface is proportional to the time-of-flight t of the pulse:
D = c · t/2,
with c being the speed of light.
Based on the known empty distance E, the level L is calculated:
L=E–D
The Levelflex has functions for interference echo suppression that can be activated by the user. They
guarantee that interference echoes from internal fixtures and struts, for example, are not interpreted
as level echoes.

Output
The Levelflex is preadjusted at the factory to the probe length ordered so that in most cases only the
application parameters that automatically adapt the device to the measuring conditions need to be
entered. For models with a current output, the factory adjustment for zero point E and span F is 4
mA and 20 mA, for digital outputs and the display module 0 % and 100 %. A linearization function
with max. 32 points, based on a table entered either manually or semi-automatically, can be
activated locally or remotely. This function allows the level to be converted into units of volume or
mass, for example.

6 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Interface measurement
When the high-frequency pulses hit the surface of the medium, only a percentage of the
transmission pulse is reflected. In the case of medium with a low DC1, in particular, the other part
enters the medium. The pulse is reflected once more at the interface point to a second medium with
a higher DC2. The distance to the interface layer can now also be determined, taking into account the
delayed time-of-flight of the pulse through the upper medium.

c = c0 = 300 000 km/s

LL

C0
DK1 (DC1) C1 =
DK1

LI

DK2 (DC2)

A0011178

2 Interface measurement with the guided radar


LL Total level
LI Interface level
R Reference point of measurement

In addition, the following general conditions must be observed for interface measurement:
• The DC of the upper medium must be known and constant 1). The DC can be determined with the
aid of the DC manual CP00019F or the "DC Values App". In addition, if the interface thickness is
known, the DC can be calculated automatically in FieldCare.
• The DC of the upper medium may not exceed 10.
• The DC difference between the upper and lower medium must be >10.
• The minimum thickness of the upper medium is 60 mm (2.4 in).
• Emulsion layers in the area of the interface can greatly attenuate the signal. However, emulsion
layers up to 50 mm (2 in) are permitted.
For the dielectric constants (DC values) of many media commonly used in industry, please refer
to:
• Dielectric constant (DC value) Compendium CP01076F
• The Endress+Hauser "DC Values app" (available for Android and iOS)

1) For FMP55: Under certain conditions measurement is possible even with a changing DC. For such cases, please contact your
Endress+Hauser sales organization.

Endress+Hauser 7
Levelflex FMP51, FMP52, FMP54

Life cycle of the product


Design
• Universal measuring principle
• Measurement is unaffected by product properties
• Hardware and software developed according to SIL IEC 61508
• Genuine, direct interface measurement
Procurement
• As the global market leader in level measurement, Endress+Hauser guarantees the security of
your investment
• Worldwide support and service
Installation
• No special tools are required
• Protection against reverse polarity
• Modern, detachable terminals
• Main electronics protected by a separate connection compartment
Commissioning
• Fast, menu-guided commissioning in just 6 steps
• Plain text display in local language reduces the risk of error or confusion
• Direct local access to all parameters
• Printed Brief Operating Instructions in the device onsite
Operation
• Multi-echo tracking: Reliable measurement thanks to self-learning echo search algorithms taking
into account the short-term and long-term history and plausibility of the detected signals to
suppress interference echoes.
• In accordance with NAMUR NE107
Maintenance
• HistoROM: Data backup for device settings and measured values
• Exact device and process diagnostics to assist fast decisions with clear information regarding
remedial action
• Intuitive, menu-guided operating concept in local language saves costs for training, maintenance
and operation
• Cover of the electronics compartment can also be opened in the hazardous area
Retirement
• Order code translation for subsequent models
• RoHS-compliant (Restriction of certain Hazardous Substances), unleaded soldering of electronic
components
• Environmentally friendly recycling approach

Measuring system General notes on probe selection


• Rod probes or coax probes should normally be used for liquids. Rope probes are used in liquids for
measuring ranges > 10 m (33 ft) (for FMP52: > 4 m (13 ft)) or if the distance from the ceiling
does not permit the installation of rigid probes.
• For interface measurement, ideally coax probes or rod probes are used in the bypass/stilling well.
• Coax probes are suitable for liquids with a viscosity of up to approx. 500 cst. The vast majority of
liquefied gases can be measured with coax probes, from a dielectric constant of 1.4. Furthermore,
installation conditions, such as nozzles, internal fixtures in the tank etc., have no effect on the
measurement when a coax probe is used. A coax probe offers maximum EMC safety when used in
plastic tanks.

Probe selection

FMP51
For level measurement and interface measurement in liquids

8 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

A0011387

3 Rod probe

Rod probe
• Maximum probe length
• 4 m (13 ft); inseparable rod probes
• 10 m (33 ft); separable rod probes
• Material:
• 316L; inseparable and separable rod probes
• Alloy C; only inseparable rod probes

A0011388

4 Rope probe with centering rod

Rope probe
• Maximum probe length
45 m (148 ft)
• Material:
• 316L
• Alloy C
• PFA>316L

A0011359

5 Coax probe

Endress+Hauser 9
Levelflex FMP51, FMP52, FMP54

Coax probe
• Maximum probe length
6 m (20 ft)
• Material:
• 316L, multiple holes
• Alloy C, one hole

FMP52
For level measurement and interface measurement in corrosive liquids

A0011357

6 Rod probe

Rod probe
• Maximum probe length
4 m (13 ft)
• Material:
PFA>316L

A0011358

7 Rope probe with centering rod

Rope probe
• Maximum probe length
45 m (148 ft)
• Material:
PFA>316L

FMP54
For level measurement and interface measurement in liquids

10 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

A0011357

8 Rod probe

Rod probe
• Maximum probe length
4 m (13 ft)
• Material:
316L

A0011358

9 Rope probe with centering rod

Rope probe
• Maximum probe length
45 m (148 ft)
• Material:
316L

A0011359

 10 Coax probe

Endress+Hauser 11
Levelflex FMP51, FMP52, FMP54

Coax probe
• Maximum probe length
6 m (20 ft)
• Material:
316L, multiple holes

12 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Input
Measured variable The measured variable is the distance between the reference point and the product surface.
The level is calculated based on "E", the empty distance entered.
Optionally, the level can be converted to other variables (volume, mass) by linearization (32 points).

Measuring range The following table describes the medium groups and the possible measuring range as a function of
the medium group.

Levelflex FMP51, FMP54

Measuring range 1)
Medium
εr Typical liquids Bare metallic Bare metallic PFA-coated Coax probes
group
Rod probes Rope probes Rope probes

1 1.4 to 1.6 Liquefied gases, e.g. N2, CO2 On request — 6 m (20 ft)

2 1.6 to 1.9 • Liquefied gas, e.g. propane • One-piece: 4 m (13 ft) 15 to 22 m 9 to 14 m 6 m (20 ft)
• Solvents • Separable: 10 m (33 ft) (49 to 72 ft) (30 to 46 ft)
• Freon
• Palm oil

3 1.9 to 2.5 Mineral oils, fuels • One-piece: 4 m (13 ft) 22 to 32 m 14 to 21 m 6 m (20 ft)
• Separable: 10 m (33 ft) (72 to 105 ft) (46 to 69 ft)

4 2.5 to 4 • Benzene, styrene, toluene • One-piece: 4 m (13 ft) 32 to 42 m 21 to 28 m 6 m (20 ft)


• Furan • Separable: 10 m (33 ft) (105 to 138 ft) (69 to 92 ft)
• Naphthalene

5 4 to 7 • Chlorobenzene, chloroform • One-piece: 4 m (13 ft) 42 to 45 m 28 to 32 m 6 m (20 ft)


• Nitrocellulose lacquers • Separable: 10 m (33 ft) (138 to 148 ft) (92 to 105 ft)
• Isocyanate, aniline

6 >7 • Aqueous solutions • One-piece: 4 m (13 ft) 45 m (148 ft) 32 to 45 m 6 m (20 ft)
• Alcohols • Separable: 10 m (33 ft) (105 to 148 ft)
• Ammonia

1) The measuring range for interface measurements is limited to 10 m (33 ft).

Endress+Hauser 13
Levelflex FMP51, FMP52, FMP54

Levelflex FMP52
1)
Measuring range
Medium group εr Typical liquids PFA-coated PFA-coated
Rod probes Rope probes

1 1.4 to 1.6 Liquefied gases, e.g. N2, CO2 — —

2 1.6 to 1.9 • Liquefied gas, e.g. propane 4 m (13 ft) 9 to 14 m (30 to 46 ft)
• Solvents
• Freon
• Palm oil

3 1.9 to 2.5 Mineral oils, fuels 4 m (13 ft) 14 to 21 m (46 to 69 ft)

4 2.5 to 4 • Benzene, styrene, toluene 4 m (13 ft) 21 to 28 m (69 to 92 ft)


• Furan
• Naphthalene

5 4 to 7 • Chlorobenzene, chloroform 4 m (13 ft) 28 to 32 m (92 to 105 ft)


• Nitrocellulose lacquers
• Isocyanate, aniline

6 >7 • Aqueous solutions 4 m (13 ft) 32 to 45 m (105 to 148 ft)


• Alcohols
• Acids, alkalis

1) The measuring range for interface measurement is limited to 10 m.

• The formation of buildup, particularly of moist products, can reduce the maximum possible
measuring range.
• Due to the high rate of diffusion of ammonia, a gas-tight bushing 2) is recommended for
measurements in this medium.

2) Always included in FMP54, optionally available for FMP51/52

14 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Blocking distance The upper blocking distance (= UB) is the minimum distance from the reference point of the
measurement (mounting flange) to the maximum level.

R
SD UB
100%
E

F
LN

0%

A0011279

 11 Definition of blocking distance and safety distance


R Reference point of measurement
LN Probe length
UB Upper blocking distance
E Empty calibration (= zero)
F Full calibration (= span)
SD Safety distance

Blocking distance (factory setting):


• For coax probes: 0 mm (0 in)
• For rod and rope probes up to 8 m (26 ft): 200 mm (8 in)
• For rod and rope probes over 8 m (26 ft): 0.025 × probe length
The specified blocking distances are factory-preset on delivery. These settings can be changed
depending on the application.
For rod and rope probes, the blocking distance can generally be reduced to 100 mm (4 in) for
media with DC > 7.
No blocking distance applies for applications in a bypass/stilling well.
A reliable measurement cannot be guaranteed within the blocking distance.
A safety distance SD can be defined in addition to the blocking distance. The device generates a
warning if the level rises to enter this safety distance.

Measuring frequency 100 MHz to 1.5 GHz


spectrum

Endress+Hauser 15
Levelflex FMP51, FMP52, FMP54

Output
Output signal HART
• Signal coding:
FSK ±0.5 mA over current signal
• Data transmission rate:
1 200 Bit/s
• Galvanic isolation:
Yes
Bluetooth® wireless technology
• Device version:
Order code 610 "Accessory mounted", option NF "Bluetooth"
• Operation / configuration:
Via the SmartBlue app
• Range under reference conditions:
> 10 m (33 ft)
• Encryption:
Encrypted communication and password encryption prevent incorrect operation by unauthorized
persons
PROFIBUS PA
• Signal coding:
Manchester Bus Powered (MBP)
• Data transmission rate:
31.25 kBit/s, Voltage mode
• Galvanic isolation:
Yes
FOUNDATION Fieldbus
• Signal coding:
Manchester Bus Powered (MBP)
• Data transmission rate:
31.25 kBit/s, Voltage mode
• Galvanic isolation:
Yes

Switch output
For HART devices, the switch output is available as an option.

• Function:
Open collector switch output
• Switching behavior:
Binary (conductive or non-conductive), switches when the programmable switch-on point/switch-
off point is reached
• Failure mode:
Non-conductive
• Electrical connection data:
U = 16 to 35 VDC, I = 0 to 40 mA
• Internal resistor:
RI < 880 Ω
The voltage drop at this internal resistor must be taken into account when planning the
configuration. For example, the resulting voltage at a connected relay must be sufficient to switch
the relay.
• Insulation voltages:
Floating, insulation voltage 1 350 VDC in relation to power supply and 500 VAC ground
• Switch point:
User-programmable, separate for switch-on point and switch-off point
• Switching delay:
User-programmable in the 0 to 100 s range, separate for switch-on point and switch-off point
• Scan rate:
Corresponds to the measuring cycle

16 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

• Signal source / device variables:


• Level linearized
• Distance
• Terminal voltage
• Electronic temperature
• Relative echo amplitude
• Diagnostic values, advanced diagnostic blocks
• Only for active interface measurement
• Signal source / device variables for active interface measurement:
• Interface linearized
• Interface distance
• Upper interface distance
• Relative interface amplitude
• Number of switch cycles:
Unlimited

Signal on alarm Depending on the interface, failure information is displayed as follows:


• Current output
• Choice of failure mode (in accordance with NAMUR Recommendation NE 43):
Minimum alarm: 3.6 mA
Maximum alarm (= factory setting): 22 mA
• Failure mode with user-configurable value: 3.59 to 22.5 mA
• Local display
• Status signal (as per NAMUR Recommendation NE 107)
• Plain text display
• Operating tool via digital communication (HART, PROFIBUS PA, FOUNDATION Fieldbus) or
service interface (CDI)
• Status signal (as per NAMUR Recommendation NE 107)
• Plain text display

Linearization The device's linearization function allows the user to convert the measured value to any length or
volume units. Linearization tables for calculating the volume in cylindrical vessels are
preprogrammed into the device. Other tables of up to 32 value pairs can be entered manually or
semi-automatically.

Galvanic isolation All circuits for the outputs are galvanically isolated from each other.

Endress+Hauser 17
Levelflex FMP51, FMP52, FMP54

Protocol-specific data HART

Manufacturer ID 17 (0x11)

Device type ID 0x1122

HART specification 7.0

Device description files (DTM, DD) Information and files under:


• www.endress.com
• www.fieldcommgroup.org

HART load min. 250 Ω

HART device variables The measured values can be freely assigned to the device variables.
Measured values for PV (primary variable)
• Level linearized
• Distance
• For active interface measurement:
• Interface
• Interface distance
• Upper interface thickness
• Relative interface amplitude
• Electronic temperature
• Relative echo amplitude
Measured values for SV, TV, QV (second, third and fourth variable)
• Level linearized
• Distance
• For active interface measurement:
• Interface linearized
• Interface distance
• Upper interface thickness
• Absolute interface amplitude
• Relative interface amplitude
• Terminal voltage
• Electronic temperature
• Absolute echo amplitude
• Relative echo amplitude
• Calculated εr value

Supported functions • Burst mode


• Additional transmitter status

Wireless HART data

Minimum starting voltage 17.5 V

Start-up current 4 mA

Starting time 80 s

Minimum operating voltage 17.5 V

Multidrop current 4.0 mA

Time for connection setup 30 s

18 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

PROFIBUS PA

Manufacturer ID 17 (0x11)

Ident number 0x1558

Profile version 3.02

GSD file Information and files under:


• www.endress.com
GSD file version • www.profibus.org

Output values Analog Input:


• Level linearized
• Distance
• For active interface measurement:
• Interface
• Interface distance
• Upper interface thickness
• Absolute interface amplitude
• Absolute interface amplitude
• Terminal voltage
• Electronic temperature
• Absolute echo amplitude
• Relative echo amplitude
• Calculated εr value
Digital Input:
• Extended diagnostic blocks
• Status output PFS Block

Input values Analog Output:


• Analog value from PLC (for sensor block, external pressure and temperature)
• Analog value from PLC to be indicated on the display
Digital Output:
• Extended diagnostic block
• Level Limiter
• Sensor Block Measurement On
• Sensor Block Save History On
• Status output

Supported functions • Identification & Maintenance


Simple device identification via control system and nameplate
• Automatic Ident Number Adoption
GSD compatibility mode with predecessor Levelflex M FMP4x
• Physical Layer Diagnostics
Installation check of the PROFIBUS segment and the Levelflex M FMP4x using
terminal voltage and message monitoring
• PROFIBUS upload/download
Reading and writing parameters is up to ten times faster with PROFIBUS upload/
download
• Condensed status
Simplest and self-explanatory diagnostic information by categorizing diagnostic
messages that occur

FOUNDATION Fieldbus

Manufacturer ID 0x452B48

Device type 0x1028

Device revision 0x01

DD revision Information and files under:


• www.endress.com
CFF revision • www.fieldcommgroup.org

Device Tester Version (ITK version) 6.0.1

ITK Test Campaign Number IT085300

Link Master capability (LAS) Yes

Endress+Hauser 19
Levelflex FMP51, FMP52, FMP54

Choice of "Link Master" and "Basic Device" Yes; Factory setting: Basic Device

Node address Factory setting: 247 (0xF7)

Supported functions The following methods are supported:


• Restart
• ENP Restart
• Setup
• Linearization
• Self Check

Virtual Communication Relationships (VCRs)

Number of VCRs 44

Number of link objects in VFD 50

Permanent entries 1

Client VCRs 0

Server VCRs 10

Source VCRs 43

Sink VCRs 0

Subscriber VCRs 43

Publisher VCRs 43

Device Link Capabilities

Slot time 4

Min. delay between PDU 8

Max. response delay 20

Transducer Blocks
Block Contents Output values

Setup Transducer Block Contains all the parameters for standard • Level or volume (channel
commissioning 1)
(depending on the block
configuration)
• Distance (channel 2)

Advanced Setup Contains all the parameters for more accurate No output values
Transducer Block measurement configuration

Display Transducer Block Contains parameters to configure the onsite display No output values

Diagnostic Transducer Contains diagnostic information No output values


Block

Advanced Diagnostic Contains parameters for advanced diagnostics No output values


Transducer Block

Expert Configuration Contains parameters that require the user to have No output values
Transducer Block in-depth knowledge of the operation of the device in
order to configure the parameters appropriately

Expert Information Contains parameters that provide information about No output values
Transducer Block the state of the device

Service Sensor Contains parameters that can only be accessed by No output values
Transducer Block Endress+Hauser Service

Service Information Contains parameters that provide Endress+Hauser No output values


Transducer Block Service with information about the state of the
device

Data Transfer Contains parameters for backing up the device No output values
Transducer Block configuration in the display module and for writing
the saved configuration to the device. Access to
these parameters is reserved for Endress+Hauser
Service.

20 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Function blocks
Block Contents Number of Number of Execution Functionality
permanent instantiatable time
blocks blocks

Resource Block The Resource Block contains all 1 0 - Enhanced


the data that uniquely identify
the device. It is an electronic
version of a nameplate of the
device.

Analog Input The AI Block receives the 2 3 25 ms Enhanced


Block measuring data from the Sensor
Block (selectable via a channel
number), and makes the data
available to other blocks at its
output.

Discrete Input The Discrete Input Block receives 1 2 20 ms Standard


Block a discrete value (e.g. indicator
that measuring range has been
exceeded) and makes the value
available for other blocks at the
output.

Multiple The Multiple Analog Output 1 0 20 ms Standard


Analog Output Block is used to transmit analog
Block values from the bus to the device.

Multiple The Multiple Discrete Output 1 0 20 ms Standard


Discrete Block is used to transmit discrete
Output Block values from the bus to the device.

PID Block The PID Block is used as a 1 1 25 ms Standard


proportional-integral-derivative
controller and can be used
universally for closed-loop-
control in the field. It enables
cascade mode and feedforward
control.

Arithmetic The Arithmetic Block is designed 1 1 25 ms Standard


Block to permit simple use of popular
measurement math functions.
The user does not have to know
how to write equations. The
math algorithm is selected by
name, chosen by the user for the
function to be performed.

Signal The Signal Characterizer Block 1 1 25 ms Standard


Characterizer has two sections, each with an
Block output value that is a non-linear
function of the input value. The
non-linear function is generated
by a single look-up table with 21
arbitrary x-y pairs.

Input Selector The Input Selector Block 1 1 25 ms Standard


Block facilitates the selection of up to
four inputs and generates an
output value based on the
configured action. This block
normally receives its inputs from
AI Blocks. The block enables the
selection of maximum, minimum,
average and ‘first good’ values.

Endress+Hauser 21
Levelflex FMP51, FMP52, FMP54

Block Contents Number of Number of Execution Functionality


permanent instantiatable time
blocks blocks

Integrator The Integrator Block integrates a 1 1 25 ms Standard


Block variable as a function of the time
or accumulates the counts from a
Pulse Input Block. The Block can
be used as a totalizer that
totalizes until a reset, or as a
batch totalizer whereby the
integrated value is compared
against a target value generated
before or during the control
routine and generates a binary
signal when the target value is
reached.

Analog Alarm 1 1 25 ms Standard


Block

Up to 20 blocks can be instantiated in the device altogether, including the blocks already
instantiated.

22 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Power supply
Terminal assignment
Terminal assignment, 2-wire: 4 to 20 mA HART

A B 2

1 +
+ 2 1 -
- 2

3 3

A0036498

 12 Terminal assignment, 2-wire: 4 to 20 mA HART


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection 4 to 20 mA, HART passive: terminals 1 and 2, without integrated overvoltage protection
2 Connection 4 to 20 mA, HART passive: terminals 1 and 2, with integrated overvoltage protection
3 Terminal for cable shield

Block diagram, 2-wire: 4 to 20 mA HART

3 4
2 5
1 6
+ +1
- -2 I

A0036499

 13 Block diagram, 2-wire: 4 to 20 mA HART


1 Active barrier for power supply (e.g. RN221N); observe terminal voltage
2 Resistor for HART communication (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display unit; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device

Terminal assignment, 2-wire: 4 to 20 mA HART, switch output

A B 3
3 + 4
+ 4 3 -
1 4 +
-
+ 2 1 -
- 2

2
1 5 5

A0036500

 14 Terminal assignment, 2-wire: 4 to 20 mA HART, switch output


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection 4 to 20 mA, HART passive: terminals 1 and 2, without integrated overvoltage protection
2 Connection, switch output (open collector): terminals 3 and 4, without integrated overvoltage protection
3 Connection, switch output (open collector): terminals 3 and 4, with integrated overvoltage protection
4 Connection 4 to 20 mA, HART passive: terminals 1 and 2, with integrated overvoltage protection
5 Terminal for cable shield

Endress+Hauser 23
Levelflex FMP51, FMP52, FMP54

Block diagram, 2-wire: 4 to 20 mA HART, switch output

3 4
2 5
1 6
+ +1
- -2 I
+3 Y
-4

7
A0036501

 15 Block diagram, 2-wire: 4 to 20 mA HART, switch output


1 Active barrier for power supply (e.g. RN221N); observe terminal voltage
2 Resistor for HART communication (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display unit; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device
7 Switch output (open collector)

Terminal assignment, 2-wire: 4 to 20 mA HART, 4 to 20 mA

A B 3
3 + 4
+ 4 3 -
1 4 +
-
+ 2 1 -
- 2

2
1 5 5

A0036500

 16 Terminal assignment, 2-wire: 4 to 20 mA HART, 4 to 20 mA


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection current output 1, 4 to 20 mA HART passive: terminals 1 and 2, without integrated overvoltage
protection
2 Connection current output 2, 4 to 20 mA: terminals 3 and 4, without integrated overvoltage protection
3 Connection current output 2, 4 to 20 mA: terminals 3 and 4, with integrated overvoltage protection
4 Connection current output 1, 4 to 20 mA HART passive: terminals 1 and 2, with integrated overvoltage
protection
5 Terminal for cable shield

24 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Block diagram, 2-wire: 4 to 20 mAHART, 4 to 20 mA

3 4
2 5
1 6
+ +1
- -2 I
+ +3 Y
- -4

7
8
A0036502

 17 Block diagram, 2-wire: 4 to 20 mAHART, 4 to 20 mA


1 Active barrier for power supply (e.g. RN221N), current output 1; observe terminal voltage
2 Resistor for HART communication (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display unit; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device
7 Analog display unit; observe maximum load
8 Active barrier for power supply (e.g. RN221N), current output 2; observe terminal voltage

Terminal assignment, 4-wire: 4 to 20 mA HART (10.4 to 48 VDC)

1
L+ 2 3
L-
+ 4
-

2
1 3

A0036516

 18 Terminal assignment, 4-wire: 4 to 20 mA HART (10.4 to 48 VDC)

1 Connection 4 to 20 mA HART (active): terminals 3 and 4


2 Connection, supply voltage: terminals 1 and 2
3 Terminal for cable shield

Block diagram, 4-wire: 4 to 20 mA HART (10.4 to 48 VDC)

3 4
2 5
1 6
+3
-4 I
L+ 1 Y
L- 2

7
A0036526

 19 Block diagram, 4-wire: 4 to 20 mA HART (10.4 to 48 VDC)

1 Evaluation unit, e.g. PLC


2 Resistor for HART communication (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display unit; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device
7 Supply voltage; observe terminal voltage, observe cable specification

Endress+Hauser 25
Levelflex FMP51, FMP52, FMP54

Terminal assignment, 4-wire: 4 to 20 mA HART (90 to 253 VAC)

1
L 2 3
N
+ 4
-

2
1 3

A0036519

 20 Terminal assignment, 4-wire: 4 to 20 mAHART (90 to 253 VAC)

1 Connection 4 to 20 mA HART (active): terminals 3 and 4


2 Connection, supply voltage: terminals 1 and 2
3 Terminal for cable shield

LCAUTION
To ensure electrical safety:
‣ Do not disconnect the protective ground connection.
‣ Disconnect the device from the supply voltage before disconnecting the protective ground.
Connect protective ground to the inner ground terminal (3) before connecting the supply
voltage. If necessary, connect the potential matching line to the outer ground terminal.
In order to ensure electromagnetic compatibility (EMC): do not ground the device exclusively
via the protective ground conductor of the supply cable. Instead, the functional grounding must
also be connected to the process connection (flange or threaded connection) or to the external
ground terminal.
An easily accessible power switch must be installed in the proximity of the device. The power
switch must be marked as a disconnector for the device (IEC/EN61010).

Block diagram, 4-wire: 4 to 20 mA HART (90 to 253 VAC)

3 4
2 5
1 6
+3
-4 I
L1 Y
N2

7
A0036527

 21 Block diagram, 4-wire: 4 to 20 mA HART (90 to 253 VAC)

1 Evaluation unit, e.g. PLC


2 Resistor for HART communication (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display unit; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device
7 Supply voltage; observe terminal voltage, observe cable specification

26 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus

A B 3
3 + 4
+ 4 3 -
1 4 +
-
+ 2 1 -
- 2

2
1 5 5

A0036500

 22 Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection, PROFIBUS PA / FOUNDATION Fieldbus: terminals 1 and 2, without integrated overvoltage
protection
2 Connection, switch output (open collector): terminals 3 and 4, without integrated overvoltage protection
3 Connection, switch output (open collector): terminals 3 and 4, with integrated overvoltage protection
4 Connection, PROFIBUS PA / FOUNDATION Fieldbus: terminals 1 and 2, with integrated overvoltage
protection
5 Terminal for cable shield

Block diagram PROFIBUS PA / FOUNDATION Fieldbus

2 3
+ 1
I
- 2

1 + 3 Y
- 4

4
A0036530

 23 Block diagram PROFIBUS PA / FOUNDATION Fieldbus


1 Cable screen; observe cable specification
2 Connection PROFIBUS PA / FOUNDATION Fieldbus
3 Measuring device
4 Switch output (open collector)

Endress+Hauser 27
Levelflex FMP51, FMP52, FMP54

Connection examples for the switch output


For HART devices, the switch output is available as an option.

3+
Ri +
-
4-
A0015909

 24 Connection of a relay

+
1
2
3+
Ri

4-
A0015910

 25 Connection to a digital input


1 Pull-up resistor
2 Digital input

For optimum interference immunity we recommend to connect an external resistor (internal


resistance of the relay or pull-up resistor) of < 1 000 Ω.

28 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

HART loop converter HMX50


The dynamic variables of the HART protocol can be converted into individual 4 to 20 mA sections
using the HART loop converter HMX50. The variables are assigned to the current output and the
measuring ranges of the individual parameters are defined in the HMX50.

HART

+1 10
+2 11 HMX50
12
-3
16
4
17
5 18
6
7
8
9

13
14

15

19
20

21
FXA195

22 23 24 HART

24V DC
A0023287

 26 Connection diagram for HART loop converter HMX50 (example: passive 2-wire device and current outputs
connected as power source)

The HART loop converter HMX50 can be acquired using the order number 71063562.
Additional documentation: TI00429F and BA00371F.

Device plug In the case of the device versions with a plug, the housing does not need to be opened to
connect the signal cable.

4 3

1 2

A0011175

 27 Pin assignment of M12 plug


1 Signal +
2 Not assigned
3 Signal -
4 Ground

Endress+Hauser 29
Levelflex FMP51, FMP52, FMP54

1 3

2 4

A0011176

 28 Pin assignment of 7/8" plug


1 Signal -
2 Signal +
3 Not assigned
4 Shielding

Supply voltage An external power supply is necessary.


Various power supply units can be ordered from Endress+Hauser.

2-wire, 4-20mA HART, passive

2-wire; 4-20mA HART 1)


"Approval" 2) Terminal voltage U at device Maximum load R, depending on the supply voltage U0 of the
power supply unit

• Non-hazardous 11.5 to 35 V 3) 4)
R [W]
• Ex nA
• Ex ic 500
• CSA GP

Ex ia / IS 11.5 to 30 V 4)

0
10 20 30 35 U0 [V]
11.5 22.5
A0035511

4) 5)
• Ex d / XP 13.5 to 30 V R [Ω]
• Ex ic[ia]
• Ex tD / DIP
500

0
U0 [V]
10 20 30
13.5 24.5
A0034969

1) Feature 020 in the product structure: option A


2) Feature 010 in the product structure
3) At ambient temperatures Ta ≤ -30 °C, a terminal voltage U ≥ 14 V is required to start the device with the minimum failure current (3.6 mA). At
ambient temperatures Ta> 60 °C, a terminal voltage U ≥ 12 V is required to start the device with the minimum failure current (3.6 mA). The
start-up current can be configured. If the device is operated with a fixed current I ≥ 4.5 mA (HART Multidrop mode), a voltage U ≥ 11.5 V in the
entire ambient temperature range suffices.
4) If the Bluetooth module is used, the minimum supply voltage increases by 2 V.
5) At ambient temperatures Ta ≤ -30 °C, a terminal voltage U ≥ 16 V is required to start the device with the minimum failure current (3.6 mA).

30 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

2-wire; 4-20 mA HART, switch output 1)


"Approval" 2) Terminal voltage U at device Maximum load R, depending on the supply voltage U0 of the
power supply unit

• Non-hazardous 13.5 to 35 V 3) 4) R [Ω]


• Ex nA
• Ex nA(ia)
500
• Ex ic
• Ex ic[ia]
• Ex d[ia] / XP
• Ex ta / DIP
• CSA GP

• Ex ia / IS 13.5 to 30 V 3) 4)
• Ex ia + Ex d[ia] / IS + XP 0
10 20 30 U0 [V]
13.5 24.5 35
A0034971

1) Feature 020 in the product structure: option B


2) Feature 010 in the product structure
3) At ambient temperatures Ta ≤ -30 °C, a terminal voltage U ≥ 16 V is required to start the device with the minimum failure current (3.6 mA).
4) If the Bluetooth module is used, the minimum supply voltage increases by 2 V.

2-wire; 4-20mA HART, 4-20mA 1)


"Approval" 2) Terminal voltage U at device Maximum load R, depending on the supply voltage U0 of the
power supply unit

All Channel 1:

13.5 to 30 V 3) 4) 5) R [Ω]

500

0
U0 [V]
10 20 30
13.5 24.5
A0034969

Channel 2:

12 to 30 V R [W]

500

0
10 20 30 U0 [V]
12 23
A0022583

1) Feature 020 in the product structure: option C


2) Feature 010 in the product structure
3) At ambient temperatures Ta ≤ -30 °C, a terminal voltage U ≥ 16 V is required to start the device with the minimum failure current (3.6 mA).
4) At ambient temperatures Ta ≤ -40 °C, the maximum terminal voltage must be limited to U ≤ 28 V.
5) If the Bluetooth module is used, the minimum supply voltage increases by 2 V.

Endress+Hauser 31
Levelflex FMP51, FMP52, FMP54

Integrated polarity Yes


reversal protection

Permitted residual ripple USS < 1 V


with f = 0 to 100 Hz

Permitted residual ripple USS < 10 mV


with f = 100 to 10000 Hz

32 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

4-wire, 4-20mA HART, active

"Power supply; output" 1) Terminal voltage U Maximum load Rmax

K: 4-wire 90-253VAC; 4-20mA HART 90 to 253 VAC (50 to 60 Hz), overvoltage 500 Ω
category II

L: 4-wire 10.4-48VDC; 4-20mA HART 10.4 to 48 VDC

1) Feature 020 in the product structure

PROFIBUS PA, FOUNDATION Fieldbus

"Power supply; output" 1) "Approval" 2) Terminal voltage


3)
E: 2-wire; FOUNDATION Fieldbus, switch output • Non-hazardous 9 to 32 V
G: 2-wire; PROFIBUS PA, switch output • Ex nA
• Ex nA[ia]
• Ex ic
• Ex ic[ia]
• Ex d[ia] / XP
• Ex ta / DIP
• CSA GP
3)
• Ex ia / IS 9 to 30 V
• Ex ia + Ex d[ia] / IS + XP

1) Feature 020 in the product structure


2) Feature 010 in the product structure
3) Input voltages up to 35 V do not destroy the device.

Polarity-dependent No

FISCO/FNICO compliant Yes


according to IEC 60079-27

Power consumption "Power supply; Output" 1) Power consumption

A: 2-wire; 4-20mA HART < 0.9 W

B: 2-wire; 4-20mA HART, switch output < 0.9 W

C: 2-wire; 4-20mA HART, 4-20mA < 2 x 0.7 W

K: 4-wire 90-253VAC; 4-20mA HART 6 VA

L: 4-wire 10,4-48VDC; 4-20mA HART 1.3 W

1) Feature 020 of the product structure

Current consumption HART

Nominal current 3.6 to 22 mA, the start-up current for multidrop mode can be parametrized (is
set to 3.6 mA on delivery)

Breakdown signal adjustable: 3.59 to 22.5 mA


(NAMUR NE43)

PROFIBUS PA

Nominal current 14 mA

Failure current FDE (Fault 0 mA


Disconnection Electronic)

Endress+Hauser 33
Levelflex FMP51, FMP52, FMP54

FOUNDATION Fieldbus

Device basic current 15 mA

Failure current FDE (Fault 0 mA


Disconnection Electronic)

FISCO

Ui 17.5 V

Ii 550 mA

Pi 5.5 W

Ci 5 nF

Li 10 μH

Power supply failure • Configuration is retained in the HistoROM (EEPROM).


• Error messages (incl. value of operated hours counter) are stored.

Potential equalization No special measures for potential equalization are required.


If the device is designed for hazardous areas, observe the information in the documentation
"Safety Instructions" (XA).

Terminals • Without integrated overvoltage protection


Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• With integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)

Cable entries Connection of the power supply and signal cables


To be selected in feature 050 "Electrical connection":
• Coupling M20, material depends on approval:
• For non-Ex, ATEX, IECEx, NEPSI Ex ia/ic:
Plastic M20x1.5 for cable ⌀5 to 10 mm (0.2 to 0.39 in)
• For Dust-Ex, FM IS, CSA IS, CSA GP, Ex ec:
• For Ex db:
No cable gland available
• Thread
• ½" NPT
• G ½"
• M20 × 1.5
• M12 plug / 7/8" plug
Only available for non-Ex, Ex ic, Ex ia

Connection of remote display FHX50

Feature 030 "Display, operation" Cable entry for connection of FHX50

L: "Prepared for display FHX50 + M12 connection" M12 socket

M: "Prepared for display FHX50 + custom connection" M12 cable gland

Cable specification • Devices without integrated overvoltage protection


Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Devices with integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)
• For ambient temperature TU≥60 °C (140 °F): use cable for temperature TU +20 K.

34 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

HART
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
• For 4-wire devices: Standard device cable is sufficient for the power line.

PROFIBUS
Use a twisted, screened two-wire cable, preferably cable type A.
For further information on the cable specifications, see Operating Instructions BA00034S
"PROFIBUS DP/PA: Guidelines for planning and commissioning", PNO Guideline 2.092
"PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).

FOUNDATION Fieldbus
Endress+Hauser recommends using twisted, shielded two-wire cables.
For further information on the cable specifications, see Operating Instructions BA00013S
"FOUNDATION Fieldbus Overview", FOUNDATION Fieldbus Guideline and IEC 61158-2 (MBP).

Overvoltage protection If the measuring device is used for level measurement in flammable liquids which requires the use of
overvoltage protection according to DIN EN 60079-14, standard for
test procedures 60060-1 (10 kA, pulse 8/20 μs), an overvoltage protection module has to be
installed.

Integrated overvoltage protection module


An integrated overvoltage protection module is available for 2-wire HART as well as PROFIBUS PA
and FOUNDATION Fieldbus devices.
Product structure: Feature 610 "Accessory mounted", option NA "Overvoltage protection".

Technical data

Resistance per channel 2 × 0.5 Ω max.

Threshold DC voltage 400 to 700 V

Threshold impulse voltage < 800 V

Capacitance at 1 MHz < 1.5 pF

Nominal arrest impulse voltage (8/20 μs) 10 kA

External overvoltage protection module


HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection.
For detailed information please refer to the following documents:
• HAW562: TI01012K
• HAW569: TI01013K

Endress+Hauser 35
Levelflex FMP51, FMP52, FMP54

Performance characteristics
Reference operating • Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)
conditions • Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)
• Humidity = 60 % ±15 %
• Reflection factor ≥ 0.8 (water surface for coax probe, metal plate for rod and rope probe with min.
1 m (40 in) diameter)
• Flange for rod or rope probe ≥ 300 mm (12 in) diameter
• Distance to obstacles ≥ 1 m (40 in)
• For interface measurement:
• Coax probe
• DC of lower medium = 80 (water)
• DC of upper medium = 2 (oil)

Reference accuracy Typical data under reference operating conditions: DIN EN IEC 61298-2 / DIN EN IEC 60770-1;
percentage values in relation to the span.

Output: digital analog 1)

Accuracy (sum of non- Level measurement: ±0.02 %


linearity, non-repeatability • Measuring distance up to 15 m (49 ft): ±2 mm (±0.08 in) 3)
and hysteresis) 2) • Measuring distance > 15 m (49 ft): ±10 mm (±0.39 in)

Interface measurement:
• Measuring distance up to 500 mm (19.7 in):
±20 mm (±0.79 in)
• Measuring distance > 500 mm (19.7 in): ±10 mm (±0.39 in)
• If thickness of upper medium < 100 mm (3.94 in):
±40 mm (±1.57 in)

Non-repeatability 4) ≤1 mm (0.04 in)

1) Add error of the analog value to the digital value.


2) If the reference conditions are not met, the offset/zero point resulting from the installation conditions can
be up to ±16 mm (±0.63 in). This additional offset/zero point can be compensated for by entering a
correction ("Level correction" parameter) during commissioning.
3) For probes with centering stars, the accuracy may deviate close to the centering stars.
4) The non-repeatability is already considered in the accuracy.

36 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Differing from this, the following measured error applies in the area of the lower probe end:

D
80 (3.15)
60 (2.36) DC = 2
40 (1.57)
DC = 80
20 (0.79)
0
-20 (-0.79)
-40 (-1.57)
-60 (-2.36)
-80 (-3.15)

50 (1.97)

100 (3.94)

150 (5.91)

250 (9.84)

300 (11.8)
A

20 (0.79)

200 (7.87)
40 (1.57)

80 (3.15)
A0021480

 29 Measured error at the probe end for rod and coax probes
A Distance from the probe end [mm(in)]
D Measured error: Sum of non-linearity, non-repeatability and hysteresis

D
80 (3.15)
60 (2.36) DC = 80
40 (1.57)
20 (0.79)
0
-20 (-0.79)
-40 (-1.57)
-60 (-2.36)
-80 (-3.15)
50 (1.97)

100 (3.94)

150 (5.91)

200 (7.87)

300 (11.8)

A
0
20 (0.79)

250 (9.84)

A0021482

 30 Measured error at the probe end for rope probes


A Distance from the probe end [mm(in)]
D Measured error: Sum of non-linearity, non-repeatability and hysteresis

Endress+Hauser 37
Levelflex FMP51, FMP52, FMP54

D
120 (4.72)
100 (3.94)
80 (3.15) DC = 2
60 (2.36) DC = 80
40 (1.57)
20 (0.79)
0
-20 (-0.79)
-40 (-1.57)
-60 (-2.36)
-80 (-3.15)
-100 (-3.94)
-120 (-4.72)

50 (1.97)

150 (5.91)

300 (11.8)
0 A
30 (0.79)

100 (3.94)

200 (7.87)

250 (9.84)
A0021483

 31 Measured error at the probe end in the case of metal centering disks (product structure: feature 610
"Accessory mounted" , version OA, OB or OC)
A Distance from the probe end [mm(in)]
D Measured error: Sum of non-linearity, non-repeatability and hysteresis

If the DC value is less than 7 in the case of rope probes, measurement is not possible in the area
of the tensioning weight (0 to 250 mm from the probe end), (lower blocking distance).

38 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

The following measured error applies in the area of the upper probe end:

R
D

40 (1.57)

30 (1.18) DC = 2
DC = 80

10 (0.39)

2 (0.08)
0 (0)
- 2 (- 0.08)

- 10 (-0.39)

- 30 (- 1.18)

- 40 (- 1.57)

100 (3.94) 200 (7.87)


A0015091

 32 Measured error at the upper probe end; engineering unit: mm (in)


D Sum of non-linearity, non-repeatability and hysteresis
R Reference point of measurement
DC Dielectric constant

Resolution • Digital: 1 mm
• Analog: 1 μA

Response time The response time can be configured. The following step response times (in accordance with DIN EN
IEC 61298-2 / DIN EN IEC 60770-1) 3) are when damping is switched off:

Level measurement

Probe length Sampling rate Response time

< 10 m (33 ft) 3.6 measurements per second < 0.8 s

< 40 m (131 ft) ≥ 2.7 measurements per second <1s

Interface measurement

Probe length Sampling rate Response time

< 10 m (33 ft) ≥ 1.1 measurements per second < 2.2 s

Influence of ambient The measurements are performed according to DIN EN IEC 61298-3 / DIN EN IEC 60770-1
temperature • Digital (HART, PROFIBUS PA, FOUNDATION Fieldbus): average TK = 0.6 mm/10 K
For FMP51 and FMP52 with a remote sensor, 4) there is an additional offset error of
±0.3 mm/10K (±0.01 in/10K) per 1 m (3.3 ft) of the remote cable.
• Analog (current output):
• Zero point (4 mA): average TK = 0.02 %/10 K
• Span (20 mA): average TK = 0.05 %/10 K

3) According to DIN EN IEC 61298-2 / DIN EN IEC 60770-1, the step response time is the time that elapses after an abrupt change in the input
signal until the change in the output signal has adopted 90% of the steady-state value for the first time.
4) Product structure: feature 600, version MB, MC or MD)

Endress+Hauser 39
Levelflex FMP51, FMP52, FMP54

Influence of gas phase High pressure reduces the speed of propagation of the measuring signals in the gas/vapor above the
medium. This effect depends on the type of gas phase and its temperature. This results in a
systematic measured error that increases with increasing distance between the reference point of
the measurement (flange) and the surface of the product. The following table shows this measured
error for a few typical gases/vapors (with regard to the distance, a positive value means that too
large a distance is being measured):

Gas phase Temperature Pressure

°C °F 1 bar (14.5 psi) 10 bar (145 psi) 50 bar (725 psi) 100 bar (1450 psi) 200 bar (2900 psi) 400 bar (5800 psi)

Air 20 68 0.00 % 0.22 % 1.2 % 2.4 % 4.9 % 9.5 %

200 392 -0.01 % 0.13 % 0.74 % 1.5 % 3.0 % 6.0 %

400 752 -0.02 % 0.08 % 0.52 % 1.1 % 2.1 % 4.2 %

Hydrogen 20 68 -0.01 % 0.10 % 0.61 % 1.2 % 2.5 % 4.9 %

200 392 -0.02 % 0.05 % 0.37 % 0.76 % 1.6 % 3.1 %

400 752 -0.02 % 0.03 % 0.25 % 0.53 % 1.1 % 2.2 %

Gas phase Temperature Pressure

°C °F 1 bar 2 bar 5 bar 10 bar 20 bar 50 bar 100 bar 200 bar
(14.5 psi) (29 psi) (72.5 psi) (145 psi) (290 psi) (725 psi) (1450 psi) (2900 psi)

Water 100 212 0.26 % - - - - - - -


vapor
(saturated 120 248 0.23 % 0.50 % - - - - - -
steam) 152 306 0.20 % 0.42 % 1.14 % - - - - -

180 356 0.17 % 0.37 % 0.99 % 2.10 % - - - -

212 414 0.15 % 0.32 % 0.86 % 1.79 % 3.9 % - - -

264 507 0.12 % 0.26 % 0.69 % 1.44 % 3.0 % 9.2 % - -

311 592 0.09 % 0.22 % 0.58 % 1.21 % 2.5 % 7.1 % 19.3 % -

366 691 0.07 % 0.18 % 0.49 % 1.01 % 2.1 % 5.7 % 13.2 % 76 %

Gas phase compensation PROFIBUS and FOUNDATION Fieldbus devices can receive the signal from an external pressure
with external pressure sensor over the bus and use it to automatically perform a pressure-dependent time-of-flight
sensor (PROFIBUS PA / correction. In the case of saturated steam in the temperature range from
FOUNDATION Fieldbus) 100 to 350 °C (212 to 662 °F), this makes it possible to reduce the distance measurement error of up
to 29 % (without compensation) to below 3 % (with compensation).

40 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Gas phase compensation At high pressures and temperatures, the speed of propagation of the microwave signals in the steam
using reference signal (polar media) is reduced above the liquid to be measured. This causes the Levelflex to display a level
(option for FMP54) that is too low →  40.
An optional version of the FMP54 is available with an automatic gas phase compensation function
that corrects this measured error (feature 540: "Application packages", option EF: "Gas phase comp.
Lref = 300mm" or EG: "Gas phase comp. Lref = 550mm"). In this version, a difference in the diameter of
the probe rod generates a reference reflection at the distance Lref from the flange. This reference
reflection must be at least 150 mm above the highest level. The current speed of propagation is
measured on the basis of this shift in the reference reflection and the level value is corrected
automatically.

(11.8/21.7)
(11.8/21.7)

300/550
300/550

Lref
Lref
Dref
iD

LN
LN
max max
150 (5.91)

150 (5.91)
min min

A0014534

 33 FMP54 with reference signal for gas phase compensation; engineering unit: mm (in)

Coax probes with reference reflection can be installed in any vessel (free in the tank or in a
bypass). Coax probes are ready-mounted and adjusted at the factory and are ready for use
without any additional parameter configuration.
The use of rod probes is only recommended if the installation of a coax probe is not possible
(e.g. if the bypass diameter is very small).
Rod probes with reference reflection are only suitable for installation in stilling wells and
bypass chambers. The diameter Dref of the probe rod in the area of the reference distance Lref
must be chosen to suit the pipe internal diameter iD, see the table below. The pipe must be
cylindrical in the area of the reference distance Lref; cross-section changes, e.g. at flange
connections, may not exceed 5% of the internal diameter iD.
In addition, following installation the settings must be checked by expert staff and corrected if
necessary.

Internal diameter iD of stilling well/bypass Diameter Dref of the rod probe in the area of the reference
length Lref

40 mm (1.57") ≤ iD < 45 mm (1.77") 22 mm (0.87")

45 mm (1.77") ≤ iD < 70 mm (2.76") 25 mm (0.98")

70 mm (2.76") ≤ iD < 100 mm (3.94") 30 mm (1.18")

Endress+Hauser 41
Levelflex FMP51, FMP52, FMP54

Restrictions for coax and rod probes

Maximum probe length LN • For rod probes:


LN ≤ 4 000 mm (157 in)
• For coax probes:
LN ≤ 6 000 mm (236 in)

Minimum probe length LN LN > Lref + 200 mm (Lref + 7.7")

Reference distance Lref 300 mm (11.8") or 550 mm (21.7"), see feature 540 of the product
structure

Maximum level in relation to the flange Lref + 150 mm


sealing surface

Minimum DC value of the medium DC > 7

Area of application
Level measurements at high pressure and measuring ranges of up to a few meters in polar media
with a dielectric constant DC > 7 (e.g. water or ammonia), which would produce a large measured
error without compensation.
The measuring accuracy under reference conditions is higher the greater the reference distance Lref
and the smaller the measuring range:

45

40

35
Lref = 300 mm
30
Delta [mm]

25

20

Lref = 550 mm
15

10

5
0 500 1000 1500 2000 2500 3000 3500 4000
D [mm]

A0014535

D Distance from product surface to lower edge of the flange


Delta Measured error

In the event of fast pressure changes, an additional error may occur as the measured reference
distance is averaged with the time constant of the level measurement. Furthermore, states of non-
equilibrium - due to heating for example - may result in density gradients within the medium and
condensation of steam at the probe. As a result, level readings may vary slightly at different
locations in the vessel. These application-related influences can increase the measured error
indicated above by a factor of up to 2 to 3.

42 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mounting
Mounting requirements Suitable mounting position

1 2 3

A 4

A0012606

 34 Installation conditions for Levelflex

Spacing requirements when mounting


• Distance (A) between the vessel wall and rod and rope probes:
• For smooth metallic walls: > 50 mm (2 in)
• For plastic walls: > 300 mm (12 in) to metallic parts outside the vessel
• For concrete walls: > 500 mm (20 in), otherwise the permitted measuring range may be
reduced.
• Distance (B) between rod probes and internal fittings (3): > 300 mm (12 in)
• When using more than one Levelflex:
Minimum distance between the sensor axes: 100 mm (3.94 in)
• Distance (C) from the end of the probe to the bottom of the vessel:
• Rope probe: > 150 mm (6 in)
• Rod probe: > 10 mm (0.4 in)
• Coax probe: > 10 mm (0.4 in)
Coax probes can be mounted at any distance to the wall and internal fixtures.

Endress+Hauser 43
Levelflex FMP51, FMP52, FMP54

Additional conditions
• When mounting outdoors, a weather protection cover (1) can be used to protect the device against
extreme weather conditions.
• In metallic vessels, preferably do not mount the probe in the center of the vessel (2), as this would
lead to increased interference echoes.
If a central mounting position cannot be avoided, it is essential to perform interference echo
suppression (mapping) after commissioning the device.
• Do not mount the probe in the filling curtain (3).
• Avoid buckling the rope probe during installation or operation (e.g. as a result of product
movement against silo wall) by selecting a suitable mounting location.
In the case of freely suspended rope probes (probe end not fixed at the bottom), the distance
between the probe rope and internal fittings, which can change due to the movement of the
product, must never be less than 300 mm (12 in). Occasional contact between the probe end
weight and the cone of the vessel, however, does not influence the measurement provided that
the dielectric constant is at least DC = 1.8.
When the housing is mounted in a recess (e.g. in a concrete ceiling), observe a minimum
distance of 100 mm (4 in) between the cover of the connection compartment/electronics
compartment and the wall. Otherwise the connection compartment/electronics compartment
will not be accessible after installation.

44 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mounting under confined conditions

Mounting with remote probe


The device version with a remote probe is suitable for applications with restricted mounting space. In
this case, the electronics housing is mounted at a separate position from the probe.

A B

6 Nm 6 Nm
(4.42 lbf ft) (4.42 lbf ft)

rmin = 100 (4) C


C
rmin = 100 (4)

6 Nm 6 Nm
(4.42 lbf ft) (4.42 lbf ft)

A0014794

A Angled plug at the probe


B Angled plug at the electronics housing
C Length of the remote cable as ordered

• Product structure, feature 600 "Probe design":


• Version MB "Sensor remote, 3m cable"
• Version MC "Sensor remote, 6m cable"
• Version MD "Sensor remote, 9m cable"
• The connecting cable is included in the delivery with these versions.
Minimum bending radius: 100 mm (4 inch)
• The mounting bracket for the electronics housing is included in the delivery with these versions.
Mounting options:
• Wall mounting
• Mounting on DN32 to DN50 (1-1/4 to 2 inch) post or pipe
• The connection cable has one straight plug and one plug angled at 90°. Depending on the local
conditions the angled plug can be connected at the probe or at the electronics housing.
The probe, electronics and connection cable are mutually compatible and bear a common serial
number. Only components with the same serial number may be connected to one another.

Endress+Hauser 45
Levelflex FMP51, FMP52, FMP54

Separable probes

14
15 Nm

A0021647

The use of separable rod probes (Ø 16 mm) is advisable in confined mounting conditions (limited
distance to the ceiling).
• Max. probe length 10 m (394 in)
• Max. lateral loading capacity 30 Nm
• Probes can be separated several times, with the individual parts having the following lengths:
• 500 mm (20 in)
• 1 000 mm (40 in)
The joints between the individual rod segments are secured by the enclosed Nord Lock washers.
Install the pre-assembled washers in pairs, cam face to cam face.

46 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Notes on the mechanical load of the probe

Tensile loading capacity of rope probes

FMP51
Rope 4mm (1/6") 316
5 kN
Rope 4mm (1/6") Alloy C
5 kN
Rope 4mm (1/6") PFA>316L
1 kN

FMP52
Rope 4mm (1/6") PFA>316
2 kN

FMP54
Rope 4mm (1/6") 316
10 kN

Lateral loading capacity (flexural strength) of rod probes

FMP51
Rod 8mm (1/3") 316L
10 Nm
Rod 12mm (1/2") 316L
30 Nm
Rod 12mm (1/2") AlloyC
30 Nm
Rod 16mm (0.63") 316L separable
30 Nm

FMP52
Rod 16mm (0.63") PFA>316L
30 Nm

FMP54
Rod 16mm (0.63") 316L
30 Nm
Rod 16mm (0.63") 316L separable
30 Nm

Lateral load (bending moment) from flow conditions


The formula for calculating the bending moment M acting on the probe:
M = cw × ρ/2 × v2 × d × L × (LN - 0.5 × L)

With:
cw: coefficient of friction
ρ [kg/m3]: density of the medium
v [m/s]: flow velocity of the medium, perpendicular to the probe rod
d [m]: diameter of the probe rod
L [m]: level

Endress+Hauser 47
Levelflex FMP51, FMP52, FMP54

LN [m]: probe length

Sample calculation

Coefficient of friction cw 0.9 (assuming turbulent flow - high Reynolds number)

Density ρ [kg/m3] 1000 (e.g. water)

Probe diameter d [m] 0.008

L = LN (unfavorable conditions)
LN
v L

d
A0014175

Bending torque [M] on rod probes, diameter 8mm (1/3”)


20.0
18.0
16.0
Bending torque [Nm]

max. bending torque


14.0
12.0
10.0
8.0
6.0
4.0
2.0
0.0 v=0.5m/s
0.4 0.8 1.2 1.6 2 2.4 2.8 3.2 3.6 4 v=0.7m/s
Probe length [LN] in meters v=1.0m/s

A0014182-EN

Lateral loading capacity (flexural strength) of coax probes

FMP51
Probe Ø 21.3 mm 316L
60 Nm
Probe Ø 42.4 mm 316L
300 Nm
Probe Ø 42.4 mm AlloyC
300 Nm

FMP54
Probe Ø 42.4 mm 316L
300 Nm

48 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Information concerning the process connection


Probes are mounted on the process connection with threaded connections or flanges. If there is
the danger with this installation that the probe end moves so much that it occasionally touches
the vessel floor or cone, the probe may need to be shortened at the lower end and fixed in place.

Threaded connection

A0015121

 35 Mounting with threaded connection; flush with the vessel ceiling

Seal
The thread and the type of seal comply with DIN 3852 Part 2, screwed plug, form A.
The following types of sealing ring can be used:
• For thread G3/4": According to DIN 7603 with dimensions 27 mm × 32 mm
• For thread G1/-1/2": According to DIN 7603 with dimensions 48 mm × 55 mm
Use a sealing ring according to this standard in form A, C or D and of a material that offers
appropriate resistance for the application.
Refer to the dimensional drawing for the length of the screwed plug:

Nozzle installation
£150 (6)
H

ø £150 (6)
A0015122

H Length of the centering rod or the rigid part of the rope probe

• Permissible nozzle diameter: ≤ 150 mm (6 in)


For larger diameters, the near-range measuring capability may be reduced.
For large nozzles, see the section "Mounting in nozzles ≥ DN300"
• Permissible nozzle height: ≤ 150 mm (6 in)
For larger heights, the near-range measuring capability may be reduced.
Larger nozzle heights are possible in special cases (on request), see sections "Centering rod for
FMP51 and FMP52" and "Rod extension/centering device HMP40 for FMP54".
• The end of the nozzle should be flush with the tank ceiling in order to avoid ringing effects.
In thermally insulated vessels, the nozzle should also be insulated in order to prevent
condensate formation.

Endress+Hauser 49
Levelflex FMP51, FMP52, FMP54

Centering rod
In the case of rope probes, it may be necessary to use a version with a centering rod so that the rope
does not come in contact with the nozzle wall during the process.
The length of the optional centering rod determines the maximum nozzle height.

Rod extension/centering device HMP40 for FMP54


For FMP54 with rope probes, the rod extension/centering device HMP40 is available as an
accessory. It must be used if the probe rope would otherwise come into contact with the lower edge
of the nozzle.
This accessory consists of the extension rod, corresponding to the nozzle height, on which a
centering disk is also mounted if the nozzles are narrow or when used in bulk solids.
This component is delivered separately from the device. Order a correspondingly shorter probe
length.
Only use centering disks with small diameters (DN40 and DN50) if there is no significant build-
up in the nozzle above the disk. The nozzle must not become clogged with product.

Mounting in nozzles ≥ DN300


If installation in nozzles ≥ 300 mm (12 in) is unavoidable, installation must be carried out in
accordance with the following diagram in order to avoid interference signals in the near range.

1
3
2

A0014199

1 Lower edge of the nozzle


2 Approximately flush with the lower edge of the nozzle (±50 mm)
3 Plate, nozzle Ø 300 mm (12 in) = plate Ø 280 mm (11 in); nozzle Ø ≥ 400 mm (16 in) = plate Ø ≥
350 mm (14 in)
4 Pipe Ø 150 to 180 mm

50 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mounting cladded flanges


Note the following for cladded flanges:
• Use the same number of flange screws as the number of flange bores provided.
• Tighten the screws with the necessary torque (see Table).
• Retighten after 24 hours or after the first temperature cycle.
• Depending on the process pressure and temperature, check and retighten the screws, where
necessary, at regular intervals.
The PTFE flange cladding normally acts simultaneously as a seal between the nozzle and the
device flange.

Flange size Number of screws Tightening torque

EN

DN40/PN40 4 35 to 55 Nm

DN50/PN16 4 45 to 65 Nm

DN50/PN40 4 45 to 65 Nm

DN80/PN16 8 40 to 55 Nm

DN80/PN40 8 40 to 55 Nm

DN100/PN16 8 40 to 60 Nm

DN100/PN40 8 55 to 80 Nm

DN150/PN16 8 75 to 115 Nm

DN150/PN40 8 95 to 145 Nm

ASME

1½"/150lbs 4 20 to 30 Nm

1½"/300lbs 4 30 to 40 Nm

2"/150lbs 4 40 to 55 Nm

2"/300lbs 8 20 to 30 Nm

3"/150lbs 4 65 to 95 Nm

3"/300lbs 8 40 to 55 Nm

4"/150lbs 8 45 to 70 Nm

4"/300lbs 8 55 to 80 Nm

6"/150lbs 8 85 to 125 Nm

6"/300lbs 12 60 to 90 Nm

JIS

10K 40A 4 30 to 45 Nm

10K 50A 4 40 to 60 Nm

10K 80A 8 25 to 35 Nm

10K 100A 8 35 to 55 Nm

10K 100A 8 75 to 115 Nm

Endress+Hauser 51
Levelflex FMP51, FMP52, FMP54

Securing the probe

Securing rope probes

B
2
A

A0012609

A Rope sag: ≥ 10 mm/(1 m probe length) [0.12 in/(1 ft probe length)]


B Reliably grounded end of probe
C Reliably insulated end of probe
1 Fastener in female thread of probe end weight
2 Insulated fastening kit

• The end of the rope probe must be secured (fixed down) under the following conditions:
If the probe would otherwise temporarily come into contact with the vessel wall, the cone, internal
fittings/beams or another part of the installation
• A female thread is provided in the probe weight to secure the end of the probe:
Rope 4 mm (1/6"), 316: M 14
• When fixed down, the end of the probe must be either reliably grounded or reliably insulated. Use
an insulated fastening kit if it is not otherwise possible to secure the probe with a reliably
insulated connection.
• To prevent an extremely high tensile load (e.g. due to thermal expansion) and the risk of the rope
breaking, the rope must be slack. Required sag: ≥ 10 mm/(1 m rope length) [ 0.12 in/(1 ft rope
length)].
Pay attention to the tensile loading capacity of rope probes.

52 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Securing rod probes


• For WHG approval: A support is required for probe lengths ≥ 3 m (10 ft).
• In general, rod probes must be secured in the event of horizontal flow (e.g. from an agitator) or
strong vibrations.
• Only secure rod probes directly at the end of the probe.

øa

øb

»50 (1.97)
3

ø<25 (1.0)

»3 (0.12)

»50 (1.97) 6

mm (in)

A0012607

Unit of measurement mm (in)


1 Probe rod, uncoated
2 Sleeve with narrow bore to ensure electrical contact between the sleeve and the rod.
3 Short metal pipe, e.g. welded in place
4 Probe rod, coated
5 Plastic sleeve, e.g. PTFE, PEEK, PPS
6 Short metal pipe, e.g. welded in place

Probe Ø 8 mm (0.31 in)


• a < Ø 14 mm (0.55 in)
• b = Ø 8.5 mm (0.34 in)
Probe Ø 12 mm (0.47 in)
• a < Ø 20 mm (0.78 in)
• b = Ø 12.5 mm (0.52 in)
Probe Ø 16 mm (0.63 in)
• a < Ø 26 mm (1.02 in)
• b = Ø 16.5 mm (0.65 in)

NOTICE
Poor grounding of the probe end may cause incorrect measurements.
‣ Use a sleeve with a narrow bore to ensure good electrical contact between the sleeve and the
probe rod.

NOTICE
Welding can damage the main electronics module.
‣ Before welding: Ground the probe rod and remove the electronics.

Endress+Hauser 53
Levelflex FMP51, FMP52, FMP54

Securing coax probes


For WHG approval: A support is required for probe lengths ≥ 3 m (10 ft).

A0012608

Coax probes can be secured (fixed) at any point in the ground tube.

54 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Special installation situations

Bypasses and stilling wells


The use of centering disks/stars/weights (available as accessories) is recommended in bypass
and stilling well applications.
As the measuring signal penetrates many plastics, installation in plastic bypasses or stilling
wells can produce incorrect results. For this reason use a metal bypass or stilling well.

A B
C
1

ød ød
2 4

(3.94)
³100
3

(3.94)
³100 ød
øD
2 ød 4
5

A0039216

 36 Engineering unit: mm (in)


A Mounting in stilling well
B Mounting in bypass
C Centering disk/centering star/centering weight
1 Metal centering disk (316L) for level measurement
2 Securing screw; torque: 25 Nm ± 5 Nm
3 Non-metal centering star (PEEK, PFA) preferred for interface measurement
4 Metal centering weight (316L) for level measurement
5 Minimum distance between probe end and lower edge of bypass 10 mm (0.4 in)

• Pipe diameter: > 40 mm (1.6 in) (for rod probes).


• A rod probe can be installed in pipes with a diameter of up to 150 mm (6 in). The use of a coax
probe is recommended for larger pipe diameters.
• Side outlets, holes, slots and welds - with a maximum inward projection of 5 mm (0.2 in) - do not
affect the measurement.
• There should not be any changes in the diameter of the pipe.
• The probe must be 100 mm (4 in) longer than the lower outlet.

Endress+Hauser 55
Levelflex FMP51, FMP52, FMP54

• The probes must not touch the pipe wall within the measuring range. Support or brace the probe if
necessary. All rope probes are prepared for bracing in vessels (tensioning weight with anchor
hole).
• If a metal centering disk is mounted at the end of the probe rod, the signal for detecting the end of
the probe is reliably defined.
Note: The non-metal centering stars made of PEEK or PFA are recommended for interface
measurements. When using metal centering disks, it is important to ensure that the lower
medium covers the centering disk at all times. Otherwise, incorrect interface measurements can
result.
• Coax probes can be used within any restrictions provided that the pipe diameter permits their
installation.
For bypasses with condensate formation (water) and a medium with a low dielectric constant
(e.g. hydrocarbons):
Over time, the bypass fills with condensate up to the lower outlet. When levels are low, the
level echo is masked by the echo of the condensate as a result. In this range, the level of the
condensate is output and the correct value is only output when levels are higher. For this
reason, ensure that the lower outlet is 100 mm (4 in) below the lowest level to be measured
and fit a metal centering disk at the level of the lower edge of the lower outlet.
In thermally insulated vessels, the bypass should also be insulated in order to prevent
condensate formation.

Assignment of centering disk/centering star/centering weight to the pipe diameter

Metal centering disk (316L)


for level measurement
Rod centering disk (Ød) 45 mm (1.77 in)
for pipe diameter (ØD)
DN50/2" to DN65/2½"
Rod centering disk (Ød) 75 mm (2.95 in)
for pipe diameter (ØD)
DN80/3" to DN100/4"
Rope centering disk (Ød) 75 mm (2.95 in)
for pipe diameter (ØD)
DN80/3" to DN100/4"

Metal centering weight (316L)


for level measurement
Rope centering weight (Ød) 45 mm (1.77 in), h 60 mm (2.36 in)
for pipe diameter (ØD)
DN50/2"
Rope centering weight (Ød) 75 mm (2.95 in), h 30 mm (1.81 in)
for pipe diameter (ØD)
DN80/3"
Rope centering weight (Ød) 95 mm (3.74 in), h 30 mm (1.81 in)
for pipe diameter (ØD)
DN100/4"

Non-metal centering star (PEEK)


For level and interface measurement, operating temperature: –60 to +250 °C (–76 to 482 °F)
Rod centering star (Ød) 48 to 95 mm (1.89 to 3.74 in)
for pipe diameter (ØD)
≥ DN50/2"

Non-metal centering star (PFA)


For level and interface measurement, operating temperature: –200 to +250 °C (–328 to +482 °F)
Rod centering star (Ød) 37 mm (1.46 in)
for pipe diameter (ØD)
≥ 40 mm (1.57 in)

56 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Horizontal cylindrical and vertical tanks

A0014141

1 Coax probe

• Any distance from wall provided occasional contact is avoided.


• Use a coax probe (1) if installing in tanks with many internal fixtures or internal fixtures located
close to the probe.

Endress+Hauser 57
Levelflex FMP51, FMP52, FMP54

Underground tanks

A0014142

In the case of nozzles with large diameters, use a coax probe to avoid reflections at the nozzle wall.

58 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mounting at an angle

LN
a

A0014145

• For mechanical reasons, the probe should be installed as vertically as possible.


• If the probe is installed at an angle, the length of the probe must be reduced depending on the
angle of installation.
• α 5 °: LNmax. 4 m (13.1 ft)
• α 10 °: LNmax. 2 m (6.6 ft)
• α 30 °: LNmax. 1 m (3.3 ft)

Endress+Hauser 59
Levelflex FMP51, FMP52, FMP54

Non-metal vessels

A0012527

1 Non-metal vessel
2 Metal sheet or metal flange

To ensure good measurement results when mounting on non-metal vessels


• Use a device with a metal flange (minimum size DN50/2").
• Alternatively: at the process connection, mount a metal sheet with a diameter of at least
200 mm (8 in) at a right angle to the probe.
A metal surface is not required at the process connection in the case of coax probes.

60 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Plastic and glass vessels: Mounting the probe on the outside wall
In the case of plastic and glass vessels, the probe can also be mounted on the outside wall under
certain conditions.

1
3

A0014150

1 Plastic or glass vessel


2 Metal plate with screw-in sleeve
3 No space between vessel wall and probe!

Requirements
• Dielectric constant of the medium: εr > 7.
• Non-conductive vessel wall.
• Maximum wall thickness (a):
• Plastic: < 15 mm (0.6 in)
• Glass: < 10 mm (0.4 in)
• No metal reinforcements on the vessel.
Note the following when mounting the device:
• Mount the probe directly on the vessel wall without any space between the wall and probe.
• To prevent any influence on the measurement, fit a plastic half pipe with a diameter of at least
200 mm (8 in), or a similar protective unit, on the probe.
• For vessel diameters less than 300 mm (12 in):
On the opposite side of the vessel, fit a grounding plate that is conductively connected to the
process connection and covers around half of the vessel's circumference.
• For vessel diameters of 300 mm (12 in) and higher:
At the process connection, fit a metal plate with a diameter of at least 200 mm (8 in) at a right
angle to the probe (see above).

Endress+Hauser 61
Levelflex FMP51, FMP52, FMP54

Vessel with thermal insulation


If process temperatures are high, the device must be included in normal vessel insulation (1) in
order to prevent the electronics heating up as a result of thermal radiation or convection. The
insulation may not go beyond the points labeled "MAX" in the drawings.

2 3

1
MAX MAX

A0014653

 37 Process connection with thread


1 Vessel insulation
2 Compact device
3 Sensor, remote

2 3
40 (1.57)

40 (1.57)
1
MAX MAX

mm (in)

A0014654

 38 Process connection with flange


1 Vessel insulation
2 Compact device
3 Sensor, remote

XT 280 °C (536 °F)* HT 450 °C (842 °F)


2 3
2 3
MAX MAX MAX MAX
1 1

A0014657

 39 Process connection with thread - sensor version XT and HT


1 Vessel insulation
2 Compact device
3 Sensor, remote
* The XT version is not recommended for saturated steam above 200 °C (392 °F); the HT version should be used
instead

62 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

XT 280 °C (536 °F)* HT 450 °C (842 °F)

2 3
2 3
MAX MAX MAX MAX
1 1

A0014658

 40 Process connection with flange - sensor version XT and HT


1 Vessel insulation
2 Compact device
3 Sensor, remote
* The XT version is not recommended for saturated steam above 200 °C (392 °F); the HT version should be used
instead

Endress+Hauser 63
Levelflex FMP51, FMP52, FMP54

Replacing a displacer system in an existing displacer chamber


FMP51 and FMP54 are a perfect replacement for a conventional displacer system in an existing
displacer chamber. Flanges that suit Fisher and Masoneilan displacer chambers are available for this
purpose (special product for FMP51; feature 100 of the product structure, options LNJ, LPJ, LQJ for
FMP54). Thanks to menu-guided local operation, commissioning the Levelflex only takes a few
minutes. Replacement is also possible when partially filled, and wet calibration is not required.
Your benefits:
• No moving parts, therefore zero-maintenance operation.
• Not affected by process influences such as temperature, density, turbulence and vibrations.
• The rod probes can be easily shortened or replaced. Therefore, the probe can also be easily
adjusted on site.

A0014153

1 Flange of the displacer chamber

Planning instructions:
• In normal cases, use a rod probe. When installing into a metal displacer chamber up to 150 mm,
you have all the advantages of a coax probe.
• Contact between the probe and the side wall must be avoided. Where necessary, use a centering
disk or centering star at the bottom end of the probe.
• The centering disk or centering star must be adapted as accurately as possible to the internal
diameter of the displacer chamber to also ensure correct operation around the probe end.
Additional information regarding interface measurement
• In the case of oil and water, the centering star should be positioned at the lower edge of the lower
outlet (water level).
• There should not be any changes in the diameter of the pipe. Use the coax probe if necessary.
• It must be ensured that rod probes do not come into contact with the wall. Where necessary, use a
centering star at the end of the probe.
• The non-metal centering stars made of PEEK or PFA are recommended for interface
measurements. When using metal centering disks, it is important to ensure that the lower
medium covers the centering disk at all times. Otherwise, incorrect interface measurements can
result.

64 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Environment
Ambient temperature Measuring device –40 to +80 °C (–40 to +176 °F)

Measuring device (option –50 to +80 °C (–58 to +176 °F) 1)


for FMP51 and FMP54)

Local display –20 to +70 °C (–4 to +158 °F), the readability of the local display may be impaired
at temperatures outside the temperature range.

Connecting cable (for –50 to +100 °C (–58 to +212 °F)


"Sensor, remote" probe
design)

Remote display FHX50 –40 to 80 °C (–40 to 176 °F)

Remote display FHX50 –50 to 80 °C (–58 to 176 °F) 1)


(option)

1) This range applies if the option JN "Transmitter ambient temperature –50 °C (–58 °F)" has been selected in
order code 580 "Test, certificate". If the temperature is permanently below –40 °C (–40 °F), higher failure
rates can be expected.

If operating outdoors in strong sunlight:


• Mount the device in the shade.
• Avoid direct sunlight, particularly in warm climatic regions.
• Use a weather protection cover (see accessories).

Ambient temperature limits The following diagrams only consider functional aspects. Additional restrictions may apply for
certified device versions. See the separate Safety Instructions for more information.

Endress+Hauser 65
Levelflex FMP51, FMP52, FMP54

In the event of temperature (Tp) at the process connection, the permitted ambient temperature (Ta)
is reduced as indicated in the following diagram (temperature derating):

Temperature derating for FMP51 with threaded connection G¾ or NPT¾

[°C] ([°F])Ta
Ta
+80 (+176)
GT20: +58(+136)
Tp
GT18: +48(+118)
A: 4...20 mA HART GT19: +29(+84)

-40 (-40)³ Tp
-40 +82 +200 [°C]
(-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +54(+129)
C: 4...20 mA HART
4...20 mA GT18: +48(+118)
K: 90...253 VAC GT19: +29(+84)
L: 10.4...48 VDC
M: Modbus
-40 (-40)³ Tp
-40 +74 +200 [°C]
(-40) (+165) (+392) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +57(+134)
GT18: +48(+118)

G 1: PROFIBUS PA
FOUNDATION Fieldbus
-40 (-40) Tp
-40 +81 +200 [°C]
(-40) (+178) (+392) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +54(+129)
GT18: +48(+118)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
-40 (-40) Tp
-40 +79 +200 [°C]
(-40) (+174) (+392) ([°F])
A0013687

GT18 = Stainless steel housing A = 1 current output Ta = Ambient temperature 3)


GT19 = Plastic housing C = 2 current outputs Tp = Temperature at the process
GT20 = Aluminum housing G1, G2 = PROFIBUS PA 1) 2) connection
K, L = 4-wire

1) G1: Switch output not used


2) G2: Switch output used
3) Ta to –50 °C (–58 °F) for order code 580 "Test, certificate" = JN "Transmitter ambient temperature
–50 °C (–58 °F)"; only available for 2-wire HART devices

66 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP51 with threaded connection G1½ or NPT1½

[°C] ([°F])Ta
Ta
+80(+176)
GT20: +57(+135)
Tp
GT18: +49(+120)

GT19: -34(-29) A: 4...20 mA HART GT19: +20(+68)


GT18: -37(-34)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +82 +200 [°C]
(-58) (-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +54(+129)
C: 4...20 mA HART
4...20 mA GT18: +49(+120)
K: 90...253 VAC +20(+68)
L: 10.4...48 VDC
GT19: -34(-29) GT19:
M: Modbus
GT18: -37(-34)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +74 +200 [°C]
(-58) (-40) (+165) (+392) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +57(+134)
GT18: +49(+120)
G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +81 +200 [°C]
(-58) (-40) (+178) (+392) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +54(+129)
2
PROFIBUS PA GT18: +49(+120)
G :
FOUNDATION Fieldbus
Switch output
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +79 +200 [°C]
(-58) (-40) (+174) (+392) ([°F])
A0014121

GT18 = Stainless steel housing A = 1 current output Ta = Ambient temperature 3)


GT19 = Plastic housing C = 2 current outputs Tp = Temperature at the process
GT20 = Aluminum housing G1, G2 = PROFIBUS PA 1) 2) connection
K, L = 4-wire

1) G1: Switch output not used


2) G2: Switch output used
3) Ta to –50 °C (–58 °F) for order code 580 "Test, certificate" = JN "Transmitter ambient temperature
–50 °C (–58 °F)"; only available for 2-wire HART devices

Endress+Hauser 67
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP51 with flange

[°C] ([°F])Ta
Ta
+80(+176)
GT20: +59(+138)
Tp
GT18: +53(+127)
GT19: -34(-29) A: 4...20 mA HART GT19: +33(+91)
GT18: -37(-34)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +82 +200 [°C]
(-58) (-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165)
4...20 mA HART GT20: +55(+131)
C:
4...20 mA GT18: +51(+124)
K: 90...253 VAC GT19: +33(+91)
GT19: -34(-29) L: 10.4...48 VDC
GT18: -37(-34) M: Modbus
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +74 +200 [°C]
(-58) (-40) (+165) (+392) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +58(+136)
GT18: +53(+127)
1
G : PROFIBUS PA
FOUNDATION Fieldbus
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +81 +200 [°C]
(-58) (-40) (+178) (+392) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +55(+131)
2
PROFIBUS PA GT18: +51(+124)
G :
FOUNDATION Fieldbus
Switch output
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +79 +200 [°C]
(-58) (-40) (+174) (+392) ([°F])
A0013689

GT18 = Stainless steel housing A = 1 current output Ta = Ambient temperature 3)


GT19 = Plastic housing C = 2 current outputs Tp = Temperature at the process
GT20 = Aluminum housing G1, G2 = PROFIBUS PA 1) 2) connection
K, L = 4-wire

1) G1: Switch output not used


2) G2: Switch output used
3) Ta to –50 °C (–58 °F) for order code 580 "Test, certificate" = JN "Transmitter ambient temperature
–50 °C (–58 °F)"; only available for 2-wire HART devices

68 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP52

T a [°C] ([°F])Ta
+80(+176)
GT20: +60(+140)
Tp GT18: +53(+127)
GT19: +38(+100)
GT19: -36(-33) A: 4...20 mA HART
GT18: -37(-35)
GT20: -38(-36)
-40(-40) Tp
-50 -40 +82 +200 [°C]
(-58) (-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +56(+133)
C: 4...20 mA HART GT18: +51(+124)
4...20 mA GT19: +38(+100)
GT19: -36(-33) K: 90...253 VAC
GT18: -37(-35)
GT20: -38(-36) L: 10.4...48 VDC
-40(-40) Tp
-50 -40 +74 +200 [°C]
(-58) (-40) (+165) (+392) ([°F])
[°C] ([°F])Ta

+81(+178)
GT20: +59(+138)
GT18: +53(+127)

G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -37(-35)
GT20: -38(-36)
-40(-40) Tp
-50 -40 +81 +200 [°C]
(-58) (-40) (+178) (+392) ([°F])
[°C] ([°F])Ta

+79(+174)
GT20: +56(+133)
GT18: +51(+124)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
GT18: -37(-35)
GT20: -38(-36)
-40(-40) Tp
-50 -40 +79 +200 [°C]
(-58) (-40) (+174) (+392) ([°F])
A0013633

GT18 = Stainless steel housing A = 1 current output Ta = Ambient temperature


GT19 = Plastic housing C = 2 current outputs Tp = Temperature at the process
GT20 = Aluminum housing G1, G2 = PROFIBUS PA 1) connection 2)
K, L = 4-wire

1) In the case of PROFIBUS PA and FOUNDATION Fieldbus, the temperature derating depends on whether
the switch output (terminals 3 and 4) is used (G2) or not (G1).
2) The process temperature should not exceed 150 °C (302 °F) for saturated steam applications. Use FMP54
for higher process temperatures.

Endress+Hauser 69
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP54 - version XT to +280 °C (+536 °F )

T a [°C] ([°F])Ta
+80(+176)
Tp GT20: +62(+144)
GT18: +56(+133)
GT19: +42(+108)
GT19: -5(+23) A: 4...20 mA HART
GT18: -16(+3)
GT20: -23(-9)
-40 (-40)³ Tp
-196 -40 +82 +280 [°C]
(-321) (-40) (+180) (+536) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +59(+138)
GT18: +52(+126)
C: 4...20 mA HART
GT19: +42(+108)
4...20 mA
GT19: -5(+23)
K: 90...253 VAC
GT18: -16(+3)
GT20: -23(-9) L: 10.4...48 VDC
-40 (-40)³ Tp
-196 -40 +74 +280 [°C]
(-321) (-40) (+165) (+536) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +61(+142)
GT18: +55(+130)
G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -16(+3)
GT20: -23(-9)
-40 (-40) Tp
-196 -40 +81 +280 [°C]
(-321) (-40) (+178) (+536) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +58(+137)
GT18: +55(+126)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
GT18: -16(+3)
GT20: -23(-9)
-40 (-40) Tp
-196 -40 +79 +280 [°C]
(-321) (-40) (+174) (+536) ([°F])
A0013631

GT18 = Stainless steel housing A = 1 current output Ta = Ambient temperature 3)


GT19 = Plastic housing C = 2 current outputs Tp = Temperature at the process
GT20 = Aluminum housing G1, G2 = PROFIBUS PA 1) 2) connection
K, L = 4-wire

1) G1: Switch output not used


2) G2: Switch output used
3) Ta to –50 °C (–58 °F) for order code 580 "Test, certificate" = JN "Transmitter ambient temperature
–50 °C (–58 °F)"; only available for 2-wire HART devices

70 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP54 - version HT to +450 °C (+842 °F )

[°C] ([°F])Ta
Ta
+80(+176)
GT20: +56(+133)
Tp GT18: +54(+129)

A: 4...20 mA HART GT19: +42(+108)


GT19: -22(-8)
GT18: -27(-17)
GT20: -28(-18)
-40 (-40)³ Tp
-196 -40 +82 +450 [°C]
(-321) (-40) (+180) (+842) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +53(+127)
C: 4...20 mA HART GT18: +51(+124)
4...20 mA GT19: +38(+100)
GT19: -22(-8) K: 90...253 VAC
GT18: -27(-17)
GT20: -28(-18) L: 10.4...48 VDC
-40 (-40)³ Tp
-196 -40 +74 +450 [°C]
(-321) (-40) (+165) (+842) ([°F])

[°C] ([°F])Ta

+81(+178)
GT20: +55(+131)
GT18: +53(+127)
G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -27(-16)
GT20: -28(-18)
-40 (-40) Tp
-196 -40 +81 +450 [°C]
(-321) (-40) (+178) (+842) ([°F])
[°C] ([°F])Ta

+79(+174)
GT20: +52(+126)
GT18: +50(+123)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
GT18: -27(-16)
GT20: -28(-18)
-40 (-40) Tp
-196 -40 +79 +450 [°C]
(-321) (-40) (+173) (+842) ([°F])
A0013632

GT18 = Stainless steel housing A = 1 current output Ta = Ambient temperature 3)


GT19 = Plastic housing C = 2 current outputs Tp = Temperature at the process
GT20 = Aluminum housing G1, G2 = PROFIBUS PA 1) 2) connection
K, L = 4-wire

1) G1: Switch output not used


2) G2: Switch output used
3) Ta to –50 °C (–58 °F) for order code 580 "Test, certificate" = JN "Transmitter ambient temperature
–50 °C (–58 °F)"; only available for 2-wire HART devices

Endress+Hauser 71
Levelflex FMP51, FMP52, FMP54

Storage temperature • Permitted storage temperature: –40 to +80 °C (–40 to +176 °F)
• Use original packaging.
• Option for FMP51 and FMP54: –50 to +80 °C (–58 to +176 °F)
This range applies if the option JN "Transmitter ambient temperature" –50 °C (–58 °F) was selected
in order code 580 "Test, Certificate". If the temperature is permanently below –40 °C (–40 °F),
higher failure rates can be expected.

Climate class DIN EN 60068-2-38 (test Z/AD)

Altitude according to • Generally up to 2 000 m (6 600 ft) above MSL.


IEC61010-1 Ed.3 • Above 2 000 m (6 600 ft) if the following conditions are met:
• Ordering feature 020 "Power supply; Output" = A, B, C, E or G (2-wire versions)
• Supply voltage U < 35 V
• Supply voltage of overvoltage category 1

Degree of protection Tested according to:


• With housing closed: IP68, NEMA6P (24 h at 1.83 m (6 ft) under water)
(also applies for the "Remote sensor" version)
• For housing: GT19 dual compartment, PBT plastic in combination with display, operation: SD02
or SD03: IP68 (24 h at 1 m (3.28 ft) under water)
• IP66, NEMA4X
• With housing open: IP20, NEMA1
• Display module: IP22, NEMA2
• For M12 plug: IP68 NEMA6P, only if the cable is plugged in and also specified according to IP68
NEMA6P

Vibration resistance DIN EN 60068-2-64 / IEC 60068-2-64: 20 to 2 000 Hz, 1 (m/s2)2/Hz

Cleaning the probe Dirt or buildup may form on the probe depending on the application. A thin, even layer has little
impact on the measurement. Thick layers can dampen the signal and reduce the measuring range.
Very uneven deposit formation or caking (e.g. due to crystallization) can result in incorrect
measurements. In such cases, use a non-contact measuring principle, or regularly inspect the probe
for contamination.
Cleaning with sodium hydroxide solution (e.g. in CIP procedures): if the coupling is wetted, larger
measurement errors can occur than under reference operating conditions. Wetting can cause
temporary incorrect measurements.

Electromagnetic Electromagnetic compatibility in accordance with all the relevant requirements outlined in the EN
compatibility (EMC) 61326 series and NAMUR Recommendation EMC (NE 21). For details, refer to the Declaration of
Conformity.
Download at www.endress.com.

Use a shielded cable for signal transmission.


Maximum measured error during EMC testing: < 0.5 % of the span.
When the probes are installed in metal and concrete vessels and when a coax probe is used:
• Interference emission according to EN 61326 - x series, Class B equipment.
• Interference immunity according to EN 61326 - x series, requirements for industry and NAMUR
Recommendation NE 21 (EMC)
When probes are installed without a shielding/metal wall, e.g. installation in plastic and wooden
silos, or if the "Remote sensor" device version is used, the effect of strong electromagnetic fields can
influence the measured value.
• Interference emission according to EN 61326 - x series, Class A equipment.
• Interference immunity: the effect of strong electromagnetic fields can influence the measured
value.

72 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Process
Process temperature range The maximum temperature permitted at the process connection is determined by the O-ring version
ordered:

Device O-ring material Process temperature Approval

FMP51 FKM (Viton GLT) –30 to +150 °C (–22 to +302 °F) FDA

–40 to +150 °C (–40 to +302 °F)


Only in conjunction with feature 610 "Accessory
mounted", feature version NC "Gas-tight gland"

EPDM (70C4 pW FKN or –40 to +120 °C (–40 to +248 °F) FDA


E7515)

FFKM (Kalrez 6375) 1) –20 to +200 °C (–4 to +392 °F) 2)

FVMQ (FVMQ 70C79) –50 to 130 °C (–58 to 260 °F)

FMP52 — –50 to +200 °C (–58 to +392 °F); fully coated FDA, 3A,
EHEDG, USP Cl.
VI 3)

FMP54 Graphite Version XT: –196 to +280 °C (–321 to +536 °F) 4)

Version HT: –196 to +450 °C (–321 to +842 °F)

1) Recommended for water vapor applications.


2) Not recommended for saturated steam over 150 °C (302 °F). Use FMP54 instead.
3) Plastic components in contact with the medium have been tested in accordance with USP <88> Class
VI-70°C
4) Not recommended for saturated steam over 200 °C (392 °F). Use the HT version instead.

For FMP52: High process temperatures (> 150 °C (302 °F)) may favor the diffusion of the
medium through the probe coating, which may reduce the operating time.
The sensor material 1.4404/316L is resistant to attack from intergranular corrosion according
to AD 2000 - instruction sheet W2 at operating temperatures up to 400 °C (752 °F) and with an
operating time of 100,000 hours (11.4 years). For higher temperatures, the suitability of the
material must be checked by the operator. Acids, in particular, may cause corrosion.
With uncoated probes, the medium temperature may be higher provided that the process
temperature specified in the table is not exceeded at the process connection.
When using rope probes, the stability of the probe rope is reduced by structural changes at
temperatures above 350 °C (662 °F), however.

Process pressure range Device Process pressure

FMP51 –1 to 40 bar (–14.5 to 580 psi)

FMP52 –1 to 40 bar (–14.5 to 580 psi)

FMP54 –1 to 400 bar (–14.5 to 5 800 psi)

Endress+Hauser 73
Levelflex FMP51, FMP52, FMP54

FMP54 - permitted process pressure depending on the process temperature

[bar] ([psi]) p

420 (6090)

400 (5800)

380 (5510)

350 (5075)

320 (4640)

300 (4350) Tp
-196 -100 0 +100 200 300 400 450 [°C]
(-321) (-148) (+32) (+212) (+392) (+572) (+752) (+842) ([°F])

A0014005

p = Permitted process pressure Tp = Process temperature

The range indicated can be reduced by the choice of process connection. The maximum working
pressure (MWP) indicated on the nameplate refers to a reference temperature of 20 °C, and of
100 °F for ASME flanges. Observe pressure-temperature dependency.
Please refer to the following standards for the pressure values permitted at higher
temperatures:
• EN 1092-1: 2007 Tab. G.4.1-x
With regard to their stability-temperature property, the materials 1.4435 and 1.4404 are
grouped together under 13E0 in EN 1092-1: 2007 Tab. G.3.1-1. The chemical composition of
the two materials can be identical.
• ASME B 16.5a - 2013 Tab. 2-2.2 F316
• ASME B 16.5a - 2013 Tab. 2.3.8 N10276
• JIS B 2220

Dielectric constant (DC) • With coax probe: DC (εr) ≥ 1.4


• Rod and rope probe: DC (εr) ≥ 1.6 (when installing in pipes DN ≤ 150 mm (6 in): DC (εr) ≥ 1.4)

Extension of the rope probes Elongation from temperature increase from 30 °C (86 °F) to 150 °C (302 °F): 2 mm / m rope length
due to temperature

74 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mechanical construction
Dimensions Dimensions of the electronics housing

R110 98 (3.86)*
78 (3.07) 90 (3.54)

07)
(4.
3.5

146 (5.75)
ø108.5 (4.27)

142 (5.59)
115.25 (4.54)
ø10
A0011666

 41 Housing GT18 (316L). Unit of measurement mm (in)


*For devices with integrated overvoltage protection.

R110
99 (3.9)*
78 (3.07) 90 (3.54)
)
.17
6 (4

144 (5.67)
ø106 (4.17)
ø10

113.5 (4.47)
A0011346

 42 GT19 housing (plastic PBT). Unit of measurement mm (in)


*For devices with integrated overvoltage protection.

R110 98 (3.86)*
78 (3.07) 90 (3.54)
07)
(4.

ø112 (4.41)

145.5 (5.73)
3.5

141.5 (5.57)
ø108.5 (4.27)
ø10

61 (2.4)

A0020751

 43 Housing GT20 (aluminum coated). Unit of measurement mm (in)


*For devices with integrated overvoltage protection.

Endress+Hauser 75
Levelflex FMP51, FMP52, FMP54

Mounting bracket dimensions

A B

ø42...60
(1.65...2.36)

52 (2)
122 (4.8) 127...140
161 (6.34) (5...5.51)
162...175
(6.38...6.89)

(3.4)
(2.8)

86
70

A0014793

 44 Mounting bracket for electronics housing. Unit of measurement mm (in)


A Wall mounting
B Post mounting

With "remote sensor" device versions (see feature 060 in the product structure), the mounting
bracket is included in the scope of delivery. However, it can also be ordered separately as an
accessory (order number: 71102216).

Dimensions of connection piece for remote probe

122(4.8)
(2.05)
52

15 (0.6)

rmin = 100 (4)

ø10
(0. .2
4)
(2.4)
61
43 (1.7)

ø82.5
(3.25)
A0023856

 45 Connection piece for remote probe; length of connecting cable: as per order. Unit of measurement mm (in)

76 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

FMP51: Dimensions of process connection (G¾,NPT¾)/probe

(0.98)
25
(2.72)
69
36 mm

R G¾
NPT¾

E F G H

7 mm 7 mm 7 mm

ø4 (0.16) ø4 (0.16)
ø8 (0.31) ø22 (0.87)
LN

150 (5.91)
20 (0.79)

12 (0.47)
4 (0.16)

2 (0.08)

M14 ø29.5 (1.16) 8 (0.31)


ø22 (0.87) ø75 (2.95)

A0012645

 46 FMP51: Process connection / probe. Unit of measurement mm (in)


B Thread ISO228 G3/4 or ANSI MNPT3/4 (feature 100)
E Rope probe 4mm or 1/6" (feature 060)
F Rope probe 4mm or 1/6"; centering disk optional (features 060 and 610)
G Rod probe 8mm or 1/3" (feature 060)
H Coax probe (feature 060); with vent opening Ø approx. 6 mm (0.24 in)
LN Probe length
R Reference point of measurement

Endress+Hauser 77
Levelflex FMP51, FMP52, FMP54

FMP51: Dimensions of process connection (G1½,NPT1½, flange)/probe

B C D
78 (3.1)

78 (3.1)

78 (3.1)
(0.98)

(1.02)
25

26
55 mm 55 mm

R
G1½ NPT1½

I J K L M N

10 mm R

(1.54)
14 mm

39
10 mm ø16 (0.63)
ø4 (0.16)
1000 (39.4)
LN

500 (19.7)
ø4 (0.16) 14 mm

ø12 (0.47)

40 (1.57)
14 mm
2.2 (0.09)
min. 50 (1.97)
max. 551 (21.7)
max. 1051 (41.4)
150 (5.91)
20 (0.79)

12 (0.47)
4 (0.16)

P
9 (0.35)

Q Q
S

M14 ø29.5 (1.16) T 2.77 (0.11) 2 (0.08)


ø22 (0.87) ø75 (2.95) ø12.7 (0.5) ø12 (0.47)
ø42.2 (1.66) ø42.4
(1.67)
A0012756

 47 FMP51: Process connection / probe. Unit of measurement mm (in)


B Thread ISO228 G1-1/2 (feature 100)
C Thread ANSI MNPT1-1/2 (feature 100)
D Flange ANSI B16.5, EN1092-1, JIS B2220 (feature 100)
I Rope probe 4mm or 1/6" (feature 060)
J Rope probe 4mm or 1/6"; centering disk optional (features 060 and 610)
K Rod probe 12mm or 1/2"; centering disk optional, see table below (features 060 and 610)
L Rod probe 16mm or 0.63in, 500mm or 1000mm, separable; centering disk optional, see table below
(features 060 and 610)
M Coax probe; AlloyC (feature 060); with vent opening Ø approx. 8 mm (0.3 in)
N Coax probe; 316L (feature 060); with vent openings Ø approx. 10 mm (0.4 in)
LN Probe length
P Thickness of centering star; for table of values, see below
Q Diameter of centering star; for table of values, see below
R Reference point of measurement
S Thickness of centering disk or centering star; for table of values, see below
T Diameter of centering disk or centering star: for table of values, see below

78 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Centering disk / centering star / centering weight


Order code 610 Meaning Thickness Diameter
"Accessory
mounted"

OA Rod centering disk 316L; pipe diameter DN80/3" + DN100/4" S = 4 mm (0.16 in) T = 75 mm (2.95 in)

OB Rod centering disk 316L; pipe diameter DN50/2" + DN65/2-1/2" S = 4 mm (0.16 in) T = 45 mm (1.77 in)

OC Rope centering disk 316L; pipe diameter DN80/3" + DN100/4" S = 4 mm (0.16 in) T = 75 mm (2.95 in)

OD Rod centering star PEEK; interface measurement; pipe diameter S =7 mm (0.28 in) T = 48 to 95 mm (1.9 to 3.7 in)
DN50/2" + DN100/4"

OE Rod centering star PFA; interface measurement; pipe diameter P = 10 mm (0.39 in) Q = 37 mm (1.46 in)
DN40/1-1/2" + DN50/2"

OK Rope centering weight 316L for DN50/2" 60 mm (2.4 in) 45 mm (1.77 in)

OL Rope centering weight 316L for DN80/3" 30 mm (1.18 in) 75 mm (2.95 in)

OM Rope centering weight 316L for DN100/4" 30 mm (1.18 in) 95 mm (3.7 in)

Note on AlloyC flanges


AlloyC flanges always have an additional thread, even if they are not used with a coax probe.
Options for order feature 100 for "Process connection" that are affected: AEM, AFM, AGM, AQM,
ARM, ASM, ATM, CEM, CFM, CGM, CQM, CRM, CSM, CTM.

M40
8 (0.31)

A0035223

 48 Dimensions of AlloyC flanges. Unit of measurement mm (in)

Endress+Hauser 79
Levelflex FMP51, FMP52, FMP54

FMP52: Dimensions of process connection/probe

A B C D E
64 91
2.52 3.58

(3.19)
(3.23)

(3.31)
(3.23)

(3.23)

81
82

84
82

82 R

ø16 ø16 (0.63) ø16 (0.63) ø16 (0.63) ø16 (0.63)


(0.63) ø68.5 (2.7)
ø50.5 (1.99)

F G

ø16 (0.63)
ø16 (0.63)
LN

ø4 (0.16)

ø37 (1.46)
118 (4.65)

ø22 (0.87)

A0012757

 49 FMP52: Process connection / probe. Unit of measurement mm (in)


A Tri-Clamp 1-1/2" (feature 100)
B Tri-Clamp 2" (feature 100)
C Tri-Clamp 3" (feature 100)
D DIN11851 (dairy pipe) DN50 (feature 100)
E Flange ANSI B16.5, EN1092-1, JIS B2220 (feature 100)
F Rod probe 16mm or 0.63in, PFA>316L (feature 060); with optional centering star (feature 610)
G Rope probe 4mm or 1/6", PFA>316 (feature 060)
LN Probe length
R Reference point of measurement

80 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

FMP54: Dimensions of process connection/probe

A B C D

60 mm 60 mm
60 mm
132 (5.2)

132 (5.2)
60 mm

342 (13.5)
315 (12.4)
267 (10.5)
240 (9.45)

159 (6.26)

159 (6.26)
ø90 ø90
ø90 ø90 (3.54) (3.54)
(3.54) (3.54)

R
37 (1.46)

37 (1.46)
8 (0.31)

8 (0.31)

8 (0.31)

8 (0.31)
14 mm 14 mm 14 mm 14 mm

M40x1 M40x1 M40x1 M40x1

E F G H I

ø16 (0.63)

40 (1.57)
14 mm 14 mm

ø16 (0.63) 1000 (39.4)


500 (19.7)
LN

ø4 (0.16) ø4 (0.16)

14 mm 14 mm
2.2 (0.09)
max. 1051 (41.4)
min. 50 (1.97)

M
150 (5.91)
20 (0.79)

12 (0.47)

L
9 (0.35)
4 (0.16)

2 (0.08)
M
N

M14 ø29.5 (1.16) O ø16 (0.63)


ø22 (0.87) ø75 (2.95) ø42.4 (1.67)

A0012778

 50 FMP54: Process connection / probe. Unit of measurement mm (in)


A Thread ISO228 G1-1/2 oder ANSI MNPT1-1/2; XT 280 °C (features 100 and 090)
B Flange ANSI B16.5, EN1092-1, JIS B2220; XT 280 °C (features 100 and 090)
C Thread ISO228 G1-1/2 or ANSI MNPT1-1/2; HT 450 °C (features 100 and 090)
D Flange ANSI B16.5, EN1092-1, JIS B2220 ; HT 450 °C (features 100 and 090)
E Rope probe 4mm or 1/6" (feature 060)
F Rope probe 4mm or 1/6"; centering disk optional (features 060 and 610)
G Rod probe 16mm or 0.63in; centering disk optional, see table below (features 060 and 610)
H Rod probe 16mm or 0.63in; 500mm or 1000mm, separable; centering disk optional, see table below (feature
060 and 610)
I Coax probe (feature 060); with vent openings Ø approx. 10 mm (0.4 in); with centering disk for "Gas phase
compensation" application package (order code 540, option EF or EG)
LN Probe length
L Thickness of centering star; for table of values, see below
M Diameter of centering star; for table of values, see below

Endress+Hauser 81
Levelflex FMP51, FMP52, FMP54

N Thickness of centering disk or centering star; for table of values, see below
O Diameter of centering disk or centering star: for table of values, see below
R Reference point of measurement

Centering disk / centering star / centering weight


Order code 610 Meaning Thickness Diameter
"Accessory
mounted"

OA Rod centering disk 316L; pipe diameter DN80/3" + DN100/4" N = 4 mm (0.16 in) O = 75 mm (2.95 in)

OB Rod centering disk 316L; pipe diameter DN50/2" + DN65/2-1/2" N = 4 mm (0.16 in) O = 45 mm (1.77 in)

OC Rope centering disk 316L; pipe diameter DN80/3" + DN100/4" N = 4 mm (0.16 in) O = 75 mm (2.95 in)

OD Rod centering star PEEK; interface measurement; pipe diameter N = 7 mm (0.28 in) O = 48 to 95 mm (1.9 to 3.7 in)
DN50/2" + DN100/4"

OE Rod centering star PFA; interface measurement; pipe diameter L = 10 mm (0.39 in) M = 37 mm (1.46 in)
DN40/1-1/2" + DN50/2"

OK Rope centering weight 316L for DN50/2" 60 mm (2.4 in) 45 mm (1.77 in)

OL Rope centering weight 316L for DN80/3" 30 mm (1.18 in) 75 mm (2.95 in)

OM Rope centering weight 316L for DN100/4" 30 mm (1.18 in) 95 mm (3.7 in)

82 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Probe length tolerances Rod and coax probes


Permitted tolerance depending on the probe length:
• < 1 m (3.3 ft) = –5 mm (–0.2 in)
• 1 to 3 m (3.3 to 9.8 ft) = –10 mm (–0.39 in)
• 3 to 6 m (9.8 to 20 ft) = –20 mm (–0.79 in)
• > 6 m (20 ft) = –30 mm (–1.18 in)
Rope probes
Permitted tolerance depending on the probe length:
• < 1 m (3.3 ft) = –10 mm (–0.39 in)
• 1 to 3 m (3.3 to 9.8 ft) = –20 mm (–0.79 in)
• 3 to 6 m (9.8 to 20 ft) = –30 mm (–1.18 in)
• > 6 m (20 ft) = –40 mm (–1.57 in)

Surface roughness of AlloyC Ra = 3.2 µm (126 µin); lower surface roughness available on request.
coated flanges
This value applies to flanges with "AlloyC>316/316L"; see product structure, feature 100 "Process
connection". For other flanges, the surface roughness corresponds to the relevant flange standard.

Shortening probes If necessary, probes can be shortened by observing the following instructions:

Shortening rod probes


Rod probes must be shortened if the distance to the vessel floor or outlet cone is less than
10 mm (0.4 in). To shorten, saw off the bottom end of the rod probe.
It is not possible to shorten FMP52 rod probes due to their coating.

Shortening rope probes


Rope probes must be shortened if the distance to the vessel floor or outlet cone is less than
150 mm (6 in).
It is not possible to shorten FMP52 rope probes due to their coating.

Shortening coax probes


Coax probes must be shortened if the distance to the vessel floor or outlet cone is less than
10 mm (0.4 in).
Coax probes can be shortened by a maximum of 80 mm (3.2 in) from below. They have
centering devices on the inside to secure the rod centrally in the pipe. A raised edge holds the
centering devices in place on the rod. It is possible to shorten the probe up to approx.
10 mm (0.4 in) below the centering device.

Endress+Hauser 83
Levelflex FMP51, FMP52, FMP54

Weight Housing
Part Weight

GT18 housing - stainless steel Approx. 4.5 kg

GT19 housing - plastic Approx. 1.2 kg

GT20 housing - aluminum Approx. 1.9 kg

FMP51 with threaded connection G¾ or NPT¾


Part Weight Part Weight

Sensor Approx. 0.8 kg Rod probe 8 mm Approx. 0.4 kg/m probe length

Rope probe 4 mm Approx. 0.1 kg/m probe length Coax probe Approx. 1.2 kg/m probe length

FMP51 with threaded connection G1½/NPT1½ or flange


Part Weight Part Weight

Sensor Approx. 1.2 kg + weight of Rod probe 16 mm Approx. 1.1 kg/m probe length
flange

Rope probe 4 mm Approx. 0.1 kg/m probe length Coax probe Approx. 3.0 kg/m probe length

Rod probe 12 mm Approx. 0.9 kg/m probe length

FMP52
Part Weight Part Weight

Sensor Approx. 1.2 kg + weight of Rope probe 4 mm Approx. 0.5 kg/m probe length
flange

Rod probe 16 mm Approx. 1.1 kg/m probe length

FMP54
Part Weight Part Weight

Sensor - version Approx. 6.7 kg + weight of Rope probe 4 mm Approx. 0.1 kg/m probe length
XT flange

Sensor - version Approx. 7.7 kg + weight of Rod probe 16 mm Approx. 1.6 kg/m probe length
HT flange

Coax probe Approx. 3.5 kg/m probe length

84 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Materials: GT18 housing


1
(stainless steel, corrosion-
resistant) 2.1

2.2

5.2
6
5.1
7

4 3 9 8
A0036037

 51 Material; GT18 housing


1 Housing; CF3M (similar to 316L/ 1.4404)
2.1 Electronics compartment cover; CF3M (similar to 316L/ 1.4404), seals; NBR, window; glass, thread coating;
graphite-based lubricant varnish
2.2 Connection compartment cover; CF3M (similar to 316L/ 1.4404), seal; NBR, thread coating; graphite-based
lubricant varnish
3 Cover lock; 316L (1.4404), A4
4 Lock at the housing neck; 316L (1.4404), A4-70
5.1 Dummy plug, gland, adapter or plug (depending on the device version); 316L (1.4404), NBR, Viton, EPDM,
PE, PBT-GF, nickel-plated brass (CuZn)
5.2 Dummy plug, gland, adapter or plug (depending on the device version); 316L (1.4404), NBR
6 Dummy plug or M12 socket (depending on the device version); 316L (1.4404)
7 Pressure relief plug; 316L (1.4404)
8 Ground terminal; 316L (1.4404), A4 (1.4571)
9 Nameplate; 316L (1.4404), A4 (1.4571)

Materials: GT19 housing


1
(plastic)
2.1

2.2

5.2
6
5.1
7

4 9 8
A0013788

 52 Material; GT19 housing


1 Housing; PBT
2.1 Electronics compartment cover; PBT-PC, seals; EPDM, window; PC, thread coating; graphite-based lubricant
varnish
2.2 Connection compartment cover; PBT, seal; EPDM, thread coating; graphite-based lubricant varnish
4 Lock at the housing neck; 316L (1.4404), A4-70
5.1 Dummy plug, gland, adapter or plug (depending on the device version); 316L (1.4404), EPDM, PE, PBT-GF,
nickel-plated brass (CuZn), PA
5.2 Dummy plug, gland, adapter or plug (depending on the device version); 316L (1.4404), EPDM, PE, PBT-GF,
galvanized steel, nickel-plated brass (CuZn), PA
6 Dummy plug; nickel-plated brass (CuZn), M12 socket; GD-Zn nickel-plated
7 Pressure relief plug; nickel-plated brass (CuZn)
8 Ground terminal; 316L (1.4404), A4 (1.4571)
9 Adhesive nameplate; plastic

Endress+Hauser 85
Levelflex FMP51, FMP52, FMP54

Materials: GT20 housing


1
(die-cast aluminum, powder-
coated) 2.1

2.2

5.2
6
5.1
7

4 3 9 8
A0036037

 53 Material; GT20 housing


1 Housing RAL 5012 (blue); AlSi10Mg (<0.1% Cu), coating; polyester
2.1 Electronics compartment cover RAL 7035 (gray); AlSi10Mg (<0.1% Cu), seals; NBR, window; glass, thread
coating; graphite-based lubricant varnish
2.2 Connection compartment cover RAL 7035 (gray); AlSi10Mg (<0.1% Cu), seals; NBR, thread coating;
graphite-based lubricant varnish
3 Cover lock; 316L (1.4404), A4
4 Lock at the housing neck; 316L (1.4404), A4-70
5.1 Dummy plug, gland, adapter or plug (depending on the device version); 316L (1.4404), EPDM, PE, PBT-GF,
nickel-plated brass (CuZn), PA
5.2 Dummy plug, gland, adapter or plug (depending on the device version); 316L (1.4404), EPDM, PE, PBT-GF,
galvanized steel, nickel-plated brass (CuZn), PA
6 Dummy plug; nickel-plated brass (CuZn), M12 socket; GD-Zn nickel-plated
7 Pressure relief plug; nickel-plated brass (CuZn)
8 Ground terminal; 316L (1.4404), A4 (1.4571)
9 Adhesive nameplate; plastic

86 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Materials: Process Endress+Hauser supplies DIN/EN flanges and process connections with threaded connection in
connection stainless steel as per AISI 316L (DIN/ EN material number 1.4404 or 14435). With regard to
their stability-temperature property, the materials 1.4404 and 1.4435 are grouped together
under 13E0 in EN 1092-1: 2007 Tab. G.3.1-1. The chemical composition of the two materials
can be identical.

Levelflex FMP51

Threaded connection Flange


No. Material
G¾, NPT¾ G1½ NPT1½ DN40 to DN200 DN40 to DN100

1.1 316L (1.4404)


1.1 1.2 1.2 Alloy C22 (2.4602)

1.1 1.1 1.1


2 2 2 ASME: 316/316L
EN: 316L (1.4404)
JIS: 316L (1.4435)
3 3 3 3 3 4
A0013850 A0013852 A0013849 A0013854 A0013910
3 Ceramic Al2O3 99.7 %

4 Cladding: Alloy C22 (2.4602)

Levelflex FMP52

Flange Dairy pipe Tri-Clamp


No. Material Approval
EN/ASME/JIS DN50 (DIN 11851) 3" 2" 1½"

1 316L (1.4404)

1 2 ASME: 316/316L
1 1 EN: 316L (1.4404)
3 1
2
3
JIS: 316L (1.4435)
5 3 3 3 316L (1.4404)
4 4 4 4 4 4 Coating 2 mm (0.08 in): PTFE (Dyneon USP Cl.VI 1)
A0013868
A0013865
A0013866 A0013867
A0013869
TFM1600)

5 304L (1.4307)

1) Plastic components in contact with the medium have been tested in accordance with USP <88> Class VI-70°C

Levelflex FMP54

Threaded connection G1½, NPT1½ Flange


No. Material
Version HT Version XT Version HT Version XT

1 316L (1.4404)

1 1 2 316L (1.4404)
1 1
3 ASME: 316/316L
EN: 316L (1.4404)
2
7 2 7 2 JIS: 316L (1.4435)
2
7 7 3
4 Alloy C22 (2.4602)
3 4
4
5
4 5 5 316L (1.4404)
4 5 6
5 6 6 Nordlock washer: 1.4547
6 A0013883
6 A0013882

A0013880 A0013881
7 Ceramic Al2O3 99.5 %, pure graphite

Endress+Hauser 87
Levelflex FMP51, FMP52, FMP54

Materials: Probe Levelflex FMP51: rod probes

Feature 060 "Probe"

• AA: 8mm 316L • AC: 12mm 316L • AL: 12mm AlloyC • BA: 16mm 316L 500mm separable
• AB: 1/3" 316L • AD: 1/2" 316L • AM: 1/2" AlloyC • BB: 0.63in 316L 20inch separable No. Material
• BC: 16mm 316L 1000mm separable
• BD: 0.63in 316L 40inch separable

1.1 316L (1.4404)

1.1 1.2 Alloy C22 (2.4602)


1.1
2 2 Connecting bolts: Alloy C22 (2.4602)
1.1 1.2 Nordlock washer: 1.4547

3 Hexagonal-headed bolt: A4-70


1.1
Nordlock washer: 1.4547

2 4 Centering star, PEEK 1)


5 5 Centering disk, 316L (1.4404) 2)

1.1 5 Centering star, PFA 3)


4

3
A0036651 A0036585 A0013912 A0036586

1) Feature 610 "Accessory mounted" = OD "Rod centering star d=48-95mm, PEEK"


2) Feature 610 "Accessory mounted" = OA "Rod centering disk d=75mm" or OB "Rod centering disk d=45mm"
3) Feature 610 "Accessory mounted" = OE "Rod centering star d=37mm, PFA"

Levelflex FMP51: rope probes

Feature 060 "Probe"

• LA: 4mm, 316L, max. 150mm nozzle • LG: 4mm, AlloyC,


• LB: 1/6", 316L, max. 6in nozzle max. 150mm
• MB: 4mm, 316L, max. 300mm nozzle nozzle
• MD: 1/6", 316L, max. 12in nozzle • LH: 1/6", AlloyC,
max. 6in nozzle
Feature 610 "Accessory mounted" • MG: 4mm, No. Material
• OK: Centering weight AlloyC, max.
d=45mm 300mm nozzle
OC: Centering disk • OL: Centering weight • MH: 1/6", AlloyC,
without option OC max. 12in nozzle
d=75mm d=75mm
• OM: Centering weight
d=95mm

1.1 316L (1.4404)


1.1 1.1 1.1 1.2 Alloy C22 (2.4602)

2 316 (1.4401)
1.1 1.2
3 Cylinder screw:
2 1.2 A4-80

2 4 Disk: 316L (1.4404)


2 2
5 Setscrew: A4-70
5 1.2 6 Bracing screw:
1.1 1.1 1.1 A2-70
3 1.1 1.2 7 Weight: 316L
4 (1.4404)
5 7 6 1.2
A0036587
7 A0036588 A0036589

A0039226

88 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Levelflex FMP51: rope probes

Feature 060 "Probe"

• LE: 4mm, PFA>316, max. 150mm nozzle


• LF: 1/6", PFA>316, max. 6in nozzle
• ME: 4mm, PFA>316, max. 300mm nozzle
• MF: 1/6", PFA>316, max. 12in nozzle No. Material
Feature 610 "Accessory mounted"

• OK: Centering weight d=45mm


• OL: Centering weight d=75mm without option OC
• OM: Centering weight d=95mm

1.1 316L (1.4404)


1.1 1.1 2 Rope: 316 (1.4401)

Coating 0.75 mm (0.03 in): PFA


1.1
5 Setscrew: A4-70
2
6 Bracing screw: A2-70
2 2 7 Weight: 316L (1.4404)

5
1.1 1.1
1.1

7 6
7 A0036588

A0039226

Levelflex FMP51: coax probes

Feature 060 "Probe"

• UA: ...mm, coax 316L • UC: ...mm, coax AlloyC


• UB: ...inch, coax 316L • UD: ...inch, coax AlloyC
No. Material
Feature 100 "Process connection"

• GDJ: thread ISO228 G3/4 all other options


• RDJ: thread ANSI MNPT3/4

1.1 1.1 316L (1.4404)

1.2 Alloy C22 (2.4602)


1.1
1.1 1.2 2.1 Rod: 316L (1.4404)

2.2 Alloy C22 (2.4602)

3 Spacer: PFA

2.1 2.1 2.2


3 3 3

A0036590 A0036591 A0036592

Endress+Hauser 89
Levelflex FMP51, FMP52, FMP54

Levelflex FMP52

Feature 060 "Probe"

• CA: rod 16mm • OA: rope 4mm, max. 150mm nozzle


• CB: rod 0.63in • OB: rope 4mm, max. 300mm nozzle No. Material
• OC: rope 1/6", max. 6in nozzle
• OD: rope 1/6", max. 12in nozzle

1 316L (1.4404)
1 1
2 Coating 2 mm (0.08 in): PFA

2 3 Rope: 316 (1.4401)

Coating 0.75 mm (0.03 in): PFA

4 Core: 316L (1.4435)


3
2 5 Centering star, PFA 1)

2
5 2

2
A0013870
A0036593

1) Feature 610 "Accessory mounted" = OE "Rod centering star d=37mm, PFA, interface measurement"

Levelflex FMP54: rope probes

Feature 060 "Probe"


• LA: rope 4mm No. Material
• LB: rope 0.63in

Feature 610 "Accessory mounted"

• OK: Centering weight d=45mm


OC: Centering disk
• OL: Centering weight d=75mm without option OC
d=75mm
• OM: Centering weight d=95mm

1 316L (1.4404)
1 1 1 2 316 (1.4401)

3 Cylinder screw: A4-80


1
4 Disk: 316L (1.4404)
2
5 Setscrew: A4-70
2 2 2 6 Bracing screw: A2-70

7 Weight: 316L (1.4404)


5
1 1 1
3 1
4
5 7 6
A0036594
7
A0036595

A0039227

90 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Levelflex FMP54: rod and coax probes

Feature 060 "Probe"

• AE: rod 16mm • BA: rod 16mm, 500mm separable • UA: ... mm, coax
• AF: rod 0.63in • BB: rod 0.63in, 20inch separable • UB: ... inch, coax No. Material
• BC: rod 16mm, 1000mm separable
• BD: rod 0.63in, 40inch separable

1 316L (1.4404) 1)

2 Connecting bolts: Alloy C22 (2.4602)


1
Nordlock washer: 1.4547
1
1 3 Hexagonal-headed bolt: A4-70

Nordlock washer: 1.4547


2
5 4 Centering star, PEEK 2)
7 6 Centering disk, 316L (1.4404) 3)
4 7
5 Centering disk, PFA 4)
3
A0036596 1 6 Rod: 316L (1.4404)

7 Spacer: ceramic Al2O3 99.5 %


A0036597

A0036598

1) In the version with gas phase compensation, is also the material of the reference rod.
2) Feature 610 "Accessory mounted" = OD "Rod centering star d=48-95mm, PEEK"
3) Feature 610 "Accessory mounted" = OA "Rod centering disk d=75mm" or OB "Rod centering disk d=45mm"
4) Feature 610 "Accessory mounted" = OE "Rod, centering star d=37mm, PFA, interface measurement"

Endress+Hauser 91
Levelflex FMP51, FMP52, FMP54

Materials: Mounting bracket

12

12

11
10

A0015143

Mounting bracket for "remote sensor" version

No. Component Material

10 Holder 316L (1.4404)

11 Round bracket 316Ti (1.4571)

Screws/nuts A4-70

Distance sleeves 316Ti (1.4571) or 316L (1.4404)

12 Half-shells 316L (1.4404)

92 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Materials: Adapter and cable


for remote sensor

4 3

A0021722

Adapter and cable for "remote sensor" version

No. Component Material

1 Cable FRNC

2 Sensor adapter 304 (1.4301)

3 Terminal 316L (1.4404)

Screw A4-70

4 Band 316 (1.4401)

Crimp sleeve Aluminum

Nameplate 304 (1.4301)

Endress+Hauser 93
Levelflex FMP51, FMP52, FMP54

Materials: weather
protection cover 1

6
4

5
A0015473

 54 Material; weather protection cover


1 Protection cap; 316L (1.4404)
2 Molded rubber part (4x); EPDM
3 Clamping screw; 316L (1.4404) + carbon fiber
4 Bracket; 316L (1.4404)
5 Ground terminal; A4, 316L (1.4404)
6 Fillister-head screw; A4-70 + washer; A4

Operability
Operating concept Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Operating languages
• English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык (Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• Bahasa Indonesia
• tiếng Việt (Vietnamese)
• čeština (Czech)
Feature 500 in the product structure determines which of these languages is preset on delivery.

Fast and safe commissioning


• Interactive wizard with graphical user interface for guided commissioning in FieldCare/DeviceCare
• Menu guidance with short explanations of the individual parameter functions
• Standardized operation at the device and in the operating tools

94 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Integrated data memory (HistoROM)


• Adoption of data configuration when electronics modules are replaced
• Up to 100 event messages recorded in the device
• Data logging with up to 1000 logged values
• A reference signal curve is saved during commissioning for later use as a reference during
operation
Efficient diagnostic behavior increases measurement availability
• Remedial measures are integrated in plain text
• Diverse simulation options and line recorder functions
Integrated Bluetooth module (option for HART devices)
• Quick and easy setup with the SmartBlue app
• No additional tools or adapters needed
• Signal curve via SmartBlue (app)
• Encrypted single point-to-point data transmission (tested by Fraunhofer Institute) and password-
protected communication via Bluetooth® wireless technology

Endress+Hauser 95
Levelflex FMP51, FMP52, FMP54

Local operation Operation Pushbuttons Touch Control


with

Order code for Option C "SD02" Option E "SD03"


"Display;
Operation"

A0036312 A0036313

Display 4-line display 4-line display


elements white background lighting; switches to red in
event of device error

Format for displaying measured variables and status variables can be individually configured

Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.

Operating local operation with 3 push buttons (, , ) external operation via touch control; 3
elements optical keys: , , 

Operating elements also accessible in various hazardous areas

Additional Data backup function


functionality The device configuration can be saved in the display module.

Data comparison function


The device configuration saved in the display module can be compared to the current device
configuration.

Data transfer function


The transmitter configuration can be transmitted to another device using the display module.

Operation with remote


1 2
display and operating
module FHX50

A0036314

 55 FHX50 operating options


1 Display and operating module SD03, optical keys; can be operated through the glass of the cover
2 Display and operating module SD02, push buttons; cover must be removed

96 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Operation via Bluetooth® Requirements


wireless technology
1

2
A0036790

 56 Device with Bluetooth module


1 Electronics housing of the device
2 Bluetooth module

This operation option is only available for devices with Bluetooth module. There are the following
options:
• The device has been ordered with a Bluetooth module:
Feature 610 "Accessory Mounted", option NF "Bluetooth"
• The Bluetooth module has been ordered as an accessory (ordering number: 71377355) and has
been mounted. See Special Documentation SD02252F.

Operation via SmartBlue (app)

1 2 3
A0034939

 57 Operation via SmartBlue (app)


1 Transmitter power supply unit
2 Smartphone / tablet with SmartBlue (app)
3 Transmitter with Bluetooth module

Endress+Hauser 97
Levelflex FMP51, FMP52, FMP54

Remote operation Via HART protocol

9
2 3 6 8

1 4 5 7
A0044334

 58 Options for remote operation via HART protocol


1 PLC (programmable logic controller)
2 Transmitter power supply unit, e.g., RN42
3 Connection for Commubox FXA195 and AMS TrexTM device communicator
4 AMS TrexTM device communicator
5 Computer with operating tool (e.g., DeviceCare/FieldCare, AMS Device View, SIMATIC PDM)
6 Commubox FXA195 (USB)
7 Field Xpert SMT70
8 Bluetooth modem with connecting cable (e.g., VIATOR)
9 Transmitter

Via PROFIBUS PA protocol

2 3

1 PROFIBUS DP

PROFIBUS PA
T

4 4 4
A0050944

1 Segment coupler
2 Computer with PROFIusb and operating tool (e.g. DeviceCare/FieldCare )
3 PLC (programmable logic controller)
4 Transmitter
5 Additional functions (valves etc.)

98 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Via FOUNDATION Fieldbus

IN

. -
FF-HSE BT
1
FF-H1 SB 2
PS BT
LD 4

3
BT

PS

FF-H1
BT

1
2

A0017188

 59 FOUNDATION Fieldbus system architecture with associated components


1 FFblue Bluetooth modem
2 Field Xpert SFX350/SFX370
3 DeviceCare/FieldCare
4 NI-FF interface card

IN Industrial network

FF-HSE High Speed Ethernet

FF-H1 FOUNDATION Fieldbus-H1

LD Linking Device FF-HSE/FF-H1

PS Bus Power Supply

SB Safety Barrier

BT Bus Terminator

Endress+Hauser 99
Levelflex FMP51, FMP52, FMP54

Via service interface (CDI)

1 2 3

A0039148

1 Computer with FieldCare/DeviceCare operating tool


2 Commubox FXA291
3 Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface)

100 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Integration into the Tank The Endress+Hauser Tank Side Monitor NRF81 has integrated communication functions for sites
Gauging System with multiple tanks, wherein each tank can have one or more sensors, e.g. radar, spot or average
temperature sensors, capacitance probes for water detection and/or pressure sensors. As the Tank
Side Monitor supports multiple protocols, it can work with virtually all industry standard tank
gauging protocols. Optional connectivity of 4-20 mA sensors, digital inputs/outputs and analog
outputs simplifies the full integration of all sensors on the tank. Use of the proven concept of the
intrinsically safe HART bus for all on-tank sensors enables extremely low wiring costs, while at the
same time ensuring maximum safety, reliability and data availability.

1 2 3

4 5

8 6

7
A0016590

 60 The complete measuring system consists of:


1 Tankvision workstation
2 Commubox FXA195 (USB) - optional
3 Computer with operating tool (ControlCare) - optional
4 Level transmitter
5 Temperature measuring device
6 Tank Side Monitor NRF81
7 Pressure measuring device
8 Tankvision Tank Scanner NXA820

Endress+Hauser 101
Levelflex FMP51, FMP52, FMP54

SupplyCare inventory SupplyCare is a web-based operating program for coordinating the flow of material and information
management software along the supply chain. SupplyCare provides a comprehensive overview of the levels of
geographically distributed tanks and silos, for instance, providing complete transparency over the
current inventory situation, regardless of time and location.
Based on the measuring and transmission technology installed onsite, the current inventory data are
collected and sent to SupplyCare. Critical levels are clearly indicated and calculated forecasts provide
additional security for material requirements planning.
The main functions of SupplyCare:
Inventory visualization
SupplyCare determines the inventory levels in tanks and silos at regular intervals. It displays current
and historical inventory data and calculated forecasts of future demand. The overview page can be
configured to suit the user's preferences.
Master data management
With SupplyCare you can create and manage the master data for locations, companies, tanks,
products and users, as well as user authorization.
Report Configurator
The Report Configurator can be used to create personalized reports quickly and easily. The reports
can be saved in a variety of formats, such as Excel, PDF, CSV and XML. The reports can be
transmitted in many ways, such as by http, ftp or e-mail.
Event management
Events, such as when levels drop below the safety stock level or plan points, are indicated by the
software. In addition, SupplyCare can also notify pre-defined users by e-mail.
Alarms
If technical problems occur, e.g. connection issues, alarms are triggered and alarm e-mails are sent to
the System Administrator and the Local System Administrator.
Delivery planning
The integrated delivery planning function automatically generates an order proposal if a pre-set
minimum inventory level is undershot. Scheduled deliveries and disposals are monitored
continuously by SupplyCare. SupplyCare notifies the user if scheduled deliveries and disposals are not
going to be met as planned.
Analysis
In the Analysis module, the most important indicators for the inflow and outflow of the individual
tanks are calculated and displayed as data and charts. Key indicators of material management are
automatically calculated and form the basis for optimizing the delivery and storage process.
Geographical visualization
All the tanks and the tank inventories are represented graphically on a map (based on Google Maps).
The tanks and inventory situations can be filtered by tank group, product, supplier or location.
Multi-language support
The multi-language user interface supports 9 languages, thereby enabling global collaboration on a
single platform. The language and settings are recognized automatically using the browser settings.

SupplyCare Enterprise
SupplyCare Enterprise runs by default as a service under Microsoft Windows on an application server
in an Apache Tomcat environment. The operators and administrators operate the application via a
Web browser from their workstations.

102 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

2
3

10
5 7 8
9

.
A0034288

 61 Example of inventory management platform with SupplyCare Enterprise SCE30B


1 SupplyCare Enterprise (via Web browser)
2 SupplyCare Enterprise installation
3 SupplyCare Enterprise on mobile devices (via Web browser)
4 Ethernet/WLAN/UMTS
5 Fieldgate FXA42
6 Power supply 24 V DC
7 Modbus TCP via Ethernet as server/client
8 Converter from Modbus to HART Multidrop
9 HART Multidrop
10 4 x 4 to 20 mA analog input (2-wire/4-wire)

Endress+Hauser 103
Levelflex FMP51, FMP52, FMP54

Cloud-based application: SupplyCare Hosting


SupplyCare Hosting is offered as a hosting service (software as a service). Here, the software is
installed within the Endress+Hauser IT infrastructure and made available to the user in the Endress
+Hauser portal.

1
3
WWW

4
10
5 7 8
9

.
A0034289

 62 Example of inventory management platform with SupplyCare Hosting SCH30


1 SupplyCare Hosting installation in Endress+Hauser data center
2 PC workstation with Internet connection
3 Warehouse locations with Internet connection via 2G/3G with FXA42 or FXA30
4 Warehouse locations with Internet connection with FXA42
5 Fieldgate FXA42
6 Power supply 24 V DC
7 Modbus TCP via Ethernet as server/client
8 Converter from Modbus to HART Multidrop
9 HART Multidrop
10 4 x 4 to 20 mA analog input (2-wire/4-wire)

With SupplyCare Hosting, users do not need to make the initial software purchase or install and run
the IT infrastructure needed. Endress+Hauser constantly update SupplyCare Hosting and enhance
the capability of the software in conjunction with the customer. The hosted version of SupplyCare is
thus always up-to-date and can be customized to meet different customer requirements. Other
services are also offered in addition to the IT infrastructure and the software that is installed in a
secure, redundant Endress+Hauser data center. These services include defined availability of the
global Endress+Hauser Service and Support Organization and defined response times in a service
event.

104 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Certificates and approvals


Current certificates and approvals that are available for the product can be selected via the Product
Configurator at www.endress.com:
1. Select the product using the filters and search field.
2. Open the product page.
3. Select Configuration.

CE mark The measuring system meets the legal requirements of the applicable EU directives. These are listed
in the corresponding EU Declaration of Conformity together with the standards applied.
The manufacturer confirms successful testing of the device by affixing to it the CE mark.

RoHS The measuring system meets the substance restrictions of the Directive on the Restriction of the Use
of Certain Hazardous Substances 2011/65/EU (RoHS 2) and the Delegated Directive (EU) 2015/863
(RoHS 3).

RCM marking The supplied product or measuring system meets the ACMA (Australian Communications and Media
Authority) requirements for network integrity, interoperability, performance characteristics as well
as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic
compatibility are met. The products bear the RCM marking on the nameplate.

A0029561

Ex-approval The devices are certified for use in hazardous areas and the relevant safety instructions are provided
in the separate "Safety Instructions" (XA, ZD) document. Reference is made to this document on the
nameplate.
The separate "Safety Instructions" documentation (XA) containing all the relevant explosion
protection data is available from your Endress+Hauser sales organization.

Dual seal according to The devices have been designed according to ANSI/ISA 12.27.01 as dual seal devices, allowing the
ANSI/ISA 12.27.01 user to waive the use and save the cost of installing external secondary process seals in the conduit
as required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC) These
instruments comply with the North-American installation practice and provide a very safe and cost-
saving installation for pressurized applications with hazardous fluids.
Further information can be found in the Safety Instructions (XA) of the relevant devices.

Functional safety Use for level monitoring (MIN, MAX, range) up to SIL 3 (homogeneous redundancy), independently
evaluated by TÜV Rheinland in accordance with IEC 61508, refer to the "Functional Safety Manual"
SD00326F for more information.

Overfill protection WHG


DIBt Z-65.16-501

Sanitary compatibility Information regarding device versions that meet the requirements of 3A Sanitary Standard No. 74
and/or are certified by the EHEDG:
SD02503F

Suitable fittings and seals must be used to ensure hygiene-compliant design according to 3A
and EHEDG specifications.
The gap-free connections can be cleaned of all residue using the typical cleaning methods
within this industry (CIP and SIP).
Wetted non-metal parts of the FMP52 comply with FDA 21 CFR 177.1550 and USP Class VI.

Endress+Hauser 105
Levelflex FMP51, FMP52, FMP54

AD2000 • For FMP51/FMP54:


The wetted material 316L (1.4435/1.4404) corresponds to AD2000 - W2/W10.
• For FMP52/FMP55:
The pressure retaining material 316L (1.4435/1.4404) corresponds to AD2000 - W2/W10.
• Declaration of Conformity: see the product structure, feature 580, version JF.

NACE MR 0175 / ISO 15156 For FMP51, FMP54:


• The wetted, metal materials (excluding ropes) meet the requirements of NACE MR 0175 / ISO
15156.
• Declaration of Conformity: see the product structure, feature 580, version JB

NACE MR 0103 For FMP51, FMP54:


• The wetted, metal materials (excluding ropes) meet the requirements of NACE MR 0103 / ISO
17495.
• The Declaration of Conformity is based on NACE MR 0175.
The hardness and intergranular corrosion have been tested, and heat treatment (solution
annealed) has been performed. The materials used therefore meet the requirements of NACE MR
0103 / ISO 17495.
• Declaration of Conformity: see the product structure, feature 580, version JE.

ASME B31.1 and B31.3 • The design, the material used, the pressure and temperature ranges and the labeling of the devices
meet the requirements of ASME B31.1 and B31.3
• Declaration of Conformity: see the product structure, feature 580, version KV.

Pressure equipment with Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not
allowable pressure ≤ fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable
200 bar (2 900 psi) pressure.

Reasons:
According to Article 2, point 5 of EU Directive 2014/68/EU, pressure accessories are defined as
"devices with an operational function and having pressure-bearing housings".
If a pressure instrument does not have a pressure-bearing housing (no identifiable pressure chamber
of its own), there is no pressure accessory present within the meaning of the Directive.

Pressure equipment with Pressure equipment designated for application in every process fluid having a pressurized volume V
allowable pressure > of <0.1 l and a max. allowable pressure PS > 200 bar (2 900 psi) shall satisfy the essential safety
200 bar (2 900 psi) requirements set out in Annex I of the Pressure Equipment Directive 2014/68/EU. According to
Article 13 pressure equipment shall be classified by category in accordance with Annex II. The
conformity assessment of the pressure equipment shall be determined by the category I under
consideration of the above-mentioned low pressurized volume. These devices shall be provided with
CE marking.

Reasons:
• Pressure Equipment Directive 2014/68/EU, Article 13 and Annex II
• Pressure equipment directive 2014/68/EU, Commission's Working Group "Pressure", Guideline
A-05

Note:
A partial examination shall be performed for pressure instruments that are part of safety equipment
for the protection of a pipe or vessel from exceeding allowable limits (equipment with safety
function in accordance with Pressure Equipment Directive 2014/68/EU, Article 2, point 4).
The conformity assessment was performed according to module A; and proof of static strength and
fatigue strength according to EN 13445 and AD2000.
The FMP54 is not suitable for use with unstable gases at nominal pressures above 200 bar (2900
psi).

Steam boiler approval The FMP54 is approved as a limiting device for high water (HW) and low water (LW) for liquids in
vessels which are subject to the requirements of EN12952-11 and EN12953-9 (certified by TÜV
Nord).
Product structure: feature 590 "Additional approval", version LX "Steam boiler approval".

106 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

For details refer to Safety Instructions SD00349F and Planning Instructions SD01071F.
Devices with a steam boiler approval always also have a SIL approval.

Marine approval Device Marine approval 1)

DNV GL ABS LR BV KR

FMP51     

FMP52     

FMP54    - -

1) See order code for 590 "Additional approval"

Radio approval Satisfies "Part 15" of the FCC rules for an unintentional radiator. All probes meet the requirements
for a Class A digital device.
In addition, coax probes and all probes in metal vessels meet the requirements for a Class B digital
device.

CRN approval Some device versions have CRN approval. Devices are CRN approved if the following two conditions
are met:
• The device has a CSA or FM approval (product structure: feature 010 "Approval")
• The device has a CRN-approved process connection according to the following table:

Feature 100 in the product structure Approval

AAJ NPS 2" Cl. 600 RF, 316/316L flange ASME B16.5

ABJ NPS 3" Cl. 600 RF, 316/316L flange ASME B16.5

AEJ NPS 1-1/2" Cl. 150 RF, 316/316L flange ASME B16.5

AEK NPS 1-1/2" Cl. 150, PTFE>316/316L flange ASME B16.5

AEM NPS 1-1/2" Cl. 150, AlloyC>316/316L flange ASME B16.5

AFJ NPS 2" Cl. 150 RF, 316/316L flange ASME B16.5

AFK NPS 2" C. 150, PTFE>316/316L flange ASME B16.5

AFM NPS 2" Cl. 150, AlloyC>316/316L flange ASME B16.5

AGJ NPS 3" Cl. 150 RF, 316/316L flange ASME B16.5

AGK NPS 3" Cl. 150, PTFE>316/316L flange ASME B16.5

AGM NPS 3" Cl. 150, AlloyC>316/316L flange ASME B16.5

AHJ NPS 4" Cl. 150 RF, 316/316L flange ASME B16.5

AHK NPS 4" Cl. 150, PTFE>316/316L flange ASME B16.5

AJJ NPS 6" Cl. 150 RF, 316/316L flange ASME B16.5

AJK NPS 6" Cl. 150, PTFE>316/316L flange ASME B16.5

AKJ NPS 8" Cl. 150 RF, 316/316L flange ASME B16.5

AOJ NPS 4" Cl. 600 RF, 316/316L flange ASME B16.5

AQJ NPS 1-1/2" Cl. 300 RF, 316/316L flange ASME B16.5

AQK NPS 1-1/2" Cl. 300, PTFE>316/316L flange ASME B16.5

AQM NPS 1-1/2" Cl. 300, AlloyC>316/316L flange ASME B16.5

ARJ NPS 2" Cl. 300 RF, 316/316L flange ASME B16.5

ARK NPS 2" Cl. 300, PTFE>316/316L flange ASME B16.5

ARM NPS 2" Cl. 300, AlloyC>316/316L flange ASME B16.5

ASJ NPS 3" Cl. 300 RF, 316/316L flange ASME B16.5

ASK NPS 3" Cl. 300, PTFE>316/316L flange ASME B16.5

Endress+Hauser 107
Levelflex FMP51, FMP52, FMP54

Feature 100 in the product structure Approval

ASM NPS 3" Cl. 300, AlloyC>316/316L flange ASME B16.5

ATJ NPS 4" Cl. 300 RF, 316/316L flange ASME B16.5

ATK NPS 4" Cl. 300, PTFE>316/316L flange ASME B16.5

ATM NPS 4" Cl.300, AlloyC>316/316L flange ASME B16.5

AZJ NPS 4" Cl. 900 RF, 316/316L flange ASME B16.5

A6J NPS 2" Cl. 1500 RF, 316/316L flange ASME B16.5

A7J NPS 3" Cl. 1500 RF, 316/316L flange ASME B16.5

A8J NPS 4" Cl. 1500 RF, 316/316L flange ASME B16.5

GGJ Thread ISO228 G1-1/2, 316L

GIJ Thread ISO228 G1-1/2, 200bar, 316L

GJJ Thread ISO228 G1-1/2, 400bar, 316L

RAJ Thread ANSI MNPT1-1/2, 200bar, 316L

RBJ Thread ANSI MNPT1-1/2, 400bar, 316L

RGJ Thread ANSI MNPT1-1/2, 316L

TAK Tri-Clamp ISO2852 DN40-51 (2"), 3A, PTFE>316L

TDK Tri-Clamp ISO2852 DN40-51 (2"), PTFE>316L

TFK Tri-Clamp ISO2852 DN70-76.1 (3"), PTFE>316L

TJK Tri-Clamp ISO2852 DN38 (1-1/2"), PTFE>316L

TLK Tri-Clamp ISO2852 DN70-76.1 (3"), 3A, PTFE>316L

TNK Tri-Clamp ISO2852 DN38 (1-1/2"), 3A, PTFE>316L

• Process connections that do not have CRN approval are not listed in this table.
• Please refer to the product structure to discover which process connections are available for a
specific device type.
• CRN-approved devices are labeled with registration number 0F14480.5C on the nameplate.

Experience FMP5x is the upgrade model of the FMP4x series.

108 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Test, certificate Feature 580 Designation Approval


"Test,
certificate"

JA 3.1 Material documentation, wetted metal parts, EN10204-3.1 inspection • FMP51


certificate • FMP54

JB Declaration of Conformity NACE MR0175, wetted metal parts • FMP51


• FMP52
• FMP54

JD 3.1 Material certificate, pressurized parts, EN10204-3.1 inspection certificate FMP52

JE Declaration of Conformity NACE MR0103, wetted metal parts • FMP51


• FMP52
• FMP54

JF Declaration of Conformity AD2000, wetted metal parts: • FMP51


Material conformity for all metal wetted/pressurized parts according to AD2000 • FMP52
(data sheets W2, W9, W10) • FMP54

JN Ambient temperature of transmitter –50 °C (–58 °F) • FMP51


• FMP54
 Devices with this option are routine tested (start-up test at
–50 °C (–58 °F)).

KD Helium leak test, internal procedure, inspection certificate • FMP51


• FMP54

KE Pressure test, internal procedure, inspection certificate • FMP51


• FMP52
• FMP54

KG 3.1 Material certificate+PMI test (XRF), internal procedure, wetted metal parts, • FMP51
EN10204-3.1 inspection certificate • FMP54

KP Penetrant testing AD2000-HP5-3(PT), wetted/pressurized metal parts, • FMP51


inspection certificate • FMP54

KQ Penetrant testing ISO23277-1 (PT), wetted/pressurized metal parts, inspection • FMP51


certificate • FMP54

KR Penetrant testing ASME VIII-1 (PT), wetted/pressurized metal parts, inspection • FMP51
certificate • FMP54

KT Welding documentation ISO, wetted/pressurized seams, declaration • FMP51


• FMP54
Consists of:
• Welding drawing
• WPQR (Welding Procedure Qualification Record) according to ISO 14613/
ISO14614
• WPS (Welding Procedure Specifications)
• WPQ (Manufacturer Declaration for Welding Professionals' Qualifications)

KU Welding documentation ASME, wetted/pressurized seams, declaration • FMP51


• FMP54
Consists of:
• Welding drawing
• WPQR (Welding Procedure Qualification Record) according to ASME BPVC
Sect. IX
• WPS (Welding Procedure Specifications)
• WPQ (Manufacturer Declaration for Welding Professionals' Qualifications)

KV Declaration of Conformity ASME B31.3: • FMP51


The design, the material used, the pressure and temperature ranges and the • FMP52
labeling of the devices meet the requirements of ASME B31.3 • FMP54

Test reports, declarations and inspection certificates are available in electronic format in the
W@M Device Viewer:
Enter the serial number from the nameplate (www.endress.com/deviceviewer)
This concerns the options for the following order codes:
• 550 "Calibration"
• 580 "Test, certificate"

Endress+Hauser 109
Levelflex FMP51, FMP52, FMP54

Product documentation on A printed (hard copy) version of test reports, declarations and inspection certificates can optionally
paper be ordered via order code 570 "Service", option I7 "Product documentation on paper". The documents
are then supplied with the product.

110 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Other standards and • EN 60529


guidelines Degrees of protection provided by enclosures (IP code)
• EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use
• IEC/EN 61326
"Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC
requirements).
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with
analog output signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics
• NAMUR NE 107
Status classification as per NE107
• NAMUR NE 131
Requirements for field devices for standard applications
• IEC61508
Functional safety of safety-related electric/electronic/programmable electronic systems

Endress+Hauser 111
Levelflex FMP51, FMP52, FMP54

Ordering information
Ordering information Detailed ordering information is available for your nearest sales organization
www.addresses.endress.com or in the Product Configurator under www.endress.com :
1. Click Corporate
2. Select the country
3. Click Products
4. Select the product using the filters and search field
5. Open the product page
The Configuration button to the right of the product image opens the Product Configurator.

Product Configurator - the tool for individual product configuration


• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop

112 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

3-point linearity protocol The following points must be considered if option F3 (3-point linearity protocol) was selected in
feature 550 ("Calibration").
The 3 points of the linearity protocol are defined as follows, depending on the selected probe:

A
1 1

2 2

LN
B
3 3

C
A0021843

A Distance from the reference point R to the first measuring point


B Measuring range
C Distance from the probe end to the third measuring point
LN Probe length
R Reference point of measurement
1 First measuring point
2 Second measuring point (in the middle between the first and third measuring point)
3 Third measuring point

Rod or coax probe 1) Separable rod probe Rope probe Rope probe
LN ≤ 6 m (20 ft) LN > 6 m (20 ft) LN ≤ 6 m (20 ft) LN > 6 m (20 ft)

Position of 1st measuring • FMP51/FMP52/FMP54 without gas phase compensation/ A = 350 mm (13.8 in) A = 350 mm (13.8 in)
point FMP55:
A = 350 mm (13.8 in)
• FMP54 with gas phase compensation, Lref = 300 mm (11 in):
A = 600 mm (23.6 in)
• FMP54 with gas phase compensation, Lref = 550 mm (21 in):
A = 850 mm (33.5 in)

Position of 2nd measuring In the middle between the In the middle between the 1st and In the middle between the In the middle between the
point 1st and 3rd measuring 3rd measuring point 1st and 3rd measuring 1st and 3rd measuring
point point point

Position of 3rd measuring Measured from bottom: Measured from top: Measured from bottom: Measured from top:
point C = 250 mm (9.84 in) A+B = 5 750 mm (226 in) C = 500 mm (19.7 in) A+B = 5 500 mm (217 in)

Minimum measuring B ≥ 400 mm (15.7 in) B ≥ 400 mm (15.7 in) B ≥ 400 mm (15.7 in) B ≥ 400 mm (15.7 in)
range

Minimum probe length LN ≥ 1 000 mm (39.4 in) LN ≥ 1 000 mm (39.4 in) LN ≥ 1 250 mm (49.2 in) LN ≥ 1 250 mm (49.2 in)

1) Also applies for separable rods

The position of the measuring points can vary by ±1 cm (±0.04 in).

• In the case of rod and rope probes, the linearity check is performed with the entire device.
• For separable rod probes, a reference rod probe is used instead of the original probe.
• In the case of coax probes, the device electronics unit is mounted on a reference rod probe
during the test and the linearity check is performed.
• The linearity check is performed under reference operating conditions.

Endress+Hauser 113
Levelflex FMP51, FMP52, FMP54

5-point linearity protocol The following points must be considered if option F4 (5-point linearity protocol) was selected in
feature 550 ("Calibration").
The 5 points of the linearity protocol are evenly distributed over the measuring range (0% - 100%).
Empty calibration (E) and Full calibration (F) must be specified in order to define the measuring
range 5).
The following restrictions must be considered when selecting E and F:

A
E
B

F
C

A0014673

A Distance from the reference point R to the 100% mark


B Measuring range
C Distance from the probe end to the 0% mark
E Empty calibration
F Full calibration
R Reference point of measurement

Sensor Minimum distance Minimum measuring


between reference point range
R and 100% mark

FMP51 A ≥ 250 mm (10 in) B ≥ 400 mm (16 in)

FMP51 A ≥ 350 mm (14 in) B ≥ 400 mm (16 in)


Rope probe with centering rod, max. 300 mm (12 in)
nozzle height 1)

FMP52 A ≥ 250 mm (10 in) B ≥ 400 mm (16 in)

FMP52 A ≥ 350 mm (14 in) B ≥ 400 mm (16 in)


Rope probe with centering rod, max. 300 mm (12 in)
nozzle height 2)

FMP54 without gas phase compensation A ≥ 250 mm (10 in) B ≥ 400 mm (16 in)

FMP54 with gas phase compensation, Lref = 300 mm A ≥ 450 mm (18 in) B ≥ 400 mm (16 in)

FMP54 with gas phase compensation, Lref = 550 mm A ≥ 700 mm (28 in) B ≥ 400 mm (16 in)

1) Product structure: feature 060 "Probe", option MB or MD


2) Product structure: feature 060 "Probe", option OB or OD

5) If (E) and (F) are not specified, probe-dependent default values will be used instead.

114 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Type of probe Minimum distance from Maximum value for


the probe end to the 0% "Empty calibration"
mark

Rod (non-separable) C ≥ 100 mm (4 in) E ≤ 3.9 m (12.8 ft)

• Coax C ≥ 100 mm (4 in) E ≤ 5.9 m (19.4 ft)


• Rod (separable)

Rope C ≥ 1 000 mm (40 in) E ≤ 23 m (75 ft)

• In the case of rod and rope probes, the linearity check is performed with the entire device.
• In the case of coax probes, the device electronics unit is mounted on a reference rod probe
during the test and the linearity check is performed.
• The linearity check is performed under reference operating conditions.
The selected values for Empty calibration and Full calibration are only used to create the
linearity protocol. Afterwards, the values are reset to the default values specific for the probe. If
values other than the default values are required, they must be ordered as a customized
parameterization →  116.

Endress+Hauser 115
Levelflex FMP51, FMP52, FMP54

Customized If the option IJ: "Customized parameterization HART", IK "Customized parameterization PA" or IL
parameterization "Customized parameterization FF" has been selected in feature 570 "Service", presettings that differ
from the default settings can be selected for the following parameters:

Parameter Communication Picklist / value range


protocol

Setup → Unit of length • HART • in


• PA • ft
• FF • mm
• m

Setup → Empty calibration • HART 0 to 45 m (0 to 147 ft)


• PA
• FF

Setup → Full calibration • HART 0 to 45 m (0 to 147 ft)


• PA
• FF

Setup → Extended setup → Curr. output 1/2 → Damping HART 0 to 999.9 s

Setup → Extended setup → Curr. output 1/2 → Failure HART • Min


mode • Max
• Last valid value

Expert → Comm. → HART config. → Burst mode HART • Off


• On

Labeling (optional) Various types of measuring point labeling can be selected in the Product Configurator.
This includes:
• Tag
• Adhesive label
• RFID TAG
• Labeling according to DIN91406, also with NFC.
Tag name
3 lines with a maximum of 18 characters per line
Labeling in the electronic nameplate (ENP)
The first 32 characters of the tag name
Labeling on the display module
The first 12 characters of the tag name

Application packages
Heartbeat Diagnostics Availability
Available in all device versions.

Function
• Continuous self-monitoring of the device.
• Diagnostic messages output to
• the local display.
• an asset management system (e.g. FieldCare/DeviceCare).
• an automation system (e.g. PLC).

Advantages
• Device condition information is available immediately and processed in time.
• The status signals are classified in accordance with VDI/VDE 2650 and NAMUR recommendation
NE 107 and contain information about the cause of the error and remedial action.

Detailed description
See the "Diagnostics and troubleshooting" section of the Operating Instructions for the device.

116 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Heartbeat Verification Availability


Available for the following versions of feature 540 "Application package":
• EH
Heartbeat Verification + Monitoring
• EJ
Heartbeat Verification

Device functionality checked on demand


• Verification of the correct functioning of the measuring device within specifications.
• The verification result provides information about the condition of the device: Passed or Failed.
• The results are documented in a verification report.
• The automatically generated report supports the obligation to demonstrate compliance with
internal and external regulations, laws and standards.
• Verification is possible without interrupting the process.

Advantages
• No onsite presence is required to use the function.
• The DTM triggers verification in the device and interprets the results. No specific knowledge is
required on the part of the user.
(DTM: Device Type Manager; controls device operation via DeviceCare, FieldCare or a DTM-based
process control system.)
• The verification report can be used to prove quality measures to a third party.
• Heartbeat Verification can replace other maintenance tasks (e.g. periodic check) or extend the
test intervals.

SIL/WHG-locked devices
Only relevant for devices with SIL or WHG approval: order code 590 ("Additional approval"), option
LA ("SIL") or LC ("WHG").
• The Heartbeat Verification module contains a wizard for the proof test which must be performed
at appropriate intervals for the following applications:
• SIL (IEC61508/IEC61511)
• WHG (German Water Resources Act)
• To perform a proof test, the device must be locked (SIL/WHG locking).
• The wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.
In the case of SIL-locked and WHG-locked devices, it is not possible to perform verification
without taking additional measures (e.g. bridging the output current) because the output
current must be simulated (Increased safety mode) or the level must be approached manually
(Expert mode) during subsequent re-locking (SIL/WHG locking).

Detailed description
SD01872F

Heartbeat Monitoring Availability


Available for the following versions of feature 540 "Application package":
EH
Heartbeat Verification + Monitoring

Function
• In addition to the verification parameters, the corresponding parameter values are also logged.
• Existing measured variables, such as the echo amplitude, are used in the Foam detection and
Build-up detection wizards.
In the Levelflex FMP5x, the Foam detection and Build-up detection wizards cannot be used
together.

Endress+Hauser 117
Levelflex FMP51, FMP52, FMP54

"Foam detection" wizard


• The Heartbeat Monitoring module contains the Foam detection wizard.
• This wizard is used to configure automatic foam detection, which detects foam on the product
surface on the basis of the reduced signal amplitude. Foam detection can be linked to a switch
output in order to control a sprinkler system, for example, which dissolves the foam.
• This wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.

"Build-up detection" wizard


• The Heartbeat Monitoring module contains the Build-up detection wizard.
• The wizard is used to configure automatic buildup detection, which detects the buildup of deposits
on the probe on the basis of the reduced signal amplitude.
• This wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.

Advantages
• Early detection of changes (trends) to ensure plant availability and product quality.
• Use of information for the proactive planning of measures (e.g. cleaning/maintenance).
• Identification of undesirable process conditions as the basis to optimizing the facility and the
processes.
• Automated control of measures to remove foam or buildup.

Detailed description
SD01872F

Accessories
Device-specific accessories Weather protection cover
The weather protection cover can be ordered together with the device via the "Accessory enclosed"
product structure.
222 (8.74)

A0015466

 63 Height. Unit of measurement mm (in)

118 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

299 (11.8) 202 (7.95)

226 (8.9)
11
12.5°×10 7.2
°

! 6.6
20 ×
A0015472

 64 Dimensions. Unit of measurement mm (in)

Material
316L
Order number for accessories:
71162242

Mounting bracket for electronics housing

Accessories Description

Mounting bracket for A B


electronics housing

ø42...60
(1.65...2.36)
52 (2)

122 (4.8) 127...140


161 (6.34) (5...5.51)
162...175
(6.38...6.89)
(3.4)
(2.8)

86
70

A0014793

 65 Mounting bracket for electronics housing; engineering unit: mm (in)


A Wall mounting
B Post mounting

 With "remote sensor" device versions (see feature 060 in the product structure), the mounting bracket is included in
the scope of delivery. However, it can also be ordered separately as an accessory (order number: 71102216).

Endress+Hauser 119
Levelflex FMP51, FMP52, FMP54

Rod extension / centering device

Accessories Description

Rod extension / centering


device HMP40
• Suitable for:
FMP54
• Permitted temperature
at lower edge of nozzle:
• without centering
disk: 1
no restriction
• with centering disk:
-40 to +150 °C
(-40 to +302 °F)
• Additional information:
SD01002F
2 3
A0013597

1 Nozzle height
2 Extension rod
3 Centering disk

010 Approval:

A Non-hazardous area

M FM DIP Cl.II Div.1 Gr.E-G N.I., Zone 21,22

P CSA DIP Cl.II Div.1 Gr.G + coal dust N.I.

S FM Cl.I, II, III Div.1 Gr.A-G N.I., Zone 0,1,2,20,21,22

U CSA Cl.I, II, III Div.1 Gr.A-G N.I., Zone 0,1,2

1 ATEX II 1G

2 ATEX II 1D

020 Extension rod; height of nozzle:

1 115mm; 150-250mm / 6-10"

2 215mm; 250-350mm / 10-14"

3 315mm; 350-450mm / 14-18"

4 415mm; 450-550mm / 18-22"

9 Special version; TSP no. to be specified

030 Centering disk:

A Not selected

B DN40 / 1-1/2", inside-d. = 40-45mm, PPS

C DN50 / 2", inside-d. = 50-57mm, PPS

D DN80 / 3", inside-d. = 80-85mm, PPS

E DN80 / 3", inside-d. = 76-78mm, PPS

G DN100 / 4", inside-d. = 100-110mm, PPS

H DN150 / 6", inside-d. = 152-164mm, PPS

J DN200 / 8", inside-d. = 210-215mm, PPS

K DN250 / 10", inside-d. = 253-269mm, PPS

Y Special version; TSP no. to be specified

120 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mounting kit, insulated

Accessories Description

Mounting kit, insulated


suitable for
• FMP51
• FMP54

A0013586

 66 Scope of delivery of mounting kit:


1 Insulation sleeve
2 Eye bolt

To secure rope probes so that they are reliably insulated.


Maximum process temperature: 150 °C (300 °F)
For rope probes 4 mm (¹⁄₆ in) or 6 mm (1/4 in) with PA>steel :
• Diameter D = 20 mm (0.8 in)
• Order number: 52014249
For rope probes 6 mm (¹⁄₄ in) or 8 mm (1/3 in) with PA>steel:
• Diameter D = 25 mm (1 in)
• Order number: 52014250
Due to the risk of electrostatic charge, the insulation sleeve is not suitable for use
in hazardous areas! In this case, the probe must be secured so that it is reliably
grounded.

 The mounting kit can also be ordered directly with the device (Levelflex
product structure, feature 620 "Accessory enclosed", version PG "mounting kit,
insulated, rope").

Endress+Hauser 121
Levelflex FMP51, FMP52, FMP54

Centering star

Accessories Description

Centering star PEEK


 48-95 mm
suitable for

7 (0.28)
• FMP51 10 (0.39)
• FMP54

3)
.6
(0
6
ø1
< 50 (1.97)
ø48 (1.89)
ø75 (2.95)
ø95 (3.74)
A0014576

The centering star is suitable for probes with a rod diameter of 16 mm (0.6 in) and
can be used in pipes from DN50 to DN100. The markings make it easer to cut to
size, ensuring that the centering star can be adjusted to the pipe diameter. See also
Operating Instructions SD02316F.
• Material of centering star: PEEK
• Material of retaining rings: PH15-7Mo (UNS S15700)
• Permitted process temperature range: –60 to +250 °C (–76 to +482 °F)
• Order number: 71069064

 Ifbypass
the centering star is used in a bypass, it must be positioned below the lower
outlet. This must be taken into account when choosing the probe
length. In general, the centering star should not be mounted more than 50
mm (1.97") above the probe tip. It is advised not to use the PEEK centering
star in the measuring range of the rod probe.

 The PEEK centering star can also be ordered directly with the device
(Levelflex product structure, feature 610 "Accessory mounted", option OD). In
this case, it is not secured to the rod using the retaining rings, but instead is
secured using a hexagonal-headed bolt (A4-70) and a Nord Lock washer
(1.4547) at the tip of the probe rod.

122 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Accessories Description

Centering star PFA

10 (0.39)
•  16.4 mm (0.65 in)
•  37 mm (1.46 in)
suitable for
• FMP51
• FMP52
• FMP54

A: ø16.4 (0.65)
B: ø37 (1.46)
A0014577

A For probe 8 mm (0.3 in)


B For probes 12 mm (0.47 in) and 16 mm (0.63 in)

The centering star is suitable for probes with a rod diameter of 8 mm (0.3 in),
12 mm (0.47 in) and 16 mm (0.63 in) (including coated rod probes) and can be
used in pipes from DN40 to DN50. See also Operating Instructions
BA00378F/00/A2.
• Material: PFA
• Permitted process temperature range: –200 to +250 °C (–328 to +482 °F)
• Order number
• Probe 8 mm (0.3 in) : 71162453
• Probe 12 mm (0.47 in): 71157270
• Probe 16 mm (0.63 in): 71069065

 product
The PFA centering star can also be ordered directly with the device (Levelflex
structure, feature 610 "Accessory mounted", option OE).

Endress+Hauser 123
Levelflex FMP51, FMP52, FMP54

Accessories Description

Centering star PEEK, ⌀

5,5 (0.22)
48 to 95 mm
(1.9 to 3.7 in)
suitable for Ø4 (0.16)
• FMP51
• FMP52
• FMP54

)
)

.2
89
(1.

(0
48

5
Ø

Ø75 (2.95)

Ø9
5(
3.7
4)

A0035182

The centering star is suitable for probes with a rope diameter of 4 mm (¹⁄₆ in)
(including coated rope probes). See also Operating Instructions SD01961F.
• Material: PEEK
• Permitted process temperature range: –60 to +250 °C (–76 to +482 °F)
• Order number
• 71373490 (1x)
• 71373492 (5x)

124 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Centering weight

Accessories Description

Centering weight 316L 0°


 45 mm (1.77 in)
12

suitable for
• FMP51
• FMP54

Ø20 (0.79)

Ø30 (1.18)

12 (0.47)
60 (2.36)
Ø5,5 (0.22)

12 (0.47)
Ø11 (0.43)
Ø45 (1.77)

A0038923

The centering weight is suitable for probes with a rope diameter of 4 mm (¹⁄₆ in)
and can be used in DN50/2" pipes.
Material: 316L
The centering weight can be ordered directly with the device (product structure
Levelflex) or as a probe without a process connection (product structure XPF0005-)
using feature 610 "Accessory mounted", version OK (for pipe DN50/2").

Endress+Hauser 125
Levelflex FMP51, FMP52, FMP54

Accessories Description

Centering weight 316L 0°


12
•  75 mm (2.95 in)
•  95 mm (3.7 in)
suitable for
• FMP51
• FMP54

Ø20 (0.79)

ØA

12 (0.47)
Ø30 (1.18)

30 (1.18)
12 (0.47)
Ø5,5 (0.22) Ø8 (0.31)
11 (0.43)
ØB
A0038924

A = 52.5 mm (2.07 in) for DN80/3" pipe


= 62.5 mm (2.47 in) for DN100/4" pipe
B = 75 mm (2.95 in) for DN80/3" pipe
= 95 mm (3.7 in) for DN100/4" pipe

The centering weight is suitable for probes with a rope diameter of 4 mm (¹⁄₆ in)
and can be used in DN80/3" or DN100/4" pipes.
Material: 316L
The centering weight can be ordered directly with the device (product structure
Levelflex) or as a probe without a process connection (product structure XPF0005-)
using feature 610 "Accessory mounted", version OL (for pipe DN80/3") or OM (for
pipe DN100/4").

Remote display FHX50

1.
3.
2.

4.

A0019128

126 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Technical data
• Material:
• Plastic PBT
• 316L/1.4404
• Aluminum
• Degree of protection: IP68 / NEMA 6P and IP66 / NEMA 4x
• Suitable for display modules:
• SD02 (push buttons)
• SD03 (touch control)
• Connecting cable:
• Cable supplied with device up to 30 m (98 ft)
• Standard cable supplied by customer up to 60 m (196 ft)
• Ambient temperature:–40 to 80 °C (–40 to 176 °F)
• Ambient temperature (option): –50 to 80 °C (–58 to 176 °F) 6)
Ordering information
• If the remote display is to be used, the device version "Prepared for display FHX50" must be
ordered.
For the FHX50, the option "Prepared for display FHX50" must be selected under "Measuring device
version".
• If a measuring device has not been ordered with the version "Prepared for display FHX50" and is to
be retrofitted with an FHX50, the version "Not prepared for display FHX50" must be ordered for the
FHX50 under "Measuring device version". In this case, a retrofit kit for the device is supplied with
the FHX50. The kit can be used to prepare the device so that the FHX50 can be used.
Use of the FHX50 may be restricted for transmitters with an approval. A device may only be
retrofitted with the FHX50 if the option "Prepared for FHX50" is listed in the associated Safety
Instructions (XA) under Basic specifications, "Display, operation".
Also pay attention to the Safety Instructions (XA) of the FHX50.
Retrofitting is not possible on transmitters with:
• An approval for use in areas with flammable dust (dust ignition-proof approval)
• Type of protection Ex nA
For details, see the "Special Documentation" SD01007F

Overvoltage protection
The surge arrester for loop-powered devices can be ordered together with the device via the
"Accessory mounted" section of the product order structure.
The surge arrester can be used for loop-powered devices.
• 1-channel devices - OVP10
• 2-channel devices - OVP20

A0021734

6) This range applies if the option JN "Transmitter ambient temperature" –50 °C (–58 °F) was selected in feature 580 "Test, Certificate". If the
temperature is permanently below –40 °C (–40 °F), higher failure rates can be expected.

Endress+Hauser 127
Levelflex FMP51, FMP52, FMP54

Technical data
• Resistance per channel: 2 × 0.5 Ωmax
• Threshold DC voltage: 400 to 700 V
• Threshold surge voltage: < 800 V
• Capacitance at 1 MHz: < 1.5 pF
• Nominal leakage current (8/20 μs): 10 kA
• Suitable for conductor cross-sections: 0.2 to 2.5 mm2 (24 to 14 AWG)
If retrofitting:
• Order number for 1-channel devices (OVP10): 71128617
• Order number for 2-channel devices (OVP20): 71128619
• The use of the OVP module may be restricted depending on the transmitter approval. A device
may only be retrofitted with the OVP module if the option NA (overvoltage protection) is listed
under Optional specifications in the Safety Instructions (XA) associated with the device.
• In order to keep the necessary safety distances when using the surge arrester module, the housing
cover also needs to be replaced when the device is retrofitted.
Depending on the housing type, the suitable cover can be ordered using the following order
number:
• Housing GT18: 71185516
• Housing GT19: 71185518
• Housing GT20: 71185517
For details, see the "Special Documentation" SD01090F

Bluetooth module BT10 for HART devices


The Bluetooth module BT10 can be ordered together with the device via the "Accessory mounted"
section of the product order structure.

A0036493

Technical data
• Quick and easy setup with the SmartBlue app
• No additional tools or adapters needed
• Signal curve via SmartBlue (app)
• Encrypted single point-to-point data transmission (tested by Fraunhofer Institute) and password-
protected communication via Bluetooth® wireless technology
• Range under reference conditions:
> 10 m (33 ft)
• When the Bluetooth module is used, the minimum supply voltage of the device increases by up to
3 V.
If retrofitting:
• Order number: 71377355
• The use of the Bluetooth module may be restricted depending on the transmitter approval. A
device may only be retrofitted with the Bluetooth module if the option NF (Bluetooth module) is
listed under Optional specifications in the Safety Instructions (XA) associated with the device.
For details, see the "Special Documentation" SD02252F

128 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Communication-specific Commubox FXA195 HART


accessories For intrinsically safe HART communication with FieldCare via the USB interface
For details, see "Technical Information" TI00404F

Commubox FXA291
Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser Common Data
Interface) and the USB port of a computer or laptop
Order number: 51516983
For details, see "Technical Information" TI00405C

HART Loop Converter HMX50


Is used to evaluate and convert dynamic HART process variables to analog current signals or limit
values
Order number: 71063562
For details, see "Technical Information" TI00429F and Operating Instructions BA00371F

WirelessHART adapter SWA70


• Is used for the wireless connection of field devices
• The WirelessHART adapter can be easily integrated into field devices and existing infrastructures,
offers data protection and transmission safety and can be operated in parallel with other wireless
networks
For details, see Operating Instructions BA00061S

Fieldgate FXA42
Fieldgates enable communication between connected 4 to 20 mA, Modbus RS485 and Modbus TCP
devices and SupplyCare Hosting or SupplyCare Enterprise. The signals are transmitted either via
Ethernet TCP/IP, WLAN or mobile communications (UMTS). Advanced automation capabilities are
available, such as an integrated Web-PLC, OpenVPN and other functions.
For details, see "Technical Information" TI01297S and Operating Instructions BA01778S.

SupplyCare Enterprise SCE30B


Inventory management software that displays the level, volume, mass, temperature, pressure,
density or other parameters of tanks. The parameters are recorded and transmitted by means of
gateways like Fieldgate FXA42, Connect Sensor FXA30B or other gateway types.
This Web-based software is installed on a local server and can also be visualized and operated with
mobile terminals such as a smartphone or tablet.
For details, see Technical Information TI01228S and Operating Instructions BA00055S

SupplyCare Hosting SCH30


Inventory management software that displays the level, volume, mass, temperature, pressure,
density or other parameters of tanks. The parameters are recorded and transmitted by means of
gateways like Fieldgate FXA42, Connect Sensor FXA30B or other gateway types.
SupplyCare Hosting is offered as a hosting service (Software as a Service, SaaS). In the Endress
+Hauser portal, the user is provided with the data over the Internet.
For details, see Technical Information TI01229S and Operating Instructions BA00050S

Field Xpert SFX350


Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It enables efficient
device configuration and diagnostics for HART and FOUNDATION Fieldbus devices in the non-Ex
area.
For details, see Operating Instructions BA01202S

Field Xpert SFX370


Field Xpert SFX370 is a mobile computer for commissioning and maintenance. It enables efficient
device configuration and diagnostics for HART and FOUNDATION Fieldbus devices in the non-Ex
area and the Ex area.
For details, see Operating Instructions BA01202S

Endress+Hauser 129
Levelflex FMP51, FMP52, FMP54

Service-specific accessories DeviceCare SFE100


Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus field devices
Technical Information TI01134S

FieldCare SFE500
FDT-based plant asset management tool
It can configure all smart field units in your system and helps you manage them. By using the status
information, it is also a simple but effective way of checking their status and condition.
Technical Information TI00028S

System components Memograph M graphic data manager


The Memograph M graphic data manager provides information on all the relevant process variables.
Measured values are recorded correctly, limit values are monitored and measuring points analyzed.
The data are stored in the 256 MB internal memory and also on a SD card or USB stick.
Technical Information TI00133R and Operating Instructions BA00247R

RN221N
Active barrier with power supply for safe separation of 4 to 20 mA standard signal circuits. Offers
bidirectional HART transmission.
Technical Information TI00073R and Operating Instructions BA00202R

RN221
Supply unit for powering two 2-wire measuring devices solely in the non-Ex area. Bidirectional
communication is possible via the HART communication jacks.
Technical Information TI00081R and Brief Operating Instructions KA00110R

Supplementary documentation
The following document types are available in the Downloads section of the Endress+Hauser website
(www.endress.com/downloads):
For an overview of the scope of the associated Technical Documentation, refer to the following:
• W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the
nameplate
• Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
matrix code on the nameplate

Brief Operating Instructions Guide that takes you quickly to the 1st measured value
(KA) The Brief Operating Instructions contain all the essential information from incoming acceptance to
initial commissioning.

Operating Instructions (BA) Your reference guide


These Operating Instructions contain all the information that is required in various phases of the life
cycle of the device: from product identification, incoming acceptance and storage, to mounting,
connection, operation and commissioning through to troubleshooting, maintenance and disposal.

Safety Instructions (XA) Depending on the approval, the following Safety Instructions (XA) are supplied with the device. They
are an integral part of the Operating Instructions.
The nameplate indicates the Safety Instructions (XA) that are relevant to the device.

Functional Safety Manual Depending on the SIL approval, the Functional Safety Manual (FY/SD) is an integral part of the
(FY/SD) Operating Instructions and applies in addition to the Operating Instructions, Technical Information
and ATEX Safety Instructions.
The different requirements that apply for the protective function are described in the
Functional Safety Manual (FY / SD).

130 Endress+Hauser
*71603692*
71603692

www.addresses.endress.com