eti max3000血凝使用手册
eti max3000血凝使用手册
We reserve the right to make changes in the course of technical development without previous
notice.
Software-copyrights
The software for the ETI-MAX 3000 has been licensed to DiaSorin
s.r.l. for worldwide distribution. The ETI-MAX 3000 software is the
intellectual property of STRATEC Biomedical Systems AG.
Intellectual property rights shall remain with STRATEC Biomedical
Systems AG.
You are entitled to use the ETI-MAX 3000 software and the printed
accompanying material at your place of work only.
Neither this manual nor any parts of it may be duplicated or transmitted in any way
without the written approval of DiaSorin s.r.l.
User´s Guide
1 System Overview
2 Performing a Test Run
3 Service and Maintenance
4 Error Messages
Contents
Contents
1 System Components.........................................................................................1
1.1 Individual Instrument Modules.....................................................................2
1.2 Connections ................................................................................................7
1.3 LED Displays ...............................................................................................8
1.4 PC Requirements ........................................................................................8
2 Software Structure and Operation...................................................................9
2.1 Basic Information about Operation..............................................................9
2.2 Menu Guidance .........................................................................................10
2.3 File Types..................................................................................................12
3 Performing a Test Run....................................................................................13
3.1 Definition and Start of Worklist..................................................................13
3.1.1 System start ......................................................................................13
3.1.2 Initialization........................................................................................13
3.1.3 Input of patient data ..........................................................................13
3.1.4 Create new worklist...........................................................................16
3.1.5 Defining several plates......................................................................18
3.1.6 Loading samples and reagents .........................................................20
3.1.7 Load other required resources ..........................................................22
3.1.8 Reagent volume check .....................................................................22
3.1.9 Plate loading......................................................................................23
3.1.10 Starting and Processing a Run .........................................................24
3.2 Result Report ............................................................................................25
3.3 Re-Loading................................................................................................28
3.3.1 Re-loading patient samples without barcode ....................................28
3.3.2 Re-loading of patient samples with barcode .....................................30
4 Loading - Unloading - Disposal .....................................................................33
4.1 Loading (instrument at hold) .....................................................................33
4.2 Unloading ..................................................................................................37
4.3 Disposal.....................................................................................................38
5 Service and Maintenance ...............................................................................39
5.1 Maintenance Schedule..............................................................................39
5.1.1 Daily Maintenance Procedure ...........................................................39
5.1.2 Weekly Maintenance Procedure .......................................................40
5.1.3 Monthly Maintenance Procedure.......................................................40
5.2 Service Plan (to be performed by authorized service engineers)..............41
5.3 Cleaning and Disinfecting the Instrument .................................................41
5.3.1 Cleaning and Disinfecting the Housing Surface................................42
5.3.2 Cleaning and Disinfecting the Interior of the Instrument ...................42
5.3.3 Maintenance of Pipettor System .......................................................42
5.3.4 Maintenance of Washer System .......................................................42
8 Contact .............................................................................................................65
Typographical Conventions
Menu items are printed in bold-face type, separated from the menu by a vertical line.
Example: File | New
III
ETI-MAX 3000 – User´s Guide
Safety Instructions
Safety Instructions
The user has to observe the information and warnings contained in this
operating manual in order to ensure safe operation of the instrument.
• If you can see that the unit has become unsafe to use, switch it off and
disconnect it from the power supply.
• If liquid gets inside the instrument (drawer/tip tray), disconnect the power
cord and clean the respective parts using appropriate cleaners (in
particular those spots that are crucial to operation). See also chapter 5
Maintenance.
• Spare fuses must match the values specified by the system
manufacturer.
Any service and maintenance work not described in the Operator's Manual
must be performed by authorized service engineers and technicians.
1 System Components
Overview The ETI-MAX 3000 is a fully automated microtiter plate analyzer performing
the complete sample processing (sample pre-dilutions, sample and reagent
dispensing, incubations, wash processes, plate transports) as well as the
photometric measurement and evaluation. The instrument is controlled via
the Windows PC software ETI-MAX 3000. This software, which was
specifically designed for this purpose, allows the user to process the pre-
defined assays of DiaSorin as well as assays programmed by the user. The
clear structure with intuitive user-guidance allows simple and quick operation
of daily routine jobs as well as programming of user-specific assays.
9 8
1 2 3 4
7 6
5
5
1a Dilution tubes
1b Tip racks
2 Patient sample &
reagent unit
3a Tip ejection station
3b Pipettor wash station
3c Pipettor station
4 Test plate 3c
compartment with
plate transport unit
1a 4
1b
2
3a 3b
Pin Groove
The software specifies which track is to be used for the respective rack. This
is indicated by a red LED. A reagent rack occupying 3 tracks must be
inserted such that the contact tappet is in contact with the lit up LED.
Each rack has to be inserted up to the limit stop. The respective LED on the
rear panel goes off (Loading/Unloading see chapter 4).
Reloading of patient and reagent racks is possible when the instrument is in
the incubation mode.
Keep the door of this unit closed during a run. It may be opened only
for reloading (see chapter 3.3). Opening the door while transfer steps
are being performed results in an immediate stop of the pipettor and
entry in the event log.
Incubator
Incubator
Photometer 6
7 Incubator
Waste
Wash buffer
Wash unit
Bottom drawer
5a) Container with wash buffer and waste bottles (Figure 5, (5a))
A maximum of 3 bottles (1 x 1 liter, 2 x 2 liters) can be used for various
buffers; another position has to be reserved for the cleaning fluid (distilled
water) to clean the washer head.
The connection fitting consists of 4 color-coded connection pairs: one tubing
and one level sensor each per bottle (Loading see chapter 4.1).
Two waste bottles are available for the wash unit. One waste bottle contains
the liquid waste which is pumped to the waste output on the rear panel of the
instrument. The second bottle serves as overflow protection.
5b) Wash unit for test plates with window (Figure 5, (5b))
The test plates are moved into the wash unit by the system. The wash
process and the height adjustment of the pipettor can be checked via the
window.
A plastic hood protects the visible working area. The closed position of
this hood is monitored by a contact switch. The ETI-MAX 3000 cannot
be operated without this hood, in order to avoid getting in contact with
the working area during a run. If these safety precautions are not
observed strictly, the operator may get hurt or contract an infection, or
the instrument may get damaged.
1.2 Connections
10 11
12
13
18
14
15
16
17
System
liquid Waste
The level sensors inform you about the status of the respective
container signal (full/empty). A corresponding message appears on
the screen. Before starting a run, check both containers.
The LED displays on the instrument front panel (next to the left and right
door) have the following meaning:
Left
o Power green, when turned on, otherwise off.
o Ready green, when ready for a run, otherwise off (e.g. in case
of Error or during initialization)
o Error yellow, when instrument error has occurred, otherwise
off.
o Drawer green, when left drawer has been cleared for loading and
unloading of dilution plates and tips, otherwise off.
Right
o Load green, when plate compartment can be opened,
otherwise off.
o Instrument green, when instrument drawer (bottom drawer, see
Figure 1 (5)) can be opened.
1.4 PC Requirements
2. Some windows show an explorer tree on the left-hand side and the
selected item on the right-hand side. Clicking on the plus sign of a folder
on the explorer tree opens the respective options, clicking on the minus
sign closes the respective folder. Clicking on an item shows the
respective page of the corresponding file.
This presentation is used for the windows:
Worklist To define the layout of test plates.
Assay Protocol To define assay process steps (pipetting, washing,
incubating, measurement, etc).
Panel Definition To assign one or several assays to one or several
plates.
*.apm File polling format setting of the import file for host systems
*.asy Assay protocol files
*.mpc Coordinate files for e.g. dilution plates
*.wor Worklist files
*.res Result files
*.ver Verification report files
*.txt Export files
*.txt ASCII patient data import files can be downloaded from a host
computer to the ETI-MAX 3000 software (patient with
associated assays)
*.tst Self-test files with information about the self-tests that were
performed
*.spe Spectrum files (contain data of a spectrum acquisition)
*.log Files documenting daily data communication PC <-> ETI-MAX
3000
*.rac Coordinate files for the racks.
3.1.2 Initialization
The system is initialized and checks all instrument
functions. The results of this instrument check is then
displayed on the screen.
Select Add Patient(s) to open the following dialog box (b). (b)
(c)
The software returns to the Patient Editor and shows the
patients that were entered new (c).
Repeat this step, if necessary, for further patient ID's.
Once you have defined all plates, confirm the dialog box
with OK.
System Status shows the working area including status System status
messages. Depending on the type of loading (barcoded,
non-barcoded samples or host connection), this
presentation differs:
Barcoded samples: sample and reagent racks are
depicted.
Non-barcoded samples and host connection: no racks are
depicted yet.
Active event log shows the communication protocol, Active event log
listing all steps of the ETI-MAX 3000.
Job list shows the allocation of patient ID’s and tests to be Job list
performed in a list.
Upon complete processing of a plate, the results are saved to a result file and displayed on the screen, which
can be printed in individual parts or as a whole.
Click on the plus sign in front of file name to open the folder tree of the result file; click on the individual folder
to display its contents in the field to the right.
By closing the result file the user can decide if the results
should be send to a host computer.
Before sending results to the host computer, individual
results can be modified by the user (all modifications are
flagged).
3.3 Re-Loading
To insert a new plate during a running worklist, this can be done by re-loading.
This is done via the Add Plate, Add Assay and Add
Patient buttons, as described in sections 3.1.4 to 3.1.5.
After definition of the additional plate panel, click OK.
Refill disposable tips, dilution tubes, wash buffer and Display after loading
distilled water as shown on the screen (see section
3.1.7).
This is done via the Add Plate, Add Assay and Add
Patient buttons, as described in sections 3.1.4 to 3.1.5.
After definition of the additional plate panel, click OK.
Pin Groove ! Pull very left drawer out. The LED Drawer must light up green.
! Place loaded tip rack – as illustrated in the drawing to the left - into the
holding device of the instrument, so that the pin sits in the groove of the
tip rack.
! Push in drawer up to the limit stop (audible click), overcoming a slight
resistance.
Please observe the position of the tip rack defined in the Load
Inserting a tip rack dialog box. Otherwise, the tip may crash!
with holding frame
The software shows all resources prior to allocation in the area Unallocated
resources.
All automatically allocated samples and reagents are crossed, all manually
allocated samples and reagents are not crossed.
Loaded reagents which are identified by a barcode and which are not needed
for the current worklist are presented in black.
Reagents that have not been identified (e.g. due to missing or illegible
barcodes) are not displayed in the reagent area, the position remains white.
The symbols of this reagents can be allocated from the field Unallocated
resources to the respective positions with the mouse.
If reagents were not identified or placed without barcode, the user has
to make sure that the manually allocated positions correspond to the
actual placement on the reagent rack. Manually allocated reagents are
identified in the result printout and in the log file.
Contact ! Insert rack on that track that is marked by the red LED on the rear panel.
tappet Place rack in front of the track and then push evenly up to the limit stop
(with the tappet in the contact opening on the rear panel).
! The LED goes off and jumps to the next position that can be loaded. (A
red flashing LED indicates that the barcodes have not been read
correctly. Pull out rack and insert it once more.) A graphical
representation of the correctly inserted rack appears on the screen. From
the pool of unallocated resources in the right-hand side of the screen, the
loaded samples are allocated to the rack in accordance with the identified
loading. Click on a sample to check the patient ID.
! Insert the next racks as described.
! After loading, close the door again (audible click).
Sample rack
4.2 Unloading
4.3 Disposal
Disposable tips
! Disposable tips are automatically ejected at the tip ejection station. Via a
slide they go directly into the plastic waste container at the instrument
front panel.
The user may perform the maintenance work described in this chapter. Any
service work described in the chapter 5.2 or not described in the operating
manual must be performed by authorized service engineers.
End of Work
Procedure
Disinfectants must not come into contact with bearings and guides, as
otherwise the greasy film may dissolve!
Disinfectants must not be used in the vicinity of circuit boards and light
barriers!
Clean and disinfect the following instrument parts of the ETI-MAX 3000 in
accordance with the maintenance plan:
• Housing surface
Interior of instrument: magazine for tip racks and dilution tubes (1),
sample and reagent unit (2), test plate area (3), test plate compartment
surface (4), drawer with wash unit and wash buffer (5)
• Manifold in washer unit (5)
Disinfectants must not come into contact with bearings and guides, as
otherwise the greasy film may dissolve!
Disinfectants must not be used in the vicinity of circuit boards and light
barriers!
• Open drawer.
• Using the supplied cleaning needle, clean the 8 dispense and the 8
aspirate needles of the manifold.
The photometer works with a special light source which – if faulty – can be
replaced by the user. If the bulb of the photometer is faulty, “Bulb blown!” is
displayed during the selftest for the photometer. The instrument is shipped
with a spare halogen bulb, which has to be replaced as described below.
Spare halogen bulbs can be purchased from DiaSorin.
• Pull bulb upward, as shown in Figure 5-5 and take it out of its holder.
(Hot! Be careful!)
• Pull bulb out of the connection socket and replace it by a new bulb. (Do
not touch the reflecting surface of the bulb!)
• Insert bulb into holder again and push in photometer drawer and fix it
with screws.
• Lock drawer stop-work after insertion of the instrument drawer again.
6 Error Messages
6.1 General Error Messages
If the ETI-MAX 3000 does not work, this is often due to minor problems
which you can deal with yourself.
This chapter describes error messages and gives instruction on error
recovery.
‘%1’ and ‘%2’ are placeholders for a system module or the designation of a
plate, a reagent or an error number.
System messages appear in the status bar of the ETI-MAX 3000 software,
error messages are displayed in a separate window, which has to be
confirmed.
Call service.
COP serial port test error
Cause:
The components are connected via serial interfaces to the control board.
Error in serial interfaces.
Action:
Call service.
ERROR: Argument error in '%1'
Cause:
Component cannot be actuated. Error in component %1.
Action:
Call service.
Error scheduling plate '...'
Cause:
Error in assay or missing resources. This error message is displayed in the
worklist for the respective plate, after the active worklist has been verified.
Action:
The worklist or the assay has to be changed accordingly.
Incubator heater %1 error
Cause:
Defect in heater foil no. %1.
Action:
Call service.
Incubator sensor %1 error
Cause:
Defect of incubator sensor no. %1.
Action:
Call service.
No disposable tips left
Cause:
Pipettor tips used up or not found.
Action:
Reload required tips at the position indicated on the monitor.
No fluid detected for '%1'
Cause:
Insufficient volume for sample ‘%1’ or reagent ‘%1’.
Action:
The status is documented in the Active event log. Confirm message with
Retry or Ignore. No further action is possible.
No respond to command '%1'
Cause:
General software error. Communication between PC and ETI-MAX 3000 is
interrupted.
Action:
Push the Retry button. If the message comes up again, push the Ignore
button, turn the PC off and start it again. Answer the questions “....worklist
...?” and “is the analyzer still running” accordingly. The system starts again
and continues with the worklist in process.
Plate ID verification error
Cause:
Plate barcode cannot be read.
Action:
Missing or faulty barcode, incorrectly selected barcode type (see Scanner
setup). Enter the plate ID manually.
Plate transport %1 motor home error
Cause:
Plate transport motor cannot find home position.
Action:
Select the Selftest function on the Utilities menu to initialize the ETI-MAX
3000 again.
Plate transport %1 motor movement error
Cause:
Error in plate transport motor no. %1.
Action:
Push the Retry button. If the error recurs, push the Ignore button.
Plate transport %1 motor not homed
Cause:
Plate transport motor cannot find home position.
Action:
Select the Selftest function on the Utilities menu to initialize the ETI-MAX
3000 again.
If the error persists, call service.
Plate transport carrier error
Cause:
Plate holder does not move far enough into the instrument, so that the light
barrier is not interrupted.
Action:
Push the Retry button. If the error recurs, push the Ignore button.
If this error recurs, the plate transport has to be adjusted new. Call service.
Plate transport EEPROM error
Cause:
EEPROM error in plate transport.
Action:
Call service.
Plate transport error
Cause:
Error in plate transport in case of increased resistance.
Action:
Push the Retry button. If the error recurs, try to locate the cause of the
resistance (blocking) and eliminate it. If no mechanical blockage is evident,
call service.
Plate transport sensor error
Cause:
Error in plate transport sensor.
Action:
Call service.
Rack scanner focusing error
Cause:
The barcode scanner of the reagent and sample racks cannot be focused.
Action:
Manual allocation of racks of samples and reagents is possible.
Call service.
Rack scanner motor error
Cause:
Motor error in scanner of reagent and sample rack. Scanner firmware does
not work correctly. Electrical or mechanical problems of scanner.
Action:
Manual allocation of racks of samples and reagents is possible.
Call service.
Rack scanner not detected
Cause:
Scanner of reagent and sample racks is not connected.
Action:
Manual allocation of racks of samples and reagents is possible.
Call service.
Reading at %1/%2
Cause:
Message: measurement at wavelengths %1/%2.
Action:
---
Reagent ... is undefined
Cause:
The reagent has not been defined.
Action:
Reagent has to be defined in the reagent database.
Enter the parameters of reagent in reagent database.
Resuming after a crash
Cause:
PC does not respond while processing the worklist and has to be restarted.
Action:
Following restart of the PC, you may be able to save the worklist and the
results. Connection to the instrument is established and the results stored
there are transferred to the PC.
The following error codes have been assigned to the individual system
modules. An error code consists of three parts: The ID number of the module,
the multi-digit error number and a letter code characterizing the errors.
Example:
ID41 25378 W means: Module 41, error code W, error number 25378.
In the following table you find this error display under the module number and
the error code.
ID 41 Virtual module
Code: W
Cause:
Waste container for system fluid is full.
Action:
Empty the waste container.
Code: F
Cause:
Not enough system fluid available.
Action:
Refill system fluid.
ID 42 Plate transport
Code: A
Cause:
Error in the command set.
Action:
Check if the correct firmware version of ETI-MAX 3000 has been installed.
Code: B
Cause:
Error of barcode scanner for plate transport.
Action:
ID 43 Colorimeter
Code: A
Cause:
Error in command set.
Action:
Check if the correct firmware version of ETI-MAX 3000 has been installed.
Code: B
Cause:
Background too high, caused by incidence of light (leakage) or colorimeter
defect.
Action:
Eliminate leakage or replace module. This work should be carried out by
service personnel.
Code: C
Cause:
Error of A/D converter.
Action:
Call service (replace PCB).
Code: Fn
Cause:
Filter error.
Action:
Replace filter.
Code: HX
Cause:
X-Motor (plate carrier) not in home position.
Action:
No action required. ETI-MAX 3000 initializes the module and tries again.
Code: HY
Cause:
Y-Motor (filter motor) not in home position.
Action:
No action required. ETI-MAX 3000 initializes the module and tries again.
Code: L
Cause:
Halogen lamp faulty.
Action:
Replace halogen lamp.
Code: MX
Cause:
X-motor (plate carrier) is blocked.
Action:
Locate cause of obstruction and remove it.
Code: MY
Cause:
Y-motor (filter motor) is blocked.
Action:
Locate cause of obstruction and remove it.
Code: P
Cause:
Error in plate carrier.
Action:
Check if the plate carrier interrupts the sensor.
Code: PO
Cause:
A/D overflow. Halogen lamp or PCB faulty.
Action:
Replace halogen lamp. If this does not help, call service (replace PCB).
Code: PU
Cause:
Signal detection too low. PCB or power supply faulty.
Action:
Call service.
Code: PW
Cause:
Colorimeter overflow. PCB faulty.
Action:
Call service.
Code: R
Cause:
Error in A/D reference voltage. PCB faulty.
Action:
Call service.
Code: Tn
Cause:
Faulty optical channel.
Action:
Check if the optical channel is damaged.
Call service.
Code: W
Cause:
EEPROM error.
Action:
Call service (replace EEPROM).
ID 44 Washer
Code: A
Cause:
Error in command set.
Action:
Check if the correct firmware version of ETI-MAX 3000 has been installed.
Code: E
Cause:
EEPROM error.
Action:
Call service (replace EEPROM).
Code: Hn
Cause:
Motor not in home position.
Action:
No action required. ETI-MAX 3000 initializes the module and tries again.
Code: Mn
Cause:
Motor blocked during movement.
Action:
Locate cause of obstruction and remove it. If necessary, call service.
Code: P
Cause:
Error of the plate carrier sensor.
Action:
Check if the plate carrier interrupts the sensor.
Code: SA
Cause:
Waste container for wash fluid full.
Action:
Empty the waste container for the wash fluid.
Code: SR
Cause:
Liquid level too low.
Action:
Refill the respective fluid container.
Code: RA
Cause:
Error in aspiration pump.
Action:
Check tubings and connections for leaks. If the error persists, call service.
Code: RD
Cause:
Error in dispensing pump.
Action:
Check tubings and connections for leaks. If the error persists, call service.
Code: RW
Cause:
Error in waste pump.
Action:
Check tubings and connections for leaks. If the error persists, call service.
Code: V
Cause:
Error pump valve.
Action:
Call service.
ID 45 Incubator
Code: A
Cause:
Error in command set.
Action:
Check if the correct firmware version of ETI-MAX 3000 has been installed.
Code: Hn
Cause:
Heater error.
Action:
Call service.
Code: P
Cause:
Error of plate carrier sensor.
Action:
Check if the plate carrier interrupts the sensor.
Call service.
Code: Tn
Cause:
Sensor error.
Action:
Call service.
COP
Code: 00D0541 0
Cause:
COP EEPROM error.
Action:
Call service (replace EEPROM).
Code: 00D05411
Cause:
COP interface error.
Action:
Call service (replace COP PCB).
Action:
Code: 00D5543D
Cause:
Scanner not connected.
Action:
Check scanner connection.
Code: 00D5543E
Cause:
Rack command could not be send.
Action:
Call service.
Code: 00D5543F
Cause:
Error in scanner motor.
Action:
Call service.
7 Technical Data
System Overview
Number of plates Up to 4 plates at once
Number of samples Max. 100 tubes, up to max. ∅ 16 mm, length up to 200 mm
Number of reagents Special racks for DiaSorin reagents
Dilution positions Positions for up to 160 dilution tubes
Number of tips Up to 5 tip racks for 300 µl or 1100 µl disposable tips
Loading Continuous loading of plates, samples, reagents and tips
Photometer
Spectral range 400 - 700 nm
Dynamic range -0.100 to 3.000 OD
Accuracy ± 0.005 or 2.5%
Linearity 0 - 2.000 OD ±1%
Detection Photodiode
Reading time < 15 seconds
Read modes OD and Kinetic mode
Filters Up to 8 positions
Pipetting System
Pipettor Liquid pipettor for disposable tips
Liquid level detection Standard
Min Max volumes 10 µl to 300 µl or
1000 µl with large tips
Accuracy < 3 % at 10 µl
< 0.5 % at 100 µl
Precision < 3 % CV at 10 µl
< 2.5 % CV at 100 µl
Features Tip detection, mixing, multi-dispensing mode
Incubation
Capacity 4 independent chambers
Temperature range Minimum temperature = RT + 5°C
Maximum temperature = 50°C
Accuracy ± 1° C mean of plate
Uniformity ± 0.7° C across plate
Washing
Capacity Max. 3 positions wash buffer + 1 distilled water
Wash head 1 x 8 with verify option
Dispense volume 200-2500 µl/well
Precision ± 5% CV at 300 µl
Residual volume < 2.5 µl in U-shaped bottom
< 4 µl in flat bottom
Fluid alarms Low reagent, waste full
Features Sweeps, soak, purge, top & bottom wash, variable pump
speeds
Software features
Operating system Windows 95
Interfaces ASTM interface
Flexible ASCII import of worklist sample ID’s
Time Scheduling Schedules 4 plates with user options
Multiple assays per plate Up to 12 assays per plate
Export options Flexible ASCII export and report
Qualitative User definable result classes
Quantitative Linear to quadratic regression, sigmoidal, and many more
Language support Multiple languages possible
QA analysis Mean, SD, CV, Standard Error and Levey-Jennings
Power Supply
Universal a.c. input 100 - 260 Vac, 47 - 63 Hz
Typically max. 500 VA
Dimensions
W x D x H in mm 1120x760x1000
Weight 120 kg
8 Contact
Italy
DiaSorin S.r.l.
Via Crescentino - 13040 Saluggia (VC)
Tel. 0039.0161487093 - Fax 0039.0161487642
Belgique/Belgie
DiaSorin S.A./N.V.
Boulevard Paepsem/Paepsemlaan, 11
1070 Bruxelles/Brussel
Tel. 0032.23402929 - Fax 0032.23402920
Brazil
DiaSorin Ltda.
Rua do Grito, 387
Conjuntos 114-115
04217-000 Ipiranga - São Paulo - SP
Tel. 0055.1169157270 - Fax 0055.1161681547
Germany
Byk & DiaSorin Diagnostics GmbH
von Hevesy-Straße, 3
D-63128 Dietzenbach
Tel. 0049.6074401400 - Fax 0049.6074401420
Spain
DiaSorin S.A.
Avda. Dr. Severo Ochoa, 40 - 2a pl.
P.A.E. Casablanca 1
28100 Alcobendas - Madrid
Tel. 0034.916623321 - Fax 0034.916624286
France
DiaSorin S.A.
Parc de Haute Technologie
Antony 2
11, rue Georges Besse
92160 Antony
Tel. 0033.155590400 - Fax 0033.155590440
United Kingdom
DiaSorin Ltd.
Charles House
Toutley Road
Wokingham, Berkshire RG41 1QN
Tel. 0044.1189364200 - Fax 0044.1189792061
USA
DiaSorin Inc.
1951 Northwestern Avenue - P.O. Box 285
Stillwater, MN 55082-0285
Tel. 001.6514399710 - Fax 001.6517797847