Testo Saveris 2 Instruction Manual
Testo Saveris 2 Instruction Manual
Instruction manual
2
Pos: 3 /BA L/Online-Hi lfe testo Saveris 2/0_Startseite/0_0_Startseite Online- Hilfe @ 15\m od_1407490185133_6.doc x @ 194776 @ 1 @ 1
1 Contents
1 Contents ....................................................................................................... 3
2 System.......................................................................................................... 5
2.1 System overview ................................................................................... 5
3 First steps .................................................................................................... 6
4 Offline Konfiguration................................................................................... 7
5 Licences ....................................................................................................... 8
6 WiFi data logger .......................................................................................... 9
6.1 Technical data ....................................................................................... 9
6.1.1 Measurement-specific data ..................................................... 9
6.1.2 General data ......................................................................... 10
6.1.3 WiFi-specific data.................................................................. 11
6.1.4 Technical data for a secure wireless LAN ............................. 11
6.2 Safety and the environment ................................................................. 12
6.2.1 General safety instructions ................................................... 12
6.2.2 Safety instructions for patients with a cardiac pacemaker or
implanted cardiac pacemaker ............................................... 12
6.2.3 Batteries ................................................................................ 13
6.2.4 Protecting the environment ................................................... 13
6.3 Wall bracket ......................................................................................... 14
6.4 Display and control elements............................................................... 16
6.4.1 Overview ............................................................................... 16
6.4.2 Display icons ......................................................................... 16
6.4.3 LED display – configuration .................................................. 17
6.4.4 LED display – operation ........................................................ 18
7 User interface ............................................................................................ 19
8 Dashboard.................................................................................................. 20
8.1 Measuring points ................................................................................. 20
8.2 Active alarms ....................................................................................... 20
9 Analysis & reports ..................................................................................... 20
9.1 Automatic reports ................................................................................ 20
10 Alarms ........................................................................................................ 22
10.1 Alarm list .............................................................................................. 22
10.2 Alarm settings ...................................................................................... 22
10.3 System warnings ................................................................................. 23
11 Configuration ............................................................................................. 25
11.1 Users ................................................................................................... 25
11.2 User roles ............................................................................................ 26
3
11.3 Account ID ........................................................................................... 26
11.4 Measuring point groups ....................................................................... 27
11.5 Areas ................................................................................................... 28
11.6 WiFi data loggers................................................................................. 28
11.7 Firmware updates ................................................................................ 29
12 User management ..................................................................................... 30
12.1 User settings........................................................................................ 30
12.2 Account information ............................................................................. 30
12.3 Change password................................................................................ 30
12.4 Logoff .................................................................................................. 30
13 Command bar ............................................................................................ 31
13.1 Open Quick Start Guide ...................................................................... 31
13.2 Open Online Help ................................................................................ 31
13.3 Open system messages ...................................................................... 31
14 System and status information ................................................................ 31
14.1 Display system status .......................................................................... 31
15 Frequently Asked Questions .................................................................... 32
16 Approvals and Certification...................................................................... 37
4
2 System
2.1 System overview
The testo Saveris 2 WiFi data logger system is the modern solution for monitoring
temperature and humidity values in storerooms and work rooms.
The system consists of WiFi data loggers and a Cloud data storage system.
The testo Saveris 2 WiFi data loggers record temperature and humidity reliably at
adjustable intervals and transmit the readings directly to the Testo Cloud via
WLAN. The readings stored in the Cloud can be analyzed at any time, anywhere,
using an internet-enabled smartphone, tablet or PC. Limit value violations are
immediately reported via e-mail, or optionally via SMS.
Similarly, WiFi data loggers can be programmed via the Testo Cloud and analyses
set up.
5
3 First steps
1. Take the data logger out of the packaging.
2. Remove the data logger from the wall bracket. (see chapter 6.3)
3. Loosen the battery cover by slightly undoing the screws on the back of the
housing.
4. Remove the insulation strips from the battery compartment.
The data logger is now activated.
5. Tighten the screws on the back of the housing back up.
The Quick Start Guide is provided to assist you when you take your first steps with
commissioning the testo Saveris 2.
It can help you with logging in WiFi data loggers.
• In the command bar, click on the icon to open the Quick Start Guide.
6
4 Offline Konfiguration
As an alternative to creating the configuration file in the Quick Start Guide with
subsequent download of the XML configuration file, the WiFi data logger can also
be configured via a PDF form.
Note
You need the Adobe Reader program (version 10 or later) to use the
PDF form correctly. If you have not installed Adobe Reader, you can
go to the following address to download it free of charge:
http://get.adobe.com/reader/.
Note
You can also save the configuration file locally on your computer and
configure other WiFi data loggers even faster by simply copying the
XML configuration file onto their external drive (Saveris 2).
7
5 Licences
When purchasing your Saveris 2 system you will receive a free basic licence.
Some measuring functions which are displayed in your Saveris 2 system are
inactive. By expanding your licence you can activate and use these functions.
In the header you can see which licence you are currently using.
Purchase licence
By clicking on the licence display in the header you will be redirected to the licence
shop. Here you can choose and purchase your licence expansion.
Or
Click on the symbol next to the inactive measuring function . This also
opens the licence shop to expand your licence.
8
6 WiFi data logger
6.1 Technical data
6.1.1 Measurement-specific data
WiFi data T1 T2 T3 H1 H2
logger
Order 0572 2031 0572 2032 0572 2033 0572 2034 0572 2035
number
Note
Saveris 2 WiFi data loggers are supplied with a factory calibration
certificate as standard. In many applications, it is recommended that
you recalibrate the loggers every 12 months. This can be carried out
by Testo Industrial Services (TIS) or other certified service providers
with the aid of easy-to-use service software. Please contact Testo for
further information.
9
6.1.2 General data
WiFi data logger T1 T2 T3 H1 H2
Order number 0572 2031 0572 2032 0572 2033 0572 2034 0572 2035
Protection class IP 65 IP 65 IP 54 IP 30 IP 54
1
Measuring cycle 1 min to 24 h (15 minutes default)
Dimensions 95 x 75 x 95 x 75 x 95 x 75 x 115 x 82 x 95 x 75 x
30.5 mm 30.5 mm 30.5 mm 31.5 mm 30.5 mm
10
WiFi data logger T1 T2 T3 H1 H2
Order number 0572 2031 0572 2032 0572 2033 0572 2034 0572 2035
Note
Please note that for EN 12830, annual checking and calibration must
be carried out as specified in EN 13486. Contact us for more
information.
Order number 0572 2031 0572 2032 0572 2033 0572 2034 0572 2035
Radio frequency 2.4 GHz (IEEE 802.11 b/g/n/x)
2 For item numbers: 0572 2001, 0572 2002, 0572 2003, 0572 2004, 0572 2005
3 For item numbers: 0572 2031, 0572 2032, 0572 2033, 0572 2034, 0572 2035
11
Ping
The default gateway, which is communicated to the probe via DHCP or manually,
must answer the PING request of the WiFi data logger.
Note
During the first configuration, it is possible to select whether
DHCP or Static IP is used (select expert mode for the
corresponding information)
12
• Maintain a distance of 20 cm between your implant and the magnetic plate
on the wall bracket.
6.2.3 Batteries
• Improper use of batteries may cause the batteries to be destroyed, or lead
to injury due to current surges, fire or escaping chemicals.
• Only use the batteries supplied in accordance with the instructions in the
instruction manual.
• Do not short-circuit the batteries.
• Do not take the batteries apart and do not modify them.
• Do not expose the batteries to heavy impacts, water, fire or temperatures in
excess of 60 °C.
• Do not store the batteries in the proximity of metal objects.
• In the event of contact with battery acid: rinse affected areas thoroughly
with water, and if necessary consult a doctor.
• Do not use any leaky or damaged batteries.
13
6.3 Wall bracket
Note
The data loggers must only be mounted vertically. Here, the
connections must point downwards. In the case of data loggers with a
display, you need to pay attention to the reading direction. Otherwise,
the measuring accuracy might be diminished.
The wall bracket ensures that the WiFi data logger is kept securely in place.
Mounting materials are not included in the scope of delivery. Select mounting
materials that are appropriate for the required mounting location.
(Suitable for: 0527 2001, 0527 2002, 0527 2003, 0527 2004, 0527 2005, 0527
2031, 0527 2032, 0527 2033, 0527 2035)
1 Wall bracket with notches for mounting materials (screws or cable ties)
2 Safety bolt
3 Lock
4 Storage compartment for USB port sealing plug
5 Storage compartment for probe socket sealing plug (left/right)
6 USB cable holder, folding: push down on the spot marked with a circle
using a screwdriver and slide cable holder downwards.
14
(Suitable for: 0572 2034)
1 Wall bracket
2 Data logger
3 Unlocking tool
4 Magnetic plate (optionally available, order number: 0554 2001)
15
6.4 Display and control elements
6.4.1 Overview
1 Display
2 Alarm LED, flashes red in the event of an alarm
3 QR code, for direct access to the instrument data in the Testo Cloud
4 Control key, to manually start data transfer
5 Battery compartment (on rear)
6 USB and probe sockets (on the bottom, instrument-specific)
16
Icon Description
WLAN signal strength 75%
Alarm message
Measurement channel 1
Measurement channel 2
Online data logger gives a short (200 Online data logger is in standby mode, there is
ms) green flash every 30 seconds. no configuration stored in the online data
logger yet.
Online data logger flashes green at one- Online data logger is in configuration mode
second intervals. and can be configured within 5 minutes after
activation.
Online data logger gives 1 long (1 sec.) Online data logger was not configured within 5
red flashes. minutes.
Online data logger gives 2 long red Online data logger is not connected to the
flashes after configuration. access point.
Online data logger gives 2 short green Configuration was successful, online data
flashes. logger is connected to the Cloud and is
switching to measuring mode.
17
6.4.4 LED display – operation
LED display Description
Online data logger which is already Online data logger is not connected to the
configured gives 2 short red flashes. access point, measurement data cannot be
transmitted to the Cloud.
Online data logger gives 1 long The online data logger has carried out a
green flash. measurement.
Online data logger gives 1 long red An alarm was detected by the online data logger.
flash.
Online data logger gives 2 short Measurement data was successfully transmitted
green flashes. to the Cloud, logger is back in the measuring
cycle.
Online data logger gives 4 long red Briefly (< 1 sec) press the button on the front of
flashes. the logger. If the online data logger flashes red 4
times again, the batteries are spent and need to
be replaced.
18
7 User interface
19
8 Dashboard
The dashboard is the home page. It provides an overview of the key system data:
Measuring points and Active alarms.
20
• Also send report via e-mail: as well as saving reports under Generated
reports, these can also be sent as e-mails. Click on the checkbox to open
the input screen for e-mail addresses.
3. Click on the Create an automatic report button.
The first report will be created on the following day.
Generated reports
A summary of the reports already generated is displayed.
Click on the icon to display more information.
• Click on the Download button to download a report.
• Click on the Edit this report series button to display and edit the settings.
Report settings
Automatic reports which have already been created are displayed in a table.
• Click on the Actions button and then on Edit to display and edit the
settings.
• Click on the Actions button and then on Delete to delete the automatic
report.
21
10 Alarms
10.1 Alarm list
Display of alarms
A summary of all triggered alarms and system warnings is displayed. Unread
alarms and system warnings are shown in bold.
The display can be filtered according to the following characteristics:
• By measuring point group/measuring point: click on the checkbox in front of
the measuring point group/measuring point.
• By date/time: click on the start date/end date and select the start date/start
time and end date/end time.
Detailed information on alarms
Click on the icon to display more information.
On displaying the detailed information, the alarm message/system warning is
marked as "read" and the alarm counter is marked down.
Click on the Mark all as read button to mark all alerts as "read".
22
The time intervals between measurements (measurement rate should be
smaller than the alarm delay (e.g. measurement rate = 5 minutes, alarm
delay = 15 minutes).
• Timed control: Define individual alarm periods for which the alarm limit
values 1 and 2, or no alarm at all, apply. To define the alarm value 1 and 2,
double-click on a time point in the table or pull open the desired time period
with the mouse. In periods in which the the table is left empty, you receive
no alarm.
If you have not defined alarm periods, the alarm limits will be active 24
hours a day.
If alarm periods have been defined, the limit value alarms are active only
in the marked period.
23
• WiFi data logger is not responding: monitor the WiFi data logger for data
transmission failure. Click on the Activate button and configure the
monitoring cycle using the slide control.
Note
The set time should be greater than the WiFi data logger's
communication cycle.
24
11 Configuration
11.1 Users
Users can be created and managed.
Standard users
By default, two users are created in the system:
• Account Owner (name can be changed), with Administrator user role (role
cannot be changed).
• Testo Support (name can be changed), with Testo User Support user role
(role cannot be changed).
25
• Details: text field for entering other user-specific information.
• Save: click on the button to save the settings.
Display Account ID x - - x
11.3 Account ID
The Account ID is the unique address of your user account in the Testo Cloud.
This is needed to configure the WiFi data loggers in order to ensure that they send
your data to the correct user account.
26
11.4 Measuring point groups
Measuring points can be organized into measuring point groups.
Assigning measuring points to a measuring point group (e.g. Room 1, Room 2,
etc.) makes the administration of multiple measuring points easier.
Creating and editing a measuring point group
• Click on the New measuring point group button to create a new
measuring point group.
Measuring point groups already created are displayed in a list.
• Click on the Actions button and then on Edit to display and edit the
settings.
The following settings can be defined and edited:
• Title: designation of the measuring point group (required field).
• Description: description of the measuring point group.
• Area: area that the measuring point group is to be assigned to.
• Measuring points: available measuring points and those assigned to the
27
11.5 Areas
Measuring point groups can be organized into areas.
Creating and editing an area
• Click on the New area button to create a new area.
Areas already created are displayed in a list.
• Click on the Actions button and then on Edit to display and edit the
settings.
The following settings can be defined and edited:
• Display name: designation of the area (required field).
• Description: description of the area.
• Save: click on the button to save the settings.
• Delete: click on the button to delete the area.
Deleting an area
• Click on the Actions button and then on Delete to delete the area.
28
the day communication cycle and the night communication cycle. Set the
communication cycle using the slide control.
• Channel name: designation of the measurement channel (required field).
Condition at delivery: Model_Serial number_Channel number
• Select unit: unit in which the readings are displayed.
• Select sensor type (only for model T2, T3): sensor type used by the
measurement channel.
• Save: click on the button to save the settings.
• Deactivate or Activate: click on the button in order to deactivate or activate
measurement channels or WiFi data loggers.
• Remove: click on the button in order to log WiFi data loggers off the
system.
29
12 User management
The user management provides information and settings options for the user
account.
Note
It is not possible to change the user name (e-mail address).
12.4 Logoff
The user is logged off the system.
• Logoff: click on the text to log off.
30
13 Command bar
13.1 Open Quick Start Guide
The Quick Start Guide can help you with logging in WiFi data loggers.
: no alarms active
31
15 Frequently Asked Questions
Can the WiFi data logger be connected to the PC using any USB cable?
We recommend that you use the USB cable supplied with the WiFi data logger to
guarantee stable data transmission. Longer USB cables are suitable for the power
supply only.
Can the WiFi data logger also be used in networks with WPA2 Enterprise
encryption?
Second-generation Saveris 2 loggers* can be used in networks with the following
WPA2 Enterprise encryption methods.
EAP-TLS, EAP-TTLS-TLS, EAP-TTLS-MSCHAPv2, EAP-TTLS-PSK, EAP-PEAP0-
TLS, EAP-PAP0-MSCHAPv2, EAP-PEAP0-PSK, EAP-PEAP1-TLS, EAP-PEAP1-
MSCHAPv2, EAP-PEAP1-PSK
To integrate the loggers into the WPA2 Enterprise network, proceed as follows:
1. Open the .PDF file stored on the logger and generate a corresponding .XML file
by selecting the programming options step by step.
2. Copy your company-specific WPA2 Enterprise certificates and the generated
.XML file to the logger's mass storage via USB using drag & drop.
3. Please pay special attention to the correct spelling and to the suffixes of
certificate names.
ca.pem, client.pem, private.key
Check whether the certificate is available in a PEM or BASE64 format. To do so, open
the certificate using a text editor and check that you can see the string “-----BEGIN
CERTIFICATE-----”. If you cannot see this string, the IT department or the user must
explicitly export the certificate from the Radius server using BASE64, or convert it using
openssl. There must be one certificate only, not a bundle, inside the file.
4. Please note that the configuration of the WiFi data logger will only be fully
transferred once the USB connector has been removed.
* Item number: 0572 2031, 0572 2032, 0572 2033, 0572 2034, 0572 2035
The XML configuration file is not being applied by the WiFi data logger, what
can I do?
Depending on the operating system, there may be difficulties with the data transfer
if the configuration file name has been changed. Leave the default file name.
The WiFi data logger's wireless connection to the access point was
interrupted, what can I do?
• Press the control key on the WiFi data logger to start searching for a WLAN
connection manually.
32
• Change the alignment or position of the WiFi data logger or the access
point (WLAN router).
The WiFi data logger is displaying the error code E03, E04, E05 or E09, what
can I do?
An error has occurred in the WiFi data logger. The error will automatically be
corrected by the firmware of the WiFi data logger. After a few seconds the error
code should no longer be displayed, you do not need to do anything.
The WiFi data logger is displaying the error code E12, what can I do?
The configuration file WifiConfig.xml indicates an error. Use the Quick Start
Guide to create a new configuration file and save this onto the WiFi data logger.
The WiFi data logger is displaying error code E20, what can I do?
A WPA2 Enterprise EAP connection needs to be configured, but no CA certificate
could be found. A CA certificate is essential.
Save the CA certificate in PEM format under the name “ca.pem” together with the
XML file or upload it in the web config.
The WiFi data logger is displaying error code E21, what can I do?
Incorrect format of the ca.pem certificate. Check whether the ca.pem certificate is
available in a PEM or BASE64 format. To do so, open the certificate using a text
editor and check that you can see the string “-----BEGIN CERTIFICATE-----”. If you
cannot see this string, the IT department or the user must explicitly export the
certificate from the Radius server using BASE64, or convert it using openssl. There
must be one certificate only, not a bundle, inside the file.
The WiFi data logger is displaying error code E22, what can I do?
A WPA2 Enterprise EAP-TLS connection needs to be configured, but no user
certificate could be found. Save the user certificate in PEM format under the name
“client.pem” together with the XML file or upload it in the web config.
The WiFi data logger is displaying error code E23, what can I do?
Incorrect format of the client.pem user certificate. Check whether the client.pem
user certificate is available in a PEM or BASE64 format. To do so, open the
certificate using a text editor and check that you can see the string “-----BEGIN
CERTIFICATE-----”. If you cannot see this string, the IT department or the user
must explicitly export the certificate from the Radius server using BASE64, or
convert it using openssl. There must be one certificate only, not a bundle, inside
the file.
Older generation Saveris2 logger, prior to 2016 (see type plate):
The most common reason for this error is low battery. Insert new
batteries into the WiFi data logger.
33
If this does not solve the problem: Reset the WiFi data logger to its
factory settings. To do this, press and hold down the control key for
> 20 s until the display goes blank.
The WiFi data logger is displaying error code E24, what can I do?
A WPA2 Enterprise EAP-TLS connection needs to be configured, but no private
key could be found. Save the private key in PEM format under the name
“private.key” together with the XML file or upload it in the web config.
The WiFi data logger is displaying error code E25, what can I do?
Incorrect format of the private.key certificate. Check whether private.key is
available in a PEM or BASE64 format. To do so, open the certificate using a text
editor and check that you can see the string “-----BEGIN CERTIFICATE-----”. If you
cannot see this string, the IT department or the user must explicitly export the
certificate from the Radius server using BASE64, or convert it using openssl. There
must be one certificate only, not a bundle, inside the file.
The WiFi data logger is displaying the error code E26, what can I do?
There are 3 possible reasons for this error:
1 The access point (WLAN router) is outside the wireless range or is switched
off. Check whether the access point is available. If necessary, change the
location of the WiFi data logger.
2 The network name (SSID) stored in the WiFi data logger is incorrect. Check
the network name of the WLAN network. Use the Quick Start Guide to
create a new configuration file with the correct network name and save this
onto the WiFi data logger.
3 The access point of the WLAN network does not use one of the following
encryption methods: WEP, WPA (TKIP), WPA2 (AES, CCMP). Configure
the access point so that one of the supported encryption methods is used.
The WiFi data logger is displaying the error code E32, what can I do?
The WiFi data logger has not obtained an IP address. There are 2 possible
reasons for this error:
1 The network password is incorrect. Check the password of the WLAN
network. Use the Quick Start Guide to create a new configuration file with
the correct password and save this onto the WiFi data logger.
2 The access point (WLAN router) has a MAC filter or does not permit the
integration of new devices. Check the settings for the access point.
The WiFi data logger is displaying the error code E35, what can I do?
The WiFi data logger has not received any reply to its test ping from the access
point (WLAN router). Make sure that a ping to the gateway is allowed within the
access point configuration.
34
The WiFi data logger is displaying error code E36, what can I do?
The DNS could not be resolved:
1 The access point (WLAN router) has no connection to the internet. Check the
access point's internet connection.
or
2 The routing within the network infrastructure is not working. Check
whether too many terminal devices are logged into the access point.
The WiFi data logger is displaying the error code E41, what can I do?
The WiFi data logger cannot obtain any current time from a time server
(pool.ntp.org).
1 The access point (WLAN router) has no connection to the internet. Check
the access point's internet connection.
2 The NTP port (123/UDP) of the access point (WLAN router) is not open.
Check whether the NTP port (123/UDP) is opened.
The WiFi data logger is displaying the error code E51, what can I do?
The WiFi data logger was not able to connect to the Testo Cloud.
1 If the WiFi data logger has already been connected to the Testo Cloud and
this connection is suddenly no longer possible: The Testo Cloud servers are
not currently accessible. The servers will be monitored and should be
accessible again within a few hours.
2 If the WiFi data logger has not yet been connected to the Testo Cloud: The
TCP ports (1883 or 8883) of the access point (WLAN router) are not open.
Check whether the TCP ports (1883 or 8883) are open in both directions.
The WiFi data logger is displaying error code E52, what can I do?
The data logger could not log into the Cloud because it is already logged into
another account. Please log the data logger out of the existing account first.
The WiFi data logger is displaying the error code E63, what can I do?
The WiFi data logger could not send any data to the Testo Cloud.
1 The internet connection was interrupted during the transmission. Check
whether there is a stable connection from the WiFi data logger to the
access point (WLAN router). Check the access point's internet connection.
The data will be transferred during the next communication cycle.
Alternatively: Initiate data transmission manually by pressing the control key
on the WiFi data logger.
2 The Testo Cloud server was not able to process the request for data
storage. The servers will be monitored and should be accessible again
within a few hours.
35
The WiFi data logger is displaying the error code E75, what can I do?
A firmware update for the WiFi data logger failed.
The internet connection was interrupted during the transmission or the data was not
received intact by the WiFi data logger for other reasons. Check whether there is a
stable connection from the WiFi data logger to the access point (WLAN router).
Check the access point's internet connection. The data will be transferred during the
next communication cycle. Alternatively: Initiate data transmission manually by
pressing the control key on the WiFi data logger.
The WiFi data logger is displaying the warning message Err AccountID, what
can I do?
The AccountID contained in the configuration file is not valid.
Use the Quick Start Guide to create a new configuration file and save this onto
the WiFi data logger.
The WiFi data logger is displaying the warning message no AccountID, what
can I do?
There is no AccountID in the configuration file.
Use the Quick Start Guide to create a new configuration file and save this onto
the WiFi data logger.
The WiFi data logger is displaying the warning message no License, what
can I do?
The WiFi data logger cannot be logged on because the number of WiFi data
loggers permitted to log on has been exceeded or your testo Saveris 2 licence has
expired.
Log off another WiFi data logger, extend or renew your testo Saveris 2 licence.
The WiFi data logger is displaying the warning message not Active, what can
I do?
The WiFi data logger has been deactivated. It is not storing, and therefore not
sending, any measurement data to the Testo Cloud.
Activate the WiFi data logger (under Configuration --> WiFi data logger) when
the WiFi data logger needs to store and send measurement data again.
36
16 Approvals and Certification
Saveris 2 T1 (2016)
Saveris 2 T2 (2016)
Product Saveris 2 T3 (2016)
Saveris 2 H1 (2016)
Saveris 2 H2 (2016)
0572 2031
0572 2032
Mat.-No. 0572 2033
0572 2034
0572 2035
21.09.2023
Date
The use of the wireless module is subject to the regulations and stipulations of the respective country
of use, and the module may only be used in countries for which a country certification has been granted.
The user and every owner has the obligation to adhere to these regulations and prerequisites for use, and
acknowledges that the re-sale, export, import etc. in particular in countries without wireless permits, is his
responsibility.
Country Comments
Australia E 1561
Belarus Authorized
Brazil T1
T2
T3
H1
H2
37
Canada Contains IC : 21461-LSD4WF0459
T1, T2, T3, H2:
IC: 6127B-0572203X
H1:
IC: 6127B-05722034
see IC Warnings
Europa + EFTA
EU countries:
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia (EE),
Finland (FI), France (FR), Greece (GR), Ireland (IE), Italy (IT), Latvia (LV),
Lithuania (LT), Luxembourg (LU), Malta (MT), Netherlands (NL), Austria
(AT), Poland (PL), Portugal (PT), Romania (RO), Sweden (SE), Slovakia
(SK), Slovenia (SI), Spain (ES), Czech Republic (CZ), Hungary (HU),
Republic of Cyprus (CY).
EFTA countries:
Iceland, Liechtenstein, Norway, Switzerland
38
Israel Authorized
Japan
Malaysia Authorized
Pakistan Authorized
South Korea
T1: MSIP-CRM-te2-05722031
T2: MSIP-CRM-te2-05722032
T3: MSIP-CRM-te2-05722033
H1: MSIP-CRM-te2-05722034
H2: MSIP-CRM-te2-05722035
see KCC Warning
Thailand Authorized
Türkiye Authorized
Vietnam Authorized
Wi-Fi-Module
Feature Values
WLAN Range 100 m
WLAN type LSD4WF0459-01D0
WLAN radio class Accord with the standard of IEEE 802.11b/g/n
Company Lierda Technology Group co., LTD
RF Band 2412-2472MHz
Transmitter Power 13.42dBm
39
IC Warnings
This instrument complies with Part 15C of the FCC Rules and Industry Canada RSS-210 (revision 8).
Commissioning is subject to the following two conditions:
(1) This instrument must not cause any harmful interference and
(2) this instrument must be able to cope with interference, even if
this has undesirable effects on operation.
Cet appareil satisfait à la partie 15C des directives FCC et au standard Industrie Canada RSS-210
(révision 8). Sa mise en service est soumise aux deux conditions suivantes :
(1) cet appareil ne doit causer aucune interférence dangereuse et
(2) cet appareil doit supporter toute interférence, y compris des interférences qui provoquerait des
opérations indésirables.
FCC Warnings
Information from the FCC (Federal Communications Commission)
Japan Information
当該機器には電波法に基づく、技術基準適合証明等を受けた特定無線設備を装着している。
KCC Warning
40
Testo SE & Co. KGaA
Celsiusstraße 2
79822 Titisee-Neustadt
Germany
Telefon: +49 7653 681-0
E-Mail: [email protected]
Internet: www.testo.com
41